Bio-Rad Bio-Plex System Validation and Calibration Tools User Manual

Bio-Rad Laboratories
2000 Alfred Nobel Dr. Hercules, CA 94547
1-800-424-6723
4110007 Rev C
Catalog # 171-203000
For use with Bio-Plex Manager Software Version 3.0 and MCV plate II
or
For use with Bio-Plex Manager Software Version 2.0 and MCV plate
For technical service, call your local Bio-Rad office, or in the US, call 1-800-4BIORAD (1-800-424-6723)
For research use only. Not for diagnostic procedures.
Table of Contents
Section 1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
Section 2 Product Description . . . . . . . . . . . . . . . . . . . . . . . . . . . .2
Section 3 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
Section 4 Storage and Handling . . . . . . . . . . . . . . . . . . . . . . . . . . .3
Section 5 Principle of Optics Validation . . . . . . . . . . . . . . . . . . . .4
Section 6 Principle of Reporter Channel Validation . . . . . . . . . . .4
Section 7 Principle of Classify Validation . . . . . . . . . . . . . . . . . . .7
Section 8 Principle of Fluidics Validation . . . . . . . . . . . . . . . . . .8
Section 9 Procedure for Bio-Plex Manager 3.0/MCV plate II . . .9
9.1 One-Step Procedure for all Validation Parameters . . . . . . . . .9
9.2 Validation of Optics Alignment . . . . . . . . . . . . . . . . . . . . . .11
9.3 Validation of Fluidics Integrity . . . . . . . . . . . . . . . . . . . . . . .13
9.4 Validation of Reporter Channel Performance . . . . . . . . . . . .15
9.5 Validation of Classify Efficiency . . . . . . . . . . . . . . . . . . . . . .16
9.6 Generating a Validation Report . . . . . . . . . . . . . . . . . . . . . .18
9.7 Validation Report Example . . . . . . . . . . . . . . . . . . . . . . . . .19
Section 10 Procedure for Bio-Plex Manager 2.0/MCV plate . . . .21
10.1 Validation of Optics Alignment . . . . . . . . . . . . . . . . . . . . . .21
10.2 Validation of Reporter Channel Performance . . . . . . . . . . . .23
10.3 Validation of Classify Efficiency . . . . . . . . . . . . . . . . . . . . . .25
10.4 Validation of Fluidics Integrity . . . . . . . . . . . . . . . . . . . . . . .26
10.5 Generating a Validation Report . . . . . . . . . . . . . . . . . . . . . .28
10.6 Validation Kit Worksheet and Report
Section 11 Troubleshooting Guide . . . . . . . . . . . . . . . . . . . . . . . . .32
Section 12 Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . .32
Section 13 Reference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .33
Form Examples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28
Section 1 Introduction
Qualification of analytical instruments is a formal process of documenting that an instrument is fit for its intended use and that it is kept maintained and calibrated. The Bio-Plex validation kit is used for operational qualification (OQ) of the Bio-Plex protein array system. The validation kit is designed to validate the operation of all of the primary components of the system and is a valuable tool that allows the user to discriminate between assay and instrumentation problems.
The Bio-Plex validation kit consists of beads to evaluate the following components of the Bio-Plex protein array system: 1) optics alignment, 2) integrity of fluidics, 3) reporter channel performance, and 4) classify efficiency. A brief definition of the parameter and the principle of each procedure is described, along with complete procedures for evaluating each of the primary components. An explanation of the potential impact of each process on a typical Bio-Plex cytokine assay is included to assist the user in assay troubleshooting and development.
NNOOTTEE:: BBiioo--PPlleexx MMaannaaggeerr 33..00 rreeqquuiirreess MMCCVV ppllaattee IIII ((CCaatt## 117711--220033003311)) ffoorr uussee wwiitthh tthhiiss VVaalliiddaattiioonn KKiitt,, sseeee SSeeccttiioonn 99..
