Specification
Classification
(electromedical
equipment)
Regulatory status European Economic Area
Dimensions and weight Dimensions Height: 210mm Width: 305mm Depth: 375mm
Power switch Marking The power switch is marked . A white LED in the center signalizes it is turned on
Power requirements Power supply 100-240 V~ Fusing 2 x T1.6 AH Marking: The mark ~ on
Environmental
conditions
Fluid Ingress In accordance with EN IEC 60601-1/UL 60601-1, the product is marked IP21 to indicate that it is provided with an enclosure that prevents entry of such an amount of
Resistance to
chemicals
Pump specification Nominal vacuum: 95 kPa Nominal free air flow rate: 50 l/min, adjustable to 40 or 60 l/min.
Thermal cut-out The temperature is observed by a sensor in the housing. If the temperature in the housing exceeds 80 °C (176 °F), the software stops the motor and turns on the yellow
Single use suction liner To prevent fluid from being drawn up to the pump the single use suction liner contains a mechanism to shut off the flow when the liner is overfilled. This mechanism is
Suction jar The suction jars can be autoclaved up to 135 ºC or disinfected. The jar is impact resistant.
Suction jar lid The suction jar lid clips and where applicable the float mechanism, are autoclavable up to 135 ºC. The inlet connection (from the endoscope) is marked with the word
Tubing Use only tubing specified in the KV-6 Instructions for Use, otherwise equipment malfunction may result.
Vacuum gauge Range: 0 to 100 kPa Accuracy: ±2.5%
Standards compliance Complies with EN ISO 10079-1 and EN IEC 60601-1 and UL 60601-1
Electro-magnetic
Compatibility
Type of protection against
electrical shock
Degree of protection
against electrical shock
Degree of protection
against explosion
Mode of operation Continuous
(EEA)
Weight 9.3 kg
Frequency 50/60Hz Power rating 120 W
Ambient Temperature Operational: +5ºC to +40ºC (+41ºF to +104ºF)
Relative humidity Operational: 30-75 %
Atmospheric pressure Operational/Storage: 70.0 to 106 kPa
falling liquid as might interfere with its safe and satisfactory operation when correctly positioned.
The external surfaces of the Olympus KV-6 are resistant to: 2% aqueous neutral detergent, 70% ethyl alcohol, isopropyl alcohol, water.
In accordance with ISO 10079-1, the unit is marked with the words ‘HIGH VACUUM, HIGH FLOW’ to indicate it has a vacuum of at least 60 kPa in 10 seconds and a
free air flow rate of 20l/min or greater.
LED. It is possible to restart the motor again as soon as the temperature in the housing drops below 70 °C (158 °F).
activated when contact is made with aspirated fluid.
PATIENT, the outlet connection (to the pump) is marked with the word VACUUM.
This product complies with the requirements of EN IEC 60601-1-2 for emissions and immunity, and as such, its operation is unlikely to
be affected by, or cause interference with, equipment meeting appropriate EMC standards. As a precaution, equipment which may be
sensitive to interference outside the limits specified by EN IEC 60601-1-2 should not be placed in close proximity to the KV-6.
Classification according to EN IEC 60601-1 / UL 60601-1: Class 1 with Type CF applied part.
In accordance with EN IEC 60601-1 and UL 60601-1, the KV-6 is marked with the symbol to indicate the provision of an adequate
degree of protection against electrical shock and that it has an applied part isolated from all other parts of the equipment.
None: the Olympus KV-6 must NOT be used in the zone of risk of flammable anaesthetic gases.
CE This mark on the product indicates compliance with Directive 93/42/EEC relating to medical devices, Class IIa.
the product indicates the
requirement for an AC
power supply
Storage: -20ºC to +50ºC (-4ºF to +122ºF)
Storage: 20-95%
SUCTION PUMP
KV-6
OLYMPUS KEYMED
KeyMed House, Stock Road, Southend-on-Sea,
Essex SS2 5QH, United Kingdom
Telephone: +44 (0)1702 616333 Fax: +44 (0)1702 465677
e-mail: keymed@keymed.co.uk www.keymed.co.uk
sales and distribution locations visit:
For a complete listing of
www.olympus.com
Printed in UK PL022/201505

SUCTION PUMP
KV-6
The KV-6 Suction Pump is the successor to the KV-5 pump. It is designed for endoscopic aspiration and for nonendoscopy applications. The unique flow change buttons allow a fast vacuum build-up with a single touch. It has a 'whisper
mode’ for quiet operation and ‘turbo mode’ for faster suction. The pump connects to an extensive fluid collection system.
Key Benefits
• High Suction Performance
The KV-6 suction pump has vacuum capabilities of up to -95kPa. There are three pre-set speed settings (40l/min, 50l/
min and 60l/min), which enables the operator the flexibility to control the rate of vacuum build-up at a touch of a button.
• Cleaning and hygiene
– Single use collection system allows easy and safe disposal of clinical waste.
– Touch sensitive buttons are easy to clean.
– Disposable bacterial and odour filters.
• Compatibility
Compatible with existing KV-5 accessories and Olympus workstations.
It is ideal for use either mounted on a procedure trolley or as a stand-alone unit. The variable suction control allows
adjustment to meet individual requirements. A high performance pump and motor combination ensure that the unit is
‘whisper quiet’ for any working environment.
Accessories Overview
Suction tubing Patient connecting tubing
K10026130
with 2 coupling
pieces (1 x 90),
7 x 12 mm,
60 cm
Disposable Liners
K10021894
K10021895
1.5 Litre
Jars for Liners
K10021893
1.5 Litre
Filters
K10026804
with 2 coupling
pieces,
7 x 12 mm,
60 cm
K10021897
K10021898
2.5 Litre
K10021896
2.5 Litre
K7505195
MAJ-103,
5 pack
K7503432
50 pack
Reusable Lids
Reusable Jars
K10026127
0.25 Litre
K7505595
(USA only)
50 pack
K7503760
1 & 2 Litre
K10026128
0.25 Litre
K7503762
1 Litre
K7503763
2 Litre
The KV-6 may be adapted to suit individual requirements by the use of optional components, to allow a single-use or
reusable suction system. The single-use liners are conveniently supplied with solidifier for safe and simple use and disposal.
Product Details
Compact design for convenient use
The KV-6's compact and easy-to-use design means that is
suitable for use in office space or the operating theatre.
Simple and easy to operate with clearly labelled
controls
A clear control panel ensures that the KV-6 suction pump
can be used safely and efficiently by all staff.
Trolley mountable for convenience
The KV-6 is ideal for use either mounted on a procedure
trolley or as a stand-alone unit for maximal flexibility of use.
Use with 1.5 or 2.5 litre jars and liners
Flexibility to use the KV-6 with either 1.5 or 2.5 litre jars and
enables its use in a variety of procedures.
Suction jar mountable on Olympus workstations
Ensuring that the suction jar can be mounted on Olympus
workstations allows compatibility with the Olympus range of
devices and equipment for the office or operating theatre.
Microbial and Odour filter for safe suction
Microbial and odour filters ensures that the fluids suctioned
are non-contaminated for maximal safety.
K10027321
Disposable
bacteria /
odour filter
10 pack
K10027322
Disposable
bacteria filter
10 pack
Clamp Holder Two or more jars
Disposable collection system Disposable / reusable collection system
K10021793
K10002819
MAJ-191
MAJ-1659
K10001115
MAJ-190
K7503556
WM-60 / D-60
K10021885
MAJ-1673
K10021188
10 Pack
K10021878
MAJ-1672
K10026130
K10026127
K10026128
K10021882
MAJ-1674