KROHNE SU600 User Manual

Page 1
Operating Instructions
SU 600
4 20 mA signal conditioning instrument
Page 2

Contents

Contents
1 About this document
1.1 Function .........................
1.2 Target group ......................
1.3 Symbolism used....................
2.1 Authorised personnel ................
2.2 Appropriate use ....................
2.3 Warning about misuse ...............
2.4 General safety instructions ............
2.5 CE conformity .....................
3 Product description
3.1 Conguration ......................
3.2 Principle of operation ................
3.3 Adjustment........................
3.4 Storage and transport ................
4 Mounting
4.1
General instructions .................
4.2 Mounting information ................
4 4 4
6 6 6 6 6
8 9 9 9
11 11
5 Connecting to power supply
5.1 Preparing the connection .............
5.2 Connection procedure ...............
5.3 Wiring plan .......................
14 15 16
6 Setup with the integrated indicating and
adjustment unit
6.1 Adjustment system ..................
6.2 Setup procedure ...................
6.3 Application example .................
18 19 24
7 Maintenance and fault rectication
7.1 Maintenance ......................
7.2 Fault rectication ...................
2 SU 600 - 4 20 mA signal conditioning
27 27
30652-EN-050525
Page 3
Contents
7.3 Instrument repair ...................
8 Dismounting
8.1 Dismounting procedure ...............
8.2 Disposal .........................
9 Supplement
9.1 Technical data .....................
9.2 Dimensions .......................
9.3 Certicate ........................
28
29 29
30 33 34
30652-EN-050525
SU 600 - 4 … 20 mA signal conditioning 3
Page 4
About this document

1 About this document

1.1 Function

This operating instructions manual has all the informa­tion you need for quick setup and safe operation of SU
600. Please read this manual before you start setup.

1.2 Target group

This operating instructions manual is directed to trained personnel. The contents of this manual should be made available to these personnel and put into practice by them.

1.3 Symbolism used

Information, tip, note
This symbol indicates helpful additional information.
Caution, warning, danger
This symbol informs you of a dangerous situation that could occur. Ignoring this cautionary note can impair the person and/or the instrument.
Ex applications
This symbol indicates special instructions for E xap­plications.
l List
The dot set in front indicates a list with no implied sequence.
à Action
This arrow indicates a single action.
4 SU 600 - 4 20 mA signal conditioning
30652-EN-050525
Page 5
1 Sequence
Numbers set in front indicate successive steps in a procedure.
About this document
30652-EN-050525
SU 600 - 4 … 20 mA signal conditioning 5
Page 6
For your safety

2 For your safety

2.1 Authorised personnel

All operations described in this operating instructions manual must be carried out only by trained specialist personnel authorised by the operator. For safety and warranty reasons, any internal work on the instruments must be carried out only by personnel authorised by the manufacturer.

2.2 Appropriate use

SU 600 is a universal signal conditioning instrument and power supply unit for connection of a 4 20 mA/HART sensor.

2.3 Warning about misuse

Inappropriate or incorrect use of the instrument can give rise to application-specific hazards, e.g. vessel overll or damage to system components through incorrect mounting or adjustment.

2.4 General safety instructions

SU 600 is a high-tech instrument requiring the strict observance of standard regulations and guidelines. The user must take note of the safety instructions in this operating instructions manual, the country-specific installation standards (e.g. the VDE regulations in Germany) as well as all prevailing safety regulations and accident prevention rules.

2.5 CE conformity

SU 600 is in CE conformity with EMC (89/336/EWG) and NSR (73/23/EWG).
6 SU 600 - 4 20 mA signal conditioning
30652-EN-050525
Page 7
For your safety
Conformity has been judged acc. to the following
standards:
l EMC:
- Emission EN 61326: 1997 (class A)
- Susceptibility EN 61326: 1997/A1: 1998
l NSR: EN 61010-1: 2001
SU 600 is designed for use in an industrial environment. Nevertheless, electromagnetic interference from elect-
rical conductors and radiated emissions must be taken into account, as is usual with a class A instrument acc. to EN 61326. If SU 600 is used in a different environment, the electromagnetic compatibility to other instruments must be ensured by suitable measures.
30652-EN-050525
SU 600 - 4 … 20 mA signal conditioning 7
Page 8
Product description

