SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. (hereinafter called
Mindray) owns the intellectual property rights to this Mindray product and this manual.
This manual may refer to information protected by copyright or patents and does not
convey any license under the patent rights or copyright of Mindray, or of others.
Mindray intends to maintain the contents of this manual as confidential information.
Disclosure of the information in this manual in any manner whatsoever without the written
permission of Mindray is strictly forbidden.
Release, amendment, reproduction, distribution, rental, adaptation, translation or any
other derivative work of this manual in any manner whatsoever without the written
permission of Mindray is strictly forbidden.
,, , , , BeneView,
WATO, BeneHeart, are the trademarks, registered or otherwise, of Mindray in China
and other countries. All other trademarks that appear in this manual are used only for
informational or editorial purposes. They are the property of their respective owners.
Applicable for
This service manual is applicable for the service engineers, authorized service personnel
and service representatives of this ultrasound system.
Responsibility on the Manufacturer
Party
Mindray is responsible for the effects on safety, reliability and performance of this product,
only if:
All installation operations, expansions, changes, modifications and repairs of this
product are conducted by Mindray authorized personnel;
The electrical installation of the relevant room complies with the applicable national
and local requirements;
The product is used in accordance with the instructions for use.
II
Page 9
Mindray's obligation or liability under this warranty does not include any transportation or
WARNING:
It is important for the hospital or organization that employs
this equipment to carry out a reasonable service/maintenance
plan. Neglect of this may result in machine breakdown or
injury of human health.
other charges or liability for direct, indirect or consequential damages or delay resulting
from the improper use or application of the product or the use of parts or accessories not
approved by Mindray or repairs by people other than Mindray authorized personnel.
This warranty shall not extend to:
Any Mindray product which has been subjected to misuse, negligence or accident;
Any Mindray product from which Mindray's original serial number tag or product
identification markings have been altered or removed;
Any products of any other manufacturers.
Company Contact
Domestic Service Department, Shenzhen Mindray Bio-medical Electronics Co., Ltd.
Mindray Building 22F, Keji 12th Road South, High-tech Industrial Park, Nanshan,
Shenzhen 518057, P.R. China
Hotline: 95105652, 4007005652
Tel: +86 755 26582888
Fax: +86 755 26492815
Statement
This service manual describes the product according to the most complete configuration;
some of the content may not apply to the product you are responsible for. If you have any
questions, please contact Mindray Customer Service Department.
Do not attempt to service this equipment unless this service manual has been consulted
and is understood. Failure to do so may result in personnel injury or product damage.
III
Page 10
Page 11
1 Preface
Signal word
Meaning
DANGER
Indicates death or serious injury may occur imminently in this
hazardous situation if not avoided.
WARNING
Indicates death or serious injury may occur potentially in this
hazardous situation if not avoided.
CAUTION
Indicates minor or moderate injury may occur potentially in this
hazardous situation if not avoided.
NOTE
Indicates property damage may occur potentially in this hazardous
situation if not avoided.
Symbol
Meaning
Location
Type-BF applied part
The ultrasound transducers connected to this system are
type-BF applied parts.
The ECG module connected to this system is Type-BF
applied part.
Above the IO panel
This chapter describes important issues related to safety precautions, as well as the
labels and icons on the ultrasound machine.
1.1 Meaning of Signal Words
In this operator’s manual, the signal words DANGER, WARNING, CAUTION and
NOTE are used regarding safety and other important instructions. The signal words and their
meanings are defined as follows. Please understand their meanings clearly before reading this
manual.
1.2 Meaning of Symbols
The meaning and location of the safety symbols and warning labels on the ultrasound machine
are described in the following tables, please read them carefully before using the system.
1.2.1 Meaning of Safety Symbols
Preface 1-1
Page 12
Caution
On the rear panel
Dangerous voltage
No.
Warning Labels
Meaning
1
Please carefully read this manual
before use device.
2
The following labels are available when the
system works with the mobile trolley.
a. Do not place the device on a sloped
surface. Otherwise the device may
slide, resulting in personal injury or the
device malfunction. Two persons are
required to move the device over a
sloped surface.
b. Do not sit on the device.
c. DO NOT push the device when the
casters are locked.
Symbol
Meaning
Location
Equipotentiality
Power panel
Power button
Upper right corner on the control
panel
Network port
IO panel
USB port
Video output
a
b
c
1.2.2 Warning Labels
1.2.3 General Symbols
This system uses the symbols listed in the following table, and their meanings are explained as
well.
1-2 Preface
Page 13
Remote port
VGA port
VGA signal output
AC indicator
Lower left corner on the control
panel
Battery indicator
Standby indicator
Lower right corner on the control
panel
Hard disk indicator
A
Transducer connector A
Rear panel
B
Transducer connector B
This product is provided with a CE
marking in accordance with the
regulations stated in Council
Directive 93 / 42 / EEC concerning
Medical Devices. The number
adjacent to the CE marking (0123)
is the number of the EU-notified
body certified for meeting the
requirements of the Directive.
DANGER
Do not operate this system in an atmosphere containing
flammable or explosive gases such as anesthetic gases, oxygen,
and hydrogen or explosive fluid such as ethanol because an
explosion may occur.
1.3 Safety Precautions
Please read the following precautions carefully to ensure the safety of the patient and the operator
when using the probes.
Preface 1-3
Page 14
1.3.1 Electric safety
WARNING:
1. Do connect the power plug of this system and power plugs
of the peripherals to wall receptacles that meet the ratings
indicated on the rating nameplate. Using a multifunctional
receptacle may affect the system grounding performance,
and cause the leakage current to exceed safety
requirements.
2.
Do not use any cables other than the cables provided with
the device by Mindray.
3.
Use the cable provided with this system to connect the
printer. Other cables may result in electric shock.
4.
Disconnect the AC power before you clean or uninstall the
ultrasound machine, otherwise, electric shock may result.
5.
Do not use this system simultaneously with equipment such
as an electrosurgical unit, high-frequency therapy
equipment, or a defibrillator, etc.; otherwise electric shock
may result.
6.
This system is not water-proof. If any water is sprayed on or
into the system, electric shock may result.
CAUTION:
1. DO NOT connect or disconnect the system’s power cord or
its accessories (e.g., a printer or a recorder) without turning
OFF the power first. This may damage the system and its
accessories or cause electric shock.
2.
Avoid electromagnetic radiation when perform performance
test on the ultrasound system.
3.
In an electrostatic sensitive environment, don’t touch the
device directly. Please wear electrostatic protecting gloves if
necessary.
4.
You should use the ECG leads provided with the ECG
module. Otherwise it may result in electric shock.
WARNING:
1. When moving the system, you should first power off the
system, fold the LCD display, disconnect the system from
other devices (including probes) and disconnect the system
from the power supply.
2.
Do not subject the transducers to knocks or drops. Use of a
defective probe may cause electric shock to the patient.
1.3.2 Mechanical Safety
1-4 Preface
Page 15
CAUTION:
1.
Do not expose the system to excessive vibration (during the
transportation) to avoid device dropping, collision, or
mechanical damage.
2.
When you place the system on the mobile trolley and move
them together, you must secure all objects on the mobile
trolley to prevent them from falling. Otherwise you should
separate the system from the mobile trolley and move them
individually. When you have to move the system with the
mobile trolley upward or downward the stairs, you must
separate them first and then move them individually.
3.
Do not move the ultrasound system if the HDD indicator is
green, sudden shake may cause the HDD in damage.
NOTE:
1.
The user is not allowed to open the covers and panel of the system, neither
device disassemble is allowed.
2.
To ensure the system performance and safety, only Mindray engineers or
engineers authorized by Mindray can perform maintenance.
3.
Only technical professionals from Mindray or engineers authorized by
Mindray after training can perform maintenance.
NOTE:
For detailed operation and other information about the ultrasound system, please
refer to the operator’s manual.
1.3.3 Personnel Safety
1.3.4 Other
Preface 1-5
Page 16
Page 17
2 Preparations for
Name
Height(mm)
Length(mm)
Width(mm)
Weight(Kg)
Value
378
415
190
8.52 (including battery)
Item
Operating Conditions
Storage & Transportation
Environment
Temperature
0℃~40℃
-20℃~55℃
Humidity
30%~85% (no
condensation)
30%~95% (no
condensation)
Atmospheric pressure
700hPa~1060hPa
700hPa~1060hPa
WARNING:
Do not use this system in the conditions other than those
specified.
NOTE:
Do not install the machine in the following locations:
Locations near heat generators
Locations with high humidity
Locations with flammable gases
Installation
2.1 Intended Use
2.1.1 Product Description:
The DP-50/ DP-50T/DP-50PT/DP-50Vet is Digital Ultrasonic Diagnostic Imaging System.
2.1.2 Dimensions and Weight
2.2 Installation Conditions
2.2.1 System Environment
Preparations for Installation 2-1
Page 18
2.2.2 Electrical Requirements
Voltage
Current
Frequency
100-240V~
1.5-0.8A
50/60Hz±3Hz
WARNING:
DO NOT connect this system to outlets with the same
circuit breakers and fuses that control the current of
devices such as life-support systems. If this system
malfunctions and generates an over-current, or when
there is an instantaneous current at power ON, the circuit
breakers and fuses of the building’s supply circuit may be
tripped.
2.2.2.1 Power Supply Requirement
2.2.2.2 Requirement of Regulated Power Supply
Due to the difference of the power supply stability of different districts, please advise
the user to adopt a regulator of good quality and performance such as an on-line UPS.
2.2.2.3 Grounding Requirements
The power cord of the system is a three-wire cable, the protective grounding terminal
of which is connected with the grounding phase of the power supply. Please ensure that
the grounding protection of the power supply works normally.
2.2.3 EMI Limitation
Ultrasound machines are susceptible to Electromagnetic Interference (EMI) from
radio frequencies, magnetic fields, and transient in the air wiring. They also generate EMI.
