Different symbols used in this manual draw your attention to safety aspects and useful tips. These symbols are explained in the following.
Warning!
The warning triangle indicates potential sources of danger which may
constitute a risk of injury for the user or a health hazard.
Caution:
The square indicates situations which may lead to malfunction, defects,
collision or damage of the system.
Note:
The hand indicates hints on the use of the system or other tips for the
user.
®
OPMI
GmbH
and Retrolux® are registered trademarks of Carl Zeiss Surgical
.
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
Contents
3
–Key to symbols 2
Functions at a glance 7
–OPMI® VARIO 8
–S88 floor stand 10
Safety 13
–Notes on installation and use 15
–Phototoxic retinal injury in eye surgery 21
–Safety devices of the suspension systems 26
–Warning labels and notes 32
Description 37
OPMI Vario Retrolux System 38
–Intended use 38
–Components of the OPMI Vario Retrolux System 38
Components of the surgical microscope 40
–Controls, displays, connections 42
–Binocular tubes and eyepieces 50
–Retrolux illumination module 54
Components of the S88 floor stand 56
–Description of the modules 57
–Design 58
–Display field with control keys 60
–Lamp housing with xenon light source 62
–Suspension arm 68
–Stand base with column 70
–Connector panel 72
–Instrument tray (option) 74
–Video monitor (option) 76
Foot control panel (option) 84
G-30-1721-enOPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
4
–Intended use 84
–Design 84
–Foot control panel with 14 functions 86
–Foot control panel with 8 functions 87
Preparations 89
Attaching the equipment 90
–Mounting the surgical microscope 90
–Attaching accessories 92
–Mounting the tube and eyepieces 94
–Mounting the Retrolux illumination module 96
–Bringing the Retrolux illumination module into its standby position
100
–Attaching the fundus imaging system to the illumination module 102
Connections 104
–Connecting the surgical microscope 104
–Mounting the light guide 104
–Aligning the X-Y coupling 108
–Strain relief device on S88 floor stand 110
–Connecting the S88 floor stand 112
Adjusting the supension system 114
–Adjusting the balance setting of the suspension arm 114
–Adjusting the limit of downward movement 116
Balancing the surgical microscope 118
Settings on the control and display panel 120
–Adjusting the suspension system 120
Adjusting the surgical microscope 121
Surgical microscope with a laser micromanipulator 123
Relocating the system 126
Operation 129
Checklist 130
Positioning the S88 floor stand 134
Using the display and key field 136
–General functions 136
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
–OPMI Vario on the suspension system, user interface
with SpeedFokus option 140
Procedure 160
What to do in an emergency 162
–Failure of a xenon lamp 162
–Failure of lamp control 164
–Failure of the zoom function 165
–Failure of the focusing function 166
–Failure of magnetic brakes 167
Maintenance / Further information 169
–Trouble-shooting 170
–Changing the xenon lamp module 178
5
–Adjusting the monitor arm 180
–Magnifications / Fields of view 182
–Care of the unit 184
–Sterilization 185
–Disinfecting the control keys 186
–Ordering data 188
–Spare parts 190
–Accessories 191
–Disposal 193
Technical data 195
–Technical data of OPMI VARIO 197
–Technical data of S88 floor stand 198
–Ambient requirements 205
–CE conformity 205
–Changes to the system 205
Index 207
G-30-1721-enOPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
6
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
Functions at a glance7
Functions at a glance
OPMI® VARIO 8
S88 floor stand 10
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
8Functions at a glance
OPMI® VARIO
1Adjusting the eyecup and the prescriptionpage 52
2Adjusting the interpupillary distancepage 50
3Adjusting the friction of vertical adjustmentpage 42
4Recentering the X-Y couplingpage 48
5Adjusting the friction of the OPMI® axis of rotationpage 48
6Manual zoom settingpage 44
7Manual focusingpage 44
8Manual setting of the illuminated field diameterpage 44
9Connecting the light guidepage 104
10 Balance setting of the microscope's
front-to-back tilt
11 Balance setting of the microscope's
lateral tilt
12 Motorized focusingpage 46
13 Motorized zoom settingpage 46
14 Display of the magnification factor of
the zoom system
15 Releasing the magnetic brakespage 46
16 Programmable release buttonspage 46
17 Switching off the electrical drive of the
focusing system
18 Connecting the mouth switchpage 42
page 42
page 42
page 42
page 44
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
Functions at a glance9
5
186
1234
7811 12131415 16 17910
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
10Functions at a glance
42153
S88 floor stand
Xenon illumination system
1Selecting a filterpage 64
2Manual activation of backup lamppage 62
3Resetting the counterpage 64
4Opening the lamp modulepage 64
5Red segment is lit - backup lamp is in usepage 62
6Control panelpage 66
7Lamp housing of xenon illumination systempage 62
8Unlocking the magnetic brakes of the suspension
system
9Limiting the suspension arm's downward movementpage 116
10 Removing/mounting the coupling for the
surgical microscope
11 Balance settingpage 68
12 Locking the stand in positionpage 70
13 Locking the suspension arm in its horizontal positionpage 68
14 Connecting the foot control panel or hand switch,
connecting the remote control connector
15 Rated voltage displaypage 72
16 Connector panelpage 72
17 Turning on the suspension systempage 72
page 68
page 68
page 72
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
Functions at a glance11
6798
1011
12913 14 15, 16, 17
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
12Functions at a glance
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
Safety
Safety13
Notes on installation and use 15
Phototoxic retinal injury in eye surgery 21
Safety devices of the suspension systems 26
Warning labels and notes 32
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
14Safety
The device described in this manual has been designed and tested in accordance with Carl Zeiss safety standards as well as German and international standards. This guarantees a high degree of instrument safety.
The system described in this user manual has been designed in compliance with the requirements of:
–EN–IEC–UL–CSA
In accordance with Directive 93/42/EEC for medical devices, the complete quality management system of the company Carl Zeiss Surgical
GmbH, 73446 Oberkochen, Germany, has been certified by DQS Deutsche Gesellschaft zur Zertifizierung von Managementsystemen GmbH, a
notified body, under registration number 250758 MP23.
–As per Directive 93/42/EEC, the unit is a Class I instrument.
–For USA: FDA classification Class I.
We would like to provide you with information about safety aspects which
must be observed when handling this device. This chapter contains a
summary of the most important information concerning matters relevant
to instrument safety.
Important safety information has been incorporated in this manual and is
marked with a warning triangle accordingly. Please give this information
your special attention.
The correct use of the system is absolutely vital for safe operation. Please
make yourself totally familiar with the contents of this manual prior to startup of the instrument. Please also observe the user manuals of any additional equipment. Further information is available from our service department or from authorized representatives.
•Please observe all applicable accident prevention regulations.
•The instrument must be connected to a special emergency backup
line supply in accordance with the regulations or directives which apply in your country.
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
Safety15
Notes on installation and use
Safe working order
•Do not operate the equipment contained in the delivery package in
–explosion-risk areas,
–the presence of inflammable anesthetics or volatile solvents such
as alcohol, benzine or similar chemicals.
•Do not station or use the instrument in damp rooms. Do not expose
the instrument to water splashes, dripping water or sprayed water.
•Switch off the unit at the power switch if you notice any smoke, sparks
or unusual noise. Do not use the unit until it has been repaired by our
service team.
•Do not place any fluid-filled containers on top of the instrument. Make
sure that no fluids can seep into the instrument.
