Miele PG 8063 Installation diagram

0 (0)
Miele PG 8063 Installation diagram

Installation plan

PG 8063

To avoid the risk of accidents or damage to the machine, it is essential to read these instructions as well as the service documents before it is installed, commissioned and used for the first time.

en - GB, IE, AU, NZ

M.-Nr. 11 486 560

Installation notes

Installation re-

This machine must be installed by a suitably qualified person with

quirements

the appropriate electrical and plumbing qualifications in accord-

 

ance with the installation instructions supplied.

 

This machine must be installed in accordance with all applicable

 

standards and guidelines, including legal requirements and health

 

and safety regulations.

 

This machine must be commissioned and operatives trained in its

 

use by Miele Service or by an approved Miele Service Partner only.

Environmental re-

Condensate can build up in the area surrounding the machine. Any

quirements

furniture and fittings in the room must therefore be suitable for pur-

 

pose.

Electrical connec-

Connection to the electrical supply must be carried out in accordance

tion

with local and national safety regulations. The power cord must be

 

protected from the risk of thermal damage.

 

It is recommended to make electrical connection via a plug and

 

socket so that electrical safety checks can be carried out easily.

 

For hard-wired machines, connection must be via a main switch to

 

be provided on site, which must completely isolate the machine from

 

the power supply with a contact gap of at least 3 mm.

 

The plug and socket as well as the main switch must be easily ac-

 

cessible after the machine has been installed.

Equipotential

For added safety the machine should be protected with a residual

bonding and

current device with trip current of 30 mA.

earthing

Equipotential bonding should be carried out if required. A screw

 

connection for equipotential bonding is provided at the back of the

 

machine. Equipotential bonding and earthing must be completed be-

 

fore the machine is commissioned.

Plumbing

This machine can be connected directly to a cold water supply or a

 

hot water supply using the supplied double check valves or to a de-

 

mineralised water supply. If a hot water supply is not available, both

 

inlet hoses (KW/WW) must be connected to the cold water supply

 

with a Y-connector. The stopcocks must be easily accessible.

2

PG 8063

Installation notes

Liquid agents: poThe liquid agent container for external dispensing must be positioned sition of external either next to or underneath the machine only. The container may be containers placed on the floor or in an adjacent cabinet. Do not place the con-

tainer on top of or above the machine. Make sure that the dispensing hose does not get kinked or trapped.

Compressed air A flexible compressed air connection must be provided on-site for reprocessing air regulators. The air quality must comply with the specifications issued by the manufacturer of the breathing apparatus. The pressure must be at least 400 kPa and should not exceed

600 kPa.

The compressed air connection must be equipped with one of the following quick-release couplings:

CEJN series 341

Parker Rectus series 25S

PG 8063

3