3 Product description

3.1 Conguration
Scope of delivery
Overview
The scope of delivery encompasses:
l Indicating and signal conditioning instrument SU 600 l Mounting set l Identication label l Documentation
- this operating instructions manual
1
2
7
6
+
0
1
F
2
E
3
D
4
5
C
6
B
7
A
8
9
-
1 2 3
4
5
Fig. 1: SU 600
1 Status indications operating relay 1 and 2 2 Status indication fail safe relay 3 Status indication operation 4 [+/-] adjustment keys 5 Insertable tag for identication of the measurement loop 6 Function switch 7 LC display
8 SU 600 - 4 20 mA signal conditioning
30652-EN-050525
Page 9

3.2 Principle of operation

Product description
Area of application
Physical principle
Power supply
SU 600 is a universal single signal conditioning instrument with integrated level switches and display for continuously measuring sensors. At the same time, it can serve as power supply unit for connected sensors.
SU 600 is designed for connection of an individual 4 20 mA sensor. The instrument is suitable for carrier
rail, panel and surface mounting.
SU 600 signal conditioning instrument can power the connected sensor and also evaluate its measurement signals. The requested parameter is indicated in the display and, in addition, outputted to the integrated current output for further processing. Hence the mea­suring signal can be transmitted to a remote display or a connected control system. In addition, two level relays for control of pumps or other actuators are integrated.
Wide-range power supply unit with 20 253 VAC/DC for world-wide use.
You will find detailed information on the power supply in the "Technical data" in the "Supplement".

3.3 Adjustment

The adjustment of SU 600 is carried out via the integrated keys and a 16-step function switch.

3.4 Storage and transport

Packaging
30652-EN-050525
SU 600 - 4 … 20 mA signal conditioning 9
Your instrument was protected by packaging during
transport. Its capacity to handle normal loads during transport is assured by a test acc. to DIN EN 24180.
The packaging of standard instruments consists of environment-friendly, recyclable cardboard. For special versions PE foam or PE foil is also used. Dispose of the packaging material via specialised recycling companies.
Page 10
Product description
Storage and transport temperature
l Storage and transport temperature see "Supplement
Technical data Ambient conditions"
l Relative humidity 20 85 %
10 SU 600 - 4 … 20 mA signal conditioning
30652-EN-050525
Page 11

4 Mounting

4.1 General instructions

Mounting
Installation location
Carrier rail mounting
You can mount theSU 600 signal conditioning instru- ment onto a carrier rail 35x7.5 acc. to DIN EN 50022, or directly on the wall by means of the three screws or in a front panel or door of a switching cabinet.

4.2 Mounting information

1 Place the adapter plate [1] to the rear of SU 600
(spring of the adapter plate downward) and fasten
the plate with screw [2](M4x6).
2 PlaceSU 600 against the carrier rail [3] from below
and push the instrument upward until it snaps in.
21 3
61
43,5
Fig. 2: Carrier rail mounting
1 Adapter plate 2 Screw (M4x6) 3 Carrier rail
Wall mounting
1 Insert the metal strap [1] from above into the housing
cut-out.
2 Fasten the instrument directly to the wall with the
three screws (ø max. 4 mm).
30652-EN-050525
SU 600 - 4 … 20 mA signal conditioning 11
Page 12
Mounting
1
105
ø 4,5*
80
Fig. 3: Wall mounting
1 Metal strap
Panel mounting
1 Remove the pluggable terminal strips from the top
and bottom.
2 Screw pin [3] into the rear of SU 600 and tighten with
a screwdriver.
3 Insert SU 600 from the front into the front panel [1].
4 Place the clamping strap [2] from the back over the
pin [3] and pull it with the knurled nut [4] against the panel [1].
30652-EN-050525
12 SU 600 - 4 … 20 mA signal conditioning
Page 13
Mounting
1
~90
Fig. 4: Panel mounting
1 Front panel 2 Terminal strap 3 Pin 4 Knurled nut
92
32 4
30652-EN-050525
SU 600 - 4 … 20 mA signal conditioning 13
Page 14
Connecting to power supply

