Zeiss OPMI PRO User manual [gb]

Page 1
OPMI® PRO magis
Surgical Microscope
Instructions for use
G-30-1292-en
Issue 3.0
Printed on 30. 07. 2001
Page 2
Page 3
Contents
Functions at a glance 5
OPMI® PRO magis surgical microscope 6 – Key for symbols 7
Safety 9
Directives and standards 10 – Notes on installation and use 11 – Warning labels and notes 14
Description 17
OPMI® PRO magis 18 – Intended use 18 – Controls, displays, connections 20 – Tiltable binocular tube and eyepieces 26
Preparations for use 31
Attaching the equipment 32 – Attaching the tube, eyepieces, objective lens and handgrips 32
Operation 35
Preparations for use 36 – Balancing the surgical microscope 36 – Adjusting the surgical microscope 41 – Checklist 42 Procedure 44 What to do in an emergency 45 – Failure of the zoom function 45 – Failure of the focusing function 46 – Failure of magnetic brakes 46
Maintenance / Further information 47
Trouble-shooting table 48
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
Printed on 30. 07. 2001
Page 4
Care of the unit 50 – Sterilization 51 – Magnifications / Fields of view 52 – Ordering data 53 – Spare parts 54 – Accessories 55 – Technical data 56 – Ambient requirements 58
Index 59
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
Printed on 30. 07. 2001
Page 5
Functions at a glance 5

Functions at a glance

OPMI® PRO magis surgical microscope 6 Key for symbols 7
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
Printed on 30. 07. 2001
Page 6
6 Functions at a glance

OPMI® PRO magis surgical microscope

1 Setting the friction of the microscope's axis of rotation 2 Setting the interpupillary distance 3 Adjusting the eyecups 4 Setting the prescription on the diopter scale 5 Display of the zoom system's magnification factor 6 Releasing the magnetic brakes (three keys each on the right and left) 7 Adjusting the friction of the front-to-back tilt motion 8 Balance setting of the front-to-back tilt of the microscope 9 Setting the friction of the lateral tilt motion 10 Balance setting of the lateral tilt of the microscope 11 Connecting the light guide 12 Adjusting the illuminated field diameter 13 Moving to the starting position of the focusing range 14 Motorized focusing (left and right on handgrip) 15 Motorized zooming (left and right on handgrip)
1
2
3
4
5
6 12
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
151411 1310987
Printed on 30. 07. 2001
Page 7

Key for symbols

Functions at a glance 7
Warning!
The warning triangle indicates potential sources of danger for per- sons which may constitute a risk of injury for the user or a health ha­zard.
Caution:
The square indicates situations which may lead to malfunction, de­fects, collision or damage of the instrument.
Note:
The hand indicates hints for the use of the instrument or other tips for the user.
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
Printed on 30. 07. 2001
Page 8
8 Functions at a glance
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
Printed on 30. 07. 2001
Page 9

Safety

Safety 9
Directives and standards 10 Notes on installation and use 11 Warning labels and notes 14
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
Printed on 30. 07. 2001
Page 10
10 Safety
The instrument described in this manual has been developed and tested in accordance with Carl Zeiss safety standards and with national and in­ternational regulations. A high degree of instrument safety is thus en­sured.
We would like to inform you on the safety aspects involved in operating the instrument. This chapter contains a summary of the most important precautions to be observed.
Further safety notes are also contained in other parts of this user's manual; they are marked with a warning triangle containing an exclama­tion mark as shown here. Please pay special attention to these safety notes.
Safety is only ensured when this instrument is operated properly. Please read through this manual carefully before turning the instrument on. Also read through the user's manuals of the other equipment used with this in­strument. You may obtain further information from our service organiza­tion or authorized representatives.

Directives and standards

The instrument described in this manual has been designed in compli­ance with the following standards:
In accordance with Directive 93/42/EEC, Annex II, Article 3, the quality­management system of Carl Zeiss has been approved by TÜV Rheinland, which is a notified body. The registration number is: 95 102 7601.
The instrument must be connected to a special emergency backup
This is a class I instrument as defined by Directive 93/42 /EEC.
Please observe all applicable accident prevention regulations.
–EN
–IEC
–UL
–CSA
line supply in accordance with the regulations or directives which ap­ply in your country.
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
Printed on 30. 07. 2001
Page 11
Safety 11

