YSI 9300 User Manual

5 (1)

®

YSI 9300 and 9500

Photometers

User Manual

®

YSI 9300 and 9500

Direct-Read Photometers

User Manual

YSI, Inc.

1725 Brannum Lane Yellow Springs, OH 45387

Tel: 800-897-4151 (+1 937-767-7241) Fax: +1 937-767-1058 E-Mail: environmental@ysi.com

 

 

Y- PT 282

 

INDEX

 

1

INTRODUCTION ..............................................................................................................

1

 

Features and Technical Specification ................................................................................

3

2

OPERATING PRINCIPLE ...................................................................................................

4

 

Powering the Photometer ..................................................................................................

4

 

Replacing the Batteries......................................................................................................

5

 

Power Supply (Model 9500 only) .......................................................................................

5

3

GENERAL PHOTOMETER OPERATION ...............................................................................

6

 

Operating Modes ...............................................................................................................

6

 

System - Quick Start..........................................................................................................

6

 

System - Full Options ........................................................................................................

7

 

View Log (9500 Only) ........................................................................................................

7

 

Back Light..........................................................................................................................

7

 

Language Options .............................................................................................................

7

 

Units..................................................................................................................................

8

 

Sample Number (9500 Only) .............................................................................................

8

 

Sample Number Increment (9500 Only).............................................................................

8

 

Dilution Factor (9500 Only) ................................................................................................

8

 

Date and Time (9500 Only)................................................................................................

8

 

Date Format (9500 Only) ...................................................................................................

8

 

Battery Level .....................................................................................................................

8

 

Locking System Mode Settings (9500 Only) ......................................................................

9

 

Rounding (9500 Only)........................................................................................................

9

 

Time-Out (9500 Only)........................................................................................................

9

 

Edit User Defined Tests (9500 Only)..................................................................................

9

 

USB (9500 Only) .............................................................................................................

10

4 INTERFACE CONNECTIONS AND DATA MEMORY (9500 ONLY) .........................................

11

5

TAKING PHOTOMETER READINGS...................................................................................

12

 

Program Numbers and Test Instructions..........................................................................

12

 

Sample Dilution (9500 Only) ............................................................................................

12

 

Blank and Sample Tubes.................................................................................................

12

 

Light Cap.........................................................................................................................

13

 

Getting the Best Results ..................................................................................................

13

 

Taking Test Readings......................................................................................................

13

 

Continuation Tests (Certain Tests Only) ..........................................................................

16

 

Favorite Tests List ...........................................................................................................

16

 

Expressing Different Chemical Forms..............................................................................

16

 

Reading in Transmittance and Absorbance .....................................................................

17

 

Timer...............................................................................................................................

17

 

User Defined Tests (9500 Only).......................................................................................

17

6

CARE AND MAINTENANCE ..............................................................................................

20

 

Cleaning the Optics .........................................................................................................

20

 

Service Requirement .......................................................................................................

20

 

Error Messages ...............................................................................................................

20

 

Photometer Up-Grade (9500 only)...................................................................................

21

 

Computer Controlled Operation (9500 Only)....................................................................

21

 

Warranty..........................................................................................................................

21

 

Limitation of Warranty......................................................................................................

22

 

Contact Information .........................................................................................................

22

1 INTRODUCTION

The YSI 9300 and 9500 direct-read photometers are designed to give long and trouble-free operation. To ensure the best results, please read this manual carefully and follow the procedures recommended. This manual covers both the 9300 and 9500 photometers. Therefore, some of the information only pertains to the 9500 as is noted in the appropriate sections.

The Photometers feature digital electronics and built-in filters. It is lightweight and portable for field or laboratory use. The instruments are rugged, durable and IP-67 rated. Additionally, the photometers are direct-reading, have automatic blank setting, automatic wavelength selection, and automatic power cut-off.

The following pages describe the use of the photometers, and give instructions for the wide range of water tests which can be performed using these instruments.

Keep the photometer clean and in good working order by adhering to the following recommendations:

Do not pour out samples or prepare the tests directly over the instrument.

Always cap the test tubes before inserting into the instrument for readings.

Wipe test tubes with a clean tissue to remove drips or condensation before placing in the photometer.

Do not leave tubes standing in the photometer test chamber. Remove the tubes immediately after each test.

Immediately wipe up any drips or spills on the instrument or in the test chamber with a clean tissue.

Keep the instrument clean. Clean the test chamber regularly using a moistened tissue or cotton ball.

Keep the instrument away from all chemicals and cleaning materials.

Keep the instrument in a clean, dry place when it is not in use. Keep it on a clean, dry bench away from chemicals, place it in a storage cupboard or keep it in a carrying case.

Keep the carrying case in a clean, dry condition. Make sure that the carrying case is dry before the case is closed up and the instrument is put away.

1

Instrument Layout

USB Interface (9500 only)

Multi-Size Cell

Graphic Display

On/Off Key

Scroll and Selection Keys

OK (Enter) Key

Test and

Number Keys

2

Features and Technical Specification

Application

For application in general water testing using Palintest

 

tablet reagent systems and Palintest Tubetests reagents.

Instrument Type

Single-beam colorimeter with built-in colour filters and pre-

 

programmed test calibrations.

