7.4.2 Parts List .................................................................................................................... 7-4
2
1 Safety
1.1 Safety Information
WARNING
Indicates a potential hazard or unsafe practice that, if not avoided, will result in
death, serious injury, or property damage.
CAUTION
Indicates a potential hazard or unsafe practice that, if not avoided, could result in
minor personal injury, product fault, or product/property damage.
NOTE
Provides application tips or other useful information to ensure that you get the
most from your product.
1.1.1 Warnings
WARNING
BP10 is intended to be used for a single patient at a time.
BP10 must be operated by medical personnel in hospitals or medical institutions.
To avoid explosion hazard, do not use the equipment in the presence of oxygen-rich
atmospheres, flammable anesthetics, or other flammable agents.
Do not use this equipment in conjunction with Electro Surgical Unit (ESU).
Before putting the system into operation, the operator must verify that the
equipment and accessories are in correct working order and operating condition.
Do not come into contact with the patient during defibrillation. Otherwise serious
injury or death could result.
Do not touch the patient and live parts simultaneously.
1-1
Do not open the equipment housing s. All servicing and future upgrades must be
carried out by trained and authorized personnel.
The physiological data displayed on the system are for reference only and cannot
be directly used for diagnostic interpretation.
Only use parts and accessories specified in this manual.
Route, wrap and secure the hose to avoid inadvertent disconnection, stumbling and
entanglement.
1.1.2 Cautions
CAUTION
Do not let BP10 directly touch the patient’s skin when the device is on. The device
temperature rises when the device is on. If the device contacts the patient’s skin for
a long time, skin burns may occur.
When disposing of the packaging material, be sure to observe the applicable waste
control regulations and keep it out of children’s reach.
Magnetic and electrical fields are capable of interfering with the proper
performance of the equipment. For this reason make sure that all external
equipment operated in the vicinity of the equipment comply with the relevant EMC
requirements. Mobile phone, X-ray equipment or MRI equipment are a possible
source of interference as they may emit higher levels of electromagnetic radiation.
Always install or carry the equipment properly to avoid damage caused by drop,
impact, strong vibration or other mechanical force.
Dry the equipment immediately in case of rain or water spray.
The system generates and uses the Radio Frequency (RF) energy. If it is not
installed correctly or not used as per the manual, RF interference to other
equipment could result.
At the end of its service life, the equipment, and its accessories, must be disposed of
in compliance with the guidelines regulating the disposal of such products to
prevent bringing potential negative consequences to the environment and human
health. If you have any questions concerning disposal of the equipment, please
contact Mindray.
1-2
1.1.3 Notes
NOTE
Place the equipment in a location where you can easily see the screen, and access
the operating controls.
Keep this manual in the vicinity of the equipment so that it can be obtained
conveniently when needed.
1.2 Equipment Symbols
See the BP10 Operator’s Manual(P/N: 046-011008-00) for information about the symbols
used on this product and its packaging.
1-3
FOR YOUR NOTES
1-4
2 Design Principles
2.1 Intended Use
BP10 NIBP module (hereinafter referred as BP10) can measure, display, review, store the
NIBP parameter for ambulating Adult and Pediatric patients over three years old, and transfer
the information to TMS60/TM80 in BeneVision Central Monitoring System.
2.2 System Connections
The figure below shows the connection system of the BP10 and its peripherals. The BP10 can
communicate with the TD60/TM80 through Bluetooth and upload the NIBP measurement
result to the telemetry box.
2-1
2.3 Peripheral Interfaces
The interfaces included in the BP10 are shown as follows:
USB interface: used to upgrade software.
Battery interface: used for power supply of BP10 and supporting the rechargeable
battery and the 2-AA battery holder.
Battery interface
USB connector
2-2
2.4 Compositions of the Main Unit
The BP10 host hardware consists of the parameter board and button board. Figure 5 shows
division of the functional block diagram of the two boards. Main functions of each board are
as follows:
M0+ parameter board: Side button detection, main pressure measurement, power
management, on/off control, pump valve control, communication with M4, and system
status monitoring.
M4 main control board: Overpressure protection, front button detection, buzzer driver,
real-time clock, bluetooth communication, USB communication, EEPROM data storage
and LCD driver.
UART
GPIO
2-3
FOR YOUR NOTES
2-4
3 Equipment Installation
3.1 Unpacking
Before unpacking, examine the packing case carefully for signs of damage. If any damage is
detected, contact the carrier. If the packing case is intact, open the packing case in a right way,
take out the BP10 NIBP Module and other components from the packing case, and check
according to the Packing List.
