IES 300
Service Manual
01.2000
V 1.xx
IES 300 No. 10321-000 Standard Version |
ISO 9001 |
CE |
|
EN 46001 |
|||
|
IES 300 No. 10321-001 UL Version
ISO 9001
EN 46001
Manual part number: 80116-441
for IES 300 International and UL Version
All rights reserved. No part of this document may be translated, stored in information retrieval systems, or transmitted in any form or by any means - electronic or mechanical, including photocopying, recording or otherwise - without the written permission of ERBE Elektromedizin.
Printed by: ERBE Elektromedizin
Printed in Germany
Copyright © ERBE Elektromedizin GmbH, Tübingen 2000
ERBE IES 300
Service Manual
Chapter |
Title ........................................................................................ |
Page |
1 |
Safety notes for servicing ..................................................................... |
7 |
2 |
Installation .............................................................................................. |
8 |
3 |
Function and configuration of unit ...................................................... |
9 |
3.1 |
Description of function............................................................................ |
10 |
3.2 |
Software function ................................................................................... |
12 |
3.3 |
Description of hardware ......................................................................... |
20 |
3.4 |
Connector pin assignment ..................................................................... |
23 |
3.5 |
Connector assignment ........................................................................... |
24 |
3.6 |
Test and measuring points ..................................................................... |
25 |
4 |
Replacement of components .............................................................. |
26 |
5 |
Elimination of operational and functional problems ........................ |
27 |
6 |
Maintenance and servicing ................................................................. |
28 |
6.1 |
General notes ........................................................................................ |
28 |
6.2 |
Inspection requirements ......................................................................... |
28 |
7 |
Spare parts and accessories .............................................................. |
29 |
7.1 |
Spare parts ............................................................................................ |
29 |
7.2 |
Accessories ............................................................................................ |
30 |
8 |
Technical data ...................................................................................... |
31 |
9 |
Unit versions ........................................................................................ |
32 |
10 |
Circuit diagrams ................................................................................... |
33 |
|
Smoke evacuation Display board .......................................................... |
33 |
|
Smoke evacuation Display board plan of component parts ................... |
34 |
|
Smoke evacuation control ...................................................................... |
35 |
|
Smoke evacuation Control - Plan of component parts ........................... |
36 |
|
Smoke evacuation wiring plan ............................................................... |
37 |
|
Smoke evacuation Filter board .............................................................. |
38 |
Safety notes for servicing, Installation • 1 / 2
WARNING |
The safety instruction WARNING denotes a danger which can cause injury |
|
to persons. |
CAUTION |
The safety instruction CAUTION denotes a danger which can cause damage |
|
to property. |
ATTENTION |
The safety instruction ATTENTION denotes a danger which can cause failure |
|
of the device. |
Applicable operating instructions
The operating instructions pertaining to the IES 300 are part of these service instructions. The knowledge fo their contents, of the kind of installation and of commissioning described there, and also of its operation are the precondition for carrying out any service activities.
Safety precautions against the threat of electric shocks
WARNING! Only connect the IES 300 using the mains cable supplied by ERBE, or one of at least the same quality, to a properly installed grounded outlet. If you use an equipment cart, this also applies to the power cord of the cart. The power cord must bear the national mark of conformity.
For safety reasons, multiple outlets and extension cables should not be used. If their use is unavoidable, they, too, must be provided with proper grounding.
WARNING! Unplug the power cord from the outlet before exchanging parts of the device or cleaning it.
WARNING! Do not plug in a mains cable which is wet into the device or into an outlet.
WARNING! Do not touch any unprotected wires or conductive surfaces while the device is disassembled and is under voltage. Never carry a grounding belt while working with a device under voltage.
WARNING! The unit is protected by mains fuses. If one of these fuses blows, the unit must not be used on patients until it has been checked by a properly trained technician. Only replacement fuses of the rating specified on the unit’s name plate may be used.
Exclusion of warranty
There are no warranty claims given for damages which have been caused by third-party accessories or by third-party consumables.
ERBE is assuming responsibility for appliances regarding safety, reliability, and function only if assembling, new settings, alterations, extensions, and repairs are carried out by ERBE or by an organisation authorized by ERBE.
The safety of the user and the problem-free operation of the appliance are ensured only when using ERBE original appliance components. If third-party accessories or third-party consumables are used, ERBE cannot grant any guarantee for the safe operation and safe function of the IES 300.
Nr.: 80116-441 |
2000 |
Art.- |
01 / |
7 / 38
1 / 2 • Safety notes for servicing, Installation
Electric connection
The IES 300 is connected to a socket outlet with earthing contact, which must be installed near the appliance.
