Metrohm 750 User Manual

CH-9101 Herisau/Switzerland Tel. ++41 71 353 85 85 Fax ++41 71 353 89 01 Internet www.metrohm.ch E-Mail info@metrohm.ch
750 Autosampler
2.750.0020 Instrument version
8.750.1023 Instructions for Use
Table of contents
Table of contents
1 Introduction............................................................................................1
1.1 Instrument description.............................................................................1
1.2 Theory of operation ..................................................................................2
1.3 Control elements.......................................................................................4
1.4 Information on the Instructions for Use................................................6
1.4.1 Organization.................................................................................... 6
1.4.2 Notation and pictograms ................................................................7
1.5 Safety notes ...............................................................................................8
1.5.1 Electrical safety ...............................................................................8
1.5.2 General precautionary rules............................................................ 8
2 Installation..............................................................................................9
2.1 Setting up the instrument........................................................................ 9
2.1.1 Packaging .......................................................................................9
2.1.2 Check..............................................................................................9
2.1.3 Location...........................................................................................9
2.2 Connection to IC system........................................................................10
2.2.1 750 Autosampler as "Master"........................................................10
2.2.2 732 IC Detector as "Master".......................................................... 11
2.2.3 Connection to 762 with «IC Net»................................................... 12
2.2.4 Tubing connection ........................................................................14
2.3 Mains connection....................................................................................15
2.3.1 Check fuses ..................................................................................15
2.3.2 Mains cable and mains connection.............................................. 16
2.3.3 Switching the instrument on/off ....................................................16
2.4 Initial checkout ........................................................................................17
3 Operation............................................................................................... 19
3.1 Fundamentals of operation ...................................................................19
3.1.1 Display and instrument dialog...................................................... 19
3.1.2 Overview of key functions............................................................. 20
3.2 System settings.......................................................................................23
750 Autosampler
3.3 Manual rinsing.........................................................................................26
3.4 Manual injection......................................................................................26
3.5 Methods ....................................................................................................28
3.6 Method procedure...................................................................................34
3.7 Priority sample.........................................................................................35
3.8 Remote control........................................................................................36
I
Inhaltsverzeichnis
4 Notes – Maintenance – Faults...................................... 39
5 Appendix ................................................................................................ 49
4.1 Operating information............................................................................ 39
4.1.1 Loading sample vials.................................................................... 39
4.1.2 Sample preparation ...................................................................... 39
4.1.3 Operating parameters................................................................... 40
4.2 Maintenance and servicing ................................................................... 41
4.2.1 General information ...................................................................... 41
4.2.2 Transfer needle replacement........................................................ 41
4.2.3 Sample preparation ...................................................................... 42
4.3 Remedying malfunctions....................................................................... 43
4.3.1 Non-reproducible injections ......................................................... 43
4.3.2 Missed injections.......................................................................... 44
4.4 Error messages.......................................................................................45
4.4.1 Entry error messages.................................................................... 45
4.4.2 System error messages................................................................ 47
4.4.3 Remote control error codes.......................................................... 48
5.1 Technical data .........................................................................................49
5.2 Interfaces.................................................................................................. 51
5.2.1 Contact closure strips................................................................... 51
5.2.2 RS232 interface............................................................................. 53
5.3 Standard equipment............................................................................... 54
5.4 Optional accessories.............................................................................. 55
5.5 Warranty and conformity....................................................................... 56
5.5.1 Warranty........................................................................................ 56
5.5.2 EU Declaration of conformity........................................................ 57
5.5.3 Certificate of conformity and system validation............................ 58
5.6 Index.......................................................................................................... 59
List of figures
Fig. 1: Autosampler 750 ............................................................................................... 1
Fig. 2: Positive displacement sample injection ............................................................2
Fig. 3: Displacement volume versus needle position .................................................. 3
Fig. 4: Front of 750 Autosampler.................................................................................. 4
Fig. 5: Rear of 750 Autosampler................................................................................... 5
Fig. 6: Connection of 750 Autosampler to 732/733 ................................................... 11
Fig. 7: Connection of 750 Autosampler to 732/733 and 762 ..................................... 13
Fig. 8: Keypad ............................................................................................................ 20
Abb. 9: Manual sample filtering ................................................................................... 39
Fig. 10: Sample filtering with FilterCaps....................................................................... 40
Fig. 11: Transfer needle replacement........................................................................... 51
Fig. 12: Valve control .................................................................................................... 57
Fig. 13: Inputs/outputs.................................................................................................. 51
II
750 Autosampler
1 Introduction
1.1 Instrument description
The 750 Autosampler is a microcomputer-operated sampling/con­troller system designed to be used with the ion chromatographic in­struments 732 IC Detector and 733 IC Separation Center. It has a ca­pacity of up to 128 samples, which are transferred automatically into the sample loop of the 733 IC Separation Center.