Peak Wavelengths

445 5nm, 495 5nm, 555 5nm, 570 5nm, 605

 

5nm and 655 5nm

Range

1 - 100%T

Accuracy

1%T

Display

Large graphic display with option of backlight.

Language

Test identification and prompts in English, French,

 

German, Spanish and Italian.

Timer

Clock and timer feature to log test results and audible

 

alarm for timing test procedure.

Units

Direct-reading of test results in mg/l, ppm, g/l or molar

 

units (mmol/l or µmol/l).

User Selectable Options

10 digit sample number entry, dilution factor, time/date,

 

date format, system lock and rounding of results.

Date Format

Date format selectable as day/month/year or month/day/

 

year.

Zeroing

Automatic zeroing on blank tube and hold blank facility for

 

series of tests. Continuation test facility without the need

 

for reblanking.

Internal Memory

Stores 500 previous readings with option to view logged

 

results on screen, or download to computer.

USB Interface

USB 1.1 full-speed, bus-powered device. Software selectable

 

between either emulation of a removable hard-drive or

 

emulation of a serial device connected via a virtual COM port.

Power

3 x 1.5V ‘AA’ alkaline batteries or via USB interface. Power

 

management system with variable length auto switch-off

 

or 'continuous' operation.

Size

250 x 150 x 70 mm

Weight

985 g (2.1 lbs)

Test Tubes

For tablet reagents - 10 ml glass test tubes, 20 mm OD

 

(YPT 595).

Cell Holder

Multi-size tube holder accepts test tubes from 12 –

 

20 mm OD and centres the tube for optimum optical

 

performance.

 

3

2 OPERATING PRINCIPLE

The YSI photometers are instruments that measure color intensity. Light is passed through a test tube containing the sample solution, and then through a colored filter onto a photodetector. Filters have been chosen so that light of a specific wavelength is selected. When the solution is completely colorless, all of the light passes through the sample. With colored samples, light is absorbed and the light which passes through the sample is proportionately reduced.

In the following test procedures, the photometer is used to measure the color which is produced when chemical reagents are reacted with the water sample. In these tests, the color intensity produced is proportional to the concentration of the parameter being tested.

The photometer is pre-programmed with calibrations for each test parameter. Different test procedures are carried out at different wavelengths to optimize the sensitivity of each test. The required wavelength is selected automatically by the instrument.

The calibrations are accessed by entering a unique program number at the start of each test procedure. This enables the instrument to select the appropriate wavelength filter automatically and allows the photodiode response to be converted to a concentration reading. The instrument thus displays a directreading of the test result.

The photometer is ideally suited for general analytical applications. The instrument can be used as a laboratory or field photometer with user-generated calibration graphs for standard analytical methods or for comparison of colored solutions.

For general analytical applications, Transmittance (test program 0), or Absorbance (test program 1) can be chosen.

Powering the Photometer

The Photometer is powered by (3) AAA batteries. The photometer features a battery indicator – see ‘System Mode’ functions. A minimum voltage of 3.0V is needed to operate the photometer. As a power-saving measure, in normal use, both the 9300 and 9500 photometers automatically switch off five minutes after the last key is pressed. The switch off period may be adjusted for the 9500 in the System mode.

In addition to the above feature, a battery-warning message will appear automatically on the display when the battery voltage becomes low. The batteries should be replaced as soon as possible after the warning message appears. Stored data in the instrument memory will not be lost during battery replacement.

4

Replacing the Batteries

The battery compartment in the base of the instrument is secured by four screws. To replace the batteries, remove the cover and install the batteries, observing the correct polarity as indicated. Use 3 x 1.5V 'AA’ alkaline batteries or equivalent. To avoid corrosion damage through leakage, remove batteries from the instrument if it is to be stored or left unused for a long period of time (> 30 days).

Power Supply (Model 9500 only)

The 9500 photometer can be powered either from alkaline batteries or via the USB socket. To use mains power, the instrument is connected using the USB Connection Cable (YPT 284) plugged to the Mains Adapter (YPT283). Alternatively, the USB connection cable can be plugged into a computer to power the 9500 from the computer.

5

3 GENERAL PHOTOMETER OPERATION

The photometer is controlled by a simple intuitive menu system:

The highlight indicates the active line or section of the screen

The and keys move the highlight through the menu choices

The and keys allow selection of options

The flashing cursor in the ‘Options’ menu at the bottom of the screen indicates the action which will occur if the [OK] button is pressed.

Operating Modes

The photometer has two distinct operating modes - the PHOTOMETER mode and the SYSTEM mode.

The PHOTOMETER mode is the normal operating mode for taking photometer readings. This mode is engaged automatically when the instrument is turned on by pressing the key.

As a power-saving measure, in normal use, both the 9300 and 9500 photometers automatically switch off five minutes after the last key is pressed. This may be adjusted for the 9500 in the System mode.

The SYSTEM mode is used to set the system options. This mode is engaged when the photometer is turned on using the key and then selecting 'System' using the and keys and pressing [OK].

Scroll through the menu box to view all the options available.

System - Quick Start

When the instrument is first used, the SYSTEM mode should be used to set the preferred operating options:

Use the and keys to scroll through the features.

Use the and keys to select the options.

Press [OK] to accept the selections and return to PHOTOMETER mode.