5 Connecting to power supply

5.1 Preparing the connection

Note safety instructions
Take note of safety
instructions for Ex applications
Select power supply
Select connection cable
Cable screening and
grounding
Always observe the following safety instructions:
l Connect only in the complete absence of line voltage l If overvoltages are expected, overvoltage arresters
should be installed.
In hazardous areas you should take note of the appropriate regulations, conformity and type approval certicates of the sensors and power supply units.
The power supply can be 20 253 VAC, 50/60 Hzor 20 253 VDC.
Power supply of SU 600 is connected with standard cable acc. to the national installation standards.
Standard two-wire cable without screening can be used to connect sensors. If electromagnetic interference is expected, screened cable must be used.
Connect the cable screen on both ends to ground potential. In the sensor, the screen must be connected directly to the internal ground terminal. The ground terminal outside on the sensor housing must be connected to the potential equalisation.
If potential equalisation currents are expected, the screen connection on SU 600 must be made via a ceramic capacitor (e.g. 1 nF, 1500 V). The low frequency potential equalisation currents are thus sup­pressed, but the protective eect against high frequency interference signals remains.
Select connection cable for Ex appli­cations
Take note of the corresponding installation regulations for Ex applications. In particular, make sure that no potential equalisation currents flow over the cable screen. In case of grounding on both sides this can be achieved by the use of a capacitor or a separate potential equalisation.
14 SU 600 - 4 … 20 mA signal conditioning
30652-EN-050525
Page 15
Connecting to power supply

5.2 Connection procedure

Proceed as follows:
1 Mount SU 600
2 Connect sensor cable to terminal 1 and 2, and where
applicable, connect the screen
3 Connect power supply (switched o) to terminal 5
and 6
4 Where applicable, connect fail safe and operating
relay as well as current output
The electrical connection is finished.
30652-EN-050525
SU 600 - 4 … 20 mA signal conditioning 15
Page 16
Connecting to power supply

5.3 Wiring plan

Overview
-
+
2
1
1 2
1
2
-
+
4
3
1
2
on
%
18 1716 15 14 13 1211 10 9 876 5
3 4 5 6
Fig. 5: Terminal assignment SU 600
1 Measurement data input, optionally available with sensor power
supply
2 Current output 3 Fail safe relay 4 Relay 2 5 Relay 1 6 Power supply
N- L1+
Information:
In the menu, it is possible to switch over between active and passive operation of the measurement data input:
l In active mode,SU 600 provides the power for the
connected sensors. Power supply and measurement data are transmitted over the same two-wire cable. This mode is provided for connection of transmitters without separate power supply (sensors in two-wire version).
16 SU 600 - 4 … 20 mA signal conditioning
30652-EN-050525
Page 17
Connecting to power supply
l In passive mode, the sensors are not powered - only
the measured value is transmitted. This input is for transmitters with their own separate power supply (sensors in four-wire version). Furthermore,SU 600 can be connected to an existing circuit like a standard current meter.
30652-EN-050525
SU 600 - 4 … 20 mA signal conditioning 17
Page 18

Setup with the integrated indicating and adjustment unit

6 Setup with the integrated indicating and
adjustment unit

6.1 Adjustment system

The integrated indicating and adjustment unit is for measured value display, adjustment and diagnosis ofSU
600. Indication and adjustment is carried out in the front via a clear LC-display and a function switch as well as two keys.
To open the cover, insert a screwdriver in the two slots on the top and turn it slightly.
1
2
7
6
+
0
1
F
2
E
3
D
4
5
C
6
B
7
A
8
9
Fig. 6: Indicating and adjustment elements
1 Status indications operating relay 1 and 2 2 Status indication fail safe relay 3 Status indication operation 4 Adjustment keys +/- 5 Insertable tag for identication of the measurement loop 6 Function switch 7 LC display
-
1 2 3
4
5
30652-EN-050525
18 SU 600 - 4 … 20 mA signal conditioning
Page 19
Setup with the integrated indicating and adjustment unit
Key functions
Set up
l [Function switch] for selection of:
- Adjustment
- Relay switching points
- Scaling
- Current output
- Integration time
- Oset correction
- Current input active/passive
l [+/-] key:
- Change value of the parameter
By pushing the [+/-] key, you change the individual parameters of the selected function. In this phase, the processed parameter is ashing. Bybriefly pushing the two keys simultaneously, you save your settings. "Save" is briey displayed.