•Do not force cable connections. If the male and female parts do not
readily connect, make sure that they are appropriate for one another.
If any of the connectors are damaged, have our service representative
repair them.
•Potential equalization: The instrument can be incorporated into potential equalization measures. For this purpose, contact our service department.
•Do not use a mobile phone in the vicinity of the equipment because
the radio interference can cause the equipment to malfunction. The effects of radio interference on medical equipment depend on a number
of various factors and are therefore entirely unforeseeable.
•Modifications and repairs on these instruments or instruments used
with them may only be performed by our service representative or by
other authorized persons.
•The manufacturer will not accept any liability for damage caused by
unauthorized persons tampering with the instrument; this will also forfeit any rights to claim under warranty.
•Over longer distances (e.g. removal, return for repair, etc), the instrument may only be transported in the original packaging or in special
return packaging. Please contact your dealer or the Carl Zeiss service
team.
•Use this instrument only for the applications described.
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
16Safety
•Only use the instrument with the accessories supplied. Should you
wish to use other accessory equipment, make sure that Carl Zeiss or
the equipment manufacturer has certified that its use will not impair
the safety of instrument.
•Only personnel who have undergone training and instruction are allowed to use this instrument. It is the responsibility of the customer or
institution operating the equipment to train and instruct all staff using
the equipment.
•Keep the user's manuals where they are easily accessible at all times
for the persons operating the instrument.
•Never look at the sun through the binocular tube, the objective lens or
an eyepiece.
•Do not pull at the light guide cable, at the power cord or at other cable
connections.
•This instrument is a high-grade technological product. To ensure optimum performance and safe working order of the instrument, its safety
must be checked once every 12 months. We recommend having this
check performed by our service representative as part of regular maintenance work.
If a failure occurs which you cannot correct using the trouble-shooting
table, attach a sign to the instrument stating it is out of order and contact our service representative.
•Observe the labels showing the symbol "Risk of crushing“!
Notes on EMC (electromagnetic compatibility)
The system meets the EMC requirements of IEC 60601-1-2. During use
of the system, the precautionary measures concerning EMC listed below
must be observed.
Only use accessories that have been approved by Carl Zeiss for this
system.
Do not use any portable or mobile high frequency communication devices
in the vicinity of the system, as this may lead to an impairment of its function.
The system complies with the limits for a Class A device concerning radio
frequency emission. However, the possibility of interference to high frequency receiving devices (e.g. TV sets or radios) being used in the surroundings cannot be ruled out. If interference of this type occurs, please
inform your Carl Zeiss Service.
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
Safety17
Connection of a laser micromanipulator
The link-up of a laser micromanipulator with the OPMI results in a medical
system for which the system manufacturer must meet the necessary requirements (approval, qualification, laser protection, etc.). Please note
the user manuals provided by the laser micromanipulator manufacturer
and laser manufacturer. Further information is available from our service
department or from authorized representatives.
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
18Safety
Connection of navigation systems (option)
The Carl Zeiss components "Surgical microscope on suspension system"
can be integrated into an external navigation system. An optional navigation interface is available.
•The manufacturer of the external navigation system (system supplier)
is responsible for the following:
–Confirmation that his navigation systems have been tested and
–Meeting all requirements (approval, qualifications, etc.) for the
–Provision of all accompanying documents required.
certified for operation with the respective Carl Zeiss surgical microscope on a suspension system in accordance with the requirements specified in the Carl Zeiss interface description "Navigation
Interface for Carl Zeiss Surgical Microscopes".
medical system created through coupling via the navigation interface.
–Ensuring that the navigation system is only connected by person-
nel who have undergone appropriate training and instruction.
–Contacting the local Carl Zeiss representative for any inquiries that
may arise.
–Implementation of a procedure that guarantees the calibration of
the surgical microscope which is absolutely vital for the use of the
Carl Zeiss components "Surgical microscope on suspension system" in combination with a connected navigation system.
This allows the calibrated surgical microscope combined with the
navigation system to be used like an optical pointer with a variable
length (corresponds to the working distance).
–Conducting complete functional testing, alignment and calibration
(landmark test) of the navigation system after every subsequent installation or exchange of navigation system components
–Incorporation of a regularly changing icon in the data injection dis-
play of surgical microscopes equipped with a data injection system, i.e. the "heartbeat" of the navigation system must be
constantly visible for the user to permit data transmission errors to
be immediately detected.
•To check the accuracy of the overall system, perform the test specified
by the navigation system manufacturer, e.g. the landmark test, also
using the surgical microscope. This allows you to ensure that the stereotactic data has been correctly generated and transmitted to the navigation system without errors.
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
Safety19
•Do not use any rotatable tube dovetail mounts when operating the surgical microscope with a connected navigation system. If mounts of this
type have been attached to the microscope, they must be carefully
locked in their central positions (tighten the knurled screw for rotation).
Requirements on the antenna design (non-Zeiss antennas)
•The antenna manufacturer must confirm that his antenna has been
tested and certified for operation with the respective Carl Zeiss surgical microscope on a suspension system in accordance with the requirements specified in the Carl Zeiss interface description
"Navigation Interface for Carl Zeiss Surgical Microscopes".
•The following requirements, in particular, apply to the antenna:
–The surgical microscope with accessories and the antenna must
not exceed the admissible total weight (see label or the chapter
"Technical data").
–The antenna must not project over the bottom edge of the micro-
scope body in the direction of the surgical field.
–The antenna must neither obstruct the movement and adjustability
of the accessories which can be used on the surgical microscope
(usually a stereo coobservation tube, a camera adapter, a face-toface tube for spinal surgery, a laser micromanipulator) nor collide
with these accessories.
–The antenna configuration must be implemented in such a way
that the localizer camera can always "see" the diodes or trackballs
when the surgical microscope and its accessories are in their usual
positions.
–When the system's internal antenna wiring is used, the EN60601-
1-2 standard (electromagnetic compatibility) must be complied
with.
–The output power of the LEDs or infrared LEDs (IRED) must al-
ways comply with the IEC 60825 laser safety standard, even in the
event of cable defects.
–The failure of one or several LEDs of the antenna may impair nav-
igation or lead to navigation failure. Take the necessary precautions.
Requirements for operation
•Our service representative or an expert authorized by us will install the
system. Please ensure that the following requirements are met for further operation:
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
20Safety
–All connecting components (details in the user's manual) which are
–All cables and plugs are in perfect condition.
–The voltage set on the instrument corresponds to the rated line
–The power cord has been plugged into a power outlet which has a
–The power cord being used is the one designed for use with this
Before every use and after re-equipping the instrument
•Make sure that all ”Requirements for operation” are fulfilled.
•Go through the checklist.
relevant to system safety have been properly connected, and the
screw connections have been firmly tightened.
voltage on the site of installation.
properly connected protective ground contact.
system.
•Re-attach or close any covers, panels or caps which have been removed or opened.
•Pay special attention to warning symbols on the instrument (triangular
warning signs with exclamation marks), labels and any parts such as
screws or surfaces painted red.
•Do not cover any ventilation openings.
For every use of the instrument
General
•Never operate the system unattended.
•Avoid looking directly into the light source, e.g. into the microscope objective lens or a light guide.
•When the illumination is on, the light guide must be connected at both
ends. Otherwise there is a risk of fire or burn injuries.
•Make sure that the instrument has been switched off before you
change the xenon lamp module. When switched on, the ignition system generates high voltage.
