Zeiss OPMI VISU 200 User manual

Page 1
OPMI
Surgical Microscope
®
VISU 200 BrightFlex™
Instructions for use
G-30-1439-en
Issue 3.0
Printed on 21. 11. 2002
Page 2
Page 3
Contents
Functions at a glance 5
VISU 200 BrightFlex™ surgical microscope 6
Key to symbols 7
Safety 9
Directives and standards 10
Notes on installation and use 11
When using a fundus imaging system (e.g. BIOM II) 14
Phototoxic retinal injury in eye surgery 14
Warning labels and notes 18
Description 19
VISU 200 BrightFlex™ Surgical Microscope 20
Intended use 20
Description of the modules 20
Illumination system 28
Controls, displays, connections 30
Hand grips (option) 38
Tube and eyepieces for main microscope 40
Tube and eyepieces for assistant's microscope 42
Preparations for use 45
Mounting the binocular tubes, eyepieces and the objective lens 46
Operation 49
Preparations for use 50
Adjusting the tilt angle 50
Setting the microscope tilt to angles greater than 15° 52
Adjusting the surgical microscope 53
Checklist 54
When using a fundus imaging system (e.g. BIOM II) 56
Printed on 21. 11. 2002
Page 4
Procedure 57
What to do in an emergency 59
Failure of zoom system 59
Failure of X-Y coupling 60
Failure of focusing system 60
Maintenance / Further information 61
Trouble-shooting table 62
Magnifications / Fields of view 64
Care of the unit 65
Sterilization 66
Ordering data 67
Spare parts 68
Accessories 68
Adapter cables 70
Technical data 71
Technical data 72
Ambient requirements 74
Index 75
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
Page 5
Functions at a glance 5

Functions at a glance

VISU 200 BrightFlex™ surgical microscope 6
Key to symbols 7
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printedon21.11.2002
Page 6
6 Functions at a glance

VISU 200 BrightFlex™ surgical microscope

1 Resetting the X-Y coupling, focus and zoom to their initial
positions Page 30 2 Adjusting the tilt of the surgical microscope Page 50 3 Setting the interpupillary distance Page 40 4 Adjusting the eyecups Page 40 5 Setting your prescription Page 40 6 Display of the magnification factor of the zoom system Page 32 7 Setting the red reflex illumination (+2° and -2°) Page 28 8 Releasing the magnetic brakes of the suspension system Page 32 9 Changing the magnification on assistant's microscope Page 34 10 Focusing of assistant's microscope Page 34 11 Setting the 6° illumination Page 36 12 Arrows indicating the focusing range Pages
13 Locking screw for assistant's microscope (vertical) Page 34 14 Selecting light stops Page 36 15 Connecting the light guide Page 36 16 Locking screw for assistant's microscope (horizontal) Page 36
30, 56
1
2
3
4
5
7 9 10 5 4 11 12 9 13 14 15 16108 8
6
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
Page 7

Key to symbols

Functions at a glance 7
Different symbols used in this user's manual draw your attention to safety aspects and useful tips. The symbols are explained in the following.
Warning!
The warning triangle indicates potential sources of danger which may constitute a risk of injury for the user or a health hazard.
Caution:
The square indicates situations which may lead to malfunction, defects, collision or damage of the instrument.
OPMI
Note: The hand indicates hints on the use of the instrument or other tips for the user.
®
®
OPMI
is a registered trademark of Carl Zeiss.
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printedon21.11.2002
Page 8
8 Functions at a glance
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
Page 9

Safety

Safety 9
Directives and standards 10
Notes on installation and use 11
When using a fundus imaging system (e.g. BIOM II) 14
Phototoxic retinal injury in eye surgery 14
Warning labels and notes 18
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printedon21.11.2002
Page 10
10 Safety
The instrument described in this manual has been developed and tested in accordance with Carl Zeiss safety standards and with national and in­ternational regulations. A high degree of instrument safety is thus en­sured.
We would like to inform you on the safety aspects involved in operating the instrument. This chapter contains a summary of the most important precautions to be observed.
Further safety notes are also contained in other parts of this user's manual; they are marked with a warning triangle containing an exclama­tion mark as shown here. Please pay special attention to these safety notes.
Safety is only ensured when this instrument is operated properly. Please read through this manual carefully before turning the instrument on. Also read through the user's manuals of the other equipment used with this in­strument. You may obtain further information from our service organiza­tion or authorized representatives.

Directives and standards

The instrument described in this manual has been designed in compli­ance with the following standards:
In accordance with Directive 93/42/EEC, the complete quality manage­ment system of the company Carl Zeiss has been certified by the DQS Deutsche Gesellschaft zur Zertifizierung von Managementsystemen mbH, a notified body, under registration number 250758 MR2.
The instrument must be connected to a special emergency backup
This is a class I instrument as defined by Directive 93/42 /EEC.
Please observe all applicable accident prevention regulations.
–EN
–IEC
–UL
–CSA
line supply in accordance with the regulations or directives which ap­ply in your country.
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
Page 11
Safety 11

