–When using a fundus imaging system (e.g. BIOM II)14
–Phototoxic retinal injury in eye surgery14
–Warning labels and notes18
Description19
VISU 200 BrightFlex™ Surgical Microscope20
–Intended use20
–Description of the modules20
–Illumination system28
–Controls, displays, connections30
–Hand grips (option)38
–Tube and eyepieces for main microscope40
–Tube and eyepieces for assistant's microscope42
Preparations for use45
–Mounting the binocular tubes, eyepieces and the objective lens46
Operation49
Preparations for use50
–Adjusting the tilt angle50
–Setting the microscope tilt to angles greater than 15°52
–Adjusting the surgical microscope53
–Checklist54
–When using a fundus imaging system (e.g. BIOM II)56
G-30-1439-enOPMI® VISU 200 BrightFlex™ Surgical MicroscopeIssue 3.0
Printed on 21. 11. 2002
Procedure57
What to do in an emergency59
–Failure of zoom system59
–Failure of X-Y coupling60
–Failure of focusing system60
Maintenance / Further information61
–Trouble-shooting table62
–Magnifications / Fields of view64
–Care of the unit65
–Sterilization66
–Ordering data67
–Spare parts68
–Accessories68
–Adapter cables70
Technical data71
–Technical data72
–Ambient requirements74
Index75
G-30-1439-enOPMI® VISU 200 BrightFlex™ Surgical MicroscopeIssue 3.0
Printed on 21. 11. 2002
Functions at a glance5
Functions at a glance
VISU 200 BrightFlex™ surgical microscope6
Key to symbols7
G-30-1439-enOPMI® VISU 200 BrightFlex™ Surgical MicroscopeIssue 3.0
Printedon21.11.2002
6Functions at a glance
VISU 200 BrightFlex™ surgical microscope
1Resetting the X-Y coupling, focus and zoom to their initial
positionsPage 30
2Adjusting the tilt of the surgical microscopePage 50
3Setting the interpupillary distancePage 40
4Adjusting the eyecupsPage 40
5Setting your prescriptionPage 40
6Display of the magnification factor of the zoom systemPage 32
7Setting the red reflex illumination (+2° and -2°)Page 28
8Releasing the magnetic brakes of the suspension system Page 32
9Changing the magnification on assistant's microscopePage 34
10 Focusing of assistant's microscopePage 34
11 Setting the 6° illuminationPage 36
12 Arrows indicating the focusing rangePages
13 Locking screw for assistant's microscope (vertical)Page 34
14 Selecting light stopsPage 36
15 Connecting the light guidePage 36
16 Locking screw for assistant's microscope (horizontal)Page 36
30, 56
1
2
3
4
5
7910 5411129 131415 161088
6
G-30-1439-enOPMI® VISU 200 BrightFlex™ Surgical MicroscopeIssue 3.0
Printed on 21. 11. 2002
Key to symbols
Functions at a glance7
Different symbols used in this user's manual draw your attention to safety
aspects and useful tips. The symbols are explained in the following.
Warning!
The warning triangle indicates potential sources of danger which may
constitute a risk of injury for the user or a health hazard.
Caution:
The square indicates situations which may lead to malfunction, defects,
collision or damage of the instrument.
OPMI
Note:
The hand indicates hints on the use of the instrument or other tips for the
user.
®
®
OPMI
is a registered trademark of Carl Zeiss.
G-30-1439-enOPMI® VISU 200 BrightFlex™ Surgical MicroscopeIssue 3.0
Printedon21.11.2002
8Functions at a glance
G-30-1439-enOPMI® VISU 200 BrightFlex™ Surgical MicroscopeIssue 3.0
Printed on 21. 11. 2002
Safety
Safety9
Directives and standards10
Notes on installation and use11
When using a fundus imaging system (e.g. BIOM II)14
Phototoxic retinal injury in eye surgery14
Warning labels and notes18
G-30-1439-enOPMI® VISU 200 BrightFlex™ Surgical MicroscopeIssue 3.0
Printedon21.11.2002
10Safety
The instrument described in this manual has been developed and tested
in accordance with Carl Zeiss safety standards and with national and international regulations. A high degree of instrument safety is thus ensured.
