BÜCHI Labortechnik AG
Meierseggstrasse 40
Postfach
CH-9230 Flawil 1
E-Mail: quality@buchi.com
BUCHI reserves the right to make changes to the manual as deemed necessary in the light of experience,
especially with respect to structure, illustrations and technical details.
This manual is copyrighted. Information from it may neither be reproduced, distributed, or used for
competitive purposes, nor made available to third parties. The manufacture of any component with the
aid of this manual without prior written agreement is also prohibited.
Büchi Labortechnik AGTable of contents
Table of contents
1About this document........................................................................................................... 6
1.1Mark-ups and symbols ..........................................................................................................6
This operation manual is applicable for all variants of the instrument.
Read this operation manual before operating the instrument and follow the
instructions to ensure safe and trouble-free operation.
Keep this operation manual for later use and pass it on to any subsequent user or
owner.
BÜCHI Labortechnik AG accepts no liability for damage, faults and malfunctions
resulting from not following this operation manual.
If you have any questions after reading this operation manual:
u Contact BÜCHI Labortechnik AG Customer Service.
https://www.buchi.com/contact
NOTE
This symbol draws attention to useful and important information.
R This character draws attention to a requirement that must be met before the
instructions below are carried out.
u This character indicates an instruction that must be carried out by the user.
ð This character indicates the result of a correctly carried out instruction.
Mark-upExplanation
Win dow
Tab
Dia log
[Button]Buttons are marked-up like this.
[Field names]Field names are marked-up like this.
[Menu / Menu item]Menus or menu items are marked-up like this.
StatusStatus is marked-up like this.
SignalSignals are marked-up like this.
1.2Trademarks
Product names and registered or unregistered trademarks that are used in this
document are used only for identification and remain the property of the owner in
each case.
Software Windows are marked-up like this.
Tabs are marked-up like this.
Dialogs are marked-up like this.
6/46
1.3Connected devices
In addition to these operating instructions, follow the instructions and specifications
in the documentation for the connected devices.
Operation Manual NIRMaster™
Büchi Labortechnik AG
2Safety
2.1Proper use
2.2Use other than intended
Safety | 2
The instrument is designed and built for laboratories and production environments
(at-line). It serves to determine the concentration of selected constituents contained
within a substance.
The instrument can be used for the following tasks:
Determination of quantifiable product properties.
Use of any kind other than that described in Chapter2.1 "Proper use", page7 and
any application that does not comply with the technical specifications (see
Chapter3.6 "Technical data", page15) constitutes use other than that intended.
In particular, the following applications are not permissible:
Use of the instrument in rooms which require Ex-protected instruments.
Use of samples, which can explode or inflame (example: explosives, etc.) due to
shock, friction, heat or spark formation.
2.3Staff qualification
Unqualified persons are unable to identify risks and are therefore exposed to greater
dangers.
The device may only be operated by suitably qualified laboratory staff.
These operating instructions are aimed at the following target groups:
Users
Users are persons that meet the following criteria:
They have been instructed in the use of the device.
They are familiar with the contents of these operating instructions and the
applicable safety regulations and apply them.
They are able on the basis of their training or professional experience to assess
the risks associated with the use of the device.
Operator
The operator (generally the laboratory manager) is responsible for the following
aspects:
The device must be correctly installed, commissioned, operated and serviced.
Only suitably qualified staff may be assigned the task of performing the
operations described in these operating instructions.
The staff must comply with the local applicable requirements and regulations for
safe and hazard-conscious working practices.
Safety-related incidents that occur while using the device should be reported to
the manufacturer (quality@buchi.com).
Operation Manual NIRMaster™
7/46
2 | Safety
Büchi Labortechnik AG
BUCHI service technicians
Service technicians authorized by BUCHI have attended special training courses and
are authorized by BÜCHI Labortechnik AG to carry out special servicing and repair
measures.
2.4Warning notices in this document
Warning notices warn you of dangers that can occur when handling the device.
There are four danger levels, each identifiable by the signal word used.
Signal wordMeaning
DANGERIndicates a danger with a high level of risk which could result in
death or serious injury if not prevented.
WARNINGIndicates a danger with a medium level of risk which could result in
death or serious injury if not prevented.
CAUTIONIndicates a danger with a low level of risk which could result in mi-
nor or medium-severity injury if not prevented.
NOTICEIndicates a danger that could result in damage to property.
2.5Warning and directive symbols
The following warning symbols are displayed in this operation manual or on the
instrument.
SymbolMeaning
General warning
Dangerous electrical voltage
Laser beam
Laser class 1
Avoid exposure to laser beam
Hot surface
8/46
Device damage
Breakable items
Operation Manual NIRMaster™
Büchi Labortechnik AG
2.6Residual risks
Safety | 2
SymbolMeaning
Magnetic field
Unplug device
Wear safety goggles
Wear protective gloves
Heavy weight, lifting requires more than one person
The instrument has been developed and manufactured using the latest technological
advances. Nevertheless, risks to persons, property or the environment can arise if
the instrument is used incorrectly.