Notes on installation and use

Safe working order
Do not operate the equipment contained in the delivery package in
explosion-risk areas,
the presence of inflammable anesthetics or volatile solvents such
as alcohol, benzine or similar chemicals.
Do not station or use the instrument in damp rooms. Do not expose the instrument to water splashes, dripping water or sprayed water.
Immediately unplug any equipment that gives off smoke, sparks or strange noises. Do not use the instrument until our service represent­ative has repaired it.
Do not place any fluid-filled containers on top of the instrument. Make sure that no fluids can seep into the instrument.
Do not force cable connections. If the male and female parts do not readily connect, make sure that they are appropriate for one another. If any of the connectors are damaged, have our service representative repair them.
Potential equalization: The instrument can be incorporated into poten­tial equalization measures. For this purpose, contact our service de­partment.
Do not use a mobile phone in the vicinity of the equipment because the radio interference can cause the equipment to malfunction. The ef­fects of radio interference on medical equipment depend on a number of various factors and are therefore entirely unforeseeable.
Modifications and repairs on these instruments or instruments used with them may only be performed by our service representative or by other authorized persons.
The manufacturer will not accept any liability for damage caused by unauthorized persons tampering with the instrument; this will also for­feit any rights to claim under warranty.
Use this instrument only for the applications described.
Only use the instrument with the accessories supplied. Should you wish to use other accessory equipment, make sure that Carl Zeiss or the equipment manufacturer has certified that its use will not impair the safety of instrument.
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printedon21.11.2002
Page 12
12 Safety
Only personnel who have undergone training and instruction are al­lowed to use this instrument. It is the responsibility of the customer or institution operating the equipment to train and instruct all staff using the equipment.
Keep the user's manuals where they are easily accessible at all times for the persons operating the instrument.
Never look at the sun through the binocular tube, the objective lens or an eyepiece.
Do not pull at the light guide cable, at the power cord or at other cable connections.
This instrument is a high-grade technological product. To ensure opti-
mum performance and safe working order of the instrument, its safety must be checked once every 12 months. We recommend having this check performed by our service representative as part of regular main­tenance work. If a failure occurs which you cannot correct using the trouble-shooting table, attach a sign to the instrument stating it is out of order and con­tact our service representative.
Requirements for operation
Our service representative or a specialist authorized by us will install the instrument. Please make sure that the following requirements for opera­tion remain fulfilled in the future:
All mechanical connections (details in the user's manual) which are
relevant to safety are properly connected and screw connections tight­ened.
All cables and plugs are in good working condition.
The voltage setting on the instrument conforms to the rated voltage of
the line supply on site.
The instrument is plugged into a power outlet which has a properly
connected protective earth contact.
Thepowercordbeingusedistheonedesignedforusewiththisin-
strument.
Before every use and after re-equipping the instrument
Make sure that all Requirements for operationare fulfilled.
Go through the checklist.
Re-attach or close any covers, panels or caps which have been re-
moved or opened.
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0
Printed on 21. 11. 2002
Page 13
Safety 13
Pay special attention to warning symbols on the instrument (triangular warning signs with exclamation marks), labels and any parts such as screws or surfaces painted red.
For every use of the instrument
Avoid looking directly into the light source, e.g. into the microscope ob­jective lens or a light guide.
Any kind of radiation has a detrimental effect on biological tissue.This also applies to the light illuminating the surgical field. Please therefore reduce the brightness and duration of illumination on the surgical field to the absolute minimum required.
When operating on the eye, always use a GG 475 protection filter to ensure that the patient's retina is not exposed to unnecessary (blue) radiation (retinal injury).
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printedon21.11.2002
Page 14
14 Safety

When using a fundus imaging system (e.g. BIOM II)

When using a fundus imaging system (e.g. BIOM II from the company Oc­ulus) which is usually installed between the surgical microscope and the patient, make sure that the patient is neither put at risk nor injured by the motorized focusing system or the movement of the stand arm.
Only use accessories expressly certified by the manufacturer for combi­nation with the surgical microscope described in this manual.
Caution!
Risk of collision!
With the fundus imaging system swung out of position, always position the microscope body in such a way that index dot (1) of the micro­scope's focus is in the middle of triangle (2) of the marking.
Select a medium magnification (e.g. 1.0).
Lower the surgical microscope towards the surgical field until you see
1
2
the patient's cornea sharply defined.
Turn the screw for limiting the downward movement clockwise as far as it will go.
It is vital that you read the users manual on the fundus imaging sys- tem (e.g. BIOM II from the company Oculus).

Phototoxic retinal injury in eye surgery

General
Several papers have been published dealing with the problems of photo­toxicity during eye surgery. A comprehensive review of these publications reveals five aspects of particular concern:
Illumination characteristics (spectral composition)
Intensity of illumination
Angle of illumination
Focus of the light source
Exposure time to light
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printed on 21. 11. 2002
Page 15
Safety 15
In the following, comments on these aspects are given and a description of how Carl Zeiss, as a manufacturer, makes allowance for them in its in­struments.
Illumination characteristics (spectral composition)
Studies on exposure of the eye to light of varying spectral composition date back to the early 1950s. These studies suggest that the potential hazard of phototoxic injury to the patient's retina can be reduced by blocking out the blue and ultraviolet light below a wavelength of 475 nm.
Carl Zeiss provides a GG 475 retina protection filter for surgical micro­scopes recommended for use in ophthalmic surgery. This reduces not only the light exposure of the patient's eye, but also that of the surgeon's.
It should be noted in this context that the use of filters inevitably leads to a change in the color of the light. The surgeon may therefore have to get used to the changed appearance of anatomical structures.
Intensity of illumination
The majority of researchers suggest that the surgeon should use the lowest light intensity necessary to guarantee good viewing during sur­gery.
Carl Zeiss has addressed this concern by providing a device for continu­ously varying the brightness of the light source. This permits the surgeon to optimally adapt the light intensity at the patient's eye to the conditions existing in each case. Carl Zeiss strongly discourages the use of xenon or other high-intensity light sources in ophthalmology.
Angle of illumination
A number of publications suggest that the microscope should be tilted to reduce the exposure of the macula to direct illumination.
Carl Zeiss ophthalmic surgical microscopes are therefore equipped with the following:
Tilting mechanism for the microscope body
Oblique illumination with brightness control
Focus of the light source
Studies show that injuries are likely to occur if the filament of the light source is imaged on the patient's retina. The peak intensity of a filament is considerably higher than the peak intensity of an even and extended light source such as a fiber guide.
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printedon21.11.2002
Page 16
16 Safety
This is the reason why fiber optic illumination is used in surgical micro­scopes from Carl Zeiss.
Exposure time to light
According to some publications, the phakic or aphakic eye should not be exposed to the light source longer than a few minutes. In every operation the exposure of the retina to light is dependent on the type and duration of surgery and on any complications which occur. It is therefore recom­mended in ophthalmic surgery to keep the light intensity as low as pos­sible, or to use a device which prevents the light from entering through the patient's pupil. Also, the surrounding light sources should not cause addi­tional strain to the patient's eye.
Carl Zeiss has provided an answer to this problem in the form of a swing­in retinal protection device for insertion into the beam path of the surgical microscope. This device ensures total eclipsing of the pupil, preventing light from entering into the patient's eye. It can be swung out when a red reflex is required.
Intensity scale
The intensity scale of our suspension system is calibrated in units of the Spectrally weighted radiance for the photochemical hazard of the phakic eye (L
L
B
)
B
is the spectral radiance integrated over the spectral range from 380 nm
to 700 nm and weighted with B(λ):
700
=
L
B
L(λ)B(λ) ∆λ
380
in which B(λ) is the spectral weighting function for the photochemical ha­zard of the retina of the phakic eye.
The value L
=500mW/cm² sr is the reference value and is defined as
B
1.0 on the intensity scale of the suspension system as recommended in
the ISO 10936-2 standard draft for surgical microscopes used in oph­thalmic applications. At this reference value, photoretinitis might be ex­pected to occur from the surgical microscope light source in a total retinal exposure time of 10 minutes. This applies to the exposure of a specific point on the retina with an uninterrupted illumination beam. In a cataract procedure, instruments such as the phacoemulsification handpiece, the
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printed on 21. 11. 2002
Page 17
Safety 17
use of fluids in the eye, manipulation and movement of the eye, among other factors, result in the interruption of the illumination from the surgical microscope light source. Such factors would be expected to significantly extend the time at which photoretinitis might occur.
In conclusion
Carl Zeiss recommends:
Use of the GG 475 eye protection filter.
Reduction of the illumination of the surgical area to the extent required
for the patient's safety and for microscopic visualization.
Tilting of the microscope body as required.
Insertion of the retinal protection device.
Maximum reduction of the exposure of the patient's eye to light from
surrounding light sources.
These measures should help the surgeon to reduce the likelihood of pho­totoxic retinal injury.
Note: The VISU 150 microscope always contains a UV blocking filter.
The use of this filter ensures that the illumination intensity lies below 50 µW/cm
2
in the range between 305 nm and 400 nm.
These measures help the surgeon to reduce the risk of phototoxic retinal injury.
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printedon21.11.2002
Page 18
18 Safety

