Read this manual before using the product in order to achieve maximum performance.
Keep this manual in a safe place after reading it so that it can be used at any time.
Danger
Provides additional information on proper operation.
This provides useful tips for the feature being descr ibed.
See "FS-N10 Series User's Manual" for details on the features of the FS-N10 Series and
detailed instructions for configuration.
Hints on Correct Use
Warning
• Do not wire the amplifier line along with power lines or high-tension lines, as the sensor may
malfunction or be damaged due to noise.
• When using a commercially available switching regulator, ground the frame ground terminal
and ground terminal.
• Do not use the FS Series outdoors, or in a place where extraneous light can enter the light-
receiving element directly.
• Due to individual dispersion characteristics and the difference in fiber unit models, the maxi-
mum sensing distance or displayed value may not be the same on all units.
• If the sensor is used for a long time with the APC function enabled and the LED is imposed
with a heavy load, the current consumption of the sensor for light emission will become constant and 'END APC' will be displayed. The sensor can still be used in this case. However,
replace the sensor if even small changes in received light intensity should be detected for
precise detection.
Precautions on Regulations and Standards
UL Certificate
This product is an UL/C-UL Listed product.
• UL File No. E301717
• Category NRKH,NRKH7
• Enclosure Type 1 (Based on UL50)
Be sure to consider the following specifications when using this product as an UL/C-UL Listed Product.
•
Use the power supply with Class 2 output defined in NFPA70 (NEC: National Electrical Code).
•
Power supply/ Control input/ Control output shall be connected to a single Class 2 source only.
•
Use with the over current protection device which is rated 30V or more and not more than 1A.
Failure to follow these instructions may lead to death or serious injury.
Failure to follow these instructions may lead to injury.
Failure to follow these instructions may lead to product damage (product malfunction, etc.)
• This product is just intended to detect the object(s). Do not
use this product for the purpose to protect a human body or
a part of human body.
• This product is not intended for use as explosion-proof
product. Do not use this product in a hazardous location
and/or potentially explosive atmosphere.
• This product uses DC power. Do not apply AC power. The
product may explode or burn if an AC voltage is applied.
FS-N10 Series Quick Start
Quick Start
Fine sensitivity adjustment
DownUp
Mode/Output *2
Press the [MODE] button once, then
to select L-on or D-onuse
Preset function
Configure easily
with a single press
Setting value
Sensitivity setting
Press once each for workpiece/no workpiece
*1 This is a channel switch on 2-output types. This is not equipped with the 0-line type.
*2 Press and hold the [MODE] button to make advanced setting changes.
Light intensity
Power select switch *
Standard ⇔MEGA (fixed)
when receiving light
1
Names of Each Part of the Main Unit and Expansion Unit
Fiber lock lever
SET button (SET)
Operation status indicators
PST indicator
DTM indicator
Digital display
Manual button
Expansion connector
MODE button (MODE)
Preset button (PRESET)
2
Cable*
Dust cover
*1 Setting to "M" locks the power mode to MEGA mode. The switch is a channel switch for the
two output type. This is not equipped with the 0-line type.
*2 On the FS-N1C, this is an M8 connector rather than a cable. On the FS-N1EN, this is
an e-CON connector.
Sub screen
(Displayed in green)
Main screen
(Displayed in red)
( )
Expansion
protective cover
Power select switch*
1
Mounting Unit
Mounting on a DIN Rail
1 Align the claw at the bottom of the main body with
the DIN rail, as shown in figure1.
While pushing the main body in the direction of
the arrow 1, push down in the direction of arrow 2.
2 To dismount the sensor, raise the main body in
the direction of the arrow 3 while pushing the
main body in the direction of arrow 1.
Installation on a Wall (Main Unit Only)
1 Attach the unit to the optional mounting bracket (OP-
73880), and secure with two M3 screws as shown in
figure2.
OP-73880
figure2
Connecting Fiber Unit
1 Open the dust cover in the direction shown by
arrow 1.
2 Move the fiber lock lever in the direction shown
by arrow 2.
