HP PageWriter 100 User manual

Page 1
PageWriter 100 HP M1772A Cardiograph
User’s Guide
m
HP Part No. M1772-91900
Printed in USA February 1996 Edition 3
E0296
PACKARD
Page 2
Notice
The information in this document is subject to change without notice.
Hewlett-Packard makes no warranty of any kind on this material, including but not limited to, the implied warranties of merchantability and fituess for a particular purpose. Hewlett-Packard shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.
This document contains proprietary information which is
protected by copyright. All rights are reserved. No part of this document may be photocopied, reproduced, 01 translated to another language without the prior written consent of Hewlett-Packard Company.
Before using the instrument, read this guide and become thoroughly familiar with the contents.
Responsibility of the Manufacturer Hewlett-Packard only considers itself responsible for
any effects on safety, reliability and performance of the equipment if:
assembly operations, extensions, re-adjustments, modifications or repairs are done by persons authorized by Hewlett-Packard, and
the electrical installation of the relevant room complies with the IEC or national requirements, and
the instrument is used according to the instructions
for use presented in this manual.
Page 3
Warning
@
As with all electronic equipment, radio frequency interference between this cardiograph and any existing RF transmitting or receiving equipment at the installation site, including electrosurgical equipment, should be evaluated carefully and any limitations noted before the equipment is placed in service. Monitoring during electrosurgery should not be attempted and monitoring electrodes should be removed from the patient to preclude the possiblity of burns. Radio frequency generation from electrosurgical equipment and close proximity transmitters may seriously degrade cardiograph performance. Hewlett-Packard assumes no liability for failures resulting from RF interference between HP medical electronics and any radio frequency generating equipment at levels exceeding those established by applicable standards.
This is to certify that this equipment is in accordance with the Radio Interference Requirements of the EMC Directive.
Warning
Use of accessories other than those recommended by Hewlett-Packard may compromise product performance and/or product safety.
THIS PRODUCT IS NOT INTENDED FOR HOME
USE.
@ Copyright 1994, 1995 Hewlett-Packard Company.
Page 4
Printing History
September 1994 April 1995 Edition 2
Edition 1
February 1996
Edition 3
iv
Page 5
Safety Summary
Safety Symbols
The following safety symbols are used on the
Marked on the cardiograph.
Cardiograph
Caution - See operating instructions.
!
A
Meets IEC type CF leakage current requirements and is defibrillator protected (Isolated ECG input).
Alternating current.
Equipotential (identifies independent protective earth conductor to the cardiograph).
Fuse.
Indicates power control for cardiograph.
Hz
Please see “Patient and Operational Safety Notes” in Chapter 1, Getting Acquainted, for further information about operating your cardiograph safely.
Indicates operating frequency in cycles per second.
V
Page 6
Conventions Used in This Manual
Warning
Caution
Note
9 IF)
I
Yb
Warning statements describe conditions or actions that can result in personal injury or loss of life.
Caution statements describe conditions or actions that can result in damage to the equipment or software.
Notes contain additional information on cardiograph usage.
@ Represents keys on ‘the front panel.
vi
Page 7
Contents
1. Getting Acquainted The Keyboard and Front Panel About Your Cardiograph
Accessories About This Manual Patient and Operational Safety Notes . . AC and DC (Battery) Operation
2. Recording an ECG Preparing the Patient Understanding When a Signal is Acquired Recording an Auto ECG
Correcting a Leads Off Condition Auto Report Filters Making Copies of Auto ECGs Recording a Manual ECG
Changing Manual Report Settings . .
Restoring the ECG Trace After
Defibrillation or Reconnecting Leads Manual Report Filters Understanding the Printed Report Choosing a Report Format
Auto Report Format Manual Report Formats An Auto Report Example Manual Report Examples
............
..........
.........
........
..........
.........
........
.....
.......
.....
.......
.......
.......
......
......
....
...
...
1-4 1-6 1-6 1-7 1-8
1-10
2-2 2-3 2-4 2-4 2-5 2-6 2-7 2-8
2-8
2-9 2-10 2-13 2-13 2-13 2-14 2-15
Contents-1
Page 8
3. Troubleshooting Checking ECG Technique Identifying ECG Problems
If the Recording Won’t Start If the Cardiograph Won’t Print a
Manual Report
4. Maintaining the Cardiograph Care and Cleaning
Cleaning the Cardiograph Cleaning the Electrodes and Cables . .
Cleaning the Digital Array Printhead
and Paper Sensor
Loading the Paper
Storing the Paper and Patient ECG
Records
Caring for the Battery
Storing the Battery Replacing the Fuses
Supplies ...............
Calling for Service
United States of America
Canada ..............
Other International Areas
..........
..........
............
..........
..........
.......
.......
.........
.......
........
.........
.....
......
......
......
3-1 3-2 3-3
3-5
4-1 4-1 4-2
4-3 4-5
4-6 4-8 4-9
4-9 4-11 4-14 4-14 4-14 4-14
Contents-2
A. Setting up Your Cardiograph for the First Time
Checking the Voltage Setting
The Battery .............
Installing the Battery
Removing the Battery Connecting the Cables Loading Paper
............ A-9
........
........
........ A-7
......
A-2 A-4 A-4 A-5
Page 9
;a, , . .,I< :., .*.
1,: ;’
.:
B. Specifications
Basic Controls . . . . . . . . . . . .
Frequency and Impulse Response . . . .
Instrument Test . . . . . . . . . . .
Patient Safety . . . . . . . . . . . .
Power and Environment . . . . . . . .
Glossary
Index
I,‘.
B-1 B-1 B-2 B-2 B-2
Contents-3
Page 10

Figures

Tables

1-1. The PageWriter 100 Cardiograph.
1-2. Bottom View of Cardiograph . . . . .
1-3. The Keyboard and Front Panel of the
Cardiograph. . . . . . . . . . .
2-1. The Auto ECG Report. . . . . . . .
2-2. An Auto 3x4 ECG with One Rhythm
Strip(3x4, 1R). . . . . . . . . .
2-3. A Manual S-Lead ECG. . . . . . . .
2-4. A Manual 6-Lead ECG. . . . . . . .
4-1. Cleaning the Digital Array Printhead. .
4-2. Loading the Paper. . . . . . . . . .
4-3. The AC Fuse Holders. . . . . . . . .
A-1. Rear View of Cardiograph . . . . . .
A-2. Removing the Battery Door. . . . . .
A-3. The Battery Compartment. . . . . .
A-4. Connecting the Power Cord. . . . . .
A-5. Connecting the Patient Cable. . . . .
A-6. Loading the Paper. . . . . . . . . .
. .
1-2 1-3
1-4
2-10 2-14
2-15 2-16
4-3 4-5
4-10
A-3 A-5 A-6 A-7 A-8 A-9
Contents-4
2-1. Leads Off Labels . . . . . . . .
2-2. Auto Filter Settings . . . . . . .
2-3. Manual Filter Settings . . . . .
2-4. Auto ECG Report Annotations . .
2-5. Calibration Signals . . . . . . .
2-6. Manual Lead Groups . . . . . .
2-5 2-5
2-9 2-11 2-12 2-13
Page 11
.:.. ‘,
3-1. ECG Problems and Solutions . . . .
3-2
Contents-5
Page 12
Getting Acquainted
Page 13

Getting Acquainted

This chapter describes to the new user the many features of the PageWriter 100 cardiograph, patient and operational safety, and AC/battery operation. The user should become familiar with this material, especially the safety information, prior to using the cardiograph.
1
Note I
!i?
See Appendix A, Setting up Your Cardiograph for the First Time for information on checking the voltage switch setting, installing the battery, connecting the cables, and loading paper. Each of these tasks must be done prior to operating the cardiograph for the first time.
Getting Acquainted 1-1
Page 14
Figure 1-l. The PageWriter 100 Cardiograph.
1-2 Getting Acquainted
A. Patient Cable
B. Cardiograph
Page 15
Figure 1-2. Bottom View of Cardiograph
A. AC Fuse Holders
B. Mounting Point for M1705B Cart
(Mounting screw included with cart)
C. Battery Door
Getting Acquainted 1-3
Page 16

