KEYENCE FS-N10 User Manual

Page 1
96141E
Digital Fiber Sensor
FS-N10 Series
User's Manual
Read this manual before use. Keep this manual in a safe place for future reference.
Page 2
Introduction
Symbols
The following symbols alert you to matters concerning the prevention of injury and product damage.
Provides additional information on proper operations.
Provides advanced and useful information for operation.
Failure to follow the instructions may lead to death or serious injury.
Failure to follow the instructions may lead to injury.
Failure to follow the instructions may lead to product damage or failure of the product.
CAUTION
DANGER
WARNING
Reference
Reference
Point
Page 3

Safety Precautions

WARNING
Hints on Correct Use
• Do not use this product in saf ety circuits for human pr otection.
• This product is not ex plosion-proof. Do not use the product in places with flammable gas, liquid, or dust.
• This product uses DC power. Do not apply AC power. The product may explode or burn if an AC voltage is applied.
•Do not wire the amplifier line along with power lines or high-tension lines, as the sensor may malfunction or be damaged due to noi se.
• When using a commercially a vailable switching regulator, ground the frame ground terminal and ground terminal.
•Do not use the FS series outdoors, or in a place where extraneous light can enter the light receiving element directly.
•Due to individual dispersion characteristics and the difference in fiber unit models, the maximum sensing distance or displayed value may not be the same on all units.
• If the sensor is used for a long time with the APC function enabled and the LED is imposed with a heavy load, the current consumption of the sensor for light emission will become constant and 'END APC' will be displayed. The sensor can still be use received light intensity should be detected for precise detection.
About UL Certification
The FS-N Series is UL and C-UL certified, and is compliant with the UL and CSA standards.
•Applicable standardsUL508 Industrial Control Equipment
•UL File No. E301717
• UL categories: NRKH, NRKH7
•Enclosure Type 1 (based on UL50 standard)
Notes on UL Certification
•The power source used with the FS-N Series must be UL Listing certified f or Class 2 output as stipulated by US National Electric Code (NEC) NFPA70.
•Power supply/C ontrol input/Control output shall be connected to a single Class 2 source only.
• Connect the FS-N12
•Use with the over current protection device which is rated 30V or more and not more than 1A.
d in this case. However, replace the sen
CAN/CSA C22.2 No.14-M05 Industrial Control Equipment
and N14 to the F
S-N11 and N13 main units for use.
sor if even small changes in
96141E
1
Page 4
Safety Precautions
MEMO
2
- Digital Fiber Sensor FS-N10 Series User's Manual -
Page 5

Manual Organization

1 2 3 4 5 6
Before Using
Installation and Connection
Basic Operation
Settings for Advanced Functions
Specifications
Appendix
Outlines the package contents and identifies part names and functions.
1
Provides procedures for installing sensor amplifiers and cables, as well as operating precautions.
2
Explains basic instructions for operating and setting the sensor amplifiers.
3
Describes settings for advanced functions of the FS-N10 Series.
4
Provides the specifications, circuit diagrams and dimensions of the FS-N10 Series.
5
Provides the troubleshooting instructions and initial settings (default values).
6
<Points for Using This Manual>
When you
cedures"
When you
When you
When you
When you
"Forgot the operation methods"
"Want to try out the FS-N10"
"Want to fully utilize the vari ous functi ons"
"Want to know the meanings of terms used"
"Want to troubleshoot the FS-N10"
- Digital Fiber Sensor FS-N10 Series User's Manual -
or
"Want to find the operation pro-
Go to pages 3-2, 4-2
Go to Chapters 2 and 3
Go to Chapters 3 and 4
Go to Chapter 6 (Index)
Go to Chapter 6 (Troubleshooting)
3
Page 6

Table of Contents

Safety Precautions...............................................................................1
Manual Organization ...........................................................................3
Table of Contents.................................................................................4
Chapter 1 Before Using
1-1 Checking the Package Contents..........................................................1-2
Sensor Amplifier.............................................................................. 1-2
List of Optional Parts....................................................................... 1-2
1-2 Part Names..........................................................................................1-3
Sensor Amplifier.............................................................................. 1-3
Display/control unit..........................................................................1-4
1-3 Model Number Description ..................................................................1-5
1-4 Fiber Units............................................................................................1-6
Chapter 2 Installation and Connection
2-1 Installing Sensor Amplifiers..................................................................2-2
Mounting the Sensor Amplifier........................................................2-2
Wiring Diagrams for Sensor Amplifiers ...........................................2-4
2-2 Connecting the Fiber Unit.................................................................... 2-6
Chapter 3 Basic Operation
3-1 Quick Reference..................................................................................3-2
3-2 Switching Output.................................................................................. 3-4
Output Switch (L-on/D-on) ..............................................................3-4
3-3 Adjusting Sensitivity.............................................................................3-5
List of Sensitivity Adjusting Methods...............................................3-5
Preset Function...............................................................................3-6
Work-Preset Function ..................................................................... 3-7
Maximum Sensitivity Preset Function.............................................3-8
Full Auto Preset Function................................................................ 3-9
Two-point Calibration.....................................................................3-11
Maximum Sensitivity Calibration...................................................3-12
Full Auto Calibration......................................................................3-13
Positioning Calibration...................................................................3-14
Other Calibration Methods ............................................................3-15
3-4 Setting the Current Received Light Intensity to 0 (Zero Shift)............ 3-16
Zero Shift Funct
Operating Principle of the Zero Shift Function.............................. 3-16
4
- Digital Fiber Sensor FS-N10 Series User's Manual -
ion.........
............................................................... 3-16
Page 7
Table of Contents
3-5
Light emission/Received light intensity adjustment (Saturation Canceling)......
Saturation Canceling function.......................................................3-18
3-6 Loading the Recommended Settings (Recipe Function) ...................3-19
Selecting Recipe ...........................................................................3-19
List of Recipes and Recommended Fiber Units............................3-20
3-7 Initialization ........................................................................................3-21
Initialization of Settings (Reset to Initial Values) ........................... 3-21
3-8 Locking in MEGA Mode ..................................................................... 3-22
MEGA Mode Lock (1-Output Type Only)............................ ...........3-22
3-9 Disabling the Key Operation .............................................................. 3-23
Key Lock........................................................................................3-23
Key Lock with PIN Number ...........................................................3-24
Chapter 4 Settings for Advanced Functions
4-1 List of Settings .....................................................................................4-2
4-2 Basic Settings............ ... ................................................................. ......4-4
Power Modes .................................................................... ..............4-4
Sensitivity Setting............................................................................4-4
4-3 Detection Settings (Func)....................................................................4-7
Output Timer.............. ... .................................................................. 4 - 7
Detection Mode... ................................................................. ...........4-8
External Input................................................................................4-17
Writing of External Input to EEPROM................................ ...........4-19
Adjusting Light Emission/Received Light Intensity (Attenuation)
Analog output scaling (FS-N11MN only).............. ... ......................4-21
Analog scaling mode (FS-N11MN only)........................................4-22
4-4 Display Settings (diSP) . .. ... ................................................................ 4-2 3
Display Reverse ............................................................................4-23
Sub Display...................................................................................4-23
Preset Saturation Function............................................................4-28
4-5 System Settings (SYS) .................................................................. ....4-3 0
APC function................................................................................. 4-30
Power Save................................................................................... 4-31
Display Gain............................................................... ...................4-32
Interference Prevention................................................................. 4-33
Common Key-Oper ations Function...............................................4-34
4-6 2-output Settings ( ).............................................................4-35
Detection Mode for Output2..........................................................4-35
Output Timer for Output2 .............................................................. 4-38
4-7 Settings Save/Recall.......................................................................... 4-39
Custom Save (Settings Save) .......................................................4-39
User Reset (Settings Recall).........................................................4-40
3-18
...4-20
- Digital Fiber Sensor FS-N10 Series User's Manual -
5
Page 8
Table of Contents
Chapter 5 Specifications
5-1 Specifications.......................................................................................5-2
1-Output Type.................................................................................. 5-2
2-Output Type.................................................................................. 5-3
Monitor Output Type........................................................................5-4
Zero Line Type ................................................................................5-5
5-2 Input/Output Circuit Diagram ...............................................................5-6
1-Output Type.................................................................................. 5-6
2-Output Type.................................................................................. 5-6
Monitor output type ......................................................................... 5-7
5-3 Dimensions..........................................................................................5-8
Chapter 6 Appendix
6-1 Troubleshooting....................................................................................6-2
Frequently Asked Questions ...........................................................6-2
Error Displays and Corrective Actions.............................................6-5
6-2 Factory Default Setting (Default Value) List .........................................6-6
6-3 List of Recipe Function Settings ..........................................................6-7
6-4 Restrictions on Each Detection Mode..................................................6-9
Restrictions for Sensitivity Settings in Each Detection Mode..........6-9
6-5 Index..................................................................................................6-10
6
- Digital Fiber Sensor FS-N10 Series User's Manual -
Page 9

Before Using

This chapter outlines the package contents and identifies part names and functions.
1-1 Checking the Package Contents ......... ... ............1-2
1-2 Part Names ........................................................1-3
1-3 Model Number Description................................ . 1-5
1-4 Fiber Units..........................................................1-6
1
- Digital Fiber Sensor FS-N10 Series User's Manual -
1-1
Page 10
1
Reference
1-1

Checking the Package Contents

Before using the unit, make sure that the following equipment and accessories are included in the package. We have thoroughly inspected the package contents before shipment. However, in the event of defective or broken items, contact your nearest KEYENCE office.
Before Using
The FS-N10 Series are amplifier units. Each amplifier unit m ust be used with a separately sold fiber unit. Select a fiber unit that suits the intended application.
"Fiber Units" (page 1-6)

Sensor Amplifier

FS-N10 Series
Sensor amplifier x 1
Instruction Manual x 1

List of Optional Parts

OP-73880
Amplifier mounting bracket (for main unit) x 1 For information on how to use, refer to "Installing Sensor Amplifiers" (page 2-2).
OP-73864
M8 connector cable (2 m) x 1 Input/Output cables needed for using the M8 connector type.
OP-26751
End unit x 2 For information on how to use, refer to "Installing Sensor Amplifiers" (page 2-2)
OP-73865
M8 connector cable (10m) x 1 Input/Output cables needed for using the M8 connector type.
Fiber cutter (OP-87098) x 1
OP-87098
Fiber cutter x 1 For information on how to use, refer to "Connecting the Fiber
.
Unit" (page 2-6).
1-2
- Digital Fiber Sensor FS-N10 Series User's Manual -
Page 11
1-2
T R
Connection Cable
*3
Expansion connector
*1
Display/control unit
Expansion connector
*1*2
Fiber lock lever
Dust cover
Fiber transmit / receive indicators
Fiber connection ports
(FS-N1غ/N1غP)
M8 connector type (FS-N1
غ
CN/N1غCP)
e-CON connector (FS-N1
غ
EN)

Part Names

Sensor Amplifier

1
Before Using
*1 Expansion protective cover is installed at the factory prior to shipment. *2 Not provided on the main unit (FS-N11 □ /N13 ).
*3 No connecting cables are provided for the Zero Line type (FS-N10).
- Digital Fiber Sensor FS-N10 Series User's Manual -
1-3
Page 12
1-2 Part Names
12

Display/control unit

1
1-output type (FS-N11/N12/N11C/N12C/N11EN/N12EN/N11MN/N10)
Before Using
-1
-2
-1
-2
-2
-1
Item Description
Operation indicator (1-output/Zero line type)
Operation indicator (2-output type)
[SET] button Set value display
(green display) Current value display
(red display) Manual button Used to adjust the setting value or select an option.
elect switch
Power s (1-output type)
Channel toggle switch (2-output type)
[PRESET] button
[MODE] button
DTM indicator
PST indicator
*
*
Indicates the current output (detection) status. Indicates the current output (detection) status of channels 1 and 2
separately. Use when setting sensitivity, etc.
"Adjusting Sensitivity" (page 3-5)
Displays a sett ing value or advanced setting item in this area of 7­segment green indicators.
Displays the current value (received light intensity) and selection items for detailed settings with red 7-seg ment display.
Changing power modes. SEL: Allows you to set a power mode using the "Changing Power
M: Fixes the power mode to the "MEGA mode".
"Locking in MEGA Mode" (page 3-22)
Toggles between channels 1 and 2 for configuring the received light intensity display or sensitivity setting.
Used for presetting or setting values or parameters.
"Preset Function" (page 3-6)
Used for toggling L-on/D-on, proceeding to advanced settings, or confirming selections.
Lights when a DATUM mode is in effect.
"DATUM1 mode" (page 4-9) "DATUM2 mode" (page 4-11)
Lights when preset value is set.
"Preset Function" (page 3-6)
Modes" function of basic setup.
-1
-2
*Not provided on the Zero Line type (FS-N10).
1-4
- Digital Fiber Sensor FS-N10 Series User's Manual -
Page 13
1-3
FS-N1
(1) (2) (3) (4)
The numbers and letters used in product names are explained below:

Model Number Description

1
(1) Amplifier type
0: Expansion unit (Zero line) 1: Main unit (1-output) 2: Expansion unit (1-output) 3: Main unit (2-output) 4: Expansion unit (2-output)
(3) Cable type
(None): 2-meter cable C : M8 connector E: e-CON connector
Model (1) Amplifier type
FS-N11N FS-N11P FS-N12N FS-N12P FS-N11CN FS-N11CP FS-N12CN FS-N12CP FS-N11EN FS-N12EN FS-N11MN
FS-N13N FS-N13P FS-N14N FS-N14P FS-N13CP FS-N14CP FS-N10
Main unit (1-output) Main unit (1-output) Expansion unit (1-output) Expansion unit (1-output) Main unit (1-output) Main unit (1-output) Expansion unit (1-output) Expansion unit (1-output) Main unit (1-output) Expansion unit (1-output) Main unit (1-output)
Main unit (2-output) Main unit (2-output) Expansion unit (2-output) Expansion unit (2-output) Main unit (2-output) Expansion unit (2-output) Expansion unit ( Zero line)
(2) Special type
(None): Standard M : Monitor output
(4) Output type
N: NPN P: PNP
(2) Special type
Standard
Monitor output
Standard
−−
(3) Cable type
2-meter cable
M8 connector
e-CON connector
2-meter cable
2-meter cable
M8 connector PNP output
(4) Output
type
NPN output PNP output NPN output PNP output NPN output PNP output NPN output PNP output
NPN output
NPN output
NPN output PNP output NPN output
PNP output
Before Using
- Digital Fiber Sensor FS-N10 Series User's Manual -
1-5
Page 14
1
Reference
1-4
The FS-N10 Series are amplifier units. They must be used in combination with separately-sold fiber units.
Before Using
Integrated bracket

Fiber Units

For fiber unit details, refer to the KEYENCE gener al catalog or contact y our nearest KEYENCE office.
Sleeve Bracket Threaded Cylinder
Narrow beam/ high power
High-flex Oil/Chemical
Retro-reflectiveArea Small spot
resistant
reflective
Liquid level LCD,
semiconductor
Definite reflective
1-6
- Digital Fiber Sensor FS-N10 Series User's Manual -
Page 15

Installation and Connection

This chapter provides procedures for installing sensor amplifiers and cables, as well as operating precautions.
2-1 Installing Sensor Amplifiers................................2-2
2-2
Connecting the Fiber Unit
.........................................2-6
2
- Digital Fiber Sensor FS-N10 Series User's Manual -
2-1
Page 16
2
(3)
(2)
(1)
Reference
WARNING
Installation and Connection
2-1

