INTRODUCTION
The 3255 Checkweigher series scales are versatile and easy to use. The large display is allows
you to easily view the weight.
The 3255 allows you to store 10 preset checkweighing target settings, provides a simple counting
feature and is battery operated using 6 “C” cell alkaline batteries (not included).
The 3255 Checkweigher series is NTEP approved at specific capacities and resolutions.
The stainless steel weighing platform and base allows you to wipe down the unit in moist
applications. Please note this unit is not intended to be used in wash down applications. Please
contact your local dealer to to see if your application meets the requirements of this scale.
The 3255 is equipped with a foldable display bracket, which allows you to lower to display
assembly when you are using the unit in portable applications.
GETTING STARTED
1. Remove the scale from it’s packaging. Included is the 120 Volt AC adapter, user manual, scale
base and display assembly.
2. To install the scale raise the display assembly up from the base, adjust the display locking
button into place and make sure the display head is locked into position.
3. To ensure proper weighing make sure the scale is level, using the level bubble on the rear of the
unit. Once level, tighten locking nuts using a 13 mm wrench.
4. Always fit the Adaptor to the scale BEFORE connecting to an electrical outlet. Disconnect from
the electrical outlet BEFORE disconnecting the adaptor from the scale.
5. If installing “C” batteries make sure you do not use the AC adapter.
For technical support contact us at 1-800- 458-7062 or fax us at 1-800-368-2042
sales@salterbrecknell.com
Page 3 of 18
Page 4
BATTERY INSTALLATION
*** WARNING ***
1. Remove Battery Cover Screws.2. Insert 6 “C” cell Alkaline batteries into
compartment, and replace cover.
When using the C3255 as a battery operated scale, do not use the AC adapter!
LOW BATTERY POWER
The 3255 battery life is estimated at 250 hours of continuous use. To conserve battery life, make
sure you have configured the scale for Auto Shutdown, see configuration section.
1. (Low battery indication) will display on LCD display when battery’s power is insufficient to
power the scale.
2. When is displayed, remove old batteries and plug in the AC adapter or replace with new
batteries.
For technical support contact us at 1-800- 458-7062 or fax us at 1-800-368-2042
sales@salterbrecknell.com
Page 5 of 18
Page 6
COUNTING PROCEDURE
The C3255 check weigher includes a simple counting function. The minimum sample size weight
required is 1/5d of scales capacity and resolution.
*Important note, for accurate counting it is suggested to use the largest sample weight as possible.
1. Make sure all weight is removed from platform.
2.If tare is required, press the Tare key prior to sampling to tare out the weight of a empty
container.
3. Press MODE key to enter into counting mode, the display will show:
4. There are four pre-defined samples sizes available in the C3255. They are 10 pieces, 25 pieces,
50 pieces and 100 pieces. Press SET key to scroll to the desired sampling quantity.
5. Once the desired sampling size is chosen, add the appropriate # of pieces onto the platform or
into the empty (tared) container.
* NOTE: If no weight is placed on the scale, and you exit the counting function by pressing the
MODE key, the scale will display an invalid # of pieces.
6. With the sample weight on the scale, the display will flash for 3-5 seconds.
7. Once the sample weight is stable, the stable enunciator will light and the scale will show the
value of the sample weight.
•
• NOTE: If the displayed value of pieces does not equal the sample chosen, repeat steps 1-7
••
and chose a higher sample size value.
8. Press the MODE key to exit counting and return to normal weighing mode. Your sample value
will be saved and stored until you store a new sample size is stored.
NOTE: The counting feature on the Model C3255 is NON TYPE approved.
For technical support contact us at 1-800- 458-7062 or fax us at 1-800-368-2042
sales@salterbrecknell.com
Page 6 of 18
Page 7
STORING CHECK WEIGHING TARGET SETTINGS
The C3255 allows you to store up to 10 target settings, using numbers 1-10. The keyboard is used
to enter in you upper and lower check weight limits. Prior to setting up your target settings, use the
mode key to scroll to the correct units of measure.
1. In normal weighing mode, press numerical key, and then press SET key to enter into a check
weighing setup mode. The display will show as below:
2. Using the numeric keypad, enter in the desired lower check weight, and press SET key. The
display will show as below:
3. Using the numeric keypad, enter in desired upper check weight, and press SET key. The
check weighing setup process is finished.
The display will briefly flash the lower and upper limits for you to make sure the setting were stored
correctly. The scale will now exit from the set-up mode and return to normal check weighing mode.
“03” Indicates you havestoredacheckweighinglimitsforkey3.
4. To set up additional check weight target setting, repeat steps 1-3.
5. To exit check weighing mode, press the MODE key to exit and return to normal weighing mode.
* Note: If you do not enter a check weight limit within 10 seconds the scale will exit target settings
and return to normal weighing mode.
If the upper limit is lower than the lower limit, the buzzer will beep-beep-beep to remind the
operator the setup is incorrect. The device will auto clear the upper check weight and return to the
last step – upper check weight setup.
