The flowmeters are supplied ready for operation.
The primary head must be installed in the pipeline as described in the installation
instructions inside the packing of the primary head.
- Installation location and connection to power (Section 1)Pages 1/1-1/6
- Electrical connection of outputs and inputs (Section 2)Pages 2/1-2/6
- Factory settings and start-up (Section 3)Pages 3/1-3/2
- Operator control of the signal converter is described in Sections 4 and 5.
Power the flowmeter. THAT’S ALL. The system is operative.
16-page pull-out condensed instructions
are located in the centrefold of these Installation and Operation Instructions.
Contents:
Installation (Sect. 1), electrical connection (Sect. 1 + 2),
start-up (Sect. 3) and operator control of the signal converter (Sect. 4)
Applicable to
Software Versions
••
IFC 010 _/ D
Display version
No. 806325.07
and
No. 317551.02
and higher
••
IFC 010 _/ B
Basic version
operator-controllable
with HHT 010
No. 806323.06
and higher
3.1M61EA4 049821Order No. 7.02140.34.00
US size: 7.02140.74.00
Contents
Versions IFC 010 signal converter0/3
Items included with supply0/3
Software history0/3
System description0/4
Product liability and warranty0/4
CE / EMC / Standards / Approvals0/4
Part A System installation and start-up1/1-3/2
1 Electrical connection: power supply1/1-1/6
Part C Special applications, functional checks, service, and order numbers6/1-9/1
6 Special applications6/1-6/2
6.1
HHT 010 hand held terminal and RS 232 adapter incl. CONFIG software (options)
6/1
6.2Stable signal outputs when measuring tube empty6/2
7
Functional checks7/1-7/11
7.1Zero check with IFC 010 /D, signal converter, Fct. 3.037/1
7.2Test of measuring range Q, Fct. 2.017/1
7.3Hardware information and error status, Fct. 2.027/2
7.4Faults and symptoms during start-up and process flow measurement7/3-7/6
7.5Test of primary head7/7
7.6Test of signal converter7/8
7.7Test of signal converter using GS 8A simulator (option)7/9-7/11
8
Service8/1-8/6
8.1Cleaning the signal converter housing8/1
8.2Replacement of power fuse(s)8/1
8.3Changeover of operating voltage on AC Versions 1, 2 and 3 (not DC Version) 8/2
8.4Replacement of electronic unit of signal converter8/2
8.5Illustrations to Sect. 8.2 to 8.78/3
8.6Turning the display PCB8/4
8.7Retrofitting the display unit8/4
8.8IFC 080 and SC 80 AS Replacement of electronic unit by IFC 0108/4
8.9Illustrations of the PCBs8/6
9
Order numbers9/1
Part D Technical data, measuring principle and block diagram10/1-12/1
10 Technical data10/1-10/5
10.1Full-scale range Q
100%
10/1
10.2Error limits at reference conditions10/2
10.3IFC 010 Signal converter10/3-10/4
10.4IFC 010 F and ZD dimensions and weights10/4
10.5Instrument name plates10/5
1
1 Measuring principle and function of the system11/1
12 Block diagram - signal converter12/1
Part E IndexE1-E2
Printed form to accompany flowmeters returned to KrohneE3
0/2
Versions IFC 010 signal converter
IFC 010_ / BBasic version (standard)
without local display and control elements.
All operating data factory-set to your order specifications.
Optionally available for operator control:
- RS 232 adapter, incl. software for DOS-PC or
- HHT hand-held terminal
IFC 010_ / DDisplay version (option)
with local display and control elements.
All operating data factory-set to your order specifications.
IFC 010 K / _ Compact flowmeter
signal converter mounted direct on primary head.
IFC 010 F / _ Signal converter in field housing,
electrical connection to primary head
via field power and signal cables.
Items included with supply
• Signal converter in the version as ordered, see above.
• These installation and operating instructions for the signal converter,
including 16-page pull-out condensed instructions for installation, electrical connection,
start-up and operator control of the signal converter.
• 2 plug connectors for connection of power supply and outputs/inputs
• for separate system version only
signal cable in the version and length as ordered (standard: signal cable A, length 10 m / 30 ft)
Software history
Display & control unitHand-held HHT 010CONFIG user software
IFC 010 _ / DIFC 010 _ / B**IMoComRS 485
SoftwareStatusSoftwareStatusSoftwareSoftware
806325.07*current806328.06current
replaces
≥ 317551.02*806325.07806328.06current
813269.00***currentCzech user interface ***
813340.00***currentSwedish user interface ***
*At least the same setting ranges and functional scope as preceding versions.
Also, customer- and application-specific add-on equipment possible, which has to be
installed and activated by the factory. Documented by enclosures to these Installation
and Operating Instructions.
**Please note: Connect HHT 010 only to devices without display and operator
control software.
***Does not contain the functional scope of the currently valid standard version;
this has been taken into account in the documentation of the respective national language.
, F Version:
V 2.00V 3.15
and higherand higher
0/3
System description
Electromagnetic flowmeters with the IFC 010 signal converter are precision instruments designed
for linear flow measurement of liquid products.
The process liquids must be electrically conductive: ≥ 5 µS/cm
(for cold demineralized water ≥ 20 µS/cm).
The full-scale range Q
DN 2.5.- 1000 / 01/10” - 40” Q
This is equivalent to a flow velocity of 0.3 - 12 m/s or 1 - 40 ft/s.
Product liability and warranty
The electromagnetic flowmeters with the IFC 010 signal converter are designed solely for
measuring the volumetric flowrate of electrically conductive, liquid process products.
These flowmeters are not certified for use in hazardous areas. Other flowmeter series are
available for such applications.
Responsibility as to suitability and intended use of these electromagnetic flowmeters rests solely
with the operator.
Improper installation and operation of the flowmeters (systems) may lead to loss of warranty.
In addition, the “General conditions of sale” forming the basis of the purchase contract are
applicable.
If flowmeters need to be returned to Krohne, please note the information given on the last-but-one
page of these Instructions. Krohne regrets that it cannot repair or check your flowmeter(s) unless
accompanied by the completed form sheet.
CE / EMC / Standards / Approvals
• Electromagnetic flowmeters with the IFC 010 signal converter meet the protection requirements
of Directive 89/336/EEC in conjunction with EN 50081-1 (1992) and EN 50082-2 (1995),
as well as Directives 73/23/EEC and 93/68/EEC in conjunction with EN 61010-1, and bear the
CE marking.
0/4
Part A System installation and start-up
1Electrical connection: power supply
1.1Important installation notes PLEASE NOTE !
1.1.1 Location
•Electrical connection in accordance with VDE 0100 ”Regulations governing heavy-currentinstallations with line voltages up to 1000 V” or equivalent national regulations.
•Do not cross or loop cables inside the terminal compartment.
•Use separate cable entries (see below) for power supply, field current cables, signal lines,
outputs and inputs.
•Protect flowmeters or switchgear cabinets with built-in devices from direct sunlight.
Fit a sunshade if necessary.
•When installed in switchgear cabinets, signal converters must be adequately cooled,
e.g. use fans or heat exchangers.
•Do not expose signal converters to intense vibration.
1.1.2 Onl
•Keep distance between primary head and signal converter as short as possible.
Refer to Sect. 1.3.4 for maximum permissible length of signal and field current cables.
•Use the supplied Krohne signal cable A (Type DS), standard length 10 m (33 ft).
•Always calibrate primary head and signal converter together. Therefore, when installing,
ensure primary constant GKL is identical; refer to instrument nameplate for the primary
head. If the GKL is not identical, set the signal converter to the GKL of the primary head.
Refer also to Sections 4.
•Dimensions of signal converter; refer to Section 10.4.
1.1.3 Cable entries
Number of cable entries: 2 for the compact flowmeters
y for separate systems/signal converters (F versions)
4 for the IFC 010 F signal converter
NOTE: Ensure gaskets are fitted correctly and maintain
the following max. torques!
1Max. torques for PG 13.5,
1/2“ NPT or 1/2“ PF adapters: 4 Nm / 2.8 ft
2Max. torques for PG 13.5 only: 3 Nm / 2.1 ft
x
lbf
x
lbf
1
3
2
3Gasket
A)PG 13.5 cable entries
These cable entries may only be used for flexible electrical cables if the relevant electrical
regulations so allow, e.g. National Electric Code (NEC).
Do not fix rigid metal conduits (IMC) or flexible plastic conduits to the PG 13.5 cable entries, refer
to “Point B, C” below (1/2” NPT or PF adapters).
B)1/2“ NPT adaptersC) 1/2“ PF adapters
For most North American systems the regulations require that electrical conductors be laid in
conduits, particularly where power voltages > 100 V AC are concerned.
In such cases, use the 1/2” NPT or 1/2” PF adapters to which flexible plastic conduits can be
screwed. Do not use rigid metal conduits (IMC)!
Lay conduits such that no moisture can penetrate into the converter housing.
Should there be risk of any condensation water forming, fill the cross-section of the conduit
around the cables at these adapters with a suitable sealing compound.
1/1
PLEASE NOTE !
Connection to power 1.2
•Rated values:
The flowmeter housings protecting the electronic equipment from dust
and moisture must always be kept closed. The selected creepage distances and
clearances have been dimensioned in conformity with VDE 0110 and IEC 664 for
contamination category 2. Supply circuits and output circuits are designed to meet the
standards of overvoltage classes III and II, respectively.
•Safet
y isolation: the flowmeters (signal converters) must be provided with an
isolating facility.
1. AC V
ersion2. AC Version
230/240 V AC (200 - 260 V AC)200 V AC (170 - 220 V AC)
switch-selectable toswitch-selectable to
115/120 V AC (100 - 130 V AC)100 V AC (85 - 110 V AC)
•Note information on instrument nameplate: supply voltage and frequency
•The PE protective ground conductor for the power supply must be connected to the
separate U-clamp terminal in the terminal compartment of the signal converter.
For exceptions (compact systems), refer to installation instructions for the primary head.
•Connection diagrams I and II for electrical connection between primary head and
signal converter: refer to Section 1.3.5.
3. AC V
ersionDC Version
48 V AC (41 - 53 V AC)24 V DC (11-32 V DC)
switch-selectable to
24 V AC (20 - 26 V AC)
•Note information on instrument nameplate: supply voltage and frequency.
•For measurement reasons, connect an FE functional ground conductor to the separate
U-clamp terminal in the terminal compartment of the signal converter.
•If connected to a functional extra-low voltage source (24 V AC / DC, 48 V AC),
provide for protective separation (PELV) in conformity with VDE 0100 / VDE 0106 or
IEC 364 / IEC 536, or equivalent national regulations.
•Connection diagrams I and II for power supply and electrical connection between primary
head and signal converter: refer to Section 1.3.5.
Connection to power
Power fuse F1
(see Sect. 8.2)
AC: 100 – 240 V
AC: 24 / 48 V
DC: 24 V
Power
LN
1L ∼ 0L∼
L+L–
for
internal
use only
U-clamp terminal
PE protective conductor
FE
functional ground
}
FE
Warning: Instrument must be properly grounded to avoid personnel shock hazard.
1/2
1.3 Electrical connection of separate primary head (F Versions)
1.3.1 General information on signal cable A and field current cable C
Use of the Krohne signal cable A with foil screen and magnetic shield will ensure proper operation
of the equipment.
• Signal cable to be solidly laid.
• Connect shields via stranded drain wires.
• Underwater and underground installation possible.
• Insulating material is flame-retardant to IEC IEC 332.1 / VDE 0742.
• Signal cables are low in halogen, unplasticized, and stay flexible at low temperatures.
Si
gnal cable A (Type DS), with double shielding
2
1 Stranded drain wire, 1st shield, 1.5 mm
2 Insulation
3 Conductor 0.5 mm
Cross-section is dependent on required length of cable, see Table in Sect. 1.3.4.
1.3.2 Grounding of primary head
All flowmeters must be properly grounded.
