KROHNE ORP8510 User Manual

Page 1
Handbook
Handbook
SMARTSENS ORP 8510
SMARTSENS ORP 8510
SMARTSENS ORP 8510SMARTSENS ORP 8510
Digital ORP sensor for the water and wastewater industries
Electronic Revision: ER 3.3.x (SW.REV. 1.x.x)
© KROHNE 11/2013 - 4002547502 - MA SMARTSENS ORP 8510 R02 en
Page 2
: IMPRINT :::::::::::::::::::::::::::::::::::::::
SMARTSENS ORP 8510
All rights reserved. It is prohibited to reproduce this documentation, or any part thereof, without the prior written authorisation of KROHNE Messtechnik GmbH.
Subject to change without notice.
Copyright 2013 by KROHNE Messtechnik GmbH - Ludwig-Krohne-Str. 5 - 47058 Duisburg (Germany)
2
www.krohne.com 11/2013 - 4002547502 - MA SMARTSENS ORP 8510 R02 en
Page 3
SMARTSENS ORP 8510
CONTENTS
1 Safety instructions 5
1.1 Intended use ..................................................................................................................... 5
1.2 Safety instructions from the manufacturer ..................................................................... 5
1.2.1 Copyright and data protection ................................................................................................ 5
1.2.2 Disclaimer ............................................................................................................................... 6
1.2.3 Product liability and warranty ................................................................................................ 6
1.2.4 Information concerning the documentation........................................................................... 6
1.2.5 Warnings and symbols used................................................................................................... 7
1.3 Safety instructions for the operator................................................................................. 7
2 Device description 8
2.1 Scope of delivery............................................................................................................... 8
2.2 Device description ............................................................................................................ 9
2.2.1 ORP sensor.............................................................................................................................. 9
2.3 Nameplate ...................................................................................................................... 10
3 Installation 11
3.1 General notes on installation ......................................................................................... 11
3.2 Storage and transport .................................................................................................... 11
3.3 Installation procedure .................................................................................................... 12
3.4 Pre-installation requirements ....................................................................................... 12
3.5 Electrical connection......................................................................................................13
3.5.1 Connecting the cable to the sensor...................................................................................... 13
3.5.2 Connecting the sensor cable ................................................................................................ 14
3.5.3 Connecting with junction box SJB 200 W.............................................................................. 14
3.6 Installing the sensor....................................................................................................... 16
3.6.1 General installation instructions.......................................................................................... 16
4 Operation 17
4.1 Calibration ...................................................................................................................... 17
4.1.1 Calibration with PACTware
4.1.2 Calibration with HART® Handheld 475 FIELD COMMUNICATOR ........................................ 18
4.2 Troubleshooting.............................................................................................................. 19
4.3 Status messages and diagnostic information................................................................ 20
TM........................................................................................................................ 17
5 Service 21
5.1 Maintenance ................................................................................................................... 21
5.1.1 Cleaning ................................................................................................................................ 21
5.1.2 Aging and re-calibration ....................................................................................................... 21
5.2 Availability of services .................................................................................................... 21
5.3 Returning the device to the manufacturer..................................................................... 22
5.3.1 General information.............................................................................................................. 22
5.3.2 Form (for copying) to accompany a returned device............................................................ 23
5.4 Disposal .......................................................................................................................... 23
www.krohne.com11/2013 - 4002547502 - MA SMARTSENS ORP 8510 R02 en
3
Page 4
CONTENTS
SMARTSENS ORP 8510
6 Technical data 24
6.1 Measuring principle........................................................................................................24
6.1.1 ORP measurement................................................................................................................ 24
6.2 Technical data................................................................................................................. 25
6.3 Dimensions ..................................................................................................................... 27
7 Description of HART interface 28
7.1 General description ........................................................................................................28
7.2 Software history ............................................................................................................. 28
7.3 Connection variants........................................................................................................29
7.3.1 Point-to-Point connection - analogue / digital mode........................................................... 30
7.4 Inputs/outputs and HART® dynamic variables and device variables............................ 31
7.5 Field Communicator 475 (FC 475).................................................................................. 31
7.5.1 Installation ............................................................................................................................ 31
7.5.2 Operation............................................................................................................................... 32
7.6 Field Device Tool / Device Type Manager (FDT/DTM).................................................... 32
7.6.1 Installation ............................................................................................................................ 32
7.7 Overview Basic-DD menu tree (positions in menu tree) ............................................... 33
7.8 Basic-DD menu tree (details for settings) ..................................................................... 34
8 Notes 37
4
www.krohne.com 11/2013 - 4002547502 - MA SMARTSENS ORP 8510 R02 en
Page 5
SMARTSENS ORP 8510
1.1 Intended use
CAUTION!
Responsibility for the use of the measuring devices with regard to suitability, intended use and corrosion resistance of the used materials against the measured fluid lies solely with the operator.
INFORMATION!
The manufacturer is not liable for any damage resulting from improper use or use for other than the intended purpose.
The intended use of SMARTSENS ORP 8510 sensor is the measurement of ORP value in water applications.
1.2 Safety instructions from the manufacturer
1.2.1 Copyright and data protection
The contents of this document have been created with great care. Nevertheless, we provide no guarantee that the contents are correct, complete or up-to-date.
SAFETY INSTRUCTIONS 1
The contents and works in this document are subject to copyright. Contributions from third parties are identified as such. Reproduction, processing, dissemination and any type of use beyond what is permitted under copyright requires written authorisation from the respective author and/or the manufacturer.
The manufacturer tries always to observe the copyrights of others, and to draw on works created in-house or works in the public domain.
The collection of personal data (such as names, street addresses or e-mail addresses) in the manufacturer's documents is always on a voluntary basis whenever possible. Whenever feasible, it is always possible to make use of the offerings and services without providing any personal data.