Possible EMI sources should be identified before the unit is installed.
These sources include: medical lasers, scanners, monitors, cauterizing guns and so
on. Besides, other devices that may result in high frequency electromagnetic interference
such as mobile phone, radio transceiver and wireless remote control toys are not allowed
to be presented or used in the room. Turn off those devices to make sure the ultrasound
system can work in a normal way.
2.3 Installation Condition
2.3.1 Space Requirements
Place the system with the necessary accessories at a proper position for convenient
use.
1. Place the system in a room with good ventilation or having an air conditioning
unit.
2. Leave at least 20cm clearance around the system to ensure effective cooling.
3. A combination lighting system in the room (dim/bright) is recommended.
4. Except the receptacle dedicated for the ultrasound system, at least 3-4 spare
2-2 Preparations for Installation
Page 19
receptacles on the wall are available for the other medical devices and peripheral
devices.
5. Power outlet and place for any external peripheral are within 2 m of each other
with peripheral within 1 m of the unit to connect cables.
2.3.2 Network Environment
Wired LAN function is supported by this ultrasound system.
Data transmission is allowed between different departments or areas with network
cable. Confirm the network devices and network conditions before the installation:
1. General information: default gateway IP address, and the other routers related
information.
2. DICOM application information: DICOM server name, DICOM port, channels, and
IP address.
2.3.3 Confirmation before Installation
Please confirm the following items before installation:
1. The video format used in the region or country where the system is installed.
2. The language used in the region or country where the system is installed.
3. The power voltage and frequency used in the region or country where the system
is installed.
4. Obstetric formulae and other measurement formulae used in the region or
country where the system is installed.
5. Other settings to be used in the region or country where the system is installed
but different from the factory settings.
6. The doctor’s habits when using the system.
Please confirm the items above prior to the installation training, and do the system
settings according to the universal setting of installed region or country.
Preparations for Installation 2-3
Page 20
Page 21
3 System Installation
3.1 Installation Tools and Duration
Tools Required: None.
Installation duration: 1 person, 0.5 hour.
3.2 Checks after Unpacking
3.2.1 Unpacking
1. Use the scissor to clip the 2 rubber belts as follows:
2. Use a knife to open the tapes:
3. Take out the operation manual:
System Installation 3-1
Page 22
4. Take out the transducer box 1:
5. Take out the accessory box:
6. Take out the transducer box2, as follows (the box is below the accessory box):
7. Hold the handle tightly and pull the machine with the surrounding foam:
3-2 System Installation
Page 23
8. Remove the surrounding protective foam to take out the main system.
3.2.2 Checking
1. After unpacking, check the objects in the container with the package list to see if
anything is in short supply or is wrong.
2. Inspect and make sure there is no damage to the machine, no indentation, no cracks. If
any, please contact Mindray Customer Service Department.Tel: 9510 5652, or
4007005652.
System Installation 3-3
Page 24
3.3 Installation of Main Unit
1
2
3
4 7 5
6
8
3.3.1 Appearance of the System
Rear view:
3-4 System Installation
Page 25
Left side:
No.
Name
Function
1
Probe holder
Sets the transducer
2
Display
Displays the image and parameters
3
Control Panel
Man-machine interface for operation control, for details,
please refer to “3.3.2.1 Control Panel”.
4
Handle
For lifting the machine
5
Transducer
sockets
Sockets connecting transducers and the main unit.
6
IO Panel
Signal input / output panel, for details, please refer to
“3.3.2.2 IO Panel”.
7
Power supply
panel
Signal input / output panel, for details, please refer to
“3.3.2.2 IO Panel”.
8
Battery cover
Covering battery
9
USB ports
For connecting USB device (2 ports)
9
System Installation 3-5
Page 26
3.3.2 Panel Introduction
No.
Name
Description
Function
1
/
Power button
Off: when system is turned off;
Green: when system is turned on by pressing this
button;
Orange: when system is in standby.
2
Esc
Exit
Press to exit the current status to the previous
status.
3
Help
/
Press to open or close the accompanying help
documents.
4
Review
/
Press to review the stored images.
5
Report
/
Press to open or close the diagnosis reports.
6
iStation
/
Press to enter or exit the patient information
management system.
7
F1~F4
User-defined
key
You can assign a function to the key.
8
Biopsy
/
Press to show or hide the biopsy guide line.
9
Setup
/
Press to open/close the setup menu.
10
Del / Press to delete the comment, etc.
11
/
Alphanumeric
keys
Same as on PC
3.3.2.1 Control Panel
3-6 System Installation
Page 27
No.
Name
Description
Function
12
Arrow
/
Press to enter or exit the arrow comment status.
13
Menu
Main menu
Press to display or hide a mode-specific parameter
menu.
14
Cine
Cine Review
Press to enter or exit the Cine Review status.
15
/
Direction key
To adjust display brightness or contrast when
pressing with <Fn> key.
16
TGC
/
Move to adjust time gain compensation.
17
Focus
Freq./THI
/
Press: to switch between Focus and Freq./THI;
Rotate: to adjust corresponding parameter
18
Depth
Zoom
/
Press: to switch between Depth and Zoom;
Rotate: to adjust corresponding parameter
19
Patient
Patient
Information
Press to open/ exit patient information screen.
20
Body Mark
/ Press to enter or exit the Body Mark status.
21
Probe
Probe switch
Press to switch Probe and Exam Type
22
Comment
/
Press to enter or exit the character comment
status.
23
End Exam
/
Press to end an exam.
24
Clear
/
Press to clear the comments or measurement
calipers on the screen.
25
Cursor
/
Press to show the cursor.
26
Nav.Rot
/
Multifunction knob
27
Quad
Quad-split
screen
Press to enter Quad mode from non-Quad mode;
Press to switch between windows in Quad mode.
28
Dual
Dual-split
screen
Press to enter Dual mode from non-Dual mode;
Press to switch between windows in Dual mode.
29
M / Press to enter the M mode.
30
B / Press to enter the B mode
31
Measure
/
Press to enter/ exit Application Measurement
32
Update
/
Measurement status: press to switch between the
fixed and active end of the caliper;
Multi-imaging mode: press to change the currently
active window.
33
Caliper
/
Press to enter/ exit Measurement
34
Gain/
iTouch
/
Rotate: to adjust B or M gain
Press: to enter/ exit iTouch
35
Save 1
/
Press to save, user-defined key
36
Save 2
/
Press to save, user-defined key
System Installation 3-7
Page 28
No.
Name
Description
Function
37
Set / Press to confirm an operation.
38
/
Trackball
Roll the trackball to change the cursor position.
39
Set / Press to confirm an operation.
40
Freeze
/
Press to freeze or unfreeze the image.
41
Print
/
Press to print: user-defined key.
42
/
Indicator 1
AC indicator
AC supply: light green;
Battery supply: light off.
43
/
Indicator 2
Battery status indicator
Charging: light in orange
Full: light in green
Discharge (electricity >20%): light in green
Discharge (electricity <20%): blinking in
orange
Discharge (electricity <5%): blinking in orange
rapidly
Non-charge/ discharge: light off
44
/
Indicator 3
Standby indicator
Standby: blinking in orange
Other status: light off
45
/
Indicator 4
HDD status indicator
Read/ write: blinking in green
Other status: light off
NOTE: DO NOT move the machine when the
indicator blinking in green. Otherwise the HDD may
be damaged by sudden shake.
3.3.2.2 IO Panel
<2><3><4>
<1>
<5><6><7>
VGA
12V
5V
3.3V
A
<8>
The IO panel is on the back of the main system.
3-8 System Installation
Page 29
No.
Symbol
Function
1
Network port
2
USB ports
3
4
Separate video output, connecting video printer or
monitor
5
Remote control port
6
VGA
VGA signal output
7
Composite video output
8
/
Power indicator
3.3.2.3 Power Supply Panel
<1>
<2>
100-240~ 50/60Hz 1.5-0.8A
No.
Name
Function
1
Power inlet
AC power inlet
2
Equipotential
terminal
Used for equipotential connection, that balances the
protective earth potentials between the system and other
electrical equipment.
3.3.3 Installation of Main Unit
3.3.3.1 Installing Battery
As shown in the following figure.
1. Open the battery cover and put the battery into the bay (Note that the battery
shall be put in correct direction).
2. Push the battery into the battery bay ①, and push to the right side as ② until it’s
tightly locked
3. Close the battery cover.
System Installation 3-9
Page 30
Battery
Lock button (right)
Open the
control panel
Figure 3-1 Installing Battery
3.3.3.2 Control Panel
Press the ① lock buttons on the both sides of the control panel as illustrated in the
following picture, open the control panel.
Figure 3-2 Control Panel Adjustment
3.3.3.3 Display Adjusting
1. Open the display as described in 3.3.3.2
2. Put the finger into the bottom of the display and pull the display to tilt the display
(30 degrees max.)
3. Ear off the screen protective film.
3-10 System Installation
Page 31
NOTE:
On the monitor, the brightness adjustment comes before contrast. After
readjusting the monitor’s contrast and brightness, adjust all preset and
peripheral settings.
Put the finger into the
bottom of the display,
and adjust the angle
Figure 3-3 Open and tilt the display
Press <Fn>+ directional keys to adjust brightness and contrast.
Brightness Adjustment:
refer to the brightness control keys; press <Fn>+ to decrease
the brightness; and press <Fn>+ to increase the brightness.
Contrast adjustment:
, refer to the contrast control keys; press <Fn>+ to decrease the
contrast; and press <Fn>+ to increase the contrast.
Restore the factory default settings
Open the page via “Setup (F10)→General” and restore the brightness/contrast to
factory default setting.
3.3.3.4 Installing Probe Holder
Fix the probe holder hanger into the groove in the right side and the rear-right side of
the main unit, and push downwards to confirm the installation.
