This equipment has been tested and found to comply with the limits for a Class A
digital device, pursuant to Part 15 of the FCC Rules . These l imit s ar e des ig ned
to provide reasonable protection against harmful interference when the
equipment is operated in a commercial environment. This equipment
generates, uses, and can radiate radio frequency energy and, if not installed
and used in accordance with the instruction manual, may cause harmful
interference to radio communications. Operation of this equipment in a
residential area is likely to cause harmful interference in which case the user
will be required to correct the interference at his own expense.
Canada
This digital apparatus does not exceed the Class A limits for radio noise
emissions from digital apparatus set out in the Radio Interference
Regulations of the Canadian Department of Communications.
Le présent appareil numérique n
les limites applicables aux appareils numériques de la Classe A prescrites
dans le Règlement sur le brouillage radioélectrique edicté par le ministère des
Communications du Canada.
’
émet pas de bruits radioélectriques dépassant
Warnings Electrical installation
For your protection, all mains (110V or 230V) equipment used
where damp or wet con dit ions may occu r, must be supplied from a
correctly fused source and protected by an approved ground fault
protection devic e (RCD, GF CI etc .).
IF IN DOUBT SEEK ADVICE FROM A QUALIFIED
ELECTRICIAN.
To avoid the pos sibility of electric shock or damage to the mac hine,
always isolate from the mai ns power supply before carryi ng out any
routine maintenance.
Cleaning the scale
Harsh abrasiv es, solvents, scouring cleane rs and alkaline cleaning
solutions, should not be used especially on the display windows.
Under no circumstances should you attempt to wip e t he i ns i de of t he
machine.
The outside of the machine may be wiped down with a clean cloth
moistened with water contai ning a small am ount of liquid soap.
EMC compliance
The following warning may be appl icabl e to your mac hine .
WARNING: This is a class A product. In a domestic environment
this product may cause radio interference in which case the user
may be required to take adequate measures.
CONTENT
Ⅰ. SPECIFICATIONS. ……………………………………………… P1
Ⅱ. KEY FUNCTION. ……………………………………………… P3
Ⅲ. CALIBRATION. ………………………………….……………… P6
Ⅳ. Examination mode………………………………………………… P9
Ⅴ. SETUP MODE. …………………………………………………… P10
Ⅵ. Details of RS232 Communication………………………………... P17
Ⅶ. Calibration resolution, division and capacity …..………………… P20
Ⅷ. Meaning of some displayed symbols. ……………………………. P23
Ⅸ. Direction of indicator with bracket. ………………………. ……… P24
Ⅹ.Key Definition summary: ………………………. ………………… P24
Ⅺ.Parameters setting summary: ………………………………... … P26
140 Indicator Operation Manual
Thank you for purchasing the 140 indicator. Please read all operating
instructions carefully before use and keep the following points in mind:
* Avoid lengthy exposure to extreme heat or cold, your scale works best when
operated at normal room temperature. Always allow the scale to acclimate to a
normal room temperature before use
* Allow sufficient warm up time. Turn the scale on and wait for a few minutes if
possible, to give the internal components a chance to stabilize before weighing.
* These electronic scales are precision instruments. Do not operate near an in-use
cell phone, radio, computer or other electronic device. These devices emit RF and
can cause unstable scale readings. If your scale ever performs poorly, try moving
the scale to a different room or location.
* Avoid using in condition of heavy vibration and airflow.
* Read the weight reading in short time after loading. The output signature of load
cell and A/D may be little influenced after weighing for a long time.
Ⅰ
. SPECIFICATIONS:
·SCALE INDICATOR:
1. Input signal range: 0mV ~ +30mV
2. Sensitivity: >0.2uV/grad
3. Internal Resolution: Approximately 520,000 counts
4. Display Resolution: can be selected between 500-100,000
When all displayed segments of LCD flashed, this indicates the batteries are
low below 4.9V and you’d better to replace batteries;
When “Lo.bAt” displayed, this indicates the batteries are low below 4.7V and
you should replace batteries immediately.
2. AC Adapter: 6VDC, 500mA, with central negative:
3. Work current: ≤25mA
(when voltage in 5Vdc-8Vdc and not include load cell’s consumption)
·OPERATION TEMPER ATURE: 20℃±15℃ STORE TEMPERATURE: -10℃-70℃
OPERATION HUMIDITY: ≤95%RH (no condensate)
·LOADCELL:
Because of more than one load cell can be used on a scale, following are
required on the load cell set to be used with this indicator,
1. Sensitivity: 0.3mV/V --- 3mV/V (must be fit to >0.2uV/display grad)
2. Input Resistor: ≥80 Ω
3. Output Resistor: <10 KΩ
·LOADCELL WIRING:
PIN 1: RED, EXCITATION +
PIN2: BLACK, EXCITATION –
PIN3: GREEN, SIGNAL -
PIN4: WHITE, SIGNAL +
Ⅱ. KEY FUNCTION:
1. FACEPLATE:
2. DISPLAY SYMBOL MEANING:
Zero◄--------The scale is at zero point and the gross weight is 0 Net ◄---------The display reading is net weight, and the tare weight is not 0. Hold -------The scale is under HOLD mode.
It displays the current live weight when flashed, and the locked reading will
be shown when does not flash and comes steady.
3. SUMMARY of KEY DEFINITION:
3.1 In normal weighing mode press and hold down keys for 3 seconds:
HOLD
PRINT
HOLD
(1)
PRINT
a. If this key is only set for HOLD (P2=0), press this key to enter or exit HOLD
mode.
b. If this key is only set for PRINT (P2=1), press this key to output the data
according to P4 setting.
c. If this key is set for both HOLD and PRINT function (P2=2), press this key to
output the data according to P4 setting.