• The grounding conductor should not transmit any interference voltages.
• Do not ground any other electrical device together with this conductor.
• The primary head is connected to ground by means of an FE functional ground conductor.
• Special information on grounding various primary heads is contained in the separate
installation instructions for primary heads.
• These instructions also contain detailed descriptions on how to use grounding rings and
how to install primary heads in metal or plastic pipes or internally coated pipelines.
1/3
Stripping (preparation) of signal cable A 1.3.3
Please note the different lengths given in the table for signal converter and primary head.
Customer-supplied materials
W
X Heat-shrinkable tubing or cable sleeve
Y Wire end sleeve to DIN 41 228: E 1.5-8
Z Wire end sleeve to DIN 41 228: E 0.5-8
Insulation tubing (PVC), Ø 2.0 - 2.5 mm (dia. 1”)
Signal cable A
bending radius
≥ 50 mm (≥ 2“)
Preparation for connection to IFC 010 F signal converter
Signal cable A
bending radius
≥ 50 mm (≥ 2“)
External shielding of signal cable A (Type DS)
Wrap stranded drain wire (7) around the mu-metal foil (6) and clamp under the shield terminal in
the signal converter terminal box (see also diagram in Sect. 1.3.5).
Cable routing in signal converter housing
see illustration in Sect. 10.4.
1/4
1.3.4 Cable lengths (max. distance between signal converter and primary head)
Abbreviations and ex
planatory notes
used in the following tables, diagrams and connection diagrams
A Signal cable A (type DS), with double shielding, see diagram for max. length
C Field current cable C, with single shielding,
type and length see Table
2
D High-temperature silicone cable, 3 × 1.5 mm
(14 AWG) Cu, with single shielding,
max. length 5 m (16 ft)
2
E High-temperature silicone cable, 2 × 1.5 mm
(14 AWG) Cu, max. length 5 m (16 ft)
L Cable length
κκ
Electrical conductivity of the process liquid
ZD Intermediate connection box required in connection with cables D and E for primary heads
ALTOFLUX IFS 4000 F, PROFIFLUX IFS 5000 F and VARIFLUX IFS 6000 F in cases where
process temperatures exceed 150 °C (302 °F)
Recommended len
gth of signal cable
for magnetic field frequency ≤ 1/6 × power frequency
Primary head
ECOFLUX IFS 1000 F
AQUAFLUX F
ALTOFLUX IFS 4000 F
PROFIFLUX IFS 5000 F
VARIFLUX IFS 6000 F
Meter size
DN mm
010.0 - 0015
025.5 - 0150
010.0 - 1000
010.0 - 0150
200.0 - 1000
002.5 - 0015
025.0 - 0100
010.5 - 0015
025.5 - 0080
Nennweite
inch
3
/8- 1/
2
1 - 06
3
/8- 40
3
/8- 06
8 - 40
1
/10-1/
2
1 - 04
1
/80- 1/
2
1- 03
Signal cable
A4
A3
A1
A2
A1
A4
A2
A4
A2
Field current cable C: max. length and min. cross-section
LengthType of cable,single shielding
Warning: Instrument must be properly grounded to avoid personnel shock hazard.
1/5
Connection diagrams I and II (power supply, converter and primary head) 1.3.5
Im
portant information PLEASE NOTE !
• The figures in brackets indicate the stranded drain wires for the shields,
see cross-sectional drawing of signal cable in Section 1.3.1.
• Electrical connection to VDE 0100 ”Regulations governing heavy-current installations with
line voltages up to 1000 V“ or equivalent national regulations.
• Power supply 24 V AC / DC:functional extra-low voltage with protective separation in
conformity with VDE 0100, Part 410 or equivalent national
regulations.
for electronic (EC) or
electromechanical (EMC) totalizers
Stromausgang
I
passive
U
ext
I= 0/4 - 20 mA
15...20V DC 20...32 VDC
U
ext
Ri0...500 Ω 250...750 Ω
passive
DC voltage,
external power source (U
ext
),
note connection polarity
External voltage source (U
ext
),
DC or AC voltage,
connection polarity arbitrary
Active mode
The current output supplies the power for
operation of the outputs.
Passive mode
External power source required for operation of
the outputs.
Pulse output P
for electronic (EC) totalizers
with and without current output I
active (and current output I
active
)
U
I
max
U
≤ 32V DC/≤ 24 VAC
ext
I
≤ 150 mA
max
(incl. status output)
5
Status output S
Load
ext
passive
U
I
(incl. pulse output)
≤ 32V DC/≤ 24 VAC
ext
≤ 150 mA
max
U
≤ 15V DC from current output
int
Operation with current output:
I
≤ 3 mA
max
Operation without current output:
I
≤ 23 mA
max
15 V
R ≤
I
max
Status output S
6
with and without current output I
active
Load
U
≤ 15V DC
int
from current output
I
≤ 3 mA
max
Operation with
current output
I
≤ 23 mA
max
Operation without
current output
2/2
3 Start-up
3.1 Power-on and measurement
•Before powering the system, please check that it has been correctly installed according
to Sect. 1 and 2.
•The flowmeter is delivered ready for operational use. All operating data have
been factory set in accordance with your specifications.
Please refer to Sect. 3.2 “factory settings”.
•Power the unit, and the flowmeter will immediately start process flow
measurement.
Basic version
•A light emitting diode (LED) under the cover of the electronic section shows the
measurement status.
LED flashing . . .
•Refer to Sect. 6.1 for operator control of the “basic version”.
play version, signal converter IFC 010 / D
Dis
•When powered, the display shows in succession: START UP and READY.
This is followed by display of the current flow rate and/or the current totalizer count
on either a continuous or alternating basis, depending on the setting under Fct. 1.04.
•Refer to Sect. 4 and 5 for operator control of the “display version”.
, signal converter IFC 010 / B
green:measurement correct, everything all right.
green/red: momentary overdriving of outputs
and/or A/D converter.
red:fatal error, parameter error or hardware fault,
please consult factory.
3/1
Factory settings 3.2
All operating data are factory set according to your order specifications.
If you have not made any particular specifications at the time of ordering, the instruments will be
delivered with the standard parameters and functions listed in the Table below.
To facilitate easy and rapid initial start-up, current output and pulse output are set to process flow
measurement in “2 flow directions”, so that the current flowrate is displayed and the volumetric
flow counted independent of the flow direction. On instruments equipped with a display,
measured values may possibly be shown with a “ – ” sign.
This factory setting for the current and pulse outputs may possibly lead to measuring errors,
particularly in the case of volume flow counting:
for example, if pumps are switched off and a “backflow” occurs which is not within the range of
the low-flow cutoff (SMU), or if separate displays and counts are required for both flow directions.
To avoid faulty measurements, therefore, it may be necessary to change the factory setting of
some or all of the following functions:
– low-flow cutoff SMU, Fct. 1.03, Sect. 5.3
– current output I, Fct. 1.05, Sect. 5.6
– pulse output P, Fct. 1.06, Sect. 5.7
– display (option), Fct. 1.04, Sect. 5.4
Instrument o
peration:
Display versions:IFC 010 _/ D, operation: refer to Sect. 4 and 5.
Basic versions:IFC 010 _/ B, operation: refer to Sect. 6.1.
Standard factory settings
FunctionSetting
1.01 Full-scale range Q
1.02 Time constant3 s, for I, S
1.03 Low-flowON: 1 %
cutoff SMUOFF: 2 %
1.04 Display (option)
flow ratem
totalizer(s)m3 or US Gal
1.05 Current output I
function2 directions
range4 - 20 mA
error message22 mA
1.06 Pulse output P
function2 directions
pulse value1 pulse/s
pulse width50 ms
1.07 Status output Pflow direction
100%
see nameplate
and display
3
/hr or US Gal/min
FunctionSetting
3.01
Language for display only
3.02 Flowmeter
diametersee nameplate
flow direction (see arrow
on primary head)
3.04 Entry codeno
3.05 User unit
English
+ direction
}
Liter/hrorUSMGal/day
3/2
Teil B IFC 010 _/ D Signal converter
4Operation of the signal converter
4.1Krohne operator control concept
1 3 6. 4 9
m3/hr
→
Measuring mode
CodE 1
Menu columnFunction columnData column
3.00 INSTALL.
When this display appears, press following keys:
→ → → ↵ ↵ ↵ ↑ ↑ ↑
3.05 USER UNIT
3.04 ENTRY CODE
3.03 ZERO SET
3.02 FLOWMETER
3.01 LANGUAGE
2.02 HARDW. INFO
Direction of movement
4/1
2.00 TEST
1.00 OPERATION
↑
↓
2.01 TEST Q
1.07 IND. OUTP. S
1.06 PULS. OUTP. P
1.05 CUR. OUTP. I
1.04 DISPLAY
1.03 L.F. CUTOFF
1.02 TIMECONST.
1.01 FULL SCALE
→
↑
↵
see
↓
→
Sect.
↵
4.4
Operating and check elements 4.2
The controls are accessible after unscrewing the 4 screws and
removing the housing cover.
Display,1st line
Display,2nd line
Display,3rd line:arrows to identify display
Flowrate
Totalizer
Overrange
+totalizer
–totalizer
Σsum totalizer (+ and –)
Ioverranging, current output I
Poverranging, pulse output P
current flowrate
Keys for operator control of signal converter
Compass field, signals actuation of a key.
4/2
4.3 Function of keys
The cursor (flashing part of display) has a grey background in the following descriptions.
o start operator control
T
Measuring mode
Operator control mode
1 3 . 5 7 1
m 3 / h r
PLEASE NOTE: When “YES” is set under Fct. 3.04 ENTRY CODE, “CodE 1 - - - - - - - - -”
appears in the display after pressing the → key.
The 9-keystroke Entry Code 1 must now be entered: → → → ↵ ↵ ↵ ↑ ↑ ↑
(each keystroke acknowledged by “
o terminate operator control
T
Press key ↵ repeatedly until one of the following menus
Fct. 1.00 OPERATION, Fct. 2.00 TEST or Fct. 3.00 INSTALL. is displayed.
F c t . 3. 0 0
INSTALL.
→
”).
*
Press key
↵
↵
F c t . 1. 0 0
O P E R AT I O N
S T O R E Y E S
Store new parameters: acknowledge by
pressing key↵ . Measuring mode continued
with the new parameters.
New parameters not to be stored:
press key ↑ to display
„STORE.NO“.
Measuring mode continued with the „old“
↵
parameters after pressing key
.
4/3
To change numbers
increase number
3 9 7. 3 5
↑
m 3 / h r
decrease number
To shift cursor (flashing position)
shift to right
3 9 7. 3 5
→
m 3 / h r
shift to left
To alter texts (units)
For units, the numerical value is
converted automatically.
3. 7 6 9 9
select next text
↑
L i t e r / S e c
select preceding text
To transfer from text (unit) to number setting
Change to number setting
←
3 9 7. 4 5
↓
m 3 / h r
3 9 7. 3 5
m 3/hr
9 3. 3 6 5
↓
U S. G a l / m i n
1 3. 5 7 1
m 3 / h r
→
←
Return to text setting
To transfer to subfunction
Subfunctions do not have a “Fct.No.“ and are identified by a “ → ”.
↵
1 D I R.
To revert to function display
1 0. 3
↵
Sec
1 3. 5 7 1
m 3/hr
→
R A N G E I
F c t. 1. 02
T I M E C O N S T.
4/4
4.4 Table of settable functions
Abbreviations used
DNNominal size, meter size
F
F
F
F
F/RForward/reverse flow in F/R mode
GKLPrimary constant
ICurrent output
PPulse output
P
P
Highest frequency of the pulse output
max
Lowest frequency of the pulse output
min
Conversion factor volume for any unit,
M
see Fct. “FACT. VOL.”
