The flowmeters are supplied ready for operation.
The primary head must be installed in the pipeline as described in the installation
instructions inside the packing of the primary head.
- Installation location and connection to power (Section 1)Pages 1/1-1/6
- Electrical connection of outputs and inputs (Section 2)Pages 2/1-2/6
- Factory settings and start-up (Section 3)Pages 3/1-3/2
- Operator control of the signal converter is described in Sections 4 and 5.
Power the flowmeter. THAT’S ALL. The system is operative.
16-page pull-out condensed instructions
are located in the centrefold of these Installation and Operation Instructions.
Contents:
Installation (Sect. 1), electrical connection (Sect. 1 + 2),
start-up (Sect. 3) and operator control of the signal converter (Sect. 4)
Applicable to
Software Versions
••
IFC 010 _/ D
Display version
No. 806325.07
and
No. 317551.02
and higher
••
IFC 010 _/ B
Basic version
operator-controllable
with HHT 010
No. 806323.06
and higher
3.1M61EA4 049821Order No. 7.02140.34.00
US size: 7.02140.74.00
Contents
Versions IFC 010 signal converter0/3
Items included with supply0/3
Software history0/3
System description0/4
Product liability and warranty0/4
CE / EMC / Standards / Approvals0/4
Part A System installation and start-up1/1-3/2
1 Electrical connection: power supply1/1-1/6
Part C Special applications, functional checks, service, and order numbers6/1-9/1
6 Special applications6/1-6/2
6.1
HHT 010 hand held terminal and RS 232 adapter incl. CONFIG software (options)
6/1
6.2Stable signal outputs when measuring tube empty6/2
7
Functional checks7/1-7/11
7.1Zero check with IFC 010 /D, signal converter, Fct. 3.037/1
7.2Test of measuring range Q, Fct. 2.017/1
7.3Hardware information and error status, Fct. 2.027/2
7.4Faults and symptoms during start-up and process flow measurement7/3-7/6
7.5Test of primary head7/7
7.6Test of signal converter7/8
7.7Test of signal converter using GS 8A simulator (option)7/9-7/11
8
Service8/1-8/6
8.1Cleaning the signal converter housing8/1
8.2Replacement of power fuse(s)8/1
8.3Changeover of operating voltage on AC Versions 1, 2 and 3 (not DC Version) 8/2
8.4Replacement of electronic unit of signal converter8/2
8.5Illustrations to Sect. 8.2 to 8.78/3
8.6Turning the display PCB8/4
8.7Retrofitting the display unit8/4
8.8IFC 080 and SC 80 AS Replacement of electronic unit by IFC 0108/4
8.9Illustrations of the PCBs8/6
9
Order numbers9/1
Part D Technical data, measuring principle and block diagram10/1-12/1
10 Technical data10/1-10/5
10.1Full-scale range Q
100%
10/1
10.2Error limits at reference conditions10/2
10.3IFC 010 Signal converter10/3-10/4
10.4IFC 010 F and ZD dimensions and weights10/4
10.5Instrument name plates10/5
1
1 Measuring principle and function of the system11/1
12 Block diagram - signal converter12/1
Part E IndexE1-E2
Printed form to accompany flowmeters returned to KrohneE3
0/2
Versions IFC 010 signal converter
IFC 010_ / BBasic version (standard)
without local display and control elements.
All operating data factory-set to your order specifications.
Optionally available for operator control:
- RS 232 adapter, incl. software for DOS-PC or
- HHT hand-held terminal
IFC 010_ / DDisplay version (option)
with local display and control elements.
All operating data factory-set to your order specifications.
IFC 010 K / _ Compact flowmeter
signal converter mounted direct on primary head.
IFC 010 F / _ Signal converter in field housing,
electrical connection to primary head
via field power and signal cables.
Items included with supply
• Signal converter in the version as ordered, see above.
• These installation and operating instructions for the signal converter,
including 16-page pull-out condensed instructions for installation, electrical connection,
start-up and operator control of the signal converter.
