The flowmeters are supplied ready for operation.
The primary head must be installed in the pipeline as described in the installation
instructions inside the packing of the primary head.
- Installation location and connection to power (Section 1)Pages 1/1-1/6
- Electrical connection of outputs and inputs (Section 2)Pages 2/1-2/6
- Factory settings and start-up (Section 3)Pages 3/1-3/2
Power the flowmeter. THAT’S ALL. The system is operative.
Operator control of the IFC 090 signal converter is described in Sections 4 and 5.
Pull-out condensed instructions are located in the centrefold of this manual, pages A – D.
3.1M61EA3 119721Order No. DIN A4: 7.02144.33.00
US size: 7.02144.73.00
Page 2
Contents
System description0/4
Product liability and waranty0/4
CE / EMC / Standards / Approvals0/4
Items included with supply0/4
Part A System installation and start-up1/1-3/2
1 Electrical connection: power supply1/1-1/6
1.1Location and important installation notes – PLEASE NOTE !1/1
1.2Connection to power1/2
1.3Electrical connection of separate primary head (F Version)1/3-1/6
1.3.1General information on signal cable A and field current cable C1/3
1.3.2Stripping (preparation) of signal converter1/3
1.3.3Grounding of primary head1/4
1.3.4Cable length (max. distance between signal converter and primary head)1/5
1.3.5Connection diagrams I and II (power supply, converter and primary head)1/6
2
2.1Combinations of outputs and inputs2/1
2.2Current output I2/1
2.3Pulse output B1 (terminals B1 / B⊥)2/2
2.4Status outputs B1 and B2 (terminals B1 / B⊥ and B2 / B⊥)2/3
2.5Control inputs B1 and B2 (terminals B1 / B⊥ and B2 / B⊥)2/4
2.6Connection diagrams for outputs and inputs2/5-2/6
Electrical connection of outputs and inputs2/1-2/6
3.
3.1Switch-on and measurement3/1
3.2Factory settings3/2
Part B IFC 090
4 Operation of the signal converter4/1-4/12
4.1Krohne operator control concept4/1
4.2Operating and check elements4/2
4.3Function of keys4/3-4/4
4.4Table of settable functions4/5-4/9
4.5Error messages in measuring mode4/10
4.6Reset totalizer and cancel error messages, RESET/QUIT menu4/11
This is equivalent to a flow velocity of 0.3 - 12 m/s or 1 - 40 ft/s.
Product liability and warranty
The electromagnetic flowmeters with the IFC 090 signal converter are designed solely for
measuring the volumetric flowrate of electrically conductive, liquid process products.
Special codes and regulations apply to their use in hazardous areas and these are referred to in
the special “Ex” Installation and Operating Instructions (supplied only with hazardous-duty
equipment).
Responsibility as to suitability and intended use of these compact electromagnetic flowmeters
rests solely with the operator.
Improper installation and operation of the flowmeters (systems) may lead to loss of warranty.
In addition, the “General conditions of sale” forming the basis of the purchase contract are
applicable.
If flowmeters need to be returned to Krohne, please note the information given on the last-but-one
page of these Instructions. Krohne regrets that it cannot repair or check your flowmeter(s) unless
accompanied by the completed form sheet.
CE / EMV / Standards / Approvals
• Electro magnetic flowmeters with the IFC 090 signal converter meet the requirements of the
EU-EMC Directives and bear the CE symbol.
• All factories and production sequences are ISO 9001 certified.
• Flowmeters are approved as hazardous-duty equipment to the harmonized European
Standards and to Factory Mutual (FM).
Further details are given in the “Ex” supplementary instructions provided only with
hazardous-duty equipment.
Items included with supply
• Signal converter as ordered
• Installation and operating instructions
• 2 plug connectors for power supply and outputs/inputs
• Special wrench for opening the housing covers
• Bar magnet to operate the display converter without opening the housing
• Additional instructions for hazardous-duty versions (applies only to hazardous-duty equipment)
0/4
Page 6
Part A System installation and start-up
1Electrical connection: power supply
1.1Location and important installation notes PLEASE NOTE !
•Electrical connection in accordance with VDE 0100 ”Regulations governing heavy-currentinstallations with line voltages up to 1000 V” or equivalent national regulations.
•Do not cross or loop cables inside the terminal compartment.
•Use separate cable entries (PG screwed conduit entries) for power supply,
field current cables, signal lines, outputs and inputs.
•Hazardous locations are subject to special regulations, see Section 6.1 and special
installation instructions for hazardous-duty (”Ex”) versions.
•Protect flowmeters or switchgear cabinets with built-in devices from direct sunlight.
Fit a sunshade if necessary.
•When installed in switchgear cabinets, signal converters must be adequately cooled,
e.g. use fans or heat exchangers.
•Do not expose signal converters to intense vibration.
Onl
y for separate systems/signal converters (F Versions)
•Keep distance between primary head and signal converter as short as possible.
Refer to Sect. 1.3.4 for maximum permissible length of signal and field current cables.
•Use the supplied Krohne signal cable A (Type DS), standard length 10 m (33 ft).
•Always calibrate primary head and signal converter together. Therefore, when installing,
ensure primary constant GK is identical; refer to instrument nameplate for the primary
head. If the GK is not identical, set the signal converter to the GK of the primary head.
Refer also to Sections 4 and 8.5.
•Dimensions of signal converter; refer to Section 10.2.
1/1
Page 7
PLEASE NOTE !
Connection to power 1.2
•Rated values:
The flowmeter housings protecting the electronic equipment from dust
and moisture must always be kept closed. The selected creepage distances and
clearances have been dimensioned in conformity with VDE 0110 and IEC 664 for
contamination category 2. Supply circuits and output circuits are designed to meet the
standards of overvoltage classes III and II, respectively.
•Safet
y isolation: the flowmeters (signal converters) must be provided with an
isolating facility.
1. AC V
ersion2. AC Version
230/240 V AC (200 - 260 V AC)200 V AC (170 - 220 V AC)
switch-selectable toswitch-selectable to
115/120 V AC (100 - 130 V AC)100 V AC (85 - 110 V AC)
•Note information on instrument nameplate: supply voltage and frequency
•The PE protective ground conductor for the power supply must be connected to the
separate U-clamp terminal in the terminal compartment of the signal converter.
For exceptions (compact systems), refer to installation instructions for the primary head.
•Connection diagrams I and II for power supply and electrical connection between
primary head and signal converter: refer to Section 1.3.5.
24 V AC
/ DC (tolerance bands: AC 20 - 27 V / DC 18 - 32 V)
•Note information on instrument nameplate: supply voltage and frequency.
•For measurement reasons, connect an FE functional ground conductor to the separate
U-clamp terminal in the terminal compartment of the signal converter.
•If connected to a functional extra-low voltage source (24 V AC / DC), provide for protective
separation (PELV) in conformity with VDE 0100 / VDE 0106 or IEC 364 / IEC 536,
or equivalent national regulations.
•Connection diagrams I and II for power supply and electrical connection between primary
head and signal converter: refer to Section 1.3.5.
Connection to power
for
internal
use only
L N
∼∼0L∼∼
1L
PE 100 -240 V AC
FE 24 V AC/DC
(PE protective ground conductor)
(FE functional ground conductor)
Warning: Instrument must be properly grounded to avoid personnel shock hazard.
1/2
Page 8
1.3 Electrical connection of separate primary head (F Versions)
1.3.1 General information on signal cable A and field current cable C
Use of the Krohne signal cable A with foil screen and magnetic shield will ensure proper operation
of the equipment.
• Signal cable to be solidly laid.
• Connect shields via stranded drain wires.
• Underwater and underground installation possible.
• Insulating material is flame-retardant to IEC IEC 332.1 / VDE 0742.
• Signal cables are low in halogen, unplasticized, and stay flexible at low temperatures.
Si
gnal cable A (Type DS), with double shielding
2
1 Stranded drain wire, 1st shield, 1.5 mm
2 Insulation
3 Conductor 0.5 mm
See Section 1.3.4 for max. permissible cable lengths
Customer-supplied materials
W
Insulation tubing (PVC), 2.0 - 2.5 mm dia. (1”)
X Heat-shrinkable tubing or cable sleeve
Y Wire end sleeve to DIN 41 228: E 1.5-8
Z Wire end sleeve to DIN 41 228: E 0.5-8
Please note:
For primary heads, stranded drain wire 1
must have the same length as stranded drain wire 7.
Signal cable A
bending radius
≥ 50 mm (≥ 2”)
1/3
Page 9
Grounding of primary head 1.3.3
• All flowmeters must be properly grounded.
• The grounding conductor should not transmit any interference voltages.
• Do not ground any other electrical device together with this conductor.
• In hazardous locations, the grounding conductor is used simultaneously for equipotential
bonding. Special grounding instructions are contained in the ”Ex” installation instructions
for hazardous-duty devices, supplied only with such devices).
• The primary head is connected to ground by means of an FE functional ground conductor.
• Special information on grounding various primary heads is contained in the separate
installation instructions for primary heads.
• These instructions also contain detailed descriptions on how to use grounding rings and
how to install primary heads in metal or plastic pipes or internally coated pipelines.
1/4
Page 10
1.3.4 Cable lengths (max. distance between signal converter and primary head)
Abbreviations and ex
planatory notes
used in the following tables, diagrams and connection diagrams
A Signal cable A (type DS), with double shielding, see diagram for max. length
C Field current cable C, with single shielding,
type and length see Table
2
D High-temperature silicone cable, 3 × 1.5 mm
(14 AWG) Cu, with single shielding,
max. length 5 m (16 ft)
2
E High-temperature silicone cable, 2 × 1.5 mm
(14 AWG) Cu, max. length 5 m (16 ft)
L Cable length
κκ
Electrical conductivity of the process liquid
ZD Intermediate connection box required in connection with cables D and E for primary heads
ALTOFLUX IFS 4000 F, PROFIFLUX IFS 5000 F and VARIFLUX IFS 6000 F in cases where
process temperatures exceed 150 °C (302 °F)
Recommended len
gth of signal cable
for magnetic field frequency ≤ 1/6 × power frequency
Primary head
ECOFLUX IFS 1000 F
AQUAFLUX F
ALTOFLUX IFS 4000 F
PROFIFLUX IFS 5000 F
VARIFLUX IFS 6000 F
Meter size
DN mm
010.0 - 0015
025.5 - 0150
010.0 - 1000
010.0 - 0150
200.0 - 1000
002.5 - 0015
025.0 - 0100
002.5 - 0015
025.5 - 0080
Nennweite
inch
3
/8- 1/
2
1 - 06
3
/8- 40
3
/8- 06
8 - 40
1
/10-1/
2
1 - 04
1
/10- 1/
2
1- 03
Signal line
Kurve
A4
A3
A1
A2
A1
A4
A2
A4
A2
Field current cable C: max. length and min. crosssection
LengthType of cable,single shielding
Process temperature < 150°C (302°F)Process temperature > 150°C (302°F)
III
Primary heads
Primary heads
1/6
Page 12
2 Electrical connection of outputs and inputs
2.1 Combinations of outputs and inputs
Assignment of the binary outputs and inputs as required,
see 3.07 Fct. “HARDWARE” and Sect. 3.2 “Factory settings”.
Current output I– active or passive mode
– internal power source for the binary outputs and inputs
Binary outputs/inputs– terminal B1:Terminals
pulse output B1
status output B1 or
control input B1
– terminal B2:
B1 B⊥⊥B2 I+ I I
status output B2 or
control input B2
Binary outputs
and inputs
Output/input combinations 1) – 6)
T
erminals: I+ / I / I
⊥⊥
B1 / B
⊥⊥
B2 / B
⊥⊥
Combination:1)IPS
2)IPC
3)ICS
4)ISC
5)IS1S2
6)IC1C2
2.2 Current output I
•The current output is galvanically isolated from all input and output circuits.
•Setting data and functions can note down on page 0/3.
Please also refer to Sect. 3.2 “Factory settings”.
⊥⊥
Current output
I = current output
P = pulse output
S = status output
C = control input
•Typical current output
I+ approx. 15 V DC positive
voltage of current output
I current sink
I⊥chassis ground,
current output
•All operating data and functions can be set.
•Display version: IFC 090 D, see Sect. 4 and 5.6, Fct. 1.05 for operator control
Basic version: IFC 090 B, see Sect. 6.2 for operator control
•The current output can also be used as an internal voltage source for the binary
outputs and inputs.
U
= 15 V DC I = 23 mA when operated without receiver instruments at the current output
int
I = 3 mA when operated with receiver instruments at the current output
Connection diagrams,
•
see Sect. 2.6: diagrams 1 2 3 6 9 10 11
2/1
Page 13
Pulse output B1 (terminals B1 / B⊥) 2.3
• The pulse output is galvanically isolated from the current output and all input circuits.
•
Setting data and functions can note down on page 0/3.
Please also refer to Sect. 3.2 “Factory settings” and Sect. 2.1 “Combinations of the
binary outputs and inputs”, Fct. 3.07 HARDWARE.
• Typical pulse output B1
B2 status output B2
or control input B2
B⊥⊥chassis ground, binary
outputs and inputs
B1 pulse output B1 (or status
output, control input)
• All operating data and functions can be set:
Display version:IFC 090 D, see Sect. 4 and 5.7, Fct. 1.06 for operator control
Basis version:IFC 090 B, see Sect. 6.2 for operator control
• The pulse output can be operated in the active or passive mode.
Active mode:The current output is the internal voltage source,
connection of electronic totalizers (EC)
Passive mode:External DC or AC voltage source required, connection of electronic (EC)
or electromechanical (EMC) totalizers
• Digital pulse division, interpulse period is non-uniform. Therefore, if frequency meters or
cycle counters are connected, allow for minimum counting interval:
2.4 Status outputs B1 and B2 (terminals B1 / B⊥ and B2 / B⊥)
•
The status outputs are galvanically isolated from the current output and all input circuits.
•
Setting data and functions can note down on page 0/3.
Please also refer to Sect. 3.2 “Factory settings” and Sect. 2.1 “Combinations of
binary outputs and inputs”, Fct. 3.07 HARDWARE.
• Typical status outputs B1 and B2
B2 status output B2
(or control input B2)
B⊥⊥chassis ground, binary
outputs and inputs
B1 status output B1 (or pulse
output, control input)
• All operating data and functions can be set:
Display version:IFC 090 D, see Sect. 4 and 5.8, Fct 1.06 or 1.07 for operator control
Basic version:IFC 090 B, see Sect. 6.2 for operator control
• The status outputs can be operated in the active or passive mode.
Active mode:The current output is the internal voltage source.
Passive mode:External DC or AC voltage source required.
•
Characteristics of the status outputs
Switch openSwitch closed
OFF (switched off)no function
ON (e.g. operation indicator)power OFFpower ON
SIGN I (F/R mode)Forward flowReverse flow
SIGN P (F/R mode)Forward flowReverse flow
TRIP POINT (limit switch)inactiveactive
AUTO RANGE (automatic range change)
high rangelow range
OVERFLOW I (I overranged)current output OKcurrent output overranged
OVERFLOW. P (P overranged)pulse output OKpulse output overranged
ALL. ERROR (all errors)errorsno error
FATAL.ERROR (fatal errors only)errorsno error
EMPTY PIPE (option)when measuringwhen measuring
Control inputs B1 and B2 (terminals B1 / B⊥ and B2 / B⊥)2.5
• The control inputs are galvanically isolated from the current output and all input circuits.
