Welch Allyn QuietTrak User manual

Page 1
Arden, North Carolina USA
Manuals
Service Manual
E
T
R
A
S Y S / D I A mmHg
Operator’s Manual
Page 2
QuietTrak Serial Number: __________________________________
Date of Purchase:_________________________________________
Dealer Name: ____________________________________________
Telephone: ___________________________________________
Dealer Representative Name:_______________________________
Telephone: ___________________________________________
Welch Allyn/Tycos
Sales Representative:__________________________________

WARNING

This device not intended for use with a defibrillator,
for neonatal use, or with any intravascular fluid system.
The CE mark on this product indicates it has been
tested to and conforms with the provisions noted
within the 93/42/EEC Medical Device Directive.
0050
European Regulatory Manager
Medical Division Welch Allyn Ltd.
Navan, Co. Meath
Republic of Ireland
Welch Allyn/Tycos QuietTrak™ ABPM Operator’s Manual
Copyright 1993 Welch Allyn/Tycos Instruments, Inc. All rights reserved.
Patent No. 5,556,073
Page 3

Contents

i About the Manual iii Quick Guide to Hookup & Operation
1 Product & Report Features
1-2 Product Features 1-4 Specifications 1-6 The Report
2 Getting Started
2-2 Unpacking Checklist 2-4 Features at a Glance 2-7 Battery Requirements 2-8 Loading the Batteries 2-9 The Power-On Self-Test 2-10 Setting Date & Time
3 Choosing Program Options
3-2 Overview of Program Options 3-7 Basic Parameters —
3-15 Advanced Parameters — 3-19 Special Parameters — 3-23 Flow Charts for Parameter Programming
3-27 Office Record of Program Selections
Step-By-Step Programming Step-by-Step Programming Step-by-Step Programming
4 Patient Hookup
4-2 Before the Patient Arrives 4-4 Preparing the ABP Monitor for Use 4-12 Sizing the Cuff to the Patient 4-14 Positioning the Cuff & Microphone 4-21 Taking the Office Runs 4-24 Verifying the Setup 4-26 Beginning the Regular Runs 4-27 Positioning the Pouch & Tubing 4-29 Instructing the Patient
Page 4
5 When the Patient Returns
5-2 Disconnecting the ABP Monitor 5-3 Ending the ABP Monitoring Session 5-5 Reviewing the Data 5-7 Printing the Data
6 Clinical Recommendations
6-2 The Clinical Usefulness of Ambulatory
Blood Pressure Monitoring 6-5 Application of ABPM to Your Practice 6-6 Patient Benefits of ABPM
7 Troubleshooting Guide
7-3 Error Messages (in alphabetical order) 7-5 Error Codes 7-7 Special Messages 7-8 Trouble Chart
8 Service, Supplies & Maintenance
8-2 Calibration Check 8-6 Maintenance & Cleaning 8-7 Warranty 8-8 Welch Allyn/Tycos Service Centers 8-9 Ordering Supplies & Accessories
Page 5

About the Manual

This Operator’s Manual is designed to help you get the most out of your Tycos QuietTrak Ambulatory Blood Pres­sure Monitor. The manual has eight tabbed sections. The first page of each section outlines the contents so you can readily find what you need.
The first section, Product & Report Features, introduces you to the product and its capabilities. The next four sec­tions (Getting Started, Choosing Program Options, Patient Hookup and When the Patient Returns) take you step-by-step through the functional operation of the ABP monitor in a logical sequence.
The last three sections, Clinical Recommendations,
Troubleshooting Guide, and Service, Supplies & Main­tenance, are informative resources which offer trouble-
shooting or special help as needed.
Recommendations for First-Time Users.
First-time users may want to begin learning about the Tycos QuietTrak ABPM by reading Product & Report Features. Or, for hands-on instruction to begin using the ABP monitor, turn to Getting Started. Next, familiarize yourself with Patient
Hookup and When the Patient Returns.
Recommendations for More Experienced Users.
For day-to-day operation, experienced operators can refer to the convenient Quick Guide to Hookup & Operation found on page iii.
For those interested in learning and customizing the ABP monitor’s optional features, Choosing Program Options guides you step­by-step in selecting the options you want.
This manual is also available in German, French, Spanish, and Italian. Contact your local Welch Allyn/Tycos Service Center for more information.
i
Page 6

Style Features of the Manual

The manual is written in a logical sequence so that you can follow the events page-by-page and step-by-step. Each step is identi­fied with a bold subheading in the left margin...
(example of style features)
Setting the Date (example of a subheading)
...and detailed instructions follow. With this format, you can rap­idly skim over the parts you are familiar with and advance to the steps you need.
The instructions are enhanced with illustrations and graphics. Each instructional step is paired with a graphic representation of the corresponding ABP monitor display:
Yes No Next Set Date ?
When the Set Date ? message appears, you are ready. Press Yes in response to the Set Date ? message. A new message
will appear. Additional comments are placed in shaded boxes to draw atten-
tion to them. They are also identified by the following symbols appearing in the margin:
Important note or caution
!
Helpful hints, notes or suggestions
*
Conditional statements
?
(example of a box frame)
The boxed messages are important. Be sure to read them.
(example of a graphic)
!
At the conclusion of each section, there are suggestions to guide you where to go next. So, regardless of where you choose to begin, the manual conveniently directs you to related topics.
Suggestion for First-Time Users:
• Learn more about the ABP monitor in Product & Report Features, Section 1, and
• Practice using the monitor in Getting Started, Section 2.
ii
Page 7
Quick Guide to Hookup & Operation
The Quick Guide briefly outlines the steps needed to operate the Tycos QuietTrak ABPM and hook up the patient. The complete instructions are found in Sections 2, 3, and 4.
1. Select appropriate cuff size.
Carefully attach microphone over brachial artery 1 to 1.5 inches (2.5 to 3.8 cm) above elbow (use microphone anchor­ing pad). Be sure the labeled
side of the microphone faces away from the arm.
2. Wrap cuff on patient’s arm and secure it with cuff anchoring pad.
AAA
AAA
AAA
AAA
3. Load four new AAA batteries.
Wait while ambulatory blood pressure monitor completes Self-Test.
iii
Page 8
4. Get ready to set patient ID :
If Set Patient ID ? is not shown, respond to displays as follows:
Review Data ? No.
Print Data ? No.
Erase Data ? Yes.
Are You Sure ? Yes.
Yes No Next Set Patient ID ?
Up Down Next 071
Ê
Set parameters, including :
• Number of periods,
• Beginning time of each period, and measurement interval
• Maximum pressure,
• Display On or Off,
• Patient Activation Button On or Off
5. Set the patient’s ID.
Answer Yes. (Use Up or Down to rotate to
the number, then Next to pro­ceed to each space, and finally to proceed to next display.)
6. Connect ABP monitor to patient’s cuff assembly.
Take three or more office runs. Begin Regular Runs. Insert ABP monitor into pouch and position
pouch on patient with belt or shoulder strap.
iviv
Page 9
7. Instruct patient…
concerning Patient Diary, Patient Activation Button, how to can­cel readings, and to remain still during readings.
8. When patient returns
1. Remove cuff assembly and ABP monitor from patient.
2. Review diary entries with patient.
3. End ABP monitoring session by pressing all three buttons simultaneously.
4. Press Yes to begin reviewing data on the ABP monitor’s display.
5. Press Next to move to Print Data ? screen.
6. Connect ABP monitor to printer and print data.
Yes No Next Review Data ?
v
Page 10
Product & Report Featur es
The features and specifications of the Tycos QuietTrak Am­bulatory Blood Pressure Monitor explained in this section will help you understand its capabilities. You may not use every feature offered, but knowing what’s available will help you optimize your monitoring procedures.
At the end of this section, the pages of the report are illus­trated and explained so that you can get the most out of the ambulatory blood pressure data.
Page Topic
1-2 Product Features 1-4 Specifications 1-6 The Report
1
1-1
Page 11
Product Features

Accurate Measurement

What makes the Tycos QuietTrak Ambulatory Blood Pressure Monitor accurate? Developed and tested by a leader in the blood pressure industry, the Tycos QuietTrak ABPM sets unbeatable standards.
• Auscultatory method — the gold standard in the physician’s office.
• Tested scientifically by Tycos and validated by leading institu­tions, to meet accuracy standards set by The Association for the Advancement of Medical Instrumentation (AAMI) and The British Hypertension Society.
• Blood pressure measurements determined by this device are equivalent to those obtained by a trained observer using the auscultation method. Measurements are within the limits prescribed by ANSI for electronic sphygmomanometers.
• Verifiable accuracy. With the Calibration Check feature, you can compare the ABP monitor’s measurement with your of­fice manometer and be assured that the results are reliable.
• Linear deflation is adjustable at 2, 3, 4, 5, or 6 mmHg per heartbeat.
1-2
S Y S / D I A mmHg
Page 12

Patient-Friendly Design

Tycos developed a small, quiet ABP monitor with the patient’s comfort in mind, and there are no electrodes to wear.
• Lightweight — only 12.5 ounces (355 gms) with batteries.
• Small package — 4.5 x 3.4 x 1.6 inches (11.43 x 8.6 x 4.1 cm).
• Quiet cuff inflation and silent deflation make the Tycos QuietTrak ABPM ideal in the ambulatory environment be­cause it is hardly noticed. That’s so the ABP monitor can do its job while the patient does his.
• Patient Activation Button is easily accessible by the patient; an opening on the pouch guides the patient’s finger to the button by touch, without having to look.
• Real time clock on the ABP monitor’s display assists the patient with diary notations.

Programmability

• Programming is quick and uncomplicated, with simple three­button control and the LCD.
• Programmable periods — The automatic measurement inter­vals are selectable from 5 to 120 minutes, with up to four different time periods.
• The Start Button allows patient activation of readings, but can be disabled.
• Measurement Parameters can be modified from one ABP monitoring procedure to the next, and customized to the patient’s needs or the type of study being done.
• The parameters you select are stored in memory so that you won’t have to reselect ABP monitoring parameters from one procedure to the next.

Data Retrieval

• All measurement data can be quickly reviewed on the QuietTrak liquid crystal display
• Direct connection to Epson compatible printer for standard hard copy report.
• Optional PC program allows customized data and report generation via a variety of printers.
1-3
Page 13

Specifications

Power Requirements:

Four alkaline AAA (1.5V) batteries (nickel cadmium optional) One lithium 3V internal battery

Dimensions:

11.43 x 8.6 x 4.1 cm (4.5 x 3.4 x 1.6 in.)

Weight:

355 gms with batteries (12.5 oz.)

Leakage Current:

Meets AAMI, UL, and IEC specifications with regard to
leakage current.