The 750 Autosampler performs up to four, fixed loop injections per vial at user-selectable time intervals. All operating parameters are entered from the front panel. Additionally, the instrument can be controlled re­motely either by an externally programmable contact closure or through the RS232 connector located on the rear panel.
1.1 Instrument description
750 Autosampler
Fig. 1: 750 Autosampler
1
1 Introduction
Sample
Needle
Cap
Sample vial
Valve
1.2 Theory of operation
Displacement principle
The 750 Autosampler operates using the positive displacement princi­ple of sample transfer. This sample transfer process into the sample loop of the 733 IC Separation Center is illustrated in Fig. 2. When an in­jection is initiated, the hollow transfer needle 55 descends into the vial. The needle tip punctures the polyethylene cap of the sample vial. The opening into the needle is located on the side to minimize plugging by cap material as the cap is punctured. The needle continues downward until the pushrod contacts the area around the puncture in the cap which forces the cap down into the vial. The piston action of the cap pressurizes the sample and forces it up through the needle and the transfer tubing 88 that is connected to the needle. From the transfer tubing, the sample flows through the injection valve and sample loop with the excess passing on through the waste tube to a waste con­tainer.
Column
Rinsing by air bubbles
The air bubble (minimum 150 µL), which is always present at the top of a properly filled and capped vial, precedes the sample stream through the plumbing. The bubble disrupts the laminar flow of the fluid in the tubing while pushing out the previous sample and solvent. This effect, in addition to the large excess volume of sample, helps to minimize sample carryover.
A B C D
Eluent
Waste
RUN FILL/LOAD
INJECT
RUN
Fig. 2: Positive displacement sample injection
2
750 Autosampler
1.2 Theory of operation
Number of injections
The number of injections per sample with the 750 Autosampler de­pends on the sample loop size. With sample loops > 10 µL max. 3, with sample loops 10 µL max. 4 injections are possible. Fig. 3 details the needle position and displacement volume at the sample delivery action. Care should always be taken that the injection volume is at least three times greater than the volume of the sample loop. This is why only a maximum of 2 injections per sample should be made with 100 µL loops; the minimum sample volume is 300 µL for 1 and 600 µL for 2 in­jections. For sample loops with a volume 10 µL up to 4 injections per sample are possible. The minimum sample volume required is 150 µL for 1, 300 µL for 2, 450 µL for 3 and 600 µL for 4 injections.
Rinsing
Rinsing will be necessary for highly concentrated samples or for sam­ples that may absorb in the tubing or in the injection valve. Rinse vials are filled with eluent or other suitable solvent. The Rinse process is shown in Fig. 3. Rinse vials can be used either after each vial (every even numbered or odd numbered vial) or at programmed vial intervals. There are two types of Rinses, the Normal Rinse and the Quick Rinse. The Normal Rinse occurs after all the injections have been made from the sample vial and the run time has elapsed. A Quick Rinse occurs immediately after an injection and is used in applications where the sample solution must not be left in the needle, transfer tubing or valve for whatever reason. Only one injection per vial can be programmed when using a Quick Rinse.