•The xenon illumination system is a high-intensity light source which if used improperly - can cause thermal injury to skin or tissue. Keep
the exposed tissue moist and provide sufficient irrigation. Carefully
monitor the effects of the illumination on the tissue, in particular in the
following cases:
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
Safety21
–during prolonged procedures on skin and tissue using objective
lenses with a short focal length (short working distance),
–during procedures on tissue with a low blood supply,
–with high brightness settings of the xenon lamp.
Warning!
Due to the high luminance of the xenon lamp and its light spectrum, which
resembles that of natural daylight, its use in ophthalmic applications is
only permissible in combination with the Retrolux illumination module.
•Any kind of radiation has a detrimental effect on biological tissue.This
also applies to the light illuminating the surgical field. Please therefore
reduce the brightness and duration of illumination on the surgical field
to the absolute minimum required.
•Using the locking pedal on the base, secure the stand in position.
Make sure that the stand is stable and cannot roll away.
After every use of the instrument
•Always use the main power switch of the instrument to turn it off.
•The main power switch must always be turned off when the instrument
is not in use.
Phototoxic retinal injury in eye surgery
General
1)-5)
Several papers
surgery have been published. A comprehensive review of these publications reveals five aspects of particular concern:
Studies on the exposure of the eye to light of varying spectral composition
date back to the early 1950s. These studies suggest that the potential
hazard of phototoxic injury to the patient's retina can be reduced by
blocking out the blue and ultraviolet light below a wavelength of 475 nm.
An important point to note here, however, is that the use of filters will inevitably change the perceived color of the light. For this reason, the physician may initially have to get used to the changed appearance of the anatomical structures.
Illumination intensity
The majority of researchers suggest that the surgeon should use the
lowest light intensity required at the patient's eye to guarantee good
viewing during surgery.
Carl Zeiss has addressed this aspect by providing its systems with a device for continuously varying the brightness of the light source. This permits the surgeon to optimally adapt the light intensity at the patient's eye
to the conditions existing in each case.
Angle of illumination
1)-4)
A number of publications
suggest that the microscope should be
tilted to reduce the exposure of the macula to direct illumination.
Carl Zeiss ophthalmic surgical microscopes are therefore equipped with
the following:
–Tilting mechanism for the microscope body
Focus of the light source
Studies show that injuries are likely to occur if the filament of the light
source is imaged on the patient's retina. The peak intensity of a filament
is considerably higher than that of an even and extended light source
such as a light guide.
This is the reason why Carl Zeiss uses fiber optic illumination in its surgical microscope systems.
Exposure time to light
According to some publications, the phakic and aphakic eye should not
be exposed to the light source longer than a few minutes. In all surgical
cases, the retinal exposure time depends on the type and duration of the
procedure and possible case complications. It is therefore recommended
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
Safety23
in ophthalmic surgery to keep the light intensity as low as possible, or to
use a device which prevents the light from entering through the patient's
pupil. It is alsorecommended to make sure that the patient's eye is not additionally exposed to the light of surrounding light sources. This problem
has been solved by Carl Zeiss by the use of a retinal protection device that
can be swung into the illumination beam path of the Retrolux illumination
module.
Brightness control
The brightness control scale of our systems has a linear structure with
values ranging from 1 to 10. The stipulations of standard ISO/DDIS
15004-2:2006(E) result in maximum radiation exposure times for the different illumination configurations as specified in the table "Maximum radiation exposure times".
The microscope light source - like any bright light source - may present a
hazard to the patient's eye both in the form of immediately visible thermal
damage to the retina as well phototoxic chemical reactions which may
lead to photoretinitis. The factors which play an important role in determining the phototoxic risk are:
–Lamp brightness.
–Spectral distribution of the light (UV and blue are more dangerous
that longer wavelengths).
–Duration of direct exposure.
–Pupil size.
–Clarity of ocular media (infants and young children, for example,
may be at a higher risk).
–Phakic status of the eye (aphakic and pseudo-aphakic eyes with-
out UV and blue filtering IOLs are at a higher risk).
–Previous exposure to bright light such as retinal photography, es-
pecially within the last 24 hrs.
The following table is intended to provide the surgeon with a guideline in
determining the potential hazard. The data has been calculated for a
worst-case scenario of direct, uninterrupted exposure of an aphakic eye
with an 8 mm dilated pupil. The calculations are based on the recom-
6)
mended occupational health daily exposure limits as defined in
. A
safety factor of 10 has been used in determining these limits.
During cataract procedures, factors such as lenticular material, instruments such as the phaco handpiece, and movement of the eye provide
interruption of the exposure and would be expected to significantly
lengthen the time before photoretinits might be expected to occur.
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
24Safety
A prospective study 7) of the effects of microscope illumination during surgery did not reveal any phototoxic retinal injuries for procedure times of
up to 30 minutes if the calculated maximum recommended exposure time
was 150 seconds. However, it was also found that at the same brightness
setting, phototoxic retinal injury could be expected after approximately
100 min.
Carl Zeiss recommends the following measures in order to minimize phototoxic risk:
•Always use the lowest possible brightness setting.
•When working on the exterior eye, use the retinal protection device to
prevent light from entering the pupil, especially when the pupil is dilated.
•Turn off the microscope illumination system or cover the patient's eye
during pauses in surgery.
The following table shows the maximum radiation exposure to the xenon
illumination.
Maximum radiation exposure times
Illumination display scaleMax. exposure time
2,53.4 minutes
5,01.5 minutes
7,51.0 minutes
100.8 minutes
Warning!
This exposure time must not be exceeded, as serious eye injury could
otherwise be caused by the xenon illumination.
Note:
A UV blocking filter is contained as a standard feature in the Retrolux illumination module on the microscope side.
This helps the surgeon to reduce the risk of phototoxic retinal injury in the
patient.
List of references
1)
H. Stiller, and B. Rassow, "Light hazards to the patient's retina from
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
Safety25
2)
American Conference of Governmental Industrial Hygienists, "Documentation of the Threshold Limit Values for physical agents. 7th Edition,"
(American Conference of Governmental Industrial Hygienists, Cincinnati,
2001).
3)
S. G. Khwarg, F. A. Linstone, S. A. Daniels, S. J. Isenberg, T. A.
Hanscom, M. Geoghegan, and B. R. Straatsma, "Incidence, risk factors,
and morphology in operating microscope light retinopathy," Am. J. Ophthalmol. 103, 255-263 (1987).
4)
G. Kleinmann, P. Hoffman, E. Schechtman, and A. Pollack, "Microscope-induced retinal phototoxicity in cataract surgery of short duration,"
Ophthalmology 109, 334-338 (2002).
5)
ISO/FDIS 15004-2:2006(E) Optical instruments -- Fundamental requirements and test methods -- Part 2: Light hazard protection
6)
David Sliney, Danielle Aron-Rosa, Francois DeLori, Franz Fankhauser,
Robert Landry, Martin Mainster, John Marshall, Bernard Rassow, Bruce
Stuck, Stephen Trokel, Teresa Motz West, and Michael Wolffe, Adjustment of guidelines for exposure of the eye to optical radiation from ocular
instruments: statement from a task group of the International Commission
on Non-Ionizing Radiation Protection (ICNIRP) APPLIED OPTICS Vol.
