Scope of supply ...................................................................................................................................... 5
Function .................................................................................................................................................5
Measuring transducer ORT 7...................................................................................................................7
Name plate / marking ............................................................................................................................. 8
Dimensions OR 52-7 ...............................................................................................................................9
Installation
OR 52-7 ................................................................................................................................................10
Example of installation .......................................................................................................................... 11
Electrical Connection
Measuring transducer ORT 7 ................................................................................................................12
Other parameter settings ...................................................................................................................... 20
6
2
Contents – continued –
Page
Start, Operation, Alarm and Function Test
Start and operation ............................................................................................................................... 21
Alarm 1 and 2 ...................................................................................................................................... 22
Functional test .....................................................................................................................................22
Data indication .....................................................................................................................................23
Operating steps – see Fig. 8 ................................................................................................................. 23
OR 52-7 ................................................................................................................................................25
OR 52-7 ................................................................................................................................................26
Maintenance
Cleaning the glass cylinder .................................................................................................................. 27
Replacing the drying cartridge ..............................................................................................................27
Replacing the light emitter/receiver ......................................................................................................27
Spare parts ........................................................................................................................................... 28
Type approval certificate ...............................................................................................................30 – 33
EG-Baumusterprüfbescheinigung (Modul B) .................................................................................... 34, 35
EC Type Examination Certificate (Module B) ....................................................................................36, 37
Declaration of Conformity .....................................................................................................................38
3
Important Notes
Usage for the intended purpose
The GESTRA Turbiscope OR 52-7 is designed for the continuous monitoring and detection of oil
contamination of water and used in conjunction with oily water separators as 15 ppm bilge alarm.
The equipment meets the requirements of the International Maritime Organization (IMO) for bilge alarm
monitors, MEPC. 107 (49).
Safety note
The equipment must only be installed and commissioned by qualified and competent staff.
Retrofitting and maintenance work must only be performed by qualified staff who – through adequate
training – have achieved a recognised level of competence.
Danger
The solenoid of the solenoid valve Clw can reach up to 80 °C during operation.
The terminal strips of the measuring transducer ORT 7 can be live during operation.
This presents the risk of electric shock.
Cut off power supply before mounting, removing or connecting the terminal strips.
4
Explanatory Notes
Scope of supply
Measuring transducer ORT 7
Measuring sensor ORG 12 with
1 Solenoid valve
1 Drying cartridge (fitted)
1 Light emitter (supplied but not fitted)
1 Spare gasket kit
Description
The oil & turbidity detector OR 52-7 can be used:
■ On board of marine vessels for monitoring the oily water separating equipment for bilge water
according to IMO Resolution MEPC. 107 (49) as 15 ppm oil content alarm.
Function
The measuring sensor is a photometric measuring device consisting of a light source (light emitter)
and a light receiver fitted with two photosensitive elements.
The constant light beam generated in the measuring sensor passes through the transparent liquid. Any
foreign matter which is not dissolved scatters the beam, with emulsified oils scattering the light mainly
in the forward direction. The intensity of the scattered light is converted into an electric current and then
used as measure of the degree of contamination.
The measuring transducer has the following functions:
■ Control of the measuring sensor and signal evaluation
■ Control of the purging valve
■ Limit ALARM 1 and ALARM 2
■ Analog actual value output 0/4 – 20 mA
■ ALARM 2 indicates a malfunction in the turbidity sensor
■ Indication of actual value / limits
■ Storage and indication of alarm, malfunction and status messages
■ Control of the bilge bypass via output ALARM 2
5
Technical Data
Measuring transducer ORG 12
Nominal size
DN 10, connection
3
/8" to DIN ISO 228
Nominal pressure
PN 10
Fluid
Water
pH value
Up to 10.5 (a pH value of 11 and above will lead to wear of the glass, depending on the temperature!)
