Leica CM1950 User Manual

Page 1
Operating Manual
Leica CM1950, V 1.1 English – 05/2007 Always keep this manual near the instrument.
Read carefully prior to working with the instrument!
Leica CM1950
Cryostat
Page 2
Page 3
1. NOTE
The information, numerical data, notes and value judgments contained in this manual represent the current state of scientific knowledge and state­of-the-art technology as we understand it follow­ing thorough investigation in this field. We are under no obligation to update the present manu­al periodically and on an ongoing basis accord­ing to the latest technical developments, nor to provide our customers with additional copies, updates etc. of this manual. For erroneous statements, drawings, technical illustrations etc. contained in this manual we ex­clude liability as far as permissible according to the national legal system applicable in each indi­vidual case. In particular, no liability whatsoever is accepted for any financial loss or consequen­tial damage caused by or related to compliance with statements or other information in this manual. Statements, drawings, illustrations and other information as regards contents or technical details of the present manual are not to be considered as warranted characteristics of our products.
These are determined only by the contract provi­sions agreed between ourselves and our customers. Leica reserves the right to change technical spec­ifications as well as manufacturing processes without prior notice. Only in this way is it possible to continuously improve the technology and man­ufacturing techniques used in our products. This document is protected under copyright laws. All copyrights to this documentation are held by Leica Biosystems Nussloch GmbH. Any reproduction of text and illustrations (or of any parts thereof) by means of print, photocopy, microfiche, web cam or other methods – includ­ing any electronic systems and media – requires express prior permission in writing by Leica Biosystems Nussloch GmbH. For the instrument serial number and year of man­ufacture, please refer to the name plate at the back of the instrument.
© Leica Biosystems Nussloch GmbH
Phone: +49 (62 24) 143-0 Fax: +49 (62 24) 143-200 eMail: histo_info@leica-microsystems.com Internet: http://www.histo-solutions.com
Leica CM1950 – Cryostat
3
Page 4
Table of contents
1. Important Information ......................................................................................................................... 6
1.1 Symbols Used in This Manual and Their Meaning ................................................................ 6
1.2 Qualification of Personnel.......................................................................................................... 6
1.3 Designated Use............................................................................................................................ 6
1.4 Instrument Type ........................................................................................................................... 6
2. Safety and Design ................................................................................................................................ 7
2.1 General Safety Instructions ....................................................................................................... 7
2.2 Unpacking and Installation ........................................................................................................ 7
2.3 Safety Features .......................................................................................................................... 10
2.4 Securing/Locking the Handwheel .......................................................................................... 11
2.5 Cleaning, Disinfection – Turning the Instrument Back On.................................................. 12
2.6 Handling Specimens – Defrosting .......................................................................................... 13
2.7 Removal of the Microtome....................................................................................................... 13
2.8 Maintenance .............................................................................................................................. 13
3. Technical Data ................................................................................................................................... 14
4. Standard Delivery .............................................................................................................................. 16
5. General Overview .............................................................................................................................. 18
5.1 Control Panel Fields and Cryostat Chamber.......................................................................... 19
6. Installation ........................................................................................................................................... 20
6.1 Site Requirements ..................................................................................................................... 20
6.2 Transport to the Site ................................................................................................................. 21
6.3 Assembling the Handwheel, Centering the Handwheel (Optional) ................................... 22
6.3.1 Locking/Unlocking the Handwheel......................................................................................... 23
6.3.2 Installing the Footswitch Dummy (Instruments with Cutting Motor) ................................ 23
6.4 Connection to Power Supply System ..................................................................................... 24
6.5 Installing Accessories/Inserting Chamber Accessories .................................................... 25
6.5.1 Installing the Adjustable Footrest (Optional) ........................................................................ 25
6.5.2 Installing the Storage Systems (Optional) ............................................................................. 26
6.5.4 Inserting the Section Waste Tray ...........................................................................................26
6.5.3 Shelf, Movable (Optional)......................................................................................................... 26
6.5.5 Installing the Heat Conductor Block, Stationary (Optional) ............................................... 27
6.5.6 Installing the Knife/Blade Holder and Adjusting the Clearance Angle............................. 27
4
Operating Manual V 1.1 – 05/2007
Page 5
7. Instrument Controls ........................................................................................................................... 30
7.1 Control Panel Fields on the CM1950 – Control Panel Field 1 .............................................. 30
Control Panel Field 2 – Electric Coarse Feed, Sectioning
and Trimming Thickness Display ............................................................................................ 32
Control panel 3 – Motorized sectioning (optional) ............................................................... 34
8. Working with the Instrument ......................................................................................................... 36
8.1 Preparing Cutting Tools, Specimen Discs and Preparation Aids ...................................... 36
8.2 Turning the Instrument On ....................................................................................................... 36
8.3 Configuring the Parameters..................................................................................................... 37
8.4 Working with the Precooled Cryostat .................................................................................... 42
8.4.1 Preparatory Work ...................................................................................................................... 42
8.4.2 Trimming with Extraction – Anti-roll Guide Installed ........................................................... 43
8.4.3 Cutting with Extraction – Anti-roll Guide Installed ............................................................... 45
9. Troubleshooting ................................................................................................................................. 48
9.1 Problems During Work ............................................................................................................. 48
10. Temperature Selection Chart (in minus °C) .............................................................................. 52
11. Optional Accessories ....................................................................................................................... 53
11.1 Ordering Information ................................................................................................................. 53
12. Maintenance and Cleaning ............................................................................................................ 66
12.1 General Maintenance ............................................................................................................... 66
12.2 Changing Fuses .......................................................................................................................... 67
12.3 Replacement of the Lamp/UVC Lamp ..................................................................................... 68
13. Decontamination Certificate (master) ........................................................................................ 70
14. EC Declaration of Conformity ........................................................................................................ 72
15. Warranty and Service ...................................................................................................................... 73
16. Peoles Republic of China ................................................................................................................ 74
Leica CM1950 – Cryostat
5
Page 6
1. Important Information
1.1 Symbols Used in This Manual and Their Meaning
Warnings appear in a gray box and are marked by a warning triangle.
Notes i.e. important user information appear in a gray box and are marked by an in­formation symbol.
Caution – UVC radiation!
(5) (Fig.5)
Figures in brackets refer to item num­bers in drawings or to the drawings themselves.
1.2 Qualification of Personnel
The Leica CM1950 may be operated by trained laboratory personnel only.
All laboratory personnel designated to operate the instrument must carefully read the present instruction manual prior to starting work with the instrument.

1.3 Designated Use

The CM1950 is a high-performance cryostat with an encapsulated microtome and separate speci­men cooling. It features a UV disinfection system, an (optional) integrated extraction system for sec­tion waste and an (optional) motor for motorized sectioning.
The cryostat is designed to produce frozen sections for biological, medical and industrial applications.
The CM1950 is suitable for in-vitro diagnostic (IVD) applications.
The instrument may only be operated within the scope of its designated use as described above and as per the instructions given in this manual.
Any other use of this instrument is considered
as improper operation.

1.4 Instrument Type

All information given in this instruction manual applies only to the instrument type indicated on the title page. A name plate, indicating the instrument serial number, is attached to the back of the instrument.
Despite chemical and/or UV-light dis­infection, personal safety precautions as per the applicable laboratory regu­lations must still be taken (i.e. safety goggles, gloves, laboratory coat and mask must be worn). This type of disinfection reduces the number of germs by at least 99.99%.
Fig. 1
6
Operating Manual V 1.1 – 05/2007
Page 7

2.1 General Safety Instructions

The CM1950 is a cryostat with an encapsulated microtome and separate specimen cooling. It is primarily used for work in the area of fast-cut diag­nostics. The displays and instrument controls are easy to operate due to their largely self-explanatory symbols. LED displays make it easy to read. The cryochamber is made of seamlessly welded, high-quality stainless steel that is free of difficult-to-access corners and thus easy to clean and disinfect. This instrument has been built and tested in accordance with the following safety regulations on electrical measuring, control, regulating and labora­tory devices. In order to maintain this condition and to ensure safe opera­tion, the operator must observe the instructions and warnings contained in this instruction manual. For current information about applicable standards, please refer to the CE declaration of conformity on our Internet site:

2.2 Unpacking and Installation

To ensure proper function of the instrument, it must be set up with a minimum distance on all sides from walls and furniture (see ‘Site Requirements’, p. 20)
The instrument may only be transported in an upright or slightly inclined position.
To ensure a safe transport with a fork lift 3 people are required: one op­erating the fork lift, and the other 2 holding the instrument on either side to prevent it from sliding down.
Before connecting to the power supply system, please check if the local voltage complies with the power rating specified on the name plate of the instrument (see also ‘Technical data’)!
Never connect the instrument to a power socket that does not have a protective conductor terminal. Length of power cable: up to 3.5 m extension possible: no
2. Safety and Design
www.histo-solutions.com
After transporting, wait at least 4 hours before turning the instrument on. This waiting period is necessary to allow the compressor oil, which may have been displaced during transport, to return into its original position. Any condensation on electrical parts that has formed due to tempera­ture differences during transport must be allowed to dry completely. Failure to comply with this can cause severe damage to the instrument!
Leica CM1950 – Cryostat
7
Page 8
2. Safety and Design
Fig. 2
When the instrument is delivered, check the tilt indicators on the packaging. If the arrowhead is blue, the shipment was transported laying flat, was tilted at too great an angle or fell over during transport. Note this on the shipping documents and check the shipment for possible damage.
1
5
5
7
5
4
22
3
8
8
8
8
8
11
4
9
10
11
Opening the packaging
1. Remove the 3 screws (1) on each of the two sides of the box (7) at the top.
9
14
17
Cover = ramp
2. Remove the 2 screws on each of the 4 hinges (2) on the front and rear side of the box.
Do NOT remove the screws of the hinges on the top lid!
3. Remove 1 screw (3) between the two hinges on the front and rear side of the box.
4. Remove the 3 screws (4) at the bottom on all four sides of the box.
5. Of the two vertical rows of screws (5), remove one on either side (which side does not matter) entirely.
Fig. 3
8
Operating Manual V 1.1 – 05/2007
Page 9
Removing the packaging
2. Safety and Design
6
6
1. Remove the cover (17) and place it on the floor with the foam parts facing downwards. - The cover will be used as a ramp later.
2. Remove the 4 foam parts (6) by pulling them upwards.
3. Open the box (7) on its side hinges like an accordion partition and remove it.
66
4. Remove the accessories carton (8) and operating manual on the front of the instrument.
Fig. 4
5. Pull off the dust cover (9) by pulling upwards.
6. Take off the foam strip (10) on the bottom rear.
Ramp assembly
13
12
Fig. 5
1. After the foam strip (10) is removed, 2 screws (12) are visible on the baseplate.
2. Swing out the cover hinges (2) at the front and rear.
3. Attach the cover (17) from the rear on the transport pallet (11). The notches (13) of the hinges (2) must be pointing towards the instrument.
4. Move the cover towards the left so that the notches (13) of the hinges slide beneath the heads of the screws (12).
Transport to the Installation Location
The instrument must be transported in an upright position.
15
16
The cover (17) now forms a ramp on which the instrument can now be rolled down off the transport pallet.
2
14
Fig. 6
Leica CM1950 Cryostat
17
Do not push the device by its hood (16). Use the transport grip
2
points (15 ). The rollers (14) must run over the cover hinges (2) at the front and rear. Danger of tipping!
1. Carefully roll the instrument backwards from the pallet.
2. Push the instrument to the installation location on the castors (14).
9
Page 10
2. Safety and Design

2.3 Safety Features

The instruction manual includes important information related to the operating safety and maintenance of the instrument. The Operating Manual is an important part of the product, and must be read carefully before startup and use and must always be kept near the instrument.
If additional requirements on accident prevention and environmental protection apply in the country of operation, this Operating Manual must be supplemented by appropriate instructions to ensure compliance with such requirements.
The instrument is equipped with the following safety devices: an emergen­cy stop switch (motorized instruments only), a handwheel lock and center­ing system (motorized instruments only), knife guard on the blade and knife holder, and a blade ejector.
The consistent use of these safety features and strict observation of the warn­ings and cautions in this manual, will safeguard the operator from accidents and/or personal injury to a great extent.
To prevent adverse health effects from UV radiation, the UV disinfec­tion cycle can be started only after the sliding window has been prop­erly closed. Closing the window activates the corresponding safety features.
Microtome knives
10
Take care when handling microtome knives/disposable blades. The cutting edge is extremely sharp and can cause severe injury!
Never leave knives and dismounted knife holders with a knife/blade mounted lying around!
Do not place a knife on a table with the cutting edge facing upwards!
We strongly recommend using the safety gloves included with the standard delivery!
Never try to catch a falling knife!
Prior to manipulating the knife and specimen or changing the specimen, always lock the handwheel and cover the cutting edge with the knife guard!
Avoid contact with cold parts of the instrument as this can cause frostbite!
Operating Manual V 1.1 – 05/2007
Page 11
Knife guard
Prior to manipulating the knife and specimen, changing the speci­men or knife, or taking a break, always lock the handwheel and cover the cutting edge with the knife guard.
The CE, CN and CN-Z knife holders feature knife guards; the glass anti-roll plate of the CE knife holder also serves as a knife guard.

