Siemens Mammomat 3000 Nova User manual

MAMMOMAT 1000/3000 Nova
SP
Installation
Installation and Start-Up Instructions
from Serial No. 9000
© Siemens AG 2003
The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable for damages. All rights, including rights created by patent grant or registration of a utility model _or_ design,_are_ reserved.
Register 3 English Print No.: SPB7-230.033.11.05.02 Doc. Gen. Date: 05.05
Replaces: SPB7-230.033.11.04.02
0 - 2 Revision
Chapter Page Revision
All All 05
Document revision level
The document corresponds to the version/revision level effective at the time of system delivery. Revi­sions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.
Disclaimer
The installation and service of equipment described herein is to be performed by qualified personnel who are employed by Siemens or one of its affiliates or who are otherwise authorized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with or autho­rized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates before attempting installation or service procedures.
M1000/3000 Nova Register 3 SPB7-230.033.11 Page 2 of 8 Siemens AG Installation and Start-Up Rev. 05 05.05 CS PS 24 Medical Solutions
Contents 0 - 3
Page
1 _______Prerequisites __________________________________________________ 1 - 1
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1 - 1
MAMMOMAT 1000 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 1
MAMMOMAT 3000 Nova . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 2
Meters and appliances, tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 3
Meters and appliances required . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 3
Tools required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 3
Important notes on start-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 4
Checking and recording for the area of application of the X-ray decree (§16 Germany)1 - 4
Checking and recording for the DHHS area of application . . . . . . . . . . . . . . . 1 - 5
Sections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 5
Note on delivery state . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 5
2 _______Protective measures ____________________________________________ 2 - 1
Protective measures. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 1
Delay times between two exposures . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 2
3 _______Preparatory work_______________________________________________ 3 - 1
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 - 1
Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 1
Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 1
Unpacking the stand. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 2
Removing the transport safeguards . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 3
Removing the swivel-arm system transport safeguard . . . . . . . . . . . . . . . 3 - 3
Transport safeguard for the lifting carriage . . . . . . . . . . . . . . . . . . . . . 3 - 3
Transport safeguard for the rotary motion . . . . . . . . . . . . . . . . . . . . . . . . 3 - 4
Ball bearings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 - 5
Protective strips for the metal curtain . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 5
Unpacking the generator and mounting the radiation shield (optional) . . . . . . . . . 3 - 6
4 _______Installing the generator and the stand _____________________________ 4 - 1
Arranging the components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 1
Notes on installations with separate generator and separate control console. . . . 4 - 1
Free-standing radiation shield (option) . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 2
Removing the generator cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 2
Installing the cable ducts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 3
Installations with separate control console . . . . . . . . . . . . . . . . . . . . . 4 - 3
Laying the cable harness . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 4
Installations with separate control console . . . . . . . . . . . . . . . . . . . . . 4 - 5
Aligning the stand . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 6
5 _______Cable connections______________________________________________ 5 - 1
EMC measures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 1
Siemens AG Register 3 SPB7-230.033.11 Page 3 of 8 M1000/3000 Nova Medical Solutions Rev. 05 05.05 CS PS 24 Installation and Start-Up
0 - 4 Contents
Page
EMC measures at the cable entry . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 1
Fitting the hose clamps and ferrite sleeves . . . . . . . . . . . . . . . . . . . . . 5 - 1
Note on separate control console and separate generator . . . . . . . . . . . . . 5 - 2
EMC measures on the bottom plate . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 3
Fitting the cables X1, X8 and X9 onto the bottom plate . . . . . . . . . . . . . . 5 - 3
Connecting the stand cable-harness . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 4
High-voltage connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 4
Connecting the high-voltage connector. . . . . . . . . . . . . . . . . . . . . . . . . 5 - 5
Connecting the cable harness to the generator . . . . . . . . . . . . . . . . . . . . 5 - 6
Unit control cable X1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 6
Filament cable X8. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 6
Power supply cable X14 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 6
AEC signal cable X10 and X11 . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 6
Rotating anode cable X9 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 7
Installations with separate control console and separate generator . . . . . . . . 5 - 8
Main voltage connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5 - 10
