This device is intended for use only by qualified medical personnel.
Use only Nihon Kohden approved products with this device. Use of non-approved products or in
a non-approved manner may affect the performance specifications of the device. This includes,
but is not limited to, batteries, recording paper, pens, extension cables, electrode leads, input
boxes and AC power.
Please read these precautions thoroughly before attempting to operate the instrument.
1. To safely and effectively use the instrument, its operation must be fully understood.
2. When installing or storing the instrument, take the following precautions:
(1) Avoid moisture or contact with water, extreme atmospheric pressure, excessive humidity and temperatures, poorly
ventilated areas, and dust, saline or sulphuric air.
(2) Place the instrument on an even, level floor. Avoid vibration and mechanical shock, even during transport.
(3) Avoid placing in an area where chemicals are stored or where there is danger of gas leakage.
(4) The power line source to be applied to the instrument must correspond in frequency and voltage to product
specifications, and have sufficient current capacity.
(5) Choose a room where a proper grounding facility is available.
3. Before Operation
(1) Check that the instrument is in perfect operating order.
(2) Check that the instrument is grounded properly.
(3) Check that all cords are connected properly.
(4) Pay extra attention when the instrument is in combination with other instruments to avoid misdiagnosis or other
problems.
(5) All circuitry used for direct patient connection must be doubly checked.
(6) Check that battery level is acceptable and battery condition is good when using battery-operated models.
4. During Operation
(1) Both the instrument and the patient must receive continual, careful attention.
(2) Turn power off or remove electrodes and/or transducers when necessary to assure the patient’s safety.
(3) Avoid direct contact between the instrument housing and the patient.
5. To Shutdown After Use
(1) Turn power off with all controls returned to their original positions.
(2) Remove the cords gently; do not use force to remove them.
(3) Clean the instrument together with all accessories for their next use.
6. The instrument must receive expert, professional attention for maintenance and repairs. When the instrument is
not functioning properly, it should be clearly marked to avoid operation while it is out of order.
7. The instrument must not be altered or modified in any way.
8. Maintenance and Inspection:
(1) The instrument and parts must undergo regular maintenance inspection at least every 6 months.
(2) If stored for extended periods without being used, make sure prior to operation that the instrument is in perfect
operating condition.
Service Manual CNS-9701i
(3) Technical information such as parts list, descriptions, calibration instructions or other information is available for
qualified user technical personnel upon request from your Nihon Kohden distributor.
9. When the instrument is used with an electrosurgical instrument, pay careful attention to the application and/or
location of electrodes and/or transducers to avoid possible burn to the patient.
10. When the instrument is used with a defibrillator, make sure that the instrument is protected against defibrillator
discharge. If not, remove patient cables and/or transducers from the instrument to avoid possible damage.
WARRANTY POLICY
Nihon Kohden Corporation (NKC) shall warrant its products against all defects in materials and workmanship for one year
from the date of delivery. However, consumable materials such as recording paper, ink, stylus and battery are excluded from
the warranty.
NKC or its authorized agents will repair or replace any products which prove to be defective during the warranty period,
provided these products are used as prescribed by the operating instructions given in the operator’s and service manuals.
No other party is authorized to make any warranty or assume liability for NKC’s products. NKC will not recognize any other
warranty, either implied or in writing. In addition, service, technical modification or any other product change performed by
someone other than NKC or its authorized agents without prior consent of NKC may be cause for voiding this warranty.
Defective products or parts must be returned to NKC or its authorized agents, along with an explanation of the failure.
Shipping costs must be pre-paid.
This warranty does not apply to products that have been modified, disassembled, reinstalled or repaired without Nihon
Kohden approval or which have been subjected to neglect or accident, damage due to accident, fire, lightning, vandalism,
water or other casualty, improper installation or application, or on which the original identification marks have been
removed.
In the USA and Canada other warranty policies may apply.
CAUTION
United States law restricts this device to sale by or on the order of a physician.
iiService Manual CNS-9701
EMC RELATED CAUTION
This equipment and/or system complies with the International Standard IEC 60601-1-2 for electromagnetic
compatibility for medical electrical equipment and/or system. However, an electromagnetic environment
that exceeds the limits or levels stipulated in the IEC 60601-1-2, can cause harmful interference to the
equipment and/or system or cause the equipment and/or system to fail to perform its intended function or
degrade its intended performance. Therefore, during the operation of the equipment and/or system, if
there is any undesired deviation from its intended operational performance, you must avoid, identify and
resolve the adverse electromagnetic effect before continuing to use the equipment and/or system.
The following describes some common interference sources and remedial actions:
1. Strong electromagnetic interference from a nearby emitter source such as an authorized radio station
or cellular phone:
Install the equipment and/or system at another location if it is interfered with by an emitter source such
as an authorized radio station. Keep the emitter source such as cellular phone away from the
equipment and/or system.
2. Radio-frequency interference from other equipment through the AC power supply of the equipment
and/or system:
Identify the cause of this interference and if possible remove this interference source. If this is not
possible, use a different power supply.
3. Effect of direct or indirect electrostatic discharge:
Make sure all users and patients in contact with the equipment and/or system are free from direct or
indirect electrostatic energy before using it.
4. Electromagnetic interference with any radio wave receiver such as radio or television:
If the equipment and/or system interferes with any radio wave receiver, locate the equipment and/or
system as far as possible from the radio wave receiver.
If the above suggested remedial actions do not solve the problem, consult your Nihon Kohden Corporation
subsidiary or distributor for additional suggestions.
The CE mark is a protected conformity mark of the European Community. The products herewith comply
with the requirements of the Medical Device Directive 93/42/EEC.
The CE mark only applies to the CNS-9701K series central monitors.
This equipment complies with EUROPEAN STANDARD EN-60601-1-2 (1993) which requires EN-55011, class
A. Class A EQUIPMENT is allowed in domestic establishments when used under the jurisdiction of a
health care professional.
Service Manual CNS-9701iii
In IEC 60601-1-2 Medical Electronic Equipment, Part 1: General Requirements for Safety, 2. Collateral
Degree of protection against harmful ingress of water
IPX1
Degree of safety of application in the presence of a flammable anaesthetic
mixture with air, oxygen or nitrous oxide
Not suitable for use in the presence of a
flammable anaesthetic mixture with air,
oxygen or nitrous oxide
Mode of operation:Continuous operation
Main unit180 (W) × 352 (H) × 405 (D) mm, approx.16 kg
LCD unit (without protrusion)457 (W) × 472 (H) × 247 (D) mm, approx.11 kg
Service Manual CNS-97011.15
1. GENERAL
Panel Descriptions
Central Monitor Main Unit
MU-971RA/RJ/RK
WARNING
While the central monitor is on, do not insert any CD-ROM or floppy
disk into it. Nihon Kohden does not warrant the normal operation of
the central monitor in the case that an application or software
installation starts automatically.
CAUTION
While the central monitor is on, do not touch unused sockets.
Failure to follow this caution may damage the monitor by static
electricity and may cause malfunction.
Front Panel
Power switch
Turns main unit on or off when the main power switch on the rear
panel is on ( ).
Power switch cover
For attaching, refer to CNS-9701 installation guide.
Hard disk lamp
Lights when the hard disk is being accessed.
Reset switch
Normally not used.
1.16Service Manual CNS-9701
1. GENERAL
Rear Panel
NETWORK socket
RECORDER socket
NOTE
••
• Do not remove the covers from the unused sockets on the rear panel
••
of the central monitor main unit.
••
• Do not touch the sockets on the rear panel of the main unit or cables
••
around the sockets. Otherwise, static electricity may cause
malfunction of the monitor.