3 Insert a fiber unit into the amplifier as indicated
by arrow 3 (approximately 14 mm).
4 Move the fiber lock lever in the direction shown
by arrow 4 to secure the fiber.
4
2
3
1
Fiber insertion sign
Included accessories
• Instruction Manual 1pc.
1
Page 2
.
Note
φ1.3
φ1.0
Adapter A
(OP-26500)
Adapter B
(OP-26501)
Cable outer diaAdapterAppearance
Note
OP-26751 (a set of two)
Workpiece
Workpiece
Workpiece
Common to
Thrubeam
and
Models
Reflective
Workpiece
• If a thin fiber unit is used, an adapter provided with the thin fiber unit will be
required.
Unless the correct adapter is connected, the thin fiber unit will be loose and
not detect targets correctly (the adapter is supplied with the fiber unit).
• To connect the coaxial reflective type fiber unit to the amplifier, connect the
single-core fiber to the transmitter side, and connect the multiple-core fiber
to the receiver side.
Other Calibration Methods
Increased Resistance to Dust and Dirt
Maximum Sensitivity Calibration
z
In the state shown below, press and hold the [SET] button for three seconds or more. Stop
pushing when "5'6" flashes.
The sensitivity is set slightly higher than the received light intensity.
Thrubeam model
Workpiece
: with
workpiece
Press and hold for 3 seconds or more
Single-core fiber
Multi-core fiber
Transmitter
Receiver
Connecting Multiple Amplifiers
Up to 16 expansion units can be connected to one main unit. However, two output types will be
treated as two main units.
Warning
• When connecting with units other than N-bus (a general term for the KEY-
• Turn the power off before connecting multiple expansion units.
• D o not touch the expansion connector with your bare hands.
1 Remove the protection covers from the main unit
and expansion unit(s).
Mount on DIN rail and install on metal surface when connecting to
multiple amplifiers or mounting main units together.
ENCE wire-saving connection system) compatible sensor amplifiers,
including the FS-N10 Series, and the network unit NU Series, consult your
nearest KEYENCE dealer.
2 Install the amplifiers on the DIN rail one at a time.
3 Slide the main unit and expansion unit(s) together.
Engage the two claws of the expansion unit with
the recesses on the main unit side until you hear/
feel a click.
4 Attach the end units (option: OP-26751) to the DIN
rail in the same way as step (2).
5 Secure the amplifiers between the end units.
Tighten the screws at the top (two screws x two
units) with a Phillips screwdriver to fix the end
units.
Reflective model
Calibrate with a Moving Workpiece
Fully Automatic Calibration
z
Press and hold the [SET] button with no workpiece in place. While “5'6” is flashing, pass a
workpiece through. (Continue pressing the [SET] button while the workpiece passes through.)
Common to
Thrubeam
and
: without
Reflective
workpiece
Models
Press and hold for 3 seconds or more
Press and hold until
"
UGV
" flashes
While pressing the [SET] button
pass through a workpiece
Settings complete
Position Workpiece
Positioning Calibration
z
Press the [SET] button with no workpiece.
Place the workpiece in the location you wish to position it. Press and hold the [SET] button
for at least 3 seconds. Release the button when "5'6" flashes.
Press [SET] button once with
no workpiece
Calibration Method
Detecting Even Small Differences
Two-point Calibration
z
Two-point calibration is the basic method of calibration.
Press the [SET] button once without the workpiece, and then press it once again with the
workpiece.
1 Press the [SET] button with no workpiece.
[SET] will be displayed on the sub-menu (green display).
2 Press the [SET] button with workpiece.
The values will be set and the submenu (green display) will flash. The values will be set to
the mid-point between the light intensity when there is no workpiece, and the light intensity
when there is a workpiece.
If "----" flashes for two seconds on the main screen, the light intensity is too small between
conditions when the workpiece is absent and when it is present. These values will be set,
but there is the possibility that detection may become unstable.