The Keyboard and Front Panel

- /qEigi!~,II,~,
Figure 1-3. The Keyboard and Front Panel of the Cardiograph.
1-4 Getting Acquainted
-
M1772-01
,
Page 17
A
Instructions
Attention:
Ij
This text describes how the front-panel lights indicate the operating status of the cardiograph:
A flashing
When the light stays on, the acquisition is complete.
Auto
light indicates that the ECG is being acquired.
B AC C Low Battery D l-1
(g giiqz-SpeedJ 0 pzzSize_) (TJ (\/) (zJ=
A flashing A flashing Manual lead group light indicates an electrode
connection problem.
Light will be on when the power cord is plugged into AC power. This also indicates that the battery, if installed, is charging. This light indicates that the cardiograph must be plugged in to recharge the battery. Switches the cardiograph between On and means the,cardiograph is off but it is still keeping the battery charged as long as the cardiograph is plugged into AC power. Sequentially changes the chart speed from 5, j 10, =+ 25, 3 50, =+ 5 mm/second. Sequentially changes the limb and chest lead sensitivity from 5, z+ 10, 3 20, * 5 mm/mV. Reduces the chest lead sensitivity to 50% of the value set by
g?CTiZSize). Turns the Baseline Wander and Noise filters on or off. See “Auto Report Filters” in Chapter 2, and “Manual Report Filters” in Chapter 2 for more information. Advances the paper to the beginning of the next page. Starts an Auto ECG recording. Starts a Manual ECG recording. Sequentially changes the lead group used to generate a Manual ECG from I II III, 3 aVR aVL aVF, j II aVF V2, j Vl-V6, j I II III. Halts any cardiograph function. Prints a copy of the last Auto ECG. If you want additional copies of an Auto ECG, you must print them before recording another Auto or Manual ECG.See “Making Copies of Auto ECGs” in Chapter 2,
Chart Speed
Recording
an
light indicates a paper supply problem.
ECG
for more information.
Standby.
Standby
Getting Acquainted 1-5
Page 18

About Your Cardiograph

Your PageWriter 100 cardiograph:
l
Acquires 12 leads simultaneously.
n
Provides selectable Manual formats.
n
Operates on a rechargeable battery. AC power charges the battery.
n
Has a digital array printer with continuous-feed paper.
n
Has a 200 sheet Z-fold paper capacity.

Accessories

Your cardiograph was shipped with one of three accessory sets, according to your geographic option:
No Electrodes - Options: ABB, ABD, ABE, ABF, ABH, ABS, ABU, ABX, ABZ, ACQ
n
Power cord
n
Patient Cable
n
1 package of paper
n
Page Writer 100 User’s Guide
n
Using the HP Page Writer 100 Cardiograph operator
training video
For electrodes, contact your local Hewlett-Packard Sales
Office or your authorized Hewlett-Packard Dealer or
Distributor. Reusable Electrodes - Options: ABG, ABK, ABM,
ABZ, AB4, AKV
n
Power cord
n
Patient Cable
n
1 package of paper
n
6 Welsh bulb electrodes
n
4 limb electrodes and straps
n
1 tube Red&@ creme
n
Page Writer 100
n
Using the HP Page Writer 100 Cardiograph operator
User's
Guide
training video
1-6 Getting Acquainted
Page 19
3 ,, I’
Disposable Electrodes - Options: ABA, ABC I Power cord
n
Patient Cable
n
1 package of paper
n
Disposable electrode starter set
n
Tab electrode adapters
n Page n
Writer 100 User’s Guide
Using the HP Page Writer 100 Cardiograph operator
training video

About This Manual

This guide contains concise operating instructions
for cardiograph users. This manual describes how to
perform the following tasks:
n
Recording an ECG
H Troubleshooting
n
Caring for and maintaining the cardiograph
n
Preparing your cardiograph for use
Getting Acquainted 1-7
Page 20

Patient and Operational Safety Notes

Your cardiograph isolates all connections to the patient from electrical ground and all other conductive circuits in the cardiograph. This reduces the possibility of hazardous currents passing from the cardiograph through the patient’s heart to ground. To ensure the patient’s safety and your own, observe the following reminders.
When operating your cardiograph from AC power, be sure it and all other electrical equipment connected to or near the patient are effectively grounded.
Use only grounded power cords (three-wire power cords with grounded plugs). Also make sure the outlet accepts the plug and is grounded. Never modify a grounded plug to fit an ungrounded outlet, i.e. removing the ground prong or ground clip to fit an ungrounded outlet. Should an ungrounded plug adapter be necessary, use a ground strap to connect the equipotential connector at the rear of the instrument to the power source ground.
The patient cable should be routed away from power cords and any other electrical equipment. Failure to do so can result in AC line frequency interference on the ECG trace.
Warning
1-8 Getting Acquainted
The HP patient cable supplied with this cardiograph, or an approved HP substitute patient cable, is an integral part of the cardiograph’s safety features. Using any other patient cable may compromise defibrillation protection as well as performance.
Only qualified personnel may service the cardiograph.
Page 21
. -. ‘, I. 1 L , :
,
I./,
,’
Warning
Caution
Do not use this cardiograph near flammable anesthetics. It is not intended for use in explosive environments.
Do not touch the patient, patient cable or cardiograph during defibrillation procedures. Death or injury may occur from the electrical shock delivered by the defibrillator.
Be sure that the electrodes or leadwire tips do not come in contact with any other conductive parts, including earth-grounded parts, especially when connecting or disconnecting electrodes to/from a patient.
The use of multiple instruments connected to the same
patient may pose a safety hazard due to the summation of leakage currents from each of the instruments. Any such combination should be evaluated by local safety personnel before being put into service.
Do not pull on the paper while a report is being printed.
This can cause distortion of the waveform and can lead to
potential misdiagnosis.
Do not block the ventilation slots located on both sides and to the rear of the cardiograph. Lack of ventilation may cause cardiograph to overheat and components to fail.
n
The Hewlett-Packard warranty is only assured if you use Hewlett-Packard approved accessories and replacement parts.
Getting Acquainted 1-9
Page 22

AC and DC (Battery) Operation

Your cardiograph requires the battery to be installed for proper operation-even if the cardiograph is plugged into AC power, it cannot print an ECG report without the
battery. For information about replacing or installing
the battery, refer to Appendix A, Setting up Your
Cardiograph for the First Time.
The following is a list of AC and battery operating instructions:
m A fully charged battery (without AC power) will print
approximately 40 Auto ECGs, or approximately 40 minutes of continuous Manual ECG information.
m The Low Battery light indicates the battery needs to
be charged. If the Low Battery light begins to flash, this indicates the cardiograph is about to shut down due to a very low battery. Plug the cardiograph into AC power.
n
From the time the Low Battery light first comes on to when the cardiograph automatically turns itself to
Standby (off), there is typically enough reserve battery capacity to print two Auto ECG reports or 2 minutes of Manual ECG data. A weak or faulty battery will reduce this time.
Note I
!I@
1-10 Getting Acquainted
H A discharged battery requires at least 5 minutes
charging time, with the cardiograph in Standby (off), to print an Auto ECG.
n
A discharged battery requires at least 10 minutes charging time, with the cardiograph in Standby (off), to print a l-minute Manual ECG.’
If the cardiograph is turned on while the battery is being charged, these charging times are doubled (10 minutes for an Auto ECG and 20 minutes for a l-minute Manual ECG).
Page 23
n
The PageWriter 100 cardiograph has a battery-saving feature: it will turn itself to Standby (off) after 30 minutes of instrument inactivity. This prevents the cardiograph from being accidentally left on for extended periods of time.
This feature is not active if all the limb electrodes are connected to a patient or if the cardiograph is plugged into AC power.
I A new. battery or a battery that has been stored for
an extended period of time requires charging (with the cardiograph in Standby (off)) for 16 hours.
n
The battery, if installed, is being charged any time the AC light is on.
n
A fully depleted battery will charge to 90% of full
capacity in 7 hours, and 100% capacity in 16 hours, as long as the cardiograph is in Standby (off) for the entire time.
m When the cardiograph is not in use, it should be
connected to AC power and left in Standby (off). This will maintain a full battery charge and prolong battery
life.
Note I
rg
The cardiograph’s battery charging circuit delivers less
power than the cardiograph uses while printing an ECG.
If a Manual ECG is being recorded, the battery charge level will continue to drop until the instrument shuts itself down or is turned to Standby (off) by the operator.
Getting Acquainted 1-1 1
Page 24
Recording an ECG
Page 25