We draw your attention to the fact that data transmission over the Internet (e.g. when communicating by e-mail) may involve gaps in security. It is not possible to protect such data completely against access by third parties.
We hereby expressly prohibit the use of the contact data published as part of our duty to publish an imprint for the purpose of sending us any advertising or informational materials that we have not expressly requested.
www.krohne.com11/2013 - 4002547502 - MA SMARTSENS ORP 8510 R02 en
5
Page 6
1 SAFETY INSTRUCTIONS
1.2.2 Disclaimer
The manufacturer will not be liable for any damage of any kind by using its product, including, but not limited to direct, indirect or incidental and consequential damages.
This disclaimer does not apply in case the manufacturer has acted on purpose or with gross negligence. In the event any applicable law does not allow such limitations on implied warranties or the exclusion of limitation of certain damages, you may, if such law applies to you, not be subject to some or all of the above disclaimer, exclusions or limitations.
Any product purchased from the manufacturer is warranted in accordance with the relevant product documentation and our Terms and Conditions of Sale.
The manufacturer reserves the right to alter the content of its documents, including this disclaimer in any way, at any time, for any reason, without prior notification, and will not be liable in any way for possible consequences of such changes.
1.2.3 Product liability and warranty
The operator shall bear responsibility for the suitability of the device for the specific purpose. The manufacturer accepts no liability for the consequences of misuse by the operator. Improper installation and operation of the devices (systems) will cause the warranty to be void. The respective "Standard Terms and Conditions" which form the basis for the sales contract shall also apply.
SMARTSENS ORP 8510
1.2.4 Information concerning the documentation
To prevent any injury to the user or damage to the device it is essential that you read the information in this document and observe applicable national standards, safety requirements and accident prevention regulations.
If this document is not in your native language and if you have any problems understanding the text, we advise you to contact your local office for assistance. The manufacturer can not accept responsibility for any damage or injury caused by misunderstanding of the information in this document.
This document is provided to help you establish operating conditions, which will permit safe and efficient use of this device. Special considerations and precautions are also described in the document, which appear in the form of underneath icons.
6
www.krohne.com 11/2013 - 4002547502 - MA SMARTSENS ORP 8510 R02 en
Page 7
SMARTSENS ORP 8510
1.2.5 Warnings and symbols used
Safety warnings are indicated by the following symbols.
DANGER!
This information refers to the immediate danger when working with electricity.
DANGER!
This warning refers to the immediate danger of burns caused by heat or hot surfaces.
DANGER!
This warning refers to the immediate danger when using this device in a hazardous atmosphere.
DANGER!
These warnings must be observed without fail. Even partial disregard of this warning can lead to serious health problems and even death. There is also the risk of seriously damaging the device or parts of the operator's plant.
SAFETY INSTRUCTIONS 1
WARNING!
Disregarding this safety warning, even if only in part, poses the risk of serious health problems. There is also the risk of damaging the device or parts of the operator's plant.
CAUTION!
Disregarding these instructions can result in damage to the device or to parts of the operator's plant.
INFORMATION!
These instructions contain important information for the handling of the device.
LEGAL NOTICE!
This note contains information on statutory directives and standards.
HANDLING
HANDLING
HANDLINGHANDLING
This symbol designates all instructions for actions to be carried out by the operator in the specified sequence.
i RESULT
RESULT
RESULTRESULT This symbol refers to all important consequences of the previous actions.
1.3 Safety instructions for the operator
WARNING!
In general, devices from the manufacturer may only be installed, commissioned, operated and maintained by properly trained and authorized personnel. This document is provided to help you establish operating conditions, which will permit safe and efficient use of this device.
www.krohne.com11/2013 - 4002547502 - MA SMARTSENS ORP 8510 R02 en
7
Page 8
2 DEVICE DESCRIPTION
2.1 Scope of delivery
INFORMATION!
Inspect the cartons carefully for damages or signs of rough handling. Report damage to the carrier and to the local office of the manufacturer.
INFORMATION!
Do a check of the packing list to make sure that you have all the elements given in the order.
INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order.
SMARTSENS ORP 8510
Figure 2-1: Standard scope of delivery
1 Ordered sensor 2 Documentation
Optional accessories
SENSOFIT FLOW 1000 series - Flow-through assemblies
SENSOFIT IMM 1000 / 2000 series - Immersion assemblies
SENSOFIT INS 1000 / 7000 series- Insertion assemblies
SMARTSENS cable VP 2.0 (various lengths)
SMARTMAC 200 W - Wall mount display with calibration and configuration functions
SD 200 W/R - Wall or rack mount indicator
SMARTBRIDGE - USB interface cable
SJB 200 W - Junction box
Consumables / Spare parts available
Various ORP solutions for sensor calibration
Various cleaning solutions
INFORMATION!
For further information contact your local sales office.
8
www.krohne.com 11/2013 - 4002547502 - MA SMARTSENS ORP 8510 R02 en
Page 9
SMARTSENS ORP 8510
2.2 Device description
2.2.1 ORP sensor
DEVICE DESCRIPTION 2
Figure 2-2: Construction of the sensor
1 Nickel plated brass body with VP 2.0 connector or
PEEK body with VP 2.0 connector (in preparation)
2 PG 13.5 thread 3 Washer 4 O-ring 5 Glass shaft 6 Diaphragm 7 Platinum electrode
www.krohne.com11/2013 - 4002547502 - MA SMARTSENS ORP 8510 R02 en
9
Page 10
2 DEVICE DESCRIPTION
2.3 Nameplate
Figure 2-3: Example for a nameplate on the sensor body
1 Manufacturer 2 Device name 3 Order code 4 Serial number 5 Approvals 6 Observe the operation and installation instruction
SMARTSENS ORP 8510
10
Figure 2-4: Example for a nameplate on the glass shaft
1 Manufacturer 2 Device name
INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order.
