Buchi ProxiMate Operation Manual

Page 1
ProxiMate™ Operation Manual
11593981 | D en
Page 2
Imprint
Product Identification: Operation Manual (Original) ProxiMate™ 11593981
Publication date: 09.2020
BÜCHI Labortechnik AG Meierseggstrasse 40 Postfach CH-9230 Flawil 1
E-Mail: quality@buchi.com
BUCHI reserves the right to make changes to the manual as deemed necessary in the light of experi­ence, especially with respect to structure, illustrations and technical details.
This manual is copyrighted. Information from it may neither be reproduced, distributed, or used for com­petitive purposes, nor made available to third parties. The manufacture of any component with the aid of this manual without prior written agreement is also prohibited.
Page 3
BÜCHI Labortechnik AG Table of contents
Table of contents
1 About this document........................................................................................................... 7
1.1 Warning notices in this document........................................................................................................ 7
1.2 Symbols............................................................................................................................................... 7
1.2.1 Warning symbols .....................................................................................................................7
1.2.2 Mark-ups and symbols.............................................................................................................8
1.3 Trademarks.......................................................................................................................................... 8
2 Safety....................................................................................................................................9
2.1 Proper use ........................................................................................................................................... 9
2.2 Use other than intended ...................................................................................................................... 9
2.3 Staff qualification ................................................................................................................................. 9
2.4 Residual risks .................................................................................................................................... 10
2.4.1 Glass and acrylic breakage....................................................................................................10
2.4.2 Faults during operation .......................................................................................................... 10
2.4.3 Instrument damage due to an incorrectly installed IP module ............................................... 10
2.4.4 Malware infection due to connections with other devices or network .................................... 10
2.4.5 Data loss ................................................................................................................................10
2.4.6 Damage to the internal memory due to incorrect shutting down of the instrument................10
2.5 Personal protective equipment .......................................................................................................... 10
2.6 Modifications...................................................................................................................................... 11
3 Product description...........................................................................................................12
3.1 Description of function ....................................................................................................................... 12
3.2 Sample presentation options ............................................................................................................. 12
3.2.1 Up view option .......................................................................................................................12
3.2.2 Down view option...................................................................................................................12
3.3 Measurement modes......................................................................................................................... 13
3.3.1 Diffuse reflection mode .......................................................................................................... 13
3.3.2 Transflectance mode .............................................................................................................13
3.4 Configuration ..................................................................................................................................... 14
3.4.1 Front view ..............................................................................................................................14
3.4.2 Rear view ...............................................................................................................................15
3.4.3 Connections (IP module removed) ........................................................................................16
3.4.4 Location of the type plate.......................................................................................................16
3.5 Scope of delivery ............................................................................................................................... 17
3.6 Type plate.......................................................................................................................................... 17
3.7 Technical data ................................................................................................................................... 17
3.7.1 ProxiMate™ ........................................................................................................................... 17
3.7.2 Ambient conditions.................................................................................................................18
3.7.3 Materials ................................................................................................................................18
4 Transport and storage ......................................................................................................20
4.1 Transport ........................................................................................................................................... 20
4.2 Storage .............................................................................................................................................. 20
4.3 Lifting the instrument ......................................................................................................................... 20
Operation Manual ProxiMate™ iii
Page 4
Table of contents BÜCHI Labortechnik AG
5 Installation..........................................................................................................................21
5.1 Before installation .............................................................................................................................. 21
5.2 Installation site................................................................................................................................... 21
5.3 Securing against earthquakes ........................................................................................................... 21
5.4 Establishing electrical connections.................................................................................................... 22
5.5 Installing a USB device...................................................................................................................... 22
5.6 Software settings ............................................................................................................................... 23
5.6.1 Apply for a licence..................................................................................................................23
5.6.2 Import Licence .......................................................................................................................24
5.6.3 Calibrating a Baseline Correction Vector (BCV) .................................................................... 24
5.6.4 Import external reference data...............................................................................................25
5.6.5 Changing the language of the instrument..............................................................................26
iv Operation Manual ProxiMate™
Page 5
BÜCHI Labortechnik AG Table of contents
6 Operation............................................................................................................................27
6.1 Layout of the control panel ................................................................................................................ 27
6.2 Function bar....................................................................................................................................... 27
6.3 Menu bar............................................................................................................................................ 28
6.3.1 Start menu .............................................................................................................................30
6.3.2 History menu..........................................................................................................................32
6.3.3 Application menu ...................................................................................................................35
6.3.4 Tools menu ............................................................................................................................ 36
6.4 Status bar .......................................................................................................................................... 38
6.5 Log in administrator mode ................................................................................................................. 39
6.6 Editing an application......................................................................................................................... 40
6.6.1 Creating a new application.....................................................................................................40
6.6.2 Changing the name of an application ....................................................................................41
6.6.3 Changing the alias of an application......................................................................................41
6.6.4 Changing the Measurement view of an application ............................................................... 42
6.6.5 Changing the Measurement mode of an application .............................................................43
6.6.6 Changing the Sample presentation of an application ............................................................43
6.6.7 Enter a description for an application.....................................................................................44
6.6.8 Changing the Rotation of an application................................................................................44
6.6.9 Changing the Measurement duration of an application .........................................................45
6.6.10 Enter a Standard Operating Procedure (SOP) for an application..........................................45
6.6.11 Changing the selection ability of an application.....................................................................46
6.7 Deleting an application ...................................................................................................................... 47
6.8 Editing a property............................................................................................................................... 47
6.8.1 Creating a new Property ........................................................................................................ 47
6.8.2 Changing the name of a property ..........................................................................................48
6.8.3 Changing the alias of a property............................................................................................48
6.8.4 Changing the Sort order of a property ...................................................................................49
6.8.5 Changing the Prediction type of a property............................................................................49
6.8.6 Changing the Initial Wavelength Range (Calibration model only).......................................... 53
6.8.7 Changing the Postpredicted Value (Calibration model only) .................................................54
6.8.8 Changing the Calibration Base (Calibration model only) ....................................................... 55
6.8.9 Changing the Display Basis (Calibration model only)............................................................55
6.8.10 Changing the Formula (Calculated Property only).................................................................56
6.8.11 Changing the Observer (Color only) ...................................................................................... 57
6.8.12 Changing the Illuminant (Color only)......................................................................................57
6.8.13 Changing the Metric (Color only) ........................................................................................... 58
6.8.14 Changing the Metric Element (Color only).............................................................................59
6.8.15 Changing the Decimal places of a property...........................................................................59
6.8.16 Changing the Unit of a property.............................................................................................60
6.8.17 Changing the Bias of a property ............................................................................................60
6.8.18 Changing the Slope of a property .......................................................................................... 61
6.8.19 Changing the Mahalanobis of a property (Calibration model only)........................................61
6.8.20 Changing the Target of a property.........................................................................................62
6.8.21 Changing the Limit maximum of a property ........................................................................... 62
6.8.22 Changing the Limit minimum of a property ............................................................................ 63
6.8.23 Changing the Tolerance maximum of a property...................................................................63
6.8.24 Changing the Tolerance minimum of a property....................................................................64
6.9 Deleting a property ............................................................................................................................ 65
6.10 Adding measurement data to a calibration ........................................................................................ 65
6.10.1 Performing a calibration update with AutoCal........................................................................65
6.10.2 Setting a reference value to a property..................................................................................66
6.10.3 Activating a measurement for AutoCal ..................................................................................66
6.11 Exporting application data ................................................................................................................. 67
6.12 Importing application data.................................................................................................................. 67
6.13 Creating a report................................................................................................................................ 68
Operation Manual ProxiMate™ v
Page 6
Table of contents BÜCHI Labortechnik AG
6.14 Exporting measurement data............................................................................................................. 68
6.15 Carrying out a measurement ............................................................................................................. 69
6.15.1 Preparing the instrument........................................................................................................69
6.15.2 Starting measurement............................................................................................................69
6.15.3 Ending measurement.............................................................................................................70
6.15.4 Shutting down the instrument ................................................................................................70
7 Cleaning and servicing .....................................................................................................71
7.1 Regular maintenance work................................................................................................................ 71
7.2 Changing the lamps........................................................................................................................... 72
7.2.1 Changing the down view lamp...............................................................................................72
7.2.2 Changing the up view lamp....................................................................................................76
7.2.3 Confirm Lamp Replacement .................................................................................................. 80
7.3 Checking the desiccant cartridge....................................................................................................... 80
7.4 Changing the desiccant cartridge ...................................................................................................... 82
7.5 Carrying out system tests .................................................................................................................. 83
7.5.1 Carrying out a Baseline Correction Vector test......................................................................83
7.5.2 Carrying out a Comprehensive System Test ......................................................................... 84
7.5.3 Carrying out an Advanced System Test ................................................................................84
7.6 Carrying out a data backup................................................................................................................ 84
7.7 Cleaning the display .......................................................................................................................... 85
8 Help with faults ..................................................................................................................86
8.1 Troubleshooting................................................................................................................................. 86
8.2 Error messages ................................................................................................................................. 86
9 Taking out of service and disposal.................................................................................. 88
9.1 Taking out of service.......................................................................................................................... 88
9.2 Disposal............................................................................................................................................. 88
9.3 Returning the instrument ................................................................................................................... 88
10 Appendix ............................................................................................................................89
10.1 Measurement results ......................................................................................................................... 89
10.2 Folder locations ................................................................................................................................. 89
10.3 Rules entering a formula.................................................................................................................... 90
10.4 Spare parts and accessories ............................................................................................................. 91
10.4.1 Accessories............................................................................................................................91
10.4.2 Spare parts ............................................................................................................................92
vi Operation Manual ProxiMate™
Page 7
BÜCHI Labortechnik AG About this document | 1
1 About this document
This operation manual is applicable for all variants of the instrument.
Read this operation manual before operating the instrument and follow the instruc­tions to ensure safe and trouble-free operation.
Keep this operation manual for later use and pass it on to any subsequent user or owner.
BÜCHI Labortechnik AG accepts no liability for damage, faults and malfunctions re­sulting from not following this operation manual.
If you have any questions after reading this operation manual:
u Contact BÜCHI Labortechnik AG Customer Service.
https://www.buchi.com/contact
1.1 Warning notices in this document
Warning notices warn you of dangers that can occur when handling the device. There are four danger levels, each identifiable by the signal word used.
Signal word Meaning
DANGER Indicates a danger with a high level of risk which could result in
death or serious injury if not prevented.
WARNING Indicates a danger with a medium level of risk which could result in
CAUTION Indicates a danger with a low level of risk which could result in mi-
NOTICE Indicates a danger that could result in damage to property.
1.2 Symbols
The following symbols are displayed in this operation manual or on the device:
1.2.1 Warning symbols
Symbol Meaning
death or serious injury if not prevented.
nor or medium-severity injury if not prevented.
General warning
Breakable items
Hot surface
Dangerous electrical voltage
Device damage
Operation Manual ProxiMate™ 7/94
Page 8
1 | About this document BÜCHI Labortechnik AG
1.2.2 Mark-ups and symbols
NOTE
This symbol draws attention to useful and important information.
R This character draws attention to a requirement that must be met before the in-
structions below are carried out.
u This character indicates an instruction that must be carried out by the user.
ð This character indicates the result of a correctly carried out instruction.
Mark-up Explanation
Win dow Software Windows are marked-up like this.
Tab Tabs are marked-up like this.
Dia log Dialogs are marked-up like this.
[Button] Buttons are marked-up like this.
[Field names] Field names are marked-up like this.
[Menu / Menu item] Menus or menu items are marked-up like this.
Status Status is marked-up like this.
Signal Signals are marked-up like this.
1.3 Trademarks
Product names and registered or unregistered trademarks that are used in this docu­ment are used only for identification and remain the property of the owner in each case.
8/94 Operation Manual ProxiMate™
Page 9
BÜCHI Labortechnik AG Safety | 2
2 Safety
2.1 Proper use
The instrument is designed and built for laboratories and production environments (at­line). It serves to determine the concentration of selected constituents contained within a substance.
The instrument can be used for the following tasks: Determination of quantifiable product properties.
2.2 Use other than intended
Use of any kind other than that described in Chapter2.1 "Proper use", page9 and any application that does not comply with the technical specifications (see Chap­ter3.7 "Technical data", page17) constitutes use other than that intended.
In particular, the following applications are not permissible:
Use of the instrument in rooms which require Ex-protected instruments.Use of samples, which can explode or inflame (example: explosives, etc.) due to
shock, friction, heat or spark formation.
2.3 Staff qualification
Unqualified persons are unable to identify risks and are therefore exposed to greater dangers.
The device may only be operated by suitably qualified laboratory staff.
These operating instructions are aimed at the following target groups:
Users
Users are persons that meet the following criteria:
They have been instructed in the use of the device.They are familiar with the contents of these operating instructions and the applica-
ble safety regulations and apply them.
They are able on the basis of their training or professional experience to assess
the risks associated with the use of the device.
Operator
The operator (generally the laboratory manager) is responsible for the following as­pects:
The device must be correctly installed, commissioned, operated and serviced.Only suitably qualified staff may be assigned the task of performing the operations
described in these operating instructions.
The staff must comply with the local applicable requirements and regulations for
safe and hazard-conscious working practices.
Safety-related incidents that occur while using the device should be reported to the
manufacturer (quality@buchi.com).
BUCHI service technicians
Service technicians authorized by BUCHI have attended special training courses and are authorized by BÜCHI Labortechnik AG to carry out special servicing and repair measures.
