This instruction manual is for the Olympus System Microscope Model CX41. To ensure the
safety, obtain optimum performance and to familiarize yourself fully with the use of this
microscope, we recommend that you study this manual thoroughly before operating the
microscope. Retain this instruction manual in an easily accessible place near the work
desk for future reference.
AX7255
Page 2
CONTENTS
Correct assembly and adjustments are critical for the microscope to exhibit its full performance. If you are going to
assemble the microscope yourself, please read Chapter 7, “ASSEMBLY” (pages 21 to 24) carefully.
IMPORTANT — Be sure to read this section for safe use of the equipment. — 1-3
3-7 Simplified Phase Contrast Ring Slits CX-PH1/PH2/PH3.......................................................16
2
2
4
3
4TROUBLESHOOTING GUIDE17- 18
5SPECIFICATIONS19
6OPTICAL CHARACTERISTICS20
7ASSEMBLY — See this section for the replacement of the light bulb. — 21-24
PROPER SELECTION OF THE POWER SUPPLY CORD....................................................25-26
Page 3
IMPORTANT
This microscope employs a UIS (Universal Infinity System) optical design, and should be used only with UIS
eyepieces, objectives and condensers, etc. (Other modules described on page 21 may also be usable with this
microscope. For details, please consult Olympus or the catalogue.) Less than optimum performance may result if
inappropriate accessories are used.
SAFETY PRECAUTIONS
1. After the equipment has been used in an observation of a specimen
that is accompanied with a potential of infection, clean the parts
coming in contact with the specimen to prevent infection.
· Moving this product is accompanied with the risk of dropping the
specimen. Be sure to remove the specimen before moving this
product.
· In case the specimen is damaged by erroneous operation, promptly
take the infection prevention measures.
· The product becomes unstable if its height is increased by an
accessory mounted on it. In this case, take anti-topping measures to
Fig. 1
prevent the specimen from being dropped when the product topples
down.
2. To avoid potential shock hazard and fire, always set the main switch @
to “ ” (OFF) and disconnect the power cord from the AC receptacle
at the rear of the microscope and from the wall outlet before replacing
the bulb. Allow the lamp housing cover ² and the bulb to cool before
touching them.
3. Install the microscope on a sturdy, level table.
The air vents on the underside of the base should never be blocked
by placing the microscope on a flexible surface such as a carpet, as
this could result in overheating and cause a fire.
4. Always use the power cord provided by Olympus. If no power cord is
provided, please select the proper power cord by referring to the
section “PROPER SELECTION OF THE POWER SUPPLY CORD” at the
end of this instruction manual. If the proper power cord is not used,
Olympus can no longer warrant the electrical safety and performance
of the equipment.
5. When installing the microscope, route the power cord away from the
microscope base. Should the power cord come in contact with the hot
microscope base, the power cord could melt and cause electric shock.
6. Connect the power cord correctly and ensure that the grounding
terminal of the power supply and that of the wall outlet are properly
connected. If the equipment is not grounded, Olympus can no longer
warrant the electrical safety and performance of the equipment.
7. Never set the main switch @ to “ ” (ON) while any metallic object is
inserted into the air vents of the microscope frame as this will result in
electrical shock, personal injury and equipment damage.
8. When the microscope is not in use or when it is malfunctioning,
disconnect the power cord plug from the AC receptacle or from the
wall outlet.
1
Page 4
Safety Symbols
The following symbols are found on the microscope. Study the meaning of the symbols and always use the equipment in
the safest possible manner.
ExplanationSymbol
IIndicates that the surface becomes hot, and should not be touched with bare hands.
Before use, carefully read the instruction manual. Improper handling could result in personal
injury to the user and/or damage to the equipment.
Indicates that the main switch is ON.
Indicates that the main switch is OFF.
CX41
Warning Label
A warning label is affixed at parts where special precaution is required when handling and using the microscope. Always
heed the warnings.
Warning label
position:
If the warning label becomes soiled, peeled off, etc., contact Olympus to have it replaced.
Getting Ready
1
Fig. 2
Base underside
(Caution for bulb replacement)
1. A microscope is a precision instrument. Handle it with care and avoid
subjecting it to sudden or severe impact.
2. Do not use the microscope where it is subjected to direct sunlight,
high temperature and humidity, dust or vibrations. (For the operating
environment conditions, refer to and adhere to the conditions specified
in Chapter 5, “SPECIFICATIONS” on page 19.)
3. The tension of the coarse focus adjustment knob should only be
adjusted by means of the tension adjustment ring.
4. Heat from the microscope is led away by natural convection.
Consequently, leave an enough space (10 cm or more) on the rear of
the microscope and ensure that the room is well ventilated.
5. When moving the microscope, carefully carry it with one hand under
the base @ and the other hand holding at the recessed handle on the
rear of the arm ² as shown in the illustration on the left.
#Damage to the microscope will occur if you hold it by the stage, X-
axis/Y-axis knob, binocular section of the observation tube, etc.
Also make sure that eyepieces, specimen, filters, etc. do not fall off.
#Sliding the microscope on the surface of the table may damage or
tear off the rubber feet and/or scratch the table top surface.