Glow lamp 12 V / 10 W BA 15s
Electric connection via 4 m connecting cable fitted with four pole connector,
protection IP 65
Light receiver
Two silicon photoelectric cells
Electric connection via 4 m connecting cable fitted with four pole connector,
protection IP 65
Purging valve
2/2 way solenoid valve
Directly controlled, normally closed
Connection:
Valve body: Brass
Internals: Ms 2.0401 / S. S. 1.4104
Internal seal: FKM
Solenoid: Voltage 24 V / 50 Hz, 43 / 24 VA
3
/8" ISO 228
Electric connection via 4 m connecting cable fitted with four pole connector,
protection IP 65
Weight
ORG 12: approx. 6.8 kg
6
Technical Data – continued –
Measuring transducer ORT 7
Inputs / outputs
2 Photocells for inputs: directly transmitted light / scattered light,
1 Voltage output for light emitter, 1 – 12 V, pulse-amplitude modulated,
1 Signal input for separator ON/OFF,
1 GPS input RS 485, 4800 Baud, NMEA 183 protocol, data record GPRMC,
3 Volt-free change-over contacts for alarm 1, 2 and purging valve (Clw),
contact material AgNi 0.15,
max. contact rating for switching voltages 24 V AC/DC, 115 V AC and 230 V AC: resistive / inductive 4 A,
1 Current output 0/4 – 20 mA as actual value output, max. load 500 ohm.
Indicators and adjustors
1 LCD display for indicating the actual value, limits, status and malfunctions,
8 Membrane keys,
4 LEDs for indicating alarm, status and malfunction
3 Green LEDs on basic circuit board for monitoring the operating voltages
Data storage
DataFlashcard 8 MB
Measuring range
0 to 25 ppm (ppm = parts per million)
Actual value output
0 or 4 mA = 0 ppm, 20 mA ≥ 25 ppm
Limits for alarm 1 and 2, adjustment range
Adjustable between 0 and 15 ppm
Delay of response for alarm 1 and 2, adjustment range
Adjustable between 0 and 15 sec.
Mains voltage
24 V +10 / –15 %, 50/60 Hz
230 V +10 / –15 %, 50/60 Hz (optional)
115 V +10 / –15 %, 50 / 60 Hz (optional)
Power consumption
50 VA
Fuse
24 V: Thermal fuse M 1.6 A 5 x 20,
230 V: Thermal fuse M 0.2 A 5 x 20
115 V: Thermal fuse M 0.4 A 5 x 20
Housing
Field case for wall mounting,
Case material: die-cast aluminium
Cable gland / Electrical connection
8 cable glands with integrated cable clamps, M 16 x 1.5
3 Three-pole screw-type terminal strips, conductor size 0.75 mm2
8 Multi-pole screw-type terminal strips, detachable, conductor size 1.5 mm
Protection
IP 65 to EN 60529
Admissible ambient temperature
0 – 55 °C
Weight
3.6 kg
2
7
Technical Data – continued –
Ser.Nr.: 3923581xxxxx
ORG12
OR 52-7
ORT7
Öl- / Trübungsmelder
Oil- / Turbitity Detector
Détecteur d’huile et de turbidité
Oil-content meter OR 52-7 (15ppm Alarm)
approved by SEE-BG
according IMO-Resolution MEPC.107(49)
Approval No. IMO/EC 320030
24 V 50/60 Hz 50VA IP 65
Before removing cover isolate from
power supplies
See installation instructions
Avant d'ouvrir le couvercle dèconnecter
complètement l'appareil
Voir instructions de montage
Vor Öffnen des Deckels Gerät
freischalten!
Betriebsanleitung beachten
Name plate / marking
Year of manufacture
Fig. 1
8
Technical Data – continued –
ON ALARM 1 ALARM2 SYSTEM
GESTRA
ORT 7
OKP
Dimensions OR 52-7
Measuring sensor ORG 12
approx. 225
approx. 200approx.170
Measuring transducer ORT 7
Fig. 2
240
260
Depth 91 mm
110
160
9
Installation
OR 52-7
Use the support and install the measuring sensor ORG 12 as close as possible to the separator in
order to keep the supply delay to a minimum. According to MEPC. 107 (49) in the event of an alarm
(oil content =/> 15 ppm) the contaminated water must be diverted from the separator to the bilge as
quickly as possible, i. e. within no more than 20 seconds.