Recording an ECG

This chapter describes how to prepare the patient for an ECG, record an ECG, understand the printed report, and change the ECG report format. Samples of the Auto and different Manual report formats are also shown.
2
Note
1 Q If the cardiograph has not been setup, refer to
Appendix A for instructions.
The basic steps and procedures for recording an ECG
are as follows:
1. If the cardiograph is not On, press (On/Standby).
2. Prepare the patient and apply the electrodes, as described in the “Preparing the Patient” section.
3. Press (Auto) or (jZZZ) to record the ECG.
4. Check the quality of the recorded ECG on the printed report.
The rest of thjs chapter discusses the details of setting up and recording ECGs and understanding the printed report.
Recording an ECG 2-1
Page 26

Preparing the Patient

Note
Note
1
Yb
1
Yb
For electrode placement information, refer to the diagram on the top of your cardiograph. For tips on proper ECG technique, see the videotape Using the HP
Page Writer 100 Cardiograph.
Proper patient preparation and electrode placement are the most important elements in producing a high quality ECG trace.
Prepare the patient by performing the following steps.
1.
Reassure and relax the patient. A calm and quiet patient produces the best ECGs.
2.
Make sure each electrode site is not covered by hair or clothing. Gently clean and abrade the surface of the skin with
3. dry gauze.
4.
Place electrodes on patient. See the notes below
regarding your type of electrodes.
The patient cable should be routed away from power cords and any other electrical equipment. Failure to do so can result in AC line frequency interference on the ECG trace.
2-2 Recording an ECG
Notes for Customers Using Reusable Electrodes Each electrode must be attached securely. Straps must
neither slide nor be so tight as to cause discomfort. The electrode paste, gel, or creme must cover an area the
size of the electrode, but must not extend beyond it, especially on the chest.
Notes for Customers Using Disposable Tab Electrodes Disposable electrodes have conductive material on one
side only, the adhesive side. The electrode tab must be
Page 27
placed between the jaws of the electrode adapter and remain flat. Do not attempt to place the jaws of.the electrode adapter so close to the circular part of the electrode that the tab of the electrode is bent or contact is made with the conductive gel. Gently tug on the electrode adapter to ensure that the electrode adapter is properly placed on the electrode.
Good and accurate placement on the first attempt should be your goal for each electrode. Each time an electrode is lifted off the skin and attached again, the conductive gel becomes weaker and less effective.
Understa
------ --
Jnding When a Signal is Acquired
Note I
VI
Your PageWriter 100 cardiograph attempts to acquire a.good signal for an Auto report before you press the
(Auto) key. Hewlett-Packard cab this Pre-acquisition.
Pre-acquisition is activated when the cardiograph is turned on and remains active until an Auto report begins to print. Pre-acquisition is deactivated by printing an Auto report to allow for copies of the Auto
report to be printed (see “Making Copies of Auto
ECGs”). Pre-acquisition is also deactivated whenever an electrode is disconnected.
Pre-acquisition is reactivated when a Manual report is
finished printing. When Pre-acquisition is active, it is important for the
patient to stay still and relaxed. This will help ensure a good signal is captured prior to printing an Auto report.
Pre-acquisition is not used for Manual ECG reports. Manual ECG reports display ECG data in real-time.
Recording an ECG 2-3
Page 28

Recording an Auto ECG

To record an Auto ECG, perform the following steps.
1. If the cardiograph is not On, press (WlL). The front panel lights are lit as the cardiograph performs a
short power-on sequence.
2. Prepare the patient and apply the electrodes.
3. Press (Auta) on the front panel. The light above the (Auto) key flashes while the
cardiograph is acquiring the ECG. (If the IAuto)
light flashes in an alternating pattern with the (Manual) lights a leads off condition is indicated. See
“Correcting a Leads Off Condition” below.) When the cardiograph begins printing the ECG, the (Auto)
light stays on. When the cardiograph has completed the report, the light turns off.
Correcting a Leads
Off Condition
A leads off condition is caused by the patient cable not being connected to the cardiograph, one or more leadwires not being connected to electrodes, or poor
contact between the patient and one or more electrodes. The cardiograph signals a leads off condition only during the acquisition of an Auto ECG. The operator can check for a leads off condition anytime an ECG report is printed by checking for traces consisting of dashed lines or by checking the upper left corner of the Auto ECG report for a leads off message. The leads off message lists the lead wires or electrodes that have a
leads off condition. Table
messages, and the lead wire and electrode to check for each one. To correct a leads off condition, perform the
following steps.
1. Check that the patient cable is properly connected.
2. Recheck the steps you performed when preparing
the patient. Pay particular attention to the proper application of the electrodes and the connections between leadwires and electrodes.
2-1
shows the possible leads off
2-4 Recording an ECG
Page 29
‘*
_>.,,_) -. ; I’
Table 2-1. Leads Off Labels
Designator
(AHA/IEC)
RL/N
RA/R LA/L LL/F Vl . . .
Cl . . .
V6/ One or more chest electrodes are not connected. For example, V2 C6 means the V2 electrode is not connected.

Auto Report Filters

Meaning
Right leg electrode not connected, or only right leg electrode is connected and all other limb electrodes are not connected.
Right arm electrode is not connected.
Left arm
Left leg elecrode is not connected.
electrode is not connectd.
The PageWriter 100 allows you to choose between two sets of filters for Auto ECG reports:
I The 0.15-40 Hz filters (0.15 Hz Baseline Wander and
40 Hz Noise) are enabled when the Filter light is on. It delivers the least amount of noise.
m The 0.15-150 Hz filters (0.15 Hz Baseline Wander and
150 Hz Noise) are enabled only when the Filter light is off. It delivers the highest fidelity signal, but unless ECG signal conditions are excellent, it can record noise.
Table 2-2 shows when each filter is enabled.
Table 2-2. Auto Filter Settings
Default (Filter light on)
0.15-40 Hz
1 Conforms to AAMI specifications
Filters Off (Filter light off)
0.15-150 Hz’
Recording an ECG 2-5
Page 30
Making Copies of
Auto ECGs
If you require additional copies of an Auto ECG, you may copy the last ECG that was recorded.
To copy your most recent Auto ECG, press the Icopy) key.
Note
3
w You must copy the ECG before the cardiograph has
been turned to Standby (off) and before another ECG
has been acquired.
n
You may change the speed (25 or 50 mm/set) prior to
printing a copy of an ECG.
w You can only print copies of Auto ECGs.
2-6 Recording an ECG
Page 31