2
Page 5
Maintenance and Storage
2
1. Clean all glass components by wiping gently with gauze. To remove fingerprints or oil smudges, wipe with gauze
slightly moistened with a mixture of ether (70%) and alcohol (30%).
Since solvents such as ether and alcohol are highly flammable, they must be handled carefully. Be sure to
keep these chemicals away from open flames or potential sources of electrical sparks -- for example,
electrical equipment that is being switched on or off. Also remember to always use these chemicals only in a
well-ventilated room.
2. Do not attempt to use organic solvents to clean the non-optical components of the microscope. To clean them, use a
lint-free, soft cloth lightly moistened with a diluted neutral detergent.
3. Do not disassemble any part of the microscope as malfunction or damage may occur.
4. When not using the microscope, ensure that the frame is cooled down and store it in a locker or cover it with a dust
cover.
5. When disposing of the microscope, check the regulations and rules of your local goverment and be sure to observe
them.
Caution
3
If the microscope is used in a manner not specified by this manual, the safety of the user may be imperiled. In
addition, the equipment may also be damaged. Always use the equipment as outlined in this instruction manual.
The following symbols are used to set off text in this instruction manual.
: Indicates that failure to follow the instructions in the warning could result in bodily harm to the
user and/or damage to equipment (including objects in the vicinity of the equipment).
# : Indicates that failure to follow the instructions could result in damage to equipment.
} : Indicates commentary (for ease of operation and maintenance).
3
This device complies with the requirements of directive 98/79/EC concerning in vitro diagnostic medical devices.
CE marking means the conformity to the directive.
NOTE : This equipment has been tested and found to comply with the limits for a Class A digital device,
pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection
against harmful interference when the equipment is operated in a commercial environment. This
equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in
accordance with the instruction manual, may cause harmful interference to radio communications.
Operation of this equipment in a residential area is likely to cause harmful interference in which case the
user will be required to correct the interference at his own expense.
FCC WARNING : Changes or modifications not expressly approved by the party responsible for compliance
could void the user's authority to operate the equipment.
Page 6
CX41
NOMENCLATURE
}The following illustration shows the CX41RF which is the microscope frame with the X-axis/Y-axis knobs provided on the
right side. The CX41LF is X-axis/Y-axis knobs are provided on the left side.
* The stage is shipped with the two transport pins locked. When using the microscope for the first time, remove the transport
lock pins before use.
}If you have not yet assembled the microscope, read Chapter 7, “ASSEMBLY” on pages 21 to 24.
Eyepieces
Diopter
adjustment ring
(Page 11)
Revolving Nosepiese
Objective
Specimen holder
(Page 9)
Transport lock pin*
Aperture iris
diaphragm knob
(Page 13)
Interpupillar distance scale
(Page 10)
Observation Tube
Dummy slider (Page 7)
Microscope Frame
Transport
lock pin*
Main switch
(Page 7)
Light intensity
control knob
(Page 7)
Condenser
centering screws
(Page 12)
Condenser
Filter holder
Place a 45 mm filter
and the U-POT polarizer
on this.
Field iris diaphragm ring
(Page 7)
Stage
Fine focus
adjustment knob
(Page 8)
Coarse focus
adjustment knob
(Page 8)
X-axis knob (Page 10)
Y-axis knob (Page 10)
4
Page 7
SUMMARY OF BRIGHTFIELD OBSERVATION
PROCEDURE
(Controls Used) (Page)
Set the main switch to “ ” (ON) and adjust
the brightness.
Place the specimen on the stage.
Engage the 10X objective in the light path.ƒRevolving nosepiece
Bring the specimen in focus.…Coarse/fine focus adjustment knobs(P. 8)
} Copy the observation procedure pages on a separate sheet and post it near your microscope.
6
Page 9
USING THE CONTROLS
3-1 Base
Turning On the Bulb(Fig. 3)
1
1. Set the main switch @ to “ ” (ON).
2. Turn the light intensity knob ² clockwise in the direction of the arrow
to make the illumination brighter or counterclockwise to make it darker.
The numbers around the knob indicates the reference voltage values.
Fig. 3
Field Iris Diaphragm(Fig. 4)
2
Using the field iris diaphragm ring @, adjust the diameter of the field
iris for objective power to the extent that it just circumscribes the field
of view. When the field iris diaphragm is stopped down to circumscribe
the field of view, it will exclude extraneous light and improve image
contrast within the field of view.
#When using the 100X objective, the field iris diaphragm image will
not be visible within the field of view. Accordingly, stop down the
diaphragm to its smallest diameter.
Fig. 4
Dummy Slider
3
The dummy slider provided with the microscope frame can be used to
accommodate the optional transmitted light analyzer (U-ANT).
By preparing a transmitted light polarizer (U-POT) and polarizing light
condenser (CH3-CDP), simple polarized light observation becomes
possible.
Also refer to the instruction manual of the CX-POL.
7
Page 10
3-2 Focusing Block
Fig. 5
Adjusting the Coarse Adjustment Knob Tension(Fig. 5)
1
1. The coarse adjustment knob tension is preadjusted for easy use.
However, if desired, one can change the tension using the tension
adjustment ring @. Applying a large flat-bladed screwdriver to any of
the grooves ² on the circumference of the ring, turning the ring
clockwise (in the direction of the arrow) increases tension, and vice
versa.
