Datex-Ohmeda Centiva-5 User manual

Page 1
Centiva/5
User’s Reference Manual
Software Revision 2.n
Page 2
User Responsibility
This Product will perform in conformity with the description thereof contained in this operating manual and accompanying labels and/or inserts, when assembled, operated, maintained and repaired in accordance with the instructions provided. This Product must be checked periodically. A defective Product should not be used. Parts that are broken, missing, plainly worn, distorted or contaminated should be replaced immediately. Should repair or replacement become necessary, Datex-Ohmeda recommends that a telephonic or written request for service advice be made to the nearest Datex-Ohmeda Field Service Support center. This Product or any of its parts should not be repaired other than in accordance with written instructions provided by Datex-Ohmeda and by Datex-Ohmeda trained personnel. The Product must not be altered. The user of this Product shall have the sole responsibility for any malfunction which results from improper use, faulty maintenance, improper repair, damage or alteration by anyone other than Datex-Ohmeda.
U.S. Federal laws restrict this device to sale by or on the order of a licensed medical practitioner. Outside the U.S.A. check local laws for any restriction that may apply.
Datex-Ohmeda products have unit serial numbers with coded logic which indicates a product group, the year of manufacture and a sequential unit number for identification.
AAA B 12345
This alpha character indicates the year of product manufacture and when the serial number was assigned. 1 = 2001, 2 = 2002 etc.
URM Centiva/5, CM-920041, rev. 2.22 TIM GmbH
Page 3
Table of Contents
1 Introduction
2 Safety
3 General description
4 Concept of operation
1.1 How to use this manual 1-2
1.2 Symbols used in this manual or on the equipment 1-3
1.3 Abbreviations used in this manual or on the equipment 1-5
2.1 Safety 2-2
3.1 General description 3-2
3.1.1 Intended use 3-2
3.1.2 Principle of operation 3-2
3.1.3 Ventilation modes 3-3
3.1.4 Additional function 3-4
3.2 Supply 3-5
3.2.1 Electrical supply 3-5
3.2.2 Gas supply 3-5
3.3 Positioning 3-6
3.4 Main components 3-7
3.5 System controls 3-8
3.5.1 Front side 3-8
3.5.2 Rear side 3-9
4.1 Concept of operation 4-2
4.1.1 Screen content 4-3
4.1.2 Soft keys 4-5
4.1.3 I : E Ratio soft key 4-6
URM Centiva/5, CM-920041, rev. 2.22
0-1
Page 4
Table of Contents
4 Concept of operation
continued
5 Preparation
4.1.4 Timing parameters 4-7
4.1.5 Graphics 4-8
4.1.9 Changing a parameter 4-13
4.1.10 Menu button 4-14
4.1.11 Selecting a menu topic 4-15
4.1.12 Activating a menu function 4-16
5.1 Electrical supply 5-2
5.1.1 Mains supply 5-2
5.1.2 Back-up battery supply 5-4
5.2 Gas supply 5-5
5.2.1 Dual gas supply 5-5
5.2.2 Single gas supply 5-6
5.3 Set up of expiration valve 5-7
5.4 Breathing gas conditioning 5-8
5.5 Breathing circuit 5-9
5.5.1 Connection with HME 5-9
5.5.2 Connection with active humidifier 5-10
5.6 Placing the device 5-12
5.6.1 Control panel 5-12
5.6.2 Entire system 5-14
5.7 Turn ON and OFF 5-15
5.8 Self test 5-16
5.8.1 Power-up test 5-16
5.8.2 System test 5-17
5.8.3 In-operation test 5-18
5.9 Select the type of patient 5-19
5.10 System test menu 5-20
5.11 Starting the ventilator 5-22
4.1.6 Numerical patient data 4-9
4.1.7 Status information 4-10
4.1.8 ComWheel 4-12
4.1.13 Submenus 4-17
4.1.14 Open a submenu 4-19
4.1.15 Change a parameter in a submenu 4-20
4.1.16 Alarm silence 4-22
0-2
URM Centiva/5, CM-920041, rev. 2.22
Page 5
Table of Contents
6 Operation
7 Cleaning and Sterilization
6.1 Adjust ventilation modes 6-2
6.1.1 Select a ventilation mode 6-2
6.1.2 Default ventilation values 6-3
6.1.3 Changing ventilation modes 6-4
6.2 Adjust ventilation parameters 6-6
6.2.1 VCV 6-6
6.2.3 Bi-Level 6-12
6.2.4 Bi-Level/PSV 6-14
6.2.5 CPAP/PSV 6-17
6.2.5 CPAP/Apnea 6-19
6.3 Additional functions 6-21
6.3.1 Standby 6-22
6.3.2 ASR (Automatic Suction routine) 6-24
6.3.5 Alarm AUTOSET 6-38
6.3.6 ARC (Airway Resistance Compensation) 6-41
6.3.9 NIV (Non Invasive Ventilation) 6-51
6.3.10 APD (Automatic Patient Detection) 6-53
6.3.11 OCV (Online O2 sensor Calibration) 6-54
6.3.12 Short cuts 6-55
7.1 General advice 7-2
7.2 Disassembly 7-3
7.2.1 Used with HME 7-3
7.2.2 Used with active humidifier 7-4
7.3 Expiration valve 7-5
7.3.1 Removing expiration valve assembly 7-5
7.3.2 Disassemble expiration valve assembly 7-7
7.4 Clean and sterilize 7-8
7.4.1 Housing and control panel 7-8
7.5 Reassembly 7-9
6.2.2 SIMV-VC 6-9
6.3.3 Parameter menu 6-32
6.3.7 NEB (Nebulizer drug delivery) 6-44
6.3.8 System test 6-50
7.4.2 Expiration valve 7-8
7.5.1 Reassemble expiration valve assembly 7-9
7.5.2 Reinsertion of expiration valve assembly 7-10
URM Centiva/5, CM-920041, rev. 2.22,
0-3
Page 6
Table of Contents
8 Alarms and messages
9 Maintenance
10 Technical data
11 Parts list
8.1 Alarm modality 8-2
8.2 Alarm messages 8-4
8.3 Other messages 8-8
9.1 General 9-2
9.2 Oxygen sensor calibration 9-3
9.2.1 Start calibration 9-3
9.2.2 Stop calibration 9-4
9.3 Expiration valve 9-5
9.3.1 Remove valve assembly 9-5
9.3.2 Replace components 9-7
9.3.3 Reassembly 9-8
9.4 Additional maintenance 9-10
9.4.1 Every twelve (12) months 9-10
9.4.2 Every twenty four (24) months 9-10
10.1 General 10-2
10.2 Supplies 10-3
10.3 Ventilator characteristics 10-4
10.4 Controls and ranges 10-8
10.6 Measurement functions 10-9
10.7 Monitoring functions 10-10
10.8 Control panel 10-13
11.1 Spare parts 11-2
11.2 Accessories 11-2
8.3.1 Messages during power up 8-8
8.3.2 Messages during system test 8-11
8.3.3 Messages during O2 calibration 8-19
8.3.4 Messages during operation 8-20
9.3.4 Reinsertion of expiration valve assembly 9-9
0-4
URM Centiva/5, CM-920041, rev. 2.22
Page 7
1 Introduction
In this section
1.1 How to use this manual 1-2
1.2 Symbols used in this manual or on the equipment 1-3
1.3 Abbreviations used in this manual or on the equipment 1-5
URM Centiva/5, CM-920041, rev. 2.22
1-1
Page 8
1 Introduction
1.1 How to use this manual
This manual describes the operation of the
The Centiva/5 ICU ventilator may be used together with additional accessories, such as trolleys, compressors etc. These accessories come with a separate operation manual.
Use this manual together with the manuals for optional accessories.
Centiva/5 ICU ventilator Release 2.2n.
1-2
URM Centiva/5, CM-920041, rev. 2.22
Page 9
1.2 Symbols used in this manual or on the equipment
Warnings and Cautions tell you about dangerous conditions that can occur if you do not follow all instructions in this manual.
WARNING
Warnings indicate a condition that can cause injury to the operator or the patient.
Cautions indicate a condition that can cause damage to the equipment. Read and follow all warnings and cautions.
Alarm silence touch key
Menu touch key
Left turn ComWheel to decrease value
Right turn ComWheel to increase value
Rotate ComWheel to change value
Push ComWheel to confirm selection (ENTER)
1 Introduction
URM Centiva/5, CM-920041, rev. 2.22
1-3
Page 10
1 Introduction
x
On (power)
Off (power)
10101
xxx
Earth ground
Protective earth ground
Equipotential
Type B equipment
Input (Expiratory port)
Output (Inspiratory port)
Attention, refer to product instructions, EN/IEC 60601-1
Serial data transfer connection
Systems with this mark comply with the European Council Directive (93/42/EEC) for Medical Devices when they are used as specified in their Operation and Maintenance Manuals. The “xxxx” is the certification number of the Notified Body used by the manufacturer’s Quality Systems.
Nebulizer gas supply
DC external power supply
1-4
URM Centiva/5, CM-920041, rev. 2.22
Page 11
1 Introduction
1.3 Abbreviations used in this manual or on the equipment
Vent. mode
VCV
SIMV - VC
Bi-Level
PSV
CPAP
CPAP/PSV
CPAP/Apnea
APD ARC ASR NEB
NIV
OCV
Endot.tube
Trach.tube
P
max
P
limit
P
insp
PS
P
peak
P
min
P
PEEP
P
aw
Rate
I:E
Ramp
PS ramp
Time window
Vt
TV
MV
I-Flow
Trigg.
Byflow
PS endflow
Flow
Ventilation mode Volume constant Controlled Ventilation Synchronized Intermittent Controlled Ventilation- volume constant Bi Level pressure controlled ventilation Pressure Support Ventilation Continuous positive airway pressure CPAP with pressure support ventilation CPAP with apnea Bi-Level back up ventilation
Automatic patient detection Airway resistance compensation Automatic suction routine Nebulizer Non Invasive Ventilation Online oxygen calibration during ventilation Endotracheal tube Tracheal tube
Safety pressure setting Pressure limit setting Inspiratory pressure setting Pressure Support pressure setting Highest airway pressure, measured within last breath Lowest airway pressure, measured within last breath Positive End Expiratory Pressure setting Airway pressure in general
Breaths per minute Inspiratory to Expiratory time ratio Ramp time between lower and upper Bi-level Ramp time between PEEP level and PS pressure level Trigger time window
t
Time in general
Tidal volume setting Tidal volume, measured Minute volume, measured Inspiratory flow setting Trigger flow required to trigger an assisted breath Bias flow End-of-flow criteria for PS pressure support Flow in general
URM Centiva/5, CM-920041, rev. 2.22
1-5
Page 12
1 Introduction
Fi-O2
O2-I
or AC or DC
Rev. x.nn
40 … 80 PSI
O
2
AIR 40 … 80 PSI
24V/3A DC
Serial Port
Inspiratory O2 concentration setting Inspiratory O2 concentration, measured
I
On (power), line supply
O
Off (power), line supply
Alternating current Direct current
Ref
Stock number
SN
Serial number Software revision number
O2 gas supply input with a pressure range of 40 … 80 PSI AIR gas supply input with a pressure range of 40 … 80 PSI
DC supply input with 24 V and a maximum current of 3 A Connection for serial data transfer
1-6
URM Centiva/5, CM-920041, rev. 2.22
Page 13
2 Safety
In this section
2.1 Safety 2-2
URM Centiva/5, CM-920041, rev. 2.22
2-1
Page 14
2 Safety
2.1 Safety
Reference
Safe Operation
Classification
Safety standards
This manual describes the operation of the
To ensure safe operation of the Centiva/5 use the system only as intended. Operators need to be familiar with this Operation Manual prior to operating the system. Only trained operators are allowed to use this system. Always ensure compliance with the requirements of this Operation manual and with the local governmental or other authority’s requirements for operating ICU ventilators.
The Centiva/5 is a device of group IIb, according to the directive 93/42/EEC, Annex IX, Rule 9 for medical products.
Classification according to EN 60601-1 for electro medical devices
EVICE OF PROTECTION CLASS I, with internal power supply. Protection
D against electrical hazard is performed by grounded operation.
EVICE TYPE B, according to safety grade for electrical hazard
D
ONTINUOUS OPERATION according to described operation modes.
C
The Centiva/5 is designed with respect to the following safety standards:
- EN 60601-1, Medical electrical equipment, General conditions
- EN 794-1, Special requirements for lung ventilators
- EN 475, Specification for warning, advice and information in anesthesia and critical care
- EN 60601-1-4, software controlled medical devices
- EN 60601-1-2, electro magnetic compatibility
URM Centiva/5, CM-920041, rev. 2.22
Release 2.2n.
2-2
URM Centiva/5, CM-920041, rev. 2.22
Page 15
2 Safety
Safety features
Patient Safety
The following safety features are built into the Centiva/5 to warn the operator in case of risk to the patient.
Device alarms
Control of Control for
Gas supply O2 supply fail, AIR supply fail
Disconnection Disconnection of patient and breathing
O2 concentration O2 concentration
Electrical energy Line supply fail, battery supply fail
Device’s control system Watchdog alarm
Patient alarms
Control of Control for
High pressure High pressure alarm
Minute volume Upper and lower minute volume alarm
Rate Upper and lower rate alarm
Airway pressure Pressure sustained
The Centiva/5 shall be operated by or on the order of a physician. The Centiva/5 shall only be operated by qualified medical personal, to ensure adequate intervention in case of a device malfunction.
