The 752Pump Unit is a 2-channel peristaltic pump which can be used
to pump two solutions simultaneously. It is particularly suitable for operating the suppressor module in the 2.733.0X30 IC Separation Center
by supplying regeneration and rinsing solutions. This application is described in detail in these Instructions for Use.
The 752 Pump Unit can either be operated manually by means of the
on/off switch for the pump or remotely controlled via the remote interface.
1.1 Instrument description
752 Pump Unit1
1 Introduction
1.2 Parts and controls
334455
Pump Unit752
POWER
REMOTEPUMP
ONON
OFFOFF
221166889977
Fig. 1: Front of the 752 Pump Unit
MetrohmMetrohm
11Pump ON/OFF
On/Off switch for the pump
(manual operation mode)
22Remote ON/OFF
On/Off switch for remote control
via remote interface
33Power on lamp
Lights up when instrument is
switched on
44Tubing cartridge
For 6.1826.050 Pump tubing
55Contact pressure lever
For adjusting the contact pressure
66Holding clamp
For locking the tubing cartridge into
place
77Snap-action lever
For releasing the tubing cartridge
88Pump drive
Roller head with contact rollers
99Mounting pin
For attaching the tubing cartridge
752 Pump Unit2
1.2 Parts and controls
1010
WARNING - Fire Hazard -
For continued protection replace only
with the same type and rating of fuse
f=50-60 Hz
S=12 VA
115V: 110-120V:
230V: 220-240V:
Type 1.752.0010 Nr.
Fuse
0,2A(T)
0,1A(T)
S
131312121111
Remote
Pump
Made by Metrohm Herisau Switzerland
15151414
Fig. 2: Rear of the 752 Pump Unit
1010Fuse holder
Changing the fuses, see section 2.2
1111Voltage selection insert with
display of voltage
1212Mains switch
For switching the instrument on/off:
I = ON0 = OFF
1313Serial number
1414Remote interface
Remote I/O lines for connection of
external devices (e.g. 732 IC Detector)
1515Mains connection plug
Mains connection, see section 2.2
752 Pump Unit3
1 Introduction
1.3 Information about the Instructions for Use
Please read through these Instructions for Use carefully before you put
the 752 Pump Unit into operation. The Instructions for Use contain
information and warnings to which the user must pay attention in order
to assure safe operation of the instruments.
1.3.1 Organization
These 8.752.1013Instructions for Use for the 752 Pump Unit provide
a comprehensive overview of the startup procedure, operation, fault
rectification and technical specifications of these instruments. The Instructions for Use are organized as follows:
Section 1Introduction
General description of instruments, parts and controls
and safety notes
Section 2Installation
Attachment of accessories, connection to IC system
Section 3Operation
Manual operation and operation via 732 IC Detector
Section 4Maintenance – Faults
Maintenance, fault rectification
Section5Appendix
Technical data, standard equipment, options, warranty,
declarations of conformity, index
To find the required information on the instruments please use either
the Table of contents or the Index at the back.
752 Pump Unit4
1.3.2 Notation and pictograms
The following notations and pictograms (symbols) are used in these Instructions for Use:
<PUMP>Switch or key
1515Part or control of 752
8989Part or control of 732/733
"range"Parameter or entry value
1.3 Information about the Instructions for Use
at 732 IC Detector
>PARAM/detector
range: 1.00 mS/cm
Display
Text in display 11 of the 732 IC
Detector
Hazard
This symbol draws attention to a
possible danger to life or of injury if
the associated directions are not
followed correctly.
Warning
This symbol draws attention to
possible damage to instruments or
instrument parts if the associated
directions are not followed correctly.
Caution
This symbol marks important information. First read the associated
directions before you continue.
Comment
This symbol marks additional information and tips.