Appropriate warnings in this manual serve to alert the user to these residual
dangers.
2.6.1Glass and acrylic breakage
Broken glass and acrylic can cause severe cuts.
Broken glass or acrylic can enter production.
u Handle the Petri Dishes and other glass and acrylic components carefully and do
not drop them.
u Always visually inspect glass and acrylic components for damage every time they
are to be used.
u Do not continue to use glass and acrylic components that are damaged.
u Always wear protective gloves when disposing of broken glass and acrylic.
2.6.2Faults during operation
If a device is damaged, sharp edges or exposed electrical wires can cause injuries.
u Regularly check device for visible damage.
u If faults occur, switch off the device immediately and inform the operator.
u Do not continue to use devices that are damaged.
2.6.3Data loss
In the event of a power failure, e.g. due to lightning or interruption of power supply,
measurement data may be lost.
u Carry out regular data backup.
2.7Personal protective equipment
Depending on the application, hazards due to heat and/or corrosive chemicals may
arise.
Operation Manual NIRMaster™
9/46
2 | Safety
Büchi Labortechnik AG
u Always wear appropriate personal protective equipment such as safety goggles,
protective clothing and gloves.
u Make sure that the personal protective equipment meets the requirements of
the safety data sheets for all chemicals used.
2.8Modifications
Unauthorized modifications can effect safety and lead to accidents.
u Use only genuine BUCHI accessories, spare parts and consumables.
u Carry out technical changes only with prior written approval from BUCHI.
u Only allow changes to be made by BUCHI service technicians.
BUCHI accepts no liability for damage, faults and malfunctions resulting from
unauthorized modifications.
10/46
Operation Manual NIRMaster™
Büchi Labortechnik AG
2
1
4
6
3Product description
3.1Description of function
Product description | 3
The instrument is a benchtop Fourier Transformation Near Infrared spectrometer
(FT-NIR) to identify and quantify analytes of interest. It generates an invisible near
infrared interferogram beam which interacts with the mole- cules of a sample,
generating a characteristic feedback. The feedback is picked up via a measurement
cell by a detector and mathematically processed via Fourier transformation into a
spectrum. This spectrum is charac- teristic for a given sample and allows
identification as well as quantification of its components.
How the interferogram is generated
An interferogram is an interference pattern of phase-shifted beams. The instrument
is a single-beam polarization interferometer, generating its interferogram in four
steps:
1. Polarization of the light source output
The polarizer
polarized light, emitted by the light source
generates a well-defined polarization output of the undefined
. Thus, only diagonally polarized
light is transmitted.
2. Beam splitting and orthogonal polarization
The polarized light enters a double refracting block (comparator) 3. Here, the
light is broken down into two, orthogonally polarized components with a small,
static phase shift.
3. Generating the ongoing phase shift
An assembly of two double refracting wedges is arranged after the comparator.
Wedge
is stationary, while wedge 5 is constantly shifted back and forwards
by a fast linear-drive. The movement and the geometric arrangement provides a
change of thickness in the light path. This leads to an ongoing phase shift
between the light beams.
4. Beam recombination and interferogram output
A second polarizer
converts the phase shifted beams into a single light output
with intensity variation – the interferogram
Operation Manual NIRMaster™
11/46
3 | Product description
875432
1
1
2
6
4
3
6
5
LASER – 633 nm
Spare lamp
1
2
3
4
5
6
7
8
1
2
3
4
5
6
3.1.1Effective NIR light path
Büchi Labortechnik AG
Number Explanation
NIR light source assembly with spare lamp and motorized parabolic reflector
First polarizer
Comparator (double refracting block)
Stationary double refracting wedge
Moving double refracting wedge
Second polarizer
Sample
Interferogram detector
3.1.2Reference laser signal path
Number Explanation
Laser output window
Second polarizator (works as a beam splitter for the laser)
Laser output-power sensor
First polarizer
Laser feedback detector
Linear motor for wedge movement
3.1.3NIRWare software suite
NIRWare is the interface program suite between the instrument and the operator.
All program parts need to be hosted on a separate PC.
12/46
Operation Manual NIRMaster™
Büchi Labortechnik AG
1
3
2
1
2
3
Product description | 3
All measurements are performed using the NIRWare Operator. The NIRWare
Operator includes the following program elements by default:
Operator interface(s)
Administrative tools
Report templates
Software to run routine analysis
Comprehensive database with analysis data and other
Advanced software and system management is done with the NIRWare
Management Console tool.
The NIRWare Management Console combines various software modules
NIRWare Application Designer to define NIRWare Operator applications.
NIRWare Sample Manager to administrate all samples and reference values.