Operation Manual ProxiMate™ 9/94
Page 10
2 | Safety BÜCHI Labortechnik AG
2.4 Residual risks
The device has been developed and manufactured using the latest technological ad­vances. Nevertheless, risks to persons, property or the environment can arise if the device is used incorrectly.
Appropriate warnings in this manual serve to alert the user to these residual dangers.
2.4.1 Glass and acrylic breakage
Broken glass and acrylic can cause severe cuts.
Broken glass or acrylic can enter production.
u Handle the Petri Dishes and other glass and acrylic components carefully and do
not drop them.
u Always visually inspect glass and acrylic components for damage every time they
are to be used.
u Do not continue to use glass and acrylic components that are damaged. u Always wear protective gloves when disposing of broken glass and acrylic.
2.4.2 Faults during operation
If a device is damaged, sharp edges or exposed electrical wires can cause injuries.
u Regularly check device for visible damage. u If faults occur, switch off the device immediately and inform the operator. u Do not continue to use devices that are damaged.
2.4.3 Instrument damage due to an incorrectly installed IP module
An incorrectly installed IP module can cause material and moisture to enter the instru­ment.
u Do not open the IP module. u Make sure that the USB sockets are covered when not in use.
2.4.4 Malware infection due to connections with other devices or network
Connections with other devices or a network can cause a malware infection to the in­strument.
u Install antivirus software and firewall on the instrument before connecting it to
other devices or network.
2.4.5 Data loss
In the event of a power failure, e.g. due to lightning or interruption of power supply, measurement data may be lost.
u Carry out regular data backup.
2.4.6 Damage to the internal memory due to incorrect shutting down of the instrument
Incorrect shutting down of the instrument can cause damage to the internal memory.
u Shut down the instrument as described. See Chapter6 "Operation", page27
2.5 Personal protective equipment
Depending on the application, hazards due to heat and/or corrosive chemicals may arise.
u Always wear appropriate personal protective equipment such as safety goggles,
protective clothing and gloves.
u Make sure that the personal protective equipment meets the requirements of the
safety data sheets for all chemicals used.
10/94 Operation Manual ProxiMate™
Page 11
BÜCHI Labortechnik AG Safety | 2
2.6 Modifications
Unauthorized modifications may impair safety and lead to accidents.
u Use only genuine BUCHI accessories, spare parts and consumables. u Technical modifications to the device or accessories should only be carried out
with the prior written approval of BÜCHI Labortechnik AG and only by authorized BUCHI technicians.
BUCHI accepts no liability whatsoever for damage arising as a result of unauthorized modifications.
Operation Manual ProxiMate™ 11/94
Page 12
3 | Product description BÜCHI Labortechnik AG
3 Product description
3.1 Description of function
ProxiMate™ is a NIR spectrometer that can be used to determine the concentration of different parameters in food and feed samples in a nondestructive way.
ProxiMate™ is supplied in different versions. Dependent on the version specified ProxiMate™ is either an NIR or combined NIR and visible spectrometer.
The instrument generates a beam of NIR and visible light which is focused onto the sample under investigation. Light reflected from the sample is collected and spatially separated by a diffraction element. The diffracted light is directed onto a diode array detector. Signals from the detector are processed and a reflectance spectrum is con­structed. This spectrum undergoes further processing to calculate the constituents re­quired.
Data processing
The NIR light interacts with the sample material in different ways, leaving a character­istic fingerprint on the spectrum. Spectra from both liquids and solids can be mea­sured with ProxiMate™. The spectra of solid samples are collected directly, liquid samples require the use of a transflectance adapter.
Application
The Application defines all of the parameters related to measurement of a particular sample type.
This includes:
the properties to be measuredthe calibrations usedthe standard operating procedure
It is possible to Import or Export a file that contains all Application data to allow the same Application to be used on a second ProxiMate™ (dependent on calibration li­cense requirements).
3.2 Sample presentation options
The choice of sample presentation of ProxiMate™ is optimised for the type of sample under analysis and for the working environment where the instrument is used.
ProxiMate™ can be configured with a choice of sample presentation options: up view and down view configurations.
3.2.1 Up view option
The up view option directs and collects light from the underside of the sample. The NIR light passes through the base of a glass petri dish before interacting with the sample under evaluation. Up view measurement has the advantage that a more con­sistent surface is presented to the ProxiMate™ ensuring accurate measurement out­put. Glass petri dishes are recommended to enable best performance. Additionally, when used in conjunction with a transflectance adapter, it is also possible to measure liquids using the up view option.
3.2.2 Down view option
With down view option light is focussed onto and collected from the top surface of the sample. In areas where glass is prohibited (such as some food production areas), the down view mode offers the advantage that NIR light does not interact with the sample container. As plastics have their own NIR spectra, changes in type of dish can influ-
12/94 Operation Manual ProxiMate™
Page 13
BÜCHI Labortechnik AG Product description | 3
1
3
2
2
1
4
3
ence the measurement output leading to perceived measurement shifts. Use of the down view mode prevents this from occurring. Additionally ProxiMate™ down view also allows the use of large volume sample dishes. These are particularly useful with samples that are inhomogeneous, as the measurement is averaged over a much larger sample area.
3.3 Measurement modes
3.3.1 Diffuse reflection mode
Fig.1: Reflection mode (up view geometry)
1 Sample 2 Sensor 3 Light
Non-translucent materials can be analyzed via diffuse reflection.
NIR light penetration is limited by the sample material. It interacts with the sample, is refracted and diffusely reflected into the sensor. The reflected rays contain the spec­tral information of the sample.
3.3.2 Transflectance mode
Fig.2: Transflectance mode (up view mode)
1 Transflectance
2 Petri dish 3 Sensor 4 Light
cover
Operation Manual ProxiMate™ 13/94
Page 14
3 | Product description BÜCHI Labortechnik AG
5
4
1
2
3
Translucent and opaque liquids can be analyzed via transflectance mode. The light penetrates the liquid, is diffusely reflected by the reference plate and passes through the sample a second time. The transflected rays contain the spectral information of the sample.
3.4 Configuration
3.4.1 Front view
Fig.3: Front view
1 Control panel 2 On/Off master switch 3 Up view window 4 Sample presentation area 5 Down view window
NOTE
The instrument is designed for stationary installation and therefore not equipped with a mains plug.
The On/Off master switch does not interrupt the electric power supply.
4 See Chapter5.4 "Establishing electrical connections", page22
Status On/Off master switch
Status Description
No light The instrument is not switched on.
Steady light The instrument is on
Flashing light The instrument shuts down
14/94 Operation Manual ProxiMate™
Page 15
BÜCHI Labortechnik AG Product description | 3
8
7
1
2
3
4
5
6
3.4.2 Rear view
Fig.4: Rear view (with Advanced IP module)
1 Cover lid down view lamp 2 Cover lid desiccant cartridge 3 USB socket 4 USB socket 5 Cable gland for network connection 6 Advanced IP module
The instrument connections are lo­cated behind the Advanced IP mod­ule. See Chapter3.4.3 "Connections (IP module removed)", page16
7 Cable gland for mains lead 8 Cooler
Operation Manual ProxiMate™ 15/94
Page 16
3 | Product description BÜCHI Labortechnik AG
6
1
2
3
4
5
7
8
9
1
3.4.3 Connections (IP module removed)
Fig.5: Connections on the rear side
1 Reserve 2 Reserve 3 RJ45 network socket 4 RJ45 network socket 5 Reserve 6 Power supply connection 7 Reserve 8 USB ports
(max. 5 W)
9 DVI port
3.4.4 Location of the type plate
Fig.6: Location of the type plate
1 Type plate
16/94 Operation Manual ProxiMate™
Page 17
BÜCHI Labortechnik AG Product description | 3
Buchi Labortechnik AG 9230 Flawil / Switzerland
Type: xxx SN: 10xxxxxxxx Volt: xxx-xxx VAC Frequ.: xx Hz Power: xxxx W Built: 20xx
Made in Switzerland
6
5
4
1
3
2
7
8
9
10
9
3.5 Scope of delivery
NOTE
The scope of delivery depends of the configuration of the purchase order.
Accessories are delivered as per the purchase order, order confirmation, and deliv­ery note.
3.6 Type plate
The type plate identifies the instrument. The type plate is located at the side of the in­strument. See Chapter3.4.4 "Location of the type plate", page16
Fig.7: Type Plate
1 Company name and address 2 Instrument name 3 Serial number 4 Input voltage range 5 Frequency 6 Power consumption maximum 7 Year of manufacture 8 Product code 9 Approvals 10 Symbol for "Do not dispose of as
3.7 Technical data
3.7.1 ProxiMate™
Specification ProxiMate™
Dimensions (W x D x H) 260 x 435 x 500 mm
Weight 23 kg
Power consumption 60 W
Frequency 50 / 60 Hz
Connection voltage 100 - 240 VAC ± 10 %
Max. power for all USB-Ports 5 W
IP Code IP69
Overvoltage category II
household waste"
Pollution degree 2
Appliance classes I
Operation Manual ProxiMate™ 17/94
Page 18
3 | Product description BÜCHI Labortechnik AG
Specification ProxiMate™
Detector NIR Thermoelectrically cooled In-
GaAs
Detector VIS Si
Wavelength range NIR 900 - 1700 nm
Resolution NIR 7.0 nm
NIR Data Resolution 3.1 nm
Wavelength range VIS 400 - 900 nm
Resolution VIS 10 nm
VIS Data Resolution 2 nm
Up view illumination spot size 8 mm
Down view illumination spot size 30 mm
Approval CE / CSA
Lamp type Tungsten-halogen
Average life (lamp)
Display 10.4 in
3.7.2 Ambient conditions
For indoor use only.
Max. altitude above sea level 2000 m
Ambient temperature 5 - 40 °C (25 °C)
Maximum relative humidity 80% for temperatures up to 31
Storage temperature max. 45 °C
3.7.3 Materials
Component Material of construction
Housing Steel 1.4301
Seals housing EPDM 50
Housing Aluminum with coating
9000 h
°C decreasing linearly to 50 % rela-
tive humidity at 40 °C
Seals housing EPDM 70
Glass up view Sapphire Al2O3
Glass down view Borofloat
Seals glasses EPDM A 75
Cooler Aluminum with coating
EPOFLON 526/4562
Sealing frame cooler HD-PE
Seals cooler EPDM
Display Glass
Frame display Steel 1.4301
18/94 Operation Manual ProxiMate™
Page 19
BÜCHI Labortechnik AG Product description | 3
Component Material of construction
Seal Display 1K MS-Polymer Körapop 225
Seals wire module FKM
Housing USB/Wi-Fi Steel 1.4301
Cover USB/Wi-Fi PE-HD
Seals USB/Wi-Fi Silicon
Operation Manual ProxiMate™ 19/94
Page 20
4 | Transport and storage BÜCHI Labortechnik AG
4 Transport and storage
4.1 Transport
NOTICE
Risk of breakage due to incorrect transportation
Make sure that the instrument is fully dismantled.
Pack every instrument components properly to prevent breakage. Use the original packaging whenever possible.
Avoid sharp movements during transit.
u After transporting, check the instrument and all glass components for damage. u Damage that has occurred in transit should be reported to the carrier. u Keep packaging for future transportation.
4.2 Storage
u Make sure that the ambient conditions are complied with (see Chapter3.7 "Tech-
nical data", page17).
u Wherever possible, store the device in its original packaging. u After storage, check the device for damage and replace if necessary.
4.3 Lifting the instrument
WARNING
Danger due to incorrect transportation
The possible consequences are crushing injuries, cuts and breakages.
u The instrument should be transported by two persons at the same time. u Lift the instrument at the points indicated.
Fig.8: Lifting the instrument
u Lift the instrument – this requires two persons lifting at the points indicated on the
bottom of the instrument.
20/94 Operation Manual ProxiMate™
Page 21
BÜCHI Labortechnik AG Installation | 5
1
2
5 Installation
5.1 Before installation
NOTICE
Instrument damage due to switching it on too early.
Switching on the instrument too early after transportation can cause damage.
u Climatize the instrument after transportation.
5.2 Installation site
The installation site must meet the following requirements:
Firm, level and vibration-free surface.Minimum space requirement: 260 mm x 435 mm x 500 mm (W x D x H).Take into account the maximum product dimensions and weight.Do not expose the instrument to any external thermal loads, such as direct solar
radiation.
NOTE
Make sure that the power supply can be disconnected at any time in an emergency.
5.3 Securing against earthquakes
Fig.9: Securing against earthquakes
1 Cooler 2 Cord
u Loop a cord around the cooler and attach it to a fixed point.
Operation Manual ProxiMate™ 21/94
Page 22
5 | Installation BÜCHI Labortechnik AG
5.4 Establishing electrical connections
The instrument is design for stationary installation and therefore not equipped with a mains plug. The electrical installation must be performed by an electrician or equiva­lent specialist in any case.
WARNING
Death or serious burns by electric current.
u Have the installation carried out by an electrician or a person with similar expert
knowledge.
u After installation, check electrical safety.
An emergency power-off device must be installed in the vicinity of the instrument with free access granted.
For details about the installation requirements, see "Electric Installation Instruction" document.
NOTE
Observe the regulatory provisions when connecting the instrument to power supply.
4 Use external main switches (e.g., emergency off) in accordance with the stan-
dards IEC 60947-1 and IEC 60947-3.
4 Use additional electrical safety features (e.g., residual-current circuit breakers) to
comply with local laws and regulations.
The power supply must fulfill the following conditions:
1. Provide the mains voltage and frequency specified on the type plate of the instru-
ment.