752 Pump Unit5
1 Introduction
1.4 Safety notes
1.4.1 Electrical safety
While electrical safety in the handling of the 752 Pump Unit is assured
in the context of the specifications IEC 1010-1 (protection class 1, degree of protection IP40), the following points should be noted:
• Mains connection
Setting the mains voltage, checking the mains fuse and the mains
connection must be effected in accordance with the instructions in
section 2.2.
• Opening the 752 Pump Unit
When the 752 Pump Unit is connected to the power supply the instrument must neither be opened nor should parts be removed from
it, otherwise there is a danger of coming into contact with components which are live. Always disconnect the instrument from all voltage
sources before you open it and ensure that the mains cable isdisconnected from mains connection 1515 !
• Protection against static charges
Electronic components are sensitive to static charging and can be
destroyed by discharges. Before you touch any of the components
inside the 752 Pump Unit, you should earth yourself and any tools you
are using by touching an earthed object (e.g. housing of the instrument or a radiator) to eliminate any static charges which exist.
1.4.2 General precautionary rules
• Handling of solvents
Check all pump tubings and all input and output lines periodically for
possible leaks. Follow the relevant instructions regarding the handling
of flammable and/or toxic solvents and their disposal.
• Regular exchange of pump tubings
Pump tubings are consumable material and must be replaced from
time to time (see section 4.4). Suitable measures must be taken so
that any leak which might occur in the pump tubing or connections
during unattended operation will cause no damage (placing the
instrument at the bottom, collection device for any liquid which may
leak out).
752 Pump Unit6
2 Installation
2.1 Setting up the instrument
2.1.1 Packaging
The 752 Pump Unit is supplied together with the separately packed
accessories in special packagings containing shock-absorbing foam
linings designed to provide excellent protection. The actual instrument
is packed in an evacuated polyethylene bag to prevent the ingress of
dust. Please store all these special packagings as only they can assure
damage-free transport of the instrument.
2.1.2 Check
2.1 Setting up the instrument
After receipt, immediately check whether the shipment is complete and
has arrived without damage (compare with delivery note and list of
accessories in section 5.2). In the case of transport damage, see
instructions in section 5.4.1 "Warranty".
2.1.3 Location
Position the instruments in the laboratory at a location convenient for
operation, free from vibrations and protected against a corrosive
atmosphere and contamination by chemicals.
2.1.4 Arrangement of the instruments
In one-channel operation with suppressor module (2.733.0X30 IC
Separation Center), the 752 Pump Unit, 709 IC Pump, 733 IC Separation Center and 732 IC Detector are best stacked on top of one another
in this order.
The 752 Pump Unit should always be placed at the bottom of the
instrument stack so that any leaks which may occur in the pump
tubing or connections cannot cause damage to the other instruments
by leakage of corrosive liquids (e.g. H2SO4).
752 Pump Unit7
2 Installation
2.2 Mains connection
Follow the instructions below for connecting to the power supply. If
the instrument is operated with the mains voltage set wrongly and/or
wrong mains fuse there is a danger of fire!
2.2.1 Setting the mains voltage
Before switching on the 752 Pump Unit for the first time, check that the
mains voltage set on the instrument (see Fig. 3) matches the local
mains voltage. If this is not the case, you must reset the mains voltage
on the instrument as follows:
Disconnect mains cable
1
Disconnect mains cable from mains connection plug 1515 of the
752 Pump Unit.
Remove fuse holder
2
Using a screwdriver, loosen fuse holder 1010 and take out com-
pletely.
Change mains voltage
3
Completely remove voltage selection insert 1111 by hand, rotate it
through 180° and reinsert it. The required mains voltage (115 or
230 V) must now be visible from the front.
Check fuses
4
Carefully take both fuses out of fuse holder 1010 and check their
specifications:
100……120 V 0.2 A (slow-blow) Metrohm-No. U.600.0009
220……240 V 0.1 A (slow-blow) Metrohm-No. U.600.0006
Insert fuses
5
Change both fuses if necessary and reinsert in fuse holder 1010.