NIRWare Administrative Tools for applications and calibrations interchange and
other administrative tasks.
NIRWare Security Designer to define users and user groups according to custom
security policies.
The BUCHI Database Manager is used for any database operations.
Optional:.
NIRWare LIMS to import and export sample information and measurement data.
NIRWare Library Designer is a powerful software module for substance identity
control using full spectral comparison. It is designed for the development of
spectral libraries tailored to individual user requirements
To connect the instrument with a PC see Chapter5.5 "Setting up the network
connection between Computer and instrument", page23.
3.2Measurement modes
3.2.1Diffuse reflection mode
Non-translucent materials can be analyzed via diffuse reflection.
NIR light penetration is limited by the
sample material. It interacts with the
sample, is refracted and diffusely reflected into the sensor. The reflected rays
contain the spectral information of the
sample.
Sample
Sensor
Light
Operation Manual NIRMaster™
13/46
3 | Product description
2
1
4
3
1
2
3
4
Primary lamp
Internal fuses
Interferometer
Interferometer
interface panel
LASER assembly
Cable
USB Connector
ON / OFF button
Add-on Connector
feedthrough
3.2.2Transflectance mode
3.3Configuration
3.3.1NIRMaster instrument
Büchi Labortechnik AG
Translucent and opaque liquids can be
analyzed via transflectance mode. The
light penetrates the liquid, is diffusely reflected by the reference plate and passes
through the sample a second time. The
transflected rays contain the spectral information of the sample.
Transflectance cover
Sample cup
Sensor
Light
14/46
3.4Scope of delivery
Fig.1: Shows the main components of the instrument at open housing.
NOTE
The scope of delivery depends of the configuration of the purchase order.
Accessories are delivered as per the purchase order, order confirmation, and
delivery note.
Operation Manual NIRMaster™
Büchi Labortechnik AG
6
7
8
11
9
10
3
2
1
5
4
3.5Type plate
Product description | 3
The type plate identifies the instrument. The type plate is located at the rear of the
instrument.
Fig.2: Type plate
1Symbol for "electronics recycling"2Initial product code
3Approvals4Symbol for "Do not dispose of as
household waste"
5Year of manufacture6Power consumption maximum
7Frequency8Input voltage range
9Serial number10 Instrument name
11 Company name and address
3.6Technical data
3.6.1NIRMaster™ Essential
SpecificationNIRMaster™ Essential
Dimensions (W x D x H)420 x 560 x 360 mm
Weight43 kg
Minimum clearance on all sides150 mm
Connection voltage100 – 240 VAC ± 10 %
Frequency50 / 60 Hz
Power consumption150 W
IP Code54
Internal fuse4A / 240V
Type of interferometerPolarisation interferometer with
Spectral range800 – 2500 nm
Resolution
Operation Manual NIRMaster™
(with boxcar apodization)
Wavenumber accuracy
(at an ambient temperature of 25°C ± 5°C)
TeO2 wedges
12 500 – 4000 cm–1
8 cm–1
± 0.2 cm–1
15/46
3 | Product description
Büchi Labortechnik AG
SpecificationNIRMaster™ Essential
Signal-to-noise ratio10000
(peak-to-peak noise of a linear
corrected base- line between
5600−6000 cm-1)
Number of scans/sec.2 - 4
Illumination spot diameter9 mm
Analog digital converter24 bit
Lamp typeTungsten-halogen
Average life
(lamp)
Type of laser12 VDC HeNe, wavelength at
Connection voltage100 – 240 VAC ± 10 %
DetectorExtended range InGaAs (temper-
Internal referenceGold plate
External referenceSpectralon®
Sample rotationEasy-spin drive, a magnetic spin-
3.6.2Ambient conditions
For indoor use only.
Max. altitude above sea level2000 m
Ambient temperature5 - 40 °C (25 °C)
6000 h
632.992 nm
ature controlled)
ner is included for high performance measurement of inhomogeneous sample
16/46
Maximum relative humidity80 % for temperatures up to 31
The installation site allows that the power supply can be disconnected at any time
in an emergency.
The installation site has a firm, level and nonslip surface.
The installation site has a fume hood.
The installation site has an own mains outlet socket for the instrument.
The installation site has enough space that cables / tubes can be routed safely.
The installation site has no obstacles (e.g. water taps, drains, etc.).
The installation site is not exposed to external thermal loads, such as direct solar
radiation.
The installation site is not located near vibration-sensitive devices.
The installation site meets the requirements for the connected devices. See
related documentation
The installation site meets the requirements of the safety data sheets for all
solvents and samples used.