Environmental Operating Requirements:

Temperature: 0° to 43.3°C (32° to 110°F) Humidity: 15% to 95% non-condensing Altitude: 150 m (500 ft.) below sea level to 1500 m (5000 ft.)
above sea level

Environmental Storage Requirements:

Temperature: -20° to 50°C (-4° to 122°F) Humidity: 15% to 95% non-condensing Altitude: 150 m (500 ft.) below sea level to 1500 m (5000 ft.)
above sea level

Measurement Ranges:

Systolic pressures 60-250 mmHg; Diastolic pressures 30-160 mmHg; Cuff pressures 0-300 mmHg; Heart rates 40-160 bpm. Up to 300 measurements may be taken on one set of AAA
alkaline batteries (based on a medium size cuff).

Measurement Cycles:

Able to store up to 300 measurements. Each measurement includes date/time/Systolic pressure/Diastolic
pressure/heart rate/initiation reason code/error code.

Cuff Inflation Time:

Typically 15 seconds; will vary with cuff size and target pressure.

Overpressure Detection:

Hardware: 300 mmHg
1-4
Page 14

Over 10 mmHg Detection:

Hardware: No longer than 180 seconds

Safety Feature for Deflation:

Can deflate from 250 mmHg to 20 mmHg in less than 4 seconds.

Deflation Rates:

Adjustable at 2, 3, 4, 5 or 6 mmHg per heartbeat

Accuracy:

Meets 1987 AAMI standards (at 3 mmHg/second deflation rate)

Pressure Accuracy:

± 3 mmHg or 2% of reading, whichever is greater.

Clock Accuracy:

± 2 seconds per day, or 1 min./month

Reliability:

50,000 typical cycles

Pressure Measurement Method:

Auscultatory (Diastolic based on fifth Korotkoff sound)

Automatic Measurement Intervals:

Selectable from 5 to 120 minutes, with up to four different
time periods.

Autozeroing:

Pressure transducer is autozeroed before each cycle.

Silent Deflation:

Continuous silent linear deflation

Liquid Crystal Display (LCD):

2 rows by 16 characters

Operator Control:

3-button control and LCD

Printer Specifications:

The QuietTrak ABP Monitor is also compatible with many other
Epson compatible printers equipped with the following serial interface specifications:
9600 BAUD, 8 data bits, 1 stop bit, No parity checking. X-on/X-off handshaking is used.
1-5
Page 15

The Report

The Tycos QuietTrak ABPM report consists of three or more pages. The actual size of the report will depend on the length of the ABP monitoring period and the number of readings taken. The report includes the following sections:
• Patient Information
• ABPM Parameter Settings
• Summary
• Hourly Averages
• Time Period Averages
• Readings Included in Averages
• Readings Excluded from Averages
The Sample Report
The basic format of the report is illustrated and detailed on the following pages.
An optional IBM compatible computer program is available which provides custom report features, and full ABP monitor programming capabilities. The customized reports which are
*
available with the program are not shown here. Ask your Welch Allyn/Tycos Dealer for more information.
1-6
Ambulatory Blood Pressure Report — The Heading Page
A
The Header appears at the top of each page and identifies
the date and time when the report is printed. The Patient ID is the patient’s unique identifying number.
B
Patient Information provides space for writing data such
as patient name, address, phone number, age, sex, height, and weight.
Additional space is provided for medication, physician name, reason for test and comments.
C
ABPM Parameter Settings list the settings programmed in
the unit during this ABP monitoring procedure.
D
The Footer identifies the serial number of the ABP monitor
that took the readings and the current revision number of the ABP monitor’s software as a reference. (It is 2.00, in this example.)
Page 16
(Sample of ABPM Report, Page 1)
1-7
Page 17
Ambulatory Blood Pressure Report — The Data Summary
A
Summary contains information about the Regular Run read-
ings, including the time and date of the first and last attempted Regular Run readings.
Number of readings expected includes all programmed
and manual readings. Number of readings used is the number of accepted
readings.
Minimum, Maximum, Average, Number of Systolic >
140 mmHg, Number of Diastolic Readings > 90: These are calculated from the accepted Regular Run readings.
B
Hourly Averages are presented in a table that includes data
from the Regular Run readings for each of the 24 one-hour inter­vals in a day. If ABP monitoring was conducted for more than a one-day period, then the hourly averages include the values ob­tained during that hour for all days of use.
The table contains these values:
Hour always begins on the hour (i.e., 07 AM) and ends after
the last minute of the hour. Num is the number of Regular Runs taken during each one-
hour period. Sys is the average Systolic value for each hour.
Dia is the average Diastolic value for each hour.
HR is the average heart rate value for each hour.
The data is also represented graphically to the right of the
data table. The average Systolic and Diastolic values are shown as “*”, and the average heart rate values are indicated with an “0”.
C
Time Period Averages provide the calculated statistics
from each ABP monitoring period.
1-8
Page 18
(Sample of ABPM Report, Page 2)
1-9
Page 19
Ambulatory Blood Pressure Report — The Readings
In the final page or pages of the report, results from each reading are listed separately. The length of the final report section varies with the total number of readings that were taken. (The example shown here was shortened so that each part can be explained on one page.)
Readings Included in Averages. The data from each reading are provided, including the day (➊), time (➋), Systolic and Dia­stolic pressures (➌), heart rate (➍), and an initiation code (➎) designating which type of reading was taken.
The data are also pr esented graphically for convenient reference.
1-10
Page 20
(Sample of ABPM Report, Page 3)
1-11
Page 21
A
Initiation Codes. Initiation code explanations are listed be-
low the table of included readings.
B
Readings Excluded from Averages. The data from any
excluded readings are presented in this table with the error code or retry reason code.
C
Error Codes Error code explanations are listed below the
table of excluded readings.
1-12
Page 22
(Sample of ABPM Report, Page 4)
1-13
Page 23
(Sample of ABPM Report, Page 5)
1-14
Page 24

Getting Started

This section identifies all of the product components in­cluded with the Tycos QuietTrak Ambulatory Blood Pres­sure Monitor. You will learn about the control keys of the ABP monitor and setup basics so that you can get started using the ABP monitor.
Page Topic
2-2 Unpacking Checklist 2-4 Features at a Glance 2-7 Battery Requirements 2-8 Loading the Batteries 2-9 The Power-On Self-Test 2-10 Setting Date & Time
2
2-1
Page 25

Unpacking Checklist

After you have unpacked the Tycos QuietTrak ABPM and the components, identify each item with the checklist that follows and inspect for missing items. Retain the shipping materials in case of shipping damage or for return, if necessary, to Welch Allyn/Tycos for repair or warranty service.
T
A
H
R
M
G
I
R
R
SIZE
X
A
M
m
c
5
.
1
3
M
R
A
T
H
G
I
R
A
Z
I
S
X
A
M
m
c
5
.
1
3
LEFT ARM
TERY
LEFT ARM
Y
R
E
T
R
A
E
2-2
Page 26

Content Checklist

A Tycos QuietTrak ABPM: The Tycos QuietTrak ABPM is a
compact, lightweight unit that is carried in a pouch and worn by the patient during the ABP monitoring period.
B Operator Manual: (This manual) Review and save this helpful
product reference.
C ABP monitor Pouch: The pouch covers the ABP monitor and
adapts to either a belt, worn at the patient’s waist, or a shoulder strap.
D Belt and Shoulder Straps: (One each) Adjustable to a wide
range of sizes.
E Medium Cuff with Bladder: (One each) F Large Cuff with Bladder: (One each) G Microphone and Pneumatic Assembly: (One each) Comes
pre-connected to the Medium size cuff with bladder.
H Batteries: (Five sets) Four AAA batteries are required for each
ABP monitoring procedure.
I Hookup Kits: (Five each) Each kit includes Patient Diary, mi-
crophone anchoring pad, and cuff anchoring pad.
J Cuff Anchoring Straps: (Three each) The straps stabilize the
cuff during the ABP monitoring period.
K Calibration T-connector: (One each) Allows connection of
Tycos QuietTrak ABPM to standard sphygmomanometer for calibration check and simultaneous arm measurement.
L Instrument Warranty Card: Fill this out today and return it to
the Welch Allyn/Tycos Service Center. This card validates your warranty.
M QuietTrak ABPM Carrying Case: (One each)
Note: Report any signs of shipping damage to the carrier. If
an item is missing or damaged, contact the Welch Allyn/ Tycos Service Center nearest you, or your distributor.
!
2-3
Page 27

Features at a Glance

The uncomplicated design of the Tycos QuietTrak ABPM is one of its best features. There are only a few parts to learn.

ABP monitor Anatomy

Top/Front View of the Tycos ABP Monitor. The three control buttons, LCD (liquid crystal display), cuff hose connector, and the microphone connector are shown here.
Three Control Buttons
2-4
S Y S / D I A mmHg
Microphone Connector
Cuff Hose Connector
LCD Display
Side View of the Tycos QuietTrak ABPM. To open
the battery compartment, apply gentle pressure with your finger and slide the door in the direction of the arrow. The door will unlatch and open.
To close the door, first press the door closed while sliding it in the direction of the arrow. Once the door is flush with the ABP monitor, allow the door to spring back in the opposite direction to latch.
Page 28
Setup Instructions
QuietTrak™Model No. 5100-01
Patient Setup Instructions
•Insert new AAA batteries prior to every setup.
•Carefully position and secure the microphone over the brachial artery, approximately 1 to 1.5 inches (2.5 to
3.8 cm) above the elbow. Position cuff with artery marker directly over microphone.
•Confirm proper setup by completing three office runs.
•Program the unit for the regular run mode.
•Instruct the patient to remain as motionless as possible during readings.
•Instruct patient on the use of the diary and patient activation button.
9302255EF7
Model Number
Serial Number
95 Old Shoals Road, Arden, NC 28704 USA 45P3245E
Bottom View of the Tycos QuietTrak ABPM. The label on the bottom of the ABP monitor lists the model number, serial number and a summary of setup instructions.
Data Output Connector
Back View of the Tycos QuietTrak ABPM. The connector on the back of the ABP monitor is for data output to a printer or an optional PC interface.