Constant surveillance of a patient connected to an ICU ventilator by trained personal is required.
Ensure alarm limits are set according to the patient’s situation and ensure alarms are activated whenever a patient is connected to the system.
Use only breathing circuits and accessories approved by the manufacturer of either the Centiva/5 or the accessory.
system
URM Centiva/5, CM-920041, rev. 2.22
2-3
Page 16
2 Safety
Patient Safety (cont.)
WARNING
Residual risk
Always perform the Centiva/5’s system test before using the ventilator.
Never perform a system test when a patient is connected to the Centiva/5.
This system test is designed to test all safety relevant features and components prior to the start of ventilation.
In case of emergency the system test may be bypassed. The system will recognize this status as “No test” and displays this status in the screen to remind the operator to perform the system test at their earliest convenience.
If the system test is not performed, the system does not know the correct breathing circuit’s specifications necessary for optimized compliance and resistance compensation. In this case the ventilator may not work with the specified accuracy and tolerances.
To reduce the risk of a device malfunction, redundant systems are integrated into the Centiva/5.
High and medium risk functions are designed to provide an alternate system in case of a first failure.
Low risk functions are tested with the power-up test each time the unit is powered up prior to operation.
The system provides for an adjustable pressure limiter in the VCV and SIMV –VC modes. In the VCV or SIMV-VC modes the P
O in Pediatric and 30 cmH2O in Adult mode.
cmH
2
The system has an additional second level safety pressure function,
, that is active in all modes. The P
P
max
Pediatric and 40 cmH
O in the Adult mode, when in the VCV or SIMV-VC
2
modes. The Pmax default is set to 18 cmH
defaults to 30cmH2O in
max
O in Pediatric and 21 cmH2O
2
in the Adult mode, when in Bi-Level, CPAP/PSV or CPAP/Apnea modes.
defaults to 20
limit
2-4
URM Centiva/5, CM-920041, rev. 2.22
Page 17
2 Safety
Risk of electrical hazard
Fire hazard
Risk of electromagnetic
interference
The Centiva/5 is designed to be operated only in electrical environments according to VDE 0107 or EN 60601-1.
Prior to cleaning, repair or maintenance always disconnect the system from the electrical supply. Prior to reconnecting the system to the electrical supply ensure the housing is dry.
Prior to connecting the device to an electrical supply check your local supply voltage to match the operation voltage of the device, which is marked on the serial plate on the rear side of the device.
Store the system only according to the given specifications for temperature and humidity.
Operate the system only according to the given specifications for temperature and humidity. In case system’s temperature is lower or higher than specified, allow the system to stabilize for one hour before operation.
Do not use materials such as Ammonia, Phenol or Acetone to clean this device.
Do not use the Centiva/5 when there is doubt in the proper function of the electrical earth ground in the installation environment
Do not use the Centiva/5 in the presence of flammable anesthetics.
Devices radiating electromagnetic fields in a strength beyond the requirements of EN 60601-1-2 can influence the safe operation of the Centiva/5 and can cause risk to the patient.
Devices radiating high frequency fields and installed near the system can influence the safe operation of the Centiva/5 and can cause risk to a patient.
Do not use cellular phones (mobile phones) within 10 meters of the Centiva/5. Otherwise this may result in a possible influence on the safe operation of the device and can cause risk to a patient.
Do not use the Centiva/5 in the presence of nuclear magnet resonance devices (NMR, MRI).
URM Centiva/5, CM-920041, rev. 2.22
2-5
Page 18
2 Safety
Accessories
Monitoring
Service and maintenance
The Centiva/5 has a wide range of accessories available.
Use only accessories that are listed in the Centiva/5 accessories list.
Prior to connecting a network or any other data management transfer system to the serial output of the Centiva/5 verify that these devices are in compliance with EN 60950 standard for data transfer safety, or EN 60601-1 medical electrical equipment.
The person connecting other systems (electrical or other) to Centiva/5 is responsible that the entire system meets the requirements of EN 60601-1.
The Centiva/5 is equipped with monitoring features that will help to monitor and control the patient’s ventilation situation and to detect changes in the ventilation parameters.
Changes in the ventilation parameters can be caused by:
changes in the patient’s status
changes in the settings
failure in operation
device malfunction
changes in the electrical supply
changes in gas supply
A resuscitator for manual ventilation shall be present to maintain ventilation in case of a device malfunction.
The Centiva/5 shall be checked at least every twelve months for safety and proper function.
Only authorized service personal shall perform repair, service and maintenance to the device.
The maintenance intervals and procedures are documented in section 9, ”Maintenance”.
2-6
URM Centiva/5, CM-920041, rev. 2.22
Page 19
2 Safety
Responsibility of the
manufacturer
Manufactured by
Manufactured for
The manufacturer is not responsible for any function change, damage or injury to patient or operator that is caused by misuse or by disregard of the safety advices mentioned in this manual.
The user or owner is responsible for the function of the system if the device is serviced, maintained or repaired by unauthorized personnel.
The user or owner is responsible for the function of the system if the device is misused or not used according to the instructions in this Operation manual .
Salvia Lifetec GmbH & Co KG Niederhöchstädter Strasse 62 61476 Kronberg Germany
Datex-Ohmeda Inc. Ohmeda Drive P.O. Box 7550 Madison, WI 53707-7550 USA Phone. +1 608 221 1551 Fax: +1 608 223 2476
0123
URM Centiva/5, CM-920041, rev. 2.22
2-7
Page 20
2 Safety
Notes
2-8
URM Centiva/5, CM-920041, rev. 2.22
Page 21
3 General description
In this section
3.1 General description 3-2
3.1.1 Intended use 3-2
3.1.2 Principle of operation 3-2
3.1.3 Ventilation modes 3-3
3.1.4 Additional function 3-4
3.2 Supply 3-5
3.2.1 Electrical supply 3-5
3.2.2 Gas supply 3-5
3.3 Positioning 3-6
3.4 Main components 3-7
3.5 System controls 3-8
3.5.1 Front side 3-8
3.5.2 Rear side 3-9
URM Centiva/5, CM-920041, rev. 2.22
3-1
Page 22
3 General description
3.1 General description
3.1.1 Intended use
3.1.2 Principle of operation
Centiva /5 is a critical care ventilator used for intensive care ventilation of humans.
This ventilator is intended to be used for adult patients and children age 1 year + or bodyweight greater than or equal to 5 kg.
This ventilator is designed to maintain and support existing and returning patient’s spontaneous breathing.
Centiva/5 is a time controlled, volume constant, pressure cycled, software controlled system.
The expiration valve is electromechanically controlled with a locking system to provide proper seating of the PEEP valve assembly.
The system uses a flow trigger and allows patient’s free breathing over a built-in free breathing valve.
The system provides for a permanent Byflow in the breathing circuit and allows patient’s spontaneous breathing in all modes.
3-2
URM Centiva/5, CM-920041, rev. 2.22
Page 23
3 General description
3.1.3 Ventilation modes
VCV (Volume Controlled Ventilation)
Controlled volume constant ventilation with the ability of patient’s spontaneous breathing between the mandatory breaths. Additional capabilities are:
- PLV - Pressure Limited Ventilation
- IRV - Inverse Ratio Ventilation
SIMV - VC (Synchronized Intermittent Mandatory Ventilation – Volume Constant) Synchronized volume constant ventilation with the ability of patient’s spontaneous breathing in between the mandatory breaths. Additional capabilities are:
- PLV - Pressure Limited Ventilation
- IRV - Inverse Ratio Ventilation
- PSV - Pressure Support Ventilation
Bi-Level (Bi-Level) Synchronized pressure controlled ventilation combined with patient’s free spontaneous breathing during the entire breathing cycle. Additional capabilities are:
- IRV - Inverse Ratio Ventilation
- PSV - Pressure Support Ventilation on lower Bi-Level pressure level
CPAP/PSV (Continuous Positive Airway Pressure/ Pressure Support Ventilation) Spontaneous breathing with environmental pressure or positive airway pressure. Additional capabilities are:
- PSV - Pressure Support Ventilation on CPAP pressure level
CPAP/Apnea (Continuous Positive Airway Pressure / Apnea back-up) Spontaneous breathing with environmental pressure or positive airway pressure with PSV - Pressure Support Ventilation on CPAP pressure level Additional capabilities are:
- Bi-Level back up ventilation mode with pre-adjusted settings if patient activates the apnea alarm.
For more details refer to section 6 “operation”.
URM Centiva/5, CM-920041, rev. 2.22
3-3
Page 24
3 General description
3.1.4. Additional functions
APD (Automatic Patient Detection)
Function that provides automatic detection of a patient being connected to the ventilator in all ventilation modes.
ARC (Airway Resistance Compensation) Function that provides adjustable inspiratory and expiratory compensation of the tube’s resistance in all spontaneous breathing modes to ease patient’s spontaneous breathing.
ASR (Automatic Suction Routine) Function that provides an automatic procedure for bronchial suction preparation with a pre-oxygenation phase, a pressure free opening of the y-piece and a post-oxygenation with automatic return to previous settings in all ventilation modes.
NEB (Nebulizer drug delivery) Function that provides drug delivery in all ventilation modes to the patient via the breathing circuit by means of a so called “jet nebulizer”.
NIV (Non-Invasive Ventilation) Function that provides leakage compensation in all ventilation modes to allow ventilation of leaking systems such as nasal or face mask , larynx mask or uncuffed tubes.
OCV (Online O2 sensor Calibration during Ventilation) Function that allows O2 sensor calibration during ventilation.
3-4
URM Centiva/5, CM-920041, rev. 2.22
Page 25
3.2 Supply
3 General description
3.2.1 Electrical supply
3.2.2 Gas supply
Special notice
3.3 Positioning
The Centiva/5 can be configured to be supplied with electrical power from a line supply voltage of 230 V or 115 V AC.
The Centiva/5 can be supplied by a 24 V DC external battery pack (option) for e.g. inner clinical transportation purposes.
As a back-up supply Centiva/5 is provided with an internal battery that supports operation for a minimum of 30 minutes (battery fully charged).
The Centiva/5 is designed to be supplied with O2 and AIR. Single gas operation with either AIR or O2 can be selected .
Ensure clean and dry medical gas supply sources.
Special notice
URM Centiva/5, CM-920041, rev. 2.22
The Centiva/5 is design to be positioned on
The control panel of Centiva/5 is tiltable and detachable. The control panel can be mounted up to a 3 meter distance from the main engine (remote control).
Ensure the Centiva/5 is securely mounted.
a trolley or cart (option)
a rail mount (option)
a ceiling pendulum (option)
3-5
Page 26
3 General description
3.4 Main Components
Main engine
Control panel
Expiration valve
The Centiva/5 is a system that can be separated into three (3) user accessible components
The main engine
The control panel
The expiration valve
These three assemblies can be separated by the user without using any tools.
3-6
URM Centiva/5, CM-920041, rev. 2.22
Page 27
3.5 System controls
I
O
A
3 General description
3.5.1 Front side
Screen
Soft keys
Paw cmH2O
Flow L/min
-20
-40
Fi02
21
Bi-Level
15
10
5
40 20
%
Rate
12
1/min
Tube : E 7.5 50%
I:E
Ramp
1:2
0.2
sec
P-Insp
10
cmH2O
Centiva/5
MV-E
Vt-E
Ppeak
Pmin
O2-I
Rate
PEEP
3
cmH2O
dult
5.2
350
14
3 21 18
PS
cmH2O
cmH2O
cmH2O
1/min
5
Liter
12
2
mL
21
Alarm lights
Silencer
25 18
%
40
8
Menu key
ComWheel
Stopper
Exp. port
Nebulizer
URM Centiva/5, CM-920041, rev. 2.22
Insp. port
ON/OFF
3-7
Page 28
3 General description
3.5.2 Rear side
Serial plate
C inl
e
t
Com p
24V/ 3A DC
2
4
V
D
Gro
und
o
r
t
co
Caution. U.S. Federal laws restrict this device to sale by o r on the order of a l icensed medical practitioner.Outside the U.S.A. check local laws for any restriction that may apply.
Line switch
Lin
e
i
nle
n.
t
O
2
g
as
A
IR
g
a
i
nl
et
s
O2 AIR
280 .. . 600 kPa or 40 ... 80 PSI
in
le
t
3-8
URM Centiva/5, CM-920041, rev. 2.22
Page 29
4 Concept of operation
In this section
4.1 Concept of operation 4-2
4.1.1 Screen content 4-3
4.1.2 Soft keys 4-5
4.1.3 I : E Ratio soft key 4-6
4.1.4 Timing parameter 4-7
4.1.5 Graphics 4-8
4.1.6 Numerical patient data 4-9
4.1.7 Status information 4-10
4.1.8 ComWheel 4-12
4.1.9 Changing a parameter 4-13
4.1.10 Menu button 4-14
4.1.11 Selecting a menu topic 4-15
4.1.12 Activating a menu function 4-16
4.1.13 Submenus 4-17
4.1.14 Open a submenu 4-19
4.1.15 Change a parameter in a submenu 4-20
4.1.16 Alarm silence 4-22
URM Centiva/5, CM-920041, rev. 2.22
4-1
Page 30
4 Concept of operation
I
O
A
4.1 Concept of operation
The Centiva/5 is a menu operated and software controlled system.
All information for control is provided by the Display.