The installation site meets the safety requirements. See Chapter2 "Safety",
page7
The installation site meets the specifications according to the technical data (e.g.
weight, dimension, etc.). See Chapter3.6 "Technical data", page15
Operation Manual NIRMaster™
17/46
4 | Transport and storage
4Transport and storage
4.1Transport
4.2Storage
Büchi Labortechnik AG
NOTICE
Risk of breakage due to incorrect transportation
u Make sure that all parts of the device are safely packed in such a way as to
prevent breakage, ideally in the original box.
u Avoid sharp movements during transit.
u After transportation, check the device for damage.
u Damage that has occurred in transit should be reported to the carrier.
u Keep packing for future transportation.
u Make sure that the ambient conditions are complied with (see Chapter3.6
"Technical data", page15).
u Wherever possible, store the device in its original packaging.
u After storage, check the device for damage and replace if necessary.
4.3Lifting the instrument
WARNING
Danger due to incorrect transportation
The possible consequences are crushing injuries, cuts and breakages.
u The instrument should be transported by two persons at the same time.
u Lift the instrument at the points indicated.
18/46
Operation Manual NIRMaster™
Büchi Labortechnik AG
Locked
Locked
4.4Lock / unlock the transportation lock
Transport and storage | 4
NOTICE
Risk of instrument damage at use with engaged transportation lock.
u Disengage the transportation lock prior use.
u Switch off the instrument.
u Unplug the power cord.
u Unlock the housing.
u Lift off the top cover.
ð The locking mechanism is accessible.
Operation Manual NIRMaster™
u Turn the knob in clockwise direction.
ð The locking mechanism is unlocked and the linear drive of the interferometer is
freely moveable.
19/46
4 | Transport and storage
Locked
Büchi Labortechnik AG
u Close the top cover.
u Lock all four latches.
u Plug in the power cord.
u Switch on the instrument.
20/46
Operation Manual NIRMaster™
Büchi Labortechnik AG
5Installation
5.1Before installation
5.1.1Electrical connections requirement
Installation | 5
NOTICE
Instrument damage due to switching it on too early.
Switching on the instrument too early after transportation can cause damage.
u Climatize the instrument after transportation.
NOTICE
For safe servicing the instrument it needs to be disconnect from the mains power
supply.
Precondition:
R The instrument is installed with a locked or secured plug.
R The instrument is installed stationary (without power plug).
u An external power switch is required.
5.2Establishing electrical connections
NOTICE
Risk of instrument damage because of not suitable power supply cables.
Not suitable power supply cables can cause bad performance or an instrument
damage
u Use only BUCHI power supply cables.
Precondition:
R The electrical installation is as specified on the
type plate.
R The electrical installation is equipped with a
proper grounding system.
R The electrical installation is equipped with
suitable fuses and electrical safety features.
R The installation site is as specified in the
technical date. See Chapter3.6 "Technical
data", page15
u Connect the power supply cable to the
connection on the instrument. See Chapter3.3
"Configuration", page14
Operation Manual NIRMaster™
21/46
5 | Installation
Büchi Labortechnik AG
u Connect the mains plug to an own mains outlet
socket.
5.2.1Connect the mains cable
Precondition:
R The cover is removed.
u Route the mains cable through the cable gland in the lower chassis.
u Connect the cable to the main socket at
mounted on the rear of the instrument lower
chassis.
5.3Connect the instrument to the PC using the supplied ethernet
cable
u Remove the cover from the top of the
instrument with the key provided. (Two fixings
on the left hand side, two fixings on the right
hand side)
u Route the ethernet cable through the cable gland in the lower chassis.
u Connect the cable to the ethernet socket on the
rear of the cream colored optical bench.
5.4Install the BUCHI NIRWare software on the computer
Install the BUCHI NIRWare software on the computer
Install BUCHI NIRWare software suite following the instruction in Chapter 4
“Software Installation” of the NIRWare Quick Guide.
22/46
Note!
Activate the relevant licences for software and any pre-calibrations that are
required.
Operation Manual NIRMaster™
Büchi Labortechnik AG
5.5Setting up the network connection between Computer and
Installation | 5
instrument
u Start the computer, and open the dialog box
Pro pert ies
u Highlight the option [Internet Protocol Version 4 (TCP/
.
Et hern et
IPv4)] from the list.
u Click the button [Properties].
ð The dialog box
IPv 4)
u Activate the radio button next to [Use the following IP
open.
Int erne t P rotocol Ve rsio n 4 (TC P/
address].
u Enter the following values.
IP address: 192.168.1.2
Subnet mask: 255.255.255.0
u Click the button [OK].
ð The dialog box
close.
IPv 4)
u Click the button [OK].
ð The dialog box
Int erne t P rotocol Ve rsio n 4 (TC P/
Eth erne t P roperti es
close.
The network settings are now correctly set up so communication between the instrument and computer is possible.
Start NIRWare operator software and verify that the Computer is correctly connected to the instrument.