For technical support contact us at 1-800- 458-7062 or fax us at 1-800-368-2042
sales@salterbrecknell.com
Page 7 of 18
Page 8
USING STORED CHECK WEIGHING TARGET SETTINGS
The Check weighing function can be used both in normal weighing mode and counting mode.
Under normal mode, it stands for weight. While in counting mode, it stands for quantity.
To use a stored check weighing target setting:
1. In normal weighing mode, press the desired numeric key and then press CALL key to enter into
check weighing mode. The scale will briefly show upper and lower check weights, for you to
validate the correct key has been chosen and activate the check weighing function. The following
is displayed if you press the 3 key, then CALL.
2. Load weight on platform.
3. If loaded weight is in the setting weighing range, LCD will show and the buzzer will
continually beep until the item is removed from the platform
.
4. If loaded weight under lower limit, LCD will show to remind the operator the loaded
weight is not within the setting range. The low bar graph segment will change to indicate how
close you are to achieving the acceptable weight.
5. If loaded weight over upper limit, LCD will show to remind the operator the loaded
weight is not within the setting range. The high bar graph segment will change to indicate how
close you are to achieving the acceptable weight.
To exit the check weighing mode, please press 0 key and then press CALL key or press the
MODE key.
Note:
Under check weighing mode, press CALL key to call out the existed upper and lower check limit.
** Shaded areas indicate C3255 capacities available
Units –
Configure each units of measure; LB, KG, oz, to On or Off. When selected units of measure is
turned off, the selection will not appear when pressing the mode key. The default setting is ON
for all units of measure.
DivS –
Select the displayed resolution of the scale for the selected capacity chosen. The default setting
is the NTEP approved capacity and resolution.
A.OFF –
To help save battery life, configure the scale to automatically shutoff after 120, 180, 240, 300
seconds or turns the setting OFF if you have the unit plugged into an AC outlet at all times. The
default setting is ON.
Capacity
LB
CaL_U –
Select the calibration weight used when calibrating your scale. The default setting is Full.
CAL_uSelect the units of measure for calibration, KG or LB. The default setting is LB
Count –
Shows raw counts of load cell and enters into calibration mode
trAd –
Configure the scale for OFF (Non-TYPE approved or sealed) or ON (Select this feature
when
calibration the unit for weights and measures sealing. With the Selection ON, only A.OFF
is configurable through the keyboard configuration menu. Press the sealing switch located
under the battery compartment to enter into all other configuration modes.
End –
Exits the configuration mode and returns to normal weighing mode.
For technical support contact us at 1-800- 458-7062 or fax us at 1-800-368-2042
sales@salterbrecknell.com
Page 9 of 18
Page 10
CONFIGURATION MODE MENU
The scale can be set as desired for the user. Press SET key for 3 seconds to enter setting function
mode. The scale will go into the following configuration mode.
ModE
Use CALL key to scroll
Use SET key to scroll or accept
setting
Unit
oz
Lb
Oz
diviS
See Table 2 for appropriate diviS
On
On
On
A. OFF
120
Off
Off
Off
Cal_ U
Full
180
Cal_u
kg
1/2
240
Count
Displays raw counts
Lb
¼
300
TrAd
OFF
OFF
End
ON
CS_ _ 6
CS_15
CS_30
CS_60
CS_60
CS_150
CS_300See Table 1 for
Mode settings
* Bold and gray Highlighted areas define factory default settings.
Table 1 – Mode Setting
Choose the appropriate mode setting for the capacity of your scale. The capacity sticker is located
on the disdlay.
For technical support contact us at 1-800- 458-7062 or fax us at 1-800-368-2042
sales@salterbrecknell.com
Page 10 of 18
Page 11
Table 2 - Division Setting Table
When setting up the Mode selection, each mode (example CS_15) has 4 separate division setting
which will give you the selection capacity and resolution for KG and LBS. Choose the correct
setting below for your needs.
Selectable
Mode
CS_66.0020 kg6 x 0.002 kg3,000 dN/AN/AN/A
CS_15
CS_30
CS_6060.020 kg60 x 0.02 kg3,000 d150 x 0.05 lb3,000 d+ / - 1 d
CS_150150.050 kg150 x 0.05 kg3000 d300 x 0.1 lb3,000 d+ / - 1 d
Division
Setting
6.0010 kg6 x 0.001 kg6,000 dN/AN/AN/A
6.0005 kg6 x 0.0005 kg12,000 dN/AN/AN/A
6.0002 kg6 x 0.0002 kg30,000 dN/AN/AN/A
15.0050 kg15 x 0.005 kg3,000 d30 x 0.01 lb3,000 d+ / - 1 d
12.0020 kg15 x 0.002 kg7,500 d30 x 0.005lb6,000 dN/A
15.0010 kg15 x 0.001 kg15,000 dN/AN/A
15.0005kg15 x 0.0005 kg30,000 dN/AN/A
30.010 kg30 x 0.01 kg3,000 d60 x 0.02 lb3,000 d+ / - 1 d
30.005 kg30 x 0.005 kg6,000 d60 x 0.01 lbN/A
30.002 kg30 x 0.002 kg15,000 dN/AN/A
30.001kg30 x 0.001 kb30,000 dN/AN/A
60.010 kg60 x 0.01 kg6,000 d150 x 0.02 lbN/A
60.005 kg60 x 0.005 kg12,000 dN/AN/AN/A
60.002 kg60 x 0.002 kg30,000dN/AN/AN/A
120.020 kg120 x 0.02 kg6,000 d300 x 0.05 lb6,000 dN/A
150.010 kg150 x 0.01 kg15,000 dN/AN/AN/A
150.005 kg150 x 0.005 kg30,000 dN/AN/AN/A
Capacity and
Resolution KG
# of divisionsCapacity and
Resolution LB
# of divisionsAccuracy
CS_300300.10 kgN/AN/A
300.05 kgN/AN/A
300.02 kgN/AN/A
300.01 kgN/AN/A
* Grey Shaded areas indicate capacities only available for Model 3255.