The sensor type is specified on the label of the sensor package and on the sensor itself.
www.krohne.com 11/2013 - 4002547502 - MA SMARTSENS ORP 8510 R02 en
Page 11
SMARTSENS ORP 8510
3.1 General notes on installation
DANGER!
For devices used in hazardous areas, additional safety notes apply; please refer to the Ex documentation.
DANGER!
All work on the electrical connections may only be carried out with the power disconnected.
DANGER!
Observe the national regulations for electrical installations!
WARNING!
During installation of the device make sure that you use ESD (electronic discharge) protection equippment.
WARNING!
Observe without fail the local occupational health and safety regulations. Any work done on the electrical components of the measuring device may only be carried out by properly trained specialists.
INSTALLATION 3
INFORMATION!
Inspect the cartons carefully for damages or signs of rough handling. Report damage to the carrier and to the local office of the manufacturer.
INFORMATION!
Do a check of the packing list to make sure that you have all the elements given in the order.
INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order.
3.2 Storage and transport
CAUTION!
Do not store the sensor tip dry. This will shorten lifetime considerably. Always store the ORP sensor tip wet in a 3 molar KCl solution when not in use. Saltless water must be avoided since this would leak the KCl ions. The original packing in which the sensor tip was delivered contains a plastic tube with KCl solution and therefore is suitable for storage and transport.
Since the ORP sensor is made out of glass it is very fragile. Avoid shocks of any kind.
Do not touch or scratch the platinum electrode.
Store the sensor in its original packaging in a dry, dust-free location. Keep it away from dirt. If
necessary, clean it as described on page 21.
www.krohne.com11/2013 - 4002547502 - MA SMARTSENS ORP 8510 R02 en
11
Page 12
3 INSTALLATION
3.3 Installation procedure
WARNING!
During installation of the device make sure that you use ESD (electronic discharge) protection equippment.
Because a new ORP sensor needs to be calibrated before it is installed into its final measuring location, it is important to follow the installation order:
1 Unpack the sensor. 2 Connect the sensor to the junction box or directly to the process control system. 3 Calibrate the sensor. 4 Install the sensor into its final measuring location.
The required steps are explained in the following sections.
3.4 Pre-installation requirements
SMARTSENS ORP 8510
CAUTION!
Never touch or scratch the platinum electrode of the sensor.
Store the sensor in its original packaging in a dry, dust-free location. Keep it away from dirt. If necessary, clean it as described on page 21
Figure 3-1: Unpacking the sensor
.
12
Unpacking the sensor
Remove by gently twisting and pulling the protective cap from the sensor 1.
Lay the sensor on a soft mat/tissue 2.
Leave the protection cap of the VP connector on the connector, as long as the sensor is not
connected to the cable.
www.krohne.com 11/2013 - 4002547502 - MA SMARTSENS ORP 8510 R02 en
Page 13
SMARTSENS ORP 8510
3.5 Electrical connection
DANGER!
For devices used in hazardous areas, additional safety notes apply; please refer to the Ex documentation.
DANGER!
All work on the electrical connections may only be carried out with the power disconnected.
DANGER!
Observe the national regulations for electrical installations!
WARNING!
Observe without fail the local occupational health and safety regulations. Any work done on the electrical components of the measuring device may only be carried out by properly trained specialists.
INSTALLATION 3
INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order.
3.5.1 Connecting the cable to the sensor
WARNING!
During installation of the device make sure that you use ESD (electronic discharge) protection equippment.
CAUTION!
Moisture inside the sensor connector must be avoided! Moisture may cause a shortcut and deliver erratic readings! If moisture has entered the connector dry it with air (e.g. hair blower).
Figure 3-2: Connecting the cable to the sensor
Connecting the cable to the sensor
Ensure that both cable and sensor connector are absolutely dry 1.
Screw the cable connector 2 on the sensor and tighten it by hand.
www.krohne.com11/2013 - 4002547502 - MA SMARTSENS ORP 8510 R02 en
13
Page 14
3 INSTALLATION
3.5.2 Connecting the sensor cable
DANGER!
All work on the electrical connections may only be carried out with the power disconnected.
INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order.
SMARTSENS cable VP 2.0
Black (inner coax shield) Ub+
White Ub-
3.5.3 Connecting with junction box SJB 200 W
SMARTSENS ORP 8510
14
Figure 3-3: SJB 200 W with SMARTMAC 200 W and with resistance R (left side). SJB 200 W with SMARTMAC 200 W and without resistance R (right side).
SJB 200 W with SMARTMAC 200 W and with resistance R
B Sensor + A Sensor +
C Sensor - C Sensor -
D Display + (SMARTMAC 200 W) D Display + (SMARTMAC 200 W)
E Display - (SMARTMAC 200 W) E Display - (SMARTMAC 200 W)
F PLC - (process control system) F PLC - (process control system)
H PLC + (process control system) I PLC + (process control system)
SJB 200 W with SMARTMAC 200 W and without resistance R
CAUTION!
´
t include the resistance R into the loop if a HART® isolation amplifier is used.
Don
www.krohne.com 11/2013 - 4002547502 - MA SMARTSENS ORP 8510 R02 en
Page 15
SMARTSENS ORP 8510
Figure 3-4: SJB 200 W and SD 200 W/R (left side). SJB 200 W for HART® communication and with resistance R (right side).
