Bio-Rad Gene Pulser Electroporation Buffer User Manual

Gene Pulser®Electroporation Buffer Reagent
Catalog # Volume 165-2676 10 x 1.8 ml 165-2677 30.0 ml
For research only. Store at 4°C.
Storage and Stability
.
Handling Precautions
Gene Pulser electroporation buffer reagent is intended for the delivery of siRNAs and plasmid DNA into cultured mammalian cells. It is nuclease-free to prevent the degradation of siRNAs or plasmid DNA. Use gloves and nuclease-free reagents, pipets, and tubes, as well as good tissue culture practices.
Contents
Electroporation buffer contains 10 x 1.8 ml (catalog #165-2676) or 30.0 ml (catalog #165-2677) of a proprietary buffer formulation. 1 ml is generally sufficient for 5–10 transfections in 0.2 cm cuvettes, 1–2 transfections in 0.4 cm cuvettes, or 5–10 transfections in a 96-well plate.
Note: One 30 ml bottle is sufficient for electroporating two entire plates (96-, 24-, or 12-well plates). You do not have to fill all wells in each plate
Overview
Electroporation is the preferred method for introducing biomolecules into cells, especially into cells that are resistant to other methods of transfection. However, this method requires an effective electroporation buffer and optimized conditions to ensure high cell viability after electroporation.
Gene Pulser electroporation buffer reagent was developed for the transfection of siRNAs and plasmid DNA into a variety of cell types, resulting in high efficiency and low cytotoxicity.
Gene Pulser electroporation buffer reagent allows transfection using a variety of protocols and conditions, can be used with any multiwell plate format or cuvette size, and can be used in Gene Pulser®II, Gene Pulser Xcell™, Gene Pulser MXcell™, and other electroporation systems.
For best results, it is important to determine the optimal amount of siRNA or plasmid DNA, and the optimal electroporation conditions for any given cell line.
Recommendations for Optimal Results
Gene Pulser electroporation buffer reagent has been developed to achieve consistent transfection efficiencies using a broad range of cell types with an easy-to-use protocol.
Optimum transfection efficiencies are achieved by adjusting:
• siRNA or plasmid DNA concentration
• Cell density at the time of transfection
• Waveform (exponential decay or square waveforms)
• Voltage
• Pulse length
Once maximum transfection efficiency has been established, the conditions should be kept constant between experiments for any particular cell line.
Bio-Rad Laboratories, Inc., 2000 Alfred Nobel Drive, Hercules, CA 94547 USA • 510-741-1000
10009178 Rev B
Optimization
Determining the optimum electroporation conditions is essential to maximize the transfection efficiency of siRNAs, to obtain the best gene silencing results, and to minimize cellular toxicity. The same is true for the delivery of plasmid DNA.
Optimize parameters for every cell line using the guidelines in Table 1.
*
See Table 2 for maximum values.
**
If using a 0.2 cm cuvette, reduce the voltage by 50%.
Transfection Protocol
1. Harvest and count the cells. a. The cells should be passaged the day before electroporation. All cell types should be harvested when they are
actively growing. If working with adherent cells, trypsinize the cells to detach them, add growth medium, and then pellet the cells. If working with suspension cells, pellet the cells.
b. After pelleting the cells, remove the medium and wash the cells once with PBS, by carefully pipetting them.
Take an aliquot and count the cells.
2. Prepare the cells for electroporation. a. Aliquot the number of cells needed to perform the experiment. For adherent cells, we recommend using 1 x 10
6
cells/ml, but we have successfully used 0.5–5 x 106cells/ml. For suspension cells, we recommend using 2–3 x 10
6
cells/ml.
b. Pellet the cells. c. Aspirate the PBS and resuspend the cells in the appropriate volume of electroporation buffer reagent (1 ml per
1 x 10
6
of adherent cells, and 1 ml per 2–3 x 106of suspension cells).
d. Add nucleic acid. For siRNA electroporation, use 10–100 nM of siRNA. For plasmid DNA electroporation, use
5–20 µg/ml.
3. Electroporate cells. a. Choose the proper vessel for your sample size (see Table 2).
b. Electroporate the sample using optimized conditions. c. Transfer cells to tissue culture dishes containing growth media. d. Incubate cells at 37°C in a humidified CO
2
incubator until ready to be assayed. After 24 hr, change the growth
medium.
4. Assess transfection efficiency. a. Fluorescently labeled siRNAs can be used to determine the transfection efficiency for siRNA delivery.
Transfection efficiency can be measured by fluorescent microscopy or flow cytometry.
b. For plasmid delivery, the transfection efficiency can be determined by electroporating plasmids expressing
reporter genes such as GFP or b-galactosidase.
Table 1. Electroporation optimization guidelines.
Parameter
Voltage Square waveforms 50–400 V (in 10 V increments) Exponential waveforms 100–350 V (in 50 V increments) Pulse length
(for square waveforms) Capacitance
(for exponential waveforms) Vessel size* 0.4 cm
Cell density Adherent cells 1–5 x 106 cells/ml Suspension cells 2–10 x106 cells/ml siRNA concentration 10–100 nM
Optimization range
10–25 ms
500–1000 µF
2
cuvette** or
96-, 24-, or 12-well electroporation plate
Table 2. Proper vessel based on sample size.
Sample Volume Vessel
100–200 µl 0.2 cm cuvette 400–800 µl 0.4 cm cuvette 100–200 µl 96-well plate 500–800 µl 24-well plate
1.0–1.5 ml 12-well plate