We would like to inform you on the safety aspects involved in operating
the instrument. This chapter contains a summary of the most important
precautions to be observed.
Further safety notes are also contained in other parts of this user's
manual; they are marked with a warning triangle containing an exclamation mark as shown here. Please pay special attention to these safety
notes.
Safety is only ensured when this instrument is operated properly. Please
read through this manual carefully before turning the instrument on. Also
read through the user's manuals of the other equipment used with this instrument. You may obtain further information from our service organization or authorized representatives.
Directives and standards
The instrument described in this manual has been designed in compliance with the following standards:
In accordance with Directive 93/42/EEC, the complete quality management system of the company Carl Zeiss has been certified by the DQS
Deutsche Gesellschaft zur Zertifizierung von Managementsystemen
mbH, a notified body, under registration number 250758 MR2.
•The instrument must be connected to a special emergency backup
•This is a class I instrument as defined by Directive 93/42 /EEC.
•Please observe all applicable accident prevention regulations.
–EN
–IEC
–UL
–CSA
line supply in accordance with the regulations or directives which apply in your country.
G-30-1439-enOPMI® VISU 200 BrightFlex™ Surgical MicroscopeIssue 3.0
Printed on 21. 11. 2002
Safety11
Notes on installation and use
Safe working order
•Do not operate the equipment contained in the delivery package in
–explosion-risk areas,
–the presence of inflammable anesthetics or volatile solvents such
as alcohol, benzine or similar chemicals.
•Do not station or use the instrument in damp rooms. Do not expose
the instrument to water splashes, dripping water or sprayed water.
•Immediately unplug any equipment that gives off smoke, sparks or
strange noises. Do not use the instrument until our service representative has repaired it.
•Do not place any fluid-filled containers on top of the instrument. Make
sure that no fluids can seep into the instrument.
•Do not force cable connections. If the male and female parts do not
readily connect, make sure that they are appropriate for one another.
If any of the connectors are damaged, have our service representative
repair them.
•Potential equalization: The instrument can be incorporated into potential equalization measures. For this purpose, contact our service department.
•Do not use a mobile phone in the vicinity of the equipment because
the radio interference can cause the equipment to malfunction. The effects of radio interference on medical equipment depend on a number
of various factors and are therefore entirely unforeseeable.
•Modifications and repairs on these instruments or instruments used
with them may only be performed by our service representative or by
other authorized persons.
•The manufacturer will not accept any liability for damage caused by
unauthorized persons tampering with the instrument; this will also forfeit any rights to claim under warranty.
•Use this instrument only for the applications described.
•Only use the instrument with the accessories supplied. Should you
wish to use other accessory equipment, make sure that Carl Zeiss or
the equipment manufacturer has certified that its use will not impair
the safety of instrument.
G-30-1439-enOPMI® VISU 200 BrightFlex™ Surgical MicroscopeIssue 3.0
Printedon21.11.2002
12Safety
•Only personnel who have undergone training and instruction are allowed to use this instrument. It is the responsibility of the customer or
institution operating the equipment to train and instruct all staff using
the equipment.
•Keep the user's manuals where they are easily accessible at all times
for the persons operating the instrument.
•Never look at the sun through the binocular tube, the objective lens or
an eyepiece.
•Do not pull at the light guide cable, at the power cord or at other cable
connections.
•This instrument is a high-grade technological product. To ensure opti-
mum performance and safe working order of the instrument, its safety
must be checked once every 12 months. We recommend having this
check performed by our service representative as part of regular maintenance work.
If a failure occurs which you cannot correct using the trouble-shooting
table, attach a sign to the instrument stating it is out of order and contact our service representative.
Requirements for operation
Our service representative or a specialist authorized by us will install the
instrument. Please make sure that the following requirements for operation remain fulfilled in the future:
–All mechanical connections (details in the user's manual) which are
relevant to safety are properly connected and screw connections tightened.
–All cables and plugs are in good working condition.
–The voltage setting on the instrument conforms to the rated voltage of
the line supply on site.