<Volax5000><Volax5001>
AC SOURCE socket
Main power switch
-240
KEYBOARD socket
-240
MOUSE socket
PROTECT KEY socket
RECORDER socket
SOUND socket
NETWORK socket
VIDEO1/VIDEO2 socket
TOUCH 1/TOUCH 2 socket
NameDescription
AC SOURCE socketConnects to a UPS via the power cord provided with the central monitor
main unit.
Main power switchTurns on or off to supply power to the main unit. Under normal
conditions keep this switch turned on ( ).
KEYBOARD socketConnects to the keyboard.
MOUSE socketConnects to the mouse.
NETWORK socketConnects to network system via a network cable.
SOUND socketConnects to the SOUND socket on the VL-971R/RK LCD unit.
PROTECT KEY socketConnects to the protection key via the connection cable provided with the
central monitor main unit. This protection key can be connected to the
protection keys for the QP-971P/972P/973P/974P program kits.
RECORDER socketConnects to the WS-971R recorder unit.
VIDEO1/VIDEO2 socketConnects to the VIDEO socket on the LCD unit via the RGB connection
cable provided with the LCD unit.
TOUCH 1/TOUCH 2 socketConnects the TOUCHSCREEN socket on the LCD unit via the
touchscreen connection cable provided with the LCD unit.
Service Manual CNS-97011.17
1. GENERAL
LCD Unit VL-971R/RK
WARNING
Do not turn off the display to which the sound cable is connected.
Otherwise alarm sound does not occur.
Front
Touchscreen display
Rear
DC IN socket
Connects to the AC adapter cord
AC adapter (100 to 240 V)
Connects to a UPS via the power
cord provided with the display
unit.
Menu/Select
Up/Right
1.18Service Manual CNS-9701
M
Video Source
S
Down/Left
SOUND socket
Connects to the SOUND socket on the
central monitor main unit via the sound
cable provided with the central monitor.
Power switch
Turns on ( ) or off ( ) the LCD
unit. Under normal conditions
keep this switch turned on.
TOUCHSCREEN socket
Connects the TOUCH 1/TOUCH 2 socket on the
central monitor main unit via the touchscreen
connection cable provided with the LCD unit.
VIDEO socket
Connects to the VIDEO1/VIDEO2 socket on the
central monitor main unit via the RGB connection
cable provided with the LCD unit.
Recorder Unit WS-971R
POWER lamp
Lights when the power is on.
CHECK RECORDER lamp
Lights when communication
between the recorder unit and
central monitor is disconnected
or when there is abnormality in
the recorder unit.
1. GENERAL
Fuse holder
Two fuses (1A) are contained.
Fuse must be replaced by
qualified service personnel.
Power switch
Turns the power ON ( ) or OFF
( ).
Usually, keep the power ON.
Paper magazine release lever
Press to open the magazine.
Paper magazine
Contains the recording paper.
Power socket
Connects to a UPS with a power
cord provided with the central
monitor main unit and supplies
power (100 to 240 V) to the
recorder unit.
PC socket
Connects to the RECORDER socket on the
rear panel of the central monitor main unit
with the recorder cable provided with the
central monitor main unit.
Service Manual CNS-97011.19
1. GENERAL
t
Composition
Options
CNS-9701A
CNS-9701J
CNS-9701K
QP-971P 12 Patient Examination Program Kit
QP-972P 16 Patient Examination Program Kit
QP-973P Full Disclosure Filing Program Kit
QP-974P 12 Lead ECG Interpretation Filing Program Ki
MU-971RACentral monitor main unit for A version
MU-971RJCentral monitor main unit for J version
MU-971RKCentral monitor main unit for K version
VL-971RLCD unit (Option)
VL-971RK
WS-971RRecorder unit (Option)
1.20Service Manual CNS-9701
Connection Diagram
1. GENERAL
Mouse
Protection key
Keyboard
LCD unit (VL-971R/RK)
Central monitor main unit
(MU-971RA/RJ/RK)
Recorder unit
WS-971R
AC adapter
UPS
Laser printer
To network system
To the AC outlet
on the wall
To the Grounding terminal
on the wall
To the AC outlet on the wall and
equipotential grounding system
Service Manual CNS-97011.21
1. GENERAL
When using two displays
Main display
Sub display
LCD unit (VL-971R/RK)
Central monitor main unit
(MU-971RA/RJ/RK)
1.22Service Manual CNS-9701
Section 2Troubleshooting
General ................................................................................................................................ 2.1
System ................................................................................................................................ 2.3
Use the troubleshooting tables to locate, identify, and solve a problem in the
instrument. The problems are divided into general problem areas. Each category
has its own troubleshooting table for fast and easy troubleshooting.
• System
• Network
• Measurement Value
• Alarm
• Admitting/Discharging
• All Beds Screen/Individual Bed Screen
• Trend Window
• Tabular Trend Window
• Hemodynamics List Window
• Arrhythmia Recall Window
• ST Recall Window
• Full Disclosure Window
• ECG 12 Lead Analysis Window
• Report Window
• Recording
• Overview Bed Screen
• Parameter Setup
• Power Problem
• Display Problem
• Touch Screen Problem
• Sound Problem
• Keyboard Problem
• Mouse Problem
• Hard Disk Problem
• CD-ROM Drive or Floppy Disk Drive Problem
• 2-channel Recorder Unit Problem
If these sections do not solve the problem, contact your Nihon Kohden
representative.
NOTE
Before contacting your Nihon Kohden representative for technical
support, please provide additional detailed information on the problem.
This will allow your Nihon Kohden representative to provide you with
the best support.
Service Manual CNS-97012.1
2. TROUBLESHOOTING
How to use the troubleshooting table
1. Determine which troubleshooting table to use.
2. In the “Problem” column, find the trouble item that matches the problem.
3. Do the action recommended in the “Action” column.
4. If the problem is not solved, do the action for the next possible cause or
criteria.
5. If none of the actions solve the problem, contact your Nihon Kohden
representative.
2.2Service Manual CNS-9701
System
TroublePossible Cause/CriteriaAction
The touch screen keys do not
function.
The mouse disappears off the
screen.
The “System Setup” key cannot
be touched.
Display position is not correct.The screen is out of
No display on the second
display
An application error occurs
when you start the central
monitor system program and the
system does not start.
Cannot display data saved in the
NFS-9000 network file server.---
The pressed position and activated
position do not match.
The central monitor is operated in
dual display mode.
The central monitor is operated in
dual display mode and the System
Setup screen is open at the other
display.
synchronization.
Settings on the Display Properties
window of Windows 2000 are not
set properly.
2. TROUBLESHOOTING
Reboot the display with the following
procedure.
1. Confirm the cable connection for the
touch screen.
2. Turn off the LCD display and then turn
it on again.
3. Perform touch screen initialization and
calibration at the Touchscreen Setting
window of the System Setup.
If the trouble remains after actions 1 to 3,
exit the central monitor program and
restart Windows 2000.
To stop dual display mode, do the
following:
1. On the Monitor Setting window of the
System Setup, set “Number of
Displays” to Single Display.
2. On the Display Properties window of
Windows 2000, set “Display2” to OFF,
then reboot the system.
VL-971R/RK LCD unit:
Set the correct screen frequency using the
smart synchronization function.
Locally purchased LCD unit:
Select the correct screen frequency
referring to the LCD unit operation
manual.
On the Display Properties window of
Windows 2000, set “Display2” to ON, then
reboot the system. (When the product is
shipped, this setting is set to OFF.)
Confirm that the Display Properties
window of Windows 2000 is set as below.
When the central monitor is operated in the
dual display mode, confirm both display 1
and 2.
Screen Area: 1280 × 1024
Colors:True color
On the Windows setting window, set
display 1 to the left and 2 to the right.