Warning labels and notes

Caution:
Observe all warning labels and notes! If any label is missing on your instrument or has become illegible, please contact us or one of our authorized representatives. We will supply the missing labels.
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printed on 21. 11. 2002
Page 19

Description

Description 19
VISU 200 BrightFlex™ Surgical Microscope 20
Intended use 20
Description of the modules 20
Illumination system 28
Controls, displays, connections 30
Hand grips (option) 38
Tube and eyepieces for main microscope 40
Tube and eyepieces for assistant's microscope 42
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printedon21.11.2002
Page 20
20 Description

VISU 200 BrightFlex™ Surgical Microscope

Intended use

The VISU 200 BrightFlexsurgical microscope has been designed for surgical procedures in the field of ophthalmology, i.e. the microscope meets the special requirements of this discipline.

Description of the modules

The VISU surgical microscope is comprised of the following modules:
1 X-Y coupling
The X-Y coupling allows fine, motorized positioning of the surgical mi­croscope in a horizontal plane. The range of travel is 40 mm x 40 mm. The speed of travel can be set on the display field of the suspension system.
The X-Y coupling is provided with a recentering mechanism. When you press the actuator button,
the X-Y coupling adopts its center position,
the focusing system of the surgical microscope is reset to its initial
position and
the zoom system will set a preselected magnification factor if the
XYZ-RESfunctionhasbeenactivated(possiblewithS8suspen­sion systems only).
You can also trigger the recentering movement using the foot control panel..
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printed on 21. 11. 2002
Page 21
Description 21
1
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printedon21.11.2002
Page 22
22 Description
2 Support arm for the surgical microscope
The support arm incorporates a tilt device. This allows the viewing di­rection of the surgical microscope to be adapted to the requirements of the surgical field. Using the knob for fine tilt, you can position the surgical microscope in a range from +180° to -180° (+ in the direction of the surgeon and - in the opposite direction). The +90° setting is ideal for surgery on patients in a seated position or lying on their side.
Caution:
Do not tilt the main microscope beyond + / - 180°,asthiscould damage the microscope cable or the light guide.
3 Main microscope
The apochromatic optics of the main microscope provide superb op­tical quality. The microscope image displays optimum contrast and ex­cellent detail recognition along with a large depth of field. The bright microscope image is a particular benefit in vitreoretinal surgery. A 1:6 ratio zoom system allows the magnification of the overall system to be set as required by the surgical procedure.
Two apochromatic objective lenses with focal lengths of 175 mm and 200 mm are available for different working distances.
A180° tiltable tube is used as a viewing device for the surgeon. The large tilt range allows work with minimum fatigue.
The standard equipment includes eyepieces with a magnification factor of 12.5x (option: 10x).
The illumination system has been designed for use in ophthalmology. A light guide directs the light from the light source in the suspension system to the surgical microscope.
6° illumination can be continuously faded in. This allows the illumina­tion intensity in the surgical field to be adjusted continuously, without changing the color of the light. The illumination of the surgical field at an angle of 6° produces an image with an outstanding impression of depth.
An additional illumination system produces an intensive red reflex even when the eye is decentered. The angle of illumination can be switched from +2° to -2°.
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printed on 21. 11. 2002
Page 23
Description 23
2
3
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printedon21.11.2002
Page 24
24 Description
To protect the patient's eye against photo-retinitis, a retinal protection de­vice is provided. This device can be swung into the illumination beam path, if no red reflex is required.
At the light source integrated in the stand, a GG 475 eye protection filter can be swung into the illumination beam path. This filter markedly reduces the exposure of the patient's and surgeon's eyes to radiation.
A video camera from our MediLive video camera line can be optionally connected to the VISU 200 surgical microscope. The light loss for the sur­geon is only 20% and cannot be perceived subjectively.
Caution!
Avoid looking directly into the light source, e.g. into the microscope ob-
jective lens or into the light guide!
When selecting the brightness level for the patient's eye, always take
caretokeepthestrainonthepatient'seyetoaminimum.
If the red reflex is not necessary, move the retinal protection device
into the beam path. Only use the retro-illumination contrast stop, if the surgical procedure requires a red reflex.
When operating on the eye, always use a GG 475 eye protection filter
to ensure that the patient's retina is not exposed to unnecessary (blue) radiation (retinal injury)!
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printed on 21. 11. 2002
Page 25
Description 25
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printedon21.11.2002
Page 26
26 Description
4 0° assistant's microscope
The assistant's microscope is an integral part of the VISU 200 Bright­Flexsurgical microscope, i.e. it cannot be separated from the main microscope. The assistant sees the same image as the main surgeon. The system's excellent image quality is also available to the assistant.
The assistant's microscope has two working positions. They are lo­cated on the right and left of the main surgeon at an angle of 90° to the main surgeon's viewing direction. No locking mechanism has been provided, allowing the assistant to move the microscope by a certain amount out of the 90° position, if necessary.
Warning!
To prevent the assistant's microscope from moving downward of its own accord when the main microscope is being tilted, the assistant's microscope must be adjusted and locked in position using screw (5) before surgery.
The assistant's microscope is equipped with a focusing system and a 5-step magnification changer. This enables the assistant to adjust his microscope image independently of the main surgeon.
The binocular tube can be turned by ± 12° about the optical axis of the assistant's microscope. In addition, the assistant's microscope can be tilted by 15°. If the assistant finds the viewing angle too steep, an op­tical wedge (option) can be installed between the microscope body and the binocular tube to permit horizontal viewing.
The standard equipment includes eyepieces with a magnification factor of 10x, providing a low initial magnification. This provides the benefit of a wide field of view and an improved overview of the surgical field. The assistant sees the red reflex in both eyepieces.
5 Locking screw for assistant's microscope
After adjusting the assistant's microscope as required, secure it in po­sition using this screw.
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printed on 21. 11. 2002
Page 27
Description 27
4
5
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printedon21.11.2002
Page 28
28 Description