All controls are located on the front panel, except the On-Off button on the lower front right and the line supply switch on the rear.
The principle for all controls is based on the S A C system.
Select
Adjust
Confirm
There are two ways to select a parameter to change, either by soft keys or by menu.
The adjustment is always made by turning the ComWheel clockwise for up and counterclockwise for down.
Confirmation of the adjusted value for the selected parameter is made by pushing the ComWheel.
Screen
Soft keys
Paw cmH2O
Flow L/min
Fi02
21
Bi-Level
15
10
5
40 20
-20
-40
%
Rate
12
1/min
Tube : E 7.5 50%
I:E
Ramp
1:2
0.2
sec
P-Insp
10
cmH2O
Centiva/5
MV-E
Vt-E
Ppeak
Pmin
O2-I
Rate
PEEP
3
cmH2O
dult
5.2
350
14
3 21 18
cmH2O
PS
5
Liter
mL
cmH2O
cmH2O
1/min
12
2
21
25 18
%
40
8
Menu key
ComWheel
4-2
URM Centiva/5, CM-920041, rev. 2.22
Page 31
4 Concept of operation
Softkeys for ventilation parameter settings
4.1.1 Screen content
Soft key parameters
Graphics
Numerical data
The Centiva/5 display shows all information on a color LCD screen. The screen is divided into four different areas.
Area # 1.
This area shows the ventilation parameter settings aligned with the specific soft key. For details refer to soft keys.
Area # 4
Graphical patient data
and alarm messages
Status information
Area # 2
Area # 1
Area # 2
This area shows the measured patient data in graphical format as well as alarm messages.
Area # 3
This area shows the
Area # 4
Graphical patient data
and alarm messages
Softkeys for ventilation parameter settings
Area # 4
Status information
Area # 2
Area # 1
Status information
measured patient data in numeric format.
Area # 2
Graphical patient data
and alarm messages
Softkeys for ventilation parameter settings
Area # 1
Area # 3
numerical
patient
data
Area # 3
numerical
patient
data
Area # 3
numerical
patient
data
URM Centiva/5, CM-920041, rev. 2.22
4-3
Page 32
4 Concept of operation
Status information
Area # 4.
Area # 4
Status information
This area shows the status information for the ventilator.
Area # 2
Graphical patient data
and alarm messages
Area # 3
numerical
patient
data
Softkeys for ventilation parameter settings
Area # 1
4-4
URM Centiva/5, CM-920041, rev. 2.22
Page 33
4 Concept of operation
%
cmH2O
cmH2O
mL
1/Min
L/Min
%
cmH2O
cmH2O
mL
1/Min
cmH2O
%
cmH2O
cmH2O
Sec.
1/Min
cmH2O
%
cmH2O
cmH2O
1/Min
cmH2O
Sec.
N-val
c-val
Val
%
cmH2O
Sec.
cmH2O
4.1.2 Soft keys
Soft keys in VCV
Soft keys in SIMV - VC
Soft keys in Bi-Level
Soft keys in CPAP/PSV
Soft keys in CPAP/Apnea
By using the soft keys the operator selects a parameter to change. Depending on the current ventilation mode the soft keys are related to different parameters. Immediately above the soft key the parameter is displayed in the screen showing the value and units of measure.
Fi O2
50
Rate
10
I : E
1:2.0
Vt
110
Plimit
30
PEEP
OFF
Fi O2
50
Plimit
30
I : E
1:2.0
Rate
10
Vt
110
PEEP
OFF
Fi O2
50
P-Insp
10
I : E
1:2.0
Rate
10
Ramp
0.20
PEEP
OFF
Fi O2
50
Fi O2
50
Rate
10
I : E
1:2.0
Ramp
0.20
Ramp
0.20
P-Insp
10
PEEP
OFF
PEEP
OFF
I-Flow
6.6
PS
10
PS
10
PS
5
PS
10
URM Centiva/5, CM-920041, rev. 2.22
The soft key shows parameter (type) and value
Type
(Val) with blue characters on yellow background.
Activating a soft key will change the background
unit
color from yellow to blue and the character from blue to yellow. When changing the parameter with the ComWheel, the current parameter is displayed as a small character on top to remind the operator
type
of the value (c-val) while making the new parameter change (N- val). After confirming the changed value the soft key returns to blue character on yellow background.
4-5
Page 34
4 Concept of operation
2.0:1
1:2.0
I : E
2.0:1
I : E
9.9:1
3:1.0
I : E
4.1.3 I : E ratio soft key
Inverse I:E ratio
Inverse I:E ratio in SIMV-VC mode
APRV I:E ratio in Bi-Level mode
Note
For standard I : E ratio settings from 1:1 to 1:9.9 the I : E ratio soft key‘s characters and background colors are identical to the other soft keys.
When changing the I : E ratio in all ventilation modes to inverse values (I>E) (range 1:1 to 4:1) the user will be notified by a change of the background color from blue to red and the character color from yellow to white. After confirming the changed value the soft key returns to blue character on yellow background.
Changing the soft key parameter rate in the SIMV­VC mode causes a change of the I:E ratio accordingly to keep the ventilation pattern similar. When this change in I : E ratio leads to inverse values (I>E) the user will be notified by a change of the background color from blue to white and the character color from yellow to red. After confirming the changed value the soft key returns to blue character on yellow background.
Only in Bi-Level mode Centiva/5 offers extreme inverse I:E ratio values to simulate APRV (airway pressure release ventilation).
When changing the I : E ratio in Bi-Level mode beyond the inverse I:E ratio level to extreme inverse I:E ratio values (range 4:1 to 9.9:1) the user will be notified by a change of the background color from red to brown. After confirming the changed value the soft key returns to blue character on yellow background.
When changing the ventilation mode from Bi-Level to any other mode, APRV I:E ratios will not transfer automatically, but the user will be notified to correct the I:E ratio accordingly by a message “I :E > 4:1 !”.
4-6
URM Centiva/5, CM-920041, rev. 2.22
Page 35
4 Concept of operation
%
cmH2O
mL
1/Min
L/Min
cmH2O
t
4.1.4 Timing parameter
The Centiva/5’s timing parameter in all ventilation modes are:
- ventilation rate (rate)
- inspiration to expiration time ratio (I:E)
TV
Fi O2
50
Plimit
30
I : E
1:2.0
Rate
10
Vt
110
PEEP
Off
Changing the I-Flow setting in the ventilation modes VCV and SIMV-VC has an impact on the timing, too. In this case the inspiratory pause will be affected by this parameter.
All timing parameters are relative parameters and do not directly set absolute times. During changing the setting of one of the timing parameters Centiva/5 will display the absolute time information on the screen in the format:
I=xx s, P=yy %, E= zz s
„I“ is inspiration time (T
„P“ is inspiratory pause in % of inspiration time (T
„E“ is expiration time (T
) in seconds
insp
) in seconds.
exsp.
pause
/ T
insp
)
After confirming made changes to one of the timing parameters this absolute time message in the screen disappears automatically.
t
I-Flow
6.6
URM Centiva/5, CM-920041, rev. 2.22
I= 2.0 s, P= 30%, E=3.0 s
4-7
Page 36
4 Concept of operation
4.1.5 Graphics
The Centiva/5 displays two graphical waveforms on the screen:
- current airway pressure versus time
- current airway flow versus time
The upper waveform shows the airway pressure and contains markers for the pressure settings, such as PEEP, Pinsp or Plimit.
The lower waveform shows the airway flow.
The display will automatically change to scale to maximize readability of the waveform.
The graphic waveform data is real time and both waveforms are synchronized on the same time base.
15
10
5
Paw cmH2O
40 20
Flow L/min
-20
-40
4-8
URM Centiva/5, CM-920041, rev. 2.22
Page 37
4 Concept of operation
Val
10
cmH2O
4.1.6 Numerical patient data
The Centiva/5 displays measured patient data in a numeric format.
The expiratory minute volume is measured over a period of 60 secs or 10 breaths.
The expiratory tidal volume is measured during each breath.
The maximum (Peak) airway pressure is measured during each breath.
The minimum (Min) airway pressure is measured during each breath.
The inspiratory O2 concentration is measured at the inspiratory port.
The total patient breath rate is measured over a period of 60secs or 10 breaths and represents spontaneous plus mechanical breaths.
The format of this numeric data field shows the parameter (Para) with the measured value (Val) and the unit of measure (unit).
If alarm limits for this particular parameter are available, the upper limit (UL) and the lower limit (LL) are displayed in the same data field.
MV-E
3.0
TV-E
361
Ppeak
14
Pmin
1
O2-I
21
Rate
Para
cmH2O
12.0
2.5
Liter
mL
21
26 18
%
40
8
1/min
UL LL
Unit
URM Centiva/5, CM-920041, rev. 2.22
4-9
Page 38
4 Concept of operation
Tube: E 7.5 40%
Battery: 28 min
Adult
No Test
4.1.7 Status information
Vent. mode
NEB
ARC
Battery
Patient type
System Test
The Centiva/5 shows status information in the top portion of the display. The following describes the status information in order from left to right appearing on the screen.
The selected ventilation mode is displayed: “VCV”, “SIMV-VC”, “Bi-Level”, “CPAP/PSV”, “CPAP/Apnea”.
When nebulization is activated, an icon is displayed during operation in nebulization mode
When Airway Resistance Compensation is activated, status information is displayed giving the selected tube (E=endotracheal, T= tracheal), size of tube (e.g. 7.5 mm) and level of compensation (e.g. 40%). If ARC is not turned ON this status information field is not present.
When the system is supplied by the internal back-up battery the remaining operation time on battery is displayed. Whenever the internal battery is being charged, the same status information appears. If the system is supplied by line (main) supply and the battery is fully charged this status information field is not present.
Status information about the selected patient type is displayed with either ”Adult” or “Pedia” (pediatric).
When the System test is not performed and the ventilator is started by bypassing the system test, the “No Test” is displayed. If the system test was performed prior to the start of the ventilation this status information field is not present.
Bi-Level
4-10
URM Centiva/5, CM-920041, rev. 2.22
Page 39
4 Concept of operation
Switch off device ? Yes: Enter, No: any button
Only O2
Single gas supply
ASR
Shut down
External DC supply
If the operator has confirmed single gas operation during the system test, this status information field displays “Only AIR” for single gas supply with AIR or “Only O2” for single gas supply with O2. If dual gas supply with O2 and AIR is confirmed this status information field is not present.
When the Automatic Suction Routine is activated the status information field reports the different phases of the ASR procedure:
If ASR is not turned ON this status information field is not present.
When the “System Open?” message is displayed on the screen and the ON-OFF button is activated for longer than 3 seconds, the shut down status information appears. This shut down information indicates that pushing the “<Enter>” button (ComWheel), powers down the Centiva/5 or by pressing any other button the system remains in the normal operation mode. This status information field overwrites all other status information fields and covers the entire status information area on the screen
When an external DC supply (e.g. battery pack) takes over the mains supply, an icon is displayed to indicate the power supply by external battery.
- “Pre oxygen” for the pre-oxygenation phase,
- “Suction” for the suction phase,
- “Post oxygen” for the post oxygenation phase.
Pre oxygen
URM Centiva/5, CM-920041, rev. 2.22
4-11
Page 40
4 Concept of operation
4.1.8 ComWheel
Special notice
The ComWheel is turned to change the selected parameter values .
Turing clockwise increases the value or moves up the cursor in a menu.
Turning counterclockwise decreases the value or moves down the cursor in a menu.
Pushing the ComWheel confirms a selection.
Confirming a selection returns the soft key background color from blue to yellow or changes the highlighted area in a menu. The changed parameter will only become valid when the ComWheel is pushed for confirmation.
When a screen message asks for “<Enter>”, the system asks the operator to push the ComWheel to confirm an action.
4-12
URM Centiva/5, CM-920041, rev. 2.22
Page 41
4 Concept of operation
%
%
4.1.9 Changing a parameter
The following shows how a parameter change is made.
This is the parameter field before we start.
Select the FiO2 soft key.
Notice the color change.
Turn the ComWheel clockwise to increase the value to 50%.
Notice the value change.
The current value is shown on the top and the new value in the middle.
Confirm the selection by pushing the ComWheel.
This is the parameter field after the change.
FiO2
30
30
30
FiO2
30
50
FiO2
FiO2
50
URM Centiva/5, CM-920041, rev. 2.22
4-13
Page 42
4 Concept of operation
A R C
Exit
4.1.10 Menu button
To select additional functions or infrequently used parameters push the menu button.
Within the Centiva/5’s screen the main menu will open. It is here that most functions are directly accessible.
Submenus can be opened for detailed parameter settings.
Stand-by A S R
ON
ON Parameter menu Alarm limits Alarm AUTOSET Vent. modes
Nebulizer
ON
System test
4-14
URM Centiva/5, CM-920041, rev. 2.22
Page 43
4 Concept of operation
4.1.11 Selecting a menu topic
To select a specific menu topic turn the ComWheel to select the menu option and push the ComWheel to confirm the selection.
Turning the ComWheel clockwise will move the cursor up the menu.
Turing the ComWheel counterclockwise will move the cursor down in the menu.
Pushing the ComWheel will confirm the selection
URM Centiva/5, CM-920041, rev. 2.22
4-15
Page 44
4 Concept of operation
A R C
Exit
A R C
Exit
4.1.12 Activating a menu function
The following shows how to activate a menu selection.
Open the main menu by pushing the menu button.
