Extech MP510, MP520, MP530 Operating Manual

0 (0)
Extech MP510, MP520, MP530 Operating Manual

User’s Guide

MultiPro™ Digital MultiMeter Series

With PC Interface

Models:

MP510

MP520

MP530

1

WARRANTY

EXTECH INSTRUMENTS CORPORATION warrants this instrument to be free of defects in parts and workmanship for three years from date of shipment (a six month limited warranty applies on sensors and cables). If it should become necessary to return the instrument for service during or beyond the warranty period, contact the Customer Service Department at (781) 890-7440 ext. 210 for authorization. A Return Authorization (RA) number must be issued before any product is returned to Extech. The sender is responsible for shipping charges, freight, insurance and proper packaging to prevent damage in transit. This warranty does not apply to defects resulting from action of the user such as misuse, improper wiring, operation outside of specification, improper maintenance or repair, or unauthorized modification. Extech specifically disclaims any implied warranties or merchantability or fitness for a specific purpose and will not be liable for any direct, indirect, incidental or consequential damages. Extech's total liability is limited to repair or replacement of the product. The warranty set forth above is inclusive and no other warranty, whether written or oral, is expressed or implied.

REPAIR AND CALIBRATION SERVICES

Extech offers complete repair and calibration services for all of the products we sell. For periodic calibration, NIST certification or repair of any Extech product, call customer service for details on services available. Extech recommends that calibration be performed on an annual basis to insure calibration integrity.

Support Hotline (781) 890-7440

Tech support: Ext. 200; Email: support@extech.com

Repair/Returns: Ext. 210; Email: repair@extech.com

Website: www.extech.com

Copyright © 2002 Extech Instruments Corporation.

All rights reserved including the right of reproduction in whole or in part in any form.

MP Series V2.2 9/03

P/N: 7M1C-0241-A000

2

Introduction

Congratulations on your purchase of Extech model MP510, MP520, or MP530 digital multimeter. Properly used, this meter will provide many years of reliable service.

MultiPro™ Meters measure AC/DC Voltage, AC/DC Current, Resistance, Frequency, Capacitance, Diode, and Continuity. The MultiPro™ also offers RS-232 PC Interface data acquisition capabilities as an option. Models MP520 and MP530 are True RMS AC sensing. In addition, the MP530 offers Temperature, Relative, Peak capture and MIN/MAX functions.

Safety

International Safety Symbols

This symbol, adjacent to another symbol or terminal, indicates the user must refer to the manual for further information.

This symbol, adjacent to a terminal, indicates that, under normal use, hazardous voltages may be present

Double insulation

Safety Precautions

1.Improper use of this meter can cause damage, shock, injury or death. Read and understand this users manual before operating the meter.

2.Make sure any covers or battery doors are properly closed and secured.

3.Always remove the test leads before replacing the battery or fuses.

4.Inspect the condition of the test leads and the meter itself for any damage before operating the meter. Repair or replace any damage before use.

5.Do not exceed the maximum rated input limits.

6.Use great care when making measurements if the voltages are greater than 25VAC rms or 35VDC. These voltages are considered a shock hazard.

7.Always discharge capacitors and remove power from the device under test before performing Capacitance, Diode, Resistance or Continuity tests.

8.Remove the battery from the meter if the meter is to be stored for3 long periods.

9.To avoid electric shock, do not measure AC current on any circuit whose voltage exceeds 500V AC.

10.Voltage checks on electrical outlets can be difficult and misleading because of the uncertainty of connection to the electrical contacts. Other means should be used to ensure that the terminals are not "live".

11.The product is intended only for indoor use

12.If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired.

13.Pollution degree: 2

PER IEC1010 OVERVOLTAGE INSTALLATION CATEGORY

OVERVOLTAGE CATEGORY I

Equipment of OVERVOLTAGE CATEGORY I is equipment for connection to circuits in which measures are taken to limit the transient overvoltages to an appropriate low level. Note – Examples include protected electronic circuits.

OVERVOLTAGE CATEGORY II

Equipment of OVERVOLTAGE CATEGORY II is energy-consuming equipment to be supplied from the fixed installation.

Note – Examples include household, office, and laboratory appliances.

OVERVOLTAGE CATEGORY III

Equipment of OVERVOLTAGE CATEGORY III is equipment in fixed installations.

Note – Examples include switches in the fixed installation and some equipment for industrial use with permanent connection to the fixed installation.

OVERVOLTAGE CATEGORY IV

Equipment of OVERVOLTAGE CATEGORY IV is for use at the origin of the installation.

Note – Examples include electricity meters and primary over-current protection equipment

Specifications

4

 

 

DC VOLTAGE

 

 

 

 

RANGE

Accuracy

50.00 mV

0.12% + 2d

500.0 mV

0.06% + 2d

5.000V, 50.00V, 500.0V, 1000V

0.08% + 2d

NMRR: >60dB @ 50/60Hz, CMRR: >120dB @ DC, 50/60Hz, Rs=1kΩ Input impedance: 10MΩ, 16pF nominal (44pF nominal for 50mV & 500mV ranges)

AC VOLTAGE

RANGE

Accuracy

50Hz/60Hz

 

50.00mV, 500.0mV, 5.000V, 50.00V, 500.0V, 1000V

0.5% + 3d

40Hz to 500Hz

 

50.00mV, 500.0mV

0.8% + 3d

5.000V, 50.00V, 500.0V

1.0% + 4d

1000V

1.2% + 4d

Up to 20kHz

 

50.00mV, 500.0mV

0.5dB*

5.000V, 50.00V, 500.0V

3dB*

1000V

Unspecified

*Specified from 30% to 100% of range CMRR: >60dB @ DC to 60Hz, Rs=1kΩ

Input Impedance: 10MΩ, 16pF nominal (44pF nominal for 50mV & 500mV ranges)

DC CURRENT

RANGE

Accuracy

Burden Voltage

500.0µA, 5000µA

0.2% + 4d

0.15mV/µA

50.00mA, 500.0mA

3.3mV/mA

5.000A, 10.00A*

 

0.03V/A

*10A continuous, 20A for 30 seconds max with a 5 minute cool down interval