Product Identification:
Operation Manual (Original) Pure Chromatography Instruments
11593990
Publication date: 05.2021
Version E
BÜCHI Labortechnik AG
Meierseggstrasse 40
Postfach
CH-9230 Flawil 1
E-Mail: quality@buchi.com
BUCHI reserves the right to make changes to the manual as deemed necessary in the light of experience,
especially with respect to structure, illustrations and technical details.
This manual is copyrighted. Information from it may neither be reproduced, distributed, or used for
competitive purposes, nor made available to third parties. The manufacture of any component with the
aid of this manual without prior written agreement is also prohibited.
Page 3
Büchi Labortechnik AGTable of contents
Table of contents
1About this document........................................................................................................... 8
1.1Warning notices in this document.......................................................................................... 8
Operation Manual Pure Chromatography Instrumentsvii
Page 8
1 | About this document
1About this document
1.1Warning notices in this document
Büchi Labortechnik AG
This operation manual is applicable for all variants of the instrument.
Read this operation manual before operating the instrument and follow the
instructions to ensure safe and trouble-free operation.
Keep this operation manual for later use and pass it on to any subsequent user or
owner.
BÜCHI Labortechnik AG accepts no liability for damage, faults and malfunctions
resulting from not following this operation manual.
If you have any questions after reading this operation manual:
u Contact BÜCHI Labortechnik AG Customer Service.
https://www.buchi.com/contact
Warning notices warn you of dangers that can occur when handling the device.
There are four danger levels, each identifiable by the signal word used.
Signal wordMeaning
DANGERIndicates a danger with a high level of risk which could result in
death or serious injury if not prevented.
WARNINGIndicates a danger with a medium level of risk which could result in
death or serious injury if not prevented.
CAUTIONIndicates a danger with a low level of risk which could result in mi-
nor or medium-severity injury if not prevented.
NOTICEIndicates a danger that could result in damage to property.
1.2Symbols
The following symbols are displayed in this operation manual or on the device:
1.2.1Warning symbols
SymbolMeaning
General warning
Dangerous electrical voltage
8/102
Flammable substances
Laser class 1
1.2.2Mark-ups and symbols
NOTE
This symbol draws attention to useful and important information.
Operation Manual Pure Chromatography Instruments
Page 9
Büchi Labortechnik AG
About this document | 1
R This character draws attention to a requirement that must be met before the
instructions below are carried out.
u This character indicates an instruction that must be carried out by the user.
ð This character indicates the result of a correctly carried out instruction.
Mark-upExplanation
Win dow
Tab
Dia log
[Button]Buttons are marked-up like this.
[Field names]Field names are marked-up like this.
[Menu / Menu item]Menus or menu items are marked-up like this.
StatusStatus is marked-up like this.
SignalSignals are marked-up like this.
1.3Trademarks
Product names and registered or unregistered trademarks that are used in this
document are used only for identification and remain the property of the owner in
each case.
1.4Connected devices
In addition to these operating instructions, follow the instructions and specifications
in the documentation for the connected devices.
Software Windows are marked-up like this.
Tabs are marked-up like this.
Dialogs are marked-up like this.
Operation Manual Pure Chromatography Instruments
9/102
Page 10
2 | Safety
Büchi Labortechnik AG
2Safety
2.1Proper use
The instrument is designed and built for laboratories.
The instrument can be used for the following tasks:
Purification
Separation of one or more compounds from a mixture
2.2Use other than intended
Use of any kind other than that described in Chapter2.1 "Proper use", page10 and
any application that does not comply with the technical specifications (See
Chapter3.5 "Technical data", page18) constitutes use other than intended.
In particular, the following applications are not permissible:
Use of the instrument in areas which require explosion-safe instruments.
Use of the instrument with solvents containing peroxides.
Use the ELSD without connected exhaust.
Use the instrument for production purposes.
2.3Staff qualification
Unqualified persons are unable to identify risks and are therefore exposed to greater
dangers.
The device may only be operated by suitably qualified laboratory staff.
These operating instructions are aimed at the following target groups:
Users
Users are persons that meet the following criteria:
They have been instructed in the use of the device.
They are familiar with the contents of these operating instructions and the
applicable safety regulations and apply them.
They are able on the basis of their training or professional experience to assess
the risks associated with the use of the device.
Operator
The operator (generally the laboratory manager) is responsible for the following
aspects:
The device must be correctly installed, commissioned, operated and serviced.
10/102
Only suitably qualified staff may be assigned the task of performing the
operations described in these operating instructions.
The staff must comply with the local applicable requirements and regulations for
safe and hazard-conscious working practices.
Safety-related incidents that occur while using the device should be reported to
the manufacturer (quality@buchi.com).
Operation Manual Pure Chromatography Instruments
Page 11
Büchi Labortechnik AG
2.4Location of warning signs on the product
Safety | 2
BUCHI service technicians
Service technicians authorized by BUCHI have attended special training courses and
are authorized by BÜCHI Labortechnik AG to carry out special servicing and repair
measures.
Fig.1: Location of the warning signs
Laser class 1
2.5Residual risks
The device has been developed and manufactured using the latest technological
advances. Nevertheless, risks to persons, property or the environment can arise if
the device is used incorrectly.
Appropriate warnings in this manual serve to alert the user to these residual
dangers.
2.5.1Faults during operation
If a device is damaged, sharp edges, moving parts or exposed electrical wires can
cause injuries.
u Regularly check device for visible damage.
u If faults occur, switch off the device immediately, unplug the power cord and
inform the operator.
u Do not continue to use devices that are damaged.
2.5.2Malware infection due to connections with other devices or network
Connections with other devices or a network can cause a malware infection to the
instrument.
u Install antivirus software and firewall on the instrument before connecting it to
other devices or network.
Operation Manual Pure Chromatography Instruments
11/102
Page 12
2 | Safety
Büchi Labortechnik AG
2.5.3Damage to the internal memory due to incorrect shutting down of the
instrument
Incorrect shutting down of the instrument can cause damage to the internal
memory.
u Shut down the instrument as described. See Chapter6 "Operation", page33
2.6Personal protective equipment
Depending on the application, hazards due to heat and/or corrosive chemicals may
arise.
u Always wear appropriate personal protective equipment such as safety goggles,
protective clothing and gloves.
u Make sure that the personal protective equipment meets the requirements of
the safety data sheets for all chemicals used.
2.7Modifications
Unauthorized modifications can effect safety and lead to accidents.
u Use only genuine BUCHI accessories, spare parts and consumables.
u Carry out technical changes only with prior written approval from BUCHI.
u Only allow changes to be made by BUCHI service technicians.
BUCHI accepts no liability for damage, faults and malfunctions resulting from
unauthorized modifications.
12/102
Operation Manual Pure Chromatography Instruments
Page 13
Büchi Labortechnik AG
Gradient valve
UV
detector
Fraction
collection
Solvent 2
Solvent 3
Solvent 4
Solvent 1
Pump
Sample
introduction
Cartridge / column
ELSD
3Product description
3.1Description of function
Product description | 3
Pure instruments are purification devices designed to purify complex samples by
either flash chromatography, prep chromatography or by both.
Flash chromatography has the ability to separate gram size samples in short
period.
Prep-HPLC has the ability to separate complex samples at high resolution.
Pure instruments allow:
4 different solvents
Injection of liquid or solid sample
Separation on a cartridge or column
Identifying the compounds by UV and/or ELS detection
Collecting the desired fractions
Pure instrument schematic:
Operation Manual Pure Chromatography Instruments
13/102
Page 14
3 | Product description
5
10
9
8
1
2
3
4
6
7
3.2Configuration
3.2.1Front view Pure C-810 / C-815
Büchi Labortechnik AG
Fig.2: Front view Pure C-810 / C-815
1Control panel2RFID reader
3On/Off switch4USB Port
5Injection port6Cartridge holder
7Flash pump8ELSD flow split valve
9ELSD nebulizer
(C-815 only)
10 Fraction collection bay
(C-815 only)
14/102
Operation Manual Pure Chromatography Instruments
Page 15
Büchi Labortechnik AG
10
9
8
7
1
2
3
4
5
6
3.2.2Front view Pure C-830 / C-835 / C-850
Product description | 3
Fig.3: Front view Pure C-830 / C-835 / C-850
1Control panel2RFID reader
3On/Off switch4USB Port
5Column holder6Prep HPLC pump
7ELSD flow split valve
(C-835 / C-850 only)
9Prep sample injection valve10Fraction collection bay
3.2.3Rear view
NOTE
All electrical connections are not limited energy.
(C-850 only)
8ELSD nebulizer
(C-835 / C-850 only)
Operation Manual Pure Chromatography Instruments
15/102
Page 16
3 | Product description
3
7
8
9
10
2
1
6
5
4
Büchi Labortechnik AG
Fig.4: Rear view
1Ventilation slot2Gradient valve
(See Chapter3.2.4 "Connections on
gradient valve", page17 )
3Exhaust
(used for instruments with ELSD
4Signal connection
(for external air supply)
only)
5USB ports6Power supply connection
7Fuse8On/Off master switch
9Pressurized air inlet10 LAN port
16/102
Operation Manual Pure Chromatography Instruments
Page 17
Büchi Labortechnik AG
6
5
4
3
21
10
9
8
7
3.2.4Connections on gradient valve
Product description | 3
Fig.5: Connections on the rear side
1Waste line2Solvent line 4
3Solvent line 34Waste level sensor
5Solvent line 26Solvent line 1
7Solvent level sensor line 18Solvent level sensor line 2
9Solvent level sensor line 310 Solvent level sensor line 4
3.3Type plate
The type plate identifies the instrument. The type plate is located at the rear of the
instrument.
Operation Manual Pure Chromatography Instruments
17/102
Page 18
3 | Product description
Buchi Labortechnik AG
9230 Flawil / Switzerland
Type: xxx
SN: 10xxxxxxxx
Volt: xxx-xxx VAC
Frequ.: xx Hz
Power: xxxx W
Built: 20xx
Made in Switzerland
6
5
4
1
3
2
7
8
9
10
9
Büchi Labortechnik AG
Fig.6: Type Plate
1Company name and address2Instrument name
3Serial number4Input voltage range
5Frequency6Power consumption maximum
7Year of manufacture8Product code
9Approvals10 Symbol for "Do not dispose of as
3.4Scope of delivery
NOTE
The scope of delivery depends of the configuration of the purchase order.