Notes on installation and use

Safe working order
Do not operate the equipment contained in the delivery package in
explosion-risk areas,
the presence of inflammable anesthetics or volatile solvents such
as alcohol, benzine or similar chemicals.
Do not station or use the instrument in damp rooms. Do not expose the instrument to water splashes, dripping water or sprayed water.
Immediately unplug any equipment that gives off smoke, sparks or strange noises. Do not use the instrument until our service represent­ative has repaired it.
Do not place any fluid-filled containers on top of the instrument. Make sure that no fluids can seep into the instrument.
Do not force cable connections. If the male and female parts do not readily connect, make sure that they are appropriate for one another. If any of the connectors are damaged, have our service representative repair them.
Potential equalization: The instrument can be incorporated into poten­tial equalization measures. For this purpose, contact our service de­partment.
Do not use a mobile phone in the vicinity of the equipment because the radio interference can cause the equipment to malfunction. The ef­fects of radio interference on medical equipment depend on a number of various factors and are therefore entirely unforeseeable.
Modifications and repairs on these instruments or instruments used with them may only be performed by our service representative or by other authorized persons.
The manufacturer will not accept any liability for damage caused by unauthorized persons tampering with the instrument; this will also for­feit any rights to claim under warranty.
Use this instrument only for the applications described.
Only use the instrument with the accessories supplied. Should you wish to use other accessory equipment, make sure that Carl Zeiss or the equipment manufacturer has certified that its use will not impair the safety of instrument.
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
Printed on 30. 07. 2001
Page 12
12 Safety
Only personnel who have undergone training and instruction are al­lowed to use this instrument. It is the responsibility of the customer or institution operating the equipment to train and instruct all staff using the equipment.
Keep the user's manuals where they are easily accessible at all times for the persons operating the instrument.
Never look at the sun through the binocular tube, the objective lens or an eyepiece.
Do not pull at the light guide cable, at the power cord or at other cable connections.
This instrument is a high-grade technological product. To ensure opti-
mum performance and safe working order of the instrument, its safety must be checked once every 12 months. We recommend having this check performed by our service representative as part of regular main­tenance work.
If a failure occurs which you cannot correct using the trouble-shooting table, attach a sign to the instrument stating it is out of order and con­tact our service representative.
Requirements for operation
Our service representative or a specialist authorized by us will install the instrument. Please make sure that the following requirements for opera­tion remain fulfilled in the future:
All mechanical connections (details in the user's manual) which are
relevant to safety are properly connected and screw connections tight­ened.
All cables and plugs are in good working condition.
The voltage setting on the instrument conforms to the rated voltage of
the line supply on site.
The instrument is plugged into a power outlet which has a properly
connected protective earth contact.
The power cord being used is the one designed for use with this in-
strument.
Before every use and after re-equipping the instrument
Make sure that all ”Requirements for operation” are fulfilled.
Go through the checklist.
Re-attach or close any covers, panels or caps which have been re­moved or opened.
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
Printed on 30. 07. 2001
Page 13
Safety 13
Pay special attention to warning symbols on the instrument (triangular warning signs with exclamation marks), labels and any parts such as screws or surfaces painted red.
For every use of the instrument
Avoid looking directly into the light source, e.g. into the microscope ob­jective lens or a light guide.
Any kind of radiation has a detrimental effect on biological tissue.This also applies to the light illuminating the surgical field. Please therefore reduce the brightness and duration of illumination on the surgical field to the absolute minimum required.
When operating on the eye, always use a GG 475 protection filter to ensure that the patient's retina is not exposed to unnecessary (blue) radiation (retinal injury).
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
Printed on 30. 07. 2001
Page 14
14 Safety

Warning labels and notes

Caution:
Observe all warning labels and notes! If any label is missing on your instrument or has become illegible, please contact us or one of our authorized representatives. We will supply the missing labels.
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
Printed on 30. 07. 2001
Page 15
Safety 15
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
Printed on 30. 07. 2001
Page 16
16 Safety
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
Printed on 30. 07. 2001
Page 17

Description

Description 17
OPMI® PRO magis 18
Intended use 18 Controls, displays, connections 20 Tiltable binocular tube and eyepieces 26
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
Printed on 30. 07. 2001
Page 18
18 Description

OPMI® PRO magis

Intended use

The OPMI® PRO magis surgical microscope is a compact and easy-to­use instrument which is ideally suited for neurosurgery, ENT applications and reconstructive and plastic surgery. It is also suitable for the use in other microsurgical disciplines.
The friction of the front-to-back and lateral tilt motion can be adjusted as required. The tilt and rotary axes of the OPMI
®
PRO magis surgical micro­scope are equipped with magnetic brakes permitting the microscope to be locked in any position.
The zoom and focusing functions of the OPMI
®
PRO magis surgical microscope are motorized, and can be directly activated on the micros­cope. A foot control panel is not required for operation of the instrument. However, you can also control the motorized functions via a foot control panel or hand control panel. The magnetic brakes can be operated from the microscope direct.
The coaxial illumination system is supplied with light via an S light guide. The diameter of the illuminated field can be continuously adjusted from a fully illuminated field ( 55 mm) to ”spot” illumination ( 11 mm) using a light stop lever. This data applies to an objective focal length of f = 200 mm. With the ”spot” setting, luminance is approx. twice as high as in the largest possible illuminated field.
You can equip the OPMI
®
PRO magis surgical microscope with all acces­sories required for documentation and coobservation. These accessories can be mounted quickly and easily.
For adjustment of the depth of field, we recommend using a double iris di­aphragm. Cat. No. 30 33 54-0000.
Caution!
If the xenon light source is used, the unit must not be used in ophthalmic applications. Severe injury to the patient's eye is possible.
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
Printed on 30. 07. 2001
Page 19
Description 19
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
Printed on 30. 07. 2001
Page 20
20 Description