Conversion factor time for any unit,
T
see Fct. 3.05 “FACT. Time”
= F
= F
max
min
/ Q
/ Q
100%
100%
max
min
Qactual flowrate
Q
100%
Q
max
Q
min
SStatus output, control input
SMULow-flow cutoff for I and P
vFlow velocity
v
max
v
min
Fct.TextDescription and settings
1.00OPERATIONOperations menu
1.01FULL SCALEFull-scale range for flowrate Q
Select unit
• m3/hr• Liter/Sec• US.Gal/min
• user unit, factory set is “Liter/hr” or “US MGal/day” (see Fct. 3.05)
Press →key to transfer to number setting.
Setting ranges
The ranges are dependent on the meter size (DN) and the flow
• xxxx PulS/user unit, factory set is “hr“ or “day“ (see Fct. 3.05)
“xxxx“ setting range is dependent on pulse width, see above.
Press ↵key to return to Fct. 1.06 “PULS.OUTP. P“.
1.07IND. OUTP. SStatus output S
• ALL ERROR• FATAL ERROR• OFF• ON
• F/R INDIC. (F/R indication for forward/reverse measurement)
• TRIP. POINT Setting range: 002 - 115 PERCENT
• EMPTY PIPE (appears only when this option is installed)
(
Press
↵
key to transfer to number setting.)
Press ↵key to return to Fct. 1.07 “IND.OUTP. S“.
min
/ Q
100%Pmax
= F
max
/ Q
100%
Fct.TextDescription and settings
2.00TESTTest menu
2.01TEST QTest measuring range Q
Precautionar
• SURE NO
• SURE YES
select value:
y query
Press ↵key to return to Fct. 2.01 “TEST Q”.
Press
↵
key, then use ↑ or ↓ key to
-110 / -100 / -50 / -10 / 0 / +10 / +50 / +100 / +110 PCT.
of set full-scale range Q
Displayed value present at outputs I and P.
Press ↵key to return to Fct. 2.01 “TEST Q”.
2.02HARDW. INFOHardware information and error status
Before consulting factory, please note down all 6 codes.
→ MODUL ADCX . X X X X X . X X
Y Y Y Y Y Y Y Y Y Y
Press ↵key to transfer to
→ MODUL IOX .X X X X X . X X
Y Y Y Y Y Y Y Y Y Y
Press ↵key to transfer to
→ MODUL DISP.X . X X X X X . X X
Y Y Y Y Y Y Y Y Y Y
Press ↵key to return to Fct. 2.02
“HARDW. INFO“.
4/7
100%
.
“MODUL IO“.
“MODUL DISP.“.
Fct.TextDescription and settings
3.00INSTALL.Installation menu
3.01LANGUAGESelect language for display texts
• GB / USA (English)• F (French)
• D (German)• others on request
Press ↵key to return to Fct. 3.01
“LANGUAGE”.
3.02FLOWMETERSet data for primary head
→ DIAMETERSelect size from meter size table
• DN 10 - 1000 mm equivalent to 3/8 - 40 inch
Select with
Press
→ FULL SCALEFull-scale range for flow Q
To set, refer to Fct. 1.01 “FULL SCALE” above.
Press
→ VALUE PPulse value (Fct. 1.06
↑
or
↓
key.
↵
key to transfer to subfunction “FULL SCALE”.
100%
↵ key to transfer to subfunction “GKL VALUE”.
“VALUE P”
) has been changed.
With the “old” pulse values the output frequency (F)
would have been exceeded or not reached.
= F
min
/ Q
100% Pmax
P
min
= F
max
/ Q
100%
→ GKL VALUESet primary constant GKL
see primary head nameplate.
ge:• 1.0000 - 9.9999
Ran
Press ↵key to transfer to subfunction
“FIELD. FREQ.”.
→ FIELD FREQ.Magnetic field frequency
Values: 1/6 or 1/18
of power frequency, see nameplate.
Press ↵key to transfer to subfunction “FLOW DIR.”;
DC units only: to transfer to subfunction
“LINE FREQ.”.
→ LINE FREQ.Normal line frequency in your country
Please note: This function is only provided for units
with DC power supply to suppress line-frequency interference.
Values:
50 Hz
and
60 Hz
Press ↵key to transfer to subfunction “FLOW DIR.”.
→ FLOW DIR.Define flow direction (in F/R mode: forward flow).
Set according to direction of arrow on primary head:
• + DIR.• – DIR.
Select using
Press ↵key to return to Fct. 3.02 “FLOWMETER”.
3.03ZERO SETZero calibration
Note: Carry out only at “0” flow and with completely filled
measuring tube!
Precautionar
• CALIB. NO
• CALIB. YES Press ↵
y query
Press ↵key to return to Fct. 3.03 “ZERO SET”.
key to start calibration.
Duration approx. 15–90 seconds, current flowrate
A “WARNING” sign appears when flowrate “> 0“;
acknowledge by pressing
displayed in the selected unit (see Fct. 1.04 “DISP. FLOW”).
↵
key.
• STORE NO (do not store new zero value)
• STORE YES (store new zero value)
Press ↵key to return to Fct. 3.03 “ZERO SET”.
3.04ENTRY CODEEntry code required to enter setting mode?
• NO(= entry with → only)
• YES (= entry with → and Code 1: → → → ↵ ↵ ↵ ↑ ↑ ↑ )
Press ↵to return to Fct. 3.04 “ENTRY CODE”.
Check new values!
↑
or
↓
key.
4/8
Fct.TextDescription and settings
3.05USER UNITSet any required unit for flowrate and counting
→ TEXT VOL.Set text for required flowrate unit (max. 5 characters)
Factory-set: “Liter” or “MGal”.
Characters assignable to each place:
• A-Z, a-z, 0-9, or “ – “ (= blank character).
Press ↵key to transfer to subfunction “FACT. VOL.”
→ FACT. VOL.Set conversion factor (FM) for volume
Factory set “1.00000” for “Liter” or “2.64172E-4” for “US MGal”
(exponent notation, here: 1x103or 2.64172x10-4).
Factor FM= volume per 1m3.
Setting range • 1.00000 E-9 to 9.99999 E+9 (= 10-9to 10+9)
Press ↵ key to transfer to subfunction “TEXT TIME”.
→ TEXT TIMESet text for required time unit (max. 3 characters)
→ FACT. TIMESet conversion factor (F
Factory-set: “hr” or “day”:
Characters assignable to each place:
• A-Z, a-z, 0-9, or “ – “ (= blank character).
Press
↵ key to transfer to subfunction “FACT. TIME”
) for time
Factory-set: “3.60000 E+3” for “hour” or “8.64000 E+4” for “day”
(exponent notation, here: 3.6 x103or 8.64 x10-4).
Set factor FTin seconds.
Setting range
• 1.00000 E-9 to 9.99999 E+9 (= 10-9to 10+9)
T
Press ↵ key to return to Fct. 3.05 “USER UNIT”.
3.06APPLICAT.Set overload point for A/D converter
→ EMPTY PIPESwitch on “empty tube” identifier option?
(appears only when this option is installed)
• YES • NO
Press ↵key to return to
Select with key ↑or ↓ .
Fct. 3.06 “APPLICAT.“.
4.5 Error messages in measuring mode
The following list gives all errors that can occur during process flow measurement.
Errors shown in display when “YES“ set in Fct. 1.04 DISPLAY, subfunction “DISP. MSG.“.
Error messagesDescription of errorError clearance
LINE INT.Power failure Note:
no counting during power failureReset totalizer if necessary.
CUR. OUTP. ICurrent output overranged.Check and if necessary correct
PULSOUTP. PPulse output overranged.Check and if necessary correct
Note:
totalizer deviation possible.elimination of cause, error message
ADCAnalog / digital Error message deleted automatically
converter overrangedafter elimination of cause.
FATAL. ERRORFatal error, all outputs setPlease consult factory.
to “min. values“
TOTALIZERTotalizer has been resetCancel error message in
EMPTY PIPEPipe has run dry.Fill pipe.
This message appears only when
the “empty pipe identifier”option is installed and the function
is switched on under Fct. 3.06
APPLICAT., submenu
“EMPTY PIPE”.
Cancel error in RESET/QUIT. menu
instrument parameters. After
elimination of cause, error message
deleted automatically.
instrument parameters. After
deleted automatically.
RESET/QUIT. menu.
4/9
Reset totalizer and cancel error messages, RESET / QUIT menu 4.6
Cancel error messages in RESET / QUIT menu
KeyDisplayDescription
↵CodE 2- -Key in entry code 2 for RESET/ QUIT
↑ →ERROR QUIT.Menu for error acknowledgement
→QUIT. NODo not delete error messages,
↑QUIT. YESDelete error messages
The cursor, flashing part of display, is shown below in bold type.
•Change measuring range of current output and value for error messages (Fct. 1.05):
•Change measuring range from 04-20 mA to 00-20 mA
•Change value for error messages from 0 mA to 22 mA
KeyDisplayDescription
→If “YES” set under Fct. 3.04 ENTRY CODE, key in the
Fct. 1.00OPERATION→Fct. 1.01FULL SCALE
4x ↑Fct. 1.05CURRENT I
→FUNCT. I
→↵RANGE IIf “REV. RANGE” appears here,
→04-20mAOld current range
↑00-20mANew current range
↵I ERROR
→0mAOld value for error messages2x ↑22mANew value for error messages
↵Fct. 1.05CURRENT I
↵Fct. 1.00OPERATION
↵STORE YES
↵- - - - - - -- - - - - / - - -Measuring range with new data for the current output
9-keystroke CODE 1 now: → → → ↵ ↵ ↵ ↑ ↑ ↑
press keys → and ↵ again.
4/10
5 Description of functions
5.1 Full-scale range Q
100%
Fct. 1.01 FULL SCALE
Press→key.
Choice of unit for full-scale range Q
100%
•m3/hr(cubic metres per hour)
•Liter/Sec(litres per second)
•US.Gal/min(US gallons per minute)
•User-defined unit, factory-set is “Liter/hr“ (litres per hour) or “US MGal/day”, see Sect. 5.12.
Select with ↑or
↓ key.
Use →key to transfer to numerical setting, 1st number (cursor) flashes.
Set full-scale range Q
100%
The setting range is dependent on meter size (DN) and flow velocity (v).
π
Q
= DN2x v
min
4
minQmax
Change flashing number (cursor) with ↑or
π
= DN2x v
4
0.0053– 33 929m3/hr
0.00147 –9 424.5Liter/Sec
0.00233 – 151 778US.Gal/min
(refer to flow table in Sect. 10.1)
max
↓ key.
Use →key to shift cursor 1 place to right.
Press
↵
key to return to Fct. 1.01 FULL SCALE
.
Note if “VALUE P” is displayed after pressing ↵ key:
PULSE/VOL. is set under Fct. 1.06 PULS B1, subfunction “SELECT P”. Due to the changed
full-scale range Q
P
= F
min
/ Q
min
, the output frequency (F) of the pulse output will be over- or undershot:
100%
100%
P
= F
max
max
/ Q
100%
Change pulse value accordingly, see Sect. 5.7 pulse output B1, Fct. 1.06.
5.2 Time constant
Fct. 1.02 TIMECONST.
Press →key.
Choice
•ALL(applies to display and all outputs)
•ONLY I+S(applies only to display, current and status output)
Select with ↑or
↓ key.
Transfer to number setting with ↵key. 1st number (cursor) flashes.
Set numerical value
•0.2 - 99.9 Sec (seconds)
Change flashing number (cursor) with the↑or
↓ key.
Use →key to shift cursor 1 place to right.
Press
↵
key to return to Fct. 1.02 TIMECONST.
5/1
Low-flow cutoff 5.3
Fct. 1.03 L.F.CUTOFF
Press →key.
Choice
•OFF(fixed trip points:ON = 0.1 % / OFF = 0.2 %,
•PERCENT(variable tripping points: ON = 1 - 19 % / OFF = 2 - 20 %)
for 100 or 1000 Hz, see Fct. 1.06, 1% or 2%)
Select with ↑or
↓ key.