Accessories are delivered as per the purchase order, order confirmation, and
delivery note.
Maximum relative humidity80% for temperatures up to 31°C
Storage temperaturemax. 45 °C
Operation Manual Pure Chromatography Instruments
Page 21
Büchi Labortechnik AG
3.5.3Material
Product description | 3
ComponentMaterials of construction
HousingPowder coated steel 1.4301
Fraction collection bayPMMA / PET
Pump headPEEK
Pump coverPP
Pump excenter housingAluminum
TubingsFEP
FittingPOM
Valve screw fittingPOM
FerruleETFE
Cone ringPOM
Radial sealPTFE
PistonsCeramic
Operation Manual Pure Chromatography Instruments
21/102
Page 22
4 | Transport and storage
4Transport and storage
4.1Transport
4.2Storage
Büchi Labortechnik AG
NOTICE
Risk of breakage due to incorrect transportation
u Make sure that all parts of the device are safely packed in such a way as to
prevent breakage, ideally in the original box.
u Avoid sharp movements during transit.
u After transportation, check the device for damage.
u Damage that has occurred in transit should be reported to the carrier.
u Keep packing for future transportation.
u Make sure that the ambient conditions are complied with (see Chapter3.5
"Technical data", page18).
u Make sure a clean solvent like ethanol or isopropanol is in the pump.
u Wherever possible, store the device in its original packaging.
u After storage, check the device, all seals and tubing for damage and replace if
necessary.
22/102
Operation Manual Pure Chromatography Instruments
Page 23
Büchi Labortechnik AG
4.3Lifting the instrument
Transport and storage | 4
WARNING
Danger due to incorrect transportation
The possible consequences are crushing injuries, cuts and breakages.
u The instrument should be transported by two persons at the same time.
u Lift the instrument at the points indicated.
u Lift the instrument – this requires two persons each lifting at one of the points
indicated on the bottom of the instrument.
Operation Manual Pure Chromatography Instruments
23/102
Page 24
5 | Installation
1
Büchi Labortechnik AG
5Installation
5.1Before installation
NOTICE
Instrument damage due to switching it on too early.
Switching on the instrument too early after transportation can cause damage.
u Climatize the instrument after transportation.
5.2Installation site
The installation site must meet the following requirements:
Firm, level surface.
Take into account the maximum product dimensions and weight. See Chapter3.5
"Technical data", page18
Clearance on each side of the instrument must be at least 200 mm.
Do not expose the instrument to any external thermal loads, such as direct solar
radiation.
Do not expose the instrument to increased electromagnetic emissions.
Electromagnetic fields in the frequency range between 200 to 300 MHz can cause
the instrument to operate incorrectly.
Make sure that the installation site meets the requirements of the safety data
sheets for all solvents and samples used.
NOTE
Make sure that the power supply can be disconnected at any time in an emergency.
5.3Securing against earthquakes
The instrument has an earthquake fixing point to protect the device against falling.
24/102
Fig.7: Securing against earthquakes
1Fixing point
u Tie the lashing mount to a fixed point using strong cord or a wire.
Operation Manual Pure Chromatography Instruments
Page 25
Büchi Labortechnik AG
5.4Establishing electrical connections
Installation | 5
NOTE
Observe the legal requirements when connecting the instrument to the power
supply.
u Use additional electrical safety features (e.g., residual-current circuit breakers) to
comply with local laws and regulations.
The power supply must fulfil the following conditions:
1. Provide the mains voltage and frequency specified on the type plate of the
instrument.
2. Be designed for the load imposed by the instruments connected.
3. Be equipped with suitable fuses and electrical safety features.
4. Be equipped with proper earthing.
NOTICE
Risk of property damage and diminished performance due to use of unsuitable
power cables.
The power supply cables supplied with the product by BUCHI precisely match the
requirements of the device. If other power cables that do not meet those
requirements are used, the device may be damaged and/or its performance
diminished.
u Use only the power supply cables supplied with the product or ordered
separately from BUCHI.
u If using any other power supply cables, make sure that they match the
specifications on the type plate.
u Make sure that all connected devices are earthed.
u Plug the power cable into the connection on the instrument. See Chapter3.2
"Configuration", page14
u Plug the mains plug into the mains outlet socket.
5.5Establishing solvent connection
NOTICE
Solvent bottles on top of the instrument.
Solvent bottles on top of the instrument can cause property damages.
u Locate the solvent bottles next to the instrument.
u Use the optional solvent bottle platform.
Operation Manual Pure Chromatography Instruments
25/102
Page 26
5 | Installation
Büchi Labortechnik AG
Precondition:
R Make sure that the instrument is not
connected to the power supply.
u Attach all solvent lines to the
instrument. Connections see
Chapter3.2.4 "Connections on gradient
valve", page17
u Put the other end of the solvent line
into the solvent bottle.
u Assign the solvent to the solvent lines.
See Chapter5.7 "Assigning solvent to
solvent lines", page26
u Install the solvent level sensor. See
Chapter5.6 "Installing the solvent level
sensor", page26
5.6Installing the solvent level sensor
Precondition:
R Make sure that the instrument is not
connected to the power supply.
u Attach all solvent level sensors to the
instrument. Connections see
Chapter3.2.4 "Connections on gradient
valve", page17
u Calibrate the solvent level sensors. See
Chapter5.8 "Calibrating the solvent
level sensor", page27
5.7Assigning solvent to solvent lines
26/102
Navigation path
➔ Tools ➔ Solvent Loading
Precondition:
R The solvent bottle is connected to the instrument. See Chapter5.5 "Establishing
solvent connection", page25
Operation Manual Pure Chromatography Instruments
Page 27
Büchi Labortechnik AG
Installation | 5
R The solvent you wish to use is part of the solvent library. See Chapter6.3 "Editing
a solvent", page53
u Navigate to the
ð The display shows the dialog box
u Tap the drop-down list next to Line 1.
Sol vent Lo ading
dialog according to the navigation path.
Sol vent Lo ading
ð The display shows a drop-down list with selectable solvents.
u Select the solvent which is connected to solvent line 1.
ð The solvent for Line 1 is assigned.
ð The drop-down list closes.
u Repeat the solvent selection for each line.
u Activate the checkbox next to the line you wish to prime.
u Tap the button [Auto Prime].
u Wait till priming finished.
u Tap the button [Close].
ð All lines are assigned with solvents.
ð The dialog box closes.
5.8Calibrating the solvent level sensor
.
Navigation path
➔ Tools ➔ Calibration and Defaults
Precondition:
R The solvent level sensor is not covered by solvent.
u Navigate to the
Cal ibra tio n
ð The display shows the dialog box
u Tap [Zero] for the related solvent line.
u Tap the button [Close].
ð The dialog box closes.
Operation Manual Pure Chromatography Instruments
dialog according to the navigation path.
Cal ibra tio n
.
27/102
Page 28
5 | Installation
1
2
Büchi Labortechnik AG
5.9Installing the exhaust (ELSD only)
u Press the exhaust gas hose (2) on the
exhaust (1).
5.10Assembling the waste bottle
NOTICE
Waste bottle on top of the instrument
Waste bottle on top of the instrument can cause property damages.
u Make sure that there is a decline between the connection on the instrument and
the waste bottle.
u Put the waste line into the waste bottle.
5.11Installing the waste level sensor
Precondition:
R Make sure that the instrument is not
connected to the power supply.
u Attach the waste level sensor to the
instrument. Connections see
Chapter3.2.4 "Connections on gradient
valve", page17
u Calibrate the waste level sensor. See
Chapter5.12 "Calibrating the waste
level sensor", page29
28/102
Operation Manual Pure Chromatography Instruments
Page 29
Büchi Labortechnik AG
5.12Calibrating the waste level sensor
Installation | 5
Navigation path
➔ Tools ➔ Calibration and Defaults
Precondition:
R The waste level sensor is not covered by liquid waste.
u Navigate to the
Cal ibra tio n
ð The display shows the dialog box
u Tap [Zero] for [Waste].
u Tap the button [Close].
dialog according to the navigation path.
Cal ibra tio n
ð The dialog box closes.
5.13Assembling the dry air supply (option)
u Connect the signal cable from the air
supply to the instrument.
u Connect the gas connection. See
Chapter5.14 "Assembling the gas
connection (option)", page30.
.
Operation Manual Pure Chromatography Instruments
29/102
Page 30
5 | Installation
Büchi Labortechnik AG
5.14Assembling the gas connection (option)
Precondition:
R The gas supply complies with the
specifications. See Chapter3.5 "Technical
data", page18
u Cut 50 mm from the gas tubing.
u Attach the cut tubing to the air filter.
u Secure the tubing in place.
u Attach the tubing from gas supply to the other
side of the air filter.
u Secure the tubing in place.
u Connect the tubing from the gas supply to the
instrument.
30/102
Operation Manual Pure Chromatography Instruments
Page 31
Büchi Labortechnik AG
5.15Assembling the solvent bottle platform (option)
Installation | 5
NOTICE
Waste bottle on top of the instrument
Waste bottle on top of the instrument can cause property damages.
u Do not place the waste bottle an the solvent bottle platform.
u Turn the solvent bottle platform upside down.
u Attach the drain line adapter to the solvent
bottle platform
u Attach the drain line to the drain line adapter.
u Attach the solvent bottle platform to the
instrument with screws.
Operation Manual Pure Chromatography Instruments
31/102
Page 32
5 | Installation
Büchi Labortechnik AG
5.16Assembling the sample loop (Prep instruments only)
u Attach the sample loop to the
instrument at the position indicated.
32/102
Operation Manual Pure Chromatography Instruments
Page 33
Büchi Labortechnik AG
5
9
8
7
1
2
3
4
6
6Operation
6.1Control panel
6.1.1Layout of control panel
Operation | 6
Fig.8: Display
No.DescriptionFunction
1[Menu] barShows the available menus.
See Chapter6.1.3 "Menu bar", page35
2[Conditions] panelShows the properties and default settings
of the installed column / cartridge.
See Chapter6.1.4 "Conditions panel",
page37
3[Wavelength] panelShows available wavelengths and scan
options.
4[Collection] panelShows collection options.
See Chapter6.1.5 "Collection panel",
page38
5[Gradient] panelShows chromatograms and gradient ta-
[View Manual]The display shows the Operation manual.
