Burkert 8223 Instruction Manual

Instruction Manual
8223
INDUCTIVE CONDUCTIVITY SENSOR
© Bürkert 2001 Subject to technical change without notice
2
8223
INTRODUCTION
Table des matières
INTRODUCTION ............................................................................................................... 2
1.1 Symbol used ..............................................................................................................3
1.2 General safety instructions .................................................................................... 3
QUICK START ................................................................................................................... 4
2.1 Quick installation guide ........................................................................................... 4
CONFIGURATION ............................................................................................................ 6
3.1 General comments ................................................................................................... 6
3.2 Measuring range ....................................................................................................... 6
3.2.1 Measuring range of the conductivity ............................................................................6
3.2.2 Measuring range of the temperature ...........................................................................8
3.3 Filtering level .............................................................................................................. 9
3.4 Temperature compensation .................................................................................10
3.4.1 Specific compensation ......................................................................................................... 10
3.4.2 Linear compensation ............................................................................................................. 11
3.5 Transmission of the temperature .........................................................................12
3.6 Calibrating the “zero conductivity” point ...........................................................12
INSTALLATION ............................................................................................................... 13
4.1 Mounting instructions ............................................................................................13
4.1.1 Temperature-pressure diagram..................................................................................... 13
4.1.2 Installation recommendations .........................................................................................13
4.2 Mounting...................................................................................................................14
4.3 Electrical connection instructions .......................................................................15
4.3.1 Mounting and connection of the EN 175301-803 connector ...........15
4.3.2 Connecting the 8223 to an external device (PLC,...) ..................................16
4.3.3 Precautions during installation and commissioning .....................................16
4.4 Examples of connections with a 8223 sensor .........................................................17
MAINTENANCE .............................................................................................................. 18
5.1 Maintenance ............................................................................................................18
5.2 If a problem occurs... .............................................................................................18
TECHNICAL DATA .........................................................................................................20
6.1 Process characteristics.........................................................................................20
6.2 Electrical characteristics .......................................................................................20
6.3 User characteristics ...............................................................................................21
6.4 Safety ........................................................................................................................21
6.5 Environment .............................................................................................................21
6.6 Conformity with standards ...................................................................................21
6.7 Dimensions (mm) ....................................................................................................22
INFORMATION ...............................................................................................................23
7.1 Standard delivery ....................................................................................................23
7.2 Ordering table for sensors 8223 ........................................................................23
7.3 Ordering table for spare parts .............................................................................23
7.4 Measurement principle ..........................................................................................24
7.5 Label description ....................................................................................................24
7.6 Declaration of conformity ......................................................................................26
3
8223
INTRODUCTION
1.1 SYMBOL USED
Indicates information which must be followed. Failure to follow the in for ­ma tion could endanger the user and affect the function of the device.
1.2 GENERAL SAFETY INSTRUCTIONS
Before installing or using this product, please read this manual and any other rel e vant documentation to ensure you fully benefit from all the ad van tag es the prod uct can offer.
- Please verify that the product is complete and free from any damage.
- It is the customer's responsibility to select an appropriate sensor for the application, ensure the unit is installed correctly, and main tain all com po nents.
- This product should only be installed or repaired by specialist staff using the correct tools.
- Please observe the relevant safety regulations throughout the operation, maintenance and repair of the product.
- Always ensure that the power supply is switched off and the pipes / tank do not contain any pressure before working on the device / system.
- This electronic device is sensitive to electrostatic discharge. To avoid any damage by immediate electrostatic discharge, pay attention to the requirements of EN 100 015-1.
- Always protect the device from electromagnetic perturbations, ultraviolet radiations and, when installed outside, from the effects of climatic conditions.
- If these instructions are ignored or the device is not used according to the specifications, no liability will be accepted and the guarantee on the device and accessories will become invalid.
4
8223
QUICK START
2.1 QUICK INSTALLATION GUIDE
Order code
OK ?
Verify that the delivered product is complete. See § 7.1
Yes
No
Unscrew the 2 plugs
Select the measuring range
using SW1
Use a coin for instance
Select the filtering level using SW1
Select the temperature com pen sa tion
using SW2 *
Select the type of current output
using SW3
Open the EN 175301-803 connector
Connect the power supply cable to
the EN 175301-803 connector
Re-mount the EN 175301-803 con-
nector and connect it to the sensor
Green
LED flashes once briefly
every 2 s and current output
indicates 4 mA?