At present, the CNS-9701 central monitor
cannot display data saved in the NFS-9000
network file server.
Service Manual CNS-97012.3
2. TROUBLESHOOTING
TroublePossible Cause/CriteriaAction
Data is deleted when the central
monitor power is turned off.
The system crashes when trying
to review the past data.
The system crashes periodically
such as once a day.
• Slow operation
• It takes time to read out the
saved data.
• Malfunction
Faulty hard diskContact your Nihon Kohden distributor or
representative. The central monitor may
lose its function as a monitor.
Clock errorIncorrect date and time
If the clock becomes incorrect after
adjusting the clock and turning the
central monitor power on again,
the back up battery is weak.
Windows is unstable.Shut down the central monitor system
Contact your Nihon Kohden distributor or
representative.
program and restart it. For a stable
operation, restart the central monitor
system program once every three months.
Network
TroublePossible Cause/CriteriaAction
A desired bed is not displayed in
the selected group.
Cannot find 8000 series beds in
the network
The lamp near the NETWORK
socket on the main unit does not
light and the main unit cannot
communicate with any
instrument connected to the
network.
The bedside monitor or multiple
patient receiver for that bed is
turned off.
The desired bed belongs to another
group.
---
The network cable is loose at the
socket.
The network circuit on the
motherboard has a failure.
The hub connected to the main unit
through the network cable has a
failure.
Turn on the bedside monitor or multiple
patient receiver.
On the Monitor Setting window, check
other groups to find the desired bed.
Check group names of each bedside
monitor and multiple patient receiver.
8000 series beds connected to the network
via the CNS-9301/9302 central monitor
belong to the same group as the CNS9301/9302 central monitor.
Check that the network cable and its
connectors at both ends have no damage
and check the continuity. After these
checks, firmly connect the cable between
the main unit and hub.
Replace the motherboard with a new one.
Replace the hub with a new one.
2.4Service Manual CNS-9701
Measurement Value
TroublePossible Cause/CriteriaAction
Heart rate is not counted
correctly.
Arrhythmia is not measured
correctly.
Heart rate and arrhythmia are
not measured correctly.
Noise on the ECG waveformHum interference on the ECG
The ST value is inaccurate.The ST analysis point is not set
Respiration is not measured
correctly.
Waveform too large or too small
QRS detection threshold is
changed.
QRS and pacemaker spikes are not
distinguished.
QRS detection is set to “Manual”.Set QRS detection to “Auto”, or confirm
ECG learning is not correct.Relearn patient’s ECG.
ECG lead that is inappropriate for
analysis is selected.
waveform.
properly.
Waveform too large or too small
Respiration sensitivity is changed.
Respiration measurement is set to
OFF.
2. TROUBLESHOOTING
Adjust the QRS detection sensitivity.
Turn pacing detection ON.
that the QRS detection lead and threshold
in manual detection are set properly.
Change the ECG lead used for analysis.
Set the hum filter to ON.
Confirm that the ST analysis point setting
is properly set for ST analysis.
Set the proper sensitivity.
Turn the respiration measurement ON.
Alarm
TroublePossible Cause/CriteriaAction
No alarm.
All alarms for a bed cannot be
turned ON/OFF together.
Each alarm setting of the Personal
Setup is set to OFF.
Arrhythmia analysis is turned OFF
at the bedside monitor and/or
central monitor.
AllalarmsaresettoOFFatthe
bedside monitor.
---
Turn the alarms ON for the bed.
Turn arrhythmia analysis ON on both the
bedside monitor and central monitor.
Turn all alarms ON at the bedside monitor.
This operation cannot be done at the
central monitor.
This operation is not available at the
central monitor.
Service Manual CNS-97012.5
2. TROUBLESHOOTING
Admitting/Discharging
TroublePossible Cause/CriteriaAction
Cannot transfer a patient’s data.
Cannot change the receiving
channel.
Cannot admit/discharge a
patient.
Cannot select a bedside monitor
as a transfer destination bedside
monitor.
The patient’s data before patient
transfer is lost.
The patient’s data before patient
transfer cannot be reviewed.
The message “Transferring” is
displayed on the screen but it
does not change to
“Completed”.
All patient’s data is lost after
patient transfer.
ST recall data is lost after
patient transfer.
You are trying to do this operation
on a central monitor where the
patient’s data is not saved.
You are trying to do this operation
on a central monitor where the
patient’s data is not saved.
The bedside monitor is turned off.Turn on the bedside monitor.
The bedside monitor is turned off.Turn on the bedside monitor.
The patient was transferred
between bedside monitors in
different CNS groups more than
once.
The central monitor where the
patient’s data is saved is turned off.
During patient transfer, the central
monitor was turned off. Or, there
was trouble on the network and the
transfer failed.
You transferred an 8000 series bed
that was a monitored bed of a
CNS-9301/9302 central monitor.
Do this on the central monitor where the
patient’s data is saved.
Do this on the central monitor where the
patient’s data is saved.
To save the patient’s data in the first
central monitor, do not transfer the patient
to a bedside monitor in a third CNS group
after the patient is transferred to a bedside
monitor in a second CNS group.
Turn on the central monitor where the
patient’s data before the patient transfer is
saved.
Admit the patient at the destination
bedside monitor. The lost data cannot be
restored.
You cannot transfer ST recall data of an
8000 series bed that was a monitored bed
of a CNS-9301/9302 central monitor.
2.6Service Manual CNS-9701
All Beds Screen/Individual Bed Screen
TroublePossible Cause/CriteriaAction
Numericdataandmessages
disappear for a moment.
Waveform is not displayed.
Numeric data is not displayed.
The desired waveform or
numeric data cannot be selected.
Some waveforms are not
displayed.
The desired parameter cannot be
displayed.
Cannot find a key to freeze a
waveform or relearn ECG.
The display is refreshed.This is
normal.
The bed is discharged.Admit the patient.
The bed is set to temporary discharge. Resume monitoring.
The bedside monitor or the multiple
patient receiver is turned off.
The currently measured lead/label is
different from the lead/label to be
displayed on the screen.
The waveform is not selected as the
displayed waveform.
The bed is not registered as a
monitored bed.
Network is disconnected.Confirm the network connection.
The bedside monitor is an 8000 series
bed connected to the network via the
CNS-9301/9302 central monitor.
The parameter of the numeric data is
not selected as the displayed
parameter.
The label of the currently measured
numericdataisdifferentfromthe
label to be displayed.
A vital alarm occurred in a parameter
that is not set to a displayed
parameter and its numeric data took
the place of a numeric data that is set
to the displayed parameter.
The waveform or numeric data is not
measured.
Some of the patient waveforms
cannot be displayed on this central
monitor while another central monitor
that is connected to a multiple patient
receiver or a signal exchanger
displays the individual bed screen of
the patient.
The parameter is a parameter that
cannot be displayed.
Those functions are not assigned to
function keys.
2. TROUBLESHOOTING
---
Turn on the bedside monitor or the
multiple patient receiver.
Confirm the setting so that the currently
measured lead/label matches the
lead/label to be displayed on the screen.
Set the waveform to the displayed
waveform.
Register the bed to the monitored bed.
Several types of waveforms cannot be
displayed on 8000 series bedside
monitors.
Set the parameter of the numeric data to
the displayed parameter.
Confirm the setting so that the label of
the currently measured numeric data
matches the label to be displayed.
The numeric data of a parameter whose
vital alarm is currently occurring has a
higher priority to be displayed.
Measure the waveform or numeric data.
Confirm the setting so that the currently
measured lead/label matches the
lead/label to be displayed on the screen.