• 2 plug connectors for connection of power supply and outputs/inputs
• for separate system version only
signal cable in the version and length as ordered (standard: signal cable A, length 10 m / 30 ft)
Software history
Display & control unitHand-held HHT 010CONFIG user software
IFC 010 _ / DIFC 010 _ / B**IMoComRS 485
SoftwareStatusSoftwareStatusSoftwareSoftware
806325.07*current806328.06current
replaces
≥ 317551.02*806325.07806328.06current
813269.00***currentCzech user interface ***
813340.00***currentSwedish user interface ***
*At least the same setting ranges and functional scope as preceding versions.
Also, customer- and application-specific add-on equipment possible, which has to be
installed and activated by the factory. Documented by enclosures to these Installation
and Operating Instructions.
**Please note: Connect HHT 010 only to devices without display and operator
control software.
***Does not contain the functional scope of the currently valid standard version;
this has been taken into account in the documentation of the respective national language.
, F Version:
V 2.00V 3.15
and higherand higher
0/3
System description
Electromagnetic flowmeters with the IFC 010 signal converter are precision instruments designed
for linear flow measurement of liquid products.
The process liquids must be electrically conductive: ≥ 5 µS/cm
(for cold demineralized water ≥ 20 µS/cm).
The full-scale range Q
DN 2.5.- 1000 / 01/10” - 40” Q
This is equivalent to a flow velocity of 0.3 - 12 m/s or 1 - 40 ft/s.
Product liability and warranty
The electromagnetic flowmeters with the IFC 010 signal converter are designed solely for
measuring the volumetric flowrate of electrically conductive, liquid process products.
These flowmeters are not certified for use in hazardous areas. Other flowmeter series are
available for such applications.
Responsibility as to suitability and intended use of these electromagnetic flowmeters rests solely
with the operator.
Improper installation and operation of the flowmeters (systems) may lead to loss of warranty.
In addition, the “General conditions of sale” forming the basis of the purchase contract are
applicable.
If flowmeters need to be returned to Krohne, please note the information given on the last-but-one
page of these Instructions. Krohne regrets that it cannot repair or check your flowmeter(s) unless
accompanied by the completed form sheet.
CE / EMC / Standards / Approvals
• Electromagnetic flowmeters with the IFC 010 signal converter meet the protection requirements
of Directive 89/336/EEC in conjunction with EN 50081-1 (1992) and EN 50082-2 (1995),
as well as Directives 73/23/EEC and 93/68/EEC in conjunction with EN 61010-1, and bear the
CE marking.
0/4
Part A System installation and start-up
1Electrical connection: power supply
1.1Important installation notes PLEASE NOTE !
1.1.1 Location
•Electrical connection in accordance with VDE 0100 ”Regulations governing heavy-currentinstallations with line voltages up to 1000 V” or equivalent national regulations.
•Do not cross or loop cables inside the terminal compartment.
•Use separate cable entries (see below) for power supply, field current cables, signal lines,
outputs and inputs.
•Protect flowmeters or switchgear cabinets with built-in devices from direct sunlight.
Fit a sunshade if necessary.
•When installed in switchgear cabinets, signal converters must be adequately cooled,
e.g. use fans or heat exchangers.
•Do not expose signal converters to intense vibration.
1.1.2 Onl
•Keep distance between primary head and signal converter as short as possible.
Refer to Sect. 1.3.4 for maximum permissible length of signal and field current cables.
•Use the supplied Krohne signal cable A (Type DS), standard length 10 m (33 ft).
•Always calibrate primary head and signal converter together. Therefore, when installing,
ensure primary constant GKL is identical; refer to instrument nameplate for the primary
head. If the GKL is not identical, set the signal converter to the GKL of the primary head.
Refer also to Sections 4.
•Dimensions of signal converter; refer to Section 10.4.
1.1.3 Cable entries
Number of cable entries: 2 for the compact flowmeters
y for separate systems/signal converters (F versions)
4 for the IFC 010 F signal converter
NOTE: Ensure gaskets are fitted correctly and maintain
the following max. torques!
1Max. torques for PG 13.5,
1/2“ NPT or 1/2“ PF adapters: 4 Nm / 2.8 ft
2Max. torques for PG 13.5 only: 3 Nm / 2.1 ft
x
lbf
x
lbf
1
3
2
3Gasket
A)PG 13.5 cable entries
These cable entries may only be used for flexible electrical cables if the relevant electrical
regulations so allow, e.g. National Electric Code (NEC).