Setting data and functions can note down on page 0/3.
•
Please also refer to Sect. 3.2 “Factory settings” and Sect. 2.1 “Combinations of
binary outputs and inputs”, Fct. 3.07 HARDWARE.
• Typical current inputs B1 and B2
B1 / B2
⊥⊥
B
• All operating data and functions can be set:
Display version:IFC 090 D, see Sect. 4 and 5.19, Fct. 1.06 and 1.07 for operator control
Basic version:IFC 090 B, see Sect. 6.2 for operator control
• The control inputs must be operated in the passive mode.
• Function of the control inputs
OFFswitched off
EXT. RANGEexternal range change
OUTP. HOLDhold value of outputs
OUTP. ZEROset outputs to “MIN.VALUES”
TOTAL.RESETreset totalizer(s)
ERROR.RESETdelete error messages
Connection diagram, see Sect. 2.6: diagram 8
2/4
Page 16
2.6 Connection diagrams for outputs and inputs
Current output
1
I
active
Ri≤ 500 Ω
Pulse output
3
P
active
for EC
Milliammeter
Totalizer
– electronic (EC)
– electromechanical (EMC)
DC voltage,
external power source (U
ext
note connection polarity
External voltage source (U
DC or AC voltage,
connection polarity arbitrary
Current output
2
I
passive
U
U
≤ 15 V DC
ext
Ri≤ 500 Ω
ext
ext
),
),
Key, N/O contact
Relay for forward/reverse
flow measurement (F/R)
and/or automatic range change (BA)
with 1 or 2 changeover contacts
Please note! This terminal is not
provided for hazardous-duty signal
converters. There is no passive current
output, see connection diagrams
2 , 3 , 6 and 11 .
Active mode
The current output supplies the power for operation
of the inputs and outputs.
Passive mode
External power source required for operation of the
inputs and outputs.
Pulse output
4
P
passive
for EC or EMC
Pulse output
5
P
active EC
passive
I
U ≤ 15 V DC from current output
I ≤ 23 mA
I ≤ 03 mA
6
operation without current output
operation with current output
Status output
S
active
(connection to B2 and/or B1)
Last
U ≤ 15 V DC from current output
I ≤ 23 mA
I ≤ 03 mA
operation without current output
operation with current output
2/5
U
≤ 32 V DC/≤ 24 VAC
ext
I≤ 150 mA
Status output
7
S
passive
to B2 and/or B1)
I
U
U
≤ 32 V DC/≤ 24 V AC
ext
I≤ 150 mA
U
ext
(
connection
Last
ext
U
≤ 32 V DC
ext
I≤ 150 mA
Control input
8
C
passive
to B2 and/or B1)
U
≤ 32 V DC/≤ 24 VAC
ext
I≤ 6 mA
(
connection
I
U
ext
Page 17
F/R flow measurement
9
I
and P
active
F/R changeover via S
R
U
≤ 32 V DC / ≤ 24 V AC
ext
I≤ 150 mA
Automatic range change (BA) with F/R flow measurement
10
/ BA changeover via S2
I
active
passive
(B1)
VV
passive
(B2)
R
passive
U
ext
I
I
(B2) / F/R changeover via S1
y type
Rela
e.g. Siemens D1
(B1)
passive
BA
V/R
U
ext
I
I
U
≤ 32 V DC/≤ 24 V AC
ext
I≤ 150 mA
Automatic range change (BA)
11
I
/ BA changeover via S
active
V
RR
high
range
(B2)
active
V
low
range
Relay type
e.g. Siemens D1
U ≤ 15 V DC
I ≤ 03 mA
high
range
low
range
Relay type
e.g. NAIS-Matsushita
type RH-C or DR-C
2/6
Page 18
3 Start-up
3.1 Switch-on and measurement
•Before powering the system, please check that it has been correctly installed according
to Sect. 1 and 2.
•The flowmeter is delivered ready for operational use. All operating data have
been factory set in accordance with your specifications.
Please refer to Sect. 3.2 “factory settings”.
•Power the unit, and the flowmeter will immediately start process flow
measurement.
Basic version
•A light emitting diode (LED) under the cover of the electronic section shows the
measurement status. Remove the cover using the special wrench.
LED flashing . . .
•Refer to Sect. 6.2 for operator control of the “basic version”.
Display version, signal converter IFC 090 / D
•When powered, the display shows in succession: START UP and READY.
This is followed by display of the current flow rate and/or the current totalizer count
on either a continuous or alternating basis, depending on the setting under Fct. 1.04.
•Refer to Sect. 4 and 5 for operator control of the “display version”.
, signal converter IFC 090 / B
green:measurement correct, everything all right.
green/ red: momentary overdriving of outputs
and/or A/D converter.
red:fatal error, parameter error or hardware fault,
please consult factory.
3/1
Page 19
Factory settings 3.2
All operating data are factory set according to your order specifications.
If you have not made any particular specifications at the time of ordering, the instruments will be
delivered with the standard parameters and functions listed in the Table below.
To facilitate easy and rapid initial start-up, current output and pulse output are set to process flow
measurement in “2 flow directions”, so that the current flowrate is displayed and the volumetric
flow counted independent of the flow direction. On instruments equipped with a display,
measured values may possibly be shown with a “ – ” sign.
This factory setting for the current and pulse outputs may possibly lead to measuring errors,
particularly in the case of volume flow counting:
for example, if pumps are switched off and a “backflow” occurs which is not within the range of
the low-flow cutoff (SMU), or if separate displays and counts are required for both flow directions.
To avoid faulty measurements, therefore, it may be necessary to change the factory setting of
some or all of the following functions:
– low-flow cutoff SMU, Fct. 1.03, Sect. 5.3
– current output I, Fct. 1.05, Sect. 5.6
– pulse output P, Fct. 1.06, Sect. 5.7
– display (option), Fct. 1.04, Sect. 5.4
For special applications, for example “pulsating flow”, see Sect. 6.
Instrument o
peration:
Display versions:IFC 090 _/ D, operation: refer to Sect. 4 and 5.
Basic versions:IFC 090 _/ B, operation: refer to Sect. 6.2.
Table of standard factory settings
FunctionSetting
1.01 Full-scale range Q
1.02 Time constant3 s, for I, S
1.03 Low-flowON: 1 %
cutoff SMUOFF: 2 %
1.04 Display (option)
flow ratem
totalizer(s)m3 or US Gal
1.05 Current output I
function2 directions
range4 - 20 mA
error message22 mA
of the electronic section using the
special wrench (supplied).
... the 3 magnetic sensors 5 and the
4
supplied bar magnet
without opening the housing.
PLEASE NOTE!
Do not damage the screw thread and the gasket, never allow dirt to accumulate, and make sure they
are well greased at all times.
Damaged gasket must be replaced immediately!
1
Display,1st line
2
Display,2nd line
Display,3rd line:arrows to identify display
3
Flowrate
Totalizer
+totalizer
current flowrate
–totalizer
Overrange
Σ
Ioverranging, current output I
sum totalizer (+ and –)
Poverranging, pulse output P
4
Keys for operator control of signal converter
5
Magnetic sensors to set the converter by means of a handheld bar magnet without opening
housing. Function of sensors same as keys .
6
Compass field, signals actuation of a key.
4
4/2
Page 22
4.3 Function of keys
The cursor (flashing part of display) has a grey background in the following descriptions.
o start operator control
T
Measuring mode
Operator control mode
1 3 . 5 7 1
m 3 / h r
PLEASE NOTE: When “YES” is set under Fct. 3.04 ENTRY CODE, “CodE 1 - - - - - - - - -”
appears in the display after pressing the → key.
The 9-keystroke Entry Code 1 must now be entered: → → → ↵ ↵ ↵ ↑ ↑ ↑
(each keystroke acknowledged by “
o terminate operator control
T
Press key ↵ repeatedly until one of the following menus
Fct. 1.00 OPERATION, Fct. 2.00 TEST or Fct. 3.00 INSTALL. is displayed.
F c t . 3. 0 0
INSTALL.
→
”).
*
Press key
↵
↵
F c t . 1. 0 0
O P E R AT I O N
S T O R E Y E S
Store new parameters: acknowledge by
pressing key↵ . Measuring mode continued
with the new parameters.
New parameters not to be stored:
press key ↑ to display
„STORE.NO“.
Measuring mode continued with the „old“
↵
parameters after pressing key
.
4/3
Page 23
To change numbers
select next number
3 9 7. 3 5
↑
m 3 / h r
To shift cursor (flashing position)
shift to right
3 9 7. 3 5
→
m 3 / h r
To alter texts (units)
For units, the numerical value is
converted automatically.
3. 7 6 9 9
select next text
↑
L i t e r / S e c
To transfer from text (unit) to number setting
Change to number setting
3 9 7. 4 5
m 3 / h r
3 9 7. 3 5
m 3/hr
9 3. 3 6 5
U S. G a l / m i n
1 3. 5 7 1
→
m 3 / h r
To transfer to subfunction
Subfunctions do not have a “Fct.No.“ and are identified by a “ → ”.
↵
1 D I R.
To revert to function display
1 0. 3
↵
Sec
1 3. 5 7 1
m 3/hr
→
R A N G E I
F c t. 1. 0 2
T IME CONST.
4/4
Page 24
4.4 Table of settable functions
Abbreviations used
B1/B2Status output, control input
DNNominal size, meter size
F
= 1/2 x pulse width [s]
max
≤ 1 kHz, if “AUTO” or “SYM.” is selected
Qactual flowrate
Q
100%
Q
max
under subfunction “PULSWIDTH”
F
F
F
F/RForward/reverse flow in F/R mode
GKPrimary constant
ICurrent output
I
0%
I
100%
PPulse output
P
P
= 10 pulse/h
min
Conversion factor volume for any unit,
M
see Fct. “FACT. VOL.”
Conversion factor time for any unit,
T
see Fct. 3.05 “FACT. Time”
Current at 0% flow
Current at 100% flow
= F
= F
max
min
/ Q
/ Q
100%
100%
max
min
Q
min
SMULow-flow cutoff for I and P
vFlow velocity
v
max
v
min
Fct.TextDescription and settings
1.00OPERATIONOperations menu
1.01FULL SCALEFull-scale range for flowrate Q
Select unit
• m3/hr• Liter/Sec•US.Gal/min
• user unit, factory set is “Liter/hr” or “US MGal/day” (see Fct. 3.05)
Press →key to transfer to number setting.
Setting ranges
The ranges are dependent on the meter size (DN) and the flow
π
velocity (v): Q
= DN2x v
min
4
Nom. dia./meter sizevmin = 0,3 m/s (1 ft/s) vmax = 12 m/s (40 ft/s)
•
DN 2.5–1000 /
1
/10”–40”: 0.0053 –033 900 m3/hr
Press ↵key to return to Fct. FULL SCALE.
→ VALUE PPulse value (Fct. 1.06
“VALUE P”
With the “old” pulse values the output frequency (F)
would have been exceeded or not reached.
P
= F
min
/ Q
100% Pmax
min
1.02TIMECONST.Time constant
Select:
Press ↵key to transfer to number setting.
• ALL (applies to display and all outputs)
• ONLY I (only display, current and status outputs)
Range:• 0.2 – 99.9 Sec
Press ↵key to return to Fct. 1.02 TIMECONST.
1.03L.F.CUTOFFLow-flow cutoff (SMU)
• OFF (fixed values: ON = 0.1% / OFF = 0.2%)
• PERCENT (variable values)ON OFF
Press →key to transfer to number setting.
Note: Cutoff “off” value must be greater than cutoff “on” value.
Press ↵key to return to Fct. 1.03 L.F. CUTOFF.
100% flow = full scale range
π
=DN2x v
4
at v
max
π
=DN2x v
4
at v
min
Max. flow velocity (12 m/s / 40 ft/s) at Q
Min. flow velocity (0.3 m/s / 1 ft/s) at Q
100%
minQmax
/ max. full-scale range (Q
max
= 12 m/s / 40 ft/s
/ min. full-scale range (Q
min
= 0.3 m/s / 1 ft/s
π
= DN2x v
4
100%
100%
100%
100%
max
0.0237 –152 000 US.Gal/min
) has been changed.
= F
max
/ Q
Check new values!
100%
1 – 19% 2 – 20%
)
)
4/5
Page 25
Fct.TextDescription and settings
1.04DISPLAYDisplay functions
→
DISP.FLOWSelect flow display
→
DISP.TOTAL.Select totalizer display
→
DISP.MSG.Additional messages required in measuring mode?
1.05CURRENT ICurrent output I
→
FUNCT. ISelect function for current output I
→
REV.RANGESet full-scale range for reverse flow of Q
→
RANGE ISelect measuring range
→
I ERRORSelect error value
1.06Output/input B1
PULS. B1Pulse output B1
STATUS B1Status output B1
CONTROL B1Control input B1
1.07Output/input B2
STATUS B2Status output B2
CONTROL B2Control input B2
• NO DISP. •
• m3/hr•PERCENT
• Liter/Sec • BARGRAPH (value and bargraph display in %)
• US.Gal/min
Press ↵key to transfer to subfunction “DISP. TOTAL.”.
• NO DISP. (totalizer switched on but not displayed)
• OFF (totalizer switched off)
• m3• Liter• US.Gal
• user unit, factory set is “Liter” or “US MGal” (see Fct. 3.05).
Press →key to transfer to format setting.
Format setting
• Auto (exponent notation)
• # . #######• ##### . ###
• ## . ######• ###### . ##
• ### . #####• ####### . #
• #### . ####• ########
Press ↵key to transfer to subfunction “DISP.MSG”.
• NO• YES (cyclic change with displays of measured values)
Press ↵key to return to Fct. 1.04 DISPLAY.
• OFF (switched off)
• 1 DIR. (1 flow direction)
• 2 DIR. (forward/reverse flow, F/R flow measurement)
Press key ↵, transfer to subfunction
if
“2 DIR.”
(appears only when “2 DIR.” selected)
• 100 PCT (same as forward flow Q
• PERCENT setting range: 005 - 150% of Q
(different value for reverse flow)
To transfer to number setting, press key→!
Press key ↵ to transfer to subfunction
• 0 - 20 mA • 4 - 20 mA (fixed ranges)
• mA (user-defined range) I
(Value I0%< I
To transfer to number setting, press key →!
Press key ↵ to transfer to subfunction
• 22 mA • 0.0 to I0%mA (variable, see above if I0%> 1 mA)
To transfer to number setting, press key →!
Press key ↵to revert to
user unit, factory set is “Liter/hr” or “US MGal/day (see Fct. 3.05)
selected, transfer to subfunction
!) 0 - 16 mA 4 - 20 mA
100%
}
Functional description of pulse output B1, status output B1
or control input B1, see next page.
}
Functional description of status output B2
or control input B2, see next page.
“RANGE I”;
“REV.RANGE”!
100%
, see Fct. 1.01)
100%
100%
“RANGE I”.
- I
0%
Fct. 1.05 CURRENT. I.
B1 = terminal, assigned as output or input,
see Fct. 3.07 “HARDWARE”
B2 = terminal, assigned as output or input,
see Fct. 3.07 “HARDWARE”
100%
“I ERROR“.
