The following instructions on how to regenerate the Metrohm Suppressor Module re­place section 4.2.5 of the 753 Instructions for Use:
Regenerating a suppressor operating at reduced capacity
If the suppressor units are exposed to certain heavy metals (e.g. iron) or organic contaminants for long periods of time, these can no longer be removed completely by the regeneration solution normally used (20 mmol/L H
). This diminishes the
2SO4
capacity of the suppressor units, which, in milder cases, results in reduced sensitiv­ity to phosphates and, in severe cases, in a strong increase in the baseline. If such capacity problems occur at one or more points, treat the suppressor units affected as follows:
1 Disconnect suppressor from IC system
Disconnect suppressor from separating column and detector.
2 Regenerate suppressor
Rinse each suppressor unit for about 15 min with one of the following solutions:
Contamination with heavy metals
1 mol/L H
2SO4
Contamination with organic cationic complexing agents
0.1 mol/L H
/ 0.1 mol/L oxalic acid / acetone 5%
2SO4
Severe contamination with organic substances
0.2 mol/L H
/ acetone 20%
2SO4
The 6.1826.050 pump tubing is made of PVC and must not be used for rinsing with solutions containing acetone. In such cases, use a different pump tubing or a different pump for rinsing purposes.
3 Connect suppressor to IC system
Reconnect suppressor to the IC system. If the capacity problems persist, replace the suppressor rotor (see section 4.2.7).
753 Suppressor Module, 8.753.1023
1
Regenerating the suppressor in case of high counterpressure
If excessive counterpressure is observed in one or more of the suppressor units, proceed as follows:
1 Disconnect suppressor from IC system
Disconnect suppressor from separating column and detector.
2 Regenerate suppressor
Connect the inlet capillary 25 marked "H
" to the 709 IC Pump using a
2SO4
6.2620.060 or 6.2744.040 coupling.
Adjust the 709 IC Pump flow rate to 0.5 mL/min and rinse the suppressor unit with 1 mol/L H
for 5 to 10 min.
2SO4
As the pressure drops, slowly increase the 709 IC Pump flow rate to 2 mL/min. The maximum pressure must not exceed 2 MPa (20 bar).
Switch suppressor to the next position using the <STEP> key.
Connect the inlet capillary 28 marked "H
O" to the 709 IC Pump using a
2
6.2620.060 or 6.2744.040 coupling.
Adjust the 709 IC Pump flow rate to 0.5 mL/min and rinse the suppressor unit with distilled H
O for 5 to 10 min.
2
As the pressure drops, slowly increase the 709 IC Pump flow rate to 2 mL/min. The maximum pressure must not exceed 2 MPa (20 bar).
3 Connect suppressor to IC system
Reconnect the suppressor to the IC system and the 753 Suppressor Module. If the pressure problems persist, replace the suppressor rotor (see section 4.2.7).
Supplement to «Cleaning the suppressor» p. 30
Supplementary to the cases listed in section 4.2.6 which may necessitate cleaning of the suppressor or replacement of connection piece 36, please also note the following case:
High counterpressure on the suppressor connection tubings
Changes in standard equipment p. 37
2 × 6.2744.010 PEEK Compression fitting (instead of 1 × 6.2744.010)
Set of 5
2 × 6.2128.180 Connecting cable (instead of 1 × 6.2128.090)
Connecting cable 762 – 752/753/754
2 × 6.2821.100 Filter unit PEEK 2 µm (new)
To avoid contamination due to abrasive particles of piston seals
New Options p. 38
6.2143.220 Connecting cable
Connecting cable 732 IC Detector – 753 Suppressor Module – 754 Dialysis Unit