Flowserve 520si User Manual

Page 1
Logix 520 si Digital Positioner
ADK041400E - 04.03
This product meets the requirements of
1
Page 2
Table of Content Page
General information...................................................................................... 3
Unpacking .............................................................................................. 4
Logix 520 si Overview .................................................................................. 5
Specifications .............................................................................................. 5
Principle of operation ................................................................................... 6
Tubing .............................................................................................. 6
Wiring .............................................................................................. 7
Cable Requirements ..................................................................................... 8
Startup .............................................................................................. 8
Configuration .................................................................................... 9
Calibration......................................................................................... 9
Use of 275 Handheld ........................................................................ 10
Use of SoftTools communication software ....................................... 10
Integration of third-party communication software .......................... 10
Error codes .............................................................................................. 12
Trouble shooting .......................................................................................... 15
Spare Part Kits ............................................................................................. 16
Mounting Kits .............................................................................................. 16
2
Page 3
1 USING FLOWSERVE VALVES, ACTUATORS
AND ACCESSORIES CORRECTLY
1.1 Using
The following instructions are designed to assist in unpacking, installing and performing maintenance as required on FLOWSERVE products. Product users and maintenance personnel should thoroughly review this bulletin prior to installing, operating or perform­ing any maintenance.
In most cases FLOWSERVE valves, actuators and accessories are designed for specific applications (e.g. with regard to medium, pressure, temperature). For this reason they should not be used in other ap­plications without first contacting the manufacturer.
1.2 Terms concerning safety
The safety terms DANGER, WARNING, CAUTION and NOTE are used in these instructions to highlight par-
ticular dangers and/or to provide additional informa­tion on aspects that may not be readily apparent.
1.3 Protective clothing
FLOWSERVE products are often used in problematic applications (e.g. extremely high pressures, danger­ous, toxic or corrosive mediums). In particular valves with bellows seals point to such applications. When performing service, inspection or repair operations always ensure, that the valve and actuator are de­pressurised and that the valve has been cleaned and is free from harmful substances. In such cases pay particular attention to personal protection (protective clothing, gloves, glasses etc.).
1.4 Qualified personnel
Qualified personnel are people who, on account of their training, experience and instruction and their knowledge of relevant standards, specifications, ac­cident prevention regulations and operating condi­tions, have been authorised by those responsible for the safety of the plant to perform the necessary work and who can recognise and avoid possible dangers.
DANGER:
indicates that death, severe personal in­jury and/or substantial property damage will occur if proper precautions are not taken.
WARNING:
jury and/or substantial property damage can occur if
STOP!
indicates that death, severe personal in-
proper precautions are not taken.
CAUTION:
indicates that minor personal injury and/ or property damage can occur if proper precautions are not taken.
NOTE:
indicates and provides additional technical information, which may not be very obvious even to qualified personnel.
Compliance with other, not particularly emphasised notes, with regard to transport, assembly, operation and maintenance and with regard to technical docu­mentation (e.g. in the operating instruction, product documentation or on the product itself) is essential, in order to avoid faults, which in themselves might directly or indirectly cause severe personal injury or property damage.
1.5 Installation
DANGER:
serial-no. and/or the tag-no. to ensure that the valve/ actuator is correct for the intended application.
Do not insulate extensions that are provided for hot or cold services.
Pipelines must be correctly aligned to ensure that the valve is not fitted under tension.
Fire protection must be provided by the user.
Before installation check the order-no,
3
Page 4
1.6 Spare parts
1.9 Valve and actuator variations
Use only FLOWSERVE original spare parts. FLOWSERVE cannot accept responsibility for any damages that occur from using spare parts or fastening materials from other manufactures. If FLOWSERVE products (especially sealing materials) have been on store for longer periods check these for corrosion or deterioration before using these products. Fire protection for FLOWSERVE products must be provided by the end user.
1.7 Service / repair
To avoid possible injury to personnel or damage to products, safety terms must be strictly adhered to. Modifying this product, substituting nonfactory parts, or using maintenance procedures other than outlined in this instruction could drastically affect performance and be hazardous to personnel and equipment, and may void existing warranties. Between actuator and valve there are moving parts. To avoid injury FLOWSERVE provides pinch-point-protection in the form of cover plates, especially where side-mounted positioners are fitted. If these plates are removed for inspection, service or repair special attention is required. After completing work the cover plates must be refitted.
These instructions cannot claim to cover all details of all possible product variations, nor in particular can they provide information for every possible example of installation, operation or maintenance. This means that the instructions normally include only the directions to be followed by qualified personal where the product is being used for is defined purpose. If there are any uncertainties in this respect particularly in the event of missing product-related information, clarification must be obtained via the appropriate FLOWSERVE sales office.
2 UNPACKING
Each delivery includes a packing slip. When unpacking, check all delivered valves and accessories using this packing slip.
Report transport damage to the carrier immediately.
In case of discrepancies, contact your nearest FLOWSERVE location.