Select the desired language from English, French, German, Spanish or Italian.

Select the desired display units from mg/l, ppm, mmol/l, µmol/l and g/l.

Set the sample number option to ‘On’ to allow the entry of a sample number during normal photometer operation (model 9500 only).

Set the sample increment option to ‘On’ to automatically increase the sample number (model 9500 only).

6

Set the dilution factor to ‘On’ or ‘Off’. If the dilution factor option is set to ‘On’, the instrument will allow the entry of a numerical factor which will be used in the calculation of the result displayed on the instrument (model 9500 only).

Select the preferred date format. The date may be shown in either Day/ Month/Year or Month/Day/Year (9500 only).

To change the date and time, select the date and time line then key in

correct setting using the numeric keys. To correct an error, use the keys to move the cursor then key in the correct data (9500 only).

System - Full Options

The Photometer features a wide range of options which may be explored at leisure to get the best results from the instrument. An explanation of the application of these options is as follows:

View Log (9500 Only)

The 9500 photometer has an internal memory which can hold up to 500 test results. Once the memory is full, each new result overwrites the oldest entry.

Select ‘View Log’ to view stored results on screen. The used to scroll through the list of stored results. The ‘Options’ menu offers several choices.

Select ‘Clear’ to empty the memory. Confirmation is requested to avoid accidentally erasing the data. Select ‘Exit’ to return to SYSTEM mode. Select ‘Download’ to transmit stored data to a PC. This option only appears if the USB mode is set to ‘COM Port’. Refer to ‘Interface Connection and Data Memory’ for further information.

Back Light

The graphic display features a backlight to enhance the display contrast. This may be switched off to conserve power when working on battery power.

Language Options

The photometer can be operated in a number of different languages. When a particular language is selected, the test names and operating commands will appear in that language. Certain tests and unit options are provided in accordance with the conventions of particular countries and are only available when the photometer is switched to the language that particular country.

7

Units

The photometer offers the choice of result expressed in mg/l, ppm, mmol/l, µmol and g/l.

Sample Number (9500 Only)

A unique number may be associated with each result record to identify it in the log. If Sample Number ‘On’ is selected, the user is offered the choice of entering a number of up to 10 digits for each sample reading. If this function is set to ‘Off’, a sample number is automatically allocated.

Sample Number Increment (9500 Only)

The sample number increment option may be used to determine whether the instrument does or does not automatically increment the sample number after each test. Incrementation of the sample number may be used when the instrument is used for carrying out a series of similar tests. Alternatively it may be preferable not to increment the number if typical use involves carrying out a number of different tests on the same sample.

Dilution Factor (9500 Only)

When samples are out of range for the test, a dilution procedure may be used. If the dilution factor option is set to ‘On’, the instrument will allow entry of a numerical factor which will be used in the calculation of the result displayed and stored in the log.

Date and Time (9500 Only)

The instrument records the date and time of each reading taken and associates this with the data record in the log. To correct the date and time on the internal clock, select the date and time display line.

Date Format (9500 Only)

The option of day/month/year or month/day/year date format is available.

Battery Level

A battery level indicator shows the power available. At least 3.0V is required for successful operation of the instrument.

8

Locking System Mode Settings (9500 Only)

It is possible to 'lock’ the system settings so that these cannot be tampered with or altered accidentally during use. This is important, for example, where it is necessary to verify that tests have actually been carried out at a particular time or date, or where procedures always require the use of a sample number or dilution factor.

The instructions for locking the settings are not included in this manual; these are provided to photometer owners or system administrators on formal request to YSI’s Technical Support department (environmental@ysi.com).

Rounding (9500 Only)

In the normal default setting, the photometer will round test results appropriately for the resolution of the test. The rounding applied differs for each parameter depending on the shape of the calibration curve. This ensures the optimum precision and accuracy of each test procedure. For normal purposes it is strongly recommended that the instrument be left in the default setting.

However, for certain analytical applications, it may be useful to switch off the rounding to display the result in unrounded form. This may be the case, for example, when carrying out statistical evaluations of test methods where it is necessary to use the data in calculation of standard deviation or distribution data.

Time-Out (9500 Only)

As a power-saving measure, in normal use, both the 9300 and 9500 photometers automatically switch off five minutes after the last key is pressed.

The 9500 photometer may be switched to ‘Long’ time-out which allows 15 minutes before shut-down or ‘Off’ which allows continuous use. This is particularly useful when powering the instrument through the USB interface.

Edit User Defined Tests (9500 Only)

Users may wish to develop their own test methods and store calibrations on the photometer. The 9500 has the facility to store up to 30 user-defined calibrations. See ‘User Defined Tests’ below for full instructions.

9

USB (9500 Only)

The USB interface allows communication between the instrument and a PC. There is a choice of two operating modes – Hard Drive and COM Port.

In Hard Drive mode, the instrument appears as a removable hard drive when connected to a PC. No additional software is required on computers running Windows 2000, ME or XP. A driver to use this option with Windows 98SE is available from YSI Technical Support Department (environmental@ysi.com).

In COM Port mode, the instrument behaves as if connected to the PC serial port via RS232. In this mode, the PC requires installation of a USB virtual COM Port driver, available from YSI Technical Support Department (environmental@ysi.com).