Operation Manual NIRMaster™
23/46
6 | Operation
Büchi Labortechnik AG
6Operation
6.1Initializing the instrument
u To start the instrument press the green power button on the left side of the
instrument for 1 – 3 seconds.
ð The indicator light at the housing front will light up yellow, to indicate the
bootup process.
u Switch on the PC with the installed NIRWare software suite.
ð The instrument is ready to log in when the indicator light will change from yellow
to green, after approximately 60 seconds.
6.1.1System warm-up
When cooled down below working temperature, the spectrometer needs a warm-up
time of approx. 15 min. The warm-up time starts as soon as the instrument is
switched on.NOTICE!Insufficient warm-up time can influence the quality of
measurement. A longer warm-up time might be required at low environmental
temperature.
6.2System Suitability Test
NOTE
Frequent tests should be performed to minimize the risk of erroneous
measurements.
The System Suitability Test (SST) serves as an internal functionality and performance
test of the spectrometer. A SST has do be done after each restart of the instrument
to monitor and verify the system performance.
The following SST values are tested:
Internal system temperatures (e.g. Laser, interferometer, detectors)
Linearity
Signal-to-noise ratio
Wavenumber stability
At the end of each SST a report protocol will be generated. The test results are
compared to values ex factory and inform the user about the system condition.
The report protocol is stored in the internal database for service and quality control
in XML-format..
The time between two SST requests can be set with the NIRWare Application
Designer. The recommended standard value between two tests is 24 hours, a testrun takes approximately 5 min. If the instrument is used in continuous operation, a
longer checking interval might be suitable.
24/46
6.2.1Temperature tests
The instrument is equipped with a number of sensors to monitor internal
temperatures. Correct operation of the instrument is ensured when the measured
internal temperatures are within specified ranges.
u Check if all actual temperature values are within the limits.
Operation Manual NIRMaster™
Büchi Labortechnik AG
6.2.2NIR linearity test
6.2.3Signal-to-noise ratio determination
Operation | 6
u Mind the ambient temperature data as listed in Chapter3.6.2 "Ambient
conditions", page16
Precondition:
R If temperature is too low:
u Let system warm up sufficiently.
Precondition:
R If temperature is too high:
u Clean filters and provide sufficient airflow.
The linearity of the NIR detector is tested by presenting five filters with known
absorbance to the instrument. The measured absorbance is compared to the known
reference value.
To pass this test, all absorbance factors must be within the tolerance.
With this test the ratio of signal power to noise power is determined. A high flux
(maximum intensity) and a low flux (10 % intensity) beam is being analyzed for a
number of wavenumber intervals.
To pass these test, the average and the minimum values must be within the
tolerance.
6.2.4Wavenumber stability test
The wavenumber stability of the instrument is checked by presenting a PMMA filter
to the instrument. The measured spectra is compared to a known reference spectra.
To pass the test the measured spectra should be within the wavenumber tolerance
limits of the reference spectra.
6.3Reference measurements
Frequent reference measurements compensate for environmental effects, which
might influence the spectra. A combination of internal and external reference
measurements provides best result quality and measurement stability.
6.3.1External reference
NOTE
External referencing needs user interaction.
In order to measure the reflectance of a sample, the measured intensity spectrum of
the sample is divided by a reference spectrum. The reference spectrum is generated
by a white standard in the sample position.
See Chapter6.3.3 "Performing a reference measurement", page26.
6.3.2Internal reference
NOTE
Internal referencing is performed by the instrument automatically.
Operation Manual NIRMaster™
25/46
6 | Operation
1
2
Büchi Labortechnik AG
The internal referencing is used to minimize the need for external referencing and
user interaction. The interval between two tests can be set via the NIRWare
Application Designer.
6.3.3Performing a reference measurement
NOTE
For proper device calibration, the external reference or the transflectance adapter
must be placed plane on top of the sample window.
Could otherwise lead to incorrect calibration.
u Do not tilt the external reference / transflectance adapter.
u The external reference must be aligned to match the cut-out of the centering
plate on top of the instrument.
u When working with the transflectance adapter, the spinner device is a
recommended accessory.
u All items / surfaces must be absolutely clean.
u The external reference and the transflectance adapter must be damage and
scratch free.
After the expiration of a predefined time interval (60 min. by default), a dialog (1)
will remind the operator to perform an external reference measurement. The
interval can be set in the Application Designer.
The software guides the user through the process:
When the external reference (2) has been positioned properly, press the
‘Checkmark’ button. The instrument will automatically measure the reference
spectrum.
26/46
Operation Manual NIRMaster™
Büchi Labortechnik AG
Operation | 6
External reference positioning with installed spinner:
External reference positioning for XL addon (without spinner):
For proper device calibration, the external reference must be placed plane on top of
the sample window of the instrument. It is not necessary to remove the spinner. In
case the spinner is installed, turn the external reference until it fits perfectly.