Operation Manual Pure Chromatography Instruments
Page 37
Büchi Labortechnik AG
1
6.1.4Conditions panel
Operation | 6
1Information button
(Information about the selected column)
The conditions area contains the following settings:
ActionOptionExplanation
[Column]
(Prep mode)
[Column]
(Flash mode)
[Flow Rate]Enter valueEdit the default flow rate.
[Duration Units]Choice of the duration
Enter valueEnter a name for the present car-
tridge.
Select column /automatic via RFID
Shows the name of the column installed in the instrument.
The following types are available:
type
minutes / column volumes
[Equilibration]Enter valueIndicates the period of time or num-
ber of column volumes that the mobile phase flows through the column
before the sample is injected.
[Run Length]Enter valueAccording to the current operation
enter the time for the separation.
Operation Manual Pure Chromatography Instruments
According to the current operation
enter the number of column volumes
required for the separation.
37/102
Page 38
6 | Operation
1
Büchi Labortechnik AG
ActionOptionExplanation
[Air Purge Time]Enter valueIndicates the period of time that air
is passed through the column after
the separation to remove mobile
phase
6.1.5Collection panel
1Tray number
The collection vial matrix corresponds to the trays. The trays are detected by the
auto recognition. The estimated number of vials required for the separation is
displayed below the solvent usage list in the lower left corner of the Setup window.
The fraction collection area contains the following settings:
ActionOptionExplanation
Fraction collection
options
Select valueThe following options are available:
[Collect Peaks] / [Collect All] / [Collect
None]
[Per-Vial Volume]View / Enter valueThe following values are changeable:
[Peak] / [Non-Peaks]
38/102
Operation Manual Pure Chromatography Instruments
Page 39
Büchi Labortechnik AG
6.1.6Gradient panel
Operation | 6
Edit the gradient and see separation details.
NOTE
According to the instrument status different options are available.
ActionOptionExplanation
[Table]Create a gradientSee Chapter "Editing the gradient in
tabular mode", page46
[Navigator]Finding out Flash sepa-
ration conditions
See Chapter10.3 "Finding out separation conditions with the navigator",
page97
[UV scan details]ViewShows the following charts:
3D (UV / Time/ Wavelength)
2D (Time / Wavelength)
Absorption maxima
All scan maxima
[Zoom]FunctionZoom the graph.
[Options]Select valuesGraph options.
[Edit]Select between the op-
tions
The graph is in edit mode. See Chapter "Editing the gradient in graphic
mode", page47
[View]The graph is in view mode.
[Zoom]Zoom the graph.
Operation Manual Pure Chromatography Instruments
(No changes possible)
39/102
Page 40
6 | Operation
Büchi Labortechnik AG
ActionOptionExplanation
[Gradient hold]FunctionThe gradient is held at the current
solvent ratio.
The gradient continues to the original end-point.
[Auto gradient hold] FunctionThe gradient will be held every time
the signal goes over the set threshold.
6.1.7Run panel
The run panel shows available functions according to the current operation.
SymbolDescriptionMeaning
[Start]Is used to start the run or re-start the
actual operation if the system has
been paused.
[Stop]Is used to terminate the operation of
the system.
[Pause]Is used to stop the actual operation.
If the system is paused due to an error, this button will change to yellow.
[Advance]Is used to advance to the next step
during equilibration.
40/102
Operation Manual Pure Chromatography Instruments
Page 41
Büchi Labortechnik AG
6.1.8Solvent selection panel
Operation | 6
Select solvents for a separation. See Chapter6.2.3 "Selecting a solvent", page45
6.1.9Detector selection panel
ActionOptionExplanation
[ELSD]Enable / Disable
Select value
Operation Manual Pure Chromatography Instruments
Action enabled:
Select between Low / High
Enter values for threshold
Action disabled:
No selection
41/102
Page 42
6 | Operation
Büchi Labortechnik AG
ActionOptionExplanation
[UV]Enable / Disable
Select value / Enter
value
[Threshold Detec-
Enter valueThe threshold defines the value
tion]
6.1.10Slope detection panel
Action enabled:
Low / High
Enter values for threshold
Action disabled:
No selection
above the fraction collector starts to
collect fractions.
ActionOptionExplanation
[Slope detection]Enable / DisableAction enabled:
6.2Editing a method
6.2.1Selecting a cartridge (Flash mode)
NOTE
The automatic reading of the cartridge information can only be done with specific
BUCHI RFID tagged cartridges.
u The indication for recognizing the cartridge is a sound.
Fraction collection triggered by slope
Action disabled:
No detection
42/102
Operation Manual Pure Chromatography Instruments
Page 43
Büchi Labortechnik AG
1
Operation | 6
1Information button
(Information about the selected column)
Navigation path
➔ Conditions panel
u Navigate to the
u Tap the input box next to [Column].
Con diti ons select ion
panel according the navigation path.
ð The display shows a menu with selectable cartridges.
u Select the cartridges you wish to use.
Operation Manual Pure Chromatography Instruments
43/102
Page 44
6 | Operation
1
Büchi Labortechnik AG
6.2.2Selecting a column (Prep mode)
1Information button
(Information about the selected column)
Navigation path
➔ Conditions panel
u Navigate to the
u Tap the input box next to [Column].
Con diti ons select ion
panel according the navigation path.
ð The display shows a dialog with an alphanumeric input box.
u Enter a name for the column.
u Tap the button [OK].
u Tap the input box next to [Flow rate].
ð The display shows a dialog box with a numeric input box.
u Enter a value for the flow rate.
u Tap the button [OK].
u Tap the input ox next to [Max. Pressure].
ð The display shows a dialog box with a numeric input box.
u Enter the maximum pressure for the column.
u Tap the button [OK].
44/102
Operation Manual Pure Chromatography Instruments
Page 45
Büchi Labortechnik AG
6.2.3Selecting a solvent
Operation | 6
Navigation path
➔ Solvent selection panel
Precondition:
R The required solvents lines are connected and assigned. See Chapter5.7
"Assigning solvent to solvent lines", page26
u Navigate to the
u Tap the drop-down list next to A: .
Sol vent se lection
ð The display shows a drop-down list with the assigned solvents.
u Tap the required solvent.
ð The solvent is selected
ð The drop-down list closes.
u Select more solvents for the mobile phase according to your needs.
6.2.4Editing the gradient
The composition of the mobile phase as a function of time can be indicated by
entering the gradient. Four solvent lines can be used to generate a binary gradient.
The solvents used to define the gradient can be altered during the separation.
panel according the navigation path.
Operation Manual Pure Chromatography Instruments
45/102
Page 46
6 | Operation
Büchi Labortechnik AG
Editing the gradient in tabular mode
The following settings are available:
ActionOptionExplanation
[Min]Enter valueEnter the time until the value in col-
umn [% 2nd] is reached.
[Solvents]Select valueSelect solvent line combinations.
[% 2nd]Enter valueEnter the percentage rate for the
second solvent.
Navigation path
➔ Gradient panel
u Navigate to the
u Tap the button [Table].
Gra dien t
ð The display shows the dialog
u Tap the cell for [Min].
panel according the navigation path.
Gra dien t T able
.
ð The display shows a dialog box with a numeric input box.
46/102
Operation Manual Pure Chromatography Instruments
Page 47
Büchi Labortechnik AG
Operation | 6
u Enter the time.
u Tap the button [OK].
u Tap the cell for [AB▼].
ð The display shows a drop-down list with solvent line combinations.
u Select the combination that you want to use.
u Tap the cell for [% 2nd].
ð The display shows a dialog box with a numeric input box.
u Enter the percentage for the second solvent.
u Tap the button [OK].
u Tap the button [Close].
ð The gradient is saved
ð The dialog Gradient table closes
ð The
Gra dien t
panel shows the set gradient.
Add additional lines to the Gradient table
u Tap the number field (e.g. 1►).
ð The display shows a drop-down list with selectable actions.
u Select if the line should be added above or below the selected line.
ð A line is added.
Deleting lines from the Gradient table
u Tap the number field (e.g. 1►).
ð The display shows a drop-down list with selectable actions.
u Select delete.
ð The line is deleted.
Editing the gradient in graphic mode
Navigation path
➔ Gradient panel
u Navigate to the
u Tap the button [Edit].
Gra dien t
ð The Display shows a drop-down list.
u Select [Edit].
panel according the navigation path.
Add steps to the Gradient graphic
u Tap on the line at the time for which you want to edit the gradient and drag it to
the desired %B, then release.
Deleting steps from the Gradient graphic
u To remove a step, drag the point to the baseline or to any gray area around the
graph until a red X is visible, then release.
Operation Manual Pure Chromatography Instruments
47/102
Page 48
6 | Operation
Büchi Labortechnik AG
Change solvent combinations
u Solvent line combinations can be accessed by clicking on the AB▼ box to reveal a
drop-down list.
6.2.5Editing detector selection
Navigation path
➔ Detector selection panel
u Navigate to the
u Activate checkbox for the detector you wish to use.
Det ecto r s electio n
panel according the navigation path.
ð The detector is activated.
6.2.6Editing the wavelength for the UV detector
48/102
Operation Manual Pure Chromatography Instruments
Page 49
Büchi Labortechnik AG
1
Operation | 6
Status CheckboxExplanationExplanation
OnThe instrument collects fractions.
MonitoringThe instrument records the data from the
UV detector but does not collect the fractions.
OffThe instrument does not record data
from the UV detector and does not collect fractions.
Navigation path
➔ Wavelength selection panel
Precondition:
R The UV detector is selected.
u Navigate to the
u Activate the checkbock next to [UV].
u Tap the input box next to [UV].
Wav elen gth
panel according the navigation path.
ð The display shows a dialog box with a numeric input box.
u Enter a value for the wavelength.
u Tap the button [OK].
ð The dialog box closes.
ð The wavelength is saved.
6.2.7Editing the fraction collection criteria
1Identifier
Operation Manual Pure Chromatography Instruments
49/102
Page 50
6 | Operation
Büchi Labortechnik AG
The following fraction collection criteria are available:
CriteriaMeaning
[Collect Peaks]The instrument collects fractions if one detector signal is
above the set threshold.
[Collect All]The instrument collects all the fractions regardless of de-
tection signals.
[Collect None]The instrument collects no fractions.
NOTE
[Peak] and [Non-Peaks] default is the maximum volume of the vial.