System Installation 3-11
Page 32
NOTE:
Before inserting the connector into the probe port, inspect the connector pin. If
the pin is bent, do not use the probe until it has been inspected / repaired /
replaced.
Fix the holder
hanger into the
groove, push
downwards
Socket
Transducer
locking lever
Figure 3-4 Installing the Probe Holder
3.3.3.5 Connecting a Transducer
As shown in the following figure.
1. Turn the locking lever of the transducer connector to the horizontal position; keep
the cable downward and insert the connector into the port, and then press in fully.
2. To lock the transducer: turn the locking lever 90° clockwise.
3. To disconnect the transducer, turn the locking lever 90° anticlockwise and pull
the transducer straight out.
Figure 3-5 Transducer connection
3-12 System Installation
Page 33
Name
Model
Footswitch
FS-81-SP-2 (1-pedal, USB port)
971-SWNOM (2-pedal, USB port)
SP-997-350 (3-pedal, USB port)
Black / white video printer
Analog: SONY UP-897MD,MITSUBISHI P93W-z
Digital: SONY UP-D898MD, SONY UP-X898MD
Color video printer
Analog: SONY UP-20, MITSUBISHI CP910E
Digital: HP Photosmart plus B210A
Graph/ text Printer
HP series
Tested: HP Deskjet 1280,HP Laserjet CM1015,
HP officejet 6000,HP officejet all-in-one 3600
External DVD R/W Drive
SE-S224Q
NOTE:
If the ultrasound system can not recognize the SONY UP-X898MD and SONY
UP-D898MD printers automatically, you may need to change the settings on
the printer: push <PUSH ENTER> to enter the main menu and select
[DIGITAL]->[DRIVER], and select [897].
USB port
Footswitch
cable
3.3.4 Installing Peripherals
The system supports the following models of peripherals.
3.3.4.1 Footswitch Installation
Take 971-SWNOM as an example: insert the USB connector to the USB ports in the
rear or left of the machine.
For settings of footswitch, please refer to 3.4.4.1.
Figure 3-6 Footswitch connection
3.3.4.2 Video Printer Installation
Analog video printer:
System Installation 3-13
Page 34
1. Connect one end of the signal line to the Video In interface of the printer, and the
Video signal cable
Remote Port
Data cord
USB port
Power cord
other to the S-Video output port in the ultrasound system IO panel.
2. Connect the Remote control line to the Remote interface in the ultrasound
system IO panel.
3. Insert the power cord to a power supply receptacle that is well grounded.
4. Please refer to 3.4.4.3 for the driver installation.
Figure 3-7 Installing Analog Video Printer
Note: Other analog video printer installation procedures are the same as those of
UP-897MD.
Digital Video Printer
Connect one terminal of the data cable of the video printer to the USB port of the
ultrasound system and the other terminal to the VIDEO IN port of the video printer, then
connect the printer with the power supply. For driver installation, please refer to 3.4.4.3.
3.3.4.3 Installing a Graph / Text Printer
A graph / text printer is configured with power cord and data cable, and is connected
to the system via the USB port. The connection method is described as follows:
1. Connect the data cable to the USB port in the ultrasound system.
2. Plug the power plug into a power supply receptacle that is well grounded.
3. Please refer to 3.4.4.3 for the driver installation.
3-14 System Installation
Page 35
3.3.4.4 Installing External DVD-R/W
Drivers preinstalled in the main board: chip group, video card, sound card and network
adapter.
Video printer driver preinstalled
HP printer driver preinstalled
USB Footswitch
Default time zone to be +8
Power on/off logo customization
Screen saver, hibernation and standby can be forbidden.
Power supply
adapter
USB cable
Take the Samsung SE-S224Q DVR as an example:
1. Connect the USB cable connector of the external DVD recorder to the USB port
in the ultrasound system;
2. Connect the adapter connector of the DVD recorder to an appropriate
receptacle.
Figure 3-8 Installing External DVD-R/W
3.4 System Configuration
3.4.1 Power ON / OFF
Connect the system power cord to the AC power, and make sure the ultrasound system
and other optional devices are correctly connected.
When the AC indicator is green, you can turn on the power button (located at the upper
right corner of the system) to initiate the system. After being normally rebooted, it will
display image interface. Or press the power button directly when the battery is of sufficient
capacity.
3.4.2 System Introduction
This product adopts Linux kernel, and compile the source code into the ultrasound
system.
Table 3-1 Functions defined by the OS
System Installation 3-15
Page 36
3.4.3 Enter Doppler
After the system is powered on after initiation (about 1min), it enters Doppler directly:
3.4.4 Function Setting
3.4.4.1 System Preset
Press the <Setup> key to enter the [Setup] menu.
Region
3-16 System Installation
Page 37
In the Region page, set the system language, date format, date, time and hospital related
information, etc.
Key Config
1) Function of keyboard keys F1-F4 and the footswitch keys (left, mid, right) are
user-defined.
2) Key brightness, key volume, trackball backlight and trackball sensitivity can be
adjusted.
Installation of Optional Devices
1) Copy the optional devices to the U disk, then insert it to the USB port of the
ultrasound system;
2) Enter [Setup], click [Maintenance] to enter [Option] menu to select the optional device
to be installed option module.
System Installation 3-17
Page 38
NOTE:
1.
You can only install one key at one time.
2.
The option un-installation function is only accessible to internal user,
for service engineer, you can log on the system as “Service” first, and
then conduct un-installation function.
NOTE:
All image parameters have been set as preset data.
3) Select the corresponding key file in the prompted “Load File” dialogue box and click
[OK].
4) After the Key file is successfully installed, you can see the option module in the
“Installed” status. Corresponding functions are available after return from preset.
3.4.4.2 Image Preset
Press the <Setup> key to enter the [Setup] menu, and click [Image Preset].
3-18 System Installation
Page 39
3.4.4.3 Print Preset
1. Press <Setup> and click [Print Preset] to set video printer, graph/text printer
parameters:
2. After connecting the local printer, Click “Printer Driver”, the system will display the
printer name and status (Ready) automatically which already installed printer driver
successfully.
3. Return “Printer Service” page, Select the corresponding service from the printer list
and increase the service.
The system integrates drivers of HP printers, after HP printers are connected, drivers
will be installed automatically (about 10s).If auto installation fails, icon will display on the
right lower corner of the screen to warn you that manual installation is necessary. The
driver installation method is as follows:
System Installation 3-19
Page 40
a) Download the ppd file from HP official website (contact R&D engineer if necessary),
NOTE:
1.
Before connect the network printer, make sure the ultrasound system
and the printer are in the same network domain, and the network is
working normally.
2.
The IP address and the server name should be valid, e.g. \\10.2.40.123
or \\5-HP, otherwise, the system will fail to connect.
3.
If the server has set accessing limitation, the system will prompt a
dialogue box to identify the user. Enter the correct user name, domain
name and password, and then click [OK].
and copy the ppd file to the storage device (USB disk as an example).
b) Connect the U disk to the USB port nearside the control panel of ultrasound system,
click the icon to pop up the screen, select the U disk to run the ppd file and finish the
installation.
Add network printer
1. In “Printer Driver” screen, click [Add Network Printer] to pop up the screen, enter the
necessary information (IP address, shared printer name, server name, domain name
and password).
2. After successful connection, the newly added network printer name will be shown in
the printer driver list.
3.4.4.4 Network Preset
Network Preset
Open “[Setup]→[Network Preset]” to enter the screen.
3-20 System Installation
Page 41
a) Please select the network type according the actually status, Select “DHCP”,
Name
Description
Current Network Adapter
To select the network connection mode
DHCP
/ Static
If “DHCP” is selected, IP address will be automatically
obtained from DNS server; if “Static” is selected (using
static IP address), you need to enter the IP address.
IP Address
IP address of the system should be at the same network
segment with the server IP.
Subnet Mask
Used to set different network segment.
Gateway
Used to set the gateway IP.
NOTE:
Only if DICOM basic option is configured, [DICOM Local], [DICOM Server],
[DICOM Service] are available.
click [OK].
b) Or, select “Static”, and input the IP address, subnet mask and gateway, then click
[OK].
DICOM Preset
1. Click [DICOM Preset] to open the DICOM preset screen.
System Installation 3-21
Page 42
NOTE:
1.
AE Title should be the same with the SCU AE Title preset in the server
(PACS/RIS/HIS).
2.
DICOM communication port in the ultrasound system should be the
same with the one in the server.
NOTE:
If the currently entered name has already existed, the system will pop up:
“The server name exists!” Click [OK] to enter another name.
DICOM Setting
Enter the AE Title of the ultrasound system, port and PDU according to the actual
situation, and then click [OK] to exit the screen.
DICOM Server Setting
a) Enter the device name and the IP address.
b) You can ping other machines to verify connection after entering the correct IP
address. Also you can check the connection of the already added server in the
list.
c) If connection is successful, click [Add] to add the service to the Service list.
2. Click [DICOM Service] to open the DICOM Service screen.
3-22 System Installation
Page 43
NOTE:
DICOM Work list can be configured only after DICOM Basic is configured,
and if DICOM Work list function is not configured, the “Work list” page is not
accessible.
Only when the system is configured with DICOM basic function module, and installed
DICOM Work list, MPPS, DICOM Structured Reporting and Query/ Retrieve modules, can
the corresponding preset settings be found in DICOM Service screen.
The DICOM Service Setting is used to set properties of DICOM services as Storage,
Print, Work list, MPPS, Storage Commitment and Query/ Retrieve.
3.4.4.5 Manage Settings
Press <Setup> to open the Setup menu, click [Manage Setting] to open the Manage
Setting screen, where you can import, export or restore the factory data.