Controls, displays, connections

1 Cable and light guide clip
2 Securing screw
3 Microscope shaft
4 Axis of microscope rotation
A magnetic brake clamps the axis of rotation as soon as you let go of the release keys for the magnetic brakes.
5 Adjusting knob
for setting the friction of the microscope's axis of rotation.
6 Knob for balance setting of the front-to-back tilt motion
Use this knob to balance the front-to-back tilt motion. The balancing procedure is described in the chapter: "Operation/Balancing the sur­gical microscope".
7 Adjusting knob
for setting the friction of the microscope's front-to-back tilt motion.
8 Knob for balance setting of the lateral tilt motion
Use this knob to balance the lateral tilt motion. The balancing proce­dure is described in the chapter: "Operation/Balancing the surgical microscope".
9 Adjusting knob
for setting the friction of the microscope's lateral tilt motion.
Warning!
Before removing the microscope or changing any microscope equip­ment, make sure to tighten the knobs (7) and (9).
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
Printed on 30. 07. 2001
Page 21
Description 21
1
2
3
4
5
6
7
8
9
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
Printed on 30. 07. 2001
Page 22
22 Description
10 Dust cover
11 Securing screw
Tighten the securing screw firmly after you have inserted the binocular tube or another module in the mount on the surgical microscope.
12 Display of magnification factor
γ of the zoom system
13 Release keys for the magnetic brakes
As long as you press one of these keys, the magnetic brakes of the surgical microscope and suspension system are released (if the sus­pension system used is equipped with magnetic brakes).
14 Focusing control rings
(on the right- and left-hand sides of the housing).
15 Magnification control rings (zoom)
(on the right- and left-hand sides of the housing).
16 Handgrips for maneuvering the surgical microscope
You can attach the handgrips to the surgical microscope at different angles.
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
Printed on 30. 07. 2001
Page 23
Description 23
11
12
13
141015 16
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
Printed on 30. 07. 2001
Page 24
24 Description
17 Cable and light guide clip
18 Instrument connector
for the power supply of the microscope.
:
Note Switch off the main power switch at the suspension system before plugging in or unplugging the instrument connector.
19 Socket for the S light guide
Insert the S light guide into the socket until it snaps in.
20 Light stop lever
The diameter of the illuminated field can be continuously adjusted from a fully illuminated field ( 55 mm) to ”spot” illumination ( 11 mm) using the light stop lever. This data applies to an objective focal length of f = 200 mm. With the ”spot” setting, luminance is approx. twice as high as in the largest possible illuminated field.
21 Reset button
Press this button to set the microscope to the starting position of the focusing range.
22 Zoom knob
Use this knob for manual setting of the magnification.
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
Printed on 30. 07. 2001
Page 25
17
Description 25
18
19 21 2220
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
Printed on 30. 07. 2001
Page 26
26 Description

Tiltable binocular tube and eyepieces

180° tiltable tube
1 PD adjustment knob
The correct position has been set when the two eyepiece images merge into one.
2 180° tiltable tube
3 Eyepiece tube
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
Printed on 30. 07. 2001
Page 27
Description 27
1
2
3
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
Printed on 30. 07. 2001
Page 28
28 Description
Widefield eyepieces with magnetic coupling
Note: When you remove these eyepieces from the tube, please note that they are fitted with a magnetic coupling. When mounted, the eyepieces display a very weak magnetic field, so that the usual rules for the handling of mag­nets must only be observed with eyepieces which have not been mounted on the microscope:
Do not place the eyepieces close to instruments where there is any
risk of magnetization.
Do not place the eyepieces on sensitive electronic units such as infu-
sion pumps, cardiac pace-makers, measuring instruments or magnet­ic data carriers such as disks, audiotapes and videotapes, or credit cards.
Always store eyepieces not used in their original packaging.
1 Eyecup
Always adjust the eyecups in such a way that you can see the full field of view.
Viewing with eyeglass-
Screw in the eyecups all the way.
es:
Viewing without eye-
glasses:
Screw out the eyecups until you see the full field of view.
2 Diopter adjustment ring
The eyepieces provide ametropia compensation between -8 D and +5 D. Eyeglass wearers who perform surgery wearing their glasses set the diopter adjustment ring to 0 D. Turn the ring until you have ob­tained the optimum setting. An integrated brake holds the ring in the position set.
3 Diopter scale
for reading the prescription set.
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
Printed on 30. 07. 2001
Page 29
Description 29
1
2
3
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
Printed on 30. 07. 2001
Page 30
30 Description
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
Printed on 30. 07. 2001
Page 31
Preparations for use 31

