OPTIFLEX 2200 C/F
1.1 Software history
"Firmware revision" agrees with NAMUR NE 53. It is a series of numbers used to record the
revision status of embedded software (firmware) in electronic equipment assemblies. It gives
data on the type of changes made and the effect that changes have on compatibility.
SAFETY INSTRUCTIONS 1
Data about software revisions is shown in menu 1.1.0 IDENT. For more data, refer to
description
of the device (given on the device nameplate) and speak to the supplier.
on page 74. If it is not possible to refer to the device menu, record the serial number
Firmware
assembly
Sensor1.21.024000357001o
HMI (LCD display option) 1.00.024000487601m
revision
Harware
revision
Changes and
compatibility
Documentation
HB OPTIFLEX 2200 R02
Function
1.2 Intended use
CAUTION!
Responsibility for the use of the measuring devices with regard to suitability, intended use and
corrosion resistance of the used materials against the measured fluid lies solely with the
operator.
INFORMATION!
The manufacturer is not liable for any damage resulting from improper use or use for other than
the intended purpose.
This TDR level transmitter measures distance, level, mass and volume of liquids, pastes,
slurries, granulates and powders.
It can be installed on tanks, silos and open pits.
www.krohne.com07/2013 - 4000668002 - HB OPTIFLEX 2200 R02 en
7
1 SAFETY INSTRUCTIONS
1.3 Certification
DANGER!
For devices used in hazardous areas, additional safety notes apply; please refer to the Ex
documentation.
In accordance with the commitment to customer service and safety, the device
described in this document meets the following safety requirements:
• Electromagnetic Compatibility (EMC) Directive 2004/108/EC in conjunction with EN 61326-1
(2006), EN 61326-2-3 (2006). SIL-approved devices agree with EN 61326-3-1 (2008) and
EN 61326-3-2 (2008).
• SIL-approved devices agree with EN 61508 (2010).
All devices are based on the CE marking and meet the requirements of NAMUR Guideline NE 21,
NE 43, NE 53 and NE 107.
OPTIFLEX 2200 C/F
1.4 Electromagnetic compatibility
The device design agrees with European Standard EN 61326-1 when installed in metallic tanks.
You can install the device on open-air tanks and tanks that are not made of metal. Refer also to
the note that follows.
CAUTION!
If you install a device with a cable probe in a non-metallic tank or open-air pit, a strong
electromagnetic field near to the device can have an unwanted effect on the accuracy. Use a
device with a coaxial probe for this type of installation.
INFORMATION!
Device operation agrees with residential-class (class B) emissions and industrial-class for
immunity requirements if:
•
the device has a single or double probe (rod or cable probe) and is used in a closed tank made
of metal or
•
the device has a coaxial probe.
8
www.krohne.com07/2013 - 4000668002 - HB OPTIFLEX 2200 R02 en
OPTIFLEX 2200 C/F
1.5 Safety instructions from the manufacturer
1.5.1 Copyright and data protection
The contents of this document have been created with great care. Nevertheless, we provide no
guarantee that the contents are correct, complete or up-to-date.
The contents and works in this document are subject to copyright. Contributions from third
parties are identified as such. Reproduction, processing, dissemination and any type of use
beyond what is permitted under copyright requires written authorisation from the respective
author and/or the manufacturer.
The manufacturer tries always to observe the copyrights of others, and to draw on works created
in-house or works in the public domain.
The collection of personal data (such as names, street addresses or e-mail addresses) in the
manufacturer's documents is always on a voluntary basis whenever possible. Whenever
feasible, it is always possible to make use of the offerings and services without providing any
personal data.
SAFETY INSTRUCTIONS 1
We draw your attention to the fact that data transmission over the Internet (e.g. when
communicating by e-mail) may involve gaps in security. It is not possible to protect such data
completely against access by third parties.
We hereby expressly prohibit the use of the contact data published as part of our duty to publish
an imprint for the purpose of sending us any advertising or informational materials that we have
not expressly requested.
1.5.2 Disclaimer
The manufacturer will not be liable for any damage of any kind by using its product, including,
but not limited to direct, indirect or incidental and consequential damages.