INSTALLATION 3
SJB 200 W and SD 200 W/R
SJB 200 W for HART® communication and with resistance R
A Sensor + B Sensor +
C Sensor - C Sensor -
D Display + (SD 200 W/R) G PLC - (process control system)
E Display - (SD 200 W/R) H PLC + (process control system)
F PLC - (process control system)
I PLC + (process control system)
Figure 3-5: SJB 200 W for HART® communication without resistance R (left side). SJB 200 W (right side).
SJB 200 W for HART® communication and without
SJB 200 W
resistance R
A Sensor + A Sensor +
C Sensor - C Sensor -
G PLC - (process control system) G PLC - (process control system)
I PLC + (process control system) I PLC + (process control system)
www.krohne.com11/2013 - 4002547502 - MA SMARTSENS ORP 8510 R02 en
15
Page 16
3 INSTALLATION
3.6 Installing the sensor
3.6.1 General installation instructions
The platinum electrode must always have full contact with the measuring medium.
The mounting position of the sensor should not deviate more than 75° from vertical position
(sensor tip pointing downwards). Doing otherwise might cause internal air bubbles to float into the platinum electrode. This would interrupt the electrical contact between the inner buffer solution and the glass surface.
SMARTSENS ORP 8510
Figure 3-6: Installation requirements
1 Measuring medium 2 Maximum deviation of 75° from vertical position
16
www.krohne.com 11/2013 - 4002547502 - MA SMARTSENS ORP 8510 R02 en
Page 17
SMARTSENS ORP 8510
4.1 Calibration
OPERATION 4
4.1.1 Calibration with PACTware
Manual hold
Start the function calibration
Activate the function manual hold
Select yes
Press Next
yes.
yesyes
Next to proceed.
NextNext
Start calibration procedure
Set value of ORP solution; Default setting: 468 mV
Press Next
Dip the sensor into the ORP solution and wait till the value is stable.
Press Next
Next to proceed.
NextNext
Next to proceed.
NextNext
i The measurement of ORP solution starts. After approx.10 seconds and a successful
calibration the following values appear: Old ORP value in mV ORP solution in mV Electrode voltage in mV New offset in mV New ORP value in mV
Press Next
Set the calibration date with DD-MM-YYYY.
Press Next
Next to proceed.
NextNext
Next to proceed.
NextNext
i The message Save values?
Select yes
Press Next
Select no
Press Next
yes to save the values.
yesyes
Next to proceed.
NextNext
no to deactivate function manual hold
nono
Next to leave the calibration menu.
NextNext
calibration in menu mode Quick Setup
calibrationcalibration
manual hold to avoid an alarm.
manual holdmanual hold
Save values? appears.
Save values?Save values?
TM
manual hold.
manual holdmanual hold
Quick Setup or Setup
Quick SetupQuick Setup
Setup.
SetupSetup
A ORP calibration is necessary in regular intervals or when installing a new ORP sensor.
If the calibration was not successful, the message Warning! Offset out of range! the value New offset
Press Next
Set the calibration date with DD-MM-YYYY.
Press Next
i The message Warning! Offset out of range!
Press Next
i The message Save values?
Select betweenyes
Press Next
Select no
Press Next
New offset.
New offsetNew offset
Next to proceed.
NextNext
Next to proceed.
NextNext
Warning! Offset out of range! appears.
Warning! Offset out of range!Warning! Offset out of range!
Next to proceed.
NextNext
Save values? appears.
Save values?Save values?
yes or no
no.
yesyes
nono
Next to proceed.
NextNext
no to deactivate the function manual hold
nono
Next to leave the calibration menu.
NextNext
manual hold.
manual holdmanual hold
www.krohne.com11/2013 - 4002547502 - MA SMARTSENS ORP 8510 R02 en
Warning! Offset out of range! appears behind
Warning! Offset out of range!Warning! Offset out of range!
17
Page 18
4 OPERATION
CAUTION!
Make sure that the sensor tip is clean and dust-free. If necessary, clean the tip as described on page 21
CAUTION!
Moisture inside the sensor connector must be avoided! Moisture may cause a shortcut and deliver erratic readings! If moisture has entered the connector dry it with air (e.g. hair blower).
4.1.2 Calibration with HART® Handheld 475 FIELD COMMUNICATOR
Manual hold
Start the function calibration
Activate the function manual hold
Select Yes
Press Enter
Start calibration procedure
Set value of ORP solution; Default setting: 468 mV. Press the key >>>> to enter the submenu. Now
Press Enter
Dip the sensor into the ORP solution and wait till the value is stable.
Press Ok
i The measurement of ORP solution starts. After approx.10 seconds and a successful
.
calibration in menu mode Quick Setup
calibrationcalibration
manual hold to avoid an alarm.
manual holdmanual hold
Yes.
YesYes
Enter to proceed.
EnterEnter
you can change the default setting.
Enter and Ok
EnterEnter
Ok to proceed.
OkOk
calibration the following values appear: Old ORP value in mV ORP solution in mV Electrode voltage in mV New offset in mV New ORP value in mV
Ok to proceed.
OkOk
Quick Setup or Setup
Quick SetupQuick Setup
Setup.
SetupSetup
SMARTSENS ORP 8510
Press Ok
Set the calibration date with DD-MM-YYYY. Press the key >>>> to enter the submenu. Now you can
Press Enter
i The message Save values?
Press yes
Press Enter
Select no
Press Enter
A ORP calibration is necessary in regular intervals or when installing a new ORP sensor.
If the calibration was not successful, the message Warning!!! offset
offset.
offsetoffset
Press Ok
Set the calibration date with DD-MM-YYYY. Press the key >>>> to enter the submenu. Now you can
Press Enter
i The messages Warning!!! Offset out of range!
Select betweenyes
Ok to proceed.
OkOk
change the calibration date .