Turning the ABP Monitor’s Power On or Off

There is no On/Off switch to operate. The ABP monitor is On when batteries are loaded and Off when the batteries are removed.
2-5
Page 29

Description of Control Buttons

The Tycos QuietTrak ABPM has only three buttons, but with them, the operator can:
• Set the date and time
• Enter patient data
• Choose programming features
• Retrieve the ABP monitoring data.
What each button does depends on the command or choice that is shown on the display screen beneath it.
Three-Button Control. You will press a button to respond to a
Yes No Next Do Office Run 1 ?
message on the display screen. The Display Screen. The bottom
line of the display screen provides a question or message, while the top line of the display offers choices. The buttons correspond with the choices.
2-6
Page 30

Battery Requirements

Replace Batteries for Every Patient Study.
Install four new AAA alkaline batteries prior to each patient hookup. Fresh batteries provide the power to inflate the arm cuff 100-300 times during a one- or multi-day period. For maximum ABP monitoring periods, fresh batteries are required.
If the ABP monitor reaches a low battery state, it automatically discontinues the measurement cycle to reserve the power it will later need to transfer the data. Partially-used batteries shorten the ABP monitoring session. Therefore, attempting an ABP moni­toring session with used batteries is not recommended.
Data Storage Protection: An internal battery of the Tycos QuietTrak ABPM powers the data’s memory storage and protects the data even if the AAA batteries wear down. Data
!
is preserved even during battery replacement.
Concerning the Use of Rechargeable Batteries:
Rechargeable NiCad batteries can be used in place of alka­line batteries. With fully-charged NiCad batteries, 100 mea-
!
surement cycles are achievable with average case usage (Medium size cuff and 180 mmHg average inflation level). The length of the ABP monitoring session could vary with differ­ent cuff sizes or with batteries which are not fully charged.
Remove the Batteries to Store the ABP monitor.
At the conclusion of an ABPM procedure, we recommend that you print the report and remove the used batteries from the unit. This will eliminate the risk of attempting a new patient study with used batteries. If you want to wait to print the report at a later time, go ahead and remove the batteries; the internal lithium battery will retain all patient data until you retrieve it or intention­ally erase the data.

Internal Battery

An internal lithium battery powers the time, date, parameters, and data retention for at least five years. The internal battery en­sures that the ABP monitor won’t lose data when you remove or replace the AAA batteries.
When the internal battery needs replacement, return the ABP monitor to the nearest Welch Allyn/Tycos Service Center.
2-7
Page 31

Loading the Batteries

If batteries were previously loaded, remove them now. Open the battery compartment, tilt the ABP monitor, and slide the used batteries into your hand. Keep the used batteries separate from your fresh battery supply so that you don’t accidentally re-insert them into the ABP monitor. (Dispose of the used batteries ac­cording to local regulations.)
AAA
AAA
AAA
AAA
Slide two fresh batteries, tab end out, into the battery chamber illus­trated with a battery tab. Slide the two remaining fresh batteries, flat end out, into the battery chamber illustrated with a flat end of a battery. Close the battery compartment door.
After powering up, the ABP monitor will briefly display a sequence of messages as it performs a Self-Test of its functions.
2-8
Page 32

The Power-On Self-Test

The Tycos QuietTrak Ambulatory Blood Pressure Monitor checks itself each time the battery door is closed. Five Self-Test mes­sages appear quickly in sequence, and then the ABP monitor pauses to wait for a command. Each Self-Test message is shown and explained here.
LCD Test: This is a quick visual test for the operator. If all ele­ments of the LCD are working, a grid pattern appears.
Autozeroing
Autozeroing: The ABP monitor checks the pressure sensor and ensures that the starting point is at zero pressure.
Jul 05 03:22P Version 02.00
Date/Time/Version: Today’s date, the current time, and the software version of the ABP monitor appear briefly.
Patient ID Num 123 45 6789
Patient ID: The patient number identifies the patient data cur­rently stored in memory.
Unit contains 000 BP readings
BP Readings: The ABP monitor displays the number of blood pressure readings currently stored in memory.
Error Messages
If the ABP monitor’s Self-Test fails, an error message ap­pears. The Troubleshooting Guide in Section 7 explains
?
each message and offers guidance.
The ABP Monitor is Ready
After completing a Self-Test, the ABP monitor pauses and waits for a command.
2-9
Page 33

Setting Date & Time

You should only need to set the date and time when you first receive your Tycos QuietTrak Ambulatory Blood Pressure Monitor, or after the internal battery is replaced. After the date and time have been set, the internal battery keeps the clock and calendar running.
To set the date and time, or to recheck that the settings are cor­rect, proceed as follows:
Always check the date and time settings prior to patient hookup, and before any readings have been taken.
!
Yes No Next Set Patient ID ?
Set Patient ID ? This message will appear if there are no
patient readings currently stored in the unit’s memory. In this case, you are clear to proceed with setting the date and time. Respond with No to this question by pressing the middle button.
The display will change to Set Date ? and you can proceed to Setting the Date.
2-10
Setting the Date
Yes No Next Set Date ?
To set the date, press Yes in response to the Set Date ? mes­sage. A new message will appear.
Up Down Next Sep 10, 1993
The new display shows today’s date, and a cursor highlights the month.
Page 34
If the month is correct, press Next to move the cursor to the next field.
OR If the month is incorrect, press Up or Down to rotate through the
calendar and select the correct month. When the correct month is showing, move the cursor to the next field by pressing Next.
After setting the month, set the day and then the year. Press Next to return to the Set Date ? display.
Yes No Next Set Date ?
You’ve just set the date, so press Next to proceed. (Or, press Yes if you want to review what you just did or re-set
the date.)
Setting the Time
Yes No Next Set Time ?
Setting the time works like setting the date. Press Yes to set the time.
Up Down Next 11:08PM
The display shows the current time, and a cursor highlights the hour (11:08PM).
If the hour shown is correct, press Next to move the cursor to the minutes.
OR If the hour is incorrect, press Up or Down to move to a later or
earlier hour. When the hour is correct, press Next to move the cursor to the minutes.
After setting the hour, set the minutes. Press Next to return to the Set Time ? menu.
2-11
Page 35
Yes No Next Set Time ?
Press Next if you would like to continue using the ABP monitor. (Or, Press Yes if you want to set the time again, or review what
you just did.)
If the displayed date and time do not remain current in subsequent ABP monitor use, then the internal lithium bat-
?
tery may need replacement. Refer to Section 7 for trouble­shooting help.
Some suggestions for continued learning:
• Review the ABP monitor’s programmable options in Choos­ing Program Options, Section 3,
• Learn how to use the ABP monitor in Patient Hookup, Section 4.
2-12
Page 36

Choosing Program Options

This section introduces the programmable parameters of the Tycos QuietTrak Ambulatory Blood Pressure Monitor, and explains step-by-step how to select them.
Flow charts illustrate the programming sequences, and can be used like road maps as you review or change the pro­gram of the ABP monitor.
The last page of the section provides space to record your programming selections for future reference.
Page Topic
3-2 Overview of Program Options 3-7 Basic Parameters —
Step-By-Step Programming
3-15 Advanced Parameters —
Step-by-Step Programming
3-19 Special Parameters —
Step-by-Step Programming
3
3-23 Flow Charts for Parameter Programming 3-27 Office Record of Program Selections
3-1
Page 37
Overview of Program Options
The Tycos QuietTrak ABPM is designed with flexible program options (parameters) to fit typical needs, as well as special clinical cases so you can adapt the ABP monitor to your needs. There are three groupings of parameters.
Basic Parameters typically are reviewed and modified from one ABP monitoring procedure to the next, customized to the patient’s needs or the type of study being done.
Advanced Parameters specify the deflation rate of the cuff, turn the dynamic pressure feature On or Off, and define the criteria for identifying data that fall outside specified levels (retry parameters).
The Advanced Parameters are preset at values optimum for most clinical conditions. To ensure maximum perform­ance of the ABP monitor during typical procedures, use the
*
default settings.
Special Parameters establish the routine of the ABP monitoring procedure, such as the format in which time is displayed, and the language of the ABP monitor’s messages. Generally, these re­main the same from one procedure to the next, unless there is a specialized procedural need.
3-2
Concerning Defaults: A default is the setting the ABP monitor has as it arrives from the factory. Each parameter has a default. If you change a parameter, your selection is
*
continually stored by the ABP monitor’s internal memory until it is changed again or until the internal battery fails.
If your ABP monitor is serviced at a service center, the ABP monitor would probably be returned to you with the default settings. For that reason, we suggest you keep a record of your selections. Space is provided at the end of this section to write down your parameter settings.
Page 38
Basic Parameters — Overview

Basic Parameters give you flexibility to customize the ABP

monitoring periods and frequency of measurements to the patient’s routine. The Basic Parameters let you decide whether or not you want the patient to activate readings or see the blood pressure measurements on the display. And, the maxi­mum cuff inflation pressure can be set in the Basic Parameters.
Because these parameters often vary from patient to patient, they are easy to review and change in the Basic Parameters.
Number of periods
Number of ABP monitoring periods; can be 1, 2, 3, or 4. Default is 2.
Beginning Time
Time of day when each period begins. Default is 7 a.m. for first period, 11 p.m. for second.
Interval
Interval of automatic measurements taken during each period. Interval is determined by the length of time between readings. Options: Manual, 5, 10, 15, 20, 30, 60, or 120 minutes. Default is 15 for first period; 30 minutes for second period.
Maximum Pressure
Maximum cuff inflation pressure, selectable in 10 mmHg incre­ments within 160-270 range. Default is 200.
Display
Turn measurement readings On or Off to patient’s view. Default is On.
Patient Activation Button
Patient Activation Button allows the patient to initiate a meas­urement with the push of a button. Turn button On or Off. Default is On.
3-3
Page 39
Advanced Parameters — Overview
The Advanced Parameters specify the deflation rate of the cuff, turn the dynamic pressure feature On or Off, and define the crite­ria for identifying data that fall outside specified levels. These parameters are adaptable when needed for specialized studies.
The Advanced Parameters are preset at values optimum for most clinical conditions. To ensure maximum perform­ance of the ABP monitor during typical procedures, use the
!
default settings.
Deflation Rate
Set deflation rate of cuff. Unit of measure is mmHg/second until pulse rate is detected, after which it becomes mmHg/heartbeat. Range is 2-6. Default is 3.
Dynamic Pressure
Turn dynamic pressure function On or Off. If On, cuff inflates to a pressure of 30 mmHg greater pressure than the previous Systolic measurement, otherwise, unit inflates to Maximum Inflation Pres­sure. Default is On.

