Xylem YSI FDO 70X IQ SW, YSI FDO 70X IQ, YSI FDO 70X IQ H User Manual

OPERATING MANUAL
FDO
®
70x IQ
FDO
®
70x IQ H
FDO
®
70x IQ SW
ba76014d03 05/2014
®
70x IQ
®
70x IQ SW
®
OPTICAL SENSOR FOR DISSOLVED OXYGEN
70x IQ H
FDO®70x IQ ...
For the most recent version of the manual, please visit www.ysi.com.
Contact YSI
1725 Brannum Lane Yellow Springs, OH 45387 USA Tel: +1 937-767-7241
800-765-4974 Email: environmental@ysi.com Internet: www.ysi.com
Copyright © 2013 Xylem Inc.
ba76014d03 05/2014
FDO®70x IQ ... Contents
FDO®70x IQ ... - Contents
1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
1.1 How to use this component operating manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-1
1.2 Sensor models . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
1.3 Recommended fields of application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
1.4 Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
2 Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
2.1 Safety information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-1
2.1.1 Hazard warnings in this operating manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-1
2.1.2 Safety information on the product . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
2.2 Safe operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
2.2.1 Authorized use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
2.2.2 Requirements for safe operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
2.2.3 Unauthorized use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
2.3 Hazardous location use (FDO
®
70x IQ H models) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
3 Commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
3.1 Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-1
3.2 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-1
3.3 Commissioning / Getting the instrument ready for measuring . . . . . . . . . . . . . . . . . . . . . 3-3
3.4 FDO
®
70x IQ ... setting table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-4
4 Measuring / Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
4.1 Measuring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-1
4.2 Function check and user calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-1
4.2.1 General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-1
4.2.2 Function check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-2
4.2.3 User calibration 4-4
4.2.4 Calibration history 4-6
4.2.5 Reactivating previous calibration data 4-7
5 Maintenance, cleaning, replacement parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
5.1 General maintenance instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-1
5.2 Handling of the sensor cap . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
5.3 Exchanging the sensor cap . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-2
5.4 Cleaning the sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-3
5.4.1 Exterior cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
5.4.2 Interior cleaning of sensor cap and sensor head 5-4
5.5 Replacement parts and accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-5
6 What to do if... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
7 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
7.1 General measurement characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
7.2 Application conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-1
ba76014e03 05/2014 0 - 1
Contents FDO®70x IQ ...
7.3 General data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-2
7.4 Instrument safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-3
7.4.1 General instrument safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-3
7.4.2 Hazardous location ratings (H models) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-3
7.5 Electrical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-3
7.6 Technical data of the SC-FDO
7.7 Technical data of the SC-FDO
®
700 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-4
®
701 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-5
8 Contact Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1
8.1 Ordering & Technical Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-1
8.2 Service Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-1
9 Indexes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-1
9.1 Explanation of the messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-1
9.1.1 Error messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-1
9.1.2 Info messages 9-3
9.2 Status info . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-4
0 - 2 ba76014e03 05/2014
FDO®70x IQ ... Overview
IQ Sensor Net Operating Manual
System
Operating
Manual
(Ring Binder)
IQ Sensor Operating
Manual
MIQ Module
Operating
Manual
MIQ Terminal
Operating
Manual
Component Operating Manuals

1Overview

1.1 How to use this component operating manual

Structure of the
IQ S
ENSORNET operating
manual
Fig. 1-1 Structure of the IQ SENSORNET operating manual
The IQ SENSOR NET operating manual has a modular structure like the IQ S
ENSORNET system itself. It consists of a system operating manual
and the operating manuals of all the components used. Please file this component operating manual into the ring binder of the
system operating manual.
ba76014e03 05/2014 1 - 1
Overview FDO®70x IQ ...
FDO 700 IQ
®
FDO 700 IQ H
®
SC-FDO 700
®
SC-FDO 701
®
Sensor basis
Sensor basis
Sensor cap:
FDO 700 IQ SW
®

1.2 Sensor models

The FDO®70x IQ ... consists of a sensor basis equipped with different sensor cap types, depending on the model:
Sensor models
SW models Compared to the standard models, the sea water models of the
Sensor model Sensor basis Sensor cap
®
FDO
FDO
FDO
FDO
FDO
FDO
700 IQ FDO®700 IQ * SC-FDO®700
®
700 IQ SW FDO®700 IQ SW * SC-FDO®700
®
701 IQ FDO®700 IQ * SC-FDO®701
®
701 IQ SW FDO®700 IQ SW * SC-FDO®701
®
700 IQ H FDO®700 IQ H ** SC-FDO®700
®
701 IQ H FDO®701 IQ H ** SC-FDO®701
* Type designation on the name plate in the plug head connector ** Type designation on the name plate on the shaft The sensor cap primarily determines the measuring characteristics of
the sensor as a whole. Thus the sensor can be adapted to the measuring job by exchanging the type of sensor cap. The specifica­tions of the individual sensor cap types are given in chapter 7
ECHNICAL DATA.
T
H models The sensor models FDO
use in Class I Division 2 hazardous locations (see chapter 2.3 for details).
sensors (SW models) are optimized concerning their resistance to corrosion in sea water and brackish water.
®
700 IQ H and FDO®701 IQ H are rated for
1 - 2 ba76014e03 05/2014
FDO®70x IQ ... Overview

1.3 Recommended fields of application

Sensor model Recommended field of application.
®
FDO FDO
FDO FDO
FDO FDO
700 IQ
®
701 IQ
®
700 IQ H
®
701 IQ H
®
700 IQ SW
®
701 IQ SW
Stationary measurements in water/wastewa­ter applications in non-hazardous locations only.
Stationary measurements in water/wastewa­ter applications in Class I Division 2 hazard­ous locations.
Stationary measurements in seawater, aquaculture.
®
The FDO
70x IQ SW is equipped with a protective hood that is prepared for the con­nection of the MSK FDO
®
CS cleaning set for compressed air-driven sensor cleaning (order information, see section
5.5 R
EPLACEMENT PARTS AND ACCESSORIES).
ba76014e03 05/2014 1 - 3
Overview FDO®70x IQ ...
67
8453
9
12

