KROHNE IFM5080KEEx User Manual


$GGLWLRQWRWKH,QVWDOODWLRQDQG2SHUDWLQJ,QVWUXFWLRQV
Magnetic inductive flowmeters
ALTOFLUX
IFM 5080 K / i-EEx
with
7.30827.32
1 Contents
In this addition only the points which differ from the standard Installation and Operating Instructions are described. This addition is only valid for compact flowmeters with the IFC090 i-EEx signal converter with intrinsically safe signal in- and outputs.
1 CONTENTS........................................................................................................................2
2 DESCRIPTION OF THE SYSTEM.....................................................................................3
3 INSTALLATION IN THE PIPELINE....................................................................................3
4 ELECTRICAL CONNECTION............................................................................................3
5 CONNECTION DIAGRAMS OF THE SIGNAL IN/OUTPUTS ............................................6
6 OPERATION OF THE SIGNAL CONVERTER...................................................................7
7 SERVICE............................................................................................................................8
7.1 REPLACEMENT OF THE POWER SUPPLY FUSE................................................8
7.2 REPLACEMENT OF THE ELECTRONIC UNIT.......................................................9
8 PART NUMBERS...............................................................................................................9
9 TECHNICAL DATA...........................................................................................................10
10 MAINTENANCE ...............................................................................................................10
11 CERTIFICATE OF CONFORMITY............... FEHLER! TEXTMARKE NICHT DEFINIERT.
2
2 Description of the system
The explosion protected compact-flowmeters differ from the standard flowmeters mainly in their "internal assembly". Several types of protection are used in the compact flowmeter IFM 5080 K / i-EEx, depending on the ordered type of the connection compartment of the signal converter housing. See following overview:
Electronic compartment signal converter housing EEx d (Flameproof enclosure)
Connection compartment signal converter housing
- Housing and connection terminals for power supply:
standard: EEx e (Increased safety) optional: EEx d (Flameproof enclosure)
- Connection terminals of signal in/outputs (Current, Pulse/status/control and Fieldbus)
always EEx ia (Intrinsic safety. category ia)
Primary heads EEx m (Encapsulation) and EEx e (Increased safety)
Internal electrode circuit EEx ib (Intrinsic safety, category ib)
3 Installation in the pipeline
The installation of the compact flowmeter in the pipeline is equal to that of the standard compact flowmeters.
4 Electrical connection
For the electrical connection of the compact flowmeter the relevant national standard codes of practice for "Electrical installations of buildings" and "Electrical apparatus for use in potentially explosive atmospheres" have to be consulted.
In the connection compartment of the IFM 5080 K/ i-EEx the terminals for the intrinsically safe signal in/outputs and the non-intrinsically safe power supply circuits are placed. The location of the connection terminals is shown in the sketch on the next page.
The cable for the signal in/outputs leading into the IFM 5080 K / i-EEx is intrinsically safe. The connection of this cable must be carried out conform the requirements specified in the relevant national standard code of practice for the installation of electrically apparatus with type of protection "i".