Apart from the operating instructions and the obligatory accident prevention directives valid in the country of use, all recognised regulations for safety and good engineering practices must be followed.
WARNING:
FLOWSERVE for repair or service FLOWSERVE must
STOP!
be provided with a certificate which confirms that the product has been decontaminated and is clean. FLOWSERVE will not accept deliveries if a certificate has not been provided (a form can be obtained from FLOWSERVE).
1.8 Storage
In most cases FLOWSERVE products are manufactured from stainless steel. Products not manufactured from stainless steel are provided with an epoxy resin coating. This means that FLOWSERVE products are well protected from corrosion. Nevertheless FLOWSERVE products must be stored adequately in a clean, dry environment. Plastic caps are fitted to protect the flange faces to prevent the ingress of foreign materials. These caps should not be removed until the valve is actually mounted into the system.
Before products are returned to
4
Page 5
3 LOGIX 520 si OVERVIEW
The Logix 520 si is a two-wire, 4-20 mA input digital valve positioner. The Logix 520 also utilizes the HART protocol to allow two-way remote communication with the positioner. The Logix 520 si positioner controls single-acting actuators with linear and rotary mountings. The Logix 520 si is completely powered by the 4-20 mA input signal. The minimum input signal required to function is 3,6 mA.
Since the positioner is insensitive to supply pressure changes and can handle supply pressures from 1,5 to 6 barg (22 to 87 psig), a supply regulator is usually not required; however, in applications where the supply pressure is higher than the
4 SPECIFICATIONS
Table 1: Input Signal
Input Signal Range 4 - 20 mA (with HART)
Compliance Voltage 12,0 VDC (with HART signal)
maximum actuator pressure rating a supply regulator is re­quired to lower the pressure to the actuator’s maximum rat­ing (not to be confused with operating range). A coalescing air filter is recommended for all applications due to the close tolerances in the positioner.
NOTE:
The air supply must conform to ISA 7.0.01or IEC 770 (a dew point at least 10 ˚C / 18 ˚F below ambient temperature, particle size below five microns – one micron recommended – and oil content not to exceed one part per million).
Table 5: Performance
Characteristics (typical)
Linearity < ± 1,0% Resolution < 0,1%
Repeatability < 0,2% Voltage Supply (maximum) 30 VDC Minimum Required 3,6 mA
Operating Current
Table 2: Stroke Output
Feedback Shaft 0° to 90° normal Rotation 0° to 40° minimum
Table 3: Supply Air
Supply Air Quality Free from moisture, oil and
dust per IEC 770 and ISA-7.0.01 Input Pressure Range 1,5 to 6,0 bar (22 – 87 psi) Air Consumption 0,08 Nm3/h @ 1,5 bar
(steady state) 0,047 SCFM @ 22 psi
0,120 Nm3/h @ 6,0 bar
0,071 SCFM @ 87 psi
Table 4: Output Signal
Output Pressure 0 - 100 % of supply air Range pressure
Output Flow Capacity 2,4 Nm3/h @ 1,5 bar (input @ pressure) 1,41 SCFM @ 22 psi
7,0 Nm3/h @ 6,0 bar
4,12 SCFM @ 90 psi
Deadband < 0,2%
Table 6: Environmental Conditions
Operating Temperature Standard -20 – ˚80 C
-4 – 178 ˚F
Low -40 – 80 ˚C
-40 – 178 ˚F
Transport and Storage -40 – 80 ˚C Temperature -40 – 178 ˚F
Operating Humidity 0 – 100% non-condensing Intrinsic Safety Logix 520 si - 15:
ATEX II1 G EEx ia IIC T6
Table 7: Physical Specifications
Housing Material Cast aluminum,
powder-painted Soft Goods Nitrile Weight 1,2 kg (2,7 lbs)
Table 8: Shipping weights
Shipping weights (incl. mounting kit)
Specification Logix 520 si with pressure gauges 1,6 kg
without pressure gauges 1,2 kg
5
Page 6
Supply Air
Filter / Regulator
for Supply Air
22 – 87 psi (1.5 – 6.0 bar)
2 Electro-pneumatic
Converter Module
1 Digital Control Circuit
Inner Loop
Piezo Control
-
Inner Loop
Position Feedback
Local User Interface
4 – 20 mA + HART
HART
+
Micro
Processor
Gain
Figure 1: Logix 520 si Principle of Operation
5 PRINCIPLE OF OPERATION
The Logix 520 si positioner is a digital positioner that incor­porates HART protocol for communication. The positioner consists of three main modules:
1. The microprocessor-based electronic control module includes HART communications and direct local user interface switches