Register the desired parameter on the
Parameters window of the System
Setup. A parameter registered at the
Parameters window of the System Setup
is displayed on setting keys even when
it is not currently measured.
Display the All Beds screen at the
central monitor where the patient’s data
is saved.
Refer to Section 15 “Parameter Priority”
for the list of parameters that can be
displayed on the central monitor.
Assign these functions to function keys.
Service Manual CNS-97012.7
2. TROUBLESHOOTING
Trend Window
TroublePossible Cause/CriteriaAction
Trend data is not displayed.The central monitor where the
Trend data is not saved.The central monitor where the
Tabular Trend Window
bed’s data is saved is turned off.
bed’sdataissavedwasturnedoff
during measurement.
Turn on the central monitor where the
bed’s data is saved.
Trend data is not created when the central
monitor where the bed’s data is saved is
turned off.
TroublePossible Cause/CriteriaAction
Tabular trend data is not
displayed.
Cannot select a parameter you
want to add to the tabular trend
NIBPnumericdataisnot
displayed.
Tabular trend data is not saved.The central monitor where the
The central monitor where the
bed’s data is saved is turned off.
The parameter is not registered at
the Parameters window of the
System Setup.
When an NIBP measurement is
delayed or remeasured and its
measurement timing does not
match the selected tabular trend
interval, that NIBP data will not
appear on the tabular trend.
bed’sdataissavedwasturnedoff
during measurement.
Hemodynamics List Window
Turn on the central monitor where the
bed’s data is saved.
Register the parameter at the Parameters
window of the System Setup.
Set the tabular trend display interval to 1
min. The NIBP tabular trend and all NIBP
measurements are displayed.
Tabular trend data is not created when the
central monitor where the bed’s data is
saved is turned off.
TroublePossible Cause/CriteriaAction
Hemodynamics data is not
displayed.
Hemodynamics data is not
saved.
2.8Service Manual CNS-9701
The central monitor where the
bed’s data is saved is turned off.
The central monitor where the
bed’sdataissavedwasturnedoff
during measurement.
The bedside monitor cannot output
hemodynamics data to the
network.
Turn on the central monitor where the
bed’s data is saved.
Hemodynamics data is not created when
the central monitor where the bed’s data is
saved is turned off.
Hemodynamics data that can be reviewed
at the central monitor is data created at the
BSM-4100 series bedside monitor version
02-09 or later.
Arrhythmia Recall Window
TroublePossible Cause/CriteriaAction
Arrhythmia recall files are not
created.
Arrhythmia recall files for a
particular arrhythmia are not
created
Arrhythmia recall files for
unnecessary arrhythmias are
created.
Arrhythmia recall files are
deleted.
Arrhythmia recall files are not
displayed.
Arrhythmia recall files are not
saved.
When the multi waveform
display is selected, the desired
waveforms cannot be displayed.
Multi waveform cannot be
displayed.
The arrhythmia analysis is set to
OFF.
Recall file creation for the
arrhythmia is set to OFF.
Recall file creation for the
arrhythmia is set to ON.
The number of arrhythmia recall
files for one bed exceeded the limit
and older files are deleted.
You touch “Delete” when several
files have a deletion mark.
The bedside monitor is turned off.Turn the bedside monitor or multiple
The central monitor where the
bed’s data is saved is turned off.
The arrhythmia types to be
displayed on the screen are not
selected.
The central monitor where the
bed’s data is saved was turned off
during measurement.
Stored waveforms and displayed
waveforms are not selected for full
disclosure waveform.
The QP-973P Full Disclosure
Filing Program Kit is not installed.
2. TROUBLESHOOTING
Turn the arrhythmia analysis to ON.
Turn the setting to ON.
Turn the setting to OFF.
Deleted files cannot be recovered.
patient receiver on.
Turn on the central monitor where the
bed’s data is saved.
Select the arrhythmias to be displayed on
the screen.
Arrhythmia recall files are not created
when the central monitor where the bed’s
data is saved is turned off.
Waveforms displayed for the multi
waveform display are synchronized with
the displayed waveform settings on the
Full Disclosure window. Set at “Stored
Waves” and “Displayed Waves” on the
Full Disclosure window.
Install the QP-973P Full Disclosure Filing
Program Kit.
ST Recall Window
TroublePossible Cause/CriteriaAction
ST recall files for 72 hours
cannot be saved.
ST recall files are not saved.
Service Manual CNS-97012.9
The interval of ST recall file
creation is set to 30 seconds.
The central monitor where the
bed’s data is saved is turned off.
The bedside monitor cannot output
ST recall files to a network.
With the creation interval of 30 seconds,
the total number of files exceeds the limit.
To save recall file for 72 hours, set the
creation interval to 1 min or more.
Turn on the central monitor where the
bed’s data is saved.
The central monitor cannot save or display
ST recall files of the following bedside
monitors:
• BSM-4100 series bedside monitor
whose version is not 02-09 or later
• 8000 series bed that is a monitored bed
of a CNS-9301/9302 central monitor.
2. TROUBLESHOOTING
Full Disclosure Window
TroublePossible Cause/CriteriaAction
Full disclosure waveform is not
displayed.
The desired full disclosure
waveform is not displayed.
The desired waveform cannot be
selected for a stored waveform.
The central monitor where the
bed’s data is saved is turned off.
The QP-973P Full Disclosure
Filing Program Kit is not installed.
The QP-973P Full Disclosure
Filing Program Kit is not installed
in the central monitor where the
bed’s data is saved.
The full disclosure waveform is
not saved.
The parameter is not measured.Measure the parameter.
Turn on the central monitor where the
bed’s data is saved.
Install the QP-973P Full Disclosure Filing
Program Kit.
Choose a central monitor with the QP973P Full Disclosure Filing Program Kit
installed for the central monitor where the
bed’s data is saved.
Choose the waveform for the stored
waveform on the Full Disclosure window.
The waveforms that are selected
for the stored waveform are not
saved.
On the Full Disclosure window,
the arrhythmia waveform is not
indicated with the preset color.
This central monitor is not the
central monitor where the bed’s
data is saved.
The central monitor where the
bed’sdataissavedwasturnedoff
during measurement.
The currently measured lead/label
is different from the lead/label to
be saved.
Arrhythmia analysis is set to OFF
at the bedside monitor and/or the
central monitor.
The arrhythmia alarm is set to OFF
at the bedside monitor and/or the
central monitor.
This setting is only available at the central
monitor where the bed’s data is saved or
the BSS-9800 bedside station (central
mode).
Full disclosure waveform is not saved
when the central monitor where the bed’s
data is saved is turned off.
Make the currently measured lead/label
match the lead/label to be saved.
Turn arrhythmia analysis to ON at the
bedside monitor and central monitor. You
cannot set color to an arrhythmia that does
not have color.
Turn the arrhythmia alarm to ON at the
bedside monitor and central monitor. You
cannot set color to an arrhythmia that does
not have color.
You cannot set color to an arrhythmia that
does not have color.
ECG 12 Lead Analysis Window
TroublePossible Cause/CriteriaAction
ECG 12 Lead analysis files are
not displayed.
2.10Service Manual CNS-9701
The QP-974P 12 Lead ECG
Interpretation Filing Program Kit is
not installed.
The bedside monitor cannot output
ECG 12 lead analysis files to a
network.
Install the QP-974P 12 Lead ECG
Interpretation Filing Program Kit.
The central monitor can display ECG 12
lead analysis files of the BSM-4100 series
bedside monitor only when the bedside
monitor version is 03-02 or later.
Report Window
TroublePossible Cause/CriteriaAction
Report printing starts
automatically.
The report printing mode is set to
“Auto”.
Recording
TroublePossible Cause/CriteriaAction
In the following recordings,
waveforms that you do not want
to record are recorded.