Yes
No
Press the push-
button for 2 s
Contact your Bürkert agent
Install the appliance on the pipe
See the con fi gu ra tion table opposite and para graph 3.2.
See the con fi gu ra tion table opposite and paragraph 3.3
See the con fi gu ra tion table opposite and paragraph 3.4
Ensure that the cable is not pow ered.
See § 4.1
Power the sensor
* or the trans mis sion of the temperature on the 4-20 mA output.
See § 3.4
See the con fi gu ra tion table opposite and paragraph 4.3.2
Drive home the connector screw.
See § 4.3.1
The sensor must be out of the fluid.
See § 3.6
Unpack the
ap pli ance
5
8223
QUICK START
1
2
3
4
4
3
2
1
Push-
button
SW3
sink ing
sourc ing
SW1
OFF<->ON
Green LED Red LED
SW2
ON<->OFF
<->
See also § 5.2 for any other LED flashing.
Measuring range
Position of the switches of SW1
12
0 to 1 mS/cm OFF OFF
0 to 10 mS/cm (default value)
OFF ON
0 to 100 mS/cm ON OFF 0 to 1 S/cm ON ON
Filtering level
Position of the switches of SW1
34
0 (no filtering) (default value)
OFF OFF
1 (min. filtering) OFF ON
2 (medium filtering) ON OFF
3 (max. filtering) ON ON
Temperature compensation
Position of the switches of SW2
1234
No compensation (default value)
OFF OFF OFF OFF
Linear compensation
0.1% ON OFF OFF OFF
0.25% OFF ON OFF OFF
0.5% ON ON OFF OFF
0.7% OFF OFF ON OFF 1% ON OFF ON OFF
1.5% OFF ON ON OFF 2% ON ON ON OFF 3% OFF OFF OFF ON 5% ON OFF OFF ON
Specific
compensation
NaOH OFF ON OFF ON
HNO
3
ON ON OFF ON
H
2SO4
OFF OFF ON ON
NaCl ON OFF ON ON
Not used OFF ON ON ON
Transmission of the
temperature to the
4-20 mA output
Position of the switches of SW2
1234
ON ON ON ON
Type of current output Position of switch SW3
Sinking (default value)
Up
Sourcing Down
Green LED Red LED Status of the sensor when operating faultless
ON ON Calibration of the "zero conductivity" in progress
Flashes once briefly
OFF
Measurement of a nil conductivity (< 2% of the full scale)
Flashes at a fequency
between 0,5 and 16 Hz
Measurement of a conductivity which is proportional to this frequency
6
8223
CONFIGURATION
3.1 GENERAL COMMENTS
Configuration is done by means of accessible switches after having unscrewed the plugs counterclockwise and removed them.
- SW1 allows the selection of:
- the measuring range (switches 1 and 2)
- the filtering level of the conductivity (switches 3 and 4)
- SW2 allows the selection of the temperature compensation or the transmission of the temperature on the 4-20 mA output.
- SW3 allows the selection of the current output mode, sinking or sourcing.
- The push-button allows the calibration of the sensor «zero conductivity» point.
3.2 MEASURING RANGE
The 4-20 mA output delivers a current which is proportional to the compensated or not compensated conductivity or to the measured temperature. For the con duc tiv i ty, the user may choose the mea sur ing range.
When the conductivity exceeds the full scale by 10%, the current output delivers a 22 mA current. As soon as the conductivity is again equal or inferior to the full scale value, the current output delivers a current between 4 and 20 mA.
