HK Pneumatic Deadweight Tester
ACCURACY
Accuracy: ±(0.025% of reading) Repeatability: ± 0.005% of reading
Temperature Coefficient: ± 0.00167% of rdg / °C outside of 23°C
WEIGHT MASSES
Pressure Range: 10 to 1500 psi /50 to 10 000 kPa
Gravity: 9.80655 m/s² (International standard)
Local Gravity (Specify) Available Units: psi, kg/cm², bar, and kPa
PRESSURE SOURCE
Source: Nitrogen or Instrument /Shop air
Maximum Supply Pressure: 1500 psi /100 bar (1000 psi and lower models)
2250 psi /150 bar (models above 1000 psi)
Minimum Supply Pressure: 150 psi /10.3 bar
Flow Rate: 3 scfh at 10 psi /85 slh at 1 bar output
100 scfh at 1500 psi /2800 slh at 100 bar output
ISA quality standard 7.3.
Or 150% of desired output pressure.
ENCLOSURE
Tester Dimensions (L x W x H): 12.5 x 11.47 x 11.5 in (31.75 x 29.1 x 29.2 cm) Tester Weight: 18 to 31.5 lb (8.2 to 14.3 kg)
Weights Case Dimensions (L x W x H): 2 x 12 x 12 in (5.1 x 30.5 x 30.5 cm) Masses: 17.5 to 61.5 lb (8 to 30 kg)
MATERIAL AND CONNECTION
Weight Masses: Stainless Steel (as standard) Ball: Ceramic
Connection: 7/16-20 37° AN4 Male
29.1
(11.47)
29.2
(11.5)
31.75
(12.5)
SS-HK Issue 1503
Page 1 of 2