1.1 Safety Information ................................................................................................................................................................................................. 1-1
2 Theory of Operation ........................................................................................................................................................ 2-1
2.1.1 Front View ...................................................................................................................................................................................................... 2-1
2.1.2 Left View ......................................................................................................................................................................................................... 2-2
2.1.3 Right View ...................................................................................................................................................................................................... 2-3
2.2 System Composition ............................................................................................................................................................................................ 2-5
2.2.1 Main Unit System Structure .................................................................................................................................................................... 2-6
2.2.2 Modular Rack and Dock System Composition ................................................................................................................................. 2-8
3.2 Preparations for Installation............................................................................................................................................................................... 3-1
3.2.1 Installation Site Requirements ............................................................................................................................................................... 3-1
4 Testing and Maintenance ................................................................................................................................................ 4-1
4.1.1 Test Equipment ............................................................................................................................................................................................ 4-1
4.1.3 Recommended Frequency ...................................................................................................................................................................... 4-2
4.3 Power-on Test .......................................................................................................................................................................................................... 4-7
4.4.2 Resp Test ......................................................................................................................................................................................................... 4-8
4.4.3 SpO
4.4.4 Temp Test ....................................................................................................................................................................................................... 4-9
4.4.6 IBP Test and Calibration ......................................................................................................................................................................... 4-10
4.4.7
Test ........................................................................................................................................................................................................ 4-9
2
Sidestream and Microstream CO2 Modules Tests and Calibration ......................................................................................... 4-11
.5 Analog Output Test ............................................................................................................................................................................................. 4-12
.7 Network Print Test ............................................................................................................................................................................................... 4-13
4
.7.1 Equipment Connection and Setup .................................................................................................................................................... 4-13
4
.7.2 Print Function Test .................................................................................................................................................................................... 4-13
.8.1 Functionality Test ...................................................................................................................................................................................... 4-14
4
.8.2 Performance Test ...................................................................................................................................................................................... 4-14
5.2 Part Replacement .................................................................................................................................................................................................. 5-1
5.3 Software Version Check ...................................................................................................................................................................................... 5-1
5.5.4 Dock or Modular Rack Failures .............................................................................................................................................................. 5-3
6.2 Preparations for Disassembly............................................................................................................................................................................ 6-1
6.3.1 Removing the Battery Door and Battery ............................................................................................................................................ 6-2
6.3.2 Installing the Battery and Battery Door .............................................................................................................................................. 6-3
6.3.3 Removing the Display Assembly .......................................................................................................................................................... 6-4
6.3.4 Removing the Rear Housing ................................................................................................................................................................... 6-5
6.3.5 Removing the Infrared Backboard ....................................................................................................................................................... 6-7
6.3.6 Removing the Main Board ....................................................................................................................................................................... 6-7
6.3.7 Removing the Parameter Board .......................................................................................................................................................... 6-10
6.3.8 Removing the NIBP Gas Circuit and NIBP Valve ............................................................................................................................. 6-12
6.3.9 Removing the NIBP Pump ..................................................................................................................................................................... 6-13
6.3.12 Removing the Interface Board .......................................................................................................................................................... 6-17
6.4 Disassembling the N1 Main Unit with Two Batteries ............................................................................................................................. 6-18
6.4.1 Removing the Battery Door and Battery .......................................................................................................................................... 6-18
6.4.2 Installing the Battery and Battery Door ............................................................................................................................................ 6-20
6.4.3 Removing the Display Assembly ........................................................................................................................................................ 6-21
2 BeneVision N1 Patient Monitor Service Manual
Page 11
7.7.2 Parts List ....................................................................................................................................................................................................... 7-11
7.8.2 Parts List ....................................................................................................................................................................................................... 7-13
7.9 Dock ......................................................................................................................................................................................................................... 7-14
7.9.1 Main Unit ..................................................................................................................................................................................................... 7-14
7.9.2 Top Shell Module of Dock ...................................................................................................................................................................... 7-16
7.9.3 Right Shell Module of Dock .................................................................................................................................................................. 7-17
7.9.4 Dock Port Module ..................................................................................................................................................................................... 7-18
7.10 Other Maintenance and Repair Parts ........................................................................................................................................................ 7-19
A Electrical Safety Inspection ............................................................................................................................................ A-1
4 BeneVision N1 Patient Monitor Service Manual
Page 12
6.4.4 Removing the Rear Housing ................................................................................................................................................................ 6-23
6.4.5 Removing the Infrared Backboard..................................................................................................................................................... 6-24
6.4.6 Removing the Main Board .................................................................................................................................................................... 6-25
6.4.7 Removing the Parameter Board ......................................................................................................................................................... 6-27
6.4.8 Removing the NIBP Gas Circuit and Secondary Battery housing ........................................................................................... 6-29
6.4.9 Removing the NIBP Pump and Valve ................................................................................................................................................ 6-30
6.4.10 Removing the Interface Board .......................................................................................................................................................... 6-31
6.5 Removing the N1 Dock ..................................................................................................................................................................................... 6-32
6.5.1 Removing the Bottom Housing .......................................................................................................................................................... 6-32
6.5.2 Removing the Right Shell Assembly ................................................................................................................................................. 6-32
6.5.3 Removing the Rotatable Plate ............................................................................................................................................................ 6-33
6.5.4 Removing the Transition Board .......................................................................................................................................................... 6-33
6.5.5 Removing the Port Module .................................................................................................................................................................. 6-34
6.5.6 Removing the Interface Board and Power Board ......................................................................................................................... 6-35
6.5.7 Removing the LED Indicator ................................................................................................................................................................ 6-36
6.6 Removing the N1 Modular Rack ................................................................................................................................................................... 6-36
6.6.1 Removing the Rear Housing ................................................................................................................................................................ 6-36
6.6.2 Removing the Unlock Rod .................................................................................................................................................................... 6-37
6.6.3 Removing the Infrared Backboard..................................................................................................................................................... 6-37
6.6.4 Removing the Transition Board .......................................................................................................................................................... 6-38
6.6.5 Removing the Contact Screws ............................................................................................................................................................ 6-38
6.7 Removing N1 Transport Dock ........................................................................................................................................................................ 6-39
6.7.1 Removing the Transition Board .......................................................................................................................................................... 6-39
6.7.2 Removing the Power Adapter ............................................................................................................................................................. 6-39
7 Parts ................................................................................................................................................................................. 7-1
7.2 Main Unit .................................................................................................................................................................................................................. 7-2
7.2.2 Parts List ......................................................................................................................................................................................................... 7-2
7.3.2 Parts List ......................................................................................................................................................................................................... 7-3
7.4.2 Parts List ......................................................................................................................................................................................................... 7-4
7.5 Main Unit Assembly .............................................................................................................................................................................................. 7-5
7.5.1 Exploded View (for CO2 Module Configuraiton) ............................................................................................................................. 7-5
7.5.2 Parts List ......................................................................................................................................................................................................... 7-5
7.5.4 Parts List ......................................................................................................................................................................................................... 7-7
7.6.2 Parts List ...................................................................................................................................................................................................... 7-10
The monitor is to be used in healthcare facilities. It can also be used during patient transport with road, rotary, and
fixed-wing ambulances.
intended for monitoring, displaying, reviewing, storing, alarming, and transferring of patients'
2.1.1 Front View
1. Alarm lamp
When a physiological alarm or technical alarm occurs, this lamp lights and flashes corresponding with the alarm priority:
High priority alarms: the lamp quickly flashes red.
Medium priority alarms: the lamp slowly flashes yellow.
Low priority alarms: the lamp lights in cyan without flashing.
2. Ambient light sensor
When screen brightness is set to auto, the system automatically adjusts screen brightness according to the strength of
ambient light.
3. Display and touchscreen
4. Battery LED
Yellow: the battery is being charged.
Green: the battery is fully charged.
Flashing green: the monitor runs on battery power.
Flashing yellow: the battery is malfunctioning.
eneVision N1 Patient Monitor Service Manual 2-1
B
Page 16
Off: no battery is installed, or the monitor is powered off and no external power is connected.
5. External power supply indicator
On: when external power supply is connected.
Off: when external power supply is not connected.
2.1.2 Left View
1. Communication interface: used for communication between the N1 and host monitor (BeneVision N series
monitor)
2. Infrared filter: used for communication between the N1 and Passport m series monitor; used for communication
between the N1 and N series monitor if the communication interface does not work.
3. Contact: used for receiving power supply from the host monitor (Passport m series monitor or BeneVision N series
monitor).
4. Multi-pin connector: connects N1 to the Modular Rack or Dock.
5. External DC power input connector: connects the N1 to the AC adapter.
NOTE
To ensure good contact, clean the contacts regularly, as dust and dirt may collect on them. When cleaning
the contacts, wipe them with cotton, dampened with alcohol. Using forceps is recommended.
In the case of water spray, dry the Multi-pin connector of the N1 before connecting the N1 to the Modular
Rack or Dock.