Illumination system

6° illumination
You can continuously fade out the 6° illumination. This results in markedly reduced reflection from the cornea, while providing nevertheless a high­contrast, informative microscope image.
For cataract surgery, we recommend opening the 6° illumination system approx. 25%.
2° illumination
You can switch the illumination from +2° to -2°. This ensures good visi­bility of the red reflex even when the patient's eye is decentered.
1 6° illumination and 2° illumination 2 2° illumination with centered patient's eye
Here, the red reflex is clearly visible in both positions (+2° and -2°).
3 2° illumination with decentered patient's eye
The red reflex is no longer visible at +2°. After switching to -2°,thered reflex is again clearly visible.
+6°
-2° +2°
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printed on 21. 11. 2002
Page 29
Description 29
1
2
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
3
Printedon21.11.2002
Page 30
30 Description

Controls, displays, connections

1 Instrument connector
2 Microscope shaft
3 X-Y coupling
4 Actuator button
recenters the X-Y coupling.
resets the focus to its initial position in the focusing range, and
the zoom system will set a preselected magnification factor if the
XYZ-RESfunctionhasbeenactivated(possiblewithS8suspen­sion systems only).
Note: Pressthisbuttontostarttherecenteringmovement.Tostopthis movement, press the button again.
You can also stop the recentering movement by briefly tipping on one of the direction keys on the foot control panel.
5 Cable and light guide clip
6 Support arm with tilt device
7 Knob
for setting the tilt angle of the surgical microscope; +180° in the direction of the surgeon,
-180° in the opposite direction.
8 Arrows indicating the focusing range
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printed on 21. 11. 2002
Page 31
Description 31
1
2
3
4
5
6
7
8
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printedon21.11.2002
Page 32
32 Description
9 Dust cover
10 Release buttons for the magnetic brakes of the stand
Only in connection with suspension systems with magnetric brakes. A second button with the same function is located on the opposite side in the knob for the additional red reflex illumination.
Button pressed magnetic brakes are released, the unit
Button released magnetic brakes are locked, the unit
11 Maneuvering grip including release button. See (10).
12 Display of the magnification factor of the zoom system
13 Securing screw
can be moved as required.
cannot be moved.
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printed on 21. 11. 2002
Page 33
Description 33
9
10
11 12 13
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printedon21.11.2002
Page 34
34 Description
14 Manual adjustment of zoom system
15 Knob for adjusting the red reflex illumination
The angle of illumination can be switched between + 2° and - 2°.
16 Release button for magnetic brakes
(A second button with the same function is located in maneuvering grip (10).
Button pressed magnetic brakes are released, the unit
Button released magnetic brakes are locked, the unit
17 Securing screw
for locking the assistant's tube within its 12° range of movement.
18 Binocular tube of the assistant's microscope
can be moved as required.
cannot be moved.
19 Focusing knob
for adjusting the assistant's microscope independently of the main surgeon.
20 Five-step manual magnification changer
21 Screw
for locking the assistant's microscope within its 15° tilt range.
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printed on 21. 11. 2002
Page 35
Description 35
14
15
16
17
18
19
20 21
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printedon21.11.2002
Page 36
36 Description
22 Light stop knob
23 Adjusting lever for the 6° illumination
for gradual fading in/out the coaxial illumination. Fading out the 6° illu­mination improves the visualization of structures in retro-illumination.
24 Connector socket for S light guide
Retro-illumination contrast stop. This stop reduces the straylight reflected from the sclera. Diameter approx. 16 mm (with objective lens f = 200 mm): Free passage. Diameter over approx. 16 mm: Partially transmitting periphery.
Free passage. The field of view is fully illuminated.
Retinal protection device. This device does not snap in, i.e. it can be moved continuously in the image.
25 Locking screw for assistant's microscope
for locking the assistant's microscope in both working positions on the right and left of the main microscope.
Warning!
To prevent the assistant's microscope from moving downward of its own accord when the main microscope is being tilted, the assistant's microscope must be adjusted and locked in position using screw (25) before surgery.
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printed on 21. 11. 2002
Page 37
Description 37
22 23 2524
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printedon21.11.2002
Page 38
38 Description

Hand grips (option)

In addition to the maneuvering grips provided, larger hand grips can be mounted. The mounting procedure is described in the instructions en­closed. This does not affect the functioning of the surgical microscope. Only the operation of the surgical microscope is slightly changed.
1 Maneuvering grip with integrated release button
By turning one of the two maneuvering grips you will activate the button for magnetic brakes of the suspension system. (Only in combination with suspension systems with magnetic brakes (e.g. S8 Floor Stand)).
Button pressed Magnetic brakes are released, the unit
can be moved as required.
Button released Magnetic brakes are locked, the unit
cannot be moved.
2 Knob for adjusting the red reflex illumination
By sliding the knob, you can switch the angle of illumination between +2° and - 2°.
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printed on 21. 11. 2002
Page 39
Description 39
1
2
1
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printedon21.11.2002
Page 40
40 Description

Tube and eyepieces for main microscope

1 180° tiltable tube
2 PD adjustment knob
The correct interpupillary distance has been set when both eyepiece images merge into one.
3 Push-in widefield eyepiece
Note: When you remove the eyepiece from the tube, please note that this eyepiece is fitted with a magnetic coupling. When mounted, the eye­pieces display a very weak magnetic field. For this reason, the usual rules for the handling of magnets must only be observed with eye­pieces which have not been mounted on the microscope:
Do not place the eyepieces close to instruments where there is any risk of magnetization.
Do not place eyepieces on sensitive electronic instruments such as in­fusion pumps, cardiac pace-makers, measuring instruments or mag­netic data carriers such as disks, audiotapes and videotapes or credit cards.
Always store the eyepieces in their original packaging when not in use.
4 Eyecup
Always adjust the eyecups in such a way that you can see the full field of view.
Viewing with
eyeglasses:
Viewing without
eyeglasses:
5 Diopter setting ring
The eyepieces provide ametropia compensation from -8 D to +5 D. If you wear your eyeglasses during work, adjust the ring to 0 D. Turn the ring until you have found the optimum setting. A built-in catch lever locks the ring in the position set.
6 Diopter scale
For reading the prescription set.
Screw in eyecups all the way.
Screw eyecups outward until you can see the full field of view.
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printed on 21. 11. 2002
Page 41
Description 41
1 2
3
4
5
6
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printedon21.11.2002
Page 42
42 Description

