Salter Brecknell C3255 User Manual

Page 1
CHECK WEIGHER SCALE
C3255
OPERATOR’S MANUAL
1000 Armstrong Drive. Fairmont, Minnesota 56031 Tel: 507-238-8702 Fax: 507-238-8271 e-mail:
For technical support contact us at 1-800- 458-7062 or fax us at 1-800-368-2042
sales@salterbrecknell.com
Page 2
INTRODUCTION
The 3255 allows you to store 10 preset checkweighing target settings, provides a simple counting feature and is battery operated using 6 “C” cell alkaline batteries (not included).
The 3255 Checkweigher series is NTEP approved at specific capacities and resolutions.
The stainless steel weighing platform and base allows you to wipe down the unit in moist applications. Please note this unit is not intended to be used in wash down applications. Please contact your local dealer to to see if your application meets the requirements of this scale.
The 3255 is equipped with a foldable display bracket, which allows you to lower to display assembly when you are using the unit in portable applications.
GETTING STARTED
1. Remove the scale from it’s packaging. Included is the 120 Volt AC adapter, user manual, scale base and display assembly.
2. To install the scale raise the display assembly up from the base, adjust the display locking button into place and make sure the display head is locked into position.
3. To ensure proper weighing make sure the scale is level, using the level bubble on the rear of the unit. Once level, tighten locking nuts using a 13 mm wrench.
4. Always fit the Adaptor to the scale BEFORE connecting to an electrical outlet. Disconnect from the electrical outlet BEFORE disconnecting the adaptor from the scale.
5. If installing “C” batteries make sure you do not use the AC adapter.
AC adapter plug
1000 Armstrong Drive. Fairmont, Minnesota 56031 Tel: 507-238-8702 Fax: 507-238-8271 e-mail:
For technical support contact us at 1-800- 458-7062 or fax us at 1-800-368-2042
sales@salterbrecknell.com
Page 2 of 18
Page 3
SPECIFICATIONS
Model # C3255
Default Capacity 15 kg x 0.005kg
30 lb x 0.01 lb
480 oz x
30 kg x 0.01 kg
60 lb x 0.02 lb
960 oz x
60 kg x 0.02 kg
150 lb x 0.05 lb
2400 oz x
150 kg x 0.05 kg
300 lb x 0.1 lb
4800 oz x
Tare Range Full scale capacity range
Units of Measure Kg, lb, oz
Calibration Weight Units Kg, lb
Display 1.2 inch 5 1/2 digital actives
Key functions ON/OFF, MODE, NET/GROSS, ZERO, TARE, CALL, SET, 0 ~ 9
Power supply 6 alkaline C-type batteries / AC adapter (15V 300mA)
Operation Temperature 100 ~ 35
Relative Humidity 250 ~ 95
0
F
0
F
Zero Widow Upon power up, 10% of scale capacity
Zero Key, 2% of scale capacity
Auto Zero Tracking - 2.5d
Dimensions Base: 17.4” L x 13.5” W x 4” H
Base with Display: 17.4” L x 13.5” W x 19.75” H
Approvals NTEP
Material Stainless Steel
1000 Armstrong Drive. Fairmont, Minnesota 56031 Tel: 507-238-8702 Fax: 507-238-8271 e-mail:
For technical support contact us at 1-800- 458-7062 or fax us at 1-800-368-2042
sales@salterbrecknell.com
Page 3 of 18
Page 4
BATTERY INSTALLATION
*** WARNING ***
1. Remove Battery Cover Screws. 2. Insert 6 “C” cell Alkaline batteries into compartment, and replace cover.
When using the C3255 as a battery operated scale, do not use the AC adapter!
LOW BATTERY POWER
The 3255 battery life is estimated at 250 hours of continuous use. To conserve battery life, make sure you have configured the scale for Auto Shutdown, see configuration section.
1. (Low battery indication) will display on LCD display when battery’s power is insufficient to power the scale.
2. When is displayed, remove old batteries and plug in the AC adapter or replace with new batteries.
1000 Armstrong Drive. Fairmont, Minnesota 56031 Tel: 507-238-8702 Fax: 507-238-8271 e-mail:
sales@salterbrecknell.com
For technical support contact us at 1-800- 458-7062 or fax us at 1-800-368-2042
Page 4 of 18
Page 5
FUNCTION KEYS and DISPLAYED ANNUNCIATORS
Stable weight enunciator
Net
Zero enunciator
Key Function
ON/OFF: Switch on / Switch off
MODE: Toggle between weight units mode (lb, oz, kg) and activate the counting mode
Gross
υ O τ
ON/OFF MODE
GROSS/NET
I
B/G
(pcs). Key is also used to exit check weighing and return to normal weighing mode.
