Anritsu S251C User Manual

MAINTENANCE MANUAL
Site Master™ Model S251C
Antenna and Cable Analyzer
Figure 1. Site Master Model S251C
1. INTRODUCTION
This manual provides maintenance instructions for the Site Master S251C Antenna and Cable Analyzer. It describes the product and provides performance verification procedures, parts replacement proce
dures, and a replaceable parts list.
2. DESCRIPTION
The Site Master (Figure 1) is a hand held S mission gain or loss), SWR/RL (standing wave ra
21
(trans
3. PERFORMANCE VERIFICATION
Paragraphs 4 through 10 contain tests that can be used to verify the performance of the Site Master model S251C having any version of firmware.
3.1. Initial Setup for Testing
1. Press and hold the press the
ter. (This sets the instrument to the factory
ON/OFF key to turn on the Site Mas
ESCAPE/CLEAR key, then
preset state.)
2. If necessary, use the Up/Down arrow key to adjust the contrast to give a readable display. Press
ENTER to accept.
490 JARVIS DRIVE ¨ MORGAN HILL, CA 95037-2809
P/N: 10580-00067
REVISION B
PRINTED: FEBRUARY 2003
COPYRIGHT 2001-2003 ANRITSU CO.
R F
S o u r c e
C o u p l e r
R F O u t
P o r t
R F I n
P o r t
I n c i d e n t
L o c a l
O s c i l l a t o r
D i s p l a y
Figure 2. Site Master Block Diagram
4. FREQUENCY ACCURACY
The following test can be used to verify the CW fre­quency accuracy of the Site Master. Measurement calibration of the Site Master is not required for this test.
a. Equipment Required:
·
Spectrum Analyzer
Anritsu Model MS2668C or equivalent
b. Procedure:
1. Press and hold the press the
ON/OFF key to turn on the Site Mas
ESCAPE/CLEAR key, then
ter. (This sets the instrument to the factory preset state.)
NOTE
Before continuing, allow a five minute warm up for the internal circuitry to stabilize.
2. Press the
3. Use the Up/Down Arrow key to highlight
SOURCE
4. Press the
MODE key.
RF
, then press ENTER.
FREQ soft key.
R e f l e c t e d
S y n c h r o n o u s
D e t e c t o r
A / D C o n v e r t e r
C P U
T r a n s m i t t e d
S e r i a l
P o r t
5. Using the keypad or Up/Down Arrow key en­ter 1000.0 MHz then press the
ENTER key.
6. Connect the RF cable from the Site Master RF Out Test Port to the RF Input on the Spec­trum Analyzer.
7. Set up the Spectrum Analyzer as follows: (a) Press
Preset
(b) Press the PRESET ALL soft key (c) Press (d) Press
(e) Press
Frequency and enter 1 GHz Span and enter 200 kHz Amplitude
(f) Press the REFERENCE LEVEL soft key and
enter 10 dB
8. If the Site Master has gone into the hold mode, press the
RUN/HOLD key to make the
measurement.
9. On the Spectrum Analyzer: (a) Press
Marker
(b) Press PEAK SEARCH. The frequency
should be 1000 MHz ±75 kHz.
2 Site Master S251C MM
5. TRANSMISSION/ISOLATION VERIFICATION
The following test can be used to verify transmission test port isolation and the accuracy of transmission measurements. Measurement calibration of the Site Master is required for this test.
TransTest Port
TestPort Extension Cable
Refl Test Port
a. Equipment Required:
10 dB Attenuator, Weinshel 1-10
·
30 dB Attenuator, Weinshel 1-30
·
Open/Short, Anritsu 22NF50
·
50 Ohm Terminations,
·
Anritsu SM/PL or 28N50-2 Anritsu SM/PLNF or 28NF50-2
Armored Test Port Extension Cable,
·
1.5 Meter, N(m) to N(m), Anritsu 15NN50-1.5A
b. Procedure:
1. Press and hold the press the
ON/OFF key to turn on the Site Mas-
ESCAPE/CLEAR key, then
ter. (This sets the instrument to the factory preset state.)
NOTE
Before continuing, allow a five minute warm up for the internal circuitry to stabilize.
2. Press the
MODE key.