Transfer to number setting using →key (only if “PERCENT” selected).
1st number (cursor) flashes.
Setting the numerical value when “PERCENT” selected
•01 to 19(cutoff “on” value, left of hyphen)
•02 to 20(cutoff “off” value, right of hyphen)
Change flashing number (cursor) with the ↑or
↓ key.
Shift cursor 1 place to right using →key.
Press
↵
key to return to Fct. 1.03 L.F.CUTOFF.
Note: The cutoff “off” value must be greater than the cutoff “on” value.
Display 5.4
Fct. 1.04 DISPLAY
Press →key.
→ DISP. FLOW = select unit for display of flowrate,
•NO DISP.(no display)
•m3/hr(cubic metres per hour)
•Liter/Sec(litres per second)
•US.Gal/min(US gallons per minute)
•user-defined unit, factory-set: “Liter/hr” (litres per hour) or “US MGal/day”, see Sect. 5.14
•PERCENT(percentage display)
•BARGRAPH(numerical value and bar graph display in %)
Select with ↑or
↓ key.
press → key
Press ↵key to transfer to subfunction “DISP. TOTAL”.
→ DISP. TOTAL = select unit for totalizer display,
•NO DISP.(no display)
•OFF(internal totalizer switched off)
• + TOTAL. • – TOTAL. • +/– TOTAL. • SUM (ΣΣ) • ALL (sequential)
Select with ↑or
↓ key.
press → key
Transfer to totalizer unit setting using ↵key.
•m3(cubic metres)
•Liter(litres)
•US.Gal(US gallons)
•user-defined unit, factory-set: “Liter” or “US MGal”, see Sect. 5.14
Select with ↑or
↓ key.
Transfer to totalizer format setting using →key.
Continuation see next page
5/2
Setting of totalizer format
•Auto (exponent notation)
•# . #######• ##### . ###
•## . ######• ###### . ##
•### . #####• ####### . #
•#### . ####• ########
Select with key↑or
Press
↵ key to transfer to subfunction “DISP. MSG”.
→
DISP. MSG. = additional messages required in measuring mode,
↓ .
press → key
•NO(no other messages)
•YES(display other messages, e.g. errors, in sequence with the measured values)
Select using the ↑or
↓ key.
Press ↵key to return to Fct. 1.04 DISPLAY.
Note: “BUSY” is displayed in the measuring mode when all displays are set to “NO DISP.” or
“NO”. Sequencing of displays is automatic. However, in the measuring mode, manual sequencing
can be carried out with the ↑ key. Return to automatic sequencing after approx. 3 minutes.
Please refer to Sect. 3.2 “factory settings”
5.5 Internal electronic totalizer
The internal electronic totalizer counts in m3, regardless of the unit set under Fct. 1.04,
subfunction “DISP. FLOW”.
The counting range is dependent upon the meter size and has been selected such that the
totalizer will count for a minimum of 1 year without overflow:
Only part of the totalizer count is shown in the display because it is not possible to output
a 14-digit number. Unit and format of the display are freely selectable, see Fct. 1.04, subfunction
“DISP. TOTAL” and Sect. 5.4. This determines which part of the count is to be displayed.
Display overflow and totalizer overflow are independent of one another.
Exam
ple
Internal count0000123 . 7654321m
3
Format, display unitXXXX . XXXXLiter
Internal count in unit0123765 . 4321000Liter
Displayed3765 . 4321Liter
Exception: When “OFF” selected, return to Fct. 1.06 PULS B1.
→ SELECT P = select pulse type,
press → key
•100 Hz
•1000 Hz
•PULSE/VOL. (pulses per unit volume, flow)
•PULSE/ZEIT(pulses per unit time for 100% flow)
Select using
Transfer to subfunction ”PULSWIDTH”
↑
or ↓ key.
with ↵key.
Note: when 100 Hz or 1000 Hz selected, return to Fct. 1.06 PULS.OUTP. P.
→ PULSWIDTH = set pulse width,
•50 mSecF
= 10 HzF
max
press →→ key
= 0.0056 Hz (= 20 Pulse / hr)
min
• 100 mSec= 5 Hz
• 200 mSec= 2.5 Hz
• 500 mSec= 1 Hz
• 501 mSec= 0,5 Hz
Select using
Transfer to subfunction ”V
↑
or ↓ key.
ALUE P” with ↵key or return to Fct. 1.06
PULS.OUT. P, depending on choice of pulse type in subfunction ”SELECT P”.
5/5
→ VALUE P = set pulse value per unit volume
(appears only when “PULSE/VOL.” set under “SELECT P”, press →
key.
• XXXX PulS/m3
• XXXX PulS/Liter
• XXXX PulS/US.Gal
• XXXX PulS/ user unit, factory-set: “Liter” or “US MGal”, see Sect. 5.12.
Select using
Transfer to number setting with
↑
or ↓ key.
→
key. 1st digit (cursor) flashes.
Set numerical value
• XXXX(setting range depends on pulse width and
full-scale range: P
Change flashing digit (cursor) with
shift cursor 1 place to right or left with
Press
↵
key to return to Fct. 1.06 PULS.OUTP. P.
min
↑
or ↓ key,
→
= F
key.
min
/ Q
100%
P
= F
max
max
or
→ VALUE P = set pulse value per unit time,
(a
ppears only when ”PULSE/TIME” has been set under “SELECT P”),
• XXXX PulSe/Sec
• XXXX PulSe/min
• XXXX PulSe/hr
• XXXX PulSe/ user unit, factory-set: “hr”, or “day”, see Sect. 5.12.
Select using
Transfer to number setting with
↑
or ↓ key.
→
key, 1st digit (cursor) flashes.
Set numerical value
• XXXX(setting range depends on pulse width)
Change flashing digit (cursor) with
shift cursor 1 place to right or left with
Press
↵
key to return to Fct. 1.06 PULS.OUTP. P.
↑
or ↓ key,
→
key.
/ Q
)
100%
press→key.
Please refer to Sect. 3.2 “factory settings”
Refer to Sect. 2.3 for connection diagrams, and to Sect. 5.14 for characteristics.
5/6
5.8 Status output S
Fct. 1.07 IND. OUTP. S
Press key →.
Select function of status outputs,
press → key
•ALL ERROR(indicates all errors)
•FATAL.ERROR(indicates fatal errors only)
•OFF(switched off, no function)
•ON(indicates that flowmeter is operative)
•F/R INDIC.(indicates direction for current and pulse outputs, F/R mode)
•EMPTY PIPE(option ”empty tube identification“)
•TRIP. POINT(setting range: 002 – 115 PERCENT of Q
, full-scale range)
100%
Transfer to number setting with ↵key, 1st digit (cursor) flashes.
Change flashing digit (cursor) with
↑
and ↓keys. Use
→
and ← keys to shift cursor 1 place to right or left.
Press
↵
key to return to Fct. 1.07 IND. OUTP. S.
• Characteristic of status outputSwitch openSwitch closed
OFF (switched off)no function
ON (e.g. operation indicator)power OFFpower ON
F/R INDIC.Forward flowReverse flow
TRIP POINT (limit switch)inactiveactive
ALL ERROR (all errors)errorsno error
FATAL.ERROR (fatal errors only)errorsno error
EMPTY PIPE (option)when measuring tubewhen measuring tube
is emptyis full
Please refer to Sect. 3.2 “factory settings”
Refer to Sect. 2.3 for connection diagrams, and to Sect. 5.14 for characteristics.
5/7
Language 5.9
Fct. 3.01 LANGUAGE
Press →key.
Select language for texts in display
•D(German)
•GB/USA (English)
•F(French)
•otherson request
Select using ↑key.
Press
↵
key to return to Fct. 3.01 LANGUAGE.
Entry code 5.10
Fct. 3.04 ENTRY CODE
Press →key.
Choice
•NO(no code, enter setting mode with → key)
•YES(enter setting mode with → key and Code 1: → → → ↵ ↵ ↵ ↑ ↑ ↑)
Select using ↑key.
Press
↵
key to return to Fct. 3.04 ENTRY CODE.
5/8
5.11 Primary head
Fct. 3.02 FLOW METER
Press →key.
→ DIAMETER = set meter size (see instrument nameplate)
press → key
Select size from table of meter sizes:
•DN 2.5 - 1000 mm equivalent to
Select using
↑
or ↓ key.
Transfer to subfunction “FULL SCALE”
→ FULL SCALE = set full-scale range,
1
/10 - 40 inch
with ↵key.
press → key.
Set as described in Sect. 5.1.
Transfer to subfunction “GKL VALUE” with ↵key.
Note: if “VALUE P” is displayed after pressing ↵ key.
PULSE/VOL. is set under Fct. 1.06 PULS.OUTP. P, subfunction “SELECT P”. Because the
full-scale range Q
undershot:P
has been changed, the output frequency (F) of the pulse output is over- or
100%
min
= F
min
/ Q
100%
P
= F
max
max
/ Q
100%
Change pulse value accordingly, see Sect. 5.7 pulse output P, Fct. 1.06.
→
GKL VALUE = set primary constant GK,
press → key.
•1.0000 - 9.9999 (note information on instrument nameplate, do not change setting !)
Change flashing digit (cursor) with
Shift cursor 1 place to right or left with
Transfer to subfunction “FIELD
→ FIELD FREQ. = set magnetic field frequency,
•1/6
•1/18
Select using
(1/6 or 1/18 of power frequency, see instrument nameplate, do not change setting)
}
↑
or ↓ key.
Transfer to subfunction “FLOW DIR.”
↑
or ↓ key.
→
or ← key.
FREQ.” with ↵key.
with ↵key.
press → key
(only for units with DC power supply, transfer to subfunction “LINE FREQ”).
→ LINE FREQ. = normal line frequency in your country,
•50 Hz
•60 Hz
→ FLOW DIR. = set flow direction,
Select using the ↑key.
Transfer to subfunction “FLOW DIR.” with ↵ key.
press → key.
press → key.
•+ DIR.(for identification of flow direction, see “+” arrow on primary head;
•- DIR.for F/R mode, identifies the “positive” flow direction)
•Refer to Sect. 2.6 for electrical connection of outputs.
•Define direction of forward (normal) flow, see Fct. 3.02, subfunction “FLOW DIR.”:
in conjunction with F/R operation, set the direction for the forward flow here.
“+” signifies the same direction as shown by the arrow on the primary head,
“–” signifies the opposite direction.
•Set the status output to ”F/R INDIC.”, see Fct. 1.07.
•Current and/or pulse output must be set to “2 DIR.”, see Fct. 1.05 and 1.06, subfunctions
“FUNCT. I” and “FUNCT. B1”.
5.14 Characteristic of outputs
ICurrent output
I
0%
I
100%
0 or 4 mA
20 mA
PPulse output
P
100%
Pulses at Q
, full-scale range
100%
Q
Q
Q
F
R
100%
1 flow direction, forward flow in F/R operation
Reverse flow in F/R operation
Full-scale range
SStatus output
switch open
switch closed
1 flow direction2 flow directions, F/R operation
I
I
max
I
100%
Q
R
%
Q
R
%
I
0%
0
P
P
115%
max
P
100%
P= 0 Hz
0
Q
F
%100 Max
Q
F
%100 Max
Q
%
Q
%
R
R
Max 100
I
I
max
I
100%
I
0%
0
P
P
115%
max
P
100%
P= 0 Hz
0
S
Q
F
%100 MaxMax 100
Q
F
%100 Max
5/11
Q
%
R
Max 100
Q
F
0
%100 Max
Fct. 3.07 APPLICAT.
Press key → .
→
EMPTY PIPE,
• YES • NO
Press
↵
key to return to Fct. 3.07 APPLICAT.
Fct. No.FunctionSettings
1.01Full-scale range
1.02Time constant
1.03Low-flow cut-off- ON:- OFF:
1.04DisplayFlow
1.05Current output IFunction
1.06Pulse output PFunction
1.07Status output S
3.01Language
3.02Primary headMeter size
3.04Entry code required ?- no- yes
3.05User-defined unit
“function switch on“.
select with ↑ or ↓ key.