6.2 Setup procedure

Setup encompasses mainly the adjustment of the measurement loop. A scaling of the measured value for the LC display and the adaptation of the relay switching points are additional settings. Additional setup steps might be, if necessary, setting an integration time to smooth the measured value or modifying the current output characteristics.
To clearly denote the measuring unit, the supplied labels can be inserted in the cover. In case several SU 600 are used, each measurement loop should be clearly labelled.
Switch on phase
After being switched on, SU 600 first of all carries out a short self-check. The following steps are carried out:
l Internal check of the electronics l indication of the rmware version l the output signal jumps briey to the set fault current
Then the actual measured value is displayed and the corresponding current is transmitted to the current output.
30652-EN-050525
SU 600 - 4 … 20 mA signal conditioning 19
Page 20
Setup with the integrated indicating and adjustment unit
Measured value indication
Function switch
Current input active/pas-
sive
The measured value indication shows the digital indication value and also an analogue bargraph. Here, it is absolutely necessary that the function switch is set to position [ 0]("OPERATE").
The following functions can be selected via the rotary switch:
l 0: Measured value display and simulation l 1: Relay 1 switching point ON l 2: Relay 1 switching point OFF l 3: Relay 2 switching point ON l 4: Relay 2 switching point OFF l 5: Decimal point position of the scaling l 6: Scaling for 100 % l 7: Scaling for 0 % l 8: Changeover current output 0/4 ... 20 mA l 9: Damping of the measured value (integration time) l A: Oset correction l B: Min. adjustment in % by modifying the level l C: Max. adjustment in % by modifying the level l D: Min. adjustment in mA without modifying the level l E: Max. adjustment in mA without modifying the level l F: Changeover current input active/passive
l In active mode, SU 600 provides the power supply
for the connected sensor. Power supply and mea­sured value transmission are carried on the same two-wire cable. This mode is provided for connection of transmitters without separate power supply (sen­sors in two-wire version).
l In passive mode, no sensors are powered, only the
measured value is transmitted. This mode is for connection of transmitters with own, separate power supply (sensor in four-wire version).
à Set the function switch to position [F] and select the
suitable mode with the [+/-] keys. Save your settings by pushing both keys simultaneously.
Oset correction
When a pressure transmitter is used, an oset correc­tion should be carried out rst. These instruments are factory-set in a certain position. If the pressure trans­mitter is now mounted in a dierent position, its
20 SU 600 - 4 … 20 mA signal conditioning
30652-EN-050525
Page 21
Setup with the integrated indicating and adjustment unit
measuring range is shifted slightly. The zero point is readjusted by carrying out the oset correction in uncovered (unpressurised) condition. The complete measuring range will be shifted by the deviating amount.
à Make sure that the pressure transmitter is completely
unpressurised, uncovered (unimmersed) and moun­ted in its nal position.
à Set the function switch to position [ A ]. The display
indicates now the actual sensor current in mA. Save the current status by pushing the [ +/- ] keys simultaneously.
Adjustment in mA without changing the level
Adjustment in % by chan­ging the level
For this adjustment procedure, two sensor current values (4 20 mA) must be entered corresponding to the levels 0 % and 100 %.
When using a pressure transmitter, an oset correction should be carried out to reach maximum accuracy. Carry out this oset correction before adjustment and with the sensor uncovered (unimmersed).
à Select now position [D] or [E] on the function switch,
enter the current values in mA for min. or max. adjustment and save your settings.
During this adjustment procedure, the current level is assigned to a certain percentage value. For this reason, percentage values that correspond to the actual filling levels must be entered for the min. and max. adjust­ment. The ideal calibration is at 0 % and 100 %. Because it is not always possible to empty or fill a vessel completely, any value can of course be entered. The greater the dierence between the two adjustment points, the more precise the measurement. It does not matter which value is entered rst.