Injection
1
2
3
4
1 Injection
per vial
2.65 cm 730
µL
2 Injections
per vial
1.63 cm 450
µL
1.02 cm 280
µL
3 Injections
per vial
1.63 cm 450
µL
0.51 cm 140
µL
0.51 cm 140
µL
4 Injections
per vial Rinse
1.12 cm 310
µL
0.51 cm 140
µL
0.51 cm 140
µL
0.51 cm 140
µL
2.65 cm 730
µL
750 Autosampler
Fig. 3: Displacement volume versus needle position
3
1 Introduction
1.3 Control elements
750 Autosampler
MANUAL
METHOD
8 9
7
5 6
4
1
2 3
0
.
11
RINSEINJECTION
FUNCTION
REM/LOCTEMPSYSTEM
RUN/STOP
F3F2F1
PRIORITY
CLEAR
ENTER
22
88
Fig. 4: Front of the 750 Autosampler
11 Display (LCD)
comprising 4 lines each of 20 characters
22 Keypad
with function, numeric and cursor keys (details see chap. 3.1.2)
33
44
556677
55 Transfer needle
66 Holder for transfer tubing
33 Sample rack (6.2041.600)
77 Holder stopper
for 128 sample vials
44 Sample vial
6.2743.000 sample vial (PP) or
6.2413.000 sample vial (glass)
88 Transfer tubing
6.1803.000 PTFE capillary for the connection of the 750 Autosampler with the injector of the 733 IC Separation Center
4
750 Autosampler
1.3 Control elements
1010
99
BCD
INJECT LOAD
IEEE
EXT PWR CONTROL
Fig. 5: Rear of the 750 Autosampler
99 Contact closure strip for valve
control
details see chap. 5.2.1
RS-232
PRINTER
Type: Nr.: U:
1313 Mains switch
switch to switch the instrument on and off:
OFF
Switzerland
Hz
f:
VA
V
S:
I = ON 0 =
1111
1212
1313
1414
1515
1010 Contact closure strip for external
control
details see chap. 5.2.1
1111 RS232 interface
details see chap. 5.2.2
1212 Model plate
with instrument and serial number, mains voltage, mains frequency, power consumption
750 Autosampler
1414 Mains connection plug
mains connection see chap. 2.3
1515 Fuse cassette
changing the fuses, see chap. 2.3.1
5
1 Introduction
1.4 Information on the Instructions for Use
Please read through these Instructions for Use carefully before you put the 750 Autosampler into operation. The Instructions for Use contain information and warnings which must be heeded by the user to assure safe operation of the instrument.
1.4.1 Organization
These 8.750.1023 Instructions for Use for the 750 Autosampler pro­vide a comprehensive overview of the installation, startup procedure, operation, fault rectification and technical specifications of the instru­ment. The Instructions for Use are organized as follows:
Section 1 Introduction
General description of instruments, control parts and safety notes
Section 2 Installation
Installation of the 750 Autosampler, connection to IC Separation Center 733, initial checkout
Section 3 Operation
Detailed description of the operation and explanation of functions of all keys
Section 4 Notes – Maintenance – Faults
Notes on operation, maintenance, fault rectification, diagnostic tests
Section 5 Appendix
Technical data, interfaces, standard equipment, options, warranty, declarations of conformity, index
To find the required information on the instruments, you will find it an advantage to use either the Table of contents or the Index at the back.
6
750 Autosampler
1.4.2 Notation and pictograms
The following notations and pictograms (symbols) are used in these In­structions for Use:
<PARAM> Key
"Run Time" Parameter or entry value
55 Control element of 750
8484 Control element of 732/733
1.4 Information on the Instructions for Use
(see chap. 1.3)
(see Instructions for Use 732)
1414 Control element of 762
READY
SELECT A FUNCTION
(see Instructions for Use 762)
Display
Text in display 11 of the 750 Auto- sampler
Hazard
This symbol draws attention to a possible danger to life or of injury if the associated directions are not followed correctly.