–The instrument is plugged into a power outlet which has a properly
Before every use and after re-equipping the instrument
•Make sure that all ”Requirements for operation” are fulfilled.
•Go through the checklist.
•Re-attach or close any covers, panels or caps which have been re-
moved or opened.
G-30-1439-enOPMI® VISU 200 BrightFlex™ Surgical MicroscopeIssue 3.0
Printed on 21. 11. 2002
Safety13
•Pay special attention to warning symbols on the instrument (triangular
warning signs with exclamation marks), labels and any parts such as
screws or surfaces painted red.
For every use of the instrument
•Avoid looking directly into the light source, e.g. into the microscope objective lens or a light guide.
•Any kind of radiation has a detrimental effect on biological tissue.This
also applies to the light illuminating the surgical field. Please therefore
reduce the brightness and duration of illumination on the surgical field
to the absolute minimum required.
•When operating on the eye, always use a GG 475 protection filter to
ensure that the patient's retina is not exposed to unnecessary (blue)
radiation (retinal injury).
G-30-1439-enOPMI® VISU 200 BrightFlex™ Surgical MicroscopeIssue 3.0
Printedon21.11.2002
14Safety
When using a fundus imaging system (e.g. BIOM II)
When using a fundus imaging system (e.g. BIOM II from the company Oculus) which is usually installed between the surgical microscope and the
patient, make sure that the patient is neither put at risk nor injured by the
motorized focusing system or the movement of the stand arm.
Only use accessories expressly certified by the manufacturer for combination with the surgical microscope described in this manual.
Caution!
Risk of collision!
•With the fundus imaging system swung out of position, always position
the microscope body in such a way that index dot (1) of the microscope's focus is in the middle of triangle (2) of the marking.
•Select a medium magnification (e.g. 1.0).
•Lower the surgical microscope towards the surgical field until you see
1
2
the patient's cornea sharply defined.
•Turn the screw for limiting the downward movement clockwise as far
as it will go.
•It is vital that you read the user’s manual on the fundus imaging sys-
tem (e.g. BIOM II from the company Oculus).
Phototoxic retinal injury in eye surgery
General
Several papers have been published dealing with the problems of phototoxicity during eye surgery. A comprehensive review of these publications
reveals five aspects of particular concern:
G-30-1439-enOPMI® VISU 200 BrightFlex™ Surgical MicroscopeIssue 3.0
Printed on 21. 11. 2002
Safety15
In the following, comments on these aspects are given and a description
of how Carl Zeiss, as a manufacturer, makes allowance for them in its instruments.
Studies on exposure of the eye to light of varying spectral composition
date back to the early 1950s. These studies suggest that the potential
hazard of phototoxic injury to the patient's retina can be reduced by
blocking out the blue and ultraviolet light below a wavelength of 475 nm.
Carl Zeiss provides a GG 475 retina protection filter for surgical microscopes recommended for use in ophthalmic surgery. This reduces not
only the light exposure of the patient's eye, but also that of the surgeon's.
It should be noted in this context that the use of filters inevitably leads to
a change in the color of the light. The surgeon may therefore have to get
used to the changed appearance of anatomical structures.
Intensity of illumination
The majority of researchers suggest that the surgeon should use the
lowest light intensity necessary to guarantee good viewing during surgery.
Carl Zeiss has addressed this concern by providing a device for continuously varying the brightness of the light source. This permits the surgeon
to optimally adapt the light intensity at the patient's eye to the conditions
existing in each case. Carl Zeiss strongly discourages the use of xenon or
other high-intensity light sources in ophthalmology.
Angle of illumination
A number of publications suggest that the microscope should be tilted to
reduce the exposure of the macula to direct illumination.
Carl Zeiss ophthalmic surgical microscopes are therefore equipped with
the following:
–Tilting mechanism for the microscope body
–Oblique illumination with brightness control
Focus of the light source
Studies show that injuries are likely to occur if the filament of the light
source is imaged on the patient's retina. The peak intensity of a filament
is considerably higher than the peak intensity of an even and extended
light source such as a fiber guide.