2.4 Securing/Locking the Handwheel

Always lock the handwheel and cover the cutting edge with the
12 o’clock position
6 o’clock position
1
2
Fig. 7
To lock the handwheel, press the lever (1) outward. Continue turning the handwheel slowly until the grip is in the upper or lower position and the handwheel is locked. Press the lever fully outward; gently rock the hand­wheel until the locking mechanism clicks into place noticeably.
To release the handwheel, press the lever (2) on the handwheel toward the cryostat housing.
knife guard prior to manipulating the knife or specimen, changing the specimen, or taking a break.
2. Safety and Design
Fig. 8
Fig. 9
Leica CM1950 – Cryostat
Centering of the handwheel (motorized instruments only)
Pull out the handwheel’s handle and position it in the middle of the handwheel. The handle automatically engages in this position.
An important safety device on the cryostat is the centering of the handwheel for motorized instruments.
Rotate the handwheel only if the refrigeration system is on and the cryochamber is cold.
11
Page 12
2. Safety and Design
2.5 Cleaning, Disinfection – Turning the Instrument Back On
It is not necessary to remove the microtome for disinfection.
The instrument has been designed for UV disinfection! Spray disinfec­tion with Leica Cryofect is also possible, thanks to the special insulation of the microtome. (Cryofect is not available in all countries!)
Remove section waste after EVERY sectioning operation and BE­FORE changing specimens. Remove the section waste using a pa­per towel soaked in Cryofect or an alcohol-based disinfectant) or remove the waste using the extraction nozzle (optional). Do not start the disinfection until after folding the section stretcher to the side! Each new specimen is a potential source of contamination.
When disinfecting the instrument, take appropriate protective measures (gloves, mask, protective clothing, etc.).
When using detergents and disinfectants please comply with the safety precautions of the disinfectant manufacturer.
The integrated glass anti-roll guide of the blade holder CE, CN and CN-Z can be cleaned either with acetone or alcohol.
Dispose of waste liquid according to the waste disposal regulations!
Do not use external heaters for drying the cryochamber. This can cause damage to the cooling system!
Do not turn the instrument on before the cryochamber is completely dry! Frost formation!
All components removed from the cryostat must be carefully dried be­fore returning them to the cryochamber!
The front panel and the slit cover of the microtome must be completely dry before turning on the instrument!
12
For more detailed information about disinfection, visit the Leica Microsystems website at
www.histo-solutions.com
Operating Manual V 1.1 – 05/2007
Page 13
2.6 Handling Specimens – Defrosting
When working with contaminated or infected material, the general safety
guidelines for laboratories must be applied.
Before defrosting the cryochamber remove all samples!
Before defrosting the specimen head, remove all samples!
Never leave samples in the cryochamber! - The instrument is not suitable for storing frozen specimens, as the refrigeration dehy­drates the specimens!
The quick freeze shelf can become very hot during the defrosting process. Do not touch it during defrosting.

2.7 Removal of the Microtome

The microtome is encapsulated and therefore does not require removal
by the user.

2.8 Maintenance

Replacing the fuses
Turn the instrument off and disconnect the power plug before replacing
the fuses.
Use only the fuse types specified in chapter 3, "Technical Data." Failure
to comply with this can cause severe damage to the instrument!
2. Safety and Design
Replacing the fluorescent lamp/UVC lamp
Turn the instrument off and disconnect the power plug before replacing
the lamps.
Leica CM1950 Cryostat
It is possible to break the UVC lamp during replacement. If this happens, the lamp change must be completed by Technical Ser­vice. If any metallic mercury is released, handle it carefully and dispose of it properly.
If both disinfection indicator lights are blinking alternately, the UV tube must be replaced!
13
Page 14
3. Technical Data
All specifications related to temperature are valid only up to an ambient temperature of 18°C to 35°C and a relative humidity of no more than 60%.
Type
Nominal voltage (±10 %) Nominal frequency Power draw Max. start-up current for 5 sec. Protective class Circuit breaker: (MDA made by Schurter) Pollution degree Overvoltage installation category Heat emission (max.) Approval
according to IEC-1010; UL 3101
-1 -2 -3
100 VAC 120 VAC 230 VAC 50/60 Hz 60 Hz 50 Hz 1900 VA 1900 VA 1900 VA 35 A eff. 35 A eff. 25 A eff. III 15A M3 15A M3 15A M3
222 II II II 1900 J/s 1900 J/s 1900 J/s CE/c_CSA_us CE/c_CSA_us CE/c_CSA_us
Microtome
Type: Rotary microtome,
encapsulated Section thickness range 1 to 100 μm Trimming range: Clinic 10 - 40μm Research 1 - 600 μm
Total specimen feed 25 mm + 1mm Vertical stroke 59 mm ±0.5mm Specimen retraction 20μm (can be deactivated) Maximum specimen size 50 x 80 mm Cutting speed Slow: 0-50 strokes/min
Fast: 0-85 strokes/min
Vmax: 85-90 strokes/min Specimen orientation 8° (x, y-axis)
For additional details, see p. 32
Electric coarse feed
Slow 300μm/s Fast 900 μm/s
For "Site Requirements," refer to item 6.1 (page 20).
Lamp
50/60 Hz-Version: Osram DULUX L 18 W/840
Cryostat
Dimensions:
Width (w/o handwheel) 700 mm Width (with handwheel) 835 mm Depth (cabinet only) 850 mm Height (total) 1215 mm Working height (armrest) 1025 mm
Weight: Weight (w/motor and suct.) 193 kg Weight (w/motor, w/o suct.) 185 kg Weight (w/o motor, w/ suct.) 183 kg Weight (w/o motor, w/o suct.) 175 kg Weight (w/ specimen head cool.) 165 kg Weight (w/o specimen head cool.) 145 kg
General information
Operating temperature range: 18 °C to 35 °C Temperature range during storage: +5 °C to +55°C Relative humidity: max. 60 %,
non-condensing
Storage humidity: < 60 %
14
Operating Manual V 1.1 – 05/2007
Page 15
3. Technical Data
Refrigeration system CM1950 , 50 Hz CM1950, 60 Hz
Cryochamber
Temperature range 0°C to -35°C ± 5 K, adjustable in 1K increments, 0 °C to -35°C ± 5 K, adjustable in 1K increments
Cooling time to -25°C approx. 5 h approx. 5 h Cooling time to -35°C approx. 8 h approx. 8 h Refrigeration capacity 690 W 690 W Cut-out pressure 25 bar 25 bar Safety factor 3 3 Refrigerant* 300 g (± 5 g) refrigerant R -404A* 300 g (± 5 g) refrigerant R -404A* Compressor oil* 0.6 l EMKARATE RL22S, ICI* 0.6 l EMKARATE RL22S, ICI*
Defrosting of cryochamber
Automatic defrosting
Programmable: Yes (hot gas defrost) Selectable time Yes (hot gas defrost) Selectable time Defrosting intervals 1 defrost in 24 h 1 defrost in 24 h
Defrost time: 12 minutes 12 minutes Automatic shutoff Defrost: at -5 °C chamber temperature at -5 °C chamber temperature
Quick-freeze shelf
Minimum temperature: - 42 °C (+ 5 K), at chamber temp. -35°C - 42 °C (+ 5 K), at chamber temp. -35°C Number of freezing stations: 15+2 15+2 Defrost Manual hot-gas defrost Manual hot-gas defrost
Peltier element
Number of freezing stations: 2 2 Max. temperature difference: 17 K, at chamber temp. of -35°C 17 K, at chamber temp. of -35°C
at ambient temperature of 20°C ambient temperature of 20°C
or manual hot gas defrost or manual hot gas defrost
Specimen cooling
Temperature range -10 to -50 °C ± 3 K -10 to -50 °C ± 3 K
Refrigeration capacity 320 W 320 W Cut-out pressure 25 bar 25 bar Safety factor 3 3 Refrigerant and quantity at 230 V/50 Hz 140 g (± 5 g), refrigerant R-404A* at 120 V/60 Hz 110 g (± 5 g), refrigerant R-404A*
at 100V/50/60 Hz 110 g (± 5 g), refrigerant R-404A*
Compressor oil* 0.4 l alpha 22, Kyodo* 0.4 l alpha 22, Kyodo*
Defrosting of specimen head
Automatic defrost: No No Manual defrost Defrost time 15 min. 15 min.
*) Refrigerant and compressor oil must be replaced by qualified, authorized service personnel only.
acc. to CECOMAF: liquid temperature 45 °C, evaporation temperature: -25 °C
Leica CM1950 – Cryostat
15
Page 16