3. Use the Up/Down Arrow key to highlight
SERTION LOSS
4. Press the
5. Press the
, then press ENTER.
AMPLITUDE key. TOP soft key.
6. On the keypad, press “0”, then press (Verify the bottom limit is set to –120 dB.)
7. Press the
LIMIT key.
IN
ENTER.
Site Master S251C
MODE
FREQ/DIST
SWEEP
AMPLITUDE
Figure 3. OSL-THRU-ISOL Calibration Setup
12. Connect a Load (28N50-2 or SM/PL) to the RF In Test Port and verify that the noise floor (isolation) is below –80 dB.
13. Press the
14. Press the
MARKER key. M1 soft key, then the EDIT soft key.
15. Using the keypad or the Up/Down Arrow key, enter 1000.0 MHz, and press
NOTE
For the following steps in the procedure, use only attenuators that have Type N connectors. The use of attenuators with other type connectors and adapters will cause measurement errors.
16. Connect the 10 dB attenuator to the RF In
-
Test Port (refer to Figure 4) and verify that the reading is –10 dB ±2.25 dB @ 1000 MHz.
17. Connect the 10 dB and 30 dB attenuators in series to the RF In Test Port and verify that the reading is –40 dB ±3.25 dB @ 1000 MHz.
Attenuator
ESCAPE CLEAR
2
1
START
AUTO
CAL
SCALE
3
4
SAVE
RECALL
SETUP
SETUP
6
5
LIMIT
MARKER
ENTER
8
7
RECALL
RUN
SAVE
DISPLAY
DISPLAY
HOLD
+
9
0
-
/
PRINT
ON
SYS
OFF
.
ENTER.
TestPort Extension Cable
8. Press the
9. Press the the
10. Press the
11. Follow the instructions on the display to per form a OSL-THRU-ISOL calibration using a 22NF50 Open/Short, 28NF50-2 or SM/PLNF Terminations, and 15NN50-1.5ATest Port Ex tension Cable (refer to Figure 3).
SINGLE LIMIT soft key. EDIT soft key, enter –80, and press
ENTER key.
START CAL key.
Figure 4. Test Setup
TransTest Port
Refl Test Port
Site Master S251C
MODE
FREQ/DIST
AMPLITUDE
SWEEP
ESCAPE CLEAR
2
1
START
AUTO
CAL
SCALE
3
4
SAVE
RECALL
SETUP
SETUP
6
5
LIMIT
MARKER
ENTER
8
7
RECALL
RUN
SAVE
DISPLAY
DISPLAY
HOLD
+
9
0
-
/
PRINT
ON
SYS
OFF
.
Site Master S251C MM 3
6. RETURN LOSS VERIFICATION
The following test can be used to verify the accuracy of return loss measurements. Measurement calibra tion of the Site Master is required for this test.
a. Equipment Required:
5. Follow the instructions on the screen to per
form a calibration using a 22N50 Open/Short
and 28N50-2 or SM/PL Termination.
6. Connect the 20 dB offset to the RF Out Test Port and verify that the reading is 20 dB ±1.7 dB
20 dB offset, Anritsu SC5270
·
6 dB offset, Anritsu SC5237
·
Open/Short, Anritsu 22N50
·
50 Ohm Termination,
·
Anritsu 28N50-2 or SM/PL
b. Procedure:
1. Press and hold the press the
ON/OFF key to turn on the Site Mas
ESCAPE/CLEAR key, then
ter. (This sets the instrument to the factory preset state.)
NOTE
Before continuing, allow a five minute warm up for the internal circuitry to stabilize
2. Press the
MODE key.
3. Use the Up/Down Arrow key to highlight
TURN LOSS
4. Press the
, then press ENTER.
START CAL key.