This disclaimer does not apply in case the manufacturer has acted on purpose or with gross
negligence. In the event any applicable law does not allow such limitations on implied warranties
or the exclusion of limitation of certain damages, you may, if such law applies to you, not be
subject to some or all of the above disclaimer, exclusions or limitations.
Any product purchased from the manufacturer is warranted in accordance with the relevant
product documentation and our Terms and Conditions of Sale.
The manufacturer reserves the right to alter the content of its documents, including this
disclaimer in any way, at any time, for any reason, without prior notification, and will not be liable
in any way for possible consequences of such changes.
www.krohne.com07/2013 - 4000668002 - HB OPTIFLEX 2200 R02 en
9
1 SAFETY INSTRUCTIONS
1.5.3 Product liability and warranty
The operator shall bear responsibility for the suitability of the device for the specific purpose.
The manufacturer accepts no liability for the consequences of misuse by the operator. Improper
installation and operation of the devices (systems) will cause the warranty to be void. The
respective "Standard Terms and Conditions" which form the basis for the sales contract shall
also apply.
1.5.4 Information concerning the documentation
To prevent any injury to the user or damage to the device it is essential that you read the
information in this document and observe applicable national standards, safety requirements
and accident prevention regulations.
If this document is not in your native language and if you have any problems understanding the
text, we advise you to contact your local office for assistance. The manufacturer can not accept
responsibility for any damage or injury caused by misunderstanding of the information in this
document.
This document is provided to help you establish operating conditions, which will permit safe and
efficient use of this device. Special considerations and precautions are also described in the
document, which appear in the form of underneath icons.
OPTIFLEX 2200 C/F
10
www.krohne.com07/2013 - 4000668002 - HB OPTIFLEX 2200 R02 en
OPTIFLEX 2200 C/F
1.5.5 Warnings and symbols used
Safety warnings are indicated by the following symbols.
DANGER!
This information refers to the immediate danger when working with electricity.
DANGER!
This warning refers to the immediate danger of burns caused by heat or hot surfaces.
DANGER!
This warning refers to the immediate danger when using this device in a hazardous atmosphere.
DANGER!
These warnings must be observed without fail. Even partial disregard of this warning can lead to
serious health problems and even death. There is also the risk of seriously damaging the device
or parts of the operator's plant.
SAFETY INSTRUCTIONS 1
WARNING!
Disregarding this safety warning, even if only in part, poses the risk of serious health problems.
There is also the risk of damaging the device or parts of the operator's plant.
CAUTION!
Disregarding these instructions can result in damage to the device or to parts of the operator's
plant.
INFORMATION!
These instructions contain important information for the handling of the device.
LEGAL NOTICE!
This note contains information on statutory directives and standards.
• HANDLING
HANDLING
HANDLINGHANDLING
This symbol designates all instructions for actions to be carried out by the operator in the
specified sequence.
iRESULT
RESULT
RESULTRESULT
This symbol refers to all important consequences of the previous actions.
1.6 Safety instructions for the operator
WARNING!
In general, devices from the manufacturer may only be installed, commissioned, operated and
maintained by properly trained and authorized personnel.
This document is provided to help you establish operating conditions, which will permit safe and
efficient use of this device.
www.krohne.com07/2013 - 4000668002 - HB OPTIFLEX 2200 R02 en
11
2 DEVICE DESCRIPTION
2.1 Scope of delivery
INFORMATION!
Do a check of the packing list to make sure that you have all the elements given in the order.
OPTIFLEX 2200 C/F
Figure 2-1: Scope of delivery
1 Signal converter and probe.
2 Probe segments. For the assembly procedure of the single rod probe, refer to
(single-piece probe)
option, only part of the probe is attached to the device. For the assembly procedure of the segmented single rod probe,
How to assemble the single rod probe (segmented probe)
refer to
mented coaxial probe, refer to
3 Quick Start
4 DVD-ROM. This contains the Handbook, the Quick Start, the Technical Datasheet and related software.
5 Strap wrench
6 Probe assembly cover
on page 29. If a single rod or coaxial probe is attached and you ordered the "segmented probe"
on page 33. For the assembly procedure of the seg-
How to assemble the segmented coaxial probe
How to assemble the single rod probe
on page 36.