Check whether there is any mechanical damage on the BP10 NIBP Module and whether all
items in the Packing List have been received. Contact us in case of any problem.
When the equipment is moved from one place to another, condensation may occur because of
temperature or humidity difference. In this case, never start the equipment before the
condensation disappears.
3.2 Preparations for Installation
3.2.1 Environmental Requirements
To avoid explosion hazard, do not use the equipment in the presence of flammable
anesthetics, vapors or liquids. The monitor shall be placed in an environment free from
vibration, dust, and corrosive. Otherwise, proper operation of the monitor cannot be
This BP10 NIBP Module is powered by two AA (LR6) alkaline batteries or Li-ion
rechargeable batteries (022-000198-00).
The rechargeable battery needs to be charged using the central charging station, as shown
below.
3.3 Installing the BP10 NIBP module
To install the NIBP module, do as follows:
1. Check the equipment and accessories according to the Packing List.
2. Check whether the equipment and accessories are damaged.
3. Install batteries. For operation details, see the BP10 Operator’s Manual (P/N:
046-011008-00).
4. Connect all accessories.
3.4 Startup
When the rechargeable lithium-ion or AA battery is properly installed into the compartment
door, BP10 will be powered on automatically.
When BP10 is powered off, press the
displays and a beep sounds.
Upon powering up, there are two situations:
If BP10 is turned on for the first time, the device will request you to set up language,
maintenance passcode, service passcode for the first time startup.
If BP10 is turned on next time, the device will directly go to the main screen.
key to turn on BP10. The startup logo screen
3-2
4 Test and Maintenance
4.1 Introduction
To ensure that the BP10 NIBP module will operate optimally, maintenance personnel must
conduct regular inspection, preventive maintenance, and test on the BP10 NIBP module. This
section describes the basic method of testing the BP10 NIBP module and recommends
appropriate test frequency and test tools for you. Maintenance personnel shall select proper
tools to conduct maintenance and test on the BP10 NIBP module as required.
The test method described in this section is mainly used to verify that the performance of the
BP10 NIBP module meets specifications. If the monitor fails to pass the test, the BP10 NIBP
module or a certain functional module of the BP10 NIBP module is faulty. In this case,
maintenance personnel shall repair or replace the faulty BP10 NIBP module. Contact the
Service Department of Mindray in case of any problem.
When necessary, maintenance personnel can ask for the circuit diagram, component list,
operating manual, calibration description, and other necessary materials that can facilitate
maintenance of the components allowed to be maintained from the manufacturer.
CAUTION
All test tasks must be executed by qualified professional maintenance personnel.
Configure and change the co ntents in the device maintenance menu with caution
because misoperation may cause data loss.
Before a test, maintenance personnel should ensure the applicability of the test
tools and connection cables. In addition, the maintenance personnel shall be
familiar with the test tools.
4.2 Startup Check
BP10 power-on detection is used to determine whether the device can be started for operation
normally. Each time when the device is powered on, the BP10 will perform detection
automatically. If no error message is displayed on the screen after powering on, the power-on
self test will pass.
4-1
4.2.1 Battery Check
The BP10 uses the 2-AA battery or rechargeable battery. Battery is detected after being
connected. For the AA battery, the device will check the battery voltage; for the rechargeable
battery, the device will check the power level, voltage and battery status.
CAUTION
Do not use batteries of different types or batteries with different amounts of
electricity at the same time.
WARNING
Use only batteries specified by the manufacturer.
Remove the batteries prior to transport or if the BP10 NIBP Module is not likely to
be used for a long time, avoiding monitor faults caused by liquid leakage of the
batteries.
Replace damaged batteries immediately when you find the damage or leakage. Do
not use faulty batteries to power the BP10 NIBP Module.
4.3 Maintenance Mode
The BP10 supports Maintenance and Service. To access these modes, press the key,
and choose [System] [Maintenance]. Password is required. Two modes support the
following functions:
Maintenance Functions: NIBP Settings, Restore Factory Defaults, Display Auto Off,
manometer (with the precision not lower than 1 mmHg).
The inspection steps are as follows:
1. Connect the BP10, standard manometer, squeeze bulb and rigid vessel as shown in the
following figure.
Monitor
Connector for NIBP
cuff
Squeeze BulbRigid Vessel
2. Before inflation, the reading on the manometer should be 0. If the reading is not 0,
disconnect the airway to make the reading return to 0.