See the section on ‘Interface Connections and Data Memory’ below for full instructions.

10

4 INTERFACE CONNECTIONS AND DATA MEMORY (9500 ONLY)

Stored data can be accessed by recall to the instrument display (see ‘View Log’). Alternatively, data can be accessed using a PC:

Connect the instrument to the computer via the USB port, using any suitable USB cable, ie YPT 746

Turn the instrument ON and select SYSTEM mode from the ‘Options’ menu

Scroll to ‘USB’ and select either ‘Hard Drive’ or ‘COM Port’.

‘Hard Drive’ – Once this option is selected, simply turning the instrument ON while it is connected to a PC will cause an extra hard drive containing the instrument files to appear on the PC. The log of test results is in text file:

‘9500_LOG.txt’. The other files shown on screen contain calibration and operating systems for use when upgrading the instrument and should not be accessed.

The log file can be copied from the instrument by dragging between windows. Once copied, the file can be opened with many text editors, word processors or spreadsheet programs.

Note: Deleting this file from the instrument’s hard drive will clear the data from the instrument memory.

COM Port’ – Once this option is selected, data can be downloaded from the instrument to the PC:

Open the ‘Virtual COM Port – HyperTerminal’ window on the computer

In the instrument SYSTEM mode, scroll to ‘View Log’ and select ‘Download’.

The data from the log will appear on the PC screen and can be transferred to other PC applications or printed.

‘Unplugged’ – Note that the ‘Hard Drive’ or ‘Com Port’ may only be selected while the instrument is being powered via its USB port. If the instrument is running on batteries and is not connected to either a PC or a YPT783 external power supply, an ‘Unplugged’ message will be displayed instead of either ‘Hard Drive’ or ‘COM Port’.

11

5 TAKING PHOTOMETER READINGS

The photometer is very simple to use. Screen prompts guide the user towards the test result. The following sections describe how to get the best results from the instrument.

Program Numbers and Test Instructions

Each test is identified by a separate program number or named key. Program numbers are shown in the test instruction sheets supplied in this manual. For some tests, a choice of different programs is offered in order to get the result in different forms (for example, for Nitrate - NO3 or Nitrate Nitrogen - NO3-N).

In certain tests, such as free chlorine and total chlorine, the test can be continued to a further stage. This is allowed for in the programming of the photometer. In these tests, once the result of the first stage is obtained, the

‘Follow-On’ option may be selected to progress the test to the next test stage or stages. The result will be calculated automatically.

These continuation programs have their own program number for reference purposes although direct access to these programs may be restricted.

Sample Dilution (9500 Only)

The photometer has a sample dilution option. This enables a factor to be entered when samples have been diluted to bring them within the measuring range of the test. For example, if a five times dilution of a sample has been made, then a dilution factor of x5 should be entered. The photometer will multiply the observed result by this factor so that the correct result for the original sample is displayed.

This option may be used in conjunction with the YSI Dilution Tube (YPT 512) which enables dilutions of x2, x3, x4, x5 and x10 to be made. Higher dilution factors may be entered but are subject to the limitation of the number of digits available of the result display for each test. When the display capabilities are exceeded, the symbol [xxx] will appear on the result display. The sample should not be diluted prior to carrying out a pH test, or a Transmittance or Absorbance reading.

Blank and Sample Tubes

A BLANK TUBE is needed each time the photometer is used. This enables the instrument to be set automatically and compensates for any inherent color in the test sample. It is important therefore to understand the meaning of the term ‘BLANK TUBE'.

12

The BLANK TUBE is a test tube filled only with the water being tested only. It is important to use the actual water to be tested to provide a true comparison for the test results.

The term 'SAMPLE TUBE' is used to describe the tube containing the water sample to which the reagents have been added in accordance with the appropriate test instructions. This tube is used to take the photometer reading.

Light Cap

A light cap is provided with the photometer. This cap fits over the test chamber and prevents stray light reaching the photodiode.

It is NOT necessary to use the light cap when using the photometer indoors or under shaded outdoor light. The light cap should be used when working outside in strong sunlight. The light cap is also recommended when carrying out turbidity-based tests such as the cyanuric acid test, under bright or variable lighting conditions. Test instructions indicate when the light cap should be used.

Getting the Best Results

Success in obtaining accurate and consistent test results will depend on the care with which test procedures are carried out. Always follow the test instructions carefully and observe the stated standing periods and temperature conditions where applicable.

Wipe test tubes free from condensation before placing in the photometer. Test tubes should always be kept in a clean condition. Wash and dry the test tubes carefully after use. Dirty tubes may be soaked in a mild detergent solution if necessary. Tubes which become stained or scratched should be discarded and replaced.

Taking Test Readings

Press key. The instrument displays the ‘Choose a Test’ menu box, with the last test program used highlighted as the active line.

The cursor will flash on the [OK] symbol of the ‘options menu' at the bottom of the screen. Press [OK] to accept this program.

To choose a different test program, either use the and keys to scroll through the menu options, or use the numeric keys to enter the Phot number of the desired test. The four most recently used tests are listed at the top of the ‘Choose a Test’ screen for convenience. Press [OK] to accept the selected program.