Install fuse holder
6
Push fuse holder 1010 back into the opening of 752 Pump Unit by
hand until it clicks into place properly.
752 Pump Unit8
1111
2.2 Mains connection
100 – 120 V220 – 240 V
1010 Fuse holder
2.2.2 Fuses
1010
230
115
1111 Voltage selection
insert with display
1212
of voltage
1212 Mains switch
1515
1515 Mains connection
plug
Fig. 3: Setting the mains voltage
Two fuses 0.2 A/slow-blow for 100…120 V or 0.1 A/slow-blow for
220…240 V are installed in the fuse holder 1010 of the 752 Pump Unit as
standard.
Ensure that the instrument is never put into operation with fuses of
another type, otherwise there is danger of fire!
For checking or changing fuses, proceed as described in section 2.2.1.
2.2.3 Mains cable and mains connection
Mains cable
The instrument is supplied with one of three mains cables:
• 6.2122.020 with plug SEV 12 (Switzerland, …)
• 6.2122.040 with plug CEE(7), VII (Germany, …)
• 6.2133.070 with plug NEMA 5-15 (USA, …)
which are three-cored and fitted with a plug with an earthing pin. If a
different plug has to be fitted, the yellow/green lead (IEC standard)
must be connected to protective earth (protection class 1).
Any break in the earthing inside or outside the instrument can make it
a hazard!
Mains connection
Plug the mains cable into mains connection plug 1515 of the 752 Pump
Unit (see Fig. 3).
2.2.4 Switching the instrument on/off
The 752 Pump Unit is switched on and off using mains switch 1212. When
the instrument is switched on lamp 33 lights up.
752 Pump Unit9
2 Installation
2.3 Connection to the IC System
2.3.1 Electrical connection to the 732 IC Detector
In order to allow remote control via 732 IC Detector the 752 Pump Unit
must be connected to the "Remote" connection of the 732 IC Detector
with the 6.2143.200 cable as shown in Fig. 4. To allow the remote control to operate, switch 22 "REMOTE" on the front of the 752 Pump Unit
must be set to "ON".
If a second, additional instrument (e.g. 752, 753, 754) is to be connected to the Remote interface of the 732 IC Detector then the
6.2143.220 cable (available as an option) should be used to connect
the 752 Pump Unit instead of the 6.2143.200 cable.
732
752
Fig. 4: Connection 752 Pump Unit – 732 IC Detector
2.3.2 Connection of the suppressor module
The Metrohm Suppressor Module MSM built into the 2.733.0X30 IC
Separation Center is described in detail in section 2.8.3 of the 732/733Manual. The 752 Pump Unit is used to supply regeneration and rinsing
solutions to the suppressor module. To avoid contamination of the
suppressor module by foreign particles or bacterial growth, it is advantageous to install an in-line filter between the 752 Pump Unit and the
inlet capillaries of the suppressor module. The two 6.2821.100 Filterunits PEEK (see section 2.6.3 of the 732/733 Manual) supplied with
the 2.733.0X30 IC Separation Center are eminently suitable for this purpose.
Connection of the suppressor module to the 752 Pump Unit is carried
out as described in section 2.8.6 of the 732/733 Manual . The most important points are repeated here; underlined numbers #### refer to op-
erating elements in the 732/733 Manual:
6.2143.200
Connect column to injector
1
see section 2.8.6 of 732/733 Manual
Rinse column
2
see section 2.8.6 of 732/733 Manual
752 Pump Unit10
2.3 Connection to the IC System
Eluent
H2O
Detector
Waste
Waste
H
SO
8989Suppressor inlet
capillary for eluent
8989
94949090
1
2
3
9191
9393
9292
Fig. 5: Connections at suppressor module
Connect column to suppressor module
3
• Cut inlet capillary 8989 marked with "Eluent" of suppressor
module 9595 (see Fig. 17 in 732/733 Manual) to the desired
length using a sharp cutting tool (e.g. razor blade).