2. Be designed for the load imposed by the instruments connected.
3. Be equipped with suitable fuses and electrical safety features.
4. Be equipped with a proper grounding.
NOTICE
Risk of property damage and diminished performance due to use of un­suitable power cables.
The power supply cables supplied with the product by BUCHI precisely match the re­quirements of the device. If other power cables that do not meet those requirements are used, the device may be damaged and/or its performance diminished.
u Use only the power supply cables supplied with the product or ordered separately
from BUCHI.
u If using any other power supply cables, make sure that they match the specifica-
tions on the type plate.
5.5 Installing a USB device
NOTE
Make sure that the USB socket is covered either with the cover lid or the protection cover.
22/94 Operation Manual ProxiMate™
Page 23
BÜCHI Labortechnik AG Installation | 5
u Open one of the cover lids for the USB socket.
u Plug USB device in the USB socket.
u Attach the protection cover.
5.6 Software settings
5.6.1 Apply for a licence
NOTE
The location of the destination folder is fixed. See Chapter10.2 "Folder locations", page89
The following data is necessary for the licence request:
Article NameSoftware licence serial number (sticker on the cover page of the Quick Guide)Company NameFirst NameLast NameSerial NumberCountryE-Mail address
Navigation path
[Licence Request]
Operation Manual ProxiMate™ 23/94
Page 24
5 | Installation BÜCHI Labortechnik AG
u Navigate to the action [Licence Request] action via the navigation path.
ð The display shows a dialog with the mandatory information needed for appli-
cation.
u Tap the button
ð The display shows a dialog with an alphanumeric input box.
u Enter your data. u Tap the button [OK].
ð The dialog box closes.
u After entering all mandatory data tap the button [Create]. u Confirm the secure question by tapping the button [OK].
ð The application file is created.
u Send the file to Buchi to apply for a licence.
E-Mail: registration@buchi.com
5.6.2 Import Licence
NOTE
Installed demo license
Delete installed demo license before importing the permanent license.
4 Start the software [Windows Explorer] on the instrument. 4 Navigate to folder options via the following navigation path: View Folder Op-
tions View
4 Activate the function [Show hidden files, folders and drives]. 4 Delete the demo license file from the following location: c:\ProgramData\BUCHI
\LicenseManager\Licences.
next to the information.
Navigation path
[Licence Import]
Precondition: R The licence file is saved on the instrument.
u Navigate to the action [Licence Import] via the navigation path.
ð The display shows a dialog for choosing the folders on the instrument.
u Navigate to the saving folder of the licence you wish to import. u Select the application. u Tap the button [OK].
ð The instrument imports the licence.
5.6.3 Calibrating a Baseline Correction Vector (BCV)
The Baseline Correction Vector reduces the influence of the sample container or pre­sentation method during measurement.
For routine sample presentation a BCV is calculated with the correct (matching) white reference standard mounted on the correct spacer. For measurements of liquids the BCV is calculated using an empty glass petri-dish and transflectance adapter.
NOTE
To ensure that the BCV is calculated correctly the white reference or sample presen­tation should be rotated 5 times during the operation.
24/94 Operation Manual ProxiMate™
Page 25
BÜCHI Labortechnik AG Installation | 5
5
4
3
2
1
Navigation path
[Calibrate BCV]
Precondition: R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
R A sample presentation is defined.
R The instrument is stabilized with at least 2 hours continuous running.
u Navigate to the action [Calibrate BCV] via the navigation path.
ð The instrument shows a dialog box with mandatory settings.
Fig.10: Dialog box Cal ibr ate BCV
1 Measurement view 2 External Reference 3 Progress (view) 4 Measurement mode 5 Sample presentation ID
u Select the name of the sample presentation you wish to calibrate. u Enter the corresponding values. u Place the white reference or empty sample presentation method in the correct po-
sition.
u Tap the button [Start]. u Follow instructions on the display during the calibration. u Confirm the instructions by tapping the [OK] button.
5.6.4 Import external reference data
NOTE
The location of the destination folder is fixed. See Chapter10.2 "Folder locations", page89
Navigation path
[Import External Refer-
ence Data]
Precondition: R The reference data file is saved on the instrument.
u Navigate to the action [Import External Reference Data] via the navigation path.
ð The display shows the dialog box Imp ort Ex ter nal Ref ere nce Da ta.
Operation Manual ProxiMate™ 25/94
Page 26
5 | Installation BÜCHI Labortechnik AG
u Tap the button next to the input box file.
ð The display shows a dialog box with the selectable reference data.
u Select the import file. u Tap the button [OK].
ð The dialog box closes.
u Tap the button [Import].
ð The external reference is imported.
5.6.5 Changing the language of the instrument
Navigation path
[General]
Precondition: R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
u Navigate to the view Gen era l via the navigation path. u Tap the function [Edit] on the function bar. u Tap the action [Selected Language].
ð The control panel shows a dialog box with selectable languages.
u Select a language. u Tap the button [OK].
ð The dialog box closes. ð The control panel shows a dialog box.
u Tap the button [OK] to confirm the dialog box. u Restart the software.
26/94 Operation Manual ProxiMate™
Page 27
BÜCHI Labortechnik AG Operation | 6
4
3
1
2
6 Operation
CAUTION
Risk of injury from glass splinters
Sharp objects can damage the display.
u Keep sharp objects away from the display.
6.1 Layout of the control panel
Fig.11: Control panel
No. Description Function
1 Status bar Shows the current status of the instru-
ment. See Chapter6.4 "Status bar", page38
2 Menu bar Shows symbols representing the menus.
3 Content area Shows current settings, submenus or ac-
4 Function bar Shows functions that can be performed
See Chapter6.3 "Menu bar", page28
tions depending on the current operation.
according to current operation. See Chapter6.2 "Function bar",
page27
6.2 Function bar
The function bar shows available functions according to the current operation.
The functions on the function bar are executed by tapping the relevant function but­tons.
Operation Manual ProxiMate™ 27/94
Page 28
6 | Operation BÜCHI Labortechnik AG
Symbol Description Meaning
[Back] The display reverts to the previous
view.
[Confirm] Confirm a measurement result.
[Switch off] The instrument shuts down.
[Select] Selects the marked application.
[Login] The display shows the dialog L ogi n.
[Multiple selection] Activates multiple selection of mea-
surements.
[Select all] Selects all measurements in the list.
NOTICE!only available if multiple selection is activated
[Report] Generates on-screen report
[Save Excel] Saves the report as Excel file.
[Save PDF] Saves the report as PDF file.
[Edit] Allows the selected item to be
edited.
[New] Creates a new application or prop-
erty.
[Delete] Deletes the selected value.
[Copy] Copies the marked application.
[Autocal] Starts the auto calibration function.
[Import] Imports an application.
[Export] Exports the marked application.
[Export measurement data]
Export the marked measurement data.
[Information] Shows information about the instru-
ment and the installed Licenses.
[Full page ] The report fits on full page.
[Width scrolling] The report fits to width scrolling.
6.3 Menu bar
The menus are represented by symbols on the menu bar. Menus are navigated via the touchscreen.
28/94 Operation Manual ProxiMate™
Page 29
BÜCHI Labortechnik AG Operation | 6
The following menus are available:
Menu symbol Meaning Actions
[Start] menu Carrying out a measurement.
See Chapter6.3.1 "Start menu", page30
[History] menu Shows the results of completed
measurements.
Creating reports.Adding sample reference values.Adding samples to calibration set.Updating calibrations using the
AutoCal function.
Exporting Measurement Data.
See Chapter6.3.2 "History menu", page32
[Application] menu Creating, editing or selecting an
application.
See Chapter6.3.3 "Application menu", page35
[Tools] menu Setting up configurations, mainte-
nance and system tests.
Viewing instrument log and other
instrument counters.
Licence requests and import func-
tions.
Setting up remote sessions.Changing Login credentials.
See Chapter6.3.4 "Tools menu", page36
[Configuration] menu Editing instrument settings.
Available in administrator mode only.
Operation Manual ProxiMate™ 29/94
Page 30
6 | Operation BÜCHI Labortechnik AG
6
5
4
3
7
8
9
2
1
6.3.1 Start menu
Fig.12: Start menu
1 View selected application 2 Order
See Chapter "Insert order", page32
3 Sample ID
4 Control button
See Chapter "Insert sample ID", page31
5 Timestamp (automatically) 6 Shows the properties of the selected
application
7 Shows the Standard Operation Pro-
cedure for the selected application
8 Note
See Chapter "Insert note", page32
9 Barcode
See Chapter "Insert barcode", page31
Control button
The [control] button can show the following symbols:
Symbol Meaning
The measurement is finished.The tested sample is in the specifications.By tapping the symbol the instrument carries out a measure-
ment.
Target value is the expected parameter for the product.
The target value is defined in the Application.
30/94 Operation Manual ProxiMate™
Page 31
BÜCHI Labortechnik AG Operation | 6
Symbol Meaning
No application available.Measurement not possible.
The measurement is finished.The predicted value is outside the tolerance.
Tolerance is a difference from the target value.
Tolerances are defined in the Application.
The measurement is finished.A calibration model is missing.Mahalanobis outlierThe predicted value is outside calibration range.The predicted value is outside of the set limit.
Limit is an absolute value.
Limits are defined in the Application.
The measurement is in progress.Cancel measurement.
Insert sample ID
The Sample ID is a label to identify the sample under analysis.
Navigation path
u Tap the button next to [Sample ID].
ð The display shows a dialog with an alphanumeric input box.
u Enter a sample number, name or code. u Tap the button [OK].
ð The sample ID is saved. ð The dialog box closes.
Insert barcode
The bar code is a label to identify the sample under analysis.
A barcode reader can be connected to the instrument. When the barcode reader is configured the user can use this label to identify the sample under analysis.
As an alternative insert a barcode manually
Operation Manual ProxiMate™ 31/94
Page 32
6 | Operation BÜCHI Labortechnik AG
Navigation path
u Tap the button next to the [Barcode].
ð The display shows a dialog with an alphanumeric input box.
u Enter the Barcode for the sample. u Tap the button [OK].
ð The barcode is saved. ð The dialog box closes.
Insert note
The Note is a label to identify the sample under analysis. With default settings the Note is not a mandatory field.
Navigation path
u Tap the button next to the [Note].
ð The display shows a dialog with an alphanumeric input box.
u Enter a note. u Tap the button [OK].
ð The note is saved. ð The dialog box closes.
Insert order
The Order is a label to identify the sample under analysis. With default settings the Order is not a mandatory field.
Navigation path
u Tap the button next to the [Order].
ð The display shows a dialog with an alphanumeric input box.
u Enter an order number or code. u Tap the button [OK].
ð The order is saved. ð The dialog box closes.
6.3.2 History menu
The [History] menu lists past measurement data.
32/94 Operation Manual ProxiMate™
Page 33
BÜCHI Labortechnik AG Operation | 6
9
8
7
6
12
13
5
42
1
3
11
10
Fig.13: Data menu
1 Shows selected filter options
2 Timestamp See Chapter "Select filter", page33 Chapter "Deselect filter", page34
3 Filter button 4 Sample ID 5 Sample properties 6 Displays spectra 7 Displays properties / spectra 8 Enlarge Note Area 9 Barcode 10 Note 11 Add to calibration dataset button 12 Application name 13 Order number
See:
Chapter6.13 "Creating a report", page68
Chapter6.14 "Exporting measurement data", page68
Chapter6.10 "Adding measurement data to a calibration", page65
Select filter
Filters allow the user to narrow the selection of sample to those that fulfill specific cri­teria.
Navigation path
u Navigate to the [History] menu via the navigation path. u Tap button filter.
ð The display shows a dialog box with the selectable filters.
Operation Manual ProxiMate™ 33/94
Page 34
6 | Operation BÜCHI Labortechnik AG
4
3
5
2
1
Fig.14: Dialog box filter
1 Application 2 Time and date 3 Order 4 Sample ID 5 Include in calibration dataset (Auto-
Cal)
u Select the filter settings according to the requirements. u Tap the button [OK].
ð The dialog box closes. ð The display shows the filtered measurements.
Deselect filter
Navigation path
u Navigate to the [History] menu via the navigation path. u Tap the button filter.
ð The display shows a dialog box with the selectable filters.
34/94 Operation Manual ProxiMate™
Page 35
BÜCHI Labortechnik AG Operation | 6
4
3
5
2
1
Fig.15: Dialog box filter
1 Application 2 Time and date 3 Order 4 Sample ID 5 Include in calibration dataset (Auto-
Cal)
u Tap the button [X] of the selected filter. u Tap the button [OK].
ð The dialog box closes. ð The display shows all available measurements.
6.3.3 Application menu
In the [Application] menu applications can be created, edited and selected.
Operation Manual ProxiMate™ 35/94
Page 36
6 | Operation BÜCHI Labortechnik AG
1
2
Fig.16: Application menu
1 Shows a list of available applications
or properties
See:
Chapter6.6 "Editing an application", page40
Chapter6.7 "Deleting an application", page47
Chapter6.8 "Editing a property", page47
Chapter6.9 "Deleting a property", page65
Chapter6.12 "Importing application data", page67
Chapter6.11 "Exporting application data", page67
6.3.4 Tools menu
The tools menu offers different tools for maintenance and application settings.
2 Shows a list of available settings.
36/94 Operation Manual ProxiMate™
Page 37
BÜCHI Labortechnik AG Operation | 6
1
Fig.17: Tools menu
1 Tools
The following tools are available:
Action Option Explanation
[NIRWise Log] View Shows a dialog with a list of mes-
[Backup NIRWise Data]
[Extended System Tests]
sages that occurred during opera­tion.