ZERO
0
TARE
CALL
SET
T
0
5
1 2
6 87
3
4
9
ZERO: Press this key to return to zero. Zero key will only operate if displayed
weight is less than 2 % of full scale capacity 2%
CALL: In checking weight mode, used to activate a saved target settings. In
configuration mode, used to scroll between setting.
B/G (Gross/Net): Used to toggle between net and gross weight. A bullet on the display
will light to indicate what mode you are in
TARE: Put a container on the scale, press this key to tare the weight. The tare value is stored.
SET: Used to enter into set-up for check point target settings. In counting
mode, used to select sampling quantity. In weighing mode, press and hold the SET key for 3 seconds to enter into configuration mode.
Numerical keys (0~9) Used to call out saved target settings, and enter upper and lower limits.
1000 Armstrong Drive. Fairmont, Minnesota 56031 Tel: 507-238-8702 Fax: 507-238-8271 e-mail:
For technical support contact us at 1-800- 458-7062 or fax us at 1-800-368-2042
sales@salterbrecknell.com
Page 5 of 18
Page 6
COUNTING PROCEDURE
The C3255 check weigher includes a simple counting function. The minimum sample size weight required is 1/5d of scales capacity and resolution.
*Important note, for accurate counting it is suggested to use the largest sample weight as possible.
1. Make sure all weight is removed from platform.
2. If tare is required, press the Tare key prior to sampling to tare out the weight of a empty
container.
3. Press MODE key to enter into counting mode, the display will show:
4. There are four pre-defined samples sizes available in the C3255. They are 10 pieces, 25 pieces,
50 pieces and 100 pieces. Press SET key to scroll to the desired sampling quantity.
5. Once the desired sampling size is chosen, add the appropriate # of pieces onto the platform or
into the empty (tared) container.
* NOTE: If no weight is placed on the scale, and you exit the counting function by pressing the MODE key, the scale will display an invalid # of pieces.
6. With the sample weight on the scale, the display will flash for 3-5 seconds.
7. Once the sample weight is stable, the stable enunciator will light and the scale will show the
value of the sample weight.
NOTE: If the displayed value of pieces does not equal the sample chosen, repeat steps 1-7
••
and chose a higher sample size value.
8. Press the MODE key to exit counting and return to normal weighing mode. Your sample value
will be saved and stored until you store a new sample size is stored.
NOTE: The counting feature on the Model C3255 is NON TYPE approved.
1000 Armstrong Drive. Fairmont, Minnesota 56031 Tel: 507-238-8702 Fax: 507-238-8271 e-mail:
For technical support contact us at 1-800- 458-7062 or fax us at 1-800-368-2042
sales@salterbrecknell.com
Page 6 of 18
Page 7
STORING CHECK WEIGHING TARGET SETTINGS
The C3255 allows you to store up to 10 target settings, using numbers 1-10. The keyboard is used to enter in you upper and lower check weight limits. Prior to setting up your target settings, use the mode key to scroll to the correct units of measure.
1. In normal weighing mode, press numerical key, and then press SET key to enter into a check weighing setup mode. The display will show as below:
2. Using the numeric keypad, enter in the desired lower check weight, and press SET key. The display will show as below:
3. Using the numeric keypad, enter in desired upper check weight, and press SET key. The check weighing setup process is finished.
The display will briefly flash the lower and upper limits for you to make sure the setting were stored correctly. The scale will now exit from the set-up mode and return to normal check weighing mode.
“03” Indicates you have stored a check weighing limits for key 3.
4. To set up additional check weight target setting, repeat steps 1-3.
5. To exit check weighing mode, press the MODE key to exit and return to normal weighing mode.
* Note: If you do not enter a check weight limit within 10 seconds the scale will exit target settings and return to normal weighing mode.
If the upper limit is lower than the lower limit, the buzzer will beep-beep-beep to remind the operator the setup is incorrect. The device will auto clear the upper check weight and return to the last step – upper check weight setup.
