The 752Pump Unit is a 2-channel peristaltic pump which can be used
to pump two solutions simultaneously. It is particularly suitable for operating the suppressor module in the 2.733.0X30 IC Separation Center
by supplying regeneration and rinsing solutions. This application is described in detail in these Instructions for Use.
The 752 Pump Unit can either be operated manually by means of the
on/off switch for the pump or remotely controlled via the remote interface.
1.1 Instrument description
752 Pump Unit1
1 Introduction
1.2 Parts and controls
334455
Pump Unit752
POWER
REMOTEPUMP
ONON
OFFOFF
221166889977
Fig. 1: Front of the 752 Pump Unit
MetrohmMetrohm
11Pump ON/OFF
On/Off switch for the pump
(manual operation mode)
22Remote ON/OFF
On/Off switch for remote control
via remote interface
33Power on lamp
Lights up when instrument is
switched on
44Tubing cartridge
For 6.1826.050 Pump tubing
55Contact pressure lever
For adjusting the contact pressure
66Holding clamp
For locking the tubing cartridge into
place
77Snap-action lever
For releasing the tubing cartridge
88Pump drive
Roller head with contact rollers
99Mounting pin
For attaching the tubing cartridge
752 Pump Unit2
1.2 Parts and controls
1010
WARNING - Fire Hazard -
For continued protection replace only
with the same type and rating of fuse
f=50-60 Hz
S=12 VA
115V: 110-120V:
230V: 220-240V:
Type 1.752.0010 Nr.
Fuse
0,2A(T)
0,1A(T)
S
131312121111
Remote
Pump
Made by Metrohm Herisau Switzerland
15151414
Fig. 2: Rear of the 752 Pump Unit
1010Fuse holder
Changing the fuses, see section 2.2
1111Voltage selection insert with
display of voltage
1212Mains switch
For switching the instrument on/off:
I = ON0 = OFF
1313Serial number
1414Remote interface
Remote I/O lines for connection of
external devices (e.g. 732 IC Detector)
1515Mains connection plug
Mains connection, see section 2.2
752 Pump Unit3
1 Introduction
1.3 Information about the Instructions for Use
Please read through these Instructions for Use carefully before you put
the 752 Pump Unit into operation. The Instructions for Use contain
information and warnings to which the user must pay attention in order
to assure safe operation of the instruments.
1.3.1 Organization
These 8.752.1013Instructions for Use for the 752 Pump Unit provide
a comprehensive overview of the startup procedure, operation, fault
rectification and technical specifications of these instruments. The Instructions for Use are organized as follows:
Section 1Introduction
General description of instruments, parts and controls
and safety notes
Section 2Installation
Attachment of accessories, connection to IC system
Section 3Operation
Manual operation and operation via 732 IC Detector
Section 4Maintenance – Faults
Maintenance, fault rectification
Section5Appendix
Technical data, standard equipment, options, warranty,
declarations of conformity, index
To find the required information on the instruments please use either
the Table of contents or the Index at the back.
752 Pump Unit4
1.3.2 Notation and pictograms
The following notations and pictograms (symbols) are used in these Instructions for Use:
<PUMP>Switch or key
1515Part or control of 752
8989Part or control of 732/733
"range"Parameter or entry value
1.3 Information about the Instructions for Use
at 732 IC Detector
>PARAM/detector
range: 1.00 mS/cm
Display
Text in display 11 of the 732 IC
Detector
Hazard
This symbol draws attention to a
possible danger to life or of injury if
the associated directions are not
followed correctly.
Warning
This symbol draws attention to
possible damage to instruments or
instrument parts if the associated
directions are not followed correctly.
Caution
This symbol marks important information. First read the associated
directions before you continue.
Comment
This symbol marks additional information and tips.
752 Pump Unit5
1 Introduction
1.4 Safety notes
1.4.1 Electrical safety
While electrical safety in the handling of the 752 Pump Unit is assured
in the context of the specifications IEC 1010-1 (protection class 1, degree of protection IP40), the following points should be noted:
• Mains connection
Setting the mains voltage, checking the mains fuse and the mains
connection must be effected in accordance with the instructions in
section 2.2.
• Opening the 752 Pump Unit
When the 752 Pump Unit is connected to the power supply the instrument must neither be opened nor should parts be removed from
it, otherwise there is a danger of coming into contact with components which are live. Always disconnect the instrument from all voltage
sources before you open it and ensure that the mains cable isdisconnected from mains connection 1515 !
• Protection against static charges
Electronic components are sensitive to static charging and can be
destroyed by discharges. Before you touch any of the components
inside the 752 Pump Unit, you should earth yourself and any tools you
are using by touching an earthed object (e.g. housing of the instrument or a radiator) to eliminate any static charges which exist.
1.4.2 General precautionary rules
• Handling of solvents
Check all pump tubings and all input and output lines periodically for
possible leaks. Follow the relevant instructions regarding the handling
of flammable and/or toxic solvents and their disposal.
• Regular exchange of pump tubings
Pump tubings are consumable material and must be replaced from
time to time (see section 4.4). Suitable measures must be taken so
that any leak which might occur in the pump tubing or connections
during unattended operation will cause no damage (placing the
instrument at the bottom, collection device for any liquid which may
leak out).
752 Pump Unit6
Loading...
+ 20 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.