à Select position [B] or [C] on the function switch, enter
the percentage values for the min. and max. adjustment and save your settings.
Relay outputs
Two operating relays are integrated inSU 600 for level detection. First of all, determine at which switching points the relays should switch on and off again.
30652-EN-050525
SU 600 - 4 … 20 mA signal conditioning 21
Page 22
Setup with the integrated indicating and adjustment unit
Furthermore you have to distinguish between the relay
modes overll and dry run protection. You switch over by exchanging the ON/OFF values of the relay.
l Overll protection: Relay is switched owhen the
max. level is exceeded (safe currentless condition), relay is switched on again when the level falls below the min. level (switch on point < switch opoint).
l Dry run protection: Relay is switched off when the
level falls below the min. level (safe currentless condition), relay is switched on again when the max. level is exceeded (switch on point > switch opoint).
à To adjust the on and oswitching points of relay 1,
set the function switch to position [1] or [2], enter the switching points for ON or OFF and save your settings. If necessary, proceed in the same way with relay 2 (position [3] or [4]).
1
off (on)
on (off)
t
LED
on
2
off
on
3
off
1 Level 2 Mode overll protection 3 Mode dry run protection
22 SU 600 - 4 … 20 mA signal conditioning
30652-EN-050525
Page 23
Setup with the integrated indicating and adjustment unit
Scaling
Damping
Scaling is the conversion of the measured value into a certain parameter or measuring unit. Instead of the percentage value, the volume can be displayed, e.g. in l. Indicating values between max.-9999 and +9999 are possible.
à First of all, determine the max. indicating range and
the number of decimal places after the decimal point
-- max. four decimal places can be displayed. Select position [ 5 ] of the function switch, place the decimal point in the required position and save your settings.
à Now select position [6] or [7] on the function switch,
enter the requested values for 100 % or 0 % and save your settings.
To suppress fluctuation in the measured value display, e.g. caused by an agitated product surface, an integra­tion time can be set. This time can be between 0 and 250 seconds. Keep in mind that this setting will increase the reaction time of the measurement, and that the reaction to quick changes of the measured value will be delayed. In general, a time of a few seconds is sucient to smooth the measured value display.
à Now select position [9] on the function switch, enter
the requested value and save your settings.
Currentoutput0/4 20 mA
The characteristics of the current output can be switched over from 4 20 mA to 0 20 mA.
à Now select position [8] on the function switch, adjust
the requested characteristics curve and save your settings.
Simulation
30652-EN-050525
SU 600 - 4 … 20 mA signal conditioning 23
To check if the settings ofSU 600 are correct, the
simulation mode can be used. Any individual measured value can be simulated and e.g. the correct behaviour of the relays and the connected instruments can be checked.
Switching over between the functions "OPERATE" and "Simulation" is done by pushing the [+/-] keys for at least
3 seconds.In the simulation mode, the set value ashes on the display. The simulation can be switched off by
Page 24
Setup with the integrated indicating and adjustment unit
pushing both keys again for approx. 3 seconds. If no key is pushed, the display switches automatically over to "OPERATE" after approx. 60 minutes and the simulation is terminated.
à To start simulation, push the [ +/- ] keys simultan-
eously until the displayed value starts ashing (approx. 3 seconds). By pushing either o f the [ +/- ] keys, you can set the requested simulation value and make your check.
Reset
With a reset, all values set by the user will be lost and are reset to factory settings.
à Interrupt the power supply of SU 600. Push the [+/-]
keys simultaneously and hold them while you switch on the power supply. The display shows "RES" and the default settings will be restored. These are the following:
l Relay switch on points: 10 % l Relay switch opoints: 100 % l Decimal point: 888.8 l Display min: 0 l Display max: 100.0 l Integration time: 0 s l Oset correction: 0 l Current output: 4 20 mA l Min. adjustment: 0 % or 4 mA l Max. adjustment: 100 % or 20 mA