The power consumption of the IES 300 is, depending on the kind of combination with the surgical appliance, max. 16 A; therefore, we recommend not to connect any further consumers to this electrical supply circuit.
For any further electrical devices which can be installed at the work place of the user a sufficient number of socket outlets with earthing contact should be provided.
Nr.: 80116-441 |
2000 |
Art.- |
01 / |
8 / 38
Function and configuration of unit, Replacement of components • 3 / 4
3 Function and configuration of unit
1 |
2 |
3
Fig. 1: overall view
(1)Front panel
(2)Main filter
(3)Filter locking
The basic IES 300 unit is consisting of:
·• Internal synchronous activation
·Automatic switch-off system
·Filter monitoring system
·Electronically controlled brushless blower
·Main filter unit (ULPA and 3-stage gas filter)
·Hose
·Prefilter set
·Suction tube
·Mains cable
Nr.: 80116-441 |
2000 |
Art.- |
01 / |
9 / 38
3 / 4 • Function and configuration of unit, Replacement of components
3.1 Description of function
The IES 300 is an electrically operated medical unit cleaning in rooms used for medical purposes the air from fumes and gas constituents arising during vaporisation of human or animal tissue. Typically, these fumes arise when using laser or electro-surgical units and normally is consisting of water vapour, aerosols, and organic gases.
The IES 300 is improving the environmental conditions in the operating theatre:
·reducion of dust load by removing respirable particles
·improvement of view
·removal of foul-smelling and partly toxic organic gases
·exhausting and filtering of dangerous bio-aerosol (filtration of viruses)
The IES 300 sucks in the fume-laden air through a device positioned at the surgical application part or through a hose hand-held close to the application part. The sucked-in air is then cleaned and the hazardous matter removed by a high-efficiency filter and returned again into the air circulation.
Cleaning of the air refers to constituents like aerosols, which are being retained by means of an ULPA high-efficiency particle filter, as well as also to organic gases, which, amongst other reasons, are being noted because of an odour, unpleasant for human beings, and which are absorbed by a special gas filter.
3.1.1 Schematic presentation of functional relationships
The opposite Figure 2 is showing the functional relationships of the IES 300.
The fumes (6) produced by the surgical unit (1) are removed through the extraction channel (7) by the extraction unit (5), consisting of the filter and the blower. The suction unit (5) is controlled by the control electronics (4) integrated in the unit, which is receiving its activation signals either from the operator’s station (8) or, however, from the current sensor (3), monitoring the power input of the surgical unit (1).
Nr.: 80116-441 |
2000 |
Art.- |
01 / |
10 / 38
Function and configuration of unit, Replacement of components • 3 / 4
1 |
3 |
2 |
|
|
9 |
6 |
4 |
8
7
5
Nr.: 80116-441 |
2000 |
Art.- |
01 / |
ElektrisElectricheconnectionVerbindu g
Electric. Netzverbindungmains connection
Suction channel
Absaugkanal
Air suction
Luftströmung
1.1.ChirurSurgicaliegerequipmentä
2.2.NetzleitungMains connectiondes Chirurgiegerof surgicaläts equipment
3.3.StromseCurrentsorsensor
4.Steuerelektronik
4.Control electronics
5.Absaugeinrichtung mit Filter u. Gebläse
5.Suction unit with filter and blower
6.Rauch an der Reaktionsstelle
6.Fumes at point of reaction
7.Absaugkanal
7.Suction channel
8.Bedienund Anzeigeeinheit
9.8.NetzversorgungOperator's and display station
9.Supply from mains
Fig. 2: Function chart
11 / 38
3 / 4 • Function and configuration of unit, Replacement of components
3.2 Software function
3.2.1 Operation-Mode
No button pressed when IES 300 is switched on.
Basic program
Display „80“ (depending on latest setting value)
LED |
(= operation suction) |
Button |
Description |
|
Increase the value |
|
Decrease the value |
|
Switch-over of setting of operation suction, base suction, run-on time |
|
Switching off equipment (also no base suction) |
|
Activate device |
Operating overview |
|
LED |
Description |
|
Operation suction (active) |
|
Base suction intensity |
|
Run on time |
Operation suction
Operation suction can be altered at any time. To do this, LED Operation suction must be lit.
Press button until LED |
Operation suction (active) switches on. |
Increase the value |
|
Decrease the value |
|
The value can be set between 20 and 100 %. This corresponds to an suction rate of approx. 150 to approx. 630 l/min.
By prolonged pressing of button |
or |
quick adjustment is effected, i. e. the value is increased or decreased very |
quickly.
When activated, there is automatic switch-over to Operation suction setting.
When there is no filter, the IES 300 cannot be activated.
Nr.: 80116-441 |
2000 |
Art.- |
01 / |
12 / 38