G-30-1439-enOPMI® VISU 200 BrightFlex™ Surgical MicroscopeIssue 3.0
Printedon21.11.2002
16Safety
This is the reason why fiber optic illumination is used in surgical microscopes from Carl Zeiss.
Exposure time to light
According to some publications, the phakic or aphakic eye should not be
exposed to the light source longer than a few minutes. In every operation
the exposure of the retina to light is dependent on the type and duration
of surgery and on any complications which occur. It is therefore recommended in ophthalmic surgery to keep the light intensity as low as possible, or to use a device which prevents the light from entering through the
patient's pupil. Also, the surrounding light sources should not cause additional strain to the patient's eye.
Carl Zeiss has provided an answer to this problem in the form of a swingin retinal protection device for insertion into the beam path of the surgical
microscope. This device ensures total eclipsing of the pupil, preventing
light from entering into the patient's eye. It can be swung out when a red
reflex is required.
Intensity scale
The intensity scale of our suspension system is calibrated in units of the
”Spectrally weighted radiance for the photochemical hazard of the phakic
eye (L
L
B
)”
B
is the spectral radiance integrated over the spectral range from 380 nm
to 700 nm and weighted with B(λ):
700
=
∑
L
B
L(λ)B(λ) ∆λ
380
in which B(λ) is the spectral weighting function for the photochemical hazard of the retina of the phakic eye.
The value L
=500mW/cm² sr is the reference value and is defined as
B
1.0 on the intensity scale of the suspension system as recommended in
the ISO 10936-2 standard draft for surgical microscopes used in ophthalmic applications. At this reference value, photoretinitis might be expected to occur from the surgical microscope light source in a total retinal
exposure time of 10 minutes. This applies to the exposure of a specific
point on the retina with an uninterrupted illumination beam. In a cataract
procedure, instruments such as the phacoemulsification handpiece, the
G-30-1439-enOPMI® VISU 200 BrightFlex™ Surgical MicroscopeIssue 3.0
Printed on 21. 11. 2002
Safety17
use of fluids in the eye, manipulation and movement of the eye, among
other factors, result in the interruption of the illumination from the surgical
microscope light source. Such factors would be expected to significantly
extend the time at which photoretinitis might occur.
In conclusion
Carl Zeiss recommends:
–Use of the GG 475 eye protection filter.
–Reduction of the illumination of the surgical area to the extent required
for the patient's safety and for microscopic visualization.
–Tilting of the microscope body as required.
–Insertion of the retinal protection device.
–Maximum reduction of the exposure of the patient's eye to light from
surrounding light sources.
These measures should help the surgeon to reduce the likelihood of phototoxic retinal injury.
Note:
The VISU 150 microscope always contains a UV blocking filter.
The use of this filter ensures that the illumination intensity lies below 50
µW/cm
2
in the range between 305 nm and 400 nm.
These measures help the surgeon to reduce the risk of phototoxic retinal
injury.
G-30-1439-enOPMI® VISU 200 BrightFlex™ Surgical MicroscopeIssue 3.0
Printedon21.11.2002
18Safety
Warning labels and notes
Caution:
Observe all warning labels and notes!
If any label is missing on your instrument or has become illegible, please
contact us or one of our authorized representatives. We will supply the
missing labels.
G-30-1439-enOPMI® VISU 200 BrightFlex™ Surgical MicroscopeIssue 3.0
Printed on 21. 11. 2002
Description
Description19
VISU 200 BrightFlex™ Surgical Microscope20
Intended use20
Description of the modules20
Illumination system28
Controls, displays, connections30
Hand grips (option)38
Tube and eyepieces for main microscope40
Tube and eyepieces for assistant's microscope42
G-30-1439-enOPMI® VISU 200 BrightFlex™ Surgical MicroscopeIssue 3.0
Printedon21.11.2002
20Description
VISU 200 BrightFlex™ Surgical Microscope
Intended use
The VISU 200 BrightFlex™ surgical microscope has been designed for
surgical procedures in the field of ophthalmology, i.e. the microscope
meets the special requirements of this discipline.