-2BeneVision N1 Patient Monitor Service Manual
2
Page 17
2.1.3 Right View
1. Power
switch
2. Sample line connector for sidestream CO2 (available when CO2 is configured)
3. Gas outlet (available when CO2 is configured)
4. ECG cable connector
5. SpO
sensor connector
2
6. NIBP cuff connector
7. Multifunctional connector: outputting analog and defib. synchronization signals
8. Speaker
9. IBP cable connector
10. Temperature probe connector
eneVision N1 Patient Monitor Service Manual 2-3
B
Page 18
2.1.4 Bottom View
1
.Clip: fastens N1 when N1 is in use with the host monitor, Dock or Modular Rack.
2.Latch: locks N1 when N1 is in use with the host monitor, Dock or Modular Rack. Pressing here releases the N1 so
that the N1 can be removed from the host monitor, Dock or Modular Rack.
3.Battery door: secures the battery.
-4BeneVision N1 Patient Monitor Service Manual
2
Page 19
2.2 System Composition
The N1 monitor system consists of the N1 main unit, Dock, Modular Rack, single-slot module, external display, AC
adapter, Transport Dock and the peripherals such as keyboard, mouse, scanner, and printer, as shown in the figure
below:
In the figure:
N1 main unit: incor
porates multiple functions such as display, control, parameter measurement, printing, and
communication. It is the core of the whole system and supports the following functions:
Integrate the basic parameter measurement module for measuring the ECG, RESP, SpO
, TEMP, NIBP, IBP,
2
CO2 and other parameters.
Connect to the Modular Rack, and to the parameter measurement modules such as the three kinds of
CO2 modules through the Modular Rack, so as to display and store module measurement results and
report alarms.
Be directly inserted into the host monitor as a submodule.
Output analog (ECG and IBP) and defib. synchronization signals.
Be powered using the DC and smart battery.
eneVision N1 Patient Monitor Service Manual 2-5
B
Page 20
Support Wi-Fi wireless communication.
Support alarm sound and light output.
Connect to the Dock to extend display, USB, LAN, host monitor communication, and power supply
interfaces.
AC adapter: provides an external power supply to the N1 main unit used separately.
Dock: secured beside the bed and connected to the N1 main unit. It can supply power to the N1 and charge
the N1 battery. It can be connected to a touchscreen, mouse, keyboard, barcode scanner, USB drive, external
host monitor, and wired network.
Modular Rack: enables N1 to connect to the mainstream CO
, sidestream CO2, microstream CO2 and PiCCO
2
modules.
Transport Dock: secures N1 in airplanes and ambulances and supplies power to N1.
2.2.1 Main Unit System Structure
The figure below shows the system structure of the N1 main unit. One should be selected from the battery B and CO2
related components.
In the figure:
Battery B exists only in the N1 main unit (without CO
The CO
module and its interface exist only in the N1 main unit (with CO2).
2
The Wi-Fi module is optional.
There are three kinds of SpO2 boards.
).
2
-6BeneVision N1 Patient Monitor Service Manual
2
Page 21
2.2.1.1 Main Board
The main board is the control core of the entire system. Its main functions are as follows:
display driver and backlight control
touchscreen driver
external display VGA drive
audio driver
integrating the alarm lamp
integrating the light sensor
wired network communication
wireless network communication (through the Wi-Fi module)
USB driver
real-time clock
data and configuration storage
DC/DC conversion and power switching
power management (power-on/off, status monitoring, etc.)
battery management (in-position detection, charge control, discharge control, status monitoring, etc.)
integrating the external power LED and battery LED
analog and defib. synchronization signals transfer
communication with the Dock
communication with the Modular Rack
communication with the host monitor
communication with the infrared communication board
communication with the multi-parameter board
communication with the CO
module
2
2.2.1.2 Multi-parameter Board
This board completes the basic parameter measurement functions, including:
parameter measurement (in the backboard mode, supporting Mindray, Masimo, and Nellcor SpO2)
2
2-channel TEMP parameter measurement
2-channel IBP parameter measurement
isolation of parameters except NIBP from the ground
isolation of ECG from other parameters
direct output of analog signals (1-channel ECG and 2-channel IBP)
direct output of defib. synchronization signals
data interaction with the main board through the serial port
BeneVision N1 Patient Monitor Service Manual 2-7
Page 22
2.2.1.3 CO
This module completes the CO2 parameter measurement function.
Module
2
2.2.1.4 Parameter Interface Board
This board provides the following functions:
ECG signal transfer and isolation of ECG signals from other signals
SpO2 signal transfer
TEMP signal transfer
IBP signal transfer
2.2.1.5 Multifunctional Connector Board
This board outputs analog signals and transfers defib. synchronization signals.
2.2.1.6 Switch Keypad Board
This board provides the power switch and power LED.
2.2.1.7 Infrared Backboard
The infrared backboard is used for infrared/contact communication between N1 and the host monitor. It receives power
supply from the host monitor and connects the multi-parameter board to the main board.
2.2.2 Modular Rack and Dock System Composition
8BeneVision N1 Patient Monitor Service Manual
2-
Page 23
The above figure shows the connection and system composition relationships between the N1, Modular Rack, and Dock.
The three can be interconnected through the pluggable 54-pin connector. N1 functions as the system core. The Modular
Rack and Dock expand the functions of N1.
When N1 is separately connected to the Modular Rack, the N main unit can communicate with the parameter module.
N1 can implement power supply and measurement parameter interaction with the parameter module through the Rack
adapter board and Rack NIOS board. The parameter module is a one-slot parameter module.
When N1 is in use with the Dock, it can be connected to the Do
board. In this way, the Dock interface board supplies power to N1. N1 connects the USB, external device connector (Mini
DB15), VGA and wired network port to the Dock response interface to implement external interface connection and
extension.
When N1, the Modular Rack, and the Dock form a whole, functions of the above two combinations are implemented at
the same time.
ck interface power board through the Dock adapter
eneVision N1 Patient Monitor Service Manual 2-9
B
Page 24
2.2.3 Extended Parameter Module Composition
N1 can be connected to the following four external modules through the Modular Rack:
Sidestream CO2
module
Microstream
CO
module
2
The figure below shows parameter module expansion through the Modular Rack:
Rack
NIOS
board
Rack
adapter
board
Seri al por t
54 pins
Three kinds of extension modules
Pluggable connector
CO
2
N1 main unit
In this application, N1 and the external module are inserted into the specified slots of the Modular Rack. The external
module is connected to the Rack adapter board through the Rack NIOS board in the Modular Rack. The battery in the N1
main unit powers the parameter module. The Rack adapter board and N1 main unit are connected through the 54-pin
pluggable connector. The Modular Rack can also be inserted into the Dock so that the Dock can supply power to it.
10BeneVision N1 Patient Monitor Service Manual
2-
Page 25
3Unpacking and Installation
3.1 Unpacking
Open the package and take out the packing list. Check that all the articles included on the packing list are available and
the quantity and specification are correct. Note that:
1. All the optional parts purchased by the customer shall also be checked.
2. Immediately contact the supplier if any item in the package is inconsistent with that listed on the packing list.
3. In case of damage during transportation, keep the packing material and immediately contact the supplier.
4. Keep the packing material until the product acceptance is completed.
3.2 Preparations for Installation
3.2.1 Installation Site Requirements
1. Ensure that the installation site meets all safety, environmental and power requirements.
2. Check that required power sockets are available.
3. Check whether a network connector is available if the monitor needs to connect to the wired network.
WARNING
When using the N1 Dock, use only the power cord supplied by the manufacturer to connect the AC power
supply. When using the monitor alone, use only the DC adapter supplied by the manufacturer to connect
the external power supply.
eneVision N1 Patient Monitor Service Manual 3-1
B
Page 26
3.2.1.1 Environmental Requirements
To avoid explosion hazard, do not use the equipment in the presence of flammable anesthetics, vapors or liquids. To
avoid impact on the normal operation, the monitor shall be placed in an environment free from vibration, dust and
corrosive substances.
Environmental specifications are as follows:
Main unit, adapter, and Transport Dock
Item Temperature (°C) Relative Humidity (Non-condensing) Atmospheric Pressure (kPa)
Work ing
environment
Storage and
transport
environment
Microstream CO2 module
Item Temperature (°C) Relative Humidity (Non-condensing) Atmospheric Pressure (kPa)
Work ing
environment
Storage and
transport
environment
0~40 5%~95%57.0~107.4
–30~70 5%~95% 16.0–107.4 (without the CO
50.0–107.4 (with the CO
0~40 15%~95%57.3~105.3
–20~60 10%~95%57.3~105.3
module)
2
module)
2
Sidestream CO2 module
Item Temperature (°C) Relative Humidity (Non-condensing) Atmospheric Pressure (kPa)
Work ing
5~40 15%~95%57.3~105.3
environment
Storage and
–20~60 10%~95%57.0~107.4
transport
environment
N1 Dock and N1 Modular Rack
Item Temperature (°C) Relative Humidity (Non-condensing) Atmospheric Pressure (kPa)
Work ing
0~40 15%~95%57.0~107.4
environment
Storage and
–20~60 10%~95%16.0~107.4
transport
environment
NOTE
Unless otherwise specified, the environmental specifications of a parameter module are the same as those
of the main unit.
-2BeneVision N1 Patient Monitor Service Manual
3
Page 27
3.2.1.2 Electrical Requirements
The monitor can be powered by an AC power supply through the N1 Dock, or it can be directly connected to an external
power supply using an adapter. The power cord must be connected only to a power socket with protective grounding.