Preparations for use

Attaching the equipment 32
Attaching the tube, eyepieces, objective lens and handgrips 32
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
Printed on 30. 07. 2001
Page 32
32 Preparations for use

Attaching the equipment

Attaching the tube, eyepieces, objective lens and handgrips

Loosen securing screw (4) by a few turns.
Remove cover (3) and store it in a safe place.
Place binocular tube (2) on the surgical microscope and tighten secu-
ring screw (4) firmly
You can install accessory equipment between the binocular tube and
the microscope body. Lock these units in position in the same way us­ing securing screw (4).
.
Insert widefield eyepieces (1) into the appropriate mounts as far as
they will go. The magnetic coupling reliably secures them in position.
Note: If a documentation device is used, one of the eyepieces can be delivered with a reticle to aid focusing. Retrofitting the reticle is only possible at the factory or by our service representative. Always use the eyepiece with the reticle on the same side of the binocular tube where the documentation device is installed.
Screw the objective lens (5) into the microscope body and tighten it
firmly
.
You can adjust the position of the handgrips (6) as required. Choose
a position which is most convenient for the surgical procedure to be performed.
Plug on the handgrips (6), making sure they snap in.
Warning! Before every use and after re-equipping the instrument, make sure that the binocular tube (2) and the objective lens (5) are securely locked in po­sition. Make sure that the securing screw (4) has been firmly
tightened!
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
Printed on 30. 07. 2001
Page 33
Preparations for use 33
1 235 64
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
Printed on 30. 07. 2001
Page 34
34 Preparations for use
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
Printed on 30. 07. 2001
Page 35

Operation

Operation 35
Preparations for use 36
Balancing the surgical microscope 36 Adjusting the surgical microscope 41 Checklist 42
Procedure 44
What to do in an emergency 45
Failure of the zoom function 45 Failure of the focusing function 46 Failure of magnetic brakes 46
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
Printed on 30. 07. 2001
Page 36
36 Operation

Preparations for use

Balancing the surgical microscope

Turn on the instrument at the power switch of the suspension system.
Note: Press one of the release keys for the magnetic brakes to set the balance of the microscope's front-to-back and lateral tilt axes.
Warning!
To permit almost effortless guidance of the surgical microscope, the sur­gical microscope and the suspension system used must be correctly bal­anced. If the system is in an extremely unbalanced state, the unit can move uncontrollably out of position. Therefore, hold the surgical micro­scope securely at its handgrips before releasing the magnetic brakes.
Balancing the front-to-back tilt motion
Hold the microscope with one hand and turn knob (2) in the "-" direc-
tion until the braking function has been deactivated.
Use knob (1) to adjust the spring force in such a way that the surgical
microscope is balanced about its front-to-back tilt axis and remains stationary in the viewing direction required.
Use knob (2) to set the friction of the front-to-back tilt motion as re-
quired.
Balancing the lateral tilt motion
Hold the microscope with one hand and turn knob (4) in the "-" direc-
tion until the braking function has been deactivated.
Use knob (3) to adjust the spring force in such a way that the surgical
microscope is balanced about its lateral tilt axis and remains stationa­ry in the viewing direction required.
Use knob (4) to set the friction of the lateral tilt motion as required.
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
Printed on 30. 07. 2001
Page 37
Operation 37
1
2
3
4
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
Printed on 30. 07. 2001
Page 38
38 Operation
Note:
Press one of the release keys for the magnetic brakes to set the balance and friction of the microscope's axis of rotation.
Adjusting the friction of the axis of rotation
Use knob (4) to set the friction of the microscope's axis of rotation as
required.
Setting the starting position of the focusing range
Press button (1) to set the surgical microscope to the starting position
of the focusing range.
Setting zoom and focus
Use the control rings for setting zoom (2) and focus (3).
Set the speed of the zoom and focusing functions on the control panel
of the suspension system.
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
Printed on 30. 07. 2001
Page 39
Operation 39
123 4
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
Printed on 30. 07. 2001
Page 40
40 Operation
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
Printed on 30. 07. 2001
Page 41
Operation 41