Press and hold [SET] button with
workpiece
Press and hold until
"
UGV
" flashes
Settings complete
Simple, User Friendly Functions
Setting the Current Value to 100.0
With the FS-N10 Series, you can set the current value to 100.0 using simple operations.
Standardizing the current value makes it possible for the sensor amplifiers to instantly differentiate reductions in light intensity and is useful in predicting the need for maintenance.
Note
• The various Preset functions listed below cannot be used when the ZeroShift function is enabled. Disable the Zero-Shift function before executing
the following functions.
• The Preset functions are not suited for transparent workpieces and other
cases of detection with low light intensity differences.
You can disable various Preset functions by pressing and holding the [PRESET]
button.
Tip
2
Page 3
Preset Function
Important
Green "PST"
lights up
Current value is ""
100.0
[
PRESET] +
3.0
.0
Workpiece
Workpiece
Thrubeam model : with workpiece
: without workpiece
Press and hold for 3 seconds or more
Press and hold for 3 seconds or more
Reflective model
Note
・
Workpiece
Press and hold until
"#WVQ" flashes
While pressing the [PRESET] button
pass through a workpiece
Settings complete
Common to Thrubeam and Reflective Models
Press the [SET] button while
pressing the [MODE] button.
z
This function adjusts the current value to "".
With light received, press the [PRESET] button. The current value is set to "".
Press [PRESET] button once
with no
workpiece
Green "PST" lights up
"
Current value is "
Setting is "
Pressing the [PRESET] button changes the setting and current value as shown below.
• Presetting with preset disabled:
The setting is changed to "". The setting can be changed via the normal calibration
method.
• Presetting with preset enabled:
Only the current value is changed to " ", and the setting is not changed.
.
Handy Uses for the Preset Function
This function is most useful when performing simple detection using a thrubeam model fiber unit (e.g. completely blocked detection, such as when all
Tip
light axes of the fiber unit are interrupted by opaque workpieces).
Work-Preset Function
z
This function adjusts the current value to "".
After executing the Preset function in a condition in which you would like "" to be displayed, executing this function in a condition in which you would like "" to be displayed, will
adjust any two points to "" and "".
The Work-Preset function can be used while the Preset function is in use
(when Preset is enabled).
Pressing the [PRESET] button and the button at the same time will set the current
value at that time to "". Values that have been set to "" with the Preset function cannot
be changed.
Full Auto Preset Function
z
This function automatically differentiates between two conditions (presence/absence of
workpiece) and adjusts the current values to "" and "". This is effective for cases when
the workpiece is moving at high-speed.
Press and hold the [PRESET] button with no workpiece in place. While "#WVQ" is flashing,
pass a workpiece through. (Continue pressing the [PRESET] button while the workpiece
passes through.)
"
• Near-maximum values while the [PRESET] button is being pr essed and held are
adjusted to "" and near-minimum values are adjusted to "" .
• The setting value is changed to ""
• The green [PST] indicator will light up.
Note
Cannot be executed when the Preset function is already being used (when
the [PST] indicator is flashing). Press and hold the [PRESET] button to disable the Preset function before executing this function.
Set Current Value to "0"
The Zero Shift Function
z
This function is primarily used with reflective models.
Press the [PRESET] button and button at the same time.
When using this function with reflective models, "" will be displayed when
there is a workpiece, and "" will be displayed when there is no workpiece,
making it easy to know when the workpiece is present or absent. Additionally,
Tip
even when with a reflective model, the background has higher light intensity
than the workpiece, if you set a condition with low light intensity to "" using
the Preset function and then using the Work-Preset function, register a condition with high light intensity as "", the background will display as "" and when
the workpiece is present, it will be displayed as "".
Maximum Sensitivity Preset Function
z
This function sets conditions that will serve as reference, to "" and adjusts conditions with
slightly high light intensity as "". This is useful when you would like to perform detection
using the background as a reference with reflective models.
In the following conditions, press and hold the [PRESET] button for 3 seconds or more then
release your finger when "#WVQ" is flashed.