(all users)
Procedure Carrying out a data backup.
(administrator only) See Chapter7.6 "Carrying out a
data backup", page84
Procedure Carrying out different system tests.
(administrator only) See Chapter7.5.2 "Carrying out a
Comprehensive System Test", page84
See Chapter7.5.3 "Carrying out an Advanced System Test", page84
[Test BCV] Procedure Carrying out a Baseline Correction
Test. (administrator only) See Chapter7.5.1 "Carrying out a
Baseline Correction Vector test", page83
[Confirm Lamp Re­placement]
Reset See Chapter7.2.3 "Confirm Lamp
Replacement", page80 (administrator only)
Operation Manual ProxiMate™ 37/94
Page 38
6 | Operation BÜCHI Labortechnik AG
Action Option Explanation
[License Request] Procedure See Chapter5.6.1 "Apply for a li-
cence", page23 (administrator only)
[Remote host] Open program The Software TeamViewer opens
for remote support. (administrator only)
[Cleaning Mode] Procedure See Chapter7.7 "Cleaning the dis-
play", page85 (all users)
[Operational Coun­ters]
View / Procedure The data is displayed depends on
the system configuration: Total Operation time / NIR Operation
Time / Lamp UP View / Reference Up View / System Temperature / Ro­tation Time / VIS Operation Time / Lamp Down View / References Down View / System Humidity
(administrator only)
[Extended System Test History]
View Shows a dialog with a further infor-
mation about the tests carried out. (administrator only)
[Calibrate BCV] Procedure See Chapter5.6.3 "Calibrating a
Baseline Correction Vector (BCV)", page24
(administrator only)
[Import External Reference Data]
Procedure See Chapter5.6.4 "Import external
reference data", page25
[Licence Import] Procedure See Chapter5.6.2 "Import Licence",
page24 (administrator only)
6.4 Status bar
The status bar shows the status of the instrument.
The following statuses are possible:
Indications on the status bar
View Status Stabilizing Spectrometer Shows the remaining time.
Ready to Measure The instruments is ready to measure.
Tempering to The instrument is warming up.
The Status bar shows the target and current in­strument temperature.
Configuring… The instrument is starting up.
The instrument is loading an application.
Measuring... The instrument is carrying out a measurement.
Adjusting Exposure Time… Instrument initialization
38/94 Operation Manual ProxiMate™
Page 39
BÜCHI Labortechnik AG Operation | 6
View Status Calibrating BCV… The instrument calibrates the Baseline Correction
Vector.
Initializing… Instrument initialization
Measuring Dark Reference… The instrument measures the internal dark refer-
ence.
Error An error occurred.
See Chapter8 "Help with faults", page86
Extended System Tests The action Extended System Test is activated. In-
strument waits for action.
Preconditioning… Instrument initialization
Standard System Test Run­ning…
The instrument carries out a Standard System Test.
System Testing… Instrument running internal tests
Ready for Configuration No application selected.
No Baseline Correction for Sample Presenta-
tion of current application defined.
Measuring White Reference... Instrument initialization
Lamp Pre-Heating The instrument is heating the lamp. The Status
bar shows the remaining time.
Symbols on the status bar
Symbol Status
The current user is logged on as administrator The instrument is in administrator mode.
The current user is logged on as operator. The instrument is in operator mode.
A warning has occurred check the [NIRWise log]. See Chapter6.3.4 "Tools menu", page36
6.5 Log in administrator mode
Navigation path
Precondition: R The instrument is in operator mode.
u Navigate to the [Tools] menu via the navigation path. u Tap the function [Login] in the function bar.
ð The display shows a dialog box with the available Users.
u Tap [Administrator].
ð The display shows a dialog with an alphanumeric input box.
u Enter the password.
Operation Manual ProxiMate™ 39/94
Page 40
6 | Operation BÜCHI Labortechnik AG
u Tap the button [OK].
ð The dialog box closes. ð The instrument is in administrator mode. ð The status bar shows the icon administrator.
6.6 Editing an application
NOTE
Applications can only be edited in Administrator mode.
6.6.1 Creating a new application
There are two ways to create an application: by copying an existing application
See Chapter "Creating a new application by copying an existing application", page40
by creating a new application
See Chapter "Creating a new application", page40
Creating a new application
Proceedings:
Navigation path
Precondition: R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
u Navigate to the [Application] menu via the navigation path. u Tap the function [Edit] on the function bar.
ð The display highlights the function [Edit] in green.
u Tap the function [Add] on the function bar.
ð The display shows a dialog with an alphanumeric input box.
u Enter a name for the application. u Tap the button [OK].
ð The dialog box closes. ð The new application is created.
Creating a new application by copying an existing application
Proceedings:
Navigation path
Precondition: R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
u Navigate to the [Application] menu via the navigation path.
40/94 Operation Manual ProxiMate™
Page 41
BÜCHI Labortechnik AG Operation | 6
u Tap the function [Edit] on the function bar.
ð The display highlights the function [Edit] in green.
u Tap the application you wish to copy.
ð The display highlights the application in green.
u Tap the function [Copy] on the function bar.
ð The display shows a dialog with an alphanumeric input box.
u Enter a name for the application. u Tap the button [OK].
ð The display shows a dialog with the properties of the copied application high-
lighted in green.
u Tap the properties you do not want to copy.
ð The disabled properties are highlighted in white.
u Tap the button [OK].
ð The dialog box closes. ð The new application is created.
6.6.2 Changing the name of an application
NOTE
It is not possible to rename an application after it has been used once.
Proceedings:
Navigation path
Precondition: R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
u Navigate to the [Application] menu via the navigation path. u Tap the function [Edit] on the function bar.
ð The display highlights the function [Edit] in green.
u Tap the name of the application that you wish to edit.
ð The display highlights the application in green.
u Tap the action [Name].
ð The display shows a dialog with an alphanumeric input box.
u Enter a name for the new application. u Tap the button [OK].
ð The dialog box closes. ð The new name is saved.
6.6.3 Changing the alias of an application
The alias function allows to give a previously defined application a local name.
Proceedings:
Navigation path
Operation Manual ProxiMate™ 41/94
Page 42
6 | Operation BÜCHI Labortechnik AG
Precondition: R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
u Navigate to the [Application] menu via the navigation path. u Tap the function [Edit] on the function bar.
ð The display highlights the function [Edit] in green.
u Tap the name of the application that you wish to edit.
ð The display highlights the application in green.
u Tap the action [Alias].
ð The display shows a dialog with an alphanumeric input box.
u Enter an alias for the application. u Tap the button [OK].
ð The dialog box closes. ð The alias is saved.
6.6.4 Changing the Measurement view of an application
Depending on instrument configuration the following measurement views are avail­able:
View Explanation
Up The application uses the up view lamp.
Radiation is directed and collected from the underside of the sample through the sample cup.
Down The application uses the down view lamp.
Radiation is directed onto and collected from the upper sur­face of the sample. The radiation does not interact with the sample cup.
Proceedings:
Navigation path
Precondition: R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
u Navigate to the [Application] menu via the navigation path. u Tap the function [Edit] on the function bar.
ð The display highlights the function [Edit] in green.
u Tap the name of the application that you wish to edit.
ð The display highlights the application in green.
u Tap the action [Measurement view].
ð The display shows a dialog box with the selectable values.
u Select a value. u Tap the button [OK].
ð The dialog box closes. ð The setting is saved.
42/94 Operation Manual ProxiMate™
Page 43
BÜCHI Labortechnik AG Operation | 6
6.6.5 Changing the Measurement mode of an application
Choose the measurement mode according to the setting for the sample presentation. See Chapter6.6.6 "Changing the Sample presentation of an application", page43
The following measurement modes are available:
Mode Explanation
Diffuse reflection The application runs the measurement in the reflection mode.
Diffuse reflectance measurement mode is used to measure solid and powdered samples.
Transflection (up-view instru-
ments only)
The application runs the measurement in the transflection mode.
Transflectance mode is used to measure liquids and gels. Transflectance mode require the use of a transflection cover.
Proceedings:
Navigation path
Precondition: R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
u Navigate to the [Application] menu via the navigation path. u Tap the function [Edit] on the function bar.
ð The display highlights the function [Edit] in green.
u Tap the name of the application that you wish to edit.
ð The display highlights the application in green.
u Tap the action [Measurement mode].
ð The display shows a dialog box with the selectable values.
u Select a value. u Tap the button [OK].
ð The dialog box closes. ð The setting is saved.
6.6.6 Changing the Sample presentation of an application
With Baseline Correction Vector (BCV) Different Sample Presentation modes can be defined to correct, for example, the effect of sample container on the spectral mea­surement.
The following presentation modes are available:
Modes Explanation
Default Apply the default settings for baseline correction.
New Calibrate an individual Baseline correction. See Chapter5.6.3
"Calibrating a Baseline Correction Vector (BCV)", page24
Individual Baseline
Individual calibrated Baseline corrections.
correction
Proceedings:
Operation Manual ProxiMate™ 43/94
Page 44
6 | Operation BÜCHI Labortechnik AG
Navigation path
Precondition: R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
u Navigate to the [Application] menu via the navigation path. u Tap the function [Edit] on the function bar.
ð The display highlights the function [Edit] in green.
u Tap the name of the application that you wish to edit.
ð The display highlights the application in green.
u Tap the action [Sample presentation].
ð The display shows a dialog box with the selectable values.
u Select a value. u Tap the button [OK].
ð The dialog box closes. ð The setting is saved.
6.6.7 Enter a description for an application
Proceedings:
Navigation path
Precondition: R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
u Navigate to the [Application] menu via the navigation path. u Tap the function [Edit] on the function bar.
ð The display highlights the function [Edit] in green.
u Tap the name of the application that you wish to edit.
ð The display highlights the application in green.
u Tap the action [Description].
ð The display shows a dialog with an alphanumeric input box.
u Enter a description for the application. u Tap the button [OK].
ð The dialog box closes. ð The description is saved.
6.6.8 Changing the Rotation of an application
The Application has the option to select whether the sample is rotated during mea­surement.
The following settings are available:
Mode Explanation
Yes The sample carrier rotates during the measurement.
No The sample carrier does not rotate during the measurement.
44/94 Operation Manual ProxiMate™
Page 45
BÜCHI Labortechnik AG Operation | 6
Proceedings:
Navigation path
Precondition: R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
u Navigate to the [Application] menu via the navigation path. u Tap the function [Edit] on the function bar.
ð The display highlights the function [Edit] in green.
u Tap the name of the application that you wish to edit.
ð The display highlights the application in green.
u Tap the action [Rotation].
ð The display shows a dialog box with the selectable values.
u Select a value. u Tap the button [OK].
ð The dialog box closes. ð The setting is saved.
6.6.9 Changing the Measurement duration of an application
The default measurement time (15 seconds) allows the sample to complete one rota­tion. Shorter measurement time does not allow a complete rotation of the sample.
Proceedings:
Navigation path
Precondition: R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
u Navigate to the [Application] menu via the navigation path. u Tap the function [Edit] on the function bar.
ð The display highlights the function [Edit] in green.
u Tap the name of the application that you wish to edit.
ð The display highlights the application in green.
u Tap the action [Measurement duration].
ð The display shows a dialog box with a numeric input box.
u Enter the value in the numeric input box. u Tap the button [OK].
ð The dialog box closes. ð The value for the duration is saved.
6.6.10 Enter a Standard Operating Procedure (SOP) for an application
Each Application has an optional Standard Operating Procedure (SOP) associated with it. A Standard Operating Procedure SOP shows instructions for the user to follow during measurement.
Operation Manual ProxiMate™ 45/94
Page 46
6 | Operation BÜCHI Labortechnik AG
Proceedings:
Navigation path
Precondition: R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
u Navigate to the [Application] menu via the navigation path. u Tap the function [Edit] on the function bar.
ð The display highlights the function [Edit] in green.
u Tap the name of the application that you wish to edit.
ð The display highlights the application in green.
u Tap the action [SOP].
ð The display shows a dialog with an alphanumeric input box.
u Enter the steps which the operator has to carry for the application. u Tap the button [OK].
ð The dialog box closes. ð The description is saved.
6.6.11 Changing the selection ability of an application
Change whether an Application can be selected by the user during operation.
The following modes are available:
Mode Explanation
Yes The application is selectable.
No The application is not selectable.
Proceedings:
Navigation path
Precondition: R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
u Navigate to the [Application] menu via the navigation path. u Tap the function [Edit] on the function bar.
ð The display highlights the function [Edit] in green.
u Tap the name of the application that you wish to edit.
ð The display highlights the application in green.
u Tap the action [Selectable].
ð The display shows a dialog box with the selectable values.
u Select a value.
46/94 Operation Manual ProxiMate™
Page 47
BÜCHI Labortechnik AG Operation | 6
u Tap the button [OK].
ð The dialog box closes. ð The setting is saved.
6.7 Deleting an application
NOTE
It is not possible to delete an application after it has been used for measurement.
NOTE
The following procedure fully deletes the Application. It is not possible to restore the Application after it has been deleted.
Proceedings:
Navigation path
Precondition: R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
u Navigate to the [Application] menu via the navigation path. u Tap the function [Edit] on the function bar.
ð The display highlights the function [Edit] in green.
u Tap the name of the application you wish to remove.
ð The display highlights the application in green.
u Tap the function [Remove] on the function bar. u Select [Yes] to confirm the action in response to the confirmation question.
ð The application is deleted.
6.8 Editing a property
NOTE
Properties can only be edited in Administrator mode.