System Installation 3-23
Page 44
Code in Figure
Function Description
A
Export the selected preset data, and save them in different files
B
Import the preset parameters, or recover the factory settings
C-
Export all preset parameters, and save them in a file
D
Import all preset data at one time
E
Recover factory settings for all preset data
3.4.4.6 System Information
NOTE:
1.
Be sure to confirm the system information before and after the
software maintenance.
2.
Ask the user to save the current system information if necessary.
In System Information screen, it displays the product configuration, the optional
installation status, software version, hardware & boards, and driver related information.
You can check the product information here.
1. Press the <Setup> and click [System Info].
2. Click [About Detail] to check the detailed board information, and you can see all
option information in the Option page.
3. Click [Save] to export the information in the format of “*.TXT”; click [Exit] to return to
the Setup menu.
3-24 System Installation
Page 45
4 Function Checking and
Testing
4.1 Preparation
4.1.1 Personnel
Function checking and testing of this part shall be carried out by Mindray service
engineers and the user together.
4.1.2 Tool
None.
4.2 Checking System Status
4.2.1 System Running Status
1. The ultrasound system can be turned on or off normally, working normally, no
any abnormal noise and phenomenon appear when the system is running.
2. The fan starts to work as soon as the system is turned on, no any abnormal
noise appear in process
3. Check basic system information such as product configuration, options,
software version, etc. Confirm all the information is normal.
4. Check the preset of the monitor, such as contrast and brightness.
5. Check if the time and date are correct, if not, please preset them again.
6. Check the log together with the user to confirm any abnormality or occasional
abnormalities when the system is running.
4.2.2 System Running Status
Check the ambient temperature and humidity. The measurements related to safety
features are particularly sensitive to humidity, measurement position and measuring
circuit.
If the insulation feature of the system deteriorates due to the increase of system
service time or system malfunctions, the fluctuation range of measurement results are
likely to increase with the increase in humidity.
Function Checking and Testing 4-1
Page 46
4.3 General exam
Control panel
Display
DVD-R/W
Peripheral devices
I/O ports
Procedure
Standard
Check the functions of all keys and knobs
Follow the direction: left to right, and up to down.
All keys and knobs are effective.
Function checking of the trackball
Press the <Freeze> key to enter the Freeze
status.
Press <Measure> to enter into measure
status, do vertical and horizontal
measurement, or do other trackball
operations.
The trackball can be rotated easily
and smoothly, the cursor responds
sensitively, the rotation direction is
the same as the direction of the
cursor.
Procedure
Standard
Adjust LCD brightness
Adjust LCD contrast
Display color temperature
adjustment
Click “Cold” or “Warm” in the
[General] page in System
Press “<Fn>+< >”, the brightness increases;
and press “<Fn>+< >”, the brightness
decreases.
Press “<Fn>+< >”, the contrast increases; and
press “<Fn>+< >”, the contrast decreases.
The LCD color temperature changes
correspondingly.
4.3.1 Check Flow
4.3.2 Checking Content
4.3.2.1 Check Control Panel
4.3.2.2 Check the Display
4-2 Function Checking and Testing
Page 47
Preset screen.
LCD maintenance
Log in as “Service”, click
[Maintenance] in the Setup
menu, then click [Monitor Test]
to go for the monitor test.
Click each functional button, the LCD responds
correctly, the standard is as follows
1 .Light-spot: 0; flash point: 0.
2. The adjoining dark dots are no more than 3 pairs,
and there is no adjoining dark dot in image area.
3.There is no adjoining dark dots of 3 or more than 3
4.The dark dots are no more than 7 and those in the
image area are no more than 2
5. The distance between bad dots is no less than
5mm.
Note: image area refers to the red the background
is black/right.
4.3.2.3 Check DVD-R/W
Procedure
Standard
DVD-R/W Drive
Check if connection is normal
Press <Eject> button on the drive
Perform DVD reading and writing.
The DVD-R/W driver can be ejected
smoothly.
Data reading and writing performs normally
without abnormal noise.
Procedure
Standard
Footswitch:
Connect the footswitch
connection, and check the
functions of footswitch according to
the functions listed in Key Config.
Press the freeze key (the right key), image is
frozen, the freeze menu is displayed; press the key
again, image is unfrozen.
Press the key (middle key), color printing starts
Press the print key (left key),B/W printing starts
Video Printer:
Check if the video printer and
ultrasound system are correctly
connected. Then check the
function of each key.
Press <Print> key, the printer begins to work, no
image print deficiency or degradation.
Switch video output port
Press <Print> key, the printer begins to work, no
image print deficiency or degradation.
Graph/ text Printer:
Check if the graph/text printer
and ultrasound system are
correctly connected. Then check
the function of each key.
Press <Print> key, the printer begins to print, no
print deficiency or degradation.
Procedure
Standard
Checking IO ports:
Video/audio interface, USB
The contents displayed on the VGA/LCD are the
4.3.2.4 Peripheral Device Checking
4.3.2.5 Checking IO Ports
Function Checking and Testing 4-3
Page 48
port have been verified,
remains are:
VGA port
Connect VGA display
Network port
USB port
same as those displayed on the ultrasound system
display, no character and image loss, no color
difference, no fluttering and flicking.
Smooth communication in both network and
connection.
USB port data storage/accessing are normal.
4.4 Function Checks
NOTE:
A complete function inspection is described here, do the checking according to
the actual system configuration.
Check each imaging
mode
Check the measurements
Check in cine mode
Check the probe
application
Image & video manage
Record and save the
exam
4.4.1 Check Flow
4.4.2 Checking Content
4.4.2.1 B Mode
In B Mode scanning, the image parameter area in the right corner of the screen will
display the real-time parameter values as follows:
4-4 Function Checking and Testing
Page 49
Display
F 2.5M
D 21.3
G 100
FR 34
IP 1
DR 65
Image
Parameters
Frequency
Depth
Gain
Frame Rate
B IP
B Dynamic
Range
Parameters that can be adjusted to optimize the B Mode image are indicated in the
The adjusting range is 0-100.The real-time value will be
displayed in the image parameter area in the upper right
corner of the screen.
Depth Adjustment
Press <Depth/zoom> to
light on the Depth
indicator, rotate the knob
Rotate clockwise to increase
Rotate anticlockwise to decrease
The adjustable depth values vary depending upon the probe
types.
TGC adjustment
Adjust through the
8-segment toggles
Push the toggle to the right to increase the corresponding
area brightness
Push the toggle to the left to decrease the corresponding area
brightness
About 1.5s after the adjustment is finished, the TGC curve
disappears.
Focus Position
Adjustment
Press
<Focus/Freq./THI.> to
light on the Focus
indicator, rotate the knob
The focus position will change in correspondence with the
knob rotates.
Frequency adjustment
Press
<Focus/Freq./THI.> to
light on the Freq./THI
indicator, rotate the knob
Rotate clockwise to increase
Rotate anticlockwise to decrease
following.
1. Control Panel
Function Checking and Testing 4-5
Page 50
2. Menu
Procedure
Standard
Acoustic power
B mode menu
[A.power]
Click [A.power] and rotate the multifunctional knob to adjust.
AP value is displayed in real time in the upper part of the screen.
You should perform exams according to actual situation and
follow the ALARA Principle.
Focus
B mode menu
[Focus Number]
Click [Focus Number] and rotate the multifunctional knob to
adjust (B mode image has max. 4 focus )
The focus position symbol is displayed on the right side of the
image.
FOV (Field of View)
B mode menu-> [FOV]
Click [FOV] and rotate the multifunctional knob to adjust
Dynamic Range
B mode menu->
[Dynamic Range]
Click [Dynamic Range], rotate the multifunctional knob to adjust,
the real time value will be displayed on the image parameter
area in the upper right of the screen. As the dynamic range
increases, the darker the image and the contrast, as well as the
noise may increase.
Line Density
B mode menu-> [Line
Density]
Click [Line Density], and rotate the multifunction knob to adjust
the parameter.
The higher the line density, the higher the resolution, and the
lower the frame rate.
iClear
B mode menu->
[iClear]
Click [iClear] to adjust, off represents no iClear is turned on, and
the bigger the value the stronger the effect. The bigger the value
the more clearly the profile of the image.
IP (Image Processing)
B mode menu-> [IP]
Click [IP] to switch among the IP groups, and the specific value
of each parameter can be preset.
iTouch
Press <Gain/iTouch>
iTouch sign will be displayed in the image area to activate
image effect auto optimization.
Click [iTouch] on the image menu to adjust gain value in
iTouch mode.
Click [iTouch Brightness] on the image menu to adjust
brightness in iTouch mode.
Image Magnification
Press <Depth/zoom> to
light on the Zoom
indicator, rotate the knob
Rotate clockwise to zoom in the image (max. magnification
factor is 10); rotate anticlockwise to zoom out the image.
Roll the trackball to change position of the magnified image.
Press <Zoom> in zoom status to exit the mode, the current
window returns to the display before zoom.
Press <B>, and then
press
Enter dual mode, and the right image is activated
Press again to switch between the windows
Press <B>, and then
press
Press 4 times, and 4 images will be displayed on the screen
with only one image activated at one time
Press again to switch among the windows
Press <B>
To enter single mode in multiple window mode, or to exit from
other modes.
4-6 Function Checking and Testing
Page 51
FR
B mode menu-> [High
FR]
In single B mode when THI is turned on, click the [High FR] item
to obtain images with high frame rates.
B mode menu-> [U/D
Flip]/[L/R Flip]
B mode menu->
[Rotate]
B mode menu-> [U/D Flip]/[L/R Flip]
Click [Rotate] to adjust.
When the image is rotated in the angle of 90° or 270°, the depth
scale is displayed on the upper part of the screen.
When you flip or rotate an image, the “M” mark will change its
position on the screen; the M mark is located in the upper left
corner of the imaging area by default.