Recording a Manual ECG

To record a Manual ECG, perform the following steps.
If the cardiograph is not On, press (On/Standby). The
1. front panel lights are lit as the cardiograph performs a
short power-on sequence. Prepare the patient and apply the electrodes.
2. Press (?J until the light above the desired lead group
3.
is lit.
4.
Choose the desired (Speed], (Sire), and (VLeads)
sensitivity.
Press (Manua) on the front panel.
5.
Inspect the ECG as it prints.
6.
If any of the traces consist of a dotted line, signifying leads off, press [Stop) and refer to “Correcting a Leads Off Condition”. If you would like to try to improve one or more
leads, press (Stopj and adjust the electrodes
accordingly. After making adjustments, press (Manual_) to restart the recording.
The cardiograph will print the ECG continuously until you press the IStop) key.
Note I
‘4$
If accurate ECG ST contours are required in Manual mode, turn off the (Filter key to activate the 0.15-150 Hz
filters. See “Manual Report Filters” for more details.
Recording an ECG 2-7
Page 32
Changing Manual
Report Settings
The following Manual report settings can be changed while a Manual report is printing:
n (Tiz-gi-Speed) n (ECG) n
(V) sensitivity
n IF;lter)
I Lead group selection ((7-J)
Each can be changed by pressing the appropriate key. There is no need to press (Stop) prior to changing settings.
Restoring the ECG
Trace After
Defibrillation or
Reconnecting Leads
After an application of a defibrillator pulse, reconnecting one or more leads, or any other time the ECG trace is off-center during a Manual report, the trace can be quickly restored by pressing the CManua/) key again.
2-8 Recording an ECG
Page 33

Manual Report Filters

The PageWriter 100 allows you to choose between two sets of filters for Manual ECG reports:
H The 0.05-40 Hz filters (0.05 Hz Baseline Wander and
40 Hz Noise) are enabled when the Filter light is on. It delivers the least amount of noise.
w The 0.05-150 Hz filters (0.05 Hz Baseline Wander and
150 Hz Noise) are enabled only when the Filter light
is off. It delivers the highest fidelity signal, but unless
ECG signal conditions are excellent, it can record
noise.
Table 2-3 shows when each filter is enabled.
Table 2-3. Manual Filter Settings
Default (Filter light on)
0.05-40 Hz
1 Conform to AAMI specifications
Filters Off (Filter light off)
0.05-150 Hz1
1
Recording an ECG 2-9
Page 34

Understanding the Printed Report

I
I I
.M
I”1 VI v*
I
I $---
“1
I
I
pJ-zcq+,+,
“6
A
Figure 2-1. The Auto ECG Report.
2-10 Recording an ECG
Page 35
Table 2-4. Auto ECG Report Annotations
1
Leads off indicators Calibration signal. See Table 2-5. Cardiograph settings for speed and for limb and
chest lead sensitivity. Filter settings: * AC filter ( w)
* Frequency range
i * Baseline Wander filter (W).
Sequence number, the total number of ECGs the cardiograph has recorded.
Description
Recording an ECG 2-11
Page 36
The following table shows how the height of the
calibration pulse indicates ECC: sensitivity.
Table 2-5. Calibration Signals ECG
Size (nun/n~V)
ECG Size [\/I Limb Leads V Leads Auto
nun/xnV Iudicator (Vl - VG)
i
off 5
5
nn r-l
Calibration Pulse
Manual
Limb Leads V Leads
i
on
5
10 ofi 10
10 on
10 5
!O off 20
!O on
20 10
2.5
10
VA--l r-t
nrl l-l
l-5l-l n
2o nn n
r-A-l
r-l
2-12
Recording an ECG
Page 37
,
Choosing Format
Auto
Report Format
Manual Report
a Report
Formats
An Auto report prints a one-page (at two-page (at 50 mm/set) summary of all 12 ECG leads. A Manual report presents a continuous printout of the selected lead group until the (Stop) key is pressed.
The Auto report format is a standard 12-lead ECG with
a lead II rhythm strip. Figure 2-2 is an example of the
auto report format.
Manual ECGs reflect the ECG waveform as it occurs and there is no significant delay in the recording.
Alternate lead groups can be selected while recording a 3- or 6-lead ECG by pressing @ until the light above the desired lead group is lit. The following table shows the available lead group choices.
Table 2-6. Manual Lead Groups
25
mm/set) or

Lead Groups

1 NE,of 1 (z%, /
Leftmost
I, II, III
2nd
3rd
Rightmost VI, V2, V3, V4, V5, V6
aVR, aVL, aVF
II,
aVF, V2
Recording an ECG 2-13
Page 38
An Auto Report
Example
The following figure is an example of the Auto ECG report format.
15 mm/r IO mm,mv c 0.15 HI - IJO “1 00101
Figure 2-2. An Auto 3x4 ECG with One Rhythm Strip(3x4, 1R).
2-14 Recording an ECG
Page 39
Manual Report
Examples
The following figures show examples of Manual ECG report formats.
Figure 2-3. A Manual 3-Lead ECG.
Recording an ECG 2-15
Page 40
n “’
15 !nm,I
IO m/m”
5 0.0s HZ - 10 H,
Figure 2-4. A Manual C-Lead ECG.
2-16 Recording an ECG
Page 41
Troubleshooting
Page 42

Troubleshooting

3
Your cardiograph is designed for reliable operation. If you have problems with an ECG, there are several things you can check before calling for service. This chapter tells how to solve basic ECG problems.

Checking ECG Technique

Many problems in taking an ECG may be related to electrode application. Review “Preparing the Patient” in Chapter 2, Recording an ECG to assure the patient leads are properly attached to the patient.
Troubleshooting 3-1
Page 43
Identifying ECG
Problems
The following table
shows
symptoms and solutions to
problems that can occur when recording an ECG.
Table 3-1. ECG Problems and Solutions
Problem
Power line AC
Interference
Wandering Baseline
Cause Possible Solutions
Poor
electrode contact.
electrodes.
Lead wires may be picking up Route electrode wires along the
Dry
or dirty Reapply electrodes. Abrade skin.
Use new electrodes.
interference. One possibility is a limbs and away from any other
poorly grounded instrument near electrical equipment. Fix other patient. equipment. Unplug cardiograph
from AC power and operate on battery only.
Patient cable is located too close to Move cardiograph further away frou the cardiograph’s power cord or patient. Unplug cardiograph from other power cords. AC power and operate on battery
only. Move other instruments further away from patient. Unplug electric bed.
Patient movement.
Electrode movement. Poor electrode Be sure that the lead wires are not contact & skin preparation. puliing on the electrodes. Reapply
Respiratory interference. Move lead wires away from areas
Reassure and relax the patient.
electrodes. Make sure the Filter light is on.
with greatest respiratory motion.
3-2 Troubleshooting
Page 44
Table 3-l. ECG Problems and Solutions (continued)
Problem
Tremor or Muscle Poor electrode placement. Poor Artifact
LLL
Intermittent or Jittery Waveform electrodes.
qj t-
t
Poor print quality (uneven contrast or blank streaks)
electrode contact. Patient is cold.
Tense, uncomfortable patient.
Tremors.
Poor electrode contact. Dry
Faulty lead wires.
Dirty printhead.
Cause
Possible
Clean the electrode site. Reapply electrodes. Be sure that the limb electrodes are placed on flat, non-muscular areas of the body. Warm the patient.
Reassure and relax the patient. Make sure the Filter light is on.
Attach electrodes high on the
extremities near the trunk. Make sure the Filter light is on.
Clean the site. Reapply electrodes.
Replace faulty patient cable.
Clean the printhead. Use HP recommended paper.
Solutions
If the Recording
Won’t Start
If you press (Auto) or (Manual) and the recording doesn’t start, investigate the
n
Is the cardiograph turned on?
At
least one light other than the AC light
be
on.
n
Is the AC power light on?
following possibilities:
should
If the cardiograph is plugged into AC power and the AC light is not on, check the two line fuses. See “Replacing the Fuses” in Chapter 4,
Maintaining
the
Cardiograph
for fuse information.
Troubleshooting 3-3
Page 45
n
Is the patient cable connected to the cardiograph?
Visually check the connection between the cardiograph and the patient cable.
n
Is the battery adequately charged?
The Low Battery light should be off.
n
Is the cardiograph out of paper? Is the paper jammed in the cardiograph?
The cardiograph will not record an ECG unless you have loaded paper or cleared the paper jam. See Appendix A!
the First Time
Setting up Your Cardiograph for
for details on loading the paper.
Reloading the paper will clear a paper jam.
w Is the paper sensor lens dirty?
Clean the paper sensor lens. See Chapter 4,
Maintaining the Cardiograph for more information.
I Is the paper door completely closed?
Open the paper door slightly and close it tightly. Listen for the door safety latch to lock.
3-4 Troubleshooting
If the cardiograph still won’t operate, perform the following steps.
1. Switch the cardiograph to
Standby
(off) with the
(0n/Standby) switch.
2. Wait 20 seconds and then switch the cardiograph back to On.
3. Press (Auto) or (Manual). If the cardiograph turns itself to
Standby
(off), the battery is not operating
properly.
If the cardiograph still won’t operate, call your local
Hewlett-Packard service representative.
Page 46
If the Cardiograph
Won’t Print a Manual
Report
n
Is the paper sensor lens dirty or obstructed?
Clean the paper sensor lens. See Chapter 4, Maintaining the Cardiograph and Figure 4-l for more information.
n
Is the cardiograph out of paper?
Load paper. See Loading the Paper in Chapter 4, Maintaining the Cardiograph.
Troubleshooting 3-5
Page 47
Maintaining the Cardiograph
Page 48