3. Press the BP10
select [NIBP Accuracy Test]. Press Start.
4. Compare the reading of the manometer with the reading of the BP10. The difference
should not be greater than 3 mmHg.
5. Raise the pressure in the rigid vessel to 50 mmHg with the squeeze bulb, and repeat step
4.
6. Raise the pressure in the rigid vessel to 200 mmHg with the squeeze bulb, and repeat step
4.
7. After the pressure test is completed, click [Stop] to exit the pressure test mode.
Main menu key and select [Maintenance], select [Service] and
Appropriate tubing
Standard
sphygmomanometer
NOTE
You can replace the standard manometer/squeeze with a blood pressure simulator
to form a test system.
4-6
q
4.4.2 NIBP Overpressure Calibration
The NIBP overpressure calibration is required at least once every two years or whenever you
doubt the NIBP reading.
Calibration tools:
T-shaped connector
Airway tube
Balloon pump
Rigid vessel (volume 500±25 ml)
Reference manometer (calibrated with accuracy not lower than 1 mmHg)
Calibration procedure:
1. Connect the equipment as shown in the following figure.
BP-10
Connector for NIBP
cuff
ueeze Bulb
S
2. Before inflation, the reading on the manometer should be 0. If the reading is not 0,
disconnect the airway to make the reading return to 0.
3. Press the BP10
select [Overpressure Calibrate]. Press Start.
4. Raise the pressure in the rigid vessel to 320 to 330 mmHg with the squeeze bulb, and
press the [Confirm] button. The interface exits after prompting "Calibration Completed!"
Main menu key and select [Maintenance], select [Service] and
Appropriate tubing
Standard
sphygmomanometer
Rigid Vessel
4-7
4.4.3 Leakage Test
The NIBP leakage test checks the integrity of the system and of the valve. It is required at
least once every two years or whenever you doubt the NIBP reading.
Test tools:
Adult cuff
Inflation tube
Cylinder
Test procedure:
1. Connect the cuff and inflation tube to the NIBP cuff connector of the BP10.
2. Wrap the cuff around the cylinder, as shown in the following figure.
Cylinder
Connector for NIBP
cuff
3. Press the BP10
select [NIBP Leakage Test]. Press Start.
4. The BP10 will automatically deflate in about 20 seconds. This means that the test is
completed. The test result is displayed on the screen. If "Airway is normal" is displayed
on the screen, the system does not have a leakage problem. If the message "NIBP
Pneumatic Leak" is displayed, it indicates that the system may have a leak. Check the
tubing and connections for leakages. If you ensure that the tubing and connections are all
correct, perform a leakage test again.
5. If fault prompt still appears, contact Mindray for maintenance.
Main menu key and select [Maintenance], select [Service] and
Air tubing
Cuff
4.4.4 Hardware Test
Press the BP10 Main menu key and select [Maintenance], select [Service] and select
[Hardware Test]. Press Start.
If “Test Successed!” is displayed continuously for three times on the screen, the system does
not have a hardware problem. If the message “Test Failed!” is displayed, contact Mindray
for maintenance.
4-8
4.4.5 Watchdog Test
Press the BP10 Main menu key and select [Maintenance], select [Service] and select
[Watchdog Test]. Press Start.
If the BP10 restarts in three seconds, the system does not have a watchdog problem. If the
BP10 does not restart or no fault prompt appears, contact Mindray for maintenance.
4.5 Upgrading the System
CAUTION
Before upgrade, disconnect the BP10 from patients and save important data in the
monitor.
During upgrade, do not remove the USB connection. Otherwise, the equipment
may malfunction
The upgrade operation can only be performed by professional maintenance
personnel.
NOTE
Before upgrade, make sure that the upgrade package is the one you desire. To
obtain the latest upgrade package, please contact the After-sales Service
Department of Mindray.
System upgrade procedure:
1. Run the network upgrade tool G-110-003608-00 on the tooling computer, select the
corresponding .tool file of G-110-003613-00, click [OK], and then select the model NIBP
PtxOD.
2. Click [Select Package] on the interface of the upgrade tool, click [Browse] on the
displayed interface, select the combined upgrade package .mpkg file prepared in section
2.3, and click [OK] after confirming that the checksum and version are correct.
4-9
1 Click to select the
upgrade package.