The main menu will appear in the screen.
Rotate the ComWheel to select the Alarm AUTOSET function.
The selected function will be highlighted in the menu.
Confirm the selection by pushing the ComWheel.
The function “Alarm AUTOSET” is activated and the menu closes automatically.
Stand-by A S R Parameter menu Alarm limits Alarm AUTOSET Vent. modes
Nebulizer
System test
Stand-by A S R Parameter menu Alarm limit s
Alarm AUTOSET
Vent. modes
Nebulizer
System test
ON ON
ON
ON ON
ON
4-16
URM Centiva/5, CM-920041, rev. 2.22
Page 45
4 Concept of operation
A R C
Exit
A R C
Exit
Buzzer
PS endflow
Rate min
Rate max
Apn.del.
4.1.13 Submenus
Parameter menu
Alarm limits
The following selections in the main menu open the submenus :
Parameter menu
Alarm limits
Ventilation modes (Vent. Modes)
Airway Resistance Compensation (ARC)
These submenus open upon selecting the topic in the main menu.
Stand-by A S R
Parameter menu
Alarm limits Alarm AUTOSET Vent. modes
Nebulizer
System test
ON ON
ON
I-flow PS ramp Trigg. ByFlow Time window
Previous menu
10.0 L/m
0.20 s
3.0 L/m
3.0 L/m
100%
Exit
60% 25%
Stand-by A S R Parameter menu
Alarm limits
Alarm AUTOSET Vent. modes
Nebulizer
System test
ON ON
ON
MV max MV min Leakage Pmax
Previous menu
1.7 L
0.5 L 25%
40cmH2O
20 Sec
40/Min
8/Min
Exit
URM Centiva/5, CM-920041, rev. 2.22
4-17
Page 46
4 Concept of operation
A R C
Exit
A R C
Exit
CPAP/Apnea
CPAP/PSV
VCV
Vent. modes
A R C
Stand-by A S R Parameter menu Alarm limits Alarm AUTOSET
Vent. modes
Nebulizer
System test
Stand-by A S R Parameter menu Alarm limits Alarm AUTOSET Vent. modes
Nebulizer
System test
ON ON
ON
ON ON
ON
SIMV - VC Bi-Level
Previous menu
Exit
Endot.tube Trach.tube Diameter 7.5 Compensat. 75 % Previous menu
OFF OFF
Exit
4-18
URM Centiva/5, CM-920041, rev. 2.22
Page 47
4 Concept of operation
A R C
Exit
CPAP/Apnea
CPAP/PSV
VCV
4.1.14 Open a submenu
Push the Menu button to open the main menu,
Select the “Vent. modes” by opening the main menu and rotating the ComWheel to the “Vent. Modes” selection.
Confirm this selection by pushing the ComWheel which opens the “Vent. Modes” submenu.
The “Vent. modes” submenu allows you to select the appropriate ventilation mode by rotating the ComWheel and confirming the selection by pushing the ComWheel.
Stand-by A S R Parameter menu Alarm limits Alarm AUTOSET
Vent. modes
Nebulizer
System test
SIMV - VC Bi-Level
Previous menu
ON ON
ON
Exit
URM Centiva/5, CM-920041, rev. 2.22
4-19
Page 48
4 Concept of operation
A R C
Exit
A R C
Exit
4.1.15 Change a
parameter in a sub menu
The following shows how to change the parameter “MV max” in the alarm limit submenu.
Open the main menu by pushing the menu button.
The main menu will appear in the screen.
Stand-by A S R Parameter menu Alarm limits Alarm AUTOSET Vent. modes
Nebulizer
System test
Select the “Alarm limits” by rotating the ComWheel.
The selected function will be highlighted in the menu.
Stand-by A S R Paramet er menu
Alarm limits
Alarm AUTOSET
Vent. m odes
Nebulizer
System test
Confirm the selection by pushing the ComWheel. The “Alarm limits” sub menu will open.
ON ON
ON
ON ON
ON
4-20
URM Centiva/5, CM-920041, rev. 2.22
Page 49
4 Concept of operation
Rate min
Rate max
Apn.del.
Rate min
Rate max
Apn.del.
Rate min
Rate max
Apn.del.
1.7 L
The “Alarm limits” submenu opens.
Select the “MV max” by rotating the ComWheel.
MV max MV min Leakage
40cmH2O
Pmax
20 Sec
40/Min
Previous menu
Exit
1.7 L
0.5 L 25%
8/Min
The selected function will be highlighted in the menu.
Confirming the selection highlights the value in the “MV max” setting.
The selected parameter value is highlighted.
MV max
MV min Leakage Pmax
Previous menu
MV max
MV min 0.5 L Leakage Pmax
Previous menu
1.7 L
0.5 L 25%
40cmH2O
20 Sec
40/Min
8/Min
Exit
25%
40cmH2O
20 Sec 40/Min
8/Min
Exit
URM Centiva/5, CM-920041, rev. 2.22
4-21
Page 50
4 Concept of operation
Rate min
Rate max
Apn.del.
2.5 L
Rate min
Rate max
Apn.del.
4.1.16 Alarm silence
Rotating the ComWheel adjusts the new value.
The MV max is changed to 2.5 L.
Confirming the selection activates the new value for the alarm limit MV max.
Selecting the “Previous menu” selection returns you to the main menu. Confirming the “exit” function will close the menu window or if left open for 30 secs the menu will automatically close.
MV max
MV min 0.5 L Leakage Pmax
Previous menu
MV max MV min Leakage Pmax
Previous menu
25%
40cmH2O
20 Sec
40/Min
8/Min
Exit
2.5 L
0.5 L 25%
40cmH2O
20 Sec
40/Min
8/Min
Exit
Pushing the alarm silence button during a high priority alarm silences the audible alarm for two (2) minutes. A clock in the screen starts displaying the remaining alarm silence time. When an alarm becomes inactive , the alarm message stays on the screen, but changes color from red to yellow and the audible alarm turns OFF. This is to remind the operator that an alarm situation had occurred . Pushing the alarm silence button again will remove the yellow alarm message from the screen.
4-22
URM Centiva/5, CM-920041, rev. 2.22
Page 51
5 Preparation
In this section
5.1 Electrical supply 5-2
5.1.1 Mains supply 5-2
5.1.2 Back-up battery supply 5-4
5.2 Gas supply 5-5
5.2.1 Dual gas supply 5-5
5.2.2 Single gas supply 5-6
5.3 Set up of expiration valve 5-7
5.4 Breathing gas conditioning 5-8
5.5 Breathing circuit 5-9
5.5.1 Connection with HME 5-9
5.5.2 Connection with active humidifier 5-10
5.6 Placing the device 5-12
5.6.1 Control panel 5-12
5.6.2 Entire system 5-14
5.7 Turn ON and OFF 5-15
5.8 Self test 5-16
5.8.1 Power-up test 5-16
5.8.2 System test 5-17
5.8.3 In-operation test 5-18
5.9 Select the type of patient 5-20
5.11 Starting the ventilator 5-22
URM Centiva/5, CM-920041, rev. 2.22
5-1
Page 52
5 Preparation
5.1 Electrical supply
5.1.1 Mains Supply
The Centiva/5 is designed to operate from electrical supplies:
Mains supply
External battery.
The system will treat all of these supply sources as a main supply.
Verify your local mains supply voltage matches the rated device’s voltage on the serial plate (rear side).
Connect the power cord to the mains inlet on the rear of the device.
24V/3A DC
Serial plate
Caution. U.S. Federal laws restrict this device to sale by or on the order of a licensed medical practitioner.Outside the U.S.A. check local laws for any restriction that may apply.
Lin
e
s
wit
ch
Line
i
nl
e
t
O2 AIR
280 ... 600 kPa or 40 ... 80 PSI
Connect the power cord to the wall outlet.
Switch ON the mains switch.
Verify the ON-OFF button on the front is lit.
5-2
URM Centiva/5, CM-920041, rev. 2.22
Page 53
5 Preparation
External battery
Place the external battery (option) below the Centiva/5 housing and lock the battery housing to the Centiva/5 housing .
Caution. U.S. Federal laws restrict this device to sale by or on the order of a l icensed medical practi tioner .Outsid e the U.S.A . chec k local laws for any restriction that may apply.
2
4
V
D
C
inle
24V/3A DC
t
O2 AIR
280 ... 600 kPa or 40 ... 80 PSI
Out: 24V/
3 A DC
In: 27,5V/ 0,25 A DC
3,15 A T
Locking
Connect the DC supply power cord from the external battery pack to the blue inlet and verify proper locking of the connector.
Special Notice
Verify the ON-OFF button on the front is lit.
Verify the external battery is fully charged prior to usage. When fully charged the battery’s charging indicator is lit permanently in green.
When powered by the external battery pack, Centiva/5 displays an battery icon in the top left of the screen
For more details to the external battery refer to the operation manual of this accessory.
URM Centiva/5, CM-920041, rev. 2.22
5-3
Page 54
5 Preparation
5.1.2 Back-up battery supply
Special notice
WARNING
The Centiva/5 is equipped with an internal back-up battery supply that provides at least 30 minutes of operation upon loss of electrical main supply.
This back-up supply will automatically switch on whenever the main supply is lost and will maintain uninterrupted operation of the Centiva/5.
When operating on the back-up battery, the system will display a message in the operation screen (no line supply).
At the same time the battery status in the status line of the screen will display the remaining operation time on battery.
When operating on back-up battery the system reduces power consumption to save energy for extended battery life by turning off the expiration valve heating system and by reducing the screen brightness.
Upon restoring line (main) supply the system automatically switches back to the line supply and changes the message in the screen (no line supply) from red to yellow.
Upon restoring line (main) supply the system automatically starts recharging the internal battery.
Whenever the Centiva/5 is supplied by any main supply (line, 24 VDC or external battery) and the system is turned ON, the internal back-up battery is recharged. The recharging process is indicated by the battery status displayed in the status line of the screen (with the timer counting up ). The re-charger is not active when the system is turned OFF.
Battery 28 min
Do not use the internal back-up battery as a main supply.
When turning the Centiva/5 ON without line (main) supply, the system powers up on the back-up battery to allow opening of the system test menu only, (e.g. to release the expiration valve for cleaning purpose). Without line (main) supply the Centiva/5 will not go into a ventilation mode. After having passed the power-up test and the system test, the back-up battery function is activated and will perform uninterrupted operation upon loss of line (main) supply.
5-4
URM Centiva/5, CM-920041, rev. 2.22
Page 55
5.2 Gas supply
5 Preparation
5.2.1 Dual gas supply
The Centiva/5 is equipped with an electronic O2/AIR mixer to provide a wide range of inspiratory oxygen concentrations. This requires the supply of both O2 and AIR gases. The gas supply range for both gases ranges from 40 to 80 PSI. The Centiva/5 is protected against overpressure up to 140 PSI. The Centiva/5 provides high peak flow capability for optimized pressure support ventilation. This capability requires a flow of 120 l/min per gas.
Verify your gas supply pressure matches the pressure range mentioned above.
24V/3A DC
Connect the O2 gas supply hose to the O2 gas inlet on the rear of the device. Connect the O2 gas supply hose to the wall outlet and check the connection for leaks.
Connect the AIR gas supply hose to the AIR gas inlet on the rear of the device. Connect the AIR gas supply hose to the wall outlet and check the connection for leaks.
Caution. U.S. Federal laws restrict this device to sale by or on the order of a licensed medical practitioner.Outside the U.S.A. check local laws for any restriction that may apply.
O2 gas
A
IR
g
a
s
i
O2 AIR
nl
280 ... 600 kPa or 40 ... 80 PSI
e
t
i
nle
t
WARNING
WARNING
URM Centiva/5, CM-920041, rev. 2.22
Use only gas supply hoses in compliance with the local standards. Incorrect connecting of O2 and AIR supply can result in risk to the patient.
Only supply the system with dry and clean gases to prevent water, oil or particles to enter the system. Dirt, water or oil can damage the system and may result in risk to the patient.
5-5
Page 56
5 Preparation
OnlyAIR
Only O2
5.2.2 Single gas supply
Special notice
The Centiva/5 needs at least 40 PSI from the gas supply for proper operation. If this pressure level is not sufficient from one gas source the system automatically switches to the gas supply which has sufficient pressure and displays an alarm message on the screen.
The Centiva/5 can operate with a single gas supply without alarms. When the system recognizes the single gas supply during the system test it prompts a message on the screen and asks for single gas operation.
In the event of a single gas supply (e.g. AIR) the following screen will appear during system test.
Restoring the O2 supply at this time in the system test removes the message automatically and the system test proceeds. Confirming the single gas operation for AIR­supply by pushing the ComWheel to <Enter> will cause the system test to proceed , but all tests that use O2 will be skipped .
After the system test is completed and ventilation has started, the system will change the default setting for FiO2 according to the selected single gas operation to avoid nuisance alarms. In the single gas AIR operation the FiO2 default is 21 %. In the single gas O2 operation the FiO2 default is 100%.
The single gas operation is displayed in the status line with "Only Air" when lacking O2 supply or “Only O2” when lacking AIR supply.
Restoring the missing gas during operation allows the unit to operate in the dual gas supply mode . The system will still display the selected single gas supply mode until a new system test is performed with both gases available.