Warning
This symbol draws attention to possible damage to instruments or instrument parts if the associated directions are not followed cor­rectly.
750 Autosampler
Caution
This symbol marks important in­formation. First read the associ­ated directions before you con­tinue.
Comment
This symbol marks additional in­formation and tips.
7
1 Introduction
1.5 Safety notes
1.5.1 Electrical safety
While electrical safety in the handling of the 750 Autosampler is assured in the context of the specifications IEC 1010-1 (protection class 1, de­gree of protection IP20), the following points should be noted:
Mains connection
Setting the mains voltage, checking the mains fuse and the mains connection must be effected in accordance with the instructions in
section 2.3.
Opening the 750 Autosampler
When the 750 Autosampler is connected to the power supply the instrument must neither be opened nor should parts be removed from it, otherwise there is a danger of coming into contact with compo­nents which are live. Always disconnect the instrument from all voltage sources before you open it and ensure that the mains cable is disconnected from mains connection plug 1414 !
Protection against static charges
Electronic components are sensitive to static charging and can be destroyed by discharges. Before you touch any of the components inside the 750 Autosampler, you should earth yourself and any tools you are using by touching an earthed object (e.g. housing of the instrument or a radiator) to eliminate any static charges which exist.
1.5.2 General precautionary rules
Handling of solvents
Check all tubing of the IC system periodically for possible leaks. Follow the relevant instructions regarding the handling of flammable and/or toxic solvents and their disposal.
8
750 Autosampler
2 Installation
2.1 Setting up the instrument
2.1.1 Packaging
The 750 Autosampler is supplied together with the separately packed accessories in special packagings containing shock-absorbing foam linings designed to provide excellent protection. The actual instrument is packed in an evacuated polyethylene bag to prevent the ingress of dust. Please store all these special packagings as only they can assure damage-free transport of the instrument.
2.1.2 Check
2.1 Setting up the instrument
2.1.3 Location
After receipt, immediately check whether the shipment is complete and has arrived without damage (compare with delivery note and list of accessories in section 5.3). In the case of transport damage, see instructions in section 5.5.1 "Warranty".
Position the instrument in the laboratory at a location convenient for op­eration, free from vibrations and protected against a corrosive atmos­phere and contamination by chemicals. The Autosampler should be lo­cated as close as possible to the IC system to minimize dead volume. When not in use the instrument should always be covered with the
6.2742.010 dust cover.
750 Autosampler
9
2 Installation
2.2 Connection to IC system
2.2.1 750 Autosampler as "Master"
When an IC system consisting of 732 IC Detector, 733 IC Separation Center, and 709 IC Pump is operated with the 750 Autosampler as "Master" the command to start the next injection after the analysis time "Run Time" has elapsed will be given by the Autosampler itself. In this case the 750 Autosampler is connected to the 733 IC Separation Cen­ter with the 6.2128.100 cable as shown in Fig. 6. In addition the 709 IC Pump can be connected to the 750 Autosampler in such a way that when the pump stops the Autosampler will also be automatically stopped (important for unattended continuous operation).
Electrical connection 750 – 733
1
Connect 6.2140.020 connector plug to contact closure strip 99
of the 750 Autosampler.
Connect one end of the 6.2128.100 cable to the connections
"LOAD NO", "LOAD COM", "INJECT NO" and "INJECT COM" on the connector plug inserted in the contact closure strip 99. The four cable ends are appropriately inscribed with "LOAD NO" (green), "LOAD COM" (yellow), "INJECT NO" (brown) and "INJECT COM" (white).
Connect the other end of the 6.2128.100 cable to the con-
nections "Ground", "Fill" and "Inject" on the terminal block 3838 (valve A) or 4646 (valve B) of the 733 IC Separation Center. The three cable ends are appropriately inscribed with "GND" (white), "FILL" (green) and "INJECT" (brown).