BBiioo--PPlleexx MMaannaaggeerr 22..00 rreeqquuiirreess MMCCVV ppllaattee ((CCaatt## 117711--220033003300)) ffoorr uussee wwiitthh tthhiiss kkiitt,, sseeee SSeeccttiioonn 1100.. BBiioo--PPlleexx MMaannaaggeerr 22..00 ddooeess nnoott aallllooww uussee ooff fflluuiiddiiccss 11 aanndd fflluuiiddiiccss 22 bbeeaaddss.. TThhiiss sstteepp iiss oommiitttteedd,, ffoollllooww SSeeccttiioonn 1100 iinnssttrruuccttiioonnss..
For research use only. Not for diagnostic procedures.
1
Section 2 Product Description
The following reagents are included in the Bio-Plex validation kit:
Reagent Quantity
Optics validation bead set: Optics beads 1 and 2 2 x 10 ml black bottles
5
(1x10
beads/ml)
Fluidics validation bead set: Fluidics bead 1 1 x 10 ml black bottle
5
(1x10
beads/ml)
Fluidics bead 2 1 x 10 ml black bottle
Reporter validation bead set: Reporter blank, 1, 2, 3, 4, and 5 6 x 10 ml white bottles
5
(1x10
beads/ml)
Classify validation bead set: Classify bead 34, 38, 54, 73, 77 5 x 10 ml black bottles
5
(1x10
beads/ml)
NNOOTTEE:: TThhee vvaalliiddaattiioonn kkiitt mmuusstt bbee uusseedd ffoolllloowwiinngg CCAALL22 ccaalliibbrraattiioonn uussiinngg tthhee RRPP11 hhiigghh ttaarrggeett vvaalluuee.. VVaalliiddaattiioonn sshhoouulldd bbee ppeerrffoorrmmeedd iimmmmeeddiiaatteellyy ffoolllloowwiinngg ccaalliibbrraattiioonn..
The following materials are required but not supplied:
Bio-Plex MCV plate
Bio-Rad catalog #171-203031 MCV plate II, use with Bio-Plex Manager 3.0, Bio-Rad catalog #171-203030 MCV plate, use with Bio-Plex Manager 2.0
Bio-Plex Protein Array System
Bio-Rad catalog #171-000001, 171-000003, or 171-000005
Bio-Plex Calibration Kit
Bio-Rad catalog #171-203060
mini vortexer sterile distilled water 70% isopropanol 10% bleach bulb pipets
2
Section 3 Specifications
General specifications for the validation kit are listed below. Certain specifications for the Bio-Plex validation kit may differ from lot to lot. For a complete listing of the current specifications, please refer to the package insert provided with your validation kit.
Parameter Specification
Optics Validation
DD mean See insert CL1 mean See insert CL1 CV% (Coefficient of Variable) 3–7% CL2 mean See insert CL2 CV% 4–8% RP1 mean See insert RP1 CV% 5–10%
Fluidics Validation
% carryover < or = 4.0%
Reporter Validation
Dynamic range 4.15–4.28 Linearity >0.995 Slope of response 0.0593–0.0799 Accuracy of response >90% Sensitivity <200 MESF
Classify Validation
Classify Bead 34, 38, 54, 73, 77 >80.0%
DD Efficiency
% Efficiency > or = 75%
Section 4 Storage and Handling
The Bio-Plex validation kit beads are stable if stored at 4°C protected from light. When using the Bio-Plex validation kit, remove beads from 4°C storage and dispense into the MCV plate. Return to 4°C storage immediately following use to preserve shelf life. All components are guaranteed for 6 months from the date of purchase when stored as specified in this manual.
3
Section 5 Principle of Optics Validation
Principle
The Bio-Plex array reader is a laser-based fluorescence detection system containing sensitive optics components. Alignment of the laser/optics system is critical for optimal instrument performance. A method for the assessment of the optics alignment is included in the validation kit. Acceptable specifications for the alignment procedure are listed in the product insert.
Impact on Assay Performance
The alignment of the optics bench of the Bio-Plex array reader is critical for proper assay performance. Misalignment of the reporter optics path can result in 1) reduced assay sensitivity or 2) poor well-to-well assay precision. Misalignment of the classification optics path can lead to 1) increased read times or 2) misclassification of one assay into another, leading to false positive or negative results. Correlation studies have been performed to determine the direct effect of misalignment on assay performance.