Tube and eyepieces for assistant's microscope

7 Push-in widefield eyepiece
Note: When you remove the eyepiece from the tube, please note that this eyepiece is fitted with a magnetic coupling. When mounted, the eye­pieces display a very weak magnetic field. For this reason, the usual rules for the handling of magnets must only be observed with eye­pieces which have not been mounted on the microscope:
Do not place the eyepieces close to instruments where there is any risk of magnetization.
Do not place eyepieces on sensitive electronic instruments such as in­fusion pumps, cardiac pace-makers, measuring instruments or mag­netic data carriers such as disks, audiotapes and videotapes or credit cards.
Always store the eyepieces in their original packaging when not in use.
8 Eyecup
Always adjust the eyecups in such a way that you can see the full field of view.
Viewing with eyeglass-
es:
Viewing w/out eyeglass-
es:
9 Diopter scale
For reading the prescription set.
10 Diopter setting ring
The eyepieces provide ametropia compensation from -8 D to +5 D.If you wear your eyeglasses during work, adjust the ring to 0 D. Turn the ring until you have found the optimum setting. A built-in catch lever locks the ring in the position set.
11 Binocular assistant's microscope
Note: Always use the assistant's microscope with the appropriate (shown) binocular assistant's tube. If you use a different tube, you will obtain an inverted image.
Screw in eyecups all the way.
Screw eyecups outward until you see the full field of view.
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printed on 21. 11. 2002
Page 43
Description 43
7
8
9
10
11
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printedon21.11.2002
Page 44
44 Description
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printed on 21. 11. 2002
Page 45
Preparations for use 45

Preparations for use

Mounting the binocular tubes, eyepieces and the objective lens 46
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printedon21.11.2002
Page 46
46 Preparations for use

Mounting the binocular tubes, eyepieces and the objective lens

Note: Always use the assistant's microscope with the appropriate binocular as­sistant's tube (shown). If you use a different tube, you will obtain an in­verted image.
Loosen securing screw (5) by a few turns.
Remove cover (1) and store it in a safe place.
Place binocular tube (2) on the surgical microscope and tighten secu-
ring screw (5) firmly.
You can install other units between the binocular tube and the micro­scope body. Lock these units in position in the same way using secur­ing screw (5).
Place binocular tube (10) on the assistant's microscope and tighten securing screw (7)firmly.
You can install other units between the binocular tube and the micro­scope body (e.g. an optical wedge, 30°). Lock these units in position in the same way using securing screw (7).
Insert 12.5x widefield eyepieces (4) into the mounts intended for them
(3) as far as they will go. The magnetic coupling reliably secures them
in position.
Insert 10x widefield eyepieces (9) into the mounts intended for them
(8) as far as they will go. The magnetic coupling reliably secures them
in position.
Note: If a documentation device is used, one of the eyepieces can be delivered with a reticle to aid focusing. Retrofitting the reticle can only be performed at the factory or by our service representative. Always use the eyepiece with reticle on the same side of the binocular tube as the documentation device is installed.
Screw objective lens (6) into the microscope body and tighten itfirmly.
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printed on 21. 11. 2002
Page 47
Preparations for use 47
Caution:
Before using and after re-equipping the unit, make sure that the two bin­ocular tubes (2 and 10) are securely locked in position. Make sure that
securing screws (5 and 7) and
objective lens (6)
have been tightened firmly
1
2
3
.
10
4
5
6
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printedon21.11.2002
9
8
7
Page 48
48 Preparations for use
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printed on 21. 11. 2002
Page 49

Operation

Operation 49
Preparations for use 50
Adjusting the tilt angle 50
Setting the microscope tilt to angles greater than 15° 52
Adjusting the surgical microscope 53
Checklist 54
When using a fundus imaging system (e.g. BIOM II) 56
Procedure 57
What to do in an emergency 59
Failure of zoom system 59
Failure of X-Y coupling 60
Failure of focusing system 60
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printedon21.11.2002
Page 50
50 Operation

Preparations for use

Adjusting the tilt angle

Using knob (1), you can position the surgical microscope in a range from +180° to -180° (+ in the direction of the surgeon and - in the opposite di­rection). The +90° setting is ideal for surgery on patients in a seated po­sition or lying on their side.
Warning!
To prevent the assistant's microscope from moving downward of its
own accord when the main microscope is being tilted, the assistant's microscope must be adjusted and locked in position using screw (2) before surgery.
Check that the assistant's microscope is firmly seated.
Do not tilt the main microscope beyond + / - 180°, as this could dam-
age the microscope cable or the light guide.
Using locking screw (2), secure the assistant's microscope in the working position set (on the left or right of the main microscope).
Turn knob (1) until the surgical microscope is in the viewing position required.
After the viewing angle has been set, the surgical microscope remains in this position. The gear drive is self-locking.
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printed on 21. 11. 2002
Page 51
Operation 51
1
2
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printedon21.11.2002
Page 52
52 Operation
Setting the microscope tilt to angles greater than 15°
Using locking screw (2), secure the assistant's microscope in the working position set (on the left or right of the main microscope).
Check that the assistant's microscope is firmly seated.
Remove the binocular tube and all accessories mounted on the assist-
ant's microscope.
Remove all accessories mounted on the objective lens (e.g. 0° or 8° assistant's microscope).
Turn knob (1) until the surgical microscope is in the viewing position required.
After the viewing angle has been set, the surgical microscope remains in this position. The gear drive is self-locking.
1
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
2
Printed on 21. 11. 2002
Page 53
Operation 53

Adjusting the surgical microscope

Bring the surgical microscope into starting position within the focusing range. Set the minimum magnification on the surgical micro­scope. Bring the surgical microscope into working position. Adjust the interpupillary distance on the binocular tube. Set the diopter scales on the eyepieces. Please note that instrument myopia may occur.
Eyepieces without
Emmetropes set the diopter scale to 0 diopter (D).
Eyeglass wearers (who perform surgery wearing their glasses)set the diopter scale to 0 di­opter (D).
Ametropes (who know their refractive powers and perform surgery without wear­ing their glasses) set the diopter scaleringtotheirrefractivepowers.
reticles:
Ametropes (who do not know their re­fractive powers and perform surgery without wearing their glasses):
Set both eyepieces on the main micro­scope to +5 D. Set both eyepieces on the assistant’s microscope to +5 D (if applicable).
Remove the binocular tube and eyepieces from the microscope body and point them at a distant ob-
*)
ject
, i.e. use them like a pair of binoculars.The image of the object is still unsharp.
Turnthediopterscaleringofone eyepiece slowly image of the object is sharp. Stop turning the ring. If necessary, repeat this procedure three times and take the average of the readings found. Adjust the second eyepiece in the same manner.
clockwise until the
Eyepieces with
Set both eyepieces on the main micro­scopeto+5D. Set both eyepieces on the assistant’s microscope to +5 D (if applicable).
Turn the diopter scale ring on the eye­piece slowly clockwise until the reticle is imaged sharply. Stop turning the ring.
Focus the surgical microscope on the object. The reticle and the object must be imaged sharply at the same time.
reticles:
Replace the tube and eyepieces on the microscope body and firmly tighten the securing screw.
Adjust the eyecups in such a way that the full field of view can be seen. Set the maximum magnification on the microscope and focus on the object. Set the working magnification required. When the magnification is changed, the focal plane is retained, but the depth of field changes.
Note
: If several surgeons use the instrument, it is advisable to draw up a table showing the individual refractive powers of each surgeon and to keep it in a handy location near the instrument.
*) CAUTION: Never
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
point the eyepieces at the sun!
Adjust the second eyepiece in the same way until the object is seen sharply.
Printedon21.11.2002
Page 54
54 Operation