Use a 10x1 mm copper or stainless steel pipe to connect the measuring sensor ORG 12/22 with the
sampling point of the separator and the clear water supply. The sampling point should be provided in a
vertical section of the separator outlet pipe so as to avoid any impairment caused by air bubbles.
In order to ensure that the arrangement for the extraction of samples gives a truly representative
sample make sure that the fluid flow always passes through the sensor under all operating conditions
and that any by-passing of the oil separator is prevented.
If it is not possible to provide the sampling point in a vertical pipe, a lateral arrangement in a horizontal
pipe is also possible.
For overboard pipes having an outside diameter of up to 75 mm a tee piece for the extraction of
samples can be used. If the size of the pipe exceeds 80 mm, make sure that the sampling arrangement extends approx. 25 % into the overboard line.
Install the vacuum breaker above the highest point of the oil separator.
Attach the measuring transducer ORT 7 to a stable vertical surface. If possible mount the indicator at
eye level for easy viewing and provide sufficient space for service and maintenance work.
The measuring transducer is designed for wall mounting. The distances between the fixing holes are
indicated on the back of the equipment.
To access the fixing holes undo the cover screws and open the cover.
10
Installation – continued –
ON ALARM1 ALARM 2 SYSTEM
GESTRA
ORT 7
OKP
Example of installation
Three-way control valveOverboard line
Separator
Pressure-
maintaining
valve
(if required)
Separator
outlet
To bilgeTo bilge
10 x 1 mm
copper or stainless steel pipe
InletOutlet
Clear water
inlet (water
pressure
≥ 2.5 bar)
Mount
vacuum
breaker
above
the oil
separator
To
bilge
Fig. 3
11
Electrical Connection
Measuring transducer ORT 7
1. Undo cover screws and open cover.
2. Detach all three-pole terminal strips from the base board and pull connecting lines through cable
glands.
3. Mark connecting lines: “Mains”, “Alarm 1”, “Alarm 2”, “Light emitter” and “Light receiver”.
4. Connect all terminal strips in accordance with the wiring diagram. Connect screens only to ORT 7.
5. Connect PE with the earthing screw in the housing.
6. Re-attach terminal strips.
7. Tighten the cable glands in order to seal the cable entry. Use the supplied sealings plugs to seal off
any unused cable glands.
8. Close housing lid and tighten cover screws.
Note
■ The light emitter/receiver and the purging valve are fitted with a 4 m long connecting
cable and only have to be connected to the measuring transducer.
■ The terminal strips on the CPU board of the ORT 7 cannot be detached.
■ Strip off approx. 25 mm of the cable insulation coating and screen of the connecting
cables for the signal input of the separator ON/OFF and for the GPS input.
Attention
■ The oil & turbidity detector has a power consumption of max. 50 VA. Please make sure
that the power supply is rated for 100 VA.
■ To protect the output relay contacts fuse electric circuits with 2.5 A (slow-blow).
■ Provide connected contactors and actuators with RC combinations (in accordance with
manufacturers’ specifications) in order to suppress interferences.
■ Make sure that the disconnecting switch (Fig. 4) is easily accessible and in the close
proximity of the equipment (EN 61010-1).
■ Mark respective switch as disconnecting device for the measuring transducer.
This switch can be dispensed with if the control cabinets are upstream of the equipment
and fitted with a disconnector for the measuring transducer.
Tools
■ Screwdriver for cross-recess screws, size 1 and 2.
■ Screwdriver for slotted screws, size 2.5, completely insulated to VDE 0680
■ Conductor end sleeve pliers
■ Insulation stripper
12
Electrical Connection – continued –
DS
1 223
3
3
3
3333
LN
0/4-20mA
R 500
2
2
2
22222
1
1
1
11111
S
D
L
Alarm2Alarm1
ORT 7
Basis
ORG
ORT 7 CPU
ABS
Separator
GPS
1
1
2233
key “+” :
Clw - On
key “-” :
Clw - Off
Clw
24VAC
Clw
1
2
2
1
3
4
1
- wire no.