Make sure that:
1. All cables and connectors are not damaged, and pins are not loose.
2. Patient cables and leadwires are properly insulated, and connectors are not loose.
The power input specifications of the adapter and Transport Dock are as follows:
00-240V~(15%, +10%)
Input voltage
Input current
Frequency
1
FSP030-RCAM-G: 1.0 ~ 0.6 A
50 Hz/60 Hz(±3Hz)
The AC po
Input voltage
Input current
Frequency
wer input specifications of the N1 Dock are as follows:
0-240V~(±10%)
10
65A ~0.35A
0.
0 Hz/60 Hz (±3Hz)
5
3.2.1.3 Turning on the Monitor
1. Check the monitor for any mechanical damage. Make sure that all external cables and accessories are properly
connected.
2. For the first use, connect the monitor to the external power supply (AC adapter, Dock, powered-on host monitor, or
Transport Dock) for a while and then turn on the monitor to activate the battery.
3. When using the monitor alone, connect the monitor and the DC adapter. If you run the patient monitor on battery
power, ensure that the battery is sufficiently charged. If the monitor is powered using the N1 Dock, check that the
power LED on the Dock is on.
4. Press the power switch to turn on the monitor.
neVision N1 Patient Monitor Service Manual 3-3
Be
Page 28
FOR YOUR NOT
ES
-4BeneVision N1 Patient Monitor Service Manual
3
Page 29
4 Testing and Maintenance
4.1 Introduction
To ensure the patient monitor always functions properly, qualified service personnel should perform regular inspection,
maintenance and test. This chapter provides a checklist of the testing procedures for the patient monitor with
recommended test equipment and frequency. The service personnel should perform the testing and maintenance
procedures as required and use appropriate test equipment.
The testing procedures provided in this chapter are intended to verify that the patient monitor meets the performance
specifications. If the patient monitor or a module fails to perform as specified in any test, repairs or replacement must be
done to correct the problem. If the problem persists, contact our Customer Service Department.
CAUTION
All tests should be performed by qualified service personnel only.
Care should be taken when changing the settings in MaintenanceandConfigurationmenus to avoid loss of
data.
Service personnel should possess a working knowledge of the test tools and make sure that test equipment
and cables are applicable.
4.1.1 Test Equipment
Required Test Equipment is listed in the specific test procedure.
4.1.2 Preventative Maintenance
It is recommended to verify accuracy and calibrate the patient monitor as needed at least once every two years (once a
year for CO
following sections.
Visual inspection
modules). Preventative maintenance involves the following test items. For detailed procedures, refer to the
2
NIBP tests and calibration
CO
BeneVision N1 Patient Monitor Service Manual 4-1
module tests and calibration
2
Page 30
4.1.3 Recommended Frequency
Check/Maintenance Item Frequency
Visual inspection 1. When the monitor is installed for the first time or is reinstalled.
Power-on test 1. When the monitor is installed for the first time or is reinstalled.
2. Following any repair or parts replacement.
EGG test and
calibration
Resp test
SpO2 test
Performance test 1. When the user suspects that the measurement is incorrect.
Module calibration
2. Following any repair or replacement of the relevant module.
3. At least once every two years.
Note: Perform CO
module tests at least once a year.
2
NIBP tests and
calibration
NIBP accuracy test
NIBP leakage test
Temp test
IBP test and calibration Performance test
Pressure calibration
Sidestream and
microstream CO
2
modules tests and
Leakage test
Performance test
Module calibration
calibration
Analog output test When the user suspects that the analog output does not function
properly.
Electrical safety tests Enclosure leakage
current test
Earth leakage test
1. Following any repair or replacement of the power module.
2. When the monitor is dropped.
3. At least twice a year or as required.
Patient leakage current
Patient auxiliary current
Battery check Functionality test 1. When the monitor is installed for the first time.
2. After the battery is replaced.
Performance test Once a year or when the battery runtime reduces significantly.
4-2 BeneVision N1 Patient Monitor Service Manual
Page 31
4.2 Preventative Maintenance Procedures
4.2.1 Visual Inspection
Inspect the monitor for obvious damage. The test is passed if the monitor has no obvious damage. Follow these
guidelines when inspecting the monitor:
Carefully inspect the enclosure, display, buttons, and knob for obvious damage.
Inspect the power cord and module accessories for obvious damage.
Inspect all external connections for loose connectors, bent pins or frayed cables.
Inspect all connectors on the monitor for loose connectors or bent pins.
Make sure that safety labels and data plates on the monitor are clearly legible.
4.2.2 NIBP Tests
4.2.2.1 NIBP Leakage Test
NOTE
You should perform the NIBP leakage test before any other NIBP concerned test and calibration.
Tools required:
NIBP cuff for adult patient
Air hose
Cylinder
Follow this procedure to perform the test:
1. Set
2. Connect the NIBP cuff to the NIBP connector on the monitor.
3. Wrap the cuff around the cylinder as shown below.
Patient Category to Adult.
Monitor
Hose
Cylinder
Cuff
NIBP cuff connector
4. Select
After about 20s, the cuff automatically deflates, which indicates that the NIBP leakage test is completed. If no message is
displayed in the NIBP parameter area, it indicates that the system has no leakage. If the message "
displayed, it indicates that the system may have a leakage. In this case, check that all connections are in good condition
and the cuff and air tubing have no leakage. Then perform the test again.
Main Menu→System→Maintenance >>→Module→NIBP→NIBP Leakage Test. The message
Leakage Testing…" is displayed in the NIBP parameter area.
"
NIBP Airway Leak" is
BeneVision N1 Patient Monitor Service Manual 4-3
Page 32
You can also perform a manual leakage test:
Balloon pump
1. Perform steps 1-4 in section 1.2.2.2 NIBP Accuracy Test.
2. Raise the pressure in the rigid vessel to 250 mmHg using the balloon pump. Then wait for 5 seconds so that the
measured values become stable.
3. Record the current pressure value, and use a timer to count time. Then record the pressure value after 60 seconds.
4. Compare the two pressure values and make sure that the difference is not greater than 6 mmHg.
4.2.2.2 NIBP Accuracy Test
Tools required:
T-shape connector
NIBP hose
Balloon pump
Rigid vessel with 500±25 ml internal volume
Standard manometer (calibrated with accuracy greater than or equal to 1 mmHg)
Follow this procedure to perform the test:
1. Connect the monitor and the test tools as shown below.
Monitor
NIBP hose
NIBP cuff connector
2. Before inflation, check whether the reading on the manometer is 0. If not, open the valve of the balloon pump to let
the whole airway open to the atmosphere. Close the valve after the reading becomes 0.
3. Select
4. Check that the reading of the manometer and that of the monitor are both 0 mmHg.
5. Raise the pressure in the rigid vessel to 50 mmHg using the balloon pump. Then wait for 10 seconds so that the
Main Menu→Maintenance → enter the required password → Module → NIBP →NIBP Accuracy Test.
measured values become stable.
Standard manometer
Rigid vessel
6. Check that the difference between the reading of the manometer and that of the monitor is less than or equal to 3
mmHg or less. If the difference is greater than 3 mmHg, contact Mindray service.
7. Raise the pressure in the rigid vessel to 200 mmHg using the balloon pump. Then wait for 10 seconds so that the
measured values become stable. Then repeat step 6.
NOTE
You can use an NIBP simulator to replace the balloon pump and the standard manometer.
You can use an appropriate cylinder and a cuff instead of the rigid vessel.
4-4 BeneVision N1 Patient Monitor Service Manual
Page 33
4.2.3 Sidestream and Microstream CO
Modules Tests
2
4.2.3.1 Leakage Test
1. Connect the module and the monitor.
2. Wait until CO2 warmup is finished and then completely block the gas inlet of the module or water trap (by using
your finger or a pinched sample line). The sidestream and microstream CO
Sidestream: After 3 seconds, the alarm message "CO
Airway Occluded" is displayed on the screen. Block
2
modules will behave as follows:
2
the gas inlet for another 60 seconds. Select Main Menu→Maintenance→ enter the required password →
Module→ CO
to check that the current CO2 flow rate is less than 10 ml/min. If the alarm message remains
2
on the screen, it indicates that the module does not leak. If the alarm message CO2 Airway Occluded
disappears, or the flow rate is greater than or equal to 10 ml/min, it indicates that the module leaks.
Microstream: After 3 seconds, the alarm message "CO
inlet for another 30 seconds. If the alarm message "
Purging" is displayed on the screen. Block the gas
2
CO2 Airway Occluded" is displayed, it indicates that the
module does not leak.
4.2.3.2 Accuracy Test
Tools required:
For microstream CO
module and sidestream CO2 module without O2 module, a gas cylinder with 5±0.03% CO2,
2
21.0% O2 and balance gas N2 (P/N 0075-00-0033-01) or a steel gas cylinder with:
CO2 concentration 3% - 7%
a/c ≤ 0.01 (where a = absolute gas concentration accuracy, c = gas concentration)
balance gas N2
For sidestream CO
module with O2 module equipped, a steel gas cylinder (P/N 0075-00-0048-01) with 6% CO2, 4%
2
Desflurane, 45% N2O, and 45% O2,
T-shape connector
Air tubing
Flowmeter
1. Connect the module and the monitor.
2. Wait until the CO
module warmup finishes. Check the airway for leakage and perform a leakage test as well to
2
make sure that the airway has no leakage.