2. The piezo valve-based electro-pneumatic converter module
3. The infinite resolution valve position sensor.
The basic positioner operation is best understood by refer­ring to Figure 1. The complete control circuit is powered by the two-wire, 4-20 mA command signal. The HART module sends and receives the superimposed FSK HART digital com­munications over the 4-20 mA signal wires providing two­way remote digital communications to the microprocessor. The analog 4-20 mA command is passed to the microproc­essor, where it is compared to the measured valve stem position. The control algorithm in the processor performs control calculations and produces an output command to the piezo valve, which drives the pneumatic amplifier. The position of the pilot valve in the pneumatic amplifier is meas­ured and relayed to the inner loop control circuit. This two­stage control provides for more responsive and tighter con­trol than is possible with a single stage control algorithm. The pneumatic amplifier controls the airflow to the actuator. The change of pressure and volume of the air in the actuator causes the valve to stroke. As the valve approaches the de­sired position, the difference between the commanded po­sition and the measured position becomes smaller and the output to the piezo is decreased. This, in turn, causes the pilot valve to close and the resulting flow to decrease, which slows the actuator movement as it approaches, the new com­manded position. When the valve actuator is at the desired
6
Pressure Regulator
Pneumatic
Piezo Valve
Amplifier
3 Valve Position
Sensor
Stroke
Control
Valve
position, the pneumatic amplifier output is held at zero, which holds the valve in a constant position.
6 TUBING POSITIONER TO ACTUATOR
After mounting has been completed, tube the positioner to the actuator using the appropriate compression fitting con­nectors:
Air connections: 1/4” NPT (standard air connection) Auxiliary power: Pressurized air or permissible gases, free
of moisture and dust in according with IEC 770 or ISA
7.0.01. Pressure range: 1,5 – 6,0 bar (22 – 87 psi) For connecting the air piping, the following notes should be
observed:
1. The positioner passageways are equipped with filters, which remove medium and coarse size dirt from the pressurized air. If necessary, they are easily accessible for cleaning.
2. Supply air should meet IEC 770 or ISA 7.0.01 require­ments. A coalescing filter should be installed in front of the supply air connection Z. Now connect the air supply to the filter, which is connected to the Logix 500 si series positioner.
3. With a maximum supply pressure of 6 bar (87 psi) a regulator is not required.
4. With an operating pressure of more than 6 bar (87 psi), a reducing regulator is required. The flow capacity of the regulator must be larger than the air consumption of the positioner (7 Nm3/hr @ 6 bar / 4,12 scfm @ 87 psi).
5. Connect the outlet connector Y of the positioner to the actuator with tubing, independent of the action (direct or reverse).
Page 7
M8
47
M6
17.7
17.7
28.6
173,5
63,5
17.7
17.7
28.6
100
28.6
1/4" NPT
ø 10
19,9
M 8
ø 10
M 20 x 1,5 or 1/2" NPT
11,9
23,9
60
96.5
Figure 2: Dimensional Drawing of the Logix 500 si Series Digital Positioner
7 WIRING AND GROUNDING GUIDELINES
Electrical connections: signal cable with cable passage (NPT,
PG13,5, or M20 x 1,5) to terminals 2 x 2,5 mm
Input signal: 4 – 20 mA
NOTE:
Observe the minimum requirements of voltage and equivalent electrical load: 12,0 VDC / 600 Ω / at 20 mA
To calculate the maximum network capacitance use the fol­lowing formula.
The performance is ensured only for a minimum input cur­rent of 3,6 mA.
For wiring, the following notes should be observed:
NOTE: The input loop current signal to the Logix 520 should be in shielded cable. Shields must be tied to a ground at only one end of the cable to provide a place for environmental electrical noise to be removed from the cable. In general, shield wire should be connected at the source. (Figure 8)
Connect the 4-20 mA current source to terminals +11 and
-12 (Figure 5).
Grounding Screw
The grounding screw, located inside the positioner cover, should be used to provide the unit with an adequate and reli­able earth ground reference. This ground should be tied to the same ground as the electrical conduit. Additionally, the electrical conduit should be earth grounded at both ends of its run. The grounded screw must not be used to terminate signal shield wires.
ø 4,8
ø 10
11,9
23,9
Compliance Voltage (Figure 4) Output compliance voltage refers to the voltage limit the cur-
rent source can provide. A current loop system consists of the current source, wiring resistance, barrier resistance (if present), and the Logix 520 si impedance. The Logix 520 si requires that the current loop system allow for a 12,0 VDC drop across the positioner at maximum loop current.
CAUTION: Never connect a voltage source directly across the positioner terminals. This could cause per­manent circuit board damage.
In order to determine if the loop will support the Logix 520 si, perform the following calculation.
Voltage = Compliance Voltage (@Current
- Current
MAX
*(R
barrier
+ R
wire
)
MAX
)
The calculated voltage must be greater than 12.0 VDC in or­der to support the Logix 520.
Example: DCS Compliance Voltage = 19 V
R
= 300
barrier
R
= 25
wire
CURRENT
= 20 mA
MAX
Voltage = 19 V - 0,020 A*(300 + 25 )
= 12,5 V
The voltage 12,5 V is greater than the required 12.0 V; there­fore, this system will support the Logix 520. The Logix 520 si has an input resistance equivalent to 600 at a 20 mA input current.