Transflectance adapter positioning with / without spinner:
Operation Manual NIRMaster™
27/46
6 | Operation
3
Büchi Labortechnik AG
When the measurement is done (3), the user will be requested to remove the
external reference.
Press the ‘Checkmark’ button to finish the procedure.
Frequency of reference measurement
An additional external reference measurement is mandatory:
after an application change
after a lamp exchange and lamp failure
after a service (when the top cover has been reinstalled and locked)
28/46
when changing the add-on or measurement setup
at every system start / restart
6.4How to perform a measurement
Detailed instructions how to perform a measurement can be found in chapter 5
"Creating a qualitative and a quantitative application" of the NIRWare Quick Guide.
Operation Manual NIRMaster™
Büchi Labortechnik AG
6.5Shutting down the instrument
Operation | 6
u Close the NIRWare program on the separate PC.
u Press the ON / OFF button on the left side of the instrument for 1 – 3 seconds to
switch it off.
Operation Manual NIRMaster™
29/46
7 | Cleaning and servicing
7Cleaning and servicing
7.1Regular maintenance work
Büchi Labortechnik AG
NOTE
Users may only carry out the servicing and cleaning operations described in this
section.
Any servicing and repair work which involves opening up the casing may only be
carried out by BUCHI service technicians.
Action
Daily
Weekly
Additional information
Yearly
7.2Cleaning the housing1Clean the surfaces in direct
contact with the sample after each measurement in
order to avoid contamination.
7.8Cleaning the add-ons and
1After each measurement
the custom sample vessels
7.6Cleaning the external refer-
1
ence
7.7Cleaning the spinner1
7.4Exchanging the outer filters1 Check if the filter is clogged
and replace latest every
6'000 h
7.5Exchanging the inner filters1 Check if the filter is clogged
and replace latest every
6'000 h
7.10Exchanging the lamp1 When defective, latest every 6'000 h
7.12Exchanging the laser unit2 When defective, latest every 20'000 hours
7.13Extended system service
2
and testing
30/46
1 - User; 2 - BUCHI service technicians
7.2Cleaning the housing
NOTE
Avoid scratches on optical surfaces due to cleaning.
Scratches can influence the measurement precision.
u Never use any objects for cleaning which might scratch the optical surfaces.
u Switch off the instrument and all directly connected electronic devices.
u Clean the housing with a soft, lint free damp cloth with mild soapy water.
Operation Manual NIRMaster™
Büchi Labortechnik AG
Cleaning and servicing | 7
u In addition, use surface disinfectant wipes in combination with similar
disinfectant spray on alcohol basis.
u Clean the sample window on top of the instrument.
u Use an antimicrobial wipe.
u Avoid fingerprints on any optical surface.
u For coated optical surfaces, use abrasion-resistant precision cloths.
7.3How to lift off the top cover
NOTICE
Ingress of particles.
Risk of instrument damage.
u Check proper seating and reliable locking before use.
u Do not operate instrument when housing is open or unlocked.
To get access to the light source and the interferometer it is necessary to remove
the top cover of the instrument.
Operation Manual NIRMaster™
31/46
7 | Cleaning and servicing
1
Büchi Labortechnik AG
u Switch off the instrument.
u Disconnect the external power supply.
u Clean the housing i.e. remove any dust between the upper and lower cover.
u Lift the four black dust flaps (1) two at each side of the instrument.
ð The latches are accessible.
u Turn the housing-key and bring the latches into horizontal position to unlock the
housing.
u Carefully lift off the top cover.
u Unplug the internal power cable.
u To reinstall the cover, re-plug the internal power cable.
u Carefully place the top cover onto the instrument.
u Turn the housing-key and bring the latches into vertical position to lock the
housing.
u Put the four black dust flaps (1) back on.
u Connect the external power supply to the instrument.
u Switch on the instrument.
32/46
Operation Manual NIRMaster™
Büchi Labortechnik AG
Cleaning and servicing | 7
7.4Exchanging the outer filters
u Pull off the filter frames located at the bottom. NOTICE!Do not tilt the
instrument. The access is possible from the side.
Operation Manual NIRMaster™
33/46
7 | Cleaning and servicing
1
3
2
7.5Exchanging the inner filters
Büchi Labortechnik AG
u Take out each filter pad.
u Insert new filter pads.
u Reassemble the filter frames.
u Make sure the ventilation slots of the filter frames show into opposite direction
to avoid a shortcut of the cooling airflow.
34/46
u Lift off the top cover.
ð The filter drawers are accessible below the fans.
u Push the drawer (1).
ð The filter holder moves out of its position.
u Pop out the sandwich filter.
u For each filter, place a new filter pad between the metal grids (2).
u Put the filters back in its holders.
u Push the drawers back into position (3).
u Close the housing.
7.6Cleaning the external reference
NOTE
Do not use any cleaning media with abrasive character which might scratch the
surface.
u For cleaning use abrasion-resistant precision cloths.