Here you can note down the settings of the signal converter !
Totalizer
Messages
Range I
Error
Selection
Pulse width
Value
GKL value
Field frequency
Power frequency
Flow direction
→ → → ↵ ↵ ↵ ↑ ↑ ↑
Application 5.15
Setting data 5.16
5/12
Part C Special applications, functional checks,
service, and order numbers
6Special applications
6.1 HHT 010 hand held terminal and RS 232 adapter incl. CONFIG software (options)
External operator control can be carried out with the following options:
– HHT 010 hand-held terminal only for IFC 010 _/ B signal converter (basic version)
– MS-DOS PC via an RS 232 adapter, incl. CONFIG software for
signal converter IFC 010 _/ B (basic version) and IFC 010 _/ D (display version)
Detailed directions supplied.
Switch off
1)Unscrew the 4 recessed head screws and remove the transparent cover.
2)Plug the HHT connection plug or the RS 232 adapter into the IMoCom Bus socket and
3)Switch on the power.
4)As described in the supplied description, change data, parameters and measured values,
5)Switch off the power.
6)Pull the HHT plug or the RS 232 adapter off the amplifier PCB.
7)Replace transparent cover and tighten down the 4 recessed head screws.
power source before opening the housing !
connect to the PC or laptop, see amplifier PCB in Sect. 8.9.
and have them called up for display.
Please refer to Sect. 3.2 “factory settings”
6/1
Stable signal outputs when measuring tube empty 6.2
Output signals can be stabilized to values as for “zero” flow to prevent random output signals
when the measuring tube is empty or when the electrodes are not wetted in the event the
measuring tube is partially full.
Display:0
Current output:0 or 4 mA, see setting in Fct. 1.05.
Pulse output:no pulses (= 0 Hz), see setting in Fct. 1.06.
Precondition:
- electrical conductivity of product ≥ 200 µS/cm (µmho/cm),
≥ 500 µS/cm for meter sizes DN 10 - 15 and 3/8“ - 1/2“
- length of signal cable ≤ 10 m / ≤ 30 ft and free of vibration at field
signal converter
- process products are homogeneous, free from solids and gas inclusions,
and do not tend to cause electrical or catalytic reactions.
Chan
ges on amplifier PCB, see illustration in Sect. 8.9.
Switch off power source before opening the housing !
Refer to Sect. 8.5 for Fig. A, B and D.
1)Unscrew the 4 recessed head screws (Fig. A) and remove transparent cover.
2)Unscrew recessed head screw (Fig. B) and remove black plastic cover.
3)Unscrew the 2 recessed head screws (Fig. D) and remove black metal cover.
4)If display unit provided, unscrew the 4 recessed head screws and fold display carefully
to side.
5)Join the two “semicircles” of points S3 and S6 on the amplifier PCB with tin solder,
see figure in Sect. 8.9.
6)Reassemble in reverse order, Items 4) to 2) above.
7)Switch on power.
8)Check setting of the low-flow cutoff SMU, Fct. 1.03, and reset if necessary:
subfunction “DISP. FLOW”.
Zero measurement in progress, duration approx. 50 seconds.
When flow “> 0” “WARNING” notice appears, confirm with ↵ key.
measuring mode.
7.2 Test of measuring range Q, Fct. 2.01
Switch off power source before opening the housing !
•For this test a measured value can be simulated in the range of -110 to +110 percent
of Q
(full-scale range set, see Fct. 1.01 FULL SCALE).
100%
•Switch on the system.
•Press the following keys for this test:
KeyDisplayDescription
→If “YES” set under Fct. 3.04 ENTRY CODE, key in
Fct. 1.00OPERATION
↑Fct. 2.00TEST
→Fct. 2.01TEST Q
→SURE NO
↑SURE YES
↵0PERCENTCurrent, pulse and status indication outputs indicate the
± 10PERCENT
↑ ± 50PERCENTSelect using ↑ or ↓ key
± 100PERCENT
± 110PERCENT
↵Fct. 2.01TEST QEnd of test, actual measured values again present at outputs.
(2x) 3x ↵- - - - - - -- - - - - / - - -Measuring mode
9-keystroke CODE 1 now: → → → ↵ ↵ ↵ ↑ ↑ ↑
corresponding values.
7/1
Hardware information and error status, Fct. 2.02 7.3
Switch off power source before opening the housing !
•Before consulting factory about errors or flow measurement problems, please invoke
Fct. 2.02 HARDW. INFO (hardware information).
•An 8-character and a 10-character status code are stored under this function in
each of 3 “windows”. These 6 status codes allow rapid and simple diagnosis of your
compact flowmeter.
•Switch on system.
•Press the following keys for display of the status codes:
KeyDisplayDescription
→If “YES” set under Fct. 3.04 ENTRY CODE,
Fct. 1.00OPERATION
↑Fct. 2.00TEST
→Fct. 2.01TEST Q
↑Fct. 2.02HARDW. INFO
key in the 9-keystroke CODE 1 now: → → → ↵ ↵ ↵ ↑ ↑ ↑
ple status code
If you need to return your flowmeter to Krohne, please refer to last
but one page of these Instructions !
7/2
7.4 Faults and symptoms during start-up and process flow measurement
•Most faults and symptoms occurring with the compact flowmeters can be eliminated
with the aid of the following tables.
•For greater clarity, faults and symptoms in the tables are divided into 2 parts and
various groups.
•Part 1
Signal converter IFC 010 B (B = basic version), without display
and without HHT or CONFIG user program (see Sect. 6.1)
Groups:LEDLED display (status messages)
ICurrent output
PPulse output
LED / I / P LED display, current output and pulse output
•Part 2
Signal converter IFC 010 D (D = display version) and
Signal converter IFC 010 B (B = basic version), without display
but with CONFIG user program (see Sect. 6.1)
Groups:DDisplay
ICurrent output
PPulse output
SStatus indication output
D / I / P / S LED display, current output,
status output
Before contacting Krohne Service, please read through the
following tables. THANK YOU !
Part 1Converter IFC 010 B (B = basic version), without display
Group LEDFault / SymptomCauseRemedial action
LED 1LED flashes red/greenOverranging of A/D converter,Reduce flowrate; if unsuccessful,
LED 2LED flashes redFatal Error, hardware and/or Replace signal converter
LED 3cyclic flashing of red LED, Hardware fault, WatchdogReplace signal converter
LED 4LED shows red continuously Hardware faultReplace signal converter
and without HHT or CONFIG operator program
current or pulse outputtest as described in Sect. 7.5.
Measuring tube drained,Fill measuring tube.
A/D converter overranged.
software error(see Sect. 8.4) or contact
Krohne Service.
approx 1 secondactivated.(see Sect. 8.4) or contact
Krohne Service.
(see Sect. 8.4) or contact
Krohne Service.
7/3
Part 1Converter IFC 010 B (B = basic version), without display
(cont’d)and without HHT or CONFIG operator program
Group IFault / SymptomCauseRemedial action
I 1Receiver instrumentConnection/polarity incorrect.Connect properly as described
I 222 mA present at current Current output I overrangedCheck device parameters
I 322 mA present at currentFatal ErrorReplace signal converter
I 4Unsteady display– Process productIncrease time constant,
Group PFault / SymptomCauseRemedial action
P 1Connected totalizer notConnection/polarity incorrectConnect properly as described
P 2Unsteady pulse rate– Process product Increase or switch on time
Group LED / I / PFault / SymptomCauseRemedial action
LED / I / P 1Red LED flashes, current output Fatal Error, hardware faultReplace signal converter
indicates “0”.in Sect. 2.3.
Receiver instrumentCheck connecting cables and
defective.receiver instrument, and
Short between current output Check connections and
and pulse output cables, see Sect. 2.3,
Current output defectiveReplace signal converter
output (fault current)and change if necessary,
output (fault current)(see Sect. 8.4) or contact
and red LED shows.Krohne Service
– conductivity too low,see Sect. 6.1, or contact
– particles/air inclusionsKrohne Service.
– too large or inhomogeneous
– Pulsating flow
– Time constant too low
counting any pulsesin Sect. 2.3
Totalizer or external Check connection cables,
voltage source defectivetotalizer and external voltage
Current output is externalCheck connection and cables
voltage source; short-circuit,(see Sect. 2.3), voltage
or current/pulse output between I+and I⊥ approx.15 V.
defective.Switch off device. Eliminate
Pulse output inactive, Switch on, see Sect. 6.1, or
see Fct. 1.06 and reportcontact Krohne Service.
on settings.
Fatal Error, red LED shows.Replace signal converter
– conductivity too low,constant, see Sect. 6.1,
– particles/air inclusionsor contact Krohne Service.
– too large or inhomogeneous
– Pulsating flow
– Time constant too low
– or switches off
indicates fault current and pulse and/or software error(see Sect. 8.4) or contact
output “0”.Krohne Service.
replace if necessary.
voltage between I+and
I⊥ approx. 15 V. Switch off
device, eliminate short-circuit,
switch device on.
(see Sect. 8.4) or contact
Krohne Service.
see Sect. 6.1, or contact
Krohne Service
source, and replace if necessary.
short-circuit and switch device
on again. If no function then
current or pulse output
defective. Replace signal
converter (see Sect. 8.4) or
contact Krohne Service.
(see Sect. 8.4) or contact
Krohne Service.
7/4
Part 2Signal converter IFC 010 D (D = display version) and
Group DDisplay shows . . .CauseRemedial action
D 1LINE INT.Power failure.Delete error message in
D 2CUR.OUTP. ICurrent output overranged.Check instrument parameters
D 3PULS.OUTP. PPulse output overranged.Check instrument parameters,
D 4ADCAnalog/digital converterError message deleted
D 5FATAL. ERRORFatal Error, Replace signal converter
D 6TOTALIZERCounts lostDelete error message in
D 7„STARTUP“Hardware fault,Replace signal converter
D 8BUSYDisplays for flow,Change setting in
D 9Unsteady display– Process productIncrease time constant,
D 10No displayPower OFF.Switch on power
Group IFault / SymptomCauseRemedial action
I 1Receiver instrument Incorrect connection/polarityConnect properly,
I 2Unsteady display– Process productIncrease time constant,
Signal converter IFC 010 B (B = basic version), without display
but with HHT or CONFIG operator program (see Sect. 6.1)
no counting duringRESET/QUIT. menu. Reset
Note:
power failure.totalizer(s) if need be.
and correct if necessary.
Error message deleted
automatically after cause
has been eliminated.
totalizer deviationcorrect if necessary, and reset
Note:
possibletotalizer(s). Error message
overranged.automatically after cause
all outputs set to “min.”(see Sect. 8.4) or contact
values.Krohne Service, having
(overflow, data error)RESET/QUIT. menu.
cyclic flashingWatchdog activated.(see Sect. 8.4) or contact
totalizers and errors disabled.Fct. 1.04.
– conductivity too low,see Sect. 6.1, or contact
– particles/air inclusionsKrohne Service.
– too large or inhomogeneous
– Pulsating flow
– Time constant too low
Check power fuse(s) Replace if defective
F1 (F1 and F2 with DC).(see Sect. 8.2).
indicates “0”.see Sect. 2.3.
Receiver instrument or currentCheck output (see Sect. 7.2)
output defective.with new milliammeter:
Current output disabled,Activate under Fct. 1.05.
see Fct. 1.05
Short-circuit between currentCheck connection and cables,
output and pulse output.see Sect. 2.3. Voltage between
– conductivity too low,see Sect. 6.1, or contact
– particles/air inclusionsKrohne Service.
– too large or inhomogeneous
– Pulsating flow
– Time constant too low
deleted automatically
after cause has been
eliminated.
has been eliminated.
first noted down hardware
information and error status,
see Fct. 2.02.