13

YSI 9300 User Manual

If the sample number option is pre-selected, then the following display will appear, for example (9500 only):

Sam ple Num ber

1024

O K

Enter or confirm the sample number (up to 10 digits), then press [OK].

1If the dilution factor option is pre-selected, then the following display will appear (9500 only):

Dilution Factor

x 1

O K

Press [OK] to accept the default value (x1, no dilution), or key in new dilution factor then press [OK].

2 The following display will now appear:

P h o t 007

C hlorine-Free / 5

Insert B lank

OK

Choos e a Te st

 

Place a BLANK TUBE in the test chamber, then press [OK].

NOTE: The instrument is designed to hold the blank setting as long as the instrument is switched on. This stage will be omitted when further tests are being carried out. However, when changing to a test which requires a colored or reagent blank, or uses a tube of a different diameter, a new blank reading is required. The ‘Insert Blank’ prompt will be displayed automatically.

If the instrument is in used continuously, it is advisable to re-blank from time to time.

14

3The instrument will be set automatically. After a few seconds the following display will appear:

P h o t 007

C hlorine-Free / 5

Insert S am ple

OK

Bla nk - Choose a Te st - Tim e

 

Place SAMPLE TUBE in the test chamber, then press [OK].

4The instrument will take the reading and display the result as follows, for example:

Phot 007

Chlorine-Free / 5

1.00

mg/l Cl2

Choose a Test

Read Blank Follow-On

The following symbols indicate the result is out of test range: Result is higher than range: >>

Result is lower than range: <<

5The ‘options menu' offers the choice to:

‘Choose a Test’ - return to the menu of test programs and select another test

‘Read’

-

read further sample tubes of the currently selected test

‘Blank’

-

re-blank the instrument

‘Follow-On’

-

carry out a continuation test if available.

15

Continuation Tests (Certain Tests Only)

1Select ‘Follow-On’ and press [OK] while the result is displayed of the currently running test. The photometer applies the previously entered sample number and dilution factor, and the ‘Insert Sample’ screen will appear.

Place SAMPLE tube in the test chamber, then press [OK].

2The instrument will take the reading and calculate the result from the combination of readings (where appropriate). The result will be displayed as follows, for example:

Phot 008

Chlorine-Total / 5

1.50

mg/l Cl2

Choose a Test Read Blank

Return

3While the test result is displayed, similar options are available as at the end of a normal test program. In order to run more samples for the same parameters, select ‘Return’ from the ‘options menu' to take the program back to the start of the first stage of a multiple test procedure.

Note: some continuation test procedures involve a standing period. The photometer may switch off automatically during this time. To avoid the instrument switching off, set for continuous operation or use the timer function to time any standing period. See Timer section. The timer will over-ride the auto switch off function.

Favorite Tests List

The four most recently used tests are listed at the top of the 'Choose a Test' screen for convenience.

Expressing Different Chemical Forms

If the test result can be expressed in different chemical forms, the chemical symbol will have flashing and to indicate this. Use the and keys to step through the options available.

Note that the log stores the result in the primary form.

16

Reading in Transmittance and Absorbance

When taking readings in Transmittance or Absorbance mode, use the and keys to step through the wavelengths until the required wavelength is reached.

Timer

The photometer features a countdown timer with alarm as an aid to carrying out test procedures. The timer can be accessed at any time by selecting

‘Timer’ from the ‘Options’ menu.

The following display will appear:

T im er

00:00

EXIT

Key in the time required in minutes and seconds using the numerical keys, then press [OK] to start the timer. The maximum time is 29 minutes and 59 seconds. Use the and keys to reposition the cursor and re-enter the time if it is keyed in incorrectly.

The timer will count down, giving an audible alarm at the end of the timed period. Press [OK] to stop the alarm.

During the timer countdown period, an ‘Options’ menu is available :

Stop

- used to abort the timing operation or to stop the alarm at

 

the end of the timed period

Exit

- used to return to the program screen to take readings. The

 

timer will continue to run and give an audible alarm at the

 

end of the period.

Exit and Read

- used to return to the program screen with the timer counting

 

down on screen. The instrument will automatically take a

 

reading at the end of the timed period and no alarm will

 

sound.

User Defined Tests (9500 Only)

Users may wish to develop their own test methods and store the calibration data on the 9500. This will allow direct reading of user tests. The 9500 has the facility to store up to 30 user-defined calibrations.

17

To program user-defined calibrations:

Turn the instrument ‘ON’, select ‘System’ menu and press [OK]. Scroll through the options to the USB entry and make sure the option is set to ‘Com Port’.

At the PC, open the HyperTerminal connection for the 9500 (contact YSI Technical Support Department to receive the virtual comport drivers for installation to PC).

At the instrument, in the ‘System’ menu, select 'Edit User Defined Tests' and press [OK]. The instrument will display the tests already downloaded, or show

‘LIST EMPTY’.

In the ‘Options’ menu, select [Add] to add a new test, or [Edit] to edit the test which is currently highlighted. Change the highlighted test with Press [OK].

The instrument displays a message box instructing the user to download the new or edited test file.

At the PC, download the calibration file from HyperTerminal using 'Transfer', 'Send Text File' and select the file to be downloaded.