1000 Armstrong Drive. Fairmont, Minnesota 56031 Tel: 507-238-8702 Fax: 507-238-8271 e-mail:
For technical support contact us at 1-800- 458-7062 or fax us at 1-800-368-2042
sales@salterbrecknell.com
Page 7 of 18
Page 8
USING STORED CHECK WEIGHING TARGET SETTINGS
The Check weighing function can be used both in normal weighing mode and counting mode. Under normal mode, it stands for weight. While in counting mode, it stands for quantity.
To use a stored check weighing target setting:
1. In normal weighing mode, press the desired numeric key and then press CALL key to enter into
check weighing mode. The scale will briefly show upper and lower check weights, for you to validate the correct key has been chosen and activate the check weighing function. The following is displayed if you press the 3 key, then CALL.
2. Load weight on platform.
3. If loaded weight is in the setting weighing range, LCD will show and the buzzer will continually beep until the item is removed from the platform
.
4. If loaded weight under lower limit, LCD will show to remind the operator the loaded weight is not within the setting range. The low bar graph segment will change to indicate how close you are to achieving the acceptable weight.
5. If loaded weight over upper limit, LCD will show to remind the operator the loaded weight is not within the setting range. The high bar graph segment will change to indicate how close you are to achieving the acceptable weight.
To exit the check weighing mode, please press 0 key and then press CALL key or press the
MODE key. Note:
Under check weighing mode, press CALL key to call out the existed upper and lower check limit.
1000 Armstrong Drive. Fairmont, Minnesota 56031 Tel: 507-238-8702 Fax: 507-238-8271 e-mail:
For technical support contact us at 1-800- 458-7062 or fax us at 1-800-368-2042
sales@salterbrecknell.com
Page 8 of 18
Page 9
CONFIGURATION MODE DEFINTIONS
Below is the definition of the configuration setting and the scale operations they affect
Mode -
Select the capacity of the scale
Mode Capacity
KG
CS_ _6 (6KG) 12 LB CS_15 (15KG) 30 LB CS_30 (30KG) 60 LB CS_60 (60KG) 150 LB CS_150 (150KG) 300 LB CS_300 (300KG) 600 LB
** Shaded areas indicate C3255 capacities available
Units –
Configure each units of measure; LB, KG, oz, to On or Off. When selected units of measure is turned off, the selection will not appear when pressing the mode key. The default setting is ON
for all units of measure.
DivS – Select the displayed resolution of the scale for the selected capacity chosen. The default setting is the NTEP approved capacity and resolution.
A.OFF –
To help save battery life, configure the scale to automatically shutoff after 120, 180, 240, 300 seconds or turns the setting OFF if you have the unit plugged into an AC outlet at all times. The
default setting is ON.
Capacity LB
CaL_U – Select the calibration weight used when calibrating your scale. The default setting is Full.
CAL_u­Select the units of measure for calibration, KG or LB. The default setting is LB
Count –
Shows raw counts of load cell and enters into calibration mode
trAd –
Configure the scale for OFF (Non-TYPE approved or sealed) or ON (Select this feature
when calibration the unit for weights and measures sealing. With the Selection ON, only A.OFF is configurable through the keyboard configuration menu. Press the sealing switch located
under the battery compartment to enter into all other configuration modes.
End –
Exits the configuration mode and returns to normal weighing mode.
.
1000 Armstrong Drive. Fairmont, Minnesota 56031 Tel: 507-238-8702 Fax: 507-238-8271 e-mail:
For technical support contact us at 1-800- 458-7062 or fax us at 1-800-368-2042
sales@salterbrecknell.com
Page 9 of 18
Page 10
CONFIGURATION MODE MENU
The scale can be set as desired for the user. Press SET key for 3 seconds to enter setting function mode. The scale will go into the following configuration mode.
ModE
Use CALL key to scroll
Use SET key to scroll or accept setting
Unit
oz
Lb
Oz
diviS
See Table 2 for appropriate diviS
On
On
On
A. OFF
120
Off
Off
Off
Cal_ U
Full
180
Cal_u
kg
1/2
240
Count
Displays raw counts
Lb
¼
300
TrAd
OFF
OFF
End
ON
CS_ _ 6
CS_15
CS_30
CS_60
CS_60
CS_150
CS_300 See Table 1 for
Mode settings
* Bold and gray Highlighted areas define factory default settings.