• Screw inlet capillary 8989 to outlet end of separating column 7676
using a 6.2744.014 Compression fitting.
9090Suppressor inlet
capillary for H2SO
9191Suppressor outlet
capillary for H2SO
9292Suppressor outlet
capillary for H2O
9393Suppressor inlet
capillary for H2O
9494Suppressor outlet
capillary for eluent
4
4
Fix column
4
• Insert one or two column holders 8484 (6.2027.030, 6.2027.040
or 6.2027.050) in mounting rails 8383 and fasten separating
column in the column holder.
Connect suppressor module to detector block
5
• Cut outlet capillary 9494 marked with "Detector" of suppressor
module 9595 to the desired length using a sharp cutting tool
(e.g. razor blade).
• Screw outlet capillary 9494 marked with "Detector" to coupling
9696 usinga 6.2744.014 Compression Fitting.
• Screw inlet capillary 8282 of detector block 8181 to other end of
coupling 9696.
Fix connection suppressor – detector block
6
• Insert one of the column holders 8484 (6.2027.030, 6.2027.040
or 6.2027.050) in mounting rail 8383 and fasten coupling 9696 in
the column holder.
Attach pump tubings to 752 Pump Unit
7
• Loosen both tubing cartridges 44 from the holding clamp 66 by
pressing down snap-action lever 77 and remove from mount-
ing pin 99 on Pump Unit 752 (see Fig. 1).
55Contact pressure lever2121Connector (6.2824.110) with filter
77Snap-action lever2222Housing for 6.2824.100 Filter unit
PEEK
1616PTFE tubing (6.1803.020)2323Connector
1717Coupling (6.2744.030)9090Suppressor inlet capillary for
H2SO
4
1818Pump tubing (6.1826.050)9393Suppressor inlet capillary for H2O
1919Stopper (white-yellow)
• Press contact pressure lever 55 on both tubing cartridges
down as far as it will go.
• Insert a length of pump tubing 1818 (6.1826.050) into each of
the tubing cartridges as shown in Fig. 6. The white-yellow
stopper 1919 must click into the corresponding holder on the
left-hand side of the tubing cartridge.
• Place the tubing cartridges on mounting pin 99 and press
down on the right-hand side until snap-action lever 77 clicks
into position on holding clamp 66. Take care that no kinks are
formed in the pump tubing.
Suppressor connection 2: H2SO
8
4
• Loosen rotary nipple screwed onto the interior side of connection 2727. Pull inlet capillary 9090 marked with "H2SO4" (see
Fig. 17 of 732/733 Manual) by hand out of the opening of
connection 2727 as far as required. Retighten nipple on the in-
terior side of connection 2727 to fix inlet capillary 9090.
752 Pump Unit12
2.3 Connection to the IC System
• Attach inlet capillary 9090 using compression fitting 2020
(6.2744.014) to the connector 2323 (see Fig. 6) of the filter unit
PEEK (see section 2.6.3 of 732/733 Manual).
• Attach a piece of PTFE tubing 1616 cut to the required length
using a compression fitting 20 20 (6.2744.014) to the connector2121 with filter at the other end of the filter unit PEEK.
• Attach the other end of the PTFE tubing 1616 using a compres-
sion fitting 2020 (6.2744.014) to the coupling 1717 (6.2744.030)
and mount it on the outlet end of the first length of pump
tubing 1818.
• Mount a coupling 17 17 (6.2744.030) to the inlet end of the first
length of pump tubing 1818. Attach a piece of the PTFE tubing1616 (6.1803.020) cut to the required length using a compression fitting 2020 (6.2744.014) to the other end of this coupling.
• Immerse the other end of the aspirating tubing in a vessel
containing regeneration solution (normally 20 mmol/L H2SO4)
and fix in place.
• Pull outlet capillary 9191 marked with "Waste" of the suppressor
module through the rear panel opening 4343, lead it to a suffi-
ciently large waste container and fix it in place.
