ZOLL AutoPulse User Manual

4 (1)
ZOLL AutoPulse User Manual

AutoPulse® Quick Reference Guide

For Adult (>18 years) Non-Traumatic Cardiac Arrest

Maximum Patient Weight 300 lbs.

Remove ALL clothing from torso (both front and back) to ensure skin-to-platform contact

Align armpits onto yellow line on Platform

1.Power up AutoPulse

2.Close chest bands

Do not twist the

 

 

 

Maintain chest bands at

chest bands

90 degrees to Platform

 

and free of obstructions

3. Press CONTINUE (green button)

4. Press START (green button) to begin compressions

To pause or stop operation press STOP (orange button)

Follow all local protocols and procedures.

Review User Guides and complete in-service training.

Troubleshooting

For Fault/User Advisory

Lift up and fully extend both chest bands.

Check both lateral and vertical patient alignment.

Verify that chest bands are not twisted, are 90 degrees to the Platform and are free of obstructions.

Press RESTART (green button) and follow

on-screen instructions to begin compressions.

If you cannot rectify problem immediately open chest bands and revert to manual CPR.

LifeBand® instructions on other side