Krohne Service.
est ok, check connection
T
cables and receiver instrument,
replace if necessary.
est faulty, current output
T
defective. Replace signal
converter (see Sect. 8.4)
or contact Krohne Service
I+ and I⊥ approx. 15 V.
Switch off device, eliminate
short-circuit, and switch
device on again.
7/5
Part 2Signal converter IFC 010 D (D = display version) and
(cont’d)Signal converter IFC 010 B (B = basic version), without display
Group PFault / SymptomCauseRemedial action
P 1Totalizer connected but doesIncorrect connection/polarityConnect properly,
P 2Unsteady pulse rate– Process productIncrease or switch on
P 3Pulse rate too high orIncorrect setting forChange setting under Fct. 1.06.
Group SFault / SymptomCauseRemedial action
S 1No functionIncorrect connection/polarityConnect properly,
Group D/I/P/SFault / SymptomCauseRemedial action
D/I/P/S1Unsteady display and– Process productIncrease time constant,
D/I/P/S 2No display and noPower OFFSwitch power on.
but with HHT or CONFIG operator program (see Sect. 6.1)
not count any pulsessee Sect. 2.3
Totalizer or external voltageCheck output (see Sect. 7.2)
source defective.with new totalizer:
Current output is externalCheck connection and cables,
voltage source, short circuitsee Sect. 2.3. Voltage between
or current / pulse outputI+and I⊥ approx. 15 V.
defectiveSwitch off device, eliminate
Pulse output is deactivated,Switch on under Fct. 1.06.
see Fct. 1.06.
– conductivity too low,time constant,
– particles/air inclusionssee Sect. 6.1, or contact
– too large or inhomogeneousKrohne Service.
– Pulsating flow
– Time constant too low
– or switched off
too low.pulse output.
of status displaysee Sect. 2.3.
Status display or outputSet status output under Fct. 1.07
defective or external to “F/R INDIC.“ (flow direction)
voltage source notand check (see Sect. 7.2)
supplying voltage.with new status display:
outputs– conductivity too low,see Sect. 6.1, or contact
– particles/air inclusionsKrohne Service.
– too large or inhomogeneous
– Pulsating flow
– Time constant too low
function of outputs
Check power fuse(s)Replace if defective,
F1 (F1 and F2 for DC).see Sect. 8.2.
est ok, check connection
T
cables and previous
totalizers and external voltage
source, and replace if
necessary.
est faulty, pulse output
T
defective, replace signal
converter (see Sect. 8.4)
or contact Krohne Service.
short circuit, switch device
on again. If no function,
then current or pulse
output defective. Replace
signal converter (see Sect.
8.4) or contact Krohne
Service.
est ok, check previous
T
status display and external
voltage source, and replace
if necessary.
est faulty, status output
T
defective, replace signal
converter (see Sect. 8.4) or
contact Krohne Service.
7/6
7.5 Test of primary head
Always switch off power source before opening the housing !
Re
quired measuring instruments and tools
-A crosstip screwdriver
-Ohmmeter with at least 6 V range
or
AC voltage/resistance measuring bridge.
Note: Accurate measurements in the area of the electrodes can only be obtained with an
AC voltage/resistance bridge. Also, the measured resistance is very heavily dependent on
the electrical conductivity of the liquid product.
Preparatory work
•Switch off power source.
•Remove transparent cover
(unscrew 4 recessed head screws)
and black plastic cover
7 8 1 2 3 4 5
Blue 9-pin plug
(connection to primary head)
unassigned
(unscrew 1 recessed head screw),
see Fig. A and B in Sect. 8.5.
•Detach blue 9-pin plug,
see Fig. D in Sect. 8.5,
field power supply (pins 7+8)
X1
X4
and signal cables (pins 1, 2, 3, 4 +5).
•Completely fill the measuring tube of the
flowmeter with the process liquid.
ActionTypical resultIncorrect result =
Resistance measurementflowmeter defective,
at the blue 9-pin plugreturn to factory for repair,
1Measure resistance30 - 150 Ωif lower:
between wiresinterwinding fault
7 and 8.
2Measure > 10 MΩif lower:
resistance between interwinding fault to PE
U-clamp terminal or FE.
in terminal compartment
(PE protective conductor
or FE functional ground)
and wires 7 and 8
between wires 1 and 3(see “Note” above).tube and repeat
and between 1 and 4Both values should bemeasurement; if still
(same measuring leadapprox. equal.too low, short-circuit in
always on wire 1 ! )electrode wires.
Jacks X1 and X4 on amplifier PCB,
see Sect. 8.9
refer to last but one page !
if higher:
wire break
if higher: break in
electrode wires or
electrodes
contaminated.
Values not equal:
break in electrode
wires or electrodes
contaminated.
7/7
Test of signal converter 7.6
Always switch off power source before opening the housing !
Re
quired measuring instruments and tools
Multimeter, DC and AC voltage, > 20 kohms / V
Crosstip screwdriver
Preparatory workMeasuring and test points on
•Switch off power source.amplifier PCB, see Sect. 8.9
•Remove transparent cover (unscrew 4MP= measuring point
recessed head screws) and black plasticTP= test point
cover (unscrew 1 recessed head screw),X1= socket connector, 20-pin
see Fig. A and B in Sect. 8.5.X3, X5 = plug connector
•If provided, remove display PCB,
see Sect. 8.7.
•Switch power source on again.
Please note: When carrying out measurements, do not produce any short-circuits between the
components !
ActionTypical resultFaulty result
Voltage measurements on
amplifier PCB, see Sect. 8.9
1between TP 1 (solder pin)15 . . . 30 V DC
and Pin 11 of X1
2between TP 1 (solder pin)30 . . . 40 V DC
and Pin 9 of X1
3between MP 5 (solder pin)19 . . . 26 V DC
and Pin 15 of X1
4between MP 5 (solder pin)-20 . . . -27 V DC
and Pin 18 of X1
5field current supply> 1.5 V AC
between Pin 7 and Pin 8
of X3
6input voltage between-10 . . . +10 V DCIf outside range, input
MP 1 and MP 5amplifier overranged,
measuring tube
empty or primary head
defective; check acc.
to Sect. 7.5.
7Short Pin 1, 2 and 3-10 . . . +10 V DCIf outside range,
of X5, measure inputsignal converter
voltage betweendefective.
MP 1 and MP 5
If measured
voltages lower,
signal converter
defective,
replace, see Sect. 8.4,
or contact Krohne
Service.
Note:A thermostatic switch is installed in the transformers of the AC versions.
In the DC version, the PCB is equipped with a thermofuse. All signal converters
contain PTR fusible links (typically 100 overload cycles). This allows cyclic
switching on and off of the signal converter under overload conditions.
The cool-down phase may be anything up to one hour.
7/8
7.7 Test of signal converter using GS 8 A simulator (option)
+
0
–
GS 8 A operating elements and accessories
Y
P
20
10
2,5
1,25
20
40
5
80
160
320
D
L
H
Electrical connection
H2
A5-to-3 pin adapter
for cable C5
BPlug for field power
supply, 2-pin
C3 /C5Plug for signal cable,
3-pin / 5-pin
DSwitch,
flow direction
HSocket for H2 plug
on cable Z
H2Plug of cable Z
LPower supply ON
PPotentiometer “zero”
X3Socket for plug B
on amplifier PCB
X5Socket for plug C3
on amplifier PCB
YSwitch,
measuring ranges
ZCable between GS 8 A
and signal converter
Z
Power
LN
1L~ 0L~
L+ L-
C 5
B
Connection of milliammeter and
electronic frequency counter,
see Sect. 2.3 “Connection of outputs”.
X 5X 3
amplifier PCB, see
Sect. 8.9
a) Switch off power source before
opening the housing !
b) Unscrew the 4 recessed head screws, see
Fig. A in Sect. 8.5, and remove transparent
cover from the signal converter housing.
c) Unscrew recessed head screw, see Fig. B in
Sect. 8.5, and remove black plastic cover.
Using GS 8 simulator
Additional adapter is required
between GS 8 and IFC 010 signal converter.
Order No. 2.10764.00.
OutputsU-clamp terminal
I⊥ I I+ P P⊥ SPE
10
Milliammeter, accuracy
A
class 0.1, Ri< 500 ohms,
range 4-20 mA
Electronic frequency
counter, input resistance
approx. 1 kohm, range
0-1 kHz, time base min. 1
second, see connection
diagrams in Sect. 2.3.2.
d) Pull off blue 7-pin plug from the amplifier PCB,
see Sect. 8.9:
socket X3 field power supply and socket X5
signal cables.
e) Connect plug B to socket X3 (2-pin) and
plug C (5-pin) to socket X5 (3-pin) by way
of adapter A (5-to-3pin).
FE
FE
Warning: Instrument must be properly grounded to avoid personnel shock hazard.
7/9
Check of set
point display
1)Switch on power source, allow at least 15 minutes’ warm-up time..
2)Set switch D (front panel of GS 8A) to “0” position.
3)Adjust zero to 0 or 4 mA with the 10-turn potentiometer P (front panel of GS 8A),
depending on setting in Fct. 1.05, deviation <±10 µA.
4)Calculate position of switch Y and displayed setpoints ”I“ and ”f“:
x
Q
K
4.1)X =
100%
2
x
GK
DN
Q
full-scale range (100%) in volumetric unit V per unit time t
100%
GKprimary constant, see instrument nameplate
DNmeter size DN im mm, not inches, see instrument nameplate
ttime in seconds (Sec), minutes (min) or hours (hr)
Vvolumetric unit
Kconstant, according to following table
Sticker on the GS 8 A primary head simulator still gives values for
“inch” flowmeter. Do not use any more !
4.2)Determine position of switch Y: Use table (front panel GS 8A) to determine value Y
which comes closest to factor X and meets condition Y ≤ X.
4.3)Calculate set
4.4)Calculate set
point reading ”I“ for current output:I = I0%+ Y(I
I
0%
I
100%
point reading ”f“ for pulse output:f =
P
X
current (0/4mA) at 0% flowrate
current (20mA) at 100% flowrate
Y
x
P
100%
X
pulses per second (Hz)
100%
100%
in Hz
- I0%) in mA
at 100% flowrate
5)Set switch D (front panel GS 8A) to position “+“ or “-“ (forward/reverse flow).
6)Set switch Y (front panel GS 8A) to the value determined by the method described above.
7)Check setpoint readings I and f, see points 4.3 and 4.4 above.
8)Deviation <1.5% of setpoint. If greater, replace signal converter, see Sect. 8.4.
9)Test of linearity: set lower Y values, readings will drop in proportion to the calculated Y
values.
10)Switch off power source after completing the test.
11)Disconnect the GS 8A.
12)Reassemble in reverse order, see points e) – b) under “electrical connection”,
see illustration in Sect. 8.5.
13)The system is ready for operation after the power source has been switched on.
Exam
ple: see overleaf
7/10
Example
Full-scale rangeQ
= 200 m3/hr (Fct. 1.01)
100%
Meter sizeDN= 80 mm = 3“ (Fct. 3.02)
Current at Q
Pulses atQ
Q
0%
100%
100%
I
0%
I
100%
P
= 4 mA
= 20 mA
= 280 pulses/hr (Fct. 1.06)
100%
}
(Fct. 1.05)
Primary head constantGKL= 3.571 (see nameplate)
Constant(Vin m3)
(tin hr)K= 7074 (see Table)
(DN in mm)
Calculation of “ X ”
Q
x K
100%
X =
GKLxDN23.571 x 80 x 80
and position of “ Y ”
200 x 7074
=
= 61.905
Y = 80, position of switch Y, see front panel GS 8A
(comes closest to X value and is smaller than X).
Calculation of set
I = I0%+Y(I
X 61.905
point readings I and f
- I0%) = 4 mA +
100%
40
(20mA - 4mA) = 14.3 mA
Deviations are permissible between 14.1 and 14.6 mA
f = Yx P
100%
X 61.905
=
40
x pulses / hr = 129.2 pulses/hr
Deviations are permissible between 127.3 and 131.1
If you need to return your flowmeter to Krohne,
please refer to last but one page of these Instructions !