Maintaining the Cardiograph

Care and Cleaning

Cleaning the
Cardiograph
Caution
a
The outside surfaces of the cardiograph and its accessories (except the patient cable) are designed to be cleaned by mild soap and water or isopropyl alcohol. The patient cable can be cleaned only with mild disinfectant OT soap and water. The patient cable cannot
be cleaned with isopropyl alcohol.
1. Unplug the power cord and ensure that the cardiograph is in Standby mode (all lights are off).
2. Wipe the external surfaces of the cardiograph with a soft cloth dampened with mild soap and water or
isopropyl alcohol. Avoid applying cleaning fluids to
the cable connectors.
Do not use any strong solvents or abrasive cleaning materials. Do not spill any liquids on the surface of the cardiograph. Immediately have the cardiograph serviced if any liquids spill on the surface of the cardiograph. Do not use the following to clean the cardiograph:
n
Acetone H Iodine-based cleaners H Phenol-based cleaners
n
Ethylene Oxide Sterilization
n
Chlorine bleach H Ammonia-based cleaners
Maintaining the Cardiograph 4-1
Page 49
Cleaning the
Electrodes and
Cables
Clean the electrodes and patient cables with a soft cloth moistened with a recommended disinfectant or cleaning agent from the following list:
Cetylcide@ (may discolor cable)
Cidex@
Lysol@ Disinfectant Lysol@ Deoderizing Cleaner (may discolor cable) Dial@ Liquid Antibacterial Soap ammonia
409@ (may discolor cable)
10% solution of Clorox@ in water (may discolor cable) Murphy@ Household Cleaner, or
Ves-phene 118.
Wring any excess moisture from the cloth before cleaning.
Caution
$
I Do not clean the patient cable with alcohol. Alcohol
can cause the plastic to become brittle and may cause the cable to fail prematurely.
n
Do not autoclave the cable or use ultrasonic cleaners.
n
The patient cable is not immersible.
m Do not use abrasive materials to clean metal
surfaces-scratches on them can cause artifacts.
n
Avoid wetting the connectors.
4-2 Maintaining the Cardiograph
Page 50
Cleaning the Digital Array Printhead and Paper Sensor
Figure 4-l. Cleaning the Digital Array Printhead.
A. Printhead
B. Paper Sensor
If the print quality is uneven, it may be due to a dirty printhead. How frequently you must clean the printhead depends on how many ECGs you print and the quality and type of Paper you use.
If the Paper fails to stop at the end of a Page, the Paper Sensor lens may be dirty.
Maintaining the Cardiograph 4-3
Page 51
To clean the printhead:
Caution
a
Touch the equipotential connector on the back of the cardiograph to discharge any static electricity stored on your skin before touching the printhead. The printhead can be damaged by static electricity.
1. From the front of the cardiograph, unlatch and open the paper door. The printhead is to the right under the paper blade and behind a brush. See Figure 4-1.
2. Wipe the printhead with a foam swab dipped in 90% isopropyl alcohol. Scrub until all visible residues are
removed.
3. Dry the printhead with a clean lint-free tissue.
To clean the paper sensor lens:
1. From the front of the cardiograph, unlatch and open the paper door. The paper sensor lens is to the right
under the printhead. See Figure 4-l.
2. Lightly wipe the paper dust off of the paper sensor lens with a dry foam swab. Do not use alcohol.
4-4 Maintaining the Cardiograph
Page 52

Loading the Paper

Figure 4-2. Loading the Paper.
A. Paper Door Latch
B. Paper Sensing Hole
C. Paper
D. Cardboard Backing
E. (-Advance) Key
The cardiograph uses continuous-feed Z-fold paper. For best results use Hewlett-Packard paper. See “Supplies” for ordering information.
Maintaining the Cardiograph 4-5
Page 53
To load the paper:
From the front of the cardiograph, release the latch on
1. the left side and slide the paper door to your left.
Remove the outer packaging on the paper stack.
2.
Place the paper stack in the compartment so that the
3. top side of the paper feeds grid side up over the top panel. The paper sensing hole will be in the lower corner of the paper. Pull a sheet halfway out and over the paper door.
4. Make sure evenly within the channel of the paper door. See Figure 4-2. Slide the paper door back into place. Make sure that
5. the door is latched. Press (On/StandbyL) to turn On the cardiograph, if
6. necessary. Press (-Advance) to advance the paper to the beginning of the next page.
that
the paper lays on the black roller
Storing the Paper and
Patient ECG Records
4-6 Maintaining the Cardiograph
Recordings on standard chemical/thermal paper decompose naturally over time. With proper storage and handling, recordings on these papers have been shown to be readable for approximately 5 years. Proper storage and handling before and after recording includes:
1. Store the paper under cool and dry conditions.
relative humidity must be below 80%.
2. Store recorded ECGs in manila folders, or in
polypropylene, or acetate, in areas as described above. below.
Storing the paper as indicated above will minimize trace fading and background development (darkening).
However, storing the paper as indicated above does not protect against trace fading or background development from the sources below.
Temperatures must not exceed 40 ’ C (104’ F) and
sheet protectors made of polyester, polyimide, Do not store ECGs in vinyl sheet protectors, as noted
Page 54
To avoid trace fading or background development, the paper must not be exposed to or come in contact with the following, either before or after recording:
l
Solvent-based adhesives, as used in mounting forms, pressure-sensitive tapes, labels, and common mending tapes. Starch- or water-based adhesives may be used.
I Plastics containing plasticisers, such as vinyl chlorides
(PVC) typically found in vinyl sheet protectors, separators and plastic envelopes. Other plasticisers include polyethylene glycol, dioctyl phthalate, and dioctyl adipate.
n
Glossy (or non-glossy) papers containing tributyl phosphate, dibutyl phthalate, or other organic solvents, such as FAX and other non-chemical/thermal
recording paper, or product literature.
n
Liquid or vaporous solvents, such as alcohols, ketones,
esters, ethers, etc. Note that many of these solvents
are found in felt-tip and other marking pens.
w Petroleum-based solvents, such as toluene, benzene,
and gasoline.
w Bright light or UV sources such as sunlight, flourescent
and related light sources.
w Chemicals containing castor oil, ammonia, some
chemicals found in common hand and face creams, or citric acid (found in fruit juices).
n
Forms containing carbon or carbonless (NCR) copy sheets.
n
Pastes, creams or gels commonly used for ECG or ultrasound tests that contain any of the above or related chemicals.
If original recordings are stored, HP recommends
that records be checked annually to determine their
integrity. However, where long term storage is desired,
Maintaining the Cardiograph 4-7
Page 55
the user should consider photocopying or microfilming, or electronic or optical storage or a fade resistant paper.
HP’s PTPTM brand thermal papers offer improved archivability. See “Supplies” for ordering information.
Caring for Battery
Caution
Note
the
a
d
Your cardiograph requires the battery to be installed for proper operation-even if the cardiograph is plugged into AC power, it cannot print an ECG report without the battery. For information about replacing or installing the battery, refer to Appendix A, Setting up Your
Cardiograph for the First Time.
The sealed lead-acid battery used in the PageWriter 100 will provide optimum life when the unit is continuously connected to AC power and fully charged after each use. A depleted battery requires 16 hours of continuous charge time to fully charge. Because it is not always possible to allow a full charge cycle between uses, the PageWriter 100 was designed to charge a depleted battery to 90% of its capacity in approximately 7 hours.
Repeated undercharging of the battery will damage the battery and reduce battery life.
Hewlett-Packard recommends that the cardiograph be plugged into AC power whenever possible to maximize battery life.
Battery life varies by how the battery is maintained and how much it is used. For improved battery life, keep the instrument plugged in when not in use. If the battery has been fully charged and requires recharging after a
few ECGs, consider replacing it. Use only HP battery,
part number M2460A.
4-8 Maintaining the Cardiograph
Page 56
Note
J
Battery should be removed from unit and placed in storage if cardiograph will not be used for more than three months without AC power.