Enter and Ok
EnterEnter
yes to save the values.
yesyes Enter to proceed.
EnterEnter
no to deactivate the function manual hold.
nono
Enter to leave the calibration menu.
EnterEnter
Ok to proceed.
OkOk
change the calibration date .
Enter and Ok
EnterEnter
Ok to proceed.
OkOk Save values? appear.
Save values?Save values?
Ok to proceed.
OkOk
Warning!!! Offset out of range! and Save values?
Warning!!! Offset out of range!Warning!!! Offset out of range!
yes or no
no.
yesyes
nono
Warning!!! appears behind the value New
Warning!!!Warning!!!
Save values? appear.
Save values?Save values?
New
New New
18
www.krohne.com 11/2013 - 4002547502 - MA SMARTSENS ORP 8510 R02 en
Page 19
SMARTSENS ORP 8510
OPERATION 4
Press Enter
Select no
Press Enter
Enter to proceed.
EnterEnter
no to deactivate the function manual hold
nono
Enter to leave the calibration menu.
EnterEnter
CAUTION!
Make sure that the sensor tip is clean and dust-free. If necessary, clean the tip as described on page 21
.
CAUTION!
Moisture inside the sensor connector must be avoided! Moisture may cause a shortcut and deliver erratic readings! If moisture has entered the connector dry it with air (e.g. hair blower).
4.2 Troubleshooting
Problem Possible cause Remedy
The ORP sensor does not deliver a signal.
The ORP sensor delivers an unstable signal.
Mechanical damage of the glass, e.g. small cracks.
Open circuit Check the cable wiring of the sensor cable on the junction box.
The diaphragm in the reference half cell does not provide good contact to the process medium due to drying up or coatings.
Exchange sensor.
Otherwise connect the sensor to the primary master e.g. PACTware
damage.
Clean the diaphragm with hot soap or acid using a soft tissue
(details on page 21).
Submerge sensor in water and increase the temperature to
+50...+60°C / +122...+140°F.
Submerge sensor in 3 molar KCl solution at ambient temperature.
The decrease in temperature will cause the reference half cell to suck in KCl solution through the diaphragm and regenerate the diaphragms functionality.
manual hold.
manual holdmanual hold
TM
FTD/DTM or to the HART® handheld to exclude a sensor
www.krohne.com11/2013 - 4002547502 - MA SMARTSENS ORP 8510 R02 en
19
Page 20
4 OPERATION
SMARTSENS ORP 8510
4.3 Status messages and diagnostic information
Measurements out of specification
Message
Message Description
MessageMessage
S01 ORP value > ORP max
S02 ORP value < ORP min
S03 Temp. value > Temp max
S04 Temp. value < Temp min
Maintenance
Message
Message Description
MessageMessage
M01 Offset < -99 mV Recalibrate the sensor. For more information refer
M02 Offset> +99 mV
M03 Maintenance interval expired
Description Action
DescriptionDescription
ORP value out of range
ORP value out of range
Temperature value out of range
Temperature value out of range
Description Action
DescriptionDescription
Action
ActionAction
Consider the measuring range limits, otherwise select a suitable sensor for the process conditions of the application.
Action
ActionAction
to
Calibration with PACTwareTM
on page 17
20
www.krohne.com 11/2013 - 4002547502 - MA SMARTSENS ORP 8510 R02 en
Page 21
SMARTSENS ORP 8510
5.1 Maintenance
5.1.1 Cleaning
INFORMATION!
Recalibrate the sensor after each cleaning procedure
Slight dirt residues or dust: Rinse the sensor tip with demineralised water and clean it with a soft tissue.
Oily and greasy coatings: Remove with a warm soap solution and rinse with demineralised water.
Hardness deposits or metal hydroxide deposits: Soak the sensor tip including diaphragm in 10% citric acid or hypochloric acid for a couple of minutes and rinse the complete glass shaft of the sensor with demineralised water.
Biological fouling: Soak the sensor tip including diaphragm in 10 % pepsin solution for minimum 3 hours and afterwards rinse the complete glass shaft of sensor with demineralised water.
5.1.2 Aging and re-calibration
SERVICE 5
During operation, but already during storage, ORP sensors age due to poisoning effects of the inner buffer system. Therefore it is important to re-calibrate the sensor in regular intervals. Consider status messages and diagnostic information.
INFORMATION!
The lifetime expectation depends heavily on the application. The right choice of the sensor type is very important.
5.2 Availability of services
The manufacturer offers a range of services to support the customer after expiration of the warranty. These include repair, maintenance, technical support and training.
INFORMATION!
For more precise information, please contact your local sales office.
www.krohne.com11/2013 - 4002547502 - MA SMARTSENS ORP 8510 R02 en
21
Page 22
5 SERVICE
5.3 Returning the device to the manufacturer
5.3.1 General information
This device has been carefully manufactured and tested. If installed and operated in accordance with these operating instructions, it will rarely present any problems.
CAUTION!
Should you nevertheless need to return a device for inspection or repair, please pay strict attention to the following points:
Due to statutory regulations on environmental protection and safeguarding the health and safety of our personnel, manufacturer may only handle, test and repair returned devices that have been in contact with products without risk to personnel and environment.
This means that the manufacturer can only service this device if it is accompanied by the following certificate (see next section) confirming that the device is safe to handle.
CAUTION!
If the device has been operated with toxic, caustic, flammable or water-endangering products, you are kindly requested:
to check and ensure, if necessary by rinsing or neutralising, that all cavities are free from such dangerous substances,
to enclose a certificate with the device confirming that is safe to handle and stating the product used.