44, No. 11, p 2162 (10 April 2005)
7)
Byrnes, G.A., Antoszyk, A.N., Mazur, D.O., Kao, T.C., Miller, S.A.,
Photic maculopathy after extracapsular cataract surgery. A prospective
study, 1992/05/01 Ophthalmology, VL - 99, IS - 5, SP - 731, EP - 737,
PB - Elsevier
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
26Safety
Safety devices of the suspension systems
1Release bar
Allows non-sterile persons to release the magnetic brakes of the suspension system.
2Adjustment screw for limiting the downward travel
Use this screw to set the minimum vertical distance (working distance)
from the surgical field. Check this setting before
dure.
3Locking knob
for locking the suspension arm in its horizontal position.
Before removing or attaching a unit (microscope, tube, etc.), move the
suspension arm into a horizontal position. Pull out the locking knob
and turn it clockwise or counterclockwise through 180°, while slightly
moving the suspension arm up and down until the lock engages.
When locked, the suspension arm can no longer suddenly spring upward when insufficient weight is attached. After attaching a unit, perform the balancing procedure.
each surgical proce-
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
Safety27
1
1
2
3
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
28Safety
Xenon illumination system
Warning!
The xenon lamp has a limited service life of 500 h.
If used beyond its maximum service life, the xenon lamp may explode.
•Replace the xenon lamp in good time.
•Reset the service hour counter to "0" after replacing the lamp.
1Switching to the backup bulb
The lamp module contains two xenon bulbs. The second bulb is used
as a backup bulb which must be swung into the illumination beam path
if the first bulb fails.
•If the xenon bulb fails, open the lamp module as follows: Press
button (7). The lamp module is slightly ejected.
•Pull out the lamp module as far as it will go.
•Turn knob (1) 180° until it snaps in place. This moves the backup bulb
into the illumination beam path.
•Push the lamp module all the way back into the lamp housing.
•Reset service hour counter (6) to "0". Use a pointed object and press
it into the recess of reset button (5).
2Indicator: backup bulb is in use
When the red segment in the knob (1) lights up, the backup bulb is in
use.
Note:
If the first bulb has failed and the backup bulb is in use, make sure to
have a backup lamp module at hand as a precaution.
3Yellow indicator lamp
Lights when the bulb has failed, or if the lamp module is defective.
After activation and ignition of the backup bulb, the yellow indicator
lamp turns off again.
4Manual function
When the manual function has been activated, all electrical control
systems are disabled. The bulb brightness is automatically adjusted to
a fixed setting.
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
Safety29
3
4
7
1
2
256
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
30Safety
Manual function
1Manual key
The Manual key permits you to switch to manual operation. The motorized control functions of the surgical microscope are deactivated.
The lamp brightness is automatically adjusted to a fixed setting, the
value being shown in the first display section.
When the manual mode is activated, the yellow LED is lit and the word
"MANUAL" blinks in the third display section.
The surgical microscope can no longer be operated via the foot control
panel, the handgrips or the display and key field.
In the manual mode, you can only switch the illumination on and off on
the foot control panel and release the magnetic brakes by pressing the
appropriate key on the surgical microscope.
The manual mode is retained even if you turn the power switch of the
instrument off and on again.
Press the Manual key once again to reactivate electronic control; the
display in the display and key field then returns to the basic mode.
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
Safety31
1
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
32Safety
Warning labels and notes
Caution:
Observe all warning labels and notes!
If any label is missing on your instrument or has become illegible, please
contact us or one of our authorized representatives. We will supply the
missing labels.
OPMI VARIO
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
000000-0000-000
Safety33
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
34Safety
1
Suspension systems - xenon illumination system
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
Safety35
176164
S88 floor stand
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
36Safety
S88 floor stand with instrument tray option
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
Description
Description37
OPMI Vario Retrolux System 38
Intended use 38
Components of the OPMI Vario Retrolux System 38
Components of the surgical microscope 40
Controls, displays, connections 42
Binocular tubes and eyepieces 50
Retrolux illumination module 54
Components of the S88 floor stand 56
Description of the modules 57
Design 58
Display field with control keys 60
Lamp housing with xenon light source 62
Suspension arm 68
Stand base with column 70
Connector panel 72
Instrument tray (option) 74
Video monitor (option) 76
Foot control panel (option) 84
Intended use 84
Design 84
Foot control panel with 14 functions 86
Foot control panel with 8 functions 87
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
38Description
OPMI Vario Retrolux System
Intended use
The OPMI® VARIO Retrolux System is a surgical microscope designed
for reconstructive and plastic surgery, for neurosurgery and multidisciplinary use. In combination with the Retrolux illumination module, it can also
be used for ophthalmic procedures.
The S88 floor stand powers and controls the motorized functions of the
surgical microscope. The hallmarks of the S88 floor stand are its superb
mobility and easy operation. Four steerable casters on the stand base
permit easy positioning in the OR. The motorized functions of the surgical
microscope can be controlled using a foot control panel.
The system has been designed for use in hospitals, clinics or other human
medicine institutions.
The system must only be operated by physicians, nurses and other OR
staff who have undergone appropriate training and observe the instructions of the user's manual. The installation conditions and the use of the
system must meet microsurgical requirements:
–low vibration
–clean environment
–avoidance of extreme mechanical stress.
Components of the OPMI Vario Retrolux System
1Surgical microscope
2Coupling
3S88 floor stand
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
Description39
3
2
1
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
40Description
Components of the surgical microscope
The surgical microscope comprises the following modules:
1Coupling
for mounting the surgical microscope on the suspension system.
2Support arm for surgical microscope
3Balancing system including magnetic brakes
This system allows balancing of the surgical microscope. When the
magnetic brakes are unlocked, the surgical microscope can be positioned almost effortlessly.
4Handgrips
for moving the surgical microscope. The buttons on the handgrips permit you to control major microscope functions such as unlocking/locking the magnetic brakes, focusing, zooming.
5Microscope body
The apochromatic optics of the microscope provide superb optical
quality. The microscope image displays optimum contrast and excellent detail recognition along with a large depth of field. The 1:6 ratio
zoom system allows the magnification of the overall system to be set
as required by the surgical procedure.
6Magnetic brake
for the vertical axis.
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
Description41
1
2
3
4
6
3
5
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
42Description
Controls, displays, connections
1Friction adjustment of the vertical axis
Use this knob to adjust the friction of the vertical axis as required.
2Balance setting of the lateral tilt motion
Use this knob to adjust the balance setting of the lateral tilt motion.
3Balance setting of the front-to-back tilt motion
Use this knob to adjust the balance setting of the front-to-back tilt motion.
4Dust cover
5Mouth switch socket
You can use the mouth switch to release or lock the magnetic brakes
of the suspension system used.
6Display window
for reading off the magnification factor γ of the zoom system.
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
Description43
6
1
3
4
2
5
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
44Description
7Focus stop button
This button permits you to deactivate the electrical drive of the focusing system. After you have pressed the focus stop button, you can
only focus manually on the surgical field using knob (9). The focus
stop button is lit. To release the stop, press the focus stop button again
(light in the button goes out).
The use of a micromanipulator for laser applications is described in
the section "Surgical microscope with laser micromanipulator", page
123.
8Zoom knob
Use this knob for manual setting of the magnification.
9Focusing knob
for manual setting of the image definition (focus, working distance)
10 Illuminated-field knob
for manual setting of the illuminated-field diameter.
11 Socket for S light guide
Insert the S light guide into the socket until it snaps in.
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
Description45
7
8
91011
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
46Description
12 Freely programmable release buttons
Specific functions of the suspension system can be assigned to these
buttons, e. g.: increasing / reducing brightness etc.