7
Page 8
Enhanced Quick-Cal Configuration Switches
LED Status Lights
Autotune
Lockout Switch
4-20 mA input with HART Communications
Terminals for Optional Limit Switches
Figure 3: Logix 520 si Local Interface
8 CABLE REQUIREMENTS
The Logix 520 si digital positioner utilizes the HART commu­nication protocol. This communication signal is superimposed on the DC 4-20 mA current signal. The two frequencies used by the HART protocol are 1200 Hz and 2200 Hz. To prevent distortion of the HART communication, cable capacitance and cable length restrictions must be calculated. The cable length must be limited if the capacitance is too high. Selecting a cable with lower capacitance/mm rating will allow longer cable runs. In addition to cable capacitance, the net­work resistance also affects the allowable cable length.
Jog Calibrate Buttons
Gain Selector
Quick-Cal Button
To control cable resistance, No. 24 AWG cable should be used for runs less than 1500 m. For cable runs longer than this, No. 20 AWG cable should be used.
Electromagnetic Compatibility
The Logix 520 si digital positioner has been designed to op­erate correctly in electromagnetic (EM) fields found in typical industrial environments. Care should be taken to prevent the positioner from being used in environments with excessively high EM field strengths (greater than 10 V/m). Portable EM devices such as hand-held two-way radios should not be used within 30 cm of the device.
65
C
(µF) < - 0,0032
network
(R
barrier
+ R
+ 390Ω)
wire
Example:
R
= 300 (if present)
barrier
R
= 50
wire
C
(300 + 50 + 390)
Max. Cable Length =
72 ρF 0.000072 µF
= =
cable
m m
65
- 0,0032 = 0,08 µF= C
C
(µF)
network
C
cable
0,08 µF
Max. Cable Length = = 1111 m
0.000072 µF/m
network
(µF)(Max)
Ensure proper wiring and shielding techniques of the control lines, and route control lines away from electromagnetic sources that may cause unwanted noise. An electromagnetic line filter can be used to further eliminate noise (FLOWSERVE Part Number 10156843).
In the event of a severe electrostatic discharge near the positioner, the device should be inspected to ensure correct operability. It may be necessary to recalibrate the Logix 520 si positioner to restore operation.
9 STARTUP
9.1 Logix 520 si Local Interface
The Logix 520 si local user interface allows the user to configure the basic operation of the positioner, tune the re­sponse, and calibrate the positioner without additional tools or configurators. The local interface consists of a Quick-Cal™ button for automatic zero and span setting, along with two jog buttons for spanning valve/actuators with no fixed inter­nal stop in the open position. There is also a switch block containing eight switches. Six of the switches are for basic configuration settings, two are for calibration options. There is also a rotaryselector switch for adjusting the positioner gain
8
Page 9
settings. For indication of operational status or alarm condi­tions there are three LEDs on the local user interface.
Initial DIP Switch Setting
Before placing the unit in service, set the DIP switches in the “Configuration” and “Cal” boxes to the desired control op­tions.
NOTE:
The switch settings in the Configuration box are activated only by pressing the Quick-Cal button or by utilizing the stroke calibration features provided by a handheld or by Flowserve PC software.
9.2 Operation of Configuration DIP Switches Setup
The first six DIP switches are for basic configuration
1. Air Action – Must be set to match the configuration of the valve/actuator mechanical configuration.
ATO
(air-to-open) – Select ATO if increasing output pres- sure from the positioner is tubed so it will cause the valve to open.
ATC
(air-to-close) – Selecting ATC if increasing output
pressure from the positioner is tubed so it will cause the valve to close.
2. Signal at Closed – Normally this will be set to 4 mA for an air-to-open actuator, and 20 mA for an air-to-close actua­tor configuration.
Selecting signal is 4 mA and fully open when the signal is 20 mA.
Selecting signal is 20 mA and fully open when the signal is 4 mA.
3. Pos. Characterization Select
proportional to the input signal. Select
is set in conjunction with the next switch, labeled
tional Pos. Char
4. Optional Pos. Char. – If the is set to optional, this switch is active with the following options:
The =% option will characterize the actuator response to the input signal based on a standard 30:1 equal percent rangability curve.
If
Custom
to a custom table that must be set-up using a properly configured HART 275 handheld or other host software.
5. Auto Tune – This switch controls whether the positioner will auto tune itself every time the Quick-Cal button is pressed
On
determine the positioner gain settings every time a Quick­Cal is performed based on the setting of the rotary switch.
If the rotary autotune switch On, a Flowserve nominal response tuning set will be calculated and used.
If the rotary the
4 mA
will make the valve fully closed when the
20 mA
will make the valve fully closed when the
Linear
if the actuator position should be directly
Optional
if another characteristic is desired, which
Op-
.
Pos. Characterization
is selected, the positioner will be characterized
enables an auto tune feature that will automatically
switch
Gain
Gain
selector switch is set to E with the
Gain
Auto Tune
selector switch is set to
switch On, progressively lower gain set-
D, C, B,
or A with
tings will be calculated and used. If the rotary
the
Auto Tune
tings will be calculated and used.