4/6
Page 26
Fct.TextDescription and settings
1.06PULS B1Pulse output B1 (see Fct. 3.07 HARDWARE)
→
FUNCT. PSelect function for pulse output P
→
SELECT PSelect pulse type
→
PULSWIDTHSelect pulse width
→
VALUE PSet pulse value per unit volume (appears only when
→
VALUE PSet pulse value per unit time (appears only when “PULSE/TIME”
• OFF (switched off)
• 1 DIR. (1 flow direction)
• 2 DIR. (forward/reverse flow, F/R flow measurement)
Press key ↵to transfer to subfunction
• PULSE/VOL. (pulses per unit volume, flowrate)
• PULSE/TIME (pulses per unit time for 100% flowrate)
Press key ↵to transfer to subfunction
• 0.01 - 1.00 Sec (only for F
• AUTO (automatic = 50% of the period duration
of the 100% output frequency)
• SYM. (symmetrical = pulse duty ratio 1:1 over total range)
• xxxx PulSe/user-defined unit, factory-set is “hr” or “day” (see Fct. 3.05)
Setting range “xxxx” is dependent on the pulse width, see above
Press key ↵to return to
user-defined unit, factory-set is “Liter” or “US M.Gal” (see Fct. 3.05)
= F
min
< 50 pulses/s)
max
/ Q
min
100%Pmax
Fct.1.06“PULS. B1”.
Fct.1.06“PULS. B1”.
“SELECT P”.
“PULSWIDTH”.
“VALUE P”.
= F
max
/ Q
100%
1.06STATUS B1Status output B1 and B2 (see Fct. 3.07 HARDWARE)
1.07STATUS B2• ALL ERROR• FATAL ERROR• OFF• ON
1.06CONTROL B1Control input B1 and B2 (see Fct. 3.07 HARDWARE)
1.07CONTROL B2• OFF• EXT.RANGE (external range change)
• SIGN. I
• SIGN. P
• OVERFLOW I
• OVERFLOW P
• AUTO RANGE (automatic range change)
Setting range 5 - 80 PERCENT (= ratio of lower to upper range from
1:20 to 1:1.25. Value must be higher than that of Fct. 1.03 L.F. CUTOFF)
• TRIP POINT: XXX- YYYXXX > YYY N/O contact
• EMPTY PIPE (signals that pipe is “empty”, only if option installed)
Press key
Press key ↵to return to
Setting range: 5 - 80 PERCENT (= ratio of lower to upper range from
1:20 to 1:1.25. Value must be greater than that of Fct. 1.03 L.F. CUTOFF).
Press
• OUTP.HOLD (hold value of outputs)
• OUTP.ZERO (set outputs to “min.values“)
• TOTAL. RESET (reset totalizers)
• ERROR. RESET (delete error messages)
Press key ↵to return to
F/R flow measurement
}
overranging
}
of outputs
0 - 150% 0 - 150% XXX < YYY N/C contact
Hysteresis ≥ 1%
(difference between XXX value and YYY value)
↵
to transfer to number setting.
Fct. 1.06 or 1.07 STATUS B1 orB2.
↵
key to transfer to number setting.
Fct. 1.06 or 1.07 CONTROL B1 or B2.
dynamic response
of outputs, see Fct. 1.02
TIMECONST.:
I = I only
}
P = ALL
4/7
Page 27
Fct.TextDescription and settings
2.00TESTTest menu
2.01TEST QTest measuring range Q
Precautionary query
• SURE NO
• SURE YES
select value:
Press ↵key to return to Fct. 2.01 “TEST Q”.
Press
↵
key, then use ↑ key to
-110 / -100 / -50 / -10 / 0 / +10 / +50 / +100 / +110 PCT.
of set full-scale range Q
Displayed value present at outputs I and P.
100%
Press ↵key to return to Fct. 2.01 “TEST Q”.
2.02HARDW. INFOHardware information and error status
Before consulting factory, please note down all 6 codes.
→ MODUL ADCX . X X X X X . X X
Y Y Y Y Y Y Y Y Y Y
Press ↵key to transfer to
→ MODUL IOX . X X X X X . X X
Y Y Y Y Y Y Y Y Y Y
Press ↵key to transfer to
→ MODUL DISP.X . X X X X X . X X
Y Y Y Y Y Y Y Y Y Y
Press ↵key to return to Fct. 2.02
“HARDW. INFO“.
3.00INSTALL.Installation menu
3.01LANGUAGESelect language for display texts
• GB / USA (English)• F (French)
• D (German)• others on request
Press ↵key to return to Fct. 3.01
“LANGUAGE”.
3.02FLOWMETERSet data for primary head
→ DIAMETERSelect size from meter size table
• DN 10 - 1000 mm equivalent to 3/8 - 40 inch
Select with
Press
→ FULL SCALEFull-scale range for flow Q
To set, refer to Fct. 1.01 “FULL SCALE” above.
Press
→ VALUE PPulse value (Fct. 1.06
↑
key.
↵
key to transfer to subfunction “FULL SCALE”.
100%
↵ key to transfer to subfunction “GK VALUE”.
“VALUE P”
) has been changed.
With the “old” pulse values the output frequency (F)
would have been exceeded or not reached.
= F
min
/ Q
100% Pmax
P
min
= F
max
/ Q
100%
→ GK VALUESet primary constant GK
see primary head nameplate.
ge:• 1.0000 - 9.9999
Ran
Press ↵key to transfer to subfunction
→ FIELD FREQ.Magnetic field frequency
Values: 1/2, 1/6, 1/18 and 1/36
of power frequency, see nameplate.
Press ↵key to transfer to subfunction “FLOW DIR.”;
DC units only: to transfer to subfunction
→ LINE FREQ.Normal line frequency in your country
Please note: This function is only provided for units
with DC power supply to suppress line-frequency interference.
Values:
50 Hz
and
60 Hz
Press ↵key to transfer to subfunction “FLOW DIR.”.
→ FLOW DIR.Define flow direction (in F/R mode: forward flow).
Set according to direction of arrow on primary head:
• + DIR.• – DIR.
Select using
Press ↵key to return to Fct. 3.02 “FLOWMETER”.
.
Check new values!
“FIELD. FREQ.”.
“LINE FREQ.”.
↑
key.
“MODUL IO“.
“MODUL DISP.“.
4/8
Page 28
Fct.TextDescription and settings
3.03ZERO SETZero calibration
Note:
Carry out only at “0” flow and with completely filled
measuring tube!
Precautionar
• CALIB. NO
• CALIB. YES Press ↵
A “WARNING” sign appears when flowrate “> 0“;
acknowledge by pressing
• STORE NO (do not store new zero value)
• STORE YES (store new zero value)
y query
Press ↵key to return to Fct. 3.03 “ZERO SET”.
key to start calibration.
Duration approx. 25 seconds, current flowrate
displayed in the selected unit (see Fct. 1.04 “DISP. FLOW”).
↵
key.
Press ↵key to return to Fct. 3.03 “ZERO SET”.
3.04ENTRY CODEEntry code required to enter setting mode?
• NO(= entry with → only)
• YES (= entry with → and Code 1: → → → ↵ ↵ ↵ ↑ ↑ ↑ )
Press ↵to return to Fct. 3.04 “ENTRY CODE”.
3.05USER UNITSet any required unit for flowrate and counting
→ TEXT VOL.Set text for required flowrate unit (max. 5 characters)
Factory-set: “Liter” or “MGal”.
Characters assi
• A-Z, a-z, 0-9, or “ – “ (= blank character).
gnable to each place:
Press ↵key to transfer to subfunction “FACT. VOL.”
→ FACT. VOL.Set conversion factor (FM) for volume
Factory set “1.00000” for “Liter” or “2.64172E-4” for “US MGal”
(exponent notation, here: 1x103or 2.64172x10-4).
Factor FM= volume per 1m3.
Setting range • 1.00000 E-9 to 9.99999 E+9 (= 10-9to 10+9)
Press ↵ key to transfer to subfunction “TEXT TIME”.
→ TEXT TIMESet text for required time unit (max. 3 characters)
→ FACT. TIMESet conversion factor (FT) for time
Factory-set: “hr” or “day”:
Characters assignable to each place:
• A-Z, a-z, 0-9, or “ – “ (= blank character).
Press
↵ key to transfer to subfunction “FACT. TIME”
Factory-set: “3.60000 E+3” for “hour” or “8.64000 E+4” for “day”
(exponent notation, here: 3.6 x103or 8.64 x10-4).
Set factor FTin seconds.
Setting range
• 1.00000 E-9 to 9.99999 E+9 (= 10-9to 10+9)
Press ↵ key to return to Fct. 3.05 “USER UNIT”.
3.06APPLICAT.Set overload point for A/D converter
→ FLOW• STEADY (150% of Q
Press key ↵to return to
with installed option
change to subfunction
→ EMPTY PIPESwitch on “empty pipe” identifier option?
(appears only when this option is installed)
• YES • NO
Select with key ↑.
Press ↵key to return to
3.07HARDWAREAssign outputs and inputs to terminals B1 and B2
→ TERM.B1Terminal B1
• PULSOUTP. • STATUSOUTP. • CONTROLINP.
) • PULSATING (1000% of Q
100%
Fct. 3.06 “APPLICAT.“,
“empty pipe“
“EMPTY PIPE“.
,
Fct. 3.06 “APPLICAT.“.
Select with key ↑.
Press key ↵ to transfer to subfunction
→ TERM.B2Terminal B2
• STATUSOUTP. • CONTROLINP.
“TERM. B2”.
Select with key ↑.
Press key ↵ to return to
Fct. 3.07 “HARDWARE”.
100%
)
4/9
Page 29
Error messages in measuring mode4.5
The following list gives all errors that can occur during process flow measurement.
Errors shown in display when “YES” set in Fct. 1.04 DISPLAY, subfunction “DISP. MSG.”:
Error messagesDescription of errorError clearance
LINE INT.Power failure Note:Cancel error in RESET-QUIT menu
OVERFLOW ICurrent output overranged.Check and if necessary correct
OVERFLOW PPulse output overranged.Check and if necessary correct
TOTALIZERTotalizer has been resetCancel error message in
FATAL. ERRORFatal error, all outputs setPlease consult factory.
EMPTY PIPEPipe has run dry.Fill pipe.
no counting during power failureReset totalizer if necessary.
instrument parameters. After
elimination of cause, error message
deleted automatically.
Note:instrument parameters. After
totalizer deviation possible.elimination of cause, error message
deleted automatically.
RESET/QUIT. menu.
converter overrangedafter elimination of cause.
to “min. values”
This message appears only when
the “empty pipe identifier”option is installed and the function
is switched on under Fct. 3.06
APPLICAT., submenue
“EMPTY PIPE”.
4/10
Page 30
4.6 Reset totalizer and cancel error messages, RESET/QUIT menu
Cancel error messages in RESET / QUIT menu
KeyDisplayDescription
- - - - - - -- - - - - / - - -Measuring mode
↵CodE 2- -Key in entry code 2 for RESET/QUIT
↑ →ERROR QUIT.Menu for error acknowledgement
→QUIT. NODo not delete error messages,
The cursor, flashing part of display, is shown below in bold type.
•Change measuring range of current output and value for error messages (Fct. 1.05):
•Change measuring range from 04-20 mA to 00-20 mA
•Change value for error messages from 0 mA to 22 mA
KeyDisplayDescription
→If ”YES“ set under Fct. 3.04 ENTRY CODE, key in the
Fct. 1.00OPERATION→Fct. 1.01FULL SCALE
4x ↑Fct. 1.05CURRENT I
→FUNCT. I
→↵RANGE IIf ”REV. RANGE“ appears here,
→04-20mAOld current range2x ↑00-20mANew current range
↵I ERROR
→0mAOld value for error messages
↑22mANew value for error messages
↵Fct. 1.05CURRENT I
↵Fct. 1.00OPERATION
↵STORE YES
↵- - - - - - -- - - - - / - - -Measuring range with new data for the current output
9-keystroke CODE 1 now: → → → ↑ ↑ ↑ ↵ ↵ ↵
press keys → and ↵ again.
4/12
Page 32
5 Description of functions
5.1 Full-scale range Q
100%
Fct. 1.01 FULL SCALE
Press→key.
Choice of unit for full-scale range Q
100%
•m3/hr(cubic metres per hour)
•Liter/Sec(litres per second)
•US.Gal/min(US gallons per minute)
•User-defined unit, factory-set is „Liter/hr“ (litres per hour) or “US MGal/day”, see Sect. 5.13.
Select with
↑ key.
Use →key to transfer to numerical setting, 1st number (cursor) flashes.
Set full-scale range Q
100%
The setting range is dependent on meter size (DN) and flow velocity (v).
Q
min
π
= DN2x v
4
minQmax
π
= DN2x v
4
0.0053– 33 929m3/hr
0.00147 –9 424.5Liter/Sec
0.00233 – 151 778US.Gal/min
(refer to flow table in Sect. 10.1)
max
Change flashing number (cursor) with ↑key.
Use
→
Press
key to shift cursor 1 place to right.
↵
key to return to Fct. 1.01 FULL SCALE
.
Note if “VALUE P” is displayed after pressing ↵ key:
PULSE/VOL. is set under Fct. 1.06 PULS B1, subfunction “SELECT P”. Due to the changed
full-scale range Q
P
= F
min
/ Q
min
, the output frequency (F) of the pulse output will be over- or undershot:
100%
100%
P
= F
max
max
/ Q
100%
Change pulse value accordingly, see Sect. 5.7 pulse output B1, Fct. 1.06.
5.2 Time constant
Fct. 1.02 TIMECONST.
Press →key.
Choice
•ALL(applies to display and all outputs)
•ONLY I(applies only to display, current and status output)
•PERCENT(variable tripping points: ON = 1 - 19 % / OFF = 2 - 20 %)
Select with ↑key.
Transfer to number setting using
→
key (only if “PERCENT” selected).
1st number (cursor) flashes.
Setting the numerical value when “PERCENT” selected
•01 to 19(cutoff “on” value, left of hyphen)
•02 to 20(cutoff “off” value, right of hyphen)
Change flashing number (cursor) with the ↑key.
Shift cursor 1 place to right using
Press
↵
key to return to Fct. 1.03 L.F.CUTOFF.
Note: The cutoff “off” value must be greater than the cutoff “on” value.
Fct. 1.04 DISPLAY
→
key.
Press →key.
Display 5.4
→ DISP. FLOW = select unit for display of flowrate,
•NO DISP.(no display)
•m3/hr(cubic metres per hour)
•Liter/Sec(litres per second)
•US.Gal/min(US gallons per minute)
•user-defined unit, factory-set: “Liter/hr” (litres per hour) or “US MGal/day”, see Sect. 5.15
•PERCENT(percentage display)
•BARGRAPH(numerical value and bar graph display in %)
press → key
Select with ↑key.
Press
↵
key to transfer to subfunction “DISP. TOTAL”.
→ DISP. TOTAL = select unit for totalizer display,
•NO DISP.(no display)
•OFF(internal totalizer switched off)
• + TOTAL. • – TOTAL. • +/– TOTAL. • SUM (ΣΣ) • ALL (sequential)
press → key
Select with ↑key.
Transfer to totalizer unit setting using
•m3(cubic metres)
•Liter(litres)
•US.Gal(US gallons)
•user-defined unit, factory-set: “Liter” or “US MGal”, see Sect. 5.12
↵
key.
Select with ↑key.
Transfer to totalizer format setting using
→
key.