2. The tension is too low if the stage drops by itself of focus is quickly lost
after adjustment with the fine adjustment knob ³. In this case, turn the
ring in the direction of the arrow to increase tension.
Pre-focusing Lever(Fig. 6)
2
CX41
Fig. 6
The pre-focusing lever ensures that the objective does not come in
contact with the specimen and simplifies focusing.
After focusing on the specimen with the coarse adjustment knob, turn
this lever @ clockwise (in the direction of the arrow) and lock; the
upper limit on coarse adjustment movement is set at the locked
position.
}Focusing using the fine adjustment knob is not affected by the pre-
focusing dial. Accordingly, after using the coarse adjustment knob to
lower the stage for changing specimens or applying immersion oil
(see section 3-6), refocusing is easily accomplished by rotating the
coarse adjustment to reach the pre-focusing position, then making fine
adjustments with the fine adjustment knob.
#When not required, leave the pre-focusing dial unlocked.
8
Page 11
3-3 Stage
Fig. 7
Placing the Specimen(Fig. 7)
1
#Releasing the curved finger with great force or suddenly releasing
your grip on the curved finger knob @ while releasing the curved
finger will crack or damage the slide glass. Always place the
specimen with great care.
Observation with Specimen Holder for Single Specimen Slide
1. Turn the coarse adjustment knob ² counterclockwise (in the direction
of the arrow) to lower the stage.
2. Open the spring-loaded curved finger ³ on the specimen holder and
place the specimen slide into the specimen holder from the front.
3. After placing the slide as far as it will go, gently release the curved
finger ³.
Observation with Specimen Holder for Two Specimen Slides
1. Place the first specimen slide as described in steps 1 and 2 above,
then place the second specimen slide so that it contacts the first
specimen slide.
2. Gently release the curved finger ³.
Observation by Placing the Specimen Slide with One Hand
Place the specimen slide at the front of the stage, then slide the
specimen slide on the stage surface to slowly and gradually open the
curved finger in the direction of the arrow. Insert the specimen slide
into the specimen holder until it is fully and properly seated in the
specimen holder.
· Cover Glass
Use cover glasses of 0.17 mm thickness in order to allow the
objectives exhibit their full performances.
· Specimen Slide
Use specimen slides of 0.9 to 1.4 mm thickness. Using thicker
specimen slides may result in inaccurate imaging of the field iris
diaphragm image on the specimen.
9
Page 12
Fig. 8
Moving the Specimen(Fig. 8)
2
Turn the upper knob which is the Y-axis knob @ to move the
specimen in the vertical direction, and turn the lower knob which is the
X-axis knob ² to move it in the horizontal direction.
#Do not use the specimen holder or stage to move the specimen,
for this will damage the rotating mechanisms of the knobs.
#When the stage and specimen holder are stopped by the pre-
focusing dial, the tension of the X-axis/Y-axis knobs increases. In
this case, do not attempt to turn the knobs to move the stage
beyond the stopped position.
CX41
3-4 Observation Tube
Fig. 9
Fig. 10
Adjusting the Interpupillar Distance(Fig. 9)
1
Be careful not have your finger caught by the clearance below the
binocular tube during interpupillary distance adjustment.
While looking through the eyepieces, adjust for binocular vision until
the left and right fields of view coincide completely. The index dot ·
indicates the interpupillary distance.
}Note your interpupillary distance so that it can be quickly duplicated.
Adjusting the Tilt (U-CTBI only)(Fig. 10)
2
}The tilting observation tube can be adjusted to a height and angle
suitable for ease of observation.
Hold the binocular section with both hands and move it up or down
as desired.
#To prevent damage to the mechanism, do not apply an excessive
force above or below the upper and lower stop positions.
Be careful not to have your finger by the clearance @ on the cover
during tilt adjustment.
10
Page 13
Fig 11
Adjusting the Diopter(Fig. 11)
3
}When using the U-CTBI, align the white marking on the right eyepiece's
diopter adjustment ring scale with the index line.
1. Looking through the right eyepiece with your right eye, rotate the
coarse and fine adjustment knobs to bring the specimen into focus.
2. Looking through the left eyepiece with your left eye, turn the diopter
adjustment ring @ to focus on the specimen.
Using the Eye Shades(Fig. 12)
4
When Wearing Eyeglasses
Use with the eye shades in the normal, folded-down position. This will
prevent the eyeglasses from contacting and scratching the eyepieces.
When Not Wearing Eyeglasses
Fig. 12
Fig. 13
Extend the folded eye shades in the direction of the arrow for efficient
use of the eyeshades by preventing extraneous light from entering
between the eyepieces and eyes.
Light Intensity Ratio of Trinocular Tube U-CTR30-2
5
The U-CTR30 does not have the light path switching facility and the
light intensity ratio is fixed at binocular tubes (50%): TV/photography
(50%).
Photomicrography/TV Observation(Fig. 13)
6
1. Using the provided Allen wrench, fully loosen the clamping screw @
on the straight photo tube mount on the trinocular tube U-CTR30-2.