3.2.1 Measuring range of the conductivity
SW1 makes it possible to select the conductivity mea sur ing range:
- a 4 mA current indicates a conductivity which is equal to 0 mSiemens/cm (less than 2% of the selected full scale)
Push-button
SW1 SW2
Green LED Red LED
SW3
Measuring range
Position of the switches of SW1
12
0 to 1 mS/cm OFF OFF
0 to 10 mS/cm (default value)
OFF ON
0 to 100 mS/cm ON OFF 0 to 1 S/cm ON ON
7
8223
CONFIGURATION
- a 20 mA current indicates a conductivity which is equal to either:
- 1 mSiemens/cm (mS/cm)
- 10 mS/cm
- 100 mS/cm
- 1 S/cm
The conductivity depending on the material the used fitting is made of and its dia­meter, the full scale value must be re-calculated using the following for mu la:
FS = FS
s
x C
F
where FS = Full Scale value to be programmed on the PLC FSs = selected Full Scale CF = Correction factor of the fitting used: see the following tables
- If a fitting with DN15, 20 or 25 is used, use the correction factor of the DN32
- If a fitting with a DN > 100 or a vessel is used, apply a correction factor = 1.
Example:
The 8223 sensor is installed in a PVDF S020 fitting with DN32.
The selected full scale is FS
s
= 10 mS/cm.
The correction factor for this fitting is C
F
= 1,113.
Thus: FS = FS
s
x CF = 10 x 1,113 = 11,13 mS/cm
DN of the
fitting
Correction factor
Weld-ends / Internal and external threads True-union / Solvent and fusion spigots
Brass Stainless steel PVDF PP PVC
DN32 0.991 0.989 1.113 1.098 1.093
DN40 0.989 0.989 1.049 1.045 1.045
DN50 0.985 0.983 1.022 1.021 1.022
DN of the
fitting
Correction factor
Weld-ends Saddles
Brass Stainless steel PVDF PP PVC
DN65 - 0.993 1.020 1.019 1.025
DN80 - 0.995 1.020 1.019 1.022
DN100 - 0.998 1.019 1.017 1.010
8
8223
CONFIGURATION
Example:
If the selected measuring range is: „0 to 10 mS/cm”. When the fluid conductivity measured by the sensor is < 2% of the full scale, the 4-20 mA output delivers a 4 mA current; When the conductivity measured by the sensor is equal to 10 mS/cm, the 4-20 mA output delivers a 20 mA current.
The curve hereunder shows the ratio between the measured conductivity and the current value delivered by the 4-20 mA output.
When the conductivity exceeds the full scale by 10% (11 mS/cm in the example), the cur­rent output delivers a 22 mA current. As soon as the conductivity is again equal or inferior to the full scale value, the current output delivers a current between 4 and 20 mA.
3.2.2 Measuring range of the temperature
- The value of the temperature which corresponds to a 4-mA current is always -10 °C
- The value of the temperature which corresponds to a 20-mA current is always 80 °C
If the 8223 has been programmed to transmit the temperature (and not the con duc ­tiv i ty) on the 4-20 mA output (all SW2 switches are set to ON), the cor re spond ing temperature/4-20 mA-output curve is the following:
The current output delivers a 22 mA current if the temperature is < -10 °C or > 80 °C.
4
20
mA
0
10
mS/cm
5
12
0,2
22
11
Green
LED Red
LED
0,5 - 16 Hz
every 2 s
every
2 s
4
20
mA
080
°C
35
12
22
-10
9
8223
CONFIGURATION
mS/cm
mS/cm
mS/cm
mS/cm
t (s)
t (s)
t (s)
t (s
)
3.3 FILTERING LEVEL
Filtering allows the attenuation of fluctuations in the conductivity. The 8223 sensor has four filtering levels, from 0 to 3:
- Level 0 corresponds to a nil filtering: the sensor indicates any conductivity variation
- Level 3 corresponds to the maximum filtering: the sensor smoothes the conductivity variations at the most
- Levels 1 and 2 correspond to intermediate filterings.
SW1 allows the selection of the filtering level:
Level 0 fil ter ing
Level 1 fil ter ing
Level 2 fil ter ing
Level 3 fil ter ing
Filtering level
Position of the switches of SW1
34
0 (no filtering) (default value)
OFF OFF
1 (min. filtering) OFF ON
2 (medium filtering) ON OFF
3 (max. filtering) ON ON
10
8223
CONFIGURATION
3.4 TEMPERATURE COMPENSATION
The conductivity varies with the temperature; To compensate the variations, the 8223 sensor measures the real conductivity and temperature of the fluid then re-calculates the conductivity corresponding to a temperature of 25 °C.