6.3 Application example

l An upright cylindrical (linear) tank has a volume of
2700 l
l The max. filling volume is 2650 l, the min. filling
volume 50 l
l The level sensor in the vessel is a pressure
transmitter (passive), outputting a standardised 4 20 mA signal
l For further processing, an additional 4 20 mA
output signal is required
30652-EN-050525
24 SU 600 - 4 … 20 mA signal conditioning
Page 25
General procedure
Setup with the integrated indicating and adjustment unit
l The full and empty adjustment was carried out
directly in the sensor. The following values are outputted:
- max. volume (display max.) 20 mA = 2650 l
- min. volume (display min.) 4 mA = 50 l
l Relay 1 should switch on an emptying pump at a
level of 90 % and switch othe pump at a level of
10 %
1 Select the following functions with the rotary switch.
2 Push one of the two [+/-] adjustment keys, the digital
indication ashes (if the rotary switch is set to "OPERATE", the [+/-] keys are inoperable).
3 Adjust the requested value with the [+/-] keys. If you
hold the key pushed, the digital indication changes the value more quickly.
4 Save your settings by pushing the [+/-] keys
simultaneously.
Current output
à Set the function switch to position [8]. Select with the
[+/-] keys the measured value output 4 20 mA or
0 20 mA. For this example, choose 4 20.
The displayed numbers have the following meanings:
l 0 – 20 = 0 … 20 mA l 4 – 20 = 4 … 20 mA
à Save the value by pushing the [+/-] keys simultan-
eously.
To scale the indication of SU 600, the lling volumes for 0 % and 100 %. The vessel does not have to be lled or
emptied.
Scaled indication at 0 %
Scaled indication at 100 %
30652-EN-050525
SU 600 - 4 … 20 mA signal conditioning 25
1 Set the function switch to position [7](display min.).
2 Set the value to 50 via the [+/-] keys.
3 Save the value by pushing both keys simultaneously.
1 Set the function switch to position [6](display max.).
2 Set the value to 2650 via the [+/-] keys.
3 Save the value by pushing both keys simultaneously.
Page 26
Setup with the integrated indicating and adjustment unit
Decimal point
Relay
Because the range of the example vessel is between 50 l to 2650 l, you will need all four positions of the digital indication.
1 Set the function switch to position [5](Decimal Point).
2 By pushing the [+/-] keys, you move the decimal
point.
3 Save the value by pushing both keys simultaneously.
l Set the function switch to position [ 1 ](relay 1 on).
For the given example, select the value 90.0 (90.0 %) with the [ +/- ] keys. The internal relay 1 then switches on when this value is reached. Save the value by pushing the two keys simultaneously.
l Set the function switch to position [ 2](relay 1 off).
Select the value 10.0 (10.0 %) with the [ +/-] keys. The internal relay 1 then switches owhen the value
falls below this value. The respective relay control lamp lights when the relay is energized. If the switching points are too close together (<0.1 %), the appropriate LED ashes. The relay takes on the safe condition. Inmode"OPERATE " an error message is outputted on the display.
l If you want to control additional relays, proceed in the
same way, as with relay 1. The switch positions for setting the respective relays are stated in the list of the function switch under "Setup procedure".
Note:
If you want to change the mode (i.e. the switching function of the relays), you have to exchange the Onand Off values.
26 SU 600 - 4 … 20 mA signal conditioning
30652-EN-050525
Page 27
Maintenance and fault rectication
7 Maintenance and fault rectication

7.1 Maintenance

When used as directed in normal operation, SU 600 is completely maintenance-free.
7.2 Fault rectication
Causes of malfunction
Fault rectication
Fault messages
SU 600 oers maximum reliability. Nevertheless faults can occur during operation. These may be caused by the following, e.g.:
l Measured value of the sensor not correct l Power supply l Interference on the cables
The rst measures to be taken are to check the input/ output signals as well as to evaluate the error messages via the display. The procedure is described below. In many cases, the causes can be determined in this way and faults can be rectied.
? E003
l CRC-error
à Carry out a reset
à Return instrument for repair
? E014
l Short-circuit
à Eliminate short-circuit
à If necessary, exchange defective sensor
? E015
l Line break
à Repair sensor cable or replace it
à Check sensor connection
30652-EN-050525
SU 600 - 4 … 20 mA signal conditioning 27
Page 28
Maintenance and fault rectication
? E016
l Empty/full adjustment reversed
à Carry out a fresh adjustment
? E017
l Adjustment span too small
à Carry out a fresh adjustment and increase the
distance between min. and max. adjustment
? E021
l Scaling span too small
à Carry out a fresh scaling, increase the distance
between min. and max. scaling.
? E110
l Relay switching points too close together
à Increase the difference between the two relay
switching points

7.3 Instrument repair

If a repair is necessary, please proceed as follows:
You can download a return form from our Internet
homepage media-lounge/PDF-Download/Specimen_e.pdf.
By doing this you help us carry out the repair quickly and without having to call back for needed information.
l Print and ll out one form per instrument l Clean the instrument and pack it damage-proof l Attach the completed form and possibly also a safety
data sheet to the instrument.
28 SU 600 - 4 20 mA signal conditioning
http://www.krohne-mar.com/leadmin/
30652-EN-050525
Page 29
Dismounting

8 Dismounting

8.1 Dismounting procedure

Warning:
Before dismounting, be aware of dangerous process conditions such as e.g. pressure in the vessel, high temperatures, corrosive or toxic products etc.
Take note of chapters "Mounting" and "Connecting to power supply" and carry out the listed steps in reverse
order.