The instrument will check the downloaded data. If it is acceptable, it will display a message box ‘Accepted’ over the downloaded data. If there are errors in the file, a list of errors will be displayed. The user should edit the calibration file to correct the errors then re-send it.

Press [OK] to accept the test. The instrument will change to the ‘User Test List’ screen (See 3), with a summary of the test displayed. Press [OK] to accept the test and write to memory. Select [Cancel] at any time to reject the calibration.

The format of the user calibration file is as follows:-

The file must start with ‘USER CALIBRATION’ and end with ‘END’.

Test Number must be between 900-929 (30 tests).

Test name, up to 18 characters.

Units. Must be one of the following - mg/l, ppm, mmol/l, µmol/l or g/l.

Wavelength. Must be one of the following – 450 nm, 500 nm, 550 nm, 570 nm, 600 nm or 650 nm.

Chemical symbols: up to 8 characters.

Data pairs. Up to 10 pairs of data in the form :-

ABSORBANCE x.xxx, CONCENTRATION (Concentration may be up to five digits).

18

An example is shown below:

USER CALIBRATION 900

Chlorine mmol/l 500nm Cl2

0,0

0.174,0.50

0.481,1.50

0.733,2.50

0.854,3.00

1.022,4.00

1.086,4.50

1.187,5.00 END

19

6 CARE AND MAINTENANCE

The photometer is designed to give long and trouble-free operation. Care must be taken, however, to avoid test solutions being spilt over the instrument, and to prevent contamination of the instrument. Spillages or moisture should be wiped off immediately with a dry cloth. Never use solvents or abrasive materials to clean the instrument. Care should be taken to keep the test chamber clean.

Cleaning the Optics

Any build-up of dirt or deposits may interrupt light transmission and affect readings.

To clean the optics, undo the two screws to remove the optics base plate. Gently clean the internal surfaces of the optics with a soft, non-abrasive cloth. Deposits may be removed with a slightly dampened cotton bud. Replace the optics base plate and re-fasten the screws.

The photometer is fitted with long-life light sources and contains no userserviceable components. If the instrument requires servicing or repair, contact YSI Technical Support Department.

Service Requirement

Servicing of the photometer is essential to ensure optimum performance. To arrange service of the instrument, contact YSI Technical Support Department or the authorized distributor who supplied the instrument. The YSI standard photometer service includes cleaning of the optical assembly, replacement of any worn parts and checking/recalibration of the instrument.

Error Messages

The photometer will display an error message in the unlikely event of a malfunction. These error messages are mainly designed to assist service staff in diagnosing instrument faults. In the event an error message appears on the photometer, contact YSI Technical Support Department for advice.

Error messages are coded 7 and 9 and both relate to blanking the instrument. If you see one of these error messages, check the operating technique and sample clarity. If you continue to get error messages, try the following:

Error 7 indicates too much light – remove the instrument from bright light and use the light cap.

Error 9 indicates not enough light – follow ‘Cleaning the Optics’ routine.

20

Photometer Up-Grade (9500 only)

It is now possible to upgrade the photometer with new test calibrations using a computer. This will ensure that users can always keep the instrument up-to- date with the latest tests. Contact YSI to request an update at environmental@ysi.com. No special computer software is required. Full instructions will be supplied with the upgrade data.

Computer Controlled Operation (9500 Only)

The photometer can be controlled from a computer using suitable control software. Such software is available from software houses or from water treatment specialists to cover specific applications. These software programs typically instruct the photometer to go through a predetermined series of tests specific to that application, and then automatically receive data from the photometer and process the test results. The internal software of the photometer is able to receive computer commands to start new sample, receive test program number, receive sample number and instigate continuation test. Programmers requiring further details should contact YSI Technical Support.

Warranty

The YSI photometers are warranted for one (1) year from date of purchase by the end user against defects in materials and workmanship, exclusive of batteries and any damaged caused by defective batteries. Within the warranty period, YSI will repair or replace, at its sole discretion, free of charge, any product that YSI determines to be covered by this warranty.

To exercise this warranty, write or call your local YSI representative, or contact YSI Customer Service at +1 937-767-7241 (800-897-4151). Send the product and proof of purchase, transportation prepaid, to the Authorized Service Center selected by YSI. Repair or replacement will be made and the product returned, transportation prepaid. Repaired or replaced products are warranted for the balance of the original warranty period, or at least 90 days from date of repair or replacement.

21

Limitation of Warranty

This Warranty does not apply to any YSI product damage or failure caused by:

1Failure to install, operate or use the product in accordance with YSI's written instructions;

2Abuse or misuse of the product;

3Failure to maintain the product in accordance with YSI's written instructions or standard industry procedure;

4Any improper repairs to the product;

5Use by you of defective or improper components or parts in servicing or repairing the product;

6Modification of the product in any way not expressly authorized by YSI.

THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING ANY WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. YSI's LIABILITY UNDER THIS WARRANTY IS LIMITED TO REPAIR OR REPLACEMENT OF THE PRODUCT, AND THIS SHALL BE YOUR SOLE AND EXCLUSIVE REMEDY FOR ANY DEFECTIVE PRODUCT COVERED BY THIS WARRANTY. IN NO EVENT SHALL YSI BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES RESULTING FROM ANY DEFECTIVE PRODUCT COVERED BY THIS WARRANTY.