Retry Parameters

The following retry parameters allow you to set boundaries for acceptable measurements. If one or more of the boundaries is violated, the unit will automatically trigger a retry reading. The purpose of the retry parameters is to identify and reject readings which are beyond the scope of “believable”, and do not reflect a true blood pressure measurement. A new reading will then be taken to replace the unacceptable data.
Max Allowable Systolic Pressure
Maximum allowable Systolic pressure can be set from 60-250 mmHg in increments of 10. Pressures exceeding this generate a retry request. Default is 250.
3-4
Page 40
Min Allowable Systolic Pressure
Minimum allowable Systolic pressure can be set from 60-240 mmHg in increments of 10. Pressures below this generate a retry request. Default is 60.
Max Allowable Change in Systolic Pressure
Maximum allowable change in Systolic pressure from one valid reading to the next can be set from 30-150 mmHg in increments of 10. Systolic pressure changes exceeding this generate a retry request. Default is 150.
Max Allowable Diastolic Pressure
Maximum allowable Diastolic pressure can be set from 40-160 mmHg in increments of 10. Pressures exceeding this generate a retry request. Default is 160.
Min Allowable Diastolic Pressure
Minimum allowable Diastolic pressure is selectable from 30-160 mmHg in increments of 10. Pressures below this generate a retry request. Default is 30.
Max Allowable Change in Diastolic Pressure
Maximum allowable change in Diastolic pressure from one valid reading to next can be set from 20-100 mmHg in increments of
10. Changes exceeding this level generate a retry request. De­fault is 100.
Max Allowable Pulse Pressure
Maximum allowable pulse pressure (difference between Systolic and Diastolic pressures) is selectable from 10-150 mmHg in in­crements of 10. Default is 150.
Min Allowable Pulse Pressure
Minimum allowable pulse pressure is selectable from 10-150 mmHg in increments of 10. Default is 10.
3-5
Page 41
Special Parameters — Overview
With the Special Parameters, you select your preference for the time format, the ABP monitor’s language, and the maximum length of the ABP monitoring period. You probably won’t change these parameters from one procedure to the next.
For best results, we recommend that the Auto Dump feature be left On unless there is a specialized procedural need to turn it Off.
*
ABP Monitor Language
This specifies the language used in ABP monitor’s display. Lan­guages available are: English, Italian, French, German and Spanish. Default is English.
Time Display
Time can be represented as a.m./p.m. time, or 24-hour time. Default is a.m./p.m.
Auto Dump
If Auto Dump feature is On, the cuff rapidly deflates at 20 mmHg below the Diastolic pressure. If Off, slow deflation continues until the cuff pressure is only 20 mmHg. This allows simultaneous same-arm efficacy studies. Default is On.
One Day Maximum
If One Day Maximum is On, the ABP monitor will discontinue taking measurements (automatic and patient-activated) when 24 hours have elapsed from the first Regular Run measurement. The default is On.
3-6
Page 42
Basic Parameters — Step-By-Step Programming
The following information will guide you step-by-step in selecting the Basic Parameters.
Ready the ABP Monitor
Load four new AAA batteries into the ABP monitor and wait as the ABP monitor completes its Self-Test.
One of three messages appears following the Self-Test. The mes­sage that comes up reflects what the ABP monitor was doing when it was last used.
One of Three Messages Appears Following the Self-Test:
Yes No Next Set Patient ID ?
Set Patient ID ? This message will appear if there are no
patient readings currently stored in the unit’s memory. In this case, you are clear to proceed with setting the Basic Parameters.
Yes No Next Review Data ?
Review Data ? This message will appear if data from a pa-
tient study remains in memory. The data will remain in the memory until it is intentionally erased.
Yes No Next Resume Readings ?
Resume Readings ? This message will appear if a 24-hour
ABP monitoring period has been interrupted. Most often this would occur if a patient removes the batteries, or opens and closes the battery door, during the 24-hour period.
This message will be displayed for one minute. If no commands are made, the unit will automatically resume the Regular Run cycles.
3-7
Page 43
If the Review Data ? or Resume Readings ? mes­sages appear following the Self-Test, it will be necessary to erase existing Regular Run patient data from memory prior
!
to changing the Basic Parameters. Refer to Section 4, page 4-5 for instructions.
You are allowed to change Basic Parameters if only Office Run data is stored.
Once you have printed out and erased any existing Regular Run data from memory, you can proceed with programming the Basic Parameters:
Yes No Next Set Patient ID ?
Set Patient ID ? Respond with No to this question by press-
ing the middle button. Then answer the sequence of questions that follow as indicated here:
Set Date ? Choose No. Set Time ? Choose No.
The display will change to Set Parameters ? and you can proceed to Reviewing or Selecting Basic Parameters.
3-8
Page 44
Reviewing or Selecting Basic Parameters
Hint: If you lose your place and need help getting back on
track, you can press either the No or Next buttons multiple times until you see Set Parameters ? in the display.
*
The Basic Parameter Flow chart provides a helpful diagram of the sequence of displays, and can be used like a road map. See page 3-24.
Yes No Next Set Parameters ?
Set Parameters ? When this display appears, you are ready
to review current parameter settings and either accept them or change them.
To review the first parameter, respond with Yes by pressing the left button. The display will change to the Set Periods ? message.
OR
The Set Periods Parameter
Yes No Next Set Periods ?
Set Periods ? This allows you to select the number of ABP
monitoring periods (1, 2, 3, or 4), define when those periods begin, and select the frequency of the automatic blood pressure measurements taken during each period.
To omit setting the periods and go on to the next Basic Param­eter, press No or Next and proceed to Setting the Maximum Pressure further in this section,
OR Press Yes to review or change the periods.
3-9
Page 45
Up Down Next # of Periods 2
Reviewing or Selecting the Number of Periods. This allows you to select the number of ABP monitoring periods (1, 2, 3, or
4). 2 is the default setting and means that two ABP monitoring periods are selected.
2 is a good choice if you plan to vary the frequency of measure­ments during two periods, such as daytime and nighttime, or working hours and non-working hours.
3 and 4 periods give you even more flexibility to customize the periods to the patient’s day: at home, at work, during leisure, and during sleep.
Select 1 period if you want an unchanging rate of automatic measurement.
Selecting the Number of Periods. To change the number shown on the ABP monitor’s display, press the Up or Down buttons to rotate through the sequence of 1 to 4. (When you r each the last number of the sequence, the sequence rolls over to the first num­ber.)
When the number of periods you want to select is displayed, press Next to accept the number. The message will change.
3-10
Up Down Next Prd 1 07AM 15m
Defining Each Period. You will need to define the ABP monitor­ing periods selected. First, you will set the beginning time of the period, then you will enter the time interval between blood pres­sure measurements taken automatically during that period.
Up Down Next Prd 1
07AM 15m
Setting the Beginning Time of the Period. A cursor high­lights the beginning time for the first period, Prd 1. Period 1 is customarily set at the patient’s normal waking time. The default setting for the beginning time is shown here in a.m./p.m. The end time of a period is automatically determined by the start time of the next period.
Page 46
Note: Time can be presented as a.m./p.m. or as 24-hour time. This is an option in Special Parameters, covered later in this section.
*
To adjust the period’s beginning time for Period 1, press the Up or Down buttons to move in one-hour increments to a later or earlier time.
Note: Period 1 always specifies the start time of the report.
*
To accept the time currently shown, press Next. The cursor will move to highlight the next field, the frequency setting.
Up Down Next Prd 1 07AM
Setting Measurement Frequency/First Period. Measurement frequency is the time interval, in minutes, between automatic readings. The choices are: MANual, 5, 10, 15, 20, 30, 60, or 120 minutes.
The default for the first period is 15. That means that the ABP monitor will initiate a blood pressure measurement every 15­minute interval during this period.
If MAN (manually-requested readings) is chosen, the automatic measurements are cancelled during the period, but the patient can initiate measurements by pressing the Patient Activation Button (if the patient activation button is not turned off).
To select the frequency of measurements, press the Up or Down button to move through the sequence until the one you want appears.
After selecting the measurement frequency, or if you want to accept the frequency currently shown, press Next.
15m
3-11
Page 47
Up Down Next Prd 2
Setting Measurement Frequency/Next Period(s). If two or more ABP monitoring periods were selected, then the second period will be displayed.
Define this period and any others selected by following the same steps you used with the first period.
Repeat the same steps to define the third and fourth periods, if needed.
Note: If the Up or Down button seems to be frozen be- cause the time won’t change, you may be trying to select the beginning time of a previous period. A design feature
*
prevents the periods from overlapping.
Up Down Next Maximum Pres 270
Setting the Maximum Pressure. This determines the maximum cuff inflation pressure, selectable in 10 mmHg increments within a range of 160-270 mmHg. The ABP monitor will never inflate above this setting.
08AM 15m
3-12
Reducing the maximum cuff pressure to match the patient’s up­per range more closely is a comfort benefit to the patient. How­ever, be careful not to set the maximum too low, so that normal blood pressure fluctuations cannot be accommodated. The Dy­namic Pressure feature (page 3-18), if utilized, will automatically adjust the inflation level to track the patient’s previous Systolic measurements. This will enhance patient comfort.
To change the pressure, press the Up or Down buttons and scroll through the pressures until the desired one appears.
Press Next to accept the displayed pressure.
Page 48
Off On Next Display On
Turning the Display On or Off. This turns the ABP monitor’s display of blood pressure readings On so they are visible to the patient, or Off so the patient does not see them.
Some patients may react to seeing their blood pressure readings and this could affect their results. For other patients, having the measurements visible may be a tool to better understand their health and to gain compliance with taking medication. The choice to display or not display may be a one-time decision to suit the clinical practice, or it can be decided patient-to-patient.
Note: Even if the display is Off, the Office Runs (test meas­urements) that follow the patient hookup will be displayed, and date and time will be displayed throughout the ABP
*
monitoring period.
Press On or Off according to your display preference. Then, press Next.
Off On Next Start Button On
Turning the Patient Activation Button On or Off. If On, the patient can initiate a blood pressure measurement by pressing the green button. If Off, the patient cannot initiate readings, and only the programmed readings are taken.
By pressing the green Patient Activation Button, the patient can cancel a measurement attempt. This feature is available even if the Patient Activation Button is Off. Normal operation will resume at the next programmed cycle time.
The Patient Activation Button can be a helpful tool with a patient who experiences transient symptoms. The patient can activate a measurement during the symptom if this feature is On. Special studies may call for measurements taken only at specified inter­vals, so the Patient Activation Button can be turned Off.
Press On or Off according to your preference. Then, press Next.
3-13
Page 49
Yes No Next Set Adv Param?
When you have completed reviewing or selecting the Basic Parameters, you can continue with the Advanced Parameters, or you can exit the parameter program.
If you are going to quit now, remove and dispose of the AAA batteries before storage so that no one attempts an ABP monitoring period with used batteries.
!
OR To exit now and return to the main menu, press Next in re-
sponse to Set Adv Param ?, then press Next again in response to Set Param ?
OR To review the Advanced Parameters, press Yes in response to
Set Adv Param ? and continue on to the next page.
3-14
Page 50
Advanced Parameters — Step-by-Step Programming
The Advanced Parameters make it possible to adapt the ABP monitor to very specialized studies. However, to ensure maximum performance in typical procedures do not change the defaults.
Yes No Next Set Adv Param ?
Reviewing or Selecting Advanced Parameters. At this dis­play, you are ready to review the Advanced Parameter settings and accept or change them.
To review the first one, press Yes.
The flow chart of Advanced Parameter Programming is an optional reference that might be helpful as you work through the display sequence. See page 3-25.
*
Up Down Next Deflation Rate 3
Setting the Deflation Rate. This determines the deflation rate of the blood pressure cuff. The unit of measure is mmHg per second until pulse rate detection, after which it becomes mmHg per heartbeat.
The range of rates is 2-6. The default setting is 3, and recom­mended for best results.
Press Up or Down to raise or lower the rate. When the desired rate appears on the display, press Next to accept it and continue.
The Retry Parameters
Up Down Next Set Retry Param ?
Set Retry Param ? The Retry Parameters set criteria for
identifying inaccurate measurement data. When a measurement meets the criteria, the ABP monitor attempts to retry the meas­urement.
3-15
Page 51
There are eight Retry Parameters to review and accept or change. The directions apply to all eight parameters. Each Retry Parameter is defined in the same sequence as it will appear on the display.
Selecting the Eight Retry Parameters. For each of the Retry Parameter displays, press Up or Down to raise or lower the setting. When the desired setting is shown in the display, press
Next to accept it and continue to the next.
Up Down Next Reject Sys > 250
The Maximum Allowable Systolic Pressure can be set from 60-250 mmHg in increments of 10. The default is 250.
Up Down Next Reject Sys < 60
The Minimum Allowable Systolic Pressure can be set from 60-240 mmHg in increments of 10. The default is 60.
Up Down Next Reject Sys Chg > 150
3-16
The Maximum Allowable Change in Systolic Pressure from one valid reading to the next can be set from 30-150 mmHg in increments of 10. The default is 150.
Up Down Next Reject Dia > 160
The Maximum Allowable Diastolic Pressure can range from 40-160 mmHg in increments of 10. The default is 160.
Up Down Next Reject Dia < 30
The Minimum Allowable Diastolic Pressure is selectable from 30-160 mmHg in increments of 10. The default is 30.
Page 52
Up Down Next Reject Dia Chg > 100
The Maximum Allowable Change in Diastolic Pressure from one valid reading to the next ranges from 20-100 mmHg in incre­ments of 10. The default is 100.
Up Down Next Reject PP > 100
The Maximum Allowable Pulse Pressure (pulse pressure is the difference between Systolic and Diastolic pressures) is select­able from 10-150 mmHg in increments of 10. The default is 150.
Up Down Next Reject PP < 10
The Minimum Allowable Pulse Pressure is selectable from 10-150 mmHg in increments of 10. The default is 10.
Yes No Next Set Retry Param ?
Completing the Retry Parameters. When the last Retry Parameter is selected and this display appears, the Retry Pa­rameters are completed.
Error codes 2, 4, 5, 8, 9, 10, 11, 12 and 13 will cause the ABP monitor to attempt a retry reading. For Manual and period intervals of 5 minutes, there are no retries. For period inter­vals of 10, 15 or 20 minutes, there is a maximum of one
*
retry per error. For period intervals of 30, 60 and 120 min­utes, there is a maximum of two retries.
Press Next to complete the one remaining Advanced Parameter.
3-17
Page 53
Off On Next Dynamic Pres On
Setting the Dynamic Pressure. If the Dynamic Pressure is turned On, the ABP monitor will inflate the cuff to 30 mmHg greater pressure than the previous Systolic measurement. This will enhance patient comfort. If Off, the cuff will inflate to the specified Maximum Inflation Pressure for every measurement. The default is On.
Make your selection and then press Next to accept it.
Yes No Next Set Adv Param ?
Completing the Advanced Parameters. When you have com­pleted reviewing or selecting the Advanced Parameters, you can exit the parameter program and return to the main menu to begin a patient hookup.
To quit using the ABP monitor, simply remove the batteries and put it away; there is no Off switch.
If you are going to quit now, remove and dispose of the AAA batteries and store the ABP monitor empty, so that no one
!
attempts an ABP monitoring period with used batteries.
3-18
OR To exit now and return to the main menu, press Next in re-
sponse to Set Adv Param ?, then press Next again in re­sponse to Set Param ?
OR To review the Special Parameters, continue on to the next page.
Page 54
Special Parameters — Step-by-Step Programming
The Special Parameters establish the routine of the ABP monitor­ing procedure, such as the time format, the ABP monitor’s lan­guage, and the maximum length of the ABP monitoring period. After your initial selection, you probably won’t need to change these parameters from one procedure to the next.
Readying the ABP Monitor. The Special Parameters are not found in the ABP monitor’s main menu, but are easily accessible.
To get to the Special Parameters, load four AAA batteries (or just open the battery compartment if the batteries are already loaded) and close the battery compartment.
As the ABP monitor begins its Self-Test, press the left button and hold it down.
Jul 05 03:22P Version 00.56
Continue holding the button down until the date/time/version display disappears and the Change Language ? menu ap­pears. Now release the button.
Yes No Next Change Language ?
The ABP monitor display shows the first of the Special Parame­ters, Change Language ? You are now ready to review and select the Special Parameters.
The flow chart of Special Parameter Programming might be a useful reference as you work through the reviewing and selecting sequence. See page 3-26.
*
3-19
Page 55
Reviewing or Selecting Special Parameters
Yes No Next Change Language ?
Selecting the ABP Monitor and Report Language. This speci­fies the language used in the ABP monitor’s display and report. There are five languages available (and corresponding Operator Manuals, too):
English, Italiano, Français, Deutsch, and Español. Press No or Next to continue to the next Special Parameter, or
press Yes if you want to change the language from the one cur­rently shown.
Once you select a language and press Next, the ABP moni­tor will display all messages in that language, so choose a language that is familiar to you.
!
If you mistakenly enter a foreign language, keep pushing the green button until Change Language ? appears again.
Up Down Next English
3-20
Press Up or Down to review the other choices or change the selection.
When the language you want is shown in the display, press
Next to accept it and continue to the next parameter.
Yes No Next AM/PM Yes
Selecting the Time Display Format. Time can be represented as a.m./p.m. time, or as 24-hour time.
The display will say AM/PM Yes if a.m./p.m. time is selected, or AM/PM No if 24-hour time is selected.
To change the display, press No for 24-hour time, or press Yes for a.m./p.m. time.
Press Next to accept the time display format as shown, then continue to the next parameter.
Page 56
Yes No Next Check Calib ?
The Calibration Check Feature. The ABP monitor’s pressure measurement accuracy and microphone sensitivity can be veri­fied with this feature. This is explained in Service, Supplies & Maintenance, Section 8.
This is not one of the Special Parameters, so press Next to continue to the next parameter.
Off On Next Auto Dump On
Selecting Auto Dump. Auto Dump initiates rapid cuff deflation at 20 mmHg below the Diastolic pressure. This is a comfort fea­ture to the patient, because the arm gets quicker relief from the cuff pressure. Auto Dump is On in the default.
If the automatic dumping feature is Off, slow deflation continues until the cuff pressure is only 20 mmHg. The ability to turn off Auto Dump is useful in the special case of simultaneous, same­arm efficacy studies. In most cases, it is preferable to turn on the Auto Dump feature.
Press On if you need to turn Auto Dump On, or press Off if you want to turn it Off.
After making your selection, press Next to continue.
Off On Next 1 Day Max On
Selecting the One Day Maximum Feature. If One Day Maxi­mum is On, the ABP monitor will discontinue taking measure­ments (automatic and patient-activated) when 24 hours have elapsed from the first measurement. The default is On.
If One Day Max is Off, the ABP monitor continues taking measure­ments until the batteries are low. If multi-day ABP monitoring is planned, then choose Off.
Press the Off or On button to choose On or Off.
3-21
Page 57
Yes No Next
Exiting the Special Parameters Menu. Press Yes if you have made the selections you wanted and are ready to return to the main menu or quit. Exiting from here continues the Power-On sequence of Self-Tests, and entry into the main menu.
OR Press No or Next if you would like to review the Special Param-
eters again.
If you are going to quit now, remove and dispose of the AAA batteries before storage. (Removing the batteries now pre­vents someone from attempting to begin the next ABP
!
monitoring period with used batteries.)
Exit ?
3-22
Page 58