Suppressor connection 3: H2O
9
• Loosen rotary nipple screwed onto the interior side of connection 2828. Pull inlet capillary 9393 marked with "H2O" (see
Fig. 17 of 732/733 Manual) by hand out of the opening of
connection 2828 as far as required. Retighten nipple on the in-
terior side of connection 2828 to fix inlet capillary 9393.
• Attach inlet capillary 9393 using compression fitting 2020
(6.2744.014) to the connector 2323 (see Fig. 6) of the second
filter unit PEEK (see section 2.6.3 of 732/733 Manual).
• Attach a piece of PTFE tubing 1616 cut to the required length
using a compression fitting 20 20 (6.2744.014) to the connector2121 with filter at the other end of the filter unit PEEK.
• Attach the other end of the PTFE tubing 1616 using a compres-
sion fitting 2020 (6.2744.014) to the coupling 1717 (6.2744.030)
and mount it on the outlet end of the second length of pump
tubing 1818.
• Mount a coupling 17 17 (6.2744.030) to the inlet end of the
second length of pump tubing 1818. Attach a piece of the PTFE
tubing 1616 (6.1803.020) cut to the required length using a
compression fitting 2020 (6.2744.014) to the other end of this
coupling.
• Immerse the other end of the aspirating tubing in a vessel
containing rinsing solution (normally dist. H2O) and fix in
place.
• Pull outlet capillary 9292 marked with "Waste" of the suppressor
module through the rear panel opening 4343, lead it to a suffi-
ciently large waste container and fix it in place.
752 Pump Unit13
2 Installation
Startup of 752 Pump Unit
10
• Set switch <REMOTE> 22 on 752 Pump Unit to "OFF".
• Switch on 752 Pump Unit with mains switch 1212 and set switch
<PUMP> 33 to "ON".
• Adjust the contact pressure for both tubing cartridges: press
contact pressure lever 55 upwards until the solutions just start
to be drawn in. Then press the contact pressure lever upwards until it clicks once more to obtain optimal contact pressure.
• Check all capillaries between the 752 Pump Unit, the sup-
pressor module and the waste container for escaping liquid. If
liquid escapes anywhere the corresponding compression fitting must be tightened further or changed.
Pump tubings are consumable material with a lifetime which depends
on the contact pressure. This is why the tubing cartridges should be
raised completely by loosening snap-action lever 77 on the right-hand
side if the pump is to remain switched off for a considerable length of
time (the set contact pressure remains unchanged).
752 Pump Unit14
3 Operation
3.1 Manual operation
The 752 Pump Unit cannot be manually operated unless the installation has been carried out properly as described in section 2 (mains
connection, attaching the pump tubing, connection to the IC system).
Switch instrument on/off
The 752 Pump Unit is switched on and off using mains switch 1212
on the rear of the instrument (see Fig. 2):
IInstrument switched on
0Instrument switched off
3.1 Manual operation
POWER
After the instrument has been switched on the power lamp 33
lights up to show that the instrument is ready for use.
Switch off remote control
REMOTE
ON
OFF
In order to allow manual operation the remote control must be
switched off.
ONRemote control switched on
OFFRemote control switched off
Switch pump on/off
PUMP
ON
OFF
The drive of Pump Unit 752 is switched on and off with the switch
<PUMP>:
ONPump switched on
OFFPump switched off
3.2 Operation via 732 IC Detector
The 752 Pump Unit cannot be remotely operated via the IC Detector
unless the installation has been carried out properly as described in
section 2 (mains connection, connection cable 732 – 752, attaching
the pump tubing, connection to the IC system).
Switch instrument on/off
The 752 Pump Unit is switched on and off using mains switch 1212
on the rear of the instrument (see Fig. 2):
IInstrument switched on
0Instrument switched off
POWER
752 Pump Unit15
After the instrument has been switched on the power lamp 33
lights up to show that the instrument is ready for use.