Connecting the incoming mains to the stand (400 V, 2-phase). . . . . . . . . . . . .5 - 11
400 V 2-phase connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5 - 11
Connecting the mains supply to the generator (400 V, 2-phase) . . . . . . . . . . . .5 - 12
Measures for changing from 2-phase to 1-phase connection . . . . . . . . . . .5 - 13
6 ______ Mains connection and power supply_______________________________6 - 1
Checks before powering up the generator . . . . . . . . . . . . . . . . . . . . . . . 6 - 1
Measuring the line resistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 1
Checking the line voltage in the generator . . . . . . . . . . . . . . . . . . . . . . . 6 - 2
Checking the supply voltages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 3
7 ______ Attaching the swivel-arm covers __________________________________7 - 1
Attaching the swivel-arm covers . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 1
Arranging the swivel-arm system . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 1
Connecting the cables to control-button boards and patient handles . . . . . . . 7 - 1
Attaching the side covers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 2
Attaching the front cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 2
8 ______ Checking the microprocessors ___________________________________8 - 1
Microprocessors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 1
9 ______ Checks without high voltage _____________________________________9 - 1
Checks without high voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 1
10 _____ Checks with high voltage _______________________________________10 - 1
Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10 - 1
kV-adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10 - 2
Filament . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10 - 4
M1000/3000 Nova Register 3 SPB7-230.033.11 Page 4 of 8 Siemens AG Installation and Start-Up Rev. 05 05.05 CS PS 24 Medical Solutions
Contents 0 - 5
Page
Checking X-ray tube high voltage, tube current and mAs values . . . . . . . . . . . 10 - 5
Oscilloscope diagrams. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 6
11 ______Calibrating and adjusting the AEC _______________________________ 11 - 1
Checking and programming with the service PC . . . . . . . . . . . . . . . . . . . 11 - 1
Preparation of backup floppy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 - 3
Offset compensation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 - 4
Configuration of the AEC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 - 6
Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 - 6
Object table group . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 - 6
General conditions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 - 6
Calibration conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 - 6
Reference cassette . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 - 6
Film . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 - 6
Mains voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 - 6
Sensitivity conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 - 6
Correction of the measured Optical Density (O.D.) . . . . . . . . . . . . . . . . 11 - 7
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 - 7
Normal developing conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 - 7
Incremental gamma . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 - 9
Overview of work routine . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 - 10
AEC Correction tables. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 - 11
DLF switch off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 - 11
Increase grid speed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 - 11
Installation of AEC correction tables. . . . . . . . . . . . . . . . . . . . . . . 11 - 12
Calibrate correction tables. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 - 14
Recalibration of an object table . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 - 19
Sensitivity correction (fine setting). . . . . . . . . . . . . . . . . . . . . . . . . . 11 - 20
Sensitivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 - 21
Copy H to D . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 - 22
12 ______Dose Calculation System _______________________________________ 12 - 1
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 - 1
Enable/Disable Dose Calculation . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 - 1
Configuring of Dose Calculation System . . . . . . . . . . . . . . . . . . . . . . . 12 - 2
Label. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 - 2
13 ______Testing ______________________________________________________ 13 - 1
Testing the AEC-function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 - 1
AEC performance test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 - 1
Testing and adjusting OPDOSE . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 - 2
14 ______Further programming __________________________________________ 14 - 1
Setting the real time clock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 - 1
Prerequisite . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 - 1
Siemens AG Register 3 SPB7-230.033.11 Page 5 of 8 M1000/3000 Nova Medical Solutions Rev. 05 05.05 CS PS 24 Installation and Start-Up
0 - 6 Contents
Page
Setting the time in the MAMMOMAT . . . . . . . . . . . . . . . . . . . . . . . . . .14 - 1
Reducing the generator power . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14 - 1
Main menu: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14 - 1
15 _____ Checking the swivel-arm system _________________________________15 - 1
Rotary motion. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15 - 1