SW2 makes it possible to choose the calculation mode of the temperature com pen sa tion.
Three calculation modes are possible, either:
- a specific compensation: 4 compensation curves are memorized by the sensor for the following solutions: NaOH (sodium hydroxyde), HNO
3
(nitric
acid), H2SO4 (sulphuric acid), and NaCl (sodium chloride).
- a linear compensation: 9 compensation factors, from 0.1 to 5% can be chosen.
- no compensation.
3.4.1 Specific compensation
The specific compensation curves for NaOH (sodium hydroxyde), HNO
3
(nitric acid) and
NaCl (sodium chloride) are valid for temperatures ranging from 10 to 70 °C.
The specific compensation curve for H
2SO4
(sulphuric acid) is valid for tem per a tures
ranging from 5 to 55 °C.
The compensation has been determined at the following concentrations:
NaOH : 1% HNO
3
: 1%
NaCl : 0,2%
H
2SO4
: 20%
- The NaCl specific curve is valid for concentrations ranging from 60 mg/l (conductivity ≅ 100 µS/cm) to 270 g/l (conductivity ≅ 220 mS/cm).
- The NaCl compensation curve can be used for some dilute solutions.
- If the fluid of your process does not correspond to any specific solution, use one of the linear compensations.
11
8223
CONFIGURATION
3.4.2 Linear compensation
If no specific compensation curve corresponds to your process, use one of the 9 linear compensation factors by using SW2.
If you do not know the compensation factor (average
α) of your process, you can deter-
mine it by doing the following:
1) Measure the conductivity of the fluid at 25 °C ( )
2) Measure the conductivity of the fluid at a temperature T ( )
3) Apply the following formula to determine the
α factor:
4) Use the compensation factor nearest to the calculated one:
25
°
T
α =
χ
χ
T
x
1
25°
χ
25°
25° T
χ
χ
25
25 C
T C
Τ
χ
T
Temperature compensation
Position of the switches of SW2
1234
No compensation (default value)
OFF OFF OFF OFF
Linear compensation
0.1% ON OFF OFF OFF
0.25% OFF ON OFF OFF
0.5% ON ON OFF OFF
0.7% OFF OFF ON OFF 1% ON OFF ON OFF
1.5% OFF ON ON OFF 2% ON ON ON OFF 3% OFF OFF OFF ON 5% ON OFF OFF ON
Specific
compensation
NaOH OFF ON OFF ON
HNO
3
ON ON OFF ON
H
2SO4
OFF OFF ON ON
NaCl ON OFF ON ON
Not used OFF ON ON ON
12
8223
CONFIGURATION
3.5 TRANSMISSION OF THE TEMPERATURE
If, instead of the conductivity, the 8223 sensor is meant to transmit the tem per a ture (be tween -10 °C and 80 °C) to the 4-20 mA output, configure SW2 as follows:
Thus the conductivity measuring range chosen with SW1 is not taken into account.
3.6 CALIBRATING THE “ZERO CONDUCTIVITY” POINT
The “zero conductivity” point of the sensor can drift in time. To check whether the sensor is correctly calibrated or not, measure the conductivity of the air (conductivity = 0).
- Dismantle the sensor from the pipe.
Ensure the finger orifice the fluid passes through is clean and dry.
- If the conductivity of the air > 0 (the current output indicates a value > 4 mA and the green LED flashes at a frequency between 0.5 and 16 Hz), press the push button for at least 2 s: both LEDs light up and the sensor starts calibrating its zero conductivity” point.
The operation can last a few minutes.
The calibration is finished when the two LEDs are not steadily lit any more. Then:
- If the red LED switches off and the green LED flashes rapidly every 2 s, the sensor is correctly calibrated.
- If the green LED lights up and the red LED flashes 3 or 4 times every 2 s, the calibration has failed: press the push-button briefly to go back to normal operating using the parameters of the previous calibration.
- If the calibration fails several times in a row, contact your Bürkert agent.