Navigation path
➔ Collection panel
u Tap the radio button next to criteria you wish to use.
ð The criteria is selected.
u Tap the input box next to [Peak].
ð The display shows a dialog box with a numeric input box.
u Enter a the volume you wish to collect.
u Tap the button [OK].
ð The dialog box closes.
ð The value for the volume is saved.
u Tap the input box next to [Non-Peaks].
ð The display shows a dialog box with a numeric input box.
u Enter the volume you wish to collect.
u Tap the button [OK].
ð The dialog box closes.
50/102
6.2.8Editing the fraction collection time
Operation Manual Pure Chromatography Instruments
Page 51
Büchi Labortechnik AG
Operation | 6
Program Collect allows to turn off fraction collection for a specific time.
The fraction collector defaults to collect the full run length unless the values from
Program Collect override the collection.
Navigation path
➔ Gradient panel
u Navigate to the
u Tap the button [Program collect].
Gra dien t
ð The display shows the dialog
u Tap the input box.
panel according the navigation path.
Pro gram co lle ct
.
ð The display shows a dialog box with a numeric input box.
u Enter the time.
u Tap the button [OK].
ð The time is saved.
ð The dialog box closes.
Add additional lines to the program collection
u Tap the number field (1►).
ð The display shows a drop-down list with selectable actions.
u Select if the line should be added above or below the selected line.
ð A line is added.
Deleting lines from the program collection
u Tap the number field (1►).
ð The display shows a drop-down list with selectable actions.
u Select delete.
ð The line is deleted.
Operation Manual Pure Chromatography Instruments
51/102
Page 52
6 | Operation
Büchi Labortechnik AG
6.2.9Editing the vapor sensor sensitivity
The vapor sensor detects solvent concentration in the ambient air.
The following sensitivity limits are available:
SensitivityMeaning
HighUsed for non-volatile solvents
MediumCompromise between the low and high setting
LowUsed for volatile or semi-volatile solvents
OffThe vapor sensor is off
Navigation path
➔ Tools ➔ Vapor Sensor and Limits
u Navigate to the
ð The display shows the dialog box
u Tap the drop-down list next [Vapor alarm sensitivity].
u Select the sensitivity value you wish to use.
u Tap the button [Close].
ð The dialog box closes.
6.2.10Saving a method
Vap or S ens ors
dialog according to the navigation path.
Vap or S ens ors
.
52/102
Navigation path
➔ File ➔ Save Method as
Operation Manual Pure Chromatography Instruments
Page 53
Büchi Labortechnik AG
Operation | 6
u Navigate to the
Sav e Me tho d as
ð The display shows the dialog box
u Tap the input box [Enter method name].
ð The display shows a dialog with an alphanumeric input box.
u Enter the name of the method.
u Tap the button [OK].
ð The dialog box closes.
u Tap the button [OK].
ð The method is saved.
ð The dialog box closes.
6.3Editing a solvent
6.3.1Adding a new solvent
dialog according to the navigation path.
Sav e Me tho d as
.
Navigation path
➔ Tools ➔ Solvent Definition
u Navigate to the
ð The display shows the dialog box
u Tap the button [Add Solvent].
ð The display shows the dialog box
u Tap the input box next to [Name].
Sol vent De finitio n
Sol vent De finitio n
Sol vent
dialog according to the navigation path.
.
.
ð The display shows a dialog with an alphanumeric input box.
u Enter the name for the solvent.
u Tap the button [OK].
ð The dialog box closes.
u Tap the input box next to [Info].
ð The display shows a dialog with an alphanumeric input box.
u Enter information to the solvent according to your requirements.
u Tap the button [OK].
ð The dialog box closes.
u Tap the button [Verify Solvents].
ð The display shows the dialog box
Operation Manual Pure Chromatography Instruments
Sol vent Ve rificat ion
.
53/102
Page 54
6 | Operation
Büchi Labortechnik AG
u Select solvent group for the defined solvent.
u Tap the button [OK].
ð The dialog box closes.
ð The solvent is added.
u Tap the button [Close].
ð The dialog box
Sol vent De finitio n
6.3.2Deleting a solvent
Navigation path
➔ Tools ➔ Solvent Definition
u Navigate to the
ð The display shows the dialog box
u Select the solvent you wish to delete.
u Tap the button [Delete Solvent].
u Answer the secure question with [Yes].
ð The solvent is deleted.
Sol vent De finitio n
closes.
dialog according to the navigation path.
Sol vent De finitio n
.
54/102
6.4Tasks during a separation
6.4.1Introducing a sample into the prep system
u Connect the syringe at the sample loop
port.
Operation Manual Pure Chromatography Instruments
Page 55
Büchi Labortechnik AG
Operation | 6
u Press the plunger.
u Wait until the run starts.
u Disconnect the syringe at the inert loop
port.
6.4.2Injecting a sample into the flash system
Injecting a sample into the flash system at the luer port
CAUTION
Not removed syringe after injection.
A not removed syringe after injection can lead to spill of solvent and injuries.
u Remove the syringe after injection.
Operation Manual Pure Chromatography Instruments
55/102
Page 56
6 | Operation
Büchi Labortechnik AG
u Connect the syringe at the luer port.
u Press the plunger.
u Disconnect the syringe at the luer port.
Injecting a sample into the flash system with a solid loader
NOTE
Removing is done in reverse sequence.
u Loosen the solvent line on the top of
the cartridge holder.
56/102
Operation Manual Pure Chromatography Instruments
Page 57
Büchi Labortechnik AG
Operation | 6
u Attach the solid loader to the cartridge
holder.
u Attach the solvent line to the solid
loader.
6.4.3Installing and removing a cartridge
NOTE
Removing is done in reverse sequence.
NOTE
Instead of a cartridge a bypass can be installed.
NOTE
The automatic reading of the cartridge information can only be done with specific
BUCHI RFID tagged cartridges.
u The indication for recognizing the cartridge is a sound.
u Hold the cartridge above the RFID
reader.
u Wait until the instrument has taken
over the cartridge data.
Operation Manual Pure Chromatography Instruments
57/102
Page 58
6 | Operation
Büchi Labortechnik AG
u Lift the clamping plate to release the
lock mechanism and pull up the upper
part of the cartridge holder.
u Install the cartridge by pushing the
cartridge inlet down into the luer
fitting at the bottom plate.
u Lower the upper part of the cartridge
holder over the luer connection.
u Align it.
u Firmly press the arm in place to ensure
a proper seal.
6.4.4Installing and removing a column
NOTE
Removing is done in reverse sequence.
58/102
Operation Manual Pure Chromatography Instruments
Page 59
Büchi Labortechnik AG
Operation | 6
u Install the column in the bottom plate
of the column holder.
u Lower down the upper part of the
column holder.
u Remove the tubing from the upper part
of the column holder.
u Attach the tubing to the column.
Operation Manual Pure Chromatography Instruments
59/102
Page 60
6 | Operation
1
2
Büchi Labortechnik AG
u Remove the lower tubing from the
upper part of the column holder.
u Attach the tubing to the column.
6.4.5Inserting the fraction collection tray
NOTE
By default the first available tube on each tray is reserved for waste.
Changing the setting. See Chapter6.4.7 "Selecting values on the dialog box Sample
Loading", page61
60/102
1Tab2Optical sensor slot
u Place the collection tubes in the collection tray.
Operation Manual Pure Chromatography Instruments
Page 61
Büchi Labortechnik AG
6.4.6Selecting an existing method
Operation | 6
u Place the trays in the fraction collector bay.
u Make sure that the tab is in the optical sensor slot.
Navigation path
➔ File ➔ Open Method
u Navigate to the
Ope n Me tho d
ð The display shows the dialog box
u Select the method you wish to use.
u The display highlights the selected method black.
u Tap the button [OK].
dialog according to the navigation path.
Ope n Me tho d
.
ð The method is selected.
ð The dialog box closes.
6.4.7Selecting values on the dialog box Sample Loading
The display shows the dialog box
separation.
The following settings are possible:
ActionExplanation
[Lower flow rate for
specified time after injection]
[Flush 2nd solvent at the
Reduce flow rate for a specified period, when the pressure
increases significantly at the start of the run after sample
injection.
Automatically cleaning the system and column.
end of the run]
Sam ple Loa din g
during the starting phase of a
[Override Start Vial]Select the start vial according to your needs.
6.5Performing a Flash separation using a method
NOTE
Pure C-810 / C-815 / C-850 only
6.5.1Preparing the instrument
Time required:approx. 30 sec.
Precondition:
R All commissioning operations have been completed. See Chapter5 "Installation",
page24
u Switch the On/Off master switch to On.
u Tap the On/Off switch.
ð The instrument is starting up.
Operation Manual Pure Chromatography Instruments
61/102
Page 62
6 | Operation
6.5.2Selecting the flash mode (Pure C-850 only)
Navigation path
➔ Mode
Mod e
u Navigate to the
u Select [Flash].
menu according to the navigation path.
ð The flash mode is selected.
6.5.3Starting Flash separation using a method
Precondition:
R The instrument is prepared. See Chapter6.5.1 "Preparing the instrument",
page61
R The required solvents are connected and assigned. See Chapter5.7 "Assigning
solvent to solvent lines", page26
R The sample is prepared.
R The waste bottle is empty.
Büchi Labortechnik AG
u Calibrate the solvent level sensor. See Chapter5.8 "Calibrating the solvent level
sensor", page27
u Calibrate the waste level sensor. See Chapter5.12 "Calibrating the waste level
sensor", page29
u Open the protection shield.
u Place the fraction collection trays inside the instrument. See Chapter6.4.5
"Inserting the fraction collection tray", page60
u Close the protection shield.
u Open an existing method. See Chapter6.4.6 "Selecting an existing method",
page61
u Tap the button [OK].
u Tap the button [Start] on the
u According the requirements adjust the file name.
u Tap the button [OK].
ð The display shows the dialog box
u According the requirements adjust the settings. See Chapter6.4.7 "Selecting
Run
panel.
Sam ple Loa din g
.
values on the dialog box Sample Loading", page61
u Follow the instructions on the display.