Flow Charts for Parameter Programming

Flow charts are diagrams that show the sequence of messages or questions that you follow while operating the ABP monitor.
The flow charts can be used like road maps to illustrate the path­ways of selecting and reviewing the parameters.
You may find it helpful to use the flow charts to accompany your progression through this section’s written instruction. The flow charts can also be used as a shortcut to the instructions to help you reset a particular parameter at a later date.
Main Menu Flow Chart
Yes
No Next
Set Patient ID ?
No/Next
No/Next
Yes
Set Date ?
Set Time ?
No Next
No/Next
No Next
No/Next
No NextYes
Erase Data ?
Yes
No NextYes
Print Data ?
No/Next
No/Next
No Next
Set Parameters ?
Yes
No Next
Do Office Run 1 ?
Basic Parameters
No/Next
No/Next
If memory contains readings
Yes Review Data ?
Yes Begin Reg Runs ?
No/Next
No NextYes
No/Next
No NextYes
If no readings are in memory
3-23
Page 59
Basic Parameter Programming Flow Chart
Main Menu
No/Next
Yes
Up
Down Next
# of Periods 2
No Next
Set Parameters ?
Yes
Up Down Prd 1 07AM 15M
Up Down Prd 2 05PM 20M
Up Down Prd 3 08PM 30M
Up Down Prd 4 11PM 60M
Next
Next
Next
Next
Next
Next
Next
Next
Next
Yes
Next
Next
Next
Yes
No Next
Set Periods ?
No/Next
Up
Down Next
Maximum Pres 200
160-270
Next
(10 mmHg Increments)
Yes Set Adv Param ?
Off Start Button On
Off Display On
No/Next
No Next
Yes
Next
On Next
Next
On Next
3-24
Advanced Parameters
Page 60
Advanced Parameter Programming Flow Chart
Yes
No Next
Set Adv Param ?
Yes
Down Next
Deflation Rate 3
(2, 3, 4, 5, 6 mmHg/sec.)
Next
Yes
Up Down Reject Sys >250
Next
60-250 (10 mmHg Increments)
Next
Yes
No Next
Set Retry Param ?
Off Dynamic Pres On
Up Reject PP <10
10-150 (10 mmHg Increments)
Next
On NextUp
No/Next
Next
Down Next
Next
Up Down Reject Sys <60
Next
60-240 (10 mmHg Increments)
Up Down Rej Sys Chg >150
Next
30-150 (10 mmHg Increments)
Up Down Reject Dia >160
40-160 (10 mmHg Increments)
Next
Up Down Reject PP >100
10-150 (10 mmHg Increments)
Up Down Rej Dia Chg >150
20-100 (10 mmHg Increments)
Up Down Reject Dia <30
30-160 (10 mmHg Increments)
NextNext
Next
NextNext
Next
NextNext
3-25
Page 61
Special Parameter Programming Flow Chart
(To access, close battery door and hold down left button during Self-Test.)
Main Menu
Yes
No Next
Yes
Change Language ?
No/Next
Up Down English
English Italiano Français Deutch Español
Pressure=+011 Microphone=0512
Next
Next
Yes No AM/PM Yes
Yes No Check Calib ?
Yes
Press any Key
Next
No/Next
No/Next
Yes No Exit ?
Next
Off On 1 Day Max On
Next
Off On Auto Dump On
Yes
NextNext
NextNext
Next
3-26
Page 62
Office Recor d of Program Selections
Keep a record of your program selections here. Note that the ABP monitor returns itself to the default settings whenever it is serviced or the internal battery is removed. Because of this, it is helpful to have this reference handy to remember your selections.
Basic Parameters
Parameters Choices Selection
Number of Periods
Beginning Time
Frequency Manual, 5, 10, 15, 20, 30, 60,
Maximum Pressure
Display On or Off. Default is On. Patient
Activation Button
1, 2, 3, or 4. Default is 2.
Default is 7 a.m. for first period; 11 p.m. for second.
or 120 minutes. Default is 15 for first period; 30 for second period.
160-270 mmHg in increments of 10. Default is 200.
On or Off. Default is On.
Period 1: Period 2: Period 3: Period 4: Period 1: Period 2: Period 3: Period 4:
3-27
Page 63