Installing Sensor Amplifiers

Mounting the Sensor Amplifier

Mounting on a DIN rail
Align the claw at the bottom of the main
1
body with the DIN rail, as shown on the right. While pushing the main body in the direc­tion of the arrow (1), push down in the direction of arrow (2).
T o release the ampl ifier, raise the amplifier
2
body in the direction of arrow (3) while pushing in the direction of arrow (1).
Installation on a wall
This method applies only when using the main unit independently. If the main unit is connected with an e xpan sion unit(s), use the method of mount­ing on a DIN rail.
Mount the amplifier on the amplifier mount-
1
ing bracket (OP-73880, sold separately), using the same manner as "Mounting on a DIN rail".
Secure the unit with two M3 screws as
2
shown in the illustration.
Connecting multiple amplifiers
Up to 16 expansion units can be connected to 1 main unit. Note, however, that the 2-output type is considered as 2 expansion units.
Mount on DIN rail and install on metal surface when connecting multiple amplifiers or mounting main units together.
2-2
- Digital Fiber Sensor FS-N10 Series User's Manual -
OP-73880
Page 17
• Contact your nearest KEYENCE office when connecting a unit
Point
Main unit Expansion unit
(3)
(2)
(1)
other than the N-bus (KEYENCE’s wire-saving system) compatible sensor amplifier, including FS-N10 Series, or the NU Serie s com­munication module.
• Turn the power off before connecting multiple expansion units.
• Do not touch the expansion conn ector with your bare hands.
• When using the FS-N10 Series as a main unit, use the products within the expansion unit’s power voltage range if the power volt­age range of the expansion unit is narrower than the FS-N10 Series.
Remove the protection covers from the main unit
1
and expansion unit(s).
Mount the main unit and expansion unit(s) on the
2
DIN rail.
2-1 Installing Sensor Amplifiers
2
Installation and Connection
3
4 5
Slide the main unit and expansion unit(s) together. Engage the 2 cla ws of the expansion unit with the recesses on the main unit side until you hear/ feel a click.
Attach the separately sold end units (OP-26751: a set of 2 units) to the DIN rail in the same manner as step (2) (Tightening torque: 0.6N•m or less).
Secure the amplifiers between the end units. Tighten the screws at the top (2 screws × 2 units) with a Phillips screwdriver to fix the end units.
- Digital Fiber Sensor FS-N10 Series User's Manual -
End unit
End unit
2-3
Page 18
2
Point
Brown
*
Output
Blue
*
Black
12 to 24 VDC
* FS-N11N/N11P only
Brown
*
Blue
*
Black
12 to 24 VDC
* FS-N13N/N13P only
Output1
External input
Output2
White Pink
Brown
Blue
12 to 24 VDC
* Connect to a device having an input impedance 10 kΩ or more.
Black
Output
Monitor output
1 to 5V
Orange
*
2-1 Installing Sensor Amplifiers

Wiring Diagrams for Sensor Amplifiers

· Be sure to turn off the power before wiring.
· Insulate each input or output cable that will not be used.
Installation and Connection
Wiring Diagrams for Cable Types
1-output type (FS-N11/N12)
2-output type (FS-N13/N14)
Monitor output type (FS-N11MN)
2-4
- Digital Fiber Sensor FS-N10 Series User's Manual -
Page 19
Wiring Diagrams for M8/e-CON Connector Types
12 to 24 VDC
* FS-N11CN/N11CP only
External input
Output
1
2
3
4
M8 connector Pin layout
12 to 24 VDC
* FS-N11EN only
External input
Output
e-CON connector Pin layout
1
2
3
4
12 to 24 VDC
*FS-N13CP only
Output1
Output2
1
2
3
4
M8 connector Pin layout
M8 connector, 1-output type (FS-N11C/N12C)
Pins and wire colors of OP-73864/OP-73865
2-1 Installing Sensor Amplifiers
2
Installation and Connection
Connected pin No.
e-CON connector, 1-output type (FS-N1 1EN/N12EN)
M8 connector, 2-output type (FS-N13CP/N 14 CP )
Pins and wire colors of OP-73864/OP-73865
Connected pin No.
Wire color
Brown
White
Blue
Black
Wire color
Brown
White
Blue
Black
Function
12 to 24 VDC External input
0V
Output
Function
12 to 24 VDC
Output2
0V
Output1
- Digital Fiber Sensor FS-N10 Series User's Manual -
2-5
Page 20
2
Fiber
(2)
(1)
Always insert fiber from the side with writing
.
Fiber cutter (OP-87098)
Point
2-2
This section provides procedures for connecting the fiber unit and operating precautions.
1

Connecting the Fiber Unit

Install the fiber unit.
After the installation, check that the transmitter and receiver of the sensor are spaced properly. Also check that the optical axis is aligned.
Installation and Connection
If the fiber unit is a free cut type, cut any excess length of the fiber.
2
(1) Insert the fiber into the hole in the cutter. (2) Bring down the blade in a single, swift motion to cut the fiber.
Failure to observe the cautions belo w may resul t in an impr operly cut surface, which may reduce the detection distance.
· When cutting the fiber unit, be sure to use a gray fiber cutter (OP-87098)
· Cut the fiber unit in a single motion without stop ping the b lade halfway through.
· Do not use the same hole twice.
Open the dust cover in the
3
direction of arrow (1).
Move the fiber lock lever down in
4
the direction of arrow (2).
2-6
- Digital Fiber Sensor FS-N10 Series User's Manual -
(1)
(3)
(2)
Page 21
2-2 Connecting the Fiber Unit
Point
Cable outer dia
1.3
1.0
Adapter
AdapterA
(OP-26500)
AdapterB
(OP-26501)
Appearance
Receiver
Position of fiber transmit / receive indicators
Transmitter
Single-core fiber
Multi-core fiber
T
R
Insert the fiber unit into the fiber connection ports on the amplifier.
5
Insert the fiber unit until it reaches the position of the transmit / receive indicators (approximately 14 mm).
Fiber
2
Fiber connection ports
Move the fiber lock lever ba ck in the direction of arrow (3) to secure the
6
fiber.
· If a thin fiber unit is used, an adapter provided wi th the thin fiber unit will be required. Make sure to use the adapter that matches the thin fiber unit.
· To connect a coaxial refl ective fiber unit to the ampl ifier, connect the single-core fiber to the transmitter (T) side, and connect the multiple-core fiber to the receiver side (R).
Fiber transmit / receive indicators
Installation and Connection
- Digital Fiber Sensor FS-N10 Series User's Manual -
2-7
Page 22
2
2-2 Connecting the Fiber Unit
MEMO
Installation and Connection
2-8
- Digital Fiber Sensor FS-N10 Series User's Manual -
Page 23

Basic Operation

This chapter explains basic instructions for operating and setting the sensor amplifier.
3-1 Quick Reference.................................................3-2
3-2 Switching Output................................................3-4
3-3 Adjusting Sensitivity........................................... 3-5
3-4 Setting the Current Received Light Intensity
to 0 (Zero Shift) ................................... ... ... .......3-16
3-5 Light emission/Received light intensity
adjustment (Saturation Canceling)................... 3-18
3-6 Loading the Recommended Settings
(Recipe Function)............................................. 3-19
3-7 Initialization.......................................................3-21
3-8 Locking in MEGA Mode....................................3-22
3-9 Disabling the Key Operation...... ... .. ... ...............3-23
3
- Digital Fiber Sensor FS-N10 Series User's Manual -
3-1
Page 24
3-1
Manual button
[MODE] button
Power select switch (1-output type))
[PRESET] button
[SET] button
Operation indicator
(1-output/Zero line type)
DTM indicator
PST indicator
Operation
indicator
(2-output type)
Channel select switch (2-output type)

Quick Reference

The main setting operations are explained according to purpose. Refer to Chapter 4 for information on advanced function settings and explanations not given below.
3
Basic Operation
Purpose Description Operation procedures
Swiching the
output
Adjusting the sensitivity and integrating the
display to "100.0"
and ".0"
Adjusting the
sensitivity
Shifting the
received light
intensity to "0"
Need to prevent
the received light
intensity from
becoming satu-
rated.
3-2
1Swiching the output. (L-on/D-on)
Set the current received light intensity
2
to "100.0". (Preset) When preset is valid, register the
received light intensity ".0".
3
(Work-preset)
Set the received light intensity slightly higher than when the setting was
4
made, to "100.0". (Maximum sensitivity preset)
Automatically register "100.0" and ".0" when workpiece passes by. (Full Auto
5
preset)
6 Cancel the various preset functions. Press and hold the [PRESET] button. 3-6
Set the setting value at the midpoint between the received light intensity
7
values when a workpiece is present and absent. (2-point calibration)
Set the setting value slightly higher than the received light intensity value
8
at which the setting was made. (Maxi­mum sensitivity calibration)
Set the setting value automatically
9
when a workpiece is passing thro ugh. (Full auto calibration)
Set the setting value to the base point
10
where the workpiece is positioned. (Positioning calibration)
11 Finely adjust the setting value directly.
Set the current display to "0". (Zero
12
shift)
13 Cancel the zero shift function. Press and hold the [PRESET] button. 3-16
Automatically adjusts to appropriate
14
light emission and light receiving sen-
sitivity.
- Digital Fiber Sensor FS-N10 Series User's Manual -
*
*
*
1. Press the [MODE] button.
2. Switch with the ( ) button. Press the [PRESET] button while the PST
indicator is OFF.
After step 7, press the [PR ESET] button + button in the state to be set as ".0".
While the PST indicator is OFF, press and hold the [PRESET] button. Reflective model: When no workpiece is present. Thrubeam/Retro­reflective model: When a workpiece is present.
Press and hold the [PRESET] button while the PST indicator is OFF.
1. Press the [SET] button once when a work- piece is present.
2. Press the [SET] button once when no work- piece is present.
Reflective model: Press and hold the [SET] but­ton when no workpiece is present. Thrubeam/ Retro-reflective mo [SET] butt
on when a workpiece is present.
Press and hold the [SET] button while the workpiece passes through.
1. Press the [SET] button once when no work­piece is present.
2. Press and hold the [SET] button at the posi­tioning point.
Press the ( ) button. Press the [PRESET] button + button when
t
he PST indicator
Briefly press [MODE] + [SET] buttons with the amount of received light in maximum-saturated condition.
12
del: Press and hold the
is OFF.
Reference
page
3-4
3-6
3-7
3-8
3-9
3-11
3-12
3-13
3-14
1-4
3-16
3-18
Page 25
Purpose Description Operation procedures
1. Press and hold the [SET] button and [PRE­SET] button.
Loading the
recommended set-
tings
Initializing the set-
tings
Displaying the out-
put 2 display
screen with the
2-output type
Switching to the
maximum received
light intensity
power mode
Preventing
incorrect
operations
Others
(Advanced
function settings,
etc.)
Saving and loading the
settings
Load the recommended settings.
15
(Recipe function)
Initializing (Restore to factory default
16
settings)
17 Switch to the output 2 display screen.
Adjust the power mode to the MEGA
18
mode.
19 Activat
20 Deactivating the key lock
21
22
23 Setting the advanc
24
25
26
27 Saving the settings (custom save)
28 Loading the settings (user reset)
ing the key lock
Activating the password-protected k ey lock
Deactivating the password-protected key lock
ed functions Press an
Setting to rescale at each preset exe­cution so that analog output is "5 V"
output in respect to "100.0".
Switching the display to extended dis­play or received light intensity hold display, etc. (sub-display)
Resetting the following values * Received light intensity hold value * Excess gain hold value * Output when output 2 is in limit set­ting detection mode
*
2. Display the LoAd screen with the ( ) button, and press the [MODE] button.
3. Select the recipe such as r-1 FALL with the () button.
4. Press the [MODE] button to execute.
1. Press and hold the [SET] button and [PRE-
SET] button.
2. Press the [MODE] button while on the rSt
screen.
3. Select init with the ( ) button.
4. Press the [MODE] button to execute.
Set the channel switch to . * The output 2 sensitivity and advanced set­tings can be modified in this state.
Set the power select s witch to .
Pr
ess and hold the [MODE] button and the
() button simultaneously. Press and hold the [MODE] button and the
() button simultaneously.
1. Press the ( ) button10 times while
holding down the [MODE] button.
2. Input the password with the ( ) button.
3. Press the [MODE] button to execute.
1. Press the ( ) button 10 times while
holding down the [MODE] button.
2. Input the password with the ( ) button.
3. Press the [MODE] button to deactivate the
key lock.
d hold the [MODE] button. 4-1
1. Press and hold the [MODE] button, but-
ton and [SET] button simultaneously.
2. Press the [MODE] button once or twice.
3. Select "Pr-A PrST" with ( ) button, and
press the [MODE] button. After setting the sub-display with the advanced
function settings, press the [MODE] button twice.
Press and hold the [MODE] button and [SET] button.
1. Press and hold
SET] button.
2. Display the SAvE screen with the ( )
button, and press the [MODE] button.
3. Select yES with the ( ) button.
4. Press the [MODE] button to execute.
1. Press and hold the [SET] button and [PRE-
SET] button.
2. Press the [MODE] button on the rSt screen.
3. Select USEr with the ( ) button.
4. Press the [MODE] button to execute.
the [SET] button and [PRE-
3-1 Quick Reference
Reference
page
3-19
3-21
4-35
3-22
3-23
3-23
3-24
3-24
4-22
4-23
4-25 4-36
4-39
4-40
3
Basic Operation
* Not available for the FS-N10 Series shipped before March 10, 2011.
- Digital Fiber Sensor FS-N10 Series User's Manual -
3-3
Page 26
3-2
Point

Switching Output

Output Switch (L-on/D-on)

This function configures when the output turns ON.
When the current received light inten-
1
sity is displayed, press the [MODE] but­ton.
3
Basic Operation
The current output condition (L-on or D­on) is displayed.
Press the button to switch the out-
2
put condition, and then press the [MODE] button.
Select "D-on" if you want to output the ON signal when the beam is blocked (a work­piece is present.) for a thrubeam or retro-reflective model. Select "L-on" if you want to output the ON signal when the beam is received (a workpiece is present.) for the reflective model.
The output condition is switched, and the current received light intensity is dis-
*2
played.
*1 If you do nothing for 3 seconds or more or press the [MODE] button, the received
light intensity display is automatically restored.
*2 When using the sub-display, the screen will switch between the current received
light intensity L-on/D-on screen  sub-display  current received light intensity and so forth each time the [MODE] button is pressed.
"Sub Display" (page 4-23)
• Area detect mode, rising edge detect mode, and falling edge detect mode are selectable by n.o./n.c. "Area detection mode" (page 4-14) "Edge detection mode" (page 4-16)
• When using a 2-output type, output can be individually selected with channel 1/2.
*1
3-4
- Digital Fiber Sensor FS-N10 Series User's Manual -
Page 27
3-3

Adjusting Sensitivity

This manual refers to the values that select ON/OFF output of the sensor amplifiers as "Set values." In addition, the manual refers to the adjustment of set values as "Adjusting Sensitivity." This section describes the sensitivity adjusting method.