1
2
Wiring diagram OR 52-7
External wiringMeasuring sensor
To first alarm
Attached in housing
cover
Solenoid valve
Fig. 4 Shown position of contacts: Alarm 1 and 2, separator OFF and power failure
Notice
Fig. 4 shows an example where the water will be run back to the ship‘s bilge in the event of an alarm
(oil content > 15 ppm) or auxiliary power failure.
With different limit settings alarm contact 1 can, for instance, be used for the first alarm and alarm
contact 2 for the main alarm.
For bilge water monitoring a pneumatic three-way valve can be controlled by the alarm contact 2 via a
solenoid valve. If the turbidity of the effluent is too high (alarm caused by ingress of oil or dirt particles
introduced into the system during start-up) the three-way valve will divert the contaminated water to
the bilge tank. The separation continues until the oil content falls below 15 ppm and the cleaned water
can be discharged overboard.
N
Earthing screw
in housing
L
PE L N
Disconnecting switch
Three-way valve
Purging valve
External
wiring
13
Functional Elements
ON ALARM 1 ALARM2 SYSTEM
GESTRA
ORT 7
OKP
Measuring sensor ORG 12
6
1
5
2
3
4
Fig. 5
14
7
8
Functional Elements – continued –
ONALARM 1 ALARM 2SYSTEM
GESTRA
ORT 7
OKP
VDR4
VDR5
VDR3
L3
P2P1P3
C24
C25
C26
R34
Measuring transducer ORT 7
9
§
"
Fig. 6
0
!
/%&
Fig. 7
$
15
Functional Elements – continued –
Key
1 Light emitter
2 Measuring cell
3 Solenoid valve Clew
4 Cleaning device
5 Drying cartridge
6 Throttling device
7 Measuring transducer ORT 7
8 Cable glands
9 LCD display
0 System LED
! Keypad
" LED Mains voltage ON
§ LED Alarm 1 and 2
$ Terminating resistor GPS interface
% Memory card
& Locking piece for Memory Card
/ Circuit board for ORT 7 CPU
16
Commissioning Procedure
Factory setting
The measuring transducer features the following factory set default values:
■ Limit alarm 1: 15 ppm
■ Limit alarm 2: 15 ppm
■ Delay of response alarm 1: 1 second
■ Delay of response alarm 2: 1 seconds
Warning
The terminal strips of the measuring transducer ORT 7 may be live during operation!
This presents the risk of severe injuries due to electric shock!
Cut off power supply before opening the housing lid!
Checking electrical connection
Before commissioning the equipment:
■ Check the connection of the supply voltage
■ The supply voltage is indicated on the name plate
■ Make sure that the connections of the signal inputs and alarm contacts are in accordance with the
wiring diagram.
Checking installation
■ Check all pipe connections for leaks and eliminate any leaks you encounter.
■ Unscrew the drying cartridge and remove the seal. Then screw the cartridge in again.
17
Commissioning Procedure – continued –
Device
Certif.
Ser.Nr.
Vers.
:ORT7
:IMO
:MEPC.107(49)
:392358
:311137.10
Flowserve Gestra
OIL CONTENT :
ppm
State:WarmUp
8
UTC10:462006-12-06
OIL CONTENT :
ppm
State:ready
8
UTC10:462006-12-06
OIL CONTENT :
ppm
State:ClearWater ON
0
UTC10:462006-12-06
First commissioning
1. Apply supply voltage to the measuring transducer.
The green LED ON will be illuminated and alarm 2 is
signaled.
2. When the equipment is switched on, the system window
appears for approx. 3 seconds, indicating the following:
Name of the equipment,
Certificate,
Serial number,
Number of the software version.
3. After approx 3 seconds the measurement window
appears, indicating the following:
Oil content (in large characters),
State: Warm up
Current time / date in UTC
4. After approx. 30 seconds the actual value in ppm is
indicated in the measurement window. If this value is
below 15 ppm the message Alarm 2 is cancelled and the
equipment state “Ready” is indicated. If, however, the
value continues to exceed 15 ppm, the message Alarm 2
remains active and the equipment state “ppm>Limit” is
indicated.
5. Press now key
indicated. The equipment state “Clear Water ON” appears.