SMARTSENS ORP 8510
22
www.krohne.com 11/2013 - 4002547502 - MA SMARTSENS ORP 8510 R02 en
Page 23
SMARTSENS ORP 8510
5.3.2 Form (for copying) to accompany a returned device
Company: Address:
Department: Name:
Tel. no.: Fax no.:
Manufacturer's order no. or serial no.:
The device has been operated with the following medium:
SERVICE 5
This medium is: water-hazardous
toxic
caustic
flammable
We checked that all cavities in the device are free from such substances.
We have flushed out and neutralized all cavities in the device.
We hereby confirm that there is no risk to persons or the environment through any residual media contained in the device when it is returned.
Date: Signature:
Stamp:
5.4 Disposal
CAUTION!
Disposal must be carried out in accordance with legislation applicable in your country.
www.krohne.com11/2013 - 4002547502 - MA SMARTSENS ORP 8510 R02 en
23
Page 24
6 TECHNICAL DATA
6.1 Measuring principle
6.1.1 ORP measurement
The oxidation reduction potential, ORP in short, is the measurement for the concentration of oxidising and reducing agents in water. Its value is influenced both by pH and temperature. ORP is a sum parameter that gives no information on the concentration of a single substance in a mixture.
ORP measurements are used to monitor chemical reactions involving electron transfer. In drinking water treatment it can be found in ozone treatment and the removal of iron, manganese and nitrate as well as in disinfection steps. In swimming pools the German DIN 19643 requires ORP measurements as a hygiene parameter and decrees maximum and minimum values for fresh water, pool water, and salt water. In wastewater treatment ORP is measured in the denitrification process and in detoxication of industrial wastewater.
The ORP sensor consists of a measuring electrode of platinum or gold and a reference of e.g. Ag/AgCl. The potential of the measuring electrode changes with the concentration of reducing and oxidising agents and is measured against the reference. The measured values can be recalculated to fit literature values based on NHE (normal hydrogen electrode) as reference.
SMARTSENS ORP 8510
24
www.krohne.com 11/2013 - 4002547502 - MA SMARTSENS ORP 8510 R02 en
Page 25
SMARTSENS ORP 8510
6.2 Technical data
INFORMATION!
The following data is provided for general applications. If you require data that is more relevant to your specific application, please contact us or your local sales office.
Additional information (certificates, special tools, software,...) and complete product documentation can be downloaded free of charge from the website (Download Center).
Measuring system
Measuring principle Potentiometric
Measuring range -1500...+1500 mV
Design
Shaft diameter 12 mm / 0.47"
Insertion length 120 mm / 4.72"
Temperature sensor Pt1000
Connector VarioPin 2.0 (VP 2.0)
TECHNICAL DATA 6
Operating conditions
Temperature range 0...+70°C / +32...+158°F
Process pressure 2 bar / 29 psi
Measuring accuracy 0.5% Conductivity Min. 150 µS/cm
Installation conditions
Ingress protection IP 68
Weight Approx. 80 g / 0.18 lb
Process connection PG 13.5
Materials
Sensor shaft Glass
Measuring electrode Platinum
Inner buffer pH 7.0
Reference Ag/AgCl/TepoxGel
Diaphragm Ceramic
Gasket EPDM (FDA)
Sensor head Nickel plated brass body with VP 2.0 connector
PEEK body with VP 2.0 connector (in preparation)
www.krohne.com11/2013 - 4002547502 - MA SMARTSENS ORP 8510 R02 en
25
Page 26
6 TECHNICAL DATA
Communication
ORP range -1500...+1500 mV
Output signal 4...20 mA (passive) Output resolution 20 µA
Field communication
Filter adjustable 1...60 seconds
Electrical connections
Power supply 15...30 VDC loop powered
Measuring range
Load Minimum 0 Ω; maximum RL = ((U
Error signal Acc. to NAMUR NE 43
®
HART
Device revision 1
Physical layer FSK
Device category Sensor, galvanically isolated
System requirements
Multidrop operation 4 mA
SMARTSENS ORP 8510
HART® 7 - FSK 1200 physical layer definition on top of the current loop
4...20 mA + HART® protocol
- 15 VDC) / 22 mA)
ext.
Upper value: 21.0 mA
HART® protocol via current output
250 Ω loop resistance for HART® communication
Approvals
CE This device fulfils the statutory requirements of the EC directives. The
manufacturer certifies successful testing of the product by applying the CE mark.
Shock resistance: IEC 60068-2-31, Environmental testing – Part 2: Test Ec
Electromagnetic compatibility: Acc. to EN 61326, NAMUR NE 21
Low voltage directive: Safety requirements for electrical equipment for measurement, control, and laboratory use in accordance with EN 61010-1:2001
INFORMATION!
For further information contact your local sales office.
26
www.krohne.com 11/2013 - 4002547502 - MA SMARTSENS ORP 8510 R02 en
Page 27
SMARTSENS ORP 8510
6.3 Dimensions
Figure 6-1: SMARTSENS PH 8510 with VP 2.0
a 104 4.09
b 120 4.72
c 12 0.47
d Ø12 Ø0.47
TECHNICAL DATA 6
Dimensions [mm] Dimensions [inch]
www.krohne.com11/2013 - 4002547502 - MA SMARTSENS ORP 8510 R02 en
27
Page 28
7 DESCRIPTION OF HART INTERFACE
7.1 General description
The open HART® protocol, which can be used for free, is integrated into the sensor for communication.
®
Devices which support the HART devices. When it comes to operating devices (Master), both manual control units (Secondary Master) and PC-supported workstations (Primary Master) are used in, for example, a control centre.
®
HART
field devices include measuring sensors, signal converters and actuators. The field
devices range from 2-wire to intrinsically safe versions for use in hazardous areas.
