Ametek CTC-155A, CTC-350A, CTC-652A, CTC-660A, CTC-155C Operating Manual

...
4.5 (2)

Reference Manual

Compact Temperature Calibrator

Jofra CTC-155/350/660 A/C

Reference Manual

Compact Temperature Calibrator

JOFRA CTC-155/350/660 A/C

Copyright 2016 AMETEK Denmark A/S

About this manual….

The structure of the manual

This reference manual is aimed at users who are familiar with AMETEK calibrators, as well as those who are not. The manual is divided into 11 sections which describe how to set up, operate, service and maintain the calibrator. The technical specifications are described and accessories may be ordered from the list of accessories.

Safety symbols

This manual contains a number of safety symbols designed to draw your attention to instructions which must be followed when using the instrument, as well as any risks involved.

Warning

Conditions and actions that may compromise the safe use of the instrument and result in considerable personal or material damage.

Caution…

Conditions and actions that may compromise the safe use of the instrument and result in slight personal or material damage.

Note…

Special situations which demand the user’s attention.

129375 00

2016-05-18

2

List of contents

1.0

Introduction .................................................................................................................................

4

 

1.1

Warranty................................................................................................................................................

5

 

1.2

Receiving the Compact Temperature Calibrator...................................................................................

6

 

1.3

Dimensioning drawing...........................................................................................................................

7

2.0

Safety instructions......................................................................................................................

8

3.0

Setting up the calibrator for use..............................................................................................

11

 

3.1

Preparing the calibrator.......................................................................................................................

11

 

3.2

Choosing an insertion tube .................................................................................................................

12

 

 

3.2.1

Standard insertion tubes ........................................................................................................

13

 

3.3

Inserting the sensor.............................................................................................................................

14

 

3.4

Programming intelligent STS sensors.................................................................................................

16

4.0

Calibrator Interface ...................................................................................................................

17

 

4.1

Keypad - Functions .............................................................................................................................

17

 

4.2

Display - Functions..............................................................................................................................

17

 

 

4.2.1 Main screen temperature values ............................................................................................

18

 

 

4.2.2 Stability of temperature values ...............................................................................................

18

 

4.3

Input/Output Connections ...................................................................................................................

19

5.0

Operating the calibrator ...........................................................................................................

20

 

5.1

Operating principle ..............................................................................................................................

20

 

 

5.1.1

System menu..........................................................................................................................

21

 

5.2

Starting the calibrator ..........................................................................................................................

23

 

5.3

Selecting a TRUE – reference sensor (C-models only) ......................................................................

23

 

5.4

Stability setting ....................................................................................................................................

25

 

5.5

Selecting the set-temperature .............................................................................................................

25

 

 

5.5.1 Editing the preset set-temperature.........................................................................................

26

 

5.6

Auto Step function...............................................................................................................................

28

 

 

5.6.1 Running an Auto Step test .....................................................................................................

28

 

 

5.6.2 The calibrator's Auto Step procedure.....................................................................................

29

 

5.7

Switch Test function ............................................................................................................................

31

 

 

5.7.1 Running a Switch Test ...........................................................................................................

31

 

 

5.7.2 The calibrator's Switch Test procedure ..................................................................................

33

6.0

Storing and transporting the calibrator ..................................................................................

34

7.0

Error messages (List of alarms) ..............................................................................................

37

8.0

Returning the calibrator for service ........................................................................................

40

9.0

Maintenance ..............................................................................................................................

42

 

9.1

Replacing the main fuses....................................................................................................................

42

 

9.2

Maintenance mode..............................................................................................................................

43

 

9.3

Cleaning ..............................................................................................................................................

47

 

9.4

Adjusting and calibrating the instrument .............................................................................................

48

 

9.5

Maintenance of STS-reference sensor ...............................................................................................

48

10.0

Technical specifications..........................................................................................................

49

11.0

List of accessories....................................................................................................................

56

129375 00

2016-05-18

3

1.0Introduction

Congratulations on your new AMETEK JOFRA CTC Calibrator!