Adjusting the surgical microscope

Bring the surgical microscope into starting position within the focusing range. Set the minimum magnification on the surgical micro­scope. Bring the surgical microscope into working position. Adjust the interpupillary distance on the binocular tube. Set the diopter scales on the eyepieces. Please note that instrument myopia may occur.
Emmetropes set the diopter scale to 0 diopter (D).
Eyeglass wearers (who perform surgery wearing their glasses) set the diopter scale to 0 di­opter (D).
Ametropes (who know their refractive powers and perform surgery without wear­ing their glasses) set the diopter scale ring to their refractive powers.
Ametropes (who do not know their refractive powers and perform surgery without wearing their glasses):
Set both eyepieces on the main microscope to +8 D. Set both eyepieces on the assistant’s microscope to +5 D (if applicable).
Eyepieces without
Remove the binocular tube and eyepieces from the microscope body and point them at a distant ob-
*)
ject
, i.e. use them like a pair of binoculars.The image of the object is still unsharp.
Turn the diopter scale ring of one eyepiece slowly image of the object is sharp. Stop turning the ring. If necessary, repeat this procedure three times and take the average of the readings found. Adjust the second eyepiece in the same manner.
reticles:
clockwise until the
Eyepieces with
Turn the diopter scale ring on the eye­piece slowly clockwise until the reticle is imaged sharply. Stop turning the ring.
Focus the surgical microscope on the object. The reticle and the object must be imaged sharply at the same time.
reticles:
Replace the tube and eyepieces on the microscope body and firmly tighten the securing screw.
Adjust the eyecups in such a way that the full field of view can be seen. Set the maximum magnification on the microscope and focus on the object. Set the working magnification required. When the magnification is changed, the focal plane is retained, but the depth of field changes.
Note
: If several surgeons use the instrument, it is advisable to draw up a table showing the individual refractive powers of each surgeon and to keep it in a handy location near the instrument.
*) CAUTION: Never
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
point the eyepieces at the sun!
Adjust the second eyepiece in the same way until the object is seen sharply.
Printed on 30. 07. 2001
Page 42
42 Operation

Checklist

Always check the following points before surgery (without the patient!):
Make sure that there is enough room between the instrument and the accessories installed.
Turn on the instrument at the power switch of the suspension system.
Release keys for magnetic brakes
The magnetic brakes of the surgical microscope and suspension system (if the suspension system used is equipped with magnetic brakes) function properly.
Balance setting and friction
The balance of the microscope's front-to-back and lateral tilt has been correctly set.
The friction of the microscope's front-to-back and lateral tilt has been set as required.
Illumination – The light guide (1) has been connected.
Zoom and focus – Proper function has been checked using the control rings (2) and (3).
Eyepieces / binocular tube – The binocular tube (4) has been securely mounted.
The securing screw (5) has been firmly tightened.
The surgical microscope and the tube are in a position convenient for
you.
The correct interpupillary distance has been set.
The eyecups have been adjusted in such a way that you can see the
full field of view.
The correct prescription has been set on the diopter scale.
Image quality is good throughout the entire magnification range.
Warning!
If a function fails, you must not use this instrument for safety reasons. Cor­rect the fault (see the "Troubleshooting table“) or contact our service dept.
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
Printed on 30. 07. 2001
Page 43
Operation 43
34 512
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
Printed on 30. 07. 2001
Page 44
44 Operation

Procedure

Turn on the instrument at the power switch of the suspension system.
Set the required brightness of the surgical field illumination on the sus­pension system.
Check the instrument using the checklist.
Swing the surgical microscope over the surgical field into an ergonom­ic position within the working distance.
Select the lowest magnification (zoom).
For coarse focusing, look through the eyepieces and lower the surgi­cal microscope until the surgical field comes into focus.
Select the highest magnification (zoom).
Look through the eyepieces and activate the focusing function until the microscope is sharply focused on the surgical field.
Select the magnification required (zoom). Look through the eyepieces of the binocular tube. Adjust the eyepieces in such a way that you can see both the edge of the field of view and the microscope image sharply.
Adjust the handgrips in a position which is most convenient for the sur­gical procedure to be performed.
Switch off the instrument when you are not using it.
Warning!
Avoid looking directly into the light source, e.g. into the microscope ob-
jective lens or into the light guide!
When operating on the eye, always use a GG 475 protection filter to
ensure that the patient’s retina is not exposed to unnecessary (blue) radiation (retinal injury)!
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
Printed on 30. 07. 2001
Page 45
Operation 45

What to do in an emergency

Failure of the zoom function

Adjust the magnification manually using the zoom knob (1).
If the zoom drive moves into an end position, press the "MANUAL" button on the control panel of your S8/S81 suspension system to switch to the manual mode. The motorized functions of the surgical microscope are deactivated.
Continue surgery by manually operating the suspension system
and the surgical microscope.
1
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
Printed on 30. 07. 2001
Page 46
46 Operation

Failure of the focusing function

Manually focus the surgical microscope by moving the suspension
arm or lift column of the suspension system up or down.