List of Sensitivity Adjusting Methods

Sensitivity adjusting method for the FS-N10 Series is broadly divided into two categories:
Preset Simple operation allows sensitivity adjustment concurrently with correction of the received light intensity to "100.0" or "0.0" This function helps preventive maintenance by eliminating dispersion of the received light intensity due to the contents of detection or individual workpieces. However, this is not suitable for detection of transparent workpiece because there is a small difference in the received light intensity between presence and absenc e of a workpiece.
Calibration Sensitivity can be adjusted by a simple operation. This function does not correct t he received light intensity. This function is applied for using the received light intensity without correction or for high-precision detection. Calibration is also available in a preset state.
How to select sensitivity adjusting method
3
Basic Operation
Sensitivity
adjusting
method
Preset
Calibration
Basics
At times like this
Basic
At times like this
Status of use Functions Contents Reference
Using thrubeam/retro-reflective models
Using reflective model
Not successfully displaying the intensity level as "100.0" and ".0" depending on the presence of a workpiece.
Mobile object moves fast
ing thrubeam/retro-reflective
Us models
Mobile body moves fast Using the unit in the environ-
ment that tends to get dirty easily.
Using the unit for positioning Using the unit for
sion detection
high-preci-
Preset
Preset to maxi­mum sensitivity
Work preset
Full automatic preset
Two-point calibration
Full automatic calibration
Maximum sensi­tivity calibration
Positioning cali­bration
Percentage cali­bration
When there is no workpiece, by sim­ply pressing the PRESET button, sensitivity adjustment is completed.
When there is no workpiece, by sim­ply holding the PRESET button, sensitivity adjustment is completed.
The status in which "100.0" or ".0" is displayed can be set at your choice
Preset is enabled by using a f ast-
rkpiece.
moving wo The setting can be established just
by pressing the [SET] button once when a workpiece is present and
absent. Calibration is enabled by using a
fast-moving workpiece. The setting seldom causes malfunc-
tion even in the environment that tends to get dirty easily.
The setting suitable f available.
Effective for making correction from external devices such as PLC.
or positioning is
- Digital Fiber Sensor FS-N10 Series User's Manual -
3-6
3-8
3-7
3-9
3-11
3-13
3-12
3-14
3-15
3-5
Page 28
3
Green PST lights up
Setting value is
"50.0"
Current value is "
100.0"
Workpiece
3-3 Adjusting Sensitivity

Preset Function

Enabling the preset function
When the PST indicator is not lit, press the [PRESET] button.
The PST indicator lights in green. The current value is set to "100.0" and the setting value is set to "50.0".
Basic Operation
Disabling the preset function
When the PST indicator is lit, press and hold the [PRESET] button.
• The PST indicator turns off, indicating that the preset function has been dis­abled.
• Once the preset function is disabled, the setting value is recalculated while retaining the ratio of the setting value and received light intensity.
3-6
- Digital Fiber Sensor FS-N10 Series User's Manual -
Page 29
3-3 Adjusting Sensitivity
Impotant
100.0
PRESET+
3.0
.0

Work-Preset Function

This function calibrates the current value to ".0". After the preset function has been executed with "100.0" displayed, and then executing this function wit h ".0" displayed, 2 random set points can be calibrated to "100.0" and ".0". * Not available for the FS-N10 Series shipped before March 10, 2011.
Enabling the work-preset function
The work-preset function can be used with the preset function (when preset is enabled).
While the preset function is enabled, press the [PRESET] button and button simultaneously.
The received light intensity at that point is set to ".0". The value set to "100.0" using the preset function does not change.
Green "PST" lights up
Current value is ".0"
Reference
Even if the received light intensity is low during preset and is high during work-preset, the value is set to "100.0" during preset and ".0" during work-
preset. When the actual received light intensity increases, the display will approach ".0". (The preset saturation level is decreased with respect to "100.0".)
3
Basic Operation
Disabling the work-preset function
When the PST indicator is lit, press and hold the [PRESET] button.
• The PST indicator turns off, indicating that the work-preset function has been disabled.
- Digital Fiber Sensor FS-N10 Series User's Manual -
3-7
Page 30
3-3 Adjusting Sensitivity
Workpiece
Setting value
Setting value
Thrubeam Model/Retro-reflective Model
Reflective Model
Received light intensity Received light intensity
No workpiece
With workpiece
No workpiece*
With workpiece*
1
1
Saturation point
"100.0"
"50.0"
".0"
"100.0" "50.0" ".0"
Saturation point
* Where detection occurs on a target having a background, the maximum sensitiv-
ity setting ignores the background. Maximum sensitivity setting is not available if the background is more reflective than the workpiece.
Point

Maximum Sensitivity Preset Function

This function calibrates the reference state to ".0" and the state at which the received light intensity is slightly higher as "100.0". This function is useful with the reflective model to detect while using the background as a reference. * Not available for the FS-N10 Series shipped before March 10, 2011.
Enabling the maximum sensitivity preset function
3
Basic Operation
When no workpiece is present for the
1
reflective model, or when a workpiece is present for thrubeam model/retro­reflective model, press and hold the [PRESET] button for 3 seconds or more while the PST indicator is OFF. Release the button when "
Calibration is complete after the setting value flashes momentarily, and then stops
(lights up). The PST indicator lights in green, and the setting value is set to "
50.0
".
• When th e amount of the light received is in saturated state (at a value higher than at the time of extended display on page 4-30), maximum sensitivity preset function cannot be implemented. ("---- ----" is indicated when operating 1.)
"Light emission/Received light intensity adjustment (Saturation
Canceling)" (page 3-18)
Auto
" flashes.
Disabling the maximum sensitivity preset function
When the PST indicator is lit, press and hold the [PRESET] button.
• The PST indicator turns off, indicating that the maxi mum sensitivity preset func- tion has been disabled.
3-8
- Digital Fiber Sensor FS-N10 Series User's Manual -
Page 31
3-3 Adjusting Sensitivity
Setting value
MAX
MIN
Time
Received
light intensity
The area near the maximum value of the received light intensity while the [PRESET] button is pressed is set as "100.0", and the area near the minimum value is set as ".0".
1
2
"100.0"
"50.0"
".0"
Point

Full Auto Preset Function

This function automatically judges 2 states (workpiece presence/absence, etc.), and calibrates the current value to "100.0" and ".0". This is useful when the detector is moving at high speed, etc. * Not available for the FS-N10 Series shipped before March 10, 2011.
Enabling the full auto preset function
When the PST indicator is OFF, con-
1
tinue pressing the [PRESET] button until "Auto" flashes while the work­piece passes through.
Workpiece
After the workpiece has completely
2
passed through, release the [PRESET] button.
Calibration is complete after the setting value flashes momentarily, and then
stops (lights up). The PST indicator lights in green, and the setting value is set to "50.0".
3
Basic Operation
• When the received light intensity is in a saturate d conditio n (more than the amount specified in "Display Gain" (page 4-32).), full-auto­matic preset function is disabled. ("---- ----" will appear during step 2.)
"Light emission/Received light intensity adjustment (Satura tio n
Canceling)" (page 3-18)
- Digital Fiber Sensor FS-N10 Series User's Manual -
3-9
Page 32
3
Point
Reference
3-3 Adjusting Sensitivity
Disabling the full auto preset function
When the PST indicator is lit, press and hold the [PRESET] button.
• The PST indicator turns off, indicating that the full auto preset function has been disabled.
Basic Operation
• Each preset function cannot be used when the following functions are set. Disable the function or change the setting before e xecuting the preset function.
• Zero shift function "Zero Shift Function" (page 3-16)
• Zero shift calibration "Adjusting Sensitivity" (page 3-5)
• Zero shift input "External Input" (page 4-17)
• DATUM 1/2 mode "Restrictions for Sensitivity Settings in Each Detection Mode" (page 6-9)
• Rising/falling edge detection mode" "Restrictions for Sen­sitivity Settings in Each Detection Mode" (page 6-9)
• The preset function is not suited for transparent workpieces such as thrubeam/retro-reflective models and other cases of detection with low light intensity differences.
• After changing any of the following settings, disable each preset function once and then execute again.
"Power Modes" (page 4-4) "Preset Saturation Function" (page 4-28) "Display Gain" (page 4-32)
• If the received light intensity raw value is 50 or less (200 or less when the light intensity is set to FuLL), the display will be "100.0" or less when the preset function is executed.
3-10
• If the [PRESET] button is pressed when the preset function is enabled
(PST indicator is lit), the current received light intensity changes to "100.0" and the setting value does not change.
• Expansion units can be preset with operations from the main unit.
"Common Key-Operations Function" (page 4-34)
• Periodic presetting is possible with signals input from an external source.
"External Input" (page 4-17)
• With the preset function, a process is carried out to ignore minute
received light intensity changes which do not affect the detection. The change amount to be ignored can be adjusted randomly.
"Preset Saturation Function" (page 4-28)
• If the sub-display is changed to "Extension", the original received light
intensity can be confirmed even when using the preset function. Extension "Sub Display" (page 4-23)
- Digital Fiber Sensor FS-N10 Series User's Manual -
Page 33
3-3 Adjusting Sensitivity
g

Two-point Calibration

Two-point calibration is the most basi c m e th od of sensitivity setting. The setting value can be established by simply pressing the [SET] button once each when a workpiece is present. and when it is absent.
Press the [SET] button once
1
when no workpiece is pres­ent.
Press the [SET] button once
2
when a workpiece is present.
Calibration is complete after the setting value flashes momentarily, and then stops (lights up).
Press the button to adjust the setting value.
Workpiece
3
Basic Operation
Reference
Step 1 or step 2 may be performed first. When performing two-point calibration on channel 2 of the 2-output type,
set the channel switch to " ".
li
Received
ht intensity
* If the difference between the 2 conditions is too small, "- - - -" flashes
1
Setting value
2
after calibration is complete. However, the setting value will still established.
- Digital Fiber Sensor FS-N10 Series User's Manual -
The setting value is set to the middle of the values at which the [SET] button was pressed first and pressed second.
Time
3-11
Page 34
3
Workpiece
Reference
Setting value Setting value
Thrubeam Model/Retro-reflective Model
Reflective Model
Received light intensity Received light intensity
No workpiece
With workpiece No workpiece*
With workpiece*
1
1
* Where detection occurs on a target having a background, the maximum sensi-
tivity setting ignores the background. Maximum sensit ivity setting is not avail- able if the background is more reflective than the workpiece.
3-3 Adjusting Sensitivity

Maximum Sensitivity Calibration

This sensitivity setting method is useful if the received light intensity is reduced by dust or dirt. The setting value is calibrated to be slightly than the light intensity received when it is determined.
1
Basic Operation
When a workpiece is absent for the reflective model, or when a workpiece is present for the thrubeam model/ retro-reflective model, press and hold the [SET] button for 3 seconds or more. Release the button when "SEt" flashes.
Calibration is complete after the setting value flashes momentarily, and then stops
(lights up). Press the button to adjust the setting
value.
When performing two-point calibration on channel 2 of the 2-output type, set the channel switch to " ".
3-12
- Digital Fiber Sensor FS-N10 Series User's Manual -
Page 35
3-3 Adjusting Sensitivity
Workpiece
g

Full Auto Calibration

This method automatically sets sensitivity using a moving workpiece. Sensitivity can be easily set by passing a workpiece without shutting down operating equipment.
While the workpiece passes through,
1
press and hold the [SET] button until "SEt" flashes.
After the workpiece has completely
2
passed through, release the [SET] but­ton.
Calibration is complete after the setting value flashes momentarily, and then stops (lights up).
3
Basic Operation
Reference
• If detection is not stable after the setting operation, for example due to vibration, press the button to adjust the setting value.
• When performing two-point calibration on channel 2 of the 2-output type, set the channel switch to " ".
li
Received
ht intensity
MIN
1
- Digital Fiber Sensor FS-N10 Series User's Manual -
2
MAX
Setting value
Time
The setting value is determined as the middle value between the maximum and minimum light intensity received values obtained while holding down the [SET] button.
3-13
Page 36
3
Workpiece
Reference
g
Calibration is complete after the setting value flashes momentarily, and then stops (lights up).
The setting value is determined as the value of light intensity received when the workpiece comes into position.
3-3 Adjusting Sensitivity

Positioning Calibration

This method is used when you want to position a workpiece.
1
Basic Operation
2
Press the [SET] button once when no workpiece is present.
Position a w orkpiece such that its edge aligns with the center of the project­ing beam. Then, press and hold the [SET] button for 3 seconds or more. Release the button when "SEt" flashes.
Press the button to adjust the setting value.
When performing two-point calibration on channel 2 of the 2-output type, set the channel switch to " ".
li
Received
3-14
ht intensity
Setting value
2
1
- Digital Fiber Sensor FS-N10 Series User's Manual -
Time
Page 37
3-3 Adjusting Sensitivity
Point

Other Calibration Methods

Percentage calibration
A setting value is established with a percentage in respect to the current received light intensity. When a sensor amplifier equipped with an external input is used, percentage tuning can be performed from an external device, such as PLC, allowing highly accurate detection of transparent workpieces and small workpieces, etc.
Refer to "Sensitivity Setting" (page 4-4) for details.
Zero shift calibration
This sensitivity setting performs the zero shift function and basic calibration (two-point calibration / maximum sensitivity calibration / full auto calibration) simultaneously. The lower of the received light intensity values specified at the time of sensitivity setting will automatically be set to "0".
Refer to "Sensitivity Setting" (page 4-4) for details.
Percentage calibration and zero shift calibration cannot be set with channel 2 of the 2-output type models.
3
Basic Operation
- Digital Fiber Sensor FS-N10 Series User's Manual -
3-15
Page 38
3-4
Green "PST" lights up
Current vallue is "
0
"
0
Received light intensity
Without workpiece (Background is detected)
200
300
400
450
650
0
Received light intensity
100
200
250
450

Setting the Current Received Light Intensity to 0 (Zero Shift)

Zero Shift Function

This function adjusts the current received light intensity display to "0". It is primar ily used with reflective sensor heads. If the received light intensity cannot be set to "0" when a reflective model is installed, this function can be used to set the received light
intensity to "0" with no workpiece present. This makes the difference in received light intensity easier to distinguish.
3
Basic Operation
Enabling the zero shift function
When the PST indicator is off, press
1
the [PRESET] button and button simultaneously.
The PST indicator lights in green, and the received light intensity changes to "0".
Disabling the zero shift function
Press and hold the [PRESET] button.
1
The PST indicator turns off, indicating that the zero shift function has been dis­abled.

Operating Principle of the Zero Shift Function

< For reflective models >
In applications for d istinguishing colors or
for detecting objects on a background, the received light intensity will not be "0" even when no workpiece is present.
3-16
Applying zero shift input to the lo wer
- Digital Fiber Sensor FS-N10 Series User's Manual -
level of received light intensity (with no workpiece present) enhances the detection display visibility.
Page 39
3-4 Setting the Current Received Light Intensity to 0 (Zero Shift)
Point
The zero shift function cannot be used in combination with the preset function. To use the zero shift function, make sure that the preset function is disabled (the PST indicator is not lit).
Reference
Zero shift of the expansion unit can be set from the main unit.
"Common Key-Oper ations Function" (page 4-34)
3
Basic Operation
- Digital Fiber Sensor FS-N10 Series User's Manual -
3-17
Page 40
3
3-5
When a fiber unit having strong power (longer detection range) is used at a close range, the received light intensity may remain unchanged from the maximum value in the display-enable d range (saturated) regardless of whether a workpiece is present or not. (Example: Numerical value does not change from 9999) In such a case, the light emission and the received light intensity can be automatically corrected to optimal state by using the saturation Canceling function.

Light emission/Received light intensity adjustment (Saturation Canceling)

Saturation Canceling function

Enabling the saturation Canceling function
Basic Operation
Press the [SET] button while holding down the [MODE] button.
1
The received light intensity corresponding to the power mode is set within the range as shown in the following table.
Power mode
HSPFINETURBO
SUPER 4095±500
ULTRAMEGA
Disabling the saturation Canceling function
Press the [SET] button while holding down the [MODE] button.
1
3-18
- Digital Fiber Sensor FS-N10 Series User's Manual -
Setting range of
received light intensity
2047±350
5000±600
Page 41
3-6
Point
The FS-N10 Series has the recommended settings (recipe) programmed in advance according to the fiber unit and applications.