1.4 Structure

Structure
Sensor cap with
memory chip
Intelligent membrane
(IQMC technology)
Fig. 1-2 Structure of the D. O. sensor (example: FDO® 70x IQ)
1 Shaft 2 Plug head connector 3 Fixing ring 4 Sensor membrane 5 Sensor cap with memory chip 6 Gold-plated contact pins for memory chip 7 Measurement window 8 Temperature sensor and locking device 9 Sensor head
A memory chip is integrated in the sensor cap. The following data is stored on the memory chip:
Type designation of the sensor capSeries numberData of the factory calibration
For each membrane the individual calibration values are determined by a factory calibration and stored on the memory chip of the sensor cap, ensuring maximum accuracy over the whole lifetime of the sensor.
1 - 4 ba76014e03 05/2014
FDO®70x IQ ... Safety

2 Safety

2.1 Safety information

2.1.1 Hazard warnings in this operating manual

The hazard warnings are defined for the following levels of danger:
DANGER
DANGER indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury.
WARNING
WARNING indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
CAUTION
CAUTION indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury.
NOTICE
NOTICE is used to address practices not related to per­sonal injury.

2.1.2 Safety information on the product

Note all labels, information signs and safety symbols on the product.
ba76014e03 05/2014 2 - 1
Safety FDO®70x IQ ...

2.2 Safe operation

2.2.1 Authorized use

®
The authorized use of the FDO sensor in the IQ S
ENSORNET.
The technical specifications according to chapter 7 T must be observed. Only operation according to the instructions in this operating manual is authorized.
Any other use is considered to be unauthorized. Unauthorized use invalidates any claims with regard to the guarantee.

2.2.2 Requirements for safe operation

Note the following points for safe operation: The product may only be op erated according to the au thorized use
specified above.
70x IQ ... comprises its use as a D.O.
ECHNICAL DATA
The product may only be supplie d with power by the energy sources
specified in this operating manual.
The product may only be operated under the environmental condi-
tions specified in this operating manual.
The product or its components may only be opened if this is required
for installation and maintenance work and described in the operating manual.

2.2.3 Unauthorized use

The product must not be put into operation if:
it is visibly damaged (e.g. after being transported)it was stored under adverse conditions for a lengthy period of time
(storing conditions, see chapter 7 T
ECHNICAL DATA)
2 - 2 ba76014e03 05/2014
FDO®70x IQ ... Safety
######
######### ###############
#################### #################
######
######
FDO 700 IQ H
28123

2.3 Hazardous location use (FDO®70x IQ H models)

DANGER
Explosion hazard. Only hazardous location rated sensor models ("Hazloc Sensor") must be used in hazardous locations. Read the name plate on the sensor shaft and verify that the sensor is rated for your specific application.
Rated sensor models The following sensor models are rated for hazardous location use:
®
FDOFDO
Hazardous location rated sensors are identified by the name plate with the rating details engraved on the sensor shaft:
700 IQ H (YSI-Order No. 207065Y)
®
701 IQ H (YSI-Order No. 207066Y)
Name plate of a Hazloc
sensor (example)
DANGER
Explosion hazard. The associated apparatus MIQ/BB1 (YSI part no. 207001Y) has to be used for connection. Refer to the following documents for proper installation: Control drawing, YSI document no. 28123
(see MIQ/BB1 operating manual)
MIQ/BB1 operating manualAll applicable electrical code regulations.
ba76014e03 05/2014 2 - 3
DANGER
Explosion hazard. This product does not meet the requirements of the directive 94/9/EC (ATEX).
Safety FDO®70x IQ ...
2 - 4 ba76014e03 05/2014
FDO®70x IQ ... Commissioning

3 Commissioning

3.1 Scope of delivery

Sensor, consisting of sensor basis and sensor capOperating manual.

3.2 Installation

Safety guidelines for
installation in a
hazardous location
(H-models only)
Explosion hazard. Only hazardous location rated sensor models ("Hazloc Sensor") must be used in hazardous locations. Verify the rating on the product name plate on the sensor shaft. The associated apparatus MIQ/BB1 (YSI part no. 207001Y) has to be used for connection. Refer to the following documents for proper installation: Control drawing, YSI document no. 28123
(see MIQ/BB1 operating manual)
MIQ/BB1 operating manualAll applicable electrical code regulations.
DANGER
Connection cable A sensor connection cable of the SACIQ or SACIQ SW type is required
to connect the sensor. The cable is available in different lengths. Com­pared to the standard model SACIQ, the SACIQ SW sensor connection cable is optimized regarding its corrosion resistance in seawater and brackish water and adapted for use in conjunction with the FDO 70x IQ SW. Information on this and other IQ S given in the YSI catalog and on the Internet.
ENSORNET accessories is
DANGER
Explosion hazard. For hazardous location use only the cable types list­ed in the control drawing must be used.
How to connect the SACIQ (SW) sensor connection cable to the terminal strip of a MIQ module is described in chapter
NSTALLATION of the IQ SENSORNET system operating
3 I manual.
ba76014e03 05/2014 3 - 1
Commissioning FDO®70x IQ ...
SACIQ
1
2
Are the plug
connections dry?
Connecting the sensor
to the sensor
connection cable
Before connecting the sensor and sensor connection cable, please make sure the plug connections are dry. If moisture gets into the plug connections, first dry the plug connections (dab them dry or blow them dry using compressed air).
Do not suspend the sensor on the sensor connection cable. Use a sensor holder or armature. Information on this and other IQ S
ENSORNET accessories is given in the YSI cata-
log and on the Internet.
1 Take the protective caps off the plug connect ions of the sensor
and the SACIQ (SW) sensor connection cable and keep them safe.
2 Plug the jack of t he SACIQ (SW) sensor connection cable onto
the plug head connector of the sensor. At the same t ime, rotate the socket so that the pin in the plug head connector (1) clicks into one of the two holes in the socket.
3 Then screw the coupling ring (2) of the sensor connection
cable onto the sensor up to the stop.
Fig. 3-1 Connecting the sensor
3 - 2 ba76014e03 05/2014
FDO®70x IQ ... Commissioning

