No p art o f this manu al may be re produce d in
any form or by any means (including electronic storage and retrieval or translation
into a foreign language) without prior agreement and written consent from Agilent
Technologies, Inc. as governed by United
States and international copyright laws.
Manual Part Number
G1978-90070
Edition
First Edition, December 2008
Printed in USA
Agilent Technologies, Inc.
5301 Stevens Creek Blvd.
Santa Clara, CA 95051 USA
Windows® and MS Windows® are U.S.
registered trademarks of Microsoft Corporation.
Windows NT
mark of Microsoft Corporation.
® is a U.S. registered trade-
Warranty
The material contained in this document is provided “as is,” and is subject to being changed, without notice,
in future editions. Further, to the maximum extent permitted by applicable
law, Agilent disclaims all warranties,
either express or implied, with regard
to this manual and any information
contained herein, including but not
limited to the implied warranties of
merchantability and fitness for a particular purpose. Agilent shall not be
liable for errors or for incidental or
consequential damages in connection with the furnishing, use, or performance of this document or of any
information contained herein. Should
Agilent and the user have a separate
written agreement with warranty
terms covering the material in this
document that conflict with these
terms, the warranty terms in the separate agreement shall control.
Technology Licenses
The hardware and/or software described in
this document are furnished under a license
and may be used or copied only in accordance with the terms of such license.
Restricted Rights Legend
U.S. Government Restricted Rights. Software and technical data rights granted to
the federal government include only those
rights customarily provided to end user customers. Agilent provides this customary
commercial license in Software and technical data pursuant to FAR 12.211 (Technical
Data) and 12.212 (Computer Software) and,
for the Department of Defense, DFARS
252.227-7015 (Technical Data - Commercial
Items) and DFARS 227.7202-3 (Rights in
Commercial Computer Software or Computer Software Documentation).
Safety Notices
CAUTION
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like
that, if not correctly performed or
adhered to, could result in damage
to the product or loss of important
data. Do not proceed beyond a
CAUTION notice until the indicated
conditions are fully understood and
met.
WARNING
A WARNING notice denotes a
hazard. It calls attention to an
operating procedure, practice, or
the like that, if not correctly performed or adhered to, could result
in personal injury or death. Do not
proceed beyond a WARNING
notice until the indicated conditions are fully understood and
met.
2Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide
In This Guide
1
2
3
This guide explains how to install, maintain and troubleshoot your
nanoelectospray ion source.
Installation
This chapter tells you how to install the multimode ion source.
Verificat ion
This chapter tells you how to install the multimode ion source.
Methods
This chapter describes basic operation and maintenance for the multimode
ion source.
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide3
4Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide
Content
1Installation 7
Installation 8
Step 1. Prepare to install 8
Step 2. Install the HV control PCA and cables 9
Changing Sources 13
To remove the multimode source 13
To convert from multimode to ESI, APCI or APPI 14
To convert from ESI, APCI or APPI to the multimode source 15
2Verification 21
To verify sensitivity for 6120 Quad - G1978B Multimode, Positive SIM
Modes 22
To verify sensitivity for 6130 and 6140 Quads - G1978B Multimode, Positive SIM
Modes 24
To verify sensitivity for 6130 and 6140 Quads - G1978B Multimode, Positive
Scan Modes 27
To verify sensitivity for the G1978B Multimode using Multiple FIA method in
Mixed Mode operation 30
To do an autotune 32
Example of multimode verification report 33
3Methods 35
To setup a method to use the multimode source 36
To create a method for positive/negative mixed mode operation 37
To create a method for alternating ESI and APCI operation 39
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide5
Contents
6Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide
Agilent G1978B Multimode Source for 6100 Series Single Quad LC/MS
Set-Up Guide
1
Installation
Installation 8
Step 1. Prepare to install 8
Step 2. Install the HV control PCA and cables 9
Changing Sources 13
To remove the multimode source 13
To convert from multimode to ESI, APCI or APPI 14
To convert from ESI, APCI or APPI to the multimode source 15
This chapter contains instructions to install the multimode source on a 6100
Series Single Quad LC/MS system, and also to remove and replace the source.
Agilent Technologies
7
1Installation
Installation
Installation
This section explains how to install the G1978B multimode source on a 6100
Series Single Quad LC/MS instrument. The G1978A multimode source is not
supported on these instruments.
The 6100 Series Single Quad LC/MS with multimode source is supported on
ChemStation B.03.01 or higher, or the ChemStation B.01.03 with a patch.
Step 1. Prepare to install
The Multimode Enablement Kit, G1978-60451, is shipped with the multimode
source. This kit needs to be installed before the multimode source is used.
Note that the multimode source and its accessories are to be installed by an
Agilent Customer Engineer.
1 Check that the Multimode Enablement Kit contains the following parts:
• Multimode Bd HV cable, p/n G1960-60858
• Multimode HV PCA, p/n G1960-61015
• Multimode Bd Power/Data Cable, p/n G1960-60873
Figure 1From left to right: G1960-60858, G1960-61015 and G1960-60873
2 Install the APCI Enablement Kit, G1947-60451, which is shipped with the
multimode source.