Failure of magnetic brakes

If the magnetic brakes fail (magnetic brakes are locked), you can manu­ally position the suspension arm including the microscope by overcoming the locking effect of the magnetic brakes.
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
Printed on 30. 07. 2001
Page 47
Maintenance / Further information 47

Maintenance / Further information

Trouble-shooting table 48 Care of the unit 50 Sterilization 51 Magnifications / Fields of view 52 Ordering data 53 Spare parts 54 Accessories 55 Technical data 56 Ambient requirements 58
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
Printed on 30. 07. 2001
Page 48
48 Maintenance / Further information

Trouble-shooting table

This instrument is a high-grade technological product. To ensure opti-
mum performance and safe working order of the instrument, its safety must be checked once every 12 months. We recommend having this check performed by our service representative as part of regular main­tenance work.
If a failure occurs which you cannot correct using the trouble-shooting table, attach a sign to the instrument stating it is out of order and con­tact our service representative.
®
OPMI
Problem Possible cause Remedy See
No function at all. Suspension system
switched off.
Line power failure. Contact in-house electri-
Surgical field illumination in­operative.
Illumination system of sus­pension system switched off.
PRO magis Surgical Microscope
Switch on suspension system.
cian.
Switch on illumination system of suspension system. See user's manual for suspension system.
Defective lamp. Change the lamp. See
user's manual for suspen­sion system.
Light guide not properly in­serted in lamp or micro-
Insert light guide as far as it will go.
scope.
Lamp module on suspen­sion system has no contact.
Insert lamp module as far as it will go.
Failure of electronics. Illuminate surgical field
using an OR illuminator. Contact service dept.
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
Printed on 30. 07. 2001
Page 49
Maintenance / Further information 49
Problem Possible cause Remedy See
Insufficient surgical field illu­mination.
Brightness level set too low. Adjust brightness using the
brightness control on the il­lumination system. See user's manual for suspen­sion system.
Light guide not properly in­serted in lamp and/or micro-
Insert light guide as far as it will go.
scope.
Defective light guide (illumi-
Contact service dept.
nation not uniform).
Zoom system inoperative. Defective motor. Adjust magnification manu-
ally using the zoom knob. Contact service dept.
Zoom always moves to upper or lower end position.
Switch suspension system to "MANUAL“ (S8/S81 sus­pension system only). Adjust magnification manu­ally using the zoom knob. Contact service dept.
Focusing system inopera­tive.
Defective motor. Adjust the working distance
manually using the focusing knob. Contact service dept.
Focus always moves to upper or lower end position.
Switch suspension system to "MANUAL“ (S8/S81 sus-
pension system only). Adjust the working distance manually using the focusing knob. Contact service dept.
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
Printed on 30. 07. 2001
Page 50
50 Maintenance / Further information

Care of the unit

Cleaning optical surfaces
The multilayer T* (T-star) coating of our optical components (e.g. eye­pieces, objective lenses) results in optimum image quality. Image quality is impaired by even slight contamination of the optics or by a fingerprint. In order to protect the internal optics from dust, the instru­ment should never be left without the objective lens, binocular tube and eyepieces installed when it is not in use. Always store objective lenses, eyepieces and accessories which are not being used in dust-free cases.
The external surfaces of optical components (eyepieces, objective lenses) should only be cleaned when required:
Do not use any chemical cleaning agents.
Blow off dust on the optical surfaces using a squeeze blower or re­move the dust using a clean and grease-free brush.
Prevention of fogging
To prevent the eyepieces from fogging, you may treat them with our anti­fogging agent. This anti-fogging agent consists of a dry, chemical solid in the form of a stick and produces a dry film when applied to the eyepieces. This film protects against fogging without causing any damage to the T* coating. The anti-fogging agent contains neither wax nor silicone; it is un­poisonous and does not scratch glass. The anti-fogging agent does not only ensure fog-free optics; it cleans and protects the eyepiece optics against dirt, grease, dust, fluff and finger­prints. Every pack of anti-fogging agent (contains 1 stick) includes instructions for use which we would ask you to observe.
Important: Spread the anti-fogging agent film with a dry cotton cloth - do not use paper!
Anti-fogging agent: Cat. No. 30 55 50- 9910
Cleaning mechanical surfaces
All mechanical surfaces of the equipment can be cleaned by wiping with a moist cloth. Do not use any aggressive or abrasive cleaning agents.
Wipe off any residue with a mixture of 50% ethyl alcohol and 50% distilled water plus a dash of household dish-washing liquid.
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
Printed on 30. 07. 2001
Page 51