Vertical adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15 - 2
Emergency stop . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15 - 3
16 _____ Field light ____________________________________________________16 - 1
Checking and adjusting the field light time . . . . . . . . . . . . . . . . . . . . . . .16 - 1
Checking the field light time . . . . . . . . . . . . . . . . . . . . . . . . . . . .16 - 1
Adjusting the field light time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16 - 2
17 _____ Checking the compression device and OPCOMP ___________________ 17 - 1
Checking the compression device . . . . . . . . . . . . . . . . . . . . . . . . . . .17 - 1
Decompression button (only with separate control console) . . . . . . . . . . . .17 - 2
Checking the OPCOMP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17 - 3
Adapting OPCOMP. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17 - 3
18 _____ Checking the exposure blocking _________________________________ 18 - 1
DLF switch on. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18 - 1
Reduce grid speed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18 - 1
Cassette loaded switch on . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18 - 1
Checking the exposure blocking . . . . . . . . . . . . . . . . . . . . . . . . . . . .18 - 2
19 _____ Checking the radiation field limitation_____________________________19 - 1
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19 - 1
Measuring procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19 - 2
Evaluation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19 - 3
Alignment radiation field/light field . . . . . . . . . . . . . . . . . . . . . . . . .19 - 3
Radiation field limitation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19 - 3
20 _____ Final procedures ______________________________________________20 - 1
Service PC and measuring instruments . . . . . . . . . . . . . . . . . . . . . . . .20 - 1
Saving the configuration file . . . . . . . . . . . . . . . . . . . . . . . . . . . .20 - 1
Reading the exposure counter . . . . . . . . . . . . . . . . . . . . . . . . . . .20 - 1
Deleting the error memory . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20 - 1
Saving the programmed values. . . . . . . . . . . . . . . . . . . . . . . . . . .20 - 1
Removing the measuring instruments . . . . . . . . . . . . . . . . . . . . . . . . .20 - 2
Checking the protective grounding resistance . . . . . . . . . . . . . . . . . . . . .20 - 2
Mounting the cable duct covers . . . . . . . . . . . . . . . . . . . . . . . . . . . .20 - 2
Mounting the stand covers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20 - 2
M1000/3000 Nova Register 3 SPB7-230.033.11 Page 6 of 8 Siemens AG Installation and Start-Up Rev. 05 05.05 CS PS 24 Medical Solutions
Contents 0 - 7
Page
MAMMOMAT stand . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 - 3
Fitting the MAMMOMAT cap . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 - 4
Fitting the cable outlet cover. . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 - 4
Mounting the front cover onto the generator . . . . . . . . . . . . . . . . . . . . . 20 - 5
Final protective grounding resistance test. . . . . . . . . . . . . . . . . . . . . . . 20 - 5
Face shield . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 - 6
Warning label on control panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 - 7
Affixing the identification labels . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 - 8
Reporting IVK status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 - 8
Other installation material . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 - 8
Handling the Installation report . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 - 9
21 ______Bolting the stand/generator to the floor ___________________________ 21 - 1
Stand and generator with integrated radiation shield and control panel. . . . . . . . 21 - 1
Installation with separate generator . . . . . . . . . . . . . . . . . . . . . . . . 21 - 2
Free-standing radiation shield (option) . . . . . . . . . . . . . . . . . . . . . . 21 - 3
22 ______Appendix ____________________________________________________ 22 - 1
Working with the service PC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 - 1
Description of the syntax used in these instructions. . . . . . . . . . . . . . . . 22 - 1
Connecting the service PC . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 - 2
Configuration of the service PC . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 - 3
Configuration of computer BIOS. . . . . . . . . . . . . . . . . . . . . . . . . . 22 - 3
Configuration of Windows® . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 - 3
Starting up and using the service PC . . . . . . . . . . . . . . . . . . . . . . . 22 - 4
Starting up and using the service PC - Dose Calculation program . . . . . . . . 22 - 5
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 - 5
Test protocol. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 - 6
kV-adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 - 6
Filament . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 - 6
AEC performance test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 - 7
M1000 or Wing 1 M3000 Nova . . . . . . . . . . . . . . . . . . . . . . . . . . 22 - 7
Wing 2 M3000 Nova . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 - 9
OPDOSE settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 - 9
23 ______Changes to previous version ____________________________________ 23 - 1
Siemens AG Register 3 SPB7-230.033.11 Page 7 of 8 M1000/3000 Nova Medical Solutions Rev. 05 05.05 CS PS 24 Installation and Start-Up
0 - 8 Contents
M1000/3000 Nova Register 3 SPB7-230.033.11 Page 8 of 8 Siemens AG Installation and Start-Up Rev. 05 05.05 CS PS 24 Medical Solutions