The APCI Enablement kit contains the following parts:
• Fast APCI HV Supply, p/n G1946-80058
8Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide
Step 2. Install the HV control PCA and cables
• Valve BD-APCI supply, p/nG1960-60802
• Valve BD-APCI Needle Interlock, p/n G1960-60856
Figure 2From left to right: G1946-80058, G1960-60802 and G1960-60856
Step 2. Install the HV control PCA and cables
1 Turn off the system power and remove the system power cord.
Keep the power cord intact if the vacuum control switch box is used. The
switch box is intended to keep the vacuum on while a CE works on the
electronics. The switch box is for CE use only.
2 Remove the CDS cover, top, side, front, and the Aux Module cover.
3 Disconnect the ribbon cable that connects the valve PCA to the
Vcap/Vchamber power supply. Then disconnect the Vcap and Vchamber
cable from the power supply.
Installation1
Figure 3Disconnecting the Vcap/Vchamber power supply from the valve PCA (left) and the Vcap/Vchamber.
4 Place the multimode HV power supply PCA in the slot between the valve
PCA and the Vcap/Vchamber. Secure the board by pressing it down into its
slot and then attach it with two screws.
5 Connect the short gray cable from valve PCA to the multimode HV power
supply.
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide9
1Installation
Step 2. Install the HV control PCA and cables
Figure 4Connecting the valve PCA to the multimode HV power supply.
6 Install the APCI HV power supply. The APCI HV power supply is located at
the end of the AUX Module.
7 Connect ribbon cable between valve PCA and Vcap/Vchamber power
supply.
Figure 5Connecting the valve PCA to the Vcap/Vchamber power supply.
8 Connect the Vcap and Vchamber cables to the Vcap/Vchamber power
supply.
10Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide
Installation1
Step 2. Install the HV control PCA and cables
Figure 6Connecting the Vcap and Vchamber cables to the power supply.
9 Connect the long ribbon cable, p/n G1960-60802, from the APCI HV power
supply to the valve PCA.
Figure 7Connecting the APCI HV power supply to the valve PCA.
10 Insert one end of the APCI Needle Interlock cable, G1960-60856, through
the slot at the front of the system and then plug it to the APCI HV
connector. Attach the other end to the chassis with the o-ring and the nut.
Figure 8Connecting the APCI HV to the chassis.
11 Insert the cable, G1960-60858, to the top slot and attach it to the chassis.
Plug the other two ends into the multimode HV PCA.
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide11
1Installation
Step 2. Install the HV control PCA and cables
Figure 9Connecting the HV PCA to the chassis.
12 Close the AUX Module cover and reconnect all cables.
13 Install the multimode source to the system and connect all connectors.
Figure 10Installing the multimode source (left) and connecting all connectors.
14 Put back the side, top, front and CDS cover.
15 Plug the system power cord back on and turn the front switch on.
The pump down process will start.
16 Start the ChemStation program.
17 Click Method and Run Control View > MSD > Spray Chamber and verify
that the source is an MM Source.
18 Go to MSD Tune window and start the Auto tune with a multimode source.
19 Verify that the multimode source is working correctly.
12Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide
Changing Sources
To remove the multimode source
Do the following steps to remove the multimode source.
1 Set the source temperatures for vaporizer heater and drying gas heater to
minimum values to cool the source.
Use the Tune > Instrument > Edit Spray Chamber menu item to display
the Edit Spray Chamber dialog box. Set the drying gas f low, nebulizer gas
flow, drying gas temperature and the vaporizer temperature to minimum
values.
Installation1
Changing Sources
WARNING
WARNING
WARNING
Do not touch the multimode source or the capillary cap. They may be very hot. Let
the parts cool before you handle them.
Never touch the source surfaces, especially when you analyze toxic substances or
when you use toxic solvents. The source has several sharp pieces which can pierce
your skin including the APCI corona needle, vaporizer sensor and counter current
electrode.
Do not insert fingers or tools through the openings on the multimode chamber. When
in use, the capillary and capillary cap are at high voltages up to 4 kV.
2 Wait approximately 20 minutes or until the source is cool.
3 Open the CDS door at the front of the MS to access the cables.
4 Disconnect the ESI high voltage charging electrode cable.
5 Disconnect the APCI Needle Interlock and multimode HV cable.
6 Unscrew the nebulizer gas line from the nebulizer.
7 Unscrew the LC sample tubing from the nebulizer.
8 Open the latch on the source and open the source.
9 Remove the multimode source from the spray chamber mount.
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide13
1Installation
To convert from multimode to ESI, APCI or APPI
10 Place the source shipping cover on the source.
To convert from multimode to ESI, APCI or APPI
WARNING
Never touch the source surfaces, especially when you analyze toxic substances or
when you use toxic solvents. The source has several sharp pieces which can pierce
your skin including the APCI corona needle, vaporizer sensor and counter current
electrode.
1 If the source to be installed is an APPI source, disconnect the multimode
high voltage PCA RS-232 serial cable from the serial port B connector of
Smart Card.
2 Unscrew and remove the multimode spray shield with the field shaping
electrodes.
3 Install the new source and the standard spray shield, making sure that the
hole in the spray shield is in the 12 o'clock position.
4 For APCI and APPI ion source, connect the vaporizer heater cable and the
APCI high voltage cable. For the APPI source, connect the RS-232 cable to
the serial port B connector of Smart Card.
5 For all sources, reconnect the nebulizer gas line tubing and the LC/MS
sample tubing.
14Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide
Loading...
+ 30 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.