Push-button
Green LED Red LED
Transmission of the
temperature to the
4-20 mA output
Position of the switches of SW2
1234
ON ON ON ON
13
8223
INSTALLATION
4.1 MOUNTING INSTRUCTIONS
4.1.1 Temperature-pressure diagram
The sensor and the fitting in which it is installed have limited operating temperatures and pressures. The diagram hereunder shows the operating range of the 8223-fitting assem­bly, for each fitting material:
A: Operating range of the assembly 8223+fitting
4.1.2 Installation recommendations
Choose a mounting position which avoids the building of air bubbles or cavities within the finger orifice.
Position 1: horizontal or vertical mounting onto a pipe. Position 2: mounting on a vessel without stirring Position 3: mounting on a vessel with stirring
- When dismounting the sensor from the pipe, take all the precautions related to this process.
- Ensure the fluid passage orifice is in the direction of the flow.
1
2
3
-10
+20 +40 +60
+80
+100
0
2 1
3
4
5
6
7
8
P [bar]
0
T [°C]
A
PVC
PP
PVDF
Metal
14
8223
INSTALLATION
4.2 MOUNTING
The 8223 conductivity sensor is installed on a pipe as follows:
- Insert the nut [4] on the fitting [6] and clip the ring [3] in the groove [5].
- Insert the sensor [1] into the fitting [6] by ensuring the gasket [2] is in its right place.
- Screw on the nut and tighten by hand only.
- Ensure the gasket material (FKM in standard) is chemically compatible
with your process.
- One EPDM gasket and one FKM gasket are additionally supplied with
the sensor.
- Make sure the gasket is not damaged when replacing it.
1
3
4
5
6
2
15
8223
INSTALLATION
1
3
2
4.3 ELECTRICAL CONNECTION INSTRUCTIONS
Always ensure that the appliance is not powered before working on it. All the connections must be disconnected.
- Use a shielded cable with a limit operating temperature > +80°C.
- Under normal con di tions of use, a shielded cable of 0.75 mm
2
section is sufficient
for signal transmission.
- Do not install the line close to high voltage or high frequency cables.
- If adjacent laying is unavoidable, keep a minimum distance of 30 cm.
- Connect the shielding to earth.
- Use a quality power supply (filtered and regulated).
- Do not open or cable the sensor when powered.
- It is essential that a 100 mA fuse be used for the power supply.
4.3.1 Mounting and connection of the EN 175301-803 connector
- Remove part [3] from part [2].
- Unscrew the cable gland [5].
- Insert cable into part [2] via cable gland.
- Wire part [3] (see below)
- Replace part [3].
- Tighten the cable gland [5].
- Place gasket [4] between the connector and the fixed connector of the 8223.
- Connect the connector to the 8223.
- Tighten screw [1] to ensure tightness and correct electrical contact
Always check the connection of the connectors to ensure correct opera­tion of the appliance.
L+
(12-30 VDC)
4-20 mA output
L-
Not connected
1
3
4
5
6
2
16
8223
INSTALLATION
4-20 mA
+
-
12-30 VDC
+
-
100 mA
I
1
3
2
SW3
4-20 mA
+
-
12-30 VDC
+
-
100 mA
I
1
3
2
SW3
4.3.2 Connecting the 8223 to an external device (PLC,...)
The 8223 sensor can be connected to a PLC or any other device which is able to use the 4-20 mA signal transmitted by the sensor.
The connection can be carried out either in sourcing or in sinking mode, as shown by the figures below:
SW3 must be configured correctly when the sensor is not powered, de ­pend ing on the connection mode chosen.
4.3.3 Precautions during in stal la tion and com mis sion ing
- When the appliance is powered and the cover is open, the protection against electrical shocks is no longer effective.
- Always verify the chemical compatibility of the materials which are in contact with the fluid to be measured.
- When cleaning the appliance use products which are chemically compatible with the materials in the appliance.
- Do not insert any object (screwdriver for instance) within the finger orifice. When the orifice is dirty, use compressed air to clean it.
PLC
PLC
Connection in sourcing mode Connection in sinking mode
Type of current output Position of switch SW3
Sinking (default value)
Up
Sourcing Down
not connected
not connected
17
8223
INSTALLATION
4.4 EXAMPLES OF CONNECTIONS WITH A 8223 SENSOR
Between the 8223 sensor and the 1067 positioner.
Between the 8223 sensor and a sole­noid valve with 8624-2 PI controller.