UNIT TARE ZERO
2
3
UNIT
(2)
Choose weighing units among kg-lb-lb: oz
Note: The weighing units that can be used are restricted by display
division, and calibration weight unit (restricted by P8, P9, and P10):
For example, if the calibration unit is “kg”; calibration display resolution
is 50kg (that means: P8=5, P9=0, P10=0), and users press UNIT key to
choose weighing units. Lb or lb:oz are not allowed to choose, since the
display resolution of 100lb or 2000oz is not available to this indicator.
(3)
TARE
Tare the weight. This function can be activated only when the scale is in
stable mode and the gross weight is not negative value.
ZERO
(4)
Zero function. When the weight is within zero range, it will active as
ZERO function and clear the tare weight. When the weight is not within
zero setting range (P13 set), the scale will show 0¯ ¯ ¯ ¯
the setting range), or 0
_ _ _ _ (zero is below the setting range).
(zero point is over
3.2. In normal weighing mode and hold down time longer than 3s:
HOLD
PRINT
HOLD
(1) Same meaning with that in 3.1
PRINT
UNIT
(2) Same meaning with that in 3.1
UNIT TARE OFF
(3) Same meaning with that in 3.1
TARE
OFF
(4) Power off the scale.
OFF
(5) +
HOLD
PRINT
Hold these two buttons to show firmware version; A/D code or input
working voltage of indicator.
OFF UNIT
(6) +
Hold these two buttons to enter setting mode when the sealed calibration
switch is on.
OFF TARE
(7) +
Hold these two buttons to enter calibration mode when the sealed
calibration switch is on.
3.3 In setting mode:
HOLD UNITS TARE OFF/ON
(1) Rotate the flashed position from right to left
(2) Change the digit on flashed position. The digit can be changed
to 0, 1, 2…9; and be flashed.
(3) Confirm receiving and storing the displayed parameters,
After the setting of the last parameter, the indicator will
not exit the setup mode, and cycles from the first parameters
for viewing or modifications.
(4) Exit from setting mode to normal work mode
3.4 In displaying A/D code or input voltage mode:
X
WT
VOL
FLT EXT
(1) No function
X
4
5
(2) Choose the weight inner code or input working voltage to be
WT
VOL
(3) When A/D code of weight is shown, press this button to choose
FLT
(4) Press this button to exit this mode, and auto-reset the indicator,
EXT
3.5 Calibration mode
HOLD UNITS TARE OFF/ON
(1) Rotate the flashed position from right to left
(2) Change the digit on flashed position. The digit can be changed
(3) Confirm receiving input data and go into next step.
(4) Exit from calibration mode to normal work mode
Ⅲ.
10% of FS weight, and the unit is same as P10 setting) for calibration.
1. Move away any weight on scale. When normal weighing mode, press and
2. When the indicator shows” CAL-?”, the scale is ready for calibration. Press
CALIBRATION:
Before calibrate the scale, you should prepare a standard weight (more than
hold down TARE and ON/OFF/ZERO buttons to enter calibration mode.
displayed. The working voltage range is 4.8V-8V. If the voltage
is not within range, it may damage the indicator. Voltage value
is shown like this “U x.xx” and its unit is V.
filtered or un-filtered weight A/D data; when is on, the data
is filtered.
display all segments of LCD, full capacity… just like power on
again and then it goes back to normal weighing mode.
to 0, 1, 2…9; and be flashed.
TARE to confirm and go to next step, or press ON/OFF/ZERO to exit the
calibration mode.
3. The 140 indicator will display “CAP.--”, that means the following data is the
full capacity according to your setting of display resolution (P7), display
division value (P8) , location of decimal point-dot in calibration unit (P9) and
capacity’s unit in calibration (P10). If the setting of FS is more than 199999
(regardless of decimal point and weight unit), the FS capacity will be shown
by first four digits and last four digits: “Hxxx” and “Lxxx”.
For example, the display resolution is selected to 100000(P7=31), the display
division is selected to 5(P8=2), the position of decimal point is selected to
one point after zero (P9=1). The calibration unit is chosen as lb (P10=1), so
the full capacity 50000.0lb will be shown as H 50 and L000.0 in lb unit. Also,
in other modes, the data will be shown as “Hxxx” and “Lxxx” when current
display data is larger than 199999 kg/lb (not include decimal point).
Press TARE to go to next step directly; press ON/OFF/ZERO to exit the
calibration mode; or after a few seconds, it will automatically to next step.
4. The scale will automatically display the setting of division. Firstly it will
display “d.--”, and then the data according to your setting of P8, P9 and P10.
You may choose division among these as below:
Ta bl e1 :
0.0001kg/lb 0.0002kg/lb 0.0005kg/lb
0.001kg/lb 0.002kg/lb 0.005 kg/lb
0.01kg/lb 0.02 kg/lb 0.05 kg/lb
0.1kg/lb 0.2 kg/lb 0.5 kg/lb
1kg/lb 2 kg/lb 5 kg/lb
10kg/lb 20 kg/lb 50 kg/lb
Press TARE to go to next step directly; press ON/OFF/ZERO to exit the
calibration mode; or after a few seconds, it will automatically to next step.
5. When ‘CAL.P0’ is displayed, that means the scale will begin to calibrate
scale’s zero-point. Move away any weight on the scale. Press TARE button
to confirm, or press ON/OFF/ZERO to exit this mode.
6. When ‘CAL.P1’ is displayed, the scale will be calibrated on second
calibration point. The default standard weight is 50%FS, and at the same, you
6
7
Loading...
+ 12 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.