Prerequisites 1

General 1

The configuration of the Mammomat installation is dependent on the customer’s choice. Up to twelve various combinations are possible as shown below:

MAMMOMAT 1000 1

MAMMOMAT 1000 - Basic unit
1 - 1
Molybdenum anode X-ray tube assembly with filter Mo 0.03
Fixed object-table arm ("Wing 2" no counterweight)
Dose Calculation System
Generator with integrated control panel and radiation shield
Molybdenum anode X-ray tube assembly with filter disc Mo 0.03/Rh 0.025
Separate generator
Separate control console
Free-standing radiation shield
Fig. 1
Siemens AG Register 3 SPB7-230.033.11 Page 1 of 6 M1000/3000 Nova Medical Solutions Rev. 05 05.05 CS PS 24 Installation and Start-Up
1 - 2 Prerequisites

MAMMOMAT 3000 Nova 1

MAMMOMAT 3000 Nova - Basic unit
Pivoting object-table arm ("Wing 1"+ "Wing 2")
Molybdenum/tungsten anode X-ray tube assembly with filter disc Mo 0.03/Rh 0.025/Rh 0.050
Dose Calculation System
Fixed object-table arm ("Wing 1" + counterweight)
Molybdenum anode X-ray tube assembly with filter disc Mo 0.03/Rh 0.025
Separate generator
Generator with integrated control panel and radiation shield
Separate control console
Free-standing radiation shield
Fig. 2
NOTE
M1000/3000 Nova Register 3 SPB7-230.033.11 Page 2 of 6 Siemens AG Installation and Start-Up Rev. 05 05.05 CS PS 24 Medical Solutions
A P40 MoW tube will be delivered irrespective of whether the cus­tomer has chosen the molybdenum/tungsten anode X-ray tube or the molybdenum anode X-ray tube. In the last case, the tungsten anode is deselected by the software.
Prerequisites 1 - 3
Unless otherwise stated, these instructions describe the stand with pivoted object-table arm and generator with integrated radiation shield and control panel. Depending on the installation configuration, some points may be omitted.

Meters and appliances, tools 1

Meters and appliances required 1

NOTE
Calibrated instruments are required.
Power Line Impedance Meter.
Protective ground-wire tester.
Oscilloscope, e.g TEKTRONIX 314.
Digital multimeter (Fluke type 8060A, part no. 97 02 101 Y4290 or Fluke type 87, part no.
97 03 976 Y4290).
Sensitometer.
Densitometer e.g PDA 81.
Service PC with connecting cable (PC-Generator), part no. 99 00 440 RE999.
Normi 7 test body (provided by the customer) or SIB phantom Type 42 001 (PTW or
INAK).
Centering cross, part no. 96 60 051 RE 999.
AEC calibration plexi:
- Three plates, "2" cm (part no. 65 61 232)
- One plate, "1" cm (part no. 65 61 224)
WAR NING
The existing ground conductor in the mains cable must under no circumstances be disconnected when operating the oscilloscope.
Life-threatening electric shock hazard exist.
For those measurements, in which any resulting ground loop may falsify the measuring result, use the differential amplifier (differ­ence measurement).

Tools required 1

Standard installation tool kit.
Torque wrench for bolting the stand/generator to the floor (optional).
Electrical screwdriver with adjustable torque is recommended.
Calculator
This symbol indicates exposure of X-ray radiation.
Siemens AG Register 3 SPB7-230.033.11 Page 3 of 6 M1000/3000 Nova Medical Solutions Rev. 05 05.05 CS PS 24 Installation and Start-Up
1 - 4 Prerequisites

Important notes on start-up 1

The MAMMOMAT is adjusted, programmed and tested in the factory, leaving the adapta- tion to the on-site mains voltage, adjustment of the AEC and the functional tests to be per­formed.
When the measurements to be made (kV, mAs, etc.) are within the tolerances stated in these instructions, this confirms that the settings made in the factory have not changed and the equipment is fully serviceable.
The measured values marked with in Chapter 10 "Checks with high voltage" shall be entered in the "start-up" column in the test certificate provided.
The service PC is required only for programming according to:
Chapter 10 "Checks with high voltage"
Chapter 11 "Calibrating and adjusting the AEC"
Chapter 12 "Dose Calculation System"
Chapter 13 "Testing"
Chapter 14 "Further programming"
Chapter 20 "Final procedures"
Description of the service PC syntax is found in Chapter 22 "Appendix".
NOTE
If the generator is switched off with the service PC connected, wait approximately 5 s before switching it on again.

Checking and recording for the area of application of the X-ray decree (§16 Germany)

In the area of application of the X-ray decree, an acceptance test certificate is supplied, with most of the data filled in by the factory.
Only:
the cut-off dose
the resolution of the film/screen systems used
the output values of the constancy test
must be determined by the owner of the equipment and recorded in the acceptance test certificate. Furthermore, the front page of the acceptance test certificate must also be completed with the operator´s data.
1
M1000/3000 Nova Register 3 SPB7-230.033.11 Page 4 of 6 Siemens AG Installation and Start-Up Rev. 05 05.05 CS PS 24 Medical Solutions
Prerequisites 1 - 5

Checking and recording for the DHHS area of application 1

In the area of application of the DHHS regulations, maintenance measurements must be made according to the "DHHS Maintenance Instructions" SPB7-230.662.03.04.02 and the "DHHS Supplements to the instructions for use" SPB7-230.661.02.03.02.