Advanced Parameters

Parameters Choices Selection
Deflation Rate Range is 2-6. Default is 3. Max Allowable
Sys Pressure Min Allowable
Sys Pressure Max Allowable
Change in Sys Pressure
Max Allowable Diastolic Pressure
Min Allowable Diastolic Pressure
Max Allowable Change in Dias Pressure
Max Allowable Pulse Pressure
Min Allowable Pulse Pressure
Dynamic Pressure
60-250 mmHg in increments of 10. Default is 250.
60-240 mmHg in increments of 10. Default is 60.
30-150 mmHg in increments of 10. Default is 150.
40-160 mmHg in increments of 10. Default is 160.
30-160 mmHg in increments of 10. Default is 30.
20-100 mmHg in increments of 10. Default is 100.
10-150 mmHg in increments of 10. Default is 150.
10-150 mmHg in increments of 10. Default is 10.
On or Off. Default is On.
3-28
Page 64

Special Parameters

Parameters Choices Selection
ABP Monitor Language
Time Display a.m./p.m. time, or 24-hour time
Auto Dump On or Off. Default is On. One Day
Maximum
English, Italiano, Français, Deutsch, and Español
On or Off. Default is On.
3-29
Page 65

Patient Hookup

This section will guide you step-by-step through the hookup procedure. The illustrations and explanations will help you learn the routine for the ABPM hookup procedure.
Page Topic
4-2 Before the Patient Arrives 4-4 Preparing the ABP Monitor for Use 4-12 Sizing the Cuff to the Patient 4-14 Positioning the Cuff & Microphone 4-21 Taking the Office Runs 4-24 Verifying the Setup 4-26 Beginning the Regular Runs 4-27 Positioning the Pouch & Tubing 4-29 Instructing the Patient
4
NOTE: If you are already familiar with the hookup and
don’t need the details provided in this section, refer to the Quick Guide to Hookup. This is on page iii, right before
*
Section 1.
Note to First Time Users: It is important to have a basic understanding of the ABP monitor before beginning a pa­tient hookup. If you have not worked with the ABP monitor
*
yet, you need to read and work through Getting Started, Section 2 before continuing in this section.
Also, if you want to customize the ABP monitoring proce­dure but are unfamiliar with selecting the ABP monitoring parameters, we suggest that you work through Basic Pa­rameters, in Section 3, before beginning a patient hookup.
4-1
Page 66

Before the Patient Arrives

When scheduling the procedure, inform the patient in general how ambulatory blood pressure ABP monitoring works, what to expect, and what to wear.
Advising the Patient What to Wear
Men — Suggest garments with loose-fitting sleeves or short
sleeves. Men may prefer to wear an undershirt beneath a dress shirt.
Women — Long-sleeve sweaters or dresses should not be worn on the day of the hookup. Blouses with loose-fitting sleeves are suggested. Loose-fitting cardigan sweaters may also be worn.
Advising the Patient of Activity Restrictions
While the Tycos QuietTrak Ambulatory Blood Pressure Monitor is designed for ambulatory use, the patient should not plan to swim or bathe during the ABP monitoring period. Also, the vibrations from power tools or heavy equipment may disrupt the ABP moni­toring attempts and should be avoided.
Carefully advise the patient to plan a day free of these activities.
4-2
Preparing the Items Needed for ABPM Hookup
Prior to the patient’s arrival, gather the items needed for the hookup procedure.
Page 67

Hookup Checklist:

• Tycos QuietTrak ABPM
• Cuff assemblies (have both Medium and Large sizes avail­able, to ensure you can properly fit the patient), including an anchoring strap
• ABP monitor pouch and belt or shoulder strap
• Four new AAA batteries
• Hookup kit, including: Patient Diary, microphone anchoring pad, and cuff anchoring pad.
Write the patient’s name, identification number, and today’s date on the diary cover. The patient identification number can be en­tered into the ambulatory blood pressure monitor (up to eleven digits in length) and will be printed in the patient’s report.
Suggestion: Record the patient’s name and a unique identi­fying number in a log book. You’ll have a record of where the ABPM is, and you’ll be able to track usage.
*
4-3
Page 68

Preparing the ABP Monitor for Use

Begin by loading fresh batteries. Remove any used batteries
from the ABP monitor. Keep the used batteries separate from your fresh battery supply so that you don’t accidently re-insert them into the ABP monitor. (Dispose of the used batteries ac­cording to local regulation.)
Load four fresh AAA batteries and close the battery compart­ment door.
The Power-On Self-Test
Wait while the ABP monitor completes the Self-Test. Verify that the date and time shown during the test are correct.
LCD Test
Autozeroing
Jul05 03:22P Version 00.56
Patient ID Num
Unit contains 000 BP readings
Autozeroing
Date/Time/Version
Patient ID
BP Readings
Error Messages
If the ABP monitor’s Self-Test fails, an error message ap­pears. The Troubleshooting Guide, Section 7 explains each message and offers help.
!
4-4
Page 69
One of Three Messages Appears Following the Self-Test:
The message that comes up following the Self-Test reflects what the ABP monitor was doing when it was last used.
Yes No Next Set Patient ID ?
Set Patient ID ? This message will appear if there are no
patient readings currently stored in the unit’s memory. In this case, you are clear to proceed with setting the Basic Parameters.
Yes No Next Review Data ?
Review Data? This message will appear if data from a previous
patient study remains in memory. The data will remain stored in the memory until it is intentionally erased. If you want to review or print the data now stored in memory, refer to Section 5 for instructions.
To proceed with clearing the data from memory and preparing for a new patient, press Next until you see the following:
Erase Data ? Choose Yes. Are You Sure ? Choose Yes.
The display will change to Set Patient ID ? and you can pro­ceed to Setting the Patient ID on the next page.
Yes No Next Resume Readings ?
Resume Readings? This message will appear if a 24-hour
ABP monitoring period has been interrupted. Most often this would occur if a patient removes the batteries, or opens and closes the battery door, during the 24-hour period.
This message will be displayed for one minute. If no commands are made, the unit will automatically resume the Regular Run cycles.
4-5
Page 70
To prepare for a new patient study, you must review and/or print out the current data and then clear the memory. Refer to Section 5 for instructions on reviewing and printing data. To proceed with clear­ing the memory, press No or Next until you see the following:
Erase Data ? Choose Yes. Are You Sure ? Choose Yes.
This will clear all data from memory and the display will change to Set Patient ID ? You can proceed to Setting the Patient ID, below.
Hint: If you accidentally pass Set Patient ID ?, continue pressing No or Next until the message appears again.
*
Setting the Patient ID
Yes No Next Set Patient ID ?
Set Patient ID ? At this display, you are ready to begin enter-
ing the patient’s identification number. Press Yes.
4-6
Up Down Next Ê
A cursor will appear in the first character space. There is room to enter up to eleven digits or spaces.
To enter the first number of the patient’s ID, press Up or Down until the number you want is shown. Press Next to accept the number and move to the next character space.
Up Down Next 5
Ê
Enter the second number of the patient’s ID using the Up or Down button in the same way.
To leave a blank space after a number, press Next while the cursor is highlighting a blank space.
*
Page 71
Continue selecting numbers and pressing Next until you have entered the entire ID number.
When all digits have been selected, press Next one or more times until the Set Patient ID ? message appears.
Yes No Next Set Patient ID ?
You’ve just set the patient ID, so press Next to proceed. (Or, press Yes to review what you just did or re-set the patient ID.)
Setting the Date
Yes No Next Set Date ?
The Set Date ? message appears next. If the date shown dur­ing the Self-Test sequence was correct, then you don’t need to set the date. Press Next to omit setting the date and go to Set­ting the Time on the next page.
Or, press Yes if you want to check or set the date, and continue from here.
Up Down Next Sep 10, 1993
The new display shows today’s date, and a cursor highlights the month.
If the month is correct, press Next to move the cursor to the next field.
OR If the month is incorrect, press Up or Down to rotate through
the calendar and select the correct month. When the correct month is showing, move the cursor to the next field by pressing Next.
After setting the month, set the day and then the year. When all fields are set, press Next to return to the Set Date ? display.
4-7
Page 72
Yes No Next Set Date ?
You’ve just set the date, so press Next to proceed. (Or, press Yes if you want to review what you just did or re-set
the date.)
Setting the Time
Yes No Next Set Time ?
Setting the time works like setting the date. If the time shown during the Self-Test sequence was correct, then press Next to omit setting the time and proceed to Setting the Parameters.
Or, press Yes to set the time, and continue from here.
Up Down Next 11:08PM
The display shows the current time, and a cursor highlights the hour. If the hour shown is correct, press Next to move the cursor to
the minutes.
4-8
OR If the hour is incorrect, press Up or Down to move to a later or
earlier hour. When the hour is correct, press Next to move the cursor to the minutes.
After setting the hour, set the minutes. Press Next.
Yes No Next Set Time ?
You’ve just set the time, so press Next to continue. (Or, press Yes if you want to set the time again, or review what
you just did.)
To change from a.m./p.m. to 24-hour clock, refer to Special Parameters in Section 3, page 3-18.
*
Page 73
Setting the Parameters
Yes No Next Set Parameters ?
Set Parameters ? When this display appears, you may review
the current parameter settings and either accept or change them. If you would like to accept the current parameter settings, press
Next and refer to Sizing the Cuff to the Patient, page 4-12. Or, to review or change the parameters, respond with Yes. The
Set Periods ? message will appear. An abbreviated guide to setting the parameters is provided here.
If you are not familiar with the ABP monitoring parameters and need more information, review Basic Parameters in Section 3.
The Set Periods Parameter
Yes No Next Set Periods ?
Set Periods ? This allows you to select the number of ABP
monitoring periods (1, 2, 3, or 4), define when those periods begin, and select the frequency of the automatic blood pressure measurements taken during each period.
To omit setting the periods and go on to the next Basic Pa­rameter, Press No or Next and proceed to Setting the Maxi­mum Pressure,
OR Press Yes to review or define each period.
Up Down Next # of Periods 2
Selecting the Number of Periods. This allows you to select the number of ABP monitoring periods (1, 2, 3, or 4).
Press Up or Down to change the number shown on the ABP monitor’s display. When the number of periods you want is displayed, press Next to accept the number.
Set the beginning time of the first period next.
4-9
Page 74
Up Down Next Prd 1
07AM 15m
Setting the Beginning Time of the Period. A cursor highlights the beginning time for the first period, Prd 1.
Select the period’s beginning time by pressing Up or Down to move to a later or earlier time.
Press Next when the correct beginning time is shown. The cur­sor will move to highlight the next field, the frequency setting.
Next, set the measurement frequency for this period.
Up Down Next Prd 1 07AM
15m
Measurement frequency is the time interval, in minutes, between automatic readings. The choices are Manual, 5, 10, 15, 20, 30, 60, or 120 minutes.
Select the frequency of measurements with the Up or Down button until the one you want appears.
After selecting the measurement frequency, or if you want to accept the frequency currently shown, press Next.
The next period that needs to be defined will be displayed.
4-10
Up Down Next Prd 2
06PM 30m
Define the next period by following the same steps you used with the first period.
Repeat the same steps to define the third and fourth periods, if needed.
Up Down Next Maximum Pres 200
Setting the Maximum Pressure. This determines the maximum cuff inflation pressure.
To change the pressure, press the Up or Down buttons and scroll through the pressures until the desired one appears.
Press Next to accept the selected pressure.
Page 75
Off On Next Display On
Turning the Display On or Off. This turns the ABP monitor’s display of blood pressure readings On so they are visible to the patient, or Off so the patient does not see them.
Press On or Off according to your display preference. Then, press Next.
Off On Next Start Button On
Turning the Patient Activation Button On or Off. If On, the patient can initiate a blood pressure measurement by pressing a button. If Off, the patient cannot initiate a blood pressure mea­surement; only the programmed readings are taken.
Press On or Off according to your preference. Then, press Next.
Yes No Next Set Adv Param ?
When you have completed reviewing or selecting the Basic Pa­rameters, you can continue with the Advanced Parameters, or you can proceed with preparing for the Office Runs. Press Next to ready the ABP monitor for Office Runs. Set the ABP monitor aside and move on to Sizing the Cuff to the Patient.
The Advanced Parameters are explained in Section 3. Here, we will assume you are accepting the Advanced Parameter settings as they are, since they generally are not changed
*
from one patient hookup to the next.
4-11
Page 76