2. Attach the U-SPT straight photo tube ² or a TV adapter on the mount
and tighten the clamping screw @.
3. Attach the photomicrography system or TV camera.
11
Page 14
Fig. 14
Using the Eyepiece Micrometer Disk (Optional) (Fig. 14)
7
}Prepare one eyepiece micrometer disk (diameter 20.4 mm, thickness 1
mm) and two 20.4-RH reticle holders (available as 2-piece set).
The field number becomes 19.6 when the reticle holders are used.
When the tilting observation tube U-CTBI is used, the field number is
fixed at 18 and does not change.
}If your eye is poor in eyesight and cannot focus the micrometer, use
eyeglasses to correct your vision. (Dioptric correction of the U-CTBI is
possible with the diopter adjustment ring of the eyepiece.)
1. Remove both eyepieces. (With the U-CTBI, remove only the right
eyepiece by loosening the clamping screw using a small Phillips
screwdriver.)
2. Hold the micrometer disk @ with the side with indications facing down
and place it into a reticle holder ².
#Be careful not to leave dirt on the micrometer disk, as it will be
noticeable during observation.
3. Screw the reticle holder ² containing the micrometer disk @ into the
bottom of an eyepiece.
Be sure to screw in all the way by hooking your nail into the notch ³
on the holder at the end.
4. Screw the other reticle holder, alone, into the other eyepiece in order to
align the field number.
5. Attach the eyepieces again.
CX41
3-5 Condenser
Fig. 15
Field iris diaphragm image
Field of view through eyepieces
Centering the Field Iris Diaphragm(Figs. 15 & 16)
1
1. With the 10X objective engaged and the specimen brought into focus,
turn the field iris diaphragm ring @ counterclockwise to stop down the
diaphragm to near its minimum size.
2. Turn the condenser height adjustment knob ² to bring the field iris
diaphragm image into focus.
3. Rotate the two auxiliary lens centering knobs ³ to adjust so that the
field iris diaphragm image is centered in the eyepiece field of view.
(Figs. 15 & 16)
4. To check centration, open the field iris diaphragm until its image
touches the perimeter of the field of view. If the image is not precisely
inscribed in the field of view, center again. (Fig. 16)
5. When used for actual observation, open the field iris diaphragm until its
image is slightly larger than the field of view.
Fig. 16
12
Page 15
Aperture iris
diaphragm image
Objective pupil
Fig 17
Fig. 18
70-80%
30-20%
Aperture Iris Diaphragm(Figs. 17 & 18)
2
· The aperture iris diaphragm determines the numerical aperture of the
illumination system. Matching the numerical aperture @ of the
illumination system with that of the objective provides better image
resolution and contrast, and also increases the depth of focus.
· Since the contrast of microscope specimens is ordinarily low, setting
the condenser aperture iris diaphragm to between 70% and 80% of
the N.A. of the objective in use is usually recommended. If necessary,
adjust the ratio by removing the eyepieces and looking into the
eyepiece sleeves while adjusting the aperture iris diaphragm knob ²
until the image shown in Fig. 17 is seen. (Fig. 18)
#If the aperture iris diaphragm is set too small, image ghost may be
observed.
Using the numerical aperture scale:
}Align the aperture iris diaphragm knob ² with the NA value @ on the
scale. The scale value should correspond to the NA number engraved
on the objective.
Since the aperture iris diaphragm knob has a certain width, align the
center line engraved on the knob with the scale indication.
When using a 100X objective, turn the aperture iris diaphragm knob ²
all the way to 0.9 on the scale. (Fig. 18)
Fig. 19
Fig. 20
Using Filters(Fig. 19)
3
}You can place a filter in the light path with either method.
· Insert a filter with diameter of 32.5 mm @ in the condenser accessory
(CH2-FH or CX-AL) attached below the condenser.
· Insert one or more filter with a diameter of 45 mm ² on the light exit
glass on the microscope base.
}For the types of the filters, please consult Olympus or its catalogues.
Using Darkfield Ring CH2-DS(Fig. 20)
4
}The CH2-DS can be attached in the same way as inserting the 32.5
mm filter.
Usage
Inserting the darkfield lens makes possible darkfield observations
using a 4X to 40X objective.
13
Page 16
5
Fig. 21
Compatibility of Option Units and Condensers
6
CX41
Using Low-Power Light Adjustment Objective CX-LA
}The CX-LA is the lens designed for providing the illumination covering
the illumination field of the 2X objective. The CX-LA can be attached
below a specified condenser (see page 23).
#The CX-LA is designed exclusively for use in observation. As the
aperture iris diaphragm cannot be seen and field iris diaphragm
cannot be used when the CX-LA is used, set both iris diaphragms
to the fully open conditions.
}When the CX-LA is attached on one position of the CX-SLC slide
condenser, observations using the 2X objective and using other
objective can be switched with a one-touch operation.
Option Unit Name
Brightfield/darkfield
phase contrast unit
CX-PCD
Darkfield condenser
CX-DCD
Simplified polarized observation set
CX-POL
Including:
· Condenser CH3-CDP
· Analyzer U-ANT
· Polarizer U-POT
Slide condenser
CX-SLC
Compatibility
The phase contrast can be centered.