3.3 Commissioning / Getting the instrument ready for measuring

Identification in the
IQ S
ENSORNET
Commissioning steps
Both the sensor cap and sensor basis (sensor without cap) have a series number of their own. The sensor has to be equipped with a sen­sor cap for a successful login to the IQ S
ENSORNET. The operable sen-
sor is displayed as follows in the list of sensors:
Model: Model designation of the sensor cap
(="SC FDO 70x")
Ser. no.: Series number of the sensor capSensor name: The series number of the sensor basis is preset
here. You can change this entry as necessary by entering a user-defined name.
1 Pull the protective cap off the sensor. 2 If required, assign a user-defined name to the sensor (see rel-
evant IQ S
ENSORNET system operating manual).
3 When using IQ SENSORNET without automatic air
pressure compensation the mean air pressure value or the location altitude has to be entered (for more detailed informa­tion see the relevant IQ S
ENSORNET system operating man-
ual).
4 Set the sensor (see section 3.4).
ba76014e03 05/2014 3 - 3
Commissioning FDO®70x IQ ...

3.4 FDO®70x IQ ... setting table

Setting Selection/values Explanation
Measuring mode Concentration
Saturation
Measuring range Concentration
Measuring range Sat-
0 ... 20.00 mg/l
0 ... 20.00 ppm
0 ... 200.0 % The measuring range is set permanently.
uration
Response time t90 150 .. 300 s
®
(with SC-FDO
700)
or
60 ... 300 s
(with SC-FDO
®
701)
Calibration Factory calibra-
tion
User calibration
active
abort
Unit of the measured value on the measured value display.
These measuring ranges are available for selection.
Response time of the signal filter. Depending on the sample matrix, the mea­sured values may fluctuate more or less. A signal filter in the sensor reduces the fluc­tuation range of the measured value. The signal filter is characterized by the Response time t90. This is the time after which 90 % of a signal change is displayed.
The setting range depends on the sensor cap type.
Determines which calibration data the mea­sured value calculation will be based on. The active calibration is displayed in the calibra­tion history.
The selection, User calibration is only dis­played if valid data of a User calibration is stored in the sensor.
active indicates that the sensor is being cali­brated.
If abort is selected, the active calibration is canceled the next time the setting table is quit with Save and quit.
Test active
abort
active indicates that the sensor is being checked.
If abort is selected, the active check is can­celed the next time the setting table is quit with Save and quit.
Temperature mode °C
°F
3 - 4 ba76014e03 05/2014
Unit of the measured temperature value (Celsius, Fahrenheit).
FDO®70x IQ ... Commissioning
Setting Selection/values Explanation
Temp. adjustment -1.5 K ... +1.5 K The temperature compensation enables the
temperature display to be balanced (shifting of the zero point by ±1.5K). Notes:
Due to the thermal capacity of the sensor,
it is necessary to place it in a container with at least 2 liters of water.
Leave the sensor in this container for at
least 15 minutes while stirring occasion­ally until the balancing can be performed. With temperature differences between the water and sensor > 10 °C, leave the sen­sor for at least 1 hour in this container while stirring occasionally until the balanc­ing can be performed.
Salinity On
Off
Determines whether the entered salinity should be taken into account.
Salinity input 2.0 ... 70.0 The entry of the salinity enables a salt con-
tent correction that compensates for the effect of salt contents > 0.1 % on the oxygen measurement. The salt content correction is recommended for measurements in salt-contaminated wastewater (salinity 2.0 corresponding to a conductivity of 3.4 mS/cm at a reference
Sensor cap data Do not download
Transmit to log
book
temperature Generates a log book message with all data
stored in the sensor cap when the settings are quit with Save and quit (see section 1.2). When opening the menu again the setting is
T
REF
= 20 °C).
reset to Do not download.
Save and quit The system confirms the saving of the set-
tings and the display switches to the next higher level.
Quit The display switches to the next higher level
without saving the new settings.
Carrying out settings Using <S>, switch from the measured value display to the main menu
of the settings. Then navigate to the setting menu (setting table) of the sensor. The exact procedure is given in the relevant IQ S
ENSORNET
system operating manual.
ba76014e03 05/2014 3 - 5
Commissioning FDO®70x IQ ...
3 - 6 ba76014e03 05/2014
FDO®70x IQ ... Measuring / Operation

4 Measuring / Operation

4.1 Measuring

WARNING
Chemical or biological hazard. Contact with the sample can be harmful to the user. Depending on the type of sample, suitable protective measures must be taken (protective clothing, gog­gles, etc.).
Note the data given in section 7.2 APPLICATION CONDITIONS, especially the minimum immersion depth of the sensor (> 50 mm). The measured value is available immediately on submersing.
If keeping the sensor clean is a problem, we recommend using the compressed air-driven cleaning system with the CH cleaning head (see section 5.5 R
AND ACCESSORIES).
EPLACEMENT PARTS
Factory calibration The FDO
When does a function
check or user
calibration make sense?
Check or calibration
medium

4.2 Function check and user calibration

4.2.1 General information

®
70x IQ ... is factory calibrated. In the recommended applica­tion (see section 1.3 R measuring characteristics of the sensor cap remain stable for the spec­ified service life. Thus, a user calibration is not usually required.
A function check or user calibration can be useful in the following spe­cial cases:
If the measured values appear to be implausible and it is assumed
that the service life of the sensor cap is over
Routinely within the framework of the company quality assurance. Select one of the following two variants depending on the air tempera-
ture at the calibration site: With air temperatures over 5 °C, the function check and user calibra-
tion ideally take place in water vapor-saturated air. To do so, posi­tion the sensor approx. 2 cm above a water surface, for example in a narrow bucket or similar container with water. The membrane must be clean and dry for this.
ECOMMENDED FIELDS OF APPLICATION), the
ba76014e03 05/2014 4 - 1
Measuring / Operation FDO®70x IQ ...
2cm
With air temperatures under 5 °C we recommend performing the
function check and user calibration not in air but in air-saturated water that has a higher temperature. You obtain air-saturated water by pouring water several times in and out of two vessels so that it sparkles.