Operation Manual NIRMaster™
Büchi Labortechnik AG
7.7Cleaning the spinner
Cleaning and servicing | 7
u If substances still adhere, clean the surfaces with alcohol.
u Avoid contact with hard objects.
u Do not drop the external reference.
u Keep it in a dry and clean place.
u The spinner can be cleaned in a dishwasher. NOTICE!Not cleaning the spinner
on a regular basis can lead to a reduced force transmission of the inner wheel
and inlay. As a result the sample cup will not rotate.
u The O-ring of the inner wheel needs replacement in case its mechanical treated.
u Replacement is recommended by a service person.
7.8Cleaning the add-ons and the custom sample vessels
u All BUCHI add-ons can be rinsed and cleaned with disinfectants on alcohol basis.
NOTICE!Spray-disinfection should be used in addition to surface disinfection
with wipes only.
u The following sample vessels can be cleaned in a dishwasher with standard
detergents:
Petri dishes
Unbreakable cup
Spinner cpl.
XL add-on
High performance cup
7.9How to reset the system fuses
NOTICE
Device fuses may be triggered occasionally (e.g. due to voltage peaks).
When fuses trigger frequently action needs to be taken.
u Inform your customer service.
u Do not try to repair or modify the fuses.
Operation Manual NIRMaster™
35/46
7 | Cleaning and servicing
Fuse ok
Fuse triggered
Büchi Labortechnik AG
Precondition:
R The primary fuses are located inside the instrument next to the interferometer
power supply socket.
u Switch off the instrument.
u To get access to the resettable fuses, lift off the top cover as described in
Chapter7.3 "How to lift off the top cover", page31.
u Push in the fuse for reset.
u Switch on the instrument.
7.10Exchanging the lamp
CAUTION
Risk of minor burns by hot lamp module.
u Do not touch hot parts or surfaces.
u Let the lamp module cool down.
u Do not touch the bulb glass.
NOTE
After lamp-module exchange, a reference spectrum should be generated to match
the individual light intensity output of the module. See Chapter6.3 "Reference
measurements", page25.
36/46
Operation Manual NIRMaster™
Büchi Labortechnik AG
1
2
3
4
5
Cleaning and servicing | 7
u Switch off the instrument.
u Lift off the top cover (1) and put it aside.
u Unplug the mains plug inside the device (2).
u Unplug the power cable (3) of the primary / secondary lamp module.
u Turn the locking lever upwards (4) to unlock the module.
u Carefully pull out the lamp (5).
u Put the new lamp in.
u Turn the locking lever downwards to lock the module.
u Connect the power cable of the primary / secondary lamp module.
u Connect the mains plug inside the device.
u Install the top cover again.
u Switch on the instrument
u Reset the lamp lifetime counter using the separate PC as described in chapter
7.8.4. of the NIRWare Operation Manual.
7.11Resetting the lamp lifetime counter
To correctly monitor the lamp lifetime the software hour counter has to be reset
after each lamp replacement. See chapter 7.8.4. of the NIRWare Operation Manual.
7.12Exchanging the laser unit
NOTE
A BUCHI service technician is required to replace the LASER unit.
Operation Manual NIRMaster™
37/46
7 | Cleaning and servicing
7.13Extended system service and testing
Büchi Labortechnik AG
Over its lifetime, the intensity of the laser light will decrease. It is continuously
monitored by a sensor to be sure levels are above a certain threshold.
Contact BUCHI customer service for the annual extended service and testing.
38/46
Operation Manual NIRMaster™
Büchi Labortechnik AG
8Help with faults
8.1Troubleshooting
Help with faults | 8
ProblemPossible causeAction
System is overheated
System does not
power up
External reference
measurement fails
Sample cup does
not rotate
Incorrect measurements results
Filter cloggedu Exchange outer / inner
filters.
Power outageu Make sure that the
instrument has power.
Built-in resettable fuses or
electronic malfunction
u Reset internal fuses.
u If error persists, call service
for repair.
See Chapter8.3 "Reference
measurement fails", page40
u Clean all optical surfaces in
the measurement path.
u Perform NADIA Chapter8.2
"Enhanced system diagnose
(NADIA)", page39.
Spinner contaminated with oil
u Clean spinner in dishwasher.
or fat
Not enough sample material
available
Sample not homogenious or
representative
u Use sufficient sample
material.
u Take a representative
sample.
u Mix the sample before
measurement.
Humidity of sample materialu Only use dry samples.
Temperature of sample material
Sample cup inadeqateu Only use from BUCHI
Malfunctions or errors which are not listed in the table are to be corrected by a
BUCHI technical engineer. In this case please refer to your local BUCHI customer
service agent.