Checklist

Always check the following points before surgery (without the patient!):
VISU 200 BrightFlex™ surgical microscope
Check that the correct rated voltage has been set for the suspension
system.
Check that all cables have been connected.
Check that the light guide has been connected.
Turn on the instrument at the power switch of the suspension system.
Note: After the system has been switched on, the S8 suspension system(oph­thalmology) automatically performs a self-test. The self-test is completed after approx. five seconds.
Zoom – Check this function using the appropriate key on the foot control panel.
Focus – Check this function using the appropriate key on the foot control panel.
Friction adjustment of the surgical microscope – Check that the friction of the surgical microscope's rotation has been
adjusted as required using the friction adjustment knob on the suspen­sion system.
Locking the assistant's microscope in position – Check that the assistant's microscope is locked in the working position
required (on the right or left of the main microscope).
Speeds of the surgical microscope's functions (only in connection with S8 suspension system (ophthalmology))
Check that the speeds of the microscope's functions have been set as
required on the S8 suspension system (ophthalmology).
Eyepieces / Tiltable binocular tubes – Check that the tiltable tube of the main microscope and the assistant's
microscope are in positions convenient for you.
Check that the correct interpupillary distance has been set.
Check that the eyecups have been adjusted in such a way that you
canseethefullfieldofview.
Check that the correct prescription has been set on the diopter scale.
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printed on 21. 11. 2002
Page 55
Operation 55
Check that image quality is the same throughout the entire magnifica-
tion range.
Accessories – Using the relevant user's manuals, check that the other equipment (il-
lumination system, video system, etc.) is working properly.
Warning!
If a function fails, you must not use this instrument for safety reasons. Cor­rect the fault (see the "Troubleshooting table) or contact our service dept.
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printedon21.11.2002
Page 56
56 Operation

When using a fundus imaging system (e.g. BIOM II)

When using a fundus imaging system (e.g. BIOM II from the company Oc­ulus) which is usually installed between the surgical microscope and the patient, make sure that the patient is neither put at risk nor injured by the motorized focusing system or the movement of the stand arm.
Only use accessories expressly certified by the manufacturer for combi­nation with the surgical microscope described in this manual.
Caution!
Risk of collision!
With the fundus imaging system swung out of position, always position the microscope body in such a way that index dot (1) of the micro­scope's focus is in the middle of triangle (2) of the marking.
Select a medium magnification (e.g. 1.0).
Lower the surgical microscope towards the surgical field until you see
1
2
the patient's cornea sharply defined.
Turn the screw for limiting the downward movement clockwise as far as it will go.
It is vital that you read the users manual on the fundus imaging sys- tem (e.g. BIOM II from the company Oculus).
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printed on 21. 11. 2002
Page 57

Procedure

Operation 57
Turn on the power switch of the suspension system.
Adjust the brightness of the surgical field illumination as required on
the stand.
Check the unit using the checklist.
Move the surgical microscope over the surgical field and bring it into a
convenient position within the working distance.
Press the activator button.
The X-Y coupling is recentered.
The focus is reset to its initial position in the focusing range.
Select the lowest magnification (zoom function on the foot control pan-
el).
For coarse focusing, look through the eyepieces and lower the surgi­cal microscope using the suspension arm until the surgical field comes into focus.
Select the highest magnification (zoom function on the foot control panel).
Look through the eyepieces and activate the focusing function on the foot control panel until the microscope is sharply focused on the sur­gical field.
Select the magnification required (zoom). Look through the eyepieces of the binocular tube. Adjust the eyepieces (including those of the as­sistant's microscope) in such a way that you can see both the edge of the field of view and the microscope image sharply. Also see Adjust­ing the surgical microscope“.
Switch the unit off when it is not in use.
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printedon21.11.2002
Page 58
58 Operation
Caution!
Avoid looking directly into the light source, e.g. into the microscope ob-
jective lens or into a fiber light guide!
When selecting the brightness level for the patient's eye, always take
caretokeepthestrainonthepatient'seyetoaminimum.
If the red reflex is not necessary, move the retinal protection device
into the beam path. Only use the retro-illumination contrast stop, if the surgical procedure requires a red reflex.
When operating on the eye, always use a GG 475 protection filter to
ensure that the patient's eye is not exposed to unnecessary (blue) ra­diation (retinal injury).
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printed on 21. 11. 2002
Page 59
Operation 59

What to do in an emergency

Failure of zoom system

Manually adjust the magnification using zoom control (1) on the micro­scope (if necessary, use a tool, e.g. screwdriver, coin, etc.).
1
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printedon21.11.2002
Page 60
60 Operation

Failure of X-Y coupling

If the X-Y coupling fails, you can manually position the surgical micro­scope utilizing the possibilities of the suspension system you are us­ing.

Failure of focusing system

Disconnect the cable between the surgical microscope and the sus­pension system (illumination continues to be on.)
Focus the surgical microscope by moving the suspension arm or the lift arm of the suspension system you are using.
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printed on 21. 11. 2002
Page 61
Maintenance / Further information 61

Maintenance / Further information

Trouble-shooting table 62
Magnifications / Fields of view 64
Care of the unit 65
Sterilization 66
Ordering data 67
Spare parts 68
Accessories 68
Adapter cables 70
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printedon21.11.2002
Page 62
62 Maintenance / Further information