- Manual recording
- Periodic recording
- Alarm recording
- Call recording
In the following recording,
waveforms that you do not want
to record are recorded.
- Dual waveform recording
In the following recording,
waveforms that you do not want
to record are recorded.
- Remote delayed waveform
recording
Alarm recording is not
performed for a particular
arrhythmia.
Vital sign alarm recording is not
performed.
not performed.
ST alarm recording is not
performed.
Periodic recording is not
performed.
Call recording is not performed.Call recording is set to OFF on the
Trend data are not printed for
the report printing.
Dual waveform recording is not
performed. (Only one waveform
is recorded/printed.)
Waveform is not printed on the
screen copy.
Waveforms displayed on the All
Beds screen are recorded.
Recorded waveform is not selected
on the Recording window of the
Setup.
Recorded waveform is selected at
the bedside monitor.
Recall file creation for the
arrhythmia is set to OFF.
Vital sign alarm recording is set to
OFF.
Arrhythmia alarm recording is set
to OFF.
Arrhythmia analysis is set to OFF
at the bedside monitor.
ST alarm recording is set to OFF.Set ST alarm recording to ON.
Periodic recording is set to OFF on
the Recording window of the
Setup.
Recording window of the Setup.
Report settings are not correct.Set the correct settings on the Report
On the Recording window of the
Setup, “Second Wave” at
“Recording Pattern” is set to
NONE.
Waveform cannot be printed on the
screen copy.
2. TROUBLESHOOTING
Set the mode to “Manual”.
---
Select the desired waveform on the
Recording window of the Setup.
---
Turn the setting to ON.
Set vital sign alarm recording to ON.
Set arrhythmia alarm recording to ON.Arrhythmia alarm recording is
Turn arrhythmia analysis to ON at the
bedside monitor.
Set periodic recording to ON.
Set call recording to ON.
window.
Set a desired waveform for “Second
Wave” at “Recording Pattern”.
---
Service Manual CNS-97012.11
2. TROUBLESHOOTING
Overview Bed Screen
TroublePossible Cause/CriteriaAction
Some waveforms are not
displayed.
The desired bed cannot be
selected for the Overview Bed
screen.
The alarm for the overview bed
does not occur.
Cannot silence an overview bed
alarm.
An overview bed alarm of an
unintended bed occurs.
When the individual bed screen of
a patient is displayed on a central
monitor that is connected to a
multiple patient receiver or a signal
exchanger, some of the patient
waveforms cannot be displayed on
the Overview Bed of another
central monitor.
The bed is an 8000 series bed.
No bedside monitor is selected at
the Overview Bed screen.
An overview bed alarm cannot be
silenced at the central monitor.
Display the All Beds screen at the central
monitor that saves the data of the bedside
monitor you want to display.
To display data of an 8000 series bed on
this central monitor, a central monitor on
the network must be monitoring the bed.
Select a desired bed as an overview bed.
To stop overview bed alarm, stop
monitoring any overview bed by touching
“Stop monitoring” on the Overview Bed
Menu screen.
Parameter Setup
TroublePossible Cause/CriteriaAction
Available settings differ among
bedside monitors.
Available settings depend on the
bedside monitor model and
connection method.
---
2.12Service Manual CNS-9701
Power Problem
ProblemPossible CauseAction
-When the main unit is
turned on, the power lamp
on the front panel does not
light.
-When the display unit is
turned on, the LCD does
not light.
-The main unit or display
unit frequently turns itself
off.
In power failure, the UPS does
not cover the power requirement
of the central monitor.
2. TROUBLESHOOTING
To replace the board or unit with a new one, refer to Section 4 “ Disassembly and
Assembly”.
AC power is not obtained from the
power cord.
The power cord is loose between
the AC outlet and AC inlet of the
main unit or display unit through
the AC adapter.
The power supply unit in the main
unit or AC adapter for the display
unit has a failure.
The battery in the UPS is
completely discharged.
UPS has a failure.Replace the UPS with a new one.
Check that the AC power is obtained from
the AC outlet on the facility wall.
Disconnect the power cord between the
AC outlet and AC inlet of the main unit or
AC adapter, and connect the power cord
between them again.
Replace the power supply unit or AC
adapter with a new one.
Replace the battery with a new one.
Service Manual CNS-97012.13
2. TROUBLESHOOTING
Display Problem
ProblemPossible CauseAction
The “NO SIGNAL” message
appears on the black screen of
the display unit.
-The screen is out of
synchronization.
-The screen is flickering.
-The screen display position
is not correct.
A "No boot device" message
appears and Windows 2000 does
not start.
CAUTION
To replace the hard disk with a new one, refer to “Hard Disk Flow” in
this section and “Replacing the Hard Disk” in Section 4 “Disassembly
and Assembly”.
The RGB cable has poor contact
betweenthedisplayunitandmain
unit.
The RGB signals are not output
from the graphic board in the main
unit.
The main unit is not turned on.
The horizontal or vertical
synchronization setting is not
adjusted properly.
The hard disk is damaged.Replace the hard disk with a new one.
The hard disk is not recognized
due to incorrect BIOS setting.
Disconnect the RGB cable and firmly
connect the RGB cable between them.
Replace the graphic board with a new one.
Check the problem items shown in the
above "Power problem" table.
Perform the synchronization adjustment at
the display unit.
Set the BIOS setting correctly.
A "HDD Problem Call Nihon
Kohden" message appears and
the hard disk lamp is lit.
The screen is only one color
such as
-The screen color is only
blue.
-The screen color is only
red.
-The screen color is only
yellow.
-The screen is not clearly
displayed.
-The screen display position
varies.
-The screen is distorted.
-The screen dims.
The hard disk is damaged.Replace the hard disk with a new one.
(To use the central monitor temporarily,
turn the main unit off and on. But note that
this problem will appear again some time.)
The RGB cable has an internal
break.
The graphic board in the main unit
has a failure.
The display unit has a failure.Replace the unit with a new one.
Adjustment*1of the display unit is
required.
2
A failure of the display unit*
graphic board of the main unit.
or
Check the continuity with a multimeter.
Replacethecablewithanewoneifthe
cable has a break.
Replace the board with a new one.
Adjust the display unit according to
"Adjustment" section.
Replace the defective board*3with a new
one.
2.14Service Manual CNS-9701
ProblemPossible CauseAction
-The display color balance
has something wrong.
-The screen background
color is not black.
2. TROUBLESHOOTING
A failure of the display unit*2or
graphic board of the main unit.
Replace the defective board*3with a new
one.
Internal break in the RGB cableCheck the continuity with a multimeter.
Replace the cable with a new one if the
cable has a break.
1
Adjustment*
required.
of the display unit is
Adjust the display unit according to
"Adjustment" section.
No screen is displayed but the
backlight lamp is not lit.
Failure of the AC adapter in the
display unit.
Check that the AC adapter outputs +12 V.
If this voltage is not output, replace the AC
adapter with a new one.
No screen is displayed but the
backlight lamp is lit.
Failure of the backlight lamp or
inverter of the LCD assembly.
If NDS logo appears at the center
of the screen when you turn on the
display unit only, the display unit
works.
If NDS logo does not appear at the
center of the screen when you turn
Replace the LCD assembly with a new
one.
Check that the RGB signals are output
from the main unit. Adjust the display unit
synchronization according to the
"Adjustment" section.
Replace the LCD control board or LCD
assembly with a new one.
on the display unit only, the
display unit does not work.
1
*
The VL-971R display unit is adjusted at the factory so that the display unit accepts the RGB signals from most of
the RADEON graphic boards used in the main unit. But some of the boards require adjustment of the display unit.