6.8.1 Creating a new Property
Proceedings:
Navigation path
Precondition: R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
u Navigate to the [Application] menu via the navigation path. u Tap the function [Edit] on the function bar.
ð The display highlights the function [Edit] in green.
u Tap the name of the application that you wish to edit.
ð The display highlights the application in green.
Operation Manual ProxiMate™ 47/94
Page 48
6 | Operation BÜCHI Labortechnik AG
u Tap one of the available properties.
ð The display shows the view Pro per ty.
u Tap the function [Add] on the function bar.
ð The display shows a dialog with an alphanumeric input box.
u Enter a name for the property. u Tap the button [OK].
ð The dialog box closes. ð The new property is created.
6.8.2 Changing the name of a property
NOTE
It is not possible to rename a property after it was used once.
Proceedings:
Navigation path
Precondition: R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
u Navigate to the [Application] menu via the navigation path. u Tap the function [Edit] on the function bar.
ð The display highlights the function [Edit] in green.
u Tap the name of the application that you wish to edit.
ð The display highlights the application in green.
u Tap the name of the property that you wish to edit.
ð The display shows the view Pro per ty.
u Tap the action [Name].
ð The display shows a dialog with an alphanumeric input box.
u Enter a name for the property. u Tap the button [OK].
ð The dialog box closes. ð The new name is saved.
6.8.3 Changing the alias of a property
The alias function allows to give a previously defined property a local name.
Proceedings:
Navigation path
Precondition: R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
u Navigate to the [Application] menu via the navigation path. u Tap the function [Edit] on the function bar.
ð The display highlights the function [Edit] in green.
48/94 Operation Manual ProxiMate™
Page 49
BÜCHI Labortechnik AG Operation | 6
u Tap the name of the application that you wish to edit.
ð The display highlights the application in green.
u Tap the name of the property that you wish to edit.
ð The display shows the view Pro per ty.
u Tap the action [Alias].
ð The display shows a dialog with an alphanumeric input box.
u Enter an alias for the property. u Tap the button [OK].
ð The dialog box closes. ð The alias is saved.
6.8.4 Changing the Sort order of a property
The sort order defines the position in which the property will be displayed within an Application when there are multiple properties.
Proceedings:
Navigation path
Precondition: R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
u Navigate to the [Application] menu via the navigation path. u Tap the function [Edit] on the function bar.
ð The display highlights the function [Edit] in green.
u Tap the name of the application that you wish to edit.
ð The display highlights the application in green.
u Tap the name of the property that you wish to edit.
ð The display shows the view Pro per ty.
u Tap the action [Sort order].
ð The display shows a dialog box with a numeric input box.
u Enter the value in the numeric input box. u Tap the button [OK].
ð The dialog box closes. ð The value for the Sort order is saved.
6.8.5 Changing the Prediction type of a property
The setting of the prediction type influences the further setting possibilities of the property.
The following settings are available:
Operation Manual ProxiMate™ 49/94
Page 50
6 | Operation BÜCHI Labortechnik AG
Prediction type Explanation Available predictions settings
[Calibration model]
Uses assigned calibration model to predict parameter value from spectrum.
The calibration model uses a chemometric model.
Decimal Places see Chap­ter6.8.15 "Changing the Deci­mal places of a property", page59
Unit see Chapter6.8.16 "Chang­ing the Unit of a property", page60
Initial Wavelength Range seeChapter6.8.6 "Changing the Initial Wavelength Range (Cali­bration model only)", page53
Postpredicted Value see Chap­ter6.8.7 "Changing the Postpre­dicted Value (Calibration model only)", page54
Slope see Chapter6.8.18 "Changing the Slope of a prop­erty", page61
Bias see Chapter6.8.17 "Changing the Bias of a prop­erty", page60
Mahalanobis see Chapter6.8.19 "Changing the Mahalanobis of a property (Calibration model only)", page61
Target see Chapter6.8.20 "Changing the Target of a prop­erty", page62
Tolerance Min. see Chap­ter6.8.24 "Changing the Toler­ance minimum of a property", page64
Tolerance Max. see Chap­ter6.8.23 "Changing the Toler­ance maximum of a property", page63
Limit Min. see Chapter6.8.22 "Changing the Limit minimum of a property", page63
Limit Max. seeChapter6.8.21 "Changing the Limit maximum of a property", page62
50/94 Operation Manual ProxiMate™
Page 51
BÜCHI Labortechnik AG Operation | 6
Prediction type Explanation Available predictions settings
[Calculated Property]
Uses define properties that are mathematical calculated from other properties. e.g. Dry Matter = 100 - Moisture
Decimal Places see Chap­ter6.8.15 "Changing the Deci­mal places of a property", page59
Unit see Chapter6.8.16 "Chang­ing the Unit of a property", page60
Formula see Chapter6.8.10 "Changing the Formula (Calcu­lated Property only)", page56
Slope see Chapter6.8.18 "Changing the Slope of a prop­erty", page61
Bias see Chapter6.8.17 "Changing the Bias of a prop­erty", page60
Target see Chapter6.8.20 "Changing the Target of a prop­erty", page62
Tolerance Min. see Chap­ter6.8.24 "Changing the Toler­ance minimum of a property", page64
Tolerance Max. see Chap­ter6.8.23 "Changing the Toler­ance maximum of a property", page63
Limit Min. seeChapter6.8.22 "Changing the Limit minimum of a property", page63
Limit Max. see Chapter6.8.21 "Changing the Limit maximum of a property", page62
Operation Manual ProxiMate™ 51/94
Page 52
6 | Operation BÜCHI Labortechnik AG
Prediction type Explanation Available predictions settings
[Color] Only for models with a visible
detector. Measures the sample color.
Decimal Places see Chap­ter6.8.15 "Changing the Deci­mal places of a property", page59
Observer see Chapter6.8.11 "Changing the Observer (Color only)", page57
Illuminate see Chapter6.8.12 "Changing the Illuminant (Color only)", page57
Metric see Chapter6.8.13 "Changing the Metric (Color only)", page58
Metric Element see Chap­ter6.8.14 "Changing the Metric Element (Color only)", page59
Slope see Chapter6.8.18 "Changing the Slope of a prop­erty", page61
Bias see Chapter6.8.17 "Changing the Bias of a prop­erty", page60
Target see Chapter6.8.20 "Changing the Target of a prop­erty", page62
Tolerance Min. see Chap­ter6.8.24 "Changing the Toler­ance minimum of a property", page64
Tolerance Max. see Chap­ter6.8.23 "Changing the Toler­ance maximum of a property", page63
Limit Min. see Chapter6.8.22 "Changing the Limit minimum of a property", page63
Limit Max. see Chapter6.8.21 "Changing the Limit maximum of a property", page62
Proceedings:
Navigation path
Precondition: R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
52/94 Operation Manual ProxiMate™
Page 53
BÜCHI Labortechnik AG Operation | 6
u Navigate to the [Application] menu via the navigation path. u Tap the function [Edit] on the function bar.
ð The display highlights the function [Edit] in green.
u Tap the name of the application that you wish to edit.
ð The display highlights the application in green.
u Tap the name of the property that you wish to edit.
ð The display shows the view Pro per ty.
u Tap the action [Prediction type].
ð The display shows a dialog box with the selectable values.
u Select a value. u Tap the button [OK].
ð The dialog box closes. ð The setting is saved.
u Editing the available prediction settings according to your needs.
6.8.6 Changing the Initial Wavelength Range (Calibration model only)
Select the wavelength range for initial calibration.
NOTE
Changes in the project file replace the selection made here.
The following settings are available:
Wavelength type Explanation
[NIR only] NIR wavelength range only (900 - 1700 nm)
[VIS only] VIS wavelength range only (400 - 900 nm)
[VIS NIR] NIR and VIS wavelength range (400 - 1700 nm)
Proceedings:
Navigation path
Precondition: R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
u Navigate to the [Application] menu via the navigation path. u Tap the function [Edit] on the function bar.
ð The display highlights the function [Edit] in green.
u Tap the name of the application that you wish to edit.
ð The display highlights the application in green.
u Tap the name of the property that you wish to edit.
ð The display shows the view Pro per ty.
u Tap the action [Initial Wavelength Range].
ð The display shows a dialog box with the selectable values.
u Select a value.
Operation Manual ProxiMate™ 53/94
Page 54
6 | Operation BÜCHI Labortechnik AG
u Tap the button [OK].
ð The dialog box closes. ð The setting is saved. ð The value for the Initial Wavelength Range is saved.
6.8.7 Changing the Postpredicted Value (Calibration model only)
The following settings are available:
Postprediction type Explanation
[Predicted Value] Calculates the values as predicted by the chemometric
model.
[Residuum] Is an indication of the applicability of the model.
The Residuum is the RMS value of the difference between pre-treated sample and model reconstructed spectra.
[Mahalanobis Distance] Calculates the value on basis of a mahalanobis calcula-
tion
[Calibration Base Stan­dard]
Converts the parameter value of the calibration to the value obtained at a different moisture content.
A property moisture is necessary for carrying out this post­prediction type.
See: Chapter6.8.8 "Changing the Calibration Base (Calibration
model only)", page55 Chapter6.8.9 "Changing the Display Basis (Calibration
model only)", page55
Proceedings:
Navigation path
Precondition: R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
u Navigate to the [Application] menu via the navigation path. u Tap the function [Edit] on the function bar.
ð The display highlights the function [Edit] in green.
u Tap the name of the application that you wish to edit.
ð The display highlights the application in green.
u Tap the name of the property that you wish to edit.
ð The display shows the view Pro per ty.
u Tap the action [Postpredicted Value].
ð The display shows a dialog box with the selectable values.
u Select a value. u Tap the button [OK].
ð The dialog box closes. ð The setting is saved. ð The value is saved.
54/94 Operation Manual ProxiMate™
Page 55
BÜCHI Labortechnik AG Operation | 6
6.8.8 Changing the Calibration Base (Calibration model only)
The following settings are available:
Type Explanation
[Dry Basis] The property value is expressed neglecting the presence
of water in the sample. The water contribution is subtracted from the calculation.
[Moisture xx %] The property value is expressed assuming xx % water
content. The amount of water is expressed as a percentage of the
total weight.
[As Is] The property value is expressed including the presence of
water in the sample. The water contribution is included in the calculation.
Proceedings:
Navigation path
Precondition: R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
R The post prediction type [Calibration Base Standard] is selected.
u Navigate to the [Application] menu via the navigation path. u Tap the function [Edit] on the function bar.
ð The display highlights the function [Edit] in green.
u Tap the name of the application that you wish to edit.
ð The display highlights the application in green.
u Tap the name of the property that you wish to edit.
ð The display shows the view Pro per ty.
u Tap the action [Calibration Base].
ð The display shows a dialog box with the selectable values.
u Select a value. u Tap the button [OK].
ð The dialog box closes. ð The setting is saved. ð The value for the calibration base is saved.
6.8.9 Changing the Display Basis (Calibration model only)
The following settings are available:
Type Explanation
[Dry Basis] The property value is expressed neglecting the presence
of water in the sample. The water contribution is subtracted from the calculation.
Operation Manual ProxiMate™ 55/94
Page 56
6 | Operation BÜCHI Labortechnik AG
Type Explanation
[Moisture xx %] The property value is expressed assuming xx % water
content. The amount of water is expressed as a percentage of the
total weight.
[As Is] The property value is expressed including the presence of
water in the sample. The water contribution is included in the calculation.
Proceedings:
Navigation path
Precondition: R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
R The post prediction type [Calibration Base Standard] is selected.
u Navigate to the [Application] menu via the navigation path. u Tap the function [Edit] on the function bar.
ð The display highlights the function [Edit] in green.
u Tap the name of the application that you wish to edit.
ð The display highlights the application in green.
u Tap the name of the property that you wish to edit.
ð The display shows the view Pro per ty.
u Tap the action [Display basis].
ð The display shows a dialog box with the selectable values.
u Select a value. u Tap the button [OK].
ð The dialog box closes. ð The setting is saved. ð The value for the Display basis is saved.
6.8.10 Changing the Formula (Calculated Property only)
Calculation formula for the prediction.
Rules how to enter a formula see Chapter10.3 "Rules entering a formula", page90
Proceedings:
Navigation path
Precondition: R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
u Navigate to the [Application] menu via the navigation path. u Tap the function [Edit] on the function bar.
ð The display highlights the function [Edit] in green.
56/94 Operation Manual ProxiMate™
Page 57
BÜCHI Labortechnik AG Operation | 6
u Tap the name of the application that you wish to edit.
ð The display highlights the application in green.
u Tap the name of the property that you wish to edit.
ð The display shows the view Pro per ty.
u Tap the action [Formula].
ð The display shows a dialog with an alphanumeric input box.
u Enter the formula. u Tap the button [OK].
ð The dialog box closes. ð The formula is saved.
6.8.11 Changing the Observer (Color only)
The settings available corresponding to the international standard CIE 1931 / CIE
1964.
Proceedings:
Navigation path
Precondition: R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
u Navigate to the [Application] menu via the navigation path. u Tap the function [Edit] on the function bar.
ð The display highlights the function [Edit] in green.
u Tap the name of the application that you wish to edit.
ð The display highlights the application in green.
u Tap the name of the property that you wish to edit.
ð The display shows the view Pro per ty.
u Tap the action [Observer].
ð The display shows a dialog box with the selectable values.
u Select a value. u Tap the button [OK].
ð The dialog box closes. ð The setting is saved. ð The value for the Observer type is saved.