4.4.2.2 M Mode
Display
V 3
IP 6
DR 65
G 45
Image
Parameters
M Speed
M IP
M Dynamic Range
M Gain
Adjustment
Items
Control Panel
Gain, TGC, Depth
Menu
Frequency, Gray Map, Focus Position, Dynamic Range, Speed,
Display Format, Colorize, Colorize Map, IP, Edge Enhance, M Soften,
Time Mark, Acoustic Power, Gray Rejection, γ, Curve, Gray Invert.
Procedure
Standard
Press <M>
Press <M> on the control panel, and roll the trackball to adjust the
sampling line.
Then press [M]
again.
Press [M] on the control panel again to enter M Mode, then you can
observe the tissue motion along with anatomical images of B Mode.
Press
<Update>
To switch between the active B image and frozen B image.
Tips:
a) Adjustment of the depth, focus position or TGC to the B Mode image will lead to
corresponding changes in M Mode image.
b) For details of other control panel adjusting parameters, please refer to
descriptions in B mode
In M Mode scanning, the image parameter area in the right corner of the screen will
display the real-time parameter values as follows:
Parameters that can be adjusted to optimize the M Mode image are indicated in the
following.
1. Control Panel
Function Checking and Testing 4-7
Page 52
2. Menu
Procedure
Standard
Speed
B mode menu-> [Speed]
Click [Speed], and rotate the multifunction knob to adjust
the parameter.
The lower the value the faster the refreshing.
Time Mark
B mode menu-> [Time Mark]
Click [Time Mark], and rotate the multifunction knob to
adjust the parameter.
Display Format
M mode menu-> [Display
Format]
There are 4 formats available for image display: L/R,
V1:1, V1:2, Full.
Tips:
a) During M Mode scanning, frequency and acoustic power of the transducer are
synchronous with that of B Mode.
b) Refer to B mode for more details.
Procedure
Standard
Press [Freeze] key to freeze an image, and
the [Cine] key indicator lights on. The system
automatically enters the manual cine
status.(Precondition: set “Status after Freeze”
to be “Cine”)
Press iStation, click [Review] or press
<Review> to open the cine file
Cine Review status
To enter cine auto cine review
status.
Roll the trackball
Manual cine review
Click image menu->[Auto Play]
Activate auto cine play status
Manually review the images until the frame which
you want to set it as start point, and then click [Set
First Frame] to set a start mark.
Set first frame:
Manually review the images until the frame which
you want to set it as end point, and then click [Set
End Frame] to set an end mark.
Set end frame
Click [Auto Play]
The cine is played within the setting
region.
Then press the [Cine] key again.
Cine review stops
Press the <Freeze> key to defreeze the
image.
Press <ESC> or <Cine>
<Freeze> backlight is off, the
image returns to the scanning
process and exits cine review.
The images are still frozen but
the system exits cine review.
4.4.2.3 Cine Review
4-8 Function Checking and Testing
Page 53
4.4.2.4 Measurement
Procedure
Standard
In B mode
Press <Measure>:
Press <Caliper> key
Enter the application measurement mode.
Enter the general measurement mode.
Click any 1-2 measurement items, the
measurement results will be displayed below the
image
Press <ESC> or the same key again.
Exits measurement.
Perform similar operations in other
modes
Application measurement functions are related to
certain application software packages.
Procedure
Standard
Press <Save1/2> in image scanning
process (preset already).
The image will be saved to corresponding patient
database and the saving icon will be displayed in
the right of the screen.
Press <Review>
Click [Exit] on the Review
screen; or press <Review> or
<ESC>.
To enter Review:
To exit Review:
Click []->[iVision] or press
user-defined key for iVision on
the control panel
Add the contents to be played
and select the demo mode,
select the to be played item and
click [Start]
After the demo is displayed, click
[Exit] or press <ESC>
Enter iVision screen:
The system begins image display
Image files are played according to file names one
by one (including the image of system-relevant
and PC-compatible format).
Exit
Press <iStation> to enter Patient
information management (iStation
page)
Saved images and information for the patient can
be checked here, and you can:
Backup/ Restore
Send to (DICOM or network storage)
Start mark
Cursor
End mark
Current frame/ Total
frames
4.4.2.5 Patient Information Management
Function Checking and Testing 4-9
Page 54
4.5 Performance Test
Resolution
Detection depth
Geometric Position
Accuracy
Blackout area test
Record the exam
images
NOTE:
The image used here is only for reference, stick to the image effect in the
real situation.
4.5.1 Test Process
4.5.2 Test Content
Requirements:
1. Display: set the contrast and brightness at the clinical application value (or the
default status)
2. Operation environment: dark room, simulating the clinical application environment.
3. Scanning techniques: contact the probe with the acoustic window of the phantom, no
spacing nor pressing.
Tips:
1. For the testing phantoms, please refer to Appendix B.
2. For the testing standards, please refer to Appendix C.
4.5.2.1 Resolution
transverse resolution
Test Step:
1. Cover the scan surface of the phantom with water or couple gel, gently contact the
probe with the scan surface, making the transverse resolution testing targets to be
displayed around the midline of the image.
2. Adjust the focus point focuses at the position where the transverse resolution testing
targets are displayed.
4-10 Function Checking and Testing
Page 55
3. Adjust parameters like gain, dynamic range, TGC, making the background tissue
unseen, just displaying the target image clearly.
4. In condition that the transverse resolution testing targets are horizontally displayed,
record the minimal distance of two targets that can be clearly recognized.
5. Repeat the operation above for the transverse resolution testing targets at other
depths.
As shown in figure below.
Axial resolution
Test Step:
1. Cover the scan surface of the phantom with water or couple gel, gently contact the
probe with the scan surface, making the longitudinal resolution testing targets to be
displayed around the midline of the image.
2. Adjust the focus point focuses at the position where the longitudinal resolution
testing targets are displayed.
3. Adjust parameters like gain, dynamic range, TGC, making the background tissue
unseen, just displaying the target image clearly.
4. Record the minimal distance of two longitudinal resolution testing targets that can be
clearly recognized.
5. Repeat the operation above for the longitudinal resolution testing targets at other
depths.
Function Checking and Testing 4-11
Page 56
NOTE:
1.
When using the convex probe, keep the transverse resolution testing
targets to be displayed near the midline.
2.
When using a linear probe with steer function, do not turn on the steer
function when perform the transverse resolution test.
3.
Zoom in the region where the targets located if necessary.
4.
The diameter of the target point at a certain depth is equal to the
transverse resolution at the depth.
NOTE:
1.
Increasing the gain will also increase the noise, and echo may be
covered.
2.
When using a linear probe, please completely contact the probe with
the scan surface, no side clearance is allowed.
3.
When using a convex or phased-array probe, make the axis targets to
be displayed at the middle of the scanning image.
4.
When system is not frozen, the fast field target information may be
similar to that of the noise, do not use this target.
4.5.2.2 Maximum Depth
Test Step:
1. Cover the scan surface of the phantom with water or couple gel, gently contact the
probe with the scan surface
2. Set the system display depth according to the expected maximum available depth of
the probe in use.
3. Adjust the focus point to the deepest, and AP at the maximum value.
4. Set gain, contrast, TGC at a greater value, but no halation nor defocus is allowed.
5. Record the depth of the furthest target (the target can be seen clearly).
As shown in figure below.
4-12 Function Checking and Testing
Page 57
NOTE:
1.
The measurement caliper should be positioned at the upper edge of
the target, not the middle nor the lower edge.
2.
The scanning plane should be vertical to the target line, that means the
scanning plane is parallel with the cross-section of the phantom
4.5.2.3 Geometric positioning accuracy
Longitudinal geometric positioning accuracy
Test Step:
1. Do adjustments as the way in testing the maximum depth.
2. Record the distance by 20mm each segment on the longitudinal targets line using
the measurement caliper;
3. Select the value with the greatest error (to 20mm), calculate the accuracy using the
formula below
As shown in figure below.
Function Checking and Testing 4-13
Page 58
NOTE:
1.
When using a linear probe, record the transverse distance by
segment.
2.
When using a convex probe, all transverse targets should be
displayed integrally in an image.
3.
The measure caliper should be posited at the upper side or lower side
of the target center.
Transverse geometric positioning accuracy
Test Step:
1. Cover the scan surface of the phantom with water or couple gel, gently contact the
probe with the scan surface
2. Adjust the depth, making the transverse targets to be displayed in the image.
3. Adjust the focus point to be posited beside the transverse targets (the standard is
not clear)
4. Adjust parameters like gain, TGC, making each transverse targets to be clearly
displayed.
5. Record the distance by 20mm each segment on the transverse targets line by using
the measurement caliper
6. Select the value with the greatest error (to 20mm), calculate the accuracy by using
the formula below
As shown in figure below.
4-14 Function Checking and Testing
Page 59
NOTE:
1.
When using a linear probe, please completely contact the probe with
the scan surface, no side clearance is allowed.
2.
For convex probe, the targets in the blackout area should be
positioned on the midline of the scanning plane.
4.5.2.4 Blackout Area
Test Step:
1. Cover the scan surface of the phantom with water or couple gel, gently contact the
probe with the scan surface
2. Adjust the depth at a lower value, and set the focus at the nearest place to the scan
surface.
3. Decrease the value of parameters like AP, Gain until the background noise just can
be seen.
4. Record the smallest depth of the target that can be seen clearly, that value is the
blackout area value.
As shown in figure below.
Function Checking and Testing 4-15
Page 60
Page 61
5 Software Upgrade and
NOTE:
Log on the system with the identity of Service before perform system
maintenance.
Maintenance
WARNING: DO NOT directly remove a USB memory device; otherwise,
the USB memory device and / or the system may be damaged.
5.1 Enter the Maintenance Window
To log on the system:
1. When access control function has not been activated: press “Ctrl+/” to show the
Login dialogue box, and then select the Service as the user name.
2. When access control function has been activated already: press “Ctrl+/” when the
Login dialogue box is displayed, and then select the Service as the user name.