Off
forces the positioner to use one of the factory preset tuning sets determined by the rotary Settings A through H are progressively higher gain predefined tuning sets. The positioner is preset to Autotune
- On at the factory. This is the recomended setting.
6. Config. Switches — By selecting will read all of the configuration switches each time a
Quick-Cal
Selecting (from the last successful calibration) before the switch was set to zeros and spans the positioner.
9.3 Calibration switches
9.3.1 Stability Switch – This switch adjusts the position control algorithm of the positioner for use with low friction control valves or high friction automated valves.
Placing the switch to the left optimizes the response for low friction, high performance control valves. This setting provides for optimum response times when used with most low friction control valves.
Placing the switch to the right optimizes the response for valves and actuators with high friction levels. This setting slightly slows the response and will normally stop limit cycling that can occur on high friction valves.
9.3.2 Setup of the Cal DIP Switch for the Quick Calibration operating mode
Select internal stop in the 100% stroke position. In mode the positioner will fully close the valve and register the 0% position and then open the valve to the stop to register the 100% position when perform­ing a self-calibration. See detailed instructions in the next section on how to perform an auto positioner calibration.
Select calibration stop in the open position. In the the positioner will fully close the valve for the 0% position and then wait for the user to set the open position using the down arrows. See the detailed instructions in the next section on how to perform a manual calibration using the
WARNING:
may stroke unexpectedly. Notify proper personnel
STOP!
that the valve will stroke, and make sure the valve is
Gain
selector switch is set to
switch On, progressively higher gain set-
Enabled
is performed to determine the configuration.
Disabled
retains the last configuration in memory
Disabled
Auto
Jog
. With this setting a
if the valve/actuator assembly has an
if the valve/actuator assembly has no
F, G,
Gain
selector switch.
, the Logix 520 si
Quick-Cal
or H with
Jog
Jog
buttons labeled with the up and
Jog
buttons.
During the Quick-Cal operation the valve
only
Auto
mode
properly isolated.
Quick-Cal Operation
The
Quick-Cal
the positioner. Pressing and holding the
button is used to locally initiate a calibration of
Quick-Cal
button for
9
Page 10
approximately three seconds will initiate the calibration. If the
Config-Switches
option is enabled, the settings of all the configuration switches are read and the operation of the positioner adjusted accordingly. The
Gain Selector
switch is also read and action will be taken to adjust the gain according to the settings of the calibration switches as described in the previous section. A briefly pressing the
Quick-Cal
Quick-Cal
can be aborted at any time by
button and the previous settings
will be retained. If the
Quick calibration
Quick-Cal
button) is set to Auto and the valve/actuator as-
switch (not to be confused with the
sembly has the necessary internal stops, the calibration will complete automatically. While the calibration is in progress you will notice a series of different lights flashing indicating the calibration progress. When the lights return to a sequence that starts with a green light, the calibration is complete. (See the appendix for an explanation of the various light sequences.)
WARNING:
When operating using local control of the
valve, the valve will not respond to external commands. Notify proper personnel that the valve
STOP!
will not respond to remote command changes, and make sure the valve is properly isolated.
If the
Quick calibration
switch is set to
Jog
, the calibration will initially close the valve then cause a small jump in the valve position. The jog calibration process will only allow
the user to manually set the span; zero position is auto­matically always set at the seat. If an elevated zero is needed
a handheld or other PC based configuration software is re­quired. The LEDs will then flash in a sequence of Y-R-R-G (yellow-red-red-green) which indicates the user must now use the
Jog
keys to manually position the valve to the desired
100% position. Now press both
Jog
buttons simultaneously to proceed to the next step. The valve will then stroke and then wait while flashing the Y-R-R-G sequence again, allow­ing the user to adjust the valve position a second time to ex­actly 100% using the positioned press both
Jog
buttons. When the stem is properly
Jog
buttons simultaneously again to register the 100% position and proceed. No more user ac­tions are required while the calibration process is completed. When the lights return to a sequence that starts with a green light the calibration is complete. (See the appendix for an ex­planation of the various light sequences.)
Factory reset — hold
Quick-Cal
button while applying power and all of the internal variables including calibration will be reset to factory defaults. The positioner must be re-calibrated after a factory reset. Tag names and other user configured limits, alarm settings, and valve information will also need to be restored. A factory reset will always reset the command source to analog 4-20 mA.
Special settings––several adjustments to the Logix 520 such as split-range, MPC, soft stop, etc. cannot be set using the local interface.
To set these parameters please refer to the appropriate com­munications tool manual.
9.4 HART 275 HANDHELD COMMUNICATOR
The Logix 520 si supports and is supported by the HART 275 Handheld Communicator. The DD and the manuals listed be­low can be obtained from the HART Communication Founda­tion or from your Flowserve representative. For more infor­mation please see the following guides:
Product Manual for the HART Communicator
• Logix 520 si Digital Positioner with HART 275 Communicator User Guide
9.5 USE OF SOFTTOOLS COMMUNICATION SOFTWARE
Flowserve corporation has written custom configuration and diagnostic software for the Logix 500 si series called SoftTools™. This software and the
SoftTools Quick Start Guide
are available from a Flowserve representative.