4.2.2 Function check

A function check is the simplest way to determine whether the sensor needs to be cleaned or user-calibrated.
Principle The function check can either be done in water vapor-saturated
air or in air-saturated water (see C
HECK OR CALIBRATION MEDIUM on
page 4-1).
Steps Generally, a check on the IQ S
tem specific details are given in the respective IQ S
ENSORNET is carried out as follows. Sys-
ENSORNET system
operating manual.
1 Switch to the measured value display with <M> and select the
®
FDO
70x IQ ... sensor.
2Press <C>.
The next step switches on the maintenance condition for the sensor. A corresponding message appears on the display.
3 Confirm the note with <OK>.
The maintenance condition is active. 4 Select the TEST procedure and press <OK>. 5 Put the sensor into the calibration position (water vapor-satu-
rated air or air-saturated water - see section 4.2.1). 6Press <OK>.
The sensor starts the check. The display switches to the mea-
sured value display. The CAL indicator flashes instead of the
main measured value. At the same time, the momentary rela-
tive slope flashes as the secondary measured value with the
addition of TEST. The process ends automatically as soon as
the measured values meet the criterion for the stability control.
With a great temperature difference between the sensor and
the environment this may take a while. Subsequently, the main
measured value and temperature are displayed.
4 - 2 ba76014e03 05/2014
FDO®70x IQ ... Measuring / Operation
7 Put the sensor in the measuring position again. 8 Wait for the measured value to b e largely stable (temperature
adjustment).
9 Switch off the maintenance condition.
Canceling the check As long as the determination of the relative slope has not yet been
started (step 6), you can quit the check with <M> or <ESC>. The running determination of the relative slope (after pressing <OK> in
step 6) can be aborted as follows:
1 Open the setting table (see section 3.4). 2 In the Test me nu item, select the abort setting and then quit the
setting table with Save and quit.
Evaluation The result of the check is entered in the log book of the sensor. If the
relative slope is outside the tolerance range (0.90 ... 1.10), the check is assessed as erroneous.
The information recorded in the log book entry is partly identical with that of the calibration history. The calibration history is described in detail in section 4.2.4 C
HISTORY.
ALIBRATION
ba76014e03 05/2014 4 - 3
Measuring / Operation FDO®70x IQ ...

4.2.3 User calibration

Principle The user calibration can either be done in water vapor-saturated
air or in air-saturated water (see C
HECK OR CALIBRATION MEDIUM on
page 4-1). With the calibration procedure, the relative slope of the sen­sor is determined. The calibration is evaluated based on the relative slope and the intensity (successful <-> unsuccessful).
The result of the user calibration is stored in the calibration record and calibration history respectively and can be viewed afterwards (see respective IQ S
ENSORNET system operating manual).
We want to point out that the factory calibration of the mem­brane is highly precise due to the IQMC technology. If you still want to carry out a user calibration, influences due to environmental conditions have to be taken into account.
Steps Generally, a user calibration on the IQ S
lows. System specific details are given in the respective IQ S system operating manual.
1 Switch to the measured value display with <M> and select the
®
FDO
70x IQ ... sensor.
2Press <C>.
The next step switches on the maintenance condition for the
sensor. A corresponding message appears on the display. 3 Confirm the note with <OK>.
The maintenance condition is active. 4 Select the CALIBRATION procedure and press <OK>. 5 Put the sensor into the calibration position (water vapor-satu-
rated air or air-saturated water - see section 4.2.1). 6Press <OK>.
The sensor determines the calibration data. The display
switches to the measured value display. The CAL indicator
flashes instead of the main measured value. At the same time,
the momentary relative slope flashes as the secondary mea-
sured value. The process ends automatically as soon as the
measured values meet the criterion for the stability control.
With a great temperature difference between the sensor and
the environment this may take a while. Subsequently, the main
measured value and temperature are displayed.
ENSORNET is carried out as fol-
ENSORNET
7 If the user calibratio n was successful, bring the sensor into the
measuring position again. 8 Wait for the measured value to be largely stable (temperat ure
adjustment). 9 Switch off the maintenance condition.
4 - 4 ba76014e03 05/2014
FDO®70x IQ ... Measuring / Operation
Canceling the user
calibration
Possible results of the
user calibration
As long as the determination of the calibration data has not yet been started (step 6), you can quit the calibration routine with <M> or <ESC>.
The running determination of calibration data (after pressing <OK> in step 6) can be aborted as follows:
1 Open the setting table (see section 3.4). 2 In the Calibration menu item, se lect the abort setting and then
quit the setting table with Save and quit.
After the user calibration was canceled the sensor again works with the calibration data that was used before the canceled user calibration.
The calibration data is evaluated by the system. A calibration can have the following results:
Display Explanation
Measured value display Sensor was successfully calibrated.
The calibration data can be viewed in the calibration history (section 4.2.4).
"----" The sensor could not be calibrated.
The sensor is blocked for further measure­ment. Notes on possible causes are given in the log book of the sensor.
ba76014e03 05/2014 4 - 5
Measuring / Operation FDO®70x IQ ...
Currently active calibration
Chronological list of the last user cali­brations