System test Rev. x.nn
No O2 gas supply Check O2 gas supply or for single gas AIR-supply press <Enter>
Centiva/5
5-6
URM Centiva/5, CM-920041, rev. 2.22
Page 57
5.3 Set up of expiration valve
The Centiva/5’s expiration valve is removable from the device for easy cleaning. Upon re-insertion it will lock in automatically for safe operation.
The valve assembly consists of three components:
Special notice
the valve block
the PEEP membrane
Verify the PEEP membrane assembly is inserted in the valve block with the note “TOP” facing you.
Verify the membrane plate (stainless steel) is positioned on top of the membrane assembly.
The expiratory valve assembly should slide easily into the mating slot on the side of the ventilator
The expiration valve assembly should be inserted smoothly . If you feel a high resistance when inserting the assembly do not force it in but check for proper assembly of the expiration valve.
5 Preparation
assembly
the membrane plate
URM Centiva/5, CM-920041, rev. 2.22
5-7
Page 58
5 Preparation
5.4 Breathing gas conditioning
An intubated patient’s physiological function for heating and humidifying the breathing gases is bypassed while on a ventilator. Therefore, it is necessary to care for this missing function by using either a passive heat and moisture exchanger (HME) or an active heated humidifier.
The use of high efficient humidification and heating with low flow resistance is preferred.
Do not use active and passive humidification at the same time in the same breathing circuit !
5-8
URM Centiva/5, CM-920041, rev. 2.22
Page 59
5.5 Breathing circuit
Insp. line
I
O
A
WARNING
5.5.1 Connection with HME
5 Preparation
The Centiva/5 ventilator can be used with all approved breathing circuits and components, noted in the Centiva/5 accessories list.
In pediatric ventilation, in general (tidal volumes less than 300 ml), the use of pediatric breathing circuits is preferred.
To avoid restriction of the ventilator’s performance, use of breathing circuits with low resistance and a flow optimized y-piece is preferred.
During the system test, compliance, resistance and leakage rate of the breathing circuit is measured and displayed.
Do not use antistatic or conductive tubes or breathing circuits. Risk of electrical hazard to the patient !
Use of a support arm (option) to hold up the breathing circuit between the device and patient is recommended.
Connect inspiratory tube to the right, front 22 mm connector (inspiratory output).
Connect expiratory tube to the left, front 22 mm connector (expiratory input).
Connect y-piece to open end of inspiratory and expiratory tube
Connect HME to the y-piece and connect patient to the HME.
Check for proper seating of all connections.
Bi-Level / ASB
15
10
5
Paw mbar
40
20 Flow L/min
-20
-40
Fi02
21
%
Rate
12
1/min
Tube : E 7. 5 50%
I:E
1:2
Ramp
0.2
Centiva/5
dult
12
MV-E
2
4.2
Liter
Vt-E
350
mL
21
Ppeak
14
cmH2O
Pmin
3
cmH2O
25
O2-I
18
21
%
40
Rate
8
12
P-Insp
sec
1/min
PEEP
P-ASB
10
3
5
cmH2O
cmH2O
cmH2O
URM Centiva/5, CM-920041, rev. 2.22
HME
Exp. line
5-9
Page 60
5 Preparation
I
O
A
I
O
A
5.5.2 Connection with active humidifier
Adjust the humidifier rail of the cart according to the height requirement of the active humidifier so that between the upper level of the humidifier and the lowest part of the drawer or
Bi-Level / ASB
15
10
5
Paw mbar
40
20 Flow L/min
-20
-40
Fi02
21
%
Rate
12
1/min
Tube : E 7.5 50%
I:E
1:2
Ramp
P-Insp
0.2
10
sec
cmH2O
plate of the cart is a minimum of 15 cm.
Release the screw in front of the rail mount to change the position of the rail and tighten it, when the correct position is found.
Place the active humidifier onto the rail and properly fix it there.
Fill humidifier chamber or use a
Bi-Level / ASB
Tube : E 7.5 50%
pre-filled humidifier chamber according to manufacturer’s advice.
15
10
5
Paw mbar
40
20 Flow L/min
-20
-40
Fi02
Rate
I:E
Ramp
21
%
P-Insp
12
1:2
0.2
10
1/min
sec
cmH2O
Connect humidifier electrical supply according to manufacturer’s advice.
Turn ON humidifier and verify proper function.
PEEP
3
cmH2O
PEEP
3
cmH2O
Centiva/5
dult
MV-E
4.2
Vt-E
350
Ppeak
14
Pmin
O2-I
21
Rate
12
P-ASB
Centiva/5
dult
MV-E
4.2
Vt-E
350
Ppeak
14
Pmin
O2-I
21
Rate
12
P-ASB
cmH2O
12
2
Liter
mL
21
cmH2O
3
cmH2O
25 18
%
40
8
1/min
5
cmH2O
12
2
Liter
mL
21
cmH2O
3
cmH2O
25 18
%
40
8
1/min
5
5-10
URM Centiva/5, CM-920041, rev. 2.22
Page 61
5 Preparation
I
O
A
Connect an inspiratory tube from the right, front 22 mm connector of the Centiva/5 to the inlet of the humidifier chamber.
Connect the inspiratory tube from to the outlet of the humidifier chamber to the y-piece.
Connect an expiratory tube from the left, front 22 mm connector of the Centiva/5 and to a water trap.
Connect another expiratory tube from the water trap to the y-piece.
Water trap
Bi-Level / ASB
15
10
5
Paw mbar
40
20 Flow L/min
-20
-40
Fi02
21
%
Rate
12
1/min
Tube : E 7.5 50%
I:E
1:2
When using active humidifiers do not use HMEs between y-piece and patient at the same time. This may result in increased breathing resistance and risk to the patient .
Special notice
Active humidifiers in general increase the resistance of the breathing system. Care should be taken to keep the resistance as low as possible.
Ramp
Centiva/5
dult
12
MV-E
2
4.2
Liter
Vt-E
350
mL
21
Ppeak
14
cmH2O
Pmin
3
cmH2O
25
O2-I
18
21
%
40
Rate
8
12
1/min
P-Insp
PEEP
0.2
sec
P-ASB
10
3
5
cmH2O
cmH2O
cmH2O
URM Centiva/5, CM-920041, rev. 2.22
5-11
Page 62
5 Preparation
I
O
5.6 Placing the device
5.6.1 Control panel
For better visibility and easy handling the Centiva/5 control panel can be adjusted to optimize the viewing angle.
To adjust the viewing angle, the upper part of the control panel can be moved forward or backward.
The control panel can be detached from the main engine and mounted a distance of up to 3 meters away from the main engine.
Ramp
0.2
Centiva/5
Adult
12
MV-E
2
4.2
Liter
Vt-E
350
mL
21
Ppeak
14
cmH2O
Pmin
3
cmH2O
25
O2-I
18
21
%
40
Rate
8
12
1/min
P-Insp
PEEP
P-ASB
10
3
sec
5
cmH2O
cmH2O
cmH2O
Bi-Level / ASB
15
10
5
Paw mbar
40
20 Flow L/min
-20
-40
Fi02
21
%
Rate
12
1/min
Tube : E 7. 5 50%
I:E
1:2
For this application use the optional extension cable and mounting accessories.
Positioning of control panel
At the main engine with adjustable viewing angle
At a wall rail mount (optional) with adjustable viewing angle
At a bed mount (optional) with adjustable viewing angle
5-12
URM Centiva/5, CM-920041, rev. 2.22
Page 63
5 Preparation
To mount the Centiva/5 control panel away from the Centiva/5 main engine , follow the instructions below.
Place the selected mounting bracket to the desired position.
Turn OFF the Centiva/5.
Move control panel to a full angled position.
Press both stainless steel buttons of the bracket mount together to release the locking mechanism.
Move the control panel carefully to the front and up.
Disconnect the control panel from the main engine.
Place the control panel onto the selected bracket.
Connect control panel and control panel bracket and secure proper seated connector.
Press both stainless steel buttons of the bracket mount together to release the locking mechanism. Insert the bracket and release the two buttons to activate the locking mechanism.
Reconnect the control panel and main engine using the extension cable.
Turn ON the Centiva/5 and check for proper function by passing the “power-up” test.
WARNING
Do not disconnect the control panel during operation. This may result in risk to the patient.
URM Centiva/5, CM-920041, rev. 2.22
5-13
Page 64
5 Preparation
H
ous
i
g
o
t
I
O
A
5.6.2 Entire system
Placement of the entire unit or main engine:
On the trolley (optional)
On other flat surfaces
On a wall rail mount (optional)
On an ICU cabinet
When placing the ventilator on the optional trolley ensure that the feet match with the trolley’s receptacles.
Ensure the locking mechanism locks in, when the device is properly placed.
To release the device pull the metal knob of locking mechanism forward and remove the device.
Hous
Bi-Level / ASB
Tube : E 7 . 5 50%
15
10
5
Paw mbar
40
20 Flow L/min
-20
-40
Fi02
Rate
I:E
21
%
t
o
fo
g
in
Ramp
12
1:2
1/min
n
i
p
g
in
k
c
o
L
Centiva/5
dult
12
MV-E
2
4.2
Liter
Vt-E
350
mL
21
Ppeak
14
cmH2O
Pmin
3
cmH2O
25
O2-I
18
21
%
40
Rate
8
12
P-Insp
0.2
sec
1/min
PEEP
P-ASB
10
3
5
cmH2O
cmH2O
cmH2O
f
n
Always ensure proper seating of the system and activated locking mechanism, when applicable.
o
5-14
URM Centiva/5, CM-920041, rev. 2.22
Page 65
5.7 Turn ON and OFF
I
O
A
5 Preparation
Turn ON
Turn OFF
The Centiva/5’s ON-OFF button is located on the lower right side of the ventilator engine .
When supplied with line (mains) power (line switch on the rear should always be turned ON) the ON-OFF button is illuminated by a green light and the unit turns ON when the ON-OFF button is pressed.
If no mains supply is present, or the mains switch is off, the ON-OFF button is not illuminated. Pressing the ON-OFF button with no mains supply activates the internal back-up battery to allow for service activities, such as removing the expiratory valve assembly . For safety reasons the Centiva/5 will not go into normal operation when the main supply is missing.
Bi-Level
15
10
5
Paw cmH2O
40
20 Flow L/min
-20
-40
Fi02
Rate
21
12
%
1/min
Tube : E 7.5 50%
I:E
Ramp
1:2
0.2
Centiva/5
dult
12
MV-E
2
5.2
Liter
Vt-E
350
mL
21
Ppeak
14
cmH2O
Pmin
3
cmH2O
25
O2-I
18
21
%
40
Rate
8
18
P-Insp
sec
1/min
PEEP
PS
10
3
5
cmH2O
cmH2O
cmH2O
The Centiva/5 can only be turned OFF when the patient is not being ventilated and is disconnected (Status: System open ?).
In the non-ventilating “System open ?” mode, or in any other non-ventilating mode, press and hold the ON­OFF button for approximately 3 secs.
Turn OFF device? Yes= Enter , No= any button
Centiva /5
System test Rev. 2.1n
System test OK. To start ventilation <Enter> Compliance 2.3 ml/cmH2O
Resistance 0.8 cmH2O/L/s Leakage 15 ml/min
Adult Pediatric
A message appears asking the user to either turn off the system by confirming with the enter key (ComWheel) or returning to the current status by pressing any other key.
ON/OFF
URM Centiva/5, CM-920041, rev. 2.22
5-15
Page 66
5 Preparation
Start system test <Enter>
I
O
A
5.8 Self test
5.8.1 Power-up test
The Centiva/5 software includes self tests that determine if the operating software is functioning properly and if the electronic and pneumatic circuits are functional. The self tests include:
Power-up test
System test
In-operation test
The power-up test starts automatically when the system is turned ON. The following screen is displayed:
System test Rev. x.nn
Test active, wait for 5 secs
Centiva/5
Message box
The power-up test takes at least 5 secs. The test time counts down on the screen.
During the power-up test, the ventilator checks the proper function of all essential electronic components. If a malfunction is identified during this power up test, the corresponding message appears in the message box area. For details of error messages refer to section 8 “Alarms and messages”.
After having passed the
System test Rev. x.nn
Complete system set up, seal y-piece
Centiva/5
power-up test, the Centiva/5 prompts with a short audible sound and moves to the “Start
Adult Pediatric
system test” screen:
Connect the breathing circuit to the system and seal the Y-piece by placing the open end onto the sealer on the exp. valve assembly.
Bi-Level
Tube : E 7.5 50%
15
10
5
Paw cmH2O
40
20 Flow L/min
-20
-40
Fi02
Rate
I:E
21
%
Ramp
12
1:2
0.2
1/min
Centiva/5
dult
12
MV-E
2
5.2
Liter
Vt-E
350
mL
21
Ppeak
14
cmH2O
Pmin
3
cmH2O
25
O2-I
18
21
%
40
Rate
8
18
1/min
P-Insp
PEEP
PS
10
3
sec
5
cmH2O
cmH2O
cmH2O
5-16
Stopper
URM Centiva/5, CM-920041, rev. 2.22
Page 67
5 Preparation
System test
5.8.2 System test
Push the ComWheel to confirm the <Enter> command.
This starts the system test.
This system test takes at least 40 secs. The test time counts down on the screen.
During the system test, the system will check the proper function of all pneumatic and electronic components together with the function of the attached breathing circuit . If any malfunction is identified during this system test, the corresponding message appears in the message box area. For details of error messages refer to section 8 “Alarms and messages”.