Continuation see next page
5/2
Page 34
Setting of totalizer format
•Auto (exponent notation)
•# . #######• ##### . ###
•## . ######• ###### . ##
•### . #####• ####### . #
•#### . ####• ########
Select with key↑.
Press
↵ key to transfer to subfunction “DISP. MSG”.
→
DISP. MSG. = additional messages required in measuring mode,
press → key
•NO(no other messages)
•YES(display other messages, e.g. errors, in sequence with the measured values)
Select using the ↑key.
Press
↵
key to return to Fct. 1.04 DISPLAY.
Note: “BUSY” is displayed in the measuring mode when all displays are set to “NO DISP.” or
“NO”. Sequencing of displays is automatic. However, in the measuring mode, manual sequencing
can be carried out with the ↑ key. Return to automatic sequencing after approx. 3 minutes.
Please refer to Sect. 3.2 “factory settings”
5.5 Internal electronic totalizer
The internal electronic totalizer counts in m3, regardless of the unit set under Fct. 1.04,
subfunction “DISP. FLOW”.
The counting range is dependent upon the meter size and has been selected such that the
totalizer will count for a minimum of 1 year without overflow:
Only part of the totalizer count is shown in the display because it is not possible to output
a 14-digit number. Unit and format of the display are freely selectable, see Fct. 1.04, subfunction
“DISP. TOTAL” and Sect. 5.4. This determines which part of the count is to be displayed.
Display overflow and totalizer overflow are independent of one another.
Exam
ple
Internal count0000123 . 7654321m
3
Format, display unitXXXX . XXXXLiter
Internal count in unit0123765 . 4321000Liter
Displayed3765 . 4321Liter
Exception: When “OFF” selected, return to Fct. 1.06 PULS B1.
→ SELECT P = select pulse type,
press → key
•PULSE/VOL. (pulses per unit volume, flowrate)
•PULSE/TIME (pulses per unit time for 100% flowrate)
Select with key ↑.
Press
↵
key to return to subfunction “PULSWIDTH”.
→ PULSWIDTH = select pulse width,
press key
→
•AUTO(automatic = 50% of the period length of the 100% output frequency)
•SYM.(symmetrical = 1:1 pulse duty ratio over entire range)
•SEC.(variable) setting range 0.01 - 1.00 SEC
Select with key ↑.
Press
→
1st number (cursor) flashes. Set numbers using keys
key to transfer to number setting.
↑ and →.
Press ↵key to transfer to subfunction “VALUE P” or return to Fct. 1.06 PULS B1,
depending on choice of pulse type in subfunction “SELECT P”.
Please note
= 10 pulses/h
F
min
F
=
max
2 x pulse width [s]
1
If “AUTO” or “SYM.” is selected under
subfunction “PULSWIDTH”
F
≤ 1 kHz !
max
5/5
Page 37
→ VALUE P = set pulse value per unit volume
(appears only when “PULSE/VOL.” set under “SELECT P”, press →
key.
• XXXX PulS/m3
• XXXX PulS/Liter
• XXXX PulS/US.Gal
• XXXX PulS/ user unit, factory-set: “Liter” or “US MGal”, see Sect. 5.13.
Select using ↑key.
Transfer to number setting with
→
key. 1st digit (cursor) flashes.
Set numerical value
• XXXX(setting range depends on pulse width and full-scale range: P
min
= F
min
/ Q
100%
P
= F
max
/ Q
max
Change flashing digit (cursor) with ↑key,
shift cursor 1 place to right or left with
Press
↵
key to return to Fct. 1.06 PULS B1.
→
key.
or
→ VALUE P = set pulse value per unit time,
(a
ppears only when ”PULSE/TIME” has been set under “SELECT P”),
press→key.
• XXXX PulSe/Sec
• XXXX PulSe/min
• XXXX PulSe/hr
• XXXX PulSe/ user unit, factory-set: “hr”, or “day”, see Sect. 5.13.
Select using ↑key.
Transfer to number setting with
→
key, 1st digit (cursor) flashes.
100%
)
Set numerical value
• XXXX(setting range depends on pulse width)
Change flashing digit (cursor) with ↑key,
shift cursor 1 place to right or left with
Press
↵
key to return to Fct. 1.06 PULS B1.
→
key.
Please refer to Sect. 3.2 “factory settings”
Refer to Sect. 2.6 for connection diagrams, and to Sect. 5.15 for characteristics.
5/6
Page 38
5.8 Status outputs B1 and B2
NOTE: Check whether under Fct. 3.07 “HARDWARE” the output terminal “B1” and/or “B2” is
defined as status output B1 and/or B2, refer also to Sect. 2.1 and Sect. 5.16.
Fct. 1.06 and/or 1.07 ST
Press key →.
Select function of status outputs,
•ALL ERROR(indicates all errors)
•FATAL.ERROR(indicates fatal errors only)
•OFF(switched off, no function)
•ON(indicates that flowmeter is operative)
•SIGN. I
•SIGN. P
•OVERFLOW Ioverranging I = I ONLY
•OVERFLOW Pof outputs P = ALL
•EMPTY PIPE(option ”empty tube identification)
•AUTO RANGE (automatic range change) Settin
(= ratio of upper to lower range, 1:20 to 1:1.25, value must be greater
than that of Fct. 1.03 “L.F.CUTOFF”, see also Sect. 5.18)
•TRIP POINT (define trip point) see also Sect. 5.17.
XXX – YYY
0 – 150% 0 – 150% N/O contact: XXX > YYY
Press ↵key to transfer to number setting, 1st digit (cursor) flashes.
Change flashing digit (cursor) with key
Press
↵
key to return to Fct. 1.06 and/or 1.07 STATUS B1 or B2.
• Characteristics of status outputsSwitch openSwitch closed
OFF (switched off)no function
ON (e.g. operation indicator)power OFFpower ON
SIGN I (F/R mode)Forward flowReverse flow
SIGN P (F/R mode)Forward flowReverse flow
TRIP POINT (limit switch)inactiveactive
AUTO RANGE
OVERFLOW I (I overranged)current output OKcurrent output overranged
OVERFLOW P (P overranged)pulse output OKpulse output overranged
ALL ERROR (all errors)errorsno error
FATAL.ERROR (fatal errors only)errorsno error
EMPTY PIPE (option)when measuring tubewhen measuring tube
(automatic range change)
ATUS B1 and/or B2
V/R mode
}
}
press → key
Dynamic response
of outputs, see Fct. 1.02, Sect. 5.2 “time constant”
}
g range 5-80 PERCENT
N/C contact: XXX < YYY
Hysteresis: difference between XXX and YYY.
↑
. Use key →to shift cursor 1 place to right.
high rangelow range
is emptyis full
Please refer to Sect. 3.2 ”factory settings“.
Connection diagrams, see Sect. 2.6.
5/7
Page 39
Control inputs B1 and B2 5.9
NOTE! Check whether under Fct. 3.07 “HARDWARE” the output terminal “B1” and/or “B2” is
defined as control input B1 and/or B2, refer also to Sect. 2.1 and Sect. 5.16.
Fct. 1.06 and 1.07 CONTROL
Press key →twice.
Select function of control inputs,
•OFF(switched off, no function)
•OUTP. HOLD(hold value of outputs)
•OUTP. ZERO(set outputs to “min. values”)
•TOTAL. RESET(reset totalizers)
•ERROR. RESET (delete/acknowledge error messages)
•EXT. RANGE(external range change for automatic range change, see also Sect. 5.19,
Press
↵
key to return to Fct. 1.06 or 1.07 CONTROL B1 or B2.
Please refer to Sect. 3.2 ”factory settings“.
Connection diagram, see Sect. 2.6.
Fct. 3.01 LANGUAGE
Press →key.
Select language for texts in display
•D(German)
•GB/USA (English)
•F(French)
•otherson request
settin
value must be greater than that of Fct. 1.03 L.F.CUTOFF)
Press ↵key to transfer to number setting, 1st digit (cursor) flashes.
Change flashing digit with key
press key → to shift cursor 1 place to the right.
B1/B2
press key↑ .
Functions also act on
}
display and totalizer
g range 5 – 80 PERCENT = ratio of low to high range 1:20 to 1:1.25,
↑
,
Language 5.10
Select using ↑key.
Press
↵
key to return to Fct. 3.01 LANGUAGE.
Entry code 5.11
Fct. 3.04 ENTRY CODE
Press →key.
Choice
•NO(no code, enter setting mode with → key)
•YES(enter setting mode with → key and Code 1: → → → ↵ ↵ ↵ ↑ ↑ ↑)
Select using ↑key.
Press
↵
key to return to Fct. 3.04 ENTRY CODE.
5/8
Page 40
5.12 Primary head
Fct. 3.02 FLOW METER
Press →key.
→ DIAMETER = set meter size (see instrument nameplate)
press → key
Select size from table of meter sizes:
•DN 2.5 - 1000 mm equivalent to
1
/10 - 40 inch
Select using ↑key.
Transfer to subfunction “FULL SCALE”
→ FULL SCALE = set full-scale range,
with ↵key.
press → key.
Set as described in Sect. 5.1.
Transfer to subfunction “GK VALUE” with ↵key.
Note: if “VALUE P” is displayed after pressing ↵ key.
PULSE/VOL. is set under Fct. 1.06 PULS B1, subfunction “SELECT P”. Because the full-scale
range Q
undershot:P
has been changed, the output frequency (F) of the pulse output is over- or
100%
min
= F
min
/ Q
100%
P
= F
max
max
/ Q
100%
Change pulse value accordingly, see Sect. 5.7 pulse output B1, Fct. 1.06.
→
GK VALUE = set primary constant GK,
press → key.
•1.0000 - 9.9999 (note information on instrument nameplate, do not change setting !)
Change flashing digit (cursor) with ↑key.
Shift cursor 1 place to right or left with
Transfer to subfunction “FIELD
FREQ.” with ↵key.
→ FIELD FREQ. = set magnetic field frequency,
→
key.
press → key
•1/2 • 1/6(1/2, 1/6, 1/18 and 1/36 of power frequency, see instrument nameplate,
•1/18 • 1/36do not change setting, exce
ptions see Sect. 6.4-6.6 !)
Select using ↑key.
Transfer to subfunction “FLOW DIR.”
with ↵key.
(only for units with DC power supply, transfer to subfunction “LINEFREQ”).
→ LINE FREQ. = normal line frequency in your country,
•50 Hz
•60 Hz
→ FLOW DIR. = set flow direction,
Select using the ↑key.
Transfer to subfunction “FLOW DIR.” with ↵ key.
press → key.
press → key.
•+ DIR.(for identification of flow direction, see “+” arrow on primary head;
•- DIR.for F/R mode, identifies the “positive” flow direction)
•Refer to Sect. 2.6 for electrical connection of outputs.
•Define direction of forward (normal) flow, see Fct. 3.02, subfunction “FLOW DIR.”:
in conjunction with F/R operation, set the direction for the forward flow here.
“+” signifies the same direction as shown by the arrow on the primary head,
“–” signifies the opposite direction.
•Set the status output to “SIGN I” or “SIGN P”, see Fct. 1.06 or 1.07, “STATUS B1 or B2”.
For dynamic response of the outputs with “SIGN I” or “SIGN P” see Sect. 5.8.
•Current and/or pulse output must be set to “2 DIR.”, see Fct. 1.05 and 1.06, subfunctions
“FUNCT. I” and “FUNCT. B1”.
5.15 Characteristic of outputs
ICurrent output
I
0%
I
100%
0 or 4 mA
20 mA
PPulse output
P
100%
Pulses at Q
, full-scale range
100%
Q
Q
Q
F
R
100%
1 flow direction, forward flow in F/R operation
Reverse flow in F/R operation
Full-scale range
SStatus output B1 or B2
switch open
switch closed
1 flow direction2 flow directions, F/R operation
I
I
max
I
100%
Q
R
%
Q
R
%
I
0%
0
P
P
115%
max
P
100%
P= 0 Hz
0
Q
F
%100 Max
Q
F
%100 Max
Q
%
Q
%
R
Max 100
R
Max 100
I
I
max
I
100%
I
0%
0
P
P
115%
max
P
100%
P= 0 Hz
0
S
Q
F
%100 Max
Q
F
%100 Max
5/11
Q
%
R
Max 100
Q
F
0
%100 Max
Page 43
Fct. 3.07 HARDWARE
Press →key.
Combinations of binary outputs and inputs 5.16
Define function of terminal B1,
•PULSOUTP.(= pulse output)
•STATUSOUTP.(= status output)
•CONTROLINP.(= control input)
Define function of terminal B2,
•STATUSOUTP.(status output)
•CONTROLINP.(control input)
Press key ↵to return to Fct. 3.07 HARDWARE.
Please note: If, for example, both output terminals (B1 and B2) are set to status output,
Fct. 1.06 or 1.07 status outputs B1 or B2
(Define operating mode of output terminals, see Sect. 5.16)
Press key → .
Set status output B1 or B2 to “TRIP POINT” by pressing key
↵
Press
Change flashing digit with key
•
Display
Setting ranges:
•
•Switchin
or to control input, their operating modes can ony be selected once.
Exam
ple: B1 and B2 are status outputs.
If status output B1 is used for the automatic range change BA, this operating mode
is not available for status output B2.
key to transfer to number setting, 1st digit (cursor) flashes.
XXX – YYY
XXX value = 0 – 150% of Q100%
YYY value = 0 – 150% of Q100%
Hysteresis ≥ 1% (= difference between XXX and YYY values)
g performance (N/O or N/C contact) can be set.
press key
press key
↑
, use key → to shift cursor 1 place to the right.
→
Select with key
press key
}
terminal B2.
→
Select with key
}
↑,
↵ to advance to
↑.
Limit switches 5.17
↑
(1 to 9 times).
N/O contact XXX value > YYY value
Contact closes
when flowrate greater than XXX value
Example: XXX = 55%
Note:When the two status outputs B1 and B2 are activated (see Sect. 5.16),
YYY = 45%
Hysteresis = 10%
YYY XXXXXX YYY
hysteresishysteresis
min. and max. values can, for example, be signalled.
The limit switches are active only in the case of forward flow.
N/C contact XXX value < YYY value
Contact opens
when flowrate greater than YYY value
Example: XXX = 45%
YYY = 55%
Hysteresis = 10%
5/12
Page 44
5.18Automatic range change BA
Automatic range change by means of status output
Fct. 1.06 or 1.07 status output B1 or B2
(Define operating mode of output terminals, see Sect. 5.16)
Press key →.
Set status output B1 or B2 to automatic range change “AUTO RANGE”
by pressing key
↑
(1 to 9 times).
Press key
Change flashing digit with key
Setting range:5 – 80 PERCENT of Q100% (= ratio between low and high range
↵
to transfer to number setting, 1st digit (cursor) flashes.
↑
, use key →to shift cursor 1 place to the right.
1 : 20 to 1 : 1.25)
Press key ↵to return to Fct. 1.06 or 1.07 status output B1 or B2.
External range change by means of control input
Fct. 1.06 or 1.07 control inputs B1 or B2
(Define operating mode of output terminals, see Sect. 5.16)
Press key →.
Set control inputs B1 or B2 to range change “EXT.RANGE” by pressing key
Press key
Change flashing digit with key
Setting range:5 – 80 PERCENT of Q100% (= ratio between low and high range
↵
to transfer to number setting, 1st digit (cursor) flashes.
↑
, use key →to shift cursor 1 place to the right.
1 : 20 to 1 : 1.25)
↑
(1 to 5 times).