INFORMATION!
No training required!
12
www.krohne.com07/2013 - 4000668002 - HB OPTIFLEX 2200 R02 en
OPTIFLEX 2200 C/F
2.2 Device description
The TDR level transmitter is designed to measure the distance, level, mass and volume of
liquids, pastes, slurries, granulates and powders.
TDR level transmitters use a probe to guide a signal to the surface of the measured product. The
device has a large choice of probes. Thus, it can measure most products in difficult conditions.
For more data, refer to
You can order these accessories:
• RS232 / HART® converter (VIATOR).
®
• USB / HART
converter (VIATOR).
INFORMATION!
For more data on accessories, refer to Accessories on page 160
2.3 Visual Check
Technical data
DEVICE DESCRIPTION 2
on page 115.
.
INFORMATION!
Inspect the cartons carefully for damages or signs of rough handling. Report damage to the
carrier and to the local office of the manufacturer.
Figure 2-2: Visual check
1 Device nameplate (for more data, refer to
2 Process connection data (size and pressure rating, material reference and heat number)
3 Gasket material data - refer to the illustration that follows
Non-Ex nameplate
on page 14)
www.krohne.com07/2013 - 4000668002 - HB OPTIFLEX 2200 R02 en
13
2 DEVICE DESCRIPTION
Figure 2-3: Symbols for the supplied gasket material (on the side of the process connection)
1 EPDM
2 Kalrez
®
6375
If the device is supplied with an FKM/FPM gasket, there is no symbol on the side of the process
connection.
2.4 Nameplates
INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order.
Check for the correct supply voltage printed on the nameplate.
OPTIFLEX 2200 C/F
2.4.1 Non-Ex nameplate
Figure 2-4: Compact (C) and remote (F) versions: Non-Ex nameplate attached to the housing
1 Cable entry size
2 Communication protocol (analog, digital, fieldbus, ...), input voltage and maximum current (fieldbus options: basic cur-
rent)
3 Degree of ingress protection (according to EN 60529 / IEC 60529)
4 Customer tag number
5 Date of manufacture
6 Order number
7 Type code (defined in order)
8 Model name and number. The last letter "X" is either:
C = compact version or
F = remote (field) version
9 Company name and address
14
www.krohne.com07/2013 - 4000668002 - HB OPTIFLEX 2200 R02 en
OPTIFLEX 2200 C/F
Figure 2-5: Remote (F) version: Non-Ex nameplate attached to the probe assembly
1 Cable entry size
2 Degree of ingress protection (according to EN 60529 / IEC 60529)
3 Date of manufacture
4 Order number
5 Type code (defined in order)
6 Model name and number. X = "F" remote (field) version
7 Company name and address
DEVICE DESCRIPTION 2
www.krohne.com07/2013 - 4000668002 - HB OPTIFLEX 2200 R02 en
15
3 INSTALLATION
3.1 General notes on installation
INFORMATION!
Inspect the cartons carefully for damages or signs of rough handling. Report damage to the
carrier and to the local office of the manufacturer.
INFORMATION!
Do a check of the packing list to make sure that you have all the elements given in the order.
INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order.
Check for the correct supply voltage printed on the nameplate.
3.2 Storage
WARNING!
Do not keep the device in a vertical position. This will damage the probe and the device will not
measure correctly.
OPTIFLEX 2200 C/F
16
Figure 3-1: Storage conditions
1 Do not bend rod and coaxial probes - support here
2 Storage temperature range: -50…+85°C/ -60…+185°F (min. -40°C/ -40°F for devices with the integrated LCD display
option)
• Store the device in a dry and dust-free location.