After having passed the system test, the system prompts the user to open the Y-piece which allows the testing of inspiratory resistance.
Disconnect the Y-piece from the sealer.
The system will recognize that the Y-piece is open automatically and will proceed with the system test.
Test active, wait for 35
System test Rev. x.nn
Open Y-piece
Centiva/5
Message
Adult
Centiva/5
Adult
URM Centiva/5, CM-920041, rev. 2.22
5-17
Page 68
5 Preparation
5.8.3 In-operation test
After measuring the inspiratory resistance, the Centiva/5 displays the following screen:
The compliance and leakage values represent the entire breathing circuit and ventilator. This value is used for the internal compliance compensation.
The resistance value represents the inspiratory resistance of the breathing circuit and ventilator to the y-piece. This value is used for the internal resistance compensation .
The nebulizer flow value represents the measured nebulizer flow of the connected nebulizer. Incase no nebulizer was connected during system, the last measured nebulizer flow is displayed with a note added (last value).
In case the expiration flow sensor was readjusted during system test the according message “Exp. flow sensor data re-adjusted” is displayed.
Values for compliance, resistance and leakage depend upon the breathing circuit used.
Values for nebulizer flow depend upon the nebulizer used (if applicable)
For details to the nebulizer test refer to section 6. 3.10. “Nebulizer”.
Filters and HMEs may significantly increase the resistance values.
System test Rev. x.nn
System test OK. To start ventilation <Enter> Compliance 2.3 ml/mbar
Resistance 0.8 mbar/L/s Leakage 15 ml/min Nebulizer flow 7,0 l/min
Exp. flow sensor data re-adjusted
Centiva /5
Adult
Do not perform the system test with a filter or HME on the breathing circuit. Inappropriate pressure and flow compensation may result.
Other breathing circuit designs, such as co-axial systems, may increase the measured resistance values.
Always use breathing circuits with low inspiratory resistance.
During operation the Centiva/5 continuously checks all safety relevant functions and parameters. If a malfunction is identified, the system will display alarms or messages on the screen. For details of error messages refer to section 8 “ Alarms and messages”.
5-18
URM Centiva/5, CM-920041, rev. 2.22
Page 69
5.9 Select the type of patient
System test
The Centiva/5 allows for the pre-adjustment of ventilator settings and alarm limits according to two different types of patients:
Adult
Pediatric
Prior to the system tests the user can select the desired patient type on the screen by pressing either of the soft keys below the patient types.
Activating the new patient type changes the color of the button in the screen from gray to yellow to prompt the change.
This pre-selection is important to establish default settings in the different ventilation modes. For details see section 6.1.2 “Default ventilation values”.
The selected patient type is displayed during system test in the screen’s lower left corner and during operation within the status information area.
The default settings provide a minimum ventilation for the selected patient type, while the patient is connected and prepared. After preparation and proper connection the settings must be adjusted according to the patient’s needs.
Centiva/5
System test OK. To start ventilation Compliance 2.3 ml/cmH2O
Resistance 0.8 cmH2O/L/s Leakage 15 ml/min
Adult
Pediatric
5 Preparation
URM Centiva/5, CM-920041, rev. 2.22
5-19
Page 70
5 Preparation
System Test
Bypass system
Unlock expiration
Calibrate O2
5.10 System test menu
Bypass system test
After having successfully passed the power-up and/or the system tests the system test menu can be activated by pressing the menu button.
The system test menu appears and allows for
Centiva/5
the selection of certain service functions. Selecting “Exit” closes the system test menu
Complete system set up, seal Y­Start system test by pressing
without any activity.
Adult
Service functions available in the system test menu are:
Bypass system test
Unlock expiration valve
Calibrate O2 sensor
The required system test is bypassed with this function.
Bypass system test
Unlock expiration valve
Calibrate O2 sensor
Exit
Exit
5-20
WARNING
Use the system test bypass only in case of emergency.
In this situation the system uses the compliance and resistance data from the last system test for all internal compensations. If the current breathing circuit differs significantly to the past circuit, differences in ventilation parameters due to changes in the compensation process are possible. This may result in risk to the patient.
Starting the ventilator without having performed the system test will result in a status display “no test” to remind the user that the system was started without a system test and to perform the system test as soon as possible.
URM Centiva/5, CM-920041, rev. 2.22
Page 71
5 Preparation
Unlock expiration valve
O2 sensor calibration
This function releases the expiration valve assembly.
Releasing the expiration valve is necessary for cleaning purposes only.
After inserting the expiration valve it will automatically lock in place and the system performs a power­up test.
Releasing the expiratory valve will work with internal back­up battery power. Locking only works with line (mains) power supply.
System test Rev. x.nn
Expiration valve unlocked Ready to be removed from device
This function starts an automatic O details refer to section 9 “Maintenance”.
Bypass system test
Unlock expiration valve
Calibra t e O2 se ns or
Exit
Centiva/5
sensor calibration procedure. For
2
URM Centiva/5, CM-920041, rev. 2.22
5-21
Page 72
5 Preparation
5.11 Starting the ventilator
After having passed the system test the Centiva/5 will go to normal operation when the <Enter> button (ComWheel) is pressed.
Push the ComWheel to start ventilation.
The Centiva/5 defaults to the ventilation mode ”Bi-Level“ with the ventilator settings and alarm limits set according to the selected patient type.
Centiva /5
System test Rev. x.nn
System test OK. To start ventilation <Enter> Compliance 2.3 ml/mbar
Resistance 0.8 mbar/L/s Leakage 15 ml/min
Nebulizer flow 7,0 l/min
Adult
5-22
URM Centiva/5, CM-920041, rev. 2.22
Page 73
6 Operation
In this section
6.1 Adjust ventilation modes 6-2
6.1.1 Select a ventilation mode 6-2
6.1.2 Default ventilation values 6-3
6.1.3 Changing ventilation modes 6-4
6.2 Adjust ventilation parameters 6-6
6.2.1 VCV 6-6
6.2.2 SIMV-VC 6-9
6.2.3 Bi-Level 6-12
6.2.4 Bi-Level/PSV 6-14
6.2.5 CPAP/PSV 6-17
6.2.5 CPAP/Apnea 6-19
6.3 Additional functions 6-21
6.3.1 Standby 6-22
6.3.2 ASR (Automatic Suction routine) 6-24
6.3.3 Parameter menu 6-27
6.3.4 Alarm lim its 6-32
6.3.5 Alarm AUTOSET 6-38
6.3.6 ARC (Airway Resistance Compensation) 6-41
6.3.7 NEB (Nebulizer drug delivery) 6-44
6.3.8 System test 6-50
6.3.9 NIV (Non Invasive Ventilation) 6-51
6.3.10 APD (Automatic Patient Detection) 6-53
6.3.11 OCV (Online O2 sensor Calibration) 6-54
6.3.12 Short cuts 6-55
URM Centiva/5, CM-920041, rev. 2.22
6-1
Page 74
6 Operation
A R C
Exit
CPAP/Apnea
CPAP/PSV
VCV
6.1 Adjust ventilation modes
6.1.1 Select a ventilation mode
The Centiva/5 uses the “Bi-Level” as the default mode.
The system uses default settings according to the selected patient type to provide a minimum level of ventilation performance.
To change the default ventilation mode to another mode open the main menu with the menu button.
Using the ComWheel select “Vent. Modes” to access the ventilation modes menu.
The “Vent. Modes” topic is highlighted.
Stand-by A S R Parameter menu Alarm limits Alarm AUTOSET
Vent. modes
Nebulizer
System test
Confirm the selection by pushing the ComWheel.
ON ON
ON
6-2
The ventilation modes menu appears.
SIMV - VC Bi-Level
Make your selection with the ComWheel for the desired ventilation mode and confirm the selection by pushing the ComWheel.
Previous menu
Exit
Adjust the ventilation parameter settings to the patient’s needs and set the alarm limits accordingly.
URM Centiva/5, CM-920041, rev. 2.22
Page 75
6 Operation
6.1.2 Default ventilation values
Pediatric
When starting the Centiva/5 the system will provide default ventilator settings to enable easy preparation of the patient.
These settings are designed for the selected patient range to maintain a minimum ventilation performance . The default ventilation mode is “Bi-Level”.
Parameter VCV SIMV-VC Bi-Level CPAP/PSV
FiO2 50 % 50 %
Rate 16 /min 16 /min
I : E 1 : 2 1 : 2
Vt 100 ml 100 ml
Ramp - -
Pinsp - -
50 %
16 /min
1 : 2
-
0.2 sec
7 cmH
2
O
Plimit 20 cmH2O 20 cmH2O - -
PEEP Off Off
I-Flow 10 l/min 10 l/min
PS - 3 cmH2O
Off
-
3 cmH
O
2
50 %
-
-
-
-
-
Off
-
3 cmH
O
2
Adult
URM Centiva/5, CM-920041, rev. 2.22
Parameter VCV SIMV-VC Bi-Level CPAP/PSV
FiO2 50 % 50 %
Rate 10 /min 10 /min
50 %
10 /min
50 %
-
I : E 1 : 2 1 : 2 1 : 2 -
Vt 500 ml 500 ml
Ramp - -
Pinsp - -
Plimit 30 cmH2O 30 cmH2O
PEEP Off Off
I-Flow 30 l/min 30 l/min
PS - 5 cmH2O
-
0.2 sec
10 cmH
-
Off
-
5 cmH
-
-
O
-
2
-
Off
-
O
2
5 cmH
2
6-3
O
Page 76
6 Operation
CPAP/Apnea
CPAP/PSV
VCV
CPAP/Apnea
CPAP/PSV
VCV
6.1.3 Changing ventilation modes
CAUTION
When changing from an active ventilation mode to another mode, the Centiva/5 will transfer the settings for identical parameters (e.g. FiO2) to the new mode.
For those non-transferable parameters algorithm are used to calculate settings for the new ventilation mode based on measured data. These algorithm are based on a calculation taking into account that the lung’s compliance and resistance will not change significantly when changing the mode.
Example:
Change from Bi-Level mode
to VCV (or SIMV-VC) mode:
FiO2 50 % - 50% Rate 10 /min - 10 /min I : E 1 : 2 - 1 : 2 Ramp 0.2 sec - N/a Pinsp 10 cmH2O - N/a Vt N/a 350 ml ( C ) 330 ml PEEP 3 cmH2O - 3 cmH2O
In Bi-Level the patient’s lung compliance is calculated and tidal volume setting for the VCV or SIMV-VC mode is calculated according to Vt set = P
insp
x C.
In case compliance calculation is not leading to stable data, the system uses the lowest value for that mode.
The data used is stored in the Centiva/5 as long as the system is not turned OFF. That provides automatic access to this data when ventilation is restarted after standby. Whenever the system is turned OFF and is turned ON again, the default settings are used.
SIMV - VC
Bi-Level
SIMV - VC
Previous menu
Exit
Bi-Level
Previous menu
Exit
Bi-Level VCV Parameter
Settings Measured Settings
Ensure all parameter settings and alarm limits are set according to the patient’s situation.
6-4
URM Centiva/5, CM-920041, rev. 2.22
Page 77
6 Operation
CPAP/Apnea
CPAP/PSV
VCV
CPAP/Apnea
CPAP/PSV
VCV
Example:
Change from VCV (or SIMV-VC) mode
to Bi-Level mode:
VCV Bi-Level Parameter
Settings Measured Settings FiO2 50 % - 50% Rate 10 /min - 10 /min I : E 1 : 2 - 1 : 2 Ramp N/a - Default Pinsp N/a ( C ) 10 cmH2O Vt 350 ml - N/a PEEP 3 cmH2O - 3 cmH2O
In VCV or SIMV-VC the patient’s lung compliance is calculated and inspiratory pressure setting for the Bi-Level mode is calculated according to P
= Vt / C.
insp
In case compliance calculation is not leading to stable data, the system uses the lowest values for that mode.
The data used is stored in the Centiva/5 as long as the system is not turned OFF. That provides automatic access to this data when ventilation is restarted after standby. Whenever the system is turned OFF and is turned ON again, the default settings are used.
SIMV - VC
Bi-Level
Previous menu
Exit
SIMV - VC
Bi-Level
Previous menu
Exit
URM Centiva/5, CM-920041, rev. 2.22
6-5
Page 78
6 Operation
t
Spontaneous breathing
6.2 Adjust ventilation parameter
6.2.1 VCV
Volume Controlled Ventilation.
This mandatory ventilation mode provides a constant minute volume to a patient, that has limited spontaneous breathing capabilities.
Paw
P max
P limit
PEEP
0
Insp. Pause
T insp T exp
60/rate
Flow
I-Flow
0
t
6-6
The desired minute volume is determined by the set tidal volume (Vt) and the set rate (rate).
The pressure and flow waveforms are influenced by the timing for inspiratory and expiratory times (I:E) and by the inspiratory flow (I-Flow).
The system automatically provides an inspiratory pause depending on timing and flow. The higher the I-Flow the longer the plateau time.
Although this is a mandatory mode, the system allows patient’s spontaneous breathing during the expiration time and therefore, allows the patient to increase the total minute volume beyond the set mandatory.
URM Centiva/5, CM-920041, rev. 2.22
Page 79
%
cmH2O
cmH2O
mL
1/Min
L/Min
Buzzer
PS endflow
Leakage compensation
CAUTION
6 Operation
Use the soft keys to adjust the ventilation settings, described above according to the patient’s needs
Fi O2
50
Rate
10
I : E
1:2.0
Vt
110
Plimit
30
PEEP
OFF
Additional adjustments for
Byflow
can be set in the parameter menu.