3 Operation
REMOTE
ON
OFF
Switch on remote control
In order to allow remote operation via 732 IC Detector the remote
control must be switched on.
ONRemote control switched on
OFFRemote control switched off
Automatic pump switch-on when the 732 IC
Detector is switched on
In order to start up the drive of the 752 Pump Unit automatically
when the 732 IC Detector is switched on alterations must be
made to the basic settings (Setup, see section 4.4.1 of 732/733Manual). The remote output lead 1 must be set to 1 (on, active,
0 V):
>SETUP/output
remote 10000000
When the 732 IC Detector is switched off the pump is also
switched off.
Switch pump on/off in a program
In the 732 IC Detector the <PROGRAM> key can be used to
create time programs with a maximum of 20 program steps (see
section 4.7.1 of 732/733 Manual). The pump can be switched on
or off at any program step by setting the remote output lead 1 to
1 (on, active, 0 V) or 0 (off, inactive, open) respectively:
Switch on:
Switch off:
>PROGRAM/edit XXX.X min
remote 1TTTTTTT
>PROGRAM/edit XXX.X min
remote 0TTTTTTT
Switch pump on/off at an event
In the 732 IC Detector the <EVENT> key can be used to
program a maximum of 4 different events (see section 4.7.3 of732/733 Manual). The pump can be switched on or off at any
event by setting the remote output lead 1 to 1 (on, active, 0 V) or
0 (off, inactive, open) respectively:
Switch on:
EVENT YY-MM-DD HH:MM:SS
remote 1TTTTTTT
Switch off:
EVENT YY-MM-DD HH:MM:SS
remote 0TTTTTTT
752 Pump Unit16
4.1 Insufficient or non-existing flow rate
4 Troubleshooting
4.1 Insufficient or non-existing flow rate
If the 752 Pump Unit is supplying insufficient liquid or no liquid at all
then this can be remedied by proceeding as follows:
Set contact pressure correctly
1
• Press contact pressure lever 55 on the tubing cartridge down
as far as it will go.
• Press contact pressure lever 55 upwards until the solution just
starts to be drawn in. Then press contact pressure lever upwards until it clicks once more to obtain optimal contact pressure.
Exchange the pump tubing
2
• If the delivery performance is still not satisfactory then the
pump tubing must be exchanged (see section 4.3).
• After the pump tubing has been exchanged press contact
pressure level 55 on the tubing cartridge down as far as it will
go.
• Press contact pressure level 55 upwards until the solution just
starts to be drawn in. Then press contact pressure lever upwards until it clicks once more to obtain optimal contact pressure.
Check the connections
3
• If the delivery performance is still not satisfactory even with
the new pump tubing then the connections must be checked.
• Loosen the first connection after the tubing cartridge in the
supply direction. If liquid is still flowing here then retighten the
connection and release the following connection and so on.
• Replace the connection at which liquid transport ceases or
exchange the connection pieces.
• If liquid transport ceases after the Filter unit PEEK, replace
filter 2121 (6.2824.110; see Fig. 6).
• If liquid transport ceases after the suppressor module then
this should be exchanged.
4.2 Leaking of tubing nipples
If liquid is leaking from the tubing nipples then these must be tightened
up further. If the fault cannot be cured in this way then the tubing nipples must be replaced.
752 Pump Unit17
4 Troubleshooting
4.3 Air bubbles in pump circuit
If air bubbles can be seen in the pump circuit then the aspirating tubing
must be immersed completely in the liquid. If this has no effect then the
connections must be checked (see section 4.2).
4.4 Exchanging the pump tubing
Pump tubings are consumable material with a limited lifetime and
should be exchanged at regular intervals (approx. every 2 weeks under
continuous use).