RE-
7. Connect the 6 dB offset to the RF Out Test Port and verify that the reading is 6 dB ±1.2 dB
7. POWER MONITOR VERIFICATION
If the Power Monitor (Option 5) is installed in the Site Master, the following test can be used to verify the accuracy of the power measurements. Measure ment calibration of the Site Master is not required for this test.
a. Equipment Required:
RF Detector, 10 MHz to 20 GHz,
·
Anritsu 560-7N50B
· 10 dB Attenuator, Weinschel 1-10
· 30 dB Attenuator, Weinschel 1-30
· RF Reference Source, 0.050 GHz,
Anritsu MA2418A
· DC Power Supply, Anritsu 2000-933
Attenuators
560-7N50B
RF Detector
MA2418AReference Source
W
50MHz, 50 0dBm
Site Master S251C
MODE
FREQ/DIST
AMPLITUDE
SWEEP
ESCAPE CLEAR
2
1
START
AUTO
CAL
SCALE
3
4
SAVE
RECALL
SETUP
SETUP
6
5
LIMIT
MARKER
ENTER
8
7
RECALL
RUN
SAVE
DISPLAY
DISPLAY
HOLD
+
9
0
-
/
PRINT
ON
SYS
OFF
.
DC Power Supply 2000-933
20mA
15-24VDC
MA2418A Reference Source
Site Master w/ Option 5 Installed
Figure 5. Power Monitor Verification
4 Site Master S251C MM
b. Procedure
1. Connect the DC power supply to the MA2418A Reference Source. (Refer to Figure
5.)
2. Connect the MA2418A Reference Source to the input of the 560-7N50B RF detector.
3. Connect the RF Detector output to the RF De tector input of the Site Master.
(Option 10A only works with the AC-DC Adapter).
2. Apply AC power to the power supply.
3. Press and hold the press the
ON/OFF key to turn on the Site Mas
ESCAPE/CLEAR key, then
ter. (This sets the instrument to the factory
preset state.)
4. Press the
SYS key.
4. Connect the DC power supply to the appropri ate line voltage to supply power to the MA2418A Reference Source.
5. Press and hold the press the
ON/OFF key to turn on the Site Mas
ESCAPE/CLEAR key, then
ter. (This sets the instrument to the factory preset state.)
6. Press the
MODE key.
7. Use the Up/Down Arrow key to highlight
POWER MONITOR, then press ENTER.
8. Press the
ZERO soft key to zero the power
monitor. When complete,
ZERO ADJ:ON is displayed in
the message area.
9. Verify that the power monitor reading is
0.0 dBm ±1 dB.
10. Connect the output of the MA2418A Refer­ence Source to the two attenuators so as to add 40 dB of attenuation (Figure 5).
11. Connect the MA2418A Reference Source and the attenuators to the input of the 560-7N50B RF detector.
5. Press the
6. Press the
OPTIONS soft key. BIAS-T soft key to activate the Bias
Tee option.
7. Verify that the display shows
15.0V, 0 mA.
8. Connect the 22NF50 short to the RF In test port.
9. Verify the test port voltage relay clicks on and off, and
FAULT is shown in the upper left cor
ner of the display.
10. Remove the short and install the 100 Ohm load.
CAUTION
The 100 Ohm resistor will get hot. Ver­ify the readings and remove the 100 Ohm load immediately.
11. Verify that the display shows
BIAS, with a
reading of 14.8 to 15.3 Volts, and 120 to 160 milliamps.
9. BIAS T (OPTION 10B) VERIFICATION a. Equipment Required:
·
12. Verify that the power monitor reading is now –40.0 dBm ±2 dB.
8. BIAS T (OPTION 10A) VERIFICATION a. Equipment Required:
·
AC-DC Adapter Power Supply 40-115
·
Open/Short, Anritsu 22N50
·
100 Ohm, 1 Watt load (100 Ohm, 1 Watt re sistor soldered from the center pin to ground of an N-type connector.)
b. Procedure:
1. Connect the AC-DC Adapter Power Supply to the battery charging port on the Site Master
Open/Short, Anritsu 22N50
·
100 Ohm, 1 Watt load (100 Ohm, 1 watt resis tor soldered from the center pin to ground of a N-type connector)
·
Anritsu T2904
b. Procedure:
1. Ensure the battery is fully charged. (Option 10B only works with the battery).
2. Press and hold the press the
ON/OFF key to turn on the Site Mas
ESCAPE/CLEAR key, then
ter. (This sets the instrument to the factory preset state.)
3. Press the
4. Press the
SYS key. OPTIONS soft key.
Site Master S251C MM 5
Loading...
+ 11 hidden pages