3. Select Main Menu →Maintenance → enter the required password → Module → CO2.
4. Connect the test system as follows:
BeneVision N1 Patient Monitor Service Manual 4-5
Page 34
Relief
valve
T-shape connector
Gas
cylinder
Tubing
Monitor
Flowmeter
5. Open the relief valve and adjust it until the flowmeter has a stable reading between 10 ml/min and 50 ml/min.
6. Check that the real-time CO
value is within 6.0±0.2% in the CO2 Maintenance menu (for microstream CO2, the
2
value is 45±2 mmHg).
7. Replace the cylinder to the steel gas cylinder with >40% O
module with O
(80%≤O
module equipped) and verify that the real-time O2 value error is within ±2% (when O2≤80%) or ±3%
2
≤100%).
2
4.2.3.3 Calibration
Tools required:
For microstream CO
21.0% O2 and balance gas N2 (P/N 0075-00-0033-01) or a steel gas cylinder with:
CO2 concentration 3% - 7%
a/c ≤ 0.01 (where a = absolute gas concentration accuracy, c = gas concentration)
balance gas N2
For sidestream CO
Desflurane, 45% N2O, and 45% O2,
T-shape connector
module and sidestream CO2 module without O2 module, a gas cylinder with 5±0.03% CO2,
2
module with O2 module equipped, a steel gas cylinder (P/N: 0075-00-0048-01) with 6% CO2, 4%
2
and balance gas N2 (applicable to sidestream CO2
2
Air tubing
Flowmeter
1. Make sure that the sidestream or microstream CO2 module has been warmed up or started up.
2. Check the airway for leakage and perform a leakage test as well to make sure that the airway has no leakage.
3. Select Main Menu→Maintenance → enter the required password →Module→CO2.
4. In the Calibrate Maintenance menu, select Zero.
5. After the zero calibration succeeds, connect the monitor and the test tools as follows:
4-6 BeneVision N1 Patient Monitor Service Manual
Page 35
Flowmeter
Relief valve
T-shape connector
Gas cylinder
6. Open the relief valve and adjust it until the flowmeter has a stable reading between 10 ml/min and 50 ml/min.
7. In the CO2 Maintenance menu, select 6% (the CO2 concentration) for CO2 calibration.
8. After the measured CO
9. Replace the cylinder to the steel gas cylinder with >40% O2 and balance gas N2(applicable to sidestream CO2
module with O2 module equipped) and calibrate O2.
If the calibration succeeds, the message "
calibration fails, the message "
and perform another calibration. If the calibration fails several times, return the module to Mindray for repair.
concentration becomes stable, select Calibrate to calibrate the CO2 module.
2
Calibration Completed!" is displayed in the Calibrate menu. If the
Calibration Failed!" is displayed. In this case, check whether the operations are correct
Tubing
Monitor
4.3 Power-on Test
This test aims to verify that the monitor can power up properly.
When the monitor is powered only by the battery, follow this procedure:
1. Install the battery on the monitor.
2. Press the power switch to turn on the monitor. The battery LED flashes.
When the monitor is powered by using the adapter, follow this procedure:
1. Install the battery on the monitor. Connect the monitor and the adapter. Then connect the adapter and the AC
power supply. In this case, the external power LED and the battery LED are on.
2. Press the power switch to turn on the monitor.
When the monitor is in use with the Dock, follow this procedure:
1. Install the battery on the monitor. Install the monitor on the Dock. Then connect the Dock and the AC power supply.
In this case, the power LED on the Dock and the battery LED on the monitor are on.
2. Press the power switch to turn on the monitor.
The monitor automatically performs a self-test upon startup. During the self-test, you can hear the alarm tone, and the
BeneVision N1 Patient Monitor Service Manual 4-7
Page 36
alarm lamp glows yellow and red. This indicates that the visible and audible alarm indicators operate properly.
4.4 Module Performance Tests
4.4.1 ECG Tests
4.4.1.1 ECG Performance Test
Tools required: Medsim300B patient simulator or other equivalent simulator
1. Connect the patient simulator with the ECG module using an ECG cable.
2. Set the patient simulator as follows: ECG sinus rhythm, HR = 80 bpm, and amplitude = 1 mV.
3. Verify that the ECG waves are displayed correctly without noise and the displayed HR value is within 80±1 bpm.
4. Disconnect each of the leads in turn and observe the corresponding lead off message displayed on the screen.
5. Set the output of the simulator to deliver a paced signal, and set
pulse marks on the monitor screen.
Paced to Yes on the monitor. Check the pace
4.4.1.2 ECG Verification
Test tool: vernier caliper
1. Select the ECG parameter window or waveform area →Filter→Diagnostic.
2. Select
3. Select
4. Compare the amplitude of the square wave with that of the scale. The difference should be less than 5%.
5. After completing the calibration, select
If necessary, you can print out the square wave and wave scale through the recorder and then measure the difference.
Main Menu →Maintenance → enter the required password → Module.
Calibration. A square wave appears on the screen and the message "ECG Calibrating" is displayed.
Stop Calibration.
4.4.2 Resp Test
Tools required:
Medsim300B patient simulator or other equivalent simulator
1. Connect the patient simulator and the monitor using a non ESU-proof cable and set lead II as the respiration lead of
the monitor.
2. Configure the simulator as follows: lead II as the respiration lead, base impedance line as 500 Ω; delta impedance as
1 Ω, respiration rate as 20 rpm.
3. Verify that the Resp wave is displayed without any distortion and the displayed Resp value is within 20±1 rpm.
4-8 BeneVision N1 Patient Monitor Service Manual
Page 37
NOTE
4.4.3 SpO
Test
2
Tools required: None
1. Connect the SpO2 sensor to the SpO2 connector on the monitor. Set Patient Category to Adult and PR Source to
SpO2 on the monitor.
2. Apply the SpO
sensor to the ring finger of a healthy person.
2
3. Check the Pleth wave and PR reading on the screen and make sure that the displayed SpO2 is within 95% and100%.
4. Remove the SpO2 sensor from your finger and check that the SpO2 sensor off alarm is triggered.
The SpO
accuracy has been verified in human experiments by comparing with the arterial blood sample reference
2
measured by a CO-oximeter. Pulse oximeter measurements are statistically distributed and about two-thirds of the
measurements are expected to fall within the specified accuracy range compared to CO-oximeter measurements.
A simulator cannot be used to assess the accuracy of a pulse oximeter monitor or a SpO
sensor. Instead, it
2
can only verify that whether the monitor is functional. The accuracy of a pulse oximeter monitor or a SpO
sensor needs to be verified by clinical data.
4.4.4 Temp Test
2
Tools required: Resistance box (with accuracy greater than or equal to 0.1 Ω)
1. Connect the two pins of any Temp connector of a module to the two ends of the resistance box using two wires.
2. Set the resistance box to 1354.9 Ω (the corresponding temperature is 37°C).
3. Verify that the displayed value is within 37±0.1°C.
4. Repeat steps 1 to 3 to verify each Temp channel of the monitor.
4.4.5 NIBP Tests
Refer to section 4.2.2NIBP Tests.
BeneVision N1 Patient Monitor Service Manual 4-9
Page 38
4.4.6 IBP Test and Calibration
4.4.6.1 Performance Test
Tools required:
Patient simulator Medsim300B, MPS450, or equivalent equipment
IBP adapter cable for test (P/N 009-002199-00 for Medsim300B; P/N 009-002198-00 for MPS450)
1. Connect the patient simulator to the IBP connector on the monitor.
2. Set the patient simulator to output 0 mmHg to an IBP channel.
3. Press the Zero key from the IBP Setup menu to perform a zero calibration.
4. Set static pressure to 200 mmHg on the patient simulator.
5. Check that the value displayed on the monitor is within 200±2 mmHg.
6. If the error is beyond ±2 mmHg, calibrate the IBP module. If the IBP module was calibrated with a dedicated
reusable IBP sensor, check the calibration together with this IBP sensor.
7. Set the patient simulator to separately output 120/80 mmHg ART signals and 120/0 mmHg LV signals to the IBP
channel and check that the IBP wave is displayed correctly.
8. Repeat the preceding steps to test all IBP channels.
4.4.6.2 Pressure Calibration
Method 1:
Tools required:
Patient simulator Medsim300B, MPS450, or equivalent equipment
IBP adapter cable for test (P/N 009-002199-00 for Medsim300B; P/N 009-002198-00 for MPS450)
1. Connect the patient simulator to the IBP connector on the monitor.
2. Set the patient simulator to 0 pressure for the target IBP channel.
3. Press the Zero key from the IBP Setup menu to perform a zero calibration.
4. Set static pressure to 200 mmHg on the patient simulator.
5. Select Main Menu→Maintenance→ enter the required password →Module→ IBP. Set the calibration
value to 200 mmHg.
6. Select the
If the calibration succeeds, the message "Calibration Completed!" is displayed. If the calibration fails, a corresponding
message is displayed.
Calibrate button next to the target IBP channel to start calibration.
