
3
ASSEMBLY
@]
Mounting
©
Refer
to
observation
1.
Remove
in
place
*
After
the
dichroic
the
cover.
©
Dummy
you
wish
When
using
mount
2.
Rotate
the
at
the
position
3.
Using
the
the
cube
toward
4.
Hold
the
it
all
the
screw
©.
5.
Remove
responding
*Use a sharp
to
lift
the
*
The
and
*
When
dummy
observation
However,
dummy
protruding
dummy
6.
Put
back
the
page
18,
section 9 for
methods.
the
cover @ from
magneticaliy.
assembly,
mirror
plates @ are
to
use
only
two
or
the
actual
mirror
of
Allen
screwdriver
clamping
the
perimeter
cube
to
be
way
onto
the
cube’s
turret
object
magnetic
alphabetic
those
at
the
performing
plates
only,
be
sure
plate
screw
plate by
the
cube
Cubes
details
the
mirror
if a cube
cube(s)
cube
the
or
exciter
mounted
one
cube,
more
cubes,
instead.
housing's
cube
is
to
in
mount
mounting
provided
screw
©.
While
of
the
turret,
mounted
the
index
such
cube
letters
turret's
in
is
housing's
with
dovetail
magnetic
fluorescence
the
heads
reversing
sticker
as
the
tip
index
on
the
index
empty
it
is
unnecessary
to
tighten
mounted.
may
index
cover
on
cubes
compatible
cube
housing
be
replaced,
filter
with
your
three
of
the
four
it
in
the
remaining
remove
the
dummy
turret @ to
sliding
remove
its
mount
place
hole.
with
the
microscope
the
it
by
index
side
©.
Then
sticker @ and
position".
of a ballpoint
sticker.
mirror
cube
housing's
positions
observation,
cube
the
If
obstruct
the
procedure
have
positions.
to
mount
cube
clamping
the
screws
turret
©.
(Figs.
with
©.
The
cover
make
sure
not
hands
when
cube
positions.
empty
platels)
the
dummy
frame,
dummy
plate @ slightiy
pulling
it
upward
facing
upward
tighten
the
affix
it
to
pen
or
mechanical
cube
no
importance
make
sure
For
transmitted
dummy
screw
are
left
loose,
rotation.
described
Mount
in
1,23)
different
is
held
to
touch
removing
When
position.
and
then
plate
@
loosen
and
slide
clamping
the
cor-
pencil
dovetail
here.
to
mount
light
plates.
where
no
the
the
steps
2-4.

3
ASSEMBLY
Attaching
Knobs
1.
If a stage
with
Insert
N
on
the
slightly
“
‘Screw
fluorescence
4.
Using
diaphragm
it a little.
σι
Rotate
ease
Mount
o
G
wounting
1.
Slide the
of
the
2.
Insert
x
Before
cooled
©
Compatible
the
and
is
mounted
the
Allen
wrench
the
two
field
fluorescence
before
they
the
field
iris
vertical
the
Allen
screwdriver,
lever's
the
operation
during
observation
the
stage.
the
fluorescence
arrow.
the
filter @ into
replacing
down.
filter
sizes
Field
the
Field
to
and
iris
diaphragm
vertical
can
be
diaphragm
illuminator
operation
knob
is
ND
Filter
vertical
the
filters,
are:
Iris
Diaphragm
Iris
Diaphragm
the
microscope,
remove
centering
illuminator.
turned
lever @ into
until
turn
the
knob © slightly
©).
When a position
found,
tighten
illuminator's
U-shaped
make
Outer
diameter
loosen
the
stage.
Turn
no
further.
the
it
stops.
clamping
the
filter
filter
holder
sure
the
32
Centering
Lever
the
clamping
knobs
(7)
into
their
the
knobs
clockwise
threaded
screw © of
counterclockwise
clamping
slider © in
filter
mm;
hole
the
that
offers
operational
screw @ firmly
the
@).
and
filter
holder
thickness
max. 6 mm
(Fig.
7)
screw
seats
©
until
(4)
in
the
field
iris
to
loosen
(Fig.
8)
direction
has