Do not fix rigid metal conduits (IMC) or flexible plastic conduits to the PG 13.5 cable entries, refer
to “Point B, C” below (1/2” NPT or PF adapters).
B)1/2“ NPT adaptersC) 1/2“ PF adapters
For most North American systems the regulations require that electrical conductors be laid in
conduits, particularly where power voltages > 100 V AC are concerned.
In such cases, use the 1/2” NPT or 1/2” PF adapters to which flexible plastic conduits can be
screwed. Do not use rigid metal conduits (IMC)!
Lay conduits such that no moisture can penetrate into the converter housing.
Should there be risk of any condensation water forming, fill the cross-section of the conduit
around the cables at these adapters with a suitable sealing compound.
1/1
PLEASE NOTE !
Connection to power 1.2
•Rated values:
The flowmeter housings protecting the electronic equipment from dust
and moisture must always be kept closed. The selected creepage distances and
clearances have been dimensioned in conformity with VDE 0110 and IEC 664 for
contamination category 2. Supply circuits and output circuits are designed to meet the
standards of overvoltage classes III and II, respectively.
•Safet
y isolation: the flowmeters (signal converters) must be provided with an
isolating facility.
1. AC V
ersion2. AC Version
230/240 V AC (200 - 260 V AC)200 V AC (170 - 220 V AC)
switch-selectable toswitch-selectable to
115/120 V AC (100 - 130 V AC)100 V AC (85 - 110 V AC)
•Note information on instrument nameplate: supply voltage and frequency
•The PE protective ground conductor for the power supply must be connected to the
separate U-clamp terminal in the terminal compartment of the signal converter.
For exceptions (compact systems), refer to installation instructions for the primary head.
•Connection diagrams I and II for electrical connection between primary head and
signal converter: refer to Section 1.3.5.
3. AC V
ersionDC Version
48 V AC (41 - 53 V AC)24 V DC (11-32 V DC)
switch-selectable to
24 V AC (20 - 26 V AC)
•Note information on instrument nameplate: supply voltage and frequency.
•For measurement reasons, connect an FE functional ground conductor to the separate
U-clamp terminal in the terminal compartment of the signal converter.
•If connected to a functional extra-low voltage source (24 V AC / DC, 48 V AC),
provide for protective separation (PELV) in conformity with VDE 0100 / VDE 0106 or
IEC 364 / IEC 536, or equivalent national regulations.
•Connection diagrams I and II for power supply and electrical connection between primary
head and signal converter: refer to Section 1.3.5.
Connection to power
Power fuse F1
(see Sect. 8.2)
AC: 100 – 240 V
AC: 24 / 48 V
DC: 24 V
Power
LN
1L ∼ 0L∼
L+L–
for
internal
use only
U-clamp terminal
PE protective conductor
FE
functional ground
}
FE
Warning: Instrument must be properly grounded to avoid personnel shock hazard.
1/2
1.3 Electrical connection of separate primary head (F Versions)
1.3.1 General information on signal cable A and field current cable C
Use of the Krohne signal cable A with foil screen and magnetic shield will ensure proper operation
of the equipment.
• Signal cable to be solidly laid.
• Connect shields via stranded drain wires.
• Underwater and underground installation possible.
• Insulating material is flame-retardant to IEC IEC 332.1 / VDE 0742.
• Signal cables are low in halogen, unplasticized, and stay flexible at low temperatures.
Si
gnal cable A (Type DS), with double shielding
2
1 Stranded drain wire, 1st shield, 1.5 mm
2 Insulation
3 Conductor 0.5 mm
Cross-section is dependent on required length of cable, see Table in Sect. 1.3.4.
1.3.2 Grounding of primary head
All flowmeters must be properly grounded.
• The grounding conductor should not transmit any interference voltages.
• Do not ground any other electrical device together with this conductor.
• The primary head is connected to ground by means of an FE functional ground conductor.
• Special information on grounding various primary heads is contained in the separate
installation instructions for primary heads.