Selecting Recipe

Loading the Recommended Settings (Recipe Function)

When a recipe is loaded, all settings other than those registered in the recipe are returned to the default values. Complete all other settings after loading the recipe.
3
Press and hold the [SET] and [PRE-
1
SET] buttons simultaneously for 3 seconds or more.
The "rSt" display flashes.
Press the button to display
2
"LoAd", and press the [MODE] button.
Press the button to select a
3
desired recipe, and press the [MODE] button.
After the settings are loaded, the screen displays "LoAd oK", which is then replaced with the current received light intensity.
Basic Operation
- Digital Fiber Sensor FS-N10 Series User's Manual -
3-19
Page 42
3-6 Loading the Recommended Settings (Recipe Function)
Reference

List of Recipes and Recommended Fiber Units

3
Basic Operation
Recipe
Fall detection
r-1 FALL)
(
Percentage calibration
r-2 SEtP)
(
Reflective background cancellation
r-3 0SEt)
(
Full light intensity (
r-4 MEGA)
Area detection (
r-5 ArEA)
0-DATUM* (
r-6 0dtM)
Recommended
units
Thrubeam model
Thrubeam model
Retro-reflective
model
Reflective model
Reflective model
*
Area model
(FU-E11/E40)
Retro-reflective
model
Description
To detect a falling workpiece. The falling edge of the received light intensity is detected, and output as one-shot. The received light intensity hold value can be con­firmed by pressing the [MODE] button twice on the output 1 screen.
To use the external input and periodically compen­sate for the reduction in received light intensity due to contamination on sensor head face. A decrease in the light-ON mode's received light intensity can be detected and output to an external source by using the 2-output type's output 2.
To use the external input and "0" the light intensity received without a workpiece present. The background received light intensity is set as 0 even when calibrating with the buttons on the display. (Zero Shift calibration)
To 0 the lower of level received light intensity values in order to display higher values of received light intensity. The display can be extended up to 5 digits by press­ing the [MODE] button twice on the output 1 screen.
This is recommended when outputting within a spe­cific received light intensity range. When the [SET] button is pressed once at the reference received light intensity, the upper/lower setting value limits will be set at ±10% from the reference received light inten­sity.
This is recommended when detecting a transparent
workpiece. The received light intensity at the state with no workpiece is set as ".0", and the displayed value is increased as the amount of light blocked by
the workpiece increases. When the [SET] button is pressed or when calibrated with external inputs, the received light intensity -5% at that point is set as the setting value.
* Not available for the FS-N10 Series shipped before March 10, 2011.
Refer to "List of Recipe Function Settings" (page 6-7) for the details on the items set with recipe loading.
3-20
- Digital Fiber Sensor FS-N10 Series User's Manual -
Page 43
3-7

Initialization of Settings (Reset to Initial Values)

The sensor amplifier can be reset to the factory default settings.
1

Initialization

Press and hold the [SET] and [PRE­SET] buttons simultaneously for 3 seconds or more.
The "rSt" display flashes.
3
Press the [MODE] button.
2
Press the button to display
3
"init".
Press the [MODE] button.
4
After the settings are initialized, the screen displays "oK", which is then replaced with the current received light intensity.
Reference
Refer to "Factory Default Setting (Default Value) List" on page 6-6.
Basic Operation
- Digital Fiber Sensor FS-N10 Series User's Manual -
3-21
Page 44
3
Reference
Basic Operation
3-8

Locking in MEGA Mode

MEGA Mode Lock (1-Output Type Only)

This function is available for the 1-output type models (FS-N11 / N12) only. The sensor amplifier can be locked in MEGA mode, such that it always operates in MEGA mode regardless of the power mode selected in the basic setup.
"Power Modes" (page 4-4)
Slide the power select switch to the
1
"M" side.
Sliding the power select switch back to the "SEL" side restores the power mode that was set before sliding the power select switch to MEGA mode.
• When the amplifier is locked in MEGA mode, the power mode may not be changed in the basic setup, as indicated by the flashing of "Loc".
"Power Modes" (page 4-4)
•Likewise, when the amplifier is in the key locked state, the power mode may not be changed, as indicated by the flashing of "Loc".
"Key Lock" (page 3-23)
3-22
- Digital Fiber Sensor FS-N10 Series User's Manual -
Page 45
3-9

Disabling the Key Operation

Key Lock

The key lock functi on di s ables all key operation to prevent unauthorized use.
Activating key lock
Press and hold the [MODE] button and
1
(or ) simu ltaneously f or 3 seconds
or more.
3
The screen displays "Loc", disabling key operation and displaying the current received light intensity.
Deactivating key lock
Press and hold the [MODE] button and
1
(or ) simu ltaneously f or 3 seconds
or more.
The screen displays "unL", enabling key operation.
Reference
• The key operations on the expansion unit can be locked from the main unit.
"Common Key-Operations Function" (page 4-34)
•By using the network unit NU series, key locks which could only be dis- abled via the network can be set. This function allo ws y ou to c hange onl y the necessary settings with a PLC and touch panel. Refer to the NU Series User's Manual for details.
Basic Operation
- Digital Fiber Sensor FS-N10 Series User's Manual -
3-23
Page 46
3
Point
3-9 Disabling the Key Operation

Key Lock with PIN Number

A PIN number can be set when activating key lock. Only users knowing the PIN number can operate the unit.
Activating key lock with a PIN number
Press (or ) 10 times while holding
1
down the [MODE] button.
Basic Operation
The screen displays "Loc 0".
Press the button to set a desired
2
number (up to 4 digits).
Press the [MODE] button.
3
The screen displays "Loc", disabling key operation and displaying the current received light intensity.
Deactivating key lock with a PIN number
Hold down the [MODE] button while
1
pressing (or ) 10 times.
The screen displays "Loc 0".
Press the button to specify the
2
PIN number, and then press the [MODE] button.
The screen displays "unL", enabling key operation.
3-24
If the PIN number is lost, contact your nearest KEYENCE office.
- Digital Fiber Sensor FS-N10 Series User's Manual -
Page 47

Settings for Advanced Functions

This chapter describes settings for advanced functions of the FS-N10 Series.
4-1 List of Settings....................................................4-2
4-2 Basic Settings.....................................................4-4
4-3 Detection Settings (Func)...................................4-7
4-4 Display Settings (diSP).....................................4-23
4-5 System Settings (SYS)..................................... 4-30
4-6 2-output Settings ( )........................... 4-35
4-7 Settings Save/Recall........................................4-39
4
- Digital Fiber Sensor FS-N10 Series User's Manual -
4-1
Page 48
4-1
MODE
MODE
MODE
MODE
StG­StG­StG-
SET
1234
1234 1234 1234 1234 1234
Std
SEtP OSEt
END
FuNc DiSP
SYS
SET SET
MODE
Press and hold for 3 seconds or more
FINE mode TURBO mode SUPER mode ULTRA mode MEGA mode HIGH SPEED mode
Normal sensitivity setting method Percentage Calibration*
1
Zero-shift calibration
Settings complete Go to detection setup mode
Go to display setup mode Go to system setup mode
Return to normal display
Fine
TurB SuPR ULTR MEGA
hSP
…(page 4-4) …(page 4-5) …(page 4-6)
…(page 4-4) …(page 4-4) …(page 4-4) …(page 4-4) …(page 4-4) …(page 4-4)
MODE
*
7
Settings complete Go to display setup mode Go to system setup mode
Return to detection setup mode
Normal (light intensity) detection mode
DATUM1 mode *
3
Timer OFF Off-delay timer *
2
On-delay timer *
2
One-shot timer *
2
DATUM2 mode *
3
Area detection mode Rising Edge Detection Mode Falling Edge Detection Mode
External input off External calibration input Preset input Zero shift input Reset input Light transmission OFF input Pause function *
4
Sleep function
Return to normal display
Adjusting light emission/received light intensity (Attenuation)
*5
*
8
Analog scaling
*6
Alternate display
…(page 4-7)
…(page 4-8)
…(page 4-9)
…(page 4-14) …(page 4-16) …(page 4-16) …(page 4-17) …(page 4-17) …(page 4-17) …(page 4-17) …(page 4-17)
…(page 4-17) …(page 4-18)
…(page 4-19)
…(page 4-7) …(page 4-7) …(page 4-7)
…(page 4-21)
…(page 4-11)
…(page 4-20)
Normal display method Reverse display
Settings complete Go to system setup mode Go to detection setup mode Return to display setup mode
Sub-display off Extension display Bar display Excess gain (%) display Light intensity hold display *
9
Excess gain hold (%) display *
9
L-on / D-on display
Disable the saturation of the Preset function
Enable the saturation of the Preset function *
10
Return to normal display
…(page 4-23) …(page 4-23)
…(page 4-23) …(page 4-24) …(page 4-24)
…(page 4-25) …(page 4-25) …(page 4-27) …(page 4-27)
…(page 4-28)
…(page 4-28)
*11
Settings complete Go to detection setup mode Go to display setup mode Return to system setup mode
Normal operation Twice the number of
interference-prevention units as
STD
(response time 2 times slower)
Disable Eco Function Enable Eco Function
Reduce power consumption (response time 4 times slower)
Disable common key operations Enable common key operations
APC function is disabled APC function is enabled
Standard current value display Maximum current value display
(4 times hysteresis)
Return to normal display
…(page 4-31)
…(page 4-31)
…(page 4-32) …(page 4-32)
…(page 4-33)
…(page 4-33)
…(page 4-34) …(page 4-34)
…(page 4-31)
…(page 4-30) …(page 4-30)

List of Settings

Basic Settings…(page 4-4)
4
Settings for Advanced Functions
Detection Settings (Func)…(page 4-7)
4-2
Display Settings (diSP)…(page 4-23)
System Settings (SYS)…(page 4-30)
- Digital Fiber Sensor FS-N10 Series User's Manual -
Page 49
4-1 List of Settings
Light intensity detection mode Limit setting output mode *
13
Warning output mode
Timer OFF Off-delay timer *
2
On-delay timer *
2
One-shot timer *
2
Settings complete
Return to normal display
…(page 4-38) …(page 4-36) …(page 4-37)
…(page 4-38) …(page 4-38) …(page 4-38) …(page 4-38)
MODE
MODE
MODE
MODE
MODE
MODE
MODE
MODE
*12
…(page 4-35)
2-output Settings ( )
*1 You can press the button to set between the range of -99P and 99P. *2 Press the button to set between the range of 1 and 9999 (ms). *3 Press the button to set the adjustment sensitivity to a range of between LEu1 and LEu3 and set the warning output level to a range of
between 0P and 100P. *4 Press the button to switch between oFF/on/KEEP. *5 Pressing the button enables you to set to the range within 1 to100.
*6 The analog output upper limit can be set between 50 and 65535. *7 FS-N11 *8 Mon
itor output type only (FS-N11MN Only). *9 Press the button to toggle between Std/P~P_/b~b_/P_b~/P~b_. *10 Press the button to set between the range of 100P and 200P.
*11 Main unit only. *12 Only 2-output types (FS-N13 / N14).
*13 Press the button to select the User (user reset) or Auto (auto reset) resetting methods.
Reference
MODE
C/N12C/N11EN/N13/N14 only
•Press the button and the button simultaneously to return to the previous setting option.
• When the button is held down, the settings menu will end and the received light intensity setting screen will appear.
4
Settings for Advanced Functions
- Digital Fiber Sensor FS-N10 Series User's Manual -
4-3
Page 50
4
Point
4-2

Basic Settings

Power Modes

Refer to page 4-2 "Basic Settings" for setting methods.
The detection stability will increase when the response time is delayed. You can select from the following 6 power modes. The response time slows down in the order of hSP, FinE, turb, Supr, ULtr, MEGA.
Display Power mode name Response time
hSP
FinE FINE mode (Default) 250 µs turb TURBO mode 500 µs SuPr SUPER mode 1 ms
Settings for Advanced Functions
Detection distance varies depending on the fiber units to be combined. For specific numerical values, refer to the catalogues and specifications of the fiber units concerned.
ULtr ULTRA mode 4 ms MEGA MEGA mode 16 ms
When using the unit in point-blank range, raising the power mode exces-
sively may attenuate the received light intensity. In such a case, either lower the power mode or use the saturation Canceling function.
"Light emission/Received light intensity adjustment (Saturation
Canceling)" (page 3-18)
HIGH SPEED mode 50 µs

Sensitivity Setting

Refer to page 4-2 "Basic Settings" for setting methods.
You can select from the following 3 sensitivity setting methods.
Display Sensitivity setting method
Std Normal sensitivity setting (without correct ion) (Default) 3-5 to 3-14 SEtP Percentage calibration 4-5 0SEt Zero shift calibration 4-6
* Refer to the followi
4-4
ng 2 pages for details.
- Digital Fiber Sensor FS-N10 Series User's Manual -
Reference
page
Page 51
Percentage calibration
900
1000
Reference
Received light intensity
When the percentage calibration target value is set to "-10%".
4-2 Basic Settings
Setup value can be configured at the ratio to the current received light intensity.
The percentage calibration target value can be set in the range of "-99P" (-99%) to 99P (99%) Default: -10P (-10%)
Sensitivity setting
In the desired reference state, press
1
the [SET] button.
Calibration is complete after the setting value flashes momentarily, and then stops (lights up).
Reference
• If the external input setting is set to "SEt" (external calibration input), peri­odic percentage calibration is possible from external devices, enabling stable detection even on workpieces having small sensitivity differences.
"External Input" (page 4-17)
• If the detection mode is set to "ArEA (area detection mode)", Hi and Lo are set simultaneously according to the received light intensity.
(Example) If the percentage calibration is set to 10P (10%) when the reference received light intensity is "100", the Hi and Lo setting values will be set as follows. Hi setting value: 110 Lo setting value: 90
Setting example
Hi 110
Received
light intensity
4
Settings for Advanced Functions
100 90Lo
- Digital Fiber Sensor FS-N10 Series User's Manual -
4-5
Page 52
4
Reference
Setting example
After two point calibration
(
Before zero shift calibration
)
Zero shift calibration is executed
500 2500 500 2500 0 2000
1000
Setting value
Setting value
1500
The light intensity received with no workpiece is subtracted from the whole.
The state with no workpiece (the lowest of received light intensity) is set to "0".
Received light intensity Received light intensity
4-2 Basic Settings
Zero shift calibration
Settings for Advanced Functions
This sensitivity setting performs the zero shift function and basic calibration (two-point calibration / maximum sensitivity calibration / full auto calibration) simultaneously. The lower of the received light intensity values specified at the time of sensitivity setting will automatically be set to "0".
The light intensity received with a workpiece present will be applied with the same amount of correction as for that received with no workpiece.
The following basic calibrations can be set while in the zero shift calibration.
•Two-point calibration
•Maximum sensitivity calibration
•Full auto ca l ibration
If
two-point calibrati
on is performed with a reflective model when the light intensity received with a workpiece is "2500" and without a workpiece is "500":
4-6
- Digital Fiber Sensor FS-N10 Series User's Manual -
Page 53
4-3
OFF
ON
With workpiece Without workpieceWithout workpiece
Control output
Timer-set period
OFF
ON
With workpiece Without workpieceWithout workpiece
Control output
Timer-set period
OFF
ON
With workpiece Without workpieceWithout workpiece
Control output
Timer-set period
OFF
ON
With workpiece Without workpieceWithout workpiece
Control output

Detection Settings (Func)

Output Timer

Refer to page 4-2 "Detect ion Set t ing s (Func)" for setting methods.
There are 3 types of timers available.
Display Function
toFF Uses no output timer. (Default)
Off-delay timer Tur ns off the output at a specified
oFFd
on-d
Shot
* Example of the light-ON (L-on) mode for the reflective model and the dark-ON (d-
on) mode for the thrubeam/retro-reflective
time after the detection signal goes off. Setting range: 1 to 9999 ms Default: 10 ms
On-delay timer Turns on the output at a specified time after the detection signal goes on. Useful if the ON duration is limited because the workpiece is shaky such as by vibration. Setting range: 1 to 9999 ms Default: 10 ms
One-shot timer Tur ns on the output and keeps it on for a specified period after the detection signal goes on. Setting range: 1 to 9999 ms Default: 10 ms
Timing chart
*
4
Settings for Advanced Functions
Reference
• When using the 2-output type, different timer functions and timer times can be set separately for channels 1 and 2.
"Output Timer for Output2" (page 4-38)
• The timer can only control on/off switching of the sensor output.
- Digital Fiber Sensor FS-N10 Series User's Manual -
4-7
Page 54
4-3 Detection Settings (Func)
Point