4. Standard Delivery

Basic instrument WITHOUT motor/WITHOUT extraction, in the specific voltage variant
1 handwheel, manual .........................................................................................0477 41346
5 specimen discs, 30 mm ...................................................................................0477 40044
1 section waste tray ...........................................................................................0477 40062
1 position holder for freeze shelf ..................................................................... 0477 40080
1 freeze shelf cover ............................................................................................0477 43763
1 tool set ............................................................................................................... 0436 43463
- 1 brush, fine ................................................................................................0183 28642
- 1 Leica brush w/magnet ........................................................................... 0183 40426
- 1 Allen key, No. 1.5 ..................................................................................... 0222 10050
- 1 Allen key, No. 2.5 ..................................................................................... 0222 04137
- 1 Allen key, No. 3.0 ..................................................................................... 0222 04138
- 1 Allen key, No. 4.0 ..................................................................................... 0222 04139
- 1 Allen key with dome head, No. 4.0 ....................................................... 0222 32131
- 1 Allen key, No. 5.0 ..................................................................................... 0222 04140
- 1 key with handle, No. 5.0 .........................................................................0194 04760
- 1 Allen key, No. 6.0 ..................................................................................... 0222 04141
- 1 Single-head wrench, No. 13 .................................................................0330 33149
- 1 Single-head wrench, No. 16 .................................................................0330 18595
1 power cable .........................................................................................................................
1 bottle of cryostat oil, type 407, 50 ml ............................................................0336 06098
1 bottle of OCT freezing compound, 125 ml ....................................................0201 08926
1 pair of safety gloves, size M, * for cryosectioning ....................................0340 29011
1 user manual plus language CD ..................................................................... 0708 37115
* Note: for the Japanese version: 100V, 50/60 Hz; 1 pair of safety gloves, size S
(0340 40859) is included.
Basic instrument WITHOUT motor and WITH extraction,
Standard scope of delivery as above, plus:
1 accessory set (extraction) .............................................................................0477 43300
- Hose adapter 1 ...........................................................................................0477 40293
- Hose adapter 2 ...........................................................................................0477 40294
- Suction nozzle ............................................................................................ 0477 40295
- Silicone hose .............................................................................................. 0477 43302
- Silicone stopper .........................................................................................0477 43304
- Chamber suction nozzle ........................................................................... 0477 43779
- Set of filters (5 pieces) .............................................................................. 0477 43792
Compare the delivered components with the parts list and your order. Should you find any discrepancies, please contact your Leica sales office without delay.
A choice of different knife holders is available for the CM1950.
16
Operating Manual V 1.1 – 05/2007
Page 17
4. Standard Delivery
Basic instrument WITH motor/WITHOUT extraction, in the specific voltage variant
1 handwheel, motorized .................................................................................... 0477 41347
5 specimen discs, 30 mm ...................................................................................0477 40044
1 section waste tray ...........................................................................................0477 40062
1 position holder for freeze shelf ..................................................................... 0477 40080
1 freeze shelf cover ............................................................................................0477 43763
1 tool set ............................................................................................................... 0436 43463
- 1 brush, fine ................................................................................................0183 28642
- 1 Leica brush w/magnet ........................................................................... 0183 40426
- 1 Allen key, No. 1.5 ..................................................................................... 0222 10050
- 1 Allen key, No. 2.5 ..................................................................................... 0222 04137
- 1 Allen key, No. 3.0 ..................................................................................... 0222 04138
- 1 Allen key, No. 4.0 ..................................................................................... 0222 04139
- 1 Allen key with dome head, No. 4.0 ....................................................... 0222 32131
- 1 Allen key, No. 5.0 ..................................................................................... 0222 04140
- 1 key with handle, No. 5.0 .........................................................................0194 04760
- 1 Allen key, No. 6.0 ..................................................................................... 0222 04141
- 1 Single-head wrench, No. 13 ................................................................. 0330 33149
- 1 Single-head wrench, No. 16 ................................................................. 0330 18595
1 power cable .........................................................................................................................
1 bottle of cryostat oil, type 407, 50 ml ............................................................0336 06098
1 footswitch dummy ........................................................................................... 0443 30420
1 bottle of OCT freezing compound, 125 ml ....................................................0201 08926
1 pair of safety gloves, size M, * for cryosectioning ....................................0340 29011
1 operating instructions plus language CD .................................................... 0708 37115
* Note: for the Japanese version: 100V, 50/60 Hz; 1 pair of safety gloves,
size S (0340 40859) is included.
Basic instrument WITH and WITH extraction, in the specific voltage variant
Standard scope of delivery as above, plus:
1 accessory set (extraction) .............................................................................0477 43300
- Hose adapter 1 ...........................................................................................0477 40293
- Hose adapter 2 ...........................................................................................0477 40294
- Suction nozzle ............................................................................................ 0477 40295
- Silicone hose .............................................................................................. 0477 43302
- Silicone stopper .........................................................................................0477 43304
- Chamber suction nozzle ........................................................................... 0477 43779
- Set of filters (5 pieces) .............................................................................. 0477 43792
Leica CM1950 – Cryostat
Compare the delivered components with the parts list and your order. Should you find any discrepancies, please contact your Leica sales office without delay.
A choice of different knife holders is available for the CM1950.
17
Page 18

5. General Overview

Cryostat chamber
(with activated
UV disinfection)
Bacterial filter
(optional, only for
instruments with filter)
Emergency-stop
switch (motorized
instruments only)
On/off switch, also circuit breaker
Footswitch dummy (motorized instruments only)
Handwheel in 12 o’clock position
Unscrew and align adjusting feet after
transport
18
Catch tank for condensate
Fig. 10
Castors for safe transport
over short distances
Align adjusting feet for secure upright position
Operating Manual V 1.1 – 05/2007
Page 19
5. General Overview

5.1 Control Panel Fields and Cryostat Chamber

1 - Control panel field 1: Extraction, temperature and time control, illumination, UV disinfection 2 - Control panel field 2: Electric coarse feed (sectioning and trimming thickness adjustment) 3 - Control panel field 3: Motorized sectioning cutting, optional (adjustment of stroke type, cutting speed etc.)
1
7
4
2
6
10a
5
10b
15
16
10
11
10c
14
3
8
9
4 - Heat conductor block, stationary (optional) 5 - Peltier element (with 2 stations) 6 - Freeze shelf, 15 positions 7 - Position holder on freeze shelf 8 - Heat and cold extractor, mobile (opt.) 9 - Shelf, movable (optional) 10 - Blade holder CE with blade ejector (a) 10b- Finger rest on the blade holder CE
Leica CM1950 – Cryostat
13
12
10c- Knife guard on the blade holder CE 11 - Extraction nozzle on the extraction hose 12 - Extraction hose for section waste 13 - Brush shelf (optional) 14 - Adapter piece for extraction hose
(the coarse filter insert is behind it)
15 - Object head, directional 16 - Waste tray
Fig. 11
19
Page 20

6. Installation

6.1 Site Requirements

The place of installation must meet the following requirements:
No direct sunlight.
Power supply at a distance no greater than approx. 3 m.
No drafts (air condition outlets etc.).
Even floor.
Vibration-free floor.
The instrument is designed for indoor use only.
Obstruction-free access to the handwheel.
The power switch/circuit breaker must be freely and easily accessible.
Room temperature always approx. 22°C.
The relative humidity must not exceed 60%.
To ensure proper function of the instrument, it must be set up while maintaining a minimum distance from walls and furniture.
Distance to walls and furniture, calculated from the cabinet:
- rear: 15 cm
- right side: 30 cm
- left side: 15 cm
No heat dissipating appliances around.
20
High room temperatures and excessive air humidity affect the cooling capacity of the cryostat and lead to ice forming in the instrument!
Plug the instrument into power sockets with ground only.
Power cable length: up to 3.5 m Extendable: no
Operating Manual V 1.1 – 05/2007
Page 21

6.2 Transport to the Site

• First, check if the location meets the conditions specified in "Site
Transport the instrument to the desired location.
Observe the following:
requirements".
The instrument must be transported in an upright position or slight­ly tilted (max. 30°)!
When tilting the instrument 2 people must counterbalance from the front side to prevent the instrument from falling down and causing severe injury!
6. Installation
When transporting the instrument on wheels (2) grip the cabinet only at the marked loca­tions (
To do so, unscrew the adjustable feet using the No. 13 open-end wrench (when subse­quently transporting the instrument on cas­tors, screw the feet back in as far as they will go). To ensure a secure upright position at its intended location, align both adjusting feet (1).
).
1
Fig. 12
Leica CM1950 – Cryostat
2
2
1
2 (covered)
Before transport or relocation, remove the filter bag from the chamber. If you fail to do this, the filter bag will thaw, then freeze solid when the instrument is reconnected. When subsequently removed, the filter will be destroyed, causing section waste to get into the bacteria filter. (Refer to p. 26.)
When not using the extraction for a long period, tightly close the opening for the extraction hose using the silicone stop­per (5, page 28) included in the scope of delivery.
21
Page 22
6. Installation
Transport with a fork lift
The instrument can be transported with a fork lift.
At the installation site, unscrew the adjustable feet (1) with the open­end wrench No. 13 (see Fig. 12). This is absolutely necessary for a stable stand.
Total width with handwheel
Fig. 13
6.3 Assembling the Handwheel
To ensure a safe transport with a fork lift 3 people are required: one operating the fork lift, and the other 2 holding the instrument on either side to prevent it from sliding down.
22
Rotate the handwheel only if the refrigeration system is on and the cryochamber is cold.
2
1
4
Insert the pin (1) of the handwheel shaft into the hole (2) of the handwheel.
Tighten the screw (4) using the No. 6 Allen key.
Place the protective cap on the screw (4).
Fig. 14
To dismount, proceed in reverse order.
Operating Manual V 1.1 – 05/2007
Page 23

6.3.1 Locking/Unlocking the Handwheel

Only rotate the handwheel when the refrigeration system is on and
12 o’clock position
6 o’clock position
the cryochamber is cold!
Always lock the handwheel prior to manipulating the knife or specimen, changing the specimen, or taking a break.
To lock the handwheel, move its handle to the 12 or 6 o’clock position. Press
Fig. 15
the lever (1) fully outward; gently rock the handwheel until the locking mech­anism clicks into place noticeably.
To release the handwheel, press the lever (2) on the handwheel toward the
1
2
Fig. 16
cryostat housing.
Centering the Handwheel (Optional)
An important safety device on the cryostat is the feature for center­ing the handwheel in motorized sectioning mode.
6. Installation
To do so, pull out the handwheel’s handle and position it in the middle of the handwheel. The handle automatically engages in this position.
Fig. 17

6.3.2 Installing the Footswitch Dummy (Instruments with Cutting Motor)

The footswitch dummy must be installed on the outer right side of the instrument (see page 16, General overview) if no footswitch (optional) is used.
If the red LED in the E-STOP field in control panel field 3 is illuminated, either:
emergency-stop function active, or
footswitch dummy (optional footswitch) connected incorrectly or not at all.
Fig. 18
Leica CM1950 Cryostat
23
Page 24
6. Installation

6.4 Connection to Power Supply System

After transporting, wait at least 4 hours before turning the instrument on. This waiting period is necessary to allow the compressor oil, which may have been displaced during transport, to re­turn into its original position. Furthermore, any condensation that has formed during this time due to temperature fluctuations must be allowed to dry completely. Failure to comply with this can cause severe damage to the instrument!
During the start-up of the compressor the nominal voltage must not drop below the values specified in the "Technical Data". Please note that the compressor requires a start-up current between 25 and 35 A. Therefore, the electric circuit at the installation site must be inspected by an electrical engineer to ensure that it meets the requirements for a smooth operation of the instrument. Failure to comply with the above will cause severe damage to the instrument!
Check mains voltage and mains frequency to comply with the specifica­tion on the type plate.
Do not connect any other consumers to this electric circuit.
Never connect the instrument to a power socket that does not have a protective conductor terminal.
Only for instruments sold in Japan
Selecting the frequency
After unpacking the instrument and setting it up at its intended location, use the lever (1) to select the frequency corresponding to the conditions of
1
Fig. 19
the existing power system.
Power plug
24
Operating Manual V 1.1 – 05/2007
Page 25

6.5 Installing Accessories/Inserting Chamber Accessories

6.5.1 Installing the Adjustable Footrest (Optional)

To install the optional footrest, unscrew the screws (1) using the No. 3 Allen key provided.
6. Installation
1
When installing the holder (2), ensure that the cutout faces downwards so that the support (3) can be hooked in.
Attach the holders (2) for the footrest on the left and right to the front wall of the housing. To do so, use the Allen screws you used earli­er. Ensure that the screws are tight.
Hook the footrest (3) into the installed holder
2
according to individual requirements (height).
Once installed, the user can adjust the height of the footrest at any time by relocating it (3) to the desired height on both sides in the
3
Fig. 20
holder (2).
Leica CM1950 – Cryostat
25
Page 26
6. Installation

6.5.2 Installing the Storage Systems (Optional)

2
Fig. 21
1
For reasons of
1
accessibility, the (optional) storage system must always be installed first.
To do so, remove the insert (1), place the frame (2) in front of the bore and tighten the screws/ washers on the cryostat housing using the No. 4 Allen key. Afterwards, insert the insert (1) into the frame and fold it up.