Sizing the Cuff to the Patient

When preparing to put the cuff on the patient, seat the patient comfortably in a private room. If the patient has long sleeves, have him or her remove the shirt so the upper arms are acces­sible for the hookup.
Determine Which Arm to Take the Readings From
In most cases, you can simply ask the patient whether he/she is right-handed or left-handed, and select the non-dominant arm for the hookup. The best arm for cuff placement is the one used less frequently, so the arm can be relaxed and motion-free during the measurements.
After identifying the arm that will be used, you need to fit the cuff. Medium and Large cuffs will fit most adult arms, and are included with the ABP monitor. A small size is available as an option from your Welch Allyn/Tycos Distributor.
The Importance of Correct Cuff Size
There are two simple methods for determining the right size cuff for the patient. Use either method.
4-12
Careful sizing of the cuff is important to the accuracy of the blood pressure readings. If the cuff is too small, the read­ings could be falsely high.
!
Page 77
Determining Cuff Size with the Cuff Markings
UM
ERY
LEFT ARM
RANGE
One way to ensure proper cuff size is to wrap the cuff around the patient’s upper arm and visually check it. The cuff is marked with two divisions that indicate “range”. When the cuff is properly fit, the cuff edge will meet the cuff at some point within the range.

Chart for Determining Cuff Size

You can also determine cuff size by measuring the patient’s arm circumference midway between the elbow and shoulder, and then use the chart below to select the correct cuff.
Cuff Size Arm Circumference
Small 12 to 21 cm Medium 18 to 32 cm Large 21 to 39 cm
4-13
Page 78

Positioning the Cuff & Microphone

Obtain a cuff assembly and pneumatic assembly in the proper size for the patient. The cuff assembly is made up of the cuff and bladder. The pneumatic assembly is made up of the air hose and the microphone assembly. If the cuff and pneumatic assembly have been taken apart for cuff washing or to change cuff sizes, you will need to reassemble them. Please follow these steps:
Inserting the Bladder into the Cuff
Slide the bladder into the cuff, making sure the bladder lies flat and unfolded.
MEDIUM
RANGE
This illustration shows the bladder insertion for Left-Arm placement.
MEDIUM
RANGE
4-14
If the cuff will be worn on the Right Arm, the bladder orientation is reversed.
Page 79
M
T
H
A
R
G
I
R
TERY
R
Z
E
I
S
X
A
M
m
c
5
.
1
3
A
LEFT ARM
Pull the bladder’s tubing through the small opening (marked Left Arm for Left-Arm placement) near the edge of the cuff. Note that there are two openings. One is for Right-Arm cuff placement, and one is for Left. (They are labeled on the face of the cuff.)
M
T
H
A
R
G
I
R
TERY
R
Z
E
I
S
X
A
M
m
c
5
.
1
3
A
LEFT ARM
Attach the tubing from the bladder to the pneumatic assembly.
4-15
Page 80
LEFT ARM
TERY
R
A
Next, thread the microphone back through the same small open­ing with the bladder tube.
M
T
H
A
R
G
I
R
TERY
R
SIZE
X
A
M
m
c
5
.
1
3
A
LEFT ARM
Push the microphone through the cuff and out through the small opening labeled Artery as shown. Make sure the microphone cable was threaded on top of the bladder, not on the side that will be next to the arm.
4-16
Page 81

Locate the Brachial Artery

1-1.5" (2.5 to 3.8 cm)
Have the patient flex the arm and locate the area between the bicep and tricep, 1 to 1.5 inches (2.5 to 3.8 cm) above the elbow on the inside of the patient’s arm. With your index and middle fingers, press into the area and feel for the pulse of the brachial artery. To confirm that you have located the artery, palpate the wrist at the same time; the pulses should beat at the same rate.

Mark the Location of the Artery

Boldly mark the location of the brachial artery with a felt pen to identify the site for microphone placement.
4-17
Page 82

Use of the Microphone Anchoring Pad

Find the adhesive microphone anchoring pad in the hookup kit. Peel the backing off the pad and insert the microphone with the label side facing microphone pad and away from arm. Posi­tion the microphone cable through the slit in the pad.
4-18
Attach the microphone anchoring pad with microphone to the brachial artery location. Firmly seal the adhesive to the arm.
Incorrect microphone placement or orientation, or incorrect cuff placement, can cause inaccurate measurements.
!
Page 83

Wrapping the Cuff

Center the circle on the cuff marked Artery over the microphone. A slight bit of white microphone pad should be showing under the edge of the cuff. Wrap the cuff snugly around the patient’s arm. Check that the cuff is not so tight that it impedes circulation or arm movement.
A S
U a n i
l o r
a
C h t
r
o N ,
n
e d r
A
Y
R
E
T
R
A
Make sure the tubing comes out of the opening at the top of cuff and follows up towards the shoulder.
4-19
Page 84

Secure the Cuff

The cuff should be secured to prevent slippage during the ABP monitoring period. Locate the cuff anchoring pad from the hookup kit and snap it onto a cuff anchor strap. Attach the fabric fastener to the top of the cuff near the shoulder. Peel off the backing of the pad and apply the anchoring pad to the patient’s arm, leaving a small amount of slack.
4-20
Lead the tubing up the arm and over the shoulder. Drape the tubing loosely around the patient’s neck to the front of the body or to either side.
Page 85

Taking the Office Runs

Connecting the ABP Monitor to the Patient’s Cuff Assembly
Attach the cuff assembly’s connector to the port on the front of the ABP monitor with a backwards (counterclockwise) twist (about a quarter turn), and then twist it clockwise until it is snugly attached to the ABP monitor. With this method, the connector should be securely locked in place.
Do not overtighten the connector; only moderate tighten­ing is required.
!
Align Red
Dots
To attach the microphone cable, align the red dot on the micro­phone cable connector with the red dot on the ABP monitor’s microphone port, and then insert the microphone. You should hear or feel a slight “click” as the microphone locks in place.
Check that the arm cuff and tubing are still positioned properly on the patient. Hold the ABP monitor or carefully place it on a nearby table.
4-21
Page 86
Taking the Office Runs
Office Runs are important for several reasons. Besides being a good test of the ABP monitor’s function and the patient hookup, the Office Runs help prepare the patient for what to expect dur­ing the ABP monitoring period. And importantly, during the Office Runs, the ABP monitor’s microphone adjusts its gain level, adapting to the patient’s own Korotkoff sounds.
For these reasons, we recommend that you complete at least three Office Runs during every patient hookup.
Yes No Next Do Office Run 1 ?
Before beginning an Office Run, ask the patient to relax the arm and to remain as motionless as possible. When the patient is ready, press Yes.
Exit Exit Start
Next, press Start to initiate the run.
Nov13 03:22P 140/82 o HR=80
4-22
The first Office Run will always inflate to the preset maximum pressure. Subsequent Office Runs will inflate 30 mmHg above the previous Systolic reading, if Dynamic Pressure is turned on.
When the run is complete, the date, time, blood pressure mea­surement, and heart rate will be displayed for two minutes. Then another Office Run can be taken. To move to Do Office Run 2Ê? before two minutes are up, press the green button.
The small o between the blood pressure reading and the heart rate (HR) indicates that the measurement was an Office Run. Symbols that appear here are called Initiation Codes.
If there were any difficulties, now is a good time to adjust the cuff, check all tubing connections, and recheck the microphone posi­tioning before taking the next run.
Page 87
Nov13 03:22P 10 140/82 o HR=80
About Error Codes: A two-digit Error Code may appear in the upper right corner of the display if a problem was encountered during the measurement. For an explanation of Error Codes, review the measurement Error Codes in the Troubleshooting Guide, Section 7.
During Office Runs, it is normal to encounter Error Codes. The unit is becoming adjusted to the Korotkoff sounds of the individual. It may be necessary to take three or more
!
Office Runs to eliminate the Error Codes.
Yes No Next Do Office Run 2 ?
When you are ready to take another Office Run, press Yes, then press Start to initiate the second run.
You can repeat as many Office Runs as needed to help the patient understand the procedure and practice relaxing for the measurement.
If the ABP monitor’s display is turned On in the Basic Parameters, show the patient how the readings will ap­pear in the display.
*
When you have completed three or more Office Runs success­fully, you can proceed to the next section, Verifying the Setup, or to Beginning the Regular Runs by pressing Next.
4-23
Page 88

Verifying the Setup

Once you have successfully completed three Office Runs, you may wish to take a simultaneous blood pressure reading to verify that the Tycos QuietTrak ABPM is set up and operating correctly.
You should not attempt to do a simultaneous reading until at least three Office Runs have been completed and no Error Codes are present.
!