13 Zoom release button
for setting the magnification factor from 0.4x 2.4x.
14 Focus release button
for continuous focusing within the working distance of 200 to 415 mm.
15 Release button for magnetic brakes
The magnetic brakes of the surgical microscope and suspension
system are released for as long as you press this button.
16 Locking the handgrips in position
Using this screw, you can lock each handgrip in almost any position.
You can swing the handgrips backward by 180° to permit a second
surgeon to operate the microscope in the 180° position.
17 Cable holder
Warning!
To permit almost effortless guidance of the surgical microscope, the microscope and the floor stand must be correctly balanced.
If the system is in an extremely unbalanced state, the unit can move uncontrollably out of position. For this reason, hold the surgical microscope
tightly at its handgrips before unlocking the magnetic brakes.
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
Description47
12
14
15
13
16
17
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
48Description
X-Y coupling (option)
Note:
•The OPMI® VARIO can be equipped (and also retrofitted) with an X-Y
coupling (2). Our service team or an authorized person will install the
X-Y coupling for you.
The X-Y coupling allows motorized fine positioning of the surgical microscope in a horizontal plane. The range of travel is 40 mm x 40 mm. The
speed of travel can be set on the display field of the suspension system.
The X-Y coupling is provided with a recentering mechanism. When you
press button (3), the X-Y coupling moves back into its center position.
You can also trigger the recentering function using the freely programmable buttons on the handgrips or foot control panel.
Knob (1) permits you to adjust the friction of the microscope's rotary axis.
See also: Preparations for use / Aligning the X-Y coupling, page 108.
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
Description49
2
3
1
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
50Description
Binocular tubes and eyepieces
180° tiltable tube
1PD adjustment knob
The correct position has been reached when the two eyepiece images
merge into one. You can read off the interpupillary distance set on the
adjustment knob.
2180° tiltable tube
3Eyepiece mount
45° inclined tube
445° inclined tube
5PD adjustment knob
The correct position has been reached when the two eyepiece images
merge into one. You can read off the interpupillary distance set on the
adjustment knob.
6Eyepiece mount
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
Description51
4
5
1
2
3
6
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
52Description
Widefield eyepieces with magnetic coupling
Note:
When you remove these eyepieces from the tube, please note that they
are fitted with a magnetic coupling. When mounted, the eyepieces display
a very weak magnetic field, so that the usual rules for the handling of magnets must only be observed with eyepieces which have not been mounted
on the microscope:
•Do not place the eyepieces close to instruments where there is any
risk of magnetization.
•Do not place the eyepieces on sensitive electronic units such as infusion pumps, cardiac pace-makers, measuring instruments or magnetic data carriers such as disks, audiotapes and videotapes, or credit
cards.
•Always store eyepieces not used in their original packaging.
1Eyecup
Always adjust the eyecups in such a way that you can see the full field
of view.
–Viewing with eyeglasses:Screw in the eyecups all the way.
–Viewing without eyeglasses: Screw out the eyecups until you
see the full field of view.
2Diopter adjustment ring
The eyepieces provide ametropia compensation between -8 D and
+5 D. Eyeglass wearers who perform surgery wearing their glasses
set the diopter adjustment ring to 0 D. Turn the ring until you have obtained the optimum setting. An integrated brake holds the ring in the
position set.
3Diopter scale
for reading the prescription set.
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
Description53
1
2
3
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
54Description
Retrolux illumination module
1Retrolux illumination module
enables the use of the OPMI Vario for ophthalmic procedures.
2Switch for swinging the retinal protection device into position
prevents light entering the patient's eye and therefore offers additional
protection against phototoxic injury.
3Dovetail
for the attachment of a fundus imaging system.
4Dovetail
for the attachment of the Retrolux illumination module
5Light guide connector
for the attachment of the light guide.
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
Description55
1234
5
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
56Description
Components of the S88 floor stand
The S88 floor stand is a carrier system for the surgical microscope. It is
used to power and control the motorized functions of the surgical microscope. The hallmarks of the floor stand are its superb mobility and easy
operation. Four steerable casters on the stand base permit easy positioning in the OR. The motorized functions of the surgical microscope can
be activated using a foot control panel.
Further useful functions include, for example:
–magnetic brakes for almost effortless positioning,
–brightness control via a foot control panel,
–reset of X-Y coupling, focus and zoom,
–user-defined basic settings for a maximum of nine users:
–speeds for focus, zoom and X-Y coupling
–and configurable buttons on the foot control panel for focus mem-
ory, XY inversion, camera release, swinging SDI in/out, triggering
an AUX signal.
Warning!
The use of OPMI Vario for ophthalmic procedures is only permissible in
combination with the Retrolux illumination module.
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
Description57
Description of the modules
The floor stand comprises an articulated arm, a stand column and a stand
base. The articulated arm comprises a carrier arm and a suspension arm.
The carrier arm contains the control unit with all electrical supply systems
required for the control of a motorized surgical microscope. These motorized functions can be activated using a foot control panel.
The suspension arm permits almost effortless positioning of the surgical
microscope. The spring force of the suspension arm can be varied in a
range from 8 to 20 kg, permitting reliable balancing of the microscope
even with heavy accessory equipment. The downward movement of the
suspension arm can be limited as required.
A maneuvering handle attached to the stand column is used to move the
stand and to attach the foot control panel. The stand column is provided
on its left and right with cable supports for winding up cables before the
unit is relocated. Four steerable casters on the stand base permit easy
positioning near the operating table. The stand base has been designed
in such a way that high stability is ensured even with unfavorable loading
of the stand. A locking pedal is provided to lock the floor stand quickly and
reliably into position.
Caution:
As the stand is very easy to maneuver, there is a tendency to underestimate its considerable weight. Therefore, move the stand slowly and carefully!
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
58Description
Design
1Control panel
2Carrier arm
3Xenon illumination system, see page 62
4Suspension arm
5Stand base
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
Description59
1
3
24
5
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
60Description
MODE
STORE MANUAL
USER
1
1
1.5
Xe
Basic mode
Display field with control keys
The display and control panel is integrated in the control unit.
The surgical microscope on the suspension system can be controlled either manually or electronically. The control software required for electronic control is installed in the electronics box of the suspension system.
You operate the software via the control and display panel, where you can
read off and reconfigure the current settings.
The control and display panel is structured as follows:
–Three display fields (LCD) with the associated keys "∇" and "Δ".
–One row of keys comprising the MODE, STORE and MANUAL keys,
and a yellow LED above the MANUAL key.
User interface
The user interface of the suspension system comprises three display
fields and keys located beside and below them.
A pair of keys "∇" and "Δ" has been assigned to every display field for
making the appropriate settings.
The control functions have been combined in several modes (menu
pages). The basic mode is always displayed in the normal operating
status.
The following is displayed in the basic mode:
–the current lamp brightness of lamp 1 (xenon) in the upper display
field,
–Xe for xenon in the middle display field,
–the current user ID in the lower display field.
Key row
Three keys and an LED are provided below the display fields.
Use the "MODE", "STORE" and "MANUAL" keys to select the different
control functions (modes).
"MODE" key and "STORE" key
The "MODE" and "STORE" keys permit you to access the different modes
of the user interface. For details, please see the chapter "Operation".
"STORE" key
Use the "STORE" key to save the current focus and zoom settings.
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
Description61
"MANUAL" key
The "MANUAL" key permits you to switch to manual operation. For details, please see the chapter "Operation".