4-10 BeneVision N1 Patient Monitor Service Manual
Page 39
T-shape
Pressure sensor
Pressure adapter cable
Method 2:
Tools required:
Standard manometer
Balloon pump
Air tubing
T-shape connector
1. Connect the 3-way stopcock, the sphygmomanometer and the balloon pump using the T-shape connector, as
shown below.
2. Turn on the 3-way stopcock to the air to zero the transducer. Then open the stopcock to the sphygmomanometer.
3-way stopcock
connector
Sphygmomanometer
3. Select Main Menu→Maintenance Maintenance→ enter the required password →Module→ IBP. In the
displayed interface, set the target calibration value of the target channel. Value range: 80 to 300 mmHg.
4. Use the balloon pump to inflate until the reading of the sphygmomanometer approximates to the preset
calibration value.
5. Adjust the calibration value in the IBP Maintenance menu until it is equal to the reading of sphygmomanometer.
6. Select the
If the calibration succeeds, the message "
message is displayed.
4.4.7 Sidestream and Microstream CO
Refer to section4.2.3 Sidestream and Microstream CO2 Modules Tests.
Calibrate button next to the target IBP channel to start calibration.
Calibration Completed!" is displayed. If the calibration fails, a corresponding
2
Modules Tests and Calibration
IBP module
BeneVision N1 Patient Monitor Service Manual 4-11
Page 40
4.5 Analog Output Test
Tools required:
Patient simulator Medsim300B, MPS450, or equivalent equipment
Oscilloscope
1. Connect the patient simulator to the monitor using an ECG or IBP cable.
2. Connect the oscilloscope to the Auxiliary Output Connector on the monitor.
3. Verify that the waves displayed on the oscilloscope are identical with those displayed on the monitor.
4.6 Electrical safety tests
WARNING
Electrical safety tests are a proven means of detecting abnormalities that, if undetected, could prove
dangerous to either the patient or the operator.
All tests can be performed using commercially available safety analyzer test equipment. Maintenance
personnel shall ensure the adaptability, functional completeness and safety of these pieces of test
equipment, and be familiar with their usage.
Electrical safety tests shall comply with the following standards: IEC 60601-1 and ANSI/AAMI ES60601-1.
In case of other stipulations in local laws and regulations, implement electrical safety tests by following
relevant stipulations.
All devices driven by AC power and connected to medical instruments in patient zones must comply with
the IEC 60601-1 standard. And electrical safety tests on these devices must be implemented in accordance
with the test interval of the patient monitor.
Use certified safety analyzer (for example, UL, CSA or AMAI) as instructed to perform relevant tests.
NOTE
Electrical safety check shall be performed after repair or routine maintenance. Ensure that all cover boards,
panels and screws are correctly installed before implementing electrical safety tests.
Electrical safety tests are used to timely detect potential electrical safety risks that might cause damage to
patients, operators or maintenance personnel. Electrical safety tests must be carried out under normal
environmental conditions (that is, normal temperature, humidity and barometric pressure).
See Appendix A Electrical Safety Inspection for electrical safety tests.
4-12 BeneVision N1 Patient Monitor Service Manual
Page 41
4.7 Network Print Test
Tools required: hub and network cable
NOTE
HP LaserJet Pro M202dw laser printer is recommended for this monitor.
4.7.1 Equipment Connection and Setup
1 Connect the monitor and network printer to a HUB using common network cables as follows:
Monitor
Network cable Network cable
HUB
2 Set the IP address:
a. Use the current Dock net setting: Select Main Menu→Maintenance→enter the password →Dock Setup. Set Net
Setting Type to Use current Dock net setting.
b. Select the IP setup tab, and set the IP address of the monitor to one in the same network segment as that of the
network printer (refer to the instructions for use of the printer).
3 Set the printer:
a. Host mode: Select Main Menu→Maintenance→User Maintenance→ enter the password →Dock Setup. Set Work
Mode to Host Mode. Select the printer setup tab. Set the IP address of the printer to one in the same network
segment as that of the monitor. Select Print Test Page to verify that the printer works properly.
b. Dock mode: Select Main Menu→Maintenance→User Maintenance→ enter the password →Dock Setup. Set
Work Mode to Dock Mode. On the User Maintenance interface, select Print→Printer, and set the IP address of
the printer. Select Print Test Page to verify that the printer works properly.
Network
printer
4.7.2 Print Function Test
1 Enter the Demo mode of the monitor.
2 Select Main Menu→Report column →Normal Report→Realtime Report. Click Print. The network printer prints
the real-time report.
BeneVision N1 Patient Monitor Service Manual 4-13
Page 42
4.8 Battery Check
4.8.1 Functionality Test
1. Remove the battery from the monitor.
2. Check that the monitor works properly when being powered by the external DC power supply or by the AC power
supply through the Dock.
3. Install the battery according to the procedures described in the Operator's Manual.
4. Disconnect the external DC power supply or the AC power supply on the Dock. Check whether the monitor still
works properly.
4.8.2 Performance Test
Refer to the related content in the Battery chapter in BeneVison N1 Patient Monitor Operator's Manual.
4-14 BeneVision N1 Patient Monitor Service Manual
Page 43
5 Troubleshooting
5.1 Overview
In this chapter, monitor problems are listed along with possible causes and recommended corrective actions. Refer to
the tables to check the monitor and identify and eliminate these problems.
The recommended solutions given in this chapter address most of the equipment problems you encounter, but do not
include any problems that may arise. Contact us if you have a problem that is not covered in this section.
5.2 Part Replacement
Printed circuit boards (PCBs) and other major parts and components in the monitor are replaceable. Once you have
located the faulty PCB, follow the steps in 6 Repair and Disassembly to replace it and check if the failure is eliminated or
if the monitor passes the relevant tests. If the failure is eliminated, it indicates that the original PCB is damaged. In this
case, send the defective PCB to us for repair. If the failure persists, reinstall the original PCB and continue
troubleshooting as directed in this chapter.
To learn about and get the parts you need to replace, refer to 7 Parts.
5.3 Software Version Check
In troubleshooting that involves software version compatibility, you need to know the monitor configuration and
software version information. For detailed information about version compatibility, contact our after-sales service
department. To check the software version, follow this procedure:
1. Select
Main Menu→System >>→Version >>. In the displayed menu, you can check the system software version
and monitor configuration.
BeneVision N1 Patient Monitor Service Manual 5-1
Page 44
2. Select Main Menu→Maintenance>>→ enter the user maintenance password →Version >>. In the displayed
menu, you can check the version information of the system software and modules.
5.4 Technical Alarm Check
Before troubleshooting the monitor, check whether the monitor displays technical alarm information. If yes, eliminate
the technical alarm first.
For technical alarm information, causes and solutions, refer to the Operator's Manual of the monitor.
5.5 Troubleshooting Guide
5.5.1 Power-On/Off Failures
Failure Description Possible Cause Troubleshooting
You cannot turn on the
monitor powered only
by the battery.
You cannot turn on the
monitor connected to
the AC adapter (when
the battery is not
available or the battery
is low or faulty).
The battery is not activated. Use the AC adapter or Dock to supply external power to activate
the battery.
The battery is low or the
battery is damaged.
The AC power supply is not
connected.
The AC adapter is faulty. If the monitor can work properly when being connected to the
The main board power supply
is faulty.
The power button PCBA is
faulty.
The main board display circuit
is faulty.
Use the AC adapter or Dock to supply external power to the
monitor. Then turn on the monitor to check the battery level and
battery status. If the battery LED flashes yellow or the system
generates the battery error alarm, the battery is damaged. In this
case, replace the battery.
Verify that the AC power supply is properly connected.
main monitor or the Dock, replace the AC adapter.
If the external power LED is off, replace the main board.
If you can turn on the monitor after connecting it to the
powered-on main monitor, replace the power button PCBA.
If the power LED is on but the display is off, replace the main
board.
5-2 BeneVision N1 Patient Monitor Service Manual
Page 45
5.5.2 Display Failures
Failure Description Possible Cause Troubleshooting
The monitor displays a
black, white, or flower
screen or displays
horizontal or vertical
lines.
The touchscreen does
not respond.
The touchscreen fails to
respond accurately.
The screen cable is loose. Check whether the screen cable is connected properly. If no,
reconnect it.
The display assembly is faulty. Replace the display assembly.
The main board is faulty. Replace the main board.
The screen cable is loose. Check whether the screen cable is connected properly. If no,
reconnect it.
The display assembly is faulty. Replace the display assembly.
The main board is faulty. Replace the main board.
The touchscreen is faulty. Replace the display assembly.
5.5.3 Alarm Failures
Failure Description Possible Cause Troubleshooting
The alarm lamp is off or
cannot be turned off
while the monitor
sounds audible alarms.
You cannot hear the
alarm.
The main board is faulty. Replace the main board.
The alarm is muted.
1. Check whether the icon is displayed in the alarm status
icon area. If yes, the alarm is muted.
2. Check whether the icon is displayed in the alarm status
icon area. If yes, the alarm is paused or disabled.
The speaker connection
cable is faulty.
The speaker is faulty. Replace the speaker.
The main board is faulty. Replace the main board.
Verify that the cable connecting the speaker to the main board is
properly connected.
5.5.4 Dock or Modular Rack Failures
Failure Description Possible Cause Troubleshooting
The parameter module
cannot be identified.