6.5.3 Shelf, Movable (Optional)

Attach the rod for the shelf to the inner front side of the cryostat housing using the provided screws (1) and the No. 3 Allen key, then attach the caps (3). (The rear side of the movable shelf has white plastic screws (2) that prevent the interior of the chamber from being scratched.) Now hook the movable shelf into the guide rod.
3
2

6.5.4 Inserting the Section Waste Tray

Fig. 23
26
Fig. 22
Before installing the knife/ blade holder base, insert the
a
section waste tray with the cutout (a) facing the user.
Operating Manual V 1.1 – 05/2007
Page 27

6.5.5 Installing the Heat Conductor Block, Stationary (Optional)

The holder (1, Fig. 24) of the heat conductor block is screwed to the left housing wall using the No. 4 Allen key provided. (Preferably, begin with
1
the bottom screw. Then, rotate the holder upwards (see arrow) and insert and tighten the top screw).
6. Installation

6.5.6 Installing the Knife/Blade Holder and Adjusting the Clearance Angle

3
1
2
4
Fig. 24
5
Now, attach the cover for the quick freeze shelf to protect the shelf from frost.
For temperature reasons, install the knife/blade holder on an appropriate base.
Set the knife or blade holder (1, Fig. 25) on the base (2), adjust the clearance angle (on the left of the knife/blade holder) to approx. 2°-5° and secure it in the bore (3) on the base (2) using the No. 4 Allen key.
Push the knife holder base (2) on the dovetail guide (4) from the front and tighten it using the clamping lever (5). To move the base, open the clamping lever only a little to prevent acciden­tal sliding in the direction of the specimen head.
If the clamping distance is not sufficient, the clamping lever (3) can be moved. To do so, pull the lever out and move it to the next position.
2
Leica CM1950 – Cryostat
When removing the knife holder base (2) from the refrigerated cryostat chamber, hold it by the grip points
a
Fig. 25
(a – front and rear) to keep your fingers from freezing.
27
Page 28
6. Installation
6.5.7 Inserting/Changing the Bacteria Filter
Fig. 26
6.5.8 Assembling the Filter Bag
1
2
A
L
3a
3
The holder for the bacterial filter (optional) is visible in the front of the instrument.
To insert the filter, hold it with one hand, press on the right of the socket, then guide the filter into the tube from the left.
To change the filter, follow the opposite proce­dure: press the filter to the right, then pull it to the left and out of the tube.
• The filter must be changed approx. every 3 months (we recommend writing the date on the filter using a marker).
The filter must be disposed of accord­ing to valid laboratory specifications.
• Set the mark (A) of the extraction opening (1) and pull it out. Plug the filter (3) into the ex­traction hose connecting piece (3a) until there is an audible click.
Now push the connected parts back into the opening in the cryostat chamber (filter first) and set it to the "closed" mark (2, Fig. 27).
28
Fig. 27
When not using the extraction, tightly close the opening for the extraction hose using the silicone stopper (E) in­cluded in the scope of delivery.
Reasons:
E
1. To prevent section waste from falling into the opening.
2. To prevent cold from escaping from the chamber.
3. To prevent moisture from penetrating into the
Fig. 28
chamber.
Operating Manual V 1.1 – 05/2007
Page 29
6. Installation
6.5.9 Installing the Section Extraction (Optional) – Use with Blade Holder CE Only
When the suction nozzles are changed, the adapt-
A
B
D
C
er (white) remains in the silicone hose. Pull off the nozzle by rotating and pulling it gently and firmly plug in the desired nozzle.
G
Fig. 29
Silicone hose (A) with hose adapter 1 (B, to the filter in the instrument), hose adapter 2 (C, for suction nozzle D or F) and suction nozzle (D) – factory pre­assembled
Silicone stopper (E)
Chamber suction nozzle (F)
Filter (G)
Plastic clip (H), for parking the chamber suction nozzle.
Mark for suction nozzle
C
H
Fig. 30
K
D
I
E
F
B
Ensure that the hose with the nozzle is not installed against its "natural" cur­vature on the pressure plate of the blade holder.
The tension acting on the hose can be minimized by turning the red ring (Fig. 31, top right) clock­wise so that the suction nozzle presses against the pressure plate (I, Fig. 31). Afterwards, fold the anti-roll guide (K) back onto the pressure plate.
The scope of delivery also includes 2 plastic clips (H). These enable comfortable "parking" of the chamber suction nozzle (F) during sec­tioning.
The clip must be glued in before switching on the refrigeration. Before doing so, briefly degrease the surface to ensure a secure hold. Preferably, the clip should be attached outside the working area, e.g. on the left inside wall of the instrument.
If the extraction is not used for a long time, it is absolutely necessary to clean the extraction hose in order to ensure maximum extraction capacity. To do so, place the hose in commercially available disinfectant or alcohol. After several cleanings, the hose must be replaced (see order information on p. 49)!
Leica CM1950 – Cryostat
If the extraction nozzle (D) is not being
A
Fig. 31
used, it can be "parked" on one of the two magnetic surfaces indicated in the interior of the instrument.
29
Page 30
7. Instrument Controls
7.1 Control Panel Fields on the CM1950 – Control Panel Field 1
Display of actual and target chamber temperatures
Fig. 32
Button to enable/disable vacuum extractor,
Extraction force intensity selectable from 1-5 (see following page for more information)
Button to enable/disable the Peltier element
Display of actual and target specimen head temperatures
+/- buttons to select chamber temperature
+/- buttons to select specimen head temperature
Max-Cool button to select the maximum low temperature of the specimen head directly (-50 °C)
+/- buttons to select real time
+/- buttons to
select defrost
Display of real time, defrost time and error messages
"Melting Snowflake" button to activate manual defrost
time
ON/OFF button of illumination
Press the key button to lock/unlock the entire keypad. (Refer to following page for activation of the specimen head)
UVC disinfection, (short time 30 min, long time 180 min)
30
EMERGENCY STOP switch to the right of control panel field 1 (motorized instruments only) For danger situations during motorized sectioning.
Immediate stop of the sectioning process – motor stops – LED in E-Stop is illuminated in red.
Turning in direction of arrow cancels the stop – LED in E-Stop goes out.
• Select single stroke (Single) or continuous stroke (Cont.) operating mode again.
Operating Manual V 1.1 – 05/2007
Page 31
7. Instrument Controls
A
Press the "VAC" button to enable the vacuum extractor. The LED in the "VAC" button is lit while the extractor is on. Press the button again to
12 o’clock
B
Position
6 o’clock
disable it.
Use the knob to adjust the intensity of the vacuum.
A Optimal area for trimming and sectioning
Trimming: Handwheel position 12 - 6 o’clock, valve open
Handwheel position 6 - 12 o’clock, valve closed
Cutting: Handwheel position 12 - 3 o’clock, valve open all the way
Handwheel position 3 - 6 o’clock, valve half open Handwheel position 6 - 12 o’clock, valve closed
B Optimal area for extraction from the chamber
To clean the chamber, turn the knob to the red range.
The strength of the required extraction force depends on the following:
• Size of the specimen
• Sectioning speed
• Section thickness used
The Peltier element provides additional cooling for the freezing stations. After the button is pressed, the display changes from "PE" to 10", indicating an additional cooling period of 10 minutes. The remaining cool­ing time is always shown in the display. Once only 4 minutes are remain­ing, a dot will appear after the "4". As of this time, the Peltier element may be switched off early by pressing the button again.
Leica CM1950 Cryostat
Caution: The specimen head and Peltier do not switch on until the chamber temperature reaches -5°C, in order to prevent icing.
If the condenser (resting phase) is off and the Peltier is activated, the digit 10 flashes until the condenser switches on again to pre­vent the Peltier from being destroyed when the condenser is not running. When the condenser starts up, the flashing starts and the 10 minutes are counted down.
(For exact instructions for using the chamber, specimen head and real time display fields, refer to the chapter on "Working with the Instrument" on page 37 of this Operating Manual.)
31
Page 32
7. Instrument Controls
Control Panel Field 2 – Electric Coarse Feed, Sectioning and Trimming Thickness Display
Illuminates in
yellow when the
specimen is in
retraction.
Toggle TRIM and
SECT (LED
active).
Press approx. 3 sec.; "on" or
"off" appears (for the
retraction). Toggle using the
"+" or "-" key
Retraction: off = 0
on = 20 μm
In manual mode.
For motorized sectioning, the re­traction value is fixed and cannot be changed (see page 39 - lower infobox).
In the "off" setting, there is no retraction in manual, automatic or rocking mode.
1
2
Fig. 33
For trimming values above a sec-
LED for trim and section thickness display
Rocking mode – only in manual mode; in rear area, i.e. approx. 12 - 3 o’clock position of the handwheel (rock handwheel back and forth a short distance)
Move specimen head backwards quickly to home position (latched)
Move specimen head backwards slowly – pressing briefly retracts the specimen 20 μm
Specimen head forwards – pressing briefly advances the specimen 20μm
Specimen head fast feed
tion thickness of 200 μm, the dis­play flashes to expressly notify the user of thick sections!
Setting section/trim thickness
Use the
- keys on control panel 2.
Range of section thickness: 1 - 100 μm Values: 1.0μm - 5.0μm in 0.5μm steps
5.0μm - 20.0 μm in 1.0 μm steps
20.0μm - 60.0 μm in 5.0 μm steps
60.0μm -100.0 μm in 10.0 μm steps
32
Range of trim thickness: 1 - 600 μm (Recommended for research applications) Values: 1.0 μm - 10.0 μm in 1.0 μm steps
10.0μm - 20.0 μm in 2.0 μm steps
20.0μm - 50.0 μm in 5.0 μm steps
50.0 μm - 100.0 μm in 10.0 μm steps
100.0 μm - 600.0 μm in 50.0 μm steps
Trimming section thickness setting range: (Recommended for clinical applications) Values: 10μm, 20μm, 30 μm, 40 μm,
Operating Manual V 1.1 – 05/2007
Page 33
Coarse feed functions
The two-speed electric coarse feed is designed to quickly position the spec­imen in relation to the knife. The buttons with double arrows move the spec­imen at 900 μm/s; the buttons with single arrows move it at 300 μm/s.
Retracting the specimen head from the knife
Pressing once starts the rapid backwards movement to the rear end
1
position (Home position). The LED (1) flashes while the specimen head is in motion.
fast
The LED (1) lights up when the rear end position (Hp.) has been reached.
The retraction can be stopped by pressing one of the coarse drive buttons.
The rapid backwards movement to the rear end position (Hp.) starts. The motion continues as long as the button is held.
A brief press of the button retracts the specimen by 20μm.
slow
7. Instrument Controls
Advancing the specimen toward the knife
Start the slow forwards movement to the knife. To feed the specimen, press and hold the button.
slow
Pressing the button briefly results in a feed motion of 20μm.
Start the fast forwards movement to the knife.
2
The LED (2) flashes while the specimen head is in motion. The LED (2) lights up when the forward end position has been reached.
fast
Manual sectioning mode
Select ROCK operating mode (LED active) – retraction must be enabled!
For sectioning, turn the handwheel a short distance (approx. 1/4 turn) forwards and back (rocking mode) – only possible at rear (handwheel in approx. 12 - 3 o'clock position). Every change in rotation direction is electronically detected and automati­cally translated into a specimen feed or retraction movement.
Leica CM1950 – Cryostat
33
Page 34
7. Instrument Controls
Control panel 3 – Motorized sectioning (optional)
Switching the sectioning mode
("CUT MODE") from continuous
stroke ("CONT") to single stroke
("SINGLE") (active)
Motor speed controller (0-100%)
Hold the button to section
at maximum speed.
Release the button to continue section-
ing at the previously selected speed
(see controller, above).
Switch from slow speed range to fast speed range: while switching on the in­strument, press and hold down the Vmax button.
Speed ranges
Slow: 0-50 strokes/min Fast: 0-85 strokes/min Vmax: 85-90 strokes/min
Sectioning modes
The instrument can be used in manual and motorized modes. The following settings are available:
Single stroke (SINGLE) or continuous stroke (CONT.) in automatic mode, as well as
ROCK (sectioning using handwheel) in manual mode.
Fig. 34
If emergency stop has been activat­ed, the cutting mode must be select­ed again.
The red LED in the E-STOP field indicates either:
• emergency-stop function active, or
• footswitch dummy (optional footswitch) connected incorrectly or not at all.
The mechanical handwheel brake is enabled when the yellow LED is lit in the M-STOP field.
The handwheel must also be locked when working on the specimen head.
Press to brake the handwheel electronically (LED lit) – specimen stops in lower (6 o’clock) position. Can be used in any position.
1. Press the buttons at the same time to start motorized sectioning.
2. To end the sectioning process, press RUN/STOP, ENABLE or BRAKE.
- Specimen head stops at bottom (for BRAKE, automatic electronic braking takes place).
3. For motorized sectioning, does not have to be unlocked; continue working by pressing both the RUN/STOP and ENABLE keys.
4. If, during manual work using handwheel, you have braked using BRAKE also use BRAKE to release!
34
When switching the instrument on, no operating mode is active for safety reasons.
Operating Manual V 1.1 – 05/2007
Page 35
7. Instrument Controls
Fig. 35
Disinfection
Duration - 30min
Duration - 180 min
UVC key - to activate / deac­tivate the disinfection cycle and/or to acknowledge in­terruption of a disinfection cycle.
The germicidal UV lamp is a simple and safe disinfection device which considerably reduces the risk of infection. Nevertheless, UV light disinfection cannot replace chemical disinfection of the cryochamber at regular intervals. UV light disinfection is effective only on those surfaces that are directly illuminated.
Specimens and section waste must be thoroughly removed from the cryochamber first (e. g. using the optional vacuum extractor, or a paper towel soaked with Cryofect or alcoholic disinfectant). Before UV disinfection, move the anti-roll guide to the side to allow complete disinfection.
To start disinfection the sliding window must be completely closed.
• Press UVC key once briefly to start the 30 min. mode
• Press UVC key once for a longer time (approx. 4sec.) to start the 180 min. mode
Leica CM1950 – Cryostat
Leica Cryofect is not available in all countries.
Opening the sliding window cancels the disinfection cycle. Press the UVC key to acknowledge this.
When the keypad lock is activated (via the key button) the UV lamp can be shut off only by opening the glass, as the UV keys are locked. The cancellation can be acknowledged only if the keypad lock is disabled. Only then can the UV lamp be switched back on.
35
Page 36

8. Working with the Instrument

8.1 Preparing Cutting Tools, Specimen Discs and Preparation Aids

The knives are extremely sharp! Handle with care! Never try to catch a falling knife!
Place working materials such as the blade box or knives (in the knife case), brush, pincers or preparation needles and, where applicable, specimen discs into the cryostat chamber.
The necessary tools and preparation aids can be precooled on the (optional) movable shelf, making them available at all times in a conve­nient position for the user.
In addition, specimen discs can be precooled and stored in the storage system (see page 26, Fig. 21).