Storing the Battery

Replacing the Fuses

To prepare the battery for storage, charge it in the cardiograph for 16 hours. Then remove it from cardiograph and store it in a cool, dry location. Recharge a lead-acid battery in storage for at least 16
hours every six months. This ensures that the battery
does not completely discharge while in storage. The
battery’s shelf life is longer with cooler temperatures, but do not store below freezing level.
To replace the AC fuses:
1. Unplug the cardiograph from AC power.
2. Turn the cardiograph bottom-side up.
3. Locate the two AC fuse holders on the bottom of the cardiograph, as shown in Figure 4-3.
4. Using a screwdriver, turn the fuse cap l/2-turn counter-clockwise. As the fuse cap is untwisted, it
extends above the surface of the cardiograph case.
5. Pull the fuse cap straight up approximately 2-l/2 cm
(l-inch), until it stops.
6. Remove the fuse. You may need to tap the fuse
holder to shake the fuse out.
7. Insert a new fuse in the holder, slide the fuse cap back
into the case. Fuse must be of the same type and rating as described on the label located next to the fuse holders.
8. Tighten the fuse cap l/2-turn clockwise.
Maintaining the Cardiograph 4-9
Page 57
9. Repeat the operation for the other AC fuse.
+/y
-'......., <,, -... . . . . . . . ~
0
0
&.O .$
A. AC Fuse holders (2)
4-10 Maintaining the Cardiograph
Figure
4-3.
The AC Fuse Holders.
Page 58

Supplies

Hewlett-Packard offers a full range of supplies for cardiographs. The following list is a collection of the most frequently ordered items. Pricing and availability of these and other supplies are available from Hewlett-Packard’s Medical Supplies Centers.
n
USA: Call l-800-225-0230
n
Outside USA: Please contact your local Hewlett-Packard Sales Office or your authorized Hewlett-Packard Dealer or Distributor.
Paper M2481A
Paper, 8.5” x 11”) 1600 sheets, with
header M2482A M2483A
Paper, 8.5” x 11”) 1600 sheets, no header
Paper, 210 x 300 mm, 1600 sheets, with
header M2484A
Paper, 210 x 300 mm, 1600 sheets, no
header M2485A
Paper, PTP brand Anti-fade, 8.5” x 11”)
1600 sheets, with header M2486A
Paper, PTP brand Anti-fade, 210 x 300
mm, 1600 sheets, with header
Maintaining the Cardiograph 4-11
Page 59
Battery M2460A
Patient Cable M2461A M2462A
Carrying Case M2463A
Redux 651-1021-010 651-1008-010 651-1024-010 40481B
Battery assembly
AHA Patient Cable with leads IEC Patient Cable with leads
Soft Carrying Case
Redux Redux Redux
Creme; 4 oz tube; 10 tubes per box Paste; 5 oz tube; 10 tubes per box Gel; 4 oz tube; 10 tubes per box
Redux Infant Creme; 5 oz bottle; 12
bottles per box
4-12 Maintaining the Cardiograph
Page 60
Electrodes 404903
404913 40421A
40424A
14030A
40420A 13943B
13943D
13944B
Lead Adapters
13946B 14282A
404983
Welsh electrode; 15mm base 5cc bulb; screw connection(IEC) Limb Plate Electrode(IEC) (4 per pack) Welsh electrodes; 15mm base 5cc bulb; push-in connection(AHA) (6 per box) Limb Plate Electrode(AHA) (4 per pack) 15” Rubber strap for limb plate electrode Disposable diagnostic pre-gelled electrode (1,000 pieces) Disposable diagnostic solid gel electrode (1,000 pieces) Disposable diagnostic solid gel electrode (1,000 pieces) (United States and Canada
only) Disposable diagnostic wet gel electrode
(300 pieces)
Universal ECG Adapter (10)
Snap Electrode Adapter for &I1 post leads, spring clip (5 per box) Snap Electrode Adapter for 4 mm banana leads, grabber (10 per box)
Cart M1705B
Fuses 21 lo-0620
2110-0930
Cart
AC fuse; 0.4 Amp 250 V Battery fuse
,
Maintaining the Cardiograph 4-13
Page 61

Calling for Service

United States of
America
HP Medical Response Center
Tel: I-800-548-8833

Canada

Other International
Areas
European Headquarters
Tel: 22 838-111
Austria, Eastern Europe
Tel: 0222125 00-333
Belgium
Tel: 02/778 35 31
Netherlands
Tel: 20/547 69 11
Norway
Tel: 2/24 60 90
Eastern Region
Tel: l-800-361-9790 Tel: l-800-387-3900
Western Region Tel: l-800-661-5626
Ceritral Region
France
Tel: (1)60 77 80 00
Gerxnany
Tel:
02102-9060
Mediterranean and Middle East
Tel: (41) 22 780 4111
Spain
Tel: l/637 00 11
Sweden
Tel: S/444 22 30
Denmark
Tel: 2/81 66 40
Finland
Tel: 90/887 21
4-14 Maintaining the Cardiograph
United Kingdom
Tel: 44 344 366 333
Italy
Tel: 2/921 22 833
Page 62
Setting up Your Cardiograph for the First Time
Page 63
Setting up Your Cardiograph for the First Time
Before using your cardiograph for the first time you must prepare it by performing the following tasks:
n
Check the voltage setting
H Install the battery
n
Connect the cables
n
Load the paper
A
Setting up Your Cardiograph for the First Time
A-1
Page 64

Checking the Voltage Setting

Your cardiograph can be set to operate at nominal line voltages of 115 or 230 Volts (see the following Note). The line voltage was set at the factory to the setting for your area. However, it is a good idea to check this setting. Refer to Figure A-l for the physical location of the voltage select switch. To check the voltage setting, perform the following steps.
1.
Locate the voltage select switch on the back of the cardiograph. See Figure A-l.
2.
Verify that the correct voltage is visible on the voltage select switch. If the voltage setting is incorrect, slide the voltage switch so the correct voltage is visible. The cardiograph operates with any line frequency from 50 to 60 Hz.
Remove and discard the label that covers the AC
3. power receptable. See Figure A-l for the location of the AC power receptable. The purpose of the label
is to remind you to check the setting of the voltage
select switch.
Caution
A-2 Setting
The cardiograph can be damaged if plugged into the incorrect voltage.
up
Your Cardiograph for the First Time
Page 65
Note
!b
1
The nominal 115 VAC voltage setting works equally well
for any voltage between loo-120 VAC. The nominal
230 VAC voltage setting works equally well for any
voltage between 220-240 VAC.
[. C-j@)
a
Figure A-l. Rear View of Cardiograph
Note I
w
Qj-JLW j
,:.:
;
I4 Ei c
A. Equipotential Connector
B. AC Power Receptacle
C. Voltage Select Switch The equipotential connector is only used when the
cardiograph must be plugged into an ungrounded outlet. See “Patient and Operational Safety Notes” in Chapter 1, Getting Acquainted for more information about using the equipotential connector.
Setting up Your Cardiograph for the First Time A-3
Page 66

The Battery

Your cardiograph requires the battery to be installed for proper operation-even if the cardiograph is plugged into AC power, it cannot operate without the battery.
Use only HP batteries, part number M2460A, in the cardiograph.