Table 1 – Mode Setting
Choose the appropriate mode setting for the capacity of your scale. The capacity sticker is located on the disdlay.
Mode Capacity
KG
CS_ _6 (6KG) 12 LB CS_15 (15KG) 30 LB CS_30 (30KG) 60 LB CS_60 (60KG) 150 LB CS_150 (150KG) 300 LB CS_300 (300KG) 600 LB
Capacity LB
1000 Armstrong Drive. Fairmont, Minnesota 56031 Tel: 507-238-8702 Fax: 507-238-8271 e-mail:
For technical support contact us at 1-800- 458-7062 or fax us at 1-800-368-2042
sales@salterbrecknell.com
Page 10 of 18
Page 11
Table 2 - Division Setting Table
When setting up the Mode selection, each mode (example CS_15) has 4 separate division setting which will give you the selection capacity and resolution for KG and LBS. Choose the correct setting below for your needs.
Selectable
Mode
CS_6 6.0020 kg 6 x 0.002 kg 3,000 d N/A N/A N/A
CS_15
CS_30
CS_60 60.020 kg 60 x 0.02 kg 3,000 d 150 x 0.05 lb 3,000 d + / - 1 d
CS_150 150.050 kg 150 x 0.05 kg 3000 d 300 x 0.1 lb 3,000 d + / - 1 d
Division Setting
6.0010 kg 6 x 0.001 kg 6,000 d N/A N/A N/A
6.0005 kg 6 x 0.0005 kg 12,000 d N/A N/A N/A
6.0002 kg 6 x 0.0002 kg 30,000 d N/A N/A N/A
15.0050 kg 15 x 0.005 kg 3,000 d 30 x 0.01 lb 3,000 d + / - 1 d
12.0020 kg 15 x 0.002 kg 7,500 d 30 x 0.005lb 6,000 d N/A
15.0010 kg 15 x 0.001 kg 15,000 d N/A N/A
15.0005kg 15 x 0.0005 kg 30,000 d N/A N/A
30.010 kg 30 x 0.01 kg 3,000 d 60 x 0.02 lb 3,000 d + / - 1 d
30.005 kg 30 x 0.005 kg 6,000 d 60 x 0.01 lb N/A
30.002 kg 30 x 0.002 kg 15,000 d N/A N/A
30.001kg 30 x 0.001 kb 30,000 d N/A N/A
60.010 kg 60 x 0.01 kg 6,000 d 150 x 0.02 lb N/A
60.005 kg 60 x 0.005 kg 12,000 d N/A N/A N/A
60.002 kg 60 x 0.002 kg 30,000d N/A N/A N/A
120.020 kg 120 x 0.02 kg 6,000 d 300 x 0.05 lb 6,000 d N/A
150.010 kg 150 x 0.01 kg 15,000 d N/A N/A N/A
150.005 kg 150 x 0.005 kg 30,000 d N/A N/A N/A
Capacity and Resolution KG
# of divisions Capacity and
Resolution LB
# of divisions Accuracy
CS_300 300.10 kg N/A N/A
300.05 kg N/A N/A
300.02 kg N/A N/A
300.01 kg N/A N/A
* Grey Shaded areas indicate capacities only available for Model 3255.
** Bold and shaded areas indicate NTEP approved capacities and resolutions only
1000 Armstrong Drive. Fairmont, Minnesota 56031 Tel: 507-238-8702 Fax: 507-238-8271 e-mail:
For technical support contact us at 1-800- 458-7062 or fax us at 1-800-368-2042
sales@salterbrecknell.com
Page 11 of 18
Page 12
CALIBRATION PROCEDURE – NON- SEALED
There are two methods of calibrating the scale. Non –sealed, when the TrAd configuration is set to off and Sealed, when the TrAd configuration setting is to on. These instructions help you calibrate the scale, when it is in the non-sealed mode.
Periodically your scale may require calibration to ensure proper weighing. You will need a pre­calibrated or accurate weight to calibrate the scale. If calibration is performed with a weight other than a pre-calibrated or accurate weight, it will cause your scale to weigh inaccurately. Scale calibration is accomplished, when needed, by the following instructions
FUNCTION SCALE RESPONSE
1. Remove all weight from the platform, and make sure the scale is in a level and stable environment.
2. Press and hold the SET key for three seconds to enter into the configuration mode.
Scale displays “Mode”
3. Press the CALL key 4 times to “CAL_U”, press the Set key. The “CAL_U” function allows you to set the capacity of the calibration weight .
