IKA KS 501 User Manual

0 (0)
®
24 760 25
IKA
IKA® HS 501 digital IKA
®
KS 501 digital
IKA® HS 501 数显型 振荡器 IKA
®
KS 501 数显型 振荡器
Table of Contents Page
Spare parts diagram 2 Wiring diagram 2 List of spare parts 3 Guarantee 3 Scheduled use 4 Unpacking 4 Setting-up the instrument 4 Starting-up 4 Attachments 5 Maintenance and cleaning 7 Technical data 7
目录 页码
零配件图 2 接线图 2 零配件清单 3 保修 3 正确使用 8 开箱 8 安装 8 开机 8 夹具 9 清洁维护 11 技术参数 11
Spare parts diagram & Wiring diagram / 零配件图和接线图
HS_KS 501 digital_092012
2
List of spare parts / 零配件清单
ldent-No. of Instrument 24 760 00
Item Designation 1 bottom plate 2 housing cover 3 shaking table 4 frame 1 5 frame 2 10 leaf spring 17 intermediate sheet metal 21 poly-V-belt 22 operating knob poti. 23 potentiometer 24 timer 25 operating knob timer 26 front part 27 identification plate 28 built-in plug
仪器ID: 24 760 00 序号 名称 1 底板 2 机壳盖 3 振荡台 4 机架 1 5 机架 2 10 片簧 17 金属间片 21 聚-V形带 22 操作旋钮电位计 23 电位计 24 计时器 25 操作旋钮计时器 26 仪器前板 27 标识牌 28 内置插头
Item Designation 29 instrument fuse 38 fork light barrier 39 PCB light barrier 40 PCB mains 46 flat strip line 47 distance holder 53 washer 54 washer 74 edge protection 86 grooved ball bearing 6004 2Z 87 grooved ball bearing 6005 2Z 88 grooved ball bearing 6008 2Z 2001 drive compl. 2002 base compl.
序号 名称 29 仪器保险丝 38 交叉式挡光板 39 PCB 挡光板 40 PCB 电源 46 平板带线 47 距离支架 53 垫片 54 垫片 74 边缘保护 86 深沟球轴承 6004 2Z 87 深沟球轴承 6005 2Z 88 深沟球轴承 6008 2Z 2001 马达组件 2002 底座组件
HS_KS 501 digital_092012
Guarantee / 保修
®
You have acquired an Original IKA
-Laboratory Instrument which meets the highest requirements of quality and disign. In accordance with the IKA®-terms of warranty, this instrument is guaranteed for a period of 24 months. In the event of any claim under guarantee, please contact your laboratory dealer.
感谢您采购IKA®公司高品质的产品。 根据IKA®公司规定本机保修两年。 保修期内若有任何问题请联系您的供货商。
3
Scheduled Use
®
By using different attachments, the IKA
-shakers HS 501 digital and KS 501 digital are suitable for mixing liquids in bottles, flasks, separating funnels, test tubes and dishes with a maximum shaking weight of 15 kg. The HS 501 digital produces a to and fro shaking motion and the KS 501 digital an orbital one.
Unpacking
Please unpack the instrument carefully and look out for any damage. It is important to recognize a transportation damage already when unpacking. If this is the case, an immediate factual report is required (mail, rail or forwarding agency).
Setting-up the Instrument
Please take care for the instrument to stand unobstructed. Attention has to be paid to the ambient conditions (temperature and humidity) listed in Technical Data. Any direct exposure to the sun should be avoided.
Safety Warning:
The instrument has to be placed upon a sturdy laboratory table with level surface and with no other objects on it. The instrument must not be operated in rooms with explosion hazards.
Starting-up
Before starting the instrument, please check the conformity between the voltage indicated on the ID-plate and your mains voltage. If they do not agree, the instrument must not be operated. The instrument has to be connected to an earthed socket.
The shaking intensity is selected on the front side potentiometer “Mot“.
Attention:
As the instrument is starting with the set speed, we recommend to set the potentiometer “Mot“ to the left-hand stop (minimum speed) before starting it.
The instrument is switched on by turning the timer rotary knob clockwise. Position I is for permanent operation. The duration of operation can be set infinitely variably
from 0 - 56 minutes. The timer rotary knob can be reset any time, yet it may not exceed position 0.
By turning the timer rotary knob to position 0, the instrument is switched off.