Sterilization

Maintenance / Further information 51
The asepsis sets supplied by Carl Zeiss contain rubber caps, sleeves and handles which you can sterilize in the autoclave. We recommend the fol­lowing program for sterilization:
Sterilization
134° C
temperature:
Sterilizing time: 20 minutes
Single, disposable drapes are available to cover the instrument.
N
ote:
Attach the drapes loosely enough to provide sufficient room for maneuver for the microscope carrier and the surgical microscope. The drapes must be particularly loose around the hand grip. The surgeon must be able to operate the controls through the drape.
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
Printed on 30. 07. 2001
Page 52
52 Maintenance / Further information

Magnifications / Fields of view

If the magnification factor γ of the zoom system is known, you can calcu- late the total magnification of the surgical microscope. The following equation is used to determine the total magnification:
f
tube
M
where:
f
the focal length of the binocular tube
tube
the focal length of the main objective lens
f
obj
γ the magnification factor of the zoom system
M
the magnification of the eyepiece
eye
-----------
T
f
obj
γ M
⋅⋅=
eye
Example:
f
= 170 mm, f
tube
= 200 mm, γ = 1.6 and M
obj
= 12.5 x.
eye
The resulting total magnification is:
170 mm
T
---------- -----------
200 mm
1.6 12.5 17.0=⋅⋅=
is known, you can
T
M
If the total magnification of the surgical microscope M use the formula:
FoVN M
FoV
to calculate the field of view diameter FoV
--------- ------------- -----------
=
D
M
eye
T
of the surgical microscope.
D
The field of view diameter is the diameter of the circular surgical field which can be seen through the eyepieces. FoVN in the above formula stands for the field of view number of the eye­piece.
By entering the value MT = 17.0 in the formula, we obtain the following field of view diameter for 12.5x eyepieces with a field of view number FoVN of 18 mm:
18 mm 12.5
FoV
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
--------- ------------- ------------
D
17.0
13.2 mm==
Printed on 30. 07. 2001
Page 53

Ordering data

Maintenance / Further information 53
OPMI® PRO magis surgical microscope
Description Cat. No.
OPMI® PRO magis surgical microscope 1025-287
180° tiltable binocular tube 30 37 91-0000
12.5x widefield eyepiece (2x) 30 55 43-0000
Objective lens f=250 mm, mount diameter 65 mm 30 26 54-0000
Dust cover 1055-278
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
Printed on 30. 07. 2001
Page 54
54 Maintenance / Further information

Spare parts

OPMI® PRO magis surgical microscope
Description Cat. No.
Dust cover 1055-278
Drape 32 60 75
Light guide, 2.5 m long, with 90° adapter 30 34 81-9225
Sterile covers for tiltable binocular tube 30 37 91-0153
Spare lamps see user's manual for S8/S81 suspen­sion system.
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
Printed on 30. 07. 2001
Page 55

Accessories

Maintenance / Further information 55
Please observe the following: Only operate the instrument with the accessories included in the delivery package. If you want to use other accessories, make sure that Carl Zeiss or the manufacturer of the accessories has proved and confirmed that these accessories meet the respective technical safety standards and can be used without risk.
®
OPMI
PRO magis surgical microscope
Description Cat. No.
X-Y coupling incl. X-Y power cable for S8/S81 1043-103
S8 coupling with shaft mount 30 59 52-8030
Foot control panel, 14 functions, 6 m cable 30 49 79-9020
Stereo bridge 1040-085
Rotatable dovetail for binocular tubes 1030-023
180° tiltable binocular tube 30 37 91-0000
45 ° inclined tube 30 37 84-0000
Straight tube 30 37 65-0000
Sleeves for push-in eyepieces on binocular tubes
30 55 42-0107
with screw threads (2x)
10x push-in widefield eyepiece (2x) 30 55 42-0000
12.5x push-in widefield eyepiece (2x) 30 55 43-0000
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
Printed on 30. 07. 2001
Page 56
56 Maintenance / Further information