3. Press <Setup> key to open the Setup menu, click [Maintenance] and then select the
target items to perform the maintenance respectively.
Software Upgrade and Maintenance 5-1
Page 62
NOTE:
Please check all the files in the restoration package before starting restoring,
or else copy the package again.
WARNING:
1.
Back up the patient data and preset data before restoring
the system.
2.
Do not cut off, shut down or restart the system during
system restoring.
5.2 System Restore
5.2.1 Restore the Operating System and Doppler
5.2.1.1 Software Introduction
Operating System restoration package includes the files as follow:
Doppler: Doppler software restoration data will be placed here. If the data exists,
ultrasound software will be restored after the operation system is restored; otherwise,
only the OS will be installed.
Sys: operation system data is placed here.
Mrcs.sh: installation program.
Version.txt: version file.
5.2.1.2 Preparation before Restoring
1. Back up preset data and patient data.
2. USB disk requirement:
a) Volume of USB disk should be more than 2G.
b) The system supports only FAT or FAT32 format USB disk.
c) If there are many drives in USB disk, installation may fail.
d) Disconnect the extra USB devices to the machine.
5-2 Software Upgrade and Maintenance
Page 63
5.2.1.3 Method
NOTE:
1.
The restoration package must be copied to USB disk root, or else it
may cause to failure of restoration;
2.
Please check all the files in the restoration package when copying, or
else copy the package again.
The operation system (“OS” for short) adopts Linux kernel and uses USB disk to
install. You can install the ultrasound software to the hardware when installing the
operation system.
1. Copy files of the system restoration package (including operation system and
Doppler) to USB disk root, and make sure that all the files of Doppler restoration
package are within the Doppler file.
2. Insert the USB disk to the ultrasound machine.
3. Power on the machine, press the button to start, and press <F4> immediately, and
when the "Mindray eRescue” displays on the screen, stop pressing F4 to enter
system installation program. Select the installation operation to enter.
Note: enter the corresponding method (1, 2 or q) and press <Enter>.
Enter 1: Start the OS restoring directly, user data will be kept when finished.
Enter 2: The HDD will be formatted and all data will be erased at first, then start
the OS restoring.
Enter q or nothing: The system does nothing and restarts.
4. Screen prompts:
If you enter q, the screen displays as below:
If you enter 1, the screen displays as below:
Software Upgrade and Maintenance 5-3
Page 64
If you enter 2 and press <Enter>, the system will ask the user to confirm if to delete all
data, enter “Y” and the system performs formatting and then goes to installation.
If you enter N here, the installation is stopped, and the system remains unchanged.
5. After the system has been installed successfully, the following information will
display on the screen, press <Enter> to restart the system.
6. After the OS is restored, Doppler restoration starts automatically. The process of
restoring refer to Doppler restoration (from the 5th to the 8th step)
5-4 Software Upgrade and Maintenance
Page 65
5.2.1.4 File Processing during Restoration
NOTE:
1.
Before executing the restoration package, please back up the patient
data as well as the user-defined preset data.
2.
When using the restoration package, key catalog in the HDD will remain
the same, and after restoration, the option key can take effect directly.
During the OS installation procedure, the machine will back up the key file in the HDD
automatically with choice 1 (Please install the key file if choose another):
1. Back up the option key file of the C drive to D:/mnt/bkcfg (temporary catalogue)
2. Format the C drive and restore the operating system.
3. Copy the option Key file to the previous catalogue in C drive, and the temp file
will be deleted.
In the OS with previous configuration data, if the ultrasound system is restored by
package without configuration data, the product configuration and the option state will
take effect as the configuration data in OS.
5.2.2 Doppler Restoration
5.2.2.1 Doppler Software Introduction
Figure 5-1 Doppler restoration Package Structure
5.2.2.2 Using Package in Doppler
1. Copy the Doppler restoration package to USB disk root (USB disk requirement is the
same as in 5.2.1.1)
2. Insert the USB disk to USB interface in the ultrasound system, the USB icon
displays on the right corner of the screen.
3. Select [Recover] in the Maintenance menu and press <Set>.
4. Switch to the USB catalog in the popped up dialogue box, and select
boot_doppler.sh file in the package and click [OK].
Software Upgrade and Maintenance 5-5
Page 66
NOTE:
If there is a package in the HDD, you can select the HDD path as well.
NOTE:
The old user preset data may be not compatible with the restored
system, if this happen, the operator must select “Y” when the system asks
whether to delete user preset data or not, otherwise, the system can’t be
restarted normally.
5. The system then will ask to restart the system. Click [OK] to enter restore procedure,
click [Cancel] will cancel the operation and returns to Doppler.
6. During the course, when the restoration is almost done, the following prompts will
display:
Enter “Y” or “y” to delete the data in D:\n7\Preset\Enter "N" and the system will not
delete data.
7. After the whole system is restored, if there is only green shutdown prompts
displayed, the restore process is normal, and press <Enter> to quit.
5-6 Software Upgrade and Maintenance
Page 67
NOTE:
If red prompts displays, please check the failed items.
Type
File
Suffix
Process
Destination
HDD
data
APD
Analyze the update file, replace the
old data with the new data
according to the update item.
C:\n7\app
EXD
C:\n7\exe
VDD
C:\n7\video
GUD
C:\n7\gui
IMD
C:\n7\Image
PST
C:\n7\Preset
PCD
C:\n7\config,
C:\usr\share\splashy\theme
s\default\poweron.png and
poweroff.png
Demo
Data
Replace the old demo with the new
demo in the resume package.
E:\n7\Demo
Driver
DRD
Decompress the update file to the
temporary directory in the hard disk,
then install the driver,
Install the drivers
C:\n7\driver
Check if every item is restored successfully, press <Enter> according to the prompts,
and the machine shut down automatically.
8. After restoration and restart, please check software version information in the path:
<Setup>-[About]
5.2.2.3 Data Processing during Restoration
Data structure and location is as the following after Doppler package restoring.
Table 5-1 data processing list
Software Upgrade and Maintenance 5-7
Page 68
NOTE:
After the Doppler is restored, product configuration data will be overwritten
by general configuration data in Doppler restore package, if necessary,
please configure the product manually.
NOTE:
[Config] is available on the Maintenance menu only if the operator logged on
the system as Service.
FPGA
FPD
Decompress the update files to the
temporary directory of the hard
disk, and write data in their
configured flash through the
corresponding drive program.
Delete the temporary directory
when finished.
Configuration flash chip in
the main board in
corresponding to logic.
Control
Panel
KUG
Decompress the update files and
cache the keyboard data to the
buffer, inspect keyboard
communication and erase the flash.
Control Panel
Power
supply
monitor
ARM
Decompress the update file, save
ARM data to the temporary
directory in the hard disk, inspect
the ARM communication and erase
the flash, and then delete the
temporary directory in the hard disk
after the installation finished.
Power supply panel
5.3 Software Maintenance
5.3.1 Product Configuration
1. Make the product configuration file—PCF file (generated by the production line or
already archived PCF file) according to specific requirements and copy to the USB
disk.
2. Connect the USB disk to the machine, click [Config] and load the file in the popped
up dialogue box.
5-8 Software Upgrade and Maintenance
Page 69
NOTE:
After Doppler restoration, please do product configuration manually
according to the practical situation.
NOTE:
[LOG] is available on the Maintenance menu only if the operator logged on
the system as Service.
3. Select correct file (PCF format), and click [OK]. After the operation is completed, the
successful prompt appears. Restart the machine according to the prompt.
Specific configuration:
a) Model configuration
b) Customized product logo, hospital logo, screen saver/hibernation icon,
active/inactive icon of dual B mode, system information, boot up/shut down
graphics.
5.3.2 Log Maintenance
5.3.2.1 Log Browse
1. Click [LOG] to open the Load File dialogue box, select the correct directory, the
default path is D:\N7\log.
Software Upgrade and Maintenance 5-9
Page 70
NOTE:
This operation can be performed by both internal and external users.
2. Select the log to be viewed, and click [OK] to open the log message.
3. Click [Export] to export the currently viewed log, or click [Close] to return to the
Maintenance menu.
5.3.2.2 Export the Log
1. Connect the USB disk.
2. Click [Export Log] on the Maintenance menu, and select the log type in the
sub-menu. Refer to 5.3.2.1
5-10 Software Upgrade and Maintenance
Page 71
3. Select the path in the Browse page to save the log, and click [OK].
NOTE:
The log can be exported to the external USB storage device only, make
sure the connection between U disk and ultrasound is normal before the
exporting.
NOTE:
This operation can be performed by both internal and external users.
NOTE:
1.
Before uploading the log, make sure the ultrasound system has been
connected to network, otherwise, the system may warn “could not
connect to server”
2.
The server (ftp.mindray.com) is already specified by the system, the
user doesn’t have to select it.
NOTE:
In normal condition, when changing display assembly or main board
assembly, the hardware will update the display data automatically. If
32-channel transmission wave forms according to scanning time sequence and
control parameters.
32-channel transmission wave forms to 32-channel high-voltage transmission
wave by driver.
Echo receiving LNA is to enlarge fixed gain.
By the control of TGC, echo receiving carries out variable gain amplification by
depth.
32-channel analog echo could be switched to 32-channel digital signal by ADC.
BF FPGA combines 32-channel echo signal to output to DSP FPGA, and DSP
FPGA is sent to signal processing to a single image which is uploaded to CPU
module by PCIE_PHY, then CPU module performs post processing.
PCI_PHY implements PCIE interface switching.
BF FPGA can communicate with the CPU module through transferring DSP
FPGA.
Hardware System 6-3
Page 78
The CPU module communicates with the probe board via BF FPGA.