Press key ↵, to return to Fct. 1.06 or 1.07 control input B1 or B2.
5/13
Page 45
Fct. 3.07 APPLICAT.
Press key → .
Applications 5.19
Set characterization for the flow,
•STEADY(flow is steady)
•PULSATING(pulsating flow, e.g. due to reciprocating pumps,
Press ↵key to return to Fct. 3.07 APPLICAT.
With built-in option “empty tube identification“ change to
subfunction “EMPTY PIPE“
→
EMPTY PIPE,
• YES • NO
Press
↵
key to return to Fct. 3.07 APPLICAT.
see also Sect. 6.4, 6.5 and 6.6
“Special-case applications”)
“function switch on“.
select with ↑ key.
press key → .
select function with ↑ key
}
5/14
Page 46
Part CSpecial applications, functional checks,
service, and order numbers
6Special applications
6.1 Use in hazardous areas
Electromagnetic flowmeters with IFC 090 signal converter are approved as electrical equipment in
conformity with the harmonized European Standards and Factory Mutual (FM).
Correspondence between the temperature classes and the temperature of the process liquid,
meter size and material of the measuring tube liner is defined in the test certificate.
Test certificate, certificate of conformity and wiring instructions are attached to the
Installation and Operating Instructions (applies only to hazardous-duty equipment).
Operator control can be carried out externally with MS-DOS PC via an RS 232 adapter
incl. CONFIG software
The Basic Version (IFC 090
can be operated with this option. Detailed instructions are supplied.
Always switch off power source before opening the housing !
1)Unscrew the cover from the electronic compartment using the special wrench.
2)If provided, remove the display unit. For this purpose, detach the two screws R and
fold display unit to the side, see illustration in Sect. 8.5.
3)Insert the RS 232 adapter (forming the connection to the PC or laptop)
into the IMoCom bus plug connector X2; for amplifier PCB, refer to Sect. 8.9.
4)Switch on the power.
5)As described in the supplied instructions, change data, parameters and measured
values or have them called up for display.
6)Switch off the power.
7)Pull off the RS 232 adapter from the amplifier PCB.
8)Secure the display unit with screws R.
9)Replace and tighten down the cover on the electronic compartment using the
special wrench.
NOTE: The screw thread and gasket of the housing cover should be well greased
at all times, always check for signs of damage and never allow dirt to accumulate.
Replace defective gasket immediately.
/ B) and the Display Version (IFC 090 / D) of the signal converter
Please refer to Sect. 3.2 “Factory settings”
6/1
Page 47
Stable signal outputs when measuring tube empty 6.3
Output signals can be stabilized to values as for “zero” flow to prevent random output signals
when the measuring tube is empty or when the electrodes are not wetted in the event the
measuring tube is partially full.
- Display:0
- Current output:0 or 4 mA, see setting in Fct. 1.05
- Pulse output:no pulses (= 0 Hz), see setting in Fct. 1.06
Precondition:
Chan
ges on amplifier PCB, see illustration in Sect. 8.9.
electrical conductivity of product: ≥ 200 µS/cm,
≥ 500 µS/cm for meter sizes DN 2.5 - 15 and
1
/10”-1/2”
Switch off power source before opening the housing!
1)Unscrew cover from the terminal compartment using the special wrench. Pull off the
two plugs for supply power (3-pin) and outputs/inputs (6-pin).
2)Unscrew cover from the electronic compartment using the special wrench.
3)If provided, remove display unit. For this purpose, detach the two screws R and fold
display unit to the side, see illustration in Sect. 8.5.
4)Carefully pull off the blue 9-pin plug X1/X4 (forming the connection to the primary head).
5)Detach the 2 recessed head screws Q and carefully remove the electronic unit.
6)Join the two “semicircles” of points S1 and S3 on the amplifier board with tin solder,
see illustration in Sect. 8.9.
7)Reassemble in reverse order, points 5) - 2) above.
8)Switch on power.
9)Check setting of the low-flow cutoff SMU, Fct. 1.03, and reset if necessary:
Display Version: IFC 090 / D, see Sect. 4 and 5.3, Fct. 1.03
Basic Version: IFC 090
/ B, see Sect. 6.2.
10) After checking and/or resetting, replace the cover on the electronic compartment and
tighten using the special wrench.
NOTE: The screw thread and gasket of the housing cover should be well greased at
all times, always check for signs of damage and never allow dirt to accumulate.
Replace defective gasket immediately.
6/2
Page 48
6.4 Pulsating flow
Application
downstream of positive-displacement pumps (reciprocating or diaphragm pumps) without
pulsation dampener
Operator control of the signal converter for the new settings
IFC 090 B (basic version)see Sect. 6.2
IFC 090 D (display version) see Sect. 4 and 5
To change settings
• Fct. 3.02 FIELD FREQ.
(change magnetic field frequency)
– Stroke frequency less than 80 strokes/min (at max. pump lift): do not change setting.
– Stroke frequency 80 - 200 strokes/min (at max. pump lift): change setting to 1/2,
only practical with IFM 5080 K and IFS 5000 F (DN 2.5-100 /
and IFM 4080 K and F IFS 4000 F (DN 10, 15, 50-100 /
1
/10”-4”)
1
/10”, 1/2”, 2”-4”),
Please consult factory where other types and meter sizes are concerned.
– Please note: given stroke frequencies close to the tripping point of 80 strokes/min, additional
measuring errors of approx. ± 0.5% of the measured value may occasionally occur.
• Fct. 3.06 APPLICA
T. (adjust overload point of the A/D converter to the application)
Change setting to “PULSATING”.
• Fct. 1.04 DISP
. FLOW (change display presentation of flow)
Change setting to “BARGRAPH” to allow better assessment of display unsteadiness.
• Fct. 1.02 TIMECONST
. (change time constant)
– Set to “ALL” and time (t) in seconds.
– Recommended:t [s] =
1000
min. strokes / min
– Example:min. number of strokes in operation = 50 strokes/minute
t [s] == 20 s
1000
50 / min
With this setting, the residual ripple of the display will amount to
approx. ± 2% of the measured value. Doubling the time constant
will reduce the residual ripple by a factor of 2.
6.5 Rapid changes in flowrate
Application
in conjunction with batching processes, fast-response control loops, etc.
Operator control of the signal converter for the new settings
IFC 090 B (basic version)see Sect. 6.2
IFC 090 D (display version) see Sect. 4 and 5
To change settings
• Fct. 1.02 TIMECONST
. (change time constant)
Setting to “ONLYI” and set time to 0.2 s.
•D
ynamic response with meter sizes DN 2.5-300 / 1/10”-12”
Dead time:approx. 0.06 s at 50 Hz line frequency
approx. 0.05 s at 60 Hz line frequency
Time constant: set as above, current output (mA) additionally plus 0.1 s
• Reducin
Fct. 3.02 FLOW METER, subfunction “FIELD FREQ.”, change to “1/2”,
only practical with IFM 5080 K and IFS 5000 F (DN 2.5-100 /
and IFM 4080 K and IFS 4000 F (DN 10, 15, 50-100 /
g the dead time by a factor of 3 (possible by changing the magnetic field frequency)
1
1
/10”-4”)
/10”, 1/2”, 2”-4”)
Please consult factory where other types and meter sizes are concerned.
6/3
Page 49
Unsteady display and outputs 6.6
Unsteady display and outputs can occur in connection with
– high solids contents,
– non-homogeneity,
– poor blending, or
– chemical reactions still in progress in the process liquid.
If, in addition, flow is also pulsating due to the use of diaphragm or reciprocating pumps,
refer to Sect. 6.4.
Operator control of the signal converter for the new settings
IFC 090 B (basic version)see Sect. 6.2
IFC 090 D (display version) see Sect. 4 and 5
To change settings
• Fct. 1.04 DISP
. FLOW (change display presentation of the flow)
Change setting to “BARGRAPH” to allow better assessment of display unsteadiness.
• Fct. 1.02 TIMECONST
. (change time constant)
– Setting to “ONLY I”, to “ALL” if pulse output too unsteady.
– Set time constant to approx. “20 s”, observe unsteadiness of display and adjust time
constant if necessary.
• Fct. 3.06 APPLICA
T. (adjust overload point of the A/D converter to the application)
Set to “PULSATING” on trial basis, if unsuccessful return to “STEADY”.
• Fct. 3.02 FIELD
FREQ. (change magnetic field frequency)
On trial basis, change setting to “1/2”; if unsuccessful return to previous setting,
usually “1/6”.
Only practical with IFM 5080 K and IFS 5000 F (DN 2.5-100 /
and IFM 4080 K and IFS 4000 F (DN 10, 15, 50-100 /
1
1
/10”-4”)
/10”, 1/2”, 2”-4”),
Consult factory where other types and meter sizes are concerned.
6/4
Page 50
6.7 HART®for IFC 090
HART®interface
The HART interface is a smart interface, in other words a communication signal superimposed on
the current output. All functions and parameters can be accessed via this interface.
The following HART features are supported by the IFC 090 signal converter:
• point-to-point connection
• multidrop (up to 15 HART devices)
The burst mode is not normally used and is therefore not supported. Further information about
HART is available from the HART Communication Foundation, of which Krohne is a member.
Electrical connection
HART connection active
to HART communicator
or smart converter
HART connection passive
4 mA
U
H
to next
HART device
power supply unit
to HART communicator
or smart converter
Factory setting for the point-to-point mode with HART
• In Menu 1.05, the “FUNCT.I” parameter must be set to “1 DIR”. or “2 DIR.”.
• IMPORTANT: In Menu 1.05, the “RANGE I” parameter must be set to “4-20 mA”, or, in the case
of variable setting, the value for “I
” must be at least 4 mA.
0%
• In Menu 3.09, the “COM1” parameter must be set to “HART” and the “ADDRESS” parameter
to “0”.
• The current output may be operated in the active or passive mode.
Factor
y setting for the multidrop mode with HART
• In Menu 1.05, the “FUNCT.I” parameter must be set to “OFF”.
• In Menu 3.09, the “COM1“ parameter must be set to “HART” and the “ADDRESS” parameter to
a value of “1-15”. This address may only be set for one device in the HART multidrop network.
• IMPORTANT: the current output may only be operated in the passive mode.
6/5
Page 51
Minimum load im
pedance
A minimum load impedance of 250 Ωis required to enable the HART signals to be modulated on
the current output. Series-connect an appropriate resistor if the devices located in the current
output circuit should not attain this value. It will then be possible via the minimum load impedance
e.g. to connect the HART communicator or smart converter in parallel without interrupting the
current output. Please note that the load impedance should not exceed a maximum of 500
.
Ω
HART o
perating tools / DD
The IFC 090 can be operated either via its local operator interface or by means of the HART
communicator or the CONFIG program. Both operating tools are available from KROHNE.
Operator control by means of the HART communicator requires a device description (DD) which
we can load for you into the communicator. We can, of course, also load the DDs of all
manufacturers who have filed their DDs with the HART Communication Foundation. If you wish to
use the IFC 090 in your operating tool, for example, please ask for the description of the HART
command used so that you can address the complete IFC 090 functionality via HART.
In the near future we shall also be supporting the ASM tool from Rosemount and the SIPROM tool
from Siemens.
Power supply units / isolation switchin
g amplifiers
You will need an appropriate power supply unit if the current output is to operate in the passive
mode. Please ensure that this power supply unit is also suitable for HART communication.
This applies equally to isolation switching amplifiers that are sometimes used in the active mode.
Additional functions for HART
Fct.TextDescription and settings
3.08LOCATIONSet any tag name (max.10 characters)
3.09COMHART communications-interface
version
Characters assi
• A - Z, a - z, 0 - 9, or “-” (= blank character)
Press ↵key to transfer to
• OFF (switched off)
• HART (switched on)
Press ↵key,
set “ADDRESS”
Range: 001 - 015
Press ↵key to transfer to
gnable to each place:
Fct. 3.08 “LOCATION”.
Select with ↑key.
}
with ↑and → keys,
Fct. 3.09 “COM”.
6/6
Page 52
7. Functional checks
7.1 Zero check with IFC 090 / D signal converter, Fct. 3.03
Switch off power source before opening the housing
•Set “zero” flow in the pipeline, but make sure that the measuring tube is completely
filled with fluid.
•Switch on the system and wait 15 minutes.
•Press the following keys for zero measurement:
KeyDisplayDescription
→If “YES” set under Fct. 3.04 ENTRY CODE, key in
Fct. 1.00OPERATION
2x ↑Fct. 3.00INSTALL.
→Fct. 3.01LANGUAGE
2x ↑Fct. 3.03ZERO SET
→CALIB. NO
↑CALIB. YES
↵0.00- - - - - / - - -Flowrate displayed in set unit, see Fct. 1.04 DISPLAY,
STORE NOIf new value not to be stored, press ↵ key (3x) 4x = return to
↑STORE YES
↵Fct. 3.03ZERO SETStore new zero value.
subfunction “DISP. FLOW”.
Zero measurement in progress, duration approx. 50 seconds.
When flow “> 0” “WARNING” notice appears, confirm with ↵ key.
measuring mode.
7.2 Test of measuring range Q, Fct. 2.01
Switch off power source before opening the housing !
•For this test a measured value can be simulated in the range of -110 to +110 percent
of Q
(full-scale range set, see Fct. 1.01 FULL SCALE).
100%
•Switch on the system.
•Press the following keys for this test:
KeyDisplayDescription
→If “YES” set under Fct. 3.04 ENTRY CODE, key in
Fct. 1.00OPERATION
↑Fct. 2.00TEST
→Fct. 2.01TEST Q
→SURE NO
↑SURE YES
↵0PERCENTCurrent, pulse and status indication outputs indicate the
± 10PERCENT
↑ ± 50PERCENTSelect using ↑ key
± 100PERCENT
± 110PERCENT
↵Fct. 2.01TEST QEnd of test, actual measured values again present at outputs.
(2x) 3x ↵- - - - - - -- - - - - / - - -Measuring mode
9-keystroke CODE 1 now: → → → ↵ ↵ ↵ ↑ ↑ ↑
corresponding values.
7/1
Page 53
Hardware information and error status, Fct. 2.02 7.3
Switch off power source before opening the housing !
•Before consulting factory about errors or flow measurement problems, please invoke
Fct. 2.02 HARDW. INFO (hardware information).
•An 8-character and a 10-character status code are stored under this function in
each of 3 “windows”. These 6 status codes allow rapid and simple diagnosis of your
compact flowmeter.
•Switch on system.
•Press the following keys for display of the status codes:
KeyDisplayDescription
→If “YES” set under Fct. 3.04 ENTRY CODE,
Fct. 1.00OPERATION
↑Fct. 2.00TEST
→Fct. 2.01TEST Q
↑Fct. 2.02HARDW. INFO
key in the 9-keystroke CODE 1 now: → → → ↵ ↵ ↵ ↑ ↑ ↑
ple status code
If you need to return your flowmeter to Krohne, please refer to last
but one page of these Instructions !
7/2
Page 54
7.4 Faults and symptoms during start-up and process flow measurement
•Most faults and symptoms occuring with the flowmeters can be eliminated
with the aid of the following tables.
•For greater clarity, faults and symptoms in the tables are divided into 2 parts and
various groups.