Trouble-shooting table

This instrument is a high-grade technological product. To ensure opti­mum performance and safe working order of the instrument, its safety must be checked once every 12 months. We recommend having this check performed by our service representative as part of regular main­tenance work. If a failure occurs which you cannot correct using the trouble-shooting table, attach a sign to the instrument stating it is out of order and con­tact our service representative.
VISU 200 BrightFlex™ Surgical Microscope
Problem Possible cause See
No function at all. Power cord of suspension
system not plugged in.
Power switch of suspension system not on.
Circuit breaker in suspen­sion system power switch activated.
Line power failure. Contact in-house electri-
Lamp module has no con­tact.
Surgical field illumination at microscope inoperative.
Illumination system of sus­pension system not switched on.
S light guide not properly in­serted in lamp or micro­scope.
Failure of suspension system electronics.
Plug in power cord. -
Press power switch. Green light in power switch must come on.
Press power switch again. -
cian.
Insert lamp module as far as it will go.
Switch on illumination system of suspension system.
Insert S light guide as far as it will go.
Illuminate surgical field using an OR illuminator. Contact service dept.
-
-
-
-
-
-
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printed on 21. 11. 2002
Page 63
Maintenance / Further information 63
Problem Possible cause See
Insufficient surgical field illu-
Brightness level set too low. Adjust brightness on sus-
mination.
Defective S light guide (no homogeneous illumination).
Focusing function inopera­tive.
Connectors on suspension system not correctly in­serted.
Focusing system always moves into upper or lower end position.
Zoom function inoperative. Connectors on suspension
system not correctly in­serted.
Zoom system always moves into upper or lower end posi­tion.
pension system or using foot control panel.
Contact service dept. The light guide must prob­ably be changed.
Plug in connectors correctly. -
Check suspension system. Move the suspension arm to adjust the focus. Contact service dept.
Plug in connectors correctly. -
Check suspension system. Manually adjust magnifica­tion using the zoom knob. Contact service dept.
-
-
-
page 34
X-Y coupling inoperative. Connectors on suspension
system not correctly in­serted.
Defective motor. Position microscope manu-
X-Y coupling always moves in opposite direction.
Function not correctly ad­justed on suspension system.
Zoom and focusing func­tions inoperative.
Functions not correctly ad­justed on suspension system.
Microscope motion too stiff. Friction adjustment knob on
suspension system tight­ened too firmly.
Plug in connectors correctly. –
­ally. Contact service dept.
Check suspension system. -
Check suspension system. -
Loosen friction adjustment. -
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printedon21.11.2002
Page 64
64 Maintenance / Further information

Magnifications / Fields of view

Using the magnification factor γ of the zoom system, you can calculate the total magnification of the surgical microscope according to the following formula:
f
tube
M
-----------
T
where:
f
is the focal length of the binocular tube
tube
is the focal length of the main objective lens
f
obj
γ is the magnification factor set on the zoom system
M
is the magnification of the eyepiece
eye
f
obj
γ M
⋅⋅=
eye
Example:
=170mm,f
f
tube
= 200 mm, γ =1.6andM
obj
eye
= 12.5 x.
The resulting total magnification is:
170 mm
T
---------- -----------
200 mm
canbecalculatedusingtheformula:
D
FoVNM
--------- ------------- -----------=
D
1.6 12.5 17.0=⋅⋅=
of the surgical microscope is known, the field-
T
eye
M
T
is the diameter of the circular area of the
D
M
If the total magnification M of-view diameter FoV
FoV
The field-of-view diameter FoV surgical field which can be seen through the eyepieces. FoV
in the above formula stands for the field-of-view number of the eye-
N
piece. This number is marked on our widefield eyepieces.
Using M
= 17.0 from the example above, the field-of-view diameter ob-
T
tained with a 12.5x eyepiece with a field-of-view number FoV is calculated as follows:
of 18 mm
N
18 mm 12.5
FoV
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
--------- ------------- ------------
D
17.0
13.2 mm==
Printed on 21. 11. 2002
Page 65

Care of the unit

Maintenance / Further information 65
Cleaning optical surfaces
The multilayer T* (T-star) coating of our optical components (e.g. eye­pieces, objective lenses) results in optimum image quality. Image quality is impaired by even slight contamination of the optics or by a fingerprint. In order to protect the internal optics from dust, the instru­ment should never be left without the objective lens, binocular tube and eyepieces installed when it is not in use. Always store objective lenses, eyepieces and accessories which are not being used in dust-free cases.
The external surfaces of optical components (eyepieces, objective lenses) should only be cleaned when required:
Do not use any chemical cleaning agents.
Blow off dust on the optical surfaces using a squeeze blower or re-
move the dust using a clean and grease-free brush.
Fogging of optical surfaces
To protect the eyepiece optics from fogging, we recommend using an anti-fogging agent.
Note: Anti-fogging agents provided by eyecare professionals for use with eye­glass lenses are also suitable for Zeiss eyepieces.
Please observe the instructions for use supplied with each anti-fog­ging agent.
Anti-fogging agents do not only ensure fog-free optics. They also clean the eyepiece optics and protect them against dirt, grease, dust, fluff and fingerprints.
Cleaning mechanical surfaces
All mechanical surfaces of the equipment can be cleaned by wiping with a moist cloth. Do not use any aggressive or abrasive cleaning agents.
Wipe off any residue with a mixture of 50% ethyl alcohol and 50% distilled water plus a dash of household dish-washing liquid.
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printedon21.11.2002
Page 66
66 Maintenance / Further information

Sterilization

The asepsis sets available from Carl Zeiss contain rubber caps, sleeves and handgrips which can be sterilized in autoclaves. We recommend the following program for sterilization:
Sterilization temperature: 134° C
Sterilization time: 10 minutes
Sterile single-use drapes are available to cover the instrument.
Note
: When draping the system, make sure there is enough slack in the drapes to allow for movement of the microscope carrier and surgical microscope. It is especially important that the drapes are completely loose around the handgrips. The surgeon must be able to operate the keys through the drape.
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printed on 21. 11. 2002
Page 67

Ordering data

Maintenance / Further information 67
VISU 200 BrightFlex™ Surgical Microscope
Description Cat.No.
VISU 200 BrightFlexsurgical microscope including
1179-101
X-Y coupling and assistant's microscope
180° tiltable tube 30 37 91-0000
10x eyepiece for assistant's microscope (2x) 30 55 42-0000
12.5x/18 B eyepiece for main microscope (2x) 30 55 43-0000
Objective lens, f = 175 mm 30 26 51-9902
Dust cover 1055-278
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printedon21.11.2002
Page 68
68 Maintenance / Further information

Spare parts

VISU 200 BrightFlex™ Surgical Microscope
Description Cat. No.
S light guide (not shown) 2.0 m 30 34 81-9020
Dust cover (not shown) 1055-278
Asepsis set (not shown) 1009.100

Accessories

Please observe the following: Only operate the instrument with the accessories included in the delivery package. If you want to use other accessories, make sure that Carl Zeiss or the manufacturer of the accessories has proved and confirmed that these accessories meet the respective technical safety standards and can be used without risk.
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printed on 21. 11. 2002
Page 69
Maintenance / Further information 69
VISU 200 BrightFlex™ Surgical Microscope
Description Cat. No.
1 Objective lens, f = 200 mm 30 26 52-9903
2 Optical wedge, 30° 1006-145
3 Sleeve (for insertion of eyepiece (4)) 30 55 42-0107
4 10x eyepiece with magnetic coupling 30 55 42-0000
5 OPMI VISU hand grips 1049- 171
Asepsis for OPMI VISU hand grips (1x included in delivery package 1049- 171)
1054- 074
12345
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printedon21.11.2002
Page 70
70 Maintenance / Further information