• Store the device in its original packing.
www.krohne.com07/2013 - 4000668002 - HB OPTIFLEX 2200 R02 en
OPTIFLEX 2200 C/F
3.3 Transport
Figure 3-2: How to hold the device
INSTALLATION 3
Figure 3-3: How to hold the device
1 Wind cable probes greater than 400 mm / 16¨ in diameter.
2 Do not hold the probe when you lift the device.
3 Do not wind the electrical cable less than 330 mm / 13¨ in diameter.
WARNING!
If you do not lift the device carefully, you can cause damage to the probe.
3.4 Pre-installation requirements
INFORMATION!
Obey the precautions that follow to make sure that the device is correctly installed.
• Make sure that there is sufficent space on all sides.
• Protect the signal converter from direct sunlight. The device has a weather protection option.
• Do not subject the signal converter to heavy vibrations.
www.krohne.com07/2013 - 4000668002 - HB OPTIFLEX 2200 R02 en
17
3 INSTALLATION
3.5 How to prepare the tank before you install the device
CAUTION!
To avoid measuring errors and device malfunction, obey these precautions.
3.5.1 Pressure and temperature ranges
OPTIFLEX 2200 C/F
Figure 3-4: Pressure and temperature ranges
1 Temperature at the process connection
The temperature at the process connection must stay in the temperature range of the gasket material unless the device is a High-Temperature version. Refer to the table "Permitted temperature ranges for gaskets" that follows and
to "Technical data" on page 116.
Devices with Hazardous Location approvals: see supplementary instructions
2 Ambient temperature for operation of the display
-20...+60°C / -4...+140°F
If the ambient temperature is not between these limits, the display screen switches off automatically
3 Ambient temperature
Non-Ex devices: -40...80°C / -40...176°F
Devices with Hazardous Location approvals: see supplementary instructions
4 Process pressure
-1...40 barg / -14.5...580 psig
WARNING!
The process connection temperature range must agree with the temperature limits of the
gasket material.
Permitted temperature ranges for gaskets
Gasket materialPermitted temperature ranges for gaskets
Standard versionHigh-Temperature version
[°C][°F][°C][°F]
FKM/FPM-40…+150-40…+302-40…+300-40…+572
Kalrez® 6375
EPDM-50…+150-58…+302-50…+250-58…+482
-20…+150-4…+302-20…+300-4…+572
18
www.krohne.com07/2013 - 4000668002 - HB OPTIFLEX 2200 R02 en
OPTIFLEX 2200 C/F
3.5.2 General information for nozzles
CAUTION!
Follow these recommendations to make sure that the device measures correctly. They have an
effect on the performance of the device.
CAUTION!
Do not put the process connection near to the product inlet. If the product that enters the tank
touches the probe, the device will measure incorrectly.
INSTALLATION 3
Figure 3-5: Do not put the device near to a product inlet
1 The device is in the correct position.
2 The device is too near to the product inlet.
3 If it is not possible to put the device in the recommended position, install a deflector pipe.
Figure 3-6: How to prevent build up of product around the process connection
1 If product particles are likely to collect in holes, a nozzle is not recommended.
2 Attach the flange directly to the tank.
3 Use a threaded connection to attach the device directly to the tank.
www.krohne.com07/2013 - 4000668002 - HB OPTIFLEX 2200 R02 en
19
3 INSTALLATION
For single cable and single rod probes:
Figure 3-7: Recommended nozzle dimensions for single rod and single cable probes
1 Recommended conditions: h ≤ d, where h is the height of the tank nozzle and d is the diameter of the tank nozzle.
2 The end of the nozzle must not have an extension into the tank. Do not install the device on a high nozzle.
CAUTION!
If the device is installed on a high nozzle, make sure that the probe does not touch the side of the
nozzle (attach the probe end, ...).
OPTIFLEX 2200 C/F
Figure 3-8: Sockets for threaded process connections
1 Recommended installation
2 The end of the socket must not have an extension into the tank
20
www.krohne.com07/2013 - 4000668002 - HB OPTIFLEX 2200 R02 en
OPTIFLEX 2200 C/F
For double cable and double rod probes:
Figure 3-9: Recommended nozzle dimensions for double rod and double cable probes
d ≥ 50 mm / 2¨, where d is the diameter of the tank nozzle
For coaxial probes:
If your device has a coaxial probe, you can ignore these installation recommendations.