•Part 1
Signal converter IFC 090 B (B = basic version), without display
and without HHT or CONFIG user program (see Sect. 6.2)
Groups:LEDLED display (status messages)
ICurrent output
PPulse output
LED / I / P LED display, current output and pulse output
•Part 2
Signal converter IFC 090 D (D = display version) and
Signal converter IFC 090 B (B = basic version), without display
but with CONFIG user program (see Sect. 6.2)
Groups:DDisplay
ICurrent output
PPulse output
SStatus indication output
CControl input
D / I / P / S LED display, current output, pulse output,
status output and display
Before contacting Krohne Service, please read through the
following tables. THANK YOU !
Part 1Converter IFC 090 B (B = basic version), without display
Group LEDFault / SymptomCauseRemedial action
LED 1LED flashes red/greenOverranging of A/D converter,
LED 2LED flashes redFatal Error, hardware and/or Replace signal converter
LED 3cyclic flashing of red LED, Hardware fault, WatchdogReplace signal converter
LED 4LED shows red continuously Hardware faultReplace signal converter
Group I
I 1Receiver instrumentConnection/polarity incorrect.Connect properly as described
I 222 mA present at currentCurrent output I overrangedCheck device parameters
and without HHT or CONFIG operator program
Reduce flowrate; if unsuccessful,
current or pulse outputtest as described in Sect. 7.5.
Measuring tube drained,Fill measuring tube.
A/D converter overranged.
software error(see Sect. 8.7) or contact
Krohne Service.
approx 1 secondactivated.(see Sect. 8.7) or contact
Krohne Service.
(see Sect. 8.7) or contact
Krohne Service.
Fault / SymptomCauseRemedial action
indicates “0”.in Sect. 2.3 + 2.7
Receiver instrumentCheck connecting cables and
defective.receiver instrument, and
replace if necessary.
Short between current output Check connections and
and pulse output cables, see Sect. 2.3 + 2.7,
voltage between I+and
I⊥ approx. 15 V. Switch off
device, eliminate short-circuit,
switch device on.
Current output defectiveReplace signal converter
(see Sect. 8.7) or contact
Krohne Service.
output (fault current)and change if necessary,
see Sect. 6.2, or contact
Krohne Service
7/3
Page 55
Part 1Converter IFC 090 B (B = basic version), without display
(cont’d)and withoutHHT or CONFIG operator program
Group IFault / SymptomCauseRemedial action
I 322 mA present at currentFatal ErrorReplace signal converter
I 4Unsteady displayElectrical conductivity of Increase time constant
I 5Receiver instruments Control input C set toChange setting
I 6“Skipping” elec.Current output set toChange hysteresis or range
I 7F/R flow measurement:Different ranges set forChange setting, see Sect. 6.2
I 8Receiver instrumentsControl input C set toChange setting,
Group P
P 1Connected totalizer not Connection/polarity incorrectConnect properly as described
P 2Unsteady pulse rateElectrical conductivity of Increase time constant, see
Group LED / I / P
LED / I / P 1Red LED flashes, current output Fatal Error, hardware fault Replace signal converter
output (fault current)(see Sect. 8.7) or contact
and red LED shows.Krohne Service
product too low(see Sect. 6.2) or contact
indicate “constant value”“hold outputs”(see Sect. 6.2) or contact
current values“automatic range change”of tripping points,
Control input C set toSwitch off or check level
“external range change”see Sect. 6.2 or
different displays even“forward and reverse flow”or contact Krohne Service
though same volume flow
rate in both directions
indicate “min. values”“set outputs to zero”see Sect. 6.2 or contact
Fault / SymptomCauseRemedial action
counting any pulsesin Sect. 2.4 + 2.7.
Totalizer or external Check connection cables,
voltage source defectivetotalizer and external voltage
Current output is externalCheck connection and cables
voltage source; short-circuit,(see Sect. 2.4 + 2.7), voltage
or current/pulse output between I+and I⊥ approx.15V.
defective.Switch off device. Eliminate
Control input CChange setting,
is set to “hold outputs”see Sect. 6.2 or contact
Pulse output inactive, Switch on, see Sect. 6.2, or
see Fct. 1.06.
Fatal Error, red LED shows.Replace signal converter
Output B1 set toChange setting,
status output orsee Sect. 6.2 or contact
control inputKrohne Service
Control input C is set toChange setting,
“set outputs to zero”see Sect. 6.2 or contact
and currently activeKrohne Service
product too low.Sect. 6.2 or contact Krohne
Fault / SymptomCauseRemedial action
indicates fault current and pulse and/or software error(see Sect. 8.7) or contact
output “0”.Krohne Service.
Krohne Service
Krohne Service
see Sect. 6.2 or contact
Krohne Service
contact Krohne Service
Krohne Service
source, and replace if necessary.
short-circuit and switch device
on again. If no function then
current or pulse output
defective. Replace signal
converter (see Sect. 8.7) or
contact Krohne Service.
Krohne Service
(see Sect. 8.7) or contact
Krohne Service.
Service.
7/4
Page 56
Part 2Signal converter IFC 090 D (D = display version) and
Group DDisplay shows . . .CausRemedial action
D 1LINE INT.Power failure.Delete error message in
D 2CUR.OUTP. ICurrent output overranged.Check instrument parameters
D 3PULS.OUTP. PPulse output overranged.Check instrument parameters,
D 4ADCAnalog/digital converterError message deleted
D 5FATAL. ERRORFatal Error,Replace signal converter
D 6TOTALIZERCounts lostDelete error message in
D 7„STARTUP“Hardware fault,Replace signal converter
D 8BUSYDisplays for flow,Change setting in
D 9Unsteady displayLow electricalIncrease time constant in
D 10No displayPower OFF.Switch on power
Signal converter IFC 090 B (B = basic version), without display
but with HHT or CONFIG operator program (see Sect. 6.2)
no counting duringRESET/QUIT. menu. Reset
Note:
power failure.totalizer(s) if need be.
and correct if necessary.
Error message deleted
automatically after cause
has been eliminated.
totalizer deviationcorrect if necessary, and reset
Note:
possibletotalizer(s). Error message
overranged.automatically after cause
all outputs set to “min.”(see Sect. 8.7) or contact
values.Krohne Service, having
(overflow, data error)RESET/QUIT. menu.
cyclic flashingWatchdog activated.(see Sect. 8.7) or contact
totalizers and errorsFct. 1.04.
disabled.
conductivity, high solidsFct. 1.02.
content, pulsating flow.
Check power (fuse(s)Replace if defective
F1 (F1 +F2 with DC).(see Sect. 8.1).
deleted automatically
after cause has been
eliminated.
has been eliminated.
first noted down hardware
information and error status,
see Fct. 2.02.
Krohne Service.
7/5
Page 57
Part 2Signal converter IFC 090 D (D = display version) and
Group IFault / SymptomCauseRemedial action
I 1Receiver instrumentIncorrect connection/polarityConnect properly
I 2Unsteady displayLow electrical conductivity,Increase time constant,
Signal converter IFC 090 B (B = basic version), without display
but with HHT or CONFIG operator program (see Sect. 6.2)
indicates “0”.see Sect. 2.3 + 2.7.
Receiver instrument or currentCheck output (see Sect. 7.2)
output defective.with new milliammeter:
Current output disabled,Activate under Fct. 1.05.
see Fct. 1.05
Short-circuit between currentCheck connection and cables,
output and pulse output.see Sect. 2.3 + 2.7.
high solids content,see Fct. 1.02, or contact
pulsating flow.Krohne Service.
est ok, check connection
T
cables and receiver instrument,
replace if necessary.
est faulty, current output
T
defective. Replace signal
converter (see Sect. 8.7)
or contact Krohne Service
Voltage between I+ and I⊥
approx. 15 V. Switch off device,
eliminate short-circuit,
and switch device on again.
7/6
Page 58
Part 2Signal converter IFC 090 D (D = display version) and
(cont’d)Signal converter IFC 090 B (B = basic version), without display
Group P
P 1Totalizer connected but doesIncorrect connection/polarityConnect properly,
P 2Unsteady pulse rateElectrical conductivity ofIncrease time constant under
P 3Pulse rate too high orIncorrect setting forChange setting under Fct. 1.06.
but with HHT or CONFIG operator program (see Sect. 6.2)
Fault / SymptomCauseRemedial action
not count any pulsessee Sect. 2.4 + 2.7.
Totalizer or external voltageCheck output (see Sect. 7.2)
source defective.with new totalizer:
Current output is externalCheck connection and cables,
voltage source, short circuitsee Sect. 2.3, 2.4 + 2.7.
or current / pulse outputVoltage between I+and I⊥
defectiveapprox. 15 V. Switch off
Pulse output is deactivated,Switch on under Fct. 1.06.
see Fct. 1.06.
product too low,Fct. 1.02, or switch on.
time constant too low or
switched off for pulse
output.
too low.pulse output.
est ok, check connection
T
cables and previous
totalizers and external voltage
source, and replace if
necessary.
est faulty, pulse output
T
defective, replace signal
converter (see Sect. 8.7)
or contact Krohne Service.
device, eliminate short circuit,
switch device on again. If no
function, then current or pulse
output defective. Replace
signal converter (see Sect.
8.7) or contact Krohne
Service.
7/7
Page 59
Part 2Signal converter IFC 090 D (D = display version) and
(cont’d)Signal converter IFC 090 B (B = basic version), without display
Group SFault / SymptomCauseRemedial action
S 1No functionIncorrect connection/polarityConnect properly,
Group D/I/P/SFault / SymptomCausRemedial action
D/I/P/S1Unsteady display andElectrical conductivityIncrease time constant under
D/I/P/S 2No display and noPower OFFSwitch power on.
Group CFault / SymptomCauseRemedial action
C 1No functionConnection is incorrectConnect properly,
but with HHT or CONFIG operator program (see Sect. 6.2)
of status displaysee Sect. 2.5 + 2.7.
Status display or output
defective or externalto “F/R INDIC.” (flow direction)
voltage source notand check (see Sect. 7.2)
supplying voltage.with new status display:
Output terminal B1 or B2Change setting under
not set to “status output”Fct. 3.07
outputsof product too low,Fct. 1.02.
function of outputs
time constant too low.
Check power fuse(s)Replace if defective,
F1 (F1 + F2 for DC).see Sect. 8.1.
Control input CCheck connection, cables
or external voltageand external voltage source,
source defectivesee Sect. 2.6 + 2.7
Output terminalChange setting under
B1 or B2 not set toFct. 3.07.
“control input”
Set status output under Fct. 1.07
est ok, check previous
T
status display and external
voltage source, and replace
if necessary.
est faulty, status output
T
defective, replace signal
converter (see Sect. 8.7) or
contact Krohne Service.
see Sect. 2.6 + 2.7
7/8
Page 60
7.5 Test of primary head
Always switch off power source before opening the housing !
Re
quired measuring instruments and tools
-Special wrench to unscrew the covers, a crosstip screwdriver
-Ohmmeter with at least 6 V range
or
AC voltage/resistance measuring bridge.
Note: Accurate measurements in the area of the electrodes can only be obtained with an
AC voltage/resistance bridge. Also, the measured resistance is very heavily dependent on
the electrical conductivity of the liquid product.
Preparatory work
•Switch off power source.
Blue 9-pin plug
(connection to primary head)
•Unscrew covers from terminal and electronic
compartments using the special wrench.
7 8 1 2 3 4 5
unassigned
If provided, remove the display unit. For this
purpose, detach the two screws R and fold
display unit to the side, see figure in Sect. 8.5.
•Pull off the blue 9-pin plug from the amplifier
PCB, see figure in Sect. 8.9, field power
supply (Pin 7+8) and signal cables
X1
X4
(Pin 1, 2, 3, 4 + 5)
•Completely fill the measuring tube of the
flowmeter with the process liquid.
ActionTypical resultIncorrect result =
Resistance measurementflowmeter defective,
at the blue 9-pin plugreturn to factory for repair,
1Measure resistance30 - 150 Ωif lower:
between wiresinterwinding fault
7 and 8.
2Measure > 10 MΩif lower:
resistance between interwinding fault to PE
U-clamp terminal or FE.
in terminal compartment
(PE protective conductor
or FE functional ground)
and wires 7 and 8
between wires 1 and 3(see “Note” above).tube and repeat
and between 1 and 4Both values should bemeasurement; if still
(same measuring leadapprox. equal.too low, short-circuit in
always on wire 1 ! )electrode wires.
Jacks X1 and X4 on amplifier PCB,
see Sect. 8.9
refer to last but one page !
if higher:
wire break
if higher: break in
electrode wires or
electrodes
contaminated.
Values not equal:
break in electrode
wires or electrodes
contaminated.
7/9
Page 61
Test of signal converter using GS 8A simulator (option) 7.6
+
0
–
GS 8A operating elements and accessories
Y
20
20
10
40
80
5
2,5
160
1,25
320
Electrical connection
H2
BPlug for field power
C5Plug for signal cable,
P
D
L
H
DSwitch,
HSocket for H2 plug
H2Plug of cable Z
LPower supply ON
PPotentiometer “zero”
X1Socket for plug B
X4Socket for plug C5
YSwitch,
ZCable between GS 8A
supply, 2-pin
5-pin
flow direction
on cable Z
on amplifier PCB
on amplifier PCB
measuring ranges
and signal converter
Z
BB⊥ B2 I+II⊥
}
binary outputs
and inputs
C 5
X 4
amplifier PCB, see
Sect. 8.9
a) Switch off power source before
opening the housing !
b) Unscrew the cover from electronic section
using the special wrench.
c) Remove screws R and fold display unit
to side, see illustration in Sect. 8.5.
B
X 1
Connection of milliammeter and
electronic frequency counter
Using GS 8 simulator
Additional adapter is required
between GS 8 and IFC 090 signal converter.
Outputs/Inputs
see Sect. 2.6.
Order No. 2.10764.00.
PowerU-clamp terminal
}
current
output
L N
10
1L~ 0L
A
d) Pull off blue 9-pin plug (X1/X4) from the
amplifier PCB, see Sect. 8.9:
socket X1 field power supply, and socket X4
signal cables.
e) Connect plug B to socket X1 (2-pin) and
plug C (5-pin) to socket X4 (5-pin).
(PE)
100-240 V AC
~
(PE protective conductor)
(FE functional ground)
Milliammeter, accuracy
class 0.1, Ri< 500 Ω,
range 4-20 mA
Electronic frequency
counter, input resistance
approx. 1 kΩ, range
0-1 kHz, time base min. 1
second, see connection
diagrams in Sect. 2.6.
(FE)
24 V AC/DC
Warning: Instrument must be properly grounded to avoid personnel shock hazard.
7/10
Page 62
Check of set
point display
1)Switch on power source, allow at least 15 minutes’ warm-up time..
2)Set switch D (front panel of GS 8A) to “0” position.
3)Adjust zero to 0 or 4 mA with the 10-turn potentiometer P (front panel of GS 8A),
depending on setting in Fct. 1.05, deviation <±10 µA.
4)Calculate position of switch Y and displayed setpoints ”I“ and ”f“:
4.1)X =
Q
100%
Q
K
100%
2
x
GK
DN
full-scale range (100%) in volumetric unit V per unit time t
x
GKprimary constant, see instrument nameplate
DNmeter size DN im mm, not inches, see instrument nameplate
ttime in seconds (Sec), minutes (min) or hours (hr)
Vvolumetric unit
Kconstant, according to following table
Sticker on the GS 8 primary head simulator still gives values for
“inch” flowmeter. Do not use any more !