®
The HART
data are superimposed over the analogue 4...20 mA signal via FSK modem. This way,
all of the connected devices can communicate digitally with one another via the HART while simultaneously transmitting the analogue signals.
When it comes to the field devices and secondary masters, the FSK or HART integrated. If a PC is used, an external modem must be connected to the serial interface (USB interface). There are, however, other connection variants which can be seen in the following connection figures.
protocol are classified as either operating devices or field
SMARTSENS ORP 8510
®
protocol
®
modem is
7.2 Software history
INFORMATION!
In the table below, "x" is a placeholder for possible multi-digit alphanumeric combinations, depending on the available version.
Release date SW version HW version
2013-04-01 1.0.x 1.0.x 1 1
HART® identification codes and revision numbers
Manufacturer ID: 69 (0x45)
Device: 195 (0xC3)
Device Revision: 1
DD Revision: 1
HART® Universal Revision:
FC 475 system SW.Rev.: ≥ 3.7
PDM version: ≥ 8.0
FDT version: ≥ 1.2
®
HART
Device Revision DD Revision
7
28
www.krohne.com 11/2013 - 4002547502 - MA SMARTSENS ORP 8510 R02 en
Page 29
SMARTSENS ORP 8510
7.3 Connection variants
The sensor is a 2-wire device with a passive 4...20 mA current output and HART® interface.
Multidrop mode is supported
Multidrop mode is supported
Multidrop mode is supportedMultidrop mode is supported In a multidrop communication system, more than 2 devices are connected to a common transmission cable.
Burst mode is not supported
Burst mode is not supported
Burst mode is not supportedBurst mode is not supported In the burst mode a slave device transfers cyclic pre-defined response telegrams, to get a higher rate of data transfer.
INFORMATION!
For detailed information about the electrical connection of the sensor for HART Connecting with junction box SJB 200 W on page 14
There are two ways of using the HART
as Point-to-Point connection and
as Multidrop connection, with 2-wire connection.
DESCRIPTION OF HART INTERFACE 7
®
, refer to
.
®
communication:
www.krohne.com11/2013 - 4002547502 - MA SMARTSENS ORP 8510 R02 en
29
Page 30
7 DESCRIPTION OF HART INTERFACE
7.3.1 Point-to-Point connection - analogue / digital mode
Point-to-Point connection between the sensor and the HART® Master.
The current output of the device is passive.
SMARTSENS ORP 8510
Figure 7-1: Point- to-Point connection
®
DD
TM
FDT/DTM
1 Primary master with e.g. PACTware 2 FSK modem
3 HART 4 SD 200 W/R or SMARTMAC 200 W (optional) 5 Sensor
6 Secondary master with HART 7 Power supply for devices (slaves) with passive current output 8 Load 250 Ω (Ohm)
®
signal
30
www.krohne.com 11/2013 - 4002547502 - MA SMARTSENS ORP 8510 R02 en
Page 31
SMARTSENS ORP 8510
DESCRIPTION OF HART INTERFACE 7
7.4 Inputs/outputs and HART® dynamic variables and device variables
PV = Primary Variable; SV = Secondary Variable; TV = Tertiary Variable
HART® dynamic variable
PV SV TV
ORP Electrode voltage Temperature
Code = device variable code
Device variables
HART® device variable
ORP value 0 linear
Electrode voltage 1 linear
Temperature 2 linear
The HART® dynamic variable PV is always connected to the HART® current output which is assigned to the ORP value.
Code Type
7.5 Field Communicator 475 (FC 475)
The Field Communicator is a hand terminal from Emerson Process Management that is
®
designed to configure HART used to integrate different devices into the Field Communicator.
7.5.1 Installation
The HART® Device Description for the sensor must be installed on the Field Communicator. Otherwise only the functions of a generic DD are available to the user and the entire device control is not possible. A "Field Communicator Easy Upgrade Programming Utility" is required to install the DDs on the Field Communicator.
The Field Communicator must be equipped with a system card with "Easy Upgrade Option". For
details consult the Field Communicator Users Manual.
and Foundation Fieldbus devices. Device Descriptions (DDs) are
www.krohne.com11/2013 - 4002547502 - MA SMARTSENS ORP 8510 R02 en
31
Page 32
7 DESCRIPTION OF HART INTERFACE
7.5.2 Operation
INFORMATION!
For more detailed information (see on page 33
Operating the sensor via the Field Communicator is very similar to manual device control using the keyboard.
Limitation: The service menu parameters for the senor are supported but a simulation for the current output is not possible.
Parameter protection for custody transfer is the same as on the SMARTMAC 200 W display. Other specific protective functions such as the passwords for the quick setup menu and the
setup menu are not supported with HART
), Menu tree for Basic DD.
®
.
7.6 Field Device Tool / Device Type Manager (FDT/DTM)
A Field Device Tool Container (FDT Container) is basically a PC program used to configure a field
®
device via HART Manager (DTM).
. To adapt to different devices, the FDT container uses a so-called Device Type
SMARTSENS ORP 8510
7.6.1 Installation
If the DTM for the sensor has not yet been installed on the FDT Container, setup is required and is available for download from the website or on CD-ROM. See the supplied documentation for information on how to install and set up the DTM.