The monitor provides no
video output, the USB
device or network
cannot be used, or the
The Rack NIOS board is
faulty.
The RACK connector
board is faulty.
The main board is faulty. Replace the main board.
The Dock interface board
is faulty.
The Dock transfer board is
faulty.
Replace the NIOS board.
Verify that the RACK connector board is securely connected to the
NIOS board, or replace the connector board.
Replace the Dock interface board.
Replace the Dock transfer board.
BeneVision N1 Patient Monitor Service Manual 5-3
Page 46
Failure Description Possible Cause Troubleshooting
monitor cannot be
connected to the main
monitor.
The Dock provides no
power output.
After the monitor is
installed on the Dock,
the external display is
black.
The touchscreen of the
external display cannot
be used.
The main board is faulty. Replace the Dock main board.
The Dock power board is
faulty.
An abnormality occurs
due to hot plug.
The connection sequence
is incorrect.
The USB connection is
abnormal.
Replace the Dock power board.
Reinstall N1.
Connect the display and the Dock first, and then install or start N1.
Reinstall or restart N1.
5.5.5 Operation Failures
Failure Description Possible Cause Troubleshooting
The monitor crashes. The system program is
faulty.
The main board is faulty. Replace the main board.
The monitor is not
identified after being
connected to the main
monitor.
After being connected to
the main monitor, the
monitor is identified but
displays no parameter
waveform.
The time is reset. The battery is faulty. Replace the battery.
The infrared communication
board is faulty.
The main board is faulty. Replace the main board.
The battery charging circuit
is faulty.
The main board is faulty. Replace the main board.
Refresh the system software.
Replace the infrared communication board.
Replace the main board.
5.5.6 Storage Failures
Failure Description Possible Cause Troubleshooting
Historical data and
holographic waveforms
cannot be stored.
N1 reports "Storage Card
Error".
5-4 BeneVision N1 Patient Monitor Service Manual
The data storage device is
faulty.
The main board is faulty. Replace the main board.
The data storage device is
faulty.
Restart N1. If the failure persists, enter the EMMC tab in the
Factory Maintenance menu to format the data storage device
or replace the main board.
Restart N1. If the failure persists, enter the EMMC tab in the
Factory Maintenance menu to format the data storage device
or replace the main board.
Page 47
5.5.7 Power Supply Failures
Failure Description Possible Cause Troubleshooting
The battery cannot be
fully charged.
The battery cannot be
charged or a battery
charging error is
reported.
The battery cannot
supply power.
The system reports that
the DC power is too high
or too low.
The system reports a
battery communication
error.
The system reports that
the battery is aged.
The system reports a
battery error.
The system reports a
high battery
temperature alarm.
The system reports a
battery off alarm.
The battery is damaged. Replace the battery.
The main board is faulty. Replace the main board.
The battery is damaged. Replace the battery.
The main board is faulty. Replace the main board.
The battery is damaged. Replace the battery.
The battery connection cable
is faulty.
The main board is faulty. Replace the main board.
The power supply is faulty. 1. Turn off the monitor and then restart it.
The battery connection cable
is faulty.
The power supply is faulty. Replace the main board.
The battery is aged. Replace the battery.
The battery is faulty. Replace the battery.
The ambient temperature is
too high.
The battery connection cable
is faulty.
Verify that the battery connection cable is properly connected
and the plug is not damaged.
2. If the failure persists, replace the main board.
Verify that the battery connection cable is properly connected
and the plug is not damaged.
Check whether the ambient temperature is too high. If yes,
reduce the ambient temperature.
Verify that the battery connection cable is properly connected
and the plug is not damaged.
NOTE
When the power module is faulty, it may cause damage to other components, for example, the monitor
suddenly crashes during startup, due to power supply protection. In this case, troubleshoot the power
module as per the procedure described in the table above.
Components of the main unit are powered by the power module. When a component malfunctions, verify
the operating voltage is correct.
If the device has two batteries, replace both batteries if required.
BeneVision N1 Patient Monitor Service Manual 5-5
Page 48
5.5.8 Wi-Fi Failures
Failure Description Possible Cause Troubleshooting
Wi-Fi connection is
unavailable.
The Wi-Fi connection is
disconnected frequently.
The monitor fails to
connect Wi-Fi.
The Wi-Fi module is faulty. Reinstall or replace the module.
The main board is faulty. Replace the main board.
The Wi-Fi- antenna falls off or
is not securely connected to
the module.
The Wi-Fi module is faulty. Replace the Wi-Fi module.
Wi-Fi signals in the region are
unstable.
The IP address is incorrect. Check whether an IP address conflict exists in the network. If yes,
Wi-Fi signals in the region are
unstable.
The Wi-Fi antenna falls off or
is disconnected from the
Wi-Fi module.
The main board is faulty. Replace the main board.
Disassemble the monitor and secure the Wi-Fi antenna.
Check the Wi-Fi connection of the hospital.
reset the IP address.
Check the Wi-Fi connection of the hospital.
Secure the Wi-Fi antenna.
5.5.9 Wired Network Failures
Failure Description Possible Cause Troubleshooting
The monitor fails to
connect to a wired
network.
The network connection is
disconnected frequently.
The monitor is connected
to a network, but the
Remote View function is
unavailable.
The network cable fails
to be connected
properly.
The IP address is
incorrect.
The network cable is
faulty.
The main board is
faulty.
The network cable fails
to be connected
properly.
The IP address is
incorrect.
The network cable fails
to be connected
properly.
There are too many
monitors to be viewed.
The network setting is
incorrect.
Make sure that the network cable is properly connected and is not
close to the power cord of high power equipment, and the length of
the network cable is less than or equal to 50 m.
Check whether an IP address conflict exists in the network. If yes,
reset the IP address.
Replace the network cable.
Replace the main board.
Make sure that the network cable is properly connected and is not
close to the power cord of high power equipment, and the length of
the network cable is less than or equal to 50 m.
Check whether an IP address conflict exists in the network. If yes,
reset the IP address.
Make sure that the network cable is properly connected and is not
close to the power cord of high power equipment, and the length of
the network cable is less than or equal to 50 m.
Confirm the maximum number of simultaneously connected
monitors according to the Operator's Manual.
Check whether an IP address conflict exists in the network. If yes,
reset the network.
5-6 BeneVision N1 Patient Monitor Service Manual
Page 49
5.5.10 Software Upgrade Failures
Failure Description Possible Cause Troubleshooting
The program cannot be
upgraded.
The connection is
faulty.
The program upgrade
package is incorrect.
The IP address of the
PC is incorrect.
The main board is
faulty.
1. Check that the network cable is connected to the network
interface on the monitor.
2. Ensure the normal operation of the network hub or switch, and
verify the hub cable or crossover cable is properly connected.
Ensure that the upgrade package is a .pkg file corresponding to the
program to be upgraded.
Configure a fixed IP address in range C as specified for the monitor.
You are not advised to upgrade a program t to when the monitor is
connected to a network with multiple PCs.
Replace the main board.
BeneVision N1 Patient Monitor Service Manual 5-7
Page 50
FOR YOUR NOTES
5-8 BeneVision N1 Patient Monitor Service Manual
Page 51
6 Repair and Disassembly
6.1 Tools
During disassembly and part replacement, the following tools may be required:
Small Phillips screwdriver (used to remove and install M2 screws; the head model is PH0)
Phillips screwdriver (used to remove and install M3 screws; when the N1 main unit face plate needs to be removed,
the rod length of the screwdriver must be greater than or equal to 115 mm)
Tweezers
Terminal screw dismounting tool (the 898-000417-00 N1 terminal assembly fixture is recommended)
Needle nose pliers
Hexagon spanner (used to remove M6 hex socket screws)
6.2 Preparations for Disassembly
Before disassembling the monitor, stop monitoring the patient, turn off the monitor, and disconnect all the accessories
and peripheral devices.
WARNING
Before disassembly, be sure to eliminate static electricity. When removing some parts with the electrostatic
sensitive marks, wear protective devices such as antistatic wristband or gloves to avoid damaging the parts.
Properly connect and route the cables and wires when reassembling the monitor to avoid short circuits.
Use specified screws to reassemble the monitor. If unfit screws are tightened by force, the monitor may be
damaged and the screws or part may fall off during use, causing unpredictable equipment damage or
human injury.
Follow the correct sequence to disassemble the monitor.
Place the removed parts by category to facilitate reassembly. Do not drop, contaminate or lose them.
To reassemble the monitor, first install the assemblies, and then the main unit. Carefully route the cables.
Make sure that the waterproof materials are properly applied during reassembly.
BeneVision N1 Patient Monitor Service Manual 6-1
Page 52
6.3 Disassembling the N1 Main Unit with the CO
6.3.1 Removing the Battery Door and Battery
1. Remove the screw cover and loosen the screw.
2. Hold the screw with your hand and open the battery door.
3. Pull out the battery connection cable and remove the battery.
NOTE
When disassembling the main unit, be sure to pull out the battery connection cable herein and remove the
battery to avoid electricity in subsequent disassembly steps.
Module
2
When reinstalling the battery, place the battery connection cable into the left B1 socket as shown below.
The battery connection cable is routed as follows. When inserting the connection cable into the socket,
align the connector to avoid damage to connector pins.