4.2.4 Calibration history

Calibration history
(available in the
IQ S
ENSORNET systems
184 XT and 2020 XT
only)
Fig. 4-1 Calibration historyFDO®70x IQ ...
The calibration history contains the following information:
Date Calibration date (Factory = factory calibration) Rel.slope Relative slope (non-dimensional) Intens. Intensity:
+ : sufficient
- : too low
Res Evaluation of the calibration
o.k. : Calibration successful. The new calibration data was taken over for measure­ment. Error : Calibration unsuccess- ful. Invalid calibration data were discarded. Measurement was continued with the last valid values.
The calibration history is stored in the sensor basis. If another sensor cap (with different series number) is mounted, the calibration history is deleted and cannot be restored afterwards.
4 - 6 ba76014e03 05/2014
FDO®70x IQ ... Measuring / Operation

4.2.5 Reactivating previous calibration data

®
The FDO
70x IQ ... enables you to reactivate the last valid user cali­bration or the factory calibration. Thus you can immediately go on measuring if a calibration procedure failed or you suspect that the cali­bration conditions were not optimally met.
Reactivating old calibration data is only a temporary solu­tion. Please take into account that the sensor might thus possibly provide incorrect measured values. Please make sure the sensor works correctly by carrying out a function check and/or recalibration.
Reactivating calibration
data
1 Open the setting table (see section 3.4). 2 In the Calibration menu item, sele ct the User calibration or Fac-
tory calibration setting and then quit the setting table with Save and quit.
ba76014e03 05/2014 4 - 7
Measuring / Operation FDO®70x IQ ...
4 - 8 ba76014e03 05/2014

FDO®70x IQ ... Maintenance, cleaning, replacement parts

5 Maintenance, cleaning, replacement
parts

5.1 General maintenance instructions

WARNING
Chemical or biological hazard. Contact with the sample can be harmful to the user. Depending on the type of sample, suitable protective measures must be taken (protective clothing, gog­gles, etc.).
Maintenance condition We recommend to switch on the maintenance condition each time the
sensor is taken out of the measuring position. This avoids unwanted reactions of linked outputs. For more detailed information on the main­tenance condition please refer to the respective IQ S operating manual.
ENSORNET system
We do not recommend unscrewing the sensor from the sensor connection cable in order to do maintenance work. Otherwise, moisture and/or dirt can get into the plug con­nection where it can cause contact problems.
If you would like to disconnect the sensor from the sensor connection cable, please note the following points:
Before disconnecting the sensor from the SACIQ (SW) sensor con-
nection cable, the sensor has to be cleaned on the outside (see sec­tion 5.4.1).
Unscrew the sensor from the SACIQ (SW) sensor connection cable.
Place a protective cap on the sensor plug head connector and on
the SACIQ (SW) sensor connection cable each time so that no mois­ture or dirt can get on the contacting surfaces.
In corrosive environments, close the socket of the sensor connec-
tion cable with the screwable SACIQ-Plug when it is dry in order to protect the electrical contacts from corrosion. The protective plug is available as an accessory (see section 5.5 R
ACCESSORIES). It is included in the standard scope of delive ry of the
SACIQ SW sensor connection cable.
EPLACEMENT PARTS AND
ba76014e03 05/2014 5 - 1
Maintenance, cleaning, replacement parts FDO®70x IQ ...

5.2 Handling of the sensor cap

Despite its exterior robustness, the sensor is an optical high precision instrument. Therefore, special care should be taken when doing any maintenance or cleaning work:
Dirt and moisture under the sensor cap can affect the functioning
and shorten the service life of the sensor cap. Therefore, make sure the working environment is clean and dry prior to removing the sen­sor cap.
Please do not touch the outer sensor membrane with your fingers.
Touch the sensor cap at the sides only (shaded area in figure on the left).
Avoid any great mechanical stress of the sensor membrane (pres-
sure, scratches).
Exposure to light, particularly daylight of the interior of the sensor
cap will, by-and-by, affect the measurement characteristics and shorten the service life of the sensor cap. Therefore, the interior of the sensor cap should not be exposed to direct sunlight. Avoid any exposure to light that exceeds the extent required for necessary maintenance and cleaning activities. Store dismantled sensor caps in a light-protected environment only.
Removing the sensor
cap

5.3 Exchanging the sensor cap

1 Pull the sensor out of the sample. 2 Clean the outside of the sensor (see section 5.4.1). 3 Unscrew the fixing ring from the sensor by hand. 4 Thoroughly clean and dry the sensor head once again. 5 Grasp the sensor cap on the sides (arrows in figure on the left)
and remove it by pulling it away from the sensor in a straight upward direction.
NOTICE
Do not push any tools or other sharp objects between the sealing surfaces. This might damage the sealing surfaces.
5 - 2 ba76014e03 05/2014
FDO®70x IQ ... Maintenance, cleaning, replacement parts
Mounting the sensor
cap
6 Check the front surface of the sensor for absolute cleanness
and clean it if necessary (see section 5.4.1). 7 Thoroughly clean the thread of the fixing ring. 8 Place the new sensor cap on the sensor so that the tempera-
ture sensor fits into the hole inside the sensor cap (see figure
opposite). 9 Put the fixing ring on the sensor head and screw it tight by
hand.

5.4 Cleaning the sensor

5.4.1 Exterior cleaning

Dirt on the sensor can affect the measuring characteristics. Biological deposits for example, consume oxygen and can, when occurring on the sensor cap membrane, impair the responding behavior and cause values that are too low. Therefore, we recommend regular visual inspections and exterior cleaning as necessary.
Pay attention to the following points for cleaning: First, thoroughly rinse the sensor with tapwater to remove loosely
adhering dirt.
Rough dirt o n the sensor shaft can be brushed off with a soft brush.
Attention:
Do not use the brush in the area of the sensor membrane.
Risk of damage!
The sensor cap including the sensor membrane should be wiped
with a soft and moist microfiber cloth.
In the case of persisting dirt you can add some household wash ing-
up liquid to the tapwater. Attention:
Never use any alcohol for clean-
ing!
If the are any salt or lime deposits, the sensor can be cleaned with
aqueous citric acid solution (10 % weight percent).
ba76014e03 05/2014 5 - 3
Maintenance, cleaning, replacement parts FDO®70x IQ ...