The field iris diaphragm can be
centered using the provided auxiliary
lens.
Centration of dark field is possible
using the centering knobs on the
condenser.
The field iris diaphragm can be
centered by preparing the CX-POL set
and attaching the auxiliary lens for the
CH3-CD condenser below the CH3CDP.
The field iris diaphragm can be
centered by attaching the CX-AL
auxiliary lens.
Remark
#The field iris diaphragm is
invisible when the 40X or
100X objective is used.
#Polarized light observation is
not available when the
standard CH3-CD condenser
is used.
#The field iris diaphragm is
invisible when the CX-LA lowpower light adjust-ment lens
is used.
14
Page 17
3-6 Immersion Objectives
1
Fig. 22
Using the Immersion Objectives(Fig. 22)
#Be sure to use the provided Olympus immersion oil.
1. Focus on the specimen by switching the objectives fro the lowest
power to highest power.
2. Before engaging the immersion objective in the light path, place a
drop of immersion oil provided with the 100X objective combination
model onto the specimen at the area to be observed.
3. Turn the revolving nosepiece to engage the immersion objective, then
focus using the fine adjustment knob.
#Since air bubbles in the oil will affect the image quality, make sure
that the oil is free of bubbles.
a. To check for bubbles, remove the eyepieces and fully open the field
and aperture iris diaphragms, then look at the exit pupil of the
objective inside the observation tube. (The pupil should appear round
and bright.)
b. To remove bubbles, turn the revolving nosepiece to move the oil
immersion objective back and forth a few times.
}If the condenser engraving shows a numerical aperture (NA) of 1.0 or
higher, the number applies only when oil is applied between the slide
glass and the top surface of the condenser. When oil is not present,
the NA is about 0.9.
4. After use, remove oil from the objective front lens by wiping with gauze
slightly moistened with an ether (70%)/alcohol (30%) mixture.
Caution in use of the immersion oil
If immersion oil penetrates in your eye or attaches to your skin,
apply the following treatment immediately.
Eye: Rinse with fresh water (for more than 15 minutes)
Skin: Wash with water and soap.
When the appearance of your eye or skin is altered or pain
continues, immediately consult the doctor.
15
Page 18
3-7 Simplified Phase Contrast Ring Slits CX-PH1/PH2/PH3
Appearance
1
CX41
Ring Slits
CX-PH1/PH2/PH3
Centering knobs
Phase Contrast Objectives
PlanCN-Ph series (10X, 20X, 40X, 100XO)
Installation
2
Attach a ring slit in the same way as a filter holder.
#Be sure to replace the objective with a phase contrast objective.
Green Filter
45G533 or 45IF550
Centering Telescope CT-5
Before adjustmentAfter adjustment
Fig. 23
Fig. 24
Operation(Figs. 23 & 24)
3
1. Turn the revolving nose piece to engage in light path the phase
contrast objective lens with the same value as the ring slit in use.
2. Place the specimen and bring it in approximate focus.
3. Remove the eyepieces and replace them with the CT-5 centering
telescope.
4. Turn the upper ring of the CT-5 to adjust the focus so that the bright
ring (ring slit) @ and dark ring (objective’s phase plate) ² are seen
clearly in the field of view. (Fig. 23)
5. Rotate the two centering knobs ³ so that the bright and dark rings
overlap concentrically. (Figs. 23 & 24)
6. Remove the CT-5, replace it with the eyepieces and start phase
contrast observation.
7. To increase the image contrast, place the green filter (45 mm diam.) on
the light exist glass of the microscope.
16
Page 19
TROUBLESHOOTING GUIDE
Under certain conditions, performance of the unit may be adversely affected by factors other than defects. If problems
occur, please review the following list and take remedial action as needed. If you cannot solve the problem after
checking the entire list, please contact your local Olympus representative for assistance.
ProblemCauseRemedyPage
1. Optical System
a) Field of view is obscured or not
evenly illuminated
Revolving nosepiece is not correctly
engaged.
Condenser is not correctly attached.Re-insert it all the way without tilt.22
Field iris diaphragm is not properly
centered.
Field iris diaphragm is stopped down
too far.
Make sure that revolving nosepiece
clicks properly into place.
Center it.
Open it to an optimum stop position.
–
12
12
Dirt/dust on objective, eyepieces,
condenser or light exit glass.
Dummy slider is not correctly
engaged.
b) Dirt or dust is visible in the field of
view.
c) Image shows diffraction.Condenser is lowered too far.Adjust the condenser height position.12
d) Visibility is poor.
· Image is not sharp.
· Contrast is poor.
· Details are indistinct.
e) One side of image is blurred.
Image seems to waver.
Dirt/dust on light exist glassClean thoroughly.
Dirt/dust on top lens of condenser
Dirt/dust on the specimen
Dirt/dust on eyepiece
Aperture iris diaphragm is stopped
down too far.
Objective is not engaged correctly in
light path.
Dirt/dust on front lens of objective.
Immersion oil is not being used with
an oil immersion objective.
Immersion oil contains bubbles.Remove bubbles.15
Recommended immersion oil is not
used.
Dirt/dust on specimen.Clean it.