I-flow PS ramp Trigg.
ByFlow
Time window
Previous menu
10.0 L/m
Exit
A leak can occur when ventilating patients. To compensate for such leakage, the Centiva/5 is equipped with an automatic leakage compensation.
This leakage compensation is always active and compensates for temporary as well as permanent leaks.
The compensation covers a maximum leak of up to 50% of the desired tidal volume.
For more details refer to section “NIV” in this chapter.
Always set the leakage alarm at an appropriate level to ensure immediate notification in the event a leak were to occur.
I-Flow
6.6
0.20 s
3.0 L/m
3.0 L/m
60% 25%
100%
URM Centiva/5, CM-920041, rev. 2.22
6-7
Page 80
6 Operation
Paw
Volume not delivered
Pressure limited
ventilation
Changes in a patient’s lung compliance or resistance can alter the peak pressure at given settings.
When in the VCV mode, the P Pressure Limited Ventilation (PLV).
setting provides the ability to perform
limit
P max
P limit
PEEP
0
Pressure limited
When P set P
is reached the flow is decelerated to limit the pressure to the
limit
level. A low priority message “Plimit reached” will appear on the
limit
screen.
When pressure limitation causes a tidal volume of less than 80% of the set tidal volume to be delivered an alarm will indicate this with the “Vt not delivered” message.
Beyond the “P for additional safety. The P
(default). When the P
P
limit
” level there is a second pressure safety barrier “P
limit
is autom atically set to 10 cmH2O over the
max
level is reached all inspiratory flow is
max
stopped and the over pressure release valve is activated, reducing the pressure to the PEEP level.
The P
safety pressure level can be adjusted manually in the alarm
max
limits menu.
t
max
6-8
URM Centiva/5, CM-920041, rev. 2.22
Page 81
6 Operation
Flow
Short
Long
Ramp
Trigg. window
Spontaneous bre athing
with ASB pressure support
P ASB
without pressure support
6.2.2 SIMV-VC
Mandatory parameters
Synchronized Intermittent Mandatory Ventilation – Volume Constant
This synchronized mandatory ventilation mode provides a constant mandatory minute volume and supports spontaneous breathing with pressure support ventilation (PSV) for patients with limited spontaneous breathing capability or to intentionally reduce the mandatory support.
Paw
P max
P limit
Spontaneous breathing
Ramp
PEEP
0
t
T insp T exp
60/rate
I-Flow
0
t
The desired mandatory minute volume is determined by the set tidal volume (Vt) and the set rate (rate).
The pressure and flow waveforms are influenced by the timing for inspiratory and expiratory times (I:E) and by the inspiratory flow (I-Flow).
The system automatically provides an inspiratory pause depending on timing and flow. The higher the I-Flow the longer the pause (plateau) time.
URM Centiva/5, CM-920041, rev. 2.22
6-9
Page 82
6 Operation
%
cmH2O
cmH2O
mL
1/Min
cmH2O
Buzzer
PS endflow
Trigger parameters
PSV parameter
Settings
To synchronize the mandatory ventilation with the spontaneous breathing activity of the patient, the Centiva/5 uses a flow trigger that detects the patient’s inspiratory effort. The sensitivity of the flow trigger (Trigg.) can be adjusted to match the patient’s needs. The lower the flow value the more sensitive the system reacts. Self triggering can occur at low trigger flow settings.
The detected flow trigger needs to occur within the trigger time window (time window) to start a synchronized mandatory breath.
Pressure support ventilation is a pressure support (PS) function that will increase the tidal volume or decrease the work of breathing)of a spontaneous breathing patient. When a pressure supported breath is initiated the ventilator delivers the required flow to increase the pressure to the set PS pressure above the PEEP level. The pressure support breath ends when the flow to the patient decreases to the set PS end flow measured in % of the peak PSV flow. 25% is the default setting.
The rise time between the PEEP level and the PS pressure level can be modified by the PS- ramp time (ramp). The shorter the ramp time the higher the initial PS flow.
The Centiva/5 allows spontaneous breathing on the PEEP pressure level if the PSV function is turned off (PS set to 0 cmH
O).
2
Use the soft keys to adjust the ventilation settings according to the patient’s needs.
6-10
Fi O2
50
Rate
10
I : E
1:2.0
Additional adjustments for
Inspiratory flow (I-Flow)
PS ramp time (PS ramp)
Trigger flow sensitivity (Trigg)
Byflow (Byflow)
Trigger time window (time window)
PS end flow (PS endflow)
can be set in the parameter menu.
PEEP
OFF
I-flow PS ramp Trigg. ByFlow Time window
Previous menu
Vt
110
Plimit
30
URM Centiva/5, CM-920041, rev. 2.22
Exit
PS
10
10.0 L/m
0.20 s
3.0 L/m
3.0 L/m 60% 25%
100%
Page 83
6 Operation
Paw
Volume not delivered
Leakage compensation
ARC
Pressure limited
ventilation
A leak can occur when ventilating patients. To compensate for such leakage, the Centiva/5 is equipped with an automatic leakage compensation.
This leakage compensation is always active and compensates for temporary as well as permanent leaks.
The compensation covers a maximum leak of up to 50% of the desired tidal volume.
For more details refer to section “NIV” in this chapter.
Always set the leakage alarm at an appropriate level to ensure immediate notification in the event a leak were to occur.
The Airway Resistance Compensation (ARC) can be activated in the SIMV-VC mode .
ARC compensates for the inspiratory and expiratory resistance of the patient’s artificial airway.
For more details refer to section “ARC” in this chapter.
In the SIMV-VC mode the Plimit setting also allows for Pressure Limited Ventilation (PLV).
URM Centiva/5, CM-920041, rev. 2.22
P max
P limit
PEEP
0
Pressure limited
t
This mode is identical with the PLV mode in VCV mode.
For more details refer to the section ”PLV” in the “VCV” section in this chapter.
6-11
Page 84
6 Operation
Without spontaneous breathing
ramp time
P insp
6.2.3 Bi-Level
Without spontaneous breathing
Bi – Level
This synchronized, time controlled, pressure cycled ventilation mode provides a CPAP with two pressure levels. The change between these two pressure levels provides a mandatory minute volume to patients with limited spontaneous breathing capability.
P max
Paw
short ramp
long ramp
PEEP
0
T insp
T exp
60/rate
Flow
t
t
6-12
The desired minute volume is determined by the set pressure difference between the upper and lower CPAP level (Pinsp) and the set rate (rate) and is dependent on the patient’s lung compliance and resistance.
The pressure and flow waveform is influenced by the timing for the inspiratory and expiratory times (I:E) and the ramp time (ramp) between the lower und upper pressure level.
The base pressure level is adjusted with the positive end expiratory pressure (PEEP) setting.
Although this is a mandatory mode, the system allows spontaneous breathing during the inspiration and expiration period which allows the patient to increase their total minute volume beyond the set mandatory minute volume.
URM Centiva/5, CM-920041, rev. 2.22
Page 85
cmH2O
cmH2O
Buzzer
PS endflow
Leakage compensation
Use the soft keys to adjust the ventilation settings, described above according to the patient’s needs.
Fi O2
50
%
Additional adjustments for
Byflow (Byflow)
can be set in the parameter menu.
Rate
10
1/Min
I : E
1:2.0
Ramp
0.20
Sec.
P-Insp
10
PEEP
OFF
I-flow
PS ramp Trigg. ByFlow Time window
Previous menu
Exit
A leak can occur when ventilating patients. To compensate for such leakage, the Centiva/5 is equipped with an automatic leakage compensation.
This leakage compensation is always active and compensates for temporary as well as permanent leaks.
For more details refer to section “NIV” in this chapter.
Always set the leakage alarm at an appropriate level to ensure immediate notification in the event a leak were to occur.
6 Operation
PS
0
cmH2O
10.0 L/m
0.20 s
3.0 L/m
3.0 L/m 60% 25%
100%
URM Centiva/5, CM-920041, rev. 2.22
6-13
Page 86
6 Operation
t t
with spontaneous
0
6.2.4 Bi-Level
With spontaneous breathing
Mandatory parameters
Bi – Level
This synchronized, time controlled, pressure cycled ventilation mode provides a CPAP with two pressure levels. The change between the two pressure levels provides a mandatory minute volume and supports spontaneous breathing activities with PSV for patients with limited spontaneous breathing capability or to intentionally reduce the mandatory support.
Paw
P max
PEEP
P insp
P S
Spontaneous breathing without pressure support
60/rate
time window
Spontaneous breathing with pressure support
Flow
The mandatory parameters are identical to the previous section “Bi-Level without spontaneous breathing”.
The desired minute volume is determined by the set pressure difference between the upper and lower CPAP level (Pinsp) and the set rate (rate) and is dependent on the patient’s lung compliance and resistance.
The pressure and flow waveform is influenced by the timing for the inspiratory and expiratory times (I:E) and the ramp time (ramp) between the lower und upper pressure level.
The base pressure level is adjusted with the positive end expiratory pressure (PEEP) setting.
6-14
URM Centiva/5, CM-920041, rev. 2.22
Page 87
6 Operation
%
cmH2O
Sec.
1/Min
cmH2O
Buzzer
PS endflow
Trigger parameters
PSV parameter
Settings
To synchronize the mandatory ventilation with the spontaneous breathing activity of the patient, the Centiva/5 uses a flow trigger to detect the patient’s inspiratory effort. The sensitivity of the flow trigger (Trigg.) can be adjusted to match the patient’s needs. The lower the flow trigger value the more sensitive the system becomes. Auto-triggering can occur at low trigger flow settings.
The detected flow trigger needs to match with the trigger time expectation window (time window) to start a synchronized mandatory breath.
Pressure support ventilation is a pressure support (PS) function that will increase the tidal volume or decrease the work of breathing of a spontaneous breathing patient. When a pressure supported breath is initiated the ventilator delivers the required flow to increase the pressure to the set PS pressure above the PEEP level. The pressure support breath ends when the flow to the patient decreases to the set PS end flow measured in % of the peak PSV flow. 25% is the default setting.
The rise time between the PEEP level and the PS pressure level can be modified by the PS- ramp time (ramp). The shorter the ramp time the higher the initial PS flow.
The Centiva/5 allows spontaneous breathing on the PEEP pressure level if the PSV function is turned off (PS set to 0 cmH
O).
2
Use the soft keys to adjust the ventilation settings, described above, according to the patient’s needs.
Fi O2
50
Rate
10
I : E
1:2.0
0.20
Ramp
P-Insp
10
cmH2O
PEEP
OFF
PS
10
Additional adjustments for
PS ramp time (PS ramp)
Trigger flow sensitivity (Trigg)
Byflow (Byflow)
Trigger time window (time window)
PS end flow (PS endflow)
can be set in the parameter menu.
I-flow
PS ramp Trigg. ByFlow Time window
Previous menu
10.0 L/m
0.20 s
3.0 L/m
3.0 L/m 60% 25%
100%
Exit
URM Centiva/5, CM-920041, rev. 2.22
6-15
Page 88
6 Operation
Leakage compensation
ARC
A leak can occur when ventilating patients. To compensate for such leakage, the Centiva/5 is equipped with an automatic leakage compensation.
This leakage compensation is always active and compensates for temporary as well as permanent leaks.
For more details refer to section “NIV” in this chapter.
Always set the leakage alarm at an appropriate level to ensure immediate notification in the event a leak were to occur.
The Airway Resistance Compensation (ARC) can be activated in the Bi-Level mode .
ARC compensates for the inspiratory and expiratory resistance of the patients artificial airway.
For more details refer to section “ARC” in this chapter.
6-16
URM Centiva/5, CM-920041, rev. 2.22
Page 89
6 Operation
P ASB
ASB endflow
6.2.5 CPAP/PSV
Trigger parameters
Continuous Positive Airway Pressure / Pressure Support Ventilation
This ventilation mode provides a set continuous positive airway pressure for a spontaneous breathing patient, with or without a set pressure support level. This mode requires a patient’s stable spontaneous breathing activity.
Paw
short ramp
long ramp
P ASB
PEEP
0
ramp time
Flow
25 %
The base pressure level is adjusted with the positive end expiratory pressure (PEEP) setting.
To detect the inspiration effort of a patient’s spontaneous breathing activity the Centiva/5 uses a flow trigger. The sensitivity of the flow trigger (Trigg.) can be adjusted to match the patient’s needs. The lower the flow trigger value the more sensitive the system becomes . Auto­triggering can occur at low trigger flow settings.
t
t
URM Centiva/5, CM-920041, rev. 2.22
6-17
Page 90
6 Operation
%
cmH2O
Sec.
cmH2O
Buzzer
PS endflow
PSV parameter
Settings
ARC
Pressure support ventilation is a pressure support (PS) function that will increase the tidal volume or decrease the work of breathing of a spontaneous breathing patient. When a pressure supported breath is initiated the ventilator delivers the required flow to increase the pressure to the set PS pressure above the PEEP level. The pressure support breath ends when the flow to the patient decreases to the set PS end flow measured in % of the peak PSV flow. 25% is the default setting.
The rise time between the PEEP level and the PS pressure level can be modified by the PS- ramp time (ramp). The shorter the ramp time the higher the initial PS flow.
The Centiva/5 allows spontaneous breathing on the PEEP pressure level if the PSV function is turned off (PS set to 0 cmH
2
O).