The working life of pump tubing depends to a considerable extent on
the contact pressure. This is why the contact pressure must be correctly set as described in section 2.3.2. If the pump is to remain
switched off for a lengthy period of time the tubing cartridges should be
raised completely by loosening snap-action lever 77 on the right-hand
side (the set contact pressure remains unchanged).
As the pump is always operated on the same side the pump tubings
6.1826.050 supplied can be used on both sides. To exchange a pump
tubing proceed as follows:
Remove old pump tubing
1
• Press contact pressure lever 55 on the tubing cartridge down
as far as it will go.
• Release tubing cartridge 44 from holding clamp 66 by pressing
down snap-action lever 77 and remove from mounting pin 99
on the 752 Pump Unit (see Fig. 1).
• Remove old pump tubing.
Insert new pump tubing
2
• Insert the new pump tubing 1818 (6.1826.050) in the tubing
cartridge as shown in Fig. 6. The white-yellow stopper 1919
must click into the corresponding holder on the left-hand side
of the tubing cartridge.
• Place the tubing cartridge on mounting pin 99 and press down
on the right-hand side until snap-action lever 77 clicks into po-
sition on holding clamp 66. Take care that no kinks are formed
in the pump tubing.
Set contact pressure
3
• Press contact pressure lever 55 upwards until the solution just
starts to be drawn in. Then press contact pressure lever upwards until it clicks once more to obtain optimal contact pressure.
752 Pump Unit18
5 Appendix
5.1 Technical data
Pump
Pump type2-channel peristaltic pump with rotational speed
Pump capacitytypically 0.5…0.6 mL/min with
Pressuremax. 4 bar (0.4 MPa)
Pumpable liquidsClear liquids with no solid contents
Pump tubing materialPVC (Tygon)
5.1 Technical data
of
20/min (50 Hz)
24/min (60 Hz)
6.1826.050 pump tubing
Mains connection
Mains voltage115 V: 100...120 V ± 10 %
230 V: 220...240 V ± 10 %
Switchable with voltage selection insert in fuse
holder (see section 2.2.1)
Mains frequency50...60 Hz
Power consumption12 VA
Fuse5 mm Ø, 20 mm length
100…120 V: 0.2 A (slow-blow)
220…240 V: 0.1 A (slow-blow)
Remote interface
Purpose
Remote control of pump (switching on/off)
Pin assignment
330
+5V
external752
Pin
current flows:on
current does not flow: off
2
+
-
5
Safety specifications
Construction / TestingAccording to IEC 1010 / EN 61010 / UL 3101-1,
protection class 1, degree of protection IP40
Safety directionsThe Instructions for Use include information and
warnings to which the user must pay attention in
order to assure safe operation of the instrument.
Material of coverPolyurethane rigid foam (PUR) with fire protection
for fire class UL94VO, CFC-free
Material of baseSteel, enamelled
Dimensions
Width260 mm
Height129 mm
Depth366 mm
Weight4.2 kg (incl. accessories)
752 Pump Unit20
5.2 Standard equipment
1.57
0.97
Subject to changes !
All dimensions are given in mm.
The 2.752.0010 Pump Unit includes the following parts:
Quant.Order No.Description
16.1803.020PTFE capillary tubing
Length = 5 m
5.2 Standard equipment
26.1826.050Pump tubing
made of PVC (Tygon); with 2 firmly
attached white-yellow stoppers;
i.d. = 0.57 mm, e.d. = 2.27 mm
16.2143.200Connecting cable
Connecting cable 732 IC Detector –
752 Pump Unit
16.2122.0X0Mains cable
to customer's specifications:
Cable socketCable connector
Type IEC 320/C 13Type SEV 12 (CH…) ...............................6.2122.020
Type IEC 320/C 13Type CEE (7), VII (D…)...........................6.2122.040
Type CEE (22), VType NEMA 5-15 (USA…) ......................6.2122.070
26.2744.014PEEK compression fitting
For connection of 6.1831.010 PEEK
capillaries or 6.1803.020 PTFE capillaries.