Rinse the measuring transducer ORG 12 for at least
15 minutes with oil-free water and operate the cleaning
device 4 2 – 3 times, Fig. 5. During the rinsing process
with oil-free water the measuring window should show
0 ppm. If this is not the case switch to programming
mode, Parameter Value Offset, and reduce the zero point
by a max. of 5 ppm.
6. To close the purging valve press key
state “Ready” or “ppm>Limit” or any other alarm message
is indicated.
Note that in the measurement window the keys and and “OK” do not have a function. If you
want to enter the programming mode press key P.
+. The purging valve opens and Alarm 2 is
–. The equipment
18
Commissioning Procedure – continued –
Parameter
act Value Offset
Value :0 ppm
= prev= next
= backOKP
= change
Parameter
act Value Offset
Value :2 ppm
= inc Para= dec Para
= confirmOKP
+
= abort
Parameter
act Value Offset
Value :2 ppm
= prev= next
= backOKP
= change
Programming mode
The following parameters can be selected and, if required, changed:
Value Offset (reduction of zero point 0-5 ppm),
Output (Output of actual value 0 / 4-20 mA),
Alarm 1 Limit Value (Limit Alarm 1, 0-15 ppm),
Alarm 2 Limit Value (Limit Alarm 2, 0-15 ppm),
Alarm 1 Time Delay (Delay of response for alarm 1, 0-15 sec),
Alarm 2 Time Delay (Delay of response for alarm 2, 0-15 sec),
Datalog Export (transferring the storage data to a PC by means of PC software).
1. Press key
system LED will be illuminated and the window showing
the parameter that was changed last appears.
2. Press
desired parameter is shown.
3. Press key
ge. To quit the programming mode without any changes
press key OK (OK = back). The window that was active
before you switched the mode will reappear.
4. After having activated the parameter change by pressing
key P use the keys +/– to adjust the value within the
permissible limits.
5. To cancel the new setting press key
value will then remain valid.
6. To confirm and activate the new value press key OK
(OK = confirm). The parameter window appears again.
7. Press key OK (OK = back) to quit the programming mode.
The window that was active before you switched the mode
will reappear.
P to enter the programming mode. The amber
or to scroll forwards / backwards until the
P (P = change) to activate the parameter chan-
P (P = abort). The old
19
Commissioning Procedure – continued –
Parameter
Output0/4 . . 20 mA
Value :4 mA
= prev= next
= backOKP
= change
Parameter
Datalog Export
Start
= prev= next
= backOKP
= change
Parameter
Alarm2Limit Value
Value :15 ppm
= prev= next
= backOK
P
= change
Parameter
Alarm2Time Delay
Value :5 sec
= prev= next
= backOKP
= change
Other parameter settings
Note
The export of data (Parameter Datalog Export) is described in a separate operating
manual.
20
Start, Operation, Alarm and Function Test
Device
Certif.
Ser.Nr.
Vers.
:ORT7
:IMO
:MEPC.107(49)
:392358
:311137.10
Flowserve Gestra
OIL CONTENT :
ppm
State:WarmUp
8
UTC10:462006-12-06
OIL CONTENT :
ppm
State:ready
8
UTC10:462006-12-06
OIL CONTENT :
ppm
State:ClearWater ON
0
UTC10:462006-12-06
Start and operation
1. Apply the supply voltage to the measuring transducer.
The green LED will be illuminated and alarm 2 is signaled.
2. When the equipment is switched on, the system window
appears for approx. 3 seconds, indicating the following:
Name of the equipment,
Certificate,
Serial number,
Number of the software version.
3. After approx 3 seconds the measurement window
appears, indicating the following:
Oil content (in large characters),
State: Warm up
Current time / date in UTC
4. After approx. 30 seconds the actual value in ppm is
indicated in the measurement window. If this value is
below 15 ppm the message Alarm 2 is cancelled and the
equipment state “Ready” is indicated. If, however, the
value continues to exceed 15 ppm, the message Alarm 2
remains active and the equipment state “ppm>Limit” is
indicated.
5. Press now key
indicated. The equipment state “Clear Water ON” appears.