32
www.krohne.com 11/2013 - 4002547502 - MA SMARTSENS ORP 8510 R02 en
Page 33
SMARTSENS ORP 8510
DESCRIPTION OF HART INTERFACE 7
7.7 Overview Basic-DD menu tree (positions in menu tree)
Main menu Submenu
1 Measuring value
1 Measuring value 1 ORP value
1 Measuring value1 Measuring value
2 Temperature
3 Electrode voltage
4 Loop current
5 Error
6 Device state
2 Quick setup
2 Quick setup 1 TAG
2 Quick setup2 Quick setup
2 Manual hold
3 I/O
4 Calibration
5 Input user password
3 Logbook
3 Logbook 1 Calibration logbook
3 Logbook3 Logbook
2 Error logbook
4 Setup
4 Setup 1 Process input
4 Setup4 Setup
2 I/O
3 I/O HART
4 Device
5 Service
5 Service 1 Service calibration
5 Service5 Service
2 Service parameter
3 Passwords
4 Sensor lock
®
www.krohne.com11/2013 - 4002547502 - MA SMARTSENS ORP 8510 R02 en
33
Page 34
7 DESCRIPTION OF HART INTERFACE
7.8 Basic-DD menu tree (details for settings)
1 Measuring values
1 ORP value
2 Temperature Display of the measured temperature value in °C / °F on process control
3 Electrode voltage Display of the measured electrode voltage in mV on process control system,
4 Loop current Display of the measured loop current in mA on process control system,
5 Error Display of status messages and diagnostic information. For further
6 Device state Display of status information icon for sensor according to NAMUR NE 107.
2 Quick setup
1 TAG Set the TAG number for the measuring loop.
2 Manual hold Activate or deactivate the manual hold function. Select between yes or no.
3 I/O Meas. value at 4 mA Set measuring value at 4 mA;
4 Calibration Start calibration procedure. For more information refer to
5 Input user password Set password
Display of the measured ORP value on process control system, HART® handheld or display.
®
system, HART
®
handheld or display.
HART
®
handheld or display.
HART
information refer to
Meas. value at 20 mA Set measuring value at 20 mA;
Time constant Set time constant (160 seconds);
page 17.
handheld or display.
Status messages and diagnostic information
Default setting: -1500 mV
Default setting: +1500 mV
Default setting: 1 second
SMARTSENS ORP 8510
on page 20.
Calibration
on
3 Logbooks
1 Calibration logbook Logbook for the last 20 calibration cycles
2 Error logbook Display errors
34
www.krohne.com 11/2013 - 4002547502 - MA SMARTSENS ORP 8510 R02 en
Page 35
SMARTSENS ORP 8510
4 Setup
1 Process input Temperature Set temperature unit °C / °F
2 I/O Meas. value at 4 mA Set measuring value at 4 mA; Default setting:
3 I/O HART TAG Set TAG for measuring loop
DESCRIPTION OF HART INTERFACE 7
Set temperature offset
Display date of offset calibration
Calibration Start calibration procedure. For more information
Maintenance interval Set maintenance interval in days (000...999 days);
Reset maintenance interval
Meas. value at 20 mA Set measuring value at 20 mA; Default setting:
Time constant Set time constant (160 seconds); Default
Message Display messages (32 packed ASCII)
Polling address Display polling address
Loop current mode Select between enable or disable to activate or
Device variables PV Display PV - ORP with
refer to
Default setting: 000
Yes / No; Default setting: No
-1500 mV
+1500 mV
setting: 1 second
Set long TAG
Display of previous long TAG
deactivate the loop current mode
SV Display SV - electrode
TV Display TV -
Calibration
on page 17.
min. and max. limits
voltage in mV
temperature in °C / °F with min. and max. limits
www.krohne.com11/2013 - 4002547502 - MA SMARTSENS ORP 8510 R02 en
35
Page 36
7 DESCRIPTION OF HART INTERFACE
4 Device Information Sensor information Order code
Commissioning Set date of commissioning
SMARTSENS ORP 8510
Device name
Serial number
HART® ID
Polling address
Manufacturer ID
Date of manufacturing
SW version
HW version
Calibration Offset
ORP solution
No. of calibrations
Operating parameters (OP)
Commissioning date
Operating hours OP time > 80°C / 176°F OP time > 110°C / 230°F
Max. temperature
5 Service
1 Service calibration Trimming at 4 mA (+/-)
Trimming at 20 mA (+/-)
2 Service parameter Reset of sensor
Load factory setting
3 Password Password protection Activate or deactivate the password protection.
Password operator Set password for operator
Password administrator
Reset password Reset all passwords (only administrator)
4 Sensor lock Select between YES
If you select YES invalid for use. The HART
YES or NO
YESYES
YES the setting is not reversible anymore and the sensor is
YESYES
Select between on and off.
Set password for administrator
NO to lock the sensor.
NONO
®
communication is not possible anymore.
36
www.krohne.com 11/2013 - 4002547502 - MA SMARTSENS ORP 8510 R02 en
Page 37
SMARTSENS ORP 8510
NOTES 8
www.krohne.com11/2013 - 4002547502 - MA SMARTSENS ORP 8510 R02 en
37
Page 38
8 NOTES
SMARTSENS ORP 8510
38
www.krohne.com 11/2013 - 4002547502 - MA SMARTSENS ORP 8510 R02 en
Page 39
SMARTSENS ORP 8510
NOTES 8
www.krohne.com11/2013 - 4002547502 - MA SMARTSENS ORP 8510 R02 en
39
Page 40
KROHNE product overview
Electromagnetic flowmeters
Variable area flowmeters
Ultrasonic flowmeters
Mass flowmeters
Vortex flowmeters
Flow controllers
Level meters
Temperature meters
Pressure meters
Analysis products
Products and systems for the oil & gas industry
Measuring systems for the marine industry
Head Office KROHNE Messtechnik GmbH Ludwig-Krohne-Straße 5 47058 Duisburg (Germany) Tel.:+49 203 301 0 Fax:+49 203 301 103 89 info@krohne.com
© KROHNE 11/2013 - 4002547502 - MA SMARTSENS ORP 8510 R02 en - Subject to change without notice.
The current list of all KROHNE contacts and addresses can be found at: www.krohne.com
Loading...