Detection Mode

Refer to page 4-2 "Detection Settings (Func)" for setting methods.
The table below lists the detec tion modes that can be selected.
4
Display Detection mode Function
Normal detection Normal detection mode (Default) -
Std
The received light intensity when there is no workpiece is always cali-
dtM1
DATUM1 mode
*
Settings for Advanced Functions
dtM2
ArEA
_|~d
~L_d
DATUM2 mode
Area detection mode Rising edge detection
mode Falling edge detection
mode
*
brated to " value is also calibrated so that the
setting value and received light inten­sity ratio is always constant.
The received light intensity when there is no workpiece is always cali­brated to " also calibrated so that the setting value and received light intensity ratio is always constant.
Detection is made only when the received light intensity goes out of a given range.
Detection is made only when the received light intensity increases.
Detection is made only when the received light intensity decreases.
100.0", and the setting
.0", and the setting value is
Reference
page
4-9
4-11
4-14
4-16 4-16
* In the DATUM mode, correction intervals can be set in the range of LEv1 to LEv3.
The default is LEv1. Also, the warning output level can be set in the range of 0P (0%) to 100P (100%). The default is 50P.
"Adjusting the correction interval" (page 4-12) "Changing the warning output level" (page 4-13)
4-8
• Refer to "Detection Mode for Output2" (page 4-35) for details on the 2-output type channel 2 detection mode.
• Refer to "Restrictions on Each Detection Mode" (page 6-9) for restrictions on the sensitivity s ett ing methods.
- Digital Fiber Sensor FS-N10 Series User's Manual -
Page 55
4-3 Detection Settings (Func)
Received light intensity
Fixed proportion
"100.0"
"5.0" (Initial value)
Setting value
0
100.0
100.0
20.0
60.0
60.0
Calibration point (Step 1)
Calibration point (Step 2)
1
2
The DTM indicator lights up.
Display changes to "
100.0".
DATUM1 mode
Refer to page 4-2 "Detection Settings (Func)" for setting methods.
In DATUM1 mode, the light intensity received without a workpiece is always co rrected to "100.0". The setting value is also corrected according to the correction amount, so that the ratio of setting value and received light intensity is kept constant, with the result of stable detection. The display of the sett ing value does not change.
The DATUM mode is effective in environments where received light intensity varies gradually, for example where the sensor head is prone to contamination or large temperature changes.
Sensitivity setting in DATUM1 mode
The sensitivity setting procedure below is an example of two-point
calibration (where the received light intensity is "100.0" when a
workpiece is absent and "20.0" when a workpiece is present) using the thrubeam/retro-reflective
model.
4
Settings for Advanced Functions
1
2
Press the [SET] button with no work­piece present.
Press the [SET] button with a work­piece present.
The light intensity received in full-light receiving state is displayed as "100.0".
- Digital Fiber Sensor FS-N10 Series User's Manual -
4-9
Page 56
4-3 Detection Settings (Func)
Point
Reference
• When the received light intensity stays lower than a given warning level, the received light intensity correction stops and the DTM indicator flashes.
"Changing the warning output level" (page 4-13)
• The light intensity correction stops if the raw received light inten­sity drops below 50 (200 when the light intensity expansio n setting is set to FuLL). The raw received light intensity can be confirmed by pressing the [MODE] button for longer than 3 seconds on the Basic screen.
"Display Gain" (page 4-32)
4
• By pressing the [PRESET] button,
Settings for Advanced Functions
the current value (received light intensity) can be forcibly calibrated to "100.0". Calibration is completed when the current value stops flash­ing. (At this time, the setting value is internally calibrated so that the received light intensity and setting value ratio stay constant.)
•You can also set the interval for executing correction, and/or the level of
received light intensity change for issuing a warning.
"Adjusting the correction interval" (page 4-12) "Changing the warning output level" (page 4-13)
•The saturation level is fixed to 101P (101%). To display a value higher
than 100.0, turn the preset saturation function OFF.
"Preset Saturation Function" (page 4-28)
4-10
- Digital Fiber Sensor FS-N10 Series User's Manual -
Page 57
4-3 Detection Settings (Func)
Display changes to "
.0".
DATUM2 mode
Refer to page 4-2 "Detection Settings (Func)" for setting methods.
In DATUM2 mode, the light intensity received without a workpiece is always co rrected to ".0". The setting value is also corrected according to the correction amount, so that the ratio of setting value and received light intensity is kept constant, with the result of stable detection. The display of the settin g value does not change.
The DATUM mode is effective in environments where received light intensity varies gradually, for example where the sensor head is prone to contamination or large temperature changes.
Sensitivity setting in DATUM2 mode
The sensitivity setting procedure below is an example of two-point
calibration (where the received light intensity is ".0" when a workpiece is absent and "80.0" when a workpiece is present) using the thrubeam/ re
tro-reflective model.
Press the [SET] button with no work-
1
piece present.
Setting value
40.0
80.0
100.0
Received light intensity
Fixed proportion
Calibration point (Step 1)
1
.0
2
Calibration point (Step 2)
The DTM indicator lights up.
".0"
"5.0" (Initial value)
4
Settings for Advanced Functions
.0
40.0
Press the [SET] button with a work-
2
piece present.
The light intensity received in the full-light receiving state is displayed as ".0".
- Digital Fiber Sensor FS-N10 Series User's Manual -
4-11
Page 58
4-3 Detection Settings (Func)
Point
Reference
Reference
• When the received light intensity stays higher than a given warn­ing level, the received light intensity correction stops and the DTM indicator flashes.
"Changing the warning output level" (page 4-13)
• The light intensity correction stops if the raw received light inten­sity drops below 50 (200 when the light intensity expansio n setting is set to Full). The raw received light intensity can be confirmed by pressing the [MODE] button for longer than 3 seconds on the Basic screen.
"Display Gain" (page 4-32)
4
• By pressing the [PRESET] button,
Settings for Advanced Functions
Adjusting the correction interval
Refer to page 4-2 "Detection Setting s (Func )" for setting methods.
A desired correction interval can be selected from the 3 levels:
HIGH SPEED 4.4 s 547.2 ms 34.2 ms FINE 13.1 s 1.64 s 102.4 ms TURBO/SUPER/ULTRA/MEGA 26.2 s 3.27 s 204.8 ms
the current value (received light intensity) can be forcibly calibrated to "0". Calibration is completed when the current value stops flash­ing. (At this time, the setting value is internally calibrated so that the received light intensity and setting value ratio stay constant.)
•You can also set the interval for executing correction, and/or the level of
received light intensity change for issuing a warning.
"Adjusting the correction interval" (page 4-12) "Changing the warning output level" (page 4-13)
Power mode
LEv1 (LEVEL1) is recommended.
LEv1
(LEVEL1)
LEv2
(LEVEL2)
LEv3
(LEVEL3)
4-12
- Digital Fiber Sensor FS-N10 Series User's Manual -
Page 59
4-3 Detection Settings (Func)
Received light intensity
Warning output
Control output
100.0
75.0
50.0
0
ON
OFF
ON
OFF
50%
100%
A warning is output if the received light intensity, when no workpiece exists, does not increase to the warning level because of beam axis deviation, etc.
Warning output level
Setting value
When the warning output level is set to 50%, dtm1, D-on:
Reference
Changing the warning output level
Refer to page 4-2 "Detecti on Settings (F unc)" f or set ting method s.
The unit goes into alarm mode when it
Display example
detects a minor level of variation of received
light intensity that does not cause the output to turn ON. If the received light intensity continues to remain between the warning level and setup level, the received light intensity correction function stops, the DTM indicator flashes, and the unit goes into alarm mode.
A level for outputting a warning can be set in the range of 0P (0%) to 100P (100%). (Default: 50P (50%)).
• 100P (100%) : The unit does not go int o alarm mode if the received light inten­sity varies.
• 50P (50%) : If no workpiece is detected, the unit goes into alarm mode when the received light intensity lowers to the intermediate value
b
etwe
en the received light int
intensity lowers even slightly.
• 0P (0%) : The unit goes into alarm mode when the level of received light
A 2-output type unit can output to the external devices that it is in the alarm output mode with 2-output setting ( ). For details, see "Warning output mode" (page 4-37).
4
Settings for Advanced Functions
ensity and the set value.
- Digital Fiber Sensor FS-N10 Series User's Manual -
4-13
Page 60
4
Reference
Point
4-3 Detection Settings (Func)
Area detection mode
Refer to page 4-2 "Detection Settings (Func)" for setting methods.
In this mode, the output signal is delivered (in N.O. condition) if the received light intensity is in the area defined by Hi (upper limit) and Lo (lower limit).
Settings for Advanced Functions
•Output-1 can be configured to be ON (N.O.) or OFF (N.C.) when the
• Area detection can be set just for output-1 scale. Output-2 scale oper-
Hi (Upper limit
setting value)
Lo (Lower limit
setting value)
0
Control output in N.O. condition
Control output in N.C. condition
ON
OFF
ON
OFF
received light intensity is above the Lo and under the Hi settings.
"Output Switch (L-on/D-on)" (page 3-4)
ates according to the "Output 2 detection mode".
"Detection Mode for Output2" (page 4-35)
4-14
Always ensure that "upper limit setting value > lower limit setting value". If "upper limit setting va lue lower limit setting value ", the received light intensity will always be output outside the Hi and Lo range.
- Digital Fiber Sensor FS-N10 Series User's Manual -
Page 61
4-3 Detection Settings (Func)
Methods for setting upper and lower limits
The upper and lower limits can be set using one of the 2 methods:
Setting sensitivity using the percentage calibration mode
See "Percentage calibration" (page 4-5).
Sensitivity setting using other calibration modes
Sensitivity is set separately for Hi and Lo.
Press to display "Hi" or "Lo".
1
The "Hi" or "Lo" value, and the current set- ting values are displayed alternately. When the [MODE] button is pressed while the display alternately flashes, the "Hi" or "Lo" display changes. If you do nothing for 3 seconds or more, the received light intensity display is auto- matically restored.
Press the [SET] button to perform sen-
2
sitivity setting (two-point calibration, maximum sensitivity setting, or other).
4
Settings for Advanced Functions
Calibration is complete after the setting value flashes momentarily, and then stops (lights up).
- Digital Fiber Sensor FS-N10 Series User's Manual -
4-15
Page 62
4-3 Detection Settings (Func)
Operation example
(1) (2) (3) (4)
OFF
ON
Control output
Received light intensity
Point
Reference
Edge detection mode
Refer to page 4-2 "Detection Settings (Func)" for setting methods.
In the edge detection mode, the sensor amplifier detects changes of received light intensity over a fixed period. This mode is used to turn on the output only when a workpiece is moving into or out of the detection range of the fiber unit.
• Turning the output ON when the received light intensity attenuates. Select "~L_d" (Falling edge detect mode)
• To turn the output ON when the received light intensity increases: Select "_|~d" (Rising edge detect mode)
4
Settings for Advanced Functions
Timing chart
When the falling edge detection mode is selected with a thrubeam model:
(1), (3):
The output is OFF because the received light intensity is not changing.
(2) :
The output is ON because the received light intensity is decreasing.
(4) :
The output remains OFF because the received light intensity is changing but is increasing.
Combining the edge detection mode and any of the functions below makes it more difficult to detect gradual changes in light intensity.
• If the interference prevention is "dobL" (double): "Interference Prevention" (page 4-33)
• If the eco mode is "FuLL" (Full): "Power Save" (page 4-31)
Sensitivity setting
Press the [SET] button with no workpiece presen t.
1
The amplifier is now configured to ignore changes in received light intensity for the dura- tion that the [SET] button is pressed. Calibration is complete after the setting value flashes momentarily , and then stops (lights up).
• When detecting workpieces run-
ning on a conveyor, the [SET] button can be pressed for a longer period to ignore the fluctuations of recei ved light intensity due to con veyor vibra- tion.
• If the setting value is so low that objects other than the workpiece of interest
are also detected, fine adjustments can be made with the button.
4-16
- Digital Fiber Sensor FS-N10 Series User's Manual -
Page 63
4-3 Detection Settings (Func)
Reference

External Input

Refer to page 4-2 "Detection Settings (Func)" for setting methods.
On sensor amplifiers with an external input option (FS-N11C/N12C/N11EN/ N12EN/N13/N14), the external input line can be controlled to utilize the following functions:
Display Function
oFF Disables use of the external input function. (Def ault)
SEt
PrSt
ShFt
rSt
LoFF
PAUS
SLEP
External calibration input
Uses the external input to perform calibration.
Preset input
*1
Executes a preset at the rising edge of the external input.
"Preset Function" (page 3-6)
Zero shift input
Executes a zero shift at the rising edge of the external input.
*2
"Setting the Current Received Light Intensity to 0 (Zero Shift)"
(page 3-16)
Reset input
Resets the display at the rising edge of the external input when using the hold function, limit setting detectio n, or counter output mode.
"Light intensity hold display" (page 4-25), "Excess gain hold (%)
display" (page 4-27)
"Limit setting output mode" (page 4-36)
Transmission OFF input
Keeps the transmission off during signal input.
Pause function
Locks the output condition during external input.
"Pause function" (page 4-18)
Sleep function
Uses the power save mode during external input.
"Sleep function" (page 4-19)
4
Settings for Advanced Functions
*1
When the preset input is selected, the zero shift function is disabled and cannot be set.
*2 When the zero shift input is selected, each preset function is disabled and cannot
be set.
• For external input cabling, see "Wiring Diagrams for Sensor Amplifiers" (page 2-4).
• Provide a short-circuit time of 2 ms or more, and 25 ms or more when external calibration input is sele cted.
- Digital Fiber Sensor FS-N10 Series User's Manual -
4-17
Page 64
4
Signal input
Received light intensity
Pause OFF
L-on
Pause ON
Pause KEEP
Signal input
Thrubeam model
OFF
ON
Output
OFF
ON
Output
OFF
ON
Output
4-3 Detection Settings (Func)
Pause function
Refer to page 4-2 "Detection Settings (F unc)" f or setting me thods .
During signal input, the output can be locked in a desired state regardless of received li ght input. There are 3 selections of output status that can be locked with the pause function.
Settings for Advanced Functions
Display Function
oFF The output is locked in OFF state during signal input. (Default)
on The output is locked in ON state during signal input.
KEEP
4-18
The output is locked in its existing state when a signal is received from the external input.
- Digital Fiber Sensor FS-N10 Series User's Manual -
Page 65
4-3 Detection Settings (Func)
Sleep function
Refer to page 4-2 "Detection Settings (Func)" for setting methods.
This function holds the sensor amplifier in the power save state (sleep mode) during input of the external signal. Once the sleep mode is entered, the following events occur.
• The display of received light intensity and setting value goes out.
• Light transmission stops.
• The output is turned off (for both L-on and D-on)
• A single segment on the digital display pulses across the display.
The normal display is restored after any key is pressed. If no operation is made for the next 4 seconds, the display unit turns back off.