4.2)Determine position of switch Y: Use table (front panel GS 8A) to determine value Y
which comes closest to factor X and meets condition Y ≤ X.
4.3)Calculate set
4.4)Calculate set
point reading ”I“ for current output:I = I0%+ Y(I
I
0%
I
100%
point reading ”f“ for pulse output:f =
P
X
current (0/4mA) at 0% flowrate
current (20mA) at 100% flowrate
Y
x
P
100%
X
pulses per second (Hz)
100%
100%
in Hz
- I0%) in mA
at 100% flowrate
5)Set switch D (front panel GS 8A) to position “+“ or “-“ (forward/reverse flow).
6)Set switch Y (front panel GS 8A) to the value determined by the method described above.
7)Check setpoint readings I and f, see points 4.3 and 4.4 above.
8)Deviation <1.5% of setpoint. If greater, replace signal converter, see Sect. 8.7.
9)Test of linearity: set lower Y values, readings will drop in proportion to the calculated Y
values.
10)Switch off power source after completing the test.
11)Disconnect the GS 8A.
12)Reassemble in reverse order, see points e) – b) under “electrical connection”,
see illustration in Sect. 8.5.
13)The system is ready for operation after the power source has been switched on.
Exam
ple: see overleaf
7/11
Page 63
Example
Full-scale rangeQ
= 200 m3/hr (Fct. 1.01)
100%
Meter sizeDN= 80 mm = 3“ (Fct. 3.02)
Current at Q
Pulses atQ
Q
0%
100%
100%
I
0%
I
100%
P
100%
= 4 mA
= 20 mA
(Fct. 1.05)
}
= 280 pulses/hr (Fct. 1.06)
Primary head constantGK= 3.571 (see nameplate)
Constant(Vin m3)
(tin hr)K= 7074 (see Table)
(DN in mm)
Calculation of “ X ”
Q
x K
100%
X =
GK x DN
2
and position of “ Y ”
200 x 7074
=
3.571 x 80 x 80
= 61.905
Y = 80, position of switch Y, see front panel GS 8A
(comes closest to X value and is smaller than X).
Calculation of set
I = I0%+Y(I
X 61.905
point readings I and f
- I0%) = 4 mA +
100%
40
(20mA - 4mA) = 14.3 mA
Deviations are permissible between 14.1 and 14.6 mA
f = Yx P
100%
X 61.905
=
40
x pulses / hr = 129.2 pulses/hr
Deviations are permissible between 127.3 and 131.1
If you need to return your flowmeter to Krohne,
please refer to last but one page of these Instructions !
(equivalent to ± 1.5 %).
pulses/hr (equivalent to ± 1.5 %).
7/12
Page 64
8.Service
8.1 Replacement of power fuses
A) Fuse F1 in AC Versions 1 and 2
Switch off power source before opening the housing !
1)Unscrew cover of electronics compartment using the special wrench.
2)If provided, remove the display unit.
3)Take out old and insert new power fuse F1.
4)Reassemble in reverse order, items 2) - 1) above.
For this purpose, detach the two screws R and fold display unit to side.
Refer to the table in Sect. 8.5 for fuse rating and order number.
B) Fuses F1 and F2 in the AC/DC V
Switch off power source before opening the housing !
1)Unscrew cover from the terminal compartment using the special wrench.
Pull off the two plugs for power (3-pin) and outputs/inputs (5-pin)
2)Unscrew cover from the electronics compartment using the special wrench.
3)If provided, remove display unit.For this purpose, detach the two screws R and fold display unit to the side.
4)Carefully pull off the blue 9-pin plug X1/X4
(forming the connection to the primary head).
5)Detach the 2 recessed head screws Q and carefully remove the electronic unit.
6)Replace power fuses F1 and F2 on the power supply PCB,
refer to Sect. 8.9 for illustration of the PCB. Refer to the table in Sect. 8.5 for fuse
ratings and order numbers.
7)Reassemble in reverse order, points 5) - 1) above.
8.2 Changeover of operating voltage on AC Versions 1 and 2
Switch off power source before opening the housing !
1)Unscrew the cover from the terminal compartment using the special wrench.
Pull off the two plugs for power (3-pin) and outputs/inputs (5-pin).
2)Unscrew the cover from the electronics compartment using the special wrench.
3)If provided, remove display unit.
For this purpose, detach the two screws R and fold display unit to the side.
4)Carefully pull off the blue 9-pin plug X1/X4
(forming the connection to the primary head).
5)Detach the 2 recessed head screws Q and carefully remove the electronic unit.
6)Transpose the voltage selector SW on the power supply PCB (see diagram in Sect. 8.9)
to obtain the required voltage according to the table in Sect. 8.5.
7)Change power fuse F1, refer to table in Sect. 8.5 for fuse rating.
8)Reassemble in reverse order, points 5) - 1) above.
ersion
.
8.3 Turning the display PCB
Switch off power source before opening the housing !
1)Unscrew cover from electronic compartment using the special wrench.
2)Detach the two screws R and carefully turn the display unit ± 90° or 180°.
3)If display turned ± 90°, reposition the screws R on the display board.
4)Reassemble in reverse order, points 2) - 1) above.
Warning: Instrument must be properly grounded to avoid personnel shock hazard.
8/1
Page 65
Retrofitting the display unit 8.4
Switch off power source before opening the housing !
1)Unscrew the cover from the electronic compartment using the special wrench.
2)Insert plug for display unit into socket X6 on the amplifier PCB,
see diagrams in Sect. 8.5 and 8.9.
3)Secure the plug by means of the supplied metal clip to prevent it from dropping out.
4)Tighten down the display PCB with screws R.
5)Switch on power.
6)Refer to Sect. 4 and 5 for operator control and display of measured values.
7)Grease the thread and gasket of the new housing cover, with recess for the display,
and tighten down using the special wrench.
Power fuses and illustrations to Sect. 8.1 to 8.4 8.5
The screw threads and gaskets of both
IMPORTANT !
F1
housing covers should be well greased at
all times, always check for signs of damage
LP-A
X4
X1
L
X6
X2
C
LP-P
Q
R
and never allow dirt to accumulate.
Replace defective gaskets immediately.
CRibbon cable, display unit
LStatus LED
LP-AAmplifier board, see Sect. 8.9
LP-PPower supply board, see Sect. 8.9
QFastening screws, electronic unit
RFastening screws, display unit
X12-pin connector, field power
X25-pin connector, IMoCom bus
X45-pin connector, electrode signals
X610-pin connector, display unit
Warning: Instrument must be properly grounded to avoid personnel shock hazard.
Power supplyVoltageFuses F1 (and F2)Location and position
RatingOrder No.of voltage selector SW
1. AC Version230/240 V AC125 mA T5.06627
115/117 V AC200 mA T5.05678
F 1
F 1
2. AC Version200 V AC125 mA T5.06627
100 V AC200 mA T5.05678
AC/DC Version 24 V AC/DCF1 + F2
1.25 A T5.09080
F 1
F 1
8/2
Page 66
8.6 Turning the converter housing of the compact flowmeters
To facilitate access to connecting, indicating and operating elements for flowmeters installed in
locations that are hard to get at, the signal converter housing can be turned through ± 90°.
Not allowed for flowmeters of hazardous-duty (Ex) design!
A
vailable versions of flowmeters with IFC 090 K signal converter
A
D
Turning the converter housing
Any faults resulting from failure to follow these instructions
scrupulously shall not be covered by our warranty !
Switch off power source before opening the housing !
1)Clamp the flowmeter firmly by the primary head housing.
2)Secure converter housing against slipping and tilting.
3)Remove the 2 hexagon socket screws connecting the two housings, and push
out the 2 plugs.
4)Carefully turn the converter housing clockwise or anti-clockwise a maximum of 90°,
but do not lift the housing. If the gasket should stick, do not attempt to lever it off.
5)To conform to the requirements of protection category IP 67, equivalent to NEMA 6,
keep connecting faces clean and tighten the 2 hexagon socket screws uniformly.
Replace the plugs in the two free holes.
B
C
E
8.7 IFC 090 Replacement of converter electronic unit
A special electronic unit is available for hazardous-duty flowmeters, see separate “Ex”
installation instructions.
Switch off power source before opening the housing !
1)Unscrew the cover from the terminal compartment using the special wrench.
2)Unscrew the cover from the electronic compartment using the special wrench.
3)If provided, remove display unit. For this purpose, detach the two screws R
4)Carefully pull off the blue 9-pin plug X1/X4
5)Detach the 2 recessed head screws Q and carefully remove the electronic unit.
6)Carefully transfer DATAPROM IC 18 on the amplifier PCB (illustration, see Sect. 8.9)
7)On the new electronic unit, check supply power and fuse F1, and if necessary change
8)Reassemble in reverse order, points 5) - 1) above.
8/3
Pull off the two plugs for power (3-pin) and outputs/inputs (5-pin).
and fold display unit to the side (illustration, see Sect. 8.8).
(forming the connection to the primary head).
from the “old” to the “new” electronic unit. When inserting, ensure direction of the IC
is correct, see Sect. 8.9 “Illustration of circuit boards”.
over or renew as described in Sect. 8.2, points 6) and 7).
Page 67
IFC 080 and SC 80 AS: Replacement of electronic unit by IFC 090 8.8
Replacement not allowed on flowmeters of hazardous-duty design !
Please consult factory.
Before dismantling the old electronic unit, please note down all converter settings and use these for
the new unit after it has been installed
.
Switch off power source before opening the housing !
1)Unscrew the cover from the terminal compartment using the special wrench anddisconnect all cables from the terminals; note down terminal assignment beforehand.
2)Unscrew the cover from the electronic compartment using the special wrench.
3)Detach the two screws R and fold display unit to the side.
4)Carefully pull off the two blue plugs: 2-pin field power cable and 5-pin signal cable
(forming connection to primary head).
5)Unscrew the 2 recessed head screws Q (size 2 screwdriver, blade length 200 mm / 8”)
and pull out “old” electronic unit.
6)On the new electronic unit check supply power and fuse F1 and change over/replace
if necessary, refer to Sect. 8.2, points 6) and 7).
7)Pull off the two plugs for power (3-pin) and outputs/inputs (6-pin) and carefully insert
the new electronic unit into the housing.
8)Detach the two screws R and fold display unit to the side.
9)Secure the electronic unit with the two screws Q.
10)On the amplifier PCB (see illustration in Sect. 8.9), insert the 2-pin plug of the field
power cable into plug connector X1 and the 5-pin plug of the signal cable into plug
connector X4. Do not kink or twist the cables.
11)Secure the display unit with screws R.
12)In the terminal compartment, press the supplied cover for the terminals into the housing
and connect the cables to the plugs (3-pin for supply power, 6-pin for inputs/outputs).
Ensure terminal assignment is correct, see Sect. 2.
Subsequently insert the plugs into connectors X3 (supply power) and X5 (outputs/inputs).
13)Replace the cover of the terminal compartment and secure using the special wrench.
14)Switch on the power. Check all settings and change if necessary. For setting and
x
operator control, refer to Sect. 4 and 5. Set the GK value (or 1/2
GKL value) for the
IFC 090, see instrument nameplate.
15)Subsequently be sure to check the zero, as described in Sect. 7.1.
16)Replace cover on the electronic compartment and secure using the special wrench.
The screw threads and gaskets of both
IMPORTANT !
F1
housing covers should be well greased at
all times, always check for signs of damage
LP-A
X4
X1
L
X6
X2
C
LP-P
Q
R
and never allow dirt to accumulate.
Replace defective gaskets immediately.
CRibbon cable, display unit
LStatus LED
LP-AAmplifier board, see Sect. 8.9
LP-PPower supply board, see Sect. 8.9
QFastening screws, electronic unit
RFastening screws, display unit
X12-pin connector, field power
X25-pin connector, IMoCom bus
X45-pin connector, electrode signals
X610-pin connector, display unit
8/4
Page 68
8.9 Illustrations of the PCBs
X 3
X 4
X 2
SW
F 1
Tr
X 3
X 4
F 1
F 2
Tr
X 2
A) Amplifier PCB, standard version
Solder points S1 and S3
IC 18DATAPROM
(sensor IC)
S1, S3for empty-tube
cut-off, see Sect. 6.3
S2, S4not used
X12-pin plug connector,
Pin 7 and 8,
see Sect. 7.5 and 7.7
X2IMoCom bus, plug
X 3
connector for connection of external
add-on equipment,
see Sect. 6.2
X324-pin socket
connector
X45-pin plug connector,
Pins 1-5, signal cable,
see Sect. 7.5 and 7.7
X610-pin plug
connector for
dispay unit,
IC 18
X 2
X 6
LED
X 1
X 4
S 1
S 3
S 4
S 2
see Sect. 8.4
B) Power supply PCB, AC Versions 1 and 2C) Power supply PCB, DC Version
F1Power fuse,
ratings see Sect. 8.5 or 9
SW Voltage selector, to change the voltage,
see Sect. 8.2
TrTransformer
8/5
F1,F2 Power fuses,
ratings see Sect. 8.5 or 9
TrTransformer
Page 69
Order numbers 9
IFC 090 electronic unit and
Power supply
unit
1. AC Version
2. AC Version
AC/DC Version
1) 5 x 20 G fuse, switching capacity 1500 A 2) TR 5, switching capacity 35 A
IFC 090 Spares and accessoriesOrder No.
Plug for supply power: all AC Versions (100-240 V AC)3.31122.02
Display unit, retrofit kit for Basic version IFC 090 K / B,
incl. cover with cut-out, clip and connecting cable1.30928.33
RS 232 adapter incl. CONFIG user software toGerman2.10531.00
operate the signal converter via MS-DOS PC or Laptop English2.10531.01
HHT hand-held terminal to operate the signal converter2.10827.00
Special wrench for opening the housing cover3.31038.10
Bar magnet to operate the display converter without opening the housing2.07053.00
Signal converter simulator GS 8A2.07068.01
Adapter for adjustment of older GS 8 simulators to the IFC 090 converter2.10764.00
O-ring gaskets for housing cover3.30870.02
Antiseize/lubricant for screw thread and O-ring gaskets on housing cover
Supply power
230/240 V AC
115/120 VAC
200 V AC
100 V AC
24 V AC/DC
power fuses
Order No.
IFC 090 D
with Display
2.10662.10
2.10662.12
2.10662.14
2.10662.13
2.10663.10
IFC 090 B
without Display
2.10662.00
2.10662.02
2.10662.04
2.10662.03
2.10663.00
Power fuses
(not for “Ex” versions!)
F1 + F2 1)
F1 + F2 1)
F1 + F2 1)
F1 + F2 1)
F1 + F2 2)
125 mA T
200 mA T
125 mA T
200 mA T
1.25 A T
5.06627
5.05678
5.06627
5.05678
5.09080
24 V AC/DC Versions3.31122.03
for outputs/inputs3.31122.01
IFC 090 D-Ex
with Display
2.10662.00
2.10662.02
2.10662.04
2.10662.03
2.10663.00
9/1
Page 70
Part D Technical Data, Measuring Principle and
Block Diagram
10 IFC 110 F technical data
10.1 Full-scale range Q
Full-scale ranges Q
100%
100%
Flow rate Q = 100%6 liter/h to 33 900 m3/h (0.03 - 156 000 US Gal/min), adjustable as
DN/2 xDN (DN = meter size)
Primary headproperly grounded and centered
0 1.6 3.2 6.4 32 36 40 v [ft/s]
Type/Meter sizeMaximum error in % of measured value (MV) ...