• The maximum value for the light intensity while the [PRESET] button is being pressed is
set to ", and light-intensity that is slightly higher than the maximum value at that time
will be adjusted to "".
• The setting value is ""
• The green [PST] indicator will light up.
Cannot be executed when the Preset function is already being used (when
the [PST] indicator is flashing). Press and hold the [PRESET] button to disable the Preset function before executing this function.
• The current value is set to "".
• Green "PST" lights up
Note
The zero shift and preset function cannot be used together. To use the
preset function, you must first disable the zero shift function.
Disable the Zero Shift Function
z
Press and hold the [PRESET] button to disable the zero shift function.
Handy Uses for the Zero Shift Function
This function is primarily used to set the current value to "0" on a reflective
model fiber unit.
Tip
When a reflective model is first installed, the light intensity is sometimes not
set to "0".
If this happens, using the zero shift function to set the value to "0" when there is
no workpiece allows for easier understanding of the difference in light intensity.
Adjusting the current intensity value when it is too large
(when saturated).
Use the Saturation Recovery Function
z
Press the [SET] button while pressing the [MODE] button.
After adjusting the light transmission level and light intensity sensitivity, the current values
will be adjusted to within the ranges listed in the table that follows.
Power modeLight intensity setting range
HSP*, FINE, TURBO2047 ± 350
SUPER4095 ± 500
ULTRA, MEGA5000 ± 600
*HIGH SPEED
Disable Saturation Recovery
z
When the saturation recovery function is enabled, press the [SET] button while pressing the
[MODE] button to cancel it.
Handy Uses for the Saturation Recovery Function
This function is useful when the intensity value is saturated after installation.
This function corrects the saturation via a simple operation, by automatically
Tip
calibrating the light transmission level and light intensity gain.
3
Page 4
Output Switch
0 V
Brown
Black
Monitor output (1 to 5V)
Orange
Load
Blue
(Control output)
12 to 24 VDC
Device with input
impedance
10 kΩ or more
Protection
circuit
Sensor main circuit
Overcurrent
protection circuit
*1
*1
DC3.3V
Sensor main circuit
Overcurrent
protection circuit
12 to 24 VDC
PLC, etc.
0V
(Short-circuit
current 1 mA
or less)
PLC, etc.
0V
(Short-circuit
current 1 mA
or less)
(Control output)
Load
123
4
M8 connector Pin layout
FS-N11CN/N12CN
e-CON connector Pin layout
FS-N11EN/N12EN
123
4
M8 connector Pin layout
1
2
3
4
Connected
pin No.
Pin - Pin and wire color table
1Brown
2
3
4
White
Blue
Black
Wire color
Initializing the Settings
Either light-ON (L-on) mode or dar k-ON (D-on) mode can be selected.
1 While the current value is displayed, press the [MODE] button once.
2 Use to switch the output (.QP/&QP), then press the [MODE] button again.
The output change completes, and the display returns to the current value.
1 Press and hold the [SET] and [PRESET] buttons simultaneously for three seconds.
Press and hold for 3 seconds or more
2 Use the to select "456", then press the [MODE] button.
3 Use the to select "K0K6", then press the [MODE] button.
After initialization is complete, the display returns to the current value.
Initial Settings
z
SettingInitial Value
Power modeFINE
Detection modeStd (Normal)
Setting value50
Output switchL-on
Using a Fiber Cutter and Cautions for Use
Using a Fiber Cutter
1 Insert the fiber into the cutter hole.
2 Bring down the blade in a single,
swift motion to cut the fiber.
Fiber
1
Always insert fiber from the side with writing
Cautions for Using a Fiber Cutter
• The fiber cutter comes with the fiber unit.
Failure to follow the cautions below could reduce the detection range.
• When cutting a fiber unit to be attached to the FS-N10 Series, be sure to use a gray fiber cut-
ter (OP-87098)
• Stopping the blade midway could cause a bad cut plane, reducing the detection range.
• Do not use the same hole twice.
