CONTENTS
Installation in the pipelinePages 4-5 and 7-8
GroundingPages 8-9
IFS 6000 F
VARIFLUX
IFS 6000 F
IFM 6010 K
IFM 6020 K
IFM 6080 K
IFM 1010 K
IFM 1080 K
IFM 6010 KIFM 6080 K
3.1M54EA1 059821Order No. DIN size 7.02143.31.00
US size 7.02143.71.00
0.73.00
Contents
System description2
Product liability and warranty2
Standards and approvals2
Items included with supply3
1 Important information for installation: PLEASE NOTE4-5
2 Suggestions for installation5
3 Instrument nameplate6
4 Flowmeter versions6
5 Installation in the pipeline and torques7
6 Limits8
7 Grounding8-9
8 Service10
9 Spares and order numbers10
10 Technical data11
11 Dimensions and weights12-14
Printed form to accompany flowmeters returned to Krohne15
System description
VARIFLUXelectromagnetic flowmeters are precision measuring instruments designed for the
linear flow measurement of process liquids.
The process liquids must be electrically conductive:≥ 05 µS/cm
≥ 20 µS/cm for demineralized cold water
The full-scale range Q
DN 2.5 - 80 /
1
/10” - 3” Q
can be set as a function of the meter size:
100%
= 0.01 - 220 m3/hr = 0.03 - 960 US Gal/min
100%
This is equivalent to a flow velocity of 0.3 - 12 m/s, or 1 - 40 ft/s.
Product liability and warranty
VARIFLUX electromagnetic flowmeters are designed solely for measuring the volumetric flowrate
of electrically conductive, liquid process products.
Special codes and regulations apply to their use in hazardous locations, and these are referred to
in the special “Ex” installation and operating instructions (supplied only with hazardous-duty
equipment).
Responsibility as to suitability and intended use of these electromagnetic flowmeters rests solely
with the operator.
Improper installation and operation of the flowmeters (systems) may lead to loss of warranty.
In addition, the “General conditions of sale” forming the basis of the purchase contract are
applicable.
If VARIFLUX flowmeters need to be returned to Krohne, please note the information given on the
last-but-one page of this manual. Krohne regret that they cannot repair or check your flowmeter(s)
unless accompanied by the completed form sheet.
Standards and approvals
Please refer to the installation and operating instructions for the signal converter.
Items included with supply (Arrangement of gaskets and grounding rings see Sect. 7).
Flanged units:with mounted grounding rings E and fitted gaskets D1.
Gaskets D2 not included, to be provided by customer.
Other connections:mounted adapter with fitted O-ring gaskets D1 and union nuts
(counterparts for the pipeline, such as tapered connectors, union nuts,
hinged clamps and gaskets D2 not included, to be provided by customer).
Inside diameter of measuring tube and pipeline
Size
1 Important information for installation: PLEASE NOTE
•Handling
Do not lift flowmeter by the
signal converter housing or
Do not set flowmeter down on signal
converter housing or terminal box.
the terminal box.
•Use only solventless detergents to clean the signal converter housing (polycarbonate).
•Temperatures
For operating pressure and vacuum load based on connection standards, see Section 6
“Limits“.
Ambient temperatureProduct temperature
Compact systems-25 to +60 °C (-13 to +140 °F)-20 to +060 °C (-4 to +140 °F)
-25 to +40 °C (-13 to +104 °F)-20 to +140 °C (-4 to +284 °F)
- EEx version-25 to +40 °C (-13 to +104 °F)-20 to +140 °C (-4 to +284 °F)
IFS 6000 F-25 to +60 °C (-13 to +140 °F)-20 to +060 °C (-4 to +140 °F)
- flange connections-25 to +40 °C (-13 to +104 °F)-20 to +180 °C (-4 to +356 °F)
- other connections-25 to +40 °C (-13 to +104 °F)-20 to +140 °C (-4 to +284 °F)
- EEx version-25 to +40 °C (-13 to +104 °F)-20 to +140 °C (-4 to +284 °F)
•Location and position as required,
but electrode axis X – • – • – • – X
Y
must be approximately horizontal in a
horizontal pipe run.
Y terminal box or converter housing
X
X
•Measuring tube must be completetly filled at all times.
•Direction of flow is arbitrary. Arrow on flowmeter can normally be ignored.
For exceptions, refer to Section ”Factory settings” in the installation and operating
instructions for the signal converter.
•Stud bolts and nuts: to fit, make sure there is sufficient room next to the pipe flanges.
•Vibration: support the pipeline on both sides of the compact flowmeter.
Level of vibration in conformity with IEC 068-2-34: 20-500 Hz, random / 2g runs / 30 minutes /
x, y, z directions.
•Do not expose to direct sunlight,
fit a sunshade if necessary, not included with flowmeter, to be provided by customer.
4
•Strong electromagnetic fields, avoid in vicinity of flowmeter.
•Straight inlet run minimum of 5 ××DN and outlet run minimum of 2 ××DN,
(DN = meter size), measured from the electrode axis.
•Vortex and corkscrew flow: increase length of inlet and outlet runs or install
flow conditioners.
•Mixing different process liquids: install flowmeter upstream of mixing point or at an
adequate distance downstream (minimum of 30 x DN), otherwise display may be unsteady.
•Plastic pipes and internally coated metal pipelines: grounding rings required,
see Section 7 ”Grounding”.
•Insulated pipeline: do not insulate flowmeter.
•Zero setting not necessary. To check, it should be possible to set ”zero” flow velocity
in the completely filled measuring tube. Shutoff valves should therefore be provided either
downsteam of the flowmeter or upsteam and downstream of the flowmeter.
2 Suggestions for installation
To avoid measuring errors due to gas/air inclusion or to pipe runing empty please observe the
following:
Highest point of pipe run
(Air bubbles collect in measuring
tube - faulty measurements!)
Preferred
locations
Downpipe
”Zero” flow
velocity.
Line drained.
Faulty
measurements!
Horizontal pipe run
Install in slightly ascending pipe section. If not
possible, assure adequate velocity to prevent air,
gas or vapor from collecting in
upper part of flow tube.
Open feed or discharge
Install meter in low section of pipe.
Downpipe over 5 m (16 ft) length
Install air valve X downstream of
flowmeter.
Long pipeline
Always install control and shutoff valves
downstream of flowmeter.
Pumps
Never install flowmeter on pump suction side.
> 5m
> 16 ft
open discharge
5