5.4.2 Interior cleaning of sensor cap and sensor head

If moisture or dirt have penetrated under the sensor cap, e.g. because the sensor cap is damaged, you can make the sensor ready for opera­tion again as follows:
NOTICE
Only use nonabrasive, alcohol-free detergents, as otherwise the optical surfaces could be damaged.
1 Remove the sensor cap (see section 5.3). 2 Clean the sensor head and sensor cap:
– Rinse all inner surfaces with tapwater
– Remove contamination containing fat and oil with warm
water and household washing-up liquid
3 Dry all surfaces with a clean, lint free cloth. A lint-free cloth
such as a microfiber cloth used to clean eyeglasses is suitable.
4 Allow the sensor and sensor cap to dry completely at a dry
location so moisture can evaporate even from corners difficult to access. When doing, so protect the inside of the sensor cap from light.
If the sensor cap is visibly damaged it has to be replaced.
5 - 4 ba76014e03 05/2014
FDO®70x IQ ... Maintenance, cleaning, replacement parts

5.5 Replacement parts and accessories

Description Model Order no.
®
Sensor cap SC-FDO Sensor cap SC-FDO
700 201 654Y
®
701 201 655Y
Components for
compressed air cleaning
system
Protective screw cap for sensor
SACIQ-Plug 480 065Y
connection cable
Protective hood against damage
MSK FDO
®
205 253Y
caused by fish, with connection for
®
the MSK FDO
Cleaning set for connection of the protective hood MSK FDO
CS cleaning set
®
to a
MSK FDO
®
CS 205 254Y
valve module or Cleaning Air Box
Description Model Order no.
Cleaning head CH 900 107Y
Active valve module (does not
MIQ/CHV PLUS 480 018Y
require a free relay output in the
ENSORNET system)
IQ S
Valve module for the
ENSORNET system 182
IQ S
DIQ/CHV 472 007Y
(external relay and compressed air supply)
Air compressor to be triggered by relay, power supply 115 V AC
Air compressor to be triggered by relay, power supply 230 V AC
Information on further IQ S in the YSI catalog and on the Internet.
Cleaning Air Box
480 017Y
- 115 VAC Cleaning Air Box
480 019Y
- 230 VAC
ENSORNET accessories is given
ba76014e03 05/2014 5 - 5
Maintenance, cleaning, replacement parts FDO®70x IQ ...
5 - 6 ba76014e03 05/2014
FDO®70x IQ ... What to do if...

6 What to do if...

Sensor does not appear
in the measured value
display and list of
sensors
Measured value
implausible
Cause Remedy
– Sensor cap not mounted or
defective
Cause Remedy
– Coating on sensor cap – Clean the outside of the sen-
– Service life of the sensor cap
over
– Dirt inside the sensor cap and
sensor head.
– Fixing ring not properly tight-
ened.
– Sensor cap untight or defec-
tive.
– Sensor cap (see section 5.3)
– Replace a defective sensor
cap as necessary
sor (see section 5.4.1)
– Check the sensor (see sec-
tion 4.2.2)
– Replace the sensor cap as
necessary (see section 5.3)
– Dismantle sensor cap
– Clean the insides of the sen-
sor cap and sensor head (see section 5.4.2)
– Mount the sensor cap cor-
rectly and tighten the fixing ring as far as it will go (see section 5.3)
– Replace a defective sensor
cap as necessary
Measured value
fluctuating heavily
Cause Remedy
– Coating on sensor cap – Clean the outside of the sen-
sor (see section 5.4.1)
– Service life of the sensor cap
over
– Dirt inside the sensor cap and
sensor head.
– Fixing ring not properly tight-
ened.
– Sensor cap untight or defec-
tive.
– See point "Measured value
implausible"
– See point "Measured value
implausible"
ba76014e03 05/2014 6 - 1
What to do if... FDO®70x IQ ...
Measured value invalid
( "----" displayed)
Incorrect temperature
display
Cause Remedy
– User calibration unsuccessful.
The sensor is blocked for mea­surement.
– As a temporary measure to
quickly restore the readiness for service: Activate the fac­tory calibration (see section
4.2.5)
– For exact measurements,
carry out a function check and repeat the user calibration if necessary.
Cause Remedy
– Temperature sensor defective – Return the sensor
6 - 2 ba76014e03 05/2014
FDO®70x IQ ... Technical data

7 Technical data

7.1 General measurement characteristics

The measurement characteristics are primarily determined by the sensor cap type. The relevant data are given in the sections 7.6 and 7.7.
Measuring principle Optical measurement based on photoluminescence.
Measurement in water According to solubility function according to ISO 5814
Measurement in salt-
contaminated
wastewater
Salinity input from 2.0 ... 70.0; corresponds to 3.4 mS/cm ... 86.2 mS/cm at T
REF
20 °C (salinity measurement according to IOT = International Oceanographic Tables)
Temperature
Temperature sensor Integrated NTC
measurement
Measuring range - 5 °C ... + 50 °C (23 ... 122 °F) Accuracy ± 0.5 K Resolution 0.1 K
Temperature
In the range -5 °C ... + 50 °C (23 ... 122 °F)
compensation

7.2 Application conditions

Pressure resistance Sensor with connected SACIQ (SW) sensor connection cable:
Max. allowed overpressure 106 Pa (10 bar) The sensor meets all requirements according to article 3(3) of 97/23/
EC ("pressure equipment directive").
Type of protection Sensor with connected SACIQ (SW) sensor connection cable:
IP 68, 10 bar (10
6
Pa)
Immersion depth Min. 10 cm; max. 100 m depth
Operating position Any
ba76014e03 05/2014 7 - 1
Technical data FDO®70x IQ ...
59.5
40.0
276
34
400
400
39.7
39.7
FDO 70x IQ SW:
FDO 70x IQ (H):
Socket SACIQ...
Socket SACIQ...