*2 Most of these problems can be caused by a failure of the LCD control board in the display unit because the RGB
outputs from the graphic board are controlled by an IC on the graphic board.
*3 To judge which board has a failure, connect another PC to the display unit or another RGB display unit to the
main unit.
Service Manual CNS-97012.15
2. TROUBLESHOOTING
Touch Screen Problem
ProblemPossible CauseAction
The touch screen does not work.
The RS232C straight cable has
poor contact between the display
unit and main unit.
The touch screen control board has
afailure.
The touch screen is damaged.
The film cable of the touch screen
has a break.
Check the continuity of the cable with a
multimeter and check the connection.
Replace the board with a new one.
Replace the touch screen with a new one.
-Effective area varies on the
touch screen.
-The touch screen sometimes
works.
The pressed position and
activated position do not match.
Sound Problem
ProblemPossible CauseAction
The main unit intermittently
generates a beep sound.
No alarm and QRS sounds.
The touch screen control board has
an error due to noise.
The display unit is moved to upper
or lower location.
The position data is changed by
noise.
A key on the keyboard is pressed
and held.
Poor contact of the speaker cable.Check the continuity of the speaker cable
Failure of the speakers in the
display unit.
Turn off the display unit and turn it on to
initialize the display unit.
If there is still the same problem or it
occurs again, replace the board with a new
one.
Calibrate the touch screen at the
Touchscreen Setting window of the System
Setup.
Release the key.
and connector connection.
Replace the speakers with new ones.
Failure of the motherboard in the
main unit.
Abnormal low noise from the
main unit.
2.16Service Manual CNS-9701
Dust or lint on the cooling fan for
the CPU or in the power supply
unit.
Replace the motherboard with a new one.
Clean the fans with an air compressor.
Keyboard Problem
ProblemPossible CauseAction
2. TROUBLESHOOTING
No keyboard keys work.
Some keyboard keys do not
work.
Mouse Problem
ProblemPossible CauseAction
When you move the mouse, the
cursor on the screen does not
move.
The keyboard cable is loose at the
socket on the main board.
The keyboard cable is damaged.Replace the keyboard with a new one.
Failure of the key switches.Replace the keyboard with a new one.
The mouse cable is loose at the
socket on the main unit.
Mouse failure.Replace the mouse with a new one.
Disconnect the cable from the socket on
the main board and connect the cable to
the socket tightly.
Disconnect the cable from the socket on
the main board and connect the cable to
the socket tightly.
Hard Disk Problem
ProblemPossible CauseAction
appears and Windows 2000 does
not start.
A "HDD Problem Call Nihon
Kohden" message appears and
the hard disk lamp is lit.
Following to the shut down
procedure in the operator's
manual, you shut down the
central monitor but the screen
does return to Windows 2000.
Metallic sound from the hard
disk.
CAUTION
To replace the hard disk with a new one, refer to “Hard Disk Flow” in
this section and “Replacing the Hard Disk” in Section 4 “Disassembly
and Assembly”.
The hard disk is damaged.Replace the hard disk with a new one.A "No boot device" message
The BIOS setting is changed by
noise so the hard disk is not
recognized.
The hard disk is damaged.Replace the hard disk with a new one.
The hard disk is damaged.Replace the hard disk with a new one.
The bearing for the spindle motor
in the hard disk is worn out.
Set the correct BIOS setting again.
(To use the central monitor temporarily,
turn the main unit off and on. But note that
this problem will appear again some time.)
Replace the hard disk with a new one.
Service Manual CNS-97012.17
2. TROUBLESHOOTING
CD-ROM Drive or Floppy Disk Drive Problem
ProblemPossible CauseAction
The access lamp lights but the
drive cannot read the data on a
disk.
The access lamp does not light
during access and the drive
cannot read the data on a disk.
The drive is dusty.Clean the drive with a cleaning disk or air
The drive has a failure.Replace the drive with a new one. After
Poor connection between the drive
and motherboard.
2-channel Recorder Unit Problem
compressor.
replacement, check that the new drive is
recognized when the main unit is turned
on.
Check the connection between the drive
and motherboard.
ProblemPossible CauseAction
The recording paper is loaded
but the “RECORDER OUT OF
PAPER” message appears on the
screen.
The printing intensity is not
even.
The paper magazine is
completely closed but the
“RECORDER DOOR OPEN”
message appears on the screen.
Missing dot on recordings from
the recorder unit.
The paper detection sensor is dirty.Clean the paper detection sensor with a
cotton swab moistened with alcohol.
The platen roller is worn out.Replace the roller with a new one.
The magazine open detection
switch has a failure.
The thermal array head has a
failure.
Replace the switch with a new one.
Replace the head with a new one.
2.18Service Manual CNS-9701
Error Messages
2. TROUBLESHOOTING
Error messages are categorized into two types, messages sent from bedside
monitors/multiple patient receiver and messages generated at the central monitor.
Messages From Bedside
Monitors/Multiple Patient
Receiver
Messages about Measurement Conditions of Each Parameter
The central monitor receives error messages from bedside monitors and multiple
patient receiver and displays them on the screen. Here is the message list. For
causes and countermeasures for each message, refer to the bedside monitor/
multiple patient receiver operator’s manual.
Some messages displayed at the bedside monitor/multiple patient receiver are not
displayed on the central monitor.
ParameterMessage
ECG
CHECK LEADS
CHG ELECTRODE
NOISE
SMALL QRS
RHYTHM CHG
LEARNING
LOW mV
HIGH mV
AUTO LEAD CHG
CANNOT ANALYZE
PAC I NG
ParameterMessage
Arrhythmia
ASYSTOLE
V. F I B
Vf/VT
EXT.TACHY
EXT.BRADY
V. TA C H Y
VPC RUN
COUPLET
EARLY VPC
MULTIFORM
BIGEMINY
FREQ.VPC
TACHYCARDIA
BRADYCARDIA
PROLONGED RR
FREQ.SVPC
NON-FUNCTION
NON-CAPTURE
PAC E FA ILURE
IRREGULAR RR
VPC
SVPC
RHYTHM CHG
NEW FORM
NEW NORMAL BEAT
Service Manual CNS-97012.19
2. TROUBLESHOOTING
ParameterMessage
SpO2
Module Failure
Too Many Params
Connector Off
Probe Off
Probe Disconnect
Probe Broken
No Pulse
Light Interfer.
Pulse Search
Check Probe Site
M (Unstable Pulse)
Small Pulse
Check Label
Probe Off
Probe Broken
Hardware Control
Noise
Cal Start
Cal Complete
ParameterMessage
RESP
Module Failure
Too Many Params
Connector Off
Check Sensor
Check Label
Power Off
Resp Off
ParameterMessage
NIBP
Module Failure
Too Many Params
Connector Off
Safety Valve Open
No pulse
Meas Time Out
Check Cuff/Hose
Systolic Over
High Cuff Pressure
Re-measuring
Please Wait
Small Pulse
Zeroing
Noise
Check Sensor
Cannot Measure
Check Cuff/Hose
2.20Service Manual CNS-9701
ParameterMessage
PRESS
Module Failure
Too Many Params
Check Label
Connector Off
Check Sensor
Out of Range
Zero Imbalance
Zero Unstable
Zero Out of Range
Zeroing Complete
Noise
ParameterMessage
Temp
Module Failure
Too Many Params
Check Label
Connector Off
Check Sensor
2. TROUBLESHOOTING
ParameterMessage
CO2
Module Failure
Too Many Params
Check Label
Connector Off
Check Sensor
Line Block
ParameterMessage
FiO2
Module Failure
Too Many Params
Check Label
Connector Off
Check Sensor
Sensor Run Down?
Cal?