8.2 Turning the Instrument On

The circuit breaker is used as power switch. The switch must be in the top position for switching on and in the bottom position for switching off. The switch must be accessible without obstruction.
Close the sliding window.
Fig. 37
36
Fig. 36
The instrument must be switched on at least 5 hours before the planned use.
To avoid frost formation always put the cover on the quick freeze shelf. Always cover the quick freeze shelf during breaks and overnight.
For installation of knife/blade holder and installation in the chamber, refer to optional accessories on page 50).
Operating Manual V 1.1 – 05/2007
Page 37

8.3 Configuring the Parameters

Turn the lamp on or off.
This button activates or deactivates the manual defrost of the cooling chamber, specimen head or freeze shelf. (For more detailed handling instructions, refer to the chapter on "Working with the Instrument" on p. 39 of this Operating Manual.)
• Pressing the key button for approx. 5 sec. locks the entire keyboard (the LEDs in the clock go out).
Pressing the key button briefly, then the "-" key in the specimen head control panel field, switches off the specimen head.
Pressing the key button briefly, then the "+" key in the specimen head control panel field, switches the specimen head back on.
8. Working with the Instrument
The instrument must be switched on at least 5 hours before the planned use.
Leica CM1950 Cryostat
Programming the temperature of the cryochamber
The temperature of the cryochamber is set and indicated on the control panel marked with the cryostat symbol.
The actual temperature is the standard indication. Briefly pressing the "+" or "-" button displays the target temperature. Set the desired value via the "+" / "-" buttons. Pushing the "+" or "-" button
for more than 1 sec. increases or decreases the chamber temperature continuously.
The actual value will be indicated 5 seconds after finishing the program­ming.
Refer to page 52 for a table with guide values. The temperature values given there are based on long-term experience, but are intended solely as guide values, as any tissue may require particu­lar adjustments.
37
Page 38
8. Working with the Instrument
Programming the specimen temperature
Select the desired temperature of the specimen.
• The specimen temperature is set and indicated on the control panel marked with the specimen head symbol.
The actual temperature is the standard indication. Briefly pressing the "+" or "-" button displays the target temperature. Set the desired value via the "+" / "-" buttons. Pushing the "+" or "-" button
for more than 1 sec. increases or decreases the specimen temperature continuously.
The actual value will be indicated 5 seconds after finishing the program­ming.
Caution: The specimen head and Peltier do not switch on until the chamber temperature reaches -5 °C, in order to prevent icing.
Specimen temperature - "Max-Cool" function
The snowflake button for the Max-Cool function is in the panel with the specimen head symbol. Pressing the button sets the lowest possible specimen head tem­perature (-50°C) as the target temperature. The instrument adjusts the maximum low temperature of the specimen head, i.e. -50°C.
Push the snowflake button again to stop the "Max-Cool" function. The temperature adjusts to the value programmed prior to activating the "Max-Cool"-function.
Alternate flashing of "LL" and the actual temperature indicates activa­tion of the Max-Cool function.
38
Setting the time
The clock time is set with the +/- buttons in the control panel marked with the clock symbol. To do so, set the current time using the "+" or "-" button below the small clock symbol. Pushing the "+" or "-" button for more than 1 sec. increases or decreases the time continuously (auto-repeat function).
Operating Manual V 1.1 – 05/2007
Page 39
8. Working with the Instrument
Programming the defrost cycle
Set the beginning of the automatic defrost cycle.
The automatic defrost cycle takes place once within 24 hours. It is set with the "+"/"-" buttons on the right of the panel with the clock
.
symbol. The two buttons are marked by a melting snowflake
Briefly press the "+" or "-" button for indication of the beginning of the
defrost cycle which has actually been set. At the same time, the LEDs between the indication of hours and minutes are flashing.
To change the beginning of the defrost cycle in steps of 15 minutes, push
the "+" or "-" key. When the "+" or "-" button is pushed for more than 1 s, the defrost time value increases or decreases continuously.
Before starting the defrost cycle remove all samples from the cryochamber!
When the automatic defrost cycle begins, the specimen head tempera-
ture adjusts to a temperature between - 10°C and - 5°C (reduction of ice formation). The specimen head cooling turns off. This is confirmed by the flashing of the decimal points on the panel for the specimen cooling. The specimen cooling (controlled to set value) automatically turns back on after 4 hours, once the chamber temperature varies by less than 5 K from the target temperature.
If you want to turn the specimen cooling back on manually before the
automatic activation sets in, push the "+" or "-" button on the panel for the specimen cooling and then the key button.
The temperature of the specimen cooling first raises to +10°C and then
adjusts to the programmed specimen temperature.
Leica CM1950 Cryostat
Manual defrost for the freeze shelf (including Peltier element)
Press the key for manual defrost
,
then the
key.
The manual defrost takes 12 min.
Push the button and then the
After defrosting the freeze shelf, carefully wipe it out, as otherwise a lot of water collects in the channel. Ice does not melt during normal defrosting.
key again to stop the manual defrost.
39
Page 40
8. Working with the Instrument
Manual defrosting of the cryochamber
Push the manual defrost button (with the melting snowflake) on the left over the key button to activate the defrost cycle of the cryochamber on demand.
Activation is confirmed by an audible signal.
Then, push the "+" or "-" button on the panel for the cryochamber temperature.
The manual defrost cycle (12 min.) is activated.
There is a flashing indication of the temperature of the cryochamber during the whole defrost cycle.
If necessary, push the manual defrost button again to deactivate the manual defrost cycle.
When the manual defrost begins, the specimen head adjusts to a tem­perature between -10°C and -5°C (reduction of ice formation). The specimen head cooling turns off. This is confirmed by the flashing of the decimal points on the panel for the specimen cooling.
Ten seconds after the manual defrost cycle has been completed, the specimen cooling turns back on.
Before starting the defrost cycle, remove all specimens from the cryochamber!
40
Manual defrosting of the specimen cooling
Push the manual defrost button (with the melting snowflake) on the left over the key button to activate the defrost cycle of the specimen head.
Activation is confirmed by an audible signal.
Then, push the "+" or "-" button on the panel for the specimen temperature.
There is a flashing indication of the specimen temperature during the defrost cycle.
For 15 minutes, the specimen head is adjusted to a temperature of 45°C.
Subsequently, the instrument adjusts to the specimen temperature which has been programmed prior to the manual defrost cycle.
If necessary, push the manual defrost button again to deactivate the manual defrost cycle.
Press the key button, then the "+" button = specimen head on Press the key button, then the "-" button = specimen head off
Operating Manual V 1.1 – 05/2007
Page 41
8. Working with the Instrument
Entering the trimming thickness
To switch from a trimming section thickness for research applica­tions (1-600 μm) to a thickness for clinical applications (10, 20, 30 or 40 μm), press and hold down the TRIM/SECT key while switching on the instrument.
Press the TRIM/SECT button. TRIM mode is active if the LED at the top
right is illuminated.
Set the desired trimming thickness using the "+" or "-" button in control
panel field 2. (Refer to page 32 for adjustable sequence of steps)
Entering the section thickness
Press the TRIM/SECT button. SECT mode is active if the LED at the bot-
tom right is illuminated.
Set the desired section thickness using the "+" or "-" button in control
panel field 2. (Refer to page 32 for adjustable sequence of steps)
Switching the retraction on or off in manual sectioning mode
Press the TRIM/SECT button for approx. 3 sec. The LED in control panel
field 2 displays on or oFF .
Leica CM1950 Cryostat
You can switch by pressing the "+" or "-" button.
"Retraction on" means a specimen retraction of 20 μm in manual mode.
During motorized sectioning, the retraction value is speed­dependent and cannot be changed by the user.
41
Page 42
8. Working with the Instrument