• These instructions also contain detailed descriptions on how to use grounding rings and
how to install primary heads in metal or plastic pipes or internally coated pipelines.
1/3
Stripping (preparation) of signal cable A 1.3.3
Please note the different lengths given in the table for signal converter and primary head.
Customer-supplied materials
W
X Heat-shrinkable tubing or cable sleeve
Y Wire end sleeve to DIN 41 228: E 1.5-8
Z Wire end sleeve to DIN 41 228: E 0.5-8
Insulation tubing (PVC), Ø 2.0 - 2.5 mm (dia. 1”)
Signal cable A
bending radius
≥ 50 mm (≥ 2“)
Preparation for connection to IFC 010 F signal converter
Signal cable A
bending radius
≥ 50 mm (≥ 2“)
External shielding of signal cable A (Type DS)
Wrap stranded drain wire (7) around the mu-metal foil (6) and clamp under the shield terminal in
the signal converter terminal box (see also diagram in Sect. 1.3.5).
Cable routing in signal converter housing
see illustration in Sect. 10.4.
1/4
1.3.4 Cable lengths (max. distance between signal converter and primary head)
Abbreviations and ex
planatory notes
used in the following tables, diagrams and connection diagrams
A Signal cable A (type DS), with double shielding, see diagram for max. length
C Field current cable C, with single shielding,
type and length see Table
2
D High-temperature silicone cable, 3 × 1.5 mm
(14 AWG) Cu, with single shielding,
max. length 5 m (16 ft)
2
E High-temperature silicone cable, 2 × 1.5 mm
(14 AWG) Cu, max. length 5 m (16 ft)
L Cable length
κκ
Electrical conductivity of the process liquid
ZD Intermediate connection box required in connection with cables D and E for primary heads
ALTOFLUX IFS 4000 F, PROFIFLUX IFS 5000 F and VARIFLUX IFS 6000 F in cases where
process temperatures exceed 150 °C (302 °F)
Recommended len
gth of signal cable
for magnetic field frequency ≤ 1/6 × power frequency
Primary head
ECOFLUX IFS 1000 F
AQUAFLUX F
ALTOFLUX IFS 4000 F
PROFIFLUX IFS 5000 F
VARIFLUX IFS 6000 F
Meter size
DN mm
010.0 - 0015
025.5 - 0150
010.0 - 1000
010.0 - 0150
200.0 - 1000
002.5 - 0015
025.0 - 0100
010.5 - 0015
025.5 - 0080
Nennweite
inch
3
/8- 1/
2
1 - 06
3
/8- 40
3
/8- 06
8 - 40
1
/10-1/
2
1 - 04
1
/80- 1/
2
1- 03
Signal cable
A4
A3
A1
A2
A1
A4
A2
A4
A2
Field current cable C: max. length and min. cross-section
LengthType of cable,single shielding
Warning: Instrument must be properly grounded to avoid personnel shock hazard.
1/5
Connection diagrams I and II (power supply, converter and primary head) 1.3.5
Im
portant information PLEASE NOTE !
• The figures in brackets indicate the stranded drain wires for the shields,
see cross-sectional drawing of signal cable in Section 1.3.1.
• Electrical connection to VDE 0100 ”Regulations governing heavy-current installations with
line voltages up to 1000 V“ or equivalent national regulations.
• Power supply 24 V AC / DC:functional extra-low voltage with protective separation in
conformity with VDE 0100, Part 410 or equivalent national
regulations.
for electronic (EC) or
electromechanical (EMC) totalizers
Stromausgang
I
passive
U
ext
I= 0/4 - 20 mA
15...20V DC 20...32 VDC
U
ext
Ri0...500 Ω 250...750 Ω
passive
DC voltage,
external power source (U
ext
),
note connection polarity
External voltage source (U
ext
),
DC or AC voltage,
connection polarity arbitrary
Active mode
The current output supplies the power for
operation of the outputs.
Passive mode
External power source required for operation of
the outputs.