2 Select the combined
upgrade package .mpkg
file prepared in section 2.3
3 Confirm that the checksum
and version are correct
3. Click the [Start] button on the interface of the upgrade tool, and use a Micro USB cable
to connect the BP10 main unit (not installed with the battery) to the tooling computer. Use
two hands to press the [Return] key and [Menu] key of the BP10 main unit at the same
time, and install the battery. The BP10 main unit starts upgrade. After upgrade, the
interface of the BP10 main unit displays "Upgrade Completed!"
4. Remove the battery and then install it again. The machine starts automatically.
4-10
5 Troubleshooting
5.1 Overview
This section classifies faults based on the components where faults occur and fault symptoms.
During troubleshooting, please refer to relevant fault list and check, locate, and rectify faults
in sequence.
The solutions recommended in this section can help to rectify most of the equipment faults
you may encounter. In case of a fault not included in this section, please contact the Mindary
Service Department.
5.2 Replacing a Component
You can replace some main components or parts of this monitor, including the circuit board
of the monitor. After determining that the circuit board of the monitor is faulty, you can
replace the faulty circuit board with a new one according to 6 Maintenance and Disassembly,
and then check whether the fault is rectified. If the fault is rectified, it indicates that the
original circuit board is damaged. In this case, deliver it back to Mindray for maintenance. If
the fault persists, install the original circuit board and continue to perform troubleshooting
based on other possible causes.
5.3 Viewing Software Version
To view the software version, press the BP10 Main menu key and select [Maintenance],
select [Service] and select [Software Version].
5.4 Viewing System Logs
To view the log record, press the BP10 Main menu key and select [Maintenance],
select [Service] and select [Service Log].
5-1
5.5 Error Codes and Corresponding Solutions
See Chapter 12 Troubleshooting in the BP10 Operator’s Manual (P/N: 046-011008-00) for
information about the error codes and corresponding solutions.
NOTE
In case of fault, you must send the faulty BP10 NIBP Module to an authorized
service center for maintenance. Otherwise, the warranty card will become invalid.
5-2
6 Maintenance and Disassembly
6.1 Tools
When disassembly or replacing components, you may need to use the following tools:
Before disassembly the monitor, make the following preparations:
Take ESD protection measures before disassembly. Do not directly touch any boards
during disassembly.
Stop monitoring patients, remove the two batteries from the monitor, and disconnect all
accessories and external devices from the monitor.
CAUTION
Maintenance personnel must take ESD protection measures before disassembling
the monitor. When disassembling some ESD-sensitive components, maintenance
personnel must wear a protection device like ESD wrist strap or ESD glove to
avoid damage on these components.
When reinstalling the monitor, install and place the connection lines properly to
protect them from being damaged, thereby avoiding short circuit.
When reinstalling the monitor, connect the airway components properly to prevent
them from being extruded, thereby avoiding airway blocking.
When reinstalling the monitor, select and use screws of appropriate models. If you
forcibly drive in a screw of inappropriate model, the monitor may be damaged. In
addition, during use after the reinstallation, the component fastened by using this
inappropriate screw may fall off, causing unpredictable product damage or
personal injury.
When disassembling the monitor, dismount various components in proper
sequence. Forcible disassembly in wrong sequence may cause an irreversible
6-1
equipment damage.
Before disassembly components, make sure that all connection lines are removed.
Do not snap the connection lines or damage the connectors during disassembly.
Separately place the disassembled screws and other parts and components based
on their types to facilitate future reinstallation and avoid drop, pollution, or loss.
The monitor has waterproof requirement. When reinstalling the monitor, ensure
that the waterproof accessories like the waterproof strip are properly installed.
6.3 Main Unit Disassembly
NOTE
When disassembling the monitor, ensure that the site for placing the monitor is
smooth and no foreign material that may scratch the anti-dazzling screen or touch
screen or damage the knob exists at the site. In addition, you have to protect the
two button stands at the front of the rear housing during disassembly.
All operations shall be performed by professional maintenance personnel.
Maintenance personnel must wear insulating gloves during the maintenance.
If an optional function is configured, relevant operations may be involved. If this
optional function is not configured, no operation is involved.
6-2
6.3.1 Separating the Front and Rear Half of the Monitor
1. After tearing off the label on the device back, remove the four M1.6×3.5 screws and two
M1.6×13.3 screws, as shown below.
Two M1.6×13.3 screws
Four M1.6×3.5 screws
2. Separate the front housing and rear housing along the device top, and then remove the
FPC between the front housing and rear housing.
6-3
Insert the shorter end of
FPC into the main control
board socket
Insert the longer end of
FPC into the parameter
board socket
pump
NOTE
During re-installation, check whether the flexible cable is installed in place.
