AutoPulse® Quick Reference Guide
For Adult (>–18 years) Non-Traumatic Cardiac Arrest
Maximum Patient Weight 300 lbs.
•Remove ALL clothing from torso (both front and back) to ensure skin-to-platform contact
•Align armpits onto yellow line on Platform
1.Power up AutoPulse
2.Close chest bands
Do not twist the |
|
|
|
Maintain chest bands at |
|||
chest bands |
90 degrees to Platform |
||
|
and free of obstructions |
3. Press CONTINUE (green button)
4. Press START (green button) to begin compressions
To pause or stop operation press STOP (orange button)
Follow all local protocols and procedures.
Review User Guides and complete in-service training.
Troubleshooting
For Fault/User Advisory
•Lift up and fully extend both chest bands.
•Check both lateral and vertical patient alignment.
•Verify that chest bands are not twisted, are 90 degrees to the Platform and are free of obstructions.
•Press RESTART (green button) and follow
on-screen instructions to begin compressions.
If you cannot rectify problem immediately open chest bands and revert to manual CPR.
LifeBand® instructions on other side |
P/N 10596-001 Rev. 10 |