Pulse output P
for electronic (EC) totalizers
with and without current output I
active (and current output I
active
)
U
I
max
U
≤ 32V DC/≤ 24 VAC
ext
I
≤ 150 mA
max
(incl. status output)
5
Status output S
Load
ext
passive
U
I
(incl. pulse output)
≤ 32V DC/≤ 24 VAC
ext
≤ 150 mA
max
U
≤ 15V DC from current output
int
Operation with current output:
I
≤ 3 mA
max
Operation without current output:
I
≤ 23 mA
max
15 V
R ≤
I
max
Status output S
6
with and without current output I
active
Load
U
≤ 15V DC
int
from current output
I
≤ 3 mA
max
Operation with
current output
I
≤ 23 mA
max
Operation without
current output
2/2
3 Start-up
3.1 Power-on and measurement
•Before powering the system, please check that it has been correctly installed according
to Sect. 1 and 2.
•The flowmeter is delivered ready for operational use. All operating data have
been factory set in accordance with your specifications.
Please refer to Sect. 3.2 “factory settings”.
•Power the unit, and the flowmeter will immediately start process flow
measurement.
Basic version
•A light emitting diode (LED) under the cover of the electronic section shows the
measurement status.
LED flashing . . .
•Refer to Sect. 6.1 for operator control of the “basic version”.
play version, signal converter IFC 010 / D
Dis
•When powered, the display shows in succession: START UP and READY.
This is followed by display of the current flow rate and/or the current totalizer count
on either a continuous or alternating basis, depending on the setting under Fct. 1.04.
•Refer to Sect. 4 and 5 for operator control of the “display version”.
, signal converter IFC 010 / B
green:measurement correct, everything all right.
green/red: momentary overdriving of outputs
and/or A/D converter.
red:fatal error, parameter error or hardware fault,
please consult factory.
3/1
Factory settings 3.2
All operating data are factory set according to your order specifications.
If you have not made any particular specifications at the time of ordering, the instruments will be
delivered with the standard parameters and functions listed in the Table below.
To facilitate easy and rapid initial start-up, current output and pulse output are set to process flow
measurement in “2 flow directions”, so that the current flowrate is displayed and the volumetric
flow counted independent of the flow direction. On instruments equipped with a display,
measured values may possibly be shown with a “ – ” sign.
This factory setting for the current and pulse outputs may possibly lead to measuring errors,
particularly in the case of volume flow counting:
for example, if pumps are switched off and a “backflow” occurs which is not within the range of
the low-flow cutoff (SMU), or if separate displays and counts are required for both flow directions.
To avoid faulty measurements, therefore, it may be necessary to change the factory setting of
some or all of the following functions:
– low-flow cutoff SMU, Fct. 1.03, Sect. 5.3
– current output I, Fct. 1.05, Sect. 5.6
– pulse output P, Fct. 1.06, Sect. 5.7
– display (option), Fct. 1.04, Sect. 5.4
Instrument o
peration:
Display versions:IFC 010 _/ D, operation: refer to Sect. 4 and 5.
Basic versions:IFC 010 _/ B, operation: refer to Sect. 6.1.
Standard factory settings
FunctionSetting
1.01 Full-scale range Q
1.02 Time constant3 s, for I, S
1.03 Low-flowON: 1 %
cutoff SMUOFF: 2 %
1.04 Display (option)
flow ratem
totalizer(s)m3 or US Gal
1.05 Current output I
function2 directions
range4 - 20 mA
error message22 mA
1.06 Pulse output P
function2 directions
pulse value1 pulse/s
pulse width50 ms
1.07 Status output Pflow direction
100%
see nameplate
and display
3
/hr or US Gal/min
FunctionSetting
3.01
Language for display only
3.02 Flowmeter
diametersee nameplate
flow direction (see arrow
on primary head)
3.04 Entry codeno
3.05 User unit
English
+ direction
}
Liter/hrorUSMGal/day
3/2
Teil B IFC 010 _/ D Signal converter
4Operation of the signal converter
4.1Krohne operator control concept
1 3 6. 4 9
m3/hr
→
Measuring mode
CodE 1
Menu columnFunction columnData column
3.00 INSTALL.
When this display appears, press following keys:
→ → → ↵ ↵ ↵ ↑ ↑ ↑
3.05 USER UNIT
3.04 ENTRY CODE
3.03 ZERO SET
3.02 FLOWMETER
3.01 LANGUAGE
2.02 HARDW. INFO
Direction of movement
4/1
2.00 TEST
1.00 OPERATION
↑
↓
2.01 TEST Q
1.07 IND. OUTP. S
1.06 PULS. OUTP. P
1.05 CUR. OUTP. I
1.04 DISPLAY
1.03 L.F. CUTOFF
1.02 TIMECONST.
1.01 FULL SCALE
→
↑
↵
see
↓
→
Sect.