Writing of External Input to EEPROM

Setting "no" to the "Writing of External Input to EEPROM" when any of "Set (External Tuning Input)", "PrSt (Preset Input)" or "ShFt (Zero Shift Input)" is selected, the setting modified by the external input will not be written to EEPROM. This prevents frequent wr iting to the EEPROM when inputting external signals and thus prevents the EEPROM from reaching the end of its service life (approx. 1 million wri
tes).
4
Settings for Advanced Functions
When received light intensity is dis-
1
played, press and hold the [MODE], [SET], and buttons simultaneously for 3 seconds or more.
To enable the restriction of writing set­tings, use the button to select "no".
Press and hold the [MODE] button.
2
The current received light intensity is displayed.
Reference
• When "no" is set, the values changed with external inputs will not be changed even after the power is turned off and on.
• When "yes" (default) is set, the settings are written to the EEPROM approx. 3 seconds after the settings are changed with external inputs.
- Digital Fiber Sensor FS-N10 Series User's Manual -
4-19
Page 66
4-3 Detection Settings (Func)
Reference

Adjusting Light Emission/Received Light Intensity (Attenuation)

Refer to page 4-2 "Detection Settings (Func)" f or setting methods .
When a fiber unit having strong power (longer detection range) is used at a close range, the received light intensity may remain unchanged from the maximum value in the display-enable d range (saturated) regardless of whether a workpiece is present or not. (Example: Numerical value does not change from 9999) In such a case, the light emission and the received light intensity can be manually corrected to optimal state by using the attenuation function.
4
The lower the value is lowered using the button, the more the received light
Settings for Advanced Functions
intensity attenuates.
Setting range: 1 to 100 (Initial value: 100)
Light emission and received light intensity can be automatically adjusted.
"Light emission/Received light intensity adjustment (Saturation
Canceling)" (page 3-18)
4-20
- Digital Fiber Sensor FS-N10 Series User's Manual -
Page 67
4-3 Detection Settings (Func)
50
1V
5V
Output voltage
4000 (Initial value)
65535
Received light intensity
Point

Analog output scaling (FS-N11MN only)

Refer to page 4-2 "Detection Settings (Func)" for setting methods.
The FS-N11MN is equipped with the analog voltage output function. In the initial setting, voltage is output within the range of 1-5 V to the displa yed received light intensity of 0 to 4000. The received light intensity corresponding to analog voltage of 5V can be changed within the range stated below:
Setting range of received light intensity corresponding to 5V: 50 to Setting range of received light intensity corresponding to 1V:
* The FS-N11MN shipped before March 10, 2011 becomes 100 - 9999.
• Received light intensity corrected with the preset function and
DATUM1/2 mode is not reflected in the analog output.
• When the work preset function, maximum sensitivity preset func-
tion or full automatic preset function is used, correctio n is made so that the light intensity that becomes ".0" will correspond to the analog voltage of 1V.
• While various preset functions are enabled, pressing the [PRESET]
button on the Analog Scaling Setting screen automatically sets analog output scaling values so that 5V will be output to the cur­rent value of "100.0".
• When the analog scaling mode setting is changed, the analog out-
put scaling values will be automatically set each time preset is per­formed. (Correction by pres etting is re flecte d in th e anal og output.)
"Analog output scaling (FS-N11MN only)" (page 4-21)
• Correction of the received light intensity using the zero shift
function is reflected in the analog output.
65535
* (Initial value:
0
fixed (cannot be changed)
4000
)
0
4
Settings for Advanced Functions
- Digital Fiber Sensor FS-N10 Series User's Manual -
4-21
Page 68
4
Reference
4-3 Detection Settings (Func)

Analog scaling mode (FS-N11MN only)

The analog output can be automatically scaled so that the 5 V is output in respect to the received light intensity "100.0". This function is used simultaneously with the various preset function. This function is useful when frequently using the preset function as the analog output can be constantly sequenced with the ".0 to 100.0" display, etc. * Not available for the FS-N10 Series shipped before March 10, 2011.
When received light intensity is dis-
1
played, press and hold the [MODE] but­ton, [SET] and button simultaneously for 3 seconds or more.
Settings for Advanced Functions
ProG YES or ProG no is displayed.
Press the [MODE] button once.
2
Using the button, select Pr-A Std
3
to not change the analog upper limit value when the preset function is exe­cuted, and select Pr-A PrSt to change the analog upper limit value sequen­tially when the preset function is exe­cuted. Then, hold down the [MODE] button.
4-22
When "Std" (default) is set, the analog output upper limit will not change even if the various preset function is used.
- Digital Fiber Sensor FS-N10 Series User's Manual -
Page 69
Normal display
Inverted display
4-4

Display Settings (diSP)

Display Reverse

Refer to page 4-2 "Display Settings (diSP)" for setting methods.
The current value and setting value displays can be inverted.
Display Function
oFF(OFF)
on(ON)Reversed display
Normal display (Default)

Sub Display

Refer to page 4-2 "Display Settings (diSP)" for setting methods.
In addition to the received light intensity and setting value, the foll owing information can be displayed on the sub display.
Operation
Display Information
oFF(OFF)
StdE
(Extension)
bAr(Bar) PEr(Percent) HLd(Hold)
HLdP
(Percent Hold)
Ldon(L-on/D-on)
Yes: Operation is possible. No: Operation is not possible. *1
For external calibration, the [SET] button can be operated to set values up to "9999".
*2
The displayed value is that before each preset function or zero shift function is executed.
*3 If the detection mode is set to DATUM1 mode or DATUM2 mode, the value before
DATUM correction is displayed.
*4 If the detection mode is set to DATUM1 mode or DATUM2 mode, the excess gain
hold display cannot be selected.
*5
If the detection mode is set to the area detection mode, rising edge detection mode, falling edge detection mode, or
*6 When set to the edge de tection mode , bar displa y, excess gain displa y, and excess
gain hold display cannot be selected.
*2
*6
*3
No sub display (Setting value / received light intensity) (Default)
Extension display (Displaying up to 5 digits)
Bar display (Bar excess gain / received light intensity)
Excess gain(%) display (Bar excess
*6
gain / excess gain) Received-light intensity hold display
(5 types selectable) Excess gain hold(%) display (5 types
*4*6
selectable)
L-on/D-on display (no/nC display)
no/nC
is displayed.
- Digital Fiber Sensor FS-N10 Series User's Manual -
*5
buttons
[SET]
Yes Yes -
*1
No
Yes Yes 4-24 Yes Yes 4-25 Yes No4-25
Yes No4-27 Yes No4-27
Reference
page
No4-24
4
Settings for Advanced Functions
4-23
Page 70
4-4 Display Settings (diSP)
Setting value/Received light intensity
Select L-on/D-on
MODE
MODE MODE
No operation for 3
sec or more
Sub-display
Reference
If the setting value is 1000:
Displaying a sub display
After selecting the desired sub-screen dis­play, press the [MODE] button twice in the normal display state (setting value / received light intensity). Press the [MODE] button once again to return to the normal display.
4
Extension display
Settings for Advanced Functions
Refer to page 4-2 "Display Settings (diSP)" for setting methods.
The received light intensity usually displayed with 4 digits can be extended to a 5-digit display. The maximum received light intensity value which can be displayed differs according
to the power mode and display gain. Refer to "Display Gain" (page 4-32) for more details.
Bar display
Refer to page 4-2 "Display Settings (diSP)" for setting methods.
Received light intensity can be displayed as a bar graph. Bars are displayed in real time in the r ang e of 85% to 115% at intervals of 5%, where 100% is the setting value. Up to 8 bars are displayed.
•The upper limit of the setting value remains a 4-digit value (9999). A larger value cannot be set.
• Each preset function, zero shift function, and each calibration function cannot be used during the extension display.
•The received light intensity calibrated with each preset function, zero shift function, each DATUM mode or edge detection mode is not applied on the extended display.
Display example
4-24
- Digital Fiber Sensor FS-N10 Series User's Manual -
Page 71
Excess gain (%) display
If the setting value is 1000 and the received light intensity is 2000 (200%):
Refer to page 4-2 "Display Settings (diSP)" for setting methods.
4-4 Display Settings (diSP)
Received light intensity is displa y ed as an excess gain* relative to the setting value. On the screen, the received light intensity is displayed as a bar graph and a percentage number. * Excess gain = (Received light intensity/Set-
ting value) x 100
Display example
Light intensity hold display
Refer to page 4-2 "Display Settings (diSP)" for setting methods.
The maximum peak values, and the minimum bottom values, of received light intensity can be displayed continuously. You can select from 5 possible display combinations:
Display Displayed values
Std (Standard)
P~P_ (Peak Max/Peak Min) b~b_ (Bottom Max/Bottom Min)
P_b~ (Peak Min/Bottom Max)
b_ (Peak Max/Bottom Min)
P~
Updates the peak or bottom value each time the cur- rent received light intensity falls below or rises above, respectively, the setting value. (Default)
Displays the maximum and minimum of peak values since the power is turned on. (Cumulative)
Displays the maximum and minimum of bottom val- ues since the power is turned on. (Cumulative)
Displays the minimum of peak values and the maxi­mum of bottom values since the power is turned on. (Cumulative)
Displays mum of bottom values since the power is turned on. (Cumulative)
the maximum of peak values and the mini-
4
Settings for Advanced Functions
How to reset the held values
To reset the peak and/or bottom values that are currently held, use one of the following procedures:
• Press and hold the [MODE] and [SET] buttons simultaneously.
• Set the external input (in) to "reset input (rSt)" and short-circuit the pink wire (pin (2)). "External Input" (page 4-17)
• Turn off the power.
- Digital Fiber Sensor FS-N10 Series User's Manual -
4-25
Page 72
4
Received light intensity
Setting value
Received light intensity
Cumulative maximum of bottom
Cumulative minimum of bottom
(2)
(1) (3)
(4)
4-4 Display Settings (diSP)
Timing chart for standard setting (Std)
[Peak value]
• Samples while the current received light inten- sity > setting value ((1), (3), etc.).
• If the current received light intensity is lower than the setting value, the previous peak value is updated.
[Bottom value]
• Samples while the current received light inten- sity < setting value ((2), (4), etc.).
• If the current received light intensity is higher than the setting value, the previous bottom value is updated.
Timing chart for peak value
Settings for Advanced Functions
[Cumulative maximum peak value]
• Samples while the current received light int en­sity > setting value ((1) to (4)).
• If the current received light intensity is lower than the setting value, the pre vious peak value and current peak value are compared. If the curren t peak value is higher, the value is updated.
[Cumulative minimum peak value]
• Samples while the current received light int en­sity > setting value ((1) to (4)).
• If the current received light intensity is lower than the setting value, the pre vious peak value and current peak value are compared. If the curren t peak value is lower, th e value is updated.
Received light intensity
Received light intensity
Peak value Received light
intensity Setting value
Bottom value
(1) (3)
(2)
Cumulative maximum of peak Received light intensity
Setting value Cumulative minimum
of peak
(2)(1) (3) (4)
(4)
Timing chart for bottom value
4-26
[Cumulative maximum bottom value]
• Samples while the current received light intensity < setting value ((1) to (4)).
• If the current received light intensity is higher than the setting value, the previous bottom value and current bottom value are com­pared. If the current bottom value is higher, the value is updated.
[Cumulative minim um bottom value]
• Samples while the current received light intensity < setting value ((1) to (4)).
• If the current received light intensity is higher than the setting value, the previous bottom value and current bottom value are com­pared. If the current bottom value is lower, the value is updated.
- Digital Fiber Sensor FS-N10 Series User's Manual -
Page 73
4-4 Display Settings (diSP)
Excess gain hold (%) display
Refer to page 4-2 "Display Settings (diSP)" for setting methods.
This function is the same as the "received-light intensity hold display", except that the value being held is displayed as excess gain instead of received light intensity.
"Light intensity hold display" (page 4-25)
L-on / D-on display
Refer to page 4-2 "Display Settings (diSP)" for setting methods.
The current output condition (L-on or D-on) can be displayed when the se ttin g value is display ed. The received light intensity is displayed in terms of received light intensity as usual. If the detection mode is set to the area detection mode, rising edge detection mode, falling edge detection mode, no/nC is displayed.
4
Settings for Advanced Functions
- Digital Fiber Sensor FS-N10 Series User's Manual -
4-27
Page 74
4
Point
Reference
4-4 Display Settings (diSP)

Preset Saturation Function

Refer to page 4-2 "Display Settings (diSP)" for setting methods.
The saturation level can be set in the range of 100P (100%) to 200P (200%) when the preset saturation function is ON. For example , if this funct ion is set to 150P, the received light intensity will be corrected to "150.0" when the preset function is enabled, and the digital display will display "100.0". The saturation level can be set to a higher value for stable detection even in environments which are prone to some contamination, such as dust. Turn off this function if you want to display values above "100.0" in applications where the preset function or the DATUM1 mode is used.
Settings for Advanced Functions
Display Function
on (ON)
oFF (OFF)
Fixes the display to "100.0" (displays no value higher than "100.0") when the actual received light intensity is above 100.0. (Default:
110P) Displays the actual received light intensity as is, even when it is
above "100.0".
Setting too high of a saturation level (specifying too large a value) decreases the detectability for small differences in light intensity or transparent workpieces.
*1 When the detection mode is set to DATUM1, the preset saturation function ON/
OFF state can be selected, but the saturation level will be fix ed to 101P (101%), and the setting screen will not appear.
*2 When the detection mode is set to DATUM2, the preset saturation function set-
ting screen will not appear.
• For the preset execution method, refer to "Preset Function" (page 3-6).
• Refer to the next page for the detailed operating principle.
4-28
- Digital Fiber Sensor FS-N10 Series User's Manual -
Page 75
Operating principles of the preset function
Reference
(Current received light intensity is 3300 in this example.)
Preset input is used to register the light intensity as "100.0".
The actual received light intensity (3300) is set to "110%" and all actual received light intensities exceeding "100%" are displayed as "100.0".
When light intensity falls below "100%", a value lower than "100.0" is displayed. (The percentage value "110%" (i.e., the saturation level) at the time of registration can be any value between 100 and 200%.)
"Preset Saturation Function" (page 4-
28)
Received light intensity
4095 3300
110.0
100.0
3000 2700
0
4-4 Display Settings (diSP)
All values in this range are 100.0.
(After preset input)
90.0
4
Settings for Advanced Functions
If the displayed light intensit y fal ls belo w "100.0", the preset input can be per­formed again to restore stable detection.
If light intensity is not displayed as "100.0" after preset input, the sensor amplifier is not in stable detection condition.
Even when the maximum sensi tivity preset function o r full auto preset func­t
ion is used, the prese
t saturation level setting function operates based on
the "100.0" point.
Received light intensity
4095 3300
2970
110.0
100.0
2700
0
(After another preset input for correction)
All values in this range are 100.0.
- Digital Fiber Sensor FS-N10 Series User's Manual -
4-29
Page 76
4
Point
Point
4-5

System Settings (SYS)

APC function

The APC (Auto Power Control) function automatically adjusts the light emis sion at a constant level. This function reduces variation in the received light intensity due to changes in ambient temperature.
• The APC function monitors and adjusts the LED light emission. Therefore, variation in the received light intensity due to contami- nation or characteristic changes of fiber unit cannot be reduced. When high-precision detection is necessary, also apply the D ATUM function and percentage calibration using external input.
Settings for Advanced Functions
DATUM function "Detection Mode" (page 4-8) Percentage Calibration "Percentage calibration" (page 4-5)
• If the unit is used for an extend period with the APC function turned ON, heavier load is imposed on the light emitting LED than when the unit is used with the APC function turned OFF. Normally, use the unit with APC turned OFF.
Screen display Functions
oFF(OFF) The APC function is disabled (Initial setting)
on(ON) The APC function is enabled
4-30
- Digital Fiber Sensor FS-N10 Series User's Manual -
Page 77
4-5 System Settings (SYS)
Point

Power Save

Refer to page 4-2 "System Settings (SYS)" for setting methods.
Power consumption can be reduced by turning off the digital display and output indicators.
Display Function
oFF (OFF) Disables the power save function. (Default)
on (ON)
FuLL (FULL)
ALL (ALL)
Immediately after the ON or FULL option is selected, the option is enabled, and the power save display results. The normal display is restored after any key is operated (except for eco ALL). If no operation is made for the next 30 seconds, the power save display returns.
Enables the power save function. (A single segment pulses across the digital display.)
Reduces power consumption. (The power save function is ON, and response time is 4 times longer.)
Eco ALL (The digital display and output indicators are all off).
This item becomes selectable by pressing and holding while FuLL is lit. The power consumption in Eco ALL is the same as that in Eco ON.
4
Settings for Advanced Functions
When FuLL is selected, response time is 4 tim es a s long as usual (i.e., as compared to when oFF or on is selected). When ALL is selected, response time is the same as usual.
Deactivating eco ALL
Press and hold the [MODE] and [PRE-
1
SET] buttons simultaneously.
The screen returns to the display of cur- rent received light intensity. If no operation is made for the next 30 seconds, the eco ALL state is restored.
- Digital Fiber Sensor FS-N10 Series User's Manual -
4-31
Page 78
4
Point
4-5 System Settings (SYS)