Rinse the measuring transducer ORG 12 a few minutes
with oil-free water and operate the cleaning device 4
2 – 3 times, Fig. 5. During the rinsing process with oil-free
water the measuring window should show 0 ppm. If this
is not the case switch to programming mode, Parameter
Value Offset, and reduce the zero point by a max. of
5 ppm.
6. To close the purging valve press key
state “Ready” or “ppm>Limit” will be indicated.
+. The purging valve opens and Alarm 2 is
–. The equipment
21
Start, Operation, Alarm and Function Test – continued –
OIL CONTENT :
ppm
State:ppm > Limit
16
UTC10:462006-12-06
Alarm 1 and 2
If the alarm limits 1 and 2 are exceeded, the adjusted time delay is activated.
LED Alarm 1 and LED Alarm 2 are flashing. After the time delay LED Alarm 1 and LED Alarm 2 are
permanently illuminated. The relay contacts Alarm 1 and Alarm 2 open.
Alarm 2 is also indicated when
■ the separator is switched off,
■ the purging valve Clw is opened and
■ if the sensor ORG 12 signals a malfunction
Depending on the reason that caused Alarm 2, the equipment
state “ppm>Limit”, “Separator OFF” or “Clear Water On” will
be indicated in the measurement window.
Functional test
Push cleaning device 4Fig. 5 slowly into the mid position in order to simulate a malfunction.
The amber system LED is flashing and the relay contact Alarm 2 opens. Error code e1 or e2 is
indicated.
Push cleaning device 4Fig. 5 back into its initial position.
The equipment must switch back to normal operation.
22
Data Storage
Memory Card
The memory card is on the CPU circuit board in the housing lid. Fig. 7. The design of the card ensures
that the recorded time, date, alarm, malfunction and status messages are stored for 18 months as
required by the IMO Resolution MEPC 107 (49).
Once the capacity limit of the card is reached, the oldest message will be overwritten, which means
that the card does not have to be exchanged.
The memory card is encoded with the serial number of the associated oil & turbidity detector OR 52-7
and, as a consequence, only this particular device can write data on the card. All other devices are
only able to read the data stored on the card.
Note
When replacing the oil & turbidity detector OR 52-7 make sure that the memory card
remains on board the ship because it contains the proof that the requirements regarding
oil separation as stipulated by the IMO Resolution were fulfilled.
To take the card out of the card holder unscrew the card locking piece &, Fig. 7.
Data indication
According to the IMO Resolution MEPC 107 (49) the time, date, alarm, malfunction and status
messages must be stored for 18 months. The measuring transducer features therefore a data logging
function.
The state of the equipment (Log MSG: Oil content, Alarm 2, Clear Water and Separator) and, in addition,
every change of state (State change: ON/OFF Alarm 2, geographic position, Clear Water and Separator)
are saved every hour.
Operating steps – see Fig. 8
1 To view the stored data in the log window press key in the measurement window. If you
want to go back press key .
2 To enter the enquiry mode press key or . Depending on the adjusted enquiry interval (dT) the
stored state messages (Log Msg) and changes (State Change) will be indicated.
3 If you want to change the interval press key + / – in the enquiry mode. The following interval
options are available: all, day, week or month.
4 Enquiry interval(dT)
dT: all = When pressing on the arrow key the next entry will be indicated.
dT: day = When pressing on the arrow key the 24
dT: week = When pressing on the arrow key the 5 x 24
dT: month = When pressing on the arrow key the 30 x 24
th
entry will be indicated.
th
entry will be indicated.
th
entry will be indicated.
5 Press key or to view all state messages (Log Msg) and changes (State Change), depending
on the adjusted enquiry interval.
6 Press key or to view only the state changes (State change), depending on the enquiry
interval.