↵
4.4
Operating and check elements 4.2
The controls are accessible after unscrewing the 4 screws and
removing the housing cover.
Display,1st line
Display,2nd line
Display,3rd line:arrows to identify display
Flowrate
Totalizer
Overrange
+totalizer
–totalizer
Σsum totalizer (+ and –)
Ioverranging, current output I
Poverranging, pulse output P
current flowrate
Keys for operator control of signal converter
Compass field, signals actuation of a key.
4/2
4.3 Function of keys
The cursor (flashing part of display) has a grey background in the following descriptions.
o start operator control
T
Measuring mode
Operator control mode
1 3 . 5 7 1
m 3 / h r
PLEASE NOTE: When “YES” is set under Fct. 3.04 ENTRY CODE, “CodE 1 - - - - - - - - -”
appears in the display after pressing the → key.
The 9-keystroke Entry Code 1 must now be entered: → → → ↵ ↵ ↵ ↑ ↑ ↑
(each keystroke acknowledged by “
o terminate operator control
T
Press key ↵ repeatedly until one of the following menus
Fct. 1.00 OPERATION, Fct. 2.00 TEST or Fct. 3.00 INSTALL. is displayed.
F c t . 3. 0 0
INSTALL.
→
”).
*
Press key
↵
↵
F c t . 1. 0 0
O P E R AT I O N
S T O R E Y E S
Store new parameters: acknowledge by
pressing key↵ . Measuring mode continued
with the new parameters.
New parameters not to be stored:
press key ↑ to display
„STORE.NO“.
Measuring mode continued with the „old“
↵
parameters after pressing key
.
4/3
To change numbers
increase number
3 9 7. 3 5
↑
m 3 / h r
decrease number
To shift cursor (flashing position)
shift to right
3 9 7. 3 5
→
m 3 / h r
shift to left
To alter texts (units)
For units, the numerical value is
converted automatically.
3. 7 6 9 9
select next text
↑
L i t e r / S e c
select preceding text
To transfer from text (unit) to number setting
Change to number setting
←
3 9 7. 4 5
↓
m 3 / h r
3 9 7. 3 5
m 3/hr
9 3. 3 6 5
↓
U S. G a l / m i n
1 3. 5 7 1
m 3 / h r
→
←
Return to text setting
To transfer to subfunction
Subfunctions do not have a “Fct.No.“ and are identified by a “ → ”.
↵
1 D I R.
To revert to function display
1 0. 3
↵
Sec
1 3. 5 7 1
m 3/hr
→
R A N G E I
F c t. 1. 02
T I M E C O N S T.
4/4
4.4 Table of settable functions
Abbreviations used
DNNominal size, meter size
F
F
F
F
F/RForward/reverse flow in F/R mode
GKLPrimary constant
ICurrent output
PPulse output
P
P
Highest frequency of the pulse output
max
Lowest frequency of the pulse output
min
Conversion factor volume for any unit,
M
see Fct. “FACT. VOL.”
Conversion factor time for any unit,
T
see Fct. 3.05 “FACT. Time”
= F
= F
max
min
/ Q
/ Q
100%
100%
max
min
Qactual flowrate
Q
100%
Q
max
Q
min
SStatus output, control input
SMULow-flow cutoff for I and P
vFlow velocity
v
max
v
min
Fct.TextDescription and settings
1.00OPERATIONOperations menu
1.01FULL SCALEFull-scale range for flowrate Q
Select unit
• m3/hr• Liter/Sec• US.Gal/min
• user unit, factory set is “Liter/hr” or “US MGal/day” (see Fct. 3.05)
Press →key to transfer to number setting.
Setting ranges
The ranges are dependent on the meter size (DN) and the flow