Contact Information

For Customer Service, Technical and Ordering Support:

Telephone: (800) 897-4151

(937) 767-7241

Monday through Friday, 8:00 AM to 5:00 PM ET

Fax:

(937)

767-9353

(orders)

 

(937)

767-1058

(technical support)

Email:

environmental@ysi.com

Mail:

YSI Incorporated

 

1725 Brannum Lane

 

Yellow Springs, OH 45387

 

USA

 

 

____________________________

22

Table of Contents

Photometer Instructions

PHOT.INST.

TABLET REAGENT SYSTEM

 

Alkalinity, Total (Alkaphot)

PHOT.2.

Alkalinity M and P (Alkaphot M/P)

PHOT.37.

Aluminum

PHOT.3.

Ammonia

PHOT.4.

Bromine

PHOT.5.

Calcium Hardness (Calcicol)

PHOT.12.

Chloride (Chloridol)

PHOT.46.

Chlorine (DPD)

PHOT.7.

Chlorine (DPD4)

PHOT.8.

Chlorine/Chloramines (DPD)

PHOT.7.1

Chlorine Dioxide LR

Phot.74.

Chlorine Dioxide HR

Phot.76.

Chlorine Dioxide (DPD Glycine Method)

PHOT.7.3.

Chlorine HR

PHOT.9.

Chromium (Chromicol)

PHOT.55.

Color

PHOT.47.

Copper (Coppercol – Total and Free)

PHOT.10.

Copper (Free)

PHOT.10.

Cyanuric Acid

PHOT.13.

Dissolved Oxygen (0.0-0.8 vials)

PHOT.49.

Dissolved Oxygen (0.0-2.0 vials)

PHOT.50.

Fluoride

PHOT.14.

Hardness, Total (Hardicol)

PHOT.15.

Hydrazine

PHOT.41.

Hydrogen Peroxide LR

PHOT.16.

Hydrogen Peroxide HR

PHOT.17.

Iron LR

PHOT.18.

Iron MR

PHOT.39.

Iron HR

PHOT.19.

Magnesium (Magnecol)

PHOT.21.

Manganese

PHOT.20.

Molybdate LR

PHOT.42.

Molybdate HR

PHOT.22.

Nickel (Nickeltest)

PHOT.53.

Nitrate (Nitratest)

PHOT.23.

Nitrite (Nitricol)

PHOT.24.

Nitrite (Nitriphot)

PHOT.43.

Organophosphonate (OP)

PHOT.44.

Ozone

PHOT.25.

pH Value

PHOT.27.

Phenol (Phenoltest)

PHOT.54.

PHMB (PHMB-PHOT)

PHOT.52.

Phosphate LR

PHOT.28.

Phosphate HR

PHOT.29.

Potassium

PHOT.30.

Silica LR

PHOT.31.

Silica HR

PHOT.56.

Sulfate

PHOT.32.

Sulfide

PHOT.33.

Sulfite (Sulfitest)

PHOT.34.

Turbidity

PHOT.48.

Zinc

PHOT.35.

Direct-Reading Photometer Program Schedule

TABLET REAGENT SYSTEM

Instruction

 

 

Photometer

Sheet

Reagent System

Parameter

Program Number

Number

 

 

 

 

 

-

-

Transmittance (%)

Phot 0

-

-

Absorbance

Phot1

PHOT.2.

Alkalinity, Total (Alkaphot)

Total Alkalinity

Phot 2

PHOT.37.

Alkalinity M (Alkaphot M)

Alkalinity M

Phot 37 (M)

PHOT.37.

Alkalinity P (Alkaphot P)

Alkalinity P

Phot 38 (P)

PHOT.3.

Aluminum

Aluminum

Phot 3

PHOT.4.

Ammonia

Ammonia Nitrogen

Phot 4

PHOT.5.

Bromine

Bromine – Total

Phot 5

Bromine – Free

Continuation test* (Phot 6)

 

 

PHOT.12.

Calcium Hardness (Calcicol)

Calcium Hardness

Phot 12

PHOT.46.

Chloride (Chloridol)

Chloride

Phot 46

PHOT.7.

Chlorine (DPD)

Chlorine - Free

Phot 7

Chlorine – Total

Continuation test* (Phot 8)

 

 

 

 

Chlorine – Free

Phot 71

PHOT.7.1.

Chlorine/Chloramines (DPD)

Monochloramine

Continuation test* (Phot 72)

 

 

Dichloramine

Continuation test* (Phot 73)

PHOT.8

Chlorine (DPD4)

Total Chlorine

Phot 8

PHOT.74.

Chlorine Dioxide LR

Chlorine Dioxide

Phot 74

PHOT.76.

Chlorine Dioxide HR

Chlorine Dioxide

Phot 76

PHOT.7.3.

Chlorine Dioxide

Chlorine Dioxide

Phot 7

 

(DPD Glycine Method)

 

 

 

PHOT.9.

Chlorine HR

Chlorine

Phot 9

 

Chromium (Chromicol)

Hexavalent

Phot 55

PHOT.55.

Chromium III – supplement

Total and Trivalent

Continuation test* (Phot 100)

 

to YPM281 above

 

 

 

PHOT.47.

Color

Color

Phot 47

PHOT.10.