Use the soft keys to adjust the ventilation settings, described above according to the patient’s needs.
Fi O2
50
Additional adjustments for
Trigger flow sensitivity (Trigg)
Byflow (Byflow)
PS end flow (PS endflow)
can be set in the parameter menu.
Ramp
0.20
PEEP
OFF
I-flow PS ramp
Trigg. ByFlow
Time window
Previous menu
10.0 L/m
Exit
The Airway Resistance Compensation (ARC) can be activated in the CPAP/PSV mode.
ARC compensates for the inspiratory and expiratory resistance of the patient’s artificial airway.
For more details refer to section “ARC” in this chapter
PS
5
0.20 s
3.0 L/m
3.0 L/m
60%
25%
100%
6-18
URM Centiva/5, CM-920041, rev. 2.22
Page 91
6 Operation
%
cmH2O
Sec.
cmH2O
1/Min
cmH2O
Buzzer
PS endflow
6.2.6 CPAP/Apnea
Setting CPAP/PSV
Continuous Positive Airway Pressure with APNEA mandatory back-up
ventilation.
This ventilation mode provides a set continuous positive airway pressure for a spontaneous breathing patient, with or without a set pressure support level. This mode requires a patient’s stable spontaneous breathing activity. For more details refer to the mode “CPAP/PSV” in this section.
In case the patient experiences an apneic episode while in this mode this ventilator will provide a back-up mode of ventilation using the
Bi – Level.
This synchronized, time controlled, pressure cycled ventilation mode provides a CPAP with two pressure levels. This change between the two pressure levels provides a mandatory minute volume and supports spontaneous breathing activities with PSV for patients with limited spontaneous breathing capability. For more details refer to the mode “Bi-Level” in this section.
When starting this mode “CPAP/PSV” is activated with the corresponding soft keys highlighted.
Use the soft keys to adjust the ventilation settings, described above according to the patient’s needs.
URM Centiva/5, CM-920041, rev. 2.22
Fi O2
50
Rate
10
I : E
1:2.0
0.20
Ramp
Additional adjustments for
Trigger flow sensitivity (Trigg)
Byflow (Byflow)
PS end flow (PS endflow)
can be set in the parameter menu.
P-Insp
10
PEEP
OFF
I-flow PS ramp
Trigg. ByFlow
Time window
Previous menu
Exit
PS
10
10.0 L/m
0.20 s
3.0 L/m
3.0 L/m
60%
25%
100%
6-19
Page 92
6 Operation
%
cmH2O
cmH2O
Sec.
1/Min
cmH2O
Buzzer
PS endflow
Setting the Back-up
ventilation mode
When Apnea is detected, the system switches automatically to the ventilation mode “Bi-Level”. The settings of this mode are displayed , but the soft keys are grayed to indicate that the settings can be set but are not active until the back up ventilation mode is activated.
Fi O2
50
Rate
10
I : E
1:2.0
0.20
Ramp
P-Insp
10
PEEP
OFF
Additional adjustments for
PS ramp time (PS ramp)
Trigger flow sensitivity (Trigg)
Byflow (Byflow)
Trigger time window (Time window)
I-flow
PS ramp Trigg. ByFlow Time window
10.0 L/m
PS end flow (PS endflow)
can be set in the parameter menu.
Previous menu
Exit
If the patient’s spontaneous breathing activity resumes the ventilator remains in the ventilation mode “Bi-Level”. The system does not automatically return to the ventilation mode “CPAP/PSV”.
To switch back to the initial mode requires the operator to re-select this ventilation mode from the ventilation menu.
PS
10
0.20 s
3.0 L/m
3.0 L/m 60% 25%
100%
6-20
URM Centiva/5, CM-920041, rev. 2.22
Page 93
6.3 Additional functions
A R C
Exit
The Centiva/5 ventilator offers additional functions to support the operation of the ventilator and to allow fine-tuning for particular needs.
The following functions are available in the main menu:
These functions will be described in detail in the order they appear in the main menu.
The Centiva/5 provides other additional functions that are permanently activated.
Stand-by
ASR (Automatic Suction Routine)
Parameter menu
Alarm limits
Alarm AUTOSET
ARC (Airway Resistance Compensation)
Nebulizer
System test
NIV (Non-Invasive Ventilation)
6 Operation
Stand-by A S R Parameter menu Alarm limits Alarm AUTOSET Vent. modes
Nebulizer
System test
ON ON
ON
URM Centiva/5, CM-920041, rev. 2.22
APD (Automatic Patient Detection)
These functions are described in this chapter.
6-21
Page 94
6 Operation
A R C
Exit
A R C
Exit
6.3.1 Stand-by
Special notice
The standby function allows for the temporary interruption of the Centiva/5 ventilator without alarming. All patient related settings are saved and available upon reconnecting the patient to the ventilator.
To activate the standby function the user must disconnect the patient from the system and wait for the “System open?” message on the screen.
Open the main menu by pushing the menu button.
Select the function “Stand-by ON” by rotating the ComWheel.
This function is only accessible when a patient is disconnected from the ventilator.
Stand-by A S R Parameter menu Alarm limits Alarm AUTOSET Vent. modes
Nebulizer
System test
Push the ComWheel to confirm the selection.
Notice the change in the menu from “Standby ON” to “Stand-by OFF”.
The status line in the screen indicates the “Stand­by” mode is active.
Ventilation is stopped and all alarms are silenced.
All ventilation settings are stored.
Stand-by
A S R Paramet er menu Alarm limits
Alarm AUTOSET
Vent. m odes
Close the main menu either with “Exit” or the menu
Nebulizer
System test
will close automatically after 30 secs.
ON ON
ON
OFF
ON
ON
6-22
URM Centiva/5, CM-920041, rev. 2.22
Page 95
6 Operation
To turn OFF the stand-by mode and to return to the former ventilation mode reconnect the patient to the Centiva/5.
When the reconnection is detected by the APD function, the stand-by function automatically is turned off and the Centiva/5 proceeds with ventilating the patient with the stored parameters .
The Automatic Patient Detection (APD) has a delay of 15 secs after turning ON the stand-by mode before becoming active to reduce false triggering of the APD function.
To manually turn OFF the stand-by mode and ventilate using current settings :
open the main menu with the menu button
select the “Stand-by OFF” with the ComWheel
confirm the selection
close the menu with “Exit”.
URM Centiva/5, CM-920041, rev. 2.22
6-23
Page 96
6 Operation
A R C
Exit
6.3.2 ASR (Automatic Suction Routine)
To perform bronchial suction the user needs access to the patient’s tube.
Using a breathing circuit with a bronchial suction access port the user can perform this therapy while in the ventilation mode .
To gain access to the tube on most breathing circuits the patient needs to be disconnected from the ventilator.
The Centiva/5 offers an Automatic Suction Routine (ASR) that provides the following sequence:
Pre-suction oxygenation phase for up to 2 minutes
Stand-by suction phase for up to 2 minutes with silenced alarms
Post-suction oxygenation phase for up to 2 minutes.
ASR can be activated in all ventilation modes via the main menu.
To activate the ASR functions open the main menu by pushing the menu button.
Select the function “ASR ON” by rotating the ComWheel.
Stand-by
A S R
Parameter menu Alarm limits Alarm AUTOSET Vent. modes
Nebulizer
System test
Confirm the selection by pushing the ComWheel.
ON
ON
ON
6-24
URM Centiva/5, CM-920041, rev. 2.22
Page 97
6 Operation
A R C
OFF
Exit
Pre-Oxygenation phase
Suction phase
Notice the change in the menu from “ASR ON” to “ASR OFF”.
The first phase of ASR starts immediately.
The status line in the screen shows “Pre-Ox 1:59” to indicate the start of the pre-oxygenation phase and to show the remaining time in this phase.
The soft key for the FiO2 setting changes from the previous setting to 100 Vol %. The soft key color changes from yellow to red to indicate that this is a temporary change during the ASR procedure.
This FiO2 setting for the pre-oxygenation of 100 Vol % can be adjusted to any desired FiO2 by the user.
Stand-by A S R Parame t er menu Alarm limits Alarm AUTOSET Vent. m odes
Nebulizer
System test
After 2 minutes the pre-oxygenation phase ends and the suction phase starts. The status line in the screen shows “Suct 1:59” to indicate the start of the suction phase and to show the remaining time in this phase.
The ventilator is stopped and the pressure is relieved to zero to allow hygienic opening of the patient’s tube connection.
The screen shows the message “System open?”.
All alarms are silenced, but the Centiva/5 gives a short audible signal to indicate the start of the suction phase. This audible signal is repeated every 5 secs to remind the operator ventilation is not active.
Within 15 secs the patient needs to be disconnected from the ventilator. If the APD does not recognize a patient disconnect within this 15 second window the suction phase will be terminated and the ventilator will go to the post-oxygenation phase.
If after 2 minutes the APD does not recognizes a reconnection of the patient the Centiva/5 starts the Apnea Alarm.
ON
ON
URM Centiva/5, CM-920041, rev. 2.22
6-25
Page 98
6 Operation
Post-Oxygenation phase
WARNING
The status line in the screen shows “Post-Ox 1:59” to indicate the start of the post-oxygenation phase and to show the remaining time in this phase.
The ventilator starts back in the set mode with the previous ventilation pattern and settings. PEEP pressure will build up again. All alarms are activated .
The soft key for FiO2 setting is still set to the value of the pre-oxygenation phase. The color of the soft key is still red to indicate this temporary change by the ASR procedure.
This FiO2 default setting for the post-oxygenation of 100 Vol % can be adjusted to any desired FiO2 by the user.
At the end of the 2 minute post-oxygenation phase the FiO2 setting returns automatically to the setting prior to the start of the ASR procedure. The ASR procedure turns off automatically.
Ensure sufficient patient oxygenation if the FiO2 pre and post suction concentration is changed from 100 Vol % to any other value.
ASR can be activated by a short cut. For more details refer to section “short cuts” in this chapter.
6-26
URM Centiva/5, CM-920041, rev. 2.22
Page 99
6 Operation
A R C
Exit
Buzzer
PS endflow
6.3.3 Parameter menu
Frequently used settings for the different ventilation modes are accessible via the soft keys on the screen. Centiva/5 offers additional setting options. To keep the interface easy to use these other settings are accessible in the parameter menu.
The parameter menu provides access to certain settings depending on the current ventilation mode. For details refer to the “Adjust ventilation parameters” in this section.
To access the parameter menu open the main menu by pushing the menu button.
Select the function “Parameter menu” by rotating the ComWheel.
Stand-by A S R
Parameter menu
Alarm limits Alarm AUTOSET Vent. modes
Nebulizer
System test
Confirm the selection by pushing the ComWheel.
The parameter menu is accessible.
ON ON
ON
URM Centiva/5, CM-920041, rev. 2.22
Close the parameter menu and main menu either with “Exit” or let the menu close automatically after 30 secs.
I-flow PS ramp Trigg. ByFlow Time window
10.0 L/m
0.20 s
3.0 L/m
3.0 L/m
Previous menu
Exit
60% 25%
100%
6-27
Page 100
6 Operation
Buzzer
PS endflow
Buzzer
PS endflow
Buzzer
PS endflow
Buzzer
PS endflow
I-flow
PS ramp
Trigg.
Byflow
Available in the ventilation mode SIMV-VC.
This allows the setting of the inspiratory flow in the range of 2 to 99.9 L/min, depending on the timing settings of the mode.
This setting is identical to the I-flow soft key setting in the ventilation mode “VCV”.
Available in the ventilation modes: SIMV-VC, Bi-Level, CPAP/PSV and CPAP/Apnea.
This setting adjusts the rate the pressure increases from the PEEP pressure to the PS pressure in the range of 0.1 to 0.5 secs. This setting is similar to the Bi-Level ramp soft key setting.
The default setting is 0.2 secs.
Available in the following ventilation modes: SIMV­VC, Bi-Level, CPAP/PSV and CPAP/Apnea.
This allows the setting of the flow trigger in the range of 0.5 to 20 L/min. The flow trigger value is the amount of flow the patient needs to remove from the breathing circuit by their inspiratory effort to trigger the ventilator.
The default setting is 3.0 L/min.
Available in the ventilation modes: VCV, SIMV-VC, Bi-Level, CPAP/PSV and CPAP/Apnea.
The setting of the Byflow is from 3.0 to 30 L/min. The Byflow permanently flows through the breathing circuit providing a “flow reservoir” for a spontaneous breathing patient. Increasing the Byflow increases the flow reservoir, but also increases the pressure drop within the breathing circuit. The default setting is 3.0 L/min.
I-flow
PS ramp Trigg. ByFlow Time window
I-flow
PS ramp
Trigg. ByFlow Time window
I-flow PS ramp
Trigg.
ByFlow Time window
I-flow PS ramp Trigg.
ByFlow
Time window
10.0 L/m
0.20 s
3.0 L/m
3.0 L/m
Previous menu
Exit
10.0 L/m
0.20 s
3.0 L/m
3.0 L/m
Previous menu
Exit
10.0 L/m
0.20 s
3.0 L/m
3.0 L/m
Previous menu
Exit
10.0 L/m
0.20 s
3.0 L/m
3.0 L/m
Previous menu
Exit
60% 25%
100%
60% 25%
100%
60% 25%
100%
60% 25%
100%
6-28
URM Centiva/5, CM-920041, rev. 2.22
Loading...