With the AMETEK JOFRA Compact Temperature Calibrator, you have chosen an extremely effective instrument which we hope will live up to all your expectations.

This CTC calibrator is a fast, timesaving, and reliable true industrial temperature calibrator designed for on-site use.

During the past several years, we have acquired extensive knowledge of industrial temperature calibration. This expertise is reflected in our products which are all designed for daily use in an industrial environment. Please note that we would be very interested in hearing from you if you have any ideas or suggestions for changes to our products.

This reference manual applies to the following instruments:

JOFRA CTC-155 A – Temperature calibrator

JOFRA CTC-155 C – Temperature calibrator with reference sensor input

JOFRA CTC-350 A – Temperature calibrator

JOFRA CTC-350 C – Temperature calibrator with reference sensor input

JOFRA CTC-660 A – Temperature calibrator

JOFRA CTC-660 C – Temperature calibrator with reference sensor input

The calibrator has the following features and functions:

Wide temperature range

Fast heating and cooling time as well as a short stabilization time

Signal input for external reference sensor, which makes it possible to improve accuracy even more.

External sensor control for running in two modes.

Multi-Information colour Display and Function keys

Useful features such as Set function, Preset mode, Auto Switch Test and Auto Stepping

IRI – Intelligent Recalibration Information

Plug and play STS reference sensors with memory chip

Broad range of inserts

Reference sensor protection

Silent mode operation

JOFRACAL calibration software

Protective carrying case with compartments for inserts, cables, manuals, plugs etc.

ISO-9001 certified

AMETEK Denmark A/S was ISO-9001 certified in September 1994 by Bureau Veritas Certification Denmark.

129375 00

2016-05-18

4

CE-label

Your new temperature calibrator bears the CE label and conforms to the

Electromagnetic Compatibility (EMC) Directive and the Low Voltage Directive.

Technical assistance

Please contact the dealer from whom you acquired the instrument if you require technical assistance.

1.1Warranty

This instrument is warranted against defects in workmanship, material and design for two (2) years from date of delivery to the extent that AMETEK will, at its sole option, repair or replace the instrument or any part thereof which is defective, provided, however, that this warranty shall not apply to instruments subjected to tampering or, abuse, or exposed to highly corrosive conditions.

THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES WHETHER EXPRESS OR IMPLIED AND AMETEK HEREBY DISCLAIMS ALL OTHER WARRANTIES, INCLUDING,

WITHOUT LIMITATION, ANY WARRANTY OF FITNESS FOR A PARTICULAR PURPOSE OR

MERCHANTABILITY. AMETEK SHALL NOT BE LIABLE FOR ANY INCIDENTAL OR CONSEQUENTIAL DAMAGES, INCLUDING, BUT NOT LIMITED TO, ANY ANTICIPATED OR

LOST PROFITS.

This warranty is voidable if the purchaser fails to follow any and all instructions, warnings or cautions in the instrument’s User Manual.

If a manufacturing defect is found, AMETEK will replace or repair the instrument or replace any defective part thereof without charge; however, AMETEK’s obligation hereunder does not include the cost of transportation, which must be borne by the customer. AMETEK assumes no responsibility for damage in transit, and any claims for such damage should be presented to the carrier by the purchaser.

129375 00

2016-05-18

5

1.2Receiving the Compact Temperature Calibrator

When you receive the instrument…

1)Unpack and check the calibrator and the accessories carefully.

2)Check the parts according to the list shown below.

If any of the parts are missing or damaged, please contact the dealer who sold you the calibrator.

You should receive:

1 CTC Calibrator

1 USB memory stick containing electronic Reference manual and software package JOFRACAL

1 mains cable

1 sets of test leads and test clips (black and red)

Thermal protection shield (CTC-660 only)

1 tool for insertion tube

1 USB cable

1 Calibration certificate (International traceable)

When reordering, please specify the part numbers according to the list of accessories, section 11.0

129375 00

2016-05-18

6

1.3Dimensioning drawing

129375 00

2016-05-18

7

2.0Safety instructions

Read this manual carefully before using the instrument!