(equivalent to ± 1.5 %).
pulses/hr (equivalent to ± 1.5 %).
7/11
Service 8
Cleaning the signal converter housing 8.1
Switch off power source before cleaning !
The housing of the signal converter (material: polycarbonate, PC) may only be cleaned with a
solventless detergent !
Replacement of power fuse(s) 8.2
A) Fuse F1 in AC Versions 1, 2 and 3
Switch off power source before opening the housing !
Refer to Sect. 8.5 for Figs. A and B.
1)Unscrew the 4 recessed head screws (Fig. A), and remove transparent cover from
signal converter housing.
2)Unscrew the recessed head screw (Fig. B) and remove the black plastic cover.
3)Take out old and insert new power fuse F1, on the left next to the green connecting
terminals. Please refer to the following table for fuse rating and order number.
4)Reassemble in reverse order, points 2) – 1) above..
B) Fuse F1 and F2 for the DC V
ersion
Switch off power source before opening the housing !
Refer to Sect. 8.5 for Figs. A to F.
1)Unscrew the 4 recessed head screws (Fig. A), and remove transparent cover.
2)Unscrew the recessed head screw (Fig. B) and remove the black plastic cover.
3)Carefully pull out the green connecting plugs (supply power and outputs) (Fig. C).
4)Unscrew the 2 recessed head screws (Fig. D) and remove the black metal cover.
5)Carefully pull out the blue 9-pin plug (connection to the primary head) (Fig. D).
6)With a screwdriver, carefully remove the 4 metal clips (Fig. E).
7)Remove the electronics unit from the housing (Fig. F) and detach the ground conductor.
8)Replace power fuses F1 and F2 on the power supply PCB, refer to Sect. 8.9 for illustration
of the PCB. Refer to the following table for fuse rating and order number.
10)Reassemble in reverse order, points 7) - 1) above.
Power PCBSupply powerFuse F1 (and F2)Location and position
RatingOrder No.of voltage selector
1. AC version230/240 V AC125 mA T5.06627
115/117 V AC200 mA T5.05678
2. AC version200 V AC125 mA T5.06627
100 V AC200 mA T5.05678
F 1
F 1
F 1
F 1
3. AC version48 V AC400 mA T5.05892
24 V AC800 mA T5.08085
DC version11-32 V DCF1 + F2
1.25 A T5.09080
F 1
F 1
Warning: Instrument must be properly grounded to avoid personnel shock hazard.
8/1
8.3 Changeover of operating voltage on AC Versions 1, 2 and 3 (not DC Version)
Switch off power source before opening the housing !
Refer to Sect. 8.5 for Figs. A-F.
1)Unscrew the 4 recessed head screws (Fig. A), and remove transparent cover.
2)Unscrew the recessed head screw (Fig. B) and remove the black plastic cover.
3)Carefully pull out the green connecting plugs (supply power and outputs) (Fig. C).
4)Unscrew the 2 recessed head screws (Fig. D) and remove the black metal cover.
5)Carefully pull out the blue 9-pin plug (connection to the primary head) (Fig. D).
6)With a screwdriver, carefully remove the 4 metal clips (Fig. E).
7)Remove the electronics unit from the housing (Fig. F) and detach the ground conductor.
8)Transpose voltage selector on the power supply PCB (illustration in Sect. 8.9)
to obtain the required voltage according to the table in Sect. 8.2.
9)Change power fuse F1, see table for fuse ratings.
10)Reassemble in reverse order, points 7) - 1) above.
8.4 Replacement of electronics unit of signal converter
Switch off power source before opening the housing !
Refer to Sect. 8.5 for Figs. A-G.
1)Unscrew the 4 recessed head screws (Fig. A), and remove transparent cover.
2) Unscrew the recessed head screw (Fig. B) and remove the black plastic cover.
3) Carefully pull out the green connecting plugs (power supply and outputs) (Fig. C).
4) Unscrew the 2 recessed head screws (Fig. D) and remove the black metal cover.
5) Carefully pull out the blue 9-pin plug (connection to the primary head) (Fig. D).
6) With a screwdriver, carefully remove the 4 metal clips (Fig. E).
7) Carefully remove the electronics unit from the housing (Fig. F) and detach the
ground conductor.
8) Carefully transpose the DATAPROM (IC 13) on the amplifier PCB (illustration in Sect. 8.9)
from the “old” to the “new” electronics unit (Fig. G). When inserting, note the direction
of the IC 13, see Sect. 8.9 “illustration of the PCBs”.
9) Check supply power and fuse F1 for the new electronics unit, and if necessary change
over or replace as described in Sect. 8.3, points 8) and 9).
10) Reassemble in reverse order, points 7) - 1) above.
8/2
Illustrations to Sect. 8.2 to 8.7 8.5
Fig. A
Fig. B
Fig. C
Fig. D
Fig. E
Fig. F
Fig. G
8/3
8.6 Turning the display PCB
Switch off power source before opening the housing !
Figs. A, B and D are given in Sect. 8.5.
1) Unscrew the 4 recessed head screws (Fig. A) and remove the transparent cover.
2) Unscrew the recessed head screw (Fig. B) and remove the black plastic cover.
3) Unscrew the 2 recessed head screws (Fig. D) and remove the black metal cover.
4) Unscrew the 4 recessed head screws on the display PCB.
5) Turn display PCB carefully.
6) Fold the ribbon cable as shown in the dia
PLEASE NOTE: The ribbon cable must lie flat between the display and amplifier PCBs
and must not exert any pressure on electronic components.
7) Reassemble in reverse order, points 4) - 1) above.
8.7 Retrofitting the display unit
Switch off power source before opening the housing !
Figs. A, B and D are given in Sect. 8.5.
1) Unscrew the 4 recessed head screws (Fig. A) and remove the transparent cover.
2) Unscrew the recessed head screw (Fig. B) and remove the black plastic cover.
3) Unscrew the 2 recessed head screws (Fig. D) and remove the black metal cover.
4) Insert the foil connector of the display unit into jack X7 on the amplifier PCB, see diagram
in Sect. 8.9. Ensure contact side is correctly positioned..
5) Carefully turn display in the desired direction.
Fold the ribbon cable as shown in the dia
PLEASE NOTE: The ribbon cable must lie flat between the display and amplifier PCBs
and must not exert any pressure on electronic components.
6) Reassemble in reverse order, see points 3) - 2) above.
7) Switch on power source.
8) For operator control and display of measured values, refer to Sect. 4 and 5.
9) Replace transparent cover and tighten down the 4 recessed head screws (Fig. A).
grams in Sect. 8.8
grams in Sect. 8.8
Socket X7 lockedSocket X7 unlocked
F
VB
X7
amplifier PCB
FRibbon cable
VBLocking clip X7
X7Socket on amplifier PCB
8/4
VB
F
X7
Directions for folding the ribbon cable on the display unit 8.8
ASocket X7 on amplifier PCB,
see Sect. 8.9
BSocket on display PCB
CContact side
DDisplay
FRibbon cable
IInsulated side
K5 keys for operator control
RReference point, power terminals
90° Bend cables 90° as shown
in drawing
Display
board
Version 2
B
F
A
D
K
I
90°
Version 2 only
F
Amplifier
board
B
F
90°
F
C
A
Version 1 / Standard IFC 010 K/D
Compact flowmeter
R
F
B
Version 3/ Standard IFC 010 F/D
Separate version
F
R
Version 4
R
F
R
B
R
A
D
K
B
A
D
K
A
D
K
8/5
8.9 Illustrations of the PCBs
A) Amplifier PCB
IC 13DATAPROM (sensor), see Sect. 8.4
MP1, MP5Measuring points, see Sect. 7.6
S3, S6for “empty tube” cut-out, see Sect. 6.2
TP1Test point, see Sect. 7.6
X120-pin socket connector, see Sect. 7.6 and 7.7
X32-pin plug connector, pin 7 and 8, field power supply, see Sect. 7.5 and 7.7
X4IMoCom Bus, plug connector for connection of RS 232 adapter, see Sect. 6.1
X55-pin plug connector, pin 1-5, signal cable, see Sect. 7.5 and 7.7
X710-pin socket (A) for foil connector (display unit), see Sect. 8.6 and 8.7.
Solder points S3 and S6
TP 1
18
X 1
15119
1
X3
X5
MP 5
7
8
S4
S3
1
2
3
4
5
S6
S5
MP2MP1
X7
X4
IC13
8/6
B) Power supply PCB, AC Versions 1, 2 and 3C) Power supply PCB, DC Version
SW
Tr
Tr
F1Power fuse,
see Sect. 8.2 or 9 for ratings
SW Voltage selector, see Sect. 8.3 for
voltage changeover
TrTransformer
F1
F1,F2 Power fuses,
see Sect. 8.2 or 9 for ratings
TrTransformer
F2
F1
9 Order numbers
Electronic unit
Power supply unitSupply powerOrder No.
IFC 010 DIFC 010 B
with displaywithout display
ersion 1230 / 240 V AC2.07494.102.07494.00
AC V
AC
Version 2200 V AC2.07494.122.07494.02
AC Version 348 V AC2.07494.342.07494.24
ersion24 V DC2.07527.102.07527.00
DC-V
115 / 117 V AC2.07494.152.07494.05
100 V AC2.07494.142.07494.04
24 V AC2.07494.582.07494.48
(11-32 V DC)
Power fuse F1 for AC and F1 and F2 for DC
Supply powerRatingOrder No.Fuse type
200 and 230 / 240 V AC125 mA T5.06627
100 and 115 / 117 V AC200 mA T5.056785 x20 G-fuse
48 V AC400 mA T5.05892switching capacity 1500 A
24 V AC800 mA T5.08085
11-32 V DC1.25 A T5.09080TR 5, switching capacity 35 A
Display unit, retrofit kit for basic version IFC 010 _/ B
incl. transparent cover and connecting cable:Order No. 1.30915.92
RS 232 adapter incl. English CONFIG operator software,
for operator control of signal converter via MS-DOS PC or laptop:Order No.
HHT hand held terminal for operator control of signal converterOrder No.
2.10531.01
2.10591.01
9/1
Part D Technical data, measuring principle and
block diagram
10 Technical data
10.1 Full-scale range Q
Full-scale ranges Q
100%
100%
Flow rate Q = 100%6 liter/h to 33 900 m3/h (0.03 - 156 000 US Gal/min), adjustable as
Current outputsame as above error limit for pulse output plus ...
0 to 20 mA:
4 to 20 mA:
± 0.05 %
}
± 0.062 %
of full-scale range in each case
}
10/2
10.3 IFC 010 Signal converter
Versions
B - versionwithout display / control unit (basic version)
D - versionwith display / control unit
Add-on equipment (option)- CONFIG software and RS 232 adapter for control via
Current output
Function- all operating data settable, galvanically isolated
Current0 - 20 mA and 4 - 20 mA
Active outputload max. 500 ohms
Passive outputexternal voltage: 15 ... 20 V DC20 ... 32 V DC
Error identification0 / 3.6 / 22 mA
Forward/reverse measurementdirection identified via status output
Pulse output
Function- all operating data settable
Pulse rate for Q = 100 %10, 100 or 1000 pulses per second (= Hz), fixed or optionally
Active output:connection: electronic totalizer (EC)
Passive output:connection electromechanical (EMC) or electronic (EC)
Pulse width50, 100, 200, 500 ms or 1 s,
Forward/reverse measurementdirection identified via status output
Status output (passive)
Functionsettable as indicator for flow direction, errors or trip point
Connectionexternal voltage: U
Time constant0.2 to 99.9 seconds, settable in increments of 0.1 second
Low-flow cutoffcutoff ”on” value: 1 to 19 % of Q
Local display (D versions only)3-line LCD
Display functionsactual flowrate, forward, reverse and sum totalizers (7-digit)
Display units: actual flowratesettable in liter/s, m
totalizersliter, m3or US gallons and 1 user-defined unit
Language of plain textsEnglish, German, French, others on request
Display:1st (top) line8-character, 7-segment numeral and sign display,
2nd (middle) line10-character, 14-segment text display
3rd (bottom) line6 markers t to identify display
MS-DOS-PC, connection to IMoCom interface
- Hand-Held-Terminal for control of basic versions
- Other bus and computer interfaces on request
load: min. ... max.0 ... 500 Ω250 ... 750 Ω
- galvanically isolated
- digital pulse division, interpulse period non-uniform,
therefore if frequency meters connected allow for minimum
counting interval:
gate time, totalizer ≥
adjustable in pulses per liter, m
P
1000
100%
[Hz]
3
or US gallons
(special version: up to 10 kHz scaling)
internal voltage: approx. 15 V DC, from current output
load rating: I
load rating: I
< 23 mA when operated without current output
max
< 23 mA when operated with current output
max
totalizers
external voltage: U
load current: I
≤ 30 V DC / ≤ 24 V AC
ext
≤ 150 mA
max
selectable with frequencies below 10 Hz
≤ 30 V DC / ≤ 24 V AC
load current: I
cutoff ”off” value: 2 to 20 % 1 % increments
ext.