8.2Enhanced system diagnose (NADIA)
Additional tests to the SST can be run via the NIRWare Automatic Diagnose (NADIA)
tool. It can be used to verify the condition of the instrument e.g. during service for
an enhanced performance check.
To run NADIA see chapter 5.9 in the NIRWare Operations Manual.
u Wait until the sample
reaches ambient
temperature.
recommended sample cups.
See Chapter10.1.2
"Accessories", page43
Operation Manual NIRMaster™
39/46
8 | Help with faults
8.3Reference measurement fails
Büchi Labortechnik AG
When the deviation between the stored reference spectrum is out of limit, the
reference measurement will fail. A dialog will show both curves and give details of
the measurement result.
Possible reasons for a reference measurement to fail:
Optical surfaces (e.g. external reference) are dirty or scratched
Environmental conditions out of operational range (e.g. condensating humidity at
optical surfaces)
Defective instrument (e.g. electronic damage)
After lamp-module exchange
What to do:
u Clean the external reference and the instrument output window.
u Place the external reference onto the output window and press the ‘Rerun’
button to repeat the measurement.
u If measurement still fails, a new reference spectrum can be stored by pressing
the ‘Checkmark’ button. NOTICE!An administrator login is required to store the
new data set. Setting a new reference spectrum should be the last option after
consequent trouble shooting (run SST).
40/46
Operation Manual NIRMaster™
Büchi Labortechnik AG
9Taking out of service and disposal
9.1Taking out of service
9.2Disposal
9.3Returning the instrument
Taking out of service and disposal | 9
u Switch off the instrument and disconnect it from the mains power supply.
u Remove all cables from the device.
The operator is responsible for proper disposal of the instrument.
u When disposing of equipment observe the local regulations and statutory
requirements regarding waste disposal.
u When disposing, observe the disposal regulations of the materials used. Materials
used see Chapter3.6 "Technical data", page15.
Before returning the instrument, contact the BÜCHI Labortechnik AG Service
Department.
https://www.buchi.com/contact
Operation Manual NIRMaster™
41/46
10 | Appendix
Büchi Labortechnik AG
10Appendix
10.1Spare parts and accessories
Use only genuine BUCHI consumables and spare parts in order to ensure correct,
safe and reliable operation of the system.
NOTE
Any modifications of spare parts or assemblies are only allowed with the prior
written permission of BUCHI.
10.1.1Spare parts
Order no.Image
Spinner11056771
External reference11056371
Outer filter holder with pads, 2 pcs11056672
Set spare filter pads 10 pcs11056773
Inner filter pads, 10 pcs11056917
Ferrit sleeve 5 mm11056714
Ferrit sleeve 6 mm11056715
Housing-key11056766
Plug bushing 3-6mm 10 pcs11056665
42/46
Plug bushing 6-9mm 10 pcs11056666
Operation Manual NIRMaster™
Büchi Labortechnik AG
Appendix | 10
Order no.Image
Set round cord 10 pcs11056667
Spinner (inlay-) ring11056772
Replacement Window HPSC046246
Primary lamp moduleN560-001
Adapter plate for gold cell
For both High Performance Transflectance Cells 11056994
or 11056995.
10.1.2Accessories
Compatibility
Sample holding accessoriesEasy-Spin
Petri dishes•
High performance cup•
Unbreakable cup•
Small plastic bags•
High performance transflectance gold cell
Sample dimensions100 mmAdaptable35 mm
*Included in system as described in Chapter3.4 "Scope of delivery", page14.
**See Chapter3.2 "Measurement modes", page13.
Magnetic
spinner*
11056793
XL* glass spacer Adapter plate
for gold cell
•
High performance cup046259
Operation Manual NIRMaster™
Order no.Image
43/46
10 | Appendix
Büchi Labortechnik AG
Order no.Image
Robust cup11055058
Glass petri dishes 10 pcs. (up view)
Not suitable for use with Transflectance Cover
Transflectance cover 0.3 mm
Not suitable for use with robust cup
11072073
041636
Transflectance cover for robust cup11055998
Pressing Stamp
To compress and compact samples like ground meat or
cheese. Stainless steel, approx. 1.8 kg.
11057584
XL-Add-On11056784
44/46
Operation Manual NIRMaster™
BUCHI Affiliates:
Europe
Quality in your hands
Switzerland/Austria
BÜCHI Labortechnik AG
CH – 9230 Flawil
T +41 71 394 63 63
F +41 71 394 64 64
buchi@buchi.com
www.buchi.com
Italy
BUCHI Italia s.r.l.
IT – 20010 Cornaredo (MI)
T +39 02 824 50 11
F +39 02 575 12 855
italia@buchi.com
www.buchi.com/it-it
NIR-Online
Germany
NIR-Online GmbH
DE – 69190 Walldor f
T +49 6227 73 26 60
F +49 6227 73 26 70
nir-online@buchi.com
www.nir-online.de