9.6 INTEGRATION OF THIRD-PARTY SOFTWARE
The Logix 520 si positioner can be used with:
– Siemens PDM™ – Fisher Rosemount AMS™
Local control of valve position — Can be done from the user interface by holding both jog buttons and then simultaneously pressing the
Quick-Cal
button. While in this mode the LEDs will flash a Y-G-R-R (yellow-green-red-red) sequence. To exit the local control mode and return to normal operation, briefly press the
Quick-Cal
button.
10
Page 11
12.0 VDC
520
si
Figure 4: Compliance Voltage
PIEZO
Y
RELAY
4-20 mA Signal
Internal Housing EARTH Terminal
Shielded Cable
Z
Limit Switch 1
Limit Switch2
Connect Shield at Source Ground
­4-20 mA Current Source
+
Figure 5: Wiring Diagram
POT.
LS1
LS2
Table 9: Connection Table
Connection Description
+11 Input+ 4..20 mA
-12 Input- 4..20 mA
+41* Limit switch 1+
-42* Limit switch 1-
+51* Limit switch 2+
-52* Limit switch 2­Y Pneumatic output signal (outlet) Z Air supply
* 0ptional
11
Page 12
10 STATUS CODES
Logix 520 si Status Condition Codes
Colors Identifier Indication and resolution G - - - Any sequence starting with a Green light flashing first is a normal operating mode and indicates that
there are no internal problems. GGGG 1 No errors, alerts, or warnings; and the unit is in analog control mode. GGGY 2 MPC active - The command is below the user set limit for tight shutoff feature. This is a normal
condition for a closed valve. The factory default setting is 1% command. To clear the condition use
handheld or Flowserve supplied software to reset the MPC if the range is incorrect or adjust the
command signal above the specified MPC value. GGYG 3 Digital command mode - The analog 4-20 mA input signal is ignored in this mode and a handheld or
Flowserve supplied software is needed to change the position command. (Note a factory reset is the only
method to reset the command back to analog control mode from the local interface if a PC or handheld
configurator is not available. A reset will cause the loss of other data.
(See page 14 for more information) GGYR 4 Initializing, or LED test mode - This sequence should only be visible for 3 sequences when powering
up the unit. GGRG 5 Cycle limit - The cycle limit set by the user has been exceeded. To clear use handheld or Flowserve
supplied software to reset. GGRY 6 Travel limit -The total accumulated travel limit set by the user has been exceeded. To clear use handheld
or Flowserve supplied software to reset. GYYR 7 Soft Stop Lower - The unit is being commanded to exceed a user defined lower travel limit and the
internal software is holding the position at the limit. To clear the condition use handheld or Flowserve
supplied software to reset the limit if more travel is needed or adjust the command signal back in the
specified range. GYRY 8 Soft Stop Upper - The unit is being commanded to exceed a user defined upper travel limit and the
internal software is holding the position at the limit. To clear the condition use handheld or Flowserve
supplied software to reset the limit if more travel is needed or adjust the command signal back in the
specified range. GRYR 9 Position Lower - The position has reached or is exceeding a user defined lower travel indicator similar to
a limit switch indicator. To clear the condition use handheld or Flowserve supplied software to reset the
indicator if more travel is needed or adjust the command signal back in the specified range. GRRY 10 Position Upper - The position has reached or is exceeding a user defined upper travel indicator similar to
a limit switch indicator. To clear the condition use handheld or Flowserve supplied software to reset the
indicator if more travel is needed or adjust the command signal back in the specified range.
12
Page 13
Logix 520 si Status Condition Codes
Colors Identifier Indication and resolution
Y - - - Any sequence starting with a yellow light indicates that the unit is in a special calibration or test mode, or
that there was a calibration problem.
YGYG 11 Signature test in progress- This is a test initiated by Flowserve supplied software that can only be
cancelled by that software.
YYYG 12 Loop Calibration in Progress - Calibration sequence controlled by a handheld or Flowserve supplied
software that can only be cancelled by that software.
YRGG 13 Stroke Calibration in Progress - Calibration sequence started either using the local Quick-Cal button or
by a handheld or Flowserve supplied software. It may be cancelled by briefly pushing the Quick-Cal button.
YGRR 14 JOG Control Mode - the unit has been placed in a local override mode where the valve can only be
stroked using the two local jog buttons. It may be cancelled by briefly pushing the Quick-Cal button.
YYYR 15 Command 0 saturated - Calibration error indicating that the 4-20 mA signal corresponding to 0%
position was out of range. Adjust the signal to the correct range and re-do the calibration. This error may be cleared by briefly pushing the Quick-Cal button, which will force the positioner to use the parameters from the last good calibration.
YYRY 16 Command 100 saturated - Calibration error indicating that the 4-20 mA signal corresponding to 100%
position was out of range. Adjust the signal to the correct range and re-do the calibration. This error may be cleared by briefly pushing the Quick-Cal button, which will force the positioner to use the parameters from the last good calibration.