23
Data Storage – continued –
OIL CONTENT :
ppm
State: ready
8
UTC 10:462006-12-06
Time
Date
Pos
Oil
Clw
Data Logger
: UTC 09:55
: 2006-12-06
: N 53° 5´ 36´´
E 008° 48´ 15´´
: 8 ppm Ala : Off
: OffSep: On
Log Msg(State change)
Time
Date
Pos
Oil
Clw
dT
Time
Date
Pos
Oil
Clw
dT
State change
Time
Date
Pos
Oil
Clw
dT
State change
Clw
dT
Clw
dT
Clw
dT
: UTC 10:00
: 2006-12-06
:0° 0´ 0´´
000° 0´ 0´´
: 1 ppm Ala : Off
: OffSep: On
: all
: UTC 09:45
: 2006-12-06
:0° 0´ 0´´
000° 0´ 0´´
: 16 ppm Ala : On
: OffSep: On
: all
: UTC 08:15
: 2006-12-06
:0° 0´ 0´´
000° 0´ 0´´
: 16 ppm Ala : On
: OffSep: On
: all
: OffSep: On
: week
: OffSep: On
: day
: OffSep: On
: month
OK : backOK : backOK : back
OK : back
OK : back
OK : back
Time
Date
Pos
Oil
Clw
dT
Log Msg +State change
Time
Date
Pos
Oil
Clw
dT
Log Msg +State change
: UTC 09:00
: 2006-12-06
: N 53° 5´ 36´´
E 008° 48´ 15´´
: 8 ppm Ala : Off
: OffSep: On
: allOK : back
: UTC 08:00
: 2006-12-06
: N 53° 5´ 36´´
E 008° 48´ 15´´
: 8 ppm Ala : Off
: OffSep: On
: allOK : back
OK
OK
OK
OK
OK
2
Device
Certif.
Ser.Nr.
Vers.
: ORT7
: IMO
: MEPC.107(49)
: 392358
: 311137.10
Flowserve Gestra
Note
If you want to view e. g. an alarm on a particular day, first chose the suitable enquiry
interval and press key
press key or to go to the event in question.
Operating flowchart
to go to this date. Then select the enquiry interval dT and
66
55
1
2
3
4
Fig. 8
24
Malfunctions
OIL CONTENT :
ppm
State:e1 Light Ctrl ov
8
UTC10:462006-12-06
OR 52-7
In the event of a malfunction the amber system LED is
flashing and the relay contact Alarm 2 is energised.
An error code is indicated in the measurement window.
The equipment is not working – no function, no indication
Fault: The fuse of the equipment is defective.
Remedy: Replace the fuse of the equipment and check the power supply.
Fault: Internal voltage supply failed.
Not all three green LEDs on the basis circuit board are illuminated.
Remedy: Check power supply or replace the measuring transducer.
Measured value increases continuously with 1ppm/day or slower
(ingress of foreign matter can be ruled out)
Fault: Accumulated dirt deposits on glass cylinder.
Remedy: Operate the cleaning device more often.
Table listing error codes
Error codePossibly caused by:Remedy
e1 Light Ctrl ov
Error codePossibly caused by:Remedy
e2 no Light rec
Error codePossibly caused by:Remedy
mem Card err
Glow lamp in light emitter contaminated Replace light emitter
Dirt deposits on glass cylinderClean glass cylinder
Excessive turbidity
Operating voltage too lowCheck voltage supply
Cleaning device not in initial positionReturn the cleaning device into the initial position
Ingress of solid particles in fluid
Glow lamp in light emitter defective Replace light emitter
Connecting line to light emitter interrupted Check connecting line
Cleaning device not it its initial positionReturn the cleaning device to its initial device
Ingress of solid particles in fluid
No or wrong memory cardInsert the right memory card
Once the fault is eliminated the equipment will return to normal operation.
Provide mechanical cleaning device upstream of
equipment
Provide mechanical cleaning device upstream of
equipment
Provide mechanical cleaning device upstream of the
equipment
25
Calibration
OR 52-7
According to the IMO Resolution MEPC 107 (49), section 4.2.11, the measuring accuracy of the 15 ppm
bilge alarm must be checked when the equipment is inspected in order to obtain the extension of the
period of validity of the IOPP certificate (5 years) as specified by the manufacturer.
For this inspection the measuring transducer ORG 12 must be exchanged.