ð Installing the cartridge. See Chapter6.4.3 "Installing and removing a cartridge",
page57
ð Introducing the sample into the system. See Chapter6.4.2 "Injecting a sample
into the flash system", page55
62/102
6.5.4Changings during a separation
NOTE
Parameters that can be edited are highlighted in green
Operation Manual Pure Chromatography Instruments
Page 63
Büchi Labortechnik AG
6.5.5Ending a Flash separation
Operation | 6
Possibilities to edit the gradient during a separation:
Change the gradient. See Chapter "Editing the gradient in graphic mode", page47
Hold the gradient. See Chapter6.1.6 "Gradient panel", page39
Auto gradient hold. See Chapter6.1.6 "Gradient panel", page39
NOTE
The separation extends automatically by 5 minutes if the baseline at the end of the
separation is not below the set threshold.
Precondition:
R The display shows the dialog box
u According to the requirements extend the separation time by 5 minutes.
Precondition:
R The display shows the dialog box
u Follow the instructions on the display.
ð Purging the instrument with air. See Chapter7.8 "Removing solvent from an
used cartridge", page78
ð Removing the cartridge. See Chapter6.4.3 "Installing and removing a cartridge",
page57
6.5.6Shutting down the instrument
Navigation path
➔ File
Precondition:
R The separation process has ended.
u Purge the instrument with purging solvent. See Chapter7.6 "Cleaning the
instrument", page77
u Navigate to the [File] menu via the navigation path.
u Tap the action [Shut down].
u Confirm the secure question with [Yes].
ð The instrument is shutting down.
Sep arat ion End
Run com ple ted
.
.
6.6Performing a Flash separation manually
NOTE
Pure C-810 / C-815 / C-850 only
6.6.1Preparing the instrument
Time required:approx. 30 sec.
Precondition:
R All commissioning operations have been completed. See Chapter5 "Installation",
page24
u Switch the On/Off master switch to On.
Operation Manual Pure Chromatography Instruments
63/102
Page 64
6 | Operation
u Tap the On/Off switch.
ð The instrument is starting up.
6.6.2Selecting the flash mode (Pure C-850 only)
Navigation path
➔ Mode
Mod e
u Navigate to the
u Select [Flash].
menu according to the navigation path.
ð The flash mode is selected.
6.6.3Starting Flash separation manually
Precondition:
R The instrument is prepared. See Chapter6.6.1 "Preparing the instrument",
page63
R The required solvents are connected and assigned. See Chapter5.7 "Assigning
solvent to solvent lines", page26
Büchi Labortechnik AG
R The sample is prepared.
R The waste bottle is empty.
u Calibrate the solvent level sensor. See Chapter5.8 "Calibrating the solvent level
sensor", page27
u Calibrate the waste level sensor. See Chapter5.12 "Calibrating the waste level
sensor", page29
u Open the protection shield.
u Place the fraction collection trays inside the instrument. See Chapter6.4.5
"Inserting the fraction collection tray", page60
u Close the protection shield.
u Select a cartridge. See Chapter6.2.1 "Selecting a cartridge (Flash mode)", page42
u Tap the drop-down list next to A: on the
Sol vent Se lection
panel.
ð The display shows a drop-down list with the assigned solvents.
u Tap the required solvent.
ð The solvent is selected
ð The drop-down list closes.
u Select more solvents for the mobile phase according to your needs.
u Edit the gradient according to your needs. See Chapter6.2.4 "Editing the
gradient", page45
64/102
u Select the sample collection in the
Col lect ion
panel. See Chapter6.2.7 "Editing
the fraction collection criteria", page49
u Select the collection criteria in the
u Tap the button [Start] on the
ð The display shows the dialog box
u According the requirements adjust the settings. See Chapter6.4.7 "Selecting
Col lect ion criter ia
Run
panel.
Sam ple Loa din g
panel.
.
values on the dialog box Sample Loading", page61
Operation Manual Pure Chromatography Instruments
Page 65
Büchi Labortechnik AG
6.6.4Changings during a separation
6.6.5Ending a Flash separation
Operation | 6
u Follow the instructions on the display.
ð Installing the cartridge. See Chapter6.4.3 "Installing and removing a cartridge",
page57
ð Introducing the sample into the system. See Chapter6.4.2 "Injecting a sample
into the flash system", page55
NOTE
Parameters that can be edited are highlighted in green
Possibilities to edit the gradient during a separation:
Change the gradient. See Chapter "Editing the gradient in graphic mode", page47
Hold the gradient. See Chapter6.1.6 "Gradient panel", page39
Auto gradient hold. See Chapter6.1.6 "Gradient panel", page39
NOTE
The separation extends automatically by 5 minutes if the baseline at the end of the
separation is not below the set threshold.
Precondition:
R The display shows the dialog box
u According to the requirements extend the separation time by 5 minutes.
Precondition:
R The display shows the dialog box
u Follow the instructions on the display.
ð Purging the instrument with air. See Chapter7.8 "Removing solvent from an
used cartridge", page78
ð Removing the cartridge. See Chapter6.4.3 "Installing and removing a cartridge",
page57
6.6.6Shutting down the instrument
Navigation path
➔ File
Precondition:
R The separation process has ended.
Sep arat ion End
Run com ple ted
.
.
u Purge the instrument with purging solvent. See Chapter7.6 "Cleaning the
instrument", page77
u Navigate to the [File] menu via the navigation path.
u Tap the action [Shut down].
u Confirm the secure question with [Yes].
ð The instrument is shutting down.
Operation Manual Pure Chromatography Instruments
65/102
Page 66
6 | Operation
Büchi Labortechnik AG
6.7Performing a Prep separation using a method
NOTE
Pure C-830 / C-835 / C-850 only
6.7.1Preparing the instrument
Time required:approx. 30 sec.
Precondition:
R All commissioning operations have been completed. See Chapter5 "Installation",
page24
u Switch the On/Off master switch to On.
u Tap the On/Off switch.
ð The instrument is starting up.
6.7.2Selecting the prep mode (Pure C-850 only)
Navigation path
➔ Mode
Mod e
u Navigate to the
u Select [Preparative].
menu according to the navigation path.
ð The prep mode is selected.
6.7.3Starting a Prep separation using a method
Precondition:
R The instrument is prepared. See Chapter6.7.1 "Preparing the instrument",
page66
R The required solvents are connected and assigned. See Chapter5.7 "Assigning
solvent to solvent lines", page26
R The sample is prepared.
R The waste bottle is empty.
u Calibrate the solvent level sensor. See Chapter5.8 "Calibrating the solvent level
sensor", page27
u Calibrate the waste level sensor. See Chapter5.12 "Calibrating the waste level
sensor", page29
u Open the protection shield.
u Place the fraction collection trays inside the instrument. See Chapter6.4.5
"Inserting the fraction collection tray", page60
u Close the protection shield.
u Open an existing method. See Chapter6.4.6 "Selecting an existing method",
page61
u Tap the button [OK].
66/102
u Tap the button [Start] on the
u According the requirements adjust the file name.
u Tap the button [OK].
ð The display shows the dialog box
Run
panel.
Sam ple Loa din g
Operation Manual Pure Chromatography Instruments
.
Page 67
Büchi Labortechnik AG
6.7.4Changings during a separation
6.7.5Ending a Prep separation
Operation | 6
u According the requirements adjust the settings. See Chapter6.4.7 "Selecting
values on the dialog box Sample Loading", page61
u Follow the instructions on the display.
ð Installing the column. See Chapter6.4.4 "Installing and removing a column",
page58
ð Introducing the sample into the system. See Chapter6.4.1 "Introducing a sample
into the prep system", page54
NOTE
Parameters that can be edited are highlighted in green
Possibilities to edit the gradient during a separation:
Change the gradient. See Chapter "Editing the gradient in graphic mode", page47
Hold the gradient. See Chapter6.1.6 "Gradient panel", page39
Auto gradient hold. See Chapter6.1.6 "Gradient panel", page39
NOTE
The separation extends automatically by 5 minutes if the baseline at the end of the
separation is not below the set threshold.
Precondition:
R The display shows the dialog box
u According to the requirements extend the separation time by 5 minutes.
6.7.6Shutting down the instrument
Navigation path
➔ File
Precondition:
R The separation process has ended.
u Removing the column. See Chapter6.4.4 "Installing and removing a column",
page58
u Purge the instrument with purging solvent. See Chapter7.6 "Cleaning the
instrument", page77
u Navigate to the [File] menu via the navigation path.
u Tap the action [Shut down].
u Confirm the secure question with [Yes].
ð The instrument is shutting down.
Sep arat ion End
.
6.8Performing a Prep separation manually
NOTE
Pure C-830 / C-835 / C-850 only
Operation Manual Pure Chromatography Instruments
67/102
Page 68
6 | Operation
6.8.1Preparing the instrument
Time required:approx. 30 sec.
Precondition:
R All commissioning operations have been completed. See Chapter5 "Installation",
page24
u Switch the On/Off master switch to On.
u Tap the On/Off switch.
ð The instrument is starting up.
6.8.2Selecting the prep mode (Pure C-850 only)
Navigation path
➔ Mode
Mod e
u Navigate to the
u Select [Preparative].
ð The prep mode is selected.
menu according to the navigation path.
Büchi Labortechnik AG
6.8.3Starting a separation
Precondition:
R The instrument is prepared. See Chapter6.8.1 "Preparing the instrument",
page68
R The required solvents are connected and assigned. See Chapter5.7 "Assigning
solvent to solvent lines", page26
R The sample is prepared.
R The waste bottle is empty.
u Calibrate the solvent level sensor. See Chapter5.8 "Calibrating the solvent level
sensor", page27
u Calibrate the waste level sensor. See Chapter5.12 "Calibrating the waste level
sensor", page29
u Open the protection shield.
u Place the fraction collection trays inside the instrument. See Chapter6.4.5
"Inserting the fraction collection tray", page60
u Close the protection shield.
u Install a column. See Chapter6.4.4 "Installing and removing a column", page58
u Tap the drop-down list next to A: on the
ð The display shows a drop-down list with the assigned solvents.
u Tap the required solvent.
ð The solvent is selected
ð The drop-down list closes.
u Select more solvents for the mobile phase according to your needs.
u Edit the gradient according to your needs. See Chapter6.2.4 "Editing the
gradient", page45
Sol vent Se lection
panel.
68/102
u Enter the required times in the
Con diti ons
Operation Manual Pure Chromatography Instruments
panel.
