Introduction

Congratulations on your purchase of the Extech PH220 waterproof pH/mV meter. This microprocessor-based device with tactile push-buttons is battery operated and ideal for field use. The PH220 simultaneously displays pH and temperature. The rubber holster which protects the meter has a magnetic backing for convenient mounting. The built-in memory (25 readings) allows STORE, RECALL, and CLEAR functionality. Careful use of this meter will provide years of reliable service.

Specifications

 

Range (Resolution)

 

 

 

Accuracy

 

 

 

 

 

 

mV (ORP)

± 1999 mV DC (1 mV)

 

 

 

± 2mV

 

 

 

 

 

 

Temperature

32 to 212oF (<100o;1 o >100o)

 

 

 

± 1.0oF

 

0 to 100oC (0.1°<100°;1°>100°)

 

 

 

± 0.5oC

 

 

 

 

 

 

pH

0.00 to 14.00pH (0.01pH)

 

 

 

± 0.02pH

 

 

 

 

Note 1: Accuracy does not include pH buffer

 

temperature coefficient

 

 

 

 

Note 2: Accuracy specified from 0.02pH to 13.98pH

Temperature Compensation

32 to 212oF (0-100oC) Automatic ‘ATC’

 

 

 

 

 

 

 

 

Power

9 volt alkaline battery

 

 

 

 

 

Display

Dual LCD display for simultaneous pH and temperature

 

readings

 

 

 

 

 

Over-range or open indication

Displays "HHH" for over-range or open input conditions

 

Displays “LLL” for under-range

 

 

 

 

 

pH Calibration points

4.00, 7.00, 10.00 (1, 2, or 3 point calibration can be

 

performed)

 

 

 

 

 

Auto Power OFF

10 minutes from last key-press (can be disabled)

 

 

 

 

 

 

 

 

 

 

 

 

Low Battery Indication

Fill bars in the battery display icon

 

 

 

 

 

 

 

 

 

represent

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

battery power status. The fewer the

 

 

bars, the weaker

 

the battery power.

 

 

 

 

 

Dimensions / Weight

4.25 x 3.00 x 1.18" (108 x 75 x 30mm) / 4.8 oz. (135g)

 

 

Accessories supplied with kit

Battery, suction cup holder, clip holder, stick electrode

 

for model PH220-S, standard electrode with 39” cable

 

for model PH220-C.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

2

Model PH220 Version 3.1 11/2005

Meter Description

1.Battery Compartment (under holster)

2.Rubber Holster

3.LCD Display

4.Function buttons

5.Electrode input

1

2

3 7.00

4

5

Basic Display Description

1.Measurement value

2.

Battery status

1

 

3.

Units of measure

 

4.

Temperature units

2

5.

Solution temperature

 

5

3

4

Note that there are other more specialized displays that appear on the LCD that have not been included in this basic operation display description. Other display icons are used and are discussed in the appropriate sections of this manual.

Order of Operations

1.Calibrate the instrument as described later in this manual.

2.Determine the test type (pH or ORP) and carefully follow the directions for the desired test.

3.Configure the meter to display in the desired units of measure for temperature (oC or oF). This is included as a step in the pH measurement instructions.

4.Pay close attention to detail, step by step instructions, and electrode cleaning/rinsing recommendations. Slight deviations in test procedures and test habits can have a significant effect on test results.

3

Model PH220 Version 3.1 11/2005