Display Gain

Refer to page 4-2 "System Settings (SYS)" for setting methods.
Received light intensity can be increased 4 times without sacrificing response time.
Std (Standard) Normal light intensity display (Default)
FuLL (Full) Full light intensity display (Hysteresis increases 4 times)
The maximum received light intensity value which can be displayed differs according to the power mode setting and display screen.
Settings for Advanced Functions
The maximum received light intensity values which can be displayed are shown below.
Display Function
Display
screen
Normal dis-
play
StdE
(Extension
display)
"Power Modes" (page 4-4) "Extension display" (page 4-24)
Display
gain
Std 4095 4095 4095 8190 9999 9999
FuLL - 9999 9999 9999 9999 9999
Std 4095 4095 4095 8190 32760 65504
FuLL - 16380 16380 32760 65504 65408
• This function is not supported in the hSP (HIGH SPEED) mode and the light intensity expansion setting screen is not displayed.
• When the preset function is used in DATUM1 or DATUM2 mode, correctible values are limited to 200.
HIGH
SPEED
FINE TURBO SUPER ULTRA MEGA
4-32
- Digital Fiber Sensor FS-N10 Series User's Manual -
Page 79
4-5 System Settings (SYS)
Point

Interference Prevention

Refer to page 4-2 "System Settings (SYS)" for setting methods.
Malfunction may result from "interference", a phenomenon in which light transmission occurs simultaneously with any other sensor amplifier. However, when the number of sensor amplifiers increases, the light transmission timing automatically shifts and prevents this interference. Setting this function to "dobL" (Double) on all of the connected amplifier units doubles the number units that do not mutually interfere.
Display Function
Std (Standard) Normal operation (Default)
dobL (Double)
• Selecting "dobL" also doubles response time over the "Std" state.
• When selecting "dobL", the main unit and its connected expansion units must all be set to "dobL".
• This function cannot be set if the eco function is set to “FuLL (Full)”.
Maximu m number of interference prevention units
Number of units prevented from interference during the connection of the main and expansion units of the FS-N10 Series is as follows.
Power mode
hSP (HIGH SPEED) 0 unit 0 unit 0 unit
FinE (FINE) 4 units 8 units
turb (TURBO) SuPr (SUPER)
ULtr (ULTRA)
MEGA (MEGA)
The interference prevention doubles over the Standard state.
Std
(Standard:
Normal
operation)
8 units 16 units
dobL
(Double: Doubling of interference
prevention units)
When the eco
function is
set to “Full”
16 units
4
Settings for Advanced Functions
Reference
Reference: Interference prevention is enabled even if connected to models other than the FS-N10 Series. For the number of units for interference pre­vention, please contact us .
- Digital Fiber Sensor FS-N10 Series User's Manual -
4-33
Page 80
4-5 System Settings (SYS)
Reference

Common Key-Operations Function

Refer to page 4-2 "System Settings (SYS)" for setting methods.
Some operations on the main unit are applied equally and simultaneously to all expansion units connected to it.
Display Function
oFF (OFF) Disables common key operations. (Default)
on (ON)Enables common key operations.
4
The operations that can be applied to all expansion units are as follows:
Settings for Advanced Functions
• Enabling/disabling the preset function "Preset Function" (page 3-6)
• Enabling/disabling the zero shift function "Zero Shift Function" (page 3-16)
• Enabling/Disabling the key lock function "Key Lock" (page 3-23)
This function is effective only with the main unit to which expansion units are connected. This function cannot be used with an independent main unit or with the expansion units.
4-34
- Digital Fiber Sensor FS-N10 Series User's Manual -
Page 81
4-6
*This function is only available for the 2-output type models (FS-N13 / N14).

2-output Settings ( )

Detection Mode for Output2

Refer to page 4-3 "2-output Settings ( )" for setting methods.
Display Function
Received-light intensity judgment mode (usual mode)
Std (Standard)
LiMt (Limit)
ALrt (Alert)
*1 If output 2 is set to Std, the DATUM1, DATUM2, area detection, and edge detection
modes are disabled in output 2.
*2 You can select the USEr (user reset) or Auto (auto reset) resetting methods after
outputting the limit setting under the following restriction:
-If output 1 is set to the D ATUM1 mode or DATUM2 mode when LiMt is selected,
the detection will be performed with the value prior to DATUM calibration.
-If output 1 is set to the area detection mode or edge detection mode, LiMt can-
not be selected with output 2.
*1
(Default)
*2
Limit setting output mode (Output 2 turns on if received light intensity for output 1 decreases).
"Limit setting output mode" (page 4-36)
War ning output mode (Output 2 turns on when an error occurs).
"Warning output mode" (page 4-37)
4
Settings for Advanced Functions
- Digital Fiber Sensor FS-N10 Series User's Manual -
4-35
Page 82
4
OFF
ON
Received light intensity
Peak hold value Limit setting for output 2
Setting value of received light intensity for output 1
Output2 (When  is selected)
Output2 (When  is selected)
Output1
OFF
ON
OFF
ON
(2)(1) (3)
4-6 2-output Settings ( )
Limit setting output mode
Refer to page 4-3 "2-output Settings ( )" for setting methods.
Setting output 2 to "LiMt" (limit setting output mode) enables detection of reductions i n received light intensity for output 1. Specifically, if the cumulative minimum of peak value for output 1 falls below the limit setting value, it is recognized as a reduction in received light intensity and therefore output 2 is turned on.
"Light intensity hold displa y" (page 4-25)
Timing chart
For period (1), the peak value is not below the limit setting value, so output 2 remains
Settings for Advanced Functions
OFF. For period (2), the peak value is below the limit setting value, so output 2 turns O N. For period (3),
the peak value again is above the limit set ting value, so output 2 turns OF F if Auto (Auto) is selected. If USEr (User) is selected, the cumulative minimum of peak values is the peak value during period (2), which is below the limit setting value, so output 2 remains ON.
How to reset after the output is activated
4-36
Display Resetting method
User reset
USEr (User)
- Press and hold the [MODE] and [SET] buttons simulta- neously.
-Execute reset using external input.
- Cycle the power.
Auto reset
- Press and hold the [MODE] and [SET] buttons simulta-
Auto (Auto)
- Digital Fiber Sensor FS-N10 Series User's Manual -
neously.
-Execute reset using external input.
- Cycle the power.
- When "peak value > setting value + hysteresis"
Page 83
4-6 2-output Settings ( )
Warning output mode
Refer to page 4-3 "2-output Settings ( )" for setting methods.
If output 2 is set to "ALrt" (warning output mode), output 2 turns on (or off if the output condition is set to N.C.) when either of the following errors occur:
• Correction error in DATUM1 or DATUM 2 mode
• END-APC error
Reference
For error recovery procedures, see "Troubleshoot ing" (page 6-2).
4
Settings for Advanced Functions
- Digital Fiber Sensor FS-N10 Series User's Manual -
4-37
Page 84
4-6 2-output Settings ( )
OFF
ON
With workpiece Without workpieceWithout workpiece
Control output
Timer-set period
OFF
ON
With workpiece Without workpieceWithout workpiece
Control output
Timer-set period
OFF
ON
With workpiece Without workpieceWithout workpiece
Control output
Timer-set period
OFF
ON
With workpiece Without workpieceWithout workpiece
Control output
Reference

Output Timer for Output2

Refer to page 4-3 "2-output Settings ( )" for setting methods.
There are 3 types of timers available.
4
Display Function
Timing chart
*
toFF Uses no output timer. (Default)
Off-delay timer
Settings for Advanced Functions
oFFd
Tur ns off the output at a specified time after the detection signal goes off. Setting range: 1 to 9999 ms Default: 10 ms
On-delay timer Tur ns on the output at a specified time after the detection signal
on-d
goes on. Useful if the ON duration is limited because the workpiece is shaky such as by vibration. Setting range: 1 to 9999 ms Default: 10 ms
One-shot timer Tur ns on the output and keeps it
Shot
on for a specified period after the detection signal goes on. Setting range: 1 to 9999 ms Default: 10 ms
4-38
* The above charts show operation examples in the light-ON (L-on) mode for the
reflective model and in the dark-ON (d-on) mode for the retro-refle ctive or thru- beam model.
• Different output timers and different timer times can be set separately for channels 1 and 2.
"Output Timer" (page 4-7)
• The timer can only control on/off switching of the sensor output.
- Digital Fiber Sensor FS-N10 Series User's Manual -
Page 85
4-7
Point

Custom Save (Settings Save)

The current setting value and advanced function settings can be saved in the sensor amplifier.
"Settings for Advanced Functions" (page 4-1)
1
2

Settings Save/Recall

Executing a custom save will override the previously saved settings with the current settings.
Press and hold the [SET] and [PRE­SET] buttons simultaneously for 3 seconds or more.
The "rSt" display flashes.
Press the button to display "SAvE", and then press the [MODE] button
4
Settings for Advanced Functions
Press the button to display "YES",
3
and then press the [MODE] button.
After the settings are saved, the screen displays "oK", which is then replaced with the current received light intensity.
- Digital Fiber Sensor FS-N10 Series User's Manual -
4-39
Page 86
4
Point
Reference
4-7 Settings Save/Recall

User Reset (Settings Recall)

The settings saved with th e c ustom save function can be recalled.
Press and hold the [SET] and [PRE-
1
SET] buttons simultaneously for 3
Settings for Advanced Functions
seconds or more.
The "rSt" display flashes.
Press the [MODE] button.
2
Executing a user reset will override all of the prev ious settings with the settings that were saved using the custom save function.
The factory settings will be recalled if a user reset is executed without pre- viously executing a custom save.
4-40
Press the button to display
3
"USEr", and then press the [MODE] button.
After a user reset is executed, the screen displays "oK", which is then replaced with the current recei ved li ght intensity.
- Digital Fiber Sensor FS-N10 Series User's Manual -
Page 87

Specifications

This chapter provides the specifications, circuit diagrams and dimensions of the FS-N10 Series.
5-1 Specifications .....................................................5-2
5-2 Input/Output Circuit Diagram..............................5-6
5-3 Dimensions.........................................................5-8
5
- Digital Fiber Sensor FS-N10 Series User's Manual -
5-1
Page 88
5-1

Specifications

1-Output Type

5
Specifications
Type 1-Output Type Input/Output configuration
Main unit/Expansion unit Main unit Expansion unit Main unit Expansion unit Main unit Expansion unit Models NPN FS-N11N FS-N12N FS-N11CN FS-N12CN FS-N11EN FS-N12EN
Number of input/out­put
Light source LED Red 4-element LED (Wavelength: 630 nm) Response time 50us(HIGH SPEED)/250us(FINE)/500us(TURBO)/1ms(SUPER)/4ms(ULTRA)/16ms(MEGA) Output toggle Light-ON/Dark-ON toggle Timer functions Timer OFF/OFF delay/ON delay/One shot Input/Out-
put specifi­cations
Expansion units Up to 16 units can be connected (total connection of 17 units including the main unit); 2-output
Protection circuit Protection against reverse power connection, output overcurrent and output surge Number of interference preven-
tion units
Rating Supply voltage 12 - 24 VDC +/- 10%, ripple (P-P) 10% or less
Environ­ment resis­tance
Case material Both main unit and cover material: Polycarbonate Case dimensions H30.3mm×W9.8mm×L71.8mm Mass Approx. 75 g Approx. 45g Approx. 22g Approx. 22g Approx. 22g Approx. 22g
PNPFS-N11P FS-N12P FS-N11CP FS-N12CP - - Control output 1-output1-output1-output1-output1-output1-output Monitor output
(1-5V) External input--1-input1-input1-input1-input
Control output (NPN)
Control output (PNP)
External inputInput time: 2 ms (ON) / 20 ms(OFF) or over. 25 ms(ON/OFF) only when toggled to external
Power consumption (NPN output type)
Power consumption (PNP output type)
Operating ambient luminance
Operating ambient temperature
Operating ambient humidity
Vibration resistance 10 to 55 Hz Compound amplitude 1.5 mm, 2 hours for each of X, Y and Z axes Shock resistance
*1 For M8 and e-CON connector types, cable length must be 30 m or less. *2 Increased by 100 mW (4.0 mA) for HIGH SPEED mode. *3 One or two units connected: -20 to +55 °C; 3 to 10 units connected: -20 to +50 °C,
11 to 16 units connected: -20 to +45 °C. When using 2-output type, one unit is counted as two units. All temperature specifications are for when the unit is mounted on a DIN rail; and installed on metal frame.
Cables
------
NPN open collector 24 V or less, maximum 100 mA or less (stand-alone)/20 mA or less (multiple
PNP open collector 24 V or less, maximum 100 mA or less (stand-alone used)/20 mA or less
Up to 16 units can be connected, but may differ depending on the power mode.
0 for HIGH SPEED; 4 for FINE and 8 for TURBO/SUPER/ULTRA/MEGA
(These numbers double when "Double" is selected.)
Normal: 900 mW or less (36 mA or less at 24 VDC; 48 mA or less at 12 VDC)
ECO ON (All): 800 mW or less (32 mA or less at 24 VDC; 39 mA or less at 12 VDC)
ECO Full: 470 mW or less (19 mA or less at 24 VDC; 23 mA or less at 12 VDC)
Normal: 950 mW or less (39 mA or less at 24 VDC; 52mA or less at 12 VDC)
ECO ON (All): 850 mW or less (35mA or less at 24 VDC; 44 mA or less at 12 VDC)
ECO Full: 520 mW or less (21mA or less at 24 VDC; 26mA or less at 12 VDC)
Incandescent lamp: 20,000 lx or less; solar light: 30,000 lx or less
500m/s
M8 connector
connections)
Residual voltage: 1V or less
(multiple connections)
Residual voltage: 1V or less
calibration input
type is treated as 2 units, however.
-20 °C to +55 °C (no freezing)
35 to 85%RH (no condensation)
2
, 3 times for each of X, Y and Z axes
*1
e-CON connector
*3
*1
*2
*2
*2
*2
5-2
- Digital Fiber Sensor FS-N10 Series User's Manual -
Page 89