6.8.12 Changing the Illuminant (Color only)
The settings available corresponding to the international standard ISO 11664-2:2007 / CIE S 014-2:2006.
Proceedings:
Navigation path
Operation Manual ProxiMate™ 57/94
Page 58
6 | Operation BÜCHI Labortechnik AG
Precondition: R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
u Navigate to the [Application] menu via the navigation path. u Tap the function [Edit] on the function bar.
ð The display highlights the function [Edit] in green.
u Tap the name of the application that you wish to edit.
ð The display highlights the application in green.
u Tap the name of the property that you wish to edit.
ð The display shows the view Pro per ty.
u Tap the action [Illuminant].
ð The display shows a dialog box with a numeric input box.
u Enter the value in the numeric input box. u Tap the button [OK].
ð The dialog box closes. ð The value for the Illuminant is saved.
6.8.13 Changing the Metric (Color only)
The following settings are available:
Type Explanation
[Lab] Calculates the metrics in the L*a*b color space.
[LCh] Calculates the metrics in the L*C*h color space.
[XYZ] Calculates the values in XYZ color space.
Proceedings:
Navigation path
Precondition: R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
u Navigate to the [Application] menu via the navigation path. u Tap the function [Edit] on the function bar.
ð The display highlights the function [Edit] in green.
u Tap the name of the application that you wish to edit.
ð The display highlights the application in green.
u Tap the name of the property that you wish to edit.
ð The display shows the view Pro per ty.
u Tap the action [Metric].
ð The display shows a dialog box with a numeric input box.
u Enter the value in the numeric input box. u Tap the button [OK].
ð The dialog box closes. ð The value for the Metric is saved.
58/94 Operation Manual ProxiMate™
Page 59
BÜCHI Labortechnik AG Operation | 6
6.8.14 Changing the Metric Element (Color only)
Available options depend on the metric selected in Chapter6.8.13 "Changing the Metric (Color only)", page58
Proceedings:
Navigation path
Precondition: R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
u Navigate to the [Application] menu via the navigation path. u Tap the function [Edit] on the function bar.
ð The display highlights the function [Edit] in green.
u Tap the name of the application that you wish to edit.
ð The display highlights the application in green.
u Tap the name of the property that you wish to edit.
ð The display shows the view Pro per ty.
u Tap the action [Metric Element].
ð The display shows a dialog box with the selectable values.
u Select a value. u Tap the button [OK].
ð The dialog box closes. ð The setting is saved. ð The value for the Metric Element is saved.
6.8.15 Changing the Decimal places of a property
Number of decimal places displayed by a property.
Proceedings:
Navigation path
Precondition: R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
u Navigate to the [Application] menu via the navigation path. u Tap the function [Edit] on the function bar.
ð The display highlights the function [Edit] in green.
u Tap the name of the application that you wish to edit.
ð The display highlights the application in green.
u Tap the name of the property that you wish to edit.
ð The display shows the view Pro per ty.
u Tap the action [Decimal places].
ð The display shows a dialog box with a numeric input box.
u Enter the value in the numeric input box.
Operation Manual ProxiMate™ 59/94
Page 60
6 | Operation BÜCHI Labortechnik AG
u Tap the button [OK].
ð The dialog box closes. ð The value for the Decimal places is saved.
6.8.16 Changing the Unit of a property
NOTE
Special characters
4 Tab the button [Shift] on the alphanumeric input box.
ð The alphanumeric input box shows the available special characters.
Navigation path
Precondition: R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
u Navigate to the [Application] menu via the navigation path. u Tap the function [Edit] on the function bar.
ð The display highlights the function [Edit] in green.
u Tap the name of the application that you wish to edit.
ð The display highlights the application in green.
u Tap the name of the property that you wish to edit.
ð The display shows the view Pro per ty.
u Tap the action [Unit].
ð The display shows a dialog with an alphanumeric input box.
u Enter a unit for the property. u Tap the button [OK].
ð The dialog box closes. ð The Unit is saved.
6.8.17 Changing the Bias of a property
A bias is a constant value.
This value is added to the result of a prediction to correct a constant deviation be­tween predicted values and reference values.
Proceedings:
Navigation path
Precondition: R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
u Navigate to the [Application] menu via the navigation path. u Tap the function [Edit] on the function bar.
ð The display highlights the function [Edit] in green.
60/94 Operation Manual ProxiMate™
Page 61
BÜCHI Labortechnik AG Operation | 6
u Tap the name of the application that you wish to edit.
ð The display highlights the application in green.
u Tap the name of the property that you wish to edit.
ð The display shows the view Pro per ty.
u Tap the action [Bias].
ð The control panel shows a dialog box with a numeric input box.
u Enter the value in the numeric input box. u Tap the button [OK].
ð The dialog box closes. ð The value for the Bias is saved.
6.8.18 Changing the Slope of a property
The slope is a factor that is used to correct proportional systematic differences be­tween the result of a measurement and the reference value.
Proceedings:
Navigation path
Precondition: R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
u Navigate to the [Application] menu via the navigation path. u Tap the function [Edit] on the function bar.
ð The display highlights the function [Edit] in green.
u Tap the name of the application that you wish to edit.
ð The display highlights the application in green.
u Tap the name of the property that you wish to edit.
ð The display shows the view Pro per ty.
u Tap the action [Slope].
ð The control panel shows a dialog box with a numeric input box.
u Enter the value in the numeric input box. u Tap the button [OK].
ð The dialog box closes. ð The value for the Slope is saved.
6.8.19 Changing the Mahalanobis of a property (Calibration model only)
The Mahalanobis distance is a measure of spectral similarity between the measured spectrum and the Dataset used in the calibration.
Proceedings:
Navigation path
Precondition: R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
Operation Manual ProxiMate™ 61/94
Page 62
6 | Operation BÜCHI Labortechnik AG
u Navigate to the [Application] menu via the navigation path. u Tap the function [Edit] on the function bar.
ð The display highlights the function [Edit] in green.
u Tap the name of the application that you wish to edit.
ð The display highlights the application in green.
u Tap the name of the property that you wish to edit.
ð The display shows the view Pro per ty.
u Tap the action [Mahalanobis].
ð The control panel shows a dialog box with a numeric input box.
u Enter the value in the numeric input box. u Tap the button [OK].
ð The dialog box closes. ð The value for the Mahalanobis is saved.
6.8.20 Changing the Target of a property
Sets the target value, relative to which the tolerances are considered.
Proceedings:
Navigation path
Precondition: R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
u Navigate to the [Application] menu via the navigation path. u Tap the function [Edit] on the function bar.
ð The display highlights the function [Edit] in green.
u Tap the name of the application that you wish to edit.
ð The display highlights the application in green.
u Tap the name of the property that you wish to edit.
ð The display shows the view Pro per ty.
u Tap the action [Target].
ð The control panel shows a dialog box with a numeric input box.
u Enter the value in the numeric input box. u Tap the button [OK].
ð The dialog box closes. ð The value for the Target is saved.
6.8.21 Changing the Limit maximum of a property
Sets the upper limit for the property.
Proceedings:
Navigation path
62/94 Operation Manual ProxiMate™
Page 63
BÜCHI Labortechnik AG Operation | 6
Precondition: R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
u Navigate to the [Application] menu via the navigation path. u Tap the function [Edit] on the function bar.
ð The display highlights the function [Edit] in green.
u Tap the name of the application that you wish to edit.
ð The display highlights the application in green.
u Tap the name of the property that you wish to edit.
ð The display shows the view Pro per ty.
u Tap the action [Limit Max.].
ð The control panel shows a dialog box with a numeric input box.
u Enter the value in the numeric input box. u Tap the button [OK].
ð The dialog box closes. ð The value for the Limit is saved.
6.8.22 Changing the Limit minimum of a property
Sets the lower limit for the property.
Proceedings:
Navigation path
Precondition: R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
u Navigate to the [Application] menu via the navigation path. u Tap the function [Edit] on the function bar.
ð The display highlights the function [Edit] in green.
u Tap the name of the application that you wish to edit.
ð The display highlights the application in green.
u Tap the name of the property that you wish to edit.
ð The display shows the view Pro per ty.
u Tap the action [Limit Min.].
ð The control panel shows a dialog box with a numeric input box.
u Enter the value in the numeric input box. u Tap the button [OK].
ð The dialog box closes. ð The value for the Limit is saved.
6.8.23 Changing the Tolerance maximum of a property
Tolerance is a difference from the target value.
Sets the upper tolerance relative to the target.
Proceedings:
Operation Manual ProxiMate™ 63/94
Page 64
6 | Operation BÜCHI Labortechnik AG
Navigation path
Precondition: R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
u Navigate to the [Application] menu via the navigation path. u Tap the function [Edit] on the function bar.
ð The display highlights the function [Edit] in green.
u Tap the name of the application that you wish to edit.
ð The display highlights the application in green.
u Tap the name of the property that you wish to edit.
ð The display shows the view Pro per ty.
u Tap the action [Tolerance Max.].
ð The control panel shows a dialog box with a numeric input box.
u Enter the value in the numeric input box. u Tap the button [OK].
ð The dialog box closes. ð The value for the Tolerance is saved.
6.8.24 Changing the Tolerance minimum of a property
Tolerance is a difference from the target value.
Sets the lower tolerance relative to the target.
Proceedings:
Navigation path
Precondition: R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
u Navigate to the [Application] menu via the navigation path. u Tap the function [Edit] on the function bar.
ð The display highlights the function [Edit] in green.
u Tap the name of the application that you wish to edit.
ð The display highlights the application in green.
u Tap the name of the property that you wish to edit.
ð The display shows the view Pro per ty.
u Tap the action [Tolerance Min.].
ð The control panel shows a dialog box with a numeric input box.
u Enter the value in the numeric input box. u Tap the button [OK].
ð The dialog box closes. ð The value for the Tolerance is saved.
64/94 Operation Manual ProxiMate™
Page 65
BÜCHI Labortechnik AG Operation | 6
6.9 Deleting a property
NOTE
It is not possible to delete a property after it has been used for measurement.
Proceedings:
Navigation path
Precondition: R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
u Navigate to the [Application] menu via the navigation path. u Tap the function [Edit] on the function bar.
ð The display highlights the function [Edit] in green.
u Tap the name of the application that you wish to edit.
ð The display highlights the application in green.
u Tap the name of the property you wish to remove.
ð The display highlights the application in green.
u Tap the function [Remove] on the function bar. u Select [Yes] to confirm the action in response to the confirmation question.
ð The property is deleted.
6.10 Adding measurement data to a calibration
6.10.1 Performing a calibration update with AutoCal
For performing the AutoCal function the following measurement data is needed:
minimum three measurementsTwo of these three measurements with different reference values
Navigation path
Precondition: R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
R Reference values are set. See Chapter6.10.2 "Setting a reference value to a
property", page66
R The measurement you wish to use is activated for AutoCal. See Chapter6.10.3
"Activating a measurement for AutoCal", page66
u Navigate to the [History] menu via the navigation path. u Tap the function [Edit] on the function bar.
ð The display highlights the function [Edit] in green.
u Tap the name of the measurement that you wish to use. u Tap the button [ AutoCal] on the function bar.
ð The display shows the dialog Aut oCa l.
Operation Manual ProxiMate™ 65/94
Page 66
6 | Operation BÜCHI Labortechnik AG
1
2 5
6
3 4
7
8
Fig.18: Dialog Aut oCa l
1 Name of statistic parameters 2 Values of statistic parameters of ac-
tual calibration
3 Values of statistic parameters of new
calibration
5 Parameter values according to actual
calibration
4 IDs of last n (default = five) measure-
ments
6 Parameter values as predicted by
new calibration
7 Button [Accept] 8 Button [Reject]
u Check the data. u Press the button [Accept].
6.10.2 Setting a reference value to a property
Navigation path
Precondition: R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
u Navigate to the [History] menu via the navigation path. u Tap the function [Edit] on the function bar.
ð The display highlights the function [Edit] in green.
u Tap the name of the measurement that you wish to edit. u Tap the property you wish to edit.
u Enter the reference value. u Tap the button [OK].
ð The display shows a dialog box with a numeric input box.
ð The reference value is set.
6.10.3 Activating a measurement for AutoCal
Navigation path
66/94 Operation Manual ProxiMate™
Page 67
BÜCHI Labortechnik AG Operation | 6
Precondition: R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
u Navigate to the [History] menu via the navigation path. u Tap the function [Edit] on the function bar.
ð The display highlights the function [Edit] in green.
u Tap the name of the measurement that you wish to use. u Tap the button [Add to calibration dataset ].
ð The measurement is activated for AutoCal.
6.11 Exporting application data
This export function allows the use of (unlicenced) applications onto another instru­ment.
NOTE
The location of the destination folder is fixed. See Chapter10.2 "Folder locations", page89
Navigation path
Precondition: R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
u Navigate to the [Application] menu via the navigation path. u Tap the application you wish to export.
ð The display highlights the application in green.
u Tap the function [Export] on the function bar.
ð The display shows a dialog with the properties of the application you wish to
export. All properties are highlighted in green.
u Tap the properties you do not want to export.
ð The disabled properties are highlighted in white.
u Tap the button [OK].
ð The dialog box closes. ð The application is exported.
6.12 Importing application data
Navigation path
Precondition: R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
u Navigate to the [Application] menu via the navigation path. u Tap the function [Import] on the function bar.
ð The display shows a dialog for choosing the folders on the instrument.
u Navigate to the saving folder of the application you wish to import.
Operation Manual ProxiMate™ 67/94
Page 68
6 | Operation BÜCHI Labortechnik AG
u Select the application.