6-2 BeneVision N1 Patient Monitor Service Manual
Page 53
6.3.2 Installing the Battery and Battery Door
1. Connect the battery to the B1 socket on the main unit through the battery connection cable.
2. Insert the battery into the battery housing with the cable facing up.
3. Route the battery connection cable. Do not squeeze the cable.
Connect battery to B2 socket
3. Secure the battery door using a M3 screw.
4. Use a screw cover to cover the screw hole.
Battery door
Route the cable.
This side up
M3 screw
Screw cover
BeneVision N1 Patient Monitor Service Manual 6-3
Page 54
6.3.3 Removing the Display Assembly
1. Use the small Phillips screwdriver to remove the two screen pins fixing the display assembly.
2. Use one hand to support the display assembly, and use the screwdriver to press the hole in the battery housing and
push out the display assembly, as shown below.
NOTE
Use one hand to support the display assembly as shown below. Otherwise, the FPC plug wire of the display
assembly will be damaged.
3. Use your hand to hold the display assembly after it is pushed out.
4. Use one hand to hold the display assembly as shown below, and the other to loosen the FPC plug wire on the
display assembly, and remove the display assembly.
6-4 BeneVision N1 Patient Monitor Service Manual
Page 55
NOTE
If the light pipe of theindicator is released after the display assembly is removed, install the light pipe at
the position shown in the figure below.
NOTE
When assembling the display assembly, press the two corners shown in the figure below, and press the
display assembly into the rear housing.
6.3.4 Removing the Rear Housing
1. Use the terminal screw dismounting tool to remove the two terminal screws fixing the rear housing.
2. Remove the four M3×5 screws fixing the rear housing.
BeneVision N1 Patient Monitor Service Manual 6-5
Page 56
NOTE
To ensure flat assembly of the face plate, when fixing the four screws on the rear housing, do as follows:
lay the N1 main unit flatwise, and turn the four screws by 2-3 circles for preliminary fixing; place the N1
main unit vertically, and use one hand to press the main unit and the other to tighten the screws.
NOTE
After assembly, check that the silicon cover shown in the figure below is assembled properly.
3. Use your hand to press the battery housing to separate the main frame, and then remove the rear housing.
6-6 BeneVision N1 Patient Monitor Service Manual
Page 57
6.3.5 Removing the Infrared Backboard
1. Loosen the two M3×5 screws fixing the infrared backboard.
2. Remove the infrared backboard.
3. Tilt the N1 main unit as shown below, and remove the two terminal nuts shown in the figure below.
6.3.6 Removing the Main Board
1. Remove the four M3×5 screws fixing the main board.
BeneVision N1 Patient Monitor Service Manual 6-7
Page 58
2. Remove the DC-in connection cable.
NOTE
Check that the waterproof jacket and connection cable are installed properly when reassembling the DC-in
connection cable.
3. Loosen the Wi-Fi antenna (when the N1 main unit is configured with the Wi-Fi function).
6-8 BeneVision N1 Patient Monitor Service Manual
Page 59
NOTE
To prevent the board from crushing the Wi-Fi antenna, route the Wi-Fi antenna according to the direction
shown in the figure when reassembling the Wi-Fi antenna.
4. Loosen the connection cable of the CO
module.
2
5. Loosen the connection cable of the power switch on the main unit.
BeneVision N1 Patient Monitor Service Manual 6-9
Page 60
6. Loosen the connection cable of the speaker.
7. Loosen the connection cable of the analog output (when the N1 main unit is configured with analog output), and
remove the main board.
6.3.7 Removing the Parameter Board
1. Loosen the four M3×5 screws fixing the main board.
2. Remove the block fixing the parameter board.
6-10 BeneVision N1 Patient Monitor Service Manual
Page 61
3. Use one hand to loosen the latch and the other to take the parameter board.
4. Loosen the NIBP pump connection cable and NIBP valve connection cable.
5. Loosen the NIBP air tubing on the parameter board, and remove the parameter board.
BeneVision N1 Patient Monitor Service Manual 6-11
Page 62
6.3.8 Removing the NIBP Gas Circuit and NIBP Valve
1. Loosen the connection between the NIBP gas circuit and interface board.
2. Loosen the connection between the NIBP gas circuit and integrated pump pipe.
3. Loosen the two M3×5 screws fixing the valve frame.
4. Remove the NIBP valve.
NOTE
When installing the NIBP valve, place the cables upwards with the quick release valve of the red cable on
the left and the slow release valve of the blue cable on the right. Lead the NIBP pump connection cable
through under the valve plate.
6-12 BeneVision N1 Patient Monitor Service Manual
Page 63
6.3.9 Removing the NIBP Pump
1. Loosen the exhaust pipe of the CO2 module.
2. Loosen the air inlet pipe of the CO2 module.
3. Loosen the zero calibration pipe of the CO
module.
2
BeneVision N1 Patient Monitor Service Manual 6-13
Page 64
NOTE
After reassembling the exhaust pipe, air inlet pipe and zero calibration pipe, clamp them in the tie mount
according to the sequence shown in the figure, and arrange them in order.
4. Loosen the four M3×5 screws fixing the CO2 module sheet.
5. Remove the CO
module sheet.
2
6. Remove the NIBP pump.
6-14 BeneVision N1 Patient Monitor Service Manual
Page 65
6.3.10 Removing the CO
1. Loosen the mushroom head with a silicone cover on the CO2 module.
2. Loosen the silicone cover head on the CO2 module.
3. Loosen the connection cable between the CO
4. Loosen the slot position of the connection cable.
Module
2
module and water trap.
2
5. Use a pair of tweezers to remove the two CO2 module rings.
BeneVision N1 Patient Monitor Service Manual 6-15
Page 66
NOTE
When reinstalling the module rings, keep the ring side with a gap upward, and clamp the inverted fastener
of the ring into the hole in the silicone cover.
6. Remove the CO
module.
2
6.3.11 Removing the CO
Module Assembly
2
1. Loosen the three M2.5×4 countersunk head screws on the cover plate of the CO2 module.
2. Remove the cover plate of the CO
module.
2
3. Remove the end cap of the CO
module gas circuit.
2
6-16 BeneVision N1 Patient Monitor Service Manual
Page 67
4. Press the latch of the CO2 module connection cable, and remove the connection cable.
5. Remove the zero calibration pipe, short circuit pipe and exhaust pipe on the CO2 module.
6. Remove the silicone cover.
6.3.12 Removing the Interface Board
1. Loosen the three M3×5 screws fixing the main frame.
2. Remove the main frame.
3. Loosen the M3×5 countersunk head screw fixing the power button.
BeneVision N1 Patient Monitor Service Manual 6-17
Page 68
4. Remove the power keypad.
5. Use one hand to loosen the latch fixing the CO
water trap and the other to remove the CO2 water trap.
2
6.4 Disassembling the N1 Main Unit with Two Batteries
6.4.1 Removing the Battery Door and Battery
1. Remove the screw cover and loosen the screw.
2. Remove the screw, and then remove the battery door.
3. Remove the battery from the primary battery housing, and remove the battery connection cable.
NOTE
When disassembling the main unit, be sure to pull out the battery connection cable herein and remove the
battery to avoid electricity in subsequent disassembly steps.
When reinstalling the battery, install the battery connection cable of the primary battery housing into the
left B1 socket. Route the battery connection cable as shown below. When inserting the connection cable
into the socket, align the connector to avoid damage to connector pins.
6-18 BeneVision N1 Patient Monitor Service Manual
Page 69
4. Remove the battery from the secondary battery housing. Remove the battery connection cable.
NOTE
When disassembling the main unit, be sure to pull out the battery connection cable herein and remove the
battery to avoid electricity in subsequent disassembly steps.
When reinstalling the battery, install the battery connection cable of the secondary battery housing into
the right B2 socket. Route the battery connection cable as shown below. When inserting the connection
cable into the socket, align the connector to avoid damage to connector pins.
BeneVision N1 Patient Monitor Service Manual 6-19
Page 70
6.4.2 Installing the Battery and Battery Door
1. Installing the secondary battery.
a. Connect the battery to the B2 socket on the main unit through the battery connection cable.
b. Insert the battery into the secondary battery housing with the cable facing up.
c. Route the battery connection cable.
Connect battery to B2 socket.
Route cable as indicated
2. Installing the primary battery.
a. Connect the battery to the B1 socket on the main unit through the battery connection cable.
b. Insert the battery into the primary battery housing with the cable facing up.
Insert the battery.
This side up.
c. Route the battery connection cable. Do not squeeze the cable.
Connect battery to B2 socket
Route the cable.
3. Secure the battery door using a M3 screw.
Insert the battery.
This side up
6-20 BeneVision N1 Patient Monitor Service Manual
Page 71
4. Use a screw cover to cover the screw hole.
Battery door
M3 screw
Screw cover
6.4.3 Removing the Display Assembly
1.Use the small Phillips screwdriver to remove the two screen pins fixing the display assembly.
2. Use one hand to support the display assembly, and use the screwdriver to press the hole in the battery housing and
push out the display assembly, as shown below.
NOTE
Use one hand to support the display assembly as shown below. Otherwise, the FPC plug wire of the display
assembly will be damaged.