Scale displays “FULL”.
4. Press the CALL key to scroll through to choose 1 / 2 and 1 / 4. See Table 3 for valid calibration
weights required
5. Once you have selected the calibration weight, press the SET key once.
Scale displays “CAL_U”.
6. Press the CALL key once
Scale displays “CAL_u”.
The “CAL_u” configuration setting allows you to configure the scale to calibrate in pounds or kilograms.
7. Press the SET key once
Scale displays “lb”
8. Press the CALL key to scroll through to choose lb and kg. Once you have selected the calibration weight, press the SET key once.
Scale displays “CAL_u”.
9. Press the CALL key once to scroll to “Count”. Press the SET key once.
Scale displays “ XXXXXX” (Raw counts At zero weight)
10. Press the TARE key once to zero the scale.
1000 Armstrong Drive. Fairmont, Minnesota 56031 Tel: 507-238-8702 Fax: 507-238-8271 e-mail:
For technical support contact us at 1-800- 458-7062 or fax us at 1-800-368-2042
sales@salterbrecknell.com
Page 12 of 18
Page 13
CALIBRATION PROCEDURE – NON- SEALED- Conitnued
FUNCTION SCALE RESPONSE
11. Load the appropriate calibration weight. “XXXXXX”
See Table 3 for valid calibration weights used, verify weight placed on platform is with acceptable span range.
12. Press the TARE key to begin calibration “-CAL-“
13. The scale will exit out of calibration mode and return to normal weighing mode. Scale will display weight used for
calibration
If calibration process does not correct any weighing issues, contact the service department for technical support at 1-800-458-7062.
Table 3 - Calibration Weight Table
“FULL” – Full capacity
Mode (Capacity KG) CS_ _6 (6KG) 12 N/A N/A N/A CS_15 (15KG) 30 10 KG 20 LB 80000-120000 CS_30 (30KG) 60 20 KG 40 LB 80000-120000 CS_60 (60KG) 150 50 KG 100 LB 100000-150000 CS_150 (150KG) CS_300 (300KG)
* Grey Shaded areas indicate capacities only available for Model 3255.
“1 / 2” – 1 / 2 Capacity
Mode (Capacity KG) CS_ _6 (6KG) 12 N/A N/A N/A CS_15 (15KG) 30 5 KG 10 LB 40000-60000 CS_30 (30KG) 60 10 KG 20 LB 40000-60000 CS_60 (60KG) 150 20 KG 50 LB 320000-60000 CS_150 (150KG) CS_300 (300KG)
Capacity LB Calibration
Weight KG
300 100 KG 200 LB 80000-120000
600 N/A N/A N/A
Capacity LB Calibration
Weight KG
300 50 KG 100 LB 40000-60000
600 N/A N/A N/A
Calibration Weight LB
Calibration Weight LB
Acceptable Span Range
Acceptable Span Range
* Grey Shaded areas indicate capacities only available for Model 3255.
1000 Armstrong Drive. Fairmont, Minnesota 56031 Tel: 507-238-8702 Fax: 507-238-8271 e-mail:
For technical support contact us at 1-800- 458-7062 or fax us at 1-800-368-2042
sales@salterbrecknell.com
Page 13 of 18
Page 14
Table 3 - Calibration Weight Table - Conitnued
“1 / 4” - 1 / 4 Capacity
Mode (Capacity KG)
Capacity LB Calibration
Weight KG
Calibration Weight LB
Acceptable
Span Range CS_ _6 (6KG) 12 N/A N/A N/A CS_15 (15KG) 30 2 KG 5 LB 16000-24000 CS_30 (30KG) 60 5 KG 10 LB 20000-30000 CS_60 (60KG) 150 10 KG 20 LB 16000-30000 CS_150
300 20 KG 50 LB 16000-24000 (150KG) CS_300
600 N/A N/A N/A (300KG)
* Grey Shaded areas indicate capacities only available for Model 3255.
SEALING YOUR SCALE FOR WEIGHTS AND MEASURES APPROVAL
If your scales requires sealing by your local weights and measures authorities, please follow the instructions shown below to seal your scale.