Dirt/dust on condenser.
Objective is not engaged correctly in
light path.
Specimen is not correctly mounted
on stage.
Clean them.
Make sure that dummy slider clicks
properly into place.
Open it.
Make sure that revolving nosepiece
clicks into place correctly.
Clean it thoroughly.3
Use immersion oil.
Use the specified immersion oil.
Make sure that revolving nosepiece
clicks into place correctly.
Place specimen correctly on top of
stage and secure it with specimen
holder.
13
15
15
3
7
3
–
3
–
9
17
Page 20
ProblemCauseRemedyPage
2. Coarse/Fine Focus Adjustment
a) Coarse adjustment knob is hard to
turn.
b) Stage drifts down by itself or focus
is lost during observation.
c) Coarse adjustment will not go all
the way up.
d) Coarse adjustment will not go all
the way down.
e) Objective makes contact with
specimen before focus is
obtained.
3. Observation Tube
Field of view of one eye does not
match that of the other.
4. Stage
Image blurs as you move specimen.
5. Objective Change
Front lens of a high power objective
comes into contact with specimen
when it is engaged after a low-power
objective.
Tension adjustment ring is
overtightened.
You are trying to raise stage with
coarse adjustment knob even
though pre-focusing lever is kept
locked.
Tension adjustment lever is too
loose.
Pre-focusing lever is keeping the
stage down.
Condenser holder is too low.
Specimen is mounted upside down.Mount specimen correctly.
Interpupillar distance is incorrect.Adjust interpupillary distance.
Incorrect diopter adjustment.Adjust diopter.11
Different eyepieces are used on left
and right.
Your view is not accustomed to
microscope observation.
Specimen is not correctly positioned
on the stage.
Specimen is mounted upside down.Mount specimen correctly.
Cover glass is too thick.Use 0.17 mm thick cover glass.
Loosen it.
Unlock pre-focusingb lever
Tighten it.
Unlock pre-focusing lever.
Raise condenser holder.
Change on eyepiece to match the
other so that both sides are the
same.
Upon looking into eyepieces, try
looking at overall field before
concentrating on specimen range.
You may also find it helpful to look
up and into distance for a moment
before looking back into microscope.
Mount specimen correctly by
applying it on stage surface and
inserting below specimen holder.
CX41
8
8
8
8
–
–
10
–
–
9
–
9
6. Electrical System
a) Bulb does not light.
b) Bulb burns out almost immediately. Wrong type of bulb is used.Use correct bulb type.22
Bulb is not mounted.Mount designated bulb.
Bulb is burnt out.Replace bulb.
Power cord plug is not connected.Connect power cord.
22
22
24
18
Page 21
SPECIFICATIONS
SPECIFICATIONS
ItemSpecification
1. Optical system
2. IlluminationIlluminator built in.
3. Focusing
4. Revolving nosepiece
5. Observation tube
UIS (Universal Infinity System) optical system
6V 30W halogen bulb (PHILIPS 5761)
(Average service time: Approximately 100 hr. when used as directed)
100-120 V/220-240 V , 0.85/0.45 A, 50/60 Hz
Stage height movement by roller guide (rack & pinion)
Stroke per rotation: 36.8 mm
Full stroke range: 25 mm
Upper limit stopped by simplified pre-focusing lever
Tension adjustment on coarse focus adjustment knob.
233(W) x 432(H) x 367.5(D) mm, approx. 6.9kg(15.2 lb.) (microscope frame only)
· Indoor use
· Altitude: Max. 2000 meters
· Ambient temperature: 5°C to 40°C (41° to 104°F)
· Maximum relative humidity: 80% for temperatures up to 31°C (88°F), decreasing linearly
through 70% at 34°C (93°F), 60% at 37°C (99°F), to 50% relative humidity at 40°C (104°F).
· Supply voltage fluctuations; Not to exceed ±10% of the normal voltage.
· Pollution degree: 2 (in accordance with IEC60664)
· Installation/Overvoltage category: II (in accordance with IEC60664)
Abbe condenser (with built-in daylight filter)
76 mm (H) x 50 mm (V)
2-slide holder
1.25 (with oil immersion)
Built in
19
Page 22
OPTICAL CHARACTERISTICS
CX41
The following table shows the optical characteristics of
combinations of eyepieces and objectives. The figure
on the right shows the performance data engraved on
the objectives.
Magnification
Mechanical
tube length
Objective designation
Numerical aperture
Color band
Characteristics
Magnification
Objective
Plan CN
Plan Achromat
(FN 22)
*Field numbers
· With the U-CBI30-2/U-CTR30-2, the field number is 20 when the WHB10X eyepieces are used. The actual field is indicated by
the figure before ( ).
· With the U-CTBI, the field number is 18 when the provided dedicated 10X eyepieces are used. The actual field is indicated by
the figure inside ( ).
4X
10X
40X
100XO
N.A.
0.10
0.25
0.65
1.25
W.D.
(mm)
18.5
10.6
0.6
0.13
Cover
Glass
Thickness
–
–
0.17
–
Resolution
(µm)
3.36
1.34
0.52
0.27
Total Mag.