YYRR 17 Command span - Calibration error indicating that the 4-20 mA signal was below the minimum calibration
span. The minimum calibration span is 1.28 mA. This error may be cleared by briefly pushing the Quick­Cal button, which will force the positioner to use the parameters from the last good calibration.
YRRG 18 Waiting for JOG set point from User - only used during Jog calibration see explanation in Quick-Cal
section of main document for operation.
YRYG 19 Setting IL Offset (in Stroke Cal) - An automatic step in the calibration process that is done with the valve
at 50% position. This must be completed for proper calibration.
YRYY 20 Feedback no-motion during calibration - Indicates that there was no motion of the actuator based on the
current stroking time configuration. Check linkages and air supply to make sure the system is properly connected. If the time out occurred because the actuator is very large then simply retry the Quick-Cal and the positioner will automatically adjust for a larger actuator by doubling the time allowed for movement. This error may be cleared by briefly pushing the Quick-Cal button, which will force the positioner to use the parameters from the last good calibration.
YRYR 21 Feedback 0 saturated - Calibration error indicating that the position sensor was out of range during the
calibration. To correct the condition, adjust the positioner mounting, linkage or feedback potentiometer to move the position sensor back into range then restart the calibration. This error may be cleared by briefly pushing the Quick-Cal button, which will force the positioner to use the parameters from the last good calibration.
YRRY 22 Feedback 100 saturated - Calibration error indicating that the position sensor was out of range during
the calibration. To correct the condition, adjust the positioner mounting, linkage or feedback potentiometer to move the position sensor back into range then restart the calibration. This error may be cleared by briefly pushing the Quick-Cal button, which will force the positioner to use the parameters from the last good calibration.
YRRR 23 Feedback span - The range of motion of the position feedback arm was too small. Check for loose
linkages and/or adjust the feedback pin to a position closer to the follower arm pivot to create a larger angle of rotation. This error may be cleared by briefly pushing the Quick-Cal button, which will force the positioner to use the parameters from the last good calibration.
YRGR 24 Feedback unstable during calibration - Check for loose linkages or loose position sensor. This error
may be cleared by briefly pushing the Quick-Cal button, which will force the positioner to use the parameters from the last good calibration.
YRGY 25 Feedback unstable setting IL Offset - Check for loose linkages or loose positioner sensor. This can also
be caused by over tightened packing, very sticky packing, or very high gain settings. This error may be cleared by briefly pushing the Quick-Cal button, which will force the positioner to use the parameters from the last good calibration.
13
Page 14
Logix 520 si Status Condition Codes
Colors Identifier Indication and resolution R - - - Any sequence starting with a red light indicates that there is an operational problem with the unit.
RGRR 26 Position Deviation - The position has exceeded user defined error band between command and
position.
RYYY 27 Hall sensor non-motion - Check to make sure the air supply is connected. This error may be cleared
by briefly pushing the Quick-Cal button, which will force the positioner to use the parameters from the last good calibration. If the positioner still does not operate replace the pneumatic relay assembly.
RYYR 28 Hall sensor lower position - Check to make sure the air supply is connected. This error may be
cleared by briefly pushing the Quick-Cal button, which will force the positioner to use the parameters from the last good calibration. If the positioner still does not operate replace the pneumatic relay assembly.
RYRY 29 Hall sensor upper position - Check to make sure the air supply is connected. This error may be
cleared by briefly pushing the Quick-Cal button, which will force the positioner to use the parameters from the last good calibration. If the positioner still does not operate replace the pneumatic relay
assembly. RRGG 30 1,23 V reference - Bad electronic assembly, replace. RRGR 31 12-bit A/D reference - Bad electronic assembly, replace. RRYG 32 Temperature limit - The internal positioner temperature is currently exceeding operational limits of
-40 ˚C (-40 ˚F) or 85 ˚C (185 ˚F). RRYY 33 Piezo voltage - Bad electronic assembly, replace. RRYR 34 Board current high - Check internal wiring and connectors for electrical shorts, if no shorts bad
electronic assembly, replace. RRRG 35 12-bit D/A reference - Bad electronic assembly, replace. RRRY 36 EEprom checksum error - The checksum of the internal data has become corrupted. Cycle power and
complete a Quick-Cal if needed. Check internal data to verify correct settings. If the error still occurs,
bad electronic assembly, replace.
13 Version number checking- The version number of the embedded code may be checked at any time except during a
calibration by holding down the button. This will not alter the operation of the unit other than to change the blink sequence to 3 blinks indicating the major version number. Holding the button will give the minor version number without affecting operation. The version codes are interpreted by adding up the numbers assigned according to the following table:
Color First blink value Second blink value Third blink value
Green 0 0 0 Yellow 9 3 1 Red 18 6 2
For example if holding the ∆ button gave a G-G-R code, and holding the ∇ button gave a Y-Y-G code then the resulting version number would be (0+0+2).(9+3+0) or version 2.12.