Technical data

Power supply The surgical microscope is powered via the sus-
pension system.
Tube / Eyepieces Widefield tube, f = 170 mm.
12.5x/18/B widefield screw-type eyepieces with integrated eyecups. This equipment provides full illumination of the field of view (diameter 55 mm).
Objective lenses f = 200 mm objective lens, mount diameter
65 mm. Options: objective lenses with different focal lengths from f = 200 mm to f = 400 mm.
Magnification Zoom system with 1:6 ratio, magnification factor γ
= 0.4x - 2.4x. Motorized or manual magnification change on microscope.
Focusing Motorized, focusing range 18 mm.
At the press of a button, the focusing drive moves to the starting position of the focusing range.
Illumination Coaxial fiber optic illumination,
supplied via S light guide from 12 V 100 W halo­gen reflector lamp in suspension system. Manually adjustable illuminated field (dia. 55 mm) to ”spot” illumination (dia. 11 mm). This data ap­plies to an objective focal length of f = 200 mm. In­creased illumination intensity in a small illuminated field. With the ”spot” setting, lumi­nance is approx. twice as high as in full-field illu­mination.
Microscope rotation Locked using magnetic brake.
Friction manually adjustable until rotation is locked. Range of rotation 310°.
Front-to-back tilt of microscope
Locked using magnetic brake. With torque com­pensation, manually adjustable using a knob. Friction manually adjustable until the tilt motion is locked. Tilt angle after balance setting for viewing direc­tion ± 20°.
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
Printed on 30. 07. 2001
Page 57
Maintenance / Further information 57
Lateral tilt of micro­scope
Locked using magnetic brake. With torque compensation, manually adjustable using a knob. Friction manually adjustable until the tilt motion is locked. Tilt angle after balance setting for position ± 20°.
Cat. No. OPMI® PRO magis surgical microscope
30 26 10- 0000
Weight of surgical
7.3 kg
microscope
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
Printed on 30. 07. 2001
Page 58
58 Maintenance / Further information

Ambient requirements

For operation Temperature
Rel. humidity Air pressure
For transportation and storage
The unit meets the essential requirements stipulated in Annex I to the 93/42/EEC Directive governing medical de­vices. The unit is marked with:
Subject to change.
Temperature Rel. humidity (without condensa­tion) Air pressure
+10 °C...+40 °C 30%...75% 700 hPa...1,060 hPa
- 40 °C...+70 °C 10%...100%
500 hPa...1,060 hPa
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
Printed on 30. 07. 2001
Page 59

Index

Index 59
Numerics
180° tiltable tube
A
Accessories 55 Accident prevention regulations 10 Adjusting the friction of axis of rotation 38 Ambient requirements 58 Anti-fogging agent 50
B
Balancing front-to-back tilt motion 36 Balancing lateral tilt motion 36 Balancing the surgical microscope 36
26
C
Care of the unit CE marking 58 Checklist 42 Cleaning optical surfaces 50 Controls, displays, connections 20
D
Description Diopter scale 28 Directive 93/42/EEC 58 Directives 10
E
Emergency backup line supply 10 Eyecup 28 Eyepieces, attachment 32
F
Failure of a function Failure of magnetic brakes 46 Failure of the focusing function 46 Failure of the zoom function 45 Fields of view 52 Focusing control rings 22 Focusing function, failure 46 Focusing range 24 Fogging, prevention 50
50
17
42
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
Printed on 30. 07. 2001
Page 60
60 Index
G
GG 475 protection filter 13
H
Handgrips 22
I
Installation, notes 11 Instrument class 10 Instrument connector 24
K
Key for symbols 7
L
Light stop lever
M
Magnetic brakes Magnetic brakes, failure of 46 Magnetic brakes, release 22 Magnification control rings 22 Magnification factor, display 22 Magnifications 52 Maintenance 47 Mechanical surfaces, cleaning 50 Mobile phones 11
24
22
N
Notes on safety 10 Notes on use 11
O
Operation 35 Operation, requirements 12 Ordering data 53
P
PD adjustment knob 26 Possible cause 48 Potential equalization 11 Preparations for use 31 Prevention of fogging 50 Problem 48 Procedure 44 Protective earth contact 12
Q
Quality management system 10
R
Re-equipping the instrument 12
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
Printed on 30. 07. 2001
Page 61
Index 61
Remedy 48 Requirements for operation 12
S
Safety 9, 10 Safety check 12, 48 Safety standards 10 Setting 38 Setting starting position of focusing range 38 Socket for S light guide 24 Spare parts 54 Standards 10 Starting position of focusing range 24 Sterilization 51
T
Technical data 56 Tiltable binocular tube and eyepieces 26 Trouble-shooting table 48 Tube and eyepieces for the main microscope 26 Tube, attachment 32
U
Use of the instrument, before 12 Use of the instrument, for every 13 Use, notes 11
W
What to do in an emergency 45 Widefield eyepieces with magnetic coupling 28
Z
Zoom and focus setting 38 Zoom control rings 22 Zoom function, failure 45 Zoom knob 24
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
Printed on 30. 07. 2001
Page 62
62 Index
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
Printed on 30. 07. 2001
Page 63
63
G-30-1292-en OPMI® PRO magis Surgical Microscope Issue 3.0
Printed on 30. 07. 2001
Page 64
Carl Zeiss
Surgical Products Division 73446 Oberkochen Germany Fax: +49 (0) 7364/20-4823 email: surgical@zeiss.de www.zeiss.de
G-30-1292-en Printed in Germany MS-TD
Loading...