Page 69
Büchi Labortechnik AG
Operation | 6
u Select the sample collection in the
the fraction collection criteria", page49
u Select the collection criteria in the
u Tap the button [Start] on the
Run
ð The display shows the dialog box
ð The instrument starts the separation.
u According the requirements adjust the settings. See Chapter6.4.7 "Selecting
values on the dialog box Sample Loading", page61
u Follow the instructions on the display.
ð Installing the column. See Chapter6.4.4 "Installing and removing a column",
page58
ð Introducing the sample into the system. See Chapter6.4.1 "Introducing a sample
into the prep system", page54
6.8.4Changings during a separation
NOTE
Parameters that can be edited are highlighted in green
Possibilities to edit the gradient during a separation:
Col lect ion
panel. See Chapter6.2.7 "Editing
Col lect ion criter ia
panel.
Sam ple Loa din g
panel.
.
Change the gradient. See Chapter "Editing the gradient in graphic mode", page47
Hold the gradient. See Chapter6.1.6 "Gradient panel", page39
Auto gradient hold. See Chapter6.1.6 "Gradient panel", page39
6.8.5Ending a Prep separation
NOTE
The separation extends automatically by 5 minutes if the baseline at the end of the
separation is not below the set threshold.
Precondition:
R The display shows the dialog box
u According to the requirements extend the separation time by 5 minutes.
6.8.6Shutting down the instrument
Navigation path
➔ File
Precondition:
R The separation process has ended.
Sep arat ion End
.
u Removing the column. See Chapter6.4.4 "Installing and removing a column",
page58
u Purge the instrument with purging solvent. See Chapter7.6 "Cleaning the
instrument", page77
u Navigate to the [File] menu via the navigation path.
u Tap the action [Shut down].
Operation Manual Pure Chromatography Instruments
69/102
Page 70
6 | Operation
Büchi Labortechnik AG
u Confirm the secure question with [Yes].
ð The instrument is shutting down.
6.9Identifying fractions
6.9.1Identifying fractions by peak
Fig.9: Identifying fractions
Navigation path
➔ Gradient panel
Precondition:
R A separation is finished.
u Navigate to the
u Tap the peak with the target value.
ð The
Col lect ion
Gra dien t
panel according the navigation path.
panel shows the corresponding vial.
70/102
Operation Manual Pure Chromatography Instruments
Page 71
Büchi Labortechnik AG
6.9.2Identifying fractions per vial
Operation | 6
Fig.10: Identifying fractions
Navigation path
➔ Collection panel
Precondition:
R A separation is finished.
u Navigate to the
u Tap the target vial.
ð The
Gra dien t
Col lect ion
panel shows the corresponding peak.
panel according the navigation path.
6.10Importing and exporting data
6.10.1Printing a run report
Navigation path
➔ File ➔ Print Run Report
Precondition:
R The instrument is in past run mode.
u Navigate to the
Pri nt R un Report
ð The display shows the Windows® print dialog.
u Select your printer.
u Tap the button [OK].
ð The report is printed.
Operation Manual Pure Chromatography Instruments
dialog according to the navigation path.
71/102
Page 72
6 | Operation
6.10.2Sending data to USB
Navigation path
➔ Run panel
Precondition:
R The instrument is in past run mode.
u Connect a USB storage device to the instrument.
u Navigate to the
u Tap the button [Data to USB].
Run
ð The instrument saves an Excel file to the USB storage device.
u Confirm the complete message.
ð The data is stored.
6.10.3Sending PDF to USB
Navigation path
➔ Run panel
Büchi Labortechnik AG
panel according the navigation path.
Precondition:
R The instrument is in past run mode.
u Connect a USB storage device to the instrument.
Run
u Navigate to the
u Tap the button [PDF to USB].
panel according the navigation path.
ð The instrument saves a PDF file to the USB storage device.
u Confirm the complete message.
ð The data is saved.
72/102
Operation Manual Pure Chromatography Instruments
Page 73
Büchi Labortechnik AG
7Cleaning and servicing
7.1Regular maintenance work
Cleaning and servicing | 7
NOTE
Users may only carry out the servicing and cleaning operations described in this
section.
Any servicing and repair work which involves opening up the casing may only be
carried out by BUCHI service technicians.
u Use only genuine BUCHI consumables and spare parts in order to ensure correct
operation of the device and preserve the warranty.
ComponentActionFrequency
Pumps and valvesu Purge the instrument with air. See
Daily
Chapter7.8 "Removing solvent from an
used cartridge", page78
u Purge the instrument with purging
solvent. See Chapter7.6 "Cleaning the
instrument", page77
Tubingu Check for leaks. If a leak is observed,
Daily
resolve the issue before continuing.
Fittingsu Inspect fittings; if solid material is
Daily
deposited on a fitting, clean and
tighten (replace) the fitting before
continuing.
Drain linesu Check all drain lines to ensure that
Daily
liquid can flow through them to the
waste container
Waste bottleu Empty the waste bottleDaily
Filtersu Check the filters in the solvent bottles
Weekly
and clean if necessary.
Fittingsu Check and if necessary tighten the
Weekly
fitting that secures the tubing from the
mobile phase reservoir manifold to the
inlet check valve housing
Datau Perform a data backupWeekly
Casingu Wipe down the casing with a damp
u If heavily soiled, use ethanol or a mild
Warning symbolsu Check that the warning symbols on the
u If they are dirty, clean them.
Displayu Wipe down the display with a damp
Operation Manual Pure Chromatography Instruments
Weekly
cloth.
detergent.
Weekly
instrument are legible.
Monthly
cloth.
73/102
Page 74
7 | Cleaning and servicing
Büchi Labortechnik AG
ComponentActionFrequency
Nebulizeru Clean the nebulizer. See Chapter7.4
"Cleaning the nebulizer", page75
Air filteru Replace the air filter.Yearly
7.2Carrying out a data backup
Monthly
Navigation path
➔ File ➔ Exit
Exi t
u Navigate to the
u Confirm the secure question.
dialog according to the navigation path.
ð The Pure software is shutting down.
ð The display shows a Windows® system.
u Connect a USB storage device to the instrument.
u Open the Windows® Explorer.
u Navigate to the data you wish to backup. See Chapter10.2 "Folder locations",
page96
u Copy the needed data to the USB storage devices.
7.3Calibrating the display
Navigation path
➔ Tools ➔ Calibration and Defaults
u Navigate to the
path.
u Tap the button [Calibrate].
u Follow the instructions on the display.
Cal ibra tio n a nd D efa ults
dialog according to the navigation
74/102
Operation Manual Pure Chromatography Instruments
Page 75
Büchi Labortechnik AG
7.4Cleaning the nebulizer
Cleaning and servicing | 7
Materials Needed:
HPLC-grade 50:50 methanol: water solution
Sonication bath
u Switch the On/Off master switch to Off.
u Disconnect the liquid inlet line from the
nebulizer.
u Disconnect the gas inlet from the nebulizer
u Remove the knurled screws.
u Remove the nebulizer from the drift tube.
u Place the nebulizer in a beaker filled with 50:50
methanol: water solution.
u Sonicate the nebulizer in an ultrasonic bath for
10 minutes.
u If the nebulizer is still completely blocked,
connect a high-pressure air line to the nebulizer
inlet to help remove the blockage.
u If the nebulizer cannot be cleaned, replace the
nebulizer.
Operation Manual Pure Chromatography Instruments
75/102
Page 76
7 | Cleaning and servicing
Büchi Labortechnik AG
u Replace the nebulizer wear band.
u Place the nebulizer to the thrift tube.
u Attach the knurled screws.
u Connect the liquid and gas inlet lines to the
nebulizer.
7.5Cleaning the check valve
NOTE
Most check valve problems can be solved by pumping a strong solution of liquid
laboratory grade detergent through the check valves at a rate of 20 mL/min for one
hour.
Pumping detergent through the check valves
Materials needed:
Liquid Laboratory Detergent
76/102
Isopropanol/Water (50/50) or Methanol/Water (50/50)
Operation Manual Pure Chromatography Instruments
Page 77
Büchi Labortechnik AG
Cleaning and servicing | 7
Navigation path
➔ Tools ➔ Manual Control ➔ Column Flushing...
Precondition:
R A bypass is installed. See Chapter6.4.4 "Installing and removing a column",
page58
u Navigate to the
Col umn Flu shing
dialog according to the navigation path.
Sonicate the check valve
Materials needed:
Open-end wrench, ½”, 9/16" x 5/16"
Torque wrench
u Switch the On/Off master switch to Off.
u Disconnect the mobile phase tubing from the inlet (bottom) of the pump head
using the 9/16” open-end wrench.
u Disconnect the mobile phase outlet tubing from the top of the pump head using
the 5/16” open end wrench.
u Remove both check valve housings from the pump head using the ½” open-end
wrench.
ð The check valve capsule is accessible.
u Sonicating the check valve for 10 min in the appropriate solvent.
u Install the check valve.
u Install the check valve housings back into the pump head.
u Tighten the check valve housing to 75 inch pounds with the ½” torque wrench.
7.6Cleaning the instrument
Navigation path
➔ Tools ➔ Manual Control ➔ Column Flushing...
Precondition:
R A purging solvent is assigned to the instrument. See Chapter5.7 "Assigning
solvent to solvent lines", page26
u Install a bypass. See Chapter6.4.4 "Installing and removing a column", page58
Operation Manual Pure Chromatography Instruments
77/102
Page 78
7 | Cleaning and servicing
Büchi Labortechnik AG
u Navigate to the
u Enter the required data according to your needs.
Col umn Flu shing
dialog according to the navigation path.
7.7Cleaning the solid loader
Navigation path
➔ Tools ➔ Manual Control ➔ Solid Loader Flushing...
u Install a bypass. See Chapter6.4.4 "Installing and removing a column", page58
u Navigate to the
Sol id L oad er Flus hin g
dialog according to the navigation
path.
7.8Removing solvent from an used cartridge
Navigation path
➔ Tools ➔ Manual Control ➔ Air Purging
Precondition:
R The cartridge is installed.
u Navigate to the
u Enter the purging time according to your needs.
u Tap the button [Column purge].
Air Pur gin g
dialog according to the navigation path.
ð The instrument and the cartridges are cleaned.