2
Fiber cutter
(OP-87098)
M8 connector Cable (Sold Separately)
For FS-N11CN/N11CP/N12CN/N12CP/N13CP/N14CP
OP-73864
(Cable length: 2 m)
OP-73865
(Cable length: 10 m)
Error Displays and Corrective Actions
Error displayCauseSolution
ErCOvercurrent in the control output.
ErEFailed to write/load the internal data. Perform initialization (p.4).
End APCLarge load on the light source.
LocThe keylock function is ON.
Consult your nearest KEYENCE office regarding error displays other than the ones listed above.
Check the load and return the current
within the rated value.
Replace the sensor if highly precise
detection is required.
For disabling (setting) method, see
"FS-N10 Series User's Manual".
Function Configuration
Basic Setting
Press and hold for 3 seconds or more
MODE
MODE
MODE
Return to normal display
*1 You can press the button to set between the range of 2 to 2.
J52
5'6
MODE
(KPG
6WT$
5W24
7.64
/')#
5VF
5'6
5'6
'0&
5V)
5V)
5V)
MODE
5'V2
15'V
(W0E
&K52
5;5
HIGH SPEED mode
FINE mode
TURBO mode
SUPER mode
ULTRA mode
MEGA mode
Normal sensitivity setting method
Percentage Calibration*
Zero-shift calibration
MODE
Settings complete
Go to detection setup mode
Go to display setup mode
Go to system setup mode
4
1
Page 5
Detection Settings
Normal (light intensity)
detection mode
DATUM1 mode *
2
Timer OFF
Off-delay timer *
1
On-delay timer *
1
One-shot timer *
1
DATUM2 mode *
2
Area detection mode
Rising Edge Detection Mode
Falling Edge Detection Mode
Settings complete
Go to display setup mode
Go to system setup mode
Return to original
detection setup mode
External input off
External calibration input
Preset input
Zero shift input
Reset input
Light transmission OFF input
Pause mode transition input *
3
Sleep mode transition input
Light emission power selection *
4
Analog scaling *
5
Return to normal display
*6
*7
MODE
Normal display method
Reverse display
Settings complete
Go to system setup mode
Go to detection setup mode
Return to original display setup mode
Sub-display off
Extended display
Bar display
Excess gain display
Light intensity hold display *
1
Excess gain hold display *
1
L-on / D-on display
Disable the saturation of the
Preset function
Enable the saturation of the
Preset function *
2
Return to normal display
Light intensity detection mode
Limit setting output mode
Auto reset
User reset
Warning output mode
Timer OFF
Off-delay timer *
1
On-delay timer *
1
One-shot timer *
1
Settings complete
Return to normal display
System Settings
*1
Return to normal display
*1 Main unit only.
Two Out put
Disable APC
Enable APC
Eco feature off
Enable eco feature
Reduce power consumption
(response time 4 times slower)
Standard current value display
Maximum current value display
(4 times hysteresis)
Normal operation
Twice the number of
interference-prevention units as
(response time 2 times slower)
Disable common key operations
Enable common key operations
Settings complete
Go to detection setup mode
Go to original display setup mode
Return to system setup mode
56&
*2
*1 Press the button to set between the range of and (ms).
*2 Press the button to set the retouch sensitivity to a range of between .'W and .'W
and set the warning output level to a range of between 12 and 2.
*3 Press the button to toggle between Q((/Q0/-''2.
*4 Can be set between the range of and .
*5 Can be set between the range of and .
*6 Can be used only for the types with the external input.
Note that these functions can be used via communication when connecting with the network
unit NU Series.
*7 Only monitor output types (FS-N11MN).
Display Settings
*1 Press the button to toggle between 5VF/2`2A/D`DA/2AD`/2`DA
*2 Press the button to set between the range of 2 and 2.
MODE
MODE
MODE
MODE
*1 Press the button to set between the range of and (ms).
MODE
*2 Only 2 output types (FS-N13N/N13P/N14N/N14P/N13CP/N14CP).
5
Page 6
Specifications
Copyright (c) 2010 KEYENCE CORPORATION. All rights reserved.