14
Page 15
11 TROUBLE SHOOTING
Logix 520 si Symptoms and Solutions
Failure Probable Cause Corrective action
No LED is blinking 1. Current source below 3,6 mA 1. Verify current source suppies at least 3,6 mA
2. Incorrect wiring polarity 2. Check wiring for correct polarity
Erratic communications 1. Current source bandwidth not 1. Maximum allowable current source rate of
limited to 25Hz change is 924 mA per second
2. Maximum cable length or cable 2. Check cable conductor size, length and impedance exceeded capacitance. Refer to ‘Cable Requirements’ on
page 11.
3. HART modem connected to PC RS-232 3. Verify laptop battery is not low port not receiving enough power
4. Interference with I.S. barrier 4. Must use HART compatible I.S. barrier
5. Current source stripping (filtering) 5. Use the HART filter (VHF) available from HART signal Flowserve (FLS part-No. 10156843)
Alternatively a 250 and a 22 µF capacitor, installed according to the following schematic drawing can be used to establish communication
22 µF
250
Control system
Unit does not respond 1. Unit is in digital command mode 1. Switch to analog command mode with handheld to analog commands communicator or Soft Tools.
2. Error occurred during calibration 2. Correct calibration error. Recalibrate
Valve position reading 1. Stem position sensor mounting 1. Reposition sensor is not what is expected is off 180 degrees
2. Stroke not calibrated 2. Calibrate stroke
3. Tight shutoff MPC (Minimum 3. No action position cutoff) is active
4. Custom characterization or soft 4. No action stops active
Position is driven fully 1. Stroke not calibrated 1. Calibrate valve stroke open or closed and will 2. Inner-loop hall sensor not connected 2. Verify hardware connections not respond to command 3. Wrong air action entered in software 3. Check ATO (Air-to-open) and ATC (Air-to-Close)
settings. Recalibrate
4. Actuator tubing backward 4. Verify ATO/ATC actuator tubing
5. Electro-pneumatic converter 5. Replace electro-pneumatic converter malfunctioning
6. Control parameter inner-loop offset 6. Adjust inner-loop and see if proper control is too high/low resumes
Sticking or hunting 1. Contamination of the 1. Check air supply for proper filtering and meeting operation of the positioner electro-pneumatic converter. ISA specifications ISA-7.0.01
2. Control tuning parameters not correct 2. Lower proportional gain settings
15
Page 16
12 SPARE PART KITS
Item No. Description Part-No.
.
1 Cover Assembly Yellow
White 2 PC Board Assembly H10014000 3 Repair kit for Logix 500-Relay Module Assembly H10011500 4 Repair kit for Potentiometer Assembly H10011600 5 Position Feedback Assembly H194500003 8 Follower Arm Assembly Max. stroke 65 mm H194512001
Max. stroke 110 mm H194512002
Mounting Kits
Item No. Description Part-No.
IEC 534 part 6 (Valtek 2000, Kämmer KA, Kämmer KP, and standard NAMUR linear valves) H192685001 – Rotary VDI/VDE 3845 (DIN ISO 5211) H194984001
Gauge Kits
Item No. Description Part-No.
–2 gauge set, 6 bar / 4 bar, brass, nickel plated H179379013 –2 gauge set, 6 bar / 4 bar, stainless steel H179379015
Regional Headquarters
Manderscheidstr. 19 1350 N. Mt. Springs Prkwy. 12 Tuas Avenue 20 45141 Essen Springville, UT 84663 Germany USA Republic of Singapore 638824 Telephone: +49 (0) 201 8919 5 Telephone: +1 801 489 8611 Telephone: +65 862 3332 Facsimile: +49 (0) 201 8919 662 Facsimile: +1 801 489 3719 Facsimile: +65 862 4940
Main Sales Offices (Europe, Middle East, Africa)
von-Braun-Straße 19a 12, av. du Québec Station Road Allee du Quartz 1 48681 Ahaus 91965, Courtaboeuf Cedex Pershore, Worcestershire CH-2300 La-Chaux-de Fonds Germany France England WR102BZ Switzerland Telephone: +49 (0) 2561 6860 Telephone: +33 (0) 1 60 923 251 Telephone: +44 (0) 1386 55 45 51 Telephone: +41 (0) 32 925 9700 Facsimile: +49 (0) 2561 68648 Facsimile: +33 (0) 1 60 923 299 Facsimile: +44 (0) 1386 55 49 68 Facsimile: +41 (0) 32 926 5422
Units 1 and 2 C/O Saleh & Abdulaziz Abahsain 26, Imvuba Road, Sebenza Ext 6 P.O. Box 209 Edenvale, Gauteng Edenglen 1613 Al Khobar 31952 South Africa Saudi Arabia Telephone: +27 11 609 2094 Telephone: 9663 857 3442 Facsimile: +27 11 609 3735 Facsimile: 9663 859 5284
©03.2002 Flowserve Corporation. Flowserve and Kämmer are trademarks of Flowserve Corporation
All data subject to change without notice
16
Loading...