78/102
Operation Manual Pure Chromatography Instruments
Page 79
Büchi Labortechnik AG
8Help with faults
8.1Faults, possible causes and solutions (general)
Help with faults | 8
MalfunctionPossible causeSolution
The instrument
does not power up
Power is not being supplied to
the system
u Verify that the power cord is
plugged in.
u Make sure that the voltage,
amperage and frequency
meet the instrument
specifications.
u Make sure that both power
switches are switched on.
u Verify that the fuse wire is
not broken and fuses are
correctly installed in the
instrument.
System shuts down
automatically
Major fluctuations in line
power are present
u Connect system to a
Uninterrupted Power Supply
line.
The touch screen is
not responsive
The touch screen is out of calibration
u Recalibrate the touch
screen.
8.2Faults, possible causes and solutions (cartridge)
MalfunctionPossible causeSolution
Cartridge is not detected
RFID tag is not facing the RFID
reader
u Turn cartridge so that RFID
tag faces RFID reader.
RFID tag is badu Use new cartridge.
8.3Faults, possible causes and solutions (solvent delivery)
MalfunctionPossible causeSolution
No solvent flowEmpty solvent bottleu Refill the solvent bottle.
Pump not primedu Prime the pump.
u Remove the check valve and
clean it by sonicating the
check valve in IPA.
u If sonication does not work
replace the check valve with
a new check valve.
Air bubbles in solvent lineu Prime the pump.
Pump seals worn outu Replace the pump seals.
Operation Manual Pure Chromatography Instruments
79/102
Page 80
8 | Help with faults
Büchi Labortechnik AG
MalfunctionPossible causeSolution
Pulsation of pumpOpen or close time of the inlet
or outlet valves are not correct
Residues in solvent
Sealing abrasion outlet valve
Inconsistent solvent
flow
Loose fitting/air leak into the
pump
Liquid leak/pump seals worn
out
Pump head temperature
reaches solvent boiling temperature, causing the pump to
lose prime and stop flow (this
is likely to occur when running
methods with highly volatile
solvents such as diethyl ether
and methylene chloride)
u Rinse the pump module with
high flowrate with ethanol
or hot distilled water.
u Change check valves.
u Find the loose fitting
between mobile phase
reservoir manifold and
pump inlet fitting and
tighten it up.
u Fix the leak/replace the
pump seals.
u Premix the solvents to
reduce solvent volatility.
u Place the highly volatile
solvent bottle in an ice bath
to eliminate boiling.
System pump pressure is higher than
expected
Blocked solvent linesu Find the blocked lines and
replace it.
Over-tightened fittingu Loosen the fitting or replace
it.
Blocked columns or fluidic
path
u Locate the component that
caused the blockage, repair,
or replace the component.
LeaksFitting connection not tightu Find the loose fitting and
tighten it up.
Damaged solvent lineu Find the damaged solvent
line and replace it.
Pump not runningPump sensor cable becomes
disconnected
u Locate the cable and
reconnect to the pump
sensor.
Pump power cable becomes
disconnected
u Locate the power cable and
reconnect to the main board
or to the pump.
Incorrect flow path Incorrect fluidic connections
to/from the mode switching
u Check/correct the fluidic
connections.
valve
80/102
Operation Manual Pure Chromatography Instruments
Page 81
Büchi Labortechnik AG
8.4Faults, possible causes and solutions (sample injection)
Help with faults | 8
MalfunctionPossible causeSolution
Leak around the injection port
(Flash mode)
Dried sample or particulate
matter interferes with syringe
fitting
Defective injection port (Luer
u Clean the injection port with
appropriate solvent or
remove particulate matter.
u Replace the injection port.
fitting) adapter
Solid loader leaking Loader hardware is not prop-
erly set up
Leak around the
Loose fittingu Find the loose fitting and
prep injection
u Verify that the loader
hardware is set up correctly.
tighten it up.
valve/sample loop
8.5Faults, possible causes and solutions (fraction collection)
MalfunctionPossible causeSolution
Liquid not being collected in fraction
tubes
Liquid not centered
in fraction tube
Incorrect fraction collection
settings
Fraction collector is not calibrated
u Verify that fraction
collection information is set
properly.
u Recalibrate the fraction
collector.
Fraction collector
(FC) arm does not
move
Fraction collector arm did not
home properly
Fraction collector arm motor is
u Reset the Fraction collector
arm.
u Tighten the motor coupler.
slipping
Fraction collector arm is obstructed
u Check for cable or burr in
the fraction collector arm
path and remove any
obstruction.
Tray not detectedRFID tag is badu Put another tray into the
same position to see if it is
recognized to confirm the
cause.
u Replace RFID tag.
Operation Manual Pure Chromatography Instruments
81/102
Page 82
8 | Help with faults
8.6Faults, possible causes and solutions (detection)
Büchi Labortechnik AG
MalfunctionPossible causeSolution
ELSD signal is low or
disappears
Sample is too volatileu Sample cannot be detected
by ELSD due to its volatility.
No or low ELSD carrier gas flow u Verify there is gas supplied
to the instrument.
u Check for leak in the gas
lines.
u Verify that there is gas flow
to the nebulizer from the
ELSD sampling valve.
u Verify that there is gas flow
to the ELSD sampling valve.
Sample stuck on columnu Use stronger solvent or
change column chemistry.
ELSD not equilibrated long
enough
u Restart run to re-stabilize
and rezero ELSD baseline.
Blocked nebulizeru Sonicate nebulizer to clean
or replace the nebulizer.
Blocked ELSD lineu Trace the blockage and
replace the blocked line.
Poor ELSD Peak
shape
Rotor and/or stator in ELSD
Sampling valve is worn, dirty,
or clogged
Blocked nebulizer or nebulizer
tubing
Nebulizer tubing is not properly connected
u Replace the rotor and/or
stator.
u Clean the nebulizer or
replace the nebulizer tubing.
u Reinstall the nebulizer
tubing properly.
82/102
Operation Manual Pure Chromatography Instruments
Page 83
Büchi Labortechnik AG
Help with faults | 8
MalfunctionPossible causeSolution
Noisy ELSD baseline Dirty or contaminated gasu Replace gas source.
u Replace filter.
Gas is not dry/high humidity
environment
Moisture trapped in the gas
lines
u Use a dry air supply.
u Use nitrogen.
u Remove moisture from lines
by purging the air flow
system with nitrogen for 5
minutes.
Mobile phase is contaminated
or contains non-volatile modi-
u Use volatile modifiers in the
mobile phase.
fiers
Solvent contains non volatile
stabilizers
u Use solvent with volatile
modifiers.
Dirty opticsu Clean the optics.
Dirty drift tubeu Clean drift tube.
Not correct installed exhaustu Install the exhaust correctly.
See Chapter5.9 "Installing
the exhaust (ELSD only)",
page28
Electronics - Preamp not
grounded properly
u Verify that preamp
grounding cable is in place.
No UV signalUV light burned outu Replace the UV light.
Low UV signalDirty flow cellu Clean the flow cell.
8.7Error messages
Error messagePossible causeSolution
Instrument Alarm:
Solvent pump: Communication to the
pump cannot be established.
Instrument Alarm:
UV detector: Lamp
ignition fails after
short douse in AUTOZERO function
and repeated High
Voltage impulse and
heater cycle fails
too.
Instrument Alarm:
UV detector: Bad
identification of
light peak caused by
low light intensity.
Instrument Alarm:
UV detector: Bad
identification of
light peak caused by
unworkable light intensity.
Service.
fan is not working or
any fan is disconnected or mechanical blocked.
Solvent List FileSolvent list file corruptedu Shutdown the Pure
software.
u Delete the solvent list.
u Restart the Pure software.
Solvent Lines File:
Invalid program execution.
Solvent Defaults
File: The solvent in-
Solvent lines file corruptedu Shutdown the Pure
software.
u Delete the solvent list.
u Restart the Pure software.
Software damagedu Uninstall and reinstall
software.
formation file and
the default solvent
information file are
missing.
Using Default Solvent List: System
defaults used instead of actual resource.
Solvent Pressure:
Pressure limit has
been exceeded.
Solvent file has been deletedu Restart the instrument.
Blockage in the lineu Remove column.
u Replace with bypass.
u Determine if column is not
the source of blockage.
u If not column is the source
of blockage, then check all
fluidic lines.
Sample crashing/precipitateu Purge lines/system with a
strong or appropriate
solvent that will dissolve the
sample
Valve stuck in incorrect position
u Reset valve position.
u Contact BUCHI Customer
Service.
86/102
Operation Manual Pure Chromatography Instruments
Page 87
Büchi Labortechnik AG
Help with faults | 8
Error messagePossible causeSolution
Solvent Pressure 2:
Pressure limit has
been exceeded.
Error in solvent line
calibration file.
Inlet Gas Pressure
Out of Range - LOW
PRESSURE
Blockage in line after the ELSD
sampling valve
u Check the fluidic lines and
components downstream
from the pressure sensor 2
(between the ELSD sampling
valve and the fraction
collector valve, waste line)
for the source of blockage.
Calibration file of solvent lines
damaged
u Shutdown the Pure
software.
u Delete the calibration list.
u Restart the Pure software.
No gas or gas flow is lowu Check the gas lines in the
system for leaks.
u Make sure gas source/tank
is available.
The pressure gauge is not set
correctly
u Check pressure gauge
setting to make sure that it
is set to deliver 2.5 L/min
gas flow (Inlet Pressure with
Air State off around 85 - 115
psi).
Inlet Gas Pressure
Out of Range - LOW
PRESSURE (Before
run starts)
Inlet gas pressure
out of range- HIGH
(During run)
Sensor pressure calibration file
Blockage in the systemu Check the gas lines in the
system for blockages.
The pressure gauge is not set
correctly
u Check pressure gauge
setting to make sure that it
is set to deliver 2.5 L/min
gas flow (Inlet Pressure with
Air State off around 85 - 115
psi).
Blockage in the nebulizer or
gas line to the nebulizer
u Check the nebulizer or
tubing to the nebulizer for
blockages
u Sonicate the nebulizer in a
suitable solvent or replace
nebulizer
Pressure sensor calibration file
damaged.
u Shutdown the Pure
software.
u Delete the pressure value
list.
u Restart the Pure software.
Operation Manual Pure Chromatography Instruments
87/102
Page 88
8 | Help with faults
Büchi Labortechnik AG
Error messagePossible causeSolution
Instrument Alarm:
Vapor
Vapor sensor no signal
Instrument Alarm:
Watchdog
Vapor sensor settings are set
too sensitive
u Change the vapor limits to a
lower sensitivity. See
Chapter6.2.9 "Editing the
vapor sensor sensitivity",
page52
Leaksu Check for any leaks in the
solvent flow paths.