Please follow the instructions and procedures described in this manual. They are designed to allow you to get the most out of your calibrator and avoid any personal injuries and/or damage to the instrument.

Disposal – WEEE Directive

These calibrators contain Electrical and Electronic circuits and must be recycled or disposed of properly (in accordance with the WEEE Directive 2002/96/EC).

Warning

About the use:

The calibrator must not be used for any purposes other than those described in this manual, as it might cause a hazard.

The calibrator has been designed for indoor use only and is not to be used in wet locations.

The calibrator is not to be used in hazardous areas, where vapour or gas leaks, etc. may constitute a danger of explosion.

The calibrator is not designed for operation in altitudes above 2000 meters.

The calibrator is a CLASS I product and must be connected to a mains outlet with a protective earth connection. Ensure the ground connection of the calibrator is properly connected to the protective earth before switching on the calibrator. Always use a mains power cable with a mains plug that connects to the protective earth.

To ensure the connection to protective earth any extension cord used must also have a protective earth conductor.

Only use a mains power cord with a current rating as specified by the calibrator and which is approved for the voltage and plug configuration in your area.

Before switching on the calibrator make sure that it is set to the voltage of the mains electricity supply.

Always position the calibrator to enable easy and quick disconnection of the power source (mains inlet socket).

The calibrator must be kept clear within an area of 20 cm on all sides and 1 metre above the calibrator due to fire hazard.

Never use heat transfer fluids such as silicone, oil, paste, etc. in the dry-block calibrators. These fluids may penetrate the calibrator and cause electrical hazard, damage or create poisonous fumes.

The calibrator must be switched off before any attempt to service the instrument is made. There are no user serviceable parts inside the calibrator.

When cleaning the well or the insertion tube, REMEMBER to wear goggles when using compressed air!

About the frontpanel:

The connectors, on the front panel of the calibrator, must NEVER be connected to a voltage source.

Thermostats connected to the switch test input must not be connected to any other voltage source during a test.

129375 00

2016-05-18

8

About insertion tubes and insulation plugs:

Never leave hot insertion tubes, which have been removed from the calibrator, unsupervised – they may constitute a fire hazard or personal injury.

If you intend to store the calibrator in the aluminium carrying case after use, you must ensure that the instrument has cooled down to a temperature below 50°C/122°F before placing it in the carrying case.

Never place a hot insertion tube in the optional carrying case.

Use only insulation plugs supplied by AMETEK Denmark A/S.

About the fuses:

The fuse box must not be removed from the power control switch until the mains cable has been disconnected.

The two main fuses must have the specified current and voltage rating and be of the specified type. The use of makeshift fuses and the short-circuiting of

fuse holders are prohibited and may cause a hazard.

Caution – Hot surface

This symbol is visible on the grid plate.

Do not touch the grid plate, the well or the insertion tube as the calibrator is heating up – they may be very hot and cause burns.

Do not touch the tip of the sensor when it is removed from the insertion tube/well

– it may be very hot and cause burns.

Do not touch the handle of the calibrator during use – it may be very hot and cause burns.

Over 50°C/122°F

If the calibrator has been heated up to temperatures above 50°C/122°F, you must wait until the instrument reaches a temperature below 50°C/122°F before you switch it off.

Do not remove the insert from the calibrator before the insert has cooled down to less than 50°C/122°F.

Caution – Cold surface

Below 0°C/32°F (applies only to the CTC-155 C models)

Do not touch the well or insertion tube when these are below 0°C/32°F - they might create frostbite.

If the calibrator has reached a temperature below 0°C/32°F, ice crystals may form on the insertion tube and the well. This, in turn, may cause the material surfaces to oxidize

To prevent this from happening the insertion tube and the well must be dried. This is done by heating up the calibrator to 100°C/212°F until all water left has evaporated.

Remove the insulation plug while heating up.

It is very important that humidity in the well and insertion tube is removed to prevent corrosion and frost expansion damages.