8.4 Working with the Precooled Cryostat

8.4.1 Preparatory Work

Lock the handwheel in the top (12 o’clock) position.
Cut the specimen to size outside of the cryostat.
Select the precooled specimen disc, cover it with freezing compound, then attach and orient the specimen.
Attach the specimen disc and specimen to the Peltier position on the freeze shelf. Activate the Peltier element and wait until the specimen is completely frozen.
Specimens that have been frozen on the Peltier element are often too cold and split apart during sectioning. Allow time for the speci­mens to become acclimated.
Insert the specimen disc into the specimen head.
Fold the anti-roll guide away and to the side. Insert a precooled knife or blade.
42
Before using them for the first time, degrease new knives using acetone or alcohol.
Approach the specimen with the knife or blade holder:
To do so, open the clamping lever of the base, approach the
specimen and close the lever again.
• Open the lever of the orientation. Orient the specimen (move it into a favorable position relative to the knife/blade) and close the lever again.
Approach the knife or blade holder using the coarse feed buttons and gentle movements of the handwheel.
If the sections are cracked, the specimen head temperature is too cold. Set a warmer temperature.
If the sections smear, the specimen head temperature is too warm. Set a colder temperature.
Operating Manual V 1.1 – 05/2007
Page 43
8. Working with the Instrument
8.4.2 Trimming with Extraction – Anti-roll Guide Installed
Remove the silicone stopper from the filter cover (and keep it in a safe
place).
E
Fig. 38
K
C
Insert the extraction hose with the black adapter.
Fold the anti-roll guide to the side and fasten the suction nozzle to the
pressure plate – see mark Fig. 39a - (using 4 magnets on the rear side of the nozzle).
Fold the anti-roll guide back into position.
Fig. 39a Fig. 39b Fig. 39c
D
I
B
The tension acting on the hose can be minimized by turning the red ring (Fig. 39, top right) clock­wise so that the suction nozzle presses against
A
Fig. 40
the pressure plate (I, Fig. 39).
Ensure that the hose with the nozzle is not installed against its "natural" cur­vature on the pressure plate of the blade holder.
Leica CM1950 – Cryostat
• Check that the anti-roll guide is parallel and correctly adjusted.
Readjust if necessary (see page 58).
Activate trimming mode.
Select the trimming thickness.
Switch on the VAC and select a low extraction level (between 1 and 2).
Start trimming by manually moving the handwheel or press the RUN/STOP and ENABLE keys simultaneously to start motorized sectioning.
43
Page 44
8. Working with the Instrument
For motorized sectioning, always begin at low speed for safety reasons.
A
Press the "VAC" button to enable the vacuum extractor. The LED in the
B
"VAC" button is lit while the extractor is on. Press the button again to disable it.
Use the knob to adjust the intensity of the vacuum.
A Optimal area for trimming and sectioning
Trimming: Handwheel position 12 - 6 o’clock, valve open
Handwheel position 6 - 12 o’clock, valve closed
Cutting: Handwheel position 12 - 3 o’clock, valve open all the way
Handwheel position 3 - 6 o’clock, valve half open Handwheel position 6 - 12 o’clock, valve closed
B Optimal area for extraction from the chamber
To clean the chamber, turn the knob to the red range.
Optimize the vacuum settings if necessary.
If the handwheel is not moved for approx. 5 sec, the valves close and the fan remains on. If the handwheel is moved for approx. 1 min, the valves close, the fan shuts off (the LED in the VAC switch goes off to prevent icing). To continue working, you now have to reactivate the VAC.
2. Trimming with Extraction – Brush Technique, finger rest installed
Remove the silicone stopper (E) from the filter cover (and keep it in a safe place).
E
Insert the extraction hose with the black adapter.
Fasten the suction nozzle to the pressure plate (using 4 magnets on the rear side of the nozzle) as far as possible towards the blade.
44
Ensure that the hose with the nozzle is not installed against its "natural" curvature on the pressure plate of the blade holder.
The tension acting on the hose can be minimized by turning the red ring (Fig. 28 on page 28) clockwise so that the suction nozzle presses against the pressure plate.
Check that the suction nozzle is seated optimally (by gently turning the handwheel)
Operating Manual V 1.1 – 05/2007
Page 45
Activate trimming mode.
Select the trimming thickness.
Switch on the VAC and select a low extraction level (between 1 and 2).
Start trimming by manually moving the handwheel or press the
RUN/STOP and ENABLE buttons simultaneously to start motorized sectioning.
Use a brush to apply the section to a precooled specimen slide, then
warm it from below using your finger.
Move the knife guard forward after removing the section.
Remove the knife or disposable blade (using the blade ejector!).
Insert the knife into the knife case.
8.4.3 Cutting with Extraction – Anti-roll Guide installed
Shut off VAC (LED in VAC button goes out).
Switch from trimming to sectioning mode (important for section stretch-
ing, as the valves work differently than in trimming mode.)
Set the desired section thickness.
Switch on the VAC and begin at level 1. If the section does not stretch
correctly, turn the VAC knob higher in small increments.
Once the desired section is on the pressure plate, switch off the VAC!
Carefully fold the anti-roll guide to the side and remove the section from the side.
8. Working with the Instrument
Leica CM1950 Cryostat
After removing the section, wipe off moisture/condensate from the pressure plate – otherwise, the next sections will become jammed.
Cutting with extraction without anti-roll guide (brush technique) is not possible, as the position of the pressure plate means that no suitable air flow is obtained.
45
Page 46
8. Working with the Instrument
A few rules:
Always begin at a low extraction level, then slowly increase it.
Do not use high extraction levels unless absolutely necessary.
Different specimen sizes require different extraction levels.
The faster the trimming or sectioning speed, the lower the extraction level should be.
The larger and/or thicker the specimen to be trimmed, the lower the extraction value.
When section specimens with a diameter of 0.5 cm, the anti-roll guide stretches the section adequately. For larger specimens, we recommend using the vacuum function.
Trimming, 1 x 1cm specimen
Trimming thickness Speed Extract. value
10μm 40% 1.5 10μm 60% 1.5 10μm 80% 1 10μm 100% 1
20μm 40% 2 20μm 60% 2 20μm 80% 1.5 20μm 100% 1
30μm 40% 2.5 30μm 60% 2.5 30μm 80% 2 30μm 100% 1.5
40μm 40% 3 40μm 60% 3 40μm 80% 2.5 40μm 100% 2
Trimming, 1 x 1cm specimen
Trimming thickness Speed Extract. value
1 μm - 5 μm 40% 2.5
60% 2 80% 1.5 100% 1.5
10μm 40% 2.5
60% 2 80% 2 100% 1.5
The values listed here are possible settings that depend on the tissue type, specimen size, section thick­ness and section speed, and are in­tended for orientation purposes only.
These settings can also vary when using different anti-roll plate!
46
Operating Manual V 1.1 – 05/2007
Page 47
After trimming or sectioning:
Specimen:
Unclamp and thaw.
Immerse in fixative for further processing.
Cleaning
Use the brush to sweep up the section waste (section waste tray) and
remove it from the cryostat (follow applicable laboratory regulations for disposal).
Or:
Clean the cryostat chamber using the chamber suction nozzle:
To do so, turn the (flat) suction nozzle of the extraction hose by holding the hose on the white adapter and removing it with a quick twist. "Park" the flat suction nozzle in a designated place in the chamber, e.g. on the right inside wall of the cryogenic chamber.
Remove the chamber suction nozzle from the plastic clip and attach it firmly to the white adapter.
Check the remaining capacity of the filter (in the interior of the chamber)
and change the filter if necessary (see page 28).
Check the bacterial filter (in the front of the unit, see page 28), change it
at least every 3 months.
Move the anti-roll guide to the side.
Spray the cryostat chamber using Cryofect.
Activate the UVC disinfection.
8. Working with the Instrument
Leica CM1950 Cryostat
47
Page 48

9. Troubleshooting

9.1 Problems During Work

Problem Causes Remedies
Frost on chamber walls and microtome
Sections smear
Sections splinter
Sections not properly flattened
Sections not properly flattened despite correct temperature and correctly aligned anti-roll plate
Cryostat is exposed to air currents (open windows and doors, air conditioning). Frost built-up by breathing into the cryochamber.
Specimen not cold enough. Anti-roll plate not yet cold enough, thus warming the section.
Specimen too cold.
Static electricity/air currents. Specimen not cold enough. Large area specimen.
Anti-roll plate poorly positioned. Anti-roll plate poorly aligned with knife edge. Incorrect clearance angle. Blunt knife.
Dirt on knife and/or anti-roll plate. Edge of anti-roll plate damaged. Blunt knife.
Change place of installation for the cryostat.
Wear mouth protection.
Select lower temperature. Wait until knife and/or anti-roll plate have reached chamber temperature.
Select higher temperature.
Remove cause. Select lower temperature. Trim the specimen parallel, increase section thickness. Reposition anti-roll plate. Align correctly.
Set correct angle. Use different part of the knife.
Clean with dry cloth or brush. Replace anti-roll plate. Use different part of the knife.
Sections curl on anti-roll plate
Scraping noise during sectioning and specimen return movement
48
Anti-roll plate does not protrude far enough beyond the knife edge.
Anti-roll plate protrudes too far beyond the knife edge and is scraping against the specimen.
Readjust correctly.
Readjust correctly.
Operating Manual V1.1 – 05/2007
Page 49
9. Troubleshooting
Problem Causes Remedies
Ridged sections
Knife damaged. Edge of anti-roll plate damaged.
Use different part of the knife. Replace anti-roll plate.
Chatter during sectioning
Condensation on anti-roll plate and knife during cleaning
Anti-roll plate damaged after adjustment
Thick-thin sections
Specimen insufficiently frozen onto the specimen disc. Specimen disc not clamped tightly. Knife not clamped tightly enough. Specimen has been sectioned too thickly and has detached from disc. Very hard, inhomogenous specimen.
Blunt knife. Knife profile inappropriate for specimen cut. Incorrect clearance angle.
Brush, forceps and/or cloth too warm.
Anti-roll plate too high above the knife edge. The adjustment was carried out in the direction of the knife.
Temperature incorrect for the tissue cut. Knife profile inappropriate for the specimen cut. Ice buildup on the knife back. Handwheel speed not uniform. Knife not clamped tightly enough. Specimen disc not clamped tightly enough. Cryocompound applied to cold specimen disc; specimen detached from disc after freezing.
Refreeze specimen onto the disc.
Check disc clamping. Check knife clamping. Refreeze specimen onto the disc.
Increase section thickness; reduce specimen surface area if ecessary. Use different part of the knife. Use knife with different profile.
Set correct angle.
Store all tools on storage shelf in the cryochamber.
Replace anti-roll plate. Be more careful next time!
Select correct temperature.
Use knife with different profile (c or d). Remove ice. Adapt speed. Check knife clamping. Check disc clamping.
Apply cryocompound to warm disc, mount specimen and freeze.
Leica CM1950 Cryostat
49
Page 50
9. Troubleshooting
Problem Causes Remedies
Blunt knife. Inappropriate section thickness. Incorrect clearance angle. Microtome not dried thoroughly enough. Dried specimen.
Use different part of the knife. Select correct section thickness. Set correct angle. Dry microtome.
Prepare new specimen.
Tissue sticks to the anti-roll plate
Flattened sections curl up when anti-roll plate is folded up
Sections tear
Cryostat non-operational
Specimen disc cannot be removed
Anti-roll plate is too warm or incorrectly positioned. Fat on the corner or edge of anti-roll plate. Anti-roll plate not correctly fixed. Rust on the knife.
Anti-roll plate too warm.
Temperature too low for the tissue cut. Blunt part, dirt, dust, frost or rust on the knife. Top edge of the anti-roll plate damaged. Hard particles in the tissue. Dirt on knife back.
Power plug not properly connected. Defective fuses, or circuit breaker has triggered.
Moisture on the underside cause the specimen to freeze to the freezing shelf or specimen head.
Cool down anti-roll plate, or reposition correctly. Remove fat from anti-roll plate.
Fix correctly. Remove rust.
Cool down anti-roll plate.
Increase temperature and wait. Remove cause.
Replace anti-roll plate.
- - ­Clean.
Check if properly connected. Replace fuses, or switch circuit breaker back on. If not possible, call technical service.
Apply concentrated alcohol to the contact point.
50
Operating Manual V1.1 – 05/2007
Page 51
9. Troubleshooting
Problem Causes Remedies
No or insufficient refrigeration of the cryochamber
Cooling system or electronic drive defective.
Call technical service.
Sliding window collects condensation
No or insufficient refrigeration of the specimen
Lamp does not work
Both disinfection LEDs flash alternately
Air humidity and room temperature too high.
Cooling system or electronic drive defective.
Lamp defective.
Switch defective.
UV radiation provided by UV tube no longer sufficient.
Image of an open-ended wrench appears due to fault to be rectified.
Comply with the requirements for the installation site.
Call technical service.
Check lamp and replace it, if necessary. Call technical service.
Replace UV tube following the manufacturer’s instructions.
Contact technical service and follow the instructions given.
Leica CM1950 Cryostat
51
Page 52

10. Temperature Selection Chart (in minus °C)

Tissue type Chamber temperature Specimen head temperature
Spleen -15°C to -20°C -11 °C
Liver -10 °C -20 °C
-15°C off to -15°C
Intestine -10°C -20°C
-15°C A*: off to -20 °C
E*: -20°C
Heart -10°C A: -20°C
E: -20 °C to -30 °C
-15°C off to -20°C
Ovaries -10°C E: -20°C
-15°C off to -15°C
Fallopian tubes -10° E: -20°C
-15°C off to -15°C
Kidney -10 °C -20°C
-15°C A: off to -15 °C
-20 °C -20 °C
Muscle -18 °C to -20 °C -15°C
Skin with fat -19 °C -32°C to -40°C
Hard fat -19°C -21°C to -25 °C
Stomach -10 °C -20 °C
-15°C off to -15°C
Brain -15°C -10°C, *E
*A = blocked, *E = complete
The temperature values given above are based on long-term experience, however, these are only approximate values, as any tissue may require particular adjustments.
52
Operating Manual V 1.1 – 05/2007
Page 53