Installing the Battery

Note
Yii
Note
3
To install the battery:
Do not remove the shrinkwrap surrounding the battery.
When you install the battery you may notice a screw hole located near the battery compartment. This screw hole is the mounting point for the optional The mounting thumbscrew is supplied with the cart.
1. Make sure the cardiograph is unplugged from AC
power.
2. Turn the cardiograph bottom-side up.
3. Slide the battery door in the direction of the arrow until it unlatches (approximately l/2 inch), as shown
in Figure A-2. Lift off the door.
4. Install the new battery in the battery compartment as shown in Figure A-3 and plug the battery connector into the cardiograph.
5. Place the battery door into its slots and slide the door in the opposite direction of the arrow until it latches, as shown in Figure A-2.
6. Turn the cardiograph top-side up.
7. Plug the cardiograph into -4C power.
8. Check that the AC indicator light is on. The unit is now in Standby (off) mode with the battery charging.
M1705B
cart.
After you finish setting up the cardiograph, it may be used on a limited basis until the battery has been fully charged. Hewlett-Packard recommends charging the
A-4 Setting up Your Cardiograph for the First Time
Page 67
r
battery as soon as possible for at least 16 hours. To charge the battery, plug the cardiograph into the wall outlet with the k-1 switch set to Standby (off).

Removing the Battery

I
MlTIo.02
Figure A-2. Removing the Battery Door.
A. Battery Door
To Remove the Battery:
1. Unplug the cardiograph from AC power.
2. Turn the cardiograph bottom-side up.
3. Slide the battery door in the direction of the arrow until it unlatches (approximately l/2 inch), as shown in Figure A-2. Lift off the door.
4. Unplug the battery connector from the cardiograph by squeezing the edges of the connector and pulling it straight out from the cardiograph.
5. Remove the battery and cable.
6. If the battery has been removed for storage, replace the battery cover by placing the battery door into its
Setting up Your Cardiograph for the First Time A-5
Page 68
slots and sliding the door in the opposite direction of the arrow until it latches, as shown in Figure A-2.
Warning
Properly dispose of or recycle depleted batteries according to local regulations. Do not disassemble, puncture or incinerate the disposed batteries.
Figure A-3. The Battery Compartment.
A. Battery
B. Battery Cable C. Battery Connector
A-6 Setting up Your Cardiograph for the First Time
Page 69

Connecting the Cables

Figure A-4. Connecting the Power Cord.
A. Power Cord
B. Voltage Select Switch
Warning
0
1. Connect the power cord to the cardiograph as shown in Figure A-4.
2. Plug the power cord into the wall outlet. If you must use an ungrounded plug adapter to plug
the power cord into the wall outlet, you must also use a ground strap to connect the equipotential connector at the rear of the cardiograph to the power source ground. Figure A-l shows the location of the equipotential connector. Refer to the “Patient and Operational Safety Notes” in Chapter 1, Getting Acquainted for more information about using the equipotential connector.
Setting up Your Cardiograph for the First Time
A-7
Page 70
Figure A-5. Connecting the Patient Cable.
A. Patient Cable
Connect the Patient cable to the front of the cardiograph
as shown in Figure A-5 and screw in both thumb-screws.
A-8 Setting up Your Cardiograph for the First Time
Page 71

Loading Paper

Figure A-6. Loading the Paper.
A. Paper Door Latch
B. Paper Sensing Hole C. Paper
D. Cardboard Backing
E. [Advance) Key
The cardiograph uses continuous-feed Z-fold paper. For best results use Hewlett-Packard paper. See “Supplies”
in Chapter 4 for ordering information.
Setting up Your Cardiograph for the First Time
A-9
Page 72
To load the paper:
From the front of the cardiograph, release the latch on
1. the left side and slide the paper door to your left. Remove the outer packaging on the paper stack.
2. Place the paper stack in the compartment so that the
3. top side of the paper feeds grid side up over the top panel. The paper sensing hole will be in the lower corner of the paper.
4.
Pull a sheet halfway out and over the paper door. Make sure that the paper lays on the black roller evenly within the channel of the paper door. See Figure A-6. Slide the paper door back into place. Make sure that
5. the door is latched. Press [?C/GG$ to turn On the cardiograph, if
6. necessary. Press (-Advance) to advance the paper to the beginning of the next page.
A-10 Setting up Your Cardiograph for the First Time
Page 73
Specifications
Page 74

Specifications

Conforms to applicable IEC, UL, AAMI, CSA
specifications.

Basic Controls

Frequency and
ECG Controls: On/Standby, Auto, Manual, Copy, Filter, Page Advance (paper feed), Stop.
ECG Format Controls: Chart Speed, ECG size, V Leads sensitivity. ECG Format Selections: Auto (3x4 with 1 rhythm lead); Manual (with 3 or 6 leads).
Sinusoidal: 0.67 to 40 Hz &lo%
Impulse Response ~~I$$~~~;;f~ecj
Displacement: Less than 0.10 mV Slope: Less than 0.30 mV/sec
Each meets or exceeds AAMI ECll-1991 standard for Diagnostic Electrocardiographic Devices.
Specifications B-l
Page 75

Instrument Test

A self-test may be started by pressing and holding both the LAuto) and @G&i) keys while turning on the cardiograph with the (On/Standby) key. The instrument performs an internal self-test. The test results are printed on the report for use by service personnel. This self-test runs continuously until the cardiograph is turned to Standby (off).

Patient Safety

Power and Environment

Patient Isolation: Less than 20 PA leakage with 120 VAC, 60 Hz or less than 50 PA leakage with 240 VAC, 50 Hz with patient cable.
Defibrillation Protection: Protected against damage from 400 joule defibrillator discharges.
Line Power: 100 to 120 VAC, 50 to 60 Hz (at 115 VAC power-switch setting); 220 to 240 VAC, 50 to 60 Hz (at 230 VAC power-switch setting); 50 VA Maximum.
Fuse: 0.4 A for both the 115 and 230 VAC switch settings.
Environmental Operating Conditions: 10 to 40” C (50 to
104” F), 15 to 80% relative humidity, non-condensing, up
to 4,550 m (15,000 ft.) altitude. Environmental Storage Conditions: 0 to 50” C (32 to
122” F), 15 to 90% relative humidity, non-condensing, up
to 4,550 m (15,000 ft.) altitude.
B-2 Specifications
Cardiograph Dimensions: 43 by 39 by 10 cm (17.0 by
15.3 by 3.9 in.).
Cardiograph Weight: 8.5 kg (19.8 lbs.).
Page 76
Glossary
Page 77

