RK Pneumatic Deadweight Tester
ACCURACY
** Accuracy: ±(0.015% of reading*), ±(0.025% of reading*), or ±(0.05% of reading) Repeatability: ± 0.005% of reading
Temperature Coefficient: ± 0.00167% of rdg / °C outside of 23°C
*Down to 10 inH2O.
±(0.05% of reading) below 10 inH2O
**See ordering information for details.
WEIGHT MASSES
Pressure Range: 0.145 to 300 psi /1 to 2000 kPa
Gravity: 9.80655 m/s² (Int’l standard)
Local Gravity (Specify)
Water Column Reference: 20°C (ISA standard)
60°F (AGA standard, specify when ordering) Available Units: psi, kg/cm², bar, kPa, inH O, cmH O, and inHg
PRESSURE SOURCE
Source: Nitrogen or instrument /shop air
Maximum Supply Pressure: 450 psi /3100 kPa Minimum Supply Pressure: 30 psi /210 kPa
Flow Rate: 1 scfh at 0.15 psi / 28 slh at 1 kPa output
60 scfh at 300 psi / 1700 slh at 2000 kPa output
ISA quality standard 7.3.
Or 150% of desired output pressure.
24.1
(9.0)
20.3
(8.0)
38.1
(15.0)
ALL DIMENSIONS IN CM (IN)
Tester Enclosure
SS-RK Issue 1503
Page 1 of 3