Calibration Error
Calibrating
ParameterMessage
FLOW
Module Failure
Too Many Params
Check Label
Connector Off
Check Sensor
Change Sensor
Calibrating
Service Manual CNS-97012.21
2. TROUBLESHOOTING
ParameterMessage
SvO2
Module Failure
Too Many Params
Check Label
OM2 Disconnected
Check OM2
OM2 Failure
Check OM2 Light
OM2 Error
Out of Range
Auto Zero
Calibration Error
ParameterMessage
CCO
Unit Failure
Too Many Params
Check Label
Connector Off
Check Catheter
Catheter Location
Meas Failure
Com Error
Alarm
Standby
Tb Out of Range
Unstable Tb
Signal Adapting
CO<1.0L/m
ParameterMessage
ANES
Unit Failure
Too Many Params
Check Label
Illegal Module
SAM Failure
Moisture Detected
Over Temp
SAM Not Installed
Com Error
GAS Liquefied
Check Sampl Line
Blocked Sampl Line
Install Aquaknot
Remove Aquaknot
Alarm
Check
Sensor Failure
O2 Sensor Failure
Calibrating(Zero)
Warming Up
DES??
Standby
Calibrating
Calibration Error
Apnea
2.22Service Manual CNS-9701
ParameterMessage
tcPO2/tcPCO2
ParameterMessage
BIS
Module Failure
Com Error
Alarm
ConnectorOff
Module Failure
Too Many Params
Check Label
Connector Off
Check Sensor
DSC Failure
Com Error
Alarm
Imp. Checking
High Impedance
2. TROUBLESHOOTING
ParameterMessage
CNIBP
Unit Failure
Connector Off
Check Sensor
Com Error
Pulse Search
Sensor Expired
Check
Out of Range
ParameterMessage
EEG
Module Failure
Too Many Params
Check Label
Connector Off
Check Sensor
Alarm
ParameterMessage
VENTAlarm
Service Manual CNS-97012.23
2. TROUBLESHOOTING
Other
TroublePossible Cause/CriteriaAction
SIGNAL LOSSWeak electric field strength of a
BATTERYWeak transmitter batteryReplace the batteries with new ones.
CALL
MONITOR OFF
OFFThe bed is discharged.---
wireless telemeter.
The CALL button on the
transmitter was pressed.
The selected bedside monitor
power is OFF or is disconnected
from the network.
The multiple patient receiver for
the transmitter is disconnected
from the network.
Check the followings:
• Remaining charge of the transmitter
battery
• Transmitter antenna lead condition
• Distance between the multiple patient
receiver and transmitter
---
---
---
SIGNAL LOSS, BATTERY and CALL are displayed only when receiving data from a transmitter.
2.24Service Manual CNS-9701
Messages From Central
g
Monitor
WS-971R Recorder Unit
TroublePossible Cause/CriteriaAction
RECORDER DOOR OPENThe recorder unit door is not
RECORDER OUT OF PAPERNo recording paper in the recorder
RECORDER POWER OFF OR
DISCONNECTED
2. TROUBLESHOOTING
Close the door completely.
closed completely.
Set recording paper in the recorder unit.
unit.
The recorder unit is disconnected.Connect the recorder unit.
The recorder unit is turned OFF.Turn the recorder unit power ON.
If there is recording paper in the recorder
unit and the message still appears, the
paper detection sensor is malfunctioning.
Clean the sensor part in the recorder unit.
Laser Printer
TroublePossible Cause/CriteriaAction
Printer is OFF LINEThe laser printer is not registered.Contact your Nihon Kohden distributor or
representative.
Printer is OUT OF PAPERNo paper in the laser printer.Set paper in the printer. Refer to the printer
manual.
Job ErrorError in the printer.Refer to the printer manual.
Paper tray is openThe laser printer cover is not
closed completely.
Other
TroublePossible Cause/CriteriaAction
OFFThe bed is discharged.---
Temporary DischargeThe bed is temporarily discharged.To resume monitoring, touch “Resume” on
Close the cover completely.
the Pause window of the Admit/Dischar
of the Personal Setup.
e
Service Manual CNS-97012.25
2. TROUBLESHOOTING
Hard Disk Replacement Flow
When you order a new hard disk for replacement, the new hard disk has the OS
(Windows 2000) and central monitor system program installed at the factory.
Therefore, you do not have to install these. Replace the old hard disk with the new
one according to the following procedure.
1. Write down the following settings.
- IP address of the personal computer used as the central monitor main unit.
- The central monitor settings which are different from the factory default
- IP address of the network printer
2. Save the printer driver software on a storage media such as floppy disk or CD-
R if you use a printer which is not recommended by us.
settings.
3. Replace the old hard disk with the new one according to the “Replacing the
Hard Disk” in Section 4 “Disassembly and Assembly”.
4. Connect the display unit, keyboard, mouse and protection key to the central
monitor main unit. Refer to “Connection Diagram of Central Monitor” in
Section 1.
5. Turn on the peripheral units and main unit. The main unit automatically
recognizes the hardware components. The hardware information stored in the
new hard disk changes to the actual hardware information. The central
monitor system program automatically runs.
6. Shut down the central monitor according to the shutdown procedure in the
operator’s manual.
7. If a restart request dialog box appears on the screen, restart Windows 2000.
8. Check that there is no restart request dialog box on the screen. If the restart
request dialog box is still there, repeat the restart of Windows 2000 until the
restart request dialog box does not appear. The number of restart request
dialog box appearances depends on the hardware composition of the central
monitor.
9. Run the central monitor system program.
10. Set the IP address on the central monitor. Refer to “Setting the IP Address on
Each Instrument” in Section 8 “Network Settings”.
11. Connect the central monitor to the network. Refer to “Network and System”
installation guide. On the central monitor, check that each instrument is
connected to the network. Refer to “Checking the Network Connection on the
Central Monitor” in Section 8 “Network Settings”.
2.26Service Manual CNS-9701
12. Install the network printer drive into the central monitor under Windows 2000.
Refer to “Installing the Laser Printer (Network Printer)” in Section 8 “Network
Settings”.
13. Assign the central monitor name under Windows 2000 and assign the group
name under the central monitor system program if necessary. Refer to
“Assigning the Group Name and Bed Name” in Section 8 “Network Settings”.
14. If some settings which are different from the factory default settings are used,
change the corresponding default settings to the necessary settings at the
central monitor. Refer to “Changing Setting of System Setup, Patient Setting
and Setup” in Section 8 “Network Settings”.
How to Recover the BIOS Initial Settings
2. TROUBLESHOOTING
When the BIOS (Basic Input/Output System) settings are lost because the lithium
battery on the motherboard is completely discharged, the following message
appears on the screen at the beginning of the boot.
BIOS Checksum Error - Default loaded
To recover the BIOS initial settings, replace the lithium battery with a new one or
press the F1 key on the keyboard each time the message appears.
If there is an electric shock such as lightning, the BIOS setting can be lost. The
above message appears on the screen if the BIOS settings are lost. To recover the
BIOS initial settings, press the F1 key on the keyboard.
If the BIOS settings are lost for any other reasons, perform the following procedure
to recover the BIOS initial settings.
1. Connect a keyboard to the central monitor main unit.
2. Turn on the main unit. The self-check program runs.
3. Press the Delete key on the keyboard during the run of the self-check program.
The BIOS setting screen appears. If the BIOS setting screen does not appear,
turn off the main unit and remove the lithium battery from the motherboard.
Start this procedure from step 2.
4. Select “Load Setup Defaults”.
5. Select “Save & Exit Setup” to exit the BIOS setting screen.
Service Manual CNS-97012.27
2. TROUBLESHOOTING
BIOS Default Setting (Vorax5000)
The BIOS setting may differ depending on the production lot.