129375 00

2016-05-18

9

Caution…

About the use:

Do not use the instrument if the internal fan is out of order.

Before cleaning the calibrator, you must switch it off, allow it to cool down and remove all cables.

About the well, insertion tube and grid plate:

The well and the insertion tube must be clean and dry before use.

Do not pour any form of liquids into the well. It might damage the well or cause a hazard.

Scratches and other damage to the insertion tubes should be avoided by storing the insertion tubes carefully when not in use.

The insertion tube must never be forced into the well. The well could be damaged as a result, and the insertion tube may get stuck.

Before using new insertion tubes for the calibration, the insertion tubes must be heated up to maximum temperature 350°C(662°F) / 660°C (1220°F) for a period of minimum 30 minutes (CTC-350/660 only).

The insertion tube must always be removed from the calibrator after use.

The humidity in the air may cause corrosion oxidation on the insertion tube inside the instrument. There is a risk that the insertion tube may get stuck if this is allowed to happen.

If the calibrator is to be transported, the insertion tube must be removed from the well to avoid damage to the instrument.

Note…

The product liability only applies if the instrument is subject to a manufacturing defect. This liability becomes void if the user fails to follow the instructions set out in this manual or uses unauthorised spare parts.

129375 00

2016-05-18

10

3.0Setting up the calibrator for use

3.1Preparing the calibrator

Warning

The calibrator has been designed for indoor use only and is not to be used in wet locations.

The calibrator is not to be used in hazardous areas, where vapour or gas leaks, etc. may constitute a danger of explosion.

The calibrator is not designed for operation in altitudes above 2000 meters.

The calibrator is a CLASS I product and must be connected to a mains outlet with a protective earth connection. Ensure the ground connection of the calibrator is properly connected to the protective earth before switching on the calibrator.

Always use a mains power cable with a mains plug that connects to the protective earth.

To ensure the connection to protective earth any extension cord used must also have a protective earth conductor.

Only use a mains power cord with a current rating as specified by the calibrator and which is approved for the voltage and plug configuration in your area.

Before switching on the calibrator make sure that it is set to the voltage of the mains electricity supply.

Always position the calibrator to enable easy and quick disconnection of the power source (mains inlet socket).

The calibrator must be kept clear within an area of 20 cm on all sides and 1 metre above the calibrator due to fire hazard.

Note…

The instrument must not be exposed to draughts.

When setting up the calibrator, you must…

Place the calibrator on an even horizontal surface where you intend to use it.

Caution…

Do not use the instrument if the internal fan is out of order.

Ensure a free supply of air to the internal fan located at the bottom of the instrument (pos. 2). The area around the calibrator should be free of draught, dirt, flammable substances etc.

Check that the fuse size corresponds to the applied voltage on (pos. 3). The fuse is contained in the power control switch (on/off switch

(230V/115V)). To check; do as follows (see Fig. 1):

129375 00

2016-05-18

11

Warning

The two main fuses must have the specified current and voltage rating and be of the specified type. The use of makeshift fuses and the short-circuiting of fuse holders are prohibited and may cause a hazard.

Fig. 1

Changing fuses

A.Open the fuse box lid using a screwdriver.

B.Take out the fuse box.

C.Remove both fuses replacing them with two new fuses. These must be identical and should correspond to the line voltage. See section 11.0.

D.Slide the fuse box back into place.

Check that the earth connection for the instrument is present and attach the cable.

Select an insertion tube with the correct bore diameter. See section 3.2 for information on how to select insertion tubes.

The calibrator is now ready for use.

3.2Choosing an insertion tube

Caution…

To get the best results out of your calibrator, the insertion tube dimensions, tolerance and material are critical. We highly advise using the JOFRA insertion tubes, as they guarantee trouble free operation. Use of other insertion tubes may reduce performance of the calibrator and cause the insertion tube to get stuck.

129375 00

2016-05-18

12

Caution…

Before using new insertion tubes for the calibration in the CTC-350/660 instruments the insertion tubes must be heated up to maximum temperature 350°C(662°F) / 660°C (1220°F) for a period of minimum 30 minutes.