** Bold and shaded areas indicate NTEP approved capacities and resolutions only
For technical support contact us at 1-800- 458-7062 or fax us at 1-800-368-2042
sales@salterbrecknell.com
Page 11 of 18
Page 12
CALIBRATION PROCEDURE – NON- SEALED
There are two methods of calibrating the scale. Non –sealed, when the TrAd configuration is set
to off and Sealed, when the TrAd configuration setting is to on. These instructions help you
calibrate the scale, when it is in the non-sealed mode.
Periodically your scale may require calibration to ensure proper weighing. You will need a precalibrated or accurate weight to calibrate the scale. If calibration is performed with a weight other
than a pre-calibrated or accurate weight, it will cause your scale to weigh inaccurately. Scale
calibration is accomplished, when needed, by the following instructions
FUNCTION SCALE RESPONSE
1. Remove all weight from the platform, and
make sure the scale is in a level and stable
environment.
2. Press and hold the SET key for three
seconds to enter into the configuration mode.
Scale displays “Mode”
3. Press the CALL key 4 times to “CAL_U”,
press the Set key. The “CAL_U” function
allows you to set the capacity of the calibration
weight .
Scale displays “FULL”.
4. Press the CALL key to scroll through to
choose 1 / 2 and 1 / 4. See Table 3 for valid calibration
weights required
5. Once you have selected the calibration
weight, press the SET key once.
Scale displays “CAL_U”.
6. Press the CALL key once
Scale displays “CAL_u”.
The “CAL_u” configuration setting allows you to configure the scale to calibrate in pounds or
kilograms.
7. Press the SET key once
Scale displays “lb”
8. Press the CALL key to scroll through to
choose lb and kg. Once you have selected
the calibration weight, press the SET key once.
Scale displays “CAL_u”.
9. Press the CALL key once to scroll to
“Count”. Press the SET key once.
Scale displays “ XXXXXX” (Raw counts
At zero weight)
* Grey Shaded areas indicate capacities only available for Model 3255.
SEALING YOUR SCALE FOR WEIGHTS AND MEASURES APPROVAL
If your scales requires sealing by your local weights and measures authorities, please follow the
instructions shown below to seal your scale.
* IMPORTANT NOTE: Once you have set the scale to the SEALED mode, access to the
configuration menu is limited to changing the “A.OFF” settings only. To enter into the
configuration menu for the other features or to calibrate your scale, you will be required to follow
the steps shown on page X, “Calibration Procedure Sealed”.
FUNCTION SCALE RESPONSE
1. Follow step 1-13 of the “Calibration Procedure Non-Sealed”.
2. Press and hold the SET key for three
seconds to enter into the configuration mode.
Scale displays “Mode”
3. Press the CALL key 7 times to “trad”,
press the Set key.
Scale displays “OFF”.
4. Press the CALL key onceScale displays “ON”.
5. Press the SET key to active the sealing modeScale displays “trAd”
6. Press the CALL key onceScale displays “_END_”.
7. Press the SET key to exit and return to normal weighing mode.
Once you have set the scale to the SEALED mode, access to the configuration menu is limited to
changing the “A.OFF” settings only. To enter into the configuration menu for the other features or
to calibrate your scale, you will be required to follow the steps shown on page X, “CalibrationProcedure Sealed”.
For technical support contact us at 1-800- 458-7062 or fax us at 1-800-368-2042
sales@salterbrecknell.com
Page 17 of 18
Page 18
ERROR MESSAGES
1. Low Battery If this message is displayed, it indicates that
the battery needs to be replaced.
Solution: Replace the battery or plug in the
AC adaptor for operations.
2. Below Zero The scale senses a weight value that is less
than the allowable limit.
Solution: Re-power on the scale again. If
the message continues, the operator should
ask the maintenance men to check the
scale.
3. Over Zero This message indicates that operator have
too much weight on the scale when operator
turned the scale on.
Solution: Remove the excess weight from
the scale platform and turn the scale off and
then back on. If the scale continues to
display this message, call for service.
4. Overload If the weighting capacity exceeds the
maximum capacity of scale, the display will
show the right message.
Solution: Remove the weight from scale
platform right away.
5. Power on zero count over calibration zero
+10%.
Solution: Turn the scale off and then back
on.
6. Power on zero count under calibration zero 10%.