Section 6 Principle of Reporter Validation
Principle
The reporter (RP1) channel is the fluorescence channel used for assay quantitation (See Bio-Plex system hardware manual for more information regarding the principle of Bio-Plex technology). Therefore, validation of this component of the Bio-Plex system is a critical part of operational qualification. R-phycoerythrin (R-PE) is the primary reporter molecule used in Bio-Plex assays. A series of beads dyed with varying intensities of a fluorochrome spectrally matched to R-phycoerythrin are used for this procedure. Each of the reporter beads has been assigned a specific intensity value corresponding to the number of fluorescent R-PE molecules. These units of fluorescent measure are known as molecules of equivalent soluble fluorescence (MESF). MESF units allow direct correlation of instrument performance to a typical assay using R-PE as the indicator molecule. The primary reporter channel performance parameters are as follows: dynamic range, linearity, accuracy of reporter channel response, sensitivity, and slope of the response. Each of these parameters is related directly to the performance of the Bio-Plex array reader and has defined acceptable specifications. Definitions for the parameters and the applicability to a typical assay performed on the Bio-Plex array reader are listed below. If any of the parameters are not within the specified range, contact Bio-Rad Technical support for assistance.
4
Dynamic Range of Reporter Channel
Definition
The dynamic range is the calculated number of decades covered by the log amplifier from the slope and the histogram scale. The available range of channels on the Bio-Plex array reader is 4.5 log amp decades or 32,767 relative linear channels. The acceptable dynamic range of fluorescence measured by the Bio-Plex array reader using the Bio-Plex reporter beads is 4.15–4.28.
Impact on Assay Performance
The dynamic range of the Bio-Plex array reader is 4.5 log amp decades or 32,767 relative linear channels. It is desirable for the range of the instrument to be greater than the range of an assay. If the dynamic range of the instrument is less than that of an assay, the range of quantitatable analyte may be limited. This parameter will guide the user in defining the instrument versus the assay dynamic range limitations.
Linearity of Reporter Channel
Definition
The reporter validation bead set is utilized to construct a plot where the reporter channel median fluorescence intensity values are plotted against the corresponding assigned MESF values. Instrument linearity is expressed as the coefficient of determination or R-squared (R
Impact on Assay Performance
2
) value. The R2value must be >0.995.
The linearity of the instrument response may directly affect a typical standard or calibration curve in a Bio-Plex assay, thereby impacting the unknown values extrapolated from that curve. If the R
2
value is not within acceptable limits, it may be necessary to realign the optics or check the response of the reporter photomultiplier tube.
Accuracy of Reporter Channel Response
Definition
The accuracy of the reporter channel response is a more stringent measurement of the linearity than the R
2
value. Simply stated, the accuracy of the reporter channel response is the percent difference that the regression line is away from the actual MESF value data points. The desired accuracy value is >90%.
Impact on Assay Performance
Since accuracy is also a measurement of the linearity of the instrument response, the same principles that apply to linearity also apply to accuracy of the reporter channel response. Accuracy values <90% could impact assay performance. The accuracy data is evaluated in combination with optics alignment to determine if the Bio-Plex array reader will perform according to specifications. It is possible for the accuracy value to fall out of specification before the linearity parameter. This is expected due to the fact that the accuracy parameter is a more sensitive measurement of linearity than the R
2
value. These data are correlated with optics alignment data as well as assay performance to determine when the array reader will not perform according to specifications.
5
Slope of the Reporter Channel Response
Definition
The slope of the regression line resulting from the plotting of reporter channel mean fluorescent values against assigned reporter channel validation bead MESF values is related to the dynamic range of the instrument. The slope of the regression line is a function of the response of the reporter channel photomultiplier tube. The acceptable range for the slope is 0.0593–0.0799.