Between the 8223 sensor and the 8630 Top Control mounted on a 2031 diaphragm valve
8223
+
-
24 V=
1
3
2
L+
L-
I
SW3
1067
4-20 mA
8223
+
-
24 V=
1
3
2
L+
L-
SW3
4
3
1
2
4
3
1
Setpoint +
Setpoint -
24 V=
+
-
9
10
7
8
65
3
4
1
2
+-+
-
+
-
I
+24 V
­SW3
1
3
2
2031+ 8630
8223
4-20 mA
L-
L+
not connected
not connected
not connected
18
8223
MAINTENANCE
5.1 MAINTENANCE
The 8223 sensor may be cleaned with water or a product which is compatible with the materials therein.
Your Bürkert supplier is available to provide you with any additional information you require.
5.2 IF A PROBLEM OCCURS...
An error is indicated by lighting up of the green LED, special flashing of the red LED and the transmission of a 22-mA current to the current output.
The following table lists possible faults that can occur and their solutions:
Problem
Green
LED
status
Red LED
status
Current
output status
Possible cause Do the following
See also
The sensor does not work
OFF OFF 0 mA
The sensor is not con­nected
Connect the sensor 4.3
The fuse of the installation is in a bad condition
Change the fuse -
The switch of the installa­tion is set to OFF
Set the switch to ON
-
The power supply has been wrong connected to the + and - terminals
Check the wiring 4.3
The power supply is not stable or < 12 VDC
Change the power supply
-
The sensor measures a wrong con­ductivity
flashes
briefly
OFF 4 mA
The "zero conductivity" point has been calibrated within the fluid or the "zero conductivity" point has drifted
carry out a calibra­tion in air
3.5
flashes OFF -
The finger is dirty
Clean the sensor finger
5.1
Air bubbles appear within the finger orifice
Follow the mountign instructions
4.1.2
The temperature compen­sation is not correct
Select an appro­priate compensation
3.4
The conductivity fluctua­tions are very important
Select a higher filte­ring level (SW1)
3.3
19
8223
MAINTENANCE
Problem
Green
LED
status
Red LED
status
Current
output
status
Possible cause Do the following
See also
The sensor transmits a nil conduc­tivity
flashes
briefly
OFF 4 mA
The chosen measuring range is not appropriate
Select the lower measuring range (SW1)
3.2
The sensor transmits no current at all
flashes flashes 0 mA
SW3 is not set correctly (sinking or sourcing)
Modify SW3 4.3.2
The current output is not connected properly
Re-connect the current output
4.3
The sensor is stopped
- an error is indicated
ON
flashes
once
briefly
every 2 s
22 mA
The conductivity > full scale + 10%
Select the next measuring range (SW1)
3.2.1
ON
flashes
twice
briefly
every 2 s
22 mA
The fluid temperature < -10 °C or > +80 °C
Bring the fluid temperature back to a value within the sensor measuring range
3.2.2
ON
flashes
3 or 4
times
briefly
every 2 s
22 mA
The "zero conductivity" point calibration failed
Press the push­button briefly. If the error persists, send the device back to Burkert.
3.6
flash simultaneously 22 mA The sensor is out of service
Send the device back to Burkert
-
20
8223
TECHNICAL DATA
6.1 PROCESS CHARACTERISTICS
Conductivity measurement
- Type of measurement Inductive conductivity measurement
- Measuring range 0 µS/cm to 1 S/cm
- Precision +/-2% of the selected measuring range (within the range 0-70 °C)
- Temperature deviation 0.2% / °C max.