2-Output Type

5-1 Specifications
Type 2-Output Type Input/Output configuration
Main unit/Expansion Units Main unit Expansion unit Main unit Expansion unit Models NPN FS-N13N FS-N14N --
Number of input/out­put
Light source LED Red 4-element LED (Wavelength: 630 nm) Response time 50us(HIGH SPEED)/250us(FINE)/500us(TURBO)/1ms(SUPER)/4ms(ULTRA)/16ms(MEGA) Output toggle Light-ON/Dark-ON toggle Timer functions Timer OFF/OFF delay/ON delay/One-shot Input/Out-
put specifi­cations
Expansion units Up to 16 units can be connected (total connection of 17 units including the main unit); 2-output
Protection circuit Protection against reverse power connection, output overcurrent and output surge Number of interference preven-
tion units
Rating Supply voltage DC12-24V±10%, ripple (P-P) 10% or less
Environ­ment resis­tance
Case material Both ma in unit and cover material: Polycarbonate Case dimensions H30.3mm×W9.8mm×L71.8mm Mass Approx. 80g Approx. 70g Approx. 22g Approx. 22g
PNPFS-N13P FS-N14P FS-N13CP FS-N14CP Control output 2-output2-output2-output2-output Monitor output
(1-5V) External input1-input1-input- -
Control output (NPN)
Control output (PNP)
External inputInput time: 2 ms(ON) / 20 ms(OFF) or more. 25 ms(ON/OFF) or more only when toggled to
Power consumption (NPN output type)
Power consumption (PNP output type)
Operating ambient luminance
Operating ambient temperature
Operating ambient humidity
Vibration resistance 10 to 55 Hz Compound amplitude 1.5 mm, two hours for each of X, Y and Z axes Shock resistance
*1 For M8 connector type, cable length must be 30 m or less. *2 Increased by 100 mW (4.0 mA) for HIGH SPEED mode. *3 One or two units connected: -20 to +55 °C; 3 to 10 units connected: -20 to +50 °C,
11 to 16 units connected: -20 to +45 °C. When using 2-output type, one unit is counted as two units. All temperature specifications are for when the unit is mounted on a DIN rail; and installed on metal frame.
NPN open collector 24 V or less, 2-output maximum 100 mA or less (used stand-alone)/20 mA or
PNP open collector 24 V or less, 2-output maximum 100 mA or less (used stand-alone)/20 mA or
ECO ON (All): 800 mW or less (32 mA or less at 24 VDC; 39 mA or less at 12 VDC)
ECO ON (All): 950 mW or less (38mA or less at 24 VDC; 47mA or less at 12 VDC)
Cable
--- -
less (multiple connections)
Residual voltage: 1V or less
less (multiple connections)
Residual voltage: 1V or less
external calibration input
type is treated as 2 units, however.
Up to 16 units can be connected, but may differ depending on the power mode.
0 for HIGH SPEED; 4 for FINE and 8 for TURBO/SUPER/ULTRA/MEGA
(These numbers double when "Double" is selected.)
Normal: 900 mW or less (36 mA or less at 24 VDC; 48 mA or less at 12 VDC)
ECO Full: 470 mW or less (19 mA or less at 24 VDC; 23 mA or less at 12 VDC) Normal: 1050 mW or less (42mA or less at 24 VDC; 56mA or less at 12 VDC)
ECO Full: 600 mW or less (24mA or less at 24 VDC; 29mA or less at 12 VDC)
Incandescent lamp: 20,000 lx or less; solar light: 30,000 lx or less
-20 °C to +55 °C (no freezing)
35 to 85% (no condensation)
2
, 3 times for each of X, Y and Z axes
500m/s
M8 connector
*3
*1
*2
*2
*2
*2
5
Specifications
- Digital Fiber Sensor FS-N10 Series User's Manual -
5-3
Page 90
5-1 Specifications

Monitor Output Type

5
Specifications
Typ e Input/Output configuration Main unit/Expansion unit Models NPN
Number of input/out­put
Light source LED Response time
Output toggle Timer functions Output
specifica­tions
PNP Control output Monitor output
(1-5V) External input
Control output (NPN)
Monitor output
50us(HIGH SPEED)/250us(FINE)/500us(TURBO)/1ms(SUPER)/4ms(ULTRA)/
Red 4-element LED (Wavelength: 630 nm)
Timer OFF/OFF delay/ON delay/One-shot
NPN open collector 24 V or less, max imum 100 mA or less (stand-alone)/20 mA or less
1 to 5 V voltage output; load resistance 10kΩ or more; repeatability +/-0.5% of F.S
Monitor output type
Cable
Main unit
FS-N11MN
-
1-output 1-output
-
16ms(MEGA)
Light-ON/Dark-ON toggle
(multiple connections)
Residual voltage: 1V or less
1 ms response time, 63% response (HIGH SPEED, FINE, TURBO)
Expansion units
Protection circuit Number of interference preven-
tion units
Rating Supply voltage
Environ­ment resis­tance
Case material Case dimensions Mass
Power consumption
Operating ambient luminance
Operating ambient temperature
Operating ambient humidity
Vibration resistance Shock resistance
*1 SUPER: 1.2ms, ULTRA:1.8ms, MEGA: 4.2ms 63% response *2 Increased by 100 mW (4.0 mA) for HIGH SPEED mode. *3 One or two units connected: -20 to +55 °C; 3 to 10 units connected: -20 to +50
11 to 16 units connected: -20 to +45 °C. When using 2-output type, one unit is counted as two units. All temperature specifications are for when the unit is mounted on a DIN rail; and installed on metal frame.
Up to 16 units can be connected (total connection of 17 units including the main unit);
2-output type is treated as 2 units, however.
Protection against reverse power connection, output overcurrent and output surge
Up to 16 units can be connected, but may differ depending on the power mode.
0 for HIGH SPEED; 4 for FINE and 8 for TURBO/SUPER/ULTRA/MEGA
(These numbers double when "Double" is selected.)
12-24 VDC +/- 10%, ripple (P-P) 10% or less
Normal: 900 mW or less (36 mA or less at 24 VDC; 48 mA or less at 12 VDC)
ECO ON (All): 800 mW or less (32 mA or less at 24 VDC; 39 mA or less at 12 VDC)
ECO Full: 470 mW or less (19 mA or less at 24 VDC; 23 mA or less at 12 VDC)
Incandescent lamp: 20,000 lx or less; solar light: 30,000 lx or less
-20 °C to +55 °C (no freezing)
*3
35 to 85% RH (no condensation)
10 to 55 Hz Compound amplitude 1.5 mm, two hours for each of X, Y and Z axes
500m/s2, 3 times for each of X, Y and Z axes
Both main unit and cover material: Poly car bonate
H30.3mm x W9.8mm x L71.8mm
Approx. 75 g
°C
,
*1
*2
*2
5-4
- Digital Fiber Sensor FS-N10 Series User's Manual -
Page 91

Zero Line Type

5-1 Specifications
Typ e Input/Output configuration Main unit/Expansion unit Models
Number of
Control output
input/out-
Monitor output
put
(1-5V)
External input Light source LED Response time
Output toggle Timer functions Expansion units Number of interference preven-
tion units
Rating Supply voltage
Power consumption
Environ-
Operating ambient ment resis-
luminance tance
Operating ambient
temperature
Operating ambient
humidity
Vibration resistance
Shock resistance Case material
Case dimensions Mass
*1 Counted as 1-output unit. when connected to the communications unit of NU Series. *2 Increased by 100 mW (4.0 mA) for HIGH SPEED mode. *3 One or two units connected: -20 to +55 °C; 3 to 10 units connected: -20 to +50
11 to 16 units connected: -20 to +45 °C. When using 2-output type, one unit is counted as two units. All temperature specifications are for when the unit is mounted on a DIN rail; and installed on metal frame.
50us(HIGH SPEED)/250us(FINE)/500us(TURBO)/1ms(SUPER)/4ms(ULTRA)/
Up to 16 units can be connected (total connection 17 units including main unit)
Up to 16 units can be connected, but may differ depending on the power mode.
0 for HIGH SPEED; 4 for FINE and 8 for TURBO/SUPER/ULTRA/MEGA
Normal: 900 mW or less (36 mA or less at 24 VDC; 48 mA or less at 12 VDC)
ECO ON (All): 800 mW or less (32 mA or less at 24 VDC; 39 mA or less at 12 VDC)
ECO Full: 470 mW or less (19 mA or less at 24 VDC; 23 mA or less at 12 VDC)
10 to 55 Hz Compound amplitude 1.5 mm, 2 hours for each of X, Y and Z axes
Red 4-element LED (Wavelength: 630 nm)
Timer OFF/OFF delay/ON delay/One-shot
(These numbers double when "Double" is selected.)
12-24 VDC +/- 10%, ripple (P-P) 10% or less
Incandescent lamp: 20,000 lx or less; solar light: 30,000 lx or less
500m/s2, 3 times for each of X, Y and Z axes
Both main unit and cover material: Polycarbonate
Zero line type
None (for communications only)
Expansion unit
FS-N10
*1
-
-
-
16ms(MEGA)
Light-ON/Dark-ON toggle
-20 °C to +55 °C (no freezing)
35 to 85% RH (no condensation)
H30.3mm x W9.8mm x L71.8mm
Approx. 20g
°C
,
*3
*2
*2
5
Specifications
- Digital Fiber Sensor FS-N10 Series User's Manual -
5-5
Page 92
5
Brown
*
Blue
*
Black
0 V
12 to 24 VDC
* FS-N11P only
Sensor main circuit
Load
Overcurrent
protection circuit
(Control output)
12 to 24VDC
DC3.3V
(4)
(Control output)
PLC, etc.
Short-circuit current 1 mA or less
(2)
(1)
*1
(3)
*1
0V
Sensor main circuit
Overcurrent protection circuit
Load
*1 FS-N11CN/N11EN only
12 to 24VDC
(4)
(Control output)
PLC, etc.
Short-circuit current 2 mA or less
(2)
(1)
*1
(3)
*1
0V
Sensor main circuit
Overcurrent protection circuit
Load
*1 FS-N11CP only
12 to 24VDC
DC3.3V
Black (Control output 1)
PLC, etc.
Short-circuit current 1 mA or less
White (Control output 2)
Pink
Brown
*1
Blue
*1
0V
Sensor main circuit
Overcurrent protection circuit
Load
Load
*1 FS-N13N only
12 to 24VDC
Black (
Control output 1)
PLC, etc.
Short-circuit current 2 mA or less
White (Control output 2)
Pink
Brown
*1
Blue
*1
0V
Sensor main circuit
Overcurrent protection circuit
Load
Load
*1 FS-N13P only
Specifications
5-2

Input/Output Circuit Diagram

1-Output Type

FS-N11N / N12N FS-N11P / N12P
12 to 24 VDC
*1
Brown
Load
Black
Overcurrent
protection circuit
Sensor main circuit
*1 FS-N11N only
FS-N11CN/N12CN/N11EN/N12EN FS-N11CP / N12CP
(Control output)
*1
Blue
0 V
M8 connector Pin layout

2-Output Type

FS-N13N/N14N FS-N13P / N14P
5-6
2
4
1
3
- Digital Fiber Sensor FS-N10 Series User's Manual -
e-CON connector Pin layout
2
1 3
4
Page 93
FS-N13CP/N14CP
0 V
12 to 24 VDC
* FS-N13CP only
Sensor main circuit
(Control output 2)
(Control output 1)
Overcurrent
protection circuit
Overcurrent
protection circuit
Load
Load
1
2
3
4
M8 connector Pin layout
p
5-2 Input/Output Circuit Diagram

Monitor output type

FS-N11MN
Protection
circuit
Overcurrent
protection circuit
Sensor main circuit
Device with in
*1
Brown Orange
Load
Black Blue
Monitor output (1 to 5V)
(Control output)
*1
12 to 24 VDC
0 V
ut impedance 10 kΩ or more
- Digital Fiber Sensor FS-N10 Series User's Manual -
5
Specifications
5-7
Page 94
5
Max. 170°
Min. 8
Cable length: 2m
When amplifier mounting bracket (OP-73880) is used
Bottom surface of amplifier mounting bracket
Counterbore: Depth:3.2
When cover is opened: 100 max.
3.4
3.9
20.7
35.4
71.8
22.9
2
2 4.4 3.4
3.4
7.2
Counterbore: Depth:2.7
6.4
25.7
15
31
25.4
(35.3)
6.7
4
30.3 (13)
(38.5)
(17.3)
30.3
3.8
6.5
9.8
9.8
Specifications
5-3

Dimensions

FS-N11□/N12□/N13□/N14□/N11MN/N10

* 1 FS-N11N(P)/N13N(P)/N11MN is not provided with an expansion connector. * 2 FS-N10 is not provided with cables.
5-8
- Digital Fiber Sensor FS-N10 Series User's Manual -
Page 95
8.6
M8 connector
3.4
20.7
35.4
71.8
22.9
223.4
3.4
7.2
6.4
4.4
25.7
15 31
25.4
9.5
(35.3)
6.7 4
30.3
(15.5)
(38.5)
(18.8)
30.3
3.8
6.5
9.8
9.8
When cover is opened: 100 max.
Max. 170°
When amplifier mounting bracket (OP-73880) is used
Bottom surface of amplifier mounting bracket
Counterbore: Depth:3.2
Counterbore:
Depth:2.7

FS-N11C□/N12C□/N13CP/N14CP

5-3 Dimensions
5
Specifications
* 1 FS-N11CN(P)/N13CP/ is not provided with an expansion connector.
- Digital Fiber Sensor FS-N10 Series User's Manual -
5-9
Page 96
5-3 Dimensions

FS-N11EN/12EN

9.8
5
Specifications
4
When cover is opened: 100 max.
6.7
35.4
76.5
6.5
25.4
Max 125°
2
3.4
Counterbore: Depth:2.7
2 4.4 3.4
Counterbore: Depth:3.2
3.4
30.3
(35.3)
20.7
When amplifier mounting bracket (OP-73880) is used
3.8
22.9
25.7
Bottom surface of amplifier mounting bracket
12.7
6.4
7.2
(10.3)
e-CON connector
30.3
38.5
15 31
* 1 FS-N11EN is not provided with an expansion connector.
5-10
- Digital Fiber Sensor FS-N10 Series User's Manual -
9.8
Page 97

End unit (OP-26751)

6
When mounted on DIN rail
20.8
(22.6)
5-3 Dimensions
9.2
35.4
53.8
5
Specifications
- Digital Fiber Sensor FS-N10 Series User's Manual -
5-11
Page 98
5
5-3 Dimensions
MEMO
Specifications
5-12
- Digital Fiber Sensor FS-N10 Series User's Manual -
Page 99

Appendix

This chapter provides troubleshooting instructions and initial settings (default values).
6-1 Troubleshooting ..................................................6-2
6-2 Factory Default Setting (Default Value) List........ 6-6
6-3 List of Recipe Function Settings......................... 6-7
6-4 Restrictions on Each Detection Mode ................6-9
6-5 Index.................................................................6-10
6
- Digital Fiber Sensor FS-N10 Series User's Manual -
6-1
Page 100
6
6-1

Troubleshooting

Frequently Asked Questions

Category Trouble Cause Action
•Review the installation environment.
• The received light intensity ca n be stably dis-
The received light intensity is incon­sistent. How can it be stabilized?
Received light intensity display
Appendix
Screen display
Output operation
The received light intensity is too low. How can it be increased?
The received light intensity remains unchanged from 4095 or 9999 even if a workpiece is placed for detec­tion.
When used un conditions of the detec the received light intensity decreases.
Nothing is displayed.
"-" scrolls from left to right on the display
The received light intensity at which the output turns ON and tur is slightly
der extremely short
different.
tion range,
It may be aff ec t ed by vibration, ther- mal characteristics caused by ther- mal changes, or the state of the workpiece surface, etc.
This may be the result of a long detection distance. When using the reflective model, the intensity of the light reflected from the workpiece may be low, or there may be con- tamination. When using t beam model or retro-reflective model, the beam axis may be devi- ated, etc.
The received light intensity has become too large.
Eco (ALL) mode is selected.
The power is cut off or the power cable is disconnected.
Eco (FuLL) mode is selected.
The sleep mode is enabled by the external input.
The output timer is set.
ns OFF
Hysteresis
he thru-
played at 100.0 by using the preset function.
"Preset Function" (page 3-6)
• When us ing the DATUM1/2 mode, the set value and received light intensity are both calibrated to maintain a set ratio.
"Detection Mode" (page 4-8)
•Review the installation environment and detec- tion distance.
• The received light intensity value can be increased by raising (delaying) the power mode.
"Power Modes" (page 4-4)
• The apparent received light intensity can be increased 4-fold by using the light intensity expansion setting.
"Display Gain" (page 4-32)
• Extend the detection range.
•Lower the light emission level and the received light intensity.
"Light emission/Received light intensity
adjustment (Saturation Canceling)" (page 3-18)
"Adjusting Light Emission/Received Light
Intensity (Attenuation)" (page 4-20)
•Disable the Eco function.
"Power Save" (page 4-31)
•Disable the Eco function.
•Disable the Eco function.
"Power Save" (page 4-31)
•Turn the external input OFF, or select a mode other than sleep mode transition input with the external input function.
"Sleep function" (page 4-19)
• Check the output timer setting.
"Output Timer" (page 4-7)
• A slight hysteresis is set to prevent chatter ing in the output. If this level of received light intensity is notable during detection, the detection toler­ance may be low. Please review the detection details.
6-2
- Digital Fiber Sensor FS-N10 Series User's Manual -
Loading...