Glossary

AC filter
A filter that screens out ECG artifact caused by power line interference. This filter is built into the cardiograph and cannot be disabled.
AHA leads
ECG lead names and identifying colors recommended by the American Heart Association. Limb leads are labelled RA, LA, LL, RL. Chest leads are labelled Vl-V6. (See IEC leads)
alternating current (AC)
Electrical current provided by wall outlets. AC may be either 60 or 50 Hz depending on country.
artifact
ECG waveform distortion that may diminish ECG quality. ECG artifact (or noise) may be caused by electrical interference, poor electrode connections, or patient movement.
Auto ECG
Twelve-lead ECG recorded and analyzed over a ten second period and printed in a predetermined format. The PageWriter 100 prints Auto reports in a 3x4 format with one rhythm strip.
baseline wander
A slow upward or downward motion on the baseline of any ECG waveform.
Glossary-1
Page 78
baseline wander filter
Hewlett-Packard term for filter which reduces baseline wander.
battery saver
Hewlett-Packard term for the cardiograph turning to Standby automatically after a preset time period to conserve power. The battery saver is factory set for 30 minutes of cardiograph inactivity.
calibration pulse
A 200 ms, 1 mV square or stepped wave pulse which appears on the printed record. The calibration pulse shows the sensitivity at which the ECG was recorded.
cycle power
To press the (On-Standby) button to Standby (off) and then back to On.
ECG report
Paper copy produced by the cardiograph when the
operator presses the (Auto) or (‘d start key. This report includes a graphic representation of the heart’s electrical activity (ECG waveforms) and identifying information.
Glossary-2
front panel
Cardiograph area that includes the control keys.
Hertz
IEC leads
(Hz)
A unit of frequency equal to one cycle per second.
Lead names and identifying colors recommended
by the International Electrotechnical Commission
standard. IEC limb electrodes are labelled R, L, F, and N. Chest electrodes are labelled Cl-C6.
Page 79
leads off
A trace consisting of a dotted line.
Manual ECG
ECG report format which runs continuously until the operator stops the recording. The ECG may show three or six lead waveforms. Some institutions and physicians may identify this format as a rhythm strip.
operator
The person who records the ECG.
patient cable
Hewlett-Packard term for the one-piece patient-lead set and instrument cable. The patient cable connects the cardiograph to the electrodes attached to the patient.
pre-acquisition
Hewlett-Packard term for acquiring 10 seconds of ECG prior to the operator pressing (Auto).
rhythm strip
Hewlett-Packard term for ten-second recording of a
particular lead that is printed at the bottom of an Auto ECG report. (See Manzlal and A&o ECG)
Standby mode
The cardiograph is off but the battery is kept charged while the unit is plugged into AC power.
standard leads
The conventional twelve lead order is I, II, III, aVR, aVL, aVF, Vl - V6.
Welsh cups
Reusable electrodes held in place with suction cups.
Glossary-3
Page 80
Index
Page 81

Index

3
3-lead Manual report, 3x4, 1R Auto report,
6
6-lead Manual report, 2-16
7
0, 2-7
A AAMI, B-1
AC
filter, Glossary-1 fuse, 4-13 fuse holders, 1-3, 4-9, 4-10 operation, 1-9 power, 1-9
power light, 1-5, 3-3 accessories, 1-6 AC operating frequency, A-2 AC voltage
operating, A-3
setting, A-2 AHA leads, Glossary-1 AHA patient cable, 4-12 alphanumeric, Glossary-1 alternating current, Glossary-1 artifact, Glossary-1
1Auto), 1-5, 2-4
Auto ECG
and ccopy_), 2-6 copying, 2-6 recording, 2-4
2-15
2-13, 2-14
index-1
Page 82
Auto filters, 2-5 automatic shut-off, 1-10 Auto report
3x4, lR, 2-13, 2-14
rhythm lead, 2-13
baseline wander, Glossary-1
B
Baseline Wander filter, 2-5, 2-9, Glossary-1 battery, 4-8, 4-12, A-4
capacity, 1-11 charging, 1-11
disposal, A-6 door, 1-3 fuse, 4-13 installing, A-4 operation, 1-9 removing, A-5 saving, 1-10 storing, 4-9
battery saver, 1-10, Glossary-2
calibration pulse, Glossary-2
C
calibration signals, 2-12 carrying case, 4-12 cart, M1705B, 4-13 Cart, M1705B
mounting point, 1-3
case, carrying, 4-12
@iGqxj, i-5
checking ECG technique, 3-1
choosing a report format, 2-13
cleaners, approved, 4-1
cleaning, 4-1
electrodes, 4-2 paper sensor, 4-4 patient cable, 4-2 printhead, 4-4
connecting patient cable, A-8 &), 1-5, 2-6 copying Auto ECGs, 2-6
Index-2
Page 83
CSA, B-1 cycle power, Glossary-2
defibrillation
D
restoring the ECG trace, 2-8
ECG
E
identifying problems, 3-2 Manual, Glossary-3 quality, 2-2 recording, 2-1 report, Glossary-2
ECG problems
AC interference, 3-2, 3-3 identifying, 3-2 intermittent waveform, 3-3
jittery waveform, 3-3
muscle artifact, 3-3 print quality, 3-3 respiratory interference, 3-2 tremor, 3-3 wandering baseline, 3-2
%!%%i$e, checking, 3-1 electrodes, 4-13
cleaning, 4-2 Welsh cups, Glossary-3
filter
F
Baseline Wander, 2-5, 2-9
Noise, 2-5, 2-9 1F;lterl, 1-5, 2-5, 2-7, 2-9 filters
Auto, 2-5
Manual, 2-9 front panel, 1-4, Glossary-2 fuse, 4-9
AC, 4-13
battery, 4-13 fuse holders
Index-3
Page 84
AC, 4-9, 4-10
fuses
replacing, 4-9
H Hertz, Glossary-2
identifying ECG problems, 3-2
I
IEC, B-1 IEC leads, Glossary-2 IEC patient cable, 4-12 installing battery, A-4
leads
L
AHA, Glossary-1 IEC, Glossary-2 standard, Glossary-3
leads off, 2-4, Glossary-2
loading paper, 45-6, A-9, A-10
Low Battery light, 1-5, 1-10
M
maintenance, 4-1 Manual
ECG, 2-7
filters, 2-9
m), 1-5, 2-7, 2-13
Manual ECG, Glossary-3 Manual Lead Groups, 2-13 Manual report
3-lead, 2-13, 2-15 6-lead, 2-13, 2-16 and defibrillation, 2-8 and m, 2-7 and &J 2-7, 2-13 changing settings, 2-8
Index-4
Page 85
Noise filter, 2-5, 2-9
N 0
$%!F%oIsZry-3
(-Advance], 1-5, 4-5, A-9
P
paper, 4-11
paper problems, 3-5
paper sensor
patient cable, 1-2, 4-12, Glossary-3
patient ECG records patient preparation, 2-2
patient safety, B-2 power cord, 1-8, A-7 pre-acquisition, Glossary-3 preparing the patient, 2-2
Printed Report printhead print problems, 3-3
,
Anti-fade, 4-11 loading, 4-5-6, A-9, A-10 storing, 4-6
cleaning, 4-4
AHA, 4-12 and abrasives, 4-2 and AC interference, 1-8 and autoclaving, 4-2 and isopropyl alcohol, 4-2 and routing, 1-8 and ultrasonic cleaners, 4-2 cleaning, connecting, A-8 IEC, 4-12 immersing, 4-2
storing, 4-6
understanding, 2-10 cleaning, 4-4
4-2
Index-5
Page 86
recording
R
Auto ECG, 2-4
ECG, 2-1
Manual ECG, 2-7 removing battery, A-5 replacing fuses, 4-9 report
ECG, Glossary-2 report fields, 2-10 report format
choosing, 2-13
rhythm strip, Glossary-3
safety, 1-8-9, B-2
S
self-test, B-2 service, 4-14 setting AC voltage, A-2 setting up, A-1 shut-off, automatic, 1-10 specifications, B-1 standard leads, Glossary-3 Standby mode, Glossary-3 Istopl, 1-5, 2-7, 2-13 storing
ECG paper, 4-6
patient ECG records 4-6
the battery, 4-9 supplies, 4-11 switch, voltage select, A-7
index-6
T troubleshooting, 3-1
U UL, B-1
understanding reports, 2-10
v
(\/Leads_], 1-5
voltage select switch, A-7
W
Welsh cups, Glossary-3
Page 87
HEWLETT PACKARD
M1772-91900
Copyright 0 1994, 1996 Hewlett-Packard Company
’ Printed in USA February 1996
Loading...