ð The display shows a dialog with the properties of the application you wish to
import. All properties are highlighted in green.
u Tap the properties you do not want to import.
ð The disabled properties are highlighted in white.
u Tap the button [OK].
ð The dialog box closes. ð The application is imported.
6.13 Creating a report
NOTE
The location of the destination folder is fixed. See Chapter10.2 "Folder locations", page89
Navigation path
u Navigate to the [History] menu via the navigation path. u Tap the name of the measurement you wish to create a report.
ð The display highlights the selected measurement in green. ð The result display shows the values
u Tap the function [Report] on the function bar.
ð The display shows the report.
u According to the requirements tap [Save Excel] or [Save PDF] on the function bar.
ð The report is saved.
6.14 Exporting measurement data
Measurement data can be exported for use with NIRWise PLUS or in general export format (JCAMP.dx or .csv).
NOTE
The location of the destination folder is fixed. See Chapter10.2 "Folder locations", page89
Navigation path
Precondition: R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
u Navigate to the [History] menu via the navigation path. u Select the measurements you wish to export. u Tap the function [Export] on the function bar.
ð The display shows a dialog with selectable values.
u Select the file format in which you wish to export the measurement data. u Tap the button [OK].
ð The measurement data is exported.
68/94 Operation Manual ProxiMate™
Page 69
BÜCHI Labortechnik AG Operation | 6
6.15 Carrying out a measurement
NOTE
Barcode / Sample ID / Order / Note
Barcode, Sample ID, Order and Note for a sample can be edited during the measure­ment process at any time.
4 See Chapter "Insert sample ID", page31 4 See Chapter "Insert barcode", page31 4 See Chapter "Insert note", page32 4 See Chapter "Insert order", page32
NOTE
Returning the test sample in the production can cause contamination.
4 Do not return test sample in the production.
NOTE
Contaminated or defective sample carrier
A contaminated or defective sample carrier causes measuring errors.
4 Do not use defective sample carriers. 4 Make sure that the sample carrier is clean.
NOTE
Wrong measurement results due to overfilled petri dishes.
The sample material can fall out from overfilled Petri dishes. This sample material can lead to an accumulation of material on the measurement window leading to in­correct measurement results.
4 Do not overfill petri dishes.
6.15.1 Preparing the instrument
Time required: up to 30 min
Precondition: R All commissioning operations have been completed. See Chapter5 "Installation",
page21
u Tap the On/Off master switch.
ð The system starts up. ð The display highlights the status bar yellow. ð Once the startup phase is complete the status bar changes from yellow to
black.
6.15.2 Starting measurement
Navigation path
Precondition: R The instrument is prepared. See Chapter6.15.1 "Preparing the instrument",
page69
u Navigate to the [Application] menu via the navigation path.
Operation Manual ProxiMate™ 69/94
Page 70
6 | Operation BÜCHI Labortechnik AG
u Tap the application you wish to use.
ð The display highlights the application in green.
u Tap the function [Select] on the function bar.
ð The display changes to the [Start] menu.
u According to the application requirements put the sample in the sample presenta-
tion area.
u Tap the [Control] button.
ð The instrument is carrying out the measurement. ð The status bar shows the status Measuring.
6.15.3 Ending measurement
Precondition: R The control button shows one of the status measurement finished.
u The symbol of the control button shows in which specifications category the sam-
ple fits.
The following specifications categories are possible:
Specifications cate­gories
Within specification x x
Tolerence x x
Limit x x
u Continue according to the specification the sample is in.
Within specification Tolerence Limit
u Tap the function [Con-
firm] on the function
u Tap the control button u Tap the function [Con-
bar.
ð The measurement
is saved.
6.15.4 Shutting down the instrument
Navigation path
Explanation see Chap­ter "Control button", page30
firm] on the function bar.
ð The measurement
saved.
Explanation see Chap­ter10.1 "Measurement results", page89
u Tap the control button u Tap the function [Con-
firm] on the function bar.
ð The measurement
saved.
u Navigate to the [Start] menu via the navigation path. u Tap the function [Switch off] on the function bar. u Answer YES to the confirmation question.
ð The instrument is shutting down.
70/94 Operation Manual ProxiMate™
Page 71
BÜCHI Labortechnik AG Cleaning and servicing | 7
7 Cleaning and servicing
NOTE
Users may only carry out the servicing and cleaning operations described in this sec­tion.
Any servicing and repair work which involves opening up the casing may only be car­ried out by BUCHI service technicians.
u Use only genuine BUCHI consumables and spare parts in order to ensure correct
operation of the device and preserve the warranty.
7.1 Regular maintenance work
NOTE
If special cleaning processes are necessary contact BUCHI Customer Service.
www.buchi.com/contact
Component Action Frequency
Sample vessel u Wipe down the sample containers
with a damp cloth.
u If heavily soiled:
use mild detergentrinse with clean waterwipe dry with a lint free cloth
Daily
Glass up view u Wipe down the glass window with a
damp cloth.
u If heavily soiled, use ethanol or a mild
detergent.
Glass down view u Wipe down the glass window with a
damp cloth.
u If heavily soiled, use ethanol or a mild
detergent.
Sample presenta­tion
u Wipe down the sample presentation
area with a damp cloth.
u If heavily soiled, use ethanol or a mild
detergent.
Data u Perform a data backup. See Chap-
ter7.6 "Carrying out a data backup", page84
Casing u Wipe down the casing with a damp
cloth.
u If heavily soiled, use ethanol or a mild
detergent.
Cooling fins u Remove dust and foreign objects
from the cooling fins using com­pressed air or a vacuum cleaner.
Daily
Daily
Daily
Weekly
Weekly
Weekly
Control panel u Carry out an Baseline Correction Vec-
Weekly tor test. See Chapter7.5.1 "Carrying out a Baseline Correction Vector test", page83
Operation Manual ProxiMate™ 71/94
Page 72
7 | Cleaning and servicing BÜCHI Labortechnik AG
Component Action Frequency
Display u Wipe down the display with a damp
Weekly cloth. See Chapter7.7 "Cleaning the display", page85
Sample carrier slid-
u Check that the sample carrier sliders
ers
Control panel u Carry out an Advanced System Test.
Desiccant cartridge u Check function of the desiccant car-
u If necessary change the desiccant
Control panel u Carry out a Comprehensive System
7.2 Changing the lamps
NOTICE
Risk of broken lamp
Monthly do not show excessive wear.
Monthly See Chapter7.5.3 "Carrying out an Advanced System Test", page84
Yearly tridge.Chapter7.3 "Checking the des­iccant cartridge", page80
cartridge. See Chapter7.4 "Changing the desiccant cartridge", page82
Yearly Test. See Chapter7.5.2 "Carrying out a Comprehensive System Test", page84
Touching the bulb or the reflector with hands can cause lamp damage.
u Do not touch the bulb with fingers.
u
Touch the lamp on the indicated areas.
7.2.1 Changing the down view lamp
u Disconnect the power supply to the
instrument.
u Wait 15 min. for lamp to cool.
72/94 Operation Manual ProxiMate™
Page 73
BÜCHI Labortechnik AG Cleaning and servicing | 7
u Open the cover lid on the top of the
instrument.
u Remove the seal and the cover lid.
u Press the ferrule. u Turn the ferrule 1/4 turn counterclock-
wise.
Operation Manual ProxiMate™ 73/94
Page 74
7 | Cleaning and servicing BÜCHI Labortechnik AG
u Remove the defective lamp from the
lamp socket.
u Disconnect the defective lamp.
u Connect the new lamp.
74/94 Operation Manual ProxiMate™
Page 75
BÜCHI Labortechnik AG Cleaning and servicing | 7
u Put the connected new lamp in the
lamp socket.
u Make sure that the screw indicated is
in the groove of the lamp socket.
u Put ferrule and spring on the lamp
socket.
u Press the ferrule. u Turn the ferrule 1/4 turn clockwise.
u Put the seal and the cover lid on the
instrument.
Operation Manual ProxiMate™ 75/94
Page 76
7 | Cleaning and servicing BÜCHI Labortechnik AG
u Attach the cover lid to the instrument.
u Change the desiccant cartridge. See
Chapter7.4 "Changing the desiccant cartridge", page82
u Confirm lamp replacement. See
Chapter7.2.3 "Confirm Lamp Re­placement", page80
u Carry out a baseline correction vector
calibration. See Chapter5.6.3 "Cali­brating a Baseline Correction Vector (BCV)", page24
7.2.2 Changing the up view lamp
u Disconnect the power supply to the
instrument.
u Wait 15 min. for lamp to cool.
u Turn the instrument around. u Make sure that the instrument cannot
tip over during the lamp replacement procedure.
76/94 Operation Manual ProxiMate™
Page 77
BÜCHI Labortechnik AG Cleaning and servicing | 7
u Open the cover lid on the bottom of
the instrument.
u Remove the seal and the cover lid.
u Press the ferrule. u Turn the ferrule 1/4 turn counterclock-
wise.
u Remove the defective lamp from the
lamp socket.
Operation Manual ProxiMate™ 77/94
Page 78
7 | Cleaning and servicing BÜCHI Labortechnik AG
u Disconnect the defective lamp.
u Connect the new lamp.
u Put the connected replacement lamp
in the lamp socket.
u Make sure that the indicated screw is
in the groove of the lamp socket.
u Put ferrule and spring on the lamp
socket.
78/94 Operation Manual ProxiMate™
Page 79
BÜCHI Labortechnik AG Cleaning and servicing | 7
u Press the ferrule. u Turn the ferrule 1/4 turn clockwise.
u Put the seal and the cover lid on the
instrument.
u Attach the cover lid with the tool pro-
vided.
u Change the desiccant cartridge. See
Chapter7.4 "Changing the desiccant cartridge", page82
u Confirm lamp replacement. See
Chapter7.2.3 "Confirm Lamp Re­placement", page80
u Carry out a baseline correction vector
calibration. See Chapter5.6.3 "Cali­brating a Baseline Correction Vector (BCV)", page24
Operation Manual ProxiMate™ 79/94
Page 80
7 | Cleaning and servicing BÜCHI Labortechnik AG
7.2.3 Confirm Lamp Replacement
Navigation path
[Confirm Lamp Replace-
ment]
Precondition:
R The lamp has been changed.
R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
u Tap the On/Off master switch.
ð The system starts up. ð The display highlights the status bar yellow. ð After completion the startup phase the display highlights the status bar black.
u Navigate to the action [Confirm Lamp Replacement] via the navigation path. u Select which lamp was replaced (up view lamp or down view lamp). u Confirm the secure question with OK.
7.3 Checking the desiccant cartridge
Color Describtion
Pink used cartridge (replacement required)
Blue unused cartridge
u Open the cover lid for the desiccant
cartridge.
u Remove the cover lid and the seal.
80/94 Operation Manual ProxiMate™
Page 81
BÜCHI Labortechnik AG Cleaning and servicing | 7
Color Describtion
u Check the color of the window of the
desiccant cartridge.
u If necessary change the desiccant
cartridge. See Chapter7.4 "Changing the desiccant cartridge", page82
u Put seal and cover lid on the instru-
ment.
u Attach the cover lid.
Operation Manual ProxiMate™ 81/94
Page 82
7 | Cleaning and servicing BÜCHI Labortechnik AG
7.4 Changing the desiccant cartridge
u Open the cover lid for the desiccant
cartridge.
u Remove the cover lid and the seal.
u Remove the used desiccant cartridge.
82/94 Operation Manual ProxiMate™
Page 83
BÜCHI Labortechnik AG Cleaning and servicing | 7
u Put a new desiccant cartridge in the
instrument.
u Put seal and cover lid on the instru-
ment.
u Attach the cover lid.
7.5 Carrying out system tests
7.5.1 Carrying out a Baseline Correction Vector test
Navigation path
[Test BCV]
Precondition: R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
Operation Manual ProxiMate™ 83/94
Page 84
7 | Cleaning and servicing BÜCHI Labortechnik AG
u Navigate to the action [Test BCV] via the navigation path.
ð The display shows the dialog Tes t B CV.
u Select the name of the sample presentation you wish to test. u Tap the button [Start]. u Follow instructions on the display during the test. u Confirm the instructions by tapping the [OK] button. u Once the test is complete, the instrument records the results in a report.
ð The display shows Baseline Correction Tests is completed.
7.5.2 Carrying out a Comprehensive System Test
Navigation path
[Extended System Test]
Precondition: R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
R External reference data for all standards are loaded.
R Performance Test Standards Kit is available.
u Navigate to the action [Extended System Test] via the navigation path.
ð The display shows the dialog Ext end ed Sys tem Tes t.
u Select check box [Comprehensive System Test] u Tap the button [Start]. u Follow instructions on the display during the test. u Confirm the instructions by tapping the [OK] button. u Once the test is complete, the instrument records the results in a report.
ð The display shows Comprehensive System Tests is completed.
7.5.3 Carrying out an Advanced System Test
Navigation path
[Extended System Test]
Precondition: R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
u Navigate to the action [Extended System Test] via the navigation path.
ð The display shows the dialog Ext end ed Sys tem Tes t.
u Select check box [Advanced System Test] u Tap the button [Start].
ð The instrument starts the test.
u Once the test is complete, the instrument records the results in a report.
ð The display shows Advanced System Tests is completed.