BeneVision N1 Patient Monitor Service Manual 6-21
Page 72
3. Use your hand to hold the display assembly after it is pushed out.
4. Use one hand to hold the display assembly as shown below, and the other to loosen the FPC plug wire on the
display assembly, and remove the display assembly.
NOTE
If the light pipe of theindicator is released after the display assembly is removed, install the light pipe at
the position shown in the figure below.
NOTE
When assembling the display assembly, press the two corners shown in the figure below, and press the
display assembly into the rear housing.
6-22 BeneVision N1 Patient Monitor Service Manual
Page 73
6.4.4 Removing the Rear Housing
1. Use the terminal screw dismounting tool to remove the two terminal screws fixing the rear housing.
2. Remove the four M3×5 screws fixing the rear housing.
NOTE
To ensure flat assembly of the face plate, when fixing the four screws on the rear housing, do as follows:
lay the N1 main unit flatwise, and turn the four screws by 2-3 circles for preliminary fixing; place the N1
main unit vertically, and use one hand to press the main unit and the other to tighten the screws.
BeneVision N1 Patient Monitor Service Manual 6-23
Page 74
NOTE
After assembly, check that the silicon cover shown in the figure below is assembled properly.
3. Use your hand to press the battery housing to separate the main frame, and then remove the rear housing.
6.4.5 Removing the Infrared Backboard
1. Loosen the two M3×5 screws fixing the infrared backboard.
2. Remove the infrared backboard.
6-24 BeneVision N1 Patient Monitor Service Manual
Page 75
3. Tilt the N1 main unit as shown below, and remove the two terminal nuts shown in the figure below.
6.4.6 Removing the Main Board
1. Remove the four M3×5 screws fixing the main board.
2. Remove the DC-in connection cable.
NOTE
Check that the waterproof jacket and connection cable are installed properly when reassembling the DC-in
connection cable.
BeneVision N1 Patient Monitor Service Manual 6-25
Page 76
3. Loosen the Wi-Fi antenna (when the N1 main unit is configured with the Wi-Fi function).
NOTE
To prevent the board from crushing the Wi-Fi antenna, route the Wi-Fi antenna according to the direction
shown in the figure when reassembling the Wi-Fi antenna.
4. Loosen the connection cable of the power switch on the main unit.
6-26 BeneVision N1 Patient Monitor Service Manual
Page 77
5. Loosen the connection cable of the speaker.
6. Loosen the connection cable of the analog output (when the N1 main unit is configured with analog output), and
remove the main board.
6.4.7 Removing the Parameter Board
1. Loosen the four M3×5 screws fixing the main board.
2. Remove the block fixing the parameter board.
BeneVision N1 Patient Monitor Service Manual 6-27
Page 78
3. Use one hand to loosen the latch and the other to take the parameter board.
4. Loosen the connection between the NIBP gas circuit and parameter board U26 sensor.
5. Loosen the connection between the NIBP gas circuit and parameter board U23 sensor.
6. Loosen the NIBP pump connection cable and NIBP valve connection cable, and remove the parameter board.
6-28 BeneVision N1 Patient Monitor Service Manual
Page 79
6.4.8 Removing the NIBP Gas Circuit and Secondary Battery housing
1. Loosen the connection between the NIBP gas circuit and interface board.
2. Loosen the connection between the NIBP gas circuit and pump.
NOTE
During reassembly, the NIBP gas circuit should be connected to the outermost gas outlet.
3. Loosen the three M3×5 screws fixing the secondary battery housing.
4. Use your hand to push the secondary battery housing to remove it from the latch.
NOTE
After reassembling the secondary battery housing, confirm that the pump connection cable and valve
connection cable have been placed in the latch.
BeneVision N1 Patient Monitor Service Manual 6-29
Page 80
5. Use a pair of tweezers to press the pump head, and remove the secondary battery housing.
6.4.9 Removing the NIBP Pump and Valve
1. Remove the NIBP pump.
2. Loosen the two clamping positions of the NIBP valve cable, and remove the valve.
NOTE
When reassembling the valve, place the quick release valve of the red cable on the left and the slow release
valve of the blue cable on the right. See the figure below for the specific placement positions.
6-30 BeneVision N1 Patient Monitor Service Manual
Page 81
6.4.10 Removing the Interface Board
1. Loosen the three M3×5 screws fixing the main frame.
2. Remove the main frame.
3. Loosen the M3×5 countersunk head screw fixing the power button.
4. Remove the power keypad.
BeneVision N1 Patient Monitor Service Manual 6-31
Page 82
6.5 Removing the N1 Dock
6.5.1 Removing the Bottom Housing
1. Remove the four M3×8 screws from the bottom housing.
2. Remove the bottom housing.
6.5.2 Removing the Right Shell Assembly
1. Remove the four screw covers from the right shell.
2. Remove the four M3×8 screws from the right shell.
3. Remove the right shell module.
6-32 BeneVision N1 Patient Monitor Service Manual
Page 83
6.5.3 Removing the Rotatable Plate
1. Remove the four M3×8 screws with pad from the rotatable plate.
2. Remove the rotatable plate.
6.5.4 Removing the Transition Board
1. Remove the six M3×6 screws from the transition board.
2. Remove the transition board plate.
3. Remove the transition board, and loosen the connection cable.
NOTE
During reassembly, route the connection cable as shown below. Do not press any cables.
BeneVision N1 Patient Monitor Service Manual 6-33
Page 84
6.5.5 Removing the Port Module
1. Remove the two M3×6 screws from the port module.
2. Remove the port module, and loosen the indicator connection cable.
NOTE
During reassembly, the sheet metal of the port module is clamped on the latch as shown below. Be sure to
properly assemble the silicone cover of the network cable connector.
6-34 BeneVision N1 Patient Monitor Service Manual
Page 85
6.5.6 Removing the Interface Board and Power Board
1. Remove the plug wire from the interface board.
2. Loosen the two M3×6 screws from the interface board.
3. Remove the interface board.
4. Remove the four M3×6 screws from the power board.
5. Remove the power board.
6. Loosen the latch on the power outlet, and remove the power connection cable.
BeneVision N1 Patient Monitor Service Manual 6-35
Page 86
6.5.7 Removing the LED Indicator
1. Remove the LED bracket.
2. Remove the LED indicator.
NOTE
During reassembly, the indicator with the yellow cable is installed on the left, and the indicator with the
red cable is installed on the right.
6.6 Removing the N1 Modular Rack
6.6.1 Removing the Rear Housing
1. Remove the five screw covers from the rear housing.
2. Loosen the five M3×8 screws on the rear housing.
3. Remove the rear housing.
NOTE
After reinstalling the rear housing, check that the unlocking button works well.
6-36 BeneVision N1 Patient Monitor Service Manual
Page 87
6.6.2 Removing the Unlock Rod
1. Loosen the M2.5×8 screw on the unlock rod.
2. Remove the unlock rod.
3. Remove the unlocking button and spring.
6.6.3 Removing the Infrared Backboard
1. Loosen the connection cable on the infrared backboard.
2. Loosen the M3×8 screw on the infrared backboard.
3. Remove the infrared backboard.
BeneVision N1 Patient Monitor Service Manual 6-37
Page 88
6.6.4 Removing the Transition Board
1. Loosen the two PT3×10 self-tapping screws on the transition board.
2. Remove the transition board assembly.
3. Loosen the two M2×8 screws fixing the transition board.
4. Remove the waterproof jacket from the transition board socket.
5. Remove the transition board.
6.6.5 Removing the Contact Screws
1. Loosen the M3 self-locking nuts from the contact screws, and then remove the contact screws, contact springs and
spring plates.
6-38 BeneVision N1 Patient Monitor Service Manual
Page 89
6.7 Removing N1 Transport Dock
6.7.1 Removing the Transition Board
1. Loosen and remove the four M3×5 countersunk head screws from the bottom of bottom plate.
2. Loosen and remove the two M3×5 countersunk head screws from the left side of bottom plate.
3. Remove the bottom plate.
6.7.2 Removing the Power Adapter
1. Remove the power cable from the power adapter.
2. Loosen the plug on the power adapter.
BeneVision N1 Patient Monitor Service Manual 6-39
Page 90
3. Remove the power adapter.
NOTE
Before reassembling the adapter, use a ruler to measure the position or refer to the position of the original
adapter, and attach a piece of black foam to the cable. In addition, attach a shockproof pad to the magnet ring.
66±1 mm
NOTE
When placing the cable into the Transport Dock, align the left edge of the foam with the position shown in
the figure below.
6-40 BeneVision N1 Patient Monitor Service Manual
Page 91
7 Parts
7.1 Introduction
This chapter lists the exploded views of the main unit and all the parts and part numbers of the repairable parts. It helps
the maintenance personnel to identify the parts during disassembly and part replacement. The part numbers listed in
the "Part Number" columns of all parts lists are used only to assist the maintenance personnel in searching for the
corresponding FRU part numbers. Please provide the FRU parts numbers if you want to purchase the spare parts.
The figure below shows the hardware architecture of the N1 main unit:
NOTE
The part numbers listed in the "Part Number" columns of all parts lists are used only to assist the
maintenance personnel in searching for the corresponding FRU part numbers. Please provide the FRU parts