DN mm
DN 2.5 – 6(1)
DN 2.5 – 6(1)
≥ DN 10
≥ DN 10
inch
1
/10” – 1/4”(1)
DN 2.5 – 6(1)
≥3/8”
≥ DN 10
v ≥ 1 m/s / ≥ 3 ft/s
≤ ± 0.5% of MV
DN 2.5 – 6(1)
≤ ± 0.3% of MV
≥ DN 10
v < 1 m/s / < 3 ft/s
≤ ± (0.4% of MV + 1 mm/s)
≤ ± (0.4% of MV + 0.04 inch/s
≤ ± (0.2% of MV + 1 mm/s)
≤ ± (0.2% of MV + 0.04 inch/s)
Curve
B
D
A
E
Current outputsame error limits as above, additionally ± 10 µA
Reproducibility and
repeatability0.1% of MV, minimum 1 mm/s / 0.04 inch/s at constant flow
External influencest
ypical valuesmaximum values
Ambient temperature
Pulse output< 0.003% of MV (2)0.015% of MV (2)
Current output< 0.013% of MV (2)0.025% of MV (2)
Power su
Load
pply< 0.023% of MV0.055% of MV (2)at 10% variation
< 0.013% of MV0.025% of MV (2)at max. permissible load,
per 1 K / 1.8°F variation
}
see Sect. 10.4
(1) IFM 6080 K and IFS 6000 F (DN 2.5 – 4 and 1/10 – 1/6”) additional error ± 0,3% of MV
(2) All Krohne signal converters undergo burn-in tests, duration minimum 20 hours at varying ambient temperatures
– 20 to + 60°C/– 4 to + 140°F. The tests are controlled by computers.
10/2
Page 72
10.4 IFC 090 Signal converter
VersionsK = contactF = separate, field housing
IFC 090 K/B and F/B (standard)Basic version, without local display and control elements
IFC 090 K/D and F/D (option)Display version, with local display and control elements
IFC 090 K/D-EExHazardous-duty version with outputs in Increased Safety
Interface (option)HART
Add-on equipment (option)CONFIG software and adapter for operator control via
Forward/reverse flow measurementflow direction identified via status output
Status output (passive)
Functionsettable as measuring range identification for BA mode,
Connectionvoltage: external, U
MS-DOS PC, connection to internal IMoCom interface
(equipment bus)
– galvanically isolated from all output and input circuits
– for active or passive mode (Ex version only active)
for Q = 100%I
for Q > 100%I
= 4 – 20 mAincrements
100%
> 20 up to 22 mA maximum
max
}
load: min. ... max.0 ... 500 Ω250 ... 750 Ω
– galvanically isolated from current output and all input circuits
– digital pulse division, interpulse period non-uniform,
therefore if frequency and cycle meters connected
allow for minimum counting interval:
gate time, totalizer ≥
P
1000
100%
[Hz]
voltage: approx. 15 V DC, from current output
load: I
load: I
voltage: external, U
load: I
variable: 10 ms – 1 s P
< 23 mA, operation without current output
max
< 23 mA, operation with current output
max
≤ 32 V DC/≤ 24 V AC
max
≤ 150 mA
ext
100%
[pulses/s] = f
max
[Hz] =
1
2xpulse width
indicator for flow direction, errors or trip point
≤ 32 V DC/≤ 24 V AC
load current: I
max
ext
≤ 150 mA
10/3
Page 73
Control input (passive)
Function– settable for range change, totalizer reset, error reset,
set outputs to min. values or hold actual output values
– initiate function by “low” or “high” control signals
Control signalsU
: 24 V AC32 V DC (any polarity)
max
low: ≤ 1,4 V≤ 2 V
high: ≥ 3 V≥ 4 V
Output/input combinationsI = current outputP = pulse output
S = status outputC = control input
the following combinations can be set:
1) IPS
2) IPC
3) ICS
4) ISC
5) IS1 S2
6) IC1 C2
Time constant0.2 – 99.9 s, adjustable in increments of 0.1 second
Low-flow cutoffCutoff “on” value: 1 – 19% of Q
Cutoff “off” value: 2 – 20% in 1% increments
}
, adjustable
100%
Local display (D version)3-field LCD
Display functionactual flowrate, forward, reverse and sum totalizers
(7-digit), or 25-character bar graph with percentage
indication and status messages
Units:Actual flowratem
3
/h, liter/s., US gallons/min or user-defined unit,
e. g. liters/day
Totalizerm3, liters, US gallons or user-defined unit,
e. g. hecto-liters or US gallons/day
(adjustable count duration up to overflow)
Language of plain textsEnglish, German, French, others on request
Display:Top field8-character, 7-segment numeral and sign display,
and symbols for key acknowledgement
Middle field10-character, 14-segment text display
Bottom field4 markers to identify display in measuring mode
Operator control– 3 keys → ↵ ↑ or
– 3 magnetic sensors and the supplied
bar magnet without opening the housing
Power supply
1. Rated voltage
Tolerance band
2. Rated voltage
Tolerance band
Frequency
Power consumption
(incl. primary head)
1. AC Version2. AC VersionAC/DC Version
StandardOptionOption
230 / 240 V200 V24 V AC24 V DC
200 – 260 V170 – 220 V20 – 27 V AC18 – 32 V DC
115 / 120 V100 V––
100 – 130 V85 – 110 V––
48 – 63 Hz48 – 63 Hz–
approx. 10 VAapprox. 10 VAapprox. 8 W
When connected to functional extra-low voltage, 24 V, safety separation (PELV)
is essential (to VDE 0100/VDE 0106 and IEC 364/IEC 536 or equivalent
national standard).
Housing
Materialdie-cast aluminium with polyurethane finish
Ambient temperature– 25 to + 60°C (– 13 to + 140°F)
Protection category (IEC 529/EN 60 529)IP 67, equivalent to NEMA 6
10/4
Page 74
10.5 Instrument nameplates
Separate signal converter in rotatable field housing
Altometer
Measuring range
Current output
Tag No.
Primary head constant
Pulse output
Compact flowmeters
Altometer
Measuring range
Current output
Tag No.
Primary head constant
Pulse output
Type designation
Series No.
Holland
Q: 0-150 m
I: 0-20 mA Ri≤ 0.5 kOhm P: 1000 p/s
Tag: 98-FT-096
Q: 0-21.2 m
I: 0-20 mA Ri≤ 0.5 kOhm P: 1000 p/s
Tag:
3
/h GK: 2.706
Type designation
Series No.
Holland
3
/h GK: 5.123
ALTOFLUX IFC090F /D/ /6
A96 6008
ALTOFLUX IFM4080K /D/ /6
A94 9872 ISO Kl. E IP 67
DN50/2”-PFA-HC4 PN 40 Bar
Cconductive rubber compound
HB 2Hastelloy B2
HC 4Hastelloy C4
INIncoloy
M4Monel 400
NiNickel
PTPlatinum
TATantalum
TITitanium
V4AStainless steel 1.4571 (SS 316 Ti)
xx/TC xx with conductive PTFE compound
(xx = base material, e.g. HC4)
Page 75
Measuring principle and function of the system 11
The flowmeter is designed for electrically conductive fluids.
Measurement is based on Faraday’s law of induction, according to which a voltage is induced in
an electrically conductive body which passes through a magnetic field. The following expression
is applicable to the voltage.:
U = K xB xv xD
where:
U = induced voltage
K = an instrument constant
B = magnetic field strength
v = mean velocity
D = pipe diameter
Thus the induced voltage is proportional to the mean flow
velocity, when the field strength is constant.
U
B
D
Inside the electromagnetic flowmeter, the fluid passes
through a magnetic field applied perpendicular to the
direction of flow. An electric voltage is induced by the
movement of the fluid (which must have a minimum
U
v
electrical conductivity). This is proportional to the mean
flow velocity and thus to the volume of flow. The induced
voltage signal is picked up by two electrodes which are in
conductive contact with the fluid and transmitted to a signal
converter for a standardized output signal.
This method of measurement offers the following advantages:
1. No pressure loss through pipe constriction or protruding parts.
2. Since the magnetic field passes through the entire flow area, the signal represents a mean
value over the pipe cross-section; therefore, only relatively short straight inlet pipes
x
DN from the electrode axis are required upstream of the primary head.
5
3. Only the pipe liner and the electrodes are in contact with the fluid.
4. Already the original signal produced is an electrical voltage which is an exact linear function
of the mean flow velocity.
5. Measurement is independent of the flow profile and other properties of the fluid.
The magnetic field of the primary head is generated by a square wave current fed from signal
converter to the field coils.
This field current alternates between positive and negative values. Alternate positive and negative
flowrate-proportional signal voltages are generated at the same frequency by the effect of the
magnetic field, which is proportional to the current. The positive and negative voltages at the
primary head electrodes are subtracted from one another in the signal converter. Subtraction
always takes place when the field current has reached its stationary value, so that constant
interference voltages or external or fault voltages changing slowly in relation to the measuring
cycle are suppressed. Power line interference voltages coupled in the primary head or in the
connecting cables are similarly suppressed.
11/1
Page 76
12 Block diagram – signal converter
1 Input amplifier
• overdrive-proof signal processing, processes flow peaks up to 20 m/s (65 ft/s) and more
rapidly and accurately
• digital signal processing and sequence control
• patented, high-resolution A/D converter, digitally controlled and monitored
• high signal-to-noise ratio through low-loss field power supply with high frequencies
and currents
2 Field power supply
• the low-loss field power supply generates the pulsed, electronically controlled DC current
for the magnetic coils of the primary head
• The high field current ensures a high signal level
3 Current output
• galvanically isolated from all other groups
• converts the digital output signal from the µP 3 microprocessor into a proportional current
4 Binary outputs and/or inputs
• galvanically isolated from other groups
• selectable input/output combinations
• pulse output (B1), passive FET optocouplers allow connection of electronic and
electromechanical totalizers
• status output (B2) for limit value, error identification, or flow direction in forward/reverse
flow mode (F/R) or measuring range identification in BA mode
• both outputs can also be used as control inputs
5 Display/operator control unit (option, D Version)
• large-size illuminated LC display
• 3 keys for operator control of the signal converter
• connection to the internal IMoCom bus
• unit can be retrofitted to basic devices (B Version)
6 IMoCom bus plug connector for connection of external control and test devices such as:
• adapter and CONFIG software for operation via MS-DOS PC
B1/B2 output/input terminals2.1, 2.6, 5.16
BA = automatic range change4.4, 5.181.06, 1.07
Block diagram IFC 09012
B version (basic)4, 6.2, 8.4,10.4
C
C = control input2.5, 2.6, 5.91.06, 1.07, 3.07
Cable length1.3.4
Changeover, power supply8.1
Characteristics of outputs5.15
Clearing error messages4.6
Coding for entry into
Hand held bar magnet4.2
Hand held terminal6.2
Hardware information7.32.02
HART6.7
Hazardous duty areas (Ex)6.1, page 0/4
I
I = current (analog) output2.3, 5.61.05
IEC technical standardspage 0/4
IMoCom bus (plug)6.2, 8.9, 12
Initial start-up3
Input (programming)4.1 et seq.
Instrument nameplates10.5
Interface RS 2326.2, 10.4
Intermediate connection box (ZD)1.3.5, 10.2
Internal electronic totalizer2.4, 5.5, 5.71.04
K
Keys4.1-4.3
Keystroke combinations for
– entry into setting level4.1-4.33.04
– error cancellation4.6
– quitting setting level4.1-4.3
– totalizer reset4.6
L
Language of display texts5.103.01
LCD display, see display4.2, 4.4, 5.41.04
Limit message4.4, 5.81.07
Line voltage, see power supply
Low-flow cutoff (SMU)4.4, 5.31.03
100%
5.71.06, 3.07
5.71.06, 3.07
4.4, 5.1, 5.141.01, 3.02
E1
Page 78
KeywordSection No. Fct. No.
M
Magnetic field frequency4.4 + 5.113.02
Magnetic sensors4.2
Main menu column4.1
Main menus4.1 to 4.31.00, 2.00, 3.00
Mass flow measurement,
see user defined unit4.4, 5.13
Measuring principle11
Menu4.1, 4.4
Meter size (DN)4.4, 5.123.02
N
Numerical format, display5.4, 5.51.04
O
Option = optional equipment6.2, 10.4
Order numbers9
Outputs
– technical data10.1-10.4
Simulator GS 8A7.6
SMU = low-flow cutoff4.4, 5.31.03
Software6.2
Spare parts, see order numbers9
Status output S2.4, 4.4, 5.81.06, 1.07, 3.07
T
T = time constant5.21.02
Technical data
– accuracies10.3
– dimensions & weights10.2
– signal converter IFC 09010.1-10.4
Terminals B1/B22.1, 2.6, 5.16
Tests, see functional checks7.1 et seq.
Time constant (T)5.21.02
Totalizer (internal electronic)2.3, 5.71.06, 3.07
Trip point2.4, 2.6, 5.81.06, 1.07
Troubleshooting, see
functional checks7.1 et seq.
U
Units for
– display4.4, 5.41.04
– flow4.4, 5.11.01
– P4.4, 5.71.06
User-defined unit4.4, 5.133.05
V
v = flow velocity4.4, 5.13.02
VDE standardspage 0/4
W
Weights, see dimensions10.2
Z
Zero check (adjustment)7.13.03
ZD, intermediate connection box1.3.5, 10.2
2.00 et seq.,
3.00 et seq.
3.07 (1.06, 1.07)
E2
Page 79
If you need to return flowmeters for testing
or repair to Krohne
Your electromagnetic flowmeter
• has been carefully manufactured and tested by a company
with ISO 9001 certification
• and volumetrically calibrated in one of the world’s most
accurate test rigs.
If installed and operated in accordance with these operating
instructions, your flowmeter will rarely present any problems.
Should you nevertheless need to return a flowmeter for
checkout or repair, please pay strict attention to the following
:
points
Due to statutory regulations concerning protection of the envi-
ronment and the health and safety of our personnel, Krohne
may only handle, test and repair returned flowmeters that
have been in contact with liquids if it is possible to do so without risk to personnel and environment. This means that
Krohne can only service your flowmeter if it is accompanied
by a certificate in line with the following model confirming that
the flowmeter is safe to handle.
If the flowmeter has been operated with toxic, caustic,
flammable or water-endangering liquids, you are kindly
requested
• to check and ensure, if necessary by rinsing or neutralizing, that all cavities in the flowmeter are free from such
dangerous substances.
(Directions on how you can find out whether the primary
head has to be opened and then flushed out or neutralized
are obtainable from Krohne on request.)
• to enclose a certificate with the flowmeter confirming that
the flowmeter is safe to handle and stating the liquid used.
Krohne regret that they cannot service your flowmeter unless
accompanied by such a certificate.
SPECIMEN certificate
Company:Address:
Department:Name:
Tel. No.:
The enclosed electromagnetic flowmeter
Type:Krohne Order No. or Series No.:
has been operated with the following liquid:
Because this liquid is
water-endangering * / toxic * / caustic * / flammable *
we have
– checked that all cavities in the flowmeter are free from such substances *
– flushed out and neutralized all cavities in the flowmeter *
(* delete if not applicable)
We confirm that there is no risk to man or environment through any residual liquid contained in this flowmeter.
Date:Signature:
Company stamp:
E 3
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.