7.6 Carrying out a data backup
NOTE
The location of the destination folder is fixed. See Chapter10.2 "Folder locations", page89
84/94 Operation Manual ProxiMate™
Page 85
BÜCHI Labortechnik AG Cleaning and servicing | 7
Navigation path
[Backup NIRWise Data]
Precondition: R The instrument is in administrator mode. See Chapter6.5 "Log in administrator
mode", page39
u Navigate to the action [Backup NIRWise Data] via the navigation path.
ð The display shows a dialog with values that can be saved.
u Select the values according to your needs. u Tap the button [Start].
ð The instrument creates a .zip file with the selected data.
u Save the data to an external data store.
7.7 Cleaning the display
The cleaning mode switches off the touch function of the touchscreen.
Navigation path
[Cleaning Mode]
u Navigate to the action [Cleaning Mode] via the navigation path.
ð The touch function of the display is locked. ð The display shows a dialog box with a numeric input box.
u Wipe down the display with a damp cloth. u Enter the displayed secure code. u Tap the button [OK].
ð The dialog box closes. ð The touch function of the display is unlocked.
Operation Manual ProxiMate™ 85/94
Page 86
8 | Help with faults BÜCHI Labortechnik AG
8 Help with faults
8.1 Troubleshooting
Problem Possible cause Action
The sample carrier does not rotate smoothly
Inaccurate Results Direct solar radiation u Make sure, that there is no
Sample presentation area is dirty
u Wipe down the sample
presentation area with mild detergent.
u Rinse the sample presen-
tation area with clean wa­ter.
u Dry the sample presenta-
tion area with a lint free cloth.
direct solar radiation.
The display is black The instrument is in standby Tap the display.
8.2 Error messages
Error code Error message Solution
1000 Unspecified error. u Restart the instrument.
1001 The communication with the in-
strument could not be estab­lished. The configured serial port is {0}.
1003 Instrument data is not available
or is not valid. Check if serial number and instrument options are set.
Sample cup not correctly posi­tioned in sample carrier
u Check that the sample cup
sits correctly in the sample carrier.
Sample cup not filled u Fill the sample cup prior to
measurement.
u Contact BUCHI Costumer
Service.
u Restart the instrument. u Contact BUCHI Costumer
Service.
u Restart the instrument. u Contact BUCHI Costumer
Service.
1004 The '{0}' view dark reference
max value ({1} cnt) is out of the
u Contact BUCHI Costumer
Service.
expected range ({2}..{3} cnt).
1005 The '{0}' view white reference
max value ({1} cnt) is outside
u Contact BUCHI Costumer
Service.
the expected range ({2}..{3} cnt).
1006 The internal reference '{0}' might
not be moving properly.
1007 The current system temperature
({0} °C) is outside the expected range ({1}..{2} °C).
u Contact BUCHI Costumer
Service.
u Move the Instrument to a lo-
cation where the ambient temperature meets specifica­tion.
86/94 Operation Manual ProxiMate™
Page 87
BÜCHI Labortechnik AG Help with faults | 8
Error code Error message Solution
1008 Adjusting the IWR level failed for
view '{0}' (NirTargetSaturation = {1}, NirTargetExposureTime = {2} µs).
u Check operation of source
lamp.
u Contact BUCHI Costumer
Service.
1009 The Peltier temperature ({0} °C)
is out of the expected range ({1}..{2} °C).
1010 The device did not respond
within given timeout of {0} ms for command '{1}'.
1011 The internal white reference sig-
nal is not valid. For more details see log files.
1500 Unknown error occurred while
generating report '{0}': {1}
1501 Unknown error occurred while
exporting report '{0}': {1}
2500 Failed to create NIRWise data
backup. Error: '{0}'.
u Contact BUCHI Costumer
Service.
u Restart the instrument. u Contact BUCHI Costumer
Service.
u Restart the instrument. u Contact BUCHI Costumer
Service.
u Restart the instrument. u Reattempt creating a report. u Contact BUCHI Costumer
Service.
u Restart the instrument. u Reattempt exporting a report. u Contact BUCHI Costumer
Service.
u Restart the instrument. u Reattempt data backup. u Contact BUCHI Costumer
Service.
2502 A critical error occured during
test run. Test run has been can­celed. For more details see log file.
u Restart the instrument. u Reattempt test. u Contact BUCHI Costumer
Service.
Operation Manual ProxiMate™ 87/94
Page 88
9 | Taking out of service and disposal BÜCHI Labortechnik AG
9 Taking out of service and disposal
9.1 Taking out of service
u Switch off the instrument and disconnect it from the mains power supply. u Remove all cables from the device.
9.2 Disposal
The operator is responsible for proper disposal of the instrument.
u When disposing of equipment observe the local regulations and statutory require-
ments regarding waste disposal.
u When disposing, observe the disposal regulations of the materials used. Materials
used see Chapter3.7 "Technical data", page17
9.3 Returning the instrument
Before returning the instrument, contact the BÜCHI Labortechnik AG Service Depart­ment.
https://www.buchi.com/contact
88/94 Operation Manual ProxiMate™
Page 89
BÜCHI Labortechnik AG Appendix | 10
xx.x %
Invalid
Outlier
! xx %
xx.x %
xx.x %
10 Appendix
10.1 Measurement results
Measurement results are displayed after a measurement in the Sta rt menu.
Measurement result Explanation
The tested sample is in the specifications.
The calibration model is missing.
Mahalanobis outlier
The predicted value is outside calibration range.
The predicted value is outside of the set limit.
10.2 Folder locations
NOTE
Hidden folders
By default settings the following folder locations are hidden.
4 Start the software [Windows Explorer] on the instrument. 4 Navigate to folder options via the following navigation path: View Folder Op-
tions View
4 Activate the function [Show hidden files, folders and drives].
Explanation Type Folder
Calibration files .cal C:\ProgramData\BUCHI\NIR-
Data files for calibration .tsv C:\ProgramData\BUCHI\NIR-
Device specific data files for calibration .tsv C:\ProgramData\BUCHI\NIR-
The predicted value is outside the tolerance.
Wise\Calibrations
Wise\Calibrations\Data
Wise\Calibrations\Local
Manually exported files with different content
LIMS system files .xml
Licence request file .xml C:\ProgramData\BUCHI\NIR-
External References .brf C:\ProgramData\BUCHI\NIR-
Operation Manual ProxiMate™ 89/94
di­verse
.csv
C:\ProgramData\BUCHI\NIR­Wise\Export\Data
C:\ProgramData\BUCHI\NIR­Wise\Export\LIMS
Wise\Export\LicenseRequests
Wise\References
Page 90
10 | Appendix BÜCHI Labortechnik AG
Explanation Type Folder
Measurement Reports .xls / .
pdf
C:\ProgramData\BUCHI\NIR­Wise\Reports
System Test Reports .pdf C:\ProgramData\BUCHI\NIR-
Wise\Reports\SystemTests
Measurement report templates .xls C:\ProgramData\BUCHI\NIR-
Wise\Templates
Licences .xml C:\ProgramData\BUCHI\Li-
censeManager\Licences
10.3 Rules entering a formula
Naming conventions for variables
ACII characters onlyUse Underlines between wordsNo numbers at the beginning of a nameNo C# keywordsNo math functions
Calculation conventions
Symbol
Operators Addition +
Subtraction -
Multiplication *
Division /
Math functions Logarithm of x Log(x)
Logarithm of x to a specified base Log(x,base)
Log base 10 of x Log10(x)
X raised to the specified power Pow(x,power)
The square root of x Sqrt(x)
Sine of x Sin(x)
Cosine of x Cos(x)
Absolute value of a double-precision
Abs(x)
floating-point number x
Rounds double-precision floating-point
Round(x)
value x to the nearest integer value
Rounds double-precision floating-point
Round(x,decimal)
value x to a specified number of decimal places
90/94 Operation Manual ProxiMate™
Page 91
BÜCHI Labortechnik AG Appendix | 10
10.4 Spare parts and accessories
Use only genuine BUCHI consumables and spare parts in order to ensure correct, safe and reliable operation of the system.
NOTE
Any modifications of spare parts or assemblies are only allowed with the prior written permission of BUCHI.
10.4.1 Accessories
Order no. Image
External White Reference 11067547
White reference spacer for down view 11067378
White reference spacer for up view 11067377
White reference locating plate 11067391
PET Large Sample Cup 11067692
FDA food approved large sample cup 11068015
Glass petri dishes 10 pcs. (up view) 041583
Plastic petri dishes 240 pcs. (down view) 11066381
Carrier for large deep sample cup plus petri dish 11067691
Carrier for small sample cup (down view) 11065472
Operation Manual ProxiMate™ 91/94
Page 92
10 | Appendix BÜCHI Labortechnik AG
Order no. Image
High Performance Sample Cup 11067399
Robust cup 11055058
Transflectance cover 0.3 mm
Not suitable for use with robust cup
Transflectance cover 2.0 mm
For measurement of crude palm oil. Not suitable for use with robust cup.
041636
11067919
Transflectance cover for robust cup 11055998
Protection cover USB-WiFi stick 11066582
Light shield (down view) 11067281
Performance test standards kit (7 pcs.) 11067545
NIRWise PLUS Chemometrics suite 11068025
10.4.2 Spare parts
Order no. Image
Spare lamp 11065441
Desiccant cartridge 11065467
92/94 Operation Manual ProxiMate™
Page 93
BÜCHI Labortechnik AG Appendix | 10
Order no. Image
Replacement Window HPSC 046246
Operation Manual ProxiMate™ 93/94
Page 94
BUCHI Affiliates:
Europe
Quality in your hands
Switzerland/Austria
BÜCHI Labortechnik AG
CH – 9230 Flawil T +41 71 394 63 63 F +41 71 394 64 64 buchi@buchi.com www.buchi.com
Italy
BUCHI Italia s.r.l.
IT – 20010 Cornaredo (MI) T +39 02 824 50 11 F +39 02 575 12 855 italia@buchi.com www.buchi.com/it-it
America
Brazil
BUCHI Brasil Ltda.
BR – Valinhos SP 13271-200 T +55 19 3849 1201 F +55 19 3849 2907 brasil@buchi.com www.buchi.com/br-pt
Benelux
BÜCHI Labortechnik GmbH
Branch Office Benelux NL – 3342 GT Hendrik-Ido-Ambacht T +31 78 684 94 29 F +31 78 684 94 30 benelux@buchi.com www.buchi.com/bx-en
Russia
BUCHI Russia/CIS
Russia 127287 Moscow T +7 495 36 36 495 russia@buchi.com www.buchi.com/ru-ru
USA/Canada
BUCHI Corporation
US – New Castle, DE 19720 T +1 877 692 8244 (Toll Free) T +1 302 652 300 0 F +1 302 652 8777 us-sales@buchi.com www.buchi.com/us-en
France
BUCHI Sarl
FR – 94656 Rungis Cedex T +33 1 56 70 62 50 F +33 1 46 86 00 31 france@buchi.com www.buchi.com/fr-fr
United Kingdom
BUCHI UK Ltd.
GB – Oldham OL9 9QL T +44 161 633 1000 F +44 161 633 1007 uk@buchi.com www.buchi.com/gb-en
Germany
BÜCHI Labortechnik GmbH
DE – 45127 Essen T +800 414 0 414 0 (Toll Free) T +49 201 747 49 0 F +49 201 747 49 20 deutschland@buchi.com www.buchi.com/de-de
Germany
BÜCHI NIR-Online
DE – 69190 Walldor f T +49 6227 73 26 60 F +49 6227 73 26 70 nir-online@buchi.com www.nir-online.de
Asia
China
BUCHI China
CN – 200233 Shanghai T +86 21 6280 3366 F +86 21 5230 8821 china@buchi.com www.buchi.com/cn-zh
Korea
BUCHI Korea Inc.
KR – Seoul 153-782 T +82 2 6718 7500 F +82 2 6718 7599 korea@buchi.com www.buchi.com/kr-ko
India
BUCHI India Private Ltd.
IN – Mumbai 400 055 T +91 22 667 75400 F +91 22 667 18986 india@buchi.com www.buchi.com/in-en
Malaysia
BUCHI Malaysia Sdn. Bhd.
MY – 47301 Petaling Jaya, Selangor T +60 3 7832 0310 F +60 3 7832 0309 malaysia@buchi.com www.buchi.com/my-en
BUCHI Support Centers:
South East Asia
BUCHI (Thailand) Ltd.
TH-Bangkok 10600 T +66 2 862 08 51 F +66 2 862 08 54 bacc@buchi.com www.buchi.com/th-th
Middle East
BÜCHI Labortechnik AG
UAE – Dubai T +971 4 313 2860 F +971 4 313 2861 middleeast@buchi.com www.buchi.com
Indonesia
PT. BUCHI Indonesia
ID – Tangerang 15321 T +62 21 537 62 16 F +62 21 537 62 17 indonesia@buchi.com www.buchi.com/id-in
Singapore
BUCHI Singapore Pte. Ltd.
SG – Singapore 609919 T +65 6565 1175 F +65 6566 7047 singapore@buchi.com www.buchi.com/sg-en
Latin America
BUCHI Latinoamérica S. de R.L. de C.V.
MX – Mexico City T +52 55 9001 5386 latinoamerica@buchi.com www.buchi.com/es-es
Japan
Nihon BUCHI K.K.
JP – Tokyo 110-00 08 T +81 3 3821 4777 F +81 3 3821 4555 nihon@buchi.com www.buchi.com/jp-ja
Thailand
BUCHI (Thailand) Ltd.
TH – Bangkok 10600 T +66 2 862 08 51 F +66 2 862 08 54 thailand@buchi.com www.buchi.com/th-th
We are represented by more than 100 distribution partners worldwide. Find your local representative at: www.buchi.com
Loading...