Insertion tubes are selected on the basis of the diameter of the sensor to be calibrated. Use the table for insertion tubes in section 3.2.1 to find the correct parts number.

Alternatively, you may order an undrilled insertion tube and drill the required hole yourself. The finished dimension should be as follows:

Sensor diameter ød+0.2 +0.05 / -0.00 mm.

Reference sensor hole ø4.2mm +0.05 / -0.00 mm

3.2.1Standard insertion tubes

Caution…

To get the best results out of your calibrator, the insertion tube dimensions, tolerance and material are critical. We highly advise using the JOFRA insertion tubes, as they guarantee trouble free operation. Use of other insertion tubes may reduce performance of the calibrator and cause the insertion tube to get stuck.

PARTS NO. FOR STANDARD INSERTION TUBES – SINGLE HOLES

 

Sensor

CTC-155

CTC-350

CTC-660

Sensor

CTC-155

CTC-350

CTC-660

 

size

size

 

 

 

 

 

 

 

 

3 mm

129407

129429

129459

Undrilled

129418

129445

129475

 

4 mm

129408

129430

129460

Undrilled/ref. hole

129419

129446

129476

 

5 mm

129409

129431

129461

1/8”

129420

129447

129477

 

6 mm

129410

129432

129462

3/16”

129421

129448

129478

 

7 mm

129411

129433

129463

1/4”

129422

129449

129479

 

8 mm

129412

129434

129464

5/16”

129423

129450

129480

 

9 mm

129413

129435

129465

3/8”

129424

129451

129481

 

10 mm

129414

129436

129466

7/16”

129425

129452

129482

 

11 mm

129415

129437

129467

1/2”

129426

129453

129483

 

12 mm

129416

129438

129468

9/16”

129427

129454

129484

 

13 mm

129417

129439

129469

5/8”

129428

129455

129485

 

14 mm

-

129440

129470

11/16”*

-

129456

129486

 

15 mm

-

129441

129471

13/16”*

-

129457

129487

 

16 mm*

-

129442

129472

3/4”*

-

129458

129488

 

18 mm*

-

129443

129473

 

 

 

 

 

20 mm*

-

129444

129474

 

 

 

 

 

 

 

 

 

 

 

 

 

The CTC-155 single-hole insertion tubes are delivered with a matching insulation plug.

*Note: Insertion tubes without reference holes.

129375 00

2016-05-18

13

PART NO. FOR STANDARD INSERTION TUBES – MULTI-HOLE

 

Description

CTC-155

CTC-350

CTC-660

 

 

 

 

 

 

 

Metric Type 1

 

 

 

 

 

Sensor size

 

 

 

 

 

3, 4, 5, 6, 9 mm and

129489

129491

129493

 

 

4mm REF

 

 

 

 

 

Inch Type 2

 

 

 

 

 

Sensor size

129490

 

 

 

 

1/8”, 3/16”, 1/4”, 3/8” and

129492

129494

 

 

4 mm REF

 

 

 

 

The CTC-155 multi-hole insertion tubes are delivered with a matching insulation plug.

3.3Inserting the sensor

Before inserting the sensor and switching on the calibrator, please note the following important warning:

Warning

Never use heat transfer fluids such as silicone, oil, paste, etc. in the dry-block calibrators.

These fluids may penetrate the calibrator and cause electrical hazard, damage or create poisonous fumes.

Never try to modify the insulation plugs to make them fit the sensor. Use only insulation plugs supplied by AMETEK Denmark A/S.

Insert the sensor as shown below in fig. 2.

Fig. 2

 

Inserting sensor and

Sensor under test

insertion tube

 

 

 

Insertion tube

Thermal protection shield

 

 

 

 

129375 00

2016-05-18

14

Check that the insulation plug fits the diameter of the sensor. Otherwise replace it (CTC-155 only).