Yellow LED above the "MANUAL" key
The yellow LED is lit when you have switched to the manual mode.
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
62Description
Lamp housing with xenon light source
Warning!
The use of OPMI Vario for ophthalmic procedures is only permissible in
combination with the Retrolux illumination module.
The suspension system is equipped with a xenon illumination system for
fiber illumination. The xenon lamp generates light whose spectrum resembles that of natural daylight. Regardless of the brightness setting, the
color temperature of the light always remains the same. Normal daylight
film without any additional conversion filters can therefore be used for
photographic documentation. The lamp housing contains two xenon
lamps. The second lamp is used as a backup lamp which must be swung
into the illumination beam path should the first lamp fail.
Ventilation grid
Do not cover the ventilation grid! For example, drapes could be covering
the grid. This can cause the lamp modules to overheat and lead to lamp
failure.
1Lamp module
2Manual activation of the backup bulb
•If the xenon bulb fails, open the lamp module as follows: Press
button (7). The lamp module is slightly ejected.
•Pull out the lamp module as far as it will go.
•Turn knob (2) 180° until it snaps in place. This moves the backup bulb
into the illumination beam path.
•Push the lamp module all the way back into the lamp housing.
•Reset service hour counter (5) to "0". Use a pointed object and press
it into the recess of reset button (6).
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
Description63
2
4
3
1
6
3
7
5
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
64Description
Note:
When inserting a new lamp module, make sure that the knob (2) is set
to "1“. If the first lamp fails, switch to the second lamp in logical sequence.
3Indicator: backup lamp is in use
When the red segment in the knob (2) lights up, the backup lamp is in
use.
4Filter selector knob
The filter knob has two positions:
0No filter
1Filter swung in
(No integrated filter in the standard configuration)
5Counter
The counter records the service hours of the xenon light source.
•Change the xenon lamps after about 500 hours of operation to prevent
them from exploding. Then reset the counter to "0" by pressing reset
button (6).
6Reset button
The reset button resets the service hour counter to "0".
7Opening the lamp module
When you press this button, the lamp module is slightly ejected.
•To change the lamp, pull out the lamp module as far as it will go. Turn
the knob (2) through 180° until it snaps in place. This moves the backup lamp into the illumination beam path.
Warning!
The xenon lamp has a limited service life of 500 h.
If used beyond its maximum service life, the xenon lamp may explode.
•Replace the xenon lamp in good time.
•Reset the service hour counter to "0" after replacing the lamp.
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
Description65
2
4
3
1
6
3
7
5
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
66Description
8Brightness control
You can adjust the brightness using the two control keys on the control
panel.
Note:
The brightness of the xenon lamp can also be adjusted by pressing
the appropriate buttons on the foot control panel.
9Yellow indicator lamp
Lights when the lamp has failed, or if the lamp module is defective.
After activation and ignition of the backup lamp, the yellow indicator
lamp goes out again.
Note:
If the first lamp has failed and the backup lamp is in use, make sure to
have a backup lamp module ready at hand as a precaution.
10 Green indicator lamp
Lights when the illumination has been switched on.
11 Selector switch:
Illumination is off.
Illumination is on.
Illumination can be switched on/off on the left-hand side of
the foot control panel.
Illumination can be switched on/off on the right-hand side of
the foot control panel.
Note:
You can adjust the selector in such a way that you can switch the illumination on/off on the right-hand and left-hand sides of the foot control
panel.
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
Description67
8
11
9
10
0
,
7
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
68Description
Suspension arm
1Lock of the cable duct
–For opening, turn a quarter turn clockwise or counterclockwise.
–For closing, press down and turn a quarter turn clockwise or coun-
terclockwise.
2Adjustment screw for limiting downward movement
Use this screw to set the minimum vertical working distance from the
surgical field. Bring the surgical microscope into its working position.
Turn the adjustment screw for limiting downward movement clockwise
as far as it will go. Adjust the downward movement limitation before
each surgical procedure.
3Balance setting screw
After mounting the surgical microscope including all
just the balance setting of the suspension arm using this screw. Balance setting is described in detail in chapter “Operation“.
accessories, ad-
4Securing screw
®
for securing the OPMI
coupling.
5Locking screw
for locking the suspension arm in a horizontal position for mounting
the surgical microscope. This prevents the suspension arm from uncontrollably moving upward when insufficient weight is attached.
6Release bar
Allows non-sterile persons to unlock the magnetic brakes of the suspension system.
Magnetic brake release buttons
The magnetic brake release buttons are located on the surgical microscope. For as long as you press one of these buttons, you can move
the articulated arm in all directions. When you let go of the button, the
magnetic brakes will lock all axes in position at the same time.
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
Description69
12
5
3
4
6
6
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
70Description
Stand base with column
1Handle
for moving the stand.
2Support
for hanging up the foot control panel during transport.
3Cable support (2x)
for winding up the power cord and the cable of the foot control panel.
4Cable deflectors
are provided to prevent cables on the floor from being run over and
damaged.
5Locking pedal
Press once to lock the stand in position.
Press a second time to release the locking pedal.
6Steerable casters
The four steerable casters on the stand base permit easy positioning
in the OR.
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
Description71
1
2
5
6
3
4
4
4
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
72Description
Connector panel
1Remote control socket
for triggering an AUX signal, e.g. to switch on/off an external device
operating at max. 24V/0.5A.
2Connector for switching component
Connection possibility for: a foot control panel, a hand control panel or
an operating chair with an appropriate foot switch.
3Potential equalization bolt
4Indicator window for rated voltage
The voltage shown here must correspond to the rated line voltage provided on the site of installation. You can adjust the sliding switch using
a suitable tool.
Warning!
Please observe the maximum current consumption of two power outlets
(4) and (5). Only connect medical devices approved by us to outlets (4)
and (5). When using other instruments, make sure that safety is guaranteed regarding admissible ground leakage currents. The admissible limit
value of the ground leakage current present in the suspension system's
power cord must not exceed 500 µA in compliance with EN60601-1/IEC
60601-1. CSA/US certification in compliance with UL 60601-1 only allows
a maximum ground leakage current of 300 µA.
5Power outlet socket
for medical devices with a maximum current consumption of 2 A.
Note:
The current of this power outlet is switched on/off using power
switch S2 (7).
6Power outlet socket
for medical devices with a maximum current consumption of 5 A.
7Power inlet socket
8Power switch S2
When the stand is on, the green indicator light in the switch is lit.
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
Description73
S2
4
5
6
7
8
2
3
9
1
9Strain relief device
The strain relief device prevents inadvertent unplugging of the following electrical connections:
–power cable,
–connecting cable for foot control panel, hand control panel or op-
erating chair with appropriate footswitch.
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
74Description
Instrument tray (option)
The S88 floor stand can be equipped or retrofitted with an instrument tray
(1). In the case of retrofitting, our service staff or an authorized person will
install the instrument tray on your suspension system.
The instrument tray (1) can carry a maximum load of 13 kg. The tray has
been designed, for example, for mounting MediLive Trio from Zeiss:
MediLive Trio is attached to the instrument tray with the aid of two stud
bolts. (The four receptacles supplied with the instrument tray are not required for mounting MediLive Trio.)
A second MediLiveTrio or other accessory equipment can be mounted on
the instrument tray using the strap provided. Please note the instrument
tray's maximum load capacity of 13 kg.
Warning!
•Make sure that the accessory equipment is positioned as securely as
possible on the instrument tray.