INSTALLATION 3
CAUTION!
Install coaxial probes in clean liquids that are not too viscous.
3.5.3 Installation requirements for concrete roofs
Figure 3-10: Installation on a concrete roof
1 The diameter, d, of the hole must be greater than the thickness, t, of the concrete.
2 If the thickness, t, of the concrete is greater than the diameter, d, of the hole, install the device in a recess.
www.krohne.com07/2013 - 4000668002 - HB OPTIFLEX 2200 R02 en
21
3 INSTALLATION
3.6 Installation recommendations for liquids
3.6.1 General requirements
OPTIFLEX 2200 C/F
Figure 3-11: Installation recommendations for liquids
1 The electromagnetic (EM) field generated by the device. It has a radius of R
objects and product flow. Refer to the table that follows.
2 If there are too many objects in the tank, install a bypass chamber or stilling well.
3 Keep the probe straight. If the probe is too long, shorten the probe length. Make sure that the device is configured with
the new probe length. For more data on the procedure, refer to
4 Empty space. Refer to the table that follows.
How to decrease the length of probes
. Make sure that the EM field is clear of
min
on page 89.
Clearance between the probe and other objects in the tank
Probe typeEmpty space (radius, R
[mm][inches]
Coaxial00
Double rod / cable1004
Single rod / cable30012
), around the probe
min
22
www.krohne.com07/2013 - 4000668002 - HB OPTIFLEX 2200 R02 en
OPTIFLEX 2200 C/F
3.6.2 How to attach probes to the bottom of the tank
If the liquid is agitated or turbulent, you can attach the probe to the bottom of the tank. The
procedure to attach the probe depends on the type of probe used.
CAUTION!
Keep the probe straight.
Double rod Ø8 mm / 0.3¨
INSTALLATION 3
Figure 3-12: How to attach a double rod probe to keep it straight
• Weld a tube with an internal diameter of 28...30 mm / 1.1...1.2¨ to the bottom of the tank.
iMake sure the tube aligns with the process connection at the top of the tank.
• Lower the probe into the tank.
• Put the end of the probe into the tube.
Double cable Ø4 mm / 0.15¨
Figure 3-13: How to attach a double cable probe to keep it straight
The probe counterweight has a hole with an M8 internal thread. You can also select the
appropriate options and attach:
1 An anchoring rod
2 A turnbuckle
For more data, contact your supplier.
www.krohne.com07/2013 - 4000668002 - HB OPTIFLEX 2200 R02 en
23
3 INSTALLATION
Single rod Ø8 mm / 0.3¨
Figure 3-14: How to attach a single rod probe to keep it straight
• Weld a tube with an internal diameter of 12 mm / 0.5¨ to the bottom of the tank.
iMake sure the tube aligns with the process connection at the top of the tank.
• Lower the probe into the tank.
• Put the end of the probe into the tube.
OPTIFLEX 2200 C/F
Single cable Ø4 mm / 0.15¨
Figure 3-15: How to attach a Ø4 mm / 0.15¨ single cable probe to keep it straight
1 Probe with threaded end
2 Probe with turnbuckle
3 Probe with chuck
4 If you chose a chuck to anchor the probe, we recommend that you fit a ferrule (metal sheath - not supplied) at the
bottom of the loop to prevent cable wear
24
The probe counterweight has a hole with an M8 internal thread. The other probe end options are
given in the illustration.
CAUTION!
If your device has a chuck, you must recalculate the probe length. For the procedure, refer to
How to decrease the length of probes on page 89
. If the device is not set to the correct probe
length, it is possible that the device will not measure correctly.
www.krohne.com07/2013 - 4000668002 - HB OPTIFLEX 2200 R02 en
OPTIFLEX 2200 C/F
Single cable Ø2 mm / 0.08¨
Figure 3-16: How to attach a Ø2 mm / 0.08¨ single cable probe to keep it straight
The probe counterweight has a hole with an M8 internal thread. You can attach the probe
counterweight to a threaded end.