Set of 2
150
400
25 pol.
2 m
26
16.2744.030PEEK Coupling
Connection between 6.2744.010 PEEK compression fitting and 6.1826.0X0 pump tubing.
Set of 4
25
18.752.1013Instructions for Use (English)
for 752 Pump Unit
5.3 Optional accessories
6.2143.220Connecting cable
Connecting cable 732 IC Detector –
752 Pump Unit – 753 Suppressor Module
752 Pump Unit21
5 Appendix
5.4 Warranty and conformity
5.4.1 Warranty
The warranty on our products is limited to defects that are traceable to
material, construction or manufacturing error which occur within 12
months from the day of delivery. In such cases the defects will be rectified in our workshops free of charge. Transport costs are to be paid by
the customer.
For day and night operation the warranty is limited to 6 months.
Glass breakage in the case of electrodes or other parts is not covered
by the warranty. Checks which we are asked to carry out during the
warranty period for reasons other than material or manufacturing faults
will be invoiced. For parts manufactured by third parties, insofar as
these constitute an appreciable part of our instrument, the warranty
stipulations of the manufacturer in question apply.
With the regard to the guarantee of accuracy, the technical specifications in the instruction manual are authoritative.
With regard to defects in material, construction or design as well as the
absence of guaranteed features, the purchaser has no rights or claims
except those mentioned above.
If damage of the packaging is evident on receipt of a consignment or if
the goods show signs of transport damage after unpacking, the carrier
must be informed immediately and a written damage report demanded.
Lack of an official damage report releases Metrohm from any liability to
pay compensation.
If any instruments and parts have to be returned, the original packaging
should be used if at all possible. This applies above all to instruments,
electrodes, burette cylinders and PTFE pistons. Before embedding
them in wood shavings or similar material, the parts must be packed in
a dustproof package (for instruments the use of a plastic bag is imperative). If open assemblies are enclosed in the scope of delivery that
are sensitive to electromagnetic voltages (e.g. data interfaces etc.)
these must be returned in the associated original protective packaging
(e.g. conductive protective bag). (Exception: assemblies with built-in
voltage source belong in a non-conductive protective packaging). For
damage which arises as a result of non-compliance with these instructions, no warranty responsibility whatsoever will be accepted by
Metrohm.
752 Pump Unit22
5.4.2 EU Declaration of conformity
EU Declaration of Conformity
The METROHM AG company, Herisau, Switzerland hereby certifies, that the instrument:
752 Pump Unit
meets the requirements of EC Directives 89/336/EWG and 73/23/EWG.
5.4 Warranty and conformity
Source of the specifications:
EN 50081-1Electromagnetic compatibility, basic specification
Emitted Interference
EN 50082-1/2Electromagnetic compatibility, basic specification
Interference Immunity
EN 61010Safety requirements for electrical laboratory measurement
and control equipment
Description of the instrument:
Two-channel peristaltic pump with remote control
Herisau, September 27, 1996
Dr. J. FrankCh. Buchmann
Development ManagerProduction and
Quality Assurance Manager
752 Pump Unit23
5 Appendix
5.4.3 Certificate of conformity and system validation
Certificate of Conformity and System Validation
This is to certify the conformity to the standard specifications for electrical appliances and accessories, as well as to the standard specifications for security and
to system validation issued by the manufacturing company.
Name of commodity: 752 Pump Unit
Name of manufacturer:Metrohm Ltd., Herisau, Switzerland
Principal technical information:Voltages:100…120, 220…240 V
Frequency:50…60 Hz
This Metrohm instrument has been built and has undergone final type testing
according to the standards:
It has also been certified by the Swiss Electrotechnical Association (SEV), which
is member of the International Certification Body (CB/IEC).
The technical specifications are documented in the instruction manual.
Metrohm Ltd. is holder of the SQS-certificate of the quality system ISO 9001 for
quality assurance in design/development, production, installation and servicing.