•Mount MediLive Trio on the instrument tray using the two stud bolts.
•If required, secure further accessory equipment on the tray using the
strap provided.
•Do not
•Remember there is a risk of collision and crushing when suspension
arm (2) is folded to its moving position. A "Risk of crushing" warning
label is therefore attached on the left and right of suspension arm (2).
•Please read the relevant user manual before starting up the accessory
equipment.
•Never pull or push at the accessory equipment (3) in order to move the
S88 floor stand. Always use handle (4) to move the S88 floor stand.
place a load of more than 13 kg on instrument tray (1).
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
Description75
1
3
4
2
1
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
76Description
Video monitor (option)
The S88 floor stand can be equipped or retrofitted with a TFT monitor (1).
In the case of retrofitting, our service staff or an authorized person will install the TFT monitor on your suspension system.
The TFT monitor (1) features a 15" screen and permits the scrub nurse
and other OR staff to follow the surgical procedure. For optimum viewing,
the TFT monitor can be precisely positioned using flexible arm (2).
Warning!
Risk of injury due to lowering of the TFT monitor!
•Ageing processes may lead to the loss of gas in the gas pressure
spring of the monitor's suspension arm, causing the TFT monitor to
move downward of its own accord.
•Compensate for the loss of gas by readjusting the gas pressure spring
as described on page 180.
If the TFT monitor continues to move downward, the gas pressure spring
is defective.
•Notify our service representative.
Warning!
Do not use the stored video sequences, video clips (cut sequences) and
single images for diagnostic purposes, as the video cameras and the
monitor have not been calibrated. The visualized images may therefore
include deviations in scale, color and shape.
This applies in particular if the sequences, clips or images are viewed on
a monitor outside the OR, as the transmission to a different monitor may
lead to changes in the display.
Note:
The background illumination of the LCD display has a limited service life.
If you notice that the display is getting darker or starts to flicker, contact
your Zeiss dealer.
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
Description77
2
1
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
78Description
Components
The principal component of the TFT monitor is the 15" screen which delivers flawless, sharp images even at low frame rates of 50 Hz.
The connectors and controls are located under cable cover (2) on the
back of the TFT monitor. To access the connectors and controls, proceed
as follows:
•Remove two screws (3) from cable cover (2) by turning them counterclockwise.
•Remove cable cover (2) by pushing it upward.
Controls
1Auto Adjust button
The Auto Adjust function permits automatic image adjustment of the
TFT monitor to obtain a sharp, optimally positioned image. Perform
the automatic adjustment when initially starting up the TFT monitor or
after making any changes to the system.
Note:
–The Auto Adjust function can only be executed if a VGA signal
source is connected.
–Always use a normal camera image for automatic image adjust-
ment. Do not use the test image which is displayed directly after
power-on or when no camera head is connected.
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
Description79
323
1
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
80Description
Connector panel
2Power supply
for power and voltage supply of the device.
Caution:
Only operate the device with the power cable included in the delivery
package.
3DVI connector
DVI stands for Digital Video Interface and is the latest technology for
digital data transmission. The cable length for this connector must not
exceed 4.5 m.
Note:
We recommend the DVI connector for the connection of a camera, as
it delivers optimum image quality with minimum flicker. The DVI cable
is included in the delivery package.
4VGA connector
VGA stands for Video Graphics Array and is an analog interface for
video data transmission between graphics cards and display devices.
5S-Video connector
S-Video - also known as separate video or Y/C - is an analog interface
that transmits brightness and color information as separate signals.
This standard provides higher video image quality than Composite
Video. The maximum cable length for this connector should not exceed 10 m. For longer cables, please use the Composite Video connector.
6Composite Video connector (1x cinch)
This is an analog interface that transmits the composite video signal
via a single channel (yellow cinch connector). This connector is particularly suitable for transmitting video signals over long distances.
7Component connector (3x cinch)
This is an analog interface that transmits the component video signal
via three channels (red, yellow, green cinch connector). Each channel
transmits one of the primary colors.
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
Description81
2345 67
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
82Description
Powering on the TFT monitor
To facilitate the operation of the TFT monitor, it is automatically activated
when the suspension system is switched on.
During the power-on process, the TFT monitor executes a power-on sequence in which the signals on the connectors (DVI, VGA, S-Video, Composite and Component) are checked. After detection of the signal available, the correct screen resolution and frame rate are set automatically.
Note:
If a signal source is present at the VGA connector, the image settings can
be optimized using the Auto Adjust function as described on page 78.
Aligining the video monitor
The best visualization is obtained when you are looking straight at the
screen of the TFT monitor.
•Swivel the monitor's carrier arm (1) and suspension arm (2) into the
horizontal position required.
Caution:
Take care not to damage the video cable!
Do not swivel the suspension arm beyond ±120°.
•Tilt the suspension arm (2) upward or downward until the required
height has been reached.
•Hold the upper corners of the TFT monitor and adjust it to the required
angle via ball joint (3).
Warning!
Risk of injury due to lowering of the TFT monitor!
•Ageing processes may lead to the loss of gas in the gas pressure
spring of the monitor's suspension arm, causing the TFT monitor to
move downward of its own accord.
•Compensate for the loss of gas by readjusting the gas pressure spring
as described on page 180.
If the TFT monitor continues to move downward, the gas pressure spring
is defective.
•Notify our service representative.
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
±90°
360°
90°
45°
2
1
3
Description83
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
84Description
Foot control panel (option)
Intended use
The foot control panel permits you to operate various functions of a suspension system or surgical microscope. The assignment of functions to
the controls on the foot control panel is described on the next page. Only
functions provided in the system configuration used (suspension system,
surgical microscope) can be controlled.
Design
The 8-function and 14-function foot control panels feature two rocker
switches (1) which are used to control the ”Zoom” and ”Focus” functions.
The switches for these functions are located on one side, allowing you to
control both directions of a function using toe/heel movement without
having to shift your foot. Bridge (3) between two rocker switches (1)
serves as a support to rest your foot on.
Connector (2) is used to connect the foot control panel to the appropriate
connector on a suspension system or a wall-mounted control panel.
The foot control panel is provided with a waterproof rubber cover.
The illustration shows the 14-function foot control panel. Depending on
the configuration involved, an 8-function foot control panel can also be
connected.
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
Description85
2
1
3
1
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
86Description
11
10
9
8
7
2
3
4
5
6
1
A
B
D
C
Foot control panel with 14 functions
The illustration shows the standard assignment of functions to the foot
control panel with 14 functions. The focus/zoom assignment can be
changed by our service staff, on request.
1Control of an X-Y coupling (if installed)
2Button A: Reducing the lamp brightness
3Button C: Function as assigned in configuration mode 1
4ZOOM: Reducing magnification, increasing the field of view
(alternatively FOCUS: Increasing the working distance)
5ZOOM: Increasing magnification, reducing the field of view
6Illumination on/off
7Illumination on/off
8FOCUS: Increasing the working distance (alternatively ZOOM:
Reducing magnification, increasing the field of view)
9FOCUS: Reducing the working distance
10 Button D: Function as assigned in configuration mode 1
11 Button B: Increasing the lamp brightness
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
Description87
10
9
8
7
6
1
2
3
4
5
D
C
Foot control panel with 8 functions
The illustration shows the standard assignment of functions to the foot
control panel with 8 functions.
The focus/zoom assignment can be changed by our service staff, on request.