11.1 Ordering Information

Specimen disc ø 20mm assembly .................................................... 0477 43739
Specimen disc ø 30mm assembly .................................................... 0477 40044
Specimen disc ø 40mm assembly .................................................... 0477 40045
Specimen disc ø 55mm assembly .................................................... 0477 40046
Specimen disc 80x 50 mm, assembly ............................................... 0477 43714
O-ring blue (10 pieces), 20 mm.......................................................... 0477 43247
O-ring red (10 pieces), 20 mm ........................................................... 0477 43248
O-ring blue (10 pieces), 30 mm.......................................................... 0477 43247
O-ring red (10 pieces), 30 mm ........................................................... 0477 43248
O-ring blue (10 pieces), 40 mm.......................................................... 0477 43249
O-ring red (10 pieces), 40 mm ........................................................... 0477 43250
O-ring blue (10 pieces), 55 mm.......................................................... 0477 43251
O-ring red (10 pieces), 55 mm ........................................................... 0477 43252
Blade holder base complete.............................................................. 0477 40351
Blade holderextension CE-BB, complete ........................................ 0477 43005
Insert for low-profile blades for blade holder CE, complete ........ 0477 42488
Finger rest, complete ........................................................................... 047740387
Anti-roll plate 70 mm - 50μm spacer ................................................ 0477 42491
Anti-roll plate 70 mm -100μm spacer................................................ 0477 42492
Anti-roll plate 70 mm -150μm spacer................................................ 0477 42493
Glass insert 70mm, polished.............................................................. 0477 42497
Knife holder base assembly............................................................... 0477 42359
Knife holder CN assembly .................................................................. 0477 42358
Knife support CN short ....................................................................... 0477 42380
Knife support CN ................................................................................. 0477 42370
Knife holder attachment CNZ, assembly ......................................... 0477 42363
Anti-roll plate assembly glass 50mm ................................................ 0419 33981
Heat and cold extractor, assembly................................................... 0477 41039
Cold extractor, assembly.................................................................... 0477 43737
Heat extractor, assembly ................................................................... 0477 43126
11. Optional Accessories
Leica CM1950 – Cryostat
53
Page 54
11. Optional Accessories
Section waste tray .............................................................................. 0477 40062
Brush shelf ........................................................................................... 0477 43036
Storage system, assembly ................................................................. 0477 42618
Shelf, movable ..................................................................................... 0477 43037
Retaining device .................................................................................. 0477 40080
Cover freezing shelf ............................................................................ 0477 43763
Specimen disc 37x 37mm ................................................................... 0477 42603
Specimen disc 28x 28mm ................................................................... 0477 42604
Bar, large .............................................................................................. 0477 42600
Bar, medium size ................................................................................. 0477 42601
Bar, small .............................................................................................. 0477 42602
Heat extractor, Dr. Peters, assembly ............................................... 0477 41338
Freezing griddle/block elevator......................................................... 0201 39119
Dispensing slides 8 pieces................................................................. 0201 39127
Foot rest assembly .............................................................................. 0477 42832
Accessory kit extraction .................................................................... 0477 43300
Bactericidal filter 350/5865................................................................. 0477 40296
Hose set, silicon 5 pieces................................................................... 0477 44469
Filter assembly 25 pieces ................................................................... 0477 44307
Cryofect, VPE Leica Cryofect 4x 350ml,
low-temp. disinfectant to -20°C......................................................... 0387 42801
Safety gloves size M ........................................................................... 0340 29011
Safety gloves size S ............................................................................ 0340 40859
Footswitch, dummy plug CM 3050 .................................................... 0443 30420
Laboratory chair on sliders (8030442) .............................................. 0710 34911
Footswitch, assembly ......................................................................... 0505 33888
Easy Dip staining container white ................................................... 0712 40150
Easy Dip staining container pink....................................................... 0712 40151
Easy Dip staining container green.................................................... 0712 40152
Easy Dip staining container yellow .................................................. 0712 40153
Easy Dip staining container blue ...................................................... 0712 40154
Easy Dip staining rack gray ............................................................... 0712 40161
54
Operating Manual V 1.1 – 05/2007
Page 55
11. Optional Accessories
Blade holder CE with anti-roll guide for low-profile (NB) and high-profile (BB)
1
6
4
4
7
12
With blade ejector (1)
With knife guard (2)
Integrated lateral shift and stable base
• Clearance angle adjustment (5) with 3 mm Allen key (see detail image 5, to left at knife holder) - recommended angle from 2° to 5°.
3
With anti-roll guide (4)
Lever (6) for the lateral shift
2
8
Lever (7) for clamping the blade
• Lever (8) for clamping the base (9) to the
9
dovetail guide in the chamber
Pressure plate (3) for section extraction
• When using low-profile blades, the ruler
(7, Fig. ) must be inserted.
5
Fig. 41a
Fig. 41
Assembling the anti-roll guide system (for the blade holder CE)
10
1. Insert the glare-minimizing glass insert into the interchangeable frame and tighten it evenly using the knurled screw (10).
2. Insert the shaft (11) of the metal frame for
6
11
10
exchangeable glass inserts from above into the bore of the swinging arm (12) in such a way that the pin rests in the notch.
3. Push the white plastic plate (14) from below
14
13
onto the shaft (11).
4. Screw the knurled nut (13) from below onto the shaft (11).
Fig. 42
The glass of the anti-roll guide can be used from all 4 sides when replace­ment is necessary (the glass stage plate can be reordered).
Leica CM1950 – Cryostat
55
Page 56
11. Optional Accessories
1
6
5
4
3
2
The anti-roll guide and pressure plate must be parallel to each other.
Fig. 43
Blade holder CE with finger rest (for NB + BB)
Blade rest (ruler) for low-profile blades
The knife guard is folded away in the detail view. Knife guard folded away
Four possible settings
1
6
7
2
Red elements on the blade or knife holders, such as the knife guard and ejector, are protective devices that may not be removed.
7
Blade ejector (1) and knife guard (2)
Integrated lateral shift and stable base
8
Clearance angle adjustment using size 3 Allen key (recommended angle between 2° and 5°)
With finger rest (6) for brush specimen
Clamping lever (8) for lateral shift must point downward to permit unhindered shifting of finger rest.
• when using BB knives, remove the knife support (7).
Conversion of blade holder with anti-roll guide to blade holder with finger rest
Screw off the anti-roll guide.
Unscrew left Allen screw using 2.5 mm Allen key and remove base of anti-roll guide.
Attach the finger rest (6) from the left, tighten the hexagon-head screw using the No. 2.5 key – be careful of the blade ejector!
If you are working with the brush technique, the knife guard must be folded upwards.
56
Fig. 44
Operating Manual V 1.1 – 05/2007
Page 57
Inserting the blade into knife holder CE
11. Optional Accessories
4
9
1. Fold the anti-roll guide system (4) to the left –
while doing so, hold the lever (11) (not the adjusting screw of the anti-roll guide) so that
8
7
the height of the anti-roll guide remains unchanged.
2. Open the clamping lever (10) by turning it coun-
11
10
terclockwise.
3. Carefully insert the blade (9) from above or
from the side between the pressure plate and the blade rest. Ensure that the blade is insert­ed so that it is centered.
Fig. 45
4. Rotate lever (10) clockwise to clamp.
5. Fold the anti-roll guide system (4) back to the
Anti-roll guide system functions as a knife guard here!
side using the lever (11).
Lateral shift for Blade Holder CE
If the sectioning results are not satisfactory, the
8
knife holder can be shifted sideways in order to use another part of the blade. To do so, follow these steps:
1. Release the clamping lever (8) by folding it
back and move the knife holder back until the desired position is reached (3-point click stop enables you to accurately find a new cutting position).
2. Fold the clamping lever (8) forwards for
clamping.
Leica CM1950 – Cryostat
Fig. 46
57
Page 58
11. Optional Accessories
Adjusting the Knife Holder with Anti-roll Guide
You can adjust the height of the anti-roll guide system using the knurled nut (1):
If you turn the nut counterclockwise, the anti­roll guide system moves toward the blade.
If you turn the nut clockwise, the anti-roll guide system moves away from the blade.
If the anti-roll guide system is in the wrong posi­tion relative to the cutter, the following problems will result:
Fig. I: The section rolls over the glass insert of the anti-roll guide system. Error: Glass insert not high enough. Remedy: Turn the knurled nut counterclockwise until the section is pushed between the
I
II
blade and anti-roll guide as shown in Fig. III.
Fig. II: Section tears and block hits the glass insert after sectioning. The section rolls over the glass insert of the anti-roll guide system.
Error: Anti-roll guide system is set too high. Remedy: Turn the knurled nut clockwise until the section is pushed between the blade
and anti-roll guide as shown in Fig. III.
1
Blade holder CE with glass anti-roll guide
Fig. 47
III
58
Generally, we recommend pre-adjusting the anti-roll guide system at a high sec­tion thickness (e.g. 10μm). Start from there and work your way down to the desired section thickness in small increments, readjusting the anti-roll guide system at each increment using the knurled nut.
Fig. 48, Knife holder CN with glass anti-roll guide
Operating Manual V 1.1 – 05/2007
Page 59
11. Optional Accessories
Cleaning the knife holder CE
35
33
34
Daily cleaning
1. Fold the anti-roll guide system (33) to the left while holding it down by the lever (34).
2. Unscrew the bolts (35) of the pressure plate.
3. Afterwards, the pressure plate (36) can be re­moved for cleaning (with alcohol or acetone).
35
36
Fig. 49
For disinfection, commercially avail­able, mild cleaners and disinfectants can be used – we recommend Leica Cry­ofect.
Cleaning the knife holder CN Daily cleaning
For daily cleaning, it is enough to fold the anti-roll guide system forwards and remove the section waste from the knife holder using a dry brush. Please use a cold brush, as otherwise the section waste will thaw and stick to the knife holder.
It is not necessary to oil the parts, such as the T-piece on the microtome base plate, clamping lever etc.
Disinfection (e.g. with Leica Cryofect)
Spray contaminated surfaces with a uniform coat of concentrate or wipe them down with a rag soaked in it, allow it to make contact for 15 minutes, then wipe it off.
Leica CM1950 – Cryostat
Fig. 50
59
Page 60
11. Optional Accessories
Knife holder CN with anti-roll guide – Moving the clamping chucks
1
not visible
3a
The specimen disc 50x80mm is suitable only for section thicknesses up to approx. 5μm (due to the large specimen size).
The large specimen disc (80 x50mm) should pref-
3a
erably be used with knife holder CN and the 16 cm C-profile steel knife.
2
The clamping jaws are factory-installed in the knife holder with a clearance of 64mm. If neces­sary, both clamping chucks can be offset with a
3
clearance of 84mm. Perform the following steps:
Use a 3 mm Allen key to loosen the screw over the clearance angle adjustment (1) and
4
4
remove the segment arch (2) from the knife holder base.
Use a 3 mm Allen key to loosen the screws (4) on the underside of the segment arch.
Carefully pull the clamping jaw (5) on the right upward and insert it in the neighboring hole
5
(6). Tighten the screws from the underside of the segment arch. Repeat on the left side.
Insert the long knife support (3) on the side
6
over the knurled screw (3a) so that the recess faces the user – turn the height adjustment
Fig. 51
knurled screws until the lower stop is reached.
The knife can now be inserted from the side and its height adjusted via the knurled screws (3a).
60
Never work with only one clamping jaw, as this does not ensure the stability required for the sectioning process. Also, a long knife will not be sufficiently protected by the knife guard in this instance.
Operating Manual V 1.1 – 05/2007
Page 61
11. Optional Accessories
19
18
19
Knife guard/lateral movement for knife holder CN
The knife guard (18) is fixed and integrated into the clamping jaws. The knife guard has handles
17
(19) that allow it to be moved. The knife guard is adequate for knives up to 22 cm in length. Always cover exposed parts of the knife blade after
16
sectioning. The anti-roll guide system can be moved side-
ways (only for the 84 mm variant). To better find the mid position, a groove (17) is provided in the
Fig. 52
shaft (16).
Knife holder CNZ with anti-roll guide
Pressure plate permits the full length of the
knife to be utilized.
Use hard metal and steel knives here.
Leica CM1950 – Cryostat
Fig. 53
The height of resharpened knives must be adjusted using knurled screws (3) (approx. 1 mm under the edge of the clamping jaw). Ensure that the knife is adjusted in parallel vertically.
61
Page 62
11. Optional accessories
70mm
Fig. 54
Fig. 55
When using low-profile blades, the blade rest (7) must be inserted into the blade holder first, then the low-profile blade (see page 50, Fig. 40).
Anti-roll guide system Anti-roll plate (with glass stage plate)
Available with various spacers:
70mm - 50μm, for section thickness: < 20μm
70mm -100μm, for section thickness: 20μm ­100 μm
70mm -150 μm, for section thickness: > 100μm
The 50 μm and 100 μm anti-roll plates are included in the standard scope of delivery of the blade holder CE.
Straightedge (blade rest)
Insert for low-profile blades for blade holder CE (0477 43005) - replacement
The blade rest is also included in the standard scope of delivery of the blade holder CE.
Fig. 56
62
Bacterial filter
Bactericidal filter 350/5865, pack of 1 unit. Rec­ommendation: replace bactericidal filter every three months.
(note date on filter when installing it)
Filter bags and bacteria filter must be disposed of according to applicable laboratory regulations for infectious material. Filters must be replaced, not cleaned.
Operating Manual V 1.1 – 05/2007
Page 63
B
Fig. 58
11. Optional Accessories
Replacement filters for extraction system
(pack of 25)
Change when vacuum power is impaired.
When using the cryostat frequently, check the
filter bag daily and replace as required.
Fig. 57
Vacuum extraction system A - Hose B - Hose adapter, black (for the filter
A
G
E
D
F
C
bag in the instrument)
C - Hose adapter, white (for suction nozzle D
or extraction nozzle F)
D - Suction nozzle – with 4 magnets (E) on
knife holder
G - Plastic clip (for parking the suction
nozzle)
Leica CM1950 – Cryostat
Storage system, complete ("hidden")
for installation in rear area of cryostat - for cold
storage of specimen discs and sectioning ac­cessories
(For assembly, refer to page 26)
Fig. 59
63
Page 64
11. Optional Accessories
Specimen discs in various dimensions
The specimen disc 50x80mm is suitable only for section thicknesses up to approx. 5μm (due to the large specimen size).
50x80mm
A
Fig. 61
Shelves
ø 55 ø 40 ø 30
C
28x28mm
37x 37mm
ø 20mm
Fig. 60
B
Dr. Peters specimen discs
D
Under­side
O-rings in various colors
for plate Ø 20mm (red or blue), 10x each
for plate Ø 30mm (red or blue), 10x each
for plate Ø 40mm (red or blue), 10x each
for plate Ø 55mm (red or blue), 10x each
for color-marking of pecimen disc
A Heat extractor with parking station, stationary
for installation in the cryostat - consisting of: support device, heat extractor and parking sta­tion
B Heat and cold extractor
suitable for two different applications: extracts cold when stored in warm environment; ex­tracts heat, when stored in cold environment.
The specimen discs of the Dr. Peters cooling system can be conveniently removed from the shelf using the underside of the heat extrac­tor (removal aid D). Slide the underside over the specimen disc in the direction of the ar­row so that the disc remains in the slot and can be removed from the rack.
Embedding well bar, large, with 3 wells, LxWxH: 30 x 30 x 7 mm Embedding well bar, medium, with 4 wells, LxWxH: 24 x 24 x 6 mm
Fig. 62
Embedding well bar, small,with 4 wells, LxWxH: 18 x 18 x 6 mm
64
Operating Manual V 1.1 – 05/2007
Page 65
Fig. 63
Fig. 64
11. Optional Accessories
Footswitch,
for use with motorized instruments only
Adjustable support,
individually height-adjustable foot rest with 5 set­tings.
Leica CM1950 – Cryostat
Brush shelf,
for use with blade holder CE
Fig. 65
Storage system, movable
for installation in front area of cryostat - for cold storage of preparation tools
Fig. 66
65
Page 66