In order to spare the sensor and its connections it is recommended to use a thermal protection shield (129264) at high temperatures (CTC-350/660 only).

Caution…

The well and the insertion tube must be clean before use.

Do not pour any form of liquids in the well. It might damage the well.

Scratches and other damage to the insertion tubes should be avoided by storing the insertion tubes carefully when not in use.

The insertion tube must never be forced into the well. The well could be damaged as a result, and the insertion tube may get stuck.

Caution – Hot surface

Do not touch the grid plate, the well or the insertion tube as the calibrator is heating up – they may be very hot and cause burns.

Do not touch the tip of the sensor when it is removed from the insertion tube/well

– it may be very hot and cause burns.

Do not touch the handle of the calibrator during use – it may be very hot and cause burns.

Do not remove the insertion tube from the calibrator before the insertion tube has cooled down to less than 50°C/122°F.

Caution – Cold surface

Below 0°C/32°F (applies only to the CTC-155 models)

Do not touch the well or insertion tube when these are below 0°C/32°F - they might create frostbite.

If the calibrator has reached a temperature below 0°C/32°F, ice crystals may form on the insertion tube and the well. This, in turn, may cause the material surfaces to oxidize

To prevent this from happening the insertion tube and the well must be dried. This is done by heating up the calibrator to 100°C/212°F until all water left has evaporated.

Remove the insulation plug while heating up.

It is very important that humidity in the well and insertion tube is removed to prevent corrosion and frost expansion damages.

129375 00

2016-05-18

15

3.4Programming intelligent STS sensors

Use the configuration software CON050 supplied with CTC to program and to update calibration information in intelligent STS sensors.

For instructions read the software manual for CON050 installed on the USB key.

129375 00

2016-05-18

16

Ametek CTC-155A, CTC-350A, CTC-652A, CTC-660A, CTC-155C Operating Manual

4.0Calibrator Interface

4.1Keypad - Functions

Function keys F1, F2, F3, F4

To operate the horizontal menu bar use the Functions keys.

System menu key

Press the key to enter the System menu.

Back key

Press the key to cancel a selection or return to previous menu.

Arrow Keys

Serve different functions depending on the mode of operation.

Navigation mode: Use the four keys to move the cursor in the desired direction.

Edit mode: The Up and Down Arrow keys scroll through the lists of options. If entering a number, the Left and Right Arrow keys move the cursor one character in the desired direction.

Action key / Enter key

Action function: Open and close edit fields or a menu button. The action key also accepts the selected option or entered value.

Enter function: Accept selected options or entered values. When a value is entered with the Enter Key the cursor selects the next configurable field in the list.

4.2Display - Functions

Reference sensor info

 

Memory reading

 

 

Warning/Error

Shows the reference sensor

 

Shows the current

 

 

symbols. See 4.2.2

selected. The serial number

 

memory selected

 

 

 

 

 

 

 

 

 

 

 

of the external reference

 

from the System

 

 

Real Time Clock and

sensor is read from the

 

menu.

 

 

date display.

intelligent reference sensor

 

 

 

 

 

 

 

 

 

 

 

 

 

and displayed in this field.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

“Time to stable”

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Selection

 

 

 

 

 

 

 

Shows the selected

 

 

 

 

 

 

 

specified stability criteria

 

 

 

 

 

 

 

and states a time when the

 

 

 

 

 

 

 

stable situation can be

 

 

 

 

 

 

 

achieved.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Process Indicator

 

Horizontal menu bar

 

 

 

Indicates the status of the

 

Provides you with the relevant menu

current process.

 

options that can be selected at the

 

 

present point. Each option can be

 

 

selected and activated by pressing the function keys (F1, F2, F3 and F4).

True temperature reading

Shows the numeric value of the temperature being measured. Can be either the internal reference sensor or an external reference sensor.

Units

Shows the unit of the current measurement.

Internal Reference temperature reading

Shows the temperature of the well.

Set temperature reading

Shows the numeric value of the current set temperature selected.

129375 00

2016-05-18

17

Loading...
+ 40 hidden pages