40X
100X
400X
1000X
Eyepieces
10X (FN*)
Depth of
3.04
0.69
Cover glass thickness
Field
Focus
(µm)
17 5. 0
28.0
of
View
5.0(4.5 )
2.0(1.8 )
0.5(0.45)
0.2(0.18)
Remark
(optional)
Glossary
Working distance (W.D.):
Numerical aperture (N.A.):
Resolving power:
Focal depth:
Field number:
Field of view diameter:
Total magnification:
The distance from the cover glass surface to the nearest point of the objective.
The N.A. value represents a performance number which can be compared to the relative
aperture (f-number) of a camera lens. The higher N.A., the higher the resolving power.
The ability to differentiate two points, i.e., the minimum distance by which the objects must be
separated in order to be revealed as two separate objects.
The depth in the image through which the focused image will appear uniformly sharp. As the
aperture iris diaphragm is stopped down, the focal depth becomes greater. The greater the N.A. of an
objective, the shorter the focal depth.
A number that represents the diameter in mm of the image of the field diaphragm that is formed by
the lens in front of it.
The actual size of the field of view in millimeters.
Equals the objective magnification multiplied by the eyepiece magnification.
20
Page 23
ASSEMBLY
7-1 Assembly Diagram
The diagram below shows the sequence of assembly of the various modules. The numbers indicate the order of assembly.
The module numbers shown in the following diagram are merely the typical examples. For the modules with which the
module numbers are not given, please consult your Olympus representative or the catalogues.
#When assembling the microscope, make sure that all parts are free of dust and dirt, and avoid scratching any parts or
touching glass surfaces.
Assembly steps enclosed in will be detailed on the subsequent pages.
}All assembly operations are possible by using the Allen wrench ( ) provided with the microscope.
1. Turn the microscope frame on its side and pull the lamp housing knob
@ on the underside of the base to open the lamp housing cover.
2. Holding the halogen bulb ² contained in the polyethylene bag to
avoid leaving fingerprints on the bulb, fully insert the contact pins into
the bulb socket ³ When properly seated, pull off the polyethylene bag.
Always use the designated bulb. Use of improper bulb may result
in fire.
Do not touch the bulb with bare hands. If fingerprints are
accidentally left on the bulb, wipe the bulb with a soft, lint-free
cloth moistened with alcohol. Using a contaminated bulb will
shorten the service life of the bulb.
3. With the lamp housing knob still pulled out, close the lamp housing
cover. Then push in the lamp housing knob to lock the cover.
#The lamp housing cover cannot be closed if the knob is pushed in
before you attempt to close the cover.
CX41
Fig. 26
Caution for Bulb Replacement During Use or Right After Use
Whenever you replace the bulb during use or right after use, first
move the main switch to “ ” (OFF), disconnect the power cord
from the wall outlet, and allow the bulb and parts around the bulb
to cool before touching.
#If the bulb burns out during an observation and needs to be
replaced, remove eyepieces, specimen, filter and other objects
likely to fall off, before tilting the microscope frame to replace the
bulb.
Mounting the Condenser(Fig. 26)
5
1. Turn the coarse adjustment knob @ to raise the stage to its highest
position.
2. Turn the condenser height adjustment knob ² to lower the condenser
holder to the position where the clamping screw ³ can be turned.
3. Insert the condenser | all the way into the mounting hole ƒ and
clamp by tightening the clamping screw ³. Positioning the condenser
with the diaphragm lever
4. Turn the condenser height adjustment knob ² to raise the condenser
holder to its highest position.
… at the front facilitates operation.
22
Page 25
Fig. 27
Fig. 28
Attaching the Condenser Accessory(Figs. 27 & 28)
6
} 32.5 mm filter @ (32.5C, 32.5G533, 32.5LB45/150/200) can be
inserted in the CH2-FH or CX-AL ².
1. Push in the condenser accessory ³ all the way into the underside of
the condenser until it clicks.
2. When attaching CX-AL, push it in so that the centering screw |
faces the front of the microscope to facilitate the manipulation.
Fig. 29
Fig. 30
Attaching the Cord Hanger (CH3-CH)(Figs. 29 & 30)
7
}By attaching the CH3-CH cord hanger on the rear of the microscope
frame, the power cord can be accommodated by winding around the
hanger.
Insert the hook ² of the cord hanger @ into the air vent groove on the
rear of the microscope by aligning the attaching position ³, and
clamp by sliding the cord hanger downward while pushing it against
the microscope with a strong force.
#Do not hold the microscope frame by the cord hanger when
carrying the microscope. Otherwise, the cord hanger may be
detached during transport, resulting in falling of the microscope
and personal injury.
Removal
To prevent electric shock, disconnect the power cord first. Also
make sure you use the provided Allen wrench, not a thin Allen
wrench.
Move the microscope frame to the edge of the table |, apply an
Allen wrench
move the entire hanger upward by pushing the Allen wrench
toward the directions 1 and 2 to remove. (Fig. 30)
ƒ onto the lower part of the cord hanger @, and
23
Page 26
Fig. 31
Connecting the Power Cord(Figs. 31 & 32)
8
Cables and cords are vulnerable when bent or twisted. Never
subject them to excessive force.