Please order a replacement for the measuring transducer ORG 12 (stock code 389 427), indicating the
serial number, in good time before the IOPP certificate expires.
Before exchanging the measuring transducer undo the central screw of the connectors for the light
emitter/receiver and for the purging valve and detach the connectors.
Remove the measuring transducer and mount the replacement equipment.
Attach the connectors to the light emitter/receiver and the purging valve and tighten the central screw.
Repeat the commissioning procedure.
26
Maintenance
Cleaning the glass cylinder
1. Push the wiper of the cleaning device
2 or Malfunction may be triggered (depending on the time delay setting) and the amber LED is
flashing.
2. Return the wiper of the cleaning device
70 mm from the equipment).
The cleaning interval (e. g. once a week) depends on the amount of dirt deposits.
Replacing the drying cartridge
Replace drying cartridge when its content turns pink. Should this happen too often check the following
items for tight shut-off:
■ O-rings of light emitter
■ Cable glands of light emitter and light receiver
■ Gaskets of connectors and drying cartridge
Replacing the light emitter/receiver
1. Undo fixing screw and remove connector from light emitter/receiver.
2. Unscrew union nut (outside) of the light emitter/receiver tube and detach the emitter/receiver from
the body of the measuring transducer.
3. Replace the light emitter/receiver. Make sure that the locating pin of the emitter/receiver fits into
the groove of the light emitter/receiver.
4. Tighten external union nut.
5. Attach connector to light emitter/receiver and tighten the fixing screw.
6. Correct the zero point when the measuring transducer in programming mode (Parameter Value
Offset).
Please observe the notes on the spare parts kit “Light”.
4 Fig. 5 up and down. When doing so Alarm 1 and/or Alarm
4 to its initial position (the plunger protrudes approx.
Attention
Do not open the measuring transducer!
Replacing the equipment fuse
1. Open the housing lid of the measuring transducer. Observe the warning note!
2. Turn the fuse carrier to the left and take it off.
3. Replace the equipment fuse.
4. Fit fuse carrier in place and turn it to the right. Close housing lid.
27
Maintenance – continued –
Warning
The terminal strips of the measuring transducer ORT 7 can be live during operation.
This presents the danger of severe injuries due to electric shock.
Always cut off power supply before mounting, removing or connecting wires to the
terminal strips.
Spare parts
The oil & turbidity detector OR 52-7 comes with a spare parts kit containing the following items:
Spare parts kit “Gaskets”, stock code 389 232
DesignationQty.
Drying cartridge1
Gasket for connector GDM 3-21
O ring 25 x 3 (housing lid / glass cylinder)2
O ring 30 x 2 (housing lid / glass cylinder)2
O ring 37 x 2 (housing lid / glass cylinder)2
O ring 48 x 2.5 (housing of light emitter / light receiver)2
If required you can order the following spare parts kit “Light”, stock code 389 333,
containing the following items:
DesignationQty.
Light emitter1
Light receiver1
2
28
Note
When ordering please state the stock code no. 389 333 for the spare parts kit and the
equipment serial number.
Please note that you have to replace both the light emitter and receiver when using
the replacement light emitter and receiver.
Decommissioning
Warning
The terminal strips of the measuring transducer ORT 7 can be live during operation.
This presents the danger of severe injuries due to electric shock.
Always cut off power supply before mounting, removing or connecting wires to the
terminal strips.
Removal
1. Cut off power supply.
2. Disconnect all wires and draw the cables through the cable gland.
3. Remove the measuring sensor and the measuring transducer.
Disposal
Dismantle the oil & turbidity detector OR 52-7 and separate the waste materials according to the
material specification. Electronic component parts such as the circuit board must be disposed of
separately. For the disposal of the control unit observe the pertinent legal regulations concerning waste
disposal.
29
Annex
Type approval certificate
30
Annex – continued –
Type approval certificate – continued –
31
Annex – continued –
Type approval certificate – continued –
32
Annex – continued –
Type approval certificate – continued –
33
Annex – continued –
EG-Baumusterprüfbescheinigung (Modul B)
34
Annex – continued –
EG-Baumusterprüfbescheinigung (Modul B) – Fortsetzung –