7.3 General data

Dimensions
Weight (without sensor
Connection technique Connection via SACIQ (SW) sensor connection cable
connection cable)
Material
®
FDO
70x IQ (H) Approx. 900 g
®
70x IQ SW Approx. 1500 g
FDO
Shaft:
–FDO
–FDO
®
70x IQ
®
70x IQ SW
V4A stainless steel 1.4571 * POM
Plug head connector housing POM Sensor head POM and PVC Sensor cap See section 7.6 or section 7.7 Plug, 3-pole ETFE (blue) Tefzel
®
Fixing ring POM
Protective hood
®
(FDO
70x IQ SW)
POM
* Stainless steel can be corrodible if there are chloride concentrations of 500 mg/L or more. For applications in such media we recommend to use the SW sensors.
(SensCheck function)
7 - 2 ba76014e03 05/2014
Automatic sensor
monitoring
Monitoring of the membrane function
FDO®70x IQ ... Technical data

7.4 Instrument safety

7.4.1 General instrument safety

Applicable norms – UL 61010-1
– CAN/CSA C22.2 # 61010.1

7.4.2 Hazardous location ratings (H models)

Applicable directives
and standards
Hazardous location
rating
In addition to the standards listed in chapter 7.4.1, the FDO
®
70x IQ H
conforms to the following directives and standards:
ANSI/ISA-12.12.01CAN/CSA C22.2 # 213
The FDO
®
70x IQ H is a non-incendive electrical equipment for use in
Class I Division 2 hazardous locations.
FDO®700 IQ H
®
FDO
701 IQ H
Class I Division 2 Groups A, B, C, D T6 Class I Zone 2
Group IIC T6 Associated apparatus MIQ/BB1, YSI part no. 207001Y Control Drawing YSI document no. 28123

7.5 Electrical data

Rated voltage (non-H models)
24 V DC
via the IQ S
details, see chapter T
DATA of the IQ SENSORNET sys-
ENSORNET (for more
ECHNICAL
tem operating manual)
Voltage range (H models)
14 V DC ... 24 V DC via the MIQ/
BB1 barrier box (=associated
apparatus)
Power consumption 0.7 W Protective class III
ba76014e03 05/2014 7 - 3
Technical data FDO®70x IQ ...

7.6 Technical data of the SC-FDO®700

Measuring ranges and
resolutions
Measurement accuracy
Repeatability ± 0.05 mg/L (ppm) according to EN ISO 15839
Response time t
D. O. partial pressure 0 ... 400 hPa
Measuring
Adjustable measuring range Resolution
mode
D. O. concentra­tion
0 ... 20.00 mg/L 0 ... 20.00 ppm
0.01 mg/L
0.01 ppm
D. O. saturation 0 ... 200.0 % 0.1 %
In the range Accuracy
< 1 mg/L (ppm) ± 0.05 mg/L (ppm) > 1 mg/L (ppm) ± 0.1 mg/L (ppm)
(90 % of the final display value after) < 150 s according to
90
EN ISO 15839
(95 % of the final display value after) < 200 s according to
t
95
EN ISO 15839
Minimum approach flow No incident flow required
Interferences Non-sensitive to against hydrogen sulfide, chlorine, and ionogenic
substances
Allowed temperature
Measuring medium - 5 °C ... + 50 °C (23 ... 122 °F)
range
Storage/transport - 10 °C ... + 50 °C (14 ... 122 °F)
Allowed pH range of the
4 ... 12
test sample
Calibration procedure Factory calibration by means of IQMC (Intelligent Membrane Calibra-
tion) procedure. User calibration possible in water vapor-saturated air or air-saturated water.
Material PMMA, PVC and silicone
Working life 2 years guaranteed with authorized use
7 - 4 ba76014e03 05/2014
FDO®70x IQ ... Technical data

7.7 Technical data of the SC-FDO®701

Measuring ranges and
resolutions
Measurement accuracy
Repeatability ± 0.05 mg/L (ppm) according to EN ISO 15839
Response time t
D. O. partial pressure 0 ... 400 hPa
Measuring
Adjustable measuring range Resolution
mode
D. O. concentra­tion
0 ... 20.00 mg/L 0 ... 20.00 ppm
0.01 mg/L
0.01 ppm
D. O. saturation 0 ... 200.0 % 0.1 %
In the range Accuracy
< 1 mg/L (ppm) ± 0.05 mg/L (ppm) > 1 mg/L (ppm) ± 0.1 mg/L (ppm)
(90 % of the final display value after) < 60 s according to
90
EN ISO 15839
(95 % of the final display value after) < 80 s according to
t
95
EN ISO 15839
Minimum approach flow No incident flow required
Interferences Non-sensitive to against hydrogen sulfide, chlorine, and ionogenic
substances
Allowed temperature
Measuring medium -5 °C ... + 40 °C (23 ... 104 °F)
range
Storage/transport - 10 °C ... + 40 °C (-33 ... 104 °F)
Allowed pH range of the
4 ... 12
test sample
Calibration procedure Factory calibration by means of IQMC (Intelligent Membrane Calibra-
tion) procedure. User calibration possible in water vapor-saturated air or air-saturated water.
Material PMMA, PVC and silicone
Working life 6 months guaranteed for authorized use under the specified environ-
mental conditions
ba76014e03 05/2014 7 - 5
Technical data FDO®70x IQ ...
7 - 6 ba76014e03 05/2014
FDO®70x IQ ... Contact Information