Coaxial Ø22 mm / 0.8¨
INSTALLATION 3
Figure 3-17: How to attach a coaxial probe to keep it straight
• Weld a tube with an internal diameter of 23...25 mm / 0.91...1¨ to the bottom of the tank.
iMake sure the tube aligns with the process connection at the top of the tank.
• Lower the probe into the tank.
• Put the end of the probe into the tube.
If this is not possible, you can attach braces to the probe.
www.krohne.com07/2013 - 4000668002 - HB OPTIFLEX 2200 R02 en
25
3 INSTALLATION
3.6.3 Installation in standpipes (stilling wells and bypass chambers)
Use a standpipe if:
• The liquid is very turbulent or agitated.
• There are too many other objects in the tank.
• The device is measuring a liquid in a tank with a floating roof.
OPTIFLEX 2200 C/F
Figure 3-18: Installation recommendations for standpipes (stilling wells and bypass chambers)
1 Stilling well
2 Bypass chamber
3 Vent
4 Level of the liquid
INFORMATION!
Stilling wells are not necessary for devices with coaxial probes. But if there is a sudden change
in diameter in the stilling well, we recommend that you install a device with a coaxial probe.
26
www.krohne.com07/2013 - 4000668002 - HB OPTIFLEX 2200 R02 en
OPTIFLEX 2200 C/F
CAUTION!
Installation requirements
•
The standpipe must be electrically conductive. If the standpipe is not made of metal, obey the
instructions for empty space around the probe. For more data, refer to General requirements
on page 22
The standpipe must be straight. There must be no changes in diameter from the device
•
process connection to the bottom of the standpipe.
•
The standpipe must be vertical.
•
Recommended surface roughness: <±0.1 mm / 0.004¨.
•
The bottom of the stilling well must be open.
•
Adjust the probe to the center of the standpipe.
•
Make sure that there are no deposits at the bottom of the standpipe which can cause
blockage of the process connections.
•
Make sure that there is liquid in the standpipe.
Floating roofs
If the device is for a tank with a floating roof, install it in a stilling well.
INSTALLATION 3
.
Figure 3-19: Floating roofs
1 Sediment
2 Support fixtures
3 Stilling well
4 Floating roof
5 Product
6 Tank
www.krohne.com07/2013 - 4000668002 - HB OPTIFLEX 2200 R02 en
27
3 INSTALLATION
3.7 Installation recommendations for solids
3.7.1 Nozzles on conical silos
We recommend that you prepare the installation when the silo is empty.
DANGER!
Risk of electrostatic discharge (ESD): The device is resistant to electrostatic discharges of up to
30 kV, but it is the responsibility of the fitter and the user to prevent ESD.
CAUTION!
Install the device at the correct location to measure level correctly and prevent too much
bending and traction. If necessary, attach the probe to the bottom of the tank.
OPTIFLEX 2200 C/F
28
Figure 3-20: Installation recommendations for solids
a ≥ 300 mm / 12¨
d ≥ 300 mm / 12¨
Figure 3-21: Do not install the probe near to a product inlet
www.krohne.com07/2013 - 4000668002 - HB OPTIFLEX 2200 R02 en
OPTIFLEX 2200 C/F
3.8 How to install the device on the tank
3.8.1 How to assemble the single rod probe (single-piece probe)
INFORMATION!
This procedure is for devices with single rod probes that are not segmented (single-piece
probes).
INSTALLATION 3
Figure 3-22: Equipment needed to assemble the device
1 Union nut
2 2 locking nuts
3 Housing assembly
4 Single rod probe
5 Tools: two 8 mm open-ended wrenches (not supplied)
6 Tools: one 7 mm open-ended wrench (not supplied)
www.krohne.com07/2013 - 4000668002 - HB OPTIFLEX 2200 R02 en
29
3 INSTALLATION
OPTIFLEX 2200 C/F
Figure 3-23: Check the order number on each component
• Make sure that the housing assembly and the single rod have the same ID numbers.
• Remove the sticker from the probe.
30
www.krohne.com07/2013 - 4000668002 - HB OPTIFLEX 2200 R02 en