≤ 150 mA
max
, adjustable in
100%
}
or 25-character bar graph with percent display and
status messages
3
/h, US gallons/min or user-defined unit,
e.g. hectoliter/day or US million gallons/day
(e.g. hectoliter), selectable overflow time
symbols for key acknowledgement
10/3
Power supply
1. Rated voltage
tolerance band
2. Rated voltage
tolerance band
Frequency
Power consumption
(incl. primary head)
1. AC-Version2. AC-Version3. AC-VersionDC-Version
StandardOptionOptionOption
230 / 240 V200 V48 V24 V
When connected to a functional extra-low voltage, 11-32 V DC, protective
separation (PELV) must be ensured (VDE 0100/VDE 0106 and IEC 364/IEC 536)
Housing
Materialpolycarbonate (PC)
Protection category (IEC 529/EN 60 529)
IFC 010 K (compact)IP 67, equivalent to NEMA 6, same as primary head
IFC 010 F (separate)IP 65, equivalent to NEMA 4 / 4X
IFC 010 F and ZD dimensions and weights 10.4
IFC 010 F
weight approx. 3.8 kg / 8.4 lb
160 (6.30”)
144 (5.67”)
9 (0.35”)
1
220 (8.66”)
2
5
34 (1.34”)
120 (4.72”)
3
258 (10.16”)
1
ZD Intermediate
connection box
Weight approx.
0.5 kg/1.1 lbs
1
dia. 9
(0.35”)
4
4
27.5
(1.08”)
dia. 9
(0.35”)
57 (4.02”)
22 (4.02”)
80
(3.15”)
PG 16
1
/2” NPT
68 (2.68”)
125 (4.92”)
1 output cable (see Sect. 2.3)
2 signal cable of primary head
(see Sect. 1.3)
3 field power cable of primary head
(see Sect. 1.3)
4 power supply cable (see Sect. 1.2)
5 internal connection
116 (4.57”)
(ssee Fig. in Sect. 8.9, plug connectors
X3 and X5)
dimensions in mm
10/4
10.5 Instrument nameplates
Separate signal converter in rotatable field housing
Altometer
Measuring range
Current output
Tag No.
Primary head constant
Pulse output
Compact flowmeters
Altometer
Measuring range
Current output
Tag No.
Primary head constant
Pulse output
Type designation
Series No.
Holland
Q: 0-150 m
I: 0-20 mA Ri≤ 0.5 kOhm P: 1000 p/s
Tag: 98-FT-096
Q: 0-21.2 m
I: 0-20 mA Ri≤ 0.5 kOhm P: 1000 p/s
Tag:
3
/h GKL: 2.706
Type designation
Series No.
Holland
3
/h GKL: 5.123
ALTOFLUX IFC090F /D/ /6
A96 6008
ALTOFLUX IFM4010K /D/ /6
A94 9872 ISO Kl. E IP 67
DN50/2”-PFA-HC4 PN 40 Bar
CConductive rubber compound
HB 2Hastelloy B2
HC 4Hastelloy C4
INIncoloy
M4Monel 400
NiNickel
PTPlatinum
TATantalum
TITitanium
V4AStainless steel 1.4571 (SS 316 Ti)
xx/TC xx with conductive PTFE compound
(xx = base material, e.g. HC4)
Measuring principle and function of the system 11
The flowmeter is designed for electrically conductive fluids.
Measurement is based on Faraday’s law of induction, according to which a voltage is induced in
an electrically conductive body which passes through a magnetic field. The following expression
is applicable to the voltage.:
xB xv x
U = K
D
where:
U = induced voltage
K = an instrument constant
B = magnetic field strength
v = mean velocity
D = pipe diameter
Thus the induced voltage is proportional to the mean flow
velocity, when the field strength is constant.
U
B
D
Inside the electromagnetic flowmeter, the fluid passes
through a magnetic field applied perpendicular to the
direction of flow. An electric voltage is induced by the
movement of the fluid (which must have a minimum
U
v
electrical conductivity). This is proportional to the mean
flow velocity and thus to the volume of flow. The induced
voltage signal is picked up by two electrodes which are in
conductive contact with the fluid and transmitted to a signal
converter for a standardized output signal.
This method of measurement offers the following advantages:
1. No pressure loss through pipe constriction or protruding parts.
2. Since the magnetic field passes through the entire flow area, the signal represents a mean
value over the pipe cross-section; therefore, only relatively short straight inlet pipes
x
DN from the electrode axis are required upstream of the primary head.
5
3. Only the tube liner and the electrodes are in contact with the fluid.
4. Already the original signal produced is an electrical voltage which is an exact linear function
of the mean flow velocity.
5. Measurement is independent of the flow profile and other properties of the fluid.
The magnetic field of the primary head is generated by a square wave current fed from the signal
converter to the field coils.
This field current alternates between positive and negative values. Alternate positive and negative
flowrate-proportional signal voltages are generated at the same frequency by the effect of the
magnetic field, which is proportional to the current. The positive and negative voltages at the
primary head electrodes are subtracted from one another in the signal converter. Subtraction
always takes place when the field current has reached its stationary value, so that constant
interference voltages or external or fault voltages changing slowly in relation to the measuring
cycle are suppressed. Power line interference voltages coupled in the primary head or in the
connecting cables are similarly suppressed.
11/1
12 Block diagram – signal converter
1 Input amplifier
• overdrive-proof signal processing, rapid and accurate
• digital signal processing and sequence control
• patented, high-resolution A/D converter, digitally controlled and monitored
• high signal-to-noise ratio through low-loss field power supply
2 Field power supply
• the low-loss field power supply generates the pulsed, electronically
controlled DC current for the magnetic coils of the primary head
3 Current output
• galvanically isolated from all other groups
• converts the digital output signal from the µP 3 microprocessor into a proportional current
4 Binary outputs
• galvanically isolated from other groups
• selectable input/output combinations
• pulse output (P), passive FET optocouplers allow connection
of electronic and electromechanical totalizers
• status output (S), for limit value, error identification, or flow direction
in forward/reverse flow mode (F/R)
5 Display/operator control unit (option, D Version)
• large-size illuminated LC display
• 3 keys for operator control of the signal converter
• connection to the internal IMoCom bus
• unit can be retrofitted to basic devices (B Version)
6 IMoCom bus plug connector
for connection of external control and test devices such as:
• HHT handheld terminal (option), display/operator control unit for operation of basic versions
• adapter and CONFIG software for operation via MS-DOS PC
I = current (analog) output2.2, 2.3, 5.61.05
IEC technical standardspage 0/4
IMoCom bus (plug)6.1, 8.9,12
Initial start-up3
Input (programming)4.1 et seq.
Instrument nameplates10.5
Interface RS 2326.1, 10.3
Intermediate connection box (ZD)1.3.5, 10.4
Internal electronic totalizer2.4, 5.5, 5.71.04
K
Keys4.1-4.3
Keystroke combinations for
– entry into setting level4.1-4.33.04
– error cancellation4.6
– quitting setting level4.1-4.3
– totalizer reset4.6
L
Language of display texts5.103.01
LCD display, see display4.2, 4.4, 5.41.04
Limit message4.4, 5.81.07
Line voltage, see power supply
Low-flow cutoff (SMU)4.4, 5.31.03
100%
4.4, 5.11.01, 3.02
E1
KeywordSection No. Fct. No.
M
Magnetic field frequency4.4 + 5.113.02
Main menu column4.1
Main menus4.1 to 4.31.00, 2.00, 3.00
Mass flow measurement,
see user defined unit4.4, 5.12
Measuring principle11
Menu4.1, 4.4
Meter size (DN)4.43.02
N
Numerical format, display5.4, 5.51.04
O
Option = optional equipment6.1, 10.3
Order numbers9
Outputs
– technical data10
Simulator GS 8A7.7
SMU = low-flow cutoff4.4, 5.31.03
Software6.1
Spare parts, see order numbers9
Status output S2.2,2.3, 4.4, 5.7 1.06, 1.07
T
T = time constant5.21.02
Technical data
– accuracies10.2
– dimensions & weights10.4
– signal converter IFC 01010.1, 10.3, 10.4
Tests, see functional checks7.1 et seq.
Time constant (T)5.21.02
Totalizer (internal electronic)2.2, 5.71.06
Trip point4.4, 5.81.06, 1.07
Troubleshooting, see
functional checks7.1 et seq.
U
Units for
– display4.4, 5.41.04
– flow4.4, 5.11.01
– P4.4, 5.71.06
User-defined unit4.4, 5.123.05
V
v = flow velocity4.4, 5.13.02
VDE standardspage 0/4,
W
Weights, see dimensions10.4
Z
Zero check (adjustment)7.13.03
ZD, intermediate connection box1.3.5, 10.4
1.1 and 2.1
et seq.
2.00 et seq.,
3.00 et seq.
E2
If you need to return flowmeters for testing
or repair to Krohne
Your electromagnetic flowmeter
• has been carefully manufactured and tested by a company
with ISO 9001 certification
• and volumetrically calibrated in one of the world’s most
accurate test rigs.
If installed and operated in accordance with these operating
instructions, your flowmeter will rarely present any problems.
Should you nevertheless need to return a flowmeter for
checkout or repair, please pay strict attention to the following
:
points
Due to statutory regulations concerning protection of the envi-
ronment and the health and safety of our personnel, Krohne
may only handle, test and repair returned flowmeters that
have been in contact with liquids if it is possible to do so without risk to personnel and environment. This means that
Krohne can only service your flowmeter if it is accompanied
by a certificate in line with the following model confirming that
the flowmeter is safe to handle.
If the flowmeter has been operated with toxic, caustic,
flammable or water-endangering liquids, you are kindly
requested
• to check and ensure, if necessary by rinsing or neutralizing, that all cavities in the flowmeter are free from such
dangerous substances.
(Directions on how you can find out whether the primary
head has to be opened and then flushed out or neutralized
are obtainable from Krohne on request.)
• to enclose a certificate with the flowmeter confirming that
the flowmeter is safe to handle and stating the liquid used.
Krohne regret that they cannot service your flowmeter unless
accompanied by such a certificate.
SPECIMEN certificate
Company:Address:
Department:Name:
Tel. No.:
The enclosed electromagnetic flowmeter
Type:Krohne Order No. or Series No.:
has been operated with the following liquid:
Because this liquid is
water-endangering * / toxic * / caustic * / flammable *
we have
– checked that all cavities in the flowmeter are free from such substances *
– flushed out and neutralized all cavities in the flowmeter *
(* delete if not applicable)
We confirm that there is no risk to man or environment through any residual liquid contained in this flowmeter.
Date:Signature:
Company stamp:
E 3
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.