IO board
PCIE
PHY
PCIE X1
SATA
HDD
VGAUSB
Video
S-
Video
Remote
Ethernet
MUL FPGA
LPC
LVDS
USB
CPU module
SATA
+12V
LVDS
Display
SMBUS
Control signal
of the inverter
Video
Encoder
Cont
rol
Data
USB
lateral
USB
System monitor
circuit
Power supply
module
ARM control signal
HDD
indicator
Control signal of
parameter board
Control panel
Main board
Status
indicator
5V
12V
3.3V
POWER
12V
5V
5VSTB
Status
indicator
Power_BTN_N
Closed status indicator on
the control panel
ARM
Power
manage
ment
FPGA
Control
signal
Power on/off
signal
Audio
Codec
Audio
power
amplifier
Speaker
ECG
6.2.2.2 Back-end of the Main Board
Back-end of the main board is the core of back-end ultrasound system, which mainly
includes back-end of the main board, CPU module, IO interface board, monitor assembly
and control panel. Mainly implements display interface, user's operation interface, power
module controlling interface and all kinds of Peripherals Supported interface.
Also provides system monitoring, running status of the monitor system, indicator of
system status and running status of the indicating system via LED.
Function describes as following:
Back-end of the main board connects to front-end of the main board via PCIE
The USB on the lateral of Main board is the output of the CPU module.
The relevant signal in HDD of the main board is output of the CPU module.
Back-end of the main board MUL FPGA implements display ports, it also
Multifunction FPGA on the back-end of the main board is to implement the
Figure 6-5 The Schematic Diagram of Back-end of Main Board
PHY.
supports display assembly, video and s-video interface of IO board.
REMOTE port on the IO panel.
6-4 Hardware System
Page 79
The CPU module implements interfaces including USB, Ethernet and VGA on
Main Board
Video S-VideoRemote
Multifunction FPGA
Video Encoder
Control
Data
VGAUSBEthernet
CPU module
System
Monitor
Circuit
Status Indicator
the IO board.
System monitor circuit connects Multifunction FPGA though smbus.
Multifunction FPGA communicates with ARM by serial port to control
transmission high-voltage, and also implements ARM update by serial port.
FPGA on the power supply module implements the machine power on/off by the
switch signal.
Standby indicator, AC present indicator and working status indicator on the
control panel are output by FPGA.
The control panel implements communication with CPU module by USB.
CPU output the audio signal to the speaker on the front cover of main unit, via
audio coder and audio amplifier.
6.2.3 IO Broad
IO board connects to the main board via connectors in order to implement the user
interface function and indicate if the power is normal.
IO board supports Video and S-Video, and the two channels of signal get by
IO board supports REMOTE signal, which is obtained by Multifunction FPGA on
USB signal, Ethernet signal and VGA signal on the IO board are obtained by
Status indicators of IO board respectively refers to12V, 5V and 3.3V powers are
Figure 6-6 Principle Diagram of IO Board
Video Encoder drived by Multifunction FPGA.
the main board.
CPU module output.
normally working.
Hardware System 6-5
Page 80
6.2.4 Ultrasound SystemMonitor
CPU
module
Multifunction
FPGA
System
monitor
curcuit
LPC
SMBUS
Temperature
sensor
Main unit front-
end FPGA
temperature
D12V
D-12V
D-5V
D2.5V
D1.8V
D1.5V
D1.2V
Voltage of
clock
battery
Back-end of main board
CPU module
temperature
Main unit fan
Power supply
module fan
Power supply
module
temperature.
Figure 6-7 Principle Diagram of Ultrasound System Monitor
Ultrasound system monitor is mainly to monitor every voltage of interior machine and
Function:
On the back-end of the main board, a special monitor IC is used to monitor
signal voltages and clock battery power that are provided to the main board by
the main unit box power supply module.
The system temperature is detected in three positions: BF FPGA, CPU module
and power supply module.
Monitoring chips control the main unit fan and power supply module fan, and
then read the fan rotating, which send to Multifunction FPGA by smbus, and to
PC module by LPC bus.
6-6 Hardware System
Page 81
6.2.5 Ultrasound System Indicator
CPU
Module
Main board
Power
status
indicator
Power-on status
indicator
Battery status
indicator
AC supply status
indicator
Standby status
indicator
Power
management
FPGA
HDD indicator
driver
5VSTB_CPU
Control panel
+12V
+5V
3.3V
ARM
driver
Power supply
module
IO Board
driver
driver
driver
Figure 6-8 Principle Diagram of Ultrasound System Indicator
Function describes as following:
There are four power status indicators on the IO panel: three are used for
indicators of +12V, +5V and 3.3V output on the DC-DC power board, the other
one is used for reserved indicator.
There are five indicators on the control panel, in which power-on status indicator
is below the power button.
AC power-on status indicator control is output by the power management FPGA
on the power module.
The hardware indicator control is output by CPU.
The status indicator of battery are controlled by the battery management ARM
on the power module which drived by 5VSTB_CPU on the main board.
The status indicators of power-on and standby are both generated by the
battery management FPGA in the power module and drived by 5VSTB_CPU on
the main board.
Hardware System 6-7
Page 82
6.2.6 Display
LVDS3.3V
Display
High-voltage
LCD
screen
Parameters
board
I2C
3.3V
Inverter board
ON/OFF
12V
Brightness
control
Main board
IO board
12V 3.3V
Multifunction FPGA
LVDS
I2C
triodetriode
Control signal
of switch
Brightness
PWM
ON/
OFF
Brightness
control
Figure 6-9 Principle Diagram of Display
Monitor mainly consists of the inverter board, LCD and the parameter board.
Function describes as following:
The inverter board produces high voltage to illuminate backlight of LCD which is
adjusted by signal control switch and brightness of the IO board output.
Color temperature and Gamma correction data memory in parameter board
used matches LCD so that the corresponding display effect shall be consistent.
Multifunction FPGA generates switch control signal which drives triode to control
inverter to open or shut off high-voltage output.
Multifunctional FPGA generates PWM signal, which drives triode to control output
high-voltage of the inverter to adjust the brightness of the display.
Multifunctional FPGA outputs LVDS digital signal of display, which is sent to screen
directly.
Multifunctional FPGA communicates with parameter via 12C signal and supports to
adjust parameters of display.
6-8 Hardware System
Page 83
6.2.7 Control Panel
5 status
indicators
TGC board
Trackball
Functional
button
Single-
encoder
board
Key
scan
LED
backlight
driver curcuit
LED
backlight
USB
Interface
chip
ADC
curcuit
Buzzer
Buzzer driver curcuit
Main
control
Trackball
scan
TGC
scan
PWM generating
curcuit
FPGA
Control panel
PCBA
Main
board
USB
+12V
+5V
Doubleencoder
board
Figure 6-10 Principle Diagram of Control Panel
panel PCBA, trackball, TGC board, encoder board and buzzer.
The control panel PCBA is the core of the control panel which comprises control
Function describes as following:
FPGA is regarded the control panel as a main controller , key,
single/dual-encoder, trackball and TGC.ect are basically scanned by FPGA;
Key LED backlight and buzzers are both controlled by FPGA, and only when
FPGA is normal can key LED backlight be on.
Five indicators on the control panel are directly drived from interior of main unit.
The control panel communicates with the CPU module by USB ports.
Hardware System 6-9
Page 84
6.3 Power System
AC Input
100-240V~,50/60Hz
Battery
14.8V,6600mAh
DC Output 19V/3.95A
Power supply module
Power
management
FPGA
DC-DC power board
AC-DC power board
PWR_BTN_N
Main board
Control
panel
System monitor
module
Temperature signal
CPU
module
ARM
Multifunction
FPGA
ARM control signal
POWER
I2C
PHV
ADDA
Closed signal of the control panel
AC in place
Power on/off control signal
Status indicator
control signal
12V、5V
battery in place
Fan
Fan monitor
12V
Display
12V
Battery
connecting
board
I2c
battery
in place
14.8V,
6600mAh
Status indicator
Restart automatically
No.
Power
description
Support circuit module or function
Remarks
1
+12V
CPU module, Display, Fan, Control panel
2
-12V
No use
3
+5V
HDD, Control panel, Front-end of the main board, Probe
board
4
-5V
Front-end of the main board
5
5Vst
CPU module
Figure 6-11 Block Diagram of Power System
The power supply configuration of the system describes as following:
The power supply module can be divided into three parts: AC-DC power board,
DC-DC power board and battery connecting board.
High-voltage programmable ARM and power management FPGA are on the DC-DC
power board.
Power output by the power supply module is transmitted to the main board, and then
it is output to control panel and display by the main board.
Battery is connected to the DC-DC board via the battery connecting board and
connecting cable of BTB sequentially.
6.3.1 Power Output of the Power Supply module
and Supporting Function Distribution
6-10 Hardware System
Page 85
6
VDD
Front-end of the main board, Back-end of the main
board, Display
7
1V2
Back-end of the main board (FPGA core voltage)
8
1V8A
Front-end of the main board (receiving chip)
9
1V8
Back-end of the main board ( video encoder, FPGA IO
voltage , DDR2)
10
2V5
Back-end of the main board ( video encoder, FPGA IO
voltage)
11
+95V
Probe board (high-voltage switch)
12
-95V
Probe board (high-voltage switch)
13
+PHV
Transmitting high-voltage
14
-PHV
Transmitting high-voltage
6.3.2 System Power-on Control
CPU
module
PWR_BTN_N
D12V
S3_N
S4_N
CPU_PWR_BTN_N
PWR_OK_N
5VSTB_CPU
Power
manag
ement
FPGA
AC-DC
Power
board
12V
AC in
place
signal
DC-DC power board
Main board
Control
panel
PWR_12V_EN
PWR_5VSTB_EN
PWR_5VSTB_OK_N
PWR_12V_OK_N
No
Controlling signal
Description
Remarks
1
PWR_BTN_N
The pulse generated by the power button on the
control panel is sent to CPU module through power
management FPGA, to power on the system.