6.3.2 Disassembling the Rear Housing
Todisassemble the ABPM peripheral board, do as follows:
1. Place the rear housing of the device flat on a table and remove the connection line on the
parameter board.
2. Remove the five M1.6×3.5 screws and the pump support sheet metal.
3. Pull up to remove the parameter board.
Buzzer line wiring direction
Two M1.6×3.5 screws to fix
support sheet metal
the
Valve line socket
Buzzer line socket
Pump line socket
Bluetooth
antenna wiring
direction
Three M1.6×3.5 screws to
fix the parameter board
6-4
6.3.3 Disassembling Pump, Valve and Integrated Airway
The following figure shows the effect after the parameter board is removed. Next, dismount
the inflation pump:
1. Remove the two M1.6×3.5 screws.
2. Pull up to remove the pump, valve and integrated airway.
The pump line passes
through the gap between
the large valve and the
plastic column, with the
red wire below the black
wire. They cannot pass
through the battery
terminal limiting column.
Two M1.6×3.5 screws
NOTE
During pump assembly, connect the pump connection line according to the
instruction in the figure.
6-5
6.3.4 Disassembling Front Housing Assembly
Remove the grounding copper foil from the figure, as shown below:
Hole alignment
Connect the grounding copper foil to the ground point
Remove the screen FPC and four M1.6×2.4 screws from the main control board
Four M1.6×3.4
cross recessed
cheese head screws
Insert the FPC of LCD
screen into the socket on
the main control board
6-6
7 Components
7.1 Introduction
This section provides theexploded view and part numbers of the monitor to illustrate the
assembling relationships between various components and parts, thereby facilitating
disassembly and replacing components and parts by maintenance personnel.
7.2 Components of the Main Unit
7.2.1 Exploded View
7-1
7.2.2 Parts List
No. P/N Description Remark
1 115 -035519-00 BP10 Front Housing Assembly /
6 115-035520-00 BP10 Rear Housing Assembly /
2 115-035521-00 M1.6×13.3 screws /
5 115-035521-00
BP-10 Maintenance Kit
7 NIBP POD internal signal FPC
(FRU)
3 0156 main unit label (CE) /
4 045-001700-00 BP10 two AA battery tray assembly /
M1.6×3.5 screws
7.3 Front Housing Assembly
7.3.1 Exploded View
7-2
7.3.2 Parts List
No. P/N Description Remark
1 115-035519-00
BP10 Front Housing
2 BP POD dust-proof mat 1
Assembly Kit (FRU)
3 BP POD dust-proof mat 2
4 LCD screen TFT 2.4" 240*320 3.3V
5 BP POD front housing (silk screen)
6 BP POD grounding copper foil
7 BP POD air inlet dustproof sponge
11 BP POD plastic button 2
12 BP POD plastic button 4
13 BP POD plastic button 1
14 BP POD plastic button 3
8 030-000661-00 Screw M1.6 x 3.4 mm /
9 115-035518-00
BP10 Main Control
Board Kit (FRU)
10 BP POD ESD shielding sheet
BP POD lens /
NIBP POD main control board
PCBA
/
7.4 Rear Housing Assembly
7.4.1 Exploded View
7-3
7.4.2 Parts List
No. P/N Description Remark
3
5
7 NIBP connector
115-035520-00
8 BP POD rear housing (silk screen)
BP10 Rear Housing
9 BP POD USB silica gel cap
Assembly Kit (FRU)
10 TP battery fastener spring piece
11 BP POD battery fixing slide fastener
12
1 082-002009-00 Air Pump DC 4.2V 300 mmHg /
2
115-035522-00
BP10 Integrated Gas
13
circuit and valve
assembly Kit (FRU)
14 Solenoid valve (1501)
4
3 BP POD connector fixing rack
115-035521-00
6 BP POD buzzer shock pad
BP10 Maintenance Kit
(FRU)
7 NIBP connector
16 M1.6×3.5 screws
15 051-001707-00 NIBP POD parameter board PCBA /
17 042-012035-00 BP POD pump support sheet metal /
BP POD connector fixing rack
Bluetooth antenna (BP) L=112mm
BP_ANT(BT)
BP POD battery buckle fixing cover
plate
BP POD integrated airway assembly
Air valve, double valve normally
open, DC3.3V 300mmHg
Buzzer Φ13*2.5 80DB terminal
/
/
/
7-4
P/N: 046-011618-00(2.0)
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.