Standard CMOS features
Advanced BIOS Features
Date/TimeCurrent date and time
IDE Primary MasterAUTO (Indicates the recognized hard
disk model)
IDE Primary SlaveAUTO (Indicates the recognized hard
disk model)
IDE Secondary MasterAUTO (IDE Secondary Master)
AUTO (Access mode)
IDE Secondary SlaveAUTO (IDE Secondary Master)
AUTO (Access mode)
Drive A1.44 MB 3.5 inch
Drive BNone
VIDEOEGA/VGA
HALT OnAll Errors
Virus WarningDisabled
CPU Internal CacheEnabled
External CacheEnabled
CPU L2 Cache ECC checkingEnabled
Processor Number FeatureEnabled
Quick POWER ON Self testEnabled
First Boot DeviceA:
Second Boot DeviceC:
Third Boot DeviceCDROM
Boot Other DeviceEnabled
Swap Floppy DiskDisabled
Boot Up Floppy SeekDisabled
Boot up NumLock StatusOff
TypematicRateSettingDisabled
Security OptionSetup
OS select for DRAM > 64MBNon-OS2
Show Logo On ScreenDisabled
Advanced Chipset Features
SDRAM CAS Latency Time3
SDRAM Cycle Time Tras/Trc7/9
SDRAM RAS-to CAS Delay3
SDRAM RAS Precharge Time3
SM I/O Buffer ControlNormal
System BIOS CashableDisabled
Video BIOS CashableDisabled
Delayed TransactionEnabled
AGP MODEAUTO
2.28Service Manual CNS-9701
2. TROUBLESHOOTING
Integrated Peripherals
On-Chip Primary PCI IDEEnabled
On-Chip Secondary PCI IDEEnabled
IDE Primary Master PIOAuto
IDE Primary Slave PIOAuto
IDE Secondary Master PIOAuto
IDE Secondary Slave PIOAuto
IDE Primary Master UDMAAuto
IDE Primary Slave UDMAAuto
IDE Secondary Master UDMAAuto
IDE Secondary Slave UDMAAuto
USB ControllerEnabled
USB Keyboard supportDisabled
Init Display FirstPCI Slot
AC97 AudioAuto
OnBoard CNR LANEnabled
IDE HDD Block ModeEnabled
POWER ON FunctionButton Only
OnBoard FDC ControllerEnabled
OnBoard Serial Port 13F8/IRQ4
OnBoard Serial Port 22F8/IRQ3
UART Mode SelectNormal
Onboard Parallel Port378/IRQ7
Parallel Port ModeSPP
AC PWR Auto RecoveryFormer Status
Game Port Address201
Midi Port Address330
Midi Port IRQ5
Power Management Setup
ACPI FunctionEnabled
ACPI Suspend TypeS1
Power ManagementUser Define
Video Off MethodV/H SYNC + Blank
Vid eo Off In SuspendYes
Suspended TypeCPU Sleep Mode
Modem Use IRQ3
Suspend ModeDisabled
HDD Power DownDisabled
Soft Off by Power ButtonInstant Off
Wake On P CI Car dBy O S
Wake On ModemDisabled
Wake On LANDisabled
Wake On RTCBy O S
Primary IDE 0Disabled
Primary IDE 1Disabled
Secondary IDE 0Disabled
Secondary IDE 1Disabled
FDD, COM, LPT PortDisabled
PCI PIRQ [A-D]#Disabled
Service Manual CNS-97012.29
2. TROUBLESHOOTING
PnP Configurations
PC Health Status
PASSWORD
Standard CMOS features
Reset Configuration DataDisable
Resources Controlled byAuto (ESCD)
PCI/VGA Palette SnoopDisabled
Assign IRQ For VGAEnable
Assign IRQ For USBEnable
Shut down TemperatureDisabled
-None
BIOS Default Setting (Vorax5001)
The BIOS setting may differ depending on the production lot.
Date/TimeCurrent date and time
IDE Primary MasterAUTO (Indicates the recognized hard
disk model)
IDE Primary SlaveAUTO (Indicates the recognized hard
disk model)
IDE Secondary MasterAUTO (IDE secondary Master)
AUTO (Access mode)
IDE Secondary SlaveAUTO (IDE Secondary slave)
AUTO (Access mode)
Drive A1.44 MB 3.5 inch
Drive BNone
VIDEOEGA/VGA
HALT OnAll Errors
Advanced BIOS Features
Virus WarningDisabled
CPU L1 & L2 CacheEnabled
Quick POWER ON Self TestEnabled
First Boot DeviceA:
Second Boot DeviceC:
Third Boot DeviceCDROM
Boot Other DeviceEnabled
Swap Floppy DiskDisabled
Boot Up Floppy SeekDisabled
Boot up NumLock StatusON
Gate A20 OptionNormal
TypeMedia Rate SettingDisabled
Security OptionSetup
APIC ModeEnabled
Report No FDD for WIN 95No
Show Logo On ScreenDisabled
2.30Service Manual CNS-9701
2. TROUBLESHOOTING
Advanced Chipset Features
Integrated Peripherals
Dram Read Thermal MgmtDisabled
System BIOS CacheableDisabled
Video BIOS CacheableDisabled
Video RAM CacheableDisabled
Memory Hole At 15M-16MDisabled
Delayed TransactionEnabled
CPU Auto Throttling Delay16Min
AGPApertureSize(MB)64
On-Chip Primary PCI IDEEnabled
IDE Primary Master PIOAuto
IDE Primary Slave PIOAuto
IDE Primary Master UDMAAuto
IDE Primary Slave UDMAAuto
On-Chip Secondary PCI IDEEnabled
IDE Secondary Master PIOAuto
IDE Secondary Slave PIOAuto
IDE Secondary Master UDMAAuto
IDE Secondary Slave UDMAAuto
USB ControllerEnabled
USB Keyboard supportDisabled
USB Mouse SupportDisabled
AC97 AUDIOAuto
AC97 MODEMAuto
Init Display FirstPCI Slot
OnBoard LAN ConrollerEnabled
IDE HDD Block ModeEnabled
POWER ON FunctionButton Only
Onboard FDC ControllerEnabled
Onboard Serial Port 13F8/IRQ4
Onboard Serial Port 22F8/IRQ3
UART Mode SelectNormal
Onboard Parallel Port378/IRQ7
Parallel Port ModeSPP
PWRON After PWR-failFormer-sts
Game Port Address201
Midi Port Address330
Midi port IRQ10
Case Opened WarningDisabled
Power Management Setup
ACPI Suspend TypeS1 (POS)
Power ManagementUser Define
Video Off MethodV/H SYNC + Blank
Vid eo Off In SuspendYes
Suspended TypeStop Grant
Modem Use IRQ3
Suspend modeDisabled
HDD Power DownDisabled
Soft-Off by PWR-BTTNInstant-Off
Wake-Up By PCI CardDisabled
Power On By RingDisabled
Resume By AlarmDisabled
Primary IDE 0Disabled
Primary IDE 1Disabled
Secondary IDE 0Disabled
Secondary IDE 1Disabled
FDD, COM, LPT PortDisabled
PCI PIRQ [A-D]Disabled
Spread SpectrumDisabled
Fixed AGP/PCI Freq.Disabled
CPU Clock133
Memory Frequency ForAuto
DRAM Timing SelectableManual
CAS Latency Time2.5
Active to Precharge Delay6
DRAM RAS# to CAS# Delay3
DRAM RAS# Precharge3
-None
2.32Service Manual CNS-9701
Section 3Board/Unit Description
Power Supply Unit (by NIPRON) ......................................................................................... 3.1