Impact on Assay Performance
The slope of the regression line is directly related to the dynamic range of the instrument. The slope yields direct information about the response of the photomultiplier tube. If the photomultiplier tube signal saturates at low fluorescence values, the dynamic range of the instrument is affected. The slope of the line impacts the dynamic range and the range in turn impacts the quantitatable range of an assay. If the validation kit yields a value for the slope that is not within specifications, assay results could be adversely affected.
Sensitivity of Reporter Channel
Definition
Every instrument has an inherent level of noise due primarily to the electronics. The sensitivity of the Bio-Plex array reader is defined as the lowest detectable signal above instrument noise. Noise can be attributed to the laser, the photomultiplier tube, the amplification electronics or the fluidics. The acceptable sensitivity using the Bio-Plex array reader is <200 MESF.
Impact on Assay Performance
The sensitivity using the Bio-Plex validation kit is expressed in terms of MESF. The fluorescence is traceable to R-PE, the primary molecule used in Bio-Plex assays. The typical background or zero standard of a Bio-Plex cytokine assay falls at a median fluorescence intensity of 100. The background of a "blank" bead from the validation kit exhibits a median fluorescence intensity of 8. A plot of the median RP1 fluorescence intensity versus the MESF units illustrates that the instrument is approximately 10 times more sensitive than a Bio-Plex cytokine assay (See Figure 1). This is a desired result, as the sensitivity of the instrument should not directly limit assay sensitivity.
6
Assay MESF
Reader sensitivity
MESF (R-PE)
RP1 Channel
Fig. 1. Assay vs. instrument sensitivity.
Assay Background
Section 7 Principle of Classify Validation
Principle
Bio-Plex technology relies on the ability of the Bio-Plex array reader to discriminate between assay beads impregnated with varying ratios of 2 fluorescent dyes. This is the concept whereby multiplexing within a single well may occur. The periodic evaluation of the classify efficiency is necessary to complete the Bio-Plex array reader qualification process. A series of beads with varying ratios of the classification dyes are analyzed on the Bio-Plex array reader and the efficiency of multiplexing is quantitated. A classify efficiency of >80% is required for optimal results. DD Efficiency is a measure of the percentage of the Classify beads that fall within the DD Gates. Greater than 75% of the beads should fall within the gates for optimal results.
Impact on Assay Performance
Inefficient classification of beads may have several potential effects on an assay. If a bead region exhibits a classify efficiency of less than 80%, the read time of a 96-well plate may be increased. The Bio-Plex array reader tabulates a specified number of defined events in each region for each well sampled. If the percentage of beads within a specific region is low, the time required to count is increased, therefore the total time to read an entire plate is prolonged. Extremely prolonged assay read times could impact well-to-well precision, since the kinetics of a sandwich assay, for example, are not 100% stable over a period of 3–5 hrs. Another potential impact of inefficient classification is the misclassification of one assay bead into another bead region. This could yield false positive or negative results for a particular assay. A DD efficiency value less than 75% may increase the read time of the assay and affect results in the same manner as a low classify efficiency.
7
Section 8 Principle of Fluidics Validation
Principle
The fluidics system of the Bio-Plex suspension array reader requires routine maintenance to prevent clogging and other malfunctions. Strict adherence to the maintenance procedures is mandatory for optimal instrument performance. An assessment of the integrity of the fluidics is automatically performed in the Fluidics Validation procedure. In the fluidics validation test, a sample of beads is analyzed followed by a sample of buffer to assess the carryover of beads from one well to another. This procedure should be performed once per week to ensure that assay results are not adversely affected. The fluidics path, including the sample needle must be completely free of debris and excess beads for optimal array reader performance.
Impact on Assay Performance
If a system is exhibiting a high level of carryover, due to valve malfunction or partially clogged sample needle, a significant percentage of beads may be carried over from one well to another. This phenomenon may adversely affect the median fluorescent intensity values. For example, if a well with a high median fluorescent intensity (FI) is read immediately prior to a well with a low median FI, the signal in the well with the low fluorescent intensity may shift upward. This phenomenon only occurs in extreme cases since the median fluorescent intensity statistic is robust and is not easily shifted by the introduction of a population of beads with a significantly different median FI.
8
Loading...
+ 25 hidden pages