Electrical connection 750 – 709 (optional)
2
Connect 6.2140.010 connector plug to contact closure strip
1010 of the 750 Autosampler.
Connect jumper P12 to the control interface of the 709 IC
Pump in such a way that the contacts at outputs 3 and 4 will be closed when the pump has stopped (see chap. 5.2 of the 709 Instructions for Use).
Use 6.2134.070 cable (option) to connect outputs 3 and 4 of
the control interface on the 709 IC Pump with connection "REMOTE STOP INPUT" and "GND" on contact closure strip 1010 of the 750 Autosampler.
Settings with 750 Autosampler as "Master"
The following settings are recommended for operating the IC system with 750 Autosampler as "Master":
750 Autosampler: "Run Time" 3.0 min
732 IC Detector: operation without program or
with "Inject" program
10
750 Autosampler
2.2 Connection to IC system
Cable 732/Evaluation System (analog signal)
Evaluation system
Cable 732/Evaluation system (start signal)
732
733
6.2115.070
709
6.2125.090
6.2125.060
6.2134.070
750
6.2128.100
INJECT LOAD
Fig. 6: Connection of 750 Autosampler at 732/733
2.2.2 732 IC Detector as "Master"
When an IC system consisting of 732 IC Detector, 733 IC Separation Center, and 709 IC Pump is operated with the 732 IC Detector as "Master" the 750 Autosampler sample change procedure will be started by a signal from the 733 IC Separation Center. The Autosampler must be connected to the 733 IC Separation Center with two cables as shown in Fig. 6. In addition the 709 IC Pump can be connected to the 750 Autosampler in such a way that when the pump stops the Auto­sampler will also be automatically stopped (important for unattended continuous operation).
Electrical connection 750 – 733
1
Connect 6.2140.020 connector plug to contact closure strip 99
and 6.2140.010 connector plug to contact closure strip 1010of the 750 Autosampler.
Connect one end of the 6.2128.100 cable to the connections
"LOAD NO", "LOAD COM", "INJECT NO" and "INJECT COM" on the connector plug inserted in the contact closure strip 99. The four cable ends are appropriately inscribed with "LOAD NO" (green), "LOAD COM" (yellow), "INJECT NO" (brown) and "INJECT COM" (white).
Connect the other end of the 6.2128.100 cable to the con-
nections "Ground", "Fill" and "Inject" on the terminal block 3838 (valve A) or 4646 (valve B) of the 733 IC Separation Center. The three cable ends are appropriately inscribed with "GND" (white), "FILL" (green) and "INJECT" (brown).
750 Autosampler
11
2 Installation
Connect one end of the 6.2115.070 cable (option) to the
connections "EXTERNAL INJECT INPUT" and "GND" on the connector plug inserted in the contact closure strip 1010.
Connect the other end of the 6.2115.070 cable to the con-
nections "Pos. Fill" on the terminal block 3838 (valve A) or 4646 (valve B) of the 733 IC Separation Center so that COM and GND are connected.
Electrical connection 750 – 709 (optional)
2
Connect jumper P12 to the control interface of the 709 IC
Pump in such a way that the contacts at outputs 3 and 4 will be closed when the pump has stopped (see chap. 5.2 of the 709 Instructions for Use).
Use 6.2134.070 cable (option) to connect outputs 3 and 4 of
the control interface on the 709 IC Pump with connection "REMOTE STOP INPUT" and "GND" on contact closure strip 1010 of the 750 Autosampler.
Settings with 732 IC Detector as "Master"
The following settings are recommended for operating the IC system with 732 IC Detector as "Master":
750 Autosampler: "Run Time" = 0.0 min
732 IC Detector: operation with cycle program with
command "valve A/B = fill" at time "0.0 min".