Sections 1

Required labels.
Radiation ON indicators.
Manual termination of exposure.
Checking the maximum adjustable mAs.
Filters in beam limiting device (BLD).
Reproducibility.
kVp-accuracy.
mAs-accuracy.
Automatic exposure control (AEC).
Coincidence of radiation field/image receptor.
Alignment of light field/X-ray field.
Illuminance of light localizer (light field).
Shut-down of motor-driven compression movement.
The result must be recorded in "DHHS Measurement certificates" SPB7-230.663.03.03.02. Some of the values to be determined can be taken from the acceptance test certificate. The radiation-field limitation is set at the factory and recorded in the acceptance test certif-
icate. If the measurements to be made concerning the radiation-field limitation are within the tol-
erances stated in these instructions, this confirms that the settings made at the factory have not changed.

Note on delivery state 1

The MAMMOMAT has been tested and adjusted at the factory, and should therefore be ready for operation after completion of the installation.
The equipment is set to 400 V, 2-phase, on delivery.
Siemens AG Register 3 SPB7-230.033.11 Page 5 of 6 M1000/3000 Nova Medical Solutions Rev. 05 05.05 CS PS 24 Installation and Start-Up
1 - 6 Prerequisites
This page intentionally left blank.
M1000/3000 Nova Register 3 SPB7-230.033.11 Page 6 of 6 Siemens AG Installation and Start-Up Rev. 05 05.05 CS PS 24 Medical Solutions

Protective measures 2

Protective measures 2

It is very important that any intervention in the equipment will start with disconnecting it from the power supply with the main circuit-breaker. Before removing or inserting any of the printed circuit board, switch of the equipment. To prevent accidental triggering of high voltage and radiation, set the switch S2 (SS) on p.c. board D702 to "OFF" (lower position, no triggering of the SS relay).
2 - 1
WAR NING
WAR NING
CAUTION
If the system is only switched off at the control panel or with S2/D711, line voltage will still be present at the generator line con­nection, line filter Z1, Z2, transformer T1, transformer T10 and p.c. board D711 (see wiring diagram).
Life-threatening electric shock hazard exists.
Disconnect mains cable and comply with the information on this page.
After shut-down of the system, there may still be 380 V DC present on the intermediate circuit.
Life-threatening electric shock hazard exists.
The voltage level will be indicated by LED V24 on PC board D710. The voltage will drop to less than 30 V within about 3 minutes, the LED goes out at about 30 V.
The p.c. boards contain electrostatic highly sensitive compo­nents.
If not regarded, the components could be damaged.
Use ESD-equipment, ground prior to making contact and place the components on a conductive surface.
WAR NING
Siemens AG Register 3 SPB7-230.033.11 Page 1 of 2 M1000/3000 Nova Medical Solutions Rev. 05 05.05 CS PS 24 Installation and Start-Up
The edges of the metal curtain of the stand are very sharp.
They may cause severe injury.
Apply the protective strips as mentioned in section "Protective strips for the metal curtain" on page 3 - 5 after removing the cov­ers from the stand. Remove the protective strips only when the covers are to be mounted or when vertical adjustment of the swivel-arm system is necessary.
2 - 2 Protective measures

Delay times between two exposures 2

Delay times listed below must be followed in order to prevent the tube from overheating.
Exposure mAs value Delay time between two exposures (seconds)
max 100 min 15 max 200 min 30 max 300 min 45 max 400 min 60 max 500 min 75
M1000/3000 Nova Register 3 SPB7-230.033.11 Page 2 of 2 Siemens AG Installation and Start-Up Rev. 05 05.05 CS PS 24 Medical Solutions

Preparatory work 3

General 3

Scope of delivery 3

The MAMMOMAT is normally packed in one crate and one cardboard box (the number of packages is dependent on the customer’s choice, however):
Crate (length 2140 mm, width 800 mm, height 1375 mm)
The crate contains:
Compression plate
Stand with X-ray tube assembly and base plate
Object table 18 cm x 24 cm (Bucky)
External diaphragm
Cover panels
Installation material
Optional: Stanchions for the lead-glass pane (radiation shield)
3 - 1
Cable ducts
Technical documents
Cardboard box (lenght 1210 mm, width 820 mm, height 700 mm)
The cardboard box contains:
Generator with high-voltage generator and base plate.
Optional: Lead-glass pane
Optional: Separate control console
Further accessories are packed in a separate cardboard box.