Description of the modules
The VISU surgical microscope is comprised of the following modules:
1X-Y coupling
The X-Y coupling allows fine, motorized positioning of the surgical microscope in a horizontal plane. The range of travel is 40 mm x 40 mm.
The speed of travel can be set on the display field of the suspension
system.
The X-Y coupling is provided with a recentering mechanism. When
you press the actuator button,
–the X-Y coupling adopts its center position,
–the focusing system of the surgical microscope is reset to its initial
position and
–the zoom system will set a preselected magnification factor if the
XYZ-RESfunctionhasbeenactivated(possiblewithS8suspension systems only).
You can also trigger the recentering movement using the foot control
panel..
G-30-1439-enOPMI® VISU 200 BrightFlex™ Surgical MicroscopeIssue 3.0
Printed on 21. 11. 2002
Description21
1
G-30-1439-enOPMI® VISU 200 BrightFlex™ Surgical MicroscopeIssue 3.0
Printedon21.11.2002
22Description
2Support arm for the surgical microscope
The support arm incorporates a tilt device. This allows the viewing direction of the surgical microscope to be adapted to the requirements
of the surgical field. Using the knob for fine tilt, you can position the
surgical microscope in a range from +180° to -180° (+ in the direction
of the surgeon and - in the opposite direction). The +90° setting is ideal
for surgery on patients in a seated position or lying on their side.
Caution:
Do not tilt the main microscope beyond + / - 180°,asthiscould
damage the microscope cable or the light guide.
3Main microscope
The apochromatic optics of the main microscope provide superb optical quality. The microscope image displays optimum contrast and excellent detail recognition along with a large depth of field. The bright
microscope image is a particular benefit in vitreoretinal surgery. A 1:6
ratio zoom system allows the magnification of the overall system to be
set as required by the surgical procedure.
Two apochromatic objective lenses with focal lengths of 175 mm and
200 mm are available for different working distances.
A180° tiltable tube is used as a viewing device for the surgeon. The
large tilt range allows work with minimum fatigue.
The standard equipment includes eyepieces with a magnification
factor of 12.5x (option: 10x).
The illumination system has been designed for use in ophthalmology.
A light guide directs the light from the light source in the suspension
system to the surgical microscope.
6° illumination can be continuously faded in. This allows the illumination intensity in the surgical field to be adjusted continuously, without
changing the color of the light. The illumination of the surgical field at
an angle of 6° produces an image with an outstanding impression of
depth.
An additional illumination system produces an intensive red reflex
even when the eye is decentered. The angle of illumination can be
switched from +2° to -2°.
G-30-1439-enOPMI® VISU 200 BrightFlex™ Surgical MicroscopeIssue 3.0
Printed on 21. 11. 2002
Description23
2
3
G-30-1439-enOPMI® VISU 200 BrightFlex™ Surgical MicroscopeIssue 3.0
Printedon21.11.2002
24Description
To protect the patient's eye against photo-retinitis, a retinal protection device is provided. This device can be swung into the illumination beam
path, if no red reflex is required.
At the light source integrated in the stand, a GG 475 eye protection filter
can be swung into the illumination beam path. This filter markedly reduces
the exposure of the patient's and surgeon's eyes to radiation.
A video camera from our MediLive video camera line can be optionally
connected to the VISU 200 surgical microscope. The light loss for the surgeon is only 20% and cannot be perceived subjectively.
Caution!
–Avoid looking directly into the light source, e.g. into the microscope ob-
jective lens or into the light guide!
–When selecting the brightness level for the patient's eye, always take
caretokeepthestrainonthepatient'seyetoaminimum.
–If the red reflex is not necessary, move the retinal protection device
into the beam path. Only use the retro-illumination contrast stop, if the
surgical procedure requires a red reflex.
–When operating on the eye, always use a GG 475 eye protection filter
to ensure that the patient's retina is not exposed to unnecessary (blue)
radiation (retinal injury)!
G-30-1439-enOPMI® VISU 200 BrightFlex™ Surgical MicroscopeIssue 3.0
Printed on 21. 11. 2002
Loading...
+ 56 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.