Solvent vapors in work areau Solvent vapors in the work
area may set on the alarm.
Use the instrument in a
hood or well ventilated area
with no open solvents near
the system.
Vapor sensor is not connected
or is defect
Timing/communication issue
in software
u Check the cable to the vapor
sensor.
If in setup mode:
u Restart the instrument.
u
If in run mode:
u Press reset.
u Restart after the run is
finish.
Solvent PressureSolvent pressure limit has
been exceeded
u Remove column.
u Replace with bypass.
u Determine if column is not
the source of blockage.
u If not column is the source
of blockage, then check all
fluidic lines.
88/102
Operation Manual Pure Chromatography Instruments
Page 89
Büchi Labortechnik AG
Help with faults | 8
Error messagePossible causeSolution
Fraction CollectorArm is obstructedu Check arm path to make
sure there are no
obstructions.
u Reset the Fraction Collector
arm. See Chapter8.8
"Resetting the Fraction
collector arm", page89
Mode Switching
Valve Failure
Prep Injection Valve
Failure
Fraction collector arm did not
home properly
u Tighten the motor coupler.
u Contact BUCHI Customer
Service.
Fraction collector arm motor is
slipping
u Contact BUCHI Customer
Service.
Home position sensor is badu Contact BUCHI Customer
Service.
Fraction collector arm motor
has no power
The Mode Switching valve
failed to switch.
u Contact BUCHI Customer
Service.
u Press Reset button to stop
the alarm.
u Contact BUCHI Customer
Service.
Prep Injection Valve failed to
switch.
Only Flash mode runs can be
performed.
u Press Reset button to stop
the alarm.
u Contact BUCHI Customer
Service.
Low Disk SpaceThe disk space on this unit is
almost full.
Solvent Safety Sensor alert
More solvent requiredu Add solvent to bottle.
Solvent safety sensor was not
calibrated
Waste Safety Sensor
alert
Waste container is fullu Empty waste container.
Waste safety sensor was not
calibrated
8.8Resetting the Fraction collector arm
Navigation path
➔ Tools ➔ Manual Control ➔ Fraction Collector Arm Reset
u Navigate to the
navigation path.
Fra ctio n C ollecto r A rm R ese t
u Back up and delete data.
u Contact BUCHI Customer
Service.
u Calibrate/zero the solvent
safety sensor properly.
u Calibrate/zero the solvent
safety sensor properly.
dialog according to the
Operation Manual Pure Chromatography Instruments
89/102
Page 90
8 | Help with faults
8.9Replacing the shuttle valve rotor
Büchi Labortechnik AG
u Confirm the secure question.
ð The Fraction collector arm is resetted.
Materials needed:
Torx wrench T10
Pure sampling valve kit
u Switch the On/Off master switch to Off.
Precondition:
R Make sure that all tubings are marked for later
installation.
u Remove all tubings.
u Loosen the screws with the Torx wrench.
u Remove the valve head.
u Remove the valve rotor.
90/102
Operation Manual Pure Chromatography Instruments
Page 91
Büchi Labortechnik AG
Help with faults | 8
u Make sure that the holes in the new rotor and
the pins inside the valve match.
u Attache the valve head to the instrument with
screws.
u Attach all tubings to the instrument.
Operation Manual Pure Chromatography Instruments
91/102
Page 92
9 | Taking out of service and disposal
9Taking out of service and disposal
9.1Taking out of service
u Switch off the instrument and disconnect it from the mains power supply.
u Remove all tubing and communication cables from the device.
9.2Disposal
The operator is responsible for proper disposal of the instrument.
u When disposing of equipment observe the local regulations and statutory
requirements regarding waste disposal.
u When disposing, observe the disposal regulations of the materials used. Materials
used see Chapter3.5 "Technical data", page18
9.3Returning the instrument
Before returning the instrument, contact the BÜCHI Labortechnik AG Service
Department.
https://www.buchi.com/contact
Büchi Labortechnik AG
92/102
Operation Manual Pure Chromatography Instruments
Page 93
Büchi Labortechnik AG
10Appendix
10.1Spare parts and accessories
10.1.1Accessories
Appendix | 10
Use only genuine BUCHI consumables and spare parts in order to ensure correct,
safe and reliable operation of the system.
NOTE
Any modifications of spare parts or assemblies are only allowed with the prior
written permission of BUCHI.
Order no.Image
Pure rack type 1
Rack for 12 x 75 mm glass tubes, 1 pc.
Pure rack type 2
Rack for 13 x 100 mm glass tubes, 1 pc.
Pure rack type 3
Rack for 16 x 125 mm glass tubes, 1 pc.
Pure rack type 4
Rack for 16 x 150 mm glass tubes, 1 pc.
Pure rack type 5
Rack for 18 x 150 mm glass tubes, 1 pc.
Pure rack type 6
Rack for 25 x 150 mm glass tubes, 1 pc.
11066672
11066673
11066674
11066675
11066676
11066677
Pure rack type 7
Rack for 9 squared bottles of 480 mL, 1 pc.
Pure rack type 8 (funnel rack)
Rack for 6 funnels
Pure rack type 9
Rack for 16 x 100 mm glass tubes, 1 pc.
Operation Manual Pure Chromatography Instruments
11068452
11069407
11069242
93/102
Page 94
10 | Appendix
Büchi Labortechnik AG
Order no.Image
Solvent bottle platform kit
Solvent bottle platform on top of the instrument. For maximum 4 bottles (volume 4 L each)
Retaining container
Retaining container for solvent bottle platform for more
safety regarding leaking
11069285
11068468
Pure Dry Air Supply unit11069026
Loading Pump
with flow rates from 2.5 to 250 mL/min, incl. tubing and fitting
Pure cartridge holder XXL
Cartridge holder for cartridges of 800 g to 5000 g
Pure column holder XL
Column holder for column diameters 50 to 70 mm
11071418
11070532
11068467
Pure nebulizer set11069464
10.1.2Sample introduction accessories
Pure Solid loader S set, incl. adapter set, sleeve, tubes (20 pcs.) and
frits (40 pcs.)
Pure Solid loader M set, incl. adapter set, sleeve, tubes (20 pcs.) and
frits (40 pcs.)
Run files.gkfrC:\Users\Public\Documents\Buchi\Pure\runs
Operation Manual Pure Chromatography Instruments
Page 97
Büchi Labortechnik AG
10.3Finding out separation conditions with the navigator
10.3.1Open the navigator
Appendix | 10
Navigation path
➔ Gradient panel
u Navigate to the
u Tap the button [Navigator].
ð The display shows the dialog
10.3.2TLC Silica
Gra dien t
panel according the navigation path.
Nav igat or
.
The TLC-Silica tool uses TLC separation data to provide a recommended gradient for
silica separations of 2 or 3 components.
Specification of the samples needed:
Two different solvent concentrations
Two separations
Silica TLC plates
Precondition:
R The display shows the dialog
navigator", page97
u Select the
u Select the radio button for two or three components.
u Select the weaker solvent at drop-down list A.
u Select the stronger solvent at drop-down list B.
u Enter the solvent B concentrations used in the TLC separations.
Operation Manual Pure Chromatography Instruments
TLC - Si lic a
tab.
Nav igat or
. See Chapter10.3.1 "Open the
97/102
Page 98
10 | Appendix
Büchi Labortechnik AG
u Measure the distance the spot moved on the silica TLC plates.
u Divide the measured value by the distance the solvent traveled.
ð This is your Rf- value.
u Enter the value in the Rf entry field.
u Select the column you wish to use from the drop-down list
ð The column is selected.
ð The display shows the default flow rate.
u Adjust the flow rate according to your needs.
u Select the radio button for Speed or Purity.
u Tap the button [Calculate].
ð The display shows the results.
u Adjust the following values if necessary:
Column
Flow rate
Speed / Purity
u Tap the button [Accept].
ð The gradient is saved for a run.
ð The dialog box closes.
10.3.3LC-C18
98/102
The LC-C18 Tool uses isocratic HPLC separation data to provide a recommended
gradient for reversed phase separations of 2 components.
Operation Manual Pure Chromatography Instruments
Page 99
Büchi Labortechnik AG
Appendix | 10
Specification of the samples needed:
Two isocratic runs on a HPLC column
Different mobile phase solvent concentrations
Retention times (t1 and t2) of the components from the chromatograms.
Precondition:
R The display shows the dialog
Nav igat or
. See Chapter10.3.1 "Open the
navigator", page97
u Select the
u Select the HPLC column used from the drop-down list.
u Select the weaker solvent at drop-down list A.
u Select the stronger solvent at drop-down list B.
u Enter the solvent concentrations used in the HPLC separations.
u Enter the retention times for each separation under each chromatograph.
u Select the column you wish to use from the drop-down list
LC- C18
tab.
ð The column is selected.
ð The display shows the default flow rate.
u Adjust the flow rate according to your needs.
u Select the radio button for Speed or Purity.
u Tap the button [Calculate].
ð The display shows the results.
u Tap the button [Accept].
ð The gradient is saved for a run.
ð The dialog box closes.
10.3.4LC-Transfer
Operation Manual Pure Chromatography Instruments
99/102
Page 100
10 | Appendix
Büchi Labortechnik AG
The LC-Transfer tool converts an HPLC gradient into a Flash Chromatography
gradient.
Specification of the sample needed:
One run on a HPLC column (in gradient elution mode)
Times %B changes (t1, t2 and t3)
Precondition:
R The display shows the dialog
Nav igat or
. See Chapter10.3.1 "Open the
navigator", page97
u Select the LC-Transfer tab.
u Select the HPLC column type used from the drop-down list.
u Enter the flow rate used in the HPLC separation.
u Enter the following HPLC gradient conditions:
lower and higher %B
times t1, t2 and t3
u Select the cartridge you wish to use for the Flash separation from the drop-down
list.
ð The cartridge is selected.
ð The display shows the default flow rate.
u Adjust the flow rate according to your needs.
u Tap the button [Calculate].
ð The display shows the results.
u Adjust the following values if necessary:
Flow rate
u Tap the button [Accept].
ð The gradient is saved for a run.
ð The dialog box closes.
100/102
Operation Manual Pure Chromatography Instruments
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.