8.2 Disposal

SU 600 consists of materials which can be recycled by specialised recycling companies. We have purposely designed the electronic modules to be easily separable. Mark the instrument as scrap and dispose of it according to government regulations (electronic scrap ordinance,
…).
Materials: see "Technical data"
If you cannot dispose of the instrument properly, please
contact us about disposal methods or return.
30652-EN-050525
SU 600 - 4 … 20 mA signal conditioning 29
Page 30
Supplement

9 Supplement

9.1 Technical data

General data
Series instrument for panel or wall mounting or
mounting on carrier rail 35x7.5 acc. to
EN 50022
Dimensions W = 96 mm (3.78 in), H = 104 mm (4.09 in),
D = 90 mm (3.54 in)
Weight approx. 400 g (0.88 lbs) Housing material plastic ABS/POM Screw terminals max. wire cross section 1.5 mm²
Power supply
Power supply 20 ... 253 VAC, 50/60 Hz; 20 ... 253 VDC Power consumption 12 VA, 4 W
Sensor input
N
umber of sensors 1x 4 20 mA
Kind of input (switch over)
- mode active sensor is powered by SU 600
- mode passive sensor has own power supply
Measured value transmission 4 … 20 mA analogue Terminal voltage mode active 19.5 … 14.5 V at 4 … 20 mA Current limitation mode active approx. 30 mA Inner resistance mode passive < 250 Ohm Detection line break <= 3.6 mA Detection shortcircuit >= 21 mA Adjustment range
- Empty adjustment 3.8 … 20.2 mA
- Full adjustment 4.1 … 20.5 mA
- min. adjustment delta 300 µ
A
Connection cable 2-wire standard cable (screening recom-
mended)
30 SU 600 - 4 … 20 mA signal conditioning
30652-EN-050525
Page 31
Supplement
Relay outputs
Quantity 2x operating relay, 1x fail safe relay Contact oating spdt Contact material AG NI 0.15 hard gold-plated Turn-on voltage min. 10 mVDC, max. 250 VAC/DC Switching current min. 10 µADC, max. 3 AAC, 1 ADC Breaking capacitance min 50 mW, max. 500 VA, max. 54 WDC Min. switching hysteresis 0.5 %
Current outputs
Quantity 1x output Range 0/4 20 mA Resolution 0.1 % or 20 µA Max
. load 500 Ohm Fault signal 22 mA Linearity error 0.1 % (relating to 20 mA) Temperature error 0.01 %/K (relating to 20 mA)
Displays
Measured value indication
- LC display (45x45 mm) digital and quasianalogue indication
- max. indicating range -9999 ... 9999
LED displays
- status indication operating vol-
1x LED green
tage
- status indication fault signal 1x LED red
- status indication operating re-
2x LED yellow
lay 1/2
Adjustment
Adjustment elements 2x keys in the front, 1x function switch
Ambient conditions
Ambient temperature -20
+60°C (-4 +140°F)
Storage and transport temperature -40 +80°C (-40 +176°F)
30652-EN-050525
SU 600 - 4 … 20 mA signal conditioning 31
Page 32
Supplement
Electrical protective measures
Protection
- Wall/rail mounting IP 20
- Panel mounting IP 40
Overvoltage category II Protection class II
Electrical separating measures
Reliable separation acc. to VDE 0106 part 1 between power supply, sensor input and digital part
- Reference voltage 250 V
- insulation resistance 3.75 kV
Galvanic separation between relay output and digital part
- Reference voltage 250 V
- insulation resistance 4 kV
32 SU 600 - 4 … 20 mA signal conditioning
30652-EN-050525
Page 33

9.2 Dimensions

Supplement
1
2
Fig. 7: Dimensions SU 600
")
")
")
8
64
/
8
/
/
3
1
5
92mm (3
111mm (4
102mm (4
69mm (2 23/32") 2mm (5/64")19mm (3/4")
%
96mm (3 25/32")
")
1
32
2
/
25
on
96mm (3
30652-EN-050525
SU 600 - 4 … 20 mA signal conditioning 33
Page 34
Supplement
9.3 Certicate
CE declaration of conformity
Fig. 8: CE declaration of conformity
34 SU 600 - 4 … 20 mA signal conditioning
30652-EN-050525
Page 35
30652-EN-050525
SU 600 - 4 20 mA signal conditioning 35
Page 36
Subject to change without notice
30652-EN-050525
Loading...