2.2.3 Connection to 762 with «IC Net»
When a modular IC system consisting of 732 IC Detector, 733 IC Separation Center, 709 IC Pump, and 762 IC Interface is operated with the PC software «IC Net» as "Master" the 750 Autosampler sample change procedure will be started by a signal from the 733 IC Separa­tion Center. The Autosampler must be connected to the 762 IC Inter­face using the 6.2134.000 cable and to the 733 IC Separation Center using the 6.2128.100 as shown in Fig. 7.
Electrical connection 750 – 762
1
Connect RS232 interface 1111 at the 750 Autosampler to one of
the RS232 interfaces 1414 or 1818 at the 762 IC Interface using the 6.2134.000 cable.
12
Electrical connection 750 – 733
2
Connect 6.2140.020 connector plug to contact closure strip 99
of the 750 Autosampler.
Connect one end of the 6.2128.100 cable to the connections
"LOAD NO", "LOAD COM", "INJECT NO" and "INJECT COM" on the connector plug inserted in the contact closure strip 99.
750 Autosampler
2.2 Connection to IC system
The four cable ends are appropriately inscribed with "LOAD NO" (green), "LOAD COM" (yellow), "INJECT NO" (brown) and "INJECT COM" (white).
Connect the other end of the 6.2128.100 cable to the con-
nections "Ground", "Fill" and "Inject" on the terminal block 3838 (valve A) or 4646 (valve B) of the 733 IC Separation Center. The three cable ends are appropriately inscribed with "GND" (white), "FILL" (green) and "INJECT" (brown).
732
6.2128.130
733
6.2125.090 (732)
6.2115.070
750
6.2128.100
6.2134.100
PC
6.2134.090
709762
6.2134.000
INJECT LOAD
Fig. 7: Connection of 750 Autosampler at 732/733 and 762
750 Autosampler
13
2 Installation
2.2.4 Tubing connection
For transferring the sample from the 750 Autosampler to the injection valve of the 733 IC Separation Center the following tubing connections must be made:
Tubing connection 750 – injection valve
1
At the 733 IC Separation Center, loosen the rotary nipple
screwed onto the interior side of connection 2222 or 2828.
Take PTFE suction tubing 8888 (see Fig. 16 and Fig. 17 of the
732/733 Instructions for Use) completely out of connection 2222 or 2828 and unscrew from connection "1" of injection valve 6868.
Cut transfer tubing 88 (6.1803.000 PTFE capillary tubing of 750
accessories) to the shortest possible length between the needle of the 750 Autosampler and injection valve 6868 of the 733 IC Separation Center.
Remove the cover from transfer needle 55 of the 750 Auto-
sampler and screw one end of transfer tubing 88 to this con- nection with the aid of a 6.2744.010 PEEK compression fit­ting.
Remove holder stopper 77 from holder 66 and pull 6.1803.000
PTFE capillary tubing so far through the opening of holder stopper 77 that the tubing length between needle connection and the holder stopper is about 30 cm. Then press holder stopper 77 fully back into holder 66.
Pull the other end of the transfer tubing 88 through the opening
2222 or 2828 of the 733 IC Separation Center and screw onto
connection "1" of injection valve 6868 using a 6.2744.010 PEEK compression fitting.
Retighten rotary nipple on the interior side of connection 2222
or 2828 to fix the capillary tubing.
14
Tubing connection injection valve – waste
2
Insert 6.2744.020 coupling (from 733 accessories) into con-
nection 2121 or 2727 of the 733 IC Separation Center.
Screw PTFE suction tubing 8888 onto the 6.2744.020 coupling
attached to connection 2121 or 2727 and lead it into the waste container.
In the case of the 2.733.0X20 IC Separation Center with two injection valves, it is possible to fill both sample loops from the same 750 Autosampler. For this, connection "1" of valve A (outlet of the sample loop) must be connected to connection "2" of valve B (inlet of the sample loop) using a 6.1803.000 PTFE Capillary tubing cut to the appropriate length.