Adapter cables

for the electrical connection of the surgical microscope to different types of suspension system.
1 EMD II Electromechanical Ceiling Mount 30 26 02- 9053
1 S3 Ceiling Mount 30 26 02- 9053
2 S4 Ceiling Mount 30 26 02- 9054
1 S23 Ceiling Mount 30 26 02- 9053
3 S6 Ceiling Mount 30 26 02- 9055
3 S61 Ceiling Mount 30 26 02- 9055
Cat. No.
1 S22 Floor Stand 30 26 02- 9053
2 S4FloorStand 302602-9054
1 S3FloorStand 302602-9053
123
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printed on 21. 11. 2002
Page 71
Technical data 71

Technical data

Technical data 72
Ambient requirements 74
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printedon21.11.2002
Page 72
72 Technical data

Technical data

VISU 200 BrightFlex™ Surgical Microscope
Optical data
Magnification Main microscope:
Focusing Motorized, focusing range 50 mm
4.9x - 29x
Motorized zoom system with apochromatic op­tics, 1:6 zoom ratio,
magnification factor γ = 0.4x - 2.4x.
At the press of a button, the focusing drive returns to its initial position.
Objective lens focal length
Tubes / Eyepieces Main microscope
Illumination Fiber illumination with two lamp modules, each
f = 175 mm (option: f = 200 mm)
:
180° tiltable binocular tube, f=170 mm. 12.5x screw-type, widefield eyepieces (option: 10x ).
Assistant's microscope
5-step magnification changer, separate fine fo­cusing, adjustable tilt.
Binocular tube, can be turned by ±12° about its optical axis,
10x widefield eyepieces with magnetic coupling.
with two 12 V 100 W halogen reflector lamps,
fully automatic lamp change in the event of lamp failure,
filter against UV exposure,
protection against IR exposure,
:
6° illumination (continuous fading),
patented 2° illumination for red reflex, usable in +2° and -2° positions
Pat.No.DE4028605C2
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printed on 21. 11. 2002
Page 73
Technical data 73
Swing-in stops Patented retinal protection device,
Pat.No.DE3339172C2
Patented field stop for reduced glare,
Pat. No. G 91 03 43 3.7
Mechanical data
Front-to-back tilt of microscope
With self-locking gear drive, manually adjustable using a knob.
Tilt angle ± 180°
Swivel range of
Swivel angle ± 90°, manually lockable assistant's micro­scope
X-Y coupling Adjustment range: max. 40 mm x 40 mm
Automatic centering at the press of a button
Weight approx. 13 kg
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printedon21.11.2002
Page 74
74 Technical data

Ambient requirements

For operation Temperature
Rel. humidity Air pressure
For transportation and storage
The unit meets the essential requirements stipulated in Annex I to the 93/42/EEC Directive governing medical de­vices. The unit is marked with:
Subject to change.
Temperature Rel. humidity (without condensation) Air pressure
+10 °C...+40 °C 30%...75% 700 hPa...1,060 hPa
-40°C...+70 °C 10%...100%
500 hPa...1,060 hPa
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printed on 21. 11. 2002
Page 75

Index

Index 75
Symbols
+90° position
Numerics
2° illumination
A
Accessories Accident prevention regulations 10 Actuator button 30 Adapter cables 70 Adjusting lever for 6° illumination 36 Adjusting the microscope tilt to angles greater than 15° 52 Adjusting the surgical microscope 53 Adjusting the tilt angle 50 Ambient requirements 74 Angle of illumination 15 Anti-fogging agent 65 Arrows indicating the focusing range 30 Assistant's microscope 0° 26
50
28
68
B
Binocular tubes, eyepieces and objective lens, mounting of Buttons for the release of the magnetic brakes 32
C
Cable and light guide clip Care of the unit 65 CE marking 74 Checklist 54 Cleaning optical surfaces 65 Composition, spectral 15 Connector socket for S light guide 36 Controls, displays, connections 30
D
Description Diopter scale 40, 42 Directive 93/42/EEC 74 Directives 10
E
Emergency backup line supply Exposure time to light 16 Eyecup 40, 42
19
30
10
46
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printed on 21. 11. 2002
Page 76
76 Index
Eyepiece optics, fog-free 65
F
Failure of a function Failure of zoom system 59 Fields of view 64 Focus of the light source 15 Focusing knob 34 Focusing system failure 60 Fogging of optical surfaces 65 Fundus imaging system 14, 56
G
GG 475 eye protection filter GG 475 protection filter 13
H
Hand grips (option)
55
17
38
I
Illumination characteristics Installation, notes 11 Instrument class 10 Instrument connector 30 Intensity of illumination 15 Intensity scale 16 ISO 10936-2 draft for surgical microscopes used in ophthalmic applicati-
16
ons
K
Key to symbols Knob for adjusting the red reflex illumination 34
L
16
LB Light source, focus 15 Light stop knob 36 Locking screw for assistant's microscope 26, 36
M
Magnification changer Magnifications 64 Main microscope 22 Main microscope, tilted 26 Maintenance 61 Manual adjustment of zoom system 34 Mechanical surfaces, cleaning 65 Microscope shaft 30 Mobile phones 11
7
15
34
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printed on 21. 11. 2002
Page 77
Index 77
N
Notes on safety Notes on use 11
O
Operation Operation, requirements 12 Optical surfaces, fogging 65 Ordering data 67
P
PD adjustment knob Phototoxic retinal injury in eye surgery 14 Potential equalization 11 Preparations for use 45 Procedure 57 Protective earth contact 12
R
Red reflex illumination Re-equipping the instrument 12 Release button for magnetic brakes 34, 38 Requirements for operation 12 Risk of collision! 14, 56
10
49
40
38
S
Safety Safety check 12, 62 Safety standards 10 Spare parts 68 Spectral composition 15 Standards 10 Sterilization 66 Support arm for surgical microscope 22 Support arm with tilt device 30
T
Technical data Time of exposure to light 16 Trouble-shooting table 62 Tube and eyepieces for assistant's microscope 42 Tube and eyepieces for main microscope 40
U
Use of the instrument, before Use of the instrument, for every 13 Use, notes 11
W
What to do in an emergency
9, 10
71, 72
12
59
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printed on 21. 11. 2002
Page 78
78 Index
X
X-Y coupling X-Y coupling failure 60
20, 30
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printed on 21. 11. 2002
Page 79
79
G-30-1439-en OPMI® VISU 200 BrightFlexSurgical Microscope Issue 3.0
Printedon21.11.2002
Page 80
Carl Zeiss
Surgical Products Division 73446 Oberkochen Germany
G-30-1439-en Printed in Germany MS-TD
Fax: +49 (0) 7364 20-4823 E-mail: surgical@zeiss.de www.zeiss.com/surgical
Loading...