1Reducing the lamp brightness
2No function
3ZOOM: Reducing magnification, increasing the field of view
(alternatively FOCUS: Increasing the working distance)
4ZOOM: Increasing magnification, reducing the field of view
5Button C: Function as assigned in configuration mode 1
6Button D: Function as assigned in configuration mode 1
7FOCUS: Increasing the working distance (alternatively ZOOM:
Reducing magnification, increasing the field of view)
8FOCUS: Reducing the working distance
9No function
10 Increasing the lamp brightness
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
88Description
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
Preparations
Preparations89
Attaching the equipment 90
Mounting the surgical microscope 90
Attaching accessories 92
Mounting the tube and eyepieces 94
Mounting the Retrolux illumination module 96
Bringing the Retrolux illumination module into its standby position 100
Attaching the fundus imaging system to the illumination module 102
Connections 104
Connecting the surgical microscope 104
Mounting the light guide 104
Aligning the X-Y coupling 108
Strain relief device on S88 floor stand 110
Connecting the S88 floor stand 112
Adjusting the supension system 114
Adjusting the balance setting of the suspension arm 114
Adjusting the limit of downward movement 116
Balancing the surgical microscope 118
Settings on the control and display panel 120
Adjusting the suspension system 120
Adjusting the surgical microscope 121
Surgical microscope with a laser micromanipulator 123
Relocating the system 126
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
90Preparations
Attaching the equipment
Mounting the surgical microscope
Warning!
The maximum
exceed 20 kg!
•Bring the suspension arm into its horizontal position, pull out locking
knob (3) and turn it clockwise or counterclockwise through 180°. At the
same time, slightly move the suspension arm up and down until the
lock snaps in. This prevents the suspension arm from uncontrollably
moving upward when insufficient weight is attached.
•Turn off the system at the power switch.
•Hook coupling (2) with the attached surgical microscope at a slight angle from above into receptacle (1) on the suspension arm, and tilt the
coupling downward into its vertical position.
•Tighten securing screw (6) firmly using a hex key.
•Insert cable clip (4) into opening (5) of the coupling.
•Perform the balance setting procedure.
Warning!
Before every use and after re-equipping the system, make sure that securing screw (6) has been firmly tightened!
weight of the microscope including accessories must not
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
Preparations91
3
2
4
1
6
5
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
92Preparations
Attaching accessories
The comprehensive accessory program permits the OPMI Vario surgical
microscope to be configured to your specific needs.
Make sure not to exceed the maximum admissible load in order to ensure
optimum mobility and reliable operation of your OPMI Vario. The weight
of the complete accessory configuration must not be higher than 9.0 kg.
For example, the combined accessory modules of configuration (1)
without the video objective lens and video camera have a weight of
3.9 kg.
Caution:
•Select your configurations so that the weight distribution is as
symmetrical as possible.
•Please observe the maximum permissible load on the suspension
system.
•The extensive removal of accessory modules may lead to unintended
tilting of the microscope body. Rebalance the surgical microscope in
this case.
•For mounting the accessory modules, proceed as described in the following section "Mounting the tube and eyepieces" () for the binocular
tube.
•Attach the accessory modules to the microscope and firmly
respective securing screw.
Warning!
•Before every use and after re-equipping the instrument, make sure
that the modules are securely locked in position.
•Make sure that all securing screws are firmly
•Accurately re-balance the surgical microscope after every reconfiguration of the system.
1Recommended configuration for lateral coobservation, left or right
Accessory modules: beam splitter, 180° tiltable tube, stereo coobservation tube with straight tube, video objective lens with video camera.
2Recommended configuration for face-to-face coobservation
Accessory modules: stereobridge, 2 x rotary adapter and 180° tiltable
tube, video objective lens with video camera.
tightened!
tighten the
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
Preparations93
2
1
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
94Preparations
Mounting the tube and eyepieces
•Loosen securing screw (4) by a few turns.
•Remove cover (3) and store it in a safe place.
•Place binocular tube (2) on the surgical microscope and tighten securing screw (4) firmly
•You can install accessory equipment between the binocular tube and
the microscope body. Lock these units in position in the same way using securing screw (4).
•Insert widefield eyepieces (1) into the appropriate mounts as far as
they will go. The magnetic coupling reliably secures them in position.
Note:
If a documentation device is used, one of the eyepieces can be delivered
with a reticle to aid focusing. Retrofitting the reticle is only possible at the
factory or by our service representative. Always use the eyepiece with the
reticle on the same side of the binocular tube as the documentation device is installed.
.
Warning!
Before using and after re-equipping the unit, make sure that binocular
tube (2) is securely locked in position. Make sure that securing screw (4)
is firmly
tightened!
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
Preparations95
1
2
3
4
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
96Preparations
Mounting the Retrolux illumination module
Mounting the illumination module
Caution:
Limited field of view and image quality caused by incorrect attachment of
the Retrolux illumination module!
Attach the Retrolux illumination module to the dovetail (1) on the underside of the OPMI Vario.
•Loosen the knurled screw (2) of the dovetail far enough to allow the
Retrolux illumination module to be pushed easily into the guideway of
the dovetail mount. The knurled screw (2) must not project into the
guideway.
•Push the Retrolux illumination module into the dovetail mount in the
direction of the arrow as far as it will go.
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
Preparations97
1
2
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
98Preparations
Locking the illumination module into position and connecting the
light guide
Caution:
Risk of damage to the thread of the knurled screw!
Do not overtighten knurled screw (2).
•Tighten the knurled screw (2) by hand in a clockwise direction.
Warning!
Danger of dazzle at the light guide!
•Switch off the OPMI Vario at the power switch before changing the
light guide.
•Remove the light guide (4) from the OPMI Vario.
•Insert the light guide (4) in the light guide connector of the Retrolux illumination module (3).
Removing the illumination module
Warning!
Danger of dazzle at the light guide!
•Switch off the OPMI Vario at the power switch before changing the
light guide.
•Remove the light guide from the Retrolux illumination module.
•Insert the light guide in the light guide connector (5) on the OPMI
Vario.
•Loosen the knurled screw by turning it counterclockwise.
•Pull the Retrolux illumination module out of the dovetail mount in the
direction of the arrow.
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
Preparations99
2
4
3
5
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
100Preparations
Bringing the Retrolux illumination module into its standby position
If you do not need the Retrolux illumination module, you can bring it into
its standby position on a dovetail mount The dovetail mount and the appropriate suspension cover must be mounted on the suspension. Proceed as follows:
Mounting the illumination module on the dovetail mount
Caution:
If not properly locked in position, the system may be damaged.
•Lock the system in position by switching it off at the power switch.
When the system is separated from line power, the magnetic brakes
lock the axes and the system remains in position.
•Remove protective foil (1) on the underside of the suspension.
•Remove screws (2) on the underside of the suspension by turning
them counterclockwise.
•Attach suspension cover (3) to the suspension and firmly tighten
screws (2) in a clockwise direction.
•Attach dovetail mount (4) by placing the groove on projection (5).
•Screw down dovetail mount (4) on projection (5). Use the accompanying hex screw for this purpose.
•Push illumination module (6) into dovetail mount (7) in the direction of
the arrow.
•Tighten the knurled screw on the illumination module.
•Turn on the system at its power switch.
Removing the illumination module from the standby position
•Loosen the knurled screw of the illumination module.
•Push illumination module (6) out of dovetail mount (7) in the direction
of the arrow.
G-30-1721-en OPMI® VARIO Retrolux System Issue 1.0
Printed on 12. 03. 2008
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.