750 Autosampler
2.3 Mains connection
Follow the instructions below for connecting to the power supply. If the instrument is operated with the mains voltage set wrongly and/or wrong mains fuse there is a danger of fire!
2.3.1 Check fuses
The selection of mains voltage (100, 120, 220 or 240 V) and mains fre­quency (50 or 60 Hz) is made automatically by the 750 Autosampler according to the mains voltage and frequency used. Nevertheless, be­fore the 750 Autosampler is switched on for the first time a check should be made as to whether the correct fuses for the intended mains supply have been inserted in the instrument. Proceed as follows:
Disconnect mains cable
1
Disconnect mains cable from mains connection plug 1414 of the 750 Autosampler (see Fig. 5).
2.3 Mains connection
Remove fuse cassette
2
Using a screwdriver, loosen fuse cassette 1515 alongside the mains connection plug 1414 and take out completely.
Remove fuse holder
3
Lift up the spring clip on the upper side of fuse cassette 1515 and remove the fuse holder completely from the fuse cassette.
Check fuses
4
Carefully take the two fuses out of the fuse holder and check their specifications:
100…120 V 2 A (rapid-action, with low rup-
turing capacity, 3AG)
230…240 V 1.25 A (rapid-action, with low rup-
turing capacity, 5××20 mm)
Insert fuses
5
Change fuses if necessary and reinsert in fuse holder.
Install fuse holder
6
Push the fuse holder into fuse cassette 1515 until the spring clip clicks into position.
750 Autosampler
Insert fuse cassette
7
Completely reinsert fuse cassette 1515 in 750 Autosampler.
15
2 Installation
2.3.2 Mains cable and mains connection
Mains cable
The instrument is supplied with one of three mains cables
6.2122.020 with plug SEV 12 (Switzerland, …)
6.2122.040 with plug CEE(7), VII (Germany, …)
6.2133.070 with plug NEMA 5-15 (USA, …)
which are three-cored and fitted with a plug with an earthing pin. If a different plug has to be fitted, the yellow/green lead (IEC standard) must be connected to protective earth (protection class 1).
Any break in the earthing inside or outside the instrument can make it a hazard!
Mains connection
Plug the mains cable into mains connection plug 1414 of the 750 Auto- sampler (see Fig. 5).
2.3.3 Switching the instrument on/off
The 750 Autosampler is switched on and off using mains switch 1313. When the instrument is switched on display 11 lights up.
16
750 Autosampler
2.4 Initial checkout
Initial checkout will prove the general operation of the instrument and show that no shipping damage or installation problem exists which would prevent normal operation. To carry out this checkout the 750 Autosampler must be connected to the 733 IC Separation Center as described in chap. 2.2 and correctly connected to the mains supply as described in chap. 2.3.
For the initial checkout a standard chromatogram for the separation column used in the 733 IC Separation Center should be recorded. In order to do this a sample vial 44 filled with at least 500 µL of a suitable standard ion solution should be placed in position 1 of sample rack 33; then proceed step by step as follows:
Switch on 750 Autosampler
1
Switch on 750 Autosampler with mains switch 1313. The following screen will be displayed after powerup and initialization:
2.4 Initial checkout
READY
SELECT A FUNCTION
Select "MANUAL INJECTION"
2
Press the <MANUAL INJECTION> key (<FUNCTION> + <8>). The following screen will appear:
MANUAL INJECTION
àà
Sample Vial 1 Needle Depth POS 1 Run Injection ââ
Enter sample vial position
3
Press the <1> key and then press the <ENTER> key. The screen will now change to:
MANUAL INJECTION Sample Vial 1
àà
Needle Depth POS 1 Run Injection ââ
750 Autosampler
Enter needle depth
4
Press the <> key two times and then press the <ENTER> key. This results in the following display:
MANUAL INJECTION Sample Vial 1 Needle Depth POS 3
àà
Run Injection ââ
17
Loading...
+ 45 hidden pages