- Answering time at a strong conductivity change < 1 s
- Sampling frequency measurement every 250 ms
Temperature measurement
- Type of measurement digital measurement
- Measuring range -10 °C to 80 °C
- Precision +/- 0.5 °C
- Answering time at a strong conductivity change 100 s
- Sampling frequency 250 ms
Process data
- Pipe interface standard Bürkert fitting
- Pressure class PN6
- Fluid temperature range -10 °C to 80 °C
- Materials in contact with the fluid Finger: PVDF, PEEK or PP Seals: FKM or EPDM
6.2 ELECTRICAL CHARACTERISTICS
Pulse rate output
- Output type current output from 4 to 20 mA (error signal at 22 mA)
- Electrical wiring Wiring in sourcing or sinking mode, can be selected by bipolar reverser
- Max. load resistance 1000 Ohms with a power supply of 30 V 690 Ohms with a power supply of 24 V 300 Ohms with a power supply of 15 V 150 Ohms with a power supply of 12 V
- Adjustment 4 conductivity ranges and 1 temperature range, can be selected by switch
Electrical connection
- Power supply voltage 12 to 30 VDC, filtered and regulated
- Maximum current consumption 50 mA max. + 22 mA for the current output
- Connector type EN 175301-803
21
8223
TECHNICAL DATA
6.3 USER CHARACTERISTICS
- Measurement indication By a green-coloured led flashing from 0.5 to 16 Hz depending on the conductivity; It flashes briefly every 2 seconds when the fluid conductivity is lower than 2% of the selected full scale
- Error message By two green and red-coloured leds (see 5.2)
- Programming the scale of measurement 4 measurement ranges selectable by means of 2 switches
- Filtering the conductivity 4 filtering levels selectable by means of 2 switches
- Temperature compensation 14 compensation types (9 pro por tion al com pen sa tion levels and 4 solution specific levels), selectable by means of 4 switches (reference tem per a ture = 25 °C)
- Calibration of «zero conductivity» point By means of a push-button
6.4 SAFETY
Electrical input and output protected against polarity reversal.
6.5 ENVIRONMENT
- Operating and storage temperatures 0 to 60 °C
- Operating and storage humidity rating < 80%, non condensated
- Housing material PEHD
- Protection rating IP65, connector plugged in and tightened
6.6 CONFORMITY WITH STANDARDS
Electromagnetic compatibility:
- Emission EN 50081-1 (1992)
- Protection EN 50082-2 (1995)
- Safety EN 61010-1
22
8223
TECHNICAL DATA
6.7 DIMENSIONS (MM)
218
46
75
200
78
95
H
DN (mm)
H (mm)
T fitting Plastic spigot
Stainless steel
spigot
15 204.3
20 201.8
25 202.0
32 205.6
40 209.4 205.3
50 215.5 210.3
65 215.5 220.3 214.3
80 225.3 221.3
100 232.3 231.3
125 242.3
150 253.3
200 274.3
23
8223
INFORMATION
7.1 STANDARD DELIVERY
The standard delivery comprises:
- an 8223 sensor
- a set with 1 EPDM gasket + 1 FKM gasket
- an EN 175301-803 connector
7.2 ORDERING TABLE FOR SENSORS 8223
7.3 ORDERING TABLE FOR SPARE PARTS
Spare part Order code
EN 175301-803 socket with cable gland (type 2508) + screw + NBR gasket
438811
EN 175301-803 socket with cable gland (type 2508) + screw + silicone gasket
156927
EN 175301-803 socket with NPT 1/2'' reduction (type 2509) + screw + NBR gasket
162673
Silicone gasket for EN 175301-803 socket
440244
Set of 2 M20x1.5 plugs + 2 gaskets
444705
Retaining ring
619205
Tightening nut
619204
Set of gaskets (1 FKM, green + 1 EPDM, black)
552111
Output type Finger material Order code
4-20 mA
PVDF
440440
PP
558767
PEEK
550335
24
8223
INFORMATION
7.4 MEASUREMENT PRINCIPLE
Conductivity is the ability of a liquid / solution to conduct an electrical current. To measure the conductivity of a solution the 8223 conductivity sensor uses the following principle:
- A voltage is connected to the primary magnetic coil.
- The induced magnetic field generates a current in the secondary magnetic coil.
- The intensity of the current is a direct function of the conductivity of the solution between the 2 magnetic coils.
7.5 LABEL DESCRIPTION
1. Measured variable
2. Type of the sensor
3. Material of the gasket
4. Material of the finger
5. Power supply data
6. Type of output
7 CE logo
8. Manufacturing code
9. Series number
10. Order code
Made in France
COND:8223-FKM -PVDF 12-30V= 50 mA 4-20mA
N=10000 440440 W41UE
132
8910
4
5 6
7
Conducting fluid
1
1
1
2
2
2
25
8223
INFORMATION
Loading...
+ 57 hidden pages