N/A
Monitor output (1 to 5 V)----------1 outputExternal input--1 input1 input1 input1 input1 input1 input---Light source LEDRed 4-element LED (wavelength 630 nm)
Response time
50 μs (HIGH SPEED)/250 μs (FINE)/500 μs (TURBO)/1 ms (SUPER)/4 ms (ULTRA)/16 ms (MEGA)
Output toggleLight-ON/dark-ON toggle
Timer functionTimer OFF, OFF delay, ON delay, One-shot
Control
output
Monitor output
External input
NPN outputNPN open collector 24 V; 1 output max: 100 mA or less; 2 output total:
PNP outputPNP open collector 24 V; 1 output max: 100 mA or less; 2 output total:
*3
1 to 5 V voltage output; load resistance 10 kΩ or more; repeat precision ± 0.5% of F.S.; 1 ms response time (HIGH SPEED, FINE, TURBO)
100 mA or less (used stand-alone)/20 mA or less (multiple connections); residual voltage 1 V or less
100 mA or less (used stand-alone)/20 mA or less (multiple connections); residual voltage 1 V or less
input time 2 ms (ON)/20 ms (OFF) or more
*5
*4
Expansion UnitsUp to 16 units can be connected (total of 17 units including the main unit). Note that the two-output type is counted as two units.
Protection circuitProtection against reverse power connection, output overcurrent, and output surge
Number of interference
prevention units
Rating
Powe r voltage
NPN
PNP
0 for HIGH SPEED; 4 for FINE; 8 for TURBO/SUPER/ULTRA/MEGA (When set to double, the number of interference-prevention units will be doubled.)
12 to 24 V DC ± 10% ripple (P-P) 10% or less
Normal:900 mW or less (36 mA max. at 24 V, 48 mA max. at 12 V)
Eco on mode:800 mW or less (32 mA max. at 24 V, 39 mA max. at 12 V)
Eco Full mode:470 mW or less (19 mA max. at 24 V, 23 mA max. at 12 V)
Normal: 950 mW or less (39 mA max. at 24 V, 52 mA max. at 12 V)
Eco on mode: 850 mW or less (35 mA max. at 24 V, 44 mA max. at 12 V)
Eco Full mode: 520 mW or less (21 mA max. at 24 V, 26 mA max. at 12 V)
Normal: 1050 mW or less (42 mA max. at 24 V,
*6
Eco on mode : 950 mW or less (38 mA max. at 24 V,
*6
56 mA max. at 12 V)
47 mA max. at 12 V)
Eco Full mode : 600 mW or less (24 mA max. at 24 V,
*6
*6
*6
*6
-
29 mA max. at 12 V)
Environme
ntal
resistance
Operating ambient
luminance
Operating ambient
temperature
Operating ambient
humidity
Vibration
resistance
Shock resistance
Incandescent lamp: 20,000 lx or less; Sunlight: 30,000 lx or less
-20 to +55 °C (no freezing)
*7
35 to 85% RH (no condensation)
10 to 55 Hz Compound amplitude 1.5 mm, 2 hours for each of X,Y,Z axis
500 m/s2 3 times for each of X,Y,Z axis
Case materialBoth main unit and expansion unit housing material: Polycarbonate
Case dimensionsH30.3mm x W9.8mm x L71.8mm
WeightApprox 75g Approx 45g Approx 22g Approx 22g Approx 22g Approx 22g Approx 80g Approx 70g Approx 22g Approx 22g Approx 75g Approx 20g
*1 Use a cable length of 30 m or less for M8 connector and e-CON connector types.
*2 Counted as 1 output when connecting with the network unit NU Series.
*3 FS-N11MN only
*4 SUPER : 1.2 ms, ULTRA : 1.8ms, MEGA : 4.2 ms
*5 Input time is 25 ms (ON)/25 ms (OFF) only when external calibration input is selected.