Copper (Coppercol)

Copper – Free

Phot 10

Copper – Total

Continuation test* (Phot 11)

 

 

PHOT.13.

Cyanuric Acid

Cyanuric Acid

Phot 13

PHOT.49.

Dissolved Oxygen

Dissolved Oxygen

Phot 49

(0.0 – 0.8 vials)

 

 

 

PHOT.50.

Dissolved Oxygen

Dissolved Oxygen

Phot 50

(0.0 – 2.0 vials)

 

 

 

PHOT.14.

Fluoride

Fluoride

Phot 14

PHOT.15.

Hardness, Total (Hardicol)

Total Hardness

Phot 15

PHOT.41.

Hydrazine

Hydrazine

Phot 41

PHOT.16.

Hydrogen Peroxide LR

Hydrogen Peroxide

Phot 16

PHOT.17.

Hydrogen Peroxide HR

Hydrogen Peroxide

Phot 17

PHOT.18.

Iron LR

Iron

Phot 18

PHOT.39.

Iron MR

Iron

Phot 39

PHOT.19.

Iron HR

Iron

Phot 19

PHOT.21.

Magnesium (Magnecol)

Magnesium

Phot 21

PHOT.20.

Manganese

Manganese

Phot 20

PHOT.42.

Molybdate LR

Molybdate

Phot 42

PHOT.22.

Molybdate HR

Molybdate

Phot 22

PHOT.53.

Nickel (Nickeltest)

Nickel

Phot 53

Instruction

 

 

Photometer

Sheet

Reagent System

Parameter

Program Number

Number

 

 

 

 

 

PHOT.23.

Nitrate (Nitratest)

Nitrate Nitrogen

Phot 23

PHOT.24.

Nitrite (Nitricol)

Nitrite Nitrogen

Phot 24

PHOT.43.

Nitrite (Nitriphot)

Sodium Nitrite

Phot 43

PHOT.44.

Organophosphonate (OP)

Org-Pho (+Phos)

Phot 44

Organophosphonate

Continuation test* (Phot 45)

 

 

PHOT.25.

Ozone

Ozone (+Chlor)

Phot 25

Ozone

Continuation test* (Phot 26)

 

 

PHOT.27.

pH Value

pH – Phenol Red

Phot 27

PHOT.54.

Phenol (Phenoltest)

Phenol

Phot 54

PHOT.52.

PHMB (PHMB-PHOT)

PHMB

Phot 52

PHOT.28.

Phosphate LR

Phosphate

Phot 28

PHOT.29.

Phosphate HR

Phosphate

Phot 29

PHOT.30.

Potassium

Potassium

Phot 30

PHOT.31.

Silica LR

Silica

Phot 31

PHOT.56.

Silica HR

Silica

Phot 56

PHOT.32.

Sulfate

Sulfate

Phot 32

PHOT.33.

Sulfide

Sulfide

Phot 33

PHOT.34.

Sulfite (Sulfitest)

Sulfite

Phot 34

PHOT.48.

Turbidity

Turbidity

Phot 48

PHOT.35.

Zinc

Zinc (+ Copper)

Phot 35

Zinc

Continuation test* (Phot 36)

 

 

*Continuation tests cannot be accessed directly

PHOT.2.AUTO

PHOTOMETER TEST INSTRUCTIONS

_________________________________________________________________________________________

Photometer Method

Automatic

Wavelength

Selection

0 – 500 mg/l CaCO3

Natural and treated waters may contain a variety of dissolved alkaline substances such as carbonates, bicarbonates, hydroxides and, to a lesser extent, borates, phosphates and silicates. In water at neutral pH the alkalinity derives mainly from the presence of bicarbonates.

Total alkalinity is an important test in determining the aggressiveness or scale forming tendency of the water. If the total alkalinity is low the water may be aggressive and cause corrosion to pipe work and structures; if the total alkalinity is high the water may more readily promote scale formation. Alkalinity control is therefore an important part of many water treatment programmes.

The YSI Alkaphot test uses a colorimetric method and covers the total alkalinity range 0 - 500 mg/l CaCO3. The test is particularly suitable for checking natural and drinking waters, swimming pool water, boiler water, etc.

Method

The YSI Alkaphot test is based on a unique colorimetric method and uses a single tablet reagent. The test is simply carried out by adding a tablet to a sample of the water. Under the conditions of the test, a distinctive range of colors from yellow, through green, to blue is produced over the alkalinity range 0 - 500 mg/l CaCO3. The color produced in the test is indicative of the alkalinity of the water and is measured using a YSI Photometer.

Reagents and Equipment

YSI Alkaphot Tablets

YSI 9300 OR 9500 Photometer

Round Test Tubes, 10 ml glass (PT 595)

Test Procedure

1Fill test tube with sample to the 10 ml mark.

2 Add one Alkaphot tablet, crush and mix until all of the particles have dissolved. 3 Stand for one minute then remix.

4Select Phot 2 on photometer.

5 Take photometer reading in usual manner (see photometer instructions).

6The result is displayed as mg/l CaCO3.

Note: To convert Total Alkalinity as CaCO3 to Total Alkalinity as HCO3- multiply result by 1.22.

____________________________

Loading...
+ 113 hidden pages