Items Needed for Simultaneous Measurement:

Accurate office sphygmomanometer
• Calibration T-connector
• Tubing adaptor
• ABP monitor and cuff assembly

Instructions for Setup

Disconnect the Tycos QuietTrak cuff and microphone assembly from the ABP monitor. Assemble the T-connector to the ABP monitor’s cuff hose (Leur) connector. Connect the QuietTrak cuff and microphone assembly to the T-connector. Attach one end of the tubing adaptor to the T-connector, and the other to a prop­erly-calibrated office manometer. Refer to setup shown below.
4-24
S Y S / D I A mmHg
A well-trained observer will now use a stethoscope to measure the patient’s blood pressure simultaneously with the QuietTrak ABP Monitor.
Page 89
Yes No Next Do Office Run 4 ?
Proceed by answering Yes to the next Office Run.
Exit Exit Start
Press Start to initiate the run.
Allow the ABP monitor time to complete its measurement cycle before concluding the reading or moving the patient.
Important
Expect to see minor variations between the reading given by the QuietTrak and the observer’s measurement. Just as no
!
two people would determine the exact same measurement, this discrepancy is unavoidable and normal. Variation be­tween the ABP monitor and the observer of ±5 mmHg or less is considered acceptable. If encountering problems, call your local Welch Allyn/Tycos Service Center for assistance.
When you have completed Setup Verification, remove the T-con­nector and reconnect the cuff to the ABP monitor. You can now proceed to Beginning the Regular Runs by pressing Next.
4-25
Page 90

Beginning the Regular Runs

Yes No Next Begin Reg Run ??
!
Prior to sending a patient out for a 24-hour study, you must put the ABP monitor into the Regular Run mode. This will initiate the preset automatic measurements.
Nov13 03:22P 140/82 o HR=80
If an Office Run measurement is still displayed, press the green button to get back to the menu.
Yes No Next Do Office Run ?
Press Next.
Yes No Next Begin Reg Run ?
4-26
To begin the Regular Runs of the ABP monitoring process, press Yes. The first measurement will be taken after the speci­fied interval for the current time period has elapsed. Times are adjusted to the hour.
Page 91

Positioning the Pouch & Tubing

Shoulder Strap or Belt? Ask the patient for any preference
towards a shoulder strap or belt. A person who is mostly going to be seated may find the strap is most comfortable, because the ABP monitor can rest in the lap. A person who will be up and down frequently may find the belt more comfortable. If pre­ferred, the patient’s own belt can be used, as long as it is narrow enough to slip through the pouch’s loops.
If using the strap, clip the strap’s fasteners to the rings on the pouch. If using a belt, slide the belt through the loop on the back of the pouch. Help the patient put on the shoulder strap or the belt.

Inserting the ABP monitor into the Pouch

Insert the ABP monitor into the pouch as shown, making sure the display is fully visible and the green Patient Activation Button is positioned directly under the open cutout locator. Close the pouch.
4-27
Page 92
Deciding Where to Place the ABP monitor
The pouch should be placed where it is comfortable for the pa­tient, and where it won’t restrict movement. Assist the patient with any needed adjustments, making sure the patient can easily reach the Patient Activation Button as needed.
The straps and tubing can be concealed inside the patient’s clothing, or can be positioned on the outside. Decide with the patient which would be preferable. Have the patient replace his or her shirt if it was removed earlier, and check that the tubing is secured but does not restrict the patient’s movement. Below are suggestions.
4-28
Page 93

Instructing the Patient

Reposition the ABP monitor if necessary so the patient can acti­vate readings, cancel readings and view the display.
Viewing the Display
Nov13 03:25P
Show the patient where the display is, and that the date and time will appear continuously.
Nov13 03:22P 140/82 o HR=80
Optional: If the display is turned On in the Basic Parameters, then point out where the blood pressure readings will appear following each measurement.
Explaining the Diary
Give the patient the diary and explain how to fill it out with time of day, symptoms/mood and activity/position. Give some examples of what you mean by:
Symptoms (such as chest pain, tenseness, feeling fine)
Activities (washing dishes, sitting to read, lying down, watching TV)

Demonstrating the Patient Activation Button

If the Patient Activation Button is turned On in the Basic Param­eters, explain how to initiate readings by pressing and releasing the button when a symptom appears. Show the patient where the button is and ask the patient to press it now to take a reading.
Advise the patient to PRESS and RELEASE the button; DO NOT hold the button down. Holding the button down
can turn the ABP monitor off. See Instructions to Patient on
!
page 4-30 for more details.
4-29
Page 94

Cancelling the Readings

The Patient Activation Button is able to cancel readings when necessary, too. This is a safety feature that is available even if the Patient Activation Button is turned Off in the Basic Parameters.
Explain to the patient that he or she can press and release the Patient Activation Button while a reading is in progress to cancel the reading. For example, a reading can be cancelled if the pa­tient gets in the middle of a task and cannot momentarily pause for the reading.

Positioning the ABP monitor at Bedtime

Explain how to detach the ABP monitor and remove the strap or belt to change clothing at bedtime. Suggest that the patient sit on the bed and carefully place the ABP monitor beside him while changing so that it doesn’t drop to the floor. For sleeping, have the patient position the ABP monitor alongside the body or near the head of the bed. It might be more comfortable and quiet if a pillow is placed over the ABP monitor; this also helps prevent the patient from rolling on it.

Instructions to the Patient

To Deactivate the ABP monitor: If the patient must remove the
ABP monitor and cuff prior to returning to the clinical site, he/she can PRESS and HOLD the Patient Activation Button for SIX SEC­ONDS. The ABP monitor will cease taking readings and can be removed from the patient’s arm. The ABP monitor stores the data for retrieval later; no data is lost.
4-30
Hold Button
to Quit
PRESS and HOLD the green Patient Activation Button for THREE SECONDS and the Hold Button to Quit message will appear.
Off
Continue holding the button for THREE MORE SECONDS, and the message Off will appear. The unit will cease taking any measurements.
Page 95
Press Button to Resume
Upon releasing the button, the message Press Button to Resume will stay on indefinitely as a reminder to the patient.
Auto Readings will Resume
To restart the regular cycle, the patient can press the green but­ton and the message Auto Readings will Resume will ap­pear. This message will stay on for two minutes. Readings will resume per the current period interval.
It is inadvisable to allow the patient to deactivate and remove the cuff and microphone assembly unless no more readings are required. Once the adhesive microphone anchoring pad is
!
removed, the patient will have difficulty in correctly reposition­ing the microphone over the brachial artery.

Scheduling the Patient’s Return

Remind the patient of the day and time to return with the ABP monitor and the diary. If the patient cannot return at the time coinciding with the end of the ABP monitoring session, explain how to turn the ABP monitor Off by pressing and holding the Patient Activation button for at least six seconds. Then the ABP monitor can be removed from the patient and safely set aside until the scheduled return date.
4-31
Page 96

Summary of Patient Do’s and Don’ts

Do’s

• Record time, symptom/mood, and activity/position in diary.
• Press Patient Activation Button to take a reading when any symptoms are felt.
• Keep arm as motionless as possible during measurement.
• Keep travel time in car or other vehicle to a minimum.
• Return on the specified date and time, and bring the diary with you.

Don’ts

• Don’t remove the cuff and microphone from original location.
• Don’t get the ABP monitor wet.
• Don’t attempt to operate a power tool or heavy equipment during a reading.
Advising the patient to keep his or her arm motionless dur­ing the measurement is crucial to attaining accurate and reliable readings. If the QuietTrak ABP Monitor detects sig-
!
nificant noise, motion or Korotkoff sounds at the end of inflation, the ABP monitor will re-inflate the cuff and start the measurement over. Re-inflations will continue until no Korotkoff sounds or noises are detected, or until maximum inflation pressure is reached.
Advise the patient to be careful while operating a vehicle. While the ABP monitor is taking a measurement, the patient’s ability to operate a vehicle may be impaired due to restriction in arm motion.
4-32
Page 97

When the Patient Returns

This section instructs the operator how to disconnect the patient and retrieve the Tycos QuietTrak ABPM report.
Page Topic
5-2 Disconnecting the ABP Monitor 5-3 Ending the ABP Monitoring Session 5-5 Reviewing the Data 5-7 Printing the Data
5
5-1
Page 98

Disconnecting the ABP Monitor

When the patient returns at the end of the ABP monitoring ses­sion, remove the ABP monitor from its pouch, disconnect the microphone cable and cuff tubing, and set the ABP monitor aside for a moment. Remove the anchoring pads and the arm cuff, then assist the patient in removing the belt or shoulder strap.
Disconnecting the microphone cable from the ABP Monitor:
To remove microphone connector, place your index finger and thumb on the knurled portion of the connector and pull straight back (away from the monitor). Do not twist!
Grasp here
5-2
Review the Diary With the Patient
Ask for the diary and review it together in case the patient now remembers something that was omitted. Add any missing details and clarify any words you don’t understand. The patient is now finished and can go home.
Page 99

Ending the ABP Monitoring Session

When the patient returns the ABP monitor, one of the follow­ing three messages will be displayed. The message that is displayed will reflect the status of the unit:
1 Day Max Limit
1 Day Max Limit will be displayed if the unit is returned
after 24 hours since the first Regular Run reading was taken, and the unit has not been turned off or the batteries removed.
Note: To program the ABP monitor to go beyond the 24­hour maximum, refer to Section 3, Choosing Program
*
Options.
Nov15 02:28P
If the 24-hour time limit has not elapsed and the Regular Runs are still in progress, the time and date will appear as normal.
If either of the above two messages are displayed, you will need to end the ABP monitoring session by pressing all three but­tons simultaneously. The display will change, and you will be presented with the Review Data ? screen. You have effec­tively ended the ABP monitoring session.
5-3
Page 100
Yes No Next Resume Readings ?
The third possible message, Resume Readings ? will appear if the patient removes the batteries prior to returning to the clini­cal site, and before the programmed ABP monitoring period had elapsed. This message will appear when batteries are inserted and the unit completes the Self-Test.
Respond with No, and the Review Data ? message will appear.
In rare circumstances, the messages Battery Failure or 300 RDG Limit may appear. Refer to Section 7, page 7-7
for an explanation. Press all three buttons to return to the
*
main menu.
Yes No Next Review Data ?
When the question Review Data ? appears, respond with Yes if you would like to review the data on the ABP monitor’s
display, or press Next and proceed to Printing the Data if you prefer to print the report now.
5-4
Loading...