12. Maintenance and Cleaning

12.1 General Maintenance

The microtome is virtually maintenance-free. To ensure a smooth operation of the instrument over several years we recommend the following:
At least once a year, have the instrument inspected by a qualified service engineer authorized by Leica.
Enter into a service contract at the end of the warranty period. For further information, please contact your local Leica service center.
Clean the instrument daily.
Every day, remove frozen section waste from the cryostat using a cold brush or use the (optional) extraction device.
Remove the section waste tray for emptying.
Remove the storage shelves and the brush shelf for cleaning.
• Remove the closed sliding window from the front by gently lifting it (see p. 63, "Replacing the Fluorescent Lamp").
Do not use organic solvents or any other aggressive substances for cleaning and disinfecting! Use only disinfectants listed in this manual such as Leica Cryofect (alcohol or other common alcohol-based disinfectants).
Emptying the condensate bottle
1
Fig. 67
66
Drain the cleaning liquid through the hose after the prescribed reagent time is over and collect it in the waste container (1).
Check the fill level of the condensate bottle (1) visible in the front panel of the instrument at regular intervals.
The bottle collects the condensate that accumulates during defrosting.
Dispose of the contents of the bottle in accordance with laboratory regulations.
Operating Manual V 1.1 – 05/2007
Page 67
12. Maintenance and Cleaning
Basically, we recommend UV disinfection (see page 35 for detailed instructions). For easy-to-use spray disinfection we recommend Leica Cryofect. (Cryofect is not available in all countries!) The cryostat has to be disinfected after each daily use.
Comply with the instructions for use! The glass anti-roll plate can remain in place during disinfection.
It is not necessary to oil the parts, such as the T-piece on the microtome base plate, clamping lever etc.
In case of visible pollution (such as dust), clean the air inlet opening
(Fig. 10, p. 18) of the condenser on the bottom right-hand side of the instrument using a brush, broom or extraction cleaner in the direction of the louvers.
Do not turn the instrument on before the cryochamber is completely dry!
Frost formation! The front panel and the slit cover of the microtome must be com-
pletely dry before turning on the instrument! Dry all parts completely before reinserting them in the cryochamber!

12.2 Changing Fuses

Leica CM1950 Cryostat
In case of power supply faults, please contact an authorized Leica service
technician immediately.
Do not carry out any repairs on your own as this will invalidate the warranty.
Repairs may only be carried out by qualified service engineers authorized by Leica.
67
Page 68
12. Maintenance and Cleaning

12.3 Replacement of the Lamp/UVC Lamp

Turn the instrument off with the automatic power fuse and pull the power plug, before replacing the lamp!
If the lamp is broken, it must be replaced by the technical service, as the replacement involves a high risk of injury. Beware of the metallic mercury in the UV lamp; handle it carefully and dispose of it properly.
A UVC lamp has an estimated service life of approx. 9,000 hours. Each on/off switching cycle reduces the lamp life by approx. one hour plus burning time (30 minutes or 180 minutes respectively). When both disinfection indicator lights are blinking alternately in control panel field 1, the UVC lamp must be replaced.
Instructions on how to replace the UVC lamp are provided together with the replacement lamp (spare part).
68
Operating Manual V 1.1 – 05/2007
Page 69
12. Maintenance and Cleaning
Slightly lift the sliding window (1) holding it by
1
2
Fig. 68
the grip (2) and pull it out to the front.
Removal of the lamp
The lamp is mounted behind a glare shield and therefore not visible.
Touch the lamp for better orientation.
Lightly tilt the fluorescent tube down to the left and pull it out of the clip (3).
Hold the lamp with both hands and pull it to the left out of the holder (4).
Only use lamps of the same specifica­tion!
Leica CM1950 – Cryostat
Fig. 69
34
Installation of the new lamp
Type: OSRAM DULUX L 18W/840.
Hold the lamp in the correct mounting posi­tion as shown and push it to the left until it engages in the holder.
Fig. 70
69
Page 70

13. Decontamination Certificate (master)

Dear Customer, Any product which is to be returned to Leica Microsystems or serviced on site, must be cleaned and decontaminated in the appropriate manner. nate for CJD and its variants and in this instance following current customer safety guidelines is acceptable. Equipment exposed to CJD samples cannot be returned to Leica Microsystems for repair. This certificate, duly completed, must be placed in the unit prior to packaging or handed over to the service engineer. Packages will not be opened nor servicing commenced until the Company or service engineer have received a satisfactory certificate. Should returned goods be considered a hazard by the Company, they will be returned immediately to the customer at his/her expense. NB: Microtome knives must be in boxes.
Description
Name/Model Fabr. No.
KAT No. Quantity
Tick Box A if applicable. Otherwise please complete all parts of B, providing further information as requested or appropriate.
Leica accepts that it is not possible to decontami-
Yes
A
B
Yes No Please provide further detail here:
Yes No Please provide further detail here:
70
This equipment has not been in contact with unfixed biological samples.
1
This equipment has been exposed internally or externally to hazardous materials as indicated below:
Blood, body fluids, pathological samples
Other biohazards
Chemicals/substances hazardous to health
Other hazards
2
This equipment has been cleaned and decontaminated:
If yes, give details of the method:
If no*, please indicate why not:
* Such equipment must not be returned without the written agreement of Leica Microsystems.
Operating Manual V 1.1 – 05/2007
Page 71
13. Decontamination Certificate (master)
3
Yes No
Important - to avoid refusal of shipment:
Place one copy in the unit prior to packaging or hand it over to the service engineer. Customer assumes all responsibility for the immediate return shipment of articles sent to Leica without proper decontam­ination documentation.
If you have any further questions, please call your local Leica organization.
Leica Internal Use: If applicable, note corresponding Job and RAN-/RGA-Number: Job Sheet No.: ____________________
SU Return Goods Authorisation: ____________________ BU Return Authorization Number: ____________________
The equipment has been prepared to ensure safe handling/transportation. Whenever possible, please use the original transportation case/box.
Signature/Date
Name
Position
eMail
Leica Biosystems Nussloch GmbH Heidelberger Str. 17-19 69226 Nussloch, Germany
Phone: ++49 (0) 6224 143 0 Fax: ++49 (0) 6224 143 200 www.leica-microsystems.com
Leica CM1950 – Cryostat
Institute
Department
Address
Phone Facsimile
71
Page 72
14. EC Declaration of Conformity
EC Declaration of Conformity
We herewith declare, in exclusive responsibility, that the
Leica CM1950 Cryostat
was developed, designed and manufactured to conform with the
Council Directive 2006/95/EC (Low Voltage),
Council Directive 89/336/EEC, Appendix I (Electromagnetic Compatibility) and
European council Directive 98/79/EC (IVD) including their amendments up to the date mentioned below.
The following harmonized standards were applied:
EN 61010-1: 2001
Safety requirements for electrical equipment for measurement, control and laboratory use ­Part 1: General requirements
EN 61010-2-101: 2002
Safety requirements for electrical equipment for measurement, control and laboratory use ­Part 2-101: Particular requirements for in vitro diagnostic (IVD) medical equipment
EN 591: 2001
Instruction for use for in vitro diagnostic instruments for professional use
EN 61326-1: 2006
Electrical equipment for measurement, control and laboratory use - EMC requirements ­Part 1: General requirements
EN 61326-2-6: 2006
Electrical equipment for measurement, control and laboratory use - EMC requirements
EN 61000-3-2: 2000
Electromagnetic compatibility (EMC) Part 3-2: Limits - Limits for harmonic current emissions
EN 61000-3-3: 1995 + A1: 2001
Part 3: Limits ­Section 3: Limitation of voltage changes, voltage fluctuations and flicker in public low-voltage supply systems, for equipment with rated current ≤ 16 A per phase Electromagnetic compatibility (EMC)
EN 418: 1992 EN 1037: 1995
Emergency stop equipment. Safety of machinery. Principles for design. Prevention of unexpected start-up.
In addition, the following in-house standards were applied:
DIN EN ISO 9001: 2000
Quality management system - requirements.
Leica Biosystems Nussloch GmbH Postfach 1120 D-69222 Nussloch
October 11, 2007
. . . . . . . . . . . . . . . . . . . . . . . . . . . .
Anne De Greef-Safft
President Biosystems Division
Page 73
Warranty
Leica Biosystems Nussloch GmbH guarantees that the contractual product delivered has been subjected to a comprehensive quality control proce­dure based on the Leica in-house testing standards, and that the product is faultless and complies with all technical specifications and/or agreed characteristics warranted.
The scope of the warranty is based on the content of the concluded agree­ment. The warranty terms of your Leica sales organization or the organiza­tion from which you have purchased the contractual product shall apply exclusively.
Technical service information
If you require technical service or replacement parts, please contact your Leica sales representative or dealer who sold the product.
Please provide the following information:
Model name and serial number of the instrument.
Location of the instrument and name of the person to contact.
Reason for the service call.
Date of delivery.

15. Warranty and Service

Shutdown and disposal of the instrument
The instrument or parts of the instrument must be disposed of in compli­ance with the local laws. Dispose of UV tube separately.
Leica CM1950 Cryostat
73
Page 74
16. People's Republic of China
74
Operating Manual V 1.1 – 05/2007
Loading...