*6 Increases 100 mW (4.0 mA) for High Speed mode
*7 One or two more units connected: -20 to +55°C; 3 to 10 more units connected: -20 to +50°C; 11 to 16 more units connected: -20 to +45°C. When using 2-outputs, one unit is counted as two units.
All temperature regulations are for when the unit is mounted on a DIN rail and installed on metal sheeting.
WARRANTIES AND DISCLAIMERS
(1) KEYENCE warrants the Products to be free of defects in materials and workmanship for a
period of one (1) year from the date of shipment. If any models or samples were shown to
Buyer, such models or samples were used merely to illustrate the general type and quality
of the Products and not to represent that the Products would necessarily conform to said
models or samples. Any Products found to be defective must be shipped to KEYENCE with
all shipping costs paid by Buyer or offered to KEYENCE for inspection and examination.
Upon examination by KEYENCE, KEYENCE, at its sole option, will refund the purchase
price of, or repair or replace at no charge any Products found to be defective. This warranty
does not apply to any defects resulting from any action of Buyer, including but not limited to
improper installation, improper interfacing, improper repair, unauthorized modification,
misapplication and mishandling, such as exposure to excessive current, heat, coldness,
moisture, vibration or outdoors air. Components which wear are not warranted.
(2) KEYENCE is pleased to offer suggestions on the use of its various Products. They are only
suggestions, and it is Buyer's responsibility to ascertain the fitness of the Products for
Buyer’s intended use. KEYENCE will not be responsible for any damages that may result
from the use of the Products.
(3) The Products and any samples ("Products/Samples") supplied to Buyer are not to be used
internally in humans, for human transportation, as safety devices or fail-safe systems,
unless their written specifications state otherwise. Should any Products/Samples be used in
such a manner or misused in any way, KEYENCE assumes no responsibility, and
additionally Buyer will indemnify KEYENCE and hold KEYENCE harmless from any liability
or damage whatsoever arising out of any misuse of the Products/Samples.
(4) OTHER THAN AS STATED HEREIN, THE PRODUCTS/SAMPLES ARE PROVIDED
WITH NO OTHER WARRANTIES WHATSOEVER. ALL EXPRESS, IMPLIED, AND
STATUTORY WARRANTIES, INCLUDING, WITHOUT LIMITATION, THE WARRANTIES
OF MERCHANTABILITY, FITNESS FOR A PARTICULAR PURPOSE, AND NONINFRINGEMENT OF PROPRIETARY RIGHTS, ARE EXPRESSLY DISCLAIMED.
IN NO EVENT SHALL KEYENCE AND ITS A FFILIATED ENTITIES BE LIABLE TO ANY
PERSON OR ENTITY FOR ANY DIRECT, INDIRECT, INCIDENTAL, PUNITIVE, SPECIAL
OR CONSEQUENTIAL DAMAGES (INCLUDING, WITHOUT LIMITATION, ANY
DAMAGES RESULTING FROM LOSS OF USE, BUSINESS INTERRUPTION, LOSS OF
INFORMATION, LOSS OR INACCURACY OF DATA, LOSS OF PROFITS, LOSS OF
SAVINGS, THE COST OF PROCUREMENT OF SUBSTITUTED GOODS, SERVICES OR
TECHNOLOGIES, OR FOR ANY MATTER ARISING OUT OF OR IN CONNECTION
WITH THE USE OR INABILITY TO USE THE PRODUCTS, EVEN IF KEYENCE OR ONE
OF ITS AFFILIATED ENTITIES WAS ADVISED OF A POSSIBLE THIRD PARTY’S
CLAIM FOR DAMAGES OR ANY OTHER CLAIM AGAINST BUYER. In some
jurisdictions, some of the foregoing warranty disclaimers or damage limitations may
not apply.
BUYER'S TRANSFER OBLIGATIONS:
If the Products/Samples purchased by Buyer are to be resold or delivered to a third party,
Buyer must provide such third party with a copy of this document, all specifications,
manuals, catalogs, leaflets and written information provided to Buyer pertaining to the
Products/Samples.
E 1110-2
6
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.