8 Contact Information

8.1 Ordering & Technical Support

Telephone
Fax
: (937) 767-1058
Email
Mail: YSI Incorporated
Internet
When placing an order please have the following information available:
YSI account number (if available) Name and Phone Number Model number or brief description Billing and shipping address Quantity Purchase Order or Credit Card
: (800) 897-4151
(937) 767-7241 Monday through Friday, 8:00 AM to 5:00 PM ET
: environmental@ysi.com
1725 Brannum Lane Yellow Springs, OH 45387 USA
: www.ysi.com

8.2 Service Information

YSI has authorized service centers throughout the United States and Internationally. For the nearest service center information, please visit
www.ysi.com
directly at 800-897-4151.
When returning a product for service, include the Product Return form with cleaning certification. The form must be completely filled out for an YSI Service Center to accept the instrument for service. The Product Return form may be downloaded at www.ysi.com ‘Support‘ tab.
and click ‘Support’ or contact YSI Technical Support
and clicking on the
ba76014e03 05/2014 8 - 1
Contact Information FDO®70x IQ ...
8 - 2 ba76014e03 05/2014
FDO®70x IQ ... Indexes

9 Indexes

9.1 Explanation of the messages

This chapter contains a list of all the message codes and related message texts that can occur in the log book of the IQ S system for the FDO
®
70x IQ ... sensor.
Information on the contents and structure of the log book, and how to call it up, is given in the L
ENSORNET system operating manual.
IQ S
OG BOOK chapter of the
The last three digits of the message code identify the source of the message:
334 = SC FDO 700 (sensor cap SC-FDO
335 = SC FDO 701 (sensor cap SC-FDO
531 = FDO700IQ (sensor basis / component class,
adapter ADA)
ENSORNET
®
700)
®
701)

9.1.1 Error messages

Message code Message text
EA133x Meas. range exceeded or undercut
* Check process * Select other meas. range
EA233x Sensor temperature too high!
* Check process and application
EA333x Sensor temperature too low!
* Check process and application
EAP33x Measurement interfered
* SensCheck: Sensor cap is missing, leaky, depleted, or defective * Clean sensor and space between cap and sensor according to op.instructions * Screw on sensor cap securely * Replace sensor cap
EAP531 Measurement interfered
* SensCheck: Sensor cap is missing, leaky, depleted, or defective * Clean sensor and space between cap and sensor according to op.instructions * Screw on sensor cap securely * Replace sensor cap
ba76014e03 05/2014 9 - 1
Indexes FDO®70x IQ ...
Message code Message text
EC833x Sensor could not be calibrated,
sensor blocked for measurement Cause: instable signal * Check temperature adjustment * Check calibration conditions (see operating manual) * Repeat calibration
EC933x Calibration error, measurement disabled
Cause: Sensor cap is missing, leaky, depleted, or defective * Clean sensor and space between cap and sensor according to op.instructions * Screw on sensor cap securely * Repeat calibration * Replace sensor cap
EI133x Operational voltage too low
* Check installation and cable lengths, Follow installation instructions * Power supply module(s) overloaded, add power supply module(s) * Check terminal and module connections * Defective components, replace components
EI233x Operational voltage too low, no operation possible
* Check installation and cable lengths, Follow installation instructions * Power supply module(s) overloaded, add power supply module(s) * Check terminal and module connections * Defective components, replace components
ES133x Component hardware defective
* Contact service
9 - 2 ba76014e03 05/2014
FDO®70x IQ ... Indexes

9.1.2 Info messages

Message code Message text
IC133x Sensor has been successfully calibrated
* For calibration data, see calibration history
IC333x Factory calibration has been activated. Make sure the sensor operates
correctly.
IC433x Last valid calibration has been activated. Make sure the sensor oper-
ates correctly.
IC533x Invalid user calibration has been replaced by last valid user calibration.
Caution! Wrong measured values possible. Carry out a new successful calibration to make sure the sensor operates correctly.es correctly.
IC633x The invalid user calibration was replaced by the factory calibration.
Caution! Wrong measured values possible. Check whether the sensor operates correctly or carry out a new successful calibration.user cali­bration.
IC733x This message displays the result of the last check (see section 4.2.2)
IS133x This message displays the information stored in the sensor cap
(see section 1.2)
ba76014e03 05/2014 9 - 3
Indexes FDO®70x IQ ...

9.2 Status info

The status info is a coded piece of information on the current status of a sensor. Each sensor sends this status info to the controller. The sta­tus info of sensors consists of 32 bits, each of which can have the value 0 or 1.
Status info,
general structure
0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15
1 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 (general) 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 (internal)
16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31
The bits 0 - 15 are reserved for general information. The bits 16 - 21 are reserved for internal service information.
You obtain the status info:
via a manual query in the Einstellungen/Settings/Service/List of all
components menu (see system operating manual)
by an automated query
– from a superordinate process control (e. g. when connected to
the Profibus)
– from the IQ Data Server (see IQ S
ENSORNET Software Pack
operating manual)
The evaluation of the status info, e.g. in the case of an auto­mated query, has to be made individually for each bit.
Status info
Status bit Explanation
Bit 0 Component hardware defective Bit 1 Sensor cap missing, untight, depleted or defective Bit 2-31 -
9 - 4 ba76014e03 05/2014
1) The tissue in plants that brings water upward from the roots;
2) a leading global water technology company.
We're 12,500 people unified in a common purpose: creating innovative solutions to meet our world's water needs. Developing new technologies that will improve the way water is used, conserved, and re-used in the future is central to our work. We move, treat, analyze, and return water to the environment, and we help people use water efficiently, in their homes, buildings, factories and farms. In more than 150 countries, we have strong, long-standing relationships with customers who know us for our powerful combination of leading product brands and applications expertise, backed by a legacy of innovation.
For more information on how Xylem can help you, go to www.xyleminc.com
Xylem |' m|zīlə
YSI 1725 Brannum Lane Yellow Springs, OH 45387 Tel: +1 937-767-7241; 800-765-4974 Fax: +1 937-767-1058 Email: environmental@ysi.com Web: www.ysi.com
©Xylem Inc
Loading...