* IMPORTANT NOTE: Once you have set the scale to the SEALED mode, access to the configuration menu is limited to changing the “A.OFF” settings only. To enter into the configuration menu for the other features or to calibrate your scale, you will be required to follow the steps shown on page X, “Calibration Procedure Sealed”.
FUNCTION SCALE RESPONSE
1. Follow step 1-13 of the “Calibration Procedure Non-Sealed”.
2. Press and hold the SET key for three seconds to enter into the configuration mode.
Scale displays “Mode”
3. Press the CALL key 7 times to “trad”, press the Set key.
Scale displays “OFF”.
4. Press the CALL key once Scale displays “ON”.
5. Press the SET key to active the sealing mode Scale displays “trAd”
6. Press the CALL key once Scale displays “_END_”.
7. Press the SET key to exit and return to normal weighing mode.
Once you have set the scale to the SEALED mode, access to the configuration menu is limited to changing the “A.OFF” settings only. To enter into the configuration menu for the other features or to calibrate your scale, you will be required to follow the steps shown on page X, “Calibration Procedure Sealed”.
1000 Armstrong Drive. Fairmont, Minnesota 56031 Tel: 507-238-8702 Fax: 507-238-8271 e-mail:
For technical support contact us at 1-800- 458-7062 or fax us at 1-800-368-2042
sales@salterbrecknell.com
Page 14 of 18
Page 15
SEALING YOUR SCALE FOR WEIGHTS AND MEASURES APPROVAL
The C3255 requires a lead wire seal to seal the scale. Place sealing mechanism here.
1000 Armstrong Drive. Fairmont, Minnesota 56031 Tel: 507-238-8702 Fax: 507-238-8271 e-mail:
For technical support contact us at 1-800- 458-7062 or fax us at 1-800-368-2042
sales@salterbrecknell.com
Page 15 of 18
Page 16
CALIBRATION PROCEDURE –SEALED
1 & 2
3
4
Breaking the seal to calibrate the scale will require you to re-certify the scale with your local authorized weights and measures inspector.
Follow the steps below to calibrate your scale:
Step
Step
1. Remove lead wire seal.
2. Remove bolts
3. Remove the two Display Mounting locks
4. Remove 4 screws
5. Remove rear display enclosure
Steps
1000 Armstrong Drive. Fairmont, Minnesota 56031 Tel: 507-238-8702 Fax: 507-238-8271 e-mail:
For technical support contact us at 1-800- 458-7062 or fax us at 1-800-368-2042
sales@salterbrecknell.com
Page 16 of 18
Page 17
CALIBRATION PROCEDURE –SEALED - Continued
A sealing label covers the access
The calibration button is located on the display main PDB, underneath the battery compartment.
hole. Remove the label and press the calibration button.
The scale will now be in the configuration mode. The display will prompt “ModE”.
You are now able to configure and calibrate the scale. Review Configuration and calibration sections to procede.
1000 Armstrong Drive. Fairmont, Minnesota 56031 Tel: 507-238-8702 Fax: 507-238-8271 e-mail:
For technical support contact us at 1-800- 458-7062 or fax us at 1-800-368-2042
sales@salterbrecknell.com
Page 17 of 18
Page 18
ERROR MESSAGES
1. Low Battery ­If this message is displayed, it indicates that the battery needs to be replaced. Solution: Replace the battery or plug in the AC adaptor for operations.
2. Below Zero ­The scale senses a weight value that is less than the allowable limit. Solution: Re-power on the scale again. If the message continues, the operator should ask the maintenance men to check the scale.
3. Over Zero ­This message indicates that operator have too much weight on the scale when operator turned the scale on. Solution: Remove the excess weight from the scale platform and turn the scale off and then back on. If the scale continues to display this message, call for service.
4. Overload ­If the weighting capacity exceeds the maximum capacity of scale, the display will show the right message. Solution: Remove the weight from scale platform right away.
5. Power on zero count over calibration zero
+10%.
Solution: Turn the scale off and then back on.
6. Power on zero count under calibration zero ­10%.
Solution: Turn the scale off and then back on.
1000 Armstrong Drive. Fairmont, Minnesota 56031 Tel: 507-238-8702 Fax: 507-238-8271 e-mail:
For technical support contact us at 1-800- 458-7062 or fax us at 1-800-368-2042
sales@salterbrecknell.com
Page 18 of 18