Make sure that the main switch is set to “ ” (OFF) before
connecting the power cord.
Always use the power cord provided by Olympus. IF no power
cord is provided with the microscope, please select the proper
power cord by referring to section “PROPER SELECTION OF THE
POWER SUPPLY CORD” at the end of this instruction manual.
1. Connect the power cord plug ² to the AC receptacle ³. (Fig. 31)
Connect the power cord to a grounded, 3-conductor power outlet
and ensure that the ground terminal of the power supply and that
of the wall outlet are properly connected. If the equipment is not
grounded, Olympus can no longer warrant the electrical safety
performance of the equipment.
2. Plug the power cord plug | into the wall outlet ƒ. (Fig. 32)
CX41
Fig. 32
24
Page 27
PROPER SELECTION OF THE POWER SUPPLY CORD
If no power supply cord is provided, please select the proper power supply cord for the equipment by referring to “Specifications"
and "Certified Cord" below.
CAUTION : In case you use a non-approved power supply cord for Olympus products, Olympus can no longer warrant
the electrical safety of the equipment.
Specifications
Voltage Rating
Current Rating
Temperature Rating
Length
Fittings Configuration
125V AC (for 100-120V AC area) or, 250V AC (for 220-240V AC area)
6A minimum
60 minimum
3.05 m maximum
Grounding type attachment pulg cap Opposite teminates in molded-on IEC configuration appliance coupling.
Table 1 Certified Cord
A power supply cord should be certified by one of the agencies listed in Table 1, or comprised of cordage marked with an
agency marking per Table 1 or marked per Table 2. The fittings are to be marked with at least one of agencies listed in Table 1.
In case you are unable to buy locally in your country the power supply cord which is approved by one of the agencies
mentioned in Table 1, please use replacements apprpved by any othere quivalent and authorized agencies in your country.
CountryAgencyCertification MarkCountryAgencyCertification Mark
ArgentinaIRAM
AustraliaSAA
AustriaÖVE
ItalyIMQ
Japan
Netherlands
JET, JQA, TÜV,
UL-APEX / MITI
KEMA
25
BelgiumCEBECNorwayNEMKO
CanadaCSA
DenmarkDEMKOSwedenSEMKO
FinlandFEI
FranceUTEUnited KingdomASTA, BSI
GermanyVDE
IrelandNSAI
SpainAEE
SwitzerlandSEV
U.S.A.UL
Page 28
Table 2 HAR Flexible Cord
APPROVAL ORGANIZATIONS AND CORDAGE HARMONIZATION MARKING METHODS
CX41
Approval Organization
Comite Electrotechnique Belge
(CEBEC)
Verband Deutscher Elektrotechniker
(VDE) e.V. Prüstelle
Union Technique de l'Electricité
(UTE)
Instltuto Itaaliano del Marchio di
Qualita’ (IMQ)
Bnitish Approvals Service for Electric
Cables (BASEC)
N.V. KEMA
SEMKO AB Svenska Elektriska
Matenelkontrollanstalter
Österreichisher Verband für
Elektrotechnik (ÖVE)
Danmarks Elektriske Materialkontroll
(DEMKO)
Printed or Embossed Harmonization
Marking (May be located on jacket
or insulation of internal wiring)
CEBEC<HAR>
<VDE><HAR>
USE<HAR>301030
IEMMEQU<HAR>
BASEC<HAR>
KEMA-KEUR<HAR>
SEMKO<HAR>
<ÖVE><HAR>
<DEMKO><HAR>
Alternative Marking Utilizing
Black-Red-Yellow Thred (Length
of color section in mm)
BlackRedYellow
103010
301010
103050
101030
103030
101050
301050
301030
National Standards Authority of Ireland
(NSAI)
Norges Elektriske Materiellkontroll
(NEMKO)
Asociacion Electrotecnica Y
Electronica Espanola (AEE)
Hellenic Organization for
Standardization (ELOT)
Instituto Portages da Qualidade
(IPQ)
Schweizerischer Elektro
Technischer Verein (SEV)
Elektriska Inspektoratet
Underwriters Laboratories Inc. (UL)
Canadian Standards Association (CSA)
<NSAI><HAR>
NEMKO<HAR>
<UNED><HAR>
ELOT<HAR>
np<HAR>
SEV<HAR>
SETI<HAR>103090
SV, SVT, SJ or SJT, 3 X 18AWG
SV, SVT, SJ or SJT, 3 X 18AWG
303050
101070
301070
303070
101090
103090
26
Page 29
MEMO
Page 30
MEMO
CX41
Page 31
ECREP
Shinjuku Monolith, 3-1, Nishi Shinjuku 2-chome, Shinjuku-ku, Tokyo, Japan
Wendenstraße 14-18, 20097 Hamburg, Germany
3500 Corporate Parkway, P.O. Box 610, Center Valley, PA 18034-0610, U.S.A.
One Corporate Drive, Orangeburg, NY 10962, U.S.A.
491B River Valley Road, #12-01/04 Valley Point Office Tower, Singapore 248373
31 Gilby Road, Mount Waverley, VIC., 3149, Australia
Blue Lagoon Drive, Suite 290 Miami, FL 33126, U.S.A.
5301
01/10
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.