2
S3_N
CPU module output effectively represents that CPU
system has been in the standby status, when
5VSTB_CPU which is controlled by power
management FPGA is in the charging status.
3
S4_N
CPU module output effectively represents that CPU
system has been in the dormancy status.
4
PWR_OK_N
The output of power supply module is transmitted to
Figure 6-12 Diagram of System Power-on Supported Circuit
Related controlling signal comment:
Hardware System 6-11
Page 86
the CPU module by power management FPGA,
which represents the +12V output of power supply
module has been powered on.
5
CPU_PWR_BTN
_N
Power management FPGA is transmitted to CPU,
which notifies CPU power on or power off.
6
PWR_12V_EN
The output of power management FPGA is
transmitted to the power supply module, which is
used to control other power of the power module
except 5VSTB_CPU and 3V3STB to be power-on ,
and also used to control +12V and +5V power-on
on the DC-DC power board.
7
PWR_5VSTB_E
N
The output of power management FPGA notifies
power to provide 5V standby power.
8 PWR_12V_OK_
N
The 12V power has been provided steadily.
9 PWR_5VSTB_O
K_N
The 5VSTB power has been provided steadily.
When battery and AC supply, the power supply module always maintains the
3V3STB output, and supports power management FPGA of the power module to be
standby status.
No matter the system is on, the system could provide AC power to 5VSTB_CPU
though DC-DC power board controlled by PWR_5VSTB_EN, after AC power is
indicated by AC online signal, and transmitted though power management FPGA of
the power supply module.
However, when powered by battery, power management FPGA of the power module
could provide power to 5VSTB_CPU though DC-DC power board controlled by
PWR_5VSTB_EN according to battery-present status if only users turn on the
system via power button of the control panel.
The detailed flow of power-on is shown as follows:
6-12 Hardware System
Page 87
Pressing power button
(control panel)
Is AC in place?
(Ppower management
FPGA)
NO
Receiving PWR_BTN_N effective
(Power management FPGA)
PWR_5VSTB_EN is valid
(Power management FPGA)
5VSTB_CPU start-up
(Power supply module)
PWR_5VSTB_OK_N is valid
(Power supply module)
CPU_PWR_BTN_N is valid
(Power management FPGA)
S3_N and S4_N is invalid
(CPU module)
PWR_12V_EN is valid
(Power management FPGA)
PWR_12V_OK_N is valid
(Power module)
CPU start-up
(CPU module)
12V and +5V start-up
(DC-DC power board)
PWR_OK_N is valid
(Power management FPGA)
YES
Indicating 12V
output of power
module is normal
Indicating
5VSTB_CPU
output of power
module f is normal
Figure 6-2 System Power-on
Hardware System 6-13
Page 88
Page 89
7 Structure and
No.
Name
No.
Name
1
Holder
5
Null turning axis of keyboard
2
Rear cover assembly of
main unit
6
Silicon cap of keyboard
3
Assembly of main unit rack
7
Top cover assembly of keyboard
4
Assembly of monitor
8
Bottom cover assembly of keyboard
Assembly/Disassembly
7.1 Structure of the Complete System
Figure 7-1 Overall Exploded View
7.2 Field Replaceable Unit
Structure and Assembly/Disassembly 7-1
Page 90
No.
Classification
Description
Order Number
Photo
Model
Compatibility
Assembly/
Disassembly
1.
Main unit
related
Main board (FRU)
801-1151-00001-00
DP-50
refer to 7.4.5
2.
1151 Main Board
Assembly(FRU
EAF328-F)
115-031630-01
DP-50
This board should
be used with
software of
version V02.03.00
(Rev15255) and
later.
3.
Main Board Assembly
(Support PW/FRU)
115-052167-00
DP-50
This board is
compatible with
Doppler in version
02.03.00
(Rev15255) for
human use,
02.03.01
(Rev15255) for
vet use, and
02.02.01
(Rev15391) for
DP-50 PT, or
former versions;
Doppler in version
03.00.00
(Rev16267) for
human use,
03.00.00
(Rev16267) for
7-2 Structure and Assembly/Disassembly
Page 91
No.
Classification
Description
Order Number
Photo
Model
Compatibility
Assembly/
Disassembly
vet use, and
03.00.00A
(Rev16342) for
DP-50 PT, or later
versions is only
compatible with
this board.
4.
CPU Module
801-1150-00002-00
DP-50
refer to 7.4.5
5.
BayTrail PC Module
(Support PW/FRU)
115-052168-00
DP-50
The old and new
PCs cannot be
replaced with
each other. If new
J1900 PC
(version:
00100010) is used
to replace the old
N270 PC (version:
SOM-6761), the
new PC, OS in
version 10.1.5 or
refer to 7.4.5
Structure and Assembly/Disassembly 7-3
Page 92
No.
Classification
Description
Order Number
Photo
Model
Compatibility
Assembly/
Disassembly
later versions, and
Doppler in version
03.00.00
(Rev16267) for
human use,
03.00.00
(Rev16267) for
vet use, and
03.00.00A
(Rev16342) for
DP-50 PT, or later
versions should
be matched.
CPU Module with
115-052168-00
has been
discontinued. If it
is out of stock,
please use
115-052168-01 to
order and restore
the latest Doppler
software and
compatible OS as
7-4 Structure and Assembly/Disassembly
Page 93
No.
Classification
Description
Order Number
Photo
Model
Compatibility
Assembly/
Disassembly
well after
replacement.
BayTrail PC Module
(Support PW/FRU)
115-052168-01
DP-50
The old and new
PCs cannot be
replaced with
each other. If new
J1900 PC
(version:
00100010) is used
to replace the old
N270 PC (version:
SOM-6761), the
new PC, OS in
version 10.1.5 or
later versions, and
Doppler in version
03.00.00
(Rev16267) for
human use,
03.00.00
(Rev16267) for
vet use, and
03.00.00A
(Rev16342) for
Structure and Assembly/Disassembly 7-5
Page 94
No.
Classification
Description
Order Number
Photo
Model
Compatibility
Assembly/
Disassembly
DP-50 PT, or later
versions should
be matched.
CPU Module with
115-052168-00
has been
discontinued. If it
is out of stock,
please use
115-052168-01 to
order and restore
the latest Doppler
software and
compatible OS as
well after
replacement.
6.
HDD
115-018460-01
DP-50
Mark the following
when apply:
software version
(Human use);
regulation type:
FDA
refer to7.4.8
7-6 Structure and Assembly/Disassembly
Page 95
No.
Classification
Description
Order Number
Photo
Model
Compatibility
Assembly/
Disassembly
115-018461-01
DP-50Vet
Mark the following
when apply:
software version
(Vet);
regulation type:
FDA
115-018462-01
DP-50
Mark the following
when apply:
software version:
human;
regulation type:
CE
115-018463-01
DP-50Vet
Mark the following
when apply:
software version:
Vet;
regulation type:
CE
115-027820-01
DP-50PT
regulation type:
CE
7.
HDD (BayTrail PC )
115-052172-00
DP-50
For CE. Please
mark out the
machine serial
number
Structure and Assembly/Disassembly 7-7
Page 96
No.
Classification
Description
Order Number
Photo
Model
Compatibility
Assembly/
Disassembly
8.
115-052173-00
DP-50 PT
For CE. Please
mark out the
machine serial
number
9.
115-052174-00
DP-50
Vet
For CE. Please
mark out the
machine serial
number
10.
115-052175-00
DP-50T
For CE. Please
mark out the
machine serial
number
11.
115-052176-00
DP-50
For FDA. Please
mark out the
machine serial
number
12.
115-052177-00
DP-50
Vet
For FDA. Please
mark out the
machine serial
number
13.
Battery connection
board
801-1150-00004-00
DP-50
refer to7.4.1
7-8 Structure and Assembly/Disassembly
Page 97
No.
Classification
Description
Order Number
Photo
Model
Compatibility
Assembly/
Disassembly
14.
IO board
801-1150-00005-00
DP-50
refer to7.4.3
15.
probe board
801-1151-00002-00
DP-50
refer to7.4.4
16.
Probe Board Assembly
(Support PW/FRU)
115-052169-00
DP-50
This probe board
is compatible with
Doppler in
02.03.00
(Rev15255) for
human use,
02.03.01
(Rev15255) for
vet use, and
02.02.01
(Rev15391) for
DP-50 PT, or
Structure and Assembly/Disassembly 7-9
Page 98
No.
Classification
Description
Order Number
Photo
Model
Compatibility
Assembly/
Disassembly
former versions in
old N270 PC
(version:
SOM-6761);
Doppler in
03.00.00
(Rev16267) for
human use,
03.00.00
(Rev16267) for
vet use, and
03.00.00A
(Rev16342) for
DP50-PT, or later
versions is only
compatible with
this probe board.
17.
Power related
power supply module
801-1150-00009-00
DP-50
refer to 7.4.2
7-10 Structure and Assembly/Disassembly
Page 99
No.
Classification
Description
Order Number
Photo
Model
Compatibility
Assembly/
Disassembly
18.
AC-DC board( FRU)
115-051844-00
DP-50
/
refer to 7.4.2
19.
DC-DC board( FRU)
115-051845-00
DP-50
/
refer to 7.4.2
20.
Battery
M05-010005-09
DP-50
refer to
7.2.5.1
Structure and Assembly/Disassembly 7-11
Page 100
No.
Classification
Description
Order Number
Photo
Model
Compatibility
Assembly/
Disassembly
21.
Control panel
related
Control panel
assembly
801-1151-00003-00
DP-50
refer to7.4.6
22.
Silicone rubber keypad
801-1151-00004-00
DP-50
refer to
7.4.6.5
23.
control panel PCBA
801-1150-00012-00
DP-50
refer to
7.4.6.5
7-12 Structure and Assembly/Disassembly
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.