Unpacking 3

As a general rule, the directional marks on the crates should be observed during trans­port, storage and unpacking.
The stand crate is bolted together, while the cardboard box is secured with plastic straps.
CAUTION
The unpacking must be carfully performed.
There’s a risk for foot injuries when handling heavy parts.
Wear safety footwear.
Siemens AG Register 3 SPB7-230.033.11 Page 1 of 8 M1000/3000 Nova Medical Solutions Rev. 05 05.05 CS PS 24 Installation and Start-Up
3 - 2 Preparatory work

Unpacking the stand 3

1
3
Fig. 1 Unpacking the stand
1
2
5
2
4
6
Unpacking
1. Open the stand crate by removing the top (1) and then the side wall (2).
2. Take out all enclosed packages (accessories, installation material, cover panels etc.) from the crate.
MAM00167
3. Remove the wooden supports.
4. Remove the remaining walls (3).
5. Upend the stand (4), with pallet (5) (two persons are required).
6. Loosen the four bolts (6) and remove the pallet (5).
M1000/3000 Nova Register 3 SPB7-230.033.11 Page 2 of 8 Siemens AG Installation and Start-Up Rev. 05 05.05 CS PS 24 Medical Solutions
Preparatory work 3 - 3

Removing the transport safeguards 3

Removing the swivel-arm system transport safeguard 3

1
2
3
MAM00081
Fig. 2 Swivel-arm system transport safeguard
1. Loosen the transport safeguard nuts (1/Fig. 2) and remove the tube head screws (2/Fig. 2).
2. Push the red transport safeguard (3/Fig. 2) inwards, away from the tube head.
NOTE
Removal of the swivel-arm system transport safeguard is done after the equipment has been powered up, see further section "Attaching the swivel-arm covers" on page 7 - 1.

Transport safeguard for the lifting carriage 3

2
1
MAM00082
Fig. 3 Transport safeguard for the lifting carriage
1. Loosen the two screws (1/Fig. 3) and remove the red transport safeguard (2/Fig.
3). Reinsert screws (1/Fig. 3), they might be needed for service purposes.
Siemens AG Register 3 SPB7-230.033.11 Page 3 of 8 M1000/3000 Nova Medical Solutions Rev. 05 05.05 CS PS 24 Installation and Start-Up
3 - 4 Preparatory work
NOTE
The lifting carriage transport safeguard shall be kept for service purposes (Fig. 5).
2. Cut and remove the cable tie securing the balancing spring during transport.

Transport safeguard for the rotary motion 3

1
MAM00083
Fig. 4 Transport safeguard for the rotary motion
1. Remove the transport safeguard screw (1/Fig. 4) with spacer to enable rotary motion. The screw and the spacer should be kept for the mounting of the lifting carriage transport safeguard.
2. Mount the red lifting carriage transport safeguard with screw and spacer (1/Fig. 4) on the left hand side of the stand, see Fig. 5. Insert the screw through the slot from the inside of the stand, see Fig. 6
MAM00314
MAM00312
MAM00313
Fig. 5 Transport safeguard mounted on the stand Fig. 6 Screw and spacer inserted from the inside of the
stand
3. Tighten the screw. Make sure that the transport safeguard is parallell to the stand.
M1000/3000 Nova Register 3 SPB7-230.033.11 Page 4 of 8 Siemens AG Installation and Start-Up Rev. 05 05.05 CS PS 24 Medical Solutions
Preparatory work 3 - 5

Ball bearings 3

The lifting carriage is equipped with twelve ball bearings (Fig. 7). Check that all twelve ball bearings are in place and that they show no sign of damage.
Fig. 7 Ball bearings

Protective strips for the metal curtain 3

1
2
MAM00084
Fig. 8 Protective strips for the metal curtain
The edges of the metal curtain are very sharp and are therefore provided with protective strips (1) on delivery. These strips must always be applied onto the edges of the metal curtain during service and maintenance work.
CAUTION
Remove the protective strips before performing vertical adjust­ment of the swivel arm.
If not removed, the protective strips could be damaged.
Make sure the protective strips are removed before adjustments.
Holders (2) for storing the protective strips when not in use, are provided on both sides of the curtain, see Fig. 8.
Siemens AG Register 3 SPB7-230.033.11 Page 5 of 8 M1000/3000 Nova Medical Solutions Rev. 05 05.05 CS PS 24 Installation and Start-Up
3 - 6 Preparatory work

Unpacking the generator and mounting the radiation shield (optional)

1
3
3
4
3
2
MAM00085
Fig. 9 Unpacking the generator
1. Open the cardboard box by cutting the plastic straps.
2. Lift out the lead-glass and put it away in a safe place.
3. Remove the protective cover from the generator.
4. Upend the generator (1/Fig. 9) and lift it off the pallet (2/Fig. 9) (two persons are required).
M1000/3000 Nova Register 3 SPB7-230.033.11 Page 6 of 8 Siemens AG Installation and Start-Up Rev. 05 05.05 CS PS 24 Medical Solutions
Preparatory work 3 - 7
Optional:
5. Insert sheet-metal strip (5/Fig. 10) (additional radiation protection) into both stanchions.
6. Fit the lead-glass (4/Fig. 10) into both vibration absorbing strips (6/Fig. 10) and fix it softly by screws (1/Fig. 10).
NOTE
The screws must be mounted before the lead-glass is inserted in the stanchions.
7. Fit the lead-glass (2/Fig. 10) into the groove of one of the stanchions (3 /Fig. 10) and keep it in position while mounting the other stanchion (3 /Fig. 10).
8. Mount the stanchions and lead-glass unit with 6 screws onto the generator (3, 4 /Fig. 9) (two persons are required).
9. Cover the holes with plastic caps (7/Fig. 10).
6
5
3
74
Fig. 10
MAM00086
Fig. 10 Mounting the radiation shield
3
2
1
1
MAM00630
3
MAM00633
Siemens AG Register 3 SPB7-230.033.11 Page 7 of 8 M1000/3000 Nova Medical Solutions Rev. 05 05.05 CS PS 24 Installation and Start-Up
3 - 8 Preparatory work
This page intentionally left blank.
M1000/3000 Nova Register 3 SPB7-230.033.11 Page 8 of 8 Siemens AG Installation and Start-Up Rev. 05 05.05 CS PS 24 Medical Solutions

Installing the generator and the stand 4

Arranging the components 4

Position the stand and the generator in accordance with installation plan and Fig. 1 below.
640
4 - 1
700
Fig. 1 Installation, top view
180
1091
700 700
1400
4
90
700
540
1
2 3
708
MAM00503
Cable duct (1) and generator (3) with radiation shield (2) can be installed optionally to the left or to the right of the the stand. The cables can also be laid underfloor. Maximum cable length between stand and generator is approximately 3.5 m.
Radiation shield (2) at right angles to the swivel-arm system and centered to the focus mark (4) on the tube-head cover according to Fig. 1.
Recommended minimum distance to wall and generator base plate respectively is 700 mm (shorter distances may be used at the customer´s desire, if in compliance with local regulations).
For the installation of the cable duct, a minimum distance of 180 mm between wall and stands is required.

Notes on installations with separate generator and separate control console

- The separate generator can be installed in any suitable place. The maximum distance of 3.5 m (maximum cable length) to the stand must be maintained, however. If separately installed, the generator must be bolted to the floor alternatively be attached to the wall by means of wall brackets, see further Chapter 21 "Bolting the stand/ generator to the floor".
- The separate control console can be installed in a suitable place (wall-mounted or mounted on a table, for example in an adjacent room with radiation-proof window), or be mounted onto the free-standing radiation shield (option). The connection cable to the generator measures 10 m.
4
Siemens AG Register 3 SPB7-230.033.11 Page 1 of 6 M1000/3000 Nova Medical Solutions Rev. 05 05.05 CS PS 24 Installation and Start-Up
4 - 2 Installing the generator and the stand

Free-standing radiation shield (option) 4

For the installation of the free-standing radiation shield, please refer to separate instruc­tions enclosed with the radiation shield.

Removing the generator cover 4

1. Pry loose the plastic strip (1) and remove the sixteen screws (2), eight on each side.
NOTICE
Pay special attention to the contact washers (there are four con­tact washers on either side).
They are needed again when reassembling the front cover to establish protective ground connection and to fulfil EMC (Electro Magnetic Compatibility) requirements.
2. Remove the front cover (3) from the generator.
2
3
MAM00091
1
Fig. 2 Removing the generator cover
M1000/3000 Nova Register 3 SPB7-230.033.11 Page 2 of 6 Siemens AG Installation and Start-Up Rev. 05 05.05 CS PS 24 Medical Solutions
Installing the generator and the stand 4 - 3

Installing the cable ducts 4

1. Position the cable ducts (1) to the stand and the generator respectively.
2. Leave sufficient space, approx 10 mm between the cable ducts (1) and the base plate of the stand (3) to allow for the mounting of the cable outlet cover (4).
3. Check that the cable duct (1) is slightly overlapped by the cable outlet cover (4).
4
3
Fig. 3 Installing the cable ducts (top view)
1
1
10 mm
6
7
1. Mark the outlines of the cable ducts (1) on the floor.
2. Cut the bottom of the cable ducts (6) and the cable duct cover (5) to proper length.
NOTICE
Fig. 3 shows a right-angled installation of the cable ducts.
5
MAM00089
If the generator is to be installed at an other angle to the stand, you just have to cut the cable ducts correspondingly.
3. Fasten the bottom plate of the cable ducts (6) to the floor according to marking, using double-sided adhesive tape.
4. At the generator, cover the cable entry not used with the enclosed cover plate (7).

Installations with separate control console 4

For the connection cable separate control console – generator, additional cable ducts are required, unless the cables are to be laid underfloor.
Siemens AG Register 3 SPB7-230.033.11 Page 3 of 6 M1000/3000 Nova Medical Solutions Rev. 05 05.05 CS PS 24 Installation and Start-Up
4 - 4 Installing the generator and the stand

Laying the cable harness 4

NOTE
Remove the transport safeguards which consist of cable ties (four around the cable harness and one at the end of each shielded cable).
1. Separate the cable harness.
2. Lay the stand cable harness in the cable duct, as shown in Fig. 4, all the way to the generator.
3. Fit the cable duct covers (6) onto the cable ducts.
1
2
3
4
MAM00092
5
6
1
2
Fig. 4 Laying the cable harness
The cable harness consists of four shielded cables, see Fig. 4.
Pos in Fig. 4 Description
1 High voltage cable H1
2 AEC signal cables X10 and X11
Unit control cable X1 Power supply cable X14
3 Anode rotation cable X9
Filament cable X8
4 Mains supply cables L1, L2
Protective ground
MAM00215
4
3
5 Decompression-button cable (only with separate control console)
M1000/3000 Nova Register 3 SPB7-230.033.11 Page 4 of 6 Siemens AG Installation and Start-Up Rev. 05 05.05 CS PS 24 Medical Solutions
Installing the generator and the stand 4 - 5

Installations with separate control console 4

Lay the connecting cable from the control console in additional cable ducts, alternatively underfloor, to the free cable entry of the generator, see further section on page 5 - 8.
The separate control console can be wall-mounted – horizontally or vertically as shown in Fig. 5 below – or be mounted on a table, either in the examination room or in an adjacent room with radiation-proof window. It is important that the control console be placed so that the operator has good view of the patient from the control console.
Ensure that the control console is attached to the wall with adequate safety margins. Dow­els and screws are not supplied. Please obtain these locally and ensure that they are suit­able for the material used in the wall. The separate control console has a weight of 4 kg.
There are two cable outlets on the separate control console; one at the back and one on the underside. On delivery, the cable is led through the outlet at the back of the control console. For wall-mounted control console, the cable shall be led through the outlet on the underside.
If the control console is to be mounted on a table, the minimum/maximum length of the screws penetrating into the control console must be observed, see Fig. 7 below.
The control console can also be mounted on the free-standing radiation shield (option) as shown in Fig. 6 below, see separate instructions enclosed with the radiation shield.
MAM00168
Horizontal mounting
Fig. 5 Wall-mounted control console
6.5 mm
Plastic cap
Wall-mounted cable duct is recommended
MAM00169
6.5 mm
Vertical mounting
Siemens AG Register 3 SPB7-230.033.11 Page 5 of 6 M1000/3000 Nova Medical Solutions Rev. 05 05.05 CS PS 24 Installation and Start-Up
4 - 6 Installing the generator and the stand
MAM00171
<
min. 10 mm – max. 30 mm
<
Tabl e t op
MAM00170
Fig. 6 Control console on free-standing radiation shield
(option)
Welded nut
M6
Fig. 7 Control console mounted on table

Aligning the stand 4

1. Align the stand vertically by means of the levelling screws (1/Fig. 8). For adjustment of the levelling screws, use a ratchet spanner 1/2" with extension (without socket).
NOTICE
2. Cover the levelling screws with the cover discs supplied.
It is important that all six screws bear against the floor.
This is to ensure optimum stability of the stand.
1
MAM00090
Fig. 8 Levelling screws
NOTE
If bolting to the floor is required, please refer to Chapter 21 "Bolt­ing the stand/generator to the floor" on page 21 - 1.
M1000/3000 Nova Register 3 SPB7-230.033.11 Page 6 of 6 Siemens AG